Measurement of the top bottom effect in scanning transmission electron microscopy of thick amorphous specimens

Size: px
Start display at page:

Download "Measurement of the top bottom effect in scanning transmission electron microscopy of thick amorphous specimens"

Transcription

1 (3 Journal of Microscopy, Vol. 100, Pt 1,January 1974, pp Received 1 January 1973; revision received 29 June 1973 Measurement of the top bottom effect in scanning transmission electron microscopy of thick amorphous specimens by P. GENTSCH, H. GILDE and L. REIMER, Physikalisches Institut der Univusitat Miinster, W. Germany, Elektronenmikroskopische Abteilung SUMMARY In Scanning Transmission Electron Microscopy (STEM) details at the top of thick specimens can be imaged with a better resolution than those at the bottom, because the scanning electron probe is broadened by multiple scattering. For quantitative measurement we used a test specimen consisting of a substrate film with an island-structured indium film and coated with polystyrene spheres of different diameters up to 1.1 pm. The fine detail contained in the indium film was observed and the beam broadening was that caused by the polystyrene spheres. This specimen was imaged with the spheres both above and below the layer, in a JEOL loob electron microscope with a scanning device. Electron energies from 20 to 100 kev were used. Images with the sphere below show a higher resolution than those with the sphere above and also those of conventional electron microscopical images at 100 kv and 200 kv. The broadening of the edges on the photographic film was recorded with a microdensitometer. The results can be compared with theoretical calculations by a Monte Carlo method. There is a good agreement with the experiments if one uses an effective aperture obtained by transmission experiments. INTRODUCTION Scanning Transmission Electron Microscopy (STEM) will be of use for thick specimen layers because at 100 kv in conventional Transmission Electron Microscopy (TEM) the resolution is limited by the chromatic aberration of the objective lens and the energy losses of the electrons passing through the specimen. The first successful experiments have been made with the TEM in the STEM mode (Koike et al., 1971; Nauta & Thompson, 1972; Thieringer, 1972). These show that there is an increase in resolution for specimens in the range of 1-2 pm thickness. This is because there is no imaging lens system below the specimen and therefore no effect due to chromatic error. The limitation in the STEM mode is caused by the diameter of the electron probe and by multiple scattering of the electrons inside the specimen. The latter results in a broadening of the electron beam with increasing penetration depth. This was postulated also by Hashimoto (1966) to explain some effects in the TEM of crystalline specimens and was called the top bottom effect (TBE). For the case of STEM in particular this 81

2 P. Gentsch, H. Gilde and L. Reimer effect causes a higher resolution for structures lying on the top of a specimen because this is scanned by the unbroadened beam; whereas details at the bottom are scanned by the broadened beam and are imaged with a lower resolution. Theoretical calculations of the TBE were made (Reimer, Gilde & Sommer, 1970; Reimer, 1972) using Monte Carlo methods. These also gave the dependence of effective electron beam broadening on the aperture. In this paper we report experiments to test this theory. EXPERIMENTAL METHOD In order to obtain the pure TBE, we used the following simple specimen structure. A thin indium film was evaporated on a carbon coated formvar supported film. The indium layer consists of small islands of flat crystals having sharp edges. This film was coated with some polystyrene spheres of different diameters simulating a typical biological specimen with that thickness. If the specimen is orientated in such a way that the spheres lie below the film, the indium islands are scanned by the unbroadened beam of about 5 nm diameter at 100 kv and 10 nm at 20 kv (Fig. la). These diameters of the scanning electron (a j Scanning \ / electron probe Fig. 1. Specimen structure and electron beam broadening in the two positions with the sphere (a) below and (b) above the indium layer. beam are obtained with a hairpin cathode. After inverting the specimen, and focusing again on the indium islands, the spheres lie above the film and we observe the additional broadening of the beam inside the sphere (Fig. lb). This effect is demonstrated in Fig. 2 showing STEM images with a 1.1 pm sphere (a) below and (b) above the indium layer. The edges of the crystals appear more diffuse in Fig. 2b. The very small indium crystals resolved in (a) cannot be detected in (b). For comparison, the same sphere is imaged with the conventional TEM mode at 100 kv in Fig. 2c. The much lower resolution caused by the energy losses of the electrons within the sphere demonstrates the advantage of the STEM mode in this range of thickness. In the TEM mode, contrary to STEM, the TBE causes a better resolution for details at the bottom of a specimen. However, the chromatic error is larger than the TBE and is also the same in the two positions of the sphere. 82

3 Top bottom eflect in STEM Therefore only one image with the sphere above the specimen is shown in Fig. 2c and d. The additional loss of resolution for details at the top of the specimen caused by the TBE should be more easily observed in the range MeV. At these higher electron energies the broadening by multiple scattering becomes comparable to or larger than the disc caused by the chromatic error (Reimer, 1972). Experiments are in progress to detect the small TBE differences in the TEM mode at 100 kv. Comparing the STEM and TEM images in Fig. 2, we notice an improved contrast range in STEM, because there is a S-shaped transfer function between video signal and viewing screen brightness. Figure 2a-c was obtained using a JEOL loob microscope with a scanning attachment. Figure 2d shows an image of the same polystyrene sphere above the specimen layer obtained with a JEOL 200A microscope at 200 kv. In comparison with Fig. 2c, a decrease of chromatic error with higher voltage can be seen. However, the sharpness of the indium edges has not reached that of the STEM mode at 100 kv. For example in STEM mode with a sphere of 1-1 pm diameter and 100 kv we measured a beam broadening of about 22 nm (see definition of r1,2 below). In TEM mode, however, at an objective aperture u = 1.75 x lo-' rad we obtained 60 nm at 100 kv and 28 nm at 200 kv (u = 2.2 x rad). Figure 3 shows a special arrangement of polystyrene spheres in the SEM and STEM mode. The SEM image (Fig. 3a) obtained by detecting the secondary electrons which are emitted at the top surface of the specimen shows, that one of the six spheres lies on the top of the other five. The SEM image was obtained at a primary energy of 100 kev and therefore shows more noise than for primary energies in the range of 20 kev. Figure 3a shows the influence of charging of the polystyrene spheres. But this charging has only an influence on the signal of the secondary electrons not on the signal of the transmitted electrons as confirmed by putting a conductive coating on the spheres. In the STEM image of the same sphere arrangement (Fig. 3b) we observe less resolution below the sphere in the middle than below the outer ones. One has to consider that there is not only a spatial broadening, but also an angular broadening at the bottom of the uppermost sphere. Therefore a further broadening of the scanning beam occurs when the electrons travel from the bottom of this sphere to the indium layer. QUANTITATIVE MEASUREMENT OF ELECTRON BEAM BROADENING To get quantitative information about the loss of resolution caused by the TBE, we made several exposures of the indium crystals and compared the sharpness of the edges seen through the spheres with that beside them. The parameters which were varied are the energy of the primary electrons (20,40,60,80 and 100 kv), the diameter of the spheres (0.234,0-481,0-71 and 1.1 pm) and the diameter of the objective diaphragm (30 pm, without diaphragm). The evaluation of the sharpness of the indium edges was made from microdensitometer records of the photographic negatives. The broadening without TBE is mainly caused by the diameter of the electron probe at the specimen. The current density within the unbroadened beam can be described by a Gaussian distribution j(r) = j o exp K-r/r0)21. (1) If the error disc of the pure multiple scattering is regarded as a Gaussian distribution with rs instead of ro, the current density in the broadened beam can also be described by a Gaussian distribution having the total characteristic parameter rt : rt2 = ro2 + r:. (2) 83

