OPERATING INSTRUCTIONS
|
|
- Kathryn Flowers
- 6 years ago
- Views:
Transcription
1 Zeiss LSM 510 M eta Confocal M icroscope OPERATING INSTRUCTIONS Starting the System: 1. Turn the black knob on the laser box one-quarter turn from Off to On. You will hear the laser cooling mechanisms engage. 2. Flip the two toggle switches on top of the CPU from O to. 3. Press the power button on the front of the CPU in the top, right corner. Allow the system to load. 4. When the log-in box appears on the computer screen, use your Histology Core Account Number and password to login. Allow Windows to load completely before opening any software. 5. Double-click the Zen 2007 software logo in the center of the screen and allow software to load. 6. When the Zen 2007 Log-in box appears, choose Start System for laser confocal imaging or Image Processing to work with previously collected image files. Allow software to load completely. 7. Click the blue Laser menu bar and make sure the pro option is highlighted. a. Turn all the lasers On if they are not already. i. The Argon/2 laser must be turned to Standby, and will change to On when it has warmed up. b. Highlight the Argon/2 laser and click Laser Information. i. When the laser status says Ready, increase the output to 50%. Imaging & Light Path Setup: 1. Open the Imaging Setup and Light Path panels by clicking the blue menu bars. If you cannot open both at the same time, right-click the blue bar and remove the check mark from in front of Auto Close Mode. a. Under Light Path, select the appropriate settings for your fluorescent dyes you need to know the excitation and emission spectral data for all your dyes. ii. Laser place checks in the boxes for the lasers you want to use. Only check the ones you need. You can adjust laser output here, if necessary. iii. Beamsplitter HT directs specific wavelengths of light from the lasers to the specimen. Choose one that includes numbers to match all the lasers you are using. iv. Beamsplitter NT1 splits fluorescent emissions from specimen to the different collection channels wavelengths below the number selected are reflected, and numbers above are transmitted. If using channel S1, choose the filter with a number lower than the emission peak for the dye you will image at channel S1, but higher than the peak of any other dye you are using. If you are not using channel S1, select the mirror Nils Hasselmo Hall 312 Church St. SE Minneapolis, MN
2 v. Beamsplitter NT2 splits fluorescent emissions from specimen between channels 2 and 3. If using both channels, select a number between the emission peaks for the dyes you are collection on each channel. If only using one channel, choose a filter that will either reflect or transmit your light to the desired channel. vi. Emission Filters for channels 2 and 3, select a long-pass (LP) or band-pass (BP) filter that will transmit only the emission wavelengths you want to capture. LP filters transmit all wavelengths above the number listed. BP filters emit only wavelengths between the numbers listed. If imaging multiple dyes simultaneously (explained later), you must choose filters for channels 2 and 3 that do not overlap. If imaging multiple dyes sequentially or only one dye, simply choose the filter that best fits the emission curve for your dye. vii. Channel S1 can be used as a customizable filter, allows you to specify the exact range of wavelengths to be collected. Move the sliding triangles together to set the range, and click the check-box to activate the channel. Use this only when collecting three or more dyes, or when there is no pre-set filter at channels 2 or 3 appropriate for your dye. viii. Assign Pseudocolors click the ChS1, Ch2, and Ch3 icons and assign a different color to each. ix. Activate Channels click the check-box next to the channel icons (ChS1, Ch2, Ch3) you want to use, and a white check mark will appear. See simultaneous and sequential modes, below, to determine how to use the different channels. Simultaneous and Sequential Imaging Modes: 1. When collecting emissions from multiple dyes on a single specimen, you must decide whether you want to excite all the dyes at once (simultaneously) or to excite and collect emitted light for each dye individually (sequentially). Either option will generate a multicolored image, but the two options are affected by non-specific excitation and color bleed-through. If your dye emission spectra are close together and you cannot configure the light path to adequately separate the collected emissions, or if your dyes are excited by more than one laser, you will want to excite each dye sequentially. If these problems do not apply, you can image more quickly using the simultaneous mode. a. Tracks: i. Simultaneous mode uses only one track, and one or more of the three collection channels are used in that track. All of the lasers you selected will turn on and excite the specimen at the same time. All dye emissions will pass through the beamsplitters at the same time and are guided to the appropriate collection channel according to the configuration you ve created. ii. Sequential mode uses multiple tracks one for each dye you are imaging. For each track, only one laser is turned on, and only one collection channel is activated. If all the tracks are activated when you scan your specimen, you will get a multicolored image where each color was collected individually Nils Hasselmo Hall 312 Church St. SE Minneapolis, MN
