Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016

Size: px
Start display at page:

Download "Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016"

Transcription

1 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Amray 1830 SEM. All users are expected to have read and understood this document. It is not a substitute for in-person training on the system and is not sufficient to qualify a user on the system. Failure to follow guidelines in this document may result in loss of privileges. 2 REFERENCE DOCUMENTS o Appropriate Tool Manuals 3 DEFINITIONS 3.1 LaB6 Filament used in SEM 4 TOOLS AND MATERIALS 4.1 General Description The AMRAY 1830 SEM will accommodate wafer pieces as well as whole wafers. Image capture is done with the Kinovea Image Capture software. 5 SAFETY PRECAUTIONS 5.1 Personal Safety Hazards Radiation-This equipment can produce low levels of radiation. Do not operate tool if any covers are not in place High Voltage-The SEM employs voltages that are dangerous and may be fatal to personnel. Do not attempt to defeat the protective interlock systems Mechanical Hazards-Drive assemblies in the chamber have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts. RIT SMFL Page 1 of 7

2 5.2 Hazards to the Tool Sample Loading - Use caution when loading samples in the chamber. If the sample is loaded too high, it may damage the final aperture. When the sample stage is pushed in, be careful not to hit the screen that covers the detector Stage Operation The stage is a manual stage. Always make small adjustments and then refocus on the sample. Check the working distance each time the stage is moved to ensure the sample does not damage the final lens Contamination - No SU8 is allowed in the SEM. Other organic materials need staff approval, require a lower kv setting and a gold coating. No adhesives are allowed in the chamber except for the carbon pads and carbon paint. Other adhesives will contaminate the column and damage the tool. Gold coated samples should not be used in other process equipment or the cleanroom Chamber Vacuum Do not open the V1 valve until the green vacuum ready light illuminates and V1 flashes on Panel 9. Always close V1 before turning off the console and never leave V1 open when the tool is unattended. 6 INSTRUCTIONS 6.1 Initial State Check Ensure valve V1 is closed, located on the left side of column. This is done by, (a)- pressing down on the metal tab and (b)-pushing in on the valve towards the column. (a) (b) RIT SMFL Page 2 of 7

3 R I T Fabrication Laboratory Revision: D Rev Date: 03/18/2016 Verify ION pump is on by checking the indicator light illuminated above [ION PUMP] button. Establish pump is operating correctly by looking at the display on the chamber column gauge and indicator lights. Ensure that the light that is illuminated for the Ion Pump is either the 2mA or 200µA and that the current is less than 500uA. (Panel 9) If none of the lights are illuminated then press the [METER SELECT] button until lights illuminate. If the Ion Pump does not indicate less than 500uA, get the technician. (Panel 9) Vacuum Control Panel SEM column with manual stage controls RIT SMFL SEM console with image capture computer Page 3 of 7

4 6.2 Operating the system Sample Loading On the column verify that the V1 valve is closed. It should be pushed in Verify that the V1 indicator light is flashing on the display below the column. (Panel 9) Press the Vent button on the panel below the column. (Panel 9) If rough pump does not shut off when [VENT] is depressed, get technician. ***********************CAUTION*********************** Do not push [METER SELECT] button Unless GREEN READY light is illuminated ******************************************************** SLOWLY open Chamber Door ***********************CAUTION*********************** Be careful when opening chamber door. Door swings on two sets of hinges. ******************************************************** Insert sample and secure with set-screw Slowly swing door closed and ensure sample is approximately 3 to 7 mm below the final lens. Use the Z knob on the stage to adjust the sample height. There is a height indicator on the knob to track the height while the sample is in the chamber Press Evacuate to pump out the chamber. (Panel 9) Preliminary Settings Press the [POWER ON] button on electronics console. (Panel 1) Select CRT on SEM keyboard. (Panel 11) Select AUTO on SEM keyboard. (Panel 11) Adjust SHIFT knobs to the 12 o clock position. (Panel 8) Adjust BEAM ALIGNMENT knobs to the 12 o clock position (Panel 3) Depress Slow Scan if it is lit To center the stage under the column, set X to 7675 and Y to 1520 RIT SMFL Page 4 of 7

