LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center
|
|
- Stuart Jacobs
- 6 years ago
- Views:
Transcription
1 LEO 912 TEM Short Manual Prepared/copyrighted by RH Berg Danforth Plant Science Center
2 Specimen holder [1] Never touch the holder (outside of the O-ring, double-headed arrow) because finger oils will contaminate this area and cause drift and column contamination The LEO carrier uses a screw to secure the grids [1]. PLEASE do not strip this screw, which is easily done and costs $ to replace. If the screw or grids are not stably attached then the image will suffer from instability and, worse case, they could fall off inside the microscope, incurring a major repair.
3 Permitted panel buttons * * * * * * * * * * * * * Only use the * highlighted buttons/knobs
4 The Green Screen Illum. Angle is an indicator of beam intensity (controlled by brightness buttons, see next page). M/LM toggles between the low mag mode (highlighted) and higher mag mode (not highlighted). Beam blanker parks the beam off the specimen, to prevent un-needed beam damage. Pressing Vac will display the vacuum system screen. When changing specimen, turn off the high voltage and press Close col valve. This isolates the vacuum-sensitive IGPs (by closing valves V3 & V9 ( ) from the specimen chamber, in case vacuum is broken during specimen insertion.
5 [1] [2] [3] Generating a Beam [4] Check the vacuum system: is it ready? (green Ready light on right panel [1]). Turn beam blanker on. Heat the filament up by pressing Filament [2]: it s status light will blink red during the ~four minute period during which the filament is heated and remains constantly lit when the filament is fully heated. Press the High voltage button [3]: its status light will blink red during voltage ramp-up and remains constantly lit when high voltage has been obtained. When changing grids there is no need to turn Filament off but High voltage should be turned off, before closing the column valve (see previous page). If vacuum is lost then you will hear valves close and Filament and High voltage lights will extinguish. The ready light will relight after high vacuum is again reached. Ill Aperture [4] controls brightness (condenser lens excitation): left button decreases brightness, right increases it. To prevent beam damage to the specimen or specimen drift from heating, keep brightness to a minimum, compensating for this by using longer exposure times for the camera.
6 Condenser lens aperture selection In our LEO TEM the system is typically operated in AIS, which selects the correct condenser aperture from a foil containing five apertures, selected by beam deflectors) dependent on magnification The third click selects the AIS aperture foil Center this with the x-y controls [arrows], at 4,000X
7 Objective lens aperture [1] For optimal contrast insert an objective lens aperture--the smaller the better, so use #3 (the third click in). The objective lens aperture selector [1] inserts the aperture. Center this (at 4,000x) with a section in the field of view by clicking image selector (lower left panel, next to mag buttons [2]) so that diffraction ( diff ) is activated (this images the shadow of the aperture) and center it with the x-y controls (arrows). The screen will be permanently damaged (burned) if the diffraction beam image is left on for long: be sure to switch back to image within a few seconds [2]
8 Selected Area aperture [2] [1] This is used to boost contrast in the low magnification mode ( M/LM on the green screen [1] is highlighted). This aperture is immediately below the objective aperture and on the column left [2]. Insert the second aperture (two clicks) and center with the X-Y centering knobs [arrows]. The aperture s shadow is not circular and centering is aimed at merely eliminating the unusual shadow that is seen.
9 Focusing [3] [2] [4] [1] Focusing [1] changes the objective lens current to alter its focal length. CW from focus is over-focus, CCW from focus is under-focus. With a new grid, position it in the eucentric plane (the correct height in the microscope) by first calibrating the objective lens (press [2], let it beep) then turn on the focusing aid ([3], the wobbler), and focus by moving the grid up or down using the Zcontrol on the console below the focus knob until the oscillations are at a minimum. Be sure to turn the wobbler off. At lower magnification, focus with the wobbler. With wobbler [3] on the image appears to oscillate in and out of focus. Adjust focus [2] to minimize this (and do not forget to turn it off). In wobbling, the illuminating beam is rocked between two tilted positions on opposite sides of the optical axis at high frequency, with the pivot point at the specimen plane. Focus above or below this produces two images so by adjusting focus to the specimen plane, at near true focus, the image shift is minimized. The fine-ness of focusing steps for [1] can be toggled between coarse and fine with button [4]. This is not available in low mag mode. For high magnification focus use the contrast dip (soft background) method.
