AxioVision 4.5 Brightfield Image Capture Procedure

Size: px
Start display at page:

Download "AxioVision 4.5 Brightfield Image Capture Procedure"

Transcription

1 AxioVision 4.5 Brightfield Image Capture Procedure 1. STARTING-UP PROCEDURE: Remove blue dust cover and place on shelf under microscope. Turn on the halogen lamp by pushing the switch at the back right of the microscope. Turn on the computer by pushing the button on the top front right side. Connect the camera computer cable to the back of the camera. Log on to the Cardiology network using your login name and password. 2. USING A CLEAN, DRY SLIDE: Clean your slide with 70% ethanol using Q-tips and Kimwipes if it has fingerprints or dust on it. Make sure AxioCam cable has been hooked up to camera for at least 15 minutes. Open the camera path. Select Position #5 on the filter cube turret (empty). 3. SETTING UP KOEHLER ILLUMINATION: Set the halogen lamp to 12 volts by turning up to the top value, "10, or push the black button beneath the voltage setting to achieve 3200 deg K (best). On the lower right side of microscope, set the neutral density filters: Set to 1.5. (% transmission) 0-6 Set to 6. (% transmission) 0-25 Set to 0. You may use 25, but it is not necessary. Diffuser: Push to the back to swing the diffuser in. Be sure that the blue swing-in filter under the stage is in place. Make sure the moveable lens under the specimen stage is swung in when using all objectives higher than 5X. Place a high contrast specimen on the stage (H&E is good), and focus with the chosen objective. Raise the condenser (under the stage) to the upper stop position. Close the luminous field diaphragm (the large round lens at the base of the microscope) by turning the outer ring until the diaphragm is visible in the field of view. Lower the condenser until the luminous field diaphragm becomes a 12-sided figure with clear edges in focus. 1

2 Open the luminous field diaphragm until it almost fills the field of view but the edges are still visible. Center the luminous field diaphragm by using the two metal centering screws on the condenser carrier. Open the luminous field diaphragm until the edges exactly disappear beyond the field of view. Set the aperture diaphragm: Remove one eyepiece from the binocular tube. While looking down the eyepiece opening, slide the knob on the front of the condenser until the aperture diaphragm fills 2/3 to 4/5 of the diameter of the objective exit pupils. Because each objective has a different field size and objective aperture, you will have to set the Koehler Illumination for each change of objective. 4. STARTING THE SOFTWARE: Start AxioVision Rel. 4.5 on the computer and activate the Live image. Select the Speed of the camera. The Slow Speed gives the best detail and the Fast is most pixilated. At the bottom of the Live-AxioCam HR screen, these display icons are shown after the Snap, Camera Speed, Exposure, White balance analysis icons: The first icon on the left is the Best Fit (usually the most vivid). The next one is the Min/Max. Next is Linear (full range of possible values). 4th icon is Gamma preset display for color closest to what you see when you look through the microscope. Use the Gamma pre-set display to perform the Shading Correction and the White Balance. You may change to other display options later. 5. SETTING THE SHADING CORRECTION: Activate the Properties window from the Live-AxioCam HR window menu, from Acquisition in the menu bar at the top of the screen or from the Workflow bar on the left of the screen Select Adjust tab and set the exposure time to 20 to 50 ms. Increase as needed. Select Frame, choose RGB (or B/W). Select Resolution, usually 1300x1030 scanned, from drop-down menu. 2

3 Select General, set the Gain factor to "1," Index to "0. Perform the Shading Correction first, each time you change objectives. Focus on a clean, clear area of the slide. If there is no clean area, you may remove the slide completely. On Live-AxioCam HR/General, click on the Shading Correction to smooth out the background. Some color may be seen at this point, but it is temporary. A message will appear if it is too dark or too bright. Increase or decrease exposure time under Live Properties/Adjust. Click on the Shading Correction box again. Repeat until the background is even with no spots or artifacts. 6. SETTING THE WHITE BALANCE: Focus on the specimen and select Adjust again. Now do a White Balance to attain the best color representation of the specimen by one of two methods: (a) Select Interactive and hold the crosshairs on a white area. This will give a readout of the values for red, green, and blue. Choose the best spot and release the mouse. (b) If there is no white area on the specimen at the region of interest, click on 3200 K first, wait about 20 sec. for color change, then click Automatic. You may also click on the White Balance icon at the bottom of the Live window. 7. ADJUSTING THE EXPOSURE: After the White Balance is done, increase the exposure as needed by increasing the exposure setting under the Adjust tab until the background is an appropriate shade of white for your specimen. You may also do an automatic exposure measurement by checking Automatic and then click Measure. Change the exposure time if necessary until the background is white enough for your application. Disable (uncheck) the automatic to use your own settings or the software will use automatic. You must repeat the SHADING CORRECTION and WHITE BALANCE every time you change objectives if you are capturing images, even if you are going back to one done previously. 3