5 6.2.3 Obtaining an Image Verify that the V1 indicator light is flashing on the display below the column. Open the V1 valve by pulling it out. (Panel 9) Press the [ ] button on the console until the desired Acceleration Potential is achieved. (Panel 3) Press the [ 10] button on the Magnification panel to 25X (Panel 7) Press the AUTO HEAT button and watch for the red light between the HEAT and BIAS knobs to go out and the green emission current lights to illuminate. Do not adjust the heat or bias knobs. (Panel 3) Emission should be approximately 10KV (50mA) 20KV (100uA) 30KV (150uA) Degauss both lenses. Press Cond Lens and then Degauss. Press Final Lens and then Degauss Adjust BEAM ALIGNMENT knobs to obtain highest brightness on CRT. (Panel 3) Brightness B and Contrast C can be varied to achieve desired image on the CRT Press the FINAL LENS button to enable focus and adjust focus by turning silver knob Zoom in or out as necessary using the Magnification buttons. 6.3 Optimizing the Image Press Wobbler to check the alignment of the final aperture. A properly aligned aperture will produce an image that pulses in and out of focus when the Wobbler is engaged. If the image shifts up and down or left and right, the final aperture needs alignment. (Panel 8) Aperture alignment is accomplished by turning knobs (1) and (2) on the front and side of the final aperture assembly until there is no up and down or left to right image shift. The aperture alignment is very sensitive and should not need much adjustment. (1) (2) RIT SMFL Page 5 of 7

6 6.3.2 Turn off the Wobbler The best way to optimize the focus and astigmatism is to zoom in on a very small spherical particle and then make the adjustments so that the small particle looks sharp and then zoom back out to your actual image. The partial field [PF] button may be pressed to look at only a small area of the screen. The size and position of this field may be shifted. This may be useful in optimizing the image due to the faster refresh for the smaller area Use the Stigmator adjustments to remove astigmatism. Adjust the x-direction to obtain the best focus and then repeat for the y-direction. (Panel 6) Press the COND LENS button to adjust the spot size. While COND LENS button is illuminated turn silver knob. When desired spot size is obtained press the DEGAUSS button and then refocus the image. System will switch back to FINAL LENS automatically. A smaller spot size will give better resolution but the image will be dimmer and grainy. Adjust the brightness and contrast. The graininess can be improved by using a slower scan to take the picture. (Panel 4) Iterate through this section again to fine tune the image. Once the Wobbler, Stigmator and Condensor Lens are optimized, you can move around the sample and only have to adjust the focus, brightness and contrast When the image looks good, use the Scan Rotation and the rotary dial to make the image level. 6.4 Picture Taking After optimizing the image, press Slow Scan and Freeze. Use the knob above the freeze button to select the scan rate (1-5); slower scan will result in a better image. Remember to take it out of Slow Scan to continue using the SEM Image capture is done with the Swiftcap Osprey program on the computer Double click on the Swiftcap Icon icon Under Tools, click on Save Snapshot Type in the desired name for the image Save to a removable drive medium only. Do not save on the PC due to the limited amount of storage. If it says Image cannot be saved, you probably exceeded the amount of storage available on the computer. 6.5 Removing a Sample Close V1. It should be pushed in. (left side of column) Press the [VENT] button. (Panel 9) If rough pump does not shut off when [VENT] is depressed, get technician Remove sample from SEM by following instruction Sample Loading Load the next sample or proceed to Shutdown. RIT SMFL Page 6 of 7

7 6.6Shutdown Set the acceleration potential down to zero or OFF Turn off electronics console by pressing the Power Off button. (Panel 1) Verify that V1 pushed in all the way and closed. (left side of column) Press Evacuate to pump out the chamber if not already done. (Panel 9) Wait until the V1 indicator light starts flashing. This indicates that the chamber is pumped down. If the chamber won t pump down, get a staff member. (Panel 9) 6.7 Errors during Run If the image is lost, return the SEM to the preliminary settings and start over If the filament is blown, no current will flow and the red light will come on and stay on. GET TECHNICIAN 7 APPROPRIATE USES OF THE TOOL 7.1 Avoid contaminating the column. Only approved samples may be used. 7.2 No out-gassing materials such as adhesives are allowed. 7.3 Organic materials must be coated with gold. 8 ATTACHMENTS 6.8 Manual for Sample Preparation in the Denton Sputterer REVISION RECORD Summary of Changes Originator Rev/Date Original Issue Sean O Brien/ A-05/25/2007 Bruce Tolleson Moved to , rewrote hazards section to match cert sheet O Brien B-09/12/2007 Renumbered sections and changed X and Y Preliminary settings Bruce Tolleson C-10/21/2013 Changed Image Capture Software to Kinovea Bruce Tolleson D-03/18/2016 RIT SMFL Page 7 of 7

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 1. Utility Requirements a. System power is supplied by two 120 VAC/20 A circuits. When doing maintenance

More information

Using the Hitachi 3400-N VP-SEM

Using the Hitachi 3400-N VP-SEM Using the Hitachi 3400-N VP-SEM Opening the Chamber to Load Specimens (This may also be done later using the software) 1. Click the AIR button on the front of the machine: 2. Wait a few minutes until you

More information

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE by Doug Bray Department of Biological Sciences University of Lethbridge Revised September, 2000 Note: The terms in bold in this document represent

More information

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES RULES All users must go through a series of standard operation procedure training. For more information contact: Longlong Liao Teaching

More information

JEOL 6500 User Manual

JEOL 6500 User Manual LOG IN to your session on the computer to the left of the microscope. Starting Conditions 1. Press Ctrl-Alt-Del and log on to the microscope computer. Click on JEOL PC SEM 6500 icon. Click yes if message

More information

R I T. Title: Wyko RST Plus. Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 05/23/06 1 SCOPE 2 REFERENCE DOCUMENTS

R I T. Title: Wyko RST Plus. Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 05/23/06 1 SCOPE 2 REFERENCE DOCUMENTS Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Wyko RST Plus. All users are expected to have read and understood this document.

