Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

Size: px
Start display at page:

Download "Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual"

Transcription

1 Scanning Electron Microscope FEI INSPECT F50 Step by step operation manual

2 Scanning Electron Microscope, FEI Inspect F50

3 FE-SEM-F Observation Flow Saving Data And Analysis Specimen preparation Error check Closing Procedure Start Observation ESEM FE-SEM-F FE-SEM-H FE-SEM-J Confirm the SEM Chamber by CCD Confirm Error Message of SEM And PC Monitor Mount Specimen on The Stub Mount Stub on The Specimen Holder of SEM Measure The Height of The Specimen holder Measure The Height of The Specimen Holder Set The Stage of SEM to The Specimen Switch On Electron Beam Insert The Specimen Holder into The SEM Chamber Select WD=12mm at 250x of The High Mag Mode Select Home Stage or Appropriate WD Adjust The Focus at The Present Condition Link (Z) to WD Gently Adjust The Actual Height of (z) Manually, for Focusing The Image With The WD You Choose Start Observation With The Low Magnification Save Images Low Vacuum / EDX BSD,STEM / EDX Switch Off Electron Beam Confirm The SEM Chamber by CCD Set The SEM Stage to The Specimen Take Out Specimen And Data Leave The Room After Verifying That The Error Message Does Not Appear

4 FE-SEM-F Observation Flow Saving Data And Analysis Specimen preparation Error check Closing Procedure Start Observation ESEM FE-SEM-F FE-SEM-H FE-SEM-J Confirm the SEM Chamber by CCD Confirm Error Message of SEM And PC Monitor Mount Specimen on The Stub Mount Stub on The Specimen Holder of SEM Measure The Height of The Specimen holder Measure The Height of The Specimen Holder Set The Stage of SEM to The Specimen Switch On Electron Beam Insert The Specimen Holder into The SEM Chamber Select WD=12mm at 250x of The High Mag Mode Select Home Stage or Appropriate WD Adjust The Focus at The Present Condition Link (Z) to WD Gently Adjust The Actual Height of (z) Manually, for Focusing The Image With The WD You Choose Start Observation With The Low Magnification Save Images Low Vacuum / EDX BSD,STEM / EDX Switch Off Electron Beam Confirm The SEM Chamber by CCD Set The SEM Stage to The Specimen Take Out Specimen And Data Leave The Room After Verifying That The Error Message Does Not Appear

5 How to Observe Your Specimen Software should be always running. Just switch on the monitor switch 1.Locate your specimen in the SEM chamber 1 Confirm SEM condition 1. Check vacuum level of SEM. Vacuum indicator (B) should be green Pressure (D) < 1.56e-4 mbar 2. Make sure the specimen holder will not touch the SEM column by CCD. C A 2 Vent Chamber 1. Click Vent(A). Wait until the chamber becomes the air condition state. The vacuum indicator (B) displayed in gray. 2. Open the door gently when the vacuum indicator (B) presented in gray High vacuum Middle of Evac / Vac Air condition 3 Mount and tighten the stub in the sample Holder using a screw B D Do not Screw Too Much 4 Check the height of stage with specimen using scale Adjust height of the sample holder including the stub. The maximum height must be below 10 of the scale. It is adjustable by turning the top of screw 2 after loosen a base of sample holder 1, or by using Z knob of the stage controller on the SEM chamber door. 5 Close the Door 1. Close the door with placing your hands and click Pump (C). 2. Waite until pressure (D) reach less than 1.56e-4mbar and the vacuum indicator (B) is presented in green. Sample holder Stub Platform Stage Base Stage base

6 FE-ESEM-F Observation Flow Saving Data And Analysis Specimen preparation Error check Closing Procedure Start Observation ESEM FE-SEM-F FE-SEM-H FE-SEM-J Confirm the SEM Chamber by CCD Confirm Error Message of SEM And PC Monitor Mount Specimen on The Stub Mount Stub on The Specimen Holder of SEM Measure The Height of The Specimen holder Measure The Height of The Specimen Holder Set The Stage of SEM to The Specimen Switch On Electron Beam Insert The Specimen Holder into The SEM Chamber Select WD=12mm at 250x of The High Mag Mode Select Home Stage or Appropriate WD Adjust The Focus at The Present Condition Link (Z) to WD Gently Adjust The Actual Height of (z) Manually, for Focusing The Image With The WD You Choose Start Observation With The Low Magnification Save Images Low Vacuum / EDX BSD,STEM / EDX Switch Off Electron Beam Confirm The SEM Chamber by CCD Set The SEM Stage to The Specimen Take Out Specimen And Data Leave The Room After Verifying That The Error Message Does Not Appear

7 6 Stage Preparation E Activate CCD monitor and make sure the sample holder will never hit the SEM column. Then click Centre position (E) in Stage tab. The sample holder moves to directly below the column, which is the centre position of SEM. * Confirm X and Y of the control panel (F) are also X=0 and Y=0. F 2. Electron Beam And Spot Size Settings H G Select appropriate acceleration voltage and spot size using pull down key or adjuster(g) and (H). *acceleration voltage and Spot Size is able to change during observation anytime during observation. High Voltage : 200 V ~ 30 kv Magnification : 30 x ~ 1,000,000 x 3. Rink Z to WD (Working Distance) I 1. Click Beam On (I).The Indicator now should be presented in green. If the indicator is presented in red, the filament may be broken. J 2. Increase magnification (J) little by little, while adjusting focus until 1000x. Magnification can be changed by pressing + and - on the keyboard; it can be double by + key, and make it half by - key. Also, magnification can be changed in the input directly. A blank column will be appeared after double-click the number above the slider. Enter the magnification of your choice. Pointer on the screen will become Arrow ( ) with right click of the mouse. Focus of the image can be adjusted by moving the arrow( ) from side to side. Same as focusing process, Stigma is adjustable by moving the mouse with right click while pressing Shift key.