4 8. FOCUSING ON THE SPECIMEN: While looking at the computer screen, adjust the fine focus knob slowly, then pause a couple seconds before continuing to adjust the focus. Right click on the live image to obtain the Spot meter/focus bar. When the green color comes to the red line, the software thinks best focus is achieved. Trust your eyes, not the focus bar. When the Spot meter is activated, a green frame appears around the live window. You may use this frame to enclose the region of interest (ROI) on which you want to focus by dragging the handles of the frame. 9. SELECTING THE OBJECTIVE AND RESOLUTION: On the Live-AxioCam HR window menu, select the Objective and Resolution from the dropdown menu. This will be saved with your image, and you will be able to add a scale bar by just selecting the scale bar icon. 10. CAPTURING THE IMAGE: Capture the image by using the "Snap" at the bottom of the Live- AxioCam HR image window, or by clicking any Camera Icon. You may do post-processing of your image before you save it, but the best practice is to save your image as captured as a.zvi. Then, after processing, save it with a new name or add "edited" to the name. Do not save images on the C Drive. Save on the D Drive or local network. 11. SAVING THE IMAGE: Save the image using File/Save As from the File menu. You may select from these four formats to store the images:.zvi This format allows you to store all data microscope, user, date, annotations and measurement together in a single file. You should save your original best image as a.zvi and export as a.tif. Keep an archival copy of both..tif With these external formats, AxioVision now stores all additional.jpg information separately from the actual image data. To do this, an.bmp XML file is created for every image file that contains other information such as additional data, annotations, and interactive measurement data. 4

5 12. USING THE SPLITTER SCREEN: Compare saved images side by side in the Splitter display: Select from the menu at the top of the screen View/Windows/ Splitter. All the opened images will be in the Gallery. In the Splitter display choose from two to eight screens. Click on Gallery. Select from the small gallery window at the top of each screen by clicking the image. Use the small red frame in the lower right corner of the Splitter display to change the size and region of the images in the Splitter Display. Right click to zoom up or back to change the size of the images. All the images in the Splitter display may be saved as a combined single image by clicking on "Create Image. 13. ADDING AND EDITING SCALE BARS: You may add a Scale Bar before or after the merged image is saved. If you selected the appropriate objective in the Live window dropdown menu, just select the scale bar icon and place it on the image. Otherwise, you must select from the top menu Measure/Scalings. To edit the Scalebar, go to the Attributes tab of Show Properties, available from the bottom menu of the image window, or from the workflow icons on the left of the screen. Highlight Scalebar. The line, color and font sizes can be edited here and saved as a default setting for all your scale bars. Undo is available from the Edit menu at the top of the screen. 14. EXPORTING IMAGES: Export by going to File/Export. An export window will appear. Export as a.tif image to a local folder, network folder, CD, or memory stick. Check "Burn in annotations" to save scalebars, etc. You will not be able to remove these annotations once they have been exported. Export as a batch by clicking on Batch, then click Add Files. Select files to be exported. The file names will appear in the window on the right. Click on Run Batch. As the files are exported, an "OK" message or an error message appears. Error messages are usually from incorrect syntax in the file name. Exported files should be saved to your folder in drive D or in a local network folder. 5

6 Set up the path before you export! 15. POST-PROCESSING YOUR IMAGES: Post-processing can be done on the saved images by selecting Show Properties and Display or by activating Processing from the menu at the top of the screen, then using its drop-down menu. The editing done with this dropdown menu takes some time to appear. There are several measuring tools available in this software that you may want to use. You may also measure your images with Image Pro 5.2 on the computer in WMB 306. You must get the USB dongle from Debby Martinson in WMB 303 or Lu Hilenski in WMB 329 to use the Image Pro software. 6

