1 Stealth dicer DISCO DFL 7340 user manual Eric Lebrasseur (Mita Lab) 2016/02/24 Please, leave this manual here. You can find an electronic version in the TakedaCR Wiki: (accessible only from inside Utokyo) This equipment separates chips by using a laser beam. The stealth dicing method forms a modified layer in the work piece, by focusing a laser inside the work piece, and then implements die separation with a machine such as a tape expander. Merits of stealth dicing: Able to suppress cutting waste because it modifies an internal portion of the work piece. Therefore, it is suitable for work pieces that are vulnerable to contamination. It is a dry process that does not require cleaning; therefore it is suitable to processes (such as MEMS) that are vulnerable to load. Greatly contributes to street reduction because the kerf width can be made narrow. If there is a small amount of resist over the dicing zone or another transparent layer, it maybe possible to dice, but if there is some material that reflects the laser beam (like metal), dicing is not possible. The 6 inch (150mm) chuck is normally set up, but it can be replaced with an 8 inch (200mm) chuck in order to process 8 inch wafers (cf. 10). There are particular cut techniques that enable circular cut (cf. 11) and dotted line cut. 1. Starting up 1.1. Power ON Switch on the tape mounter (red button) (it needs about 10min to reach 50 C) Distribution Board B2P-7: Stealth Dicer Breaker ON On the wall behind the machine: Open the Air and Nitrogen valves, Open the two cooling water valves, both at the same time (close in horizontal position and open in vertical position) On the left side of the machine: Breaker ON On the front side of the machine: Turn the key to position START until POWER LAMP turns on Wait until the soft finish to start up 1.2. Initialization of the machine From now on, unless otherwise noted, we will talk about the front side screen. This screen is a touch screen, so click means tap actually Click on "System initialization" (" システムイニシャル ") Wait until the message "System Initialization Completed" (" システムイニシャル終了 ") appears on the top part of the screen Click on "Diode" (" ダイオード "): set ON Laser Setup (" レーザセットアップ ") "Laser Idling" (" レーザアイドリング ") START When the laser idling is completed (~10min), an alarm will ring. Click on "Alarm/Clear" (" アラーム クリア ") Click "Enter" "Exit" to return to the laser setup screen Laser Setup (" レーザセットアップ ") "Full Power Check" (" フルパワーチェック ") "START/STOP". Write the result in the log then click Exit (if the "output of the processing spot" ( 加工点出力 ) is less than 1.2W, contact the person in charge) Antenna maintenance ( アッテネータメンテナンス ) Normally not used. If you change some parameters here (mostly sample thickness), set them back after use. Default parameters are: Sample thickness > 100µm Checkboxes BP calibration (BP キャリブレーション ) and Antenna calibration ( アッテネータキャリブレーション ) are checked 1 / 5

2 Set the device data according to Power Setup Be sure to have set up and selected the device data before next step. If not, do it now (cf. 2) Laser setup ( レーザセットアップ ) "Power Setup" (" パワーセットアップ ") "START/STOP" When finished, click "EXIT" several times to go back to the top menu. 2. Setting the device data To enter numbers, push the button "123" at the bottom right of the screen. Then a numeric keypad will appear. If you push the button "abc", an alphabetic keypad will appear. You can change between capital and small letters by pressing "Shift" From the top menu, click "Device Data" (" デバイス データ ") 2.2. Select the data file that you want to use and click "Enter" (to create a new data file, chose an existing data file, then click "copy" (" コピー ") and indicate the copy folder and device ID, maximum of 8 characters) 2.3. Set the work shape and size Work shape (" ワーク形状 "): "ROUND" for a circle, "SQUARE" for a rectangle Work size (" ワークサイズ "): Set the size of the sample 2.4. Click "Street Cut Sequence" (" ストリートカットシーケンス ") to set the depth of the laser exposure If you check "Valid/Invalid" (" 有効 無効 "), the cut will become valid. The first one at the left should always be checked with the items fixed to 0W/0.00mm Amount of Defocus (" デフォーカス量 "): it expresses the depth of the laser exposure (0 is the surface. The higher is the value, the deeper is the position. Unit is mm). However, this value is not the real depth: in the case of Si the value should be multiplied by 4 to get the real depth (to take into account the refractive index of the material) Click "Sub index" (" サブインデックス ") to set the number of cutting lines, the space between the cutting lines etc (in the case of a circular cut, ignore the sub index and set up the circular cut according to 11) Number of cutting lines (" カット本数 "): set the number of lines to dice. If the number is set to 0, if at all possible it will cut within the sample size Cutting direction (" カット方向 "): "REAR" means "from your position to the inside part of the machine" (on the screen, from the bottom to the top), "FRONT" means "from inside the machine to your position" (on the screen, from the top to the bottom) Cutting sequence (Channel) ( カット順序 ( チャンネル )): set the channel that you want to use. "12" means that you will use CH1 (horizontal) and CH2 (vertical). "1" means that you will use only one channel CH1 (horizontal), that is you will cut in only one direction Y Index ("Y インデックス "): the space between the cutting lines 2.6. When the setup is finished click "Enter". The setup will be saved Click several times on "Exit" to return to the top menu. 3. Setting the sample 3.1. Setting the sample on the plastic film. This section describes how to fix the wafer on the plastic film using the tape mounter equipment which is placed on front of the dicing saw. For this purpose, we use 6inch metallic frames which are stored on the dicing saw Switch on the "POWER" of the tape mounter and wait until the temperature reaches 50 C Switch on the vacuum pump under the desk Adjust the metallic frame on the tape mounter by matching it to the pins Switch on the 4 breakers in front of the machine Pull the film until the front of the machine and fix it to the metallic frame using the roller Cut the film along the metallic frame with a cutter and throw away the unneeded film Turn the metallic frame upside down Put the sample in the center of the frame. In the case of a wafer, the flat should be placed on your side Improve the adhesion of the film to the wafer by rubbing it with your fingers, from the backside. If the air bubbles are two big, you can cut it with the cutter and flatten Setting the sample in the dicer Click "Shutter" (" シャッタ "): it turns OFF the light blue square at the top left of the shutter button. 2 / 5

