JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES
|
|
- Dominick Day
- 5 years ago
- Views:
Transcription
1 JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES
2 RULES All users must go through a series of standard operation procedure training. For more information contact: Longlong Liao Teaching Assistant 206 Wilhelm Hall Ames Laboratory lliao@iastate.edu Phone: (515) All users must read and UNDERSTAND this standard operating procedure manual. All users must FILL in the logbook for every usage. No one is allowed to tamper with the safety measures, interlock or vacuum systems. ANY USER WHO SHOWS DISREGARD FOR THE SAFETY PRECAUTIONS WILL HAVE THEIR SEM PRIVILEGES REVOKED. You should reserve time prior to using instrument. To schedule a session visit: Please be advised that you must use your scheduled time. You can delete a previously scheduled period 24 hours in advance with no penalty. If you schedule an SEM session and do not attend, you ll still be charged for the scheduled time. You can check out the room key from Karen Knight (2220 Hoover Hall).
3 Don t forget to return the keys in time. After-hours privileges will only be given after the user has acquired at least 30 hours of SEM time beyond training and has demonstrated competence in operating the equipment. After-hours privileges require special permission from Scott Chumbley. Please save your pictures in your directory on the ftp server. Do not save any pictures in SEM or EDS computers. Any data saved on the SEM or EDS computers will be deleted. You can reach your saved data from The user is responsible for cleaning the sample holders and sample stages after every usage. All problems and concerns must be reported to the lab supervisor immediately. Please see the contact list for phone and information.
4 Three Options to Move the Stage 1) First Option Joystick 2) Second Option Mouse 3) Third Option Stage Tab
5 Standard Operating Procedure for JSM-6060 LV SEM -SECONDARY ELECTRON IMAGING Specimen Preparation 1. Start with wearing gloves. 2. Choose one of the main sample holders and proper inset(s) for this holder. 3. According to your specimen conditions (irregular shape, nonconductive, powder, oily and dirty specimen, etc.) you may want to use copper or double side sticky carbon tapes, liquid carbon paint, carbon paste, gold or carbon sputtering, ultrasonic cleaning methods. If you are not sure which method fits best for your specimen, contact your lab supervisor (see contacts). 4. Please prepare your specimen in the specimen preparation area (Fig.1). Under no conditions are you to remove any specimen preparation tools from the SEM laboratory. 5. If your specimen has an irregular shape, draw a schematic of its cross-section and determine the highest position of the specimen. You can put a mark (like a small portion of carbon tape) at the highest position. This will help you while you are changing the working distance.
6 Fig. 1 Specimen preparation area and drawers showing preparation tools. Login to Computer 1. Start by filling out the log book. 2. Log in to the computer (ctrl + alt + del). If this will not allow you access to the SEM computer please first hit ctrl + alt Type your user name and password; do not select dial-up connections. 4. Double click on JSM 6060 icon and login using your ISU account. You should type a correct account number. 5. Wait until you see the SEM console on the screen (This takes app. 1 minute). Loading the Specimen 1. Change your gloves. Your gloves, specimen holder and specimen should be clean and dry before you load your specimen.
7 Figure 2. Z-height adjustment knob (at lowest position). 2. Check the Z-direction (Fig.2). Make sure that the stage Z is set to its minimum position so your specimen won t hit the lens or BEI detector. If the Z was not left in this condition, please note in the logbook. 3. Press and hold the VENT button (Fig.3) for 2 seconds until the button lights. Fig. 3 VENT and EVAC buttons.