7 16. SHUTTING-DOWN PROCEDURE: Exit the AxioVision Rel. 4.5 from the File menu. Check to see if the next person is signed up to do fluorescence imaging. If so, then leave the lamp turned on and the camera plugged in. Log off the computer but do not turn the computer off. If you are the last person to use the system today, shut down the computer. Turn off the computer now a one step process. From the start menu, choose SHUTDOWN. The new monitor can be left on. Disconnect the camera computer cable. Swing a low power objective (5X or 10X) into the light path. Turn off the power supply for the mercury lamp by pushing the on/off switch. Put the blue dust cover back on the microscope, but not over the hot lamp. CLEAN UP THE AREA. You may log on the local Cardiology network and save directly to your own local shared folder. Please delete your files from the computer! 7

Olympus IX71 Microscope and DP71 Camera Instructions

Olympus IX71 Microscope and DP71 Camera Instructions Olympus IX71 Microscope and DP71 Camera Instructions Microscopy in Medicine (MiM) Core Emory University Department of Medicine 1 Olympus IX71 Image Capture Procedure 2 3 1. STARTING-UP PROCEDURE: Remove

More information

Zeiss Axioskop II. The AIF's "routine" light microscope. (Installed 8/24/04)AxioCam installed July 11th 2005

Zeiss Axioskop II. The AIF's routine light microscope. (Installed 8/24/04)AxioCam installed July 11th 2005 Zeiss Axioskop II The AIF's "routine" light microscope. (Installed 8/24/04)AxioCam installed July 11th 2005 Featuring: Phase Contrast Darkfield DIC/Nomarski Brightfield Fluorescent filters for Dapi, FITC,Rhodamine

More information

Simplified Instructions: Zeiss Brightfield Microscope S1000

Simplified Instructions: Zeiss Brightfield Microscope S1000 Contents General Microscope Set-Up Adjust Illumination Focus Condenser Open Software Image Capture Settings Shading & Color Corrections Image Capture & Viewing Scaling and Measurements Synopsis of Other

More information

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions Starting up: Schematic of Olympus BX-61. 1. Turn on Olympus microscope power box (left of microscope) with toggle switch on

More information

Zeiss Deconvolution Microscope: A Quick Guide

Zeiss Deconvolution Microscope: A Quick Guide Zeiss Deconvolution Microscope: A Quick Guide Start-up Uncover microscope. Do not put dust cover on the floor. Plug in both cameras. The default camera is the AxioCam HRm (monochrome camera) for fluorescence

More information

Characterization Microscope Nikon LV150

Characterization Microscope Nikon LV150 Characterization Microscope Nikon LV150 Figure 1: Microscope Nikon LV150 Introduction This upright optical microscope is designed for investigating up to 150 mm (6 inch) semiconductor wafers but can also

More information

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer Brightfield Microscopy and Image Acquisition on Spotcam1 by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer Log onto the computer. Enter your username and password to log onto the server.

More information

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE University of Virginia ITC Academic Computing Health Sciences CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE Introduction The Olympus BH-2 microscope in ACHS s microscope lab has objectives from

More information

Zeiss Axio Imager.A1 manual

Zeiss Axio Imager.A1 manual Zeiss Axio Imager.A1 manual Power-up protocol 1. Mercury lamp 2. Power strip on shelf 3. Computer The Mercury lamp should always be first-on and last-off. This prevents any electrical surges caused by

More information

Olympus Time-lapse Microscope Basic operation

Olympus Time-lapse Microscope Basic operation Olympus Time-lapse Microscope Basic operation To start up the microscope 1. Switch on the Olympus UCB. (label as ) 1 Power Switch 2 2. Switch on the MT10. (label as ) Power Switch Page 1 of 18 3 3. Switch

More information

Nikon E800 Microscope. Operating Instructions

Nikon E800 Microscope. Operating Instructions Nikon E800 Microscope Operating Instructions B Watson 12/2005 Table of contents: 1. The Nikon E800 Microscope 2. Turning the system ON and OFF 3. Selecting the light path 4. Operating in transmitted light

More information

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope Nikon Eclipse Ti A1-A Confocal Operating Manual Start-up 1. Turn on Excite Fluorescent light power supply- metal halide. a. Cool down as for mercury bulb b. Wheel closed liquid light guide 2. Turn on power

More information

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 CIAN LSM1 or LSM2 short instructions, version 1.4, September 2014 page 1 of 6 SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 Before starting To work with LSM1

More information

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Operations Guide Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Nikon Eclipse 90i Operations Guide STARTUP Startup Powering Up Fluorescence

More information

Using the Nikon TE2000 Inverted Microscope

Using the Nikon TE2000 Inverted Microscope Wellcome Trust Centre for Human Genetics Molecular Cytogenetics and Microscopy Core Using the Nikon TE2000 Inverted Microscope Fluorescence image acquisition using Scanalytic s IPLab software and the B&W