8 FE-SEM-F Observation Flow Saving Data And Analysis Specimen preparation Error check Closing Procedure Start Observation ESEM FE-SEM-F FE-SEM-H FE-SEM-J Confirm the SEM Chamber by CCD Confirm Error Message of SEM And PC Monitor Mount Specimen on The Stub Mount Stub on The Specimen Holder of SEM Measure The Height of The Specimen holder Measure The Height of The Specimen Holder Set The Stage of SEM to The Specimen Switch On Electron Beam Insert The Specimen Holder into The SEM Chamber Select WD=12mm at 250x of The High Mag Mode Select Home Stage or Appropriate WD Adjust The Focus at The Present Condition Link (Z) to WD Gently Adjust The Actual Height of (z) Manually, for Focusing The Image With The WD You Choose Start Observation With The Low Magnification Save Images Low Vacuum / EDX BSD,STEM / EDX Switch Off Electron Beam Confirm The SEM Chamber by CCD Set The SEM Stage to The Specimen Take Out Specimen And Data Leave The Room After Verifying That The Error Message Does Not Appear

9 4. After adjust focus, actual height (Z) MUST be linked to WD. Click (K). 5. Confirm the indication (K) is now displayed as (L) and WD and Z are also similar distance. WD(Working Distance) is distance between a just focus position on the specimen and SEM Column. K L L Before Link Height (z) and WD are different After Link Height (z) and WD are similar 4. Actual Observation, Capture and Saving Images A. Search An Area of Interest Track mode (effective at low magnification) : A yellow ball appears on the monitor by pressing a middle button of the mouse. Move it to any direction by a cursor. Get mode : An area ( or subject) of interest can be located to the centre of the monitor by left double click of the mouse. Arrow in the keyboard : An observation area moves to the direction of arrow for approx. 80 % of a field of view. Beam Shift Mode / effective at high magnification : a hand mark, which is represented as shift of electron beam, appears on the monitor by pressing a left button of the mouse while pressing [Ctrl]. A field of view can be moved by beam shift, without actual stage moving. B. Adjusting Focus and Stigma Use the Selected area frame (M) to adjust focus and stigma. The selected area frame allows changing magnification within the frame area only. Therefore, focus and stigma can make a fine adjustment while minimizing beam irradiation damage on the surface. 1. Adjust focus precisely. M 2. Adjust stigma if required. Cross-type arrow appears after right -click of the mouse while pressing Shift key. Move mouse in left and right for X-axis stigma while pressing Shift key, and then, move it in top and bottom for y-axis stigma. 3. Adjust focus again. Repeat process 1 and 2, until obtaining the finely adjusted image.

10 FE-SEM-F Observation Flow Saving Data And Analysis Specimen preparation Error check Closing Procedure Start Observation ESEM FE-SEM-F FE-SEM-H FE-SEM-J Confirm the SEM Chamber by CCD Confirm Error Message of SEM And PC Monitor Mount Specimen on The Stub Mount Stub on The Specimen Holder of SEM Measure The Height of The Specimen holder Measure The Height of The Specimen Holder Set The Stage of SEM to The Specimen Switch On Electron Beam Insert The Specimen Holder into The SEM Chamber Select WD=12mm at 250x of The High Mag Mode Select Home Stage or Appropriate WD Adjust The Focus at The Present Condition Link (Z) to WD Gently Adjust The Actual Height of (z) Manually, for Focusing The Image With The WD You Choose Start Observation With The Low Magnification Save Images Low Vacuum / EDX BSD,STEM / EDX Switch Off Electron Beam Confirm The SEM Chamber by CCD Set The SEM Stage to The Specimen Take Out Specimen And Data Leave The Room After Verifying That The Error Message Does Not Appear

11 C. Adjusting Contrast and Brightness Click Auto contrast (N) Or, use sliding bar(o) of the user interface. Arrows (o-2) can change current condition finely, but knob at the centre (o-1) can change it coarsely. O-1 N O-2 D. Changing Scan Speed P Slow scan :Turtle (P) / Fast scan :Rabbit (Q) Scan speed can be changed by clicking Turtle or Rabbit. Setting of scanning speed can be changed following, Scan (R) Preference (S) Scanning (T) Q R T S E. Filter for imaging Select Live, Average, Integrate Live : Real-time imaging Average : Default setting / Fast scan will be useful during observation. However, in order to capture a fine image, Number of Flame should be set up four to 16. Integrate : The integrate mode is effective if materials are easily affected by the electrostatic charge. Scanning will be paused once automatically after integrated the set number of times.

