Simplified Instructions: Olympus Widefield Microscope S1230
|
|
- Rodger Reeves
- 5 years ago
- Views:
Transcription
1 Contents General Microscope Operation Simple Image Capture Multi-Wavelength Capture Z-Series Timelapse Combining Capture Modes Synopsis of Other Functions Pages Simplified Instructions: Olympus Widefield Microscope S1230 This is an Overview Only intended to refresh specific items learned in the Hands-On Training Use only in conjunction with your own much more detailed notes taken during your training
2 1, Reserve Time on Scheduler 2, Remove Dust Cover Please do NOT place on the table beside the camera. Use shelf or drawers.
3 3, Power-Up Locate Lamp Power Supply (to Left of System) Ensure Power to All Other Components is Off Before Turning Lamp On
4 3, Power-Up Turn on Power to Lamp (Depress Top of Red Button)
5 3, Power-Up Ignite Lamp (Lift Toggle to Manual, Release) please do not set toggle to Auto Auto decreases lamp life
6 3, Power-Up Locate Power Conditioner for Other Components on Upper Shelf
7 3, Power-Up Turn on Power to Other Components Please Do NOT Touch the Individual Power Switches For these Components
8 3, Power-Up Turn on Computer Computer is Under Monitor
9 4, Access Software Use Windows Log-In/Password provided to you at training Click on Meta Imaging Series icon (some may have a direct short cut to Metamorph on their Desktop)
10 5, Open Metamorph (double-click icon) (may be on desktop for some)
11 6, Viewing Sample Move Desired Objective into Position
12 6, Viewing Sample Indicate Objective in Metamorph Using pull-down menu (important for correct scale-bar after images collected)
13 6, Viewing Sample Clean Objective
14 6, Viewing Sample Direct Emissions to Ocular Viewing
15 6, Viewing Sample For Brightfield Viewing, Turn on halogen lamp
16 6, Viewing Sample For Brightfield Viewing, Turn on halogen lamp Filters on top Condensor
17 6, Viewing Sample For Brightfield Viewing, Slider changes intensity of halogen lamp
18 6, Viewing Sample For Fluorescence Viewing Turn off Halogen Lamp Indicate Filter Set using pull-down menu Note Mirror information ( m3 etc.)
19 6, Viewing Sample For Fluorescence Viewing Move Correct Mirror into Position (mechanical shutter must be open)
20 6, Viewing Sample For Fluorescence Viewing Toggle Shutter On/Off
21 6, Viewing Sample For Fluorescence Viewing Focus using indicated knob ONLY (Do NOT use knobs on microscope) Objective Up (toward slide) is clockwise
22 6, Viewing Sample For Fluorescence Viewing Toggle between Course (default), medium and fine focus
23 Remember to keep exposure to a minimum to prevent photobleaching i.e. keep shutter off as much as possible 6, Viewing Sample For Fluorescence Viewing Move Stage with joystick left toggle switch is for course to fine (default) stage movements
24 7, Simple Image Capture Direct Emissions to Camera ( Side Port, SP)
25 7, Simple Image Capture Under Acquire Pull-Down, Open Acquire Control (this control sets standard image collection parameters)
26 7, Simple Image Capture In Acquire Control, Display Set Exposure Time NEVER change anything except Items discussed below!!!
27 7, Simple Image Capture In Acquire Control, Display Set Binning (note, can set Live binning differently)
28 7, Simple Image Capture In Acquire Control, Display Set intensity where white is for viewing (preference is 4095) (can use autoscale also) NEVER change image gamma from 1
29 7, Simple Image Capture In Acquire Control, Special Set Gain Do NOT change anything else
30 7, Simple Image Capture In Acquire Control Click Show Live to Start Image Capture Keep on Full Chip, there is no reason to collect smaller size
31 7, Simple Image Capture Use continuously update Live image to focus your image on the screen (change exposure times accordingly) Stop Live toggle shuts down screen
32 7, Simple Image Capture Icons on Show Live (or Captured Image) Set Display Size, Color or Display Intensity
33 7, Simple Image Capture Click Acquire to Capture Image Exactly as Seen Under Show Live
34 8, Image Saving Under File : Save Save Your Captured Image to Your File on the Server
35 8, Image Saving To Save a Series with Same Base Name a, Open Setup Sequential File Names
36 8, Image Saving To Save a Series with Same Base Name b, Type In Base Name
37 8, Image Saving To Save a Series with Same Base Name c, Set Number of Digits in Name (image number is starting number) (output style example is shown)
38 8, Image Saving To Save a Series with Same Base Name d, Set Image Type Strongly recommend Metamorph TIFF for single images (grouped images- stacks-discussed later)
39 8, Image Saving To Save a Series with Same Base Name e, Select Server and Folder Location (can create a new Folder in location with New ) Click OK to finish Naming Set-up
40 8, Image Saving To Save a Series with Same Base Name f, Save Using Sequential File Name Saves files to designated location with Name 01, 02, 03, etc.