4 P. Gentsch, H. Gilde and L. Reimer Fig. 2. Images of an indium layer and the same polystyrene sphere with 1.1 pm diameter in the following operation modes: (a) 100 kv, STEM, sphere below the indium layer; (b) 100 kv, STEM, sphere above; (c) 100 kv, TEM, sphere above, 01 = 1.75 x rad; (d) 200 kv, TEM, sphere above, 01 = 2.2 x rad.

5 Top bottom effect in STEM 85

6 P. Gentsch, H. Gilde and L. Reimer 86 Fig. 3. (a) SEM, and (b) STEM images of spheres at 100 kv.

7 Top bottom effect in STEM Taking a photometer record in the x-direction from the edge of an indium crystal one has to expect an intensity record Z(x) (Fig. 4), which can be described by integrating j(xyy) in the y-direction from the limits - co to + co and in the x-direction from 0 to x: I(x) = j: s I 2 AXY y)dxdy = { J j, exp ( - (x2 + y2)lro2)dxdy = Io(42>ro2+(X/~o> = Zo+(x/ro>. (3) +(x/r,) is the Gaussian error function with lim,,,+(x/r,) = 1. Z(x) gives the number of electrons within a Gaussian distribution which have an x-coordinate lower than x (dotted area in Fig. 4). 1 I Fig. 4. Integration of a Gaussian distribution (equation 3). We now define a characteristic width xliz of the Z(x) curve, which is chosen in such a way that half of all electrons in the Gaussian distribution fall between the marks x = _+0.5x1,, (Fig. 4). This means: (~I2>ro2+(X1,2Ir*> = I012 = (a/4)ro2. Solving for x1 one gets x1 rs = 0.48 To. For a Gaussian distribution x1 is proportional to To. The values of x,, without beam broadening are obtained from intensity records of the indium crystals beside the polystyrene spheres and the values xt, 1,2 with beam broadening from records of crystals below the centre of the spheres. Analogous to equation (2) we have xt2, 112 = xo2, x,2, 112. (4) One now can calculate the broadening xs, caused by the TBE from the experimental values x,, 112 and xt, 2. In the following xs, 112 is attributed to x ~,~. The results of these measurements are shown in Fig. 5 in which the values x1 are plotted as a function of sphere diameter, t, with the accelerating voltage as a parameter. In order to avoid the sharpness of the indium edges having an influence on x1,2 many measurements were made from different crystals for one data point in Fig. 5. The dependence of the diameter of the objective diaphragm could not be detected, although Monte Carlo calculations show that a very strong effect should be expected (section D). 87

8 P. Gentsch, H. Gilde and L. Reimer Sphere diameter f (pm) Fig. 5. Observed broadening of an electron probe by multiple scattering as a function of specimen thickness (density p = 1.05 g/cm3) and beam voltage. +, 20 kv; a, 40 kv; u, 60 kv; 0, 80 kv; 0, 100 kv. In addition to the parameter x~,~, the TBE can also be discussed in terms of rl;z which is defined in such a way that half of the transmitted electrons fall into the area m-1,z2 around the direction of the incident probe. One can easily show that the relation between and xli2 is given by rliz = 1.75~~~~. (5) COMPARISON WITH MONTE CARL0 CALCULATIONS The trajectories of 50,000 electrons passing through specimens of different mass thickness pt (p density, t thickness) have been simulated by a Monte Carlo program (Reimer et al., 1970; Reimer & Gilde, 1969). The electrons which started in a direction normal to the surface, and with coordinates (x, y, z) = (0, 0, 0) are distributed over a certain area at the bottom of the specimen. This local distribution can be characterized by the distance xli2 described above (Fig. 4). The value of xli2 will decrease with decreasing observation aperture a, because the probability of large angle scattering is greater among the highly displaced electrons. The results of the Monte Carlo calculations with different primary energies can be plotted in one diagram by using reduced parameters m for the mass thickness and for the beam broadening. IY~, characteristic angle (f(~?~) = f (8 = 0")/2, f(8) scattering amplitude); A = xj(1 + n), total mean free path between elastic and inelastic scattering events ; x,, mean free path for elastic scattering; n = uinel/uel, ratio of inelastic and elastic total cross sections (n = 2-5 as a value for carbon). 88

9 Top bottom effect in STEM Information about x, and 8, can be obtained by electron transmission experiments (Reimer & Sommer, 1968) and fitting the experiments to Lenz theory (Lenz, 1954). We used the following values from measurements on carbon: Eo (kev) xr (pg/cm2) , rad) Results of the Monte Carlo calculations are plotted in Fig. 6b. It can be seen from this figure that to a first approximation, is proportional to m and therefore from equation 6: x1/2 pt2 = (pt> p. 9 rn m i rn Reduced thickness rn., Fig. 6. Beam broadening in reduced parameters; (a) experiment and (b) Monte Carlo calculations. f, 20 kv; 40 kv; o, 60 kv; 0, 80 kv; 0, 100 kv. 89