12 FE-SEM-F Observation Flow Saving Data And Analysis Specimen preparation Error check Closing Procedure Start Observation ESEM FE-SEM-F FE-SEM-H FE-SEM-J Confirm the SEM Chamber by CCD Confirm Error Message of SEM And PC Monitor Mount Specimen on The Stub Mount Stub on The Specimen Holder of SEM Measure The Height of The Specimen holder Measure The Height of The Specimen Holder Set The Stage of SEM to The Specimen Switch On Electron Beam Insert The Specimen Holder into The SEM Chamber Select WD=12mm at 250x of The High Mag Mode Select Home Stage or Appropriate WD Adjust The Focus at The Present Condition Link (Z) to WD Gently Adjust The Actual Height of (z) Manually, for Focusing The Image With The WD You Choose Start Observation With The Low Magnification Save Images Low Vacuum / EDX BSD,STEM / EDX Switch Off Electron Beam Confirm The SEM Chamber by CCD Set The SEM Stage to The Specimen Take Out Specimen And Data Leave The Room After Verifying That The Error Message Does Not Appear

13 V F. Capture images U 1. Adjust focus and stigma with appropriate scan speed and magnification. 2. Click Camera (U), or Press F2 button. Camera mode allows capturing images with the fixed scan speed, which is set up in Scan Preference. After each capturing, image on the monitor will be frozen. Pixel size of captured image is 1024 x 884; appropriate exposure time is 30 to 90 seconds. 3. After capturing process, Save as menu (V) appeared automatically. For image saving, you must create your own holder in the ESEM datastation (W). And then, save images. Do not create a holder in the SEM PC except the ESEM datastation. *Make sure those checkbox were selected W 4. Take out your data with your USB. and remove your date from the ESEM datastation immediately. Please make sure you saved all of your data in your USB. Data in the ESEM datastation will be deleted without prior notice.

14 FE-SEM-F Observation Flow Saving Data And Analysis Specimen preparation Error check Closing Procedure Start Observation ESEM FE-SEM-F FE-SEM-H FE-SEM-J Confirm the SEM Chamber by CCD Confirm Error Message of SEM And PC Monitor Mount Specimen on The Stub Mount Stub on The Specimen Holder of SEM Measure The Height of The Specimen holder Measure The Height of The Specimen Holder Set The Stage of SEM to The Specimen Switch On Electron Beam Insert The Specimen Holder into The SEM Chamber Select WD=12mm at 250x of The High Mag Mode Select Home Stage or Appropriate WD Adjust The Focus at The Present Condition Link (Z) to WD Gently Adjust The Actual Height of (z) Manually, for Focusing The Image With The WD You Choose Start Observation With The Low Magnification Save Images Low Vacuum / EDX BSD,STEM / EDX Switch Off Electron Beam Confirm The SEM Chamber by CCD Set The SEM Stage to The Specimen Take Out Specimen And Data Leave The Room After Verifying That The Error Message Does Not Appear

15 5. Closing Procedure 1 1. Click Beam On 1 button to switch off the electron beam. 2. Confirm the stage height of SEM that will never hit the SEM column by CCD Click Vent 2 and, wait until the chamber becomes the air condition state. The vacuum indicator (B) displayed in gray. 3. Gently open the door and take out the specimen 4. Close the door and then click Pump 4. Wait until the chamber becomes the high vacuum condition, such as <1.56e mbar. 3 It is compulsory that confirming following conditions before you leave the room. PC monitor is showing inside of the chamber by CCD Vacuum indicator should be presented in green Keep running the SEM software Switch off the PC monitor only No error message or alarm are appeared

16 Closing Procedure Saving Data And Analysis FE-SEM-F Observation Flow - BSD, STEM and EDX Error check Specimen preparation Start Observation Save Images BSD STEM EDX Need to Change Detector * Contact us Analyze your Specimen Saving images and data Switch Off Electron Beam Confirm The SEM Chamber by CCD Take Out Specimen And Data Leave The Room After Verifying That The Error Message Does Not Appear

17 FE-SEM-F Observation Flow - BSD, STEM and EDX Cautionary note BSD STEM To attach detector is necessary. Please contact our staff. Both detectors fix to the bottom of the SEM column. So, you have to be seriously careful because you may give serious damages to both detectors if your specimen are not flat. Therefore, you have to confirm the actual position of stage through the CCD camera often, especially, when you want to change the stage position. Recommended observation condition BSD STEM Aperture Size 6 6 Acceleration Voltage Spot Size Higher than normal observation. You must find out the appropriate condition depending on your material, which the BSD and the STEM detectors are able to detect sufficient signals. Same as normal observation WD EDX Cooling the detector is necessary before starting analysis. It is very important to extend the detector s service life, cooling must be switched off. life Detector EDX SDD(Silicon Drift Detector) Aperture Size 6 Acceleration voltage Spot Size Theoretically, an appropriate acceleration voltage, which is 2-3 times higher than the energies of K-shell or L-shell of elements of interests, should be selected. Same as normal observation WD 10 Please contact our staff when you are not able to obtain sufficient signals during your experiments.