3 Open the door at the right side of the equipment and put the film with the sample on it on the chuck. CH1 correspond to the direction from the front side to the back side of the equipment while CH2 correspond to the direction from the left side to the right side Close the door. Click "C/T vacuum" ("C/T バキューム ") and "Shutter" (" シャッタ "): check that the squares at the top left of the buttons turn on (light blue color). If it doesn't, check if there are no air bubbles at the back of the sample. 4. Laser focus 4.1. Move to the position of the cutting line In the top menu, click "Manual operation" (" マニュアルオペレーション ") and "Autocut" (" オートカット ") Click "Just Focus" Click "Move the position of the alignment camera" (" アライメントカメラ位置移動 ") Adjustment of the microscope focus: "Change magnification" (" 倍率変更 "): change magnification from "HI" (for high magnification) to "LO" (for low magnification) Click "Intensity" (" 光量 ") and adjust the light to a suitable intensity. Then "Exit" Click "Focus" (" フォーカス ") and "Autofocus" (" オートフォーカス "). When the autofocus is completed, click "Exit" Move to the position of your cutting line (refer to the explanations at the end of this section about how to change the position) Click "Exit" to return to the focus adjustment screen Adjust the focus the laser In the laser adjustment screen, click "Focus" (" フォーカス ") then "Full". A kind of square target like shape should appear. If not, try with Work surface Click "SLD ON/OFF": check that the light hit the sample inside the green square Click "Sensor" (" センサー ") and "LASER" Check that the light quantity is more than 1.0V. It will not cut at a lower quantity. For a silicon wafer, we expect a quantity larger than 2V Click "Exit" to go back to the laser adjustment screen Click "Save Laser Focus" (" レーザ焦点保存 ") than "Enter" and "Exit" to go back to the autocut screen Click "Sensor" (" センサー "): the sensor window disappears. 5. Manual alignment 5.1. Click "Manual Alignment" (" マニュアルアライメント ") 5.2. Align CH1 (CH1 direction is from the front to the back of the machine. On the screen it correspond to the direction from the left to the right) Change magnification to "LO" Adjustment of the microscope focus (refer to 4.1.4) Align the rotation (θ) Displace the screen to an extremity of a cutting line (or a line parallel to the cutting line) and align it with the line on the screen Click "θ adjustment" ("θ 合わせ "): the screen will move to the opposite extremity of the line Align the cutting line with the line on the screen Click "θ adjustment": the screen return to the first position. The rotation is aligned Change magnification to "HI" Align the rotation again Align the first cutting line with the alignment line on the screen 5.3. Click "Enter" and align the CH2 (if there is any CH2) Do the same procedure as for CH1, and finish with the alignment of the first cutting line with the line on the screen Click "Enter" 3 / 5

4 6. Start the cutting Start the cutting by clicking "START". When the cutting is completed, press "Exit". 7. Sample unloading and separation 7.1. Unloading the sample Click the button "C/T vacuum" ("C/T バキューム ") (vacuum will be turned of and the shutter will open) Open the door on the right side of the machine, unload the sample and close the door If you need to cut another sample, go back to Sample separation The sample may separate only by pushing with the finger from the backside, but it can be done more properly by using the film expander. This section describes how to use the film expander There is a frame with a black part and a white part. Separate the two parts and set the white part on the film expander chuck Set the sample (frame + film + sample) on the chuck Turn the power "ON" and click "Operation screen" (" 操作画面 ") Click "Start rising" (" 上昇開始 "): the film will expand as the chuck rise and the chips will separate Put the black frame around the sample to fix the film in the expended state Take the sample out of the frame Click "Return to Origin" (" 原点復帰 ") to move the chuck back to its original position Turn the power "OFF" 8. Shutting down 8.1. Set back the 6inch chuck if you used the 8inch chuck 8.2. Click "Diode" (" ダイオード ") to turn it OFF 8.3. Turn the key in front of the machine to OFF and wait until it shut down 8.4. Turn the breaker OFF on the left side of the machine 8.5. On the wall behind the machine: Close the Air and Nitrogen valves, Close the two cooling water valves, both at the same time 8.6. Distribution Board B2P-7: Stealth Dicer Breaker OFF 8.7. Fill in the paper log and the ipod log 9. Annex 1: How to move the position on the screen. There are two ways to do that: 9.1. Directly by touching the screen: There are two modes for that: the JOG mode and the IMMEDIATE mode. The current mode is displayed on the top right of the screen. Click on the top right side of the screen to toggle the mode JOG mode: the point where we touch the screen move to the center of the screen IMMEDIATE mode: a circle representing the sample appears on the screen, and when we touch a point on the sample, the screen will move to that position By using the displacement panel: Click the button " 分割されたウェハのボタン " at the bottom right of the screen to make the panel appear "IND.": by touching the arrow key, it will move by the distance defined by the Y index set in "SCR IND.": by touching the arrow key, it will move by one third of the screen "SCAN": to move precisely. If you press for a long time, the displacement speed will increase. 10. Annex 2: change the chuck In order to cut 8 inch samples, it is necessary to replace the 6 inch chuck with the 8 inch chuck Machine maintenance ( マシンメンテナンス ) Change frame size ( フレームサイズ変更 ) Change the chuck table ( チャックテーブル ) size from 6 to 8 inch Change the frame ( フレーム ) from 2-6 to Press F3 (C/T 取り外し ON/OFF). 4 / 5