More information

Motorized Axio Observer Start-up instructions

Motorized Axio Observer Start-up instructions Start-up instructions 1. If using fluorescence turn on Fluorescent light source. TL light Source (Hal 100) 2. Turn on microscope using switch on lower left side of the microscope. 3. If imaging, turn on

More information

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope ZEN 2012 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 If you require fluorescence imaging,

More information

Zeiss AxioImager.Z2 Brightfield Protocol

Zeiss AxioImager.Z2 Brightfield Protocol Zeiss AxioImager.Z2 Brightfield Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge

More information

Nikon E800 Operating Instructions.

Nikon E800 Operating Instructions. Nikon E800 Operating Instructions. You can request electronic copies of this manual by contacting lshats@jhsph.edu Copies are also available on the JHU MMI Department web site. Please send your comments

More information

Olympus Digital Microscope Camera (DP70) checklist

Olympus Digital Microscope Camera (DP70) checklist Smith College - July 2005 Olympus Digital Microscope Camera (DP70) checklist CONTENT, page no. Camera Information, 1 Startup, 1 Retrieve an Image, 2 Microscope Setup, 2 Capture, 3 Preview. 3 Color Balans,

More information

Leica DB LB Research microscope and Studo Lite Imaging software

Leica DB LB Research microscope and Studo Lite Imaging software Leica DB LB Research microscope and Studo Lite Imaging software Room B523 User Guide Molecular Imaging Unit University of Helsinki www.miu.helsinki.fi 9.4.2008 1 GENERAL USER INFORMATION... 1 2 SETTINGS

More information

Zeiss LSM 880 Protocol

Zeiss LSM 880 Protocol Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Nikon E800 Operating Instructions.

Nikon E800 Operating Instructions. Nikon E800 Operating Instructions. You can request electronic copies of this manual by contacting imaging@fhcrc.org. Copies are also available on the Scientific Imaging web site. Please send your comments

More information

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000 The ideal K-12 science microscope solution User Guide for use with the Nova5000 NovaScope User Guide Information in this document is subject to change without notice. 2009 Fourier Systems Ltd. All rights

More information

QUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Imperial College London

QUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Imperial College London Imperial College London Facility for Imaging by Light Microscopy QUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Observing Life As It Happens

More information

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue

More information

EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY

EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY TURN ON THE FOLLOWING EQUIPMENT The fluorescent light (if needed) The power strip for the microscope and accessories The CoolSNAP HQ camera on the right (Turn

More information

INSTRUCTIONS FOR COURSE WORK 4 (AxioVert) Instructor: Anne Vaahtokari (MIU) 1. Purpose of the work

INSTRUCTIONS FOR COURSE WORK 4 (AxioVert) Instructor: Anne Vaahtokari (MIU) 1. Purpose of the work INSTRUCTIONS FOR COURSE WORK 4 (AxioVert) Instructor: Anne Vaahtokari (MIU) 1. Purpose of the work In this work, you will get familiar with an inverted epifluorescence microscope. Also, you will learn

More information

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue

More information

Takeoff Guide Tutorial to get started with the AxioVision Imaging System (Based on Release 4.8 June 2009)

Takeoff Guide Tutorial to get started with the AxioVision Imaging System (Based on Release 4.8 June 2009) Takeoff Guide Tutorial to get started with the AxioVision Imaging System (Based on Release 4.8 June 2009) This Takeoff Guide has been prepared for Carl Zeiss by Imaging Associates Ltd. www.imas.co.uk Copyright

More information

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software September 2007 Check website for most current User Guide

More information

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol Nikon Eclipse Ti-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol Please sign on the log sheet before switching on system. Turn on system Turn on A only if confocal mode or laser

More information

Zeiss Axioplan 2 imaging microscope and Axiovision software

Zeiss Axioplan 2 imaging microscope and Axiovision software Zeiss Axioplan 2 imaging microscope and Axiovision software Microscopes 1 and 2 in room B501b User Guide Molecular Imaging Unit University of Helsinki www.miu.helsinki.fi 20.5.2010 1 GENERAL... 1 1.1...