Scanning Electron Microscope in Our Facility

Scanning Electron Microscope in Our Facility SEM Training Scanning Electron Microscope in Our Facility Specifications Table SEM ESEM FE-SEM-F FE-SEM-J FE-SEM-H FE-SEM-CZ Device name TM3030 Inspect S50 Inspect F50 JSM-7600 S-4700 Marlin compact Company

More information

Using the Hitachi 3400-N VP-SEM

Using the Hitachi 3400-N VP-SEM Using the Hitachi 3400-N VP-SEM Opening the Chamber to Load Specimens (This may also be done later using the software) 1. Click the AIR button on the front of the machine: 2. Wait a few minutes until you

More information

FE-SEM SU-8020 Operating manual (Preliminary version)

FE-SEM SU-8020 Operating manual (Preliminary version) FE-SEM SU-8020 Operating manual (Preliminary version) 2016/04/11 Seimitsu Bunseki sitsu lab. Starting up 1.Turn on the Display switch. Windows OS is starting up 2. Select the user SU-8000. 3. Click the

More information

JEOL 6500 User Manual

JEOL 6500 User Manual LOG IN to your session on the computer to the left of the microscope. Starting Conditions 1. Press Ctrl-Alt-Del and log on to the microscope computer. Click on JEOL PC SEM 6500 icon. Click yes if message

More information

FEI Quanta 200 ESEM Basic instructions

FEI Quanta 200 ESEM Basic instructions FEI Quanta 200 ESEM Basic instructions Desktop and then start the UI. If the computer has restarted and you need to login, Username: supervisor and Password: supervisor Log-in to the Microscope using the

More information

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Warning Always adjust your specimen height before closing the chamber door to make sure your specimen will not hit the bottom of the lens;

More information

SEM Training Notebook

SEM Training Notebook SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside December 21, 2017 (rev. 3.4) 1 Before you begin Complete

More information

SEM Training Notebook

SEM Training Notebook SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside March 8, 2018 (rev. 3.5) 1 Before you begin Complete

More information

Model SU3500 Scanning Electron Microscope

Model SU3500 Scanning Electron Microscope Model SU3500 Scanning Electron Microscope Modified and Parts taken from Hitachi Easy Operation Guide. Before using the Model SU3500 SEM, be sure to read the [GENERAL SAFETY GUIDELINES] in the instruction

More information

Dickinson College Department of Geology

Dickinson College Department of Geology Dickinson College Department of Geology Title: Equipment: BASIC OPERATION OF THE SCANNING ELECTRON MICROSCOPE (SEM) JEOL JSM-5900 SCANNING ELECTRON MICROSCOPE Revision: 2.2 Effective Date: 1/29/2003 Author(s):

More information

1.1. Log on to the TUMI system (you cannot proceed further until this is done).

1.1. Log on to the TUMI system (you cannot proceed further until this is done). FEI DB235 SEM mode operation Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Sample loading 1.1. Log on to the TUMI system (you cannot proceed further until this is done). 1.2. The FIB software (xp)

More information

SEM OPERATION IN LOW VACUUM MODE

SEM OPERATION IN LOW VACUUM MODE SEM OPERATION IN LOW VACUUM MODE Instructions for JEOL 5800 LV The EVAC light of the SEM specimen chamber should be already lit when you approach the SEM & the SEM will have been left in the high vacuum

More information

JEOL 6700 User Manual 05/18/2009

JEOL 6700 User Manual 05/18/2009 JEOL 6700 User Manual 05/18/2009 LOG IN to your session on the computer to the right of the microscope. Starting Conditions 1. Click the button and read the Penning Gauge to ensure that the microscope

More information

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400. Smith College August 2009 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check 1 Startup 1 Specimen Insertion 2 Filament Saturation 2 Beam Alignment

More information

1. Preliminary sample preparation

1. Preliminary sample preparation FEI Helios NanoLab 600 standard operating procedure Nicholas G. Rudawski ngr@ufl.edu (352) 392 3077 (office) (805) 252-4916 (cell) Last updated: 03/02/18 What this document provides: an overview of basic

More information

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 1. Utility Requirements a. System power is supplied by two 120 VAC/20 A circuits. When doing maintenance

More information

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center LEO 912 TEM Short Manual Prepared/copyrighted by RH Berg Danforth Plant Science Center Specimen holder [1] Never touch the holder (outside of the O-ring, double-headed arrow) because finger oils will contaminate

More information

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only Version 1.0 Prepared by D. Turnbull February 21, 2007. Please submit any omissions to the Author Note: This SEM is a recent

More information

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES RULES All users must go through a series of standard operation procedure training. For more information contact: Longlong Liao Teaching

More information

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course)

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course) JEOL 7000F BASIC OPERATING INSTRUCTIONS-Ver.-2.0 Note: This is minimal operation checklist and does not replace the other reference manuals. Read the manual for Specimen Exchange (JEOL 7000 Specimen Exchange

More information

ZEISS EVO SOP. May 2017 ELECTRON OPTICS

ZEISS EVO SOP. May 2017 ELECTRON OPTICS ZEISS EVO SOP May 2017 ELECTRON OPTICS The patented EVO column is the area of the SEM, where electrons are emitted, accelerated, deflected, focused, and scanned. Main characteristics of the EVO optics

More information

2. Raise HT to 200kVby following the procedure explained in 1.6.

2. Raise HT to 200kVby following the procedure explained in 1.6. JEOL 2100 MANUAL Quick check list 1. If needed, fill the reservoir with LN2 2. Raise HT to 200kVby following the procedure explained in 1.6. 3. Insert specimen holder into TEM (Insert holder in airlock,

More information

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. Smith College August 2005 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check, 1 Specimen Insertion, 1 Startup, 2 Filament Saturation, 2 Beam Alignment,

More information

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE by Doug Bray Department of Biological Sciences University of Lethbridge Revised September, 2000 Note: The terms in bold in this document represent

More information

RAITH e-line OPERATING INSTRUCTIONS

RAITH e-line OPERATING INSTRUCTIONS RAITH e-line OPERATING INSTRUCTIONS 1) LOADING A SAMPLE a. Start the system i. On the Column PC (Right side monitor [R]), select the SmartSEM icon to on the desktop to begin the column software. ii. On

More information

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Amray 1830 SEM. All users are expected to have read and understood this document.