3 b. Simultaneous Mode setup: i. In the Imaging Setup panel, make sure there is only one track, and that it is activated by a white check mark. If other tracks are also present, click on them to highlight one at a time and delete them by clicking the minus sign button at the bottom of the panel. ii. Follow the instructions above to set up your lasers, beamsplitters, filters, etc. in the Light Path panel as appropriate for your dyes. When finished, your Imaging Setup panel should resemble the figure to the right. The white vertical lines represent the lasers you ve selected, and the colored bars represent the wavelengths of light that will be collected by each channel. The panel to the right is an example of a simultaneous mode setup for Cy3, FITC, and DAPI in channels S1, 3, and 2, respectively. c. Sequential Mode setup: Laser Scanning & Image Collection: i. First configure the Light Path panel just as you would for simultaneous mode you have only one track at this point. All beamsplitters and filters must be identical for each of the tracks you will use, because the microscope cannot switch filters during image acquisition. All channels will appear under Track 1, and your configuration will be the same as for simultaneous mode, above. ii. In the Imaging Setup panel, click the plus sign button to add a second and third track. When you do add a track, it duplicates the exact settings you ve chosen for the first track. The configurations must be identical for each track because the microscope cannot switch optics during image acquisition. iii. Click the Track 1 bar to highlight it in the Imaging Setup panel the track-specific light path settings will appear in the Light Path panel, below. Configure Track 1 to image the dye with the longest emission wavelength. 1. In the Light Path panel, select the laser that will excite the dye with the highest emission wavelength, and unclick all other lasers. 2. Activate only the collection channel for this dye, and remove the check marks from the other channel icons to de-activate them. iv. Configure the other track(s) the same way, using Track 2 for the dye with the second-longest emission wavelength, and Track 3 for the dye with the shortest emission wavelength. Make sure all tracks are activated with a check mark in the Imaging Setup panel. The panel shown above is an example of a sequential mode setup for Cy3, FITC, and DAPI in Tracks 1, 2, and 3, respectively. 1. At the top right corner of the Zen software window, there are three buttons: Ocular, Camera, and LSM. a. Ocular Mode: you can view your specimen through the binocular lenses, illuminated by a mercury bulb for fluorescent reflected light, or a standard halogen lamp for transmitted light. The Light Path panel shows the configuration for reflected and transmitted light. Use this mode to find your region of interest and choose desired magnification. You cannot do confocal laser scanning or view your specimen on the computer screen in this mode. b. Camera Mode: currently, no camera is attached to this scope. Ignore this function. c. LSM Mode: you can scan your specimen with lasers to capture confocal images. Laser safety shields prevent you from viewing your specimen through the binocular lenses in this mode. You must be in this mode to view the confocal light path configuration and optimize laser-scanning settings Nils Hasselmo Hall 312 Church St. SE Minneapolis, MN
4 2. Find your region of interest: a. To find your region of interest on the slide, click the Ocular button. In the Light Path panel, click the light source button on the lower left corner this controls the reflected light. Click the On button to turn on the mercury lamp, and then increase the light intensity to a desired level. b. Use the buttons on the microscope to open the shutter, change light filters and magnification, and navigate the slide in x-y directions. c. When you have found the region you want to image at the desired magnification and it is in focus, close the shutter and then click the LSM button in the upper right corner of the screen. You will hear the laser safety shutters moving into place. The Light Path panel will change to show the confocal laser configuration you created. 3. Optimize signal intensity: a. In the Acquisition Mode panel, make the following adjustments: i. Frame Size: adjust change manually, use a pre-set ratio by clicking X*Y, or click Optimal to find the best resolution for your objective. This setting determines the number of digital pixels calculated for the scanned area, and is limited by the objective s numerical aperture. ii. Speed: this is how fast/slow the laser scans across your specimen. Slower speed gives the most smooth, crisp detail but can take extremely long, especially when doing Z-stacks. Faster speeds produce slightly more pixelated images but are faster. Pixelation can be smoothed by averaging (explained below). Additionally, fluorescent molecules respond better to (and do not photobleach as easily with) short bursts of excitation repeated several times (fast speed w/ averaging) compared to long, single periods of excitation (slow speed w/o averaging). We recommend you click Max and use that speed. iii. Averaging: when you first start scanning your specimen in order to optimize laser settings, do not average set the number to 1. Once you have adjusted the gain, offset, and laser intensity to