RENISHAW INVIA RAMAN SPECTROMETER

Size: px
Start display at page:

Download "RENISHAW INVIA RAMAN SPECTROMETER"

Transcription

1 STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw invia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples to infer the chemistry and structure of the material of interest. Location: 381 Chemistry Research Laboratory Primary Staff Contact: Weiqiang Ding (304) cell Secondary Staff Contact: Harley Hart (412) cell The Shared Research Facilities are operated for the benefit of all researchers. If you encounter any problems with this piece of equipment, please contact the staff member listed above immediately. There is never a penalty for asking questions. If the equipment is not behaving exactly the way it should, contact a staff member. WARNING: This system uses a Class IIIb laser and emits visible and invisible radiation. Laser safety goggles MUST be worn when operating the laser as an eye hazard exists. START-UP NOTE: The RAMAN spectrometer, microscope, and laser should be powered down when not in use. If the power has been left on to any of these items, please inform a MCF staff member. 1. Turn on the computer if it is not already on and log in to the computer. 2. Log in your session on the FOM. Write down your name and sample/laser information on the log book located on the sample preparation bench beside the computer cart. 3. Turn ON the laser safety warning light (switch by the door next to the room light switch) to warn any individuals who need to enter the lab during your session. 4. Turn ON the orange microscope power switch on the left-side of the optical microscope base (Figure 1a). 5. Turn ON the power toggle switch on the front-side of the PRIOR control box located to the left of the microscope (Figure 1b). 6. Toggle the power switch on the right-hand side of the RAMAN spectrometer set-up to ON (Figure 2a). Allow 20 minutes for the CCD detector to cool down to its operating temperature. 1

2 (a) (b) Figure 1. (a) Optical microscope power; and (b) motor control box power. (a) (b) Figure 2. (a) Raman spectrometer power switch (marked); (b) Motor reference options window. 7. Start the WIRE 3.4 software on the computer desktop. 8. The software will give options as to which motors to reference (Figure 2b). Select Reference All Motors and then click OK. The tool will initialize all motors of the spectrometer. If the motors fail to reference, contact a MCF staff member for help. LASER SETUP 1. Check and remove any item placed in the path of the laser beam on the optical table (Figure 3). Such item will cause dangerous reflections of the laser beam that poses a serious eye hazard. 2

3 Figure 3. Laser setup on optical table. 2. If you need to use 532 nm laser, first find and put on a pair of laser safety glasses for 532 nm laser (orange colored lenses). Laser safety glasses are stored on the sample preparation bench beside the computer cart. 3. Switch the 532 nm laser power supply toggle switch to the ON position (located on the optical table, Figure 4a). The power light indicator on the front of the power supply will light up. 4. Wait 30 Seconds to allow the power supply to stabilize. 5. Turn on the laser power supply key to the ON position. The laser light indicator will light up (Figure 4b). Allow 3 minutes for the laser to stabilize. (a) (b) Figure nm laser power supply: (a) power switch in the back; (b) laser in the ON position. 6. At the bottom of the Wire 3.4 program, select the 1800 l/mm grating and 532 nm laser edge (Figure 5). Figure 5. Grating and laser edge selection at bottom of the Wire software program. 3

4 SWITCHING LASER WARNING: Do NOT use 785 nm laser unless you are trained by MCF staff to do so. 1. To use 785 nm NIR laser, find and put on a pair of laser safety glasses for 785 nm laser (green colored lenses). Laser safety glasses are stored on the sample preparation bench beside the computer cart. 7. Turn on the laser power key at the back of the red laser (Figure 6a). 8. Wait 30 seconds to allow the power supply to stabilize. 9. Turn on the red laser by press the Laser ON button (Figure 6b). (a) (b) Figure 6. (a) 785 nm laser power on (red LED); (b) 785 nm laser on (green LED). 10. At the bottom of the Wire program, select the 1200 l/mm grating and 785 nm laser edge (Figure 7). A Change Spectrometer Lens Set window pops up for replacing the A(1)B(1)C(1) lenses with New Lens Set. Figure 7. Grating and laser edge selection options. 11. Find the Raman front door lock key (Figure 7a) in the black Raman toolbox on optical table. Use the key to open up the Raman instrument front door (Figure 7a). 12. Locate the accessory box labeled A(1)B(1)C(1) and accessory box labeled A(3) (Figure 7b) on the optical table beside the spectroscope. 4

5 (a) (b) Figure 7. (a) Open Raman enclosure lock with key; (b) Accessory boxes for A1/B1/C1 lens (for 532 nm laser) and A3/B3/C3 lens (for 785 nm laser). 13. WARNING: Do NOT perform the following steps unless you are trained by MCF staff to do so. Remove the cover of the grating chamber. Find and remove the A(1) B(1) C(1) lens (Figure 8). Use extreme care while removing C(1) lens to avoiding hitting the prism. Store them in the box labeled A(1)B(1)C(1). 14. Mount A(3) B(3) C(3) lenses at their specific positions (Figure 8). 15. Put the grating chamber cover back. Close the instrument door and lock the door with the key. A C B Figure 8. Raman instrument chamber with A(1), B(1) and C(1) lenses installed for 532 nm laser. 5