8 4. Wait until VENT light stops blinking. 5. Pull the chamber towards you. 6. Insert the specimen holder onto the stage dovetail. (The flat side of the holder should be aligned with the dovetail) (Fig.4). Fig. 4 Specimen holder in the chamber. (The flat on the specimen holder goes in first) 7. Check the distance between your specimen surface and the lens (or BEI detector). Your specimen should not touch any of them. 8. Slowly close the chamber drawer. 9. Hold the door while activating the pre-pump sequence by pressing EVAC button (Fig 3). Hold the button for 2 sec until the button lights and the pump starts. 10. Wait until the HT READY indication appears in the HT icon on the screen. Typical Operation 1. If you previously created a username, click on the user tab and chose your own username and click OK. (If not you can use the general user tab.) 2. Check the scan rate, signal, magnification, and spot size before you hit the HT
9 button. You can select a fast scan (Scan 2), SEI signal, low magnification (X30 or lower), and spot size for starting. 3. Hit the HT Ready button. 4. If the screen is all dark or white, first try the ACB (Auto Contrast and Brightness) button to set a proper level of contrast and brightness. 5. Use coarse/fine focus to focus your sample. 6. If you are authorized to change the Z stage level, find the highest position of your specimen using your drawing and/or marker. Focus on this spot before changing Z. Be mindful of the actual distance between this spot and the lens at all times. If you are not authorized to change Z-stage level, consult the instructor. 7. Take a picture on Scan 2 or Scan 3 by clicking the FREEZE button. (Scan 4 automatically freezes). Use the SAVE button to save your picture in your folder. If Necessary 1. If the image shows directionality of focus (Fig. 5), you should correct astigmatism. This is only apparent if you are at a high magnification (> 3000x). Fig. 5 (a) Astigmatism corrected, b) Directionality seen by going over focus.
10 2. Using the FOCUS control (mouse or knob), go over and under focus while looking for directionality in the image at high magnification (> 3000x). 3. Obtain the best focus. (The image may not be sharp but has no directionality). 4. Adjust the STIGX and STIGY controls (mouse or knobs), as if they were focusing controls. Try to obtain a sharp image. 5. Check the focus and repeat steps 2-5 if necessary. If Necessary 1. If the image shifts while focusing (Fig. 6), you should align the Objective Aperture. Fig. 6 (a) in focus image with Objective Aperture is aligned well, (b) over focus image with bad Objective Aperture alignment showing a slight image shift.
11 2. Set spot size to ~30 and center a feature with good contrast on the screen. 3. Increase your magnification to ~10,000X. 4. Adjust Contrast and Brightness, and Focus your image. 5. Turn on the OL Wobbler (Under Tools menu). 6. If the image moves back and forth, adjust OL aperture knobs (Fig.7) to minimize image shift. These knobs are sensitive. Only small adjustments are necessary. Fig. 7 OL Wobbler knobs 7. Turn off the OL Wobbler. Shut Down 1. Check the scan rate, signal, magnification, and spot size before you hit the HT button. You should select Scan 2, SEI signal, low magnification (X30 or lower), and spot size.
12 2. If you changed the Z-direction stage, find the highest position of your specimen. (Use your drawing and/or marker to determine it properly). Focus on this spot and carefully decrease the Z-stage to its minimum position (Fig.2). 3. Press HT button to turn off high tension. 4. Hit EXIT to close the software. 5. Wear your gloves. 6. Press and hold the VENT button (Fig.3) for about 2 seconds until it lights. 7. Pull the chamber towards you. Check the distance between your specimen surface and the lens (or BEI detector). Your specimen should not touch any of them. 8. Slowly close the chamber drawer. 9. Hold the door while activating the pre-pump sequence by pressing the EVAC button (Fig 3) until it lights and begins blinking. 10. Log out from the computer. 11. Finish writing your logbook entries. 12. DO NOT LEAVE the lab until EVAC light stays lit. Specimen chamber should be kept under vacuum. 13. Clean up your mess and put any specimen preparation tools in order.
13 Low Vacuum Mode 1. Set your working distance to 8 mm 12 mm for imaging. If you are not authorized to change Z-stage level, consult the supervisor. Be mindful of the actual distance between this spot and the lens at all times. 2. Press LV button. Low Vacuum Control tab (Fig.11) will appear on the screen. Fig.10 Vacuum Control Buttons 3. Click on the target pressure you want to reach. (133 Pascal = 1 Torr) (Fig.11) Fig.11 Low Vacuum Control 4. In low vacuum mode image will only be BEI. Use SHADOW mode to emphasize topography.