More information

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Warning Always adjust your specimen height before closing the chamber door to make sure your specimen will not hit the bottom of the lens;

More information

General information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument.

General information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument. JEOL JSM-7400F Table of contents General information.. 3 The operation panel. 4 The different sample holders and inserting the samples.. 5 Turning on the beam... 6 Stage map control... 8 Correcting astigmatism...

More information

Model SU3500 Scanning Electron Microscope

Model SU3500 Scanning Electron Microscope Model SU3500 Scanning Electron Microscope Modified and Parts taken from Hitachi Easy Operation Guide. Before using the Model SU3500 SEM, be sure to read the [GENERAL SAFETY GUIDELINES] in the instruction

More information

SEM Training Notebook

SEM Training Notebook SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside March 8, 2018 (rev. 3.5) 1 Before you begin Complete

More information

User Operation of JEOL 1200 EX II

User Operation of JEOL 1200 EX II **Log onto Computer** Open item program Start Up Procedure User Operation of JEOL 1200 EX II 1. If scope is not running, locate an electron microscopy technician (EMT) to find out why not. 2. Turn up brightness

More information

FE-SEM SU-8020 Operating manual (Preliminary version)

FE-SEM SU-8020 Operating manual (Preliminary version) FE-SEM SU-8020 Operating manual (Preliminary version) 2016/04/11 Seimitsu Bunseki sitsu lab. Starting up 1.Turn on the Display switch. Windows OS is starting up 2. Select the user SU-8000. 3. Click the

More information

SEM OPERATION IN LOW VACUUM MODE

SEM OPERATION IN LOW VACUUM MODE SEM OPERATION IN LOW VACUUM MODE Instructions for JEOL 5800 LV The EVAC light of the SEM specimen chamber should be already lit when you approach the SEM & the SEM will have been left in the high vacuum

More information

2. Raise HT to 200kVby following the procedure explained in 1.6.

2. Raise HT to 200kVby following the procedure explained in 1.6. JEOL 2100 MANUAL Quick check list 1. If needed, fill the reservoir with LN2 2. Raise HT to 200kVby following the procedure explained in 1.6. 3. Insert specimen holder into TEM (Insert holder in airlock,

More information

SEM Training Notebook

SEM Training Notebook SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside December 21, 2017 (rev. 3.4) 1 Before you begin Complete

More information

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400. Smith College August 2009 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check 1 Startup 1 Specimen Insertion 2 Filament Saturation 2 Beam Alignment

More information

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier). JEOL 200CX operating procedure Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Specimen loading 1.1. Unlock the TUMI system. 1.2. Load specimen(s) into the holder. If using the double tilt holder, ensure

More information

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course)

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course) JEOL 7000F BASIC OPERATING INSTRUCTIONS-Ver.-2.0 Note: This is minimal operation checklist and does not replace the other reference manuals. Read the manual for Specimen Exchange (JEOL 7000 Specimen Exchange

More information

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. Smith College August 2005 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check, 1 Specimen Insertion, 1 Startup, 2 Filament Saturation, 2 Beam Alignment,

More information

1. Preliminary sample preparation

1. Preliminary sample preparation FEI Helios NanoLab 600 standard operating procedure Nicholas G. Rudawski ngr@ufl.edu (352) 392 3077 (office) (805) 252-4916 (cell) Last updated: 03/02/18 What this document provides: an overview of basic

More information

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual Scanning Electron Microscope FEI INSPECT F50 Step by step operation manual Scanning Electron Microscope, FEI Inspect F50 FE-SEM-F Observation Flow Saving Data And Analysis Specimen preparation Error check

More information

JEOL 6700 User Manual 05/18/2009

JEOL 6700 User Manual 05/18/2009 JEOL 6700 User Manual 05/18/2009 LOG IN to your session on the computer to the right of the microscope. Starting Conditions 1. Click the button and read the Penning Gauge to ensure that the microscope

More information

1.1. Log on to the TUMI system (you cannot proceed further until this is done).

1.1. Log on to the TUMI system (you cannot proceed further until this is done). FEI DB235 SEM mode operation Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Sample loading 1.1. Log on to the TUMI system (you cannot proceed further until this is done). 1.2. The FIB software (xp)

More information

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Follow the procedures below when you use the Hitachi 7100 TEM. Starting Session 1. Turn on the cold

More information

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below.