6 SYSTEM CHECK 1. Turn the objectives to the back to open up the sample stage area (Figure 9). 2. Lower the stage of the microscope using the Z-position adjustment knobs on the side of the microscope. 3. Place the silicon reference standard (stored in the black box on the computer cart) on the stage. Use the stage clip to secure the glass slide. Figure 9. Lowered sample stage with objectives in the rear. 4. Move the 5x objective into position so that it is above the silicon reference sample. NOTE: Whenever the sample objective is changed on the turret, the objective setting needs to be changed in the Sample Review window. This will adjust the scale on the video window (see Figure 6b). 5. On the filter wheels of the microscope, adjust the filter wheels to the following positions (Figure 10a): Upper filter wheel = 2 Lower filter wheel = 1 This allows for viewing of the sample using the white light source of the microscope and video camera (Figure 10b). (a) (b) Figure 10. (a) Filter wheels in positions 2 and 1; (b) Focused silicon reference sample in video window. 6

7 6. Focus on the surface of the silicon using height adjustment knob on the side of the microscope. It may be necessary to adjust the F-stop wheel on the bottom left of the microscope to reduce the light intensity. 7. Once the sample is in focus, rotate the objective turret to the 20x position and focus on the sample. Then rotate the turret to the 50x objective and focus the sample. WARNING: Adjust the stage height SLOWLY so as NOT to crash the objective onto the sample, which will damage the objective. 8. In the software program, select Measurement > NEW Measurement > 532nm-1800 set-up. This set-up is the saved measurement set-up for the silicon reference sample. 9. On the filter wheels of the microscope, adjust the filter wheels to the following positions to activate Raman collection channel (Figure 11a): Upper filter wheel = ARROW Lower filter wheel = 4 The laser spot will appear in the crosshairs (Figure 11b). Using the stage height adjustment knob, focus the laser spot. (a) (b) Figure 11. (a) Filter wheels positions for measurement; (b) Centered laser spot in the Video window. NOTE: If the laser spot is outside of the crosshairs, DO NOT ADJUST THE CROSSHAIR POSITION. Contact a MFCF staff for help in adjusting the laser position. 9. Select Tools > Calibration > Quick Calibration and the tool will set the offset to the silicon reference peak to 521 l/mm (Figure 12). 7

8 Figure 12. Silicon reference sample spectral scan. 10. After the calibration, lower the stage height using the adjustment knobs on the side of the microscope and rotate the objective turret so that an empty objective position is above the sample. 11. Release sample clip and remove the silicon reference sample and place it back into the storage box. The tool is calibrated for the processing of samples. SPECTRAL ACQUISITION There are several types of measurements available for users are as follows: Spectral Acquisition standard method for spectral acquisition Image Acquisition collection of filter spectra. Mapping Acquisition collection of spectral acquisitions over an area of the sample. If the mapping technique is desired to be used, contact a Shared Facilities staff member. NOTE: The following SOP is for a Spectral Acquisition. For instruction on the Mapping and Image Acquisition techniques, please contact a MFCF staff member. 1. Place sample onto a glass microscope slide located on the sample prep table. 2. Turn the objective turret so that scope objectives are in the rear. This is to prevent accidental crashing of the objectives into the sample. 3. Lower the stage of the microscope using the stage height adjustment knobs on the side of the microscope. Pull back the sample clip on the stage, and place the glass slide on the stage. Release the stage clip to secure the glass slide. 4. Move the 5x objective into position so that it is above the sample. NOTE: Whenever the sample objective is changed on the turret, then the objective setting needs to be changed in the Sample Review window. This will adjust the scale on the video window. 5. On the filter wheels of the microscope, adjust the filter wheels to the following positions: 8

9 Upper filter wheel = 2 Lower filter wheel = 1 This allows for viewing of the sample using the white light source of the microscope and video camera. 6. Using the stage height adjustment knob on the side of the microscope, focus on the surface of the sample. It may be necessary to adjust the F-stop on the bottom left of the microscope in order to focus on the surface of clear or highly reflective samples. 7. Once the sample is in focus, rotate the objective turret to the 20x position and focus the sample using the stage height adjustment knob. Then rotate the turret to the 50x objective and focus the sample using the stage height adjustment knob. CAUTION: Adjust the stage height slowly so as to NOT crash the objective into the sample as damage to the objective can occur. 8. Select Measurement > NEW > Spectral Acquisition and set the scan to the desired parameters in the Spectral Acquisition Setup window: NOTE: The APPLY button must be pressed to save any changes made to the acquisition parameters. 1). RANGE TAB (a) (b) Figure 13. Range Tab windows for (a) Static scan and (b) Extended scan. 9

10 Grating Scan: i. Static Scan: The user sets the center point of the Spectrum Range and the system will set the scan range 500 cm -1 on the center point. This scan can have an exposure time of less than one second per accumulation. ii. Extended Scan: The user sets the upper and lower limit of the Spectrum Range and the spectrometer will scan continuously over the range. This scan is limited to exposure times of 10 seconds or greater. Confocality: sets the sample volume of the sample collected i. Standard: Uses a larger volume to increase signal strength. ii. High: Reduces the volume and signal strength to increase depth resolution. Configuration: i. Laser name: 532nm notch ii. Grating Type: 1800 l/mm (vis) 2) ACQUISITION TAB Figure 14. Acquisition Tab window. Title and Description: Name the acquisition and give a description of the scan. This information will be saved with the acquisition. Exposure Time: This is the amount of exposure time at the detector where longer exposure times will improve the signal-to-noise ratio. Accumulations: This is the number of scan repetitions added together to improve signalto-noise ratios and for the removal of cosmic rays. If the cosmic ray removal box is selected, then two additional scans will be processed in addition to the number of accumulations selected 10