5 10.2. Open the lateral door Extend the 4 parts that maintain the frame to 8 inch position. There is a button at the side of the parts. Push this button and push the part outward until reaching the notch Take out the 6 inch chuck, and set the 8 inch chuck Close the door and press F3 (C/T 取り外し ON/OFF) Click Focus maintenance ( フォーカスメンテナンス ). An error occurs. Ignore and click again Set the light intensity to 100% Set the measurement marks to 5 or 9 and click Enter Click Execute measurement ( 測定実行 ) and wait few minutes until it is finished If the autofocus fail because the light is to intense, lower a little the light intensity and try again Exit x4 to return to top menu Proceed with usual cutting procedure After cutting, be sure to re-install the 6 inch chuck and execute the Focus maintenance procedure. Notice that the 4 parts that maintain the frame should be set back to the second notch. 11. Annex 3: circular cut A circular cut is needed when cutting a 4 inch wafer out of a 6 inch wafer for instance. Warnings: The circular chuck is centered around the chuck table rotation axis. After the frame is set on the chuck, it is not possible to align. When circle cut is selected, the sub index setup is ignored. When cutting 2 inch wafer out of 4 inch wafer, there is better chance of success if you do it in two steps: 4 inch to 3 inch, then 3 inch to 2 inch (the smaller the size, the less chance of success) In the screen displayed after selecting Device data : Click Z compensation data ( Z 補正データ ) Set the edge detection voltage ( エッジ検出電圧 ) to 0.1V Click Enter Exit to go back to device data screen Click Circle cut ( サークルカット ) at the right Check the circle cut check-box Circle cut data ( サークルカットデータ ) (CH1) Set the circle size ( サークルサイズ ) to the diameter of the circle that you want to cut Set the circle cut speed ( サークルカット速度 ) to 90deg/s Splitting cut data ( 分割カットデータ ) (CH2) Set the number of splitting line ( 分割本数 ) (4~8, typically 4) Click Enter Exit to go back to device data screen Set the rest of the device data as usual (except the sub index) 12. Remarks, troubleshooting E1245 SD エンジンでエラーが発生しました (code=402) Just ignore and continue by pressing 続行 ( ぞっこう ) (resuming). It might occur several times. This error may be due to the presence of resist on the surface, or to a non horizontal position of the sample etc. leading to bad focusing of the laser. If this error occurs too often, the laser engine might be out of order. Contact the person in charge Chiller problem Check the water level of the chiller behind the machine. If no more water, refill with pure water The maximum number of items in the street cut sequence menu is 50. If you have a thick sample and increase this number above 50, the defocus values closest to the surface (from no.50) will be automatically deleted. 5 / 5

DISCO DICING SAW SOP. April 2014 INTRODUCTION

DISCO DICING SAW SOP. April 2014 INTRODUCTION DISCO DICING SAW SOP April 2014 INTRODUCTION The DISCO Dicing saw is an essential piece of equipment that allows cleanroom users to divide up their processed wafers into individual chips. The dicing saw

More information

Installation Manual WIND TRANSDUCER

Installation Manual WIND TRANSDUCER Installation Manual WIND TRANSDUCER Model FI-5001/FI-5001L This manual provides the instructions for how to install the Wind Transducer FI-5001/FI- 5001L. For connection to the instrument, see the operator

More information

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and

More information

MAXYM Mortiser Operating Manual

MAXYM Mortiser Operating Manual MAXYM Mortiser Operating Manual Rev 2.112/16/02 Copyright MAXYM Technologies Inc. Table of Contents Visual Tour 1-2 Operating the Maxym Mortiser 3 Starting the Mortiser 3 Touch Screen Description 3 Mortise

More information

Micro Automation- Model 1006 Dicing Saw Instructions. Serial # Rev 2 ( R.DeVito) Location Chase 1

Micro Automation- Model 1006 Dicing Saw Instructions. Serial # Rev 2 ( R.DeVito) Location Chase 1 Micro Automation- Model 1006 Dicing Saw Instructions Serial # Rev 2 (12-23-05 R.DeVito) Location Chase 1 Dicing Saw Instructions (Revised 8/9/03 - K.J) 1. On the Log Sheet sign in, including Name and Date.

More information

University of Minnesota Nano Fabrication Center Standard Operating Procedure Equipment Name:

University of Minnesota Nano Fabrication Center Standard Operating Procedure Equipment Name: Equipment Name: Coral Name: Nanoimprinter Revision Number: 1.1 Model: NX-B200 Revisionist: M. Fisher Location: Bay 4 Date: 2/12/2010 1 Description Nanonex NX-B200 nanoimprinter is another method of transfer

More information

SKF Shaft Alignment Tool Horizontal machines app

SKF Shaft Alignment Tool Horizontal machines app SKF Shaft Alignment Tool Horizontal machines app Short flex couplings Instructions for use Table of contents 1. Using the Horizontal shaft alignment app... 2 1.1 How to change the app language...2 1.2

More information

Figure 1 The Raith 150 TWO

Figure 1 The Raith 150 TWO RAITH 150 TWO SOP Figure 1 The Raith 150 TWO LOCATION: Raith 150 TWO room, Lithography area, NanoFab PRIMARY TRAINER: SECONDARY TRAINER: 1. OVERVIEW The Raith 150 TWO is an ultra high resolution, low voltage

More information

Dicing Saw. Standard Operating Procedure. Revision: 1.0 Last Updated: Apr.15/2015, Revised by Mohamad Rezaei

Dicing Saw. Standard Operating Procedure. Revision: 1.0 Last Updated: Apr.15/2015, Revised by Mohamad Rezaei Dicing Saw Standard Operating Procedure Revision: 1.0 Last Updated: Apr.15/2015, Revised by Mohamad Rezaei Overview This document will provide a detailed operation procedure of the Dicing Saw. Formal Training

More information

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop 1. Log in usage using the SMIF web site 2. Turn power

More information

Multi-bit Sigma-Delta TDC Architecture for Digital Signal Timing Measurement

Multi-bit Sigma-Delta TDC Architecture for Digital Signal Timing Measurement IEEE International ixed-signals, Sensors, and Systems Test Workshop, Taipei, 22 ulti-bit Sigma-Delta TDC Architecture for Digital Signal Timing easurement S. emori,. Ishii, H. Kobayashi, O. Kobayashi T.