41 9, Multi-Image Capture Set Basic Image Capture Parameter Using the Acquire Box Open Acquire Multiple Wavelengths
42 9, Multi-Image Capture Set the Number of Different Wavelengths to Capture
43 9, Multi-Image Capture For Each Wavelength, Indicate Filter Set (Note: You MUST Choose Compatible Wavelengths that Use a Single Mirror)
44 9, Multi-Image Capture Manually Position the Correct Mirror Enter Exposure Times (Usually Determined Using Show Live) (Can Also Click Acquire to See Test Image for Each Wavelength)
45 9, Multi-Image Capture Click Top Acquire to Capture All Images
46 9, Multi-Image Capture Toggle Between Images in the Stack Using the Arrows. Other Image Display Consideration As Described under 7, Simple Image Capture
47 9, Multi-Image Capture Saving As Described under 7, Simple Image Capture Except That Save as a Metamorph Stack
48 10, Capture a Z-Series Open Dialog Under Acquire Again, Basic Parameters Set Under Acquire Drop-Down
49 10, Capture a Z-Series Set a, Number of Z-sections to collect
50 10, Capture a Z-Series Set b, Number of Steps between Each Z-section (positive steps move from bottom to top)
51 10, Capture a Z-Series Set c, to start where last focused i.e. Current Position Others are set positions on stage: Do NOT use those
52 10, Capture a Z-Series Set d, Planes x Spacing Do NOT Use Others
53 10, Capture a Z-Series Set e, to Collect as a Stack using f, the Desired Filter Set (Use Correct Mirror) F, Exposure Time Set Under Acquire
54 10, Capture a Z-Series g, OK Captures Z-series
55 10, Capture a Z-Series Toggle Between Images in the Stack Using the Arrows. Other Image Display Consideration As Described under 7, Simple Image Capture
56 10, Capture a Z-Series Saving As Described under 7, Simple Image Capture Except That Save as a Metamorph Stack
57 11, Capture a Timelapse Open Dialog Under Acquire Again, Basic Parameters Set Under Acquire Drop-Down
58 11, Capture a Timelapse a, Set Time Between Start of Each Collection (If Duration of Collection Longer than Time, It Is Not Permissible)
59 11, Capture a Timelapse b, Set Number of Timepoints to Collect c, Set Filters with Correct Mirror d, Select Exposure Time e, Click OK to Start Collection f, Save as a Stack
60 12, Combining Any of Multi-λ, Z or Time-Series Open Apps: Multi Dimensional Acq.