10 Energy filter entrance aperture [1] [2] This is opposite from the top of the electron spectrometer [1], on the column right [2]. The first aperture is selected and centered with the x-y knobs. There is no need to remove this aperture.
11 Energy filter slit aperture [1] This is the next aperture down from the energy filter entrance aperture, but is on the column left [1]. It images electrons elastically scattered at small angles (elastic brightfield imaging, AKA zero loss imaging because inelastically scattered electrons are blocked by the aperture). This aperture produces a major improvement in image contrast, but at the cost of reduced image brightness. Insert the second aperture and center, first with the smaller right hand knob and then with the other, if needed [arrows, note the left is the smallest/outermost of the three]. The aperture shadow is not a circle unless it is nearly centered and will not always center at the screen center. Centering alters the size of the shadow. Adjust it so that the shadow becomes circular and illumination is as small as possible without cropping the image captured by the camera. If the area of illumination is too large then a (a sometimes subtle) dark spot will appear within the illuminated area and this is undesirable (distorts the image in that area). As magnification is changed often this aperture will need to be adjusted as it tends to close (screen becomes black) with magnification increase. The slit width of this aperture can be adjusted and this affects the spectral resolution of electron energy loss spectra (EELS).
12 Choice of magnification and camera operation Use the specimen carrier mechanical selector (on the left of the column) to release the clamp and select the first grid (look on the screen and also there will be a bit of steel showing on the holder on the column right) or the second. The LEO has a low mag mode M/LM mode on the green screen, used to select a decent section and region therein, and to image large areas. 80x-2,000x. Use the second SA (selected area) aperture for contrast (this aperture selector is immediately below the objective selector, and on the left). Go to the hi mag mode by first inserting the objective aperture, then touching M/LM so that it is not highlighted. Center the objective aperture. Select a mag that images the area of interest and also provides the desired resolution. The CCD camera is sufficient size (2k x 2k pixels) to produce a large image, but images a relatively small field of the image seen on the phosphor screen. To increase the field of capture make a montage ( MIA ) with the caveat that you will likely have to massage the data to get the montage put together decently. Minimize specimen heating/contamination by blanking the beam when it is not needed (this uses beam deflectors to point it away from the specimen). To preserve filament life turn the beam off if it is not being used for a while). Exposure time is controlled using the camera control menu (open this under the SIS Image drop down menu).
13 Shut down When you are through with the microscope, remove your grids and store the specimen holder inside the microscope. Confirm that high vacuum is reached (green Ready lamp). Be sure to turn off filament heating and high voltage. Confirm that this happens by waiting the ~4 minutes for the heating lamp to stop blinking and switch off. Both red lights should be off. Turn down the green screen brightness. Turn off the sodium vapor room light.
14 Signs of Trouble The most common problem you are likely to have is centering apertures. Come get me to help you if you are having problems do this rather than move the aperture so far out of alignment that it is difficult to find or that vacuum is broken. Find or call staff ( , Howard Berg on call 24/7) if you experience any problems. Specimen grid holder screw does not tighten. It may be stripped. Do not insert the carrier into the microscope if the grids are not firmly secured and if the screw is loose. Ready light is not lit green: there is not adequate high vacuum for generating the electron beam. This could happen when inserting or removing apertures and the vacuum loss in this case is usually minimal high vacuum could return in a few minutes, signaled by the green light. Then ramp up high voltage and filament heating. More fundamental measures must be taken if the vacuum is much poorer.
MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I
MSE 595T Basic Transmission Electron Microscopy TEM Imaging - I Purpose The purpose of this lab is to: 1. Make fine adjustments to the microscope alignment 2. Obtain a diffraction pattern 3. Obtain an
More informationMSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope
MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope Last updated on 1/8/2018 Jinsong Wu, jinsong-wu@northwestern.edu Aims: The aim of this lab is to familiarize you with basic TEM alignment
More informationUser Operation of JEOL 1200 EX II
**Log onto Computer** Open item program Start Up Procedure User Operation of JEOL 1200 EX II 1. If scope is not running, locate an electron microscopy technician (EMT) to find out why not. 2. Turn up brightness
More informationTransmission Electron Microscopy 9. The Instrument. Outline
Transmission Electron Microscopy 9. The Instrument EMA 6518 Spring 2009 02/25/09 Outline The Illumination System The Objective Lens and Stage Forming Diffraction Patterns and Images Alignment and Stigmation
More informationJEOL JEM-1400 Transmission Electron Microscope Operating Instructions
JEOL JEM-1400 Transmission Electron Microscope Operating Instructions Anti-contamination device Objective aperture Objective aperture translation knobs Specimen holder Pump/air switch Left hand control
More informationBasic Users Manual for Tecnai-F20 TEM
Basic Users Manual for Tecnai-F20 TEM NB: This document contains my personal notes on the operating procedure of the Tecnai F20 and may be used as a rough guide for those new to the microscope. It may
More information1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).
FEI Tecnai F20 S/TEM: imaging in TEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 01/21/18 1. Filling the cold trap (if needed) 1.1. Prior to use, the cold trap needs
More informationScanning Electron Microscope FEI INSPECT F50. Step by step operation manual
Scanning Electron Microscope FEI INSPECT F50 Step by step operation manual Scanning Electron Microscope, FEI Inspect F50 FE-SEM-F Observation Flow Saving Data And Analysis Specimen preparation Error check
More informationMSE 460 TEM Lab 4: Bright/Dark Field Imaging Operation
MSE 460 TEM Lab 4: Bright/Dark Field Imaging Operation Last updated on 1/8/2018 Jinsong Wu, jinsong-wu@northwestern.edu Aims: The aim of this lab is to familiarize you with bright/dark field imaging operation.
More informationInstructions for Tecnai a brief start up manual
Instructions for Tecnai a brief start up manual Version 3.0, 8.12.2015 Manual of Tecnai 12 transmission electron microscope located at Aalto University's Nanomicroscopy Center. More information of Nanomicroscopy
More information1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).
JEOL 200CX operating procedure Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Specimen loading 1.1. Unlock the TUMI system. 1.2. Load specimen(s) into the holder. If using the double tilt holder, ensure
More informationFull-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right
Quick Guide to Operating FEI Titan Themis G2 200 (S)TEM: TEM mode Susheng Tan Nanoscale Fabrication and Characterization Facility, University of Pittsburgh Office: M104/B01 Benedum Hall, 412-383-5978,
More informationOperating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah
Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Follow the procedures below when you use the Hitachi 7100 TEM. Starting Session 1. Turn on the cold
More informationPlease follow these instructions for use of the Philips CM100 TEM. Adopted from website below.
Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. http://staff.washington.edu/wpchan/if/cm100_inst.shtml Instructions for the Philips CM100 TEM and peripherals
More informationFEI Tecnai G 2 F20 Operating Procedures
FEI Tecnai G 2 F20 Operating Procedures 1. Startup (1) Sign-up in the microscope log-sheet. Please ensure you have written an account number for billing. (2) Log in to the computer: Login to your account
More informationStandard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014
Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 1. Utility Requirements a. System power is supplied by two 120 VAC/20 A circuits. When doing maintenance
More information2. Raise HT to 200kVby following the procedure explained in 1.6.
JEOL 2100 MANUAL Quick check list 1. If needed, fill the reservoir with LN2 2. Raise HT to 200kVby following the procedure explained in 1.6. 3. Insert specimen holder into TEM (Insert holder in airlock,
More informationOPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge
OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE by Doug Bray Department of Biological Sciences University of Lethbridge Revised September, 2000 Note: The terms in bold in this document represent
More informationThis document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the digital camera.