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. http://staff.washington.edu/wpchan/if/cm100_inst.shtml Instructions for the Philips CM100 TEM and peripherals

More information

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope Last updated on 1/8/2018 Jinsong Wu, jinsong-wu@northwestern.edu Aims: The aim of this lab is to familiarize you with basic TEM alignment

More information

JEOL JEM-1400 Transmission Electron Microscope Operating Instructions

JEOL JEM-1400 Transmission Electron Microscope Operating Instructions JEOL JEM-1400 Transmission Electron Microscope Operating Instructions Anti-contamination device Objective aperture Objective aperture translation knobs Specimen holder Pump/air switch Left hand control

More information

Amray 3600 FESEM. Standard Operating Procedure. v2.2 modified by Bryan Cord. General Notes...3. Sample Loading...5. System Loading...

Amray 3600 FESEM. Standard Operating Procedure. v2.2 modified by Bryan Cord. General Notes...3. Sample Loading...5. System Loading... Amray 3600 FESEM Standard Operating Procedure v2.2 modified 5.13.13 by Bryan Cord Contents General Notes...3 Sample Loading...5 System Loading...8 Imaging...12 Saving Data...16 System Unloading...18 Troubleshooting...20

More information

Dickinson College Department of Geology

Dickinson College Department of Geology Dickinson College Department of Geology Title: Equipment: BASIC OPERATION OF THE SCANNING ELECTRON MICROSCOPE (SEM) JEOL JSM-5900 SCANNING ELECTRON MICROSCOPE Revision: 2.2 Effective Date: 1/29/2003 Author(s):

More information

RAITH e-line OPERATING INSTRUCTIONS

RAITH e-line OPERATING INSTRUCTIONS RAITH e-line OPERATING INSTRUCTIONS 1) LOADING A SAMPLE a. Start the system i. On the Column PC (Right side monitor [R]), select the SmartSEM icon to on the desktop to begin the column software. ii. On

More information

1. Specimen Holder Removal, Loading, and Insertion

1. Specimen Holder Removal, Loading, and Insertion OPERATION OF THE PHILIPS CM-200 FEG-TEM When not in use, the CM-200 should be in the MICROSCOPE ON configuration with the HIGH TENSION ON (illuminates green when the high tension is on).. The microscope

More information

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack). FEI Tecnai F20 S/TEM: imaging in TEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 01/21/18 1. Filling the cold trap (if needed) 1.1. Prior to use, the cold trap needs

More information

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only Version 1.0 Prepared by D. Turnbull February 21, 2007. Please submit any omissions to the Author Note: This SEM is a recent

More information

Strata DB235 FESEM FIB

Strata DB235 FESEM FIB Strata DB235 FESEM FIB Standard Operating Procedure Revision: 5.0 Last Updated: August 16/2016, revised by Li Yang Overview This document will provide a detailed operation procedure of the Focused Ion

More information

Instructions for Tecnai a brief start up manual

Instructions for Tecnai a brief start up manual Instructions for Tecnai a brief start up manual Version 3.0, 8.12.2015 Manual of Tecnai 12 transmission electron microscope located at Aalto University's Nanomicroscopy Center. More information of Nanomicroscopy

More information

FEI Quanta 200 ESEM Basic instructions

FEI Quanta 200 ESEM Basic instructions FEI Quanta 200 ESEM Basic instructions Desktop and then start the UI. If the computer has restarted and you need to login, Username: supervisor and Password: supervisor Log-in to the Microscope using the

More information

Basic Users Manual for Tecnai-F20 TEM

Basic Users Manual for Tecnai-F20 TEM Basic Users Manual for Tecnai-F20 TEM NB: This document contains my personal notes on the operating procedure of the Tecnai F20 and may be used as a rough guide for those new to the microscope. It may

More information

2 How to operate the microscope/obtain an image

2 How to operate the microscope/obtain an image Morgagni Operating Instructions 50079 010912 2-1 2 ow to operate the microscope/obtain an image 2.1 Starting the microscope 2.1.1 Starting the microscope with several manually-operated steps 1. Turn on

More information

SOP: EDAX Eagle III Microspot XRF

SOP: EDAX Eagle III Microspot XRF SOP: EDAX Eagle III Microspot XRF Page 1 of 6 SOP: EDAX Eagle III Microspot XRF 1. Scope 1.1 This document describes the standard operating procedure (SOP) for the EDAX Eagle III Microspot XRF. This X-ray