More information

Heidelberg µpg 101 Laser Writer

Heidelberg µpg 101 Laser Writer Heidelberg µpg 101 Laser Writer Standard Operating Procedure Revision: 3.0 Last Updated: Aug.1/2012, Revised by Nathanael Sieb Overview This document will provide a detailed operation procedure of the

More information

Rotary Fixture M/V/X CLASS LASER SYSTEMS. Installation and Operation Instructions

Rotary Fixture M/V/X CLASS LASER SYSTEMS. Installation and Operation Instructions Rotary Fixture M/V/X CLASS LASER SYSTEMS Installation and Operation Instructions 02/01/2000 Introduction The Rotary Fixture controls in the Printer Driver are used along with the optional Rotary Fixture

More information

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack). FEI Tecnai F20 S/TEM: imaging in TEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 01/21/18 1. Filling the cold trap (if needed) 1.1. Prior to use, the cold trap needs

More information

User Operation of JEOL 1200 EX II

User Operation of JEOL 1200 EX II **Log onto Computer** Open item program Start Up Procedure User Operation of JEOL 1200 EX II 1. If scope is not running, locate an electron microscopy technician (EMT) to find out why not. 2. Turn up brightness

More information

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope The procedures given below were written specifically for the FEI Tecnai G 2 Sphera microscope. Modifications will need to

More information

アルゴリズムの設計と解析. 教授 : 黄潤和 (W4022) SA: 広野史明 (A4/A8)

アルゴリズムの設計と解析. 教授 : 黄潤和 (W4022) SA: 広野史明 (A4/A8) アルゴリズムの設計と解析 教授 : 黄潤和 (W4022) rhuang@hosei.ac.jp SA: 広野史明 (A4/A8) fumiaki.hirono.5k@stu.hosei.ac.jp Divide and Conquer Dynamic Programming L3. 動的計画法 Dynamic Programming What is dynamic programming? Dynamic

More information

MEASUREMENT CAMERA USER GUIDE

MEASUREMENT CAMERA USER GUIDE How to use your Aven camera s imaging and measurement tools Part 1 of this guide identifies software icons for on-screen functions, camera settings and measurement tools. Part 2 provides step-by-step operating

More information

MicroPG 101 Pattern Generator Standard Operating Procedure Draft v.0.2

MicroPG 101 Pattern Generator Standard Operating Procedure Draft v.0.2 Tool owner: Roman Akhmechet, romana@princeton.edu, x 8-0468 Backup: David Barth, dbarth@princeton.edu MicroPG 101 Pattern Generator Standard Operating Procedure Draft v.0.2 QUICK GUIDE PROCEDURE OVERVIEW

More information

3 안전을위한주의사항 AAH-02B3W. Product Composition & Specifications. Product Manual. Cautions for Safety. Cautions for Safety. Cautions.

3 안전을위한주의사항 AAH-02B3W. Product Composition & Specifications. Product Manual. Cautions for Safety. Cautions for Safety. Cautions. 1 Product Product Composition & Specifications 2 Basic Composition of the Product Main Body of the Product Fixing Bracket on the wall Knob Bolt(2pcs) AAH-02B3W Dimensions and color of the product are subject

More information

4. Contact arrangement 回路形式 1 poles 1 throws 1 回路 1 接点 (Details of contact arrangement are given in the assembly drawings 回路の詳細は製品図による )

4. Contact arrangement 回路形式 1 poles 1 throws 1 回路 1 接点 (Details of contact arrangement are given in the assembly drawings 回路の詳細は製品図による ) 1/6 SKHLAJA010 For reference 参考 1. General 一般事項 1.1 Application 適用範囲 This specification is applied to TACT switches which have no keytop. この規格書は キートッフ なしのタクトスイッチについて適用する 1.2 Operating temperature range

More information

HEX02 EMBOSSING SYSTEM

HEX02 EMBOSSING SYSTEM HEX02 EMBOSSING SYSTEM LOCATION: Hot Embossing Area PRIMARY TRAINER: 1. Scott Munro (2-4826, smunro@ualberta.ca) OVERVIEW The hot embosser is available to users who require polymer mold fabrication. This

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual Buehler EcoMet TM 300 Polisher Version 1.0 Page 1 of 19 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness Standard 2.2 Possible Polishing Materials 2.3

More information

Omochi rabbit amigurumi pattern

Omochi rabbit amigurumi pattern Omochi rabbit amigurumi pattern うさぎのあみぐるみ Materials Yarn: 1 main color (MC) and 1 contrasting color (CC), as needed. You can make this rabbit toy with any yarn weight, but the yarn colors used per one

More information

P Z N V S T I. センサ信号入力仕様 Input signal type. 1 ~ 5 V 4 ~ 20 ma 1 ~ 5 V 4 ~ 20 ma 1 ~ 5 V 4 ~ 20 ma 1 ~ 5 V 4 ~ 20 ma

P Z N V S T I. センサ信号入力仕様 Input signal type. 1 ~ 5 V 4 ~ 20 ma 1 ~ 5 V 4 ~ 20 ma 1 ~ 5 V 4 ~ 20 ma 1 ~ 5 V 4 ~ 20 ma 自由な位置設置を可能に! 分離型器 表示倍率 種類 kinds of display multiplier 見やすい 3 桁 表示 Full 3-digit red marking patible with EMC directive 3 mm 角小型フェイス Small 3 mm square face 電流出力は除く Except for Current output type 分離型なのでを自由にセット

More information

RF4C050AP. V DSS -20V R DS(on) (Max.) 26mΩ I D ±10A P D 2W. Pch -20V -10A Middle Power MOSFET Datasheet

RF4C050AP. V DSS -20V R DS(on) (Max.) 26mΩ I D ±10A P D 2W. Pch -20V -10A Middle Power MOSFET Datasheet RF4C050AP Pch -20V -10A Middle Power MOSFET Datasheet V DSS -20V R DS(on) (Max.) 26mΩ I D ±10A l 外形図 HUML2020L8 P D 2W DFN2020-8S l 特長 1) 低オン抵抗 2) 小型ハイパワーパッケージ (HUML2020L8) 3) 鉛フリー対応済み RoHS 準拠 4) ハロゲンフリー

More information

車載カメラにおける信号機認識および危険運転イベント検知 Traffic Light Recognition and Detection of Dangerous Driving Events from Surveillance Video of Vehicle Camera

車載カメラにおける信号機認識および危険運転イベント検知 Traffic Light Recognition and Detection of Dangerous Driving Events from Surveillance Video of Vehicle Camera 車載カメラにおける信号機認識および危険運転イベント検知 Traffic Light Recognition and Detection of Dangerous Driving Events from Surveillance Video of Vehicle Camera * 関海克 * 笠原亮介 * 矢野友章 Haike GUAN Ryosuke KASAHARA Tomoaki YANO 要旨

More information

RENISHAW INVIA RAMAN SPECTROMETER

RENISHAW INVIA RAMAN SPECTROMETER STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw invia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples

More information

Laser Application DAL7020 DFL7020 DFL7161 DFL7160 DFL7341 DFL7360FH DFL7361 DFL7560L. Ablation Process. Stealth Dicing.