10 P. Gentsch, H. GiIde and L. Reimer At equal mass thickness pt electron beam broadening will thus only be important for materials with low density p, like aluminium, or biological sections. The results of experimental TBE values of Fig. 5 are plotted in Fig. 6a with reduced parameters. At first sight there seems to be only slight agreement between Fig. 6a and Fig. 6b. Firstly the Monte Carlo calculations show a strong dependence on the size of the objective diaphragm, whereas experimental values are independent within the accuracy of measurement. Secondly, the whole shape of the 5,,,(m) curve through the experimental points of Fig. 6a does not correspond to any one of the Monte Carlo curves in Fig. 6b. For this reason the effective aperture aeff in the STEM mode must be independent of the objective diaphragm. However it must depend on the specimen thickness and the energy of the primary electrons. The first point is to be expected, because in the STEM mode a large illumination I I c P 5 o c-,o-l t- TEM 10 * 10-3 Aperture a 10-I rod Fig. 7. Electron transmission measurements of the effective objective aperture aeff in STEM at 100 kv. aperture (18 = 2 x rad) is used. Furthermore the focal length of the objective lens is much shorter than in TEM (with the specimen at the same position) and therefore the first diffraction image is no longer in the plane of the objective diaphragm. As far as the second point is concerned, at the moment we do not know the reason for this dependence of aeff upon specimen thickness. Perhaps it is important that in STEM mode, the intermediate lens is used, corresponding to a 9000-fold TEM magnification, to image the transmitted electrons through the pole pieces onto the detector below the viewing screen. By this an unknown selection of scattering angles can occur. In order to establish the value of aeff, we performed electron transmission experiments with the same specimen as in the TBE measurements, but without an indium layer. We measured the ratio T = j/jo of current densities through the sphere (j) and beside it (jo) at 100 kv as a function of three sphere diameters and three objective diaphragms in the TEM mode (illumination angle rad). One can obtain an accurate value of the objective aperture by taking a selected area diffraction diagram of an evaporated gold film and further exposure of the photographic plates with the diaphragm. The TEM calibration curves obtained 90

11 Top bottom efsect in STEM by transmission experiments were used to determine the effective STEM aperture. The values of T in STEM were measured with the same spheres and diagraphragms as in TEM above using a Y-modulation linescan across the sphere (Fig. 7). As expected, transmission experiments show that T and therefore the effective observation apertures in the STEM mode are nearly independent of diaphragm size, but increase with decreasing thickness. This method of obtaining effective apertures by transmission is justified because the same Monte Carlo calculations lead to the right value of transmission. The transmission at reduced thickness my reduced aperture a/s, and different electron energies can also be plotted in one diagram like that of the reduced beam broadening (1/2 (Fig. 8a, b in Reimer et al., 1970). In Fig. 8, the beam broadening parameter xllg is plotted as a function of sphere diameter t at 100 kv. The heavy curve represents the experimental results, the other curves Monte Carlo results for the indicated apertures. The three points are Monte Carlo results at the effective apertures obtained by interpolation of the thin curves. In this way we get a good agreement between theory and experiment. 15'[ - Experiment / o Sphere diameter I (prn) Fig. 8. Comparison of experimental beam broadening (heavy curve) and Monte Carlo results (3 points) when using aeff from Fig. 7. The thin curves are Monte Carlo results with the indicated apertures. ACKNOWLEDGMENTS We should like to thank the Deutsche Forschungsgemeinschaft for the loan of a JEOL loob microscope and Mr Schwaab who helped us to take the 200 kv micrographs on a JEOL 200A. References Hashimoto, H. (1966) Proc. AMU-ANL Workshop on High Voltage Electron Microscopy, 68. Koike, S., Ueno, K., Suzuki, M., Matsuo, T., Aim, S. &z Shibatomi, K. (1971) High resolution scanning device for the JEM-IOOB electron microscope.jeo1 News, 9e, No. 3,

12 P. Gentsch, H. Gilde and L. Reimer Lenz, F. (1954) Zur Streuung mittelschneller Elektronen in kleinste Winkel. Z. Naturf. 9a, 185. Nauta, R. & Thompson, M.N. (1972) A new scanning electron microscope attachment for the Philips EM 300G. Proc. Fifth Eur. Congr. Electron. Microsc., Manchester, 46. Reimer, L. (1972) Physical limits in transmission scanning electron microscopy of thick specimens. Proc. Fifth Ann. SEM Symposium, 197. IIT Research Institute Chicago. Reimer, L. & Gilde, H. (1969) Die Verbreiterung des Elektronenstrahls durch Mehrfachstreuung und ihre Bedeutung fur die Raster-Elektronenmikroskopie. Beitr. elektronenmikroskop. Direktabbildung von Oberflachen, 2, 81. Reimer, L., Gilde, H. & Sommer, K.H. (1970) Die Verbreiterung eines Elektronenstrahles ( kv) durch Mehrfachstreuung. Optik 30, 590. Reimer, L. & Sommer, K.H. (1968) Messungen und Berechnungen zum elektronenmikroskopischen Streukontrast fur 17 bis 1200 kev Elektronen. Z. Naturf. 23a, Thieringer, H.M. (1972) Transmission scanning electron microscopy and energy analysis with the Siemens Elmiskop 101. Proc. Fifth Eur. Congr. Electron Microsc., Manchester,

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES Shortly after the experimental confirmation of the wave properties of the electron, it was suggested that the electron could be used to examine objects

More information

Chapter 1. Basic Electron Optics (Lecture 2)

Chapter 1. Basic Electron Optics (Lecture 2) Chapter 1. Basic Electron Optics (Lecture 2) Basic concepts of microscope (Cont ) Fundamental properties of electrons Electron Scattering Instrumentation Basic conceptions of microscope (Cont ) Ray diagram

More information

S200 Course LECTURE 1 TEM

S200 Course LECTURE 1 TEM S200 Course LECTURE 1 TEM Development of Electron Microscopy 1897 Discovery of the electron (J.J. Thompson) 1924 Particle and wave theory (L. de Broglie) 1926 Electromagnetic Lens (H. Busch) 1932 Construction

More information

NanoSpective, Inc Progress Drive Suite 137 Orlando, Florida

NanoSpective, Inc Progress Drive Suite 137 Orlando, Florida TEM Techniques Summary The TEM is an analytical instrument in which a thin membrane (typically < 100nm) is placed in the path of an energetic and highly coherent beam of electrons. Typical operating voltages

More information

Special Invited Review Scanning transmission electron microscopy*

Special Invited Review Scanning transmission electron microscopy* (0 Journal of Microscopy, Vol. 100, Pt 3, April 1974, pp. 247-259. Received 22 October 1973 Special Invited Review Scanning transmission electron microscopy* by ALBERT V. C R E w E, Departments of Physics

More information

ELECTRON MICROSCOPY. 13:10 16:00, Oct. 6, 2008 Institute of Physics, Academia Sinica. Tung Hsu

ELECTRON MICROSCOPY. 13:10 16:00, Oct. 6, 2008 Institute of Physics, Academia Sinica. Tung Hsu ELECTRON MICROSCOPY 13:10 16:00, Oct. 6, 2008 Institute of Physics, Academia Sinica Tung Hsu Department of Materials Science and Engineering National Tsing Hua University Hsinchu 300, TAIWAN Tel. 03-5742564

More information

NANO 703-Notes. Chapter 9-The Instrument

NANO 703-Notes. Chapter 9-The Instrument 1 Chapter 9-The Instrument Illumination (condenser) system Before (above) the sample, the purpose of electron lenses is to form the beam/probe that will illuminate the sample. Our electron source is macroscopic