More information

SCIENTIFIC INSTRUMENT NEWS. Introduction. Design of the FlexSEM 1000

SCIENTIFIC INSTRUMENT NEWS. Introduction. Design of the FlexSEM 1000 SCIENTIFIC INSTRUMENT NEWS 2017 Vol. 9 SEPTEMBER Technical magazine of Electron Microscope and Analytical Instruments. Technical Explanation The FlexSEM 1000: A Scanning Electron Microscope Specializing

More information

JEOL 5DII. Operation introduction. By Serge Charlebois

JEOL 5DII. Operation introduction. By Serge Charlebois JEOL 5DII Operation introduction By Serge Charlebois July 2003 General procedure Loading the cassette in the load lock Selecting EOS mode, table and aperture Setting and maximising current Observation

More information

ZEISS EVO SOP. May 2017 ELECTRON OPTICS

ZEISS EVO SOP. May 2017 ELECTRON OPTICS ZEISS EVO SOP May 2017 ELECTRON OPTICS The patented EVO column is the area of the SEM, where electrons are emitted, accelerated, deflected, focused, and scanned. Main characteristics of the EVO optics

More information

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right Quick Guide to Operating FEI Titan Themis G2 200 (S)TEM: TEM mode Susheng Tan Nanoscale Fabrication and Characterization Facility, University of Pittsburgh Office: M104/B01 Benedum Hall, 412-383-5978,

More information

Operation Guide. Hitachi S-3400N. Variable Pressure Scanning Electron Microscope. with. Deben Peltier Coolstage

Operation Guide. Hitachi S-3400N. Variable Pressure Scanning Electron Microscope. with. Deben Peltier Coolstage Operation Guide Hitachi S-3400N Variable Pressure Scanning Electron Microscope with Deben Peltier Coolstage www.deben.co.uk www.taltos.stanford.edu www.hitachi-hta.com Index Main Unit 3 Electron Optical

More information

Protective Equipment Nitrile gloves for handling sample holder and safety glasses for filling liquid nitrogen dewar.

Protective Equipment Nitrile gloves for handling sample holder and safety glasses for filling liquid nitrogen dewar. Emergency Information: 1. Medical Emergencies: Contact 911 and McGill Security 514.398.3000 2. Leave TEM as is. Do NOT shut down the vacuum system. 3. If possible, turn off High Tension and Close Column

More information

FEI Tecnai G 2 F20 Operating Procedures

FEI Tecnai G 2 F20 Operating Procedures FEI Tecnai G 2 F20 Operating Procedures 1. Startup (1) Sign-up in the microscope log-sheet. Please ensure you have written an account number for billing. (2) Log in to the computer: Login to your account

More information

Standard Operating Procedure Hitachi UHR CFE SU8230 SEM

Standard Operating Procedure Hitachi UHR CFE SU8230 SEM Standard Operating Procedure Hitachi UHR CFE SU8230 Yale West Campus Materials Characterization Core ywcmatsci.yale.edu ESC II, Room E119F 810 West Campus Drive West Haven, CT 06516 Version 1.1, May 2016

More information

Figure 1 The Raith 150 TWO

Figure 1 The Raith 150 TWO RAITH 150 TWO SOP Figure 1 The Raith 150 TWO LOCATION: Raith 150 TWO room, Lithography area, NanoFab PRIMARY TRAINER: SECONDARY TRAINER: 1. OVERVIEW The Raith 150 TWO is an ultra high resolution, low voltage

More information

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope The procedures given below were written specifically for the FEI Tecnai G 2 Sphera microscope. Modifications will need to

More information

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and

More information

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center LEO 912 TEM Short Manual Prepared/copyrighted by RH Berg Danforth Plant Science Center Specimen holder [1] Never touch the holder (outside of the O-ring, double-headed arrow) because finger oils will contaminate

More information

Introduction: Why electrons?

Introduction: Why electrons? Introduction: Why electrons? 1 Radiations Visible light X-rays Electrons Neutrons Advantages Not very damaging Easily focused Eye wonderful detector Small wavelength (Angstroms) Good penetration Small

More information

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface.

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface. FEI Tecnai F20 S/TEM: imaging in STEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 05/10/18 This document assumes the user is already familiar with basic operation

More information

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I MSE 595T Basic Transmission Electron Microscopy TEM Imaging - I Purpose The purpose of this lab is to: 1. Make fine adjustments to the microscope alignment 2. Obtain a diffraction pattern 3. Obtain an

More information

Functions of the SEM subsystems

Functions of the SEM subsystems Functions of the SEM subsystems Electronic column It consists of an electron gun and two or more electron lenses, which influence the path of electrons traveling down an evacuated tube. The base of the

More information

SAMUEL ROBERTS NOBLE ELECTRON MICROSCOPY LABORATORY. Operating Procedures for the Zeiss 9 S-2. Transmission Electron Microscope