11 Objective: This is the objective selected in the Sample Review window and must be changed every time the user changes the objective manually on the microscope. Laser Power: This is the percentage of laser power being used for the scan where higher powers translate into better signal-noise ratios. Start with lower laser powers on initial scans if the sample is susceptible to being damaged by the laser. Cosmic Ray Removal: Select this box to have the spectrometer take two additional scans to remove random cosmic peaks Close Shutter on Completion: Select this box to close the laser at the end of the acquisition and limit laser exposure on the sample. 3) FILE TAB Autosave: Select this box so that the acquisition will be saved to the folder with filename designated by the user. Select the AUTO INCREMENT box to save every scan acquisition with a sequential numbered filename. 4) TIMING TAB Figure 15. File Tab window. Sample Bleaching: This is used to decrease fluorescence from the sample by exposing the sample to the sample to the laser prior to taking a measurement. Select this box and set the time if this issue is occurring on the sample. Time Series Measurements: This allows for the user to set a specified time in between scans. 11

12 5) TEMPERATURE Figure 16. Timing Tab window. This is currently NOT an option to users as a temperature stage is required. 6) FOCUS TRACK TAB This is used to maintain the laser focus position for time and mapping acquisitions. See Module #7 of user manual (located on the computer desktop) and/or contact a MFCF staff member for additional information. 7) ADVANCED TAB Scan Type: When an Extended Scan has been selected in the RANGE TAB, then the user has the option of using a Step Scan or a Synchroscan (or Continuous Scan). The recommended option is the Synchroscan, but if saturation of the detector is occurring, then the Step option should be selected so exposure time under 10s can be used. Camera Gain: Select the HIGH sensitivity option. Pinhole: The system has a motorized pin-hole assembly and this function is used to for improving the profile of the laser beam. The option IN is recommended. Laser Focus: This option controls the beam expander. A setting of 100% is a completely defocused beam, while 0% is a tightly focused beam. If the sample being used is subject to damage by the laser, then a highly defocused beam is recommended to reduce the power density at the sample. Image Capture: This feature can be used to have white light images of the sample taken before or after an acquisition. 12

13 Figure 17. Advanced Tab window. 9. On the filter wheels of the microscope, adjust the filter wheels to the following positions: Upper filter wheel = ARROW Lower filter wheel = 4 The laser spot will appear in the crosshairs (Figure 18). Using the stage height adjustment knob, focus the laser spot. Figure 18. Centered laser spot in Video window. NOTE: If the laser spot is outside of the crosshairs, DO NOT ADJUST THE CROSSHAIR POSITION. The laser position needs adjusted and a Shared Facilities staff member needs to be contacted. 10. Turn off the room light (light switch located by the door). 11. Choose the RUN icon on the toolbar to run the measurement. The acquisition can be stopped at any time by pressing the ABORT icon. 12. If another acquisition is required and spectral acquisition settings need to be changed, select Measurement > Set-up Measurement option. The scan parameters from the previous acquisition will still be stored in the Spectral Acquisition Set-up window. If the scan parameters need to be saved for later use, select Measurement > Save Measurement. 13

14 13. To remove the sample from the microscope stage, lower the stage height using the adjustment knobs on the side of the microscope and rotate the objectives turret so that an empty objective position is above the sample. 14. Release the sample clip and remove the microscope slide from the stage. 15. See Module #3 of user manual for data manipulation techniques under the Analyze tab of the Wire 3.2 software. 16. If another sample is to be processed, then proceed back to step #1 of this section. If all necessary acquisitions, have been acquired, then proceed to the SYSTEM SHUT DOWN section of the SOP. NOTE: Users should make sure that they copy their data off the computer in a timely manner. The Shared Research Facilities is a multi-user facility and therefore cannot guarantee that the saved file will not be modified or deleted. DATA OPTIMIZATION 1. To improve signal to noise ratio a. Increase exposure time b. Increase number of accumulations 2. To eliminate strong background due to fluorescence a. Decrease laser power b. Quench fluorescence by exposing sample to incident laser light for a period of time c. Change excitation wavelength 3. To avoid saturated signal a. Reduce laser power b. Reduce exposure time 4. To avoid laser ablation on the sample a. Use a lower magnification objective lens b. Defocus the laser spot c. Reduce laser power 14

15 DATA ANALYSIS NOTE: Below lists two basic data processing procedures. Please refer to M3- Data manipulation module (on computer desktop) for additional information or contact a MFCF staff member. 1. Baseline subtraction 1) Open the spectrum in the Viewer. 2) Select Processing > Subtract Baseline. A new Viewer opens with the spectrum in the top half and the results of any baseline subtraction in the lower (Figure 15). 3) The default baseline is fitted between the two end points of the spectrum. Use left mouse click to add more points to the baseline. 4) Use right mouse click to get the context menu. Select Properties > Cubic Spline Interpolation from the menu. 5) Select Accept from the context menu. 6) You will be asked to keep the correction. NOTE: If you select Yes, you will overwrite the original file. You may need to keep a backup or rename the file. 2. Smoothing 1) Select Processing > Smooth. Figure 19. Baseline subtraction example. 2) A new window will open with the sample spectrum at the top and the result spectrum below (Figure 16). 3) To modify the degree of smoothing, select Properties from the context menu (right mouse click) to see the Smooth Properties window. 4) You can save the resulting smoothed spectrum by select Processing > Smooth > Accept > Current dataset. 15

16 Figure 20. Curve smoothing example 3. Zap Use Zap function to remove stray bands such as cosmic rays from the spectrum. 1) Open the spectrum of interest. 2) Select Processing > Zap from the menu. A new Viewer will open with the sample spectrum at the top and the result spectrum below (Figure 17). 3) The upper spectrum has a zap region between two vertical black lines. Grab each line in turn and adjust the position of the zap region to just enclose the band to be removed. You may need to use the zoom function to isolate the band. Figure 21. Zap function example. 16