More information

Zeiss AxioObserver with ApoTome

Zeiss AxioObserver with ApoTome Zeiss AxioObserver with ApoTome Quick Start User Guide LSU Health Sciences Center-Shreveport Research Core Facility (RCF) Microscopy Table of Contents 1 Start up the system.. Page 3 2 Touch screen controller

More information

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 Leica DMI AF6000LX Table of contents START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 ACQUIRE MODULE 6 SETTING THE LIGHTPATH 6

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS Zeiss LSM 510 M eta Confocal M icroscope OPERATING INSTRUCTIONS Starting the System: 1. Turn the black knob on the laser box one-quarter turn from Off to On. You will hear the laser cooling mechanisms

More information

Things to check before start-up.

Things to check before start-up. Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your

More information

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5. Smith College August 2010 Operating Checklist for using the Laser Scanning Confocal Microscope Leica TCS SP5. CONTENT, page no. Startup, 1 Initial set-up, 1 Software, 2 Microscope Specimen observation

More information

Nikon TE300 Eclipse Wide-Field Microscope

Nikon TE300 Eclipse Wide-Field Microscope Nikon TE300 Eclipse Wide-Field Microscope User Guide LSU Health Science Center-Shreveport Research Core Facility 1 User manual for Nikon Elements software Equipment: Nikon TE300 Eclipse microscope Photometrics

More information

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 710 Confocal Microscope Standard Operation Protocol LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.

More information

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com User Manual Digital Compound Binocular LED Microscope Model MD82ES10 MicroscopeNet.com Table of Contents i. Caution... 1 ii. Care and Maintenance... 2 1. Components Illustration... 3 2. Installation...

More information

QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Imperial College London

QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Imperial College London Imperial College London Facility for Imaging by Light Microscopy QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Observing Life As It Happens Startup procedure...

More information

Digital Design and Communication Teaching (DiDACT) University of Sheffield Department of Landscape. Adobe Photoshop CS5 INTRODUCTION WORKSHOPS

Digital Design and Communication Teaching (DiDACT) University of Sheffield Department of Landscape. Adobe Photoshop CS5 INTRODUCTION WORKSHOPS Adobe INTRODUCTION WORKSHOPS WORKSHOP 1 - what is Photoshop + what does it do? Outcomes: What is Photoshop? Opening, importing and creating images. Basic knowledge of Photoshop tools. Examples of work.

More information

LSM 780 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch

More information

Using the AmScope Microscope Cameras

Using the AmScope Microscope Cameras Using the AmScope Microscope Cameras Part 1 Setup. In order to use the camera, you will need: a) the camera system; b) a computer running the camera software. The camera system is contained in a Pelican

More information

Cell Biology and Bioimaging Core

Cell Biology and Bioimaging Core Cell Biology and Bioimaging Core Leica TCS SP5 Operating Instructions Starting up the instrument 1. First, log in the log book located on the confocal desk. Include your name, your lab s PI, an account

More information

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL START-UP On the Switchbox, turn both black switches to the ON position. Wait for the microscope to boot up completely (watch the screen on the side of the microscope).

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, EL6000 fluorescent light source for the microscope stand. 2. Turn on the Scanner Power

More information

Photography Protocol. Software: ZEN Blue 2012

Photography Protocol. Software: ZEN Blue 2012 Photography Protocol Hardware: Zeiss Axio Imager.A2 or Imager.M2 microscope 10x eye piece 20x / 0.45 dry lens 40x / 0.75 dry lens 63x / 0.95 dry lens 100x / 1.4 oil lens AxioCam MRc camera Software: ZEN

More information

User Guide for TWAIN / DirectX interface for GRYPHAX USB 3.0 cameras

User Guide for TWAIN / DirectX interface for GRYPHAX USB 3.0 cameras User Guide for TWAIN / DirectX interface for GRYPHAX USB 3.0 cameras The TWAIN & DirectX driver for PROGRES GRYPHAX USB 3.0 cameras enables user to operate with TWAIN and DirectX supported 3 rd party software

More information

User manual for Nikon Elements software

User manual for Nikon Elements software User manual for Nikon Elements software Equipment: Nikon TE300 Eclipse microscope ANDOR Neo/Zyla B&W camera (default) DS Fi2 color camera Sign in on the sign in sheet; please use both your given name and

More information

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL MIF ZEISS LSM510 CONFOCAL USER PROTOCOL START-UP Turn on the Mercury Bulb Power Supply (if needed). Power-on the Control Box. Turn on the computer. Open the LSM 510 software. Choose Scan New Images and

More information

Olympus xcellence Software - basic user guide

Olympus xcellence Software - basic user guide Olympus xcellence Software - basic user guide This is a basic overview of setting up time lapse experiments using Olympus's xcellence software on BIU's IX81 inverted phase contrast system - the software