SAMUEL ROBERTS NOBLE ELECTRON MICROSCOPY LABORATORY. Operating Procedures for the Zeiss 9 S-2. Transmission Electron Microscope 1 SAMUEL ROBERTS NOBLE ELECTRON MICROSCOPY LABORATORY Operating Procedures for the Zeiss 9 S-2 Transmission Electron Microscope Prepared by Dr. Scott D. Russell Department of Botany and Microbiology September,

More information

JEOL JEM 2010 TRAINING TRANSMISSION ELECTRON MICROSCOPE USER MANUAL

JEOL JEM 2010 TRAINING TRANSMISSION ELECTRON MICROSCOPE USER MANUAL JEOL JEM 2010 TRAINING TRANSMISSION ELECTRON MICROSCOPE USER MANUAL Version 5.1 EM Facility CMSE-SEF Massachusetts Institution of Technology TABLE OF CONTENTS 1. Specifications...2 1.1 Performance...2

More information

Zeiss AxioImager.Z2 Brightfield Protocol

Zeiss AxioImager.Z2 Brightfield Protocol Zeiss AxioImager.Z2 Brightfield Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge

More information

S200 Course LECTURE 1 TEM

S200 Course LECTURE 1 TEM S200 Course LECTURE 1 TEM Development of Electron Microscopy 1897 Discovery of the electron (J.J. Thompson) 1924 Particle and wave theory (L. de Broglie) 1926 Electromagnetic Lens (H. Busch) 1932 Construction

More information

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+)

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+) Standard Operating Procedure of Atomic Force Microscope (Anasys afm+) The Anasys Instruments afm+ system incorporates an Atomic Force Microscope which can scan the sample in the contact mode and generate

More information

CHA ELECTRON BEAM EVAPORATOR SYSTEM OPERATING MANUAL Version: 2.0 August 2016

CHA ELECTRON BEAM EVAPORATOR SYSTEM OPERATING MANUAL Version: 2.0 August 2016 CHA ELECTRON BEAM EVAPORATOR SYSTEM OPERATING MANUAL Version: 2.0 August 2016 UNIVERSITY OF TEXAS AT ARLINGTON Nanotechnology Research Center TABLE OF CONTENTS 1.0 DESCRIPTION.... 3 2.0 SAFETY....... 3

More information

Introduction to Scanning Electron Microscopy

Introduction to Scanning Electron Microscopy Introduction to Scanning Electron Microscopy By: Brandon Cheney Ant s Leg Integrated Circuit Nano-composite This document was created as part of a Senior Project in the Materials Engineering Department

More information

Scanning electron microscope

Scanning electron microscope Scanning electron microscope 6 th CEMM workshop Maja Koblar, Sc. Eng. Physics Outline The basic principle? What is an electron? Parts of the SEM Electron gun Electromagnetic lenses Apertures Chamber and

More information

RENISHAW INVIA RAMAN SPECTROMETER

RENISHAW INVIA RAMAN SPECTROMETER STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw invia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples

More information

STEM alignment procedures

STEM alignment procedures STEM alignment procedures Step 1. ASID alignment mode 1. Write down STD for TEM, and then open the ASID control window from dialogue. Also, start Simple imager viewer program on the Desktop. 2. Click on

More information

Standard Operating Procedure Hitachi UHR CFE SU8230 SEM

Standard Operating Procedure Hitachi UHR CFE SU8230 SEM Standard Operating Procedure Hitachi UHR CFE SU8230 Yale West Campus Materials Characterization Core ywcmatsci.yale.edu ESC II, Room E119F 810 West Campus Drive West Haven, CT 06516 Version 1.5, February

More information

CM20 USER GUIDE. Duncan Alexander, CIME 2010

CM20 USER GUIDE. Duncan Alexander, CIME 2010 CM20 USER GUIDE Duncan Alexander, CIME 2010 CM20 START UP AND CHECK LIST 2 SPECIMEN EXCHANGE 5 - REMOVING SAMPLE HOLDER 6 - INSERTING SAMPLE HOLDER 7 TURNING ON HT 8 STARTING THE FILAMENT 9 GUN TILT ALIGNMENT

More information

Cressington 108 Auto/SE Sputter Coater Standard Operating Procedures (S.O.P)

Cressington 108 Auto/SE Sputter Coater Standard Operating Procedures (S.O.P) Cressington 108 Auto/SE Sputter Coater Standard Operating Procedures (S.O.P) The Cressington sputter system is designed for only one purpose which is the deposition of gold onto a sample to reduce charging

More information

1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height.