More information

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right Quick Guide to Operating FEI Titan Themis G2 200 (S)TEM: TEM mode Susheng Tan Nanoscale Fabrication and Characterization Facility, University of Pittsburgh Office: M104/B01 Benedum Hall, 412-383-5978,

More information

Basic Users Manual for Tecnai-F20 TEM

Basic Users Manual for Tecnai-F20 TEM Basic Users Manual for Tecnai-F20 TEM NB: This document contains my personal notes on the operating procedure of the Tecnai F20 and may be used as a rough guide for those new to the microscope. It may

More information

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack). FEI Tecnai F20 S/TEM: imaging in TEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 01/21/18 1. Filling the cold trap (if needed) 1.1. Prior to use, the cold trap needs

More information

The user should already be familiar with operation of the instrument in STEM mode, use of the Microscope Control interface, and TIA.

The user should already be familiar with operation of the instrument in STEM mode, use of the Microscope Control interface, and TIA. FEI Tecnai F20 S/TEM: EDS system operation Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 01/22/18 The user should already be familiar with operation of the instrument in

More information

FEI Tecnai G 2 F20 Operating Procedures

FEI Tecnai G 2 F20 Operating Procedures FEI Tecnai G 2 F20 Operating Procedures 1. Startup (1) Sign-up in the microscope log-sheet. Please ensure you have written an account number for billing. (2) Log in to the computer: Login to your account

More information

User Operation of JEOL 1200 EX II

User Operation of JEOL 1200 EX II **Log onto Computer** Open item program Start Up Procedure User Operation of JEOL 1200 EX II 1. If scope is not running, locate an electron microscopy technician (EMT) to find out why not. 2. Turn up brightness

More information

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface.

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface. FEI Tecnai F20 S/TEM: imaging in STEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 05/10/18 This document assumes the user is already familiar with basic operation

More information

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000 The ideal K-12 science microscope solution User Guide for use with the Nova5000 NovaScope User Guide Information in this document is subject to change without notice. 2009 Fourier Systems Ltd. All rights

More information

Nanoscale Fabrication & Characterization Facility. Raith e-line EBL Users Guide (updated:aug 2 nd, 2017)

Nanoscale Fabrication & Characterization Facility. Raith e-line EBL Users Guide (updated:aug 2 nd, 2017) Nanoscale Fabrication & Characterization Facility Raith e-line EBL Users Guide (updated:aug 2 nd, 2017) Overview: The Raith e-line EBL system is designed to write features with critical dimensions as small

More information

Figure 1 The Raith 150 TWO

Figure 1 The Raith 150 TWO RAITH 150 TWO SOP Figure 1 The Raith 150 TWO LOCATION: Raith 150 TWO room, Lithography area, NanoFab PRIMARY TRAINER: SECONDARY TRAINER: 1. OVERVIEW The Raith 150 TWO is an ultra high resolution, low voltage

More information

Amray 3600 FESEM. Standard Operating Procedure. v2.2 modified by Bryan Cord. General Notes...3. Sample Loading...5. System Loading...

Amray 3600 FESEM. Standard Operating Procedure. v2.2 modified by Bryan Cord. General Notes...3. Sample Loading...5. System Loading... Amray 3600 FESEM Standard Operating Procedure v2.2 modified 5.13.13 by Bryan Cord Contents General Notes...3 Sample Loading...5 System Loading...8 Imaging...12 Saving Data...16 System Unloading...18 Troubleshooting...20

More information

Zeiss AxioImager.Z2 Brightfield Protocol

Zeiss AxioImager.Z2 Brightfield Protocol Zeiss AxioImager.Z2 Brightfield Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge

More information

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope The procedures given below were written specifically for the FEI Tecnai G 2 Sphera microscope. Modifications will need to

More information

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the digital camera.

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the digital camera. FEI Tecnai F20 S/TEM: acquiring diffraction patterns Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 10/18/17 This document assumes the user is already familiar with basic

More information

Jeol JEM Responsible personell: Endy ( ) Online booking is compulsory!

Jeol JEM Responsible personell: Endy ( ) Online booking is compulsory! Jeol JEM 1230 Responsible personell: Endy (45279377) Online booking is compulsory! After training you will have access to working alone on the instrument. All insertion of samples is done by responsible

More information

JEOL JEM-1400 Transmission Electron Microscope Operating Instructions

JEOL JEM-1400 Transmission Electron Microscope Operating Instructions JEOL JEM-1400 Transmission Electron Microscope Operating Instructions Anti-contamination device Objective aperture Objective aperture translation knobs Specimen holder Pump/air switch Left hand control

More information

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below.

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. http://staff.washington.edu/wpchan/if/cm100_inst.shtml Instructions for the Philips CM100 TEM and peripherals

More information

General information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument.

General information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument. JEOL JSM-7400F Table of contents General information.. 3 The operation panel. 4 The different sample holders and inserting the samples.. 5 Turning on the beam... 6 Stage map control... 8 Correcting astigmatism...