61 12, Combining Any of Multi-λ, Z or Time-Series a, Under Main, Check Types of Collections to Combine (Appropriate Interface Boxes Appear)
62 12, Combining Any of Multi-λ, Z or Time-Series This Dialog Contains Parameters Also Found Under Acquire (But Not Gain) (Recommend to Set Under Acquire And Let Auto-populate)
63 12, Combining Any of Multi-λ, Z or Time-Series b, Collections Under This Dialog Are Saved With the Name Specified in the Specified Folder
64 12, Combining Any of Multi-λ, Z or Time-Series c, Enter Number of Wavelengths and Exposure Time for Each Wavelength
65 12, Combining Any of Multi-λ, Z or Time-Series d, Enter Z-Series Settings (Multiple Stage Positions in Main Menu can be use to set multiple x,y positions)
66 12, Combining Any of Multi-λ, Z or Time-Series e, Enter Timelapse Settings (Minimum Time Interval Determined by λ, x-y and Z-Settings)
67 12, Combining Any of Multi-λ, Z or Time-Series f, Acquire All Settings
68 12, Combining Any of Multi-λ, Z or Time-Series g, Use Review Multi Dim. Data To See Results
69 There are a Wide Variety of Functions Available in Metamorph Not Detailed In This Short Synopsis Applications Under File Shown Here
70 There are a Wide Variety of Functions Available in Metamorph Not Detailed In This Short Synopsis Applications Under Edit Shown Here
71 There are a Wide Variety of Functions Available in Metamorph Not Detailed In This Short Synopsis Applications Under Regions Useful, But Not Shown Here
72 There are a Wide Variety of Functions Available in Metamorph Not Detailed In This Short Synopsis Applications Under Stack Shown Here
73 There are a Wide Variety of Functions Available in Metamorph Not Detailed In This Short Synopsis Applications Under Device Shown Here UNDER NOT CIRCUMSTANCE DOES ANYBODY TOUCH CONFIGURE ILLUMINATION OR MAGNIFICATION. DON T EVEN LOOK AT THOSE
74 There are a Wide Variety of Functions Available in Metamorph Not Detailed In This Short Synopsis Applications Under Device Shown Here Can Use Stage to Mark Stage Positions
75 There are a Wide Variety of Functions Available in Metamorph Not Detailed In This Short Synopsis Applications Under Display Shown Here Scale Image is Very Useful to Change Display Settings to a Consistent Level
76 There are a Wide Variety of Functions Available in Metamorph Not Detailed In This Short Synopsis Applications Under Display Shown Here Overlay Image Combine Stacks Into an Overlaid Image
77 There are a Wide Variety of Functions Available in Metamorph Not Detailed In This Short Synopsis Applications Under Process Shown Here Image Manipulation, Such As Background Subtraction and Much More
78 There are a Wide Variety of Functions Available in Metamorph Not Detailed In This Short Synopsis Applications Under Log Shown Here Log is Used to Record Data Into An Excel Sheet
79 There are a Wide Variety of Functions Available in Metamorph Not Detailed In This Short Synopsis Applications Under Measure Shown Here
80 There are a Wide Variety of Functions Available in Metamorph Not Detailed In This Short Synopsis Applications Under Apps Shown Here
81 There are a Wide Variety of Functions Available in Metamorph Not Detailed In This Short Synopsis Applications Under Journal Shown Here Journals are Very Useful For Combining Applications Seemlessly Together
82 There are a Wide Variety of Functions Available in Metamorph Not Detailed In This Short Synopsis Use Window to Find Open Windows
83 There are a Wide Variety of Functions Available in Metamorph Not Detailed In This Short Synopsis Go to Help to Obtain Instructions for Any of The Prior Functions
Widefield 1. Switching on
Widefield 1 Switching on 1. Ignite DG5 lamp - must be switched on first (if previous user has switched off, wait 30 min before igniting) 2. Wait 5s and then turn on the main DG5 controller switch. 3. DG5
More informationSimplified Instructions: Zeiss Brightfield Microscope S1000
Contents General Microscope Set-Up Adjust Illumination Focus Condenser Open Software Image Capture Settings Shading & Color Corrections Image Capture & Viewing Scaling and Measurements Synopsis of Other
More informationZeiss Axio Imager.A1 manual
Zeiss Axio Imager.A1 manual Power-up protocol 1. Mercury lamp 2. Power strip on shelf 3. Computer The Mercury lamp should always be first-on and last-off. This prevents any electrical surges caused by
More informationNikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol
Nikon Eclipse Ti-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol Please sign on the log sheet before switching on system. Turn on system Turn on A only if confocal mode or laser
More informationZeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)
University of Chicago Integrated Light Microscopy Core Dr. Vytas Bindokas, Director http://digital.bsd.uchicago.edu By: Christine Labno, Assistant Director Room: AB-129 Phone: 4-9040 Zeiss Axiovert 135
More informationGuide to Confocal 5. Starting session
Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope
More informationZeiss LSM 880 Protocol
Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.