More information

Chapter 2 Instrumentation for Analytical Electron Microscopy Lecture 7. Chapter 2 CHEM Fall L. Ma

Chapter 2 Instrumentation for Analytical Electron Microscopy Lecture 7. Chapter 2 CHEM Fall L. Ma Chapter 2 Instrumentation for Analytical Electron Microscopy Lecture 7 Outline Electron Sources (Electron Guns) Thermionic: LaB 6 or W Field emission gun: cold or Schottky Lenses Focusing Aberration Probe

More information

Transmission electron Microscopy

Transmission electron Microscopy Transmission electron Microscopy Image formation of a concave lens in geometrical optics Some basic features of the transmission electron microscope (TEM) can be understood from by analogy with the operation

More information

(Refer Slide Time: 00:10)

(Refer Slide Time: 00:10) Fundamentals of optical and scanning electron microscopy Dr. S. Sankaran Department of Metallurgical and Materials Engineering Indian Institute of Technology, Madras Module 03 Unit-6 Instrumental details

More information

Chapter 4 Imaging Lecture 17

Chapter 4 Imaging Lecture 17 Chapter 4 Imaging Lecture 17 d (110) Imaging Imaging in the TEM Diffraction Contrast in TEM Image HRTEM (High Resolution Transmission Electron Microscopy) Imaging STEM imaging Imaging in the TEM What is

More information

Electron

Electron Electron 1897: Sir Joseph John Thomson (1856-1940) discovered corpuscles small particles with a charge-to-mass ratio over 1000 times greater than that of protons. Plum pudding model : electrons in a sea

More information

MODULE I SCANNING ELECTRON MICROSCOPE (SEM)

MODULE I SCANNING ELECTRON MICROSCOPE (SEM) MODULE I SCANNING ELECTRON MICROSCOPE (SEM) Scanning Electron Microscope (SEM) Initially, the plan of SEM was offered by H. Stintzing in 1927 (a German patent application). His suggested procedure was

More information

Indiana University JEM-3200FS

Indiana University JEM-3200FS Indiana University JEM-3200FS Installation Specification Model: JEM 3200FS Serial Number: EM 15000013 Objective Lens Configuration: High Resolution Pole Piece (HRP) JEOL Engineer: Michael P. Van Etten

More information

SECONDARY ELECTRON DETECTION

SECONDARY ELECTRON DETECTION SECONDARY ELECTRON DETECTION CAMTEC Workshop Presentation Haitian Xu June 14 th 2010 Introduction SEM Raster scan specimen surface with focused high energy e- beam Signal produced by beam interaction with

More information

No part of this material may be reproduced without explicit written permission.

No part of this material may be reproduced without explicit written permission. This material is provided for educational use only. The information in these slides including all data, images and related materials are the property of : Robert M. Glaeser Department of Molecular & Cell

More information

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name:

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name: EE119 Introduction to Optical Engineering Spring 2003 Final Exam Name: SID: CLOSED BOOK. THREE 8 1/2 X 11 SHEETS OF NOTES, AND SCIENTIFIC POCKET CALCULATOR PERMITTED. TIME ALLOTTED: 180 MINUTES Fundamental

More information

Transmission Electron Microscopy 9. The Instrument. Outline

Transmission Electron Microscopy 9. The Instrument. Outline Transmission Electron Microscopy 9. The Instrument EMA 6518 Spring 2009 02/25/09 Outline The Illumination System The Objective Lens and Stage Forming Diffraction Patterns and Images Alignment and Stigmation

More information

Confocal Imaging Through Scattering Media with a Volume Holographic Filter

Confocal Imaging Through Scattering Media with a Volume Holographic Filter Confocal Imaging Through Scattering Media with a Volume Holographic Filter Michal Balberg +, George Barbastathis*, Sergio Fantini % and David J. Brady University of Illinois at Urbana-Champaign, Urbana,

More information

PROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, Edited by

PROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, Edited by X - R A Y M I C R O S C O P Y A N D M I C R O R A D I O G R A P H Y PROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, 1956 Edited by V. E. COSSLETT Cavendish Laboratory, University

More information

Physics 431 Final Exam Examples (3:00-5:00 pm 12/16/2009) TIME ALLOTTED: 120 MINUTES Name: Signature:

Physics 431 Final Exam Examples (3:00-5:00 pm 12/16/2009) TIME ALLOTTED: 120 MINUTES Name: Signature: Physics 431 Final Exam Examples (3:00-5:00 pm 12/16/2009) TIME ALLOTTED: 120 MINUTES Name: PID: Signature: CLOSED BOOK. TWO 8 1/2 X 11 SHEET OF NOTES (double sided is allowed), AND SCIENTIFIC POCKET CALCULATOR

More information

Applications of Optics

Applications of Optics Nicholas J. Giordano www.cengage.com/physics/giordano Chapter 26 Applications of Optics Marilyn Akins, PhD Broome Community College Applications of Optics Many devices are based on the principles of optics

More information

Chapter Ray and Wave Optics

Chapter Ray and Wave Optics 109 Chapter Ray and Wave Optics 1. An astronomical telescope has a large aperture to [2002] reduce spherical aberration have high resolution increase span of observation have low dispersion. 2. If two

More information

Introduction of New Products

Introduction of New Products Field Emission Electron Microscope JEM-3100F For evaluation of materials in the fields of nanoscience and nanomaterials science, TEM is required to provide resolution and analytical capabilities that can

More information

(1) Research Institute for Scientific Measurements, Tohoku University, Katahira 2-1-1, Aoba-ku,

(1) Research Institute for Scientific Measurements, Tohoku University, Katahira 2-1-1, Aoba-ku, 351 Classification Physics Abstracts 07.80 Performance of a new high-resolution electron energy-loss spectroscopy microscope Masami Thrauchi(1), Ryuichi Kuzuo(1), Futami Satoh(1), Michiyoshi Thnaka(1),

More information

Low Voltage Electron Microscope

Low Voltage Electron Microscope LVEM5 Low Voltage Electron Microscope Nanoscale from your benchtop LVEM5 Delong America DELONG INSTRUMENTS COMPACT BUT POWERFUL The LVEM5 is designed to excel across a broad range of applications in material

More information

Functions of the SEM subsystems

Functions of the SEM subsystems Functions of the SEM subsystems Electronic column It consists of an electron gun and two or more electron lenses, which influence the path of electrons traveling down an evacuated tube. The base of the

More information

Observing Microorganisms through a Microscope LIGHT MICROSCOPY: This type of microscope uses visible light to observe specimens. Compound Light Micros

Observing Microorganisms through a Microscope LIGHT MICROSCOPY: This type of microscope uses visible light to observe specimens. Compound Light Micros PHARMACEUTICAL MICROBIOLOGY JIGAR SHAH INSTITUTE OF PHARMACY NIRMA UNIVERSITY Observing Microorganisms through a Microscope LIGHT MICROSCOPY: This type of microscope uses visible light to observe specimens.