17 4. Curve-fitting 1) Select Analysis > Curve Fit to open the curve fit window. 2) Zoom into a region that contains the band and some baseline data at either side. 3) Use the mouse to position the approximate centre of the band. 4) Click to add the band and repeat for the centers of other bands if there are multiple bands. 5) You can adjust the center or width markers for each band using the mouse and cursor to get a better rough fit. 6) Use mouse right click to get the menu and select Start Fit to fit the peak (Figure 18). 7) You can save the curve fit file from the context menu. Figure 22. Curve fitting example. 5. Peak pick 1) Select Analysis > Peak pick to open the curve fit window (Figure 19). 2) You may adjust the thresholds by selecting Analysis > Autoset thresholds. Figure 23. Peak pick example. 17

18 SYSTEM SHUT DOWN 1. Turn OFF the laser. 2. Turn the laser power supply key to the OFF position. The power light indicator on the front of the power supply will turn off. 3. If you used 785 nm laser, set the laser and grating back to 532 nm/1800l/mm in the WIRE 3.4 program. Follow the accessory switch procedure to replace the A(3)B(3)C(3) lens with A(1)B(1)C(1) lens. 4. Toggle the power switch on the right side of the RAMAN spectrometer set-up to OFF. 5. Turn OFF the power switch on the front-side of the PRIOR control box. 6. Turn OFF the power switch on the left-side of the white light source for the microscope. 7. CLOSE the WIRE 3.4 software on the computer desktop. 8. Check that all required parameters are recorded in the log book. Sign out on the log book. 9. Clean up work area by doing the following: a. Safety glasses are returned to the proper case on the sample preparation bench next to the computer cart. b. The silicon reference sample is placed back in the storage case located on top of the computer cart. 10. Properly dispose of any used materials i.e. wipes, glass slides, etc. 11. Sign out of the FOM. Report any problems in the log off window. 12. Turn off the Laser Warning light (switch by the door) white exiting the lab. 18

19 EMERGENCY PROCEDURES If no one is available and the machine is not acting as expected, the user should do the following: Press ABORT on the WIRE 3.2 software. Turn the key to the laser power supply to the OFF position. Turn off the x-rays by pressing the red OFF button on the power controller. Do not leave the machine running in an abnormal state. If the machine cannot be placed in the default state, immediately contact: Primary Staff Contact: Weiqiang Ding (304) Secondary Staff Contact: Harley Hart (412) Then, if possible, the user should stay by the tool while trying to contact a Shared Facilities staff member. If it becomes necessary to leave the tool then the user should leave a large, legible note at the RAMAN stating the tool is DOWN. The user should then complete a Tool Issue Report form and submit this form to ESB G75D. The user should also add comment using the FOM software indicating the tool status. If a dangerous situation is evident (smoke, fire, sparks, etc), ONLY if it is safe to do so, the user should turn off power to the system by switching the toggle switch on the right side of the tool and the laser power supply key and toggle switch to the OFF position. The user should then notify all other persons within the Hodges Hall 325 lab to evacuate and leave the lab immediately. The user should then contact proper emergency personnel from a safe place. 19

Renishaw InVia Raman microscope

Renishaw InVia Raman microscope Laser Spectroscopy Labs Renishaw InVia Raman microscope Operation instructions 1. Turn On the power switch, system power switch is located towards the back of the system on the right hand side. Wait ~10

More information

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and

More information

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop 1. Log in usage using the SMIF web site 2. Turn power

More information

WITec Alpha 300R Quick Operation Summary October 2018

WITec Alpha 300R Quick Operation Summary October 2018 WITec Alpha 300R Quick Operation Summary October 2018 This document is frequently updated if you feel information should be added, please indicate that to the facility manager (currently Philip Carubia,

More information

Renishaw InVia Quick Operation Summary October 2018

Renishaw InVia Quick Operation Summary October 2018 Renishaw InVia Quick Operation Summary October 2018 This document is frequently updated if you feel information should be added, please indicate that to the facility manager (currently Philip Carubia,

More information

KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017)

KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017) KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017) 1 Introduction You must be trained to operate the Laser-scanning confocal microscope (LSCM) independently.

More information

Leica SPEII confocal microscope. Short Manual

Leica SPEII confocal microscope. Short Manual Leica SPEII confocal microscope Short Manual Switching ON sequence: 1. Turn on the Workstation under the bench (top, far right). 2. Turn on the Supply Unit - Laser box (big green switch first and then

More information

IBIL setup operation manual for SynerJY software version

IBIL setup operation manual for SynerJY software version IBIL setup operation manual for SynerJY software version 1.8.5.0 Manual version 1.0, 31/10/2008 Author: Carlos Marques Equipment Managers: Carlos Marques, +351219946084, cmarques@itn.pt Luís Alves, +351219946112,

More information

Things to check before start-up.

Things to check before start-up. Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your

More information

RENISHAW RAMAN MICROSCOPE STANDARD OPERATING PROCEDURE

RENISHAW RAMAN MICROSCOPE STANDARD OPERATING PROCEDURE RENISHAW RAMAN MICROSCOPE STANDARD OPERATING PROCEDURE 09DJ00 CHRIS BUXEY (TECHNICIAN) X2715 C.BUXEY@SURREY.AC.UK V5.0 OCTOBER 2008 RENISHAW MICRORAMAN STANDARD OPERATING PROCEDURE 1.0 Foreword Chris Buxey:

More information

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120)

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120) Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120) Please contact Dr. Amanda Henkes for training requests and assistance: 979-862-5959, amandahenkes@tamu.edu Hardware LN 2 FTIR FTIR camera 1

More information

Zeiss LSM 780 Protocol

Zeiss LSM 780 Protocol Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

TRAINING MANUAL. Olympus FV1000

TRAINING MANUAL. Olympus FV1000 TRAINING MANUAL Olympus FV1000 September 2014 TABLE OF CONTENTS A. Start-Up Procedure... 1 B. Visual Observation under the Microscope... 1 C. Image Acquisition... 4 A brief Overview of the Settings...