More information

Instructions for Tecnai a brief start up manual

Instructions for Tecnai a brief start up manual Instructions for Tecnai a brief start up manual Version 3.0, 8.12.2015 Manual of Tecnai 12 transmission electron microscope located at Aalto University's Nanomicroscopy Center. More information of Nanomicroscopy

More information

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier). JEOL 200CX operating procedure Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Specimen loading 1.1. Unlock the TUMI system. 1.2. Load specimen(s) into the holder. If using the double tilt holder, ensure

More information

Operation Guide. Hitachi S-3400N. Variable Pressure Scanning Electron Microscope. with. Deben Peltier Coolstage

Operation Guide. Hitachi S-3400N. Variable Pressure Scanning Electron Microscope. with. Deben Peltier Coolstage Operation Guide Hitachi S-3400N Variable Pressure Scanning Electron Microscope with Deben Peltier Coolstage www.deben.co.uk www.taltos.stanford.edu www.hitachi-hta.com Index Main Unit 3 Electron Optical

More information

Olympus Digital Microscope Camera (DP70) checklist

Olympus Digital Microscope Camera (DP70) checklist Smith College - July 2005 Olympus Digital Microscope Camera (DP70) checklist CONTENT, page no. Camera Information, 1 Startup, 1 Retrieve an Image, 2 Microscope Setup, 2 Capture, 3 Preview. 3 Color Balans,

More information

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Follow the procedures below when you use the Hitachi 7100 TEM. Starting Session 1. Turn on the cold

More information

This procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface.

This procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface. Tescan MIRA3 SEM: EDS using EDAX TEAM Nicholas G. Rudawski ngr@ufl.edu Cell: (805) 252-4916 Office: (352) 392-3077 Last updated: 12/04/17 This procedure assumes the user is already familiar with basic

More information

Standard Operating Procedure Hitachi UHR CFE SU8230 SEM

Standard Operating Procedure Hitachi UHR CFE SU8230 SEM Standard Operating Procedure Hitachi UHR CFE SU8230 Yale West Campus Materials Characterization Core ywcmatsci.yale.edu ESC II, Room E119F 810 West Campus Drive West Haven, CT 06516 Version 1.1, May 2016

More information

Zeiss LSM 880 Protocol

Zeiss LSM 880 Protocol Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Zeiss LSM 780 Protocol

Zeiss LSM 780 Protocol Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

MEASUREMENT CAMERA USER GUIDE

MEASUREMENT CAMERA USER GUIDE How to use your Aven camera s imaging and measurement tools Part 1 of this guide identifies software icons for on-screen functions, camera settings and measurement tools. Part 2 provides step-by-step operating

More information

Protective Equipment Nitrile gloves for handling sample holder and safety glasses for filling liquid nitrogen dewar.

Protective Equipment Nitrile gloves for handling sample holder and safety glasses for filling liquid nitrogen dewar. Emergency Information: 1. Medical Emergencies: Contact 911 and McGill Security 514.398.3000 2. Leave TEM as is. Do NOT shut down the vacuum system. 3. If possible, turn off High Tension and Close Column

More information

SCIENTIFIC INSTRUMENT NEWS. Introduction. Design of the FlexSEM 1000

SCIENTIFIC INSTRUMENT NEWS. Introduction. Design of the FlexSEM 1000 SCIENTIFIC INSTRUMENT NEWS 2017 Vol. 9 SEPTEMBER Technical magazine of Electron Microscope and Analytical Instruments. Technical Explanation The FlexSEM 1000: A Scanning Electron Microscope Specializing

More information

COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual

COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual 1 COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual ABOUT THIS MICROSCOPE (room HG01.240) The JEM-1400Flash

More information

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE University of Virginia ITC Academic Computing Health Sciences CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE Introduction The Olympus BH-2 microscope in ACHS s microscope lab has objectives from

More information

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I MSE 595T Basic Transmission Electron Microscopy TEM Imaging - I Purpose The purpose of this lab is to: 1. Make fine adjustments to the microscope alignment 2. Obtain a diffraction pattern 3. Obtain an

More information

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com User Manual Digital Compound Binocular LED Microscope Model MD82ES10 MicroscopeNet.com Table of Contents i. Caution... 1 ii. Care and Maintenance... 2 1. Components Illustration... 3 2. Installation...

More information

Talos on-line help User interface 1 Software version 1.6 and higher

Talos on-line help User interface 1 Software version 1.6 and higher Talos on-line help User interface 1 Talos on-line help manual -- User interface Table of Contents 1 User Interface... 3 1.1 View modes... 4 1.2 Workset tabs... 4 1.3 Control panels... 5 1.4 Popup panels...

More information

Strata DB235 FESEM FIB

Strata DB235 FESEM FIB Strata DB235 FESEM FIB Standard Operating Procedure Revision: 5.0 Last Updated: August 16/2016, revised by Li Yang Overview This document will provide a detailed operation procedure of the Focused Ion

More information

2 How to operate the microscope/obtain an image

2 How to operate the microscope/obtain an image Morgagni Operating Instructions 50079 010912 2-1 2 ow to operate the microscope/obtain an image 2.1 Starting the microscope 2.1.1 Starting the microscope with several manually-operated steps 1. Turn on

More information

Tecnai on-line help User interface 1 Tecnai F20 Tecnai F30 User interface Software version 2.1.8/3.0