More information

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. JY/T

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. JY/T Translated English of Chinese Standard: JY/T011-1996 www.chinesestandard.net Sales@ChineseStandard.net INDUSTRY STANDARD OF THE JY PEOPLE S REPUBLIC OF CHINA General rules for transmission electron microscopy

More information

ELECTRON MICROSCOPY. 14:10 17:00, Apr. 3, 2007 Department of Physics, National Taiwan University. Tung Hsu

ELECTRON MICROSCOPY. 14:10 17:00, Apr. 3, 2007 Department of Physics, National Taiwan University. Tung Hsu ELECTRON MICROSCOPY 14:10 17:00, Apr. 3, 2007 Department of Physics, National Taiwan University Tung Hsu Department of Materials Science and Engineering National Tsinghua University Hsinchu 300, TAIWAN

More information

Introduction to Transmission Electron Microscopy (Physical Sciences)

Introduction to Transmission Electron Microscopy (Physical Sciences) Introduction to Transmission Electron Microscopy (Physical Sciences) Centre for Advanced Microscopy Program 9:30 10:45 Lecture 1 Basics of TEM 10:45 11:00 Morning tea 11:00 12:15 Lecture 2 Diffraction

More information

ELECTRON MICROSCOPY AN OVERVIEW

ELECTRON MICROSCOPY AN OVERVIEW ELECTRON MICROSCOPY AN OVERVIEW Anjali Priya 1, Abhishek Singh 2, Nikhil Anand Srivastava 3 1,2,3 Department of Electrical & Instrumentation, Sant Longowal Institute of Engg. & Technology, Sangrur, India.

More information

Low Voltage Electron Microscope

Low Voltage Electron Microscope LVEM 25 Low Voltage Electron Microscope fast compact powerful Delong America FAST, COMPACT AND POWERFUL The LVEM 25 offers a high-contrast, high-throughput, and compact solution with nanometer resolutions.

More information

Applied Optics. , Physics Department (Room #36-401) , ,

Applied Optics. , Physics Department (Room #36-401) , , Applied Optics Professor, Physics Department (Room #36-401) 2290-0923, 019-539-0923, shsong@hanyang.ac.kr Office Hours Mondays 15:00-16:30, Wednesdays 15:00-16:30 TA (Ph.D. student, Room #36-415) 2290-0921,

More information

Scanning Electron Microscopy. EMSE-515 F. Ernst

Scanning Electron Microscopy. EMSE-515 F. Ernst Scanning Electron Microscopy EMSE-515 F. Ernst 1 2 Scanning Electron Microscopy Max Knoll Manfred von Ardenne Manfred von Ardenne Principle of Scanning Electron Microscopy 3 Principle of Scanning Electron

More information

LVEM 25. Low Voltage Electron Mictoscope. fast compact powerful

LVEM 25. Low Voltage Electron Mictoscope. fast compact powerful LVEM 25 Low Voltage Electron Mictoscope fast compact powerful FAST, COMPACT AND POWERFUL The LVEM 25 offers a high-contrast, high-throughput, and compact solution with nanometer resolutions. All the benefits

More information

BMB/Bi/Ch 173 Winter 2018

BMB/Bi/Ch 173 Winter 2018 BMB/Bi/Ch 73 Winter 208 Homework Set 2 (200 Points) Assigned -7-8, due -23-8 by 0:30 a.m. TA: Rachael Kuintzle. Office hours: SFL 229, Friday /9 4:00-5:00pm and SFL 220, Monday /22 4:00-5:30pm. For the

More information

Chapter 2 The Study of Microbial Structure: Microscopy and Specimen Preparation

Chapter 2 The Study of Microbial Structure: Microscopy and Specimen Preparation Chapter 2 The Study of Microbial Structure: Microscopy and Specimen Preparation 1 Lenses and the Bending of Light light is refracted (bent) when passing from one medium to another refractive index a measure

More information

Chapter 25. Optical Instruments

Chapter 25. Optical Instruments Chapter 25 Optical Instruments Optical Instruments Analysis generally involves the laws of reflection and refraction Analysis uses the procedures of geometric optics To explain certain phenomena, the wave

More information

Introduction to Electron Microscopy-II

Introduction to Electron Microscopy-II Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built the first electron microscope in 1931 (Nobel Prize to Ruska in 1986)

More information

Scanning electron microscope

Scanning electron microscope Scanning electron microscope 5 th CEMM workshop Maja Koblar, Sc. Eng. Physics Outline The basic principle? What is an electron? Parts of the SEM Electron gun Electromagnetic lenses Apertures Detectors

More information

INTRODUCTION THIN LENSES. Introduction. given by the paraxial refraction equation derived last lecture: Thin lenses (19.1) = 1. Double-lens systems

INTRODUCTION THIN LENSES. Introduction. given by the paraxial refraction equation derived last lecture: Thin lenses (19.1) = 1. Double-lens systems Chapter 9 OPTICAL INSTRUMENTS Introduction Thin lenses Double-lens systems Aberrations Camera Human eye Compound microscope Summary INTRODUCTION Knowledge of geometrical optics, diffraction and interference,

More information

Low-energy Electron Diffractive Imaging for Three dimensional Light-element Materials

Low-energy Electron Diffractive Imaging for Three dimensional Light-element Materials Low-energy Electron Diffractive Imaging for Three dimensional Light-element Materials Hitachi Review Vol. 61 (2012), No. 6 269 Osamu Kamimura, Ph. D. Takashi Dobashi OVERVIEW: Hitachi has been developing

More information

SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS

SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS Robert Edward Lee Electron Microscopy Center Department of Anatomy and Neurobiology Colorado State University P T R Prentice Hall, Englewood Cliffs,

More information

Observing Microorganisms through a Microscope

Observing Microorganisms through a Microscope 2016/2/19 PowerPoint Lecture Presentations prepared by Bradley W. Christian, McLennan Community College CHAPTER 3 Observing Microorganisms through a Microscope 1 Figure 3.2 Microscopes and Magnification.