FEI Tecnai F20 S/TEM: acquiring diffraction patterns Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 10/18/17 This document assumes the user is already familiar with basic
More information1. Specimen Holder Removal, Loading, and Insertion
OPERATION OF THE PHILIPS CM-200 FEG-TEM When not in use, the CM-200 should be in the MICROSCOPE ON configuration with the HIGH TENSION ON (illuminates green when the high tension is on).. The microscope
More information2 How to operate the microscope/obtain an image
Morgagni Operating Instructions 50079 010912 2-1 2 ow to operate the microscope/obtain an image 2.1 Starting the microscope 2.1.1 Starting the microscope with several manually-operated steps 1. Turn on
More informationJEOL 6500 User Manual
LOG IN to your session on the computer to the left of the microscope. Starting Conditions 1. Press Ctrl-Alt-Del and log on to the microscope computer. Click on JEOL PC SEM 6500 icon. Click yes if message
More informationBasic Operating Instructions for Strata Dual Beam 235 FIB/SEM
Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Warning Always adjust your specimen height before closing the chamber door to make sure your specimen will not hit the bottom of the lens;
More informationSection 1: TEM parts and functions... 2
Introduction The set of instructions below are written by Charlie Sanabria within the first few sessions of his TEM training process, and are intended for anyone interested in viewing the TEM operation
More informationProtective Equipment Nitrile gloves for handling sample holder and safety glasses for filling liquid nitrogen dewar.
Emergency Information: 1. Medical Emergencies: Contact 911 and McGill Security 514.398.3000 2. Leave TEM as is. Do NOT shut down the vacuum system. 3. If possible, turn off High Tension and Close Column
More informationProcedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope
Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope The procedures given below were written specifically for the FEI Tecnai G 2 Sphera microscope. Modifications will need to
More informationTecnai T12 Operating Procedures
Tecnai T12 Operating Procedures I. Initial Procedures 1 II. Accelerating Voltage 3 III. Specimen Loading and Holder Insertion/Removal 3 IV. Emission Current 7 V. Alignment 7 VI. Camera Control and Imaging
More information05/20/14 1. Philips CM200T. Standby Condition
05/20/14 1 Philips CM200T Standby Condition HT and filament off, HT setting at 200kV. RESET HOLDER, center sample tilt knobs, and remove sample. Mag ~ 5-10kX Objective and SA apertures out, C2 aperture
More informationCOMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual
1 COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual ABOUT THIS MICROSCOPE (room HG01.240) The JEM-1400Flash
More information1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height.
JEOL 2010F operating procedure Covers operation in STEM mode (See separate procedures for operation in TEM mode and operation of EDS system) Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 NOTE: this operating
More informationHow to use the Jeol 1010 TEM of GI (Liesbeth own GI version)
How to use the Jeol 1010 TEM of GI (Liesbeth own GI version) 1.Load the specimen Load a grid into the rod holder: USE ONLY THE TOP POSITION (blue arrow), Specimen selection on 1 (The rear one is only a
More informationSTEM alignment procedures
STEM alignment procedures Step 1. ASID alignment mode 1. Write down STD for TEM, and then open the ASID control window from dialogue. Also, start Simple imager viewer program on the Desktop. 2. Click on
More informationNANO 703-Notes. Chapter 9-The Instrument
1 Chapter 9-The Instrument Illumination (condenser) system Before (above) the sample, the purpose of electron lenses is to form the beam/probe that will illuminate the sample. Our electron source is macroscopic
More informationUsing the Hitachi 3400-N VP-SEM
Using the Hitachi 3400-N VP-SEM Opening the Chamber to Load Specimens (This may also be done later using the software) 1. Click the AIR button on the front of the machine: 2. Wait a few minutes until you
More informationCheck that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course)
JEOL 7000F BASIC OPERATING INSTRUCTIONS-Ver.-2.0 Note: This is minimal operation checklist and does not replace the other reference manuals. Read the manual for Specimen Exchange (JEOL 7000 Specimen Exchange
More informationMotorized Axio Observer Start-up instructions
Start-up instructions 1. If using fluorescence turn on Fluorescent light source. TL light Source (Hal 100) 2. Turn on microscope using switch on lower left side of the microscope. 3. If imaging, turn on
More informationFE-SEM SU-8020 Operating manual (Preliminary version)
FE-SEM SU-8020 Operating manual (Preliminary version) 2016/04/11 Seimitsu Bunseki sitsu lab. Starting up 1.Turn on the Display switch. Windows OS is starting up 2. Select the user SU-8000. 3. Click the
More informationJEOL 6700 User Manual 05/18/2009
JEOL 6700 User Manual 05/18/2009 LOG IN to your session on the computer to the right of the microscope. Starting Conditions 1. Click the button and read the Penning Gauge to ensure that the microscope
More informationJeol JEM Responsible personell: Endy ( ) Online booking is compulsory!