Tecnai on-line help User interface 1 Tecnai F20 Tecnai F30 User interface Software version 2.1.8/3.0 Tecnai on-line help User interface 1 Tecnai on-line help manual -- User interface Table of Contents 1 User Interface...5 1.1 View modes...6 1.2 Toolbar...6 1.3 Workset tabs...7 1.4 Control panels...7 1.5

More information

CHAPTER1: QUICK START...3 CAMERA INSTALLATION... 3 SOFTWARE AND DRIVER INSTALLATION... 3 START TCAPTURE...4 TCAPTURE PARAMETER SETTINGS... 5 CHAPTER2:

CHAPTER1: QUICK START...3 CAMERA INSTALLATION... 3 SOFTWARE AND DRIVER INSTALLATION... 3 START TCAPTURE...4 TCAPTURE PARAMETER SETTINGS... 5 CHAPTER2: Image acquisition, managing and processing software TCapture Instruction Manual Key to the Instruction Manual TC is shortened name used for TCapture. Help Refer to [Help] >> [About TCapture] menu for software

More information

CELL-IQ IMAGEN USER MANUAL

CELL-IQ IMAGEN USER MANUAL CELL-IQ IMAGEN USER MANUAL FOR CELL-IQ, FOR YOUR RESEARCH NEEDS 2010 Chip-Man Technologies Ltd. All rights reserved. Protected by copyright law and international treaties. Contents 0 Introduction 2 Cell-IQ

More information

Confocal Raman Microscopy (WITec Alpha 300R)

Confocal Raman Microscopy (WITec Alpha 300R) Confocal Raman Microscopy (WITec Alpha 300R) Please refer to Witec Alpha300R Confocal Raman Microscope User Manual for the details of the operating procedure. Sample preparation 1. Attach your sample on

More information

Digital Microscope. User Manual

Digital Microscope. User Manual Digital Microscope User Manual Features The digital microscope provides 10~200X adjustable magnification range. The build-in high-performance white LED can illuminate the object without using any auxiliary

More information

Titan on-line help manual -- User Interface

Titan on-line help manual -- User Interface 1 Titan on-line help manual -- User Interface Table of Contents 1 User Interface... 6 1.1 View modes... 7 1.2 Workset tabs... 7 1.3 Control panels... 7 1.4 Popup panels... 9 1.5 Display... 12 1.5.1 Binding

More information

Operating Procedures for MICROCT1 Nikon XTH 225 ST

Operating Procedures for MICROCT1 Nikon XTH 225 ST Operating Procedures for MICROCT1 Nikon XTH 225 ST Ensuring System is Ready (go through to ensure all windows and tasks below have been completed either by you or someone else prior to mounting and scanning

More information

Schottky Emission VP FE-SEM

Schottky Emission VP FE-SEM Schottky Emission VP FE-SEM Variable Pressure The Scanning Electron Microscope (SEM) has played an important role for many years for research and development of advanced materials in the leading edge of

More information

Manual for BMS E1 eplan series, compound microscope

Manual for BMS E1 eplan series, compound microscope Manual for BMS E1 eplan series, compound microscope The compound microscope allows it to study, at cell level, structures of textures of botanical and zoological nature. (e.g. slides of roots, leaves and

More information

Scanning Electron Microscope. Instructions for Use

Scanning Electron Microscope. Instructions for Use Scanning Electron Microscope Instructions for Use The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders are liable for damages.

More information

Bruker Dimension Icon AFM Quick User s Guide

Bruker Dimension Icon AFM Quick User s Guide Bruker Dimension Icon AFM Quick User s Guide March 3, 2015 GLA Contacts Jingjing Jiang (jjiang2@caltech.edu 626-616-6357) Xinghao Zhou (xzzhou@caltech.edu 626-375-0855) Bruker Tech Support (AFMSupport@bruker-nano.com

More information

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop 1. Log in usage using the SMIF web site 2. Turn power

More information

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer Brightfield Microscopy and Image Acquisition on Spotcam1 by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer Log onto the computer. Enter your username and password to log onto the server.

More information

1. Specimen Holder Removal, Loading, and Insertion

1. Specimen Holder Removal, Loading, and Insertion OPERATION OF THE PHILIPS CM-200 FEG-TEM When not in use, the CM-200 should be in the MICROSCOPE ON configuration with the HIGH TENSION ON (illuminates green when the high tension is on).. The microscope

More information

Digital Portable Overhead Document Camera LV-1010

Digital Portable Overhead Document Camera LV-1010 Digital Portable Overhead Document Camera LV-1010 Instruction Manual 1 Content I Product Introduction 1.1 Product appearance..3 1.2 Main functions and features of the product.3 1.3 Production specifications.4

More information

SOP: EDAX Eagle III Microspot XRF

SOP: EDAX Eagle III Microspot XRF SOP: EDAX Eagle III Microspot XRF Page 1 of 6 SOP: EDAX Eagle III Microspot XRF 1. Scope 1.1 This document describes the standard operating procedure (SOP) for the EDAX Eagle III Microspot XRF. This X-ray

More information

How to use the Jeol 1010 TEM of GI (Liesbeth own GI version)

How to use the Jeol 1010 TEM of GI (Liesbeth own GI version) How to use the Jeol 1010 TEM of GI (Liesbeth own GI version) 1.Load the specimen Load a grid into the rod holder: USE ONLY THE TOP POSITION (blue arrow), Specimen selection on 1 (The rear one is only a

More information

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope ZEN 2012 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 If you require fluorescence imaging,

More information

Scanning electron microscope

Scanning electron microscope Scanning electron microscope 6 th CEMM workshop Maja Koblar, Sc. Eng. Physics Outline The basic principle? What is an electron? Parts of the SEM Electron gun Electromagnetic lenses Apertures Chamber and

More information

R I T. Title: Wyko RST Plus. Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 05/23/06 1 SCOPE 2 REFERENCE DOCUMENTS

R I T. Title: Wyko RST Plus. Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 05/23/06 1 SCOPE 2 REFERENCE DOCUMENTS Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Wyko RST Plus. All users are expected to have read and understood this document.