Laser Application DAL7020 DFL7020 DFL7161 DFL7160 DFL7341 DFL7360FH DFL7361 DFL7560L. Ablation Process. Stealth Dicing. Laser Application Ablation Process Stealth Dicing Laser Lift Off DAL7020 DFL7020 DFL7161 DFL7160 DFL7341 DFL7360FH DFL7361 DFL7560L ABLATION PROCESS DISCO s laser application lineup supports miniaturized

More information

MAXYM Dovetailer Operating Manual

MAXYM Dovetailer Operating Manual MAXYM Dovetailer Operating Manual 1 2 Visual Tour Front View Touch Screen Blow Off Control Power Switch Air Pressure Control Air Clamp Controls Stop Button Start Cycle Button Top Table Air Clamp Controls

More information

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 1. Utility Requirements a. System power is supplied by two 120 VAC/20 A circuits. When doing maintenance

More information

User Manual. This User Manual will guide you through the steps to set up your Spike and take measurements.

User Manual. This User Manual will guide you through the steps to set up your Spike and take measurements. User Manual (of Spike ios version 1.14.6 and Android version 1.7.2) This User Manual will guide you through the steps to set up your Spike and take measurements. 1 Mounting Your Spike 5 2 Installing the

More information

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000 The ideal K-12 science microscope solution User Guide for use with the Nova5000 NovaScope User Guide Information in this document is subject to change without notice. 2009 Fourier Systems Ltd. All rights

More information

Considerations. image solid color for tracing make sure your traced shapes are a single solid color - Black RGB 0,0,0 or #000000

Considerations. image solid color for tracing make sure your traced shapes are a single solid color - Black RGB 0,0,0 or #000000 2D Design Object Description Form Fit and Function (FFF) Design for laser cut elements.bmp or.jpg border at least 300 dpi allow 1/8 margin for boarder of the material image solid color for tracing make

More information

RAITH e-line OPERATING INSTRUCTIONS

RAITH e-line OPERATING INSTRUCTIONS RAITH e-line OPERATING INSTRUCTIONS 1) LOADING A SAMPLE a. Start the system i. On the Column PC (Right side monitor [R]), select the SmartSEM icon to on the desktop to begin the column software. ii. On

More information

SUSS Mask Aligner. Purpose: To expose photoresist on a wafer using a photomask

SUSS Mask Aligner. Purpose: To expose photoresist on a wafer using a photomask SUSS Mask Aligner Purpose: To expose photoresist on a wafer using a photomask Overview This SOP will go over how to use the machine for basic exposures. This will include commonly used controls and frequently

More information

Basic Users Manual for Tecnai-F20 TEM

Basic Users Manual for Tecnai-F20 TEM Basic Users Manual for Tecnai-F20 TEM NB: This document contains my personal notes on the operating procedure of the Tecnai F20 and may be used as a rough guide for those new to the microscope. It may

More information

Decisions in games Minimax algorithm α-β algorithm Tic-Tac-Toe game

Decisions in games Minimax algorithm α-β algorithm Tic-Tac-Toe game Decisions in games Minimax algorithm α-β algorithm Tic-Tac-Toe game 1 Games Othello Chess TicTacToe 2 Games as search problems Game playing is one of the oldest areas of endeavor in AI. What makes games

More information

Speed of light E Introduction

Speed of light E Introduction Notice: All measurements and calculated values must be presented with SI units with an appropriate number of significant digits. Uncertainties required only when explicitly asked for. 1.0 Introduction

More information

FEI Tecnai G 2 F20 Operating Procedures

FEI Tecnai G 2 F20 Operating Procedures FEI Tecnai G 2 F20 Operating Procedures 1. Startup (1) Sign-up in the microscope log-sheet. Please ensure you have written an account number for billing. (2) Log in to the computer: Login to your account

More information

Module 1C: Adding Dovetail Seams to Curved Edges on A Flat Sheet-Metal Piece

Module 1C: Adding Dovetail Seams to Curved Edges on A Flat Sheet-Metal Piece 1 Module 1C: Adding Dovetail Seams to Curved Edges on A Flat Sheet-Metal Piece In this Module, we will explore the method of adding dovetail seams to curved edges such as the circumferential edge of a

More information

LinuxCNC Help for the Sherline Machine CNC System

LinuxCNC Help for the Sherline Machine CNC System WEAR YOUR SAFETY GLASSES FORESIGHT IS BETTER THAN NO SIGHT READ INSTRUCTIONS BEFORE OPERATING LinuxCNC Help for the Sherline Machine CNC System LinuxCNC Help for Programming and Running 1. Here is a link

More information

Xyron Professional 2500 Laminating System

Xyron Professional 2500 Laminating System Xyron Professional 2500 Laminating System Instruction Manual Provided By http://www.mybinding.com http://www.mybindingblog.com U S E R S G U I D E 2500 Adhesive Application & Laminating System The XM2500

More information

1 Radar calibration at first installation

1 Radar calibration at first installation 1 Radar calibration at first installation At first installation is necessary to properly calibrate the radar. This includes: Heading Line Transmission Trigger Delay Antenna Parking Position (only MDS 9

More information

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual Scanning Electron Microscope FEI INSPECT F50 Step by step operation manual Scanning Electron Microscope, FEI Inspect F50 FE-SEM-F Observation Flow Saving Data And Analysis Specimen preparation Error check

More information

CNC Using the FlexiCam CNC and HMI Software. Guldbergsgade 29N, P0 E: T:

CNC Using the FlexiCam CNC and HMI Software. Guldbergsgade 29N, P0 E: T: CNC Using the FlexiCam CNC and HMI Software Guldbergsgade 29N, P0 E: makerlab@kea.dk T: +46 46 03 90 This grey box is the NC controller. Let s start by turning the red switch to the ON position, then press

More information

Zeiss 880 Training Notes Zen 2.3

Zeiss 880 Training Notes Zen 2.3 Zeiss 880 Training Notes Zen 2.3 1 Turn on the HXP 120V Lamp 2 Turn on Main Power Switch Turn on the Systems PC Switch Turn on the Components Switch. 3 4 5 Turn on the PC and log into your account. Start