Jeol JEM 1230 Responsible personell: Endy (45279377) Online booking is compulsory! After training you will have access to working alone on the instrument. All insertion of samples is done by responsible
More informationIntroduction: Why electrons?
Introduction: Why electrons? 1 Radiations Visible light X-rays Electrons Neutrons Advantages Not very damaging Easily focused Eye wonderful detector Small wavelength (Angstroms) Good penetration Small
More informationTherefore, all descriptions and illustrations in this instruction manual, including all specifications are subject to change without notice.
We are constantly endeavouring to improve our instruments and to adapt them to the requirements of modern research techniques and testing methods. This involves modification to the mechanical structure
More informationFEI Falcon Direct Electron Detector. Best Practice Document
FEI Falcon Direct Electron Detector Best Practice Document 2 1. Introduction FEI Falcon Direct Electron Detector Best Practice Application Guide The FEI Falcon Detector is based on direct electron detection
More informationBX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions
BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions Starting up: Schematic of Olympus BX-61. 1. Turn on Olympus microscope power box (left of microscope) with toggle switch on
More informationOperating F20/F30 with SerialEM
Chen Xu xuchen@brandeis.ede $BrandeisEM: ~emdoc-xml/en_us.iso8859-1/articles/operating-f20-or-f30/article.xml, 1 2013-01-19 01:42:20 xuchen Exp$ This is a quick check list for the Tecnai F20 or Tecnai
More informationThis document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface.
FEI Tecnai F20 S/TEM: imaging in STEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 05/10/18 This document assumes the user is already familiar with basic operation
More informationJEOL 2010 FasTEM & DigitalMicrograph User's Guide
JEOL 2010 FasTEM & DigitalMicrograph User's Guide Electron Microscopy Laboratory Instititute of Materials Science University of Connecticut The purpose of this manual is to remind you of the essential
More informationSTANDARD OPERATING PROCEDURE: JEOL TEM-2100
STANDARD OPERATING PROCEDURE: JEOL TEM-2100 Purpose of this Instrument: Essential tool for structural characterization of natural or synthesized nanostructures. Location: WVU - Engineering Sciences Building
More informationINSTRUCTIONS JEM-2010F FIELD-EMISSION TRANSMISSION ELECTRON MICROSCOPE WITH STEM CAPABILITY
INSTRUCTIONS JEM-2010F FIELD-EMISSION TRANSMISSION ELECTRON MICROSCOPE WITH STEM CAPABILITY August 2011 PRELIMINARIES OPERATION 1. Ensure that EMISSION and HT are on: The HT READY and FEG READY lights
More informationHoriba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer
Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and
More informationJEOL JEM 2010 TRAINING TRANSMISSION ELECTRON MICROSCOPE USER MANUAL
JEOL JEM 2010 TRAINING TRANSMISSION ELECTRON MICROSCOPE USER MANUAL Version 5.1 EM Facility CMSE-SEF Massachusetts Institution of Technology TABLE OF CONTENTS 1. Specifications...2 1.1 Performance...2
More informationCM20 USER GUIDE. Duncan Alexander, CIME 2010
CM20 USER GUIDE Duncan Alexander, CIME 2010 CM20 START UP AND CHECK LIST 2 SPECIMEN EXCHANGE 5 - REMOVING SAMPLE HOLDER 6 - INSERTING SAMPLE HOLDER 7 TURNING ON HT 8 STARTING THE FILAMENT 9 GUN TILT ALIGNMENT
More information1. Preliminary sample preparation
FEI Helios NanoLab 600 standard operating procedure Nicholas G. Rudawski ngr@ufl.edu (352) 392 3077 (office) (805) 252-4916 (cell) Last updated: 03/02/18 What this document provides: an overview of basic
More informationNanoSpective, Inc Progress Drive Suite 137 Orlando, Florida
TEM Techniques Summary The TEM is an analytical instrument in which a thin membrane (typically < 100nm) is placed in the path of an energetic and highly coherent beam of electrons. Typical operating voltages
More informationTitle: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016
Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Amray 1830 SEM. All users are expected to have read and understood this document.
More informationBrightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer
Brightfield Microscopy and Image Acquisition on Spotcam1 by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer Log onto the computer. Enter your username and password to log onto the server.
More informationUser instructions Compound laboratory microscope
KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-mail: info@kern-sohn.com User instructions Compound laboratory microscope Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com KERN
More informationUser Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com
User Manual Digital Compound Binocular LED Microscope Model MD82ES10 MicroscopeNet.com Table of Contents i. Caution... 1 ii. Care and Maintenance... 2 1. Components Illustration... 3 2. Installation...
More informationSTEM Spectrum Imaging Tutorial
STEM Spectrum Imaging Tutorial Gatan, Inc. 5933 Coronado Lane, Pleasanton, CA 94588 Tel: (925) 463-0200 Fax: (925) 463-0204 April 2001 Contents 1 Introduction 1.1 What is Spectrum Imaging? 2 Hardware 3
More informationFilter & Spectrometer Electron Optics
Filter & Spectrometer Electron Optics Parameters Affecting Practical Performance Daniel Moonen & Harold A. Brink Did Something Go Wrong? 30 20 10 0 500 600 700 800 900 1000 1100 ev 1 Content The Prism
More informationUser instructions Compound laboratory microscope
KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-mail: info@kern-sohn.com User instructions Compound laboratory microscope Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com KERN
More information--> Buy True-PDF --> Auto-delivered in 0~10 minutes. JY/T
Translated English of Chinese Standard: JY/T011-1996 www.chinesestandard.net Sales@ChineseStandard.net INDUSTRY STANDARD OF THE JY PEOPLE S REPUBLIC OF CHINA General rules for transmission electron microscopy
More informationTecnai on-line help manual --
Tecnai on-line help Alignments 1 Tecnai on-line help manual -- Alignments Table of Contents 1 Alignments in the Tecnai microscope...5 2 Alignment procedures...6 3 Introduction to electron optics...11 3.1
More informationECEN 4606, UNDERGRADUATE OPTICS LAB
ECEN 4606, UNDERGRADUATE OPTICS LAB Lab 2: Imaging 1 the Telescope Original Version: Prof. McLeod SUMMARY: In this lab you will become familiar with the use of one or more lenses to create images of distant
More informationNikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire
Nikon Ti-E Microscope Manual Rightmire Hall Ohio State University Director: Tony Brown Rightmire 060 292-1205 brown.2302@osu.edu Facility Manager: Paula Monsma Rightmire 062 293-0939 292-1367 monsma.1@osu.edu
More informationSerialEM Help Index. ABC Amber HLP Converte r T rial version,
SerialEM Help Index Genera l Topics Introduction to SerialEM Mouse and Keyboard Controls Control Panels Image Acquisition Acquiring Tilt Series Montaging Low Dose Mode Energy Filtering Using the Navigator
More informationTransmission electron Microscopy
Transmission electron Microscopy Image formation of a concave lens in geometrical optics Some basic features of the transmission electron microscope (TEM) can be understood from by analogy with the operation
More informationComponents of the Microscope
Swift M3 Microscope The Swift M3 is a versatile microscope designed for both microscopic (high magnification, small field of view) and macroscopic (low magnification, large field of view) applications.
More informationSEM OPERATION IN LOW VACUUM MODE
SEM OPERATION IN LOW VACUUM MODE Instructions for JEOL 5800 LV The EVAC light of the SEM specimen chamber should be already lit when you approach the SEM & the SEM will have been left in the high vacuum
More informationOperating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.