14 If charging occurs: 5. (a) Adjust air to determine the optimum pressure when charging stops. Too much air may result in too little signal. (b) Adjust the probe current (spot size). Too low a probe current may result in too little a signal (c) Adjust your accelerating voltage. Too low a kv may result in too little a signal. Fig.12 JEOL digital control panel at LV mode Shut Down 1. Switch your vacuum back to High Vacuum.. 2. Check the scan rate, magnification, and spot size before you hit the HT button. You should select Scan 2, low magnification (X30 or lower), and spot size. 3. Follow normal sample removal procedure (See Secondary Electron Imaging Mode).
JEOL 6500 User Manual
LOG IN to your session on the computer to the left of the microscope. Starting Conditions 1. Press Ctrl-Alt-Del and log on to the microscope computer. Click on JEOL PC SEM 6500 icon. Click yes if message
More informationJEOL 6700 User Manual 05/18/2009
JEOL 6700 User Manual 05/18/2009 LOG IN to your session on the computer to the right of the microscope. Starting Conditions 1. Click the button and read the Penning Gauge to ensure that the microscope
More informationUsing the Hitachi 3400-N VP-SEM
Using the Hitachi 3400-N VP-SEM Opening the Chamber to Load Specimens (This may also be done later using the software) 1. Click the AIR button on the front of the machine: 2. Wait a few minutes until you
More informationTitle: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016
Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Amray 1830 SEM. All users are expected to have read and understood this document.
More informationSEM OPERATION IN LOW VACUUM MODE
SEM OPERATION IN LOW VACUUM MODE Instructions for JEOL 5800 LV The EVAC light of the SEM specimen chamber should be already lit when you approach the SEM & the SEM will have been left in the high vacuum
More informationBasic Users Manual for Tecnai-F20 TEM
Basic Users Manual for Tecnai-F20 TEM NB: This document contains my personal notes on the operating procedure of the Tecnai F20 and may be used as a rough guide for those new to the microscope. It may
More informationOperating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.
Smith College August 2005 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check, 1 Specimen Insertion, 1 Startup, 2 Filament Saturation, 2 Beam Alignment,
More informationOperating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.
Smith College August 2009 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check 1 Startup 1 Specimen Insertion 2 Filament Saturation 2 Beam Alignment
More informationDickinson College Department of Geology
Dickinson College Department of Geology Title: Equipment: BASIC OPERATION OF THE SCANNING ELECTRON MICROSCOPE (SEM) JEOL JSM-5900 SCANNING ELECTRON MICROSCOPE Revision: 2.2 Effective Date: 1/29/2003 Author(s):
More informationStandard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014
Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 1. Utility Requirements a. System power is supplied by two 120 VAC/20 A circuits. When doing maintenance
More informationFEI Quanta 200 ESEM Basic instructions
FEI Quanta 200 ESEM Basic instructions Desktop and then start the UI. If the computer has restarted and you need to login, Username: supervisor and Password: supervisor Log-in to the Microscope using the
More informationCheck that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course)
JEOL 7000F BASIC OPERATING INSTRUCTIONS-Ver.-2.0 Note: This is minimal operation checklist and does not replace the other reference manuals. Read the manual for Specimen Exchange (JEOL 7000 Specimen Exchange
More information1.1. Log on to the TUMI system (you cannot proceed further until this is done).
FEI DB235 SEM mode operation Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Sample loading 1.1. Log on to the TUMI system (you cannot proceed further until this is done). 1.2. The FIB software (xp)
More informationSEM Training Notebook
SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside December 21, 2017 (rev. 3.4) 1 Before you begin Complete
More informationBasic Operating Instructions for Strata Dual Beam 235 FIB/SEM
Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Warning Always adjust your specimen height before closing the chamber door to make sure your specimen will not hit the bottom of the lens;
More informationSEM Training Notebook
SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside March 8, 2018 (rev. 3.5) 1 Before you begin Complete
More informationScanning Electron Microscope FEI INSPECT F50. Step by step operation manual
Scanning Electron Microscope FEI INSPECT F50 Step by step operation manual Scanning Electron Microscope, FEI Inspect F50 FE-SEM-F Observation Flow Saving Data And Analysis Specimen preparation Error check
More informationGeneral information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument.