More information

CER7027B / CER7032B / CER7042B / CER7042BA / CER7052B CER8042B / CER8065B CER1042B / CER1065B CER1242B / CER1257B / CER1277B

CER7027B / CER7032B / CER7042B / CER7042BA / CER7052B CER8042B / CER8065B CER1042B / CER1065B CER1242B / CER1257B / CER1277B 一般機器用 For Consumer Products 汎用パワーインダクタ Common Power Inductors CER-B series RoHS CER727B / CER732B / CER742B / CER742BA / CER752B CER842B / CER865B CER42B / CER65B CER242B / CER257B / CER277B 特徴 DC-DC コンバータ用インダクタとして最適

More information

D80 を使用したオペレーション GSL システム周波数特性 アンプコントローラー設定. Arc 及びLine 設定ラインアレイスピーカーを2 から7 までの傾斜角度に湾曲したアレイセクションで使用する場合 Arcモードを用います Lineモード

D80 を使用したオペレーション GSL システム周波数特性 アンプコントローラー設定. Arc 及びLine 設定ラインアレイスピーカーを2 から7 までの傾斜角度に湾曲したアレイセクションで使用する場合 Arcモードを用います Lineモード D8 を使用したオペレーション GSL システム周波数特性 アンプコントローラー設定 Arc 及びLine 設定ラインアレイスピーカーを2 から7 までの傾斜角度に湾曲したアレイセクションで使用する場合 Arcモードを用います Lineモード アンプ1 台あたりの最大スピーカー数 SL-SUB SL-GSUB - - - - は 3つ以上の連続した から1 までの傾斜設定のロングスローアレイセクションで使用する場合に用います

More information

Radial Polarization Converter With LC Driver USER MANUAL

Radial Polarization Converter With LC Driver USER MANUAL ARCoptix Radial Polarization Converter With LC Driver USER MANUAL Arcoptix S.A Ch. Trois-portes 18 2000 Neuchâtel Switzerland Mail: info@arcoptix.com Tel: ++41 32 731 04 66 Principle of the radial polarization

More information

研究開発評価に関する国際的な視点や国際動向

研究開発評価に関する国際的な視点や国際動向 第 1 部 文部科学省平成 28 年度研究開発評価シンポジウム 大綱的指針の改定を踏まえた新しい研究開発評価へ向けて 講演 : 国の研究開発評価に関する大綱的指針 を踏まえた研究開発評価の推進について 研究開発評価に関する国際的な視点や国際動向 東京, 全日通霞が関ビルディング 8 階大会議室 2017 年 3 月 22 日 伊地知寛博 *1 *1 成城大学社会イノベーション学部教授 アウトライン

More information

Operating Procedures for MICROCT1 Nikon XTH 225 ST

Operating Procedures for MICROCT1 Nikon XTH 225 ST Operating Procedures for MICROCT1 Nikon XTH 225 ST Ensuring System is Ready (go through to ensure all windows and tasks below have been completed either by you or someone else prior to mounting and scanning

More information

Finding Offsets for Multiple Spindles/Air Drills

Finding Offsets for Multiple Spindles/Air Drills 888-680-4466 ShopBotTools.com Finding Offsets for Multiple Spindles/Air Drills Copyright 2016 ShopBot Tools, Inc. page 1 Copyright 2016 ShopBot Tools, Inc. page 2 Introduction This document explains how

More information

SCITEX Dual Roll Kit. User s guide

SCITEX Dual Roll Kit. User s guide SCITEX Dual Roll Kit User s guide 2011 Hewlett-Packard Development Company, L.P. First edition Legal notices The information contained herein is subject to change without notice. The only warranties for

More information

This procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface.

This procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface. Tescan MIRA3 SEM: EDS using EDAX TEAM Nicholas G. Rudawski ngr@ufl.edu Cell: (805) 252-4916 Office: (352) 392-3077 Last updated: 12/04/17 This procedure assumes the user is already familiar with basic

More information

U N I T. 1. What are Maxine and Debbie talking about? They are talking about. 2. What doesn t Maxine like? She doesn t like. 3. What is a shame?

U N I T. 1. What are Maxine and Debbie talking about? They are talking about. 2. What doesn t Maxine like? She doesn t like. 3. What is a shame? 1. Conversation: U N I T 1. What are Maxine and Debbie talking about? They are talking about 2. What doesn t Maxine like? She doesn t like 3. What is a shame? 4. Whose fault is it and why? 5. What did

More information

LSM 780 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch

More information

DOC. NO. FT02000-S E FULLY AUTOMATIC PROBER SPECIFICATIONS TOKYO SEIMITSU CO., LTD.

DOC. NO. FT02000-S E FULLY AUTOMATIC PROBER SPECIFICATIONS TOKYO SEIMITSU CO., LTD. DOC. NO. FT02000-S 0 0 1- E1 1 1-18 -9 7 FULLY AUTOMATIC PROBER SPECIFICATIONS TOKYO SEIMITSU CO., LTD. TOKYO, JAPAN Contents of this document may be subject to change without notice. No part of this document

More information

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center LEO 912 TEM Short Manual Prepared/copyrighted by RH Berg Danforth Plant Science Center Specimen holder [1] Never touch the holder (outside of the O-ring, double-headed arrow) because finger oils will contaminate

More information

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Introduction of Fluoresence Confocal Microscopy The first confocal microscope was invented by Princeton

More information

Effective Utilization of Patent Information in Japanese global companies

Effective Utilization of Patent Information in Japanese global companies Effective Utilization of Patent Information in Japanese global companies ATIS (Association of Technical Information Services) Member: IHI Corporation Intellectual property Dept. IP STRATEGY G. ATSUSHI

More information

OPERATION MANUAL MIMAKI ENGINEERING CO., LTD.