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS Zeiss LSM 510 M eta Confocal M icroscope OPERATING INSTRUCTIONS Starting the System: 1. Turn the black knob on the laser box one-quarter turn from Off to On. You will hear the laser cooling mechanisms

More information

Confocal Raman Microscopy (WITec Alpha 300R)

Confocal Raman Microscopy (WITec Alpha 300R) Confocal Raman Microscopy (WITec Alpha 300R) Please refer to Witec Alpha300R Confocal Raman Microscope User Manual for the details of the operating procedure. Sample preparation 1. Attach your sample on

More information

Zeiss LSM 880 Protocol

Zeiss LSM 880 Protocol Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 710 Confocal Microscope Standard Operation Protocol LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.

More information

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5. Smith College August 2010 Operating Checklist for using the Laser Scanning Confocal Microscope Leica TCS SP5. CONTENT, page no. Startup, 1 Initial set-up, 1 Software, 2 Microscope Specimen observation

More information

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO TRAINING MANUAL Multiphoton Microscopy LSM 510 META-NLO September 2010 Multiphoton Microscopy Training Manual Multiphoton microscopy is only available on the LSM 510 META-NLO system. This system is equipped

More information

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL START-UP On the Switchbox, turn both black switches to the ON position. Wait for the microscope to boot up completely (watch the screen on the side of the microscope).

More information

LSM 780 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch

More information

Title: Leica SP5 Confocal User Manual

Title: Leica SP5 Confocal User Manual Title: Leica SP5 Confocal User Manual Date of first issue: 23/10/2015 Date of review: Version: Admin For assistance or to report an issue Office: CG07 or 05 Email: Igmm-imaginghelpdesk@igmm.ed.ac.uk Website:

More information

Guide to Operating the Bruker FT-IR Microscope I. Basic Spectrum Acquisition with Vertex 70

Guide to Operating the Bruker FT-IR Microscope I. Basic Spectrum Acquisition with Vertex 70 Guide to Operating the Bruker FT-IR Microscope I. Basic Spectrum Acquisition with Vertex 70 Susheng Tan, Ph.D. NanoScale Fabrication and Characterization Facility, University of Pittsburgh SB60 Benedum

More information

Microscopy from Carl Zeiss

Microscopy from Carl Zeiss Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path

More information

Zeiss 880 Training Notes Zen 2.3

Zeiss 880 Training Notes Zen 2.3 Zeiss 880 Training Notes Zen 2.3 1 Turn on the HXP 120V Lamp 2 Turn on Main Power Switch Turn on the Systems PC Switch Turn on the Components Switch. 3 4 5 Turn on the PC and log into your account. Start

More information

Zeiss AxioImager.Z2 Brightfield Protocol

Zeiss AxioImager.Z2 Brightfield Protocol Zeiss AxioImager.Z2 Brightfield Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge

More information

Instructions for the Experiment

Instructions for the Experiment Instructions for the Experiment Excitonic States in Atomically Thin Semiconductors 1. Introduction Alongside with electrical measurements, optical measurements are an indispensable tool for the study of

More information

LSM 510 Training Notes

LSM 510 Training Notes LSM 510 Training Notes Turning on the system Turn on the arc lamp, found on the bench top left of the microscope. This supplies light for epifluorescence for viewing your samples through the microscope.

More information

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL MIF ZEISS LSM510 CONFOCAL USER PROTOCOL START-UP Turn on the Mercury Bulb Power Supply (if needed). Power-on the Control Box. Turn on the computer. Open the LSM 510 software. Choose Scan New Images and

More information

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Introduction of Fluoresence Confocal Microscopy The first confocal microscope was invented by Princeton

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Version : 07/08/0 Leica SP8 TCS Users Manual Start up:. Turn the PC Microscope, Scanner Power, Laser Power, and the Laser Emission key to on (bottom right of desk).. Turn on the fluorescent lamp (top left

More information

Ocean Optics R-2000 Raman Spectrometer Setup and Operating Instructions Arlen Viste and Deanna Donohoue April 2000 Update 2003, DEW

Ocean Optics R-2000 Raman Spectrometer Setup and Operating Instructions Arlen Viste and Deanna Donohoue April 2000 Update 2003, DEW Ocean Optics R-2000 Raman Spectrometer Setup and Operating Instructions Arlen Viste and Deanna Donohoue April 2000 Update 2003, DEW References Raman Systems R-2000 Operating Manual, Version 1.6, Ocean

More information

Guide to Confocal 5. Starting session

Guide to Confocal 5. Starting session Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope

More information

Cell Biology and Bioimaging Core

Cell Biology and Bioimaging Core Cell Biology and Bioimaging Core Leica TCS SP5 Operating Instructions Starting up the instrument 1. First, log in the log book located on the confocal desk. Include your name, your lab s PI, an account

More information

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling Zeiss LSM 510 Confocor III Training Notes Center for Cell Analysis & Modeling Confocor 3 Start Up Go to System Module Turn on Main Switch, System/ PC, and Components Switches Do you need the arc lamp?