Smith College August 2009 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check 1 Startup 1 Specimen Insertion 2 Filament Saturation 2 Beam Alignment
More informationDIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands
DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands Differential Interference Contrast (DIC) imaging is a technique used to increase contrast in brightfield images. In confocal systems,
More informationZeiss Axio Imager.A1 manual
Zeiss Axio Imager.A1 manual Power-up protocol 1. Mercury lamp 2. Power strip on shelf 3. Computer The Mercury lamp should always be first-on and last-off. This prevents any electrical surges caused by
More informationSEM Training Notebook
SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside December 21, 2017 (rev. 3.4) 1 Before you begin Complete
More informationFEI Titan Image Corrected STEM
05/03/16 1 FEI Titan 60-300 Image Corrected STEM Standby Condition HT setting at 300kV, Col. Valves Closed RESET Holder and remove sample. Mag ~ 5-10kX Objective and SA apertures out, C2 aperture at 150µm
More informationDickinson College Department of Geology
Dickinson College Department of Geology Title: Equipment: BASIC OPERATION OF THE SCANNING ELECTRON MICROSCOPE (SEM) JEOL JSM-5900 SCANNING ELECTRON MICROSCOPE Revision: 2.2 Effective Date: 1/29/2003 Author(s):
More informationSEM Training Notebook
SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside March 8, 2018 (rev. 3.5) 1 Before you begin Complete
More informationSystems Biology. Optical Train, Köhler Illumination
McGill University Life Sciences Complex Imaging Facility Systems Biology Microscopy Workshop Tuesday December 7 th, 2010 Simple Lenses, Transmitted Light Optical Train, Köhler Illumination What Does a
More informationJSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES
JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES RULES All users must go through a series of standard operation procedure training. For more information contact: Longlong Liao Teaching
More informationSOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only
SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only Version 1.0 Prepared by D. Turnbull February 21, 2007. Please submit any omissions to the Author Note: This SEM is a recent
More informationS200 Course LECTURE 1 TEM
S200 Course LECTURE 1 TEM Development of Electron Microscopy 1897 Discovery of the electron (J.J. Thompson) 1924 Particle and wave theory (L. de Broglie) 1926 Electromagnetic Lens (H. Busch) 1932 Construction
More informationML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X
JAPAN DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30 (a) Field Iris Control Lever (c) Filter Slots EYEPIECES, KHW10X ANALYZER CONTROL LEVER (b) Aperture Iris Control Lever LIGHT SOURCE HOUSING VERTICAL
More informationDevelopment of JEM-2800 High Throughput Electron Microscope
Development of JEM-2800 High Throughput Electron Microscope Mitsuhide Matsushita, Shuji Kawai, Takeshi Iwama, Katsuhiro Tanaka, Toshiko Kuba and Noriaki Endo EM Business Unit, JEOL Ltd. Electron Optics
More informationRENISHAW RAMAN MICROSCOPE STANDARD OPERATING PROCEDURE
RENISHAW RAMAN MICROSCOPE STANDARD OPERATING PROCEDURE 09DJ00 CHRIS BUXEY (TECHNICIAN) X2715 C.BUXEY@SURREY.AC.UK V5.0 OCTOBER 2008 RENISHAW MICRORAMAN STANDARD OPERATING PROCEDURE 1.0 Foreword Chris Buxey:
More informationCAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE
University of Virginia ITC Academic Computing Health Sciences CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE Introduction The Olympus BH-2 microscope in ACHS s microscope lab has objectives from
More informationManual for BMS E1 eplan series, compound microscope
Manual for BMS E1 eplan series, compound microscope The compound microscope allows it to study, at cell level, structures of textures of botanical and zoological nature. (e.g. slides of roots, leaves and
More informationSAMUEL ROBERTS NOBLE ELECTRON MICROSCOPY LABORATORY. Operating Procedures for the Zeiss 9 S-2. Transmission Electron Microscope
1 SAMUEL ROBERTS NOBLE ELECTRON MICROSCOPY LABORATORY Operating Procedures for the Zeiss 9 S-2 Transmission Electron Microscope Prepared by Dr. Scott D. Russell Department of Botany and Microbiology September,
More informationZeiss AxioImager.Z2 Brightfield Protocol
Zeiss AxioImager.Z2 Brightfield Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge
More informationOperating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.