More information

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09) University of Chicago Integrated Light Microscopy Core Dr. Vytas Bindokas, Director http://digital.bsd.uchicago.edu By: Christine Labno, Assistant Director Room: AB-129 Phone: 4-9040 Zeiss Axiovert 135

More information

Instructions for Making On-Line Reservations for Microscopes in NB11-204

Instructions for Making On-Line Reservations for Microscopes in NB11-204 Instructions for Making On-Line Reservations for Microscopes in NB11-204 1. Log into Mail using Mail.swmed.edu 2. Log in using your university id and password. 3. Click the Calendar Tab at the top right

More information

LSM 800 Confocal Microscope Standard Operation Protocol

LSM 800 Confocal Microscope Standard Operation Protocol LSM 800 Confocal Microscope Standard Operation Protocol Turning on the system 1. Switch on the Main switch (labeled 1 and 2 ) mounted on the wall. 2. Turn the Laser Key (labeled 3 ) 90 clockwise for power

More information

COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual

COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual 1 COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual ABOUT THIS MICROSCOPE (room HG01.240) The JEM-1400Flash

More information

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands Differential Interference Contrast (DIC) imaging is a technique used to increase contrast in brightfield images. In confocal systems,

More information

Zeiss 880 Training Notes Zen 2.3

Zeiss 880 Training Notes Zen 2.3 Zeiss 880 Training Notes Zen 2.3 1 Turn on the HXP 120V Lamp 2 Turn on Main Power Switch Turn on the Systems PC Switch Turn on the Components Switch. 3 4 5 Turn on the PC and log into your account. Start

More information

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire Nikon Ti-E Microscope Manual Rightmire Hall Ohio State University Director: Tony Brown Rightmire 060 292-1205 brown.2302@osu.edu Facility Manager: Paula Monsma Rightmire 062 293-0939 292-1367 monsma.1@osu.edu

More information

Zeiss LSM 780 Protocol

Zeiss LSM 780 Protocol Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, Fluorescent Light for the microscope stand. 2. Turn on the Scanner Power (1) on the front

More information

Widefield 1. Switching on

Widefield 1. Switching on Widefield 1 Switching on 1. Ignite DG5 lamp - must be switched on first (if previous user has switched off, wait 30 min before igniting) 2. Wait 5s and then turn on the main DG5 controller switch. 3. DG5

More information

Everest System / Slidebook Operating Procedures

Everest System / Slidebook Operating Procedures Everest System / Slidebook Operating Procedures NOTICE: This guide is meant to supplement training, not replace it. All users must be trained first hand by a core employee. Training of others in your lab

More information

Optika ISview. Image acquisition and processing software. Instruction Manual

Optika ISview. Image acquisition and processing software. Instruction Manual Optika ISview Image acquisition and processing software Instruction Manual Key to the Instruction Manual IS is shortened name used for OptikaISview Square brackets are used to indicate items such as menu

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Version : 07/08/0 Leica SP8 TCS Users Manual Start up:. Turn the PC Microscope, Scanner Power, Laser Power, and the Laser Emission key to on (bottom right of desk).. Turn on the fluorescent lamp (top left

More information

USING LEICA AS LASER MICRODISSECTION (LMD6000) MICROSCOPE Written By Jungim Hur

USING LEICA AS LASER MICRODISSECTION (LMD6000) MICROSCOPE Written By Jungim Hur USING LEICA AS LASER MICRODISSECTION (LMD6000) MICROSCOPE Written By Jungim Hur Digital Video Camera Eyepieces Laser module Laser safety UV shield Specimen holder Smart move control LEICA CTR6500 electronics

More information

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes Turn on Main Switch, System PC and Components Switches 780 Start up sequence Do you need the argon laser (458, 488, 514 nm lines)? Yes Turn on the laser s main power switch and

More information

Microscopy from Carl Zeiss

Microscopy from Carl Zeiss Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path

More information

Simplified Instructions: Olympus Widefield Microscope S1230

Simplified Instructions: Olympus Widefield Microscope S1230 Contents General Microscope Operation Simple Image Capture Multi-Wavelength Capture Z-Series Timelapse Combining Capture Modes Synopsis of Other Functions Pages 2-23 24-40 41-47 48-56 57-59 60-68 69-83

More information

Operating Instructions for Zeiss LSM 510

Operating Instructions for Zeiss LSM 510 Operating Instructions for Zeiss LSM 510 Location: GNL 6.312q (BSL3) Questions? Contact: Maxim Ivannikov, maivanni@utmb.edu 1 Attend A Complementary Training Before Using The Microscope All future users

More information

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide Matyas Molnar, Biovis 2016 Starting the microscpe 1. Check the microscope if everything looks clean and normal. If not, report it in the logbook.