More information

Low Voltage Electron Microscope. Nanoscale from your benchtop LVEM5. Delong America

Low Voltage Electron Microscope. Nanoscale from your benchtop LVEM5. Delong America LVEM5 Low Voltage Electron Microscope Nanoscale from your benchtop LVEM5 Delong America DELONG INSTRUMENTS COMPACT BUT POWERFUL The LVEM5 is designed to excel across a broad range of applications in material

More information

The extended-focus, auto-focus and surface-profiling techniques of confocal microscopy

The extended-focus, auto-focus and surface-profiling techniques of confocal microscopy JOURNAL OF MODERN OPTICS, 1988, voi,. 35, NO. 1, 145-154 The extended-focus, auto-focus and surface-profiling techniques of confocal microscopy C. J. R. SHEPPARD and H. J. MATTHEWS University of Oxford,

More information

Atomic Resolution Imaging with a sub-50 pm Electron Probe

Atomic Resolution Imaging with a sub-50 pm Electron Probe Atomic Resolution Imaging with a sub-50 pm Electron Probe Rolf Erni, Marta D. Rossell, Christian Kisielowski, Ulrich Dahmen National Center for Electron Microscopy, Lawrence Berkeley National Laboratory

More information

A Tutorial on Electron Microscopy

A Tutorial on Electron Microscopy A Tutorial on Electron Microscopy Jian-Min (Jim) Zuo Mat. Sci. Eng. and Seitz-Materials Research Lab., UIUC Outline of This Tutorial I. Science and opportunities of electron microscopy II. The basic TEM,

More information

ELECTRON MICROSCOPY. 09:10 12:00, Oct. 27, 2006 Institute of Physics, Academia Sinica. Tung Hsu

ELECTRON MICROSCOPY. 09:10 12:00, Oct. 27, 2006 Institute of Physics, Academia Sinica. Tung Hsu ELECTRON MICROSCOPY 09:10 12:00, Oct. 27, 2006 Institute of Physics, Academia Sinica Tung Hsu Department of Materials Science and Engineering National Tsinghua University Hsinchu 300, TAIWAN Tel. 03-5742564

More information

attosnom I: Topography and Force Images NANOSCOPY APPLICATION NOTE M06 RELATED PRODUCTS G

attosnom I: Topography and Force Images NANOSCOPY APPLICATION NOTE M06 RELATED PRODUCTS G APPLICATION NOTE M06 attosnom I: Topography and Force Images Scanning near-field optical microscopy is the outstanding technique to simultaneously measure the topography and the optical contrast of a sample.

More information

A Portable Scanning Electron Microscope Column Design Based on the Use of Permanent Magnets

A Portable Scanning Electron Microscope Column Design Based on the Use of Permanent Magnets SCANNING VOL. 20, 87 91 (1998) Received October 8, 1997 FAMS, Inc. Accepted with revision November 9, 1997 A Portable Scanning Electron Microscope Column Design Based on the Use of Permanent Magnets A.

More information

Laser Speckle Reducer LSR-3000 Series

Laser Speckle Reducer LSR-3000 Series Datasheet: LSR-3000 Series Update: 06.08.2012 Copyright 2012 Optotune Laser Speckle Reducer LSR-3000 Series Speckle noise from a laser-based system is reduced by dynamically diffusing the laser beam. A

More information

Exam 4. Name: Class: Date: Multiple Choice Identify the choice that best completes the statement or answers the question.

Exam 4. Name: Class: Date: Multiple Choice Identify the choice that best completes the statement or answers the question. Name: Class: Date: Exam 4 Multiple Choice Identify the choice that best completes the statement or answers the question. 1. Mirages are a result of which physical phenomena a. interference c. reflection

More information

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy, KTH Applied Physics Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy, 2009-06-05, 8-13, FB51 Allowed aids: Compendium Imaging Physics (handed out) Compendium Light Microscopy

More information

High Resolution Transmission Electron Microscopy (HRTEM) Summary 4/11/2018. Thomas LaGrange Faculty Lecturer and Senior Staff Scientist

High Resolution Transmission Electron Microscopy (HRTEM) Summary 4/11/2018. Thomas LaGrange Faculty Lecturer and Senior Staff Scientist Thomas LaGrange Faculty Lecturer and Senior Staff Scientist High Resolution Transmission Electron Microscopy (HRTEM) Doctoral Course MS-637 April 16-18th, 2018 Summary Contrast in TEM images results from

More information

Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers.

Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers. Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers. Finite-difference time-domain calculations of the optical transmittance through

More information

Microscopy Techniques that make it easy to see things this small.

Microscopy Techniques that make it easy to see things this small. Microscopy Techniques that make it easy to see things this small. What is a Microscope? An instrument for viewing objects that are too small to be seen easily by the naked eye. Dutch spectacle-makers Hans

More information

EXPRIMENT 3 COUPLING FIBERS TO SEMICONDUCTOR SOURCES

EXPRIMENT 3 COUPLING FIBERS TO SEMICONDUCTOR SOURCES EXPRIMENT 3 COUPLING FIBERS TO SEMICONDUCTOR SOURCES OBJECTIVES In this lab, firstly you will learn to couple semiconductor sources, i.e., lightemitting diodes (LED's), to optical fibers. The coupling

More information

5. The Scanning Electron Microscope

5. The Scanning Electron Microscope Physical Principles of Electron Microscopy 5. The Scanning Electron Microscope Ray Egerton University of Alberta and National Institute of Nanotechnology Edmonton, Canada www.tem-eels.ca regerton@ualberta.ca

More information

Introduction to Electron Microscopy

Introduction to Electron Microscopy Introduction to Electron Microscopy Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built the first electron microscope in 1931 (Nobel Prize to Ruska in 1986)

More information

Chapter 18 Optical Elements

Chapter 18 Optical Elements Chapter 18 Optical Elements GOALS When you have mastered the content of this chapter, you will be able to achieve the following goals: Definitions Define each of the following terms and use it in an operational

More information

Recent results from the JEOL JEM-3000F FEGTEM in Oxford

Recent results from the JEOL JEM-3000F FEGTEM in Oxford Recent results from the JEOL JEM-3000F FEGTEM in Oxford R.E. Dunin-Borkowski a, J. Sloan b, R.R. Meyer c, A.I. Kirkland c,d and J. L. Hutchison a a b c d Department of Materials, Parks Road, Oxford OX1

More information

Scanning Electron Microscopy SEM. Warren Straszheim, PhD MARL, 23 Town Engineering

Scanning Electron Microscopy SEM. Warren Straszheim, PhD MARL, 23 Town Engineering Scanning Electron Microscopy SEM Warren Straszheim, PhD MARL, 23 Town Engineering wesaia@iastate.edu 515-294-8187 How it works Create a focused electron beam Accelerate it Scan it across the sample Map

More information

Very short introduction to light microscopy and digital imaging

Very short introduction to light microscopy and digital imaging Very short introduction to light microscopy and digital imaging Hernan G. Garcia August 1, 2005 1 Light Microscopy Basics In this section we will briefly describe the basic principles of operation and

More information

Use of Back Scattered Ionizing Radiation for Measurement of Thickness of the Catalytic Agent Active Material

Use of Back Scattered Ionizing Radiation for Measurement of Thickness of the Catalytic Agent Active Material 18th World Conference on Nondestructive Testing, 16- April 1, Durban, South Africa Use of Back Scattered Ionizing Radiation for Measurement of Thickness of the Catalytic Agent Active Material Boris V.