JEOL JSM-7400F Table of contents General information.. 3 The operation panel. 4 The different sample holders and inserting the samples.. 5 Turning on the beam... 6 Stage map control... 8 Correcting astigmatism...
More informationOperating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah
Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Follow the procedures below when you use the Hitachi 7100 TEM. Starting Session 1. Turn on the cold
More informationModel SU3500 Scanning Electron Microscope
Model SU3500 Scanning Electron Microscope Modified and Parts taken from Hitachi Easy Operation Guide. Before using the Model SU3500 SEM, be sure to read the [GENERAL SAFETY GUIDELINES] in the instruction
More informationFE-SEM SU-8020 Operating manual (Preliminary version)
FE-SEM SU-8020 Operating manual (Preliminary version) 2016/04/11 Seimitsu Bunseki sitsu lab. Starting up 1.Turn on the Display switch. Windows OS is starting up 2. Select the user SU-8000. 3. Click the
More information1. Preliminary sample preparation
FEI Helios NanoLab 600 standard operating procedure Nicholas G. Rudawski ngr@ufl.edu (352) 392 3077 (office) (805) 252-4916 (cell) Last updated: 03/02/18 What this document provides: an overview of basic
More informationJEOL JEM-1400 Transmission Electron Microscope Operating Instructions
JEOL JEM-1400 Transmission Electron Microscope Operating Instructions Anti-contamination device Objective aperture Objective aperture translation knobs Specimen holder Pump/air switch Left hand control
More informationUser Operation of JEOL 1200 EX II
**Log onto Computer** Open item program Start Up Procedure User Operation of JEOL 1200 EX II 1. If scope is not running, locate an electron microscopy technician (EMT) to find out why not. 2. Turn up brightness
More information2. Raise HT to 200kVby following the procedure explained in 1.6.
JEOL 2100 MANUAL Quick check list 1. If needed, fill the reservoir with LN2 2. Raise HT to 200kVby following the procedure explained in 1.6. 3. Insert specimen holder into TEM (Insert holder in airlock,
More informationOPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge
OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE by Doug Bray Department of Biological Sciences University of Lethbridge Revised September, 2000 Note: The terms in bold in this document represent
More information1. Specimen Holder Removal, Loading, and Insertion
OPERATION OF THE PHILIPS CM-200 FEG-TEM When not in use, the CM-200 should be in the MICROSCOPE ON configuration with the HIGH TENSION ON (illuminates green when the high tension is on).. The microscope
More informationZEISS EVO SOP. May 2017 ELECTRON OPTICS
ZEISS EVO SOP May 2017 ELECTRON OPTICS The patented EVO column is the area of the SEM, where electrons are emitted, accelerated, deflected, focused, and scanned. Main characteristics of the EVO optics
More informationFEI Tecnai G 2 F20 Operating Procedures
FEI Tecnai G 2 F20 Operating Procedures 1. Startup (1) Sign-up in the microscope log-sheet. Please ensure you have written an account number for billing. (2) Log in to the computer: Login to your account
More informationFull-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right
Quick Guide to Operating FEI Titan Themis G2 200 (S)TEM: TEM mode Susheng Tan Nanoscale Fabrication and Characterization Facility, University of Pittsburgh Office: M104/B01 Benedum Hall, 412-383-5978,
More informationInstructions for Tecnai a brief start up manual
Instructions for Tecnai a brief start up manual Version 3.0, 8.12.2015 Manual of Tecnai 12 transmission electron microscope located at Aalto University's Nanomicroscopy Center. More information of Nanomicroscopy
More informationMCR Scanning Electron Microscopy Laboratory Portfolio
SUNY College of Environmental Science and Forestry Digital Commons @ ESF N.C. Brown Center for Ultrastructure Studies Fall 2016 MCR 484 - Scanning Electron Microscopy Laboratory Portfolio Timothy Gervascio
More informationRAITH e-line OPERATING INSTRUCTIONS
RAITH e-line OPERATING INSTRUCTIONS 1) LOADING A SAMPLE a. Start the system i. On the Column PC (Right side monitor [R]), select the SmartSEM icon to on the desktop to begin the column software. ii. On
More informationThis procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface.