More information

LEICA TCS SP5 AOBS TANDEM USER MANUAL

LEICA TCS SP5 AOBS TANDEM USER MANUAL LEICA TCS SP5 AOBS TANDEM USER MANUAL STARTING THE SYSTEM...2 THE LAS AF SOFTWARE...3 THE «ACQUIRE» MENU...5 CHOOSE AND CREATE A SETTING...6 THE CONTROL PANEL...8 THE DMI6000B MICROSCOPE...10 ACQUIRE ONE

More information

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Training Guide for Leica SP8 Confocal/Multiphoton Microscope Training Guide for Leica SP8 Confocal/Multiphoton Microscope LAS AF v3.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON power switch for epifluorescence

More information

Guide to Confocal 5. Starting session

Guide to Confocal 5. Starting session Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope

More information

Introduction to 3D Printing. Activity 1: Design a keychain using computer-aided design software

Introduction to 3D Printing. Activity 1: Design a keychain using computer-aided design software Introduction to 3D Printing Activity 1: Design a keychain using computer-aided design software 1 In this activity we ll design a keychain name tag and learn the fundamentals of computer-aided design, the

More information

Image Pro Ultra. Tel:

Image Pro Ultra.  Tel: Image Pro Ultra www.ysctech.com info@ysctech.com Tel: 510.226.0889 Instructions for installing YSC VIC-USB and IPU For software and manual download, please go to below links. http://ysctech.com/support/ysc_imageproultra_20111010.zip

More information

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION Olympus VS120-L100 Slide Scanner Standard Operating Procedure Startup 1) Red power bar switch (behind monitor) 2) Computer 3) Login: UserVS120 account (no password) 4) Double click: WAIT FOR INITIALIZATION

More information

Guide to AmScope SZM Series Dissecting Scope for Sample Imaging with MU Series Camera

Guide to AmScope SZM Series Dissecting Scope for Sample Imaging with MU Series Camera Guide to AmScope SZM Series Dissecting Scope for Sample Imaging with MU Series Camera The following guide is an adaptation of the AmScope Manual for use with DNA barcoding samples. Please refer to the

More information

ISCapture User Guide. advanced CCD imaging. Opticstar

ISCapture User Guide. advanced CCD imaging. Opticstar advanced CCD imaging Opticstar I We always check the accuracy of the information in our promotional material. However, due to the continuous process of product development and improvement it is possible

More information

Apple Photos Quick Start Guide

Apple Photos Quick Start Guide Apple Photos Quick Start Guide Photos is Apple s replacement for iphoto. It is a photograph organizational tool that allows users to view and make basic changes to photos, create slideshows, albums, photo

More information

Brief manual how to start and close the Leica sp2 Confocal. (TCS SP2 AOBS system mounted on a DM IRE2)

Brief manual how to start and close the Leica sp2 Confocal. (TCS SP2 AOBS system mounted on a DM IRE2) Brief manual how to start and close the Leica sp2 Confocal (TCS SP2 AOBS system mounted on a DM IRE2) A. Switching on hardware B. Acquiring and saving images C. Switching off the microscope D. Good working

More information

Guide to Operating the Bruker FT-IR Microscope I. Basic Spectrum Acquisition with Vertex 70

Guide to Operating the Bruker FT-IR Microscope I. Basic Spectrum Acquisition with Vertex 70 Guide to Operating the Bruker FT-IR Microscope I. Basic Spectrum Acquisition with Vertex 70 Susheng Tan, Ph.D. NanoScale Fabrication and Characterization Facility, University of Pittsburgh SB60 Benedum

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Leica SP8 TCS Users Manual Follow the procedure for start up and log on as posted in the lab. Please log on with your account only and do not share your password with anyone. We track and confirm usage

More information

Standard Operating Procedure

Standard Operating Procedure CENTER FOR NANOSCALE SCIENCE AND ENGINEERING Standard Operating Procedure Microscope Software Brian Wajdyk Page 1 of 6 Important Images are not to be saved to the computer. They will be deleted without