More informationOverview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide
Operation 58. UltraVIEW VoX Getting Started Guide Overview This chapter outlines the basic methods used to operate the UltraVIEW VoX system. About other software Volocity places great demands on the computer
More informationZeiss AxioImager.Z2 Brightfield Protocol
Zeiss AxioImager.Z2 Brightfield Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge
More informationSimplified Instructions: IXMicro High Throughput Microscope S1225
Simplified Instructions: IXMicro High Throughput Microscope S1225 This is an Overview Only intended to refresh specific items learned in the Hands-On Training Use only in conjunction with your own much
More informationWidefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software
Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software September 2007 Check website for most current User Guide
More informationUse of the HSW5 Spinning Disk Confocal Microscope Updated last May 25, 2010 OK
Use of the HSW5 Spinning Disk Confocal Microscope Updated last May 25, 2010 OK Getting Started: 2 Starting Micromanager and Loading a Configuration 3 The Main Micromanager GUI 3 Configuration Settings
More informationOlympus IX71 Microscope and DP71 Camera Instructions
Olympus IX71 Microscope and DP71 Camera Instructions Microscopy in Medicine (MiM) Core Emory University Department of Medicine 1 Olympus IX71 Image Capture Procedure 2 3 1. STARTING-UP PROCEDURE: Remove
More informationContents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING
Operations Guide Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Nikon Eclipse 90i Operations Guide STARTUP Startup Powering Up Fluorescence
More informationTRAINING MANUAL. Olympus FV1000
TRAINING MANUAL Olympus FV1000 September 2014 TABLE OF CONTENTS A. Start-Up Procedure... 1 B. Visual Observation under the Microscope... 1 C. Image Acquisition... 4 A brief Overview of the Settings...
More informationSPINNING DISK CSU-X1 USER MANUAL
SPINNING DISK CSU-X1 USER MANUAL Starting the temperature controller... 2 Starting the CO2 controller... 3 Start the spinning disk... 4 Sample observation with the oculars... 5 Spatial sampling, Pixel
More informationLeica SPEII confocal microscope. Short Manual
Leica SPEII confocal microscope Short Manual Switching ON sequence: 1. Turn on the Workstation under the bench (top, far right). 2. Turn on the Supply Unit - Laser box (big green switch first and then
More informationStandard Operating Procedure (SOP) for Shared Equipment: Spinning Disk Confocal Microscope
Standard Operating Procedure (SOP) for Shared Equipment: Spinning Disk Confocal Microscope This document is to be used as a supplementary guide and not as a replacement for formal training. DO NOT operate
More informationZeiss LSM 780 Protocol
Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.
More informationNasmyth Ultraview Vox User Protocol
Nasmyth Ultraview Vox User Protocol Switch on all wall sockets labelled Nasmyth, switch camera on (power supply located on table behind monitor), switch on laser switch in laser rack, switch computer on
More informationLeica SP8 TCS Users Manual
Version : 07/08/0 Leica SP8 TCS Users Manual Start up:. Turn the PC Microscope, Scanner Power, Laser Power, and the Laser Emission key to on (bottom right of desk).. Turn on the fluorescent lamp (top left
More informationTraining Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope
Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Verify that main power switches on the
More informationOlympus Digital Microscope Camera (DP70) checklist
Smith College - July 2005 Olympus Digital Microscope Camera (DP70) checklist CONTENT, page no. Camera Information, 1 Startup, 1 Retrieve an Image, 2 Microscope Setup, 2 Capture, 3 Preview. 3 Color Balans,
More informationOPERATING INSTRUCTIONS
Zeiss LSM 510 M eta Confocal M icroscope OPERATING INSTRUCTIONS Starting the System: 1. Turn the black knob on the laser box one-quarter turn from Off to On. You will hear the laser cooling mechanisms
More informationDante (Microscope) & Beatrice (Guide) Orth Lab
Dante (Microscope) & Beatrice (Guide) Orth Lab Olympus IX81 Widefield Microscope User Guide v. 1.2 (11/2014) Objectives 4x/0.13NA UPLFLN Semi Apo 10x/0.4NA PH UPLAPO Plan Apo 20x/0.8NA PH UPLAPO Plan Apo
More informationLSM 710 Confocal Microscope Standard Operation Protocol
LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.