Tescan MIRA3 SEM: EDS using EDAX TEAM Nicholas G. Rudawski ngr@ufl.edu Cell: (805) 252-4916 Office: (352) 392-3077 Last updated: 12/04/17 This procedure assumes the user is already familiar with basic
More informationFigure 1 The Raith 150 TWO
RAITH 150 TWO SOP Figure 1 The Raith 150 TWO LOCATION: Raith 150 TWO room, Lithography area, NanoFab PRIMARY TRAINER: SECONDARY TRAINER: 1. OVERVIEW The Raith 150 TWO is an ultra high resolution, low voltage
More informationScanning electron microscope
Scanning electron microscope 6 th CEMM workshop Maja Koblar, Sc. Eng. Physics Outline The basic principle? What is an electron? Parts of the SEM Electron gun Electromagnetic lenses Apertures Chamber and
More informationCOMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual
1 COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual ABOUT THIS MICROSCOPE (room HG01.240) The JEM-1400Flash
More informationThe user should already be familiar with operation of the instrument in STEM mode, use of the Microscope Control interface, and TIA.
FEI Tecnai F20 S/TEM: EDS system operation Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 01/22/18 The user should already be familiar with operation of the instrument in
More informationProtective Equipment Nitrile gloves for handling sample holder and safety glasses for filling liquid nitrogen dewar.
Emergency Information: 1. Medical Emergencies: Contact 911 and McGill Security 514.398.3000 2. Leave TEM as is. Do NOT shut down the vacuum system. 3. If possible, turn off High Tension and Close Column
More information1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).
JEOL 200CX operating procedure Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Specimen loading 1.1. Unlock the TUMI system. 1.2. Load specimen(s) into the holder. If using the double tilt holder, ensure
More informationMSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I
MSE 595T Basic Transmission Electron Microscopy TEM Imaging - I Purpose The purpose of this lab is to: 1. Make fine adjustments to the microscope alignment 2. Obtain a diffraction pattern 3. Obtain an
More informationProcedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope
Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope The procedures given below were written specifically for the FEI Tecnai G 2 Sphera microscope. Modifications will need to
More informationSOP: EDAX Eagle III Microspot XRF
SOP: EDAX Eagle III Microspot XRF Page 1 of 6 SOP: EDAX Eagle III Microspot XRF 1. Scope 1.1 This document describes the standard operating procedure (SOP) for the EDAX Eagle III Microspot XRF. This X-ray
More informationJEOL 5DII. Operation introduction. By Serge Charlebois
JEOL 5DII Operation introduction By Serge Charlebois July 2003 General procedure Loading the cassette in the load lock Selecting EOS mode, table and aperture Setting and maximising current Observation
More information1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).
FEI Tecnai F20 S/TEM: imaging in TEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 01/21/18 1. Filling the cold trap (if needed) 1.1. Prior to use, the cold trap needs
More informationSTANDARD OPERATING PROCEDURE: JEOL TEM-2100
STANDARD OPERATING PROCEDURE: JEOL TEM-2100 Purpose of this Instrument: Essential tool for structural characterization of natural or synthesized nanostructures. Location: WVU - Engineering Sciences Building
More informationPlease follow these instructions for use of the Philips CM100 TEM. Adopted from website below.
Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. http://staff.washington.edu/wpchan/if/cm100_inst.shtml Instructions for the Philips CM100 TEM and peripherals
More informationZeiss AxioImager.Z2 Brightfield Protocol
Zeiss AxioImager.Z2 Brightfield Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge
More informationJeol JEM Responsible personell: Endy ( ) Online booking is compulsory!