OPERATION MANUAL MIMAKI ENGINEERING CO., LTD. OPERATION MANUAL MIMAKI ENGINEERING CO., LTD. http://www.mimaki.co.jp/ E-mail:traiding@mimaki.co.jp D200674 About FineCut for CorelDRAW Thank you very much for purchasing a product of Mimaki. FineCut,

More information

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Follow the procedures below when you use the Hitachi 7100 TEM. Starting Session 1. Turn on the cold

More information

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope ZEN 2009 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn Chameleon TiS laser key from Standby

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Version : 07/08/0 Leica SP8 TCS Users Manual Start up:. Turn the PC Microscope, Scanner Power, Laser Power, and the Laser Emission key to on (bottom right of desk).. Turn on the fluorescent lamp (top left

More information

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Verify that main power switches on the

More information

IMPORTANT SAFETY INSTRUCTIONS Regulatory Safety Information

IMPORTANT SAFETY INSTRUCTIONS Regulatory Safety Information HELSINKI 取扱説明書 IMPORTANT SAFETY INSTRUCTIONS Regulatory Safety Information 1 Read these instructions. 2 Keep these instructions. 3 Heed all warnings. 4 Follow all instructions. 5 Do not use this apparatus

More information

Instructions for Tecnai a brief start up manual

Instructions for Tecnai a brief start up manual Instructions for Tecnai a brief start up manual Version 3.0, 8.12.2015 Manual of Tecnai 12 transmission electron microscope located at Aalto University's Nanomicroscopy Center. More information of Nanomicroscopy

More information

CNC Chucker Lathe P/N 6600, 6610, and 6620

CNC Chucker Lathe P/N 6600, 6610, and 6620 WEAR YOUR SAFETY GLASSES FORESIGHT IS BETTER THAN NO SIGHT READ INSTRUCTIONS BEFORE OPERATING PRODUCT DESCRIPTION 6600 CNC Chucker w/3c headstock, ball screws, high-torque stepper motors & PC w/4-axis

More information

Quick Start Guide. Contents

Quick Start Guide. Contents 1 Quick Start Guide Contents Powering on the Machine Login/Password Entry Jaw Set Up High Security Cut by Code High Security Jaw Set Up Edge Cut Cut by Code Edge Cut Cut by Decode Cutter Replacement Tracer

More information

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES RULES All users must go through a series of standard operation procedure training. For more information contact: Longlong Liao Teaching

More information

We recommend downloading the latest core installer for our software from our website. This can be found at:

We recommend downloading the latest core installer for our software from our website. This can be found at: Dusk Getting Started Installing the Software We recommend downloading the latest core installer for our software from our website. This can be found at: https://www.atik-cameras.com/downloads/ Locate and

More information

Bruker Dimension Icon AFM Quick User s Guide

Bruker Dimension Icon AFM Quick User s Guide Bruker Dimension Icon AFM Quick User s Guide March 3, 2015 GLA Contacts Jingjing Jiang (jjiang2@caltech.edu 626-616-6357) Xinghao Zhou (xzzhou@caltech.edu 626-375-0855) Bruker Tech Support (AFMSupport@bruker-nano.com

More information

SPS1M-EVK. SPS1M-EVK Battery Free Wireless Sensor Handheld Evaluation System EVAL BOARD USER S MANUAL

SPS1M-EVK. SPS1M-EVK Battery Free Wireless Sensor Handheld Evaluation System EVAL BOARD USER S MANUAL SPS1M-EVK SPS1M-EVK Battery Free Wireless Sensor Handheld Evaluation System EVAL BOARD USER S MANUAL Introduction This guide describes how to use the Handheld Evaluation System to carry out sensor measurements

More information

記号 / 定格 /Ratings. B. 電気的特性 /Electrical Characteristics 測定条件 /Measure Condition (Tc = 25 ±3 ) 記号 / 測定条件 /Measure Condition

記号 / 定格 /Ratings. B. 電気的特性 /Electrical Characteristics 測定条件 /Measure Condition (Tc = 25 ±3 ) 記号 / 測定条件 /Measure Condition 種別 /Type 用途 /Application 構造 /Structure シリコン MOS 形集積回路 /Silicon MOSFET Type Integrated Circuit スイッチング電源制御用 /For a Switching Power Supply Control CMOS 形 /CMOS Type 等価回路 /Equivalent Circuit 添付図 /See Figure

More information

Physics 248 Spring 2009 Lab 1: Interference and Diffraction

Physics 248 Spring 2009 Lab 1: Interference and Diffraction Name Section Physics 248 Spring 2009 Lab 1: Interference and Diffraction Your TA will use this sheet to score your lab. It is to be turned in at the end of lab. You must clearly explain your reasoning

More information

PH75A280-* RELIABILITY DATA 信頼性データ

PH75A280-* RELIABILITY DATA 信頼性データ RELIABILITY DATA 信頼性データ TDKLambda C2725701 INDEX PAGE 1.MTBF 計算値 Calculated Values of MTBF R1 2. 部品ディレーティング Components Derating R3 3. 主要部品温度上昇値 Main Components Temperature Rise T List R5 4. アブノーマル試験 Abnormal

More information

Maintenance Manual for Auto Lab

Maintenance Manual for Auto Lab Version 1.1 9/30/2012 Maintenance Manual for Auto Lab Lubricate X axis There are two locations for lubrication at the back side of the gantry as shown in the below picture. The left one (from the backside

More information

JEOL 6700 User Manual 05/18/2009

JEOL 6700 User Manual 05/18/2009 JEOL 6700 User Manual 05/18/2009 LOG IN to your session on the computer to the right of the microscope. Starting Conditions 1. Click the button and read the Penning Gauge to ensure that the microscope

More information

DigiScope II v3 TM Aperture Scope User s Manual

DigiScope II v3 TM Aperture Scope User s Manual DigiScope II v3 TM Aperture Scope User s Manual Welcome Thank you for choosing DigiScope II v3 TM Aperture scope! The DigiScope II v3 TM Aperture Scope is an exciting new device to Capture and record the

More information

Installation tutorial for Console Customs Xbox 360 MaxFire LITE rapid fire Mod Chip.