More information

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Operations Guide Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Nikon Eclipse 90i Operations Guide STARTUP Startup Powering Up Fluorescence

More information

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s LSM 5 MP, LSM 510 and LSM 510 META M i c r o s c o p y f r o m C a r l Z e i s s Quick Guide Laser Scanning Microscopes LSM Software ZEN 2007 August 2007 We make it visible. Contents Page Contents... 1

More information

SOP: EDAX Eagle III Microspot XRF

SOP: EDAX Eagle III Microspot XRF SOP: EDAX Eagle III Microspot XRF Page 1 of 6 SOP: EDAX Eagle III Microspot XRF 1. Scope 1.1 This document describes the standard operating procedure (SOP) for the EDAX Eagle III Microspot XRF. This X-ray

More information

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017. Nikon AZ100 Laser Scanning Macro Confocal Microscope Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin May 2017 Contents 1 Introduction 2 2 Hardware - Startup 2 3 Software/Operation 4 3.1 Multidimensional

More information

Figure 1 The Raith 150 TWO

Figure 1 The Raith 150 TWO RAITH 150 TWO SOP Figure 1 The Raith 150 TWO LOCATION: Raith 150 TWO room, Lithography area, NanoFab PRIMARY TRAINER: SECONDARY TRAINER: 1. OVERVIEW The Raith 150 TWO is an ultra high resolution, low voltage

More information

LSM 800 Confocal Microscope Standard Operation Protocol

LSM 800 Confocal Microscope Standard Operation Protocol LSM 800 Confocal Microscope Standard Operation Protocol Turning on the system 1. Switch on the Main switch (labeled 1 and 2 ) mounted on the wall. 2. Turn the Laser Key (labeled 3 ) 90 clockwise for power

More information

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Training Guide for Leica SP8 Confocal/Multiphoton Microscope Training Guide for Leica SP8 Confocal/Multiphoton Microscope LAS AF v3.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON power switch for epifluorescence

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, EL6000 fluorescent light source for the microscope stand. 2. Turn on the Scanner Power

More information

LEICA TCS SP5 AOBS TANDEM USER MANUAL

LEICA TCS SP5 AOBS TANDEM USER MANUAL LEICA TCS SP5 AOBS TANDEM USER MANUAL STARTING THE SYSTEM...2 THE LAS AF SOFTWARE...3 THE «ACQUIRE» MENU...5 CHOOSE AND CREATE A SETTING...6 THE CONTROL PANEL...8 THE DMI6000B MICROSCOPE...10 ACQUIRE ONE

More information

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide Matyas Molnar, Biovis 2016 Starting the microscpe 1. Check the microscope if everything looks clean and normal. If not, report it in the logbook.

More information

Leica Sp5 II Confocal User Guide

Leica Sp5 II Confocal User Guide Leica Sp5 II Confocal User Guide Turning on the Confocal System (instructions are posted in the room) 1. Turn on Laser Power Button 2. Turn Key to On position 3. Turn on Scanner Power Button 4. Turn on

More information

LSM 510 Meta Training Notes

LSM 510 Meta Training Notes LSM 510 Meta Training Notes Turning on the system Turn on X-Cite power supply. This supplies light for epifluorescence for viewing your samples through the microscope. Turn on the remote control switch.

More information

UV-Vis-NIR Spectrophotometer Quick Start Guide

UV-Vis-NIR Spectrophotometer Quick Start Guide UV-Vis-NIR Spectrophotometer Quick Start Guide The following instructions are provided as a Quick Start Guide for powering up, running measurements, and shutting down the Lambda 950 UV-Vis Spectrophotometer.

More information

This procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface.

This procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface. Tescan MIRA3 SEM: EDS using EDAX TEAM Nicholas G. Rudawski ngr@ufl.edu Cell: (805) 252-4916 Office: (352) 392-3077 Last updated: 12/04/17 This procedure assumes the user is already familiar with basic

More information

Bruker Optical Profilometer SOP Revision 2 01/04/16 Page 1 of 13. Bruker Optical Profilometer SOP

Bruker Optical Profilometer SOP Revision 2 01/04/16 Page 1 of 13. Bruker Optical Profilometer SOP Page 1 of 13 Bruker Optical Profilometer SOP The Contour GT-I, is a versatile bench-top optical surface-profiling system that can measure a wide variety of surfaces and samples. Contour GT optical profilers

More information

Point Calibration. July 3, 2012

Point Calibration. July 3, 2012 Point Calibration July 3, 2012 The purpose of the Point Calibration process is to generate a map of voltages (for galvos) or motor positions of the pointing device to the voltages or pixels of the reference

More information

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE University of Virginia ITC Academic Computing Health Sciences CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE Introduction The Olympus BH-2 microscope in ACHS s microscope lab has objectives from

More information

Standard Operating Procedure

Standard Operating Procedure Standard Operating Procedure Nanosurf Atomic Force Microscopy Operation Facility NCCRD Nanotechnology Center for Collaborative Research and Development Department of Chemistry and Engineering Physics The

More information

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL START THE SYSTEM... 2 START ZEN SOFTWARE... 3 SET THE TEMPERATURE AND THE CO2 CONTROLLERS... OBSERVATION AT OCULARS... 5 STATIF PRESENTATION... 6 ACQUIRE ONE

More information

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: )

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: ) University of Minnesota College of Science and Engineering Characterization Facility PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: 2012.10.17) The following instructions

More information

R I T. Title: Wyko RST Plus. Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 05/23/06 1 SCOPE 2 REFERENCE DOCUMENTS

R I T. Title: Wyko RST Plus. Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 05/23/06 1 SCOPE 2 REFERENCE DOCUMENTS Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Wyko RST Plus. All users are expected to have read and understood this document.