More informationTitle: Leica SP5 Confocal User Manual
Title: Leica SP5 Confocal User Manual Date of first issue: 23/10/2015 Date of review: Version: Admin For assistance or to report an issue Office: CG07 or 05 Email: Igmm-imaginghelpdesk@igmm.ed.ac.uk Website:
More informationLeica SP8 Resonant Confocal. Quick-Start Guide
Leica SP8 Resonant Confocal Quick-Start Guide Contents Start-up Preparing for Imaging Part 1 On the scope Part 2 Software interface Part 3 Heat & CO2 incubation Part 4 Other hardware options Shut-down
More informationOlympus Time-lapse Microscope Basic operation
Olympus Time-lapse Microscope Basic operation To start up the microscope 1. Switch on the Olympus UCB. (label as ) 1 Power Switch 2 2. Switch on the MT10. (label as ) Power Switch Page 1 of 18 3 3. Switch
More informationLSM 510 Meta Training Notes
LSM 510 Meta Training Notes Turning on the system Turn on X-Cite power supply. This supplies light for epifluorescence for viewing your samples through the microscope. Turn on the remote control switch.
More informationEverest System / Slidebook Operating Procedures
Everest System / Slidebook Operating Procedures NOTICE: This guide is meant to supplement training, not replace it. All users must be trained first hand by a core employee. Training of others in your lab
More informationUsing the Nikon TE2000 Inverted Microscope
Wellcome Trust Centre for Human Genetics Molecular Cytogenetics and Microscopy Core Using the Nikon TE2000 Inverted Microscope Fluorescence image acquisition using Scanalytic s IPLab software and the B&W
More informationZeiss 880 Training Notes Zen 2.3
Zeiss 880 Training Notes Zen 2.3 1 Turn on the HXP 120V Lamp 2 Turn on Main Power Switch Turn on the Systems PC Switch Turn on the Components Switch. 3 4 5 Turn on the PC and log into your account. Start
More informationLSM 510 Training Notes
LSM 510 Training Notes Turning on the system Turn on the arc lamp, found on the bench top left of the microscope. This supplies light for epifluorescence for viewing your samples through the microscope.
More informationLSM 780 Confocal Microscope Standard Operation Protocol
LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch
More informationZeiss 780 Training Notes
Zeiss 780 Training Notes Turn on Main Switch, System PC and Components Switches 780 Start up sequence Do you need the argon laser (458, 488, 514 nm lines)? Yes Turn on the laser s main power switch and
More informationTraining Guide for Leica SP8 Confocal/Multiphoton Microscope
Training Guide for Leica SP8 Confocal/Multiphoton Microscope LAS AF v3.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON power switch for epifluorescence
More informationSTART-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7
Leica DMI AF6000LX Table of contents START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 ACQUIRE MODULE 6 SETTING THE LIGHTPATH 6
More informationLeica SP8 TCS Users Manual
Leica SP8 TCS Users Manual Follow the procedure for start up and log on as posted in the lab. Please log on with your account only and do not share your password with anyone. We track and confirm usage
More informationThings to check before start-up.
Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your
More informationBrightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer
Brightfield Microscopy and Image Acquisition on Spotcam1 by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer Log onto the computer. Enter your username and password to log onto the server.
More informationUser manual for Olympus SD-OSR spinning disk confocal microscope
User manual for Olympus SD-OSR spinning disk confocal microscope Ved Prakash, PhD. Research imaging specialist Imaging & histology core University of Texas, Dallas ved.prakash@utdallas.edu Once you open
More informationQUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Imperial College London
Imperial College London Facility for Imaging by Light Microscopy QUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Observing Life As It Happens
More informationLEICA TCS SP5 AOBS TANDEM USER MANUAL
LEICA TCS SP5 AOBS TANDEM USER MANUAL STARTING THE SYSTEM...2 THE LAS AF SOFTWARE...3 THE «ACQUIRE» MENU...5 CHOOSE AND CREATE A SETTING...6 THE CONTROL PANEL...8 THE DMI6000B MICROSCOPE...10 ACQUIRE ONE
More informationTraining Guide for Carl Zeiss LSM 510 META Confocal Microscope
Training Guide for Carl Zeiss LSM 510 META Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON Components and System/PC switches
More informationMAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION
Olympus VS120-L100 Slide Scanner Standard Operating Procedure Startup 1) Red power bar switch (behind monitor) 2) Computer 3) Login: UserVS120 account (no password) 4) Double click: WAIT FOR INITIALIZATION
More informationTraining Guide for Carl Zeiss AxioZoom V16 Stereo Microscope
Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope ZEN 2012 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 If you require fluorescence imaging,
More informationMicroscopy from Carl Zeiss
Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path
More informationDiskovery Spinning Disk Guide
Diskovery Spinning Disk Guide qbi.microscopy@uq.edu.au Getting started The microscope and its peripherals (Fig. 1a) should always be turned on, but if they are not, turn them on in the following way: 1.