Smith College August 2005 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check, 1 Specimen Insertion, 1 Startup, 2 Filament Saturation, 2 Beam Alignment,
More informationMicroscopy. The dichroic mirror is an important component of the fluorescent scope: it reflects blue light while transmitting green light.
Microscopy I. Before coming to lab Read this handout and the background. II. Learning Objectives In this lab, you'll investigate the physics of microscopes. The main idea is to understand the limitations
More informationUsing the Microscope for a NANSLO Remote Web-based Science Lab Activity
Using the Microscope for a NANSLO Remote Web-based Science Lab Activity MICROSCOPE RWSL LAB INTERFACE INSTRUCTIONS The Remote Web-based Science Lab (RWSL) microscope is a high quality digital microscope
More informationQuick Operation Guide
Quick Operation Guide Power ON Mounting specimens Set the specimen on the sample holder, and install the sample holder to the holder frame. Attach the holder frame to the XY stage. Type of holder Main
More informationNikon E800 Operating Instructions.
Nikon E800 Operating Instructions. You can request electronic copies of this manual by contacting lshats@jhsph.edu Copies are also available on the JHU MMI Department web site. Please send your comments
More informationBe aware that there is no universal notation for the various quantities.
Fourier Optics v2.4 Ray tracing is limited in its ability to describe optics because it ignores the wave properties of light. Diffraction is needed to explain image spatial resolution and contrast and
More informationFigure 1 The Raith 150 TWO
RAITH 150 TWO SOP Figure 1 The Raith 150 TWO LOCATION: Raith 150 TWO room, Lithography area, NanoFab PRIMARY TRAINER: SECONDARY TRAINER: 1. OVERVIEW The Raith 150 TWO is an ultra high resolution, low voltage
More informationAdvanced Polarising Microscope Operating Instructions
PriorLux POL Advanced Polarising Microscope Operating Instructions q~ääé=çñ=`çåíéåíë= 1. Introduction 2 2. Unpacking 2 3. Specifications 3 4. Component Parts 4 5. Electrical Connections & Safety 5 6.
More informationOptics Day 3 Kohler Illumination (Philbert Tsai July 2004) Goal : To build an bright-field microscope with a Kohler illumination pathway
Optics Day 3 Kohler Illumination (Philbert Tsai July 2004) Goal : To build an bright-field microscope with a Kohler illumination pathway Prepare the Light source and Lenses Set up Light source Use 3 rail
More informationModel SU3500 Scanning Electron Microscope
Model SU3500 Scanning Electron Microscope Modified and Parts taken from Hitachi Easy Operation Guide. Before using the Model SU3500 SEM, be sure to read the [GENERAL SAFETY GUIDELINES] in the instruction
More informationRIGAKU VariMax Dual Part 0 Startup & Shutdown Manual
i RIGAKU VariMax Dual Part 0 Startup & Shutdown Manual X-ray Laboratory, Nano-Engineering Research Center, Institute of Engineering Innovation, School of Engineering, The University of Tokyo Figure 0:
More informationSCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS
SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS Robert Edward Lee Electron Microscopy Center Department of Anatomy and Neurobiology Colorado State University P T R Prentice Hall, Englewood Cliffs,
More informationCharacterization Microscope Nikon LV150
Characterization Microscope Nikon LV150 Figure 1: Microscope Nikon LV150 Introduction This upright optical microscope is designed for investigating up to 150 mm (6 inch) semiconductor wafers but can also
More informationRAITH e-line OPERATING INSTRUCTIONS
RAITH e-line OPERATING INSTRUCTIONS 1) LOADING A SAMPLE a. Start the system i. On the Column PC (Right side monitor [R]), select the SmartSEM icon to on the desktop to begin the column software. ii. On
More information