More information

Everest System / Slidebook Operating Procedures

Everest System / Slidebook Operating Procedures Everest System / Slidebook Operating Procedures NOTICE: This guide is meant to supplement training, not replace it. All users must be trained first hand by a core employee. Training of others in your lab

More information

Introduction of New Products

Introduction of New Products Field Emission Electron Microscope JEM-3100F For evaluation of materials in the fields of nanoscience and nanomaterials science, TEM is required to provide resolution and analytical capabilities that can

More information

Tecnai T12 Operating Procedures

Tecnai T12 Operating Procedures Tecnai T12 Operating Procedures I. Initial Procedures 1 II. Accelerating Voltage 3 III. Specimen Loading and Holder Insertion/Removal 3 IV. Emission Current 7 V. Alignment 7 VI. Camera Control and Imaging

More information

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION Olympus VS120-L100 Slide Scanner Standard Operating Procedure Startup 1) Red power bar switch (behind monitor) 2) Computer 3) Login: UserVS120 account (no password) 4) Double click: WAIT FOR INITIALIZATION

More information

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope Last updated on 1/8/2018 Jinsong Wu, jinsong-wu@northwestern.edu Aims: The aim of this lab is to familiarize you with basic TEM alignment

More information

RENISHAW INVIA RAMAN SPECTROMETER

RENISHAW INVIA RAMAN SPECTROMETER STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw invia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples

More information

Standard Operating Procedure Hitachi UHR CFE SU8230 SEM

Standard Operating Procedure Hitachi UHR CFE SU8230 SEM Standard Operating Procedure Hitachi UHR CFE SU8230 Yale West Campus Materials Characterization Core ywcmatsci.yale.edu ESC II, Room E119F 810 West Campus Drive West Haven, CT 06516 Version 1.5, February

More information

INTRODUCTION We believe that every laboratory working in the field of nanotechnology needs an SEM, therefore we would like to introduce to you our IEM

INTRODUCTION We believe that every laboratory working in the field of nanotechnology needs an SEM, therefore we would like to introduce to you our IEM INTRODUCTION We believe that every laboratory working in the field of nanotechnology needs an SEM, therefore we would like to introduce to you our IEM series of SEM. In short space of time, our device

More information

GXCapture 8.1 Instruction Manual

GXCapture 8.1 Instruction Manual GT Vision image acquisition, managing and processing software GXCapture 8.1 Instruction Manual Contents of the Instruction Manual GXC is the shortened name used for GXCapture Square brackets are used to

More information

Renishaw InVia Raman microscope

Renishaw InVia Raman microscope Laser Spectroscopy Labs Renishaw InVia Raman microscope Operation instructions 1. Turn On the power switch, system power switch is located towards the back of the system on the right hand side. Wait ~10

More information

JEOL 5800LV SCANNING ELECTRON MICROSCOPE OPERATOR'S MANUAL

JEOL 5800LV SCANNING ELECTRON MICROSCOPE OPERATOR'S MANUAL JEOL 5800LV SCANNING ELECTRON MICROSCOPE OPERATOR'S MANUAL Michael N. Spilde Christopher Adcock Original version: October 1999 Revised: January 2006 Department of Earth and Planetary Sciences and Institute

More information

Quick and simple installation and no maintenance needed. 3 Times More affordable Than a normal SEM. Obtaining results in less than 4 minutes

Quick and simple installation and no maintenance needed. 3 Times More affordable Than a normal SEM. Obtaining results in less than 4 minutes INTRODUCTION We believe that every laboratory working in the field of nanotechnology needs an SEM, therefore we would like to introduce to you our IEM series of SEM. In short space of time, our device

More information

ScanArray Overview. Principle of Operation. Instrument Components

ScanArray Overview. Principle of Operation. Instrument Components ScanArray Overview The GSI Lumonics ScanArrayÒ Microarray Analysis System is a scanning laser confocal fluorescence microscope that is used to determine the fluorescence intensity of a two-dimensional

More information

Using Cura for the first time

Using Cura for the first time Cura User manual Using Cura for the first time Initial startup When you start Cura you will be greeted by the following wizard: This wizard will guide you through your initial checkup and calibration of

More information

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions Starting up: Schematic of Olympus BX-61. 1. Turn on Olympus microscope power box (left of microscope) with toggle switch on

More information

STEM alignment procedures

STEM alignment procedures STEM alignment procedures Step 1. ASID alignment mode 1. Write down STD for TEM, and then open the ASID control window from dialogue. Also, start Simple imager viewer program on the Desktop. 2. Click on

More information

Scanning electron microscope

Scanning electron microscope Scanning electron microscope 5 th CEMM workshop Maja Koblar, Sc. Eng. Physics Outline The basic principle? What is an electron? Parts of the SEM Electron gun Electromagnetic lenses Apertures Detectors

More information