Jeol JEM 1230 Responsible personell: Endy (45279377) Online booking is compulsory! After training you will have access to working alone on the instrument. All insertion of samples is done by responsible
More informationScanning Electron Microscopy Laboratory Portfolio
SUNY College of Environmental Science and Forestry Digital Commons @ ESF N.C. Brown Center for Ultrastructure Studies Fall 2016 Scanning Electron Microscopy Laboratory Portfolio Kensey Portman SUNY College
More informationScanning electron microscope
Scanning electron microscope 5 th CEMM workshop Maja Koblar, Sc. Eng. Physics Outline The basic principle? What is an electron? Parts of the SEM Electron gun Electromagnetic lenses Apertures Detectors
More informationStandard Operating Procedure Hitachi UHR CFE SU8230 SEM
Standard Operating Procedure Hitachi UHR CFE SU8230 Yale West Campus Materials Characterization Core ywcmatsci.yale.edu ESC II, Room E119F 810 West Campus Drive West Haven, CT 06516 Version 1.1, May 2016
More informationOperation Guide. Hitachi S-3400N. Variable Pressure Scanning Electron Microscope. with. Deben Peltier Coolstage
Operation Guide Hitachi S-3400N Variable Pressure Scanning Electron Microscope with Deben Peltier Coolstage www.deben.co.uk www.taltos.stanford.edu www.hitachi-hta.com Index Main Unit 3 Electron Optical
More informationSTEM alignment procedures
STEM alignment procedures Step 1. ASID alignment mode 1. Write down STD for TEM, and then open the ASID control window from dialogue. Also, start Simple imager viewer program on the Desktop. 2. Click on
More informationAmray 3600 FESEM. Standard Operating Procedure. v2.2 modified by Bryan Cord. General Notes...3. Sample Loading...5. System Loading...
Amray 3600 FESEM Standard Operating Procedure v2.2 modified 5.13.13 by Bryan Cord Contents General Notes...3 Sample Loading...5 System Loading...8 Imaging...12 Saving Data...16 System Unloading...18 Troubleshooting...20
More information2 How to operate the microscope/obtain an image
Morgagni Operating Instructions 50079 010912 2-1 2 ow to operate the microscope/obtain an image 2.1 Starting the microscope 2.1.1 Starting the microscope with several manually-operated steps 1. Turn on
More informationOperating Procedures for MICROCT1 Nikon XTH 225 ST
Operating Procedures for MICROCT1 Nikon XTH 225 ST Ensuring System is Ready (go through to ensure all windows and tasks below have been completed either by you or someone else prior to mounting and scanning
More informationMSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope
MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope Last updated on 1/8/2018 Jinsong Wu, jinsong-wu@northwestern.edu Aims: The aim of this lab is to familiarize you with basic TEM alignment
More informationThis document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface.
FEI Tecnai F20 S/TEM: imaging in STEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 05/10/18 This document assumes the user is already familiar with basic operation
More information1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height.
JEOL 2010F operating procedure Covers operation in STEM mode (See separate procedures for operation in TEM mode and operation of EDS system) Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 NOTE: this operating
More informationScanning Electron Microscopy Laboratory Portfolio
SUNY College of Environmental Science and Forestry Digital Commons @ ESF N.C. Brown Center for Ultrastructure Studies Fall 2016 Scanning Electron Microscopy Laboratory Portfolio Marissa Lanzatella SUNY
More informationFire CR Calibration Guide
1 Fire CR Calibration Guide This reference guide will guide you through the steps to complete the calibration for the Fire CR.. Getting Started: 1. Click on the Opal Icon on the Desktop. Figure 1 2. Once
More informationStandard Operating Procedure Hitachi UHR CFE SU8230 SEM
Standard Operating Procedure Hitachi UHR CFE SU8230 Yale West Campus Materials Characterization Core ywcmatsci.yale.edu ESC II, Room E119F 810 West Campus Drive West Haven, CT 06516 Version 1.5, February
More informationLEICA TCS SP5 AOBS TANDEM USER MANUAL
LEICA TCS SP5 AOBS TANDEM USER MANUAL STARTING THE SYSTEM...2 THE LAS AF SOFTWARE...3 THE «ACQUIRE» MENU...5 CHOOSE AND CREATE A SETTING...6 THE CONTROL PANEL...8 THE DMI6000B MICROSCOPE...10 ACQUIRE ONE
More informationBrightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer
Brightfield Microscopy and Image Acquisition on Spotcam1 by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer Log onto the computer. Enter your username and password to log onto the server.
More informationOPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop
OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop 1. Log in usage using the SMIF web site 2. Turn power
More informationDevelopment of JEM-2800 High Throughput Electron Microscope
Development of JEM-2800 High Throughput Electron Microscope Mitsuhide Matsushita, Shuji Kawai, Takeshi Iwama, Katsuhiro Tanaka, Toshiko Kuba and Noriaki Endo EM Business Unit, JEOL Ltd. Electron Optics
More informationZeiss LSM 880 Protocol
Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.
More informationZeiss LSM 780 Protocol
Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.
More informationSOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only
SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only Version 1.0 Prepared by D. Turnbull February 21, 2007. Please submit any omissions to the Author Note: This SEM is a recent
More informationNikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol
Nikon Eclipse Ti-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol Please sign on the log sheet before switching on system. Turn on system Turn on A only if confocal mode or laser
More informationLEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center
LEO 912 TEM Short Manual Prepared/copyrighted by RH Berg Danforth Plant Science Center Specimen holder [1] Never touch the holder (outside of the O-ring, double-headed arrow) because finger oils will contaminate
More informationOlympus LEXT OLS 4000 Confocal Laser Microscope
Olympus LEXT OLS 4000 Confocal Laser Microscope The Olympus LEXT OLS4000 is a confocal microscope capable of taking high-resolution 3D images. The magnification (Optical and Digital) of this microscope
More informationUniversity of Minnesota Nano Fabrication Center Standard Operating Procedure Equipment Name:
Equipment Name: Coral Name: Nanoimprinter Revision Number: 1.1 Model: NX-B200 Revisionist: M. Fisher Location: Bay 4 Date: 2/12/2010 1 Description Nanonex NX-B200 nanoimprinter is another method of transfer
More informationJEOL 5800LV SCANNING ELECTRON MICROSCOPE OPERATOR'S MANUAL
JEOL 5800LV SCANNING ELECTRON MICROSCOPE OPERATOR'S MANUAL Michael N. Spilde Christopher Adcock Original version: October 1999 Revised: January 2006 Department of Earth and Planetary Sciences and Institute
More informationJEOL 2010 FasTEM & DigitalMicrograph User's Guide
JEOL 2010 FasTEM & DigitalMicrograph User's Guide Electron Microscopy Laboratory Instititute of Materials Science University of Connecticut The purpose of this manual is to remind you of the essential
More information05/20/14 1. Philips CM200T. Standby Condition
05/20/14 1 Philips CM200T Standby Condition HT and filament off, HT setting at 200kV. RESET HOLDER, center sample tilt knobs, and remove sample. Mag ~ 5-10kX Objective and SA apertures out, C2 aperture
More informationLeica DMi8A Quick Guide
Leica DMi8A Quick Guide 1 Optical Microscope Quick Start Guide The following instructions are provided as a Quick Start Guide for powering up, running measurements, and shutting down Leica s DMi8A Inverted
More informationHoriba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer
Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and
More informationBruker Dimension Icon AFM Quick User s Guide
Bruker Dimension Icon AFM Quick User s Guide March 3, 2015 GLA Contacts Jingjing Jiang (jjiang2@caltech.edu 626-616-6357) Xinghao Zhou (xzzhou@caltech.edu 626-375-0855) Bruker Tech Support (AFMSupport@bruker-nano.com
More informationRENISHAW INVIA RAMAN SPECTROMETER
STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw invia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples
More informationCAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE
University of Virginia ITC Academic Computing Health Sciences CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE Introduction The Olympus BH-2 microscope in ACHS s microscope lab has objectives from
More informationScanning Electron Microscopy Student Image Portfolio
SUNY College of Environmental Science and Forestry Digital Commons @ ESF N.C. Brown Center for Ultrastructure Studies Fall 12-7-2016 Scanning Electron Microscopy Student Image Portfolio Matthew DaRin SUNY
More informationStandard Operating Procedure
Standard Operating Procedure Nanosurf Atomic Force Microscopy Operation Facility NCCRD Nanotechnology Center for Collaborative Research and Development Department of Chemistry and Engineering Physics The
More informationPANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: )
University of Minnesota College of Science and Engineering Characterization Facility PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: 2012.10.17) The following instructions
More informationNanoscale Fabrication & Characterization Facility. Raith e-line EBL Users Guide (updated:aug 2 nd, 2017)
Nanoscale Fabrication & Characterization Facility Raith e-line EBL Users Guide (updated:aug 2 nd, 2017) Overview: The Raith e-line EBL system is designed to write features with critical dimensions as small
More informationStandard Instructions for the Bruker D8 Advance Diffractometer, EPFL Valais Bragg Brentano and GID (Reflection)
Standard Instructions for the Bruker D8 Advance Diffractometer, EPFL Valais Bragg Brentano and GID (Reflection) For any questions regarding the X-ray facility, contact: Pascal Schouwink pascal.schouwink@epfl.ch
More informationSHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014
CIAN LSM1 or LSM2 short instructions, version 1.4, September 2014 page 1 of 6 SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 Before starting To work with LSM1
More informationSPECTRALIS Training Guide
SPECTRALIS Training Guide SPECTRALIS Diagram 1 SPECTRALIS Training Guide Table of Contents 1. Entering Patient Information & Aligning the Patient a. Start Up/Shut Down the System... 4 b. Examine a New
More informationSHORT GUIDE TO LASER MICRODISSECTION USING THE PALM COMBI SYSTEM
SHORT GUIDE TO LASER MICRODISSECTION USING THE PALM COMBI SYSTEM Turning ON the PALM DuoFlex Combi system 1. Turn on the three power point switches on the wall. From right to left: mercury lamp, microscope
More informationRenishaw InVia Raman microscope
Laser Spectroscopy Labs Renishaw InVia Raman microscope Operation instructions 1. Turn On the power switch, system power switch is located towards the back of the system on the right hand side. Wait ~10
More informationLSM 800 Confocal Microscope Standard Operation Protocol
LSM 800 Confocal Microscope Standard Operation Protocol Turning on the system 1. Switch on the Main switch (labeled 1 and 2 ) mounted on the wall. 2. Turn the Laser Key (labeled 3 ) 90 clockwise for power
More informationAtomic Force Microscopy (Bruker MultiMode Nanoscope IIIA)
Atomic Force Microscopy (Bruker MultiMode Nanoscope IIIA) This operating procedure intends to provide guidance for general measurements with the AFM. For more advanced measurements or measurements with
More informationTecnai T12 Operating Procedures
Tecnai T12 Operating Procedures I. Initial Procedures 1 II. Accelerating Voltage 3 III. Specimen Loading and Holder Insertion/Removal 3 IV. Emission Current 7 V. Alignment 7 VI. Camera Control and Imaging
More informationOlympus Digital Microscope Camera (DP70) checklist
Smith College - July 2005 Olympus Digital Microscope Camera (DP70) checklist CONTENT, page no. Camera Information, 1 Startup, 1 Retrieve an Image, 2 Microscope Setup, 2 Capture, 3 Preview. 3 Color Balans,
More informationBX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions
BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions Starting up: Schematic of Olympus BX-61. 1. Turn on Olympus microscope power box (left of microscope) with toggle switch on
More informationKulicke & Soffa 4700AD Instructions A.Atalar (April 2017)
Kulicke & Soffa 4700AD Instructions A.Atalar (April 2017) 4700AD model wire bonder can act like a wedge or ball-bonder. In the user manual refer to 4523AD for wedge bonding and 4524AD for ball-bonding.
More informationAxioVision 4.5 Brightfield Image Capture Procedure
AxioVision 4.5 Brightfield Image Capture Procedure 1. STARTING-UP PROCEDURE: Remove blue dust cover and place on shelf under microscope. Turn on the halogen lamp by pushing the switch at the back right
More informationScanning Electron Microscope. Instructions for Use
Scanning Electron Microscope Instructions for Use The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders are liable for damages.
More informationALTURA EDS. Rev. 0915
ALTURA EDS Rev. 0915 Enable the Oxford PC Enable the Altura-EDS under Dual Beam Tools in Coral. Or enter your NETID and password directly into the Oxford PC. Warning: Ion-milling, GISs/microprobe, and
More information