More information

FIRST INDIRECT X-RAY IMAGING TESTS WITH AN 88-mm DIAMETER SINGLE CRYSTAL

FIRST INDIRECT X-RAY IMAGING TESTS WITH AN 88-mm DIAMETER SINGLE CRYSTAL FERMILAB-CONF-16-641-AD-E ACCEPTED FIRST INDIRECT X-RAY IMAGING TESTS WITH AN 88-mm DIAMETER SINGLE CRYSTAL A.H. Lumpkin 1 and A.T. Macrander 2 1 Fermi National Accelerator Laboratory, Batavia, IL 60510

More information

Compact OAM Microscope for Edge Enhancement of Biomedical and Object Samples

Compact OAM Microscope for Edge Enhancement of Biomedical and Object Samples Compact OAM Microscope for Edge Enhancement of Biomedical and Object Samples Richard Gozali, 1 Thien-An Nguyen, 1 Ethan Bendau, 1 Robert R. Alfano 1,b) 1 City College of New York, Institute for Ultrafast

More information

The Resolution in the Electron Microscopy

The Resolution in the Electron Microscopy Volume 3, Issue, February 1 ISSN 319-87 The Resolution in the Electron Microscopy ABSTRACT Benefit from the group's equations, especially the resolution limits in the transmission electron microscope (TEM)

More information

plasmonic nanoblock pair

plasmonic nanoblock pair Nanostructured potential of optical trapping using a plasmonic nanoblock pair Yoshito Tanaka, Shogo Kaneda and Keiji Sasaki* Research Institute for Electronic Science, Hokkaido University, Sapporo 1-2,

More information

p q p f f f q f p q f NANO 703-Notes Chapter 5-Magnification and Electron Sources

p q p f f f q f p q f NANO 703-Notes Chapter 5-Magnification and Electron Sources Chapter 5-agnification and Electron Sources Lens equation Let s first consider the properties of an ideal lens. We want rays diverging from a point on an object in front of the lens to converge to a corresponding

More information

Reflection! Reflection and Virtual Image!

Reflection! Reflection and Virtual Image! 1/30/14 Reflection - wave hits non-absorptive surface surface of a smooth water pool - incident vs. reflected wave law of reflection - concept for all electromagnetic waves - wave theory: reflected back

More information

Project Staff: Feng Zhang, Prof. Jianfeng Dai (Lanzhou Univ. of Tech.), Prof. Todd Hasting (Univ. Kentucky), Prof. Henry I. Smith

Project Staff: Feng Zhang, Prof. Jianfeng Dai (Lanzhou Univ. of Tech.), Prof. Todd Hasting (Univ. Kentucky), Prof. Henry I. Smith 3. Spatial-Phase-Locked Electron-Beam Lithography Sponsors: No external sponsor Project Staff: Feng Zhang, Prof. Jianfeng Dai (Lanzhou Univ. of Tech.), Prof. Todd Hasting (Univ. Kentucky), Prof. Henry

More information

Design and Application of a Quadrupole Detector for Low-Voltage Scanning Electron Mcroscopy

Design and Application of a Quadrupole Detector for Low-Voltage Scanning Electron Mcroscopy SCANNING Vol. 8, 294-299 (1986) 0 FACM. Inc. Received: August 29, 1986 Original Paper Design and Application of a Quadrupole Detector for Low-Voltage Scanning Electron Mcroscopy R. Schmid and M. Brunner"

More information

Tangents. The f-stops here. Shedding some light on the f-number. by Marcus R. Hatch and David E. Stoltzmann

Tangents. The f-stops here. Shedding some light on the f-number. by Marcus R. Hatch and David E. Stoltzmann Tangents Shedding some light on the f-number The f-stops here by Marcus R. Hatch and David E. Stoltzmann The f-number has peen around for nearly a century now, and it is certainly one of the fundamental

More information

OPTICAL PRINCIPLES OF MICROSCOPY. Interuniversity Course 28 December 2003 Aryeh M. Weiss Bar Ilan University

OPTICAL PRINCIPLES OF MICROSCOPY. Interuniversity Course 28 December 2003 Aryeh M. Weiss Bar Ilan University OPTICAL PRINCIPLES OF MICROSCOPY Interuniversity Course 28 December 2003 Aryeh M. Weiss Bar Ilan University FOREWORD This slide set was originally presented at the ISM Workshop on Theoretical and Experimental

More information

MCR Scanning Electron Microscopy Laboratory Portfolio

MCR Scanning Electron Microscopy Laboratory Portfolio SUNY College of Environmental Science and Forestry Digital Commons @ ESF N.C. Brown Center for Ultrastructure Studies Fall 2016 MCR 484 - Scanning Electron Microscopy Laboratory Portfolio Timothy Gervascio

More information

2.Components of an electron microscope. a) vacuum systems, b) electron guns, c) electron optics, d) detectors. Marco Cantoni 021/

2.Components of an electron microscope. a) vacuum systems, b) electron guns, c) electron optics, d) detectors. Marco Cantoni 021/ 2.Components of an electron microscope a) vacuum systems, b) electron guns, c) electron optics, d) detectors, 021/693.48.16 Centre Interdisciplinaire de Microscopie Electronique CIME Summary Electron propagation

More information

MASSACHUSETTS INSTITUTE OF TECHNOLOGY. 2.71/2.710 Optics Spring 14 Practice Problems Posted May 11, 2014

MASSACHUSETTS INSTITUTE OF TECHNOLOGY. 2.71/2.710 Optics Spring 14 Practice Problems Posted May 11, 2014 MASSACHUSETTS INSTITUTE OF TECHNOLOGY 2.71/2.710 Optics Spring 14 Practice Problems Posted May 11, 2014 1. (Pedrotti 13-21) A glass plate is sprayed with uniform opaque particles. When a distant point

More information

Scanning Electron Microscopy Student Image Portfolio

Scanning Electron Microscopy Student Image Portfolio SUNY College of Environmental Science and Forestry Digital Commons @ ESF N.C. Brown Center for Ultrastructure Studies Fall 12-7-2016 Scanning Electron Microscopy Student Image Portfolio Matthew DaRin SUNY

More information

TEM theory Basic optics, image formation and key elements

TEM theory Basic optics, image formation and key elements Workshop series of Chinese 3DEM community Get acquainted with Cryo-Electron Microscopy: First Chinese Workshop for Structural Biologists TEM theory Basic optics, image formation and key elements Jianlin