1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height. JEOL 2010F operating procedure Covers operation in STEM mode (See separate procedures for operation in TEM mode and operation of EDS system) Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 NOTE: this operating

More information

STANDARD OPERATING PROCEDURE: JEOL TEM-2100

STANDARD OPERATING PROCEDURE: JEOL TEM-2100 STANDARD OPERATING PROCEDURE: JEOL TEM-2100 Purpose of this Instrument: Essential tool for structural characterization of natural or synthesized nanostructures. Location: WVU - Engineering Sciences Building

More information

Tecnai T12 Operating Procedures

Tecnai T12 Operating Procedures Tecnai T12 Operating Procedures I. Initial Procedures 1 II. Accelerating Voltage 3 III. Specimen Loading and Holder Insertion/Removal 3 IV. Emission Current 7 V. Alignment 7 VI. Camera Control and Imaging

More information

Heidelberg µpg 101 Laser Writer

Heidelberg µpg 101 Laser Writer Heidelberg µpg 101 Laser Writer Standard Operating Procedure Revision: 3.0 Last Updated: Aug.1/2012, Revised by Nathanael Sieb Overview This document will provide a detailed operation procedure of the

More information

Zeiss LSM 880 Protocol

Zeiss LSM 880 Protocol Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

05/20/14 1. Philips CM200T. Standby Condition

05/20/14 1. Philips CM200T. Standby Condition 05/20/14 1 Philips CM200T Standby Condition HT and filament off, HT setting at 200kV. RESET HOLDER, center sample tilt knobs, and remove sample. Mag ~ 5-10kX Objective and SA apertures out, C2 aperture

More information

Zeiss LSM 780 Protocol

Zeiss LSM 780 Protocol Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

How to use the Jeol 1010 TEM of GI (Liesbeth own GI version)

How to use the Jeol 1010 TEM of GI (Liesbeth own GI version) How to use the Jeol 1010 TEM of GI (Liesbeth own GI version) 1.Load the specimen Load a grid into the rod holder: USE ONLY THE TOP POSITION (blue arrow), Specimen selection on 1 (The rear one is only a

More information

SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS

SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS Robert Edward Lee Electron Microscopy Center Department of Anatomy and Neurobiology Colorado State University P T R Prentice Hall, Englewood Cliffs,

More information

Section 1: TEM parts and functions... 2

Section 1: TEM parts and functions... 2 Introduction The set of instructions below are written by Charlie Sanabria within the first few sessions of his TEM training process, and are intended for anyone interested in viewing the TEM operation

More information

Scanning Electron Microscope. Instructions for Use

Scanning Electron Microscope. Instructions for Use Scanning Electron Microscope Instructions for Use The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders are liable for damages.

More information

Nanoscale Fabrication & Characterization Facility. Raith e-line EBL Users Guide (updated:aug 2 nd, 2017)

Nanoscale Fabrication & Characterization Facility. Raith e-line EBL Users Guide (updated:aug 2 nd, 2017) Nanoscale Fabrication & Characterization Facility Raith e-line EBL Users Guide (updated:aug 2 nd, 2017) Overview: The Raith e-line EBL system is designed to write features with critical dimensions as small

More information

A510S Operation Manual

A510S Operation Manual A510S Operation Manual REV 1.1 1 Table of Contents 1 General Information 1-1 Description 1-2 Potential Operational Hazards 1-3 Technical Specifications 1-4 Instrument Overview 1-5 Function Summary 2 How

More information

KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017)

KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017) KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017) 1 Introduction You must be trained to operate the Laser-scanning confocal microscope (LSCM) independently.

More information

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the digital camera.

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the digital camera. FEI Tecnai F20 S/TEM: acquiring diffraction patterns Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 10/18/17 This document assumes the user is already familiar with basic

More information

A Beginner s Guide To Exposure

A Beginner s Guide To Exposure A Beginner s Guide To Exposure What is exposure? A Beginner s Guide to Exposure What is exposure? According to Wikipedia: In photography, exposure is the amount of light per unit area (the image plane

More information

Components of the Microscope

Components of the Microscope Swift M3 Microscope The Swift M3 is a versatile microscope designed for both microscopic (high magnification, small field of view) and macroscopic (low magnification, large field of view) applications.