More informationMIF ZEISS LSM510 CONFOCAL USER PROTOCOL
MIF ZEISS LSM510 CONFOCAL USER PROTOCOL START-UP Turn on the Mercury Bulb Power Supply (if needed). Power-on the Control Box. Turn on the computer. Open the LSM 510 software. Choose Scan New Images and
More informationOPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop
OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop 1. Log in usage using the SMIF web site 2. Turn power
More informationLSM 800 Confocal Microscope Standard Operation Protocol
LSM 800 Confocal Microscope Standard Operation Protocol Turning on the system 1. Switch on the Main switch (labeled 1 and 2 ) mounted on the wall. 2. Turn the Laser Key (labeled 3 ) 90 clockwise for power
More informationCh. 1 - Installation Guidelines
Ch. 1 - Installation Guidelines Table of Contents Ch. 1 - Installation Guidelines Introduction... 8 The Image-Pro Driver Interface... 8 Installing the SPOT Image-Pro Driver... 8 Image-Pro Driver Supplement
More informationQuick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center
Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue
More informationUsing Autofocus in NIS-Elements
Using Autofocus in NIS-Elements Overview This technical note provides an overview of the available autofocus routines in NIS-Elements, and describes the necessary steps for using the autofocus functions.
More informationZEISS LSM510META confocal manual
ZEISS LSM510META confocal manual Switching on the system 1) Switch on the Remote Control button located on the table to the right of the microscope. This is the main switch for the whole system including
More informationQuick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center
Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue
More informationZeiss Deconvolution Microscope: A Quick Guide
Zeiss Deconvolution Microscope: A Quick Guide Start-up Uncover microscope. Do not put dust cover on the floor. Plug in both cameras. The default camera is the AxioCam HRm (monochrome camera) for fluorescence
More informationb. Turn the power switch and key to on position for blue laser.
OLYMPUS FLUOVIEW 300 CONFOCAL MICOSCOPE OPERATION PROCEDURE 1. Turn ON microscope in this order: 1) Turn on mercury lamp (Note: once the mercury lamp is turned off, DO NOT turn it back on for at least
More informationQUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Imperial College London
Imperial College London Facility for Imaging by Light Microscopy QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Observing Life As It Happens Startup procedure...
More informationCell Biology and Bioimaging Core
Cell Biology and Bioimaging Core Leica TCS SP5 Operating Instructions Starting up the instrument 1. First, log in the log book located on the confocal desk. Include your name, your lab s PI, an account
More informationBX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions
BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions Starting up: Schematic of Olympus BX-61. 1. Turn on Olympus microscope power box (left of microscope) with toggle switch on
More informationTitle: Nikon A1R Confocal User Manual
Title: Nikon A1R Confocal User Manual Date of first issue: 23/10/2015 Date of review: Version: Admin For assistance or to report an issue Office: CG.07 or CG.05 Email: Igmm-imaginghelpdesk@igmm.ed.ac.uk
More informationQuick Operation Guide
Quick Operation Guide Power ON Mounting specimens Set the specimen on the sample holder, and install the sample holder to the holder frame. Attach the holder frame to the XY stage. Type of holder Main
More informationAxioscan - Startup. 1. Turn on the Axioscan (button to the left) and turn on the computer. 2. Log on and start the ZEN Blue software from the desktop
Axioscan - Startup 1. Turn on the Axioscan (button to the left) and turn on the computer 2. Log on and start the ZEN Blue software from the desktop 3. Press ZEN slidescan and Start System 4. Start by changing
More informationEPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY
EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY TURN ON THE FOLLOWING EQUIPMENT The fluorescent light (if needed) The power strip for the microscope and accessories The CoolSNAP HQ camera on the right (Turn
More informationMIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL
MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL START-UP On the Switchbox, turn both black switches to the ON position. Wait for the microscope to boot up completely (watch the screen on the side of the microscope).