More information

Scanning electron microscope

Scanning electron microscope Scanning electron microscope 6 th CEMM workshop Maja Koblar, Sc. Eng. Physics Outline The basic principle? What is an electron? Parts of the SEM Electron gun Electromagnetic lenses Apertures Chamber and

More information

Secondary Electron Detector

Secondary Electron Detector Secondary Electron Detector Fig. 17 Everhart-Thornley Detector (Fig. 7-9, p. 215, Bozzola and Russell) Secondary electrons (SE) are attracted to Faraday cage because of its positive charge. Detector surface

More information

Basics of Light Microscopy and Metallography

Basics of Light Microscopy and Metallography ENGR45: Introduction to Materials Spring 2012 Laboratory 8 Basics of Light Microscopy and Metallography In this exercise you will: gain familiarity with the proper use of a research-grade light microscope

More information

Introduction to Scanning Electron Microscopy

Introduction to Scanning Electron Microscopy Introduction to Scanning Electron Microscopy By: Brandon Cheney Ant s Leg Integrated Circuit Nano-composite This document was created as part of a Senior Project in the Materials Engineering Department

More information

X-ray backscattering: Variable irradiation geometry facilitates new insights

X-ray backscattering: Variable irradiation geometry facilitates new insights 18 th World Conference of Non Destructive Testing, 16-20 April 2012, Durban, South Africa X-ray backscattering: Variable irradiation geometry facilitates new insights Norma WROBEL 1, Kurt OSTERLOH 1, Mirko

More information

Nanotechnology in Consumer Products

Nanotechnology in Consumer Products Nanotechnology in Consumer Products Advances in Transmission Electron Microscopy Friday, April 21, 2017 October 31, 2014 The webinar will begin at 1pm Eastern Time Click here to watch the webinar recording

More information

A Parallel Radial Mirror Energy Analyzer Attachment for the Scanning Electron Microscope

A Parallel Radial Mirror Energy Analyzer Attachment for the Scanning Electron Microscope 142 doi:10.1017/s1431927615013288 Microscopy Society of America 2015 A Parallel Radial Mirror Energy Analyzer Attachment for the Scanning Electron Microscope Kang Hao Cheong, Weiding Han, Anjam Khursheed

More information

Optical design of a high resolution vision lens

Optical design of a high resolution vision lens Optical design of a high resolution vision lens Paul Claassen, optical designer, paul.claassen@sioux.eu Marnix Tas, optical specialist, marnix.tas@sioux.eu Prof L.Beckmann, l.beckmann@hccnet.nl Summary:

More information

The diffraction of light

The diffraction of light 7 The diffraction of light 7.1 Introduction As introduced in Chapter 6, the reciprocal lattice is the basis upon which the geometry of X-ray and electron diffraction patterns can be most easily understood

More information

IMAGE SENSOR SOLUTIONS. KAC-96-1/5" Lens Kit. KODAK KAC-96-1/5" Lens Kit. for use with the KODAK CMOS Image Sensors. November 2004 Revision 2

IMAGE SENSOR SOLUTIONS. KAC-96-1/5 Lens Kit. KODAK KAC-96-1/5 Lens Kit. for use with the KODAK CMOS Image Sensors. November 2004 Revision 2 KODAK for use with the KODAK CMOS Image Sensors November 2004 Revision 2 1.1 Introduction Choosing the right lens is a critical aspect of designing an imaging system. Typically the trade off between image

More information

Measurement of the Modulation Transfer Function (MTF) of a camera lens. Laboratoire d Enseignement Expérimental (LEnsE)

Measurement of the Modulation Transfer Function (MTF) of a camera lens. Laboratoire d Enseignement Expérimental (LEnsE) Measurement of the Modulation Transfer Function (MTF) of a camera lens Aline Vernier, Baptiste Perrin, Thierry Avignon, Jean Augereau, Lionel Jacubowiez Institut d Optique Graduate School Laboratoire d

More information

ECEN 4606, UNDERGRADUATE OPTICS LAB

ECEN 4606, UNDERGRADUATE OPTICS LAB ECEN 4606, UNDERGRADUATE OPTICS LAB Lab 2: Imaging 1 the Telescope Original Version: Prof. McLeod SUMMARY: In this lab you will become familiar with the use of one or more lenses to create images of distant

More information

A few concepts in TEM and STEM explained

A few concepts in TEM and STEM explained A few concepts in TEM and STEM explained Martin Ek November 23, 2011 1 Introduction This is a collection of short, qualitative explanations of key concepts in TEM and STEM. Most of them are beyond what

More information

UNIVERSITY OF FLORIDA 1978

UNIVERSITY OF FLORIDA 1978 THE EFFECT OF CRYSTAL DEFECTS ON MICRODIFFRACTION PATTERNS BY JOHN BEVERLY WARREN A DISSERTATION PRESENTED TO THE GRADUATE COUNCIL OF THE UNIVERSITY OF FLORIDA IN PARTIAL FULFILLMENT OF THE REQUIREMENTS

More information

EE119 Introduction to Optical Engineering Fall 2009 Final Exam. Name:

EE119 Introduction to Optical Engineering Fall 2009 Final Exam. Name: EE119 Introduction to Optical Engineering Fall 2009 Final Exam Name: SID: CLOSED BOOK. THREE 8 1/2 X 11 SHEETS OF NOTES, AND SCIENTIFIC POCKET CALCULATOR PERMITTED. TIME ALLOTTED: 180 MINUTES Fundamental

More information

New 500 kv Electron Microscope

New 500 kv Electron Microscope New 500 kv Electron Microscope B. T ADANO, H. KIMURA, S. KATAGIRI, M. NISHIGAKI Hitachi Central Research Laboratory, Kokubunji, Tokyo and R. UYEDA, Y. SAKAKI, S. MARUSE, K. MIHAMA, Y. KAMIYA Nagoya University,

More information

OPTICAL SYSTEMS OBJECTIVES

OPTICAL SYSTEMS OBJECTIVES 101 L7 OPTICAL SYSTEMS OBJECTIVES Aims Your aim here should be to acquire a working knowledge of the basic components of optical systems and understand their purpose, function and limitations in terms

More information

DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS. GUI Simulation Diffraction: Focused Beams and Resolution for a lens system

DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS. GUI Simulation Diffraction: Focused Beams and Resolution for a lens system DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS GUI Simulation Diffraction: Focused Beams and Resolution for a lens system Ian Cooper School of Physics University of Sydney ian.cooper@sydney.edu.au DOWNLOAD

More information

Modulation Transfer Function

Modulation Transfer Function Modulation Transfer Function The resolution and performance of an optical microscope can be characterized by a quantity known as the modulation transfer function (MTF), which is a measurement of the microscope's

More information