More information

Scanning electron microscope

Scanning electron microscope Scanning electron microscope 5 th CEMM workshop Maja Koblar, Sc. Eng. Physics Outline The basic principle? What is an electron? Parts of the SEM Electron gun Electromagnetic lenses Apertures Detectors

More information

ALTURA EDS. Rev. 0915

ALTURA EDS. Rev. 0915 ALTURA EDS Rev. 0915 Enable the Oxford PC Enable the Altura-EDS under Dual Beam Tools in Coral. Or enter your NETID and password directly into the Oxford PC. Warning: Ion-milling, GISs/microprobe, and

More information

User instructions Compound laboratory microscope

User instructions Compound laboratory microscope KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-mail: info@kern-sohn.com User instructions Compound laboratory microscope Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com KERN

More information

NeoScope. Simple Operation to 40,000. Table Top SEM. Serving Advanced Technology

NeoScope. Simple Operation to 40,000. Table Top SEM. Serving Advanced Technology Table Top SEM Simple Operation to 40,000 Serving Advanced Technology From 10 to 40,000 Table Top SEM Notebook PC version Just plug it to a wall outlet after placing it on a table Desktop PC version Option

More information

Olympus LEXT OLS 4000 Confocal Laser Microscope

Olympus LEXT OLS 4000 Confocal Laser Microscope Olympus LEXT OLS 4000 Confocal Laser Microscope The Olympus LEXT OLS4000 is a confocal microscope capable of taking high-resolution 3D images. The magnification (Optical and Digital) of this microscope

More information

Bruker Optical Profilometer SOP Revision 2 01/04/16 Page 1 of 13. Bruker Optical Profilometer SOP

Bruker Optical Profilometer SOP Revision 2 01/04/16 Page 1 of 13. Bruker Optical Profilometer SOP Page 1 of 13 Bruker Optical Profilometer SOP The Contour GT-I, is a versatile bench-top optical surface-profiling system that can measure a wide variety of surfaces and samples. Contour GT optical profilers

More information

Allegro Home Office Assembly Instructions. Tipping Restraint. Bun Foot W/ Leveler. 4 pcs.

Allegro Home Office Assembly Instructions. Tipping Restraint. Bun Foot W/ Leveler. 4 pcs. email: info@riverside-furniture.com Allegro Home Office Assembly Instructions Components and Hardware List Page 1 of 5 Made in China Shelf Pin Wood Shelf File Rod & Clips 1/2"Wood Screw A 16 pcs. B 4 pcs.

More information

Introduction of New Products

Introduction of New Products Field Emission Electron Microscope JEM-3100F For evaluation of materials in the fields of nanoscience and nanomaterials science, TEM is required to provide resolution and analytical capabilities that can

More information

Schottky Emission VP FE-SEM

Schottky Emission VP FE-SEM Schottky Emission VP FE-SEM Variable Pressure The Scanning Electron Microscope (SEM) has played an important role for many years for research and development of advanced materials in the leading edge of

More information

ENSC 470/894 Lab 3 Version 6.0 (Nov. 19, 2015)

ENSC 470/894 Lab 3 Version 6.0 (Nov. 19, 2015) ENSC 470/894 Lab 3 Version 6.0 (Nov. 19, 2015) Purpose The purpose of the lab is (i) To measure the spot size and profile of the He-Ne laser beam and a laser pointer laser beam. (ii) To create a beam expander

More information

Kit for building your own THz Time-Domain Spectrometer

Kit for building your own THz Time-Domain Spectrometer Kit for building your own THz Time-Domain Spectrometer 16/06/2016 1 Table of contents 0. Parts for the THz Kit... 3 1. Delay line... 4 2. Pulse generator and lock-in detector... 5 3. THz antennas... 6

More information

QUICK-START FOR UNIVERSAL VLS 4.6 LASER! FRESH 21 SEPTEMBER 2017

QUICK-START FOR UNIVERSAL VLS 4.6 LASER! FRESH 21 SEPTEMBER 2017 QUICK-START FOR UNIVERSAL VLS 4.6 LASER! FRESH 21 SEPTEMBER 2017 The laser is quite safe to use, but it is powerful; using it requires your full caution, attention and respect. Some rules of the road:

More information

CATHODOLUMINESCENCE IMAGING UNIVERSITY OF ST ANDREWS

CATHODOLUMINESCENCE IMAGING UNIVERSITY OF ST ANDREWS CATHODOLUMINESCENCE IMAGING AT THE UNIVERSITY OF ST ANDREWS Instruction Guide by Adrian Finch 2 Noddy s Guide to Using the CL at St Andrews (SHORT version) 1. Switch on plugs at Wall (Microscope, Luminoscope

More information

INSTRUCTIONS JEM-2010F FIELD-EMISSION TRANSMISSION ELECTRON MICROSCOPE WITH STEM CAPABILITY

INSTRUCTIONS JEM-2010F FIELD-EMISSION TRANSMISSION ELECTRON MICROSCOPE WITH STEM CAPABILITY INSTRUCTIONS JEM-2010F FIELD-EMISSION TRANSMISSION ELECTRON MICROSCOPE WITH STEM CAPABILITY August 2011 PRELIMINARIES OPERATION 1. Ensure that EMISSION and HT are on: The HT READY and FEG READY lights

More information

COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual

COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual 1 COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual ABOUT THIS MICROSCOPE (room HG01.240) The JEM-1400Flash

More information