More informationTraining Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope
Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope ZEN 2009 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn Chameleon TiS laser key from Standby
More informationQuick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s
LSM 5 MP, LSM 510 and LSM 510 META M i c r o s c o p y f r o m C a r l Z e i s s Quick Guide Laser Scanning Microscopes LSM Software ZEN 2007 August 2007 We make it visible. Contents Page Contents... 1
More informationZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide
ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide Matyas Molnar, Biovis 2016 Starting the microscpe 1. Check the microscope if everything looks clean and normal. If not, report it in the logbook.
More informationLeica TCS SP8 Quick Start Guide
Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, EL6000 fluorescent light source for the microscope stand. 2. Turn on the Scanner Power
More informationOperation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009
Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Introduction of Fluoresence Confocal Microscopy The first confocal microscope was invented by Princeton
More informationZeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling
Zeiss LSM 510 Confocor III Training Notes Center for Cell Analysis & Modeling Confocor 3 Start Up Go to System Module Turn on Main Switch, System/ PC, and Components Switches Do you need the arc lamp?
More informationCAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE
University of Virginia ITC Academic Computing Health Sciences CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE Introduction The Olympus BH-2 microscope in ACHS s microscope lab has objectives from
More informationMetaMorph Imaging Handbook Update 6/4/13
MetaMorph Imaging Handbook Update 6/4/13 Startup FIRST turn on mercury lamp (Fluorescence) Computer and monitor Qimaging Camera (on top) Uniblitz Shutters-2 Halogen Lamp (Transmitted Light) Computer Login
More informationSOP: Polyvar Met Light Microscope
SOP Polyvar Met Light Microscope Page 1 of 8 SOP: Polyvar Met Light Microscope 1. Scope 1.1 This document describes the standard operating procedure (SOP) for the Polyvar Met Light Microscope. 2. Table
More informationLeica DB LB Research microscope and Studo Lite Imaging software
Leica DB LB Research microscope and Studo Lite Imaging software Room B523 User Guide Molecular Imaging Unit University of Helsinki www.miu.helsinki.fi 9.4.2008 1 GENERAL USER INFORMATION... 1 2 SETTINGS
More informationOlympus xcellence Software - basic user guide
Olympus xcellence Software - basic user guide This is a basic overview of setting up time lapse experiments using Olympus's xcellence software on BIU's IX81 inverted phase contrast system - the software
More informationNikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope
Nikon Eclipse Ti A1-A Confocal Operating Manual Start-up 1. Turn on Excite Fluorescent light power supply- metal halide. a. Cool down as for mercury bulb b. Wheel closed liquid light guide 2. Turn on power
More informationpersonal DELTAVISION (pdv)
GUIDELINES AND HINTS Version 1.3 (March 2015) personal DELTAVISION (pdv) Epifluorescence microscope from Applied Precision Inc.: The microscope can be found in room 1.320. For details see the architectural
More informationLeica Sp5 II Confocal User Guide
Leica Sp5 II Confocal User Guide Turning on the Confocal System (instructions are posted in the room) 1. Turn on Laser Power Button 2. Turn Key to On position 3. Turn on Scanner Power Button 4. Turn on
More informationHoriba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer
Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and
More informationNikon A1Rsi Confocal Start-Up Sequence
1. Turn the key on the Nikon LUN-V Laser Launch. Nikon A1Rsi Confocal Start-Up Sequence 2. Press the button the left side of the A1Rsi Controller unit. 3. Turn on the power strip underneath the microscope.
More informationBefore you start, make sure that you have a properly calibrated system to obtain high-quality images.