More information

Leica DMi8A Quick Guide

Leica DMi8A Quick Guide Leica DMi8A Quick Guide 1 Optical Microscope Quick Start Guide The following instructions are provided as a Quick Start Guide for powering up, running measurements, and shutting down Leica s DMi8A Inverted

More information

AxioVision User's Guide. Release 4.1

AxioVision User's Guide. Release 4.1 AxioVision User's Guide Release 4.1 Number of this manual: B 48-0038 e 10.2003 Date of issue: 10.2003 Carl Zeiss Vision draws the User's attention to the fact that the information and references contained

More information

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands Differential Interference Contrast (DIC) imaging is a technique used to increase contrast in brightfield images. In confocal systems,

More information

Leica DMI 4000 tutorial

Leica DMI 4000 tutorial Leica DMI 4000 tutorial Before using the Leica DMI 4000, You will need to put down your name on the reservation system = 1 Welcome to the Leica DMI4000 Microscope tutorial How to start up the system (p.3)

More information

TRAINING MANUAL. Olympus FV1000

TRAINING MANUAL. Olympus FV1000 TRAINING MANUAL Olympus FV1000 September 2014 TABLE OF CONTENTS A. Start-Up Procedure... 1 B. Visual Observation under the Microscope... 1 C. Image Acquisition... 4 A brief Overview of the Settings...

More information

GlobiScope Analysis Software for the Globisens QX7 Digital Microscope. Quick Start Guide

GlobiScope Analysis Software for the Globisens QX7 Digital Microscope. Quick Start Guide GlobiScope Analysis Software for the Globisens QX7 Digital Microscope Quick Start Guide Contents GlobiScope Overview... 1 Overview of home screen... 2 General Settings... 2 Measurements... 3 Movie capture...

More information

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Follow the procedures below when you use the Hitachi 7100 TEM. Starting Session 1. Turn on the cold

More information

GXCapture 8.1 Instruction Manual

GXCapture 8.1 Instruction Manual GT Vision image acquisition, managing and processing software GXCapture 8.1 Instruction Manual Contents of the Instruction Manual GXC is the shortened name used for GXCapture Square brackets are used to

More information

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide Nikon C1si Spectral Laser Scanning Confocal Microscope User Guide Contents: C1Si Turn-On/ShutDown Procedures... 2 Overview... 4 Setup for epi-illumination to view through the eyepieces:... 5 Setup for

More information

Internal Medicine Imaging Core Emory University Department of Medicine

Internal Medicine Imaging Core Emory University Department of Medicine Internal Medicine Imaging Core Emory University Department of Medicine 1 OPERATION OF THE ZEISS LSM 510 META YOU MUST SIGN UP TO USE THE MICROSCOPE OR COMPUTER EVERY TIME NO EXCEPTIONS Before attempting

More information

Title: Leica SP5 Confocal User Manual

Title: Leica SP5 Confocal User Manual Title: Leica SP5 Confocal User Manual Date of first issue: 23/10/2015 Date of review: Version: Admin For assistance or to report an issue Office: CG07 or 05 Email: Igmm-imaginghelpdesk@igmm.ed.ac.uk Website:

More information

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s LSM 5 MP, LSM 510 and LSM 510 META M i c r o s c o p y f r o m C a r l Z e i s s Quick Guide Laser Scanning Microscopes LSM Software ZEN 2007 August 2007 We make it visible. Contents Page Contents... 1

More information

Progeny Imaging Veterinary

Progeny Imaging Veterinary Progeny Imaging Veterinary User Guide V1.14 and higher 00-02-1605 Rev. K1 ECN: ECO052875 Revision Date: 5/17/2017 Contents 1. About This Manual... 6 How to Use this Guide... 6 Text Conventions... 6 Getting

More information

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope The procedures given below were written specifically for the FEI Tecnai G 2 Sphera microscope. Modifications will need to

More information

CHAPTER1: QUICK START...3 CAMERA INSTALLATION... 3 SOFTWARE AND DRIVER INSTALLATION... 3 START TCAPTURE...4 TCAPTURE PARAMETER SETTINGS... 5 CHAPTER2:

CHAPTER1: QUICK START...3 CAMERA INSTALLATION... 3 SOFTWARE AND DRIVER INSTALLATION... 3 START TCAPTURE...4 TCAPTURE PARAMETER SETTINGS... 5 CHAPTER2: Image acquisition, managing and processing software TCapture Instruction Manual Key to the Instruction Manual TC is shortened name used for TCapture. Help Refer to [Help] >> [About TCapture] menu for software

More information