Installation tutorial for Console Customs Xbox 360 MaxFire LITE rapid fire Mod Chip. Installation tutorial for Console Customs Xbox 360 MaxFire LITE rapid fire Mod Chip. This tutorial is designed to aid you in installation of a console customs MaxFire LITE modchip. This tutorial covers

More information

2 How to operate the microscope/obtain an image

2 How to operate the microscope/obtain an image Morgagni Operating Instructions 50079 010912 2-1 2 ow to operate the microscope/obtain an image 2.1 Starting the microscope 2.1.1 Starting the microscope with several manually-operated steps 1. Turn on

More information

Powder diffractometer operation instructions D8 Advance with a Cu Kα sealed tube and Lynxeye MARCH 30, 2016

Powder diffractometer operation instructions D8 Advance with a Cu Kα sealed tube and Lynxeye MARCH 30, 2016 Powder diffractometer operation instructions D8 Advance with a Cu Kα sealed tube and Lynxeye MARCH 30, 2016 1 SAFETY. For radiation safety materials refer to http://www.ehs.wisc.edu/raddevices.htm. There

More information

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Operations Guide Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Nikon Eclipse 90i Operations Guide STARTUP Startup Powering Up Fluorescence

More information

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only Version 1.0 Prepared by D. Turnbull February 21, 2007. Please submit any omissions to the Author Note: This SEM is a recent

More information

Module 1G: Creating a Circle-Based Cylindrical Sheet-metal Lateral Piece with an Overlaying Lateral Edge Seam And Dove-Tail Seams on the Top Edge

Module 1G: Creating a Circle-Based Cylindrical Sheet-metal Lateral Piece with an Overlaying Lateral Edge Seam And Dove-Tail Seams on the Top Edge Inventor (10) Module 1G: 1G- 1 Module 1G: Creating a Circle-Based Cylindrical Sheet-metal Lateral Piece with an Overlaying Lateral Edge Seam And Dove-Tail Seams on the Top Edge In Module 1A, we have explored

More information

Omni LED Bulb. Illustration( 实际安装, 설치사례, 設置事例 ) Bulb, Downlight OBB. OBB-i15W OBB-i20W OBB-i25W OBB-i30W OBB-i35W. Omni LED.

Omni LED Bulb. Illustration( 实际安装, 설치사례, 設置事例 ) Bulb, Downlight OBB. OBB-i15W OBB-i20W OBB-i25W OBB-i30W OBB-i35W. Omni LED. CR2000 CH2000 CH2500 CD800 CD1500 CD3000 CD4000 CT2000 CT2500 CT3000 CT5000 OBB Street CT8000 CD800S CD1500S CD3000S CD4000S Illustration( 实际安装, 설치사례, 設置事例 ) OBB OBB-i15W OBB-i20W OBB-i25W OBB-i30W OBB-i35W

More information

Adhesive Application & Laminating System

Adhesive Application & Laminating System U S E R S G U I D E Adhesive Application & Laminating System The XM2500 is an economical and easy to use document finishing system. This versatile and non-electric system has been designed to laminate

More information

Eddy current flaw detector «Eddycon C»

Eddy current flaw detector «Eddycon C» ULTRACON-SERVICE LLC Eddy current flaw detector «Eddycon C» Quick start guide CONTENTS P. 1 CONTROLLERS OF EDDYCON C FLAW DETECTOR... 3 2 SETTINGS OF «TEST» MENU... 5 3 INSTRUCTIONS FOR USE... 8 3.1 THRESHOLD

More information

Flange Design & Maintenance

Flange Design & Maintenance Flange Design & Maintenance Back Flange S.S. Bushing S.S. Front Flange Al. + Hard Anodize Nut - S.S. Good Damaged Dicing Seminar Blade Gang Assembly s Dual blade Spacer Blade Wobbling Flange deflection

More information

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Warning Always adjust your specimen height before closing the chamber door to make sure your specimen will not hit the bottom of the lens;

More information

CHA ELECTRON BEAM EVAPORATOR SYSTEM OPERATING MANUAL Version: 2.0 August 2016

CHA ELECTRON BEAM EVAPORATOR SYSTEM OPERATING MANUAL Version: 2.0 August 2016 CHA ELECTRON BEAM EVAPORATOR SYSTEM OPERATING MANUAL Version: 2.0 August 2016 UNIVERSITY OF TEXAS AT ARLINGTON Nanotechnology Research Center TABLE OF CONTENTS 1.0 DESCRIPTION.... 3 2.0 SAFETY....... 3

More information

RDworks and Basic Rabbit Laser Operation Tutorial

RDworks and Basic Rabbit Laser Operation Tutorial RDworks and Basic Rabbit Laser Operation Tutorial File Setup Layer color is how the Rabbit Software designates different operations so, regardless of design software, you will want to organize geometry

More information

TROUBLE-SHOOTING: Error States

TROUBLE-SHOOTING: Error States TROUBLE-SHOOTING: Error States Please note, there is much commonality between the different models of LabelStation and therefore it is advisable to read the comments on other models if you cannot find

More information

SPECTRALIS Training Guide

SPECTRALIS Training Guide SPECTRALIS Training Guide SPECTRALIS Diagram 1 SPECTRALIS Training Guide Table of Contents 1. Entering Patient Information & Aligning the Patient a. Start Up/Shut Down the System... 4 b. Examine a New

More information

1. Preliminary sample preparation

1. Preliminary sample preparation FEI Helios NanoLab 600 standard operating procedure Nicholas G. Rudawski ngr@ufl.edu (352) 392 3077 (office) (805) 252-4916 (cell) Last updated: 03/02/18 What this document provides: an overview of basic

More information

Direct Printing from PictBridge-compatible Printers

Direct Printing from PictBridge-compatible Printers Direct Printing from PictBridge-compatible Printers This camera is compatible with PictBridge for direct printing with PictBridge-compatible printers. Connect the camera to a PictBridge-compatible printer

More information

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling Zeiss LSM 510 Confocor III Training Notes Center for Cell Analysis & Modeling Confocor 3 Start Up Go to System Module Turn on Main Switch, System/ PC, and Components Switches Do you need the arc lamp?

More information

SOP: EDAX Eagle III Microspot XRF

SOP: EDAX Eagle III Microspot XRF SOP: EDAX Eagle III Microspot XRF Page 1 of 6 SOP: EDAX Eagle III Microspot XRF 1. Scope 1.1 This document describes the standard operating procedure (SOP) for the EDAX Eagle III Microspot XRF. This X-ray

More information

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below.

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. http://staff.washington.edu/wpchan/if/cm100_inst.shtml Instructions for the Philips CM100 TEM and peripherals

More information