CONTENT Step 1: Optimizing your Workspace for Acquisition... 1 Step 2: Tracing the Region of Interest... 2 Step 3: Camera (& Multichannel) Settings... 3 Step 4: Acquiring a Background Image (Brightfield)...
More informationOperating Instructions for Zeiss LSM 510
Operating Instructions for Zeiss LSM 510 Location: GNL 6.312q (BSL3) Questions? Contact: Maxim Ivannikov, maivanni@utmb.edu 1 Attend A Complementary Training Before Using The Microscope All future users
More informationFixed cell DSU Spinning Disk Confocal and Slidebook 5.0 Quick Guide. Light Microscopy Core Facility University of Chicago Vers 1.
Fixed cell DSU Spinning Disk Confocal and Slidebook 5.0 Quick Guide Light Microscopy Core Facility University of Chicago Vers 1.0, February 2009 By: Vytas Bindokas, Ph.D., Core Director The fixed cell
More informationStandard Operating Procedure
CENTER FOR NANOSCALE SCIENCE AND ENGINEERING Standard Operating Procedure Microscope Software Brian Wajdyk Page 1 of 6 Important Images are not to be saved to the computer. They will be deleted without
More informationNikon E800 Operating Instructions.
Nikon E800 Operating Instructions. You can request electronic copies of this manual by contacting imaging@fhcrc.org. Copies are also available on the Scientific Imaging web site. Please send your comments
More informationCharacterization Microscope Nikon LV150
Characterization Microscope Nikon LV150 Figure 1: Microscope Nikon LV150 Introduction This upright optical microscope is designed for investigating up to 150 mm (6 inch) semiconductor wafers but can also
More informationZeiss LSM 510 Multiphoton Confocal Microscope
Zeiss LSM 510 Multiphoton Confocal Microscope Quick Start User Guide LSU Health Sciences Research Core Facility Table of Contents 1 Safety... Page 3 2 Turn On the System... Page 4 3 Start Up the ZEN Software.
More informationScanArray Overview. Principle of Operation. Instrument Components
ScanArray Overview The GSI Lumonics ScanArrayÒ Microarray Analysis System is a scanning laser confocal fluorescence microscope that is used to determine the fluorescence intensity of a two-dimensional
More informationImage Processing Tutorial Basic Concepts
Image Processing Tutorial Basic Concepts CCDWare Publishing http://www.ccdware.com 2005 CCDWare Publishing Table of Contents Introduction... 3 Starting CCDStack... 4 Creating Calibration Frames... 5 Create
More informationSHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014
CIAN LSM1 or LSM2 short instructions, version 1.4, September 2014 page 1 of 6 SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 Before starting To work with LSM1
More informationAxioscan - Startup. 1. Turn on the Axioscan on (button to the left) and turn on the computer
Axioscan - Startup 1. Turn on the Axioscan on (button to the left) and turn on the computer 2. Log on and start the ZEN Blue software from the desktop 3. Press ZEN slidescan and Start System 4. Start by
More informationZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL
ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL START THE SYSTEM... 2 START ZEN SOFTWARE... 3 SET THE TEMPERATURE AND THE CO2 CONTROLLERS... OBSERVATION AT OCULARS... 5 STATIF PRESENTATION... 6 ACQUIRE ONE
More informationLeica TCS SP8 Quick Start Guide
Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, Fluorescent Light for the microscope stand. 2. Turn on the Scanner Power (1) on the front
More informationJeol JEM Responsible personell: Endy ( ) Online booking is compulsory!
Jeol JEM 1230 Responsible personell: Endy (45279377) Online booking is compulsory! After training you will have access to working alone on the instrument. All insertion of samples is done by responsible
More informationMicroscope Confocal LSM510 META
Microscope Confocal LSM510 META Welcome to the Zeiss LSM 510 Meta Confocal tutorial. Before using the LSM 510 META, Log off any other computer that is open with your personal login. You will need to put
More informationDIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands
DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands Differential Interference Contrast (DIC) imaging is a technique used to increase contrast in brightfield images. In confocal systems,
More informationGXCapture 8.1 Instruction Manual
GT Vision image acquisition, managing and processing software GXCapture 8.1 Instruction Manual Contents of the Instruction Manual GXC is the shortened name used for GXCapture Square brackets are used to
More information