More information

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes Turn on Main Switch, System PC and Components Switches 780 Start up sequence Do you need the argon laser (458, 488, 514 nm lines)? Yes Turn on the laser s main power switch and

More information

Fast Laser Raman Microscope RAMAN

Fast Laser Raman Microscope RAMAN Fast Laser Raman Microscope RAMAN - 11 www.nanophoton.jp Fast Raman Imaging A New Generation of Raman Microscope RAMAN-11 developed by Nanophoton was created by combining confocal laser microscope technology

More information

ENSC 470/894 Lab 3 Version 6.0 (Nov. 19, 2015)

ENSC 470/894 Lab 3 Version 6.0 (Nov. 19, 2015) ENSC 470/894 Lab 3 Version 6.0 (Nov. 19, 2015) Purpose The purpose of the lab is (i) To measure the spot size and profile of the He-Ne laser beam and a laser pointer laser beam. (ii) To create a beam expander

More information

ISCapture User Guide. advanced CCD imaging. Opticstar

ISCapture User Guide. advanced CCD imaging. Opticstar advanced CCD imaging Opticstar I We always check the accuracy of the information in our promotional material. However, due to the continuous process of product development and improvement it is possible

More information

ScanArray Overview. Principle of Operation. Instrument Components

ScanArray Overview. Principle of Operation. Instrument Components ScanArray Overview The GSI Lumonics ScanArrayÒ Microarray Analysis System is a scanning laser confocal fluorescence microscope that is used to determine the fluorescence intensity of a two-dimensional

More information

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope Nikon Eclipse Ti A1-A Confocal Operating Manual Start-up 1. Turn on Excite Fluorescent light power supply- metal halide. a. Cool down as for mercury bulb b. Wheel closed liquid light guide 2. Turn on power

More information

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Verify that main power switches on the

More information

Leica DMi8A Quick Guide

Leica DMi8A Quick Guide Leica DMi8A Quick Guide 1 Optical Microscope Quick Start Guide The following instructions are provided as a Quick Start Guide for powering up, running measurements, and shutting down Leica s DMi8A Inverted

More information

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide Operation 58. UltraVIEW VoX Getting Started Guide Overview This chapter outlines the basic methods used to operate the UltraVIEW VoX system. About other software Volocity places great demands on the computer

More information

Fast Laser Raman Microscope RAMAN

Fast Laser Raman Microscope RAMAN Fast Laser Raman Microscope RAMAN - 11 www.nanophoton.jp Fast Raman Imaging A New Generation of Raman Microscope RAMAN-11 developed by Nanophoton was created by combining confocal laser microscope technology

More information

Quick Start MMRC Nicolet 6700 Revised 10/24/2018

Quick Start MMRC Nicolet 6700 Revised 10/24/2018 Do Not Remove Quick Start MMRC Nicolet 6700 Revised 10/24/2018 1. Fill out the logbook before you begin. 2. Before you start the Nicolet should be on with purge gas flowing. 3. Check that the purge gas

More information

The operation manual of spotlight 300 IR microscope

The operation manual of spotlight 300 IR microscope The operation manual of spotlight 300 IR microscope Make sure there is no sample under the microscope and then click spotlight on the desktop to open the software. You can do imaging with the image mode

More information

Nasmyth Ultraview Vox User Protocol

Nasmyth Ultraview Vox User Protocol Nasmyth Ultraview Vox User Protocol Switch on all wall sockets labelled Nasmyth, switch camera on (power supply located on table behind monitor), switch on laser switch in laser rack, switch computer on

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, Fluorescent Light for the microscope stand. 2. Turn on the Scanner Power (1) on the front

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Leica SP8 TCS Users Manual Follow the procedure for start up and log on as posted in the lab. Please log on with your account only and do not share your password with anyone. We track and confirm usage

More information

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

3. are adherent cells (ie. cells in suspension are too far away from the coverslip) Before you begin, make sure your sample... 1. is seeded on #1.5 coverglass (thickness = 0.17) 2. is an aqueous solution (ie. fixed samples mounted on a slide will not work - not enough difference in refractive

More information

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: Confocal imaging on the Leica TCS SP8 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: 4) Do not touch the microscope while the software is initializing. Choose your options: Turn

More information

Ph 3455 The Photoelectric Effect

Ph 3455 The Photoelectric Effect Ph 3455 The Photoelectric Effect Required background reading Tipler, Llewellyn, section 3-3 Prelab Questions 1. In this experiment you will be using a mercury lamp as the source of photons. At the yellow

More information

Fastest high definition Raman imaging. Fastest Laser Raman Microscope RAMAN

Fastest high definition Raman imaging. Fastest Laser Raman Microscope RAMAN Fastest high definition Raman imaging Fastest Laser Raman Microscope RAMAN - 11 www.nanophoton.jp Observation A New Generation in Raman Observation RAMAN-11 developed by Nanophoton was newly created by

More information

Widefield 1. Switching on

Widefield 1. Switching on Widefield 1 Switching on 1. Ignite DG5 lamp - must be switched on first (if previous user has switched off, wait 30 min before igniting) 2. Wait 5s and then turn on the main DG5 controller switch. 3. DG5

More information

Simplified Instructions: Olympus Widefield Microscope S1230

Simplified Instructions: Olympus Widefield Microscope S1230 Contents General Microscope Operation Simple Image Capture Multi-Wavelength Capture Z-Series Timelapse Combining Capture Modes Synopsis of Other Functions Pages 2-23 24-40 41-47 48-56 57-59 60-68 69-83

More information

Characterization Microscope Nikon LV150

Characterization Microscope Nikon LV150 Characterization Microscope Nikon LV150 Figure 1: Microscope Nikon LV150 Introduction This upright optical microscope is designed for investigating up to 150 mm (6 inch) semiconductor wafers but can also

More information

CONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application

CONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application CONFIGURING Your Spectroscopy System For PEAK PERFORMANCE A guide to selecting the best Spectrometers, s, and s for your application Spectral Measurement System Spectral Measurement System Spectrograph

More information

ThermaViz. Operating Manual. The Innovative Two-Wavelength Imaging Pyrometer

ThermaViz. Operating Manual. The Innovative Two-Wavelength Imaging Pyrometer ThermaViz The Innovative Two-Wavelength Imaging Pyrometer Operating Manual The integration of advanced optical diagnostics and intelligent materials processing for temperature measurement and process control.

More information

Practical work no. 3: Confocal Live Cell Microscopy

Practical work no. 3: Confocal Live Cell Microscopy Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside

More information

RAITH e-line OPERATING INSTRUCTIONS

RAITH e-line OPERATING INSTRUCTIONS RAITH e-line OPERATING INSTRUCTIONS 1) LOADING A SAMPLE a. Start the system i. On the Column PC (Right side monitor [R]), select the SmartSEM icon to on the desktop to begin the column software. ii. On

More information

Nikon SIM-E & A1-R System

Nikon SIM-E & A1-R System Nikon SIM-E & A1-R System USER GUIDE LSU Health Sciences Center Shreveport Research Core Facility June 01 2017 Chaowei Shang 1 Table of Content 1. Start Up the System... Page 3 Hardware and microscope

More information

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions Starting up: Schematic of Olympus BX-61. 1. Turn on Olympus microscope power box (left of microscope) with toggle switch on

More information

Table of Contents. Lesson 1 Getting Started

Table of Contents. Lesson 1 Getting Started NX Lesson 1 Getting Started Pre-reqs/Technical Skills Basic computer use Expectations Read lesson material Implement steps in software while reading through lesson material Complete quiz on Blackboard

More information

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide Nikon C1si Spectral Laser Scanning Confocal Microscope User Guide Contents: C1Si Turn-On/ShutDown Procedures... 2 Overview... 4 Setup for epi-illumination to view through the eyepieces:... 5 Setup for

More information

CONFOCAL MICROSCOPE CM-1

CONFOCAL MICROSCOPE CM-1 CONFOCAL MICROSCOPE CM-1 USER INSTRUCTIONS Scientific Instruments Dr. J.R. Sandercock Im Grindel 6 Phone: +41 44 776 33 66 Fax: +41 44 776 33 65 E-Mail: info@jrs-si.ch Internet: www.jrs-si.ch 1. Properties

More information

1. Preliminary sample preparation

1. Preliminary sample preparation FEI Helios NanoLab 600 standard operating procedure Nicholas G. Rudawski ngr@ufl.edu (352) 392 3077 (office) (805) 252-4916 (cell) Last updated: 03/02/18 What this document provides: an overview of basic

More information

Heidelberg µpg 101 Laser Writer

Heidelberg µpg 101 Laser Writer Heidelberg µpg 101 Laser Writer Standard Operating Procedure Revision: 3.0 Last Updated: Aug.1/2012, Revised by Nathanael Sieb Overview This document will provide a detailed operation procedure of the

More information

SEM Training Notebook

SEM Training Notebook SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside December 21, 2017 (rev. 3.4) 1 Before you begin Complete

More information

Scanning Procedure Using SilverFast Ai Software (6.6.1r2b) 1

Scanning Procedure Using SilverFast Ai Software (6.6.1r2b) 1 last modified 6.16.2015 Scanning Procedure Using SilverFast Ai Software (6.6.1r2b) 1 PREP In scanner notebook, fill out Incoming Checklist column on a Scanner Equipment Checklist sheet. Use flash drive

More information

Using the Nikon TE2000 Inverted Microscope

Using the Nikon TE2000 Inverted Microscope Wellcome Trust Centre for Human Genetics Molecular Cytogenetics and Microscopy Core Using the Nikon TE2000 Inverted Microscope Fluorescence image acquisition using Scanalytic s IPLab software and the B&W

More information

Using the Microscope for a NANSLO Remote Web-based Science Lab Activity

Using the Microscope for a NANSLO Remote Web-based Science Lab Activity Using the Microscope for a NANSLO Remote Web-based Science Lab Activity MICROSCOPE RWSL LAB INTERFACE INSTRUCTIONS The Remote Web-based Science Lab (RWSL) microscope is a high quality digital microscope

More information

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each lens with a 1 Airy unit pinhole setting. Many water lenses

More information

Standard Operating Procedure

Standard Operating Procedure CENTER FOR NANOSCALE SCIENCE AND ENGINEERING Standard Operating Procedure Microscope Software Brian Wajdyk Page 1 of 6 Important Images are not to be saved to the computer. They will be deleted without

More information

ZEISS LSM510META confocal manual

ZEISS LSM510META confocal manual ZEISS LSM510META confocal manual Switching on the system 1) Switch on the Remote Control button located on the table to the right of the microscope. This is the main switch for the whole system including

More information

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 1. Utility Requirements a. System power is supplied by two 120 VAC/20 A circuits. When doing maintenance

More information

Raman Spectroscopy of TiO2

Raman Spectroscopy of TiO2 Raman Spectroscopy of TiO2 Experiment #4 Characterization of Materials (96.445/545) Meg Noah Meg Noah 1 of 8 10/21/2010 Objective The purpose of this laboratory is to identify two different phases (rutile

More information

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software September 2007 Check website for most current User Guide

More information

Contents. Introduction

Contents. Introduction Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path and lasers... 12 Scanning

More information