the desired parameters, use averaging to smooth out your image and remove any unstable fluorescence signals. Increase the number from 2 to 8 (above 8 is not valuable) the higher you go, the more times the laser will scan over the same area. Change the mode to Line this means the laser repeats the multiple scans line by line, rather than doing the whole frame at once. This reduces the effect of any slight movement or light intensity changes. iv. Bit Depth: this determines the range of grayscale values that can be assigned to your image. Default setting is 8 Bit, but 12Bit is recommended for publication-quality images (caution: much larger file size!). v. Leave all other settings as they are shown in figure, above, unless you have had advanced training. b. In the Channels panel, make the following adjustments: i. Pinhole: This controls the size of the diaphragm through which your specimen s fluorescent emission is collected, and is a function of the wavelength of the emitted light and the numerical aperture of the objective. It must be adjusted any time you change objectives. Click the tab for the dye with the longest emission wavelength. Then click the 1 AU button this automatically calculates the appropriate pinhole size for your objective given the wavelength of your dye. Look at the measurement (above shows 12.7 µm section ). Then click the tab for the other dyes, and change the pinhole size until the section thickness matches that of the longest wavelength dye Nils Hasselmo Hall 312 Church St. SE Minneapolis, MN
5 c. Click the Find button in the upper left corner of the Zen window this function estimates the approximate gain and amplitude offset necessary for your specimen. The lasers will scan your specimen one time, adjust these two controls, and produce a multicolored image. The light levels may be close to correct, but they will most likely require further adjustment. d. To make fine adjustments to your image s signal intensities after using the Find function, click the Continuous button. The lasers will repeatedly scan your image, and you can make changes to different parameters and watch the effect in the image window. You want to do this quickly, especially if your specimen is sensitive to photo bleaching. You can also adjust the focus at this time using the focus knob on the TFT screen. i. Split View: - click this tab along the left side of the live image to see each color channel and the merged channel in four separate panels. ii. Signal Intensity range indicator: Once the continuous scanning begins, go to the controls panel under the image and click the 3-color square under the Merged button. This changes the imaged from color panels to grayscale panels with minimum and maximum intensities indicated by blue and red, respectively. Do not click the drop-down menu arrow this will change the assigned pseudocolor. iii. Laser Transmission [%]: Increase laser output slightly above You do not want to increase this too much, or it will photobleach your signal. Use the controls below to further adjust intensity. iv. Gain: adjust level individually for each dye by clicking the dye s channel tab and then increasing or decreasing the gain value. Optimal level will show positive signal approaching white, but with very little to no red pixels. Red pixels indicate over saturation, which is considered false data. v. e. Amplitude Offset: adjust level individually for each dye as above. Optimal level will show background levels approaching black, with very little to no blue pixels. Blue pixels indicate lack of background signal, and if this is reduced too much, it is considered false data. Once your parameters are set and your specimen is in focus, click Stop. If you are scanning at the highest speed possible, you now want to turn on the averaging function, as described above. When you are ready to collect your image, click Single. Your image will appear in the display as a pseudocolored image Nils Hasselmo Hall 312 Church St. SE Minneapolis, MN
6 4. Save Images: a. In the Open Images panel on the right side of the Zen window, click the box containing your image and then click the Save button under the image thumbnail, or select File and Save As from the Zen menu. b. Save your file as a.lsm document this automatically saves all acquisition parameters including objective and magnification, light path settings, laser intensity, gain, amplitude, etc. However,.lsm format is not compatible with software other than Zen, so you will also need to export your image to.tif format see below! c. Save your document to your own removable USB device or to your AHC-network server. CD and DVD discs are available for purchase from the HMCF staff. You may temporarily save your work in the data (D:/) drive of the confocal computer, however these images are not backed up and are not guaranteed to be protected they could be lost permanently if you do not transfer them! d. If you want to open your image in software other than Zen 2007, you must also export the image as a.tif file. 5. Shut Down System: i. In the Zen window, double-click the image you want to export, so that it appears in the center image panel. Click the Split tab so that you can see all individual channels separately. ii. From the Zen menu, click File and Export. In the Export window, choose Format: Tagged Image File and Data: Full Resolution Image Window Single Plane (or choose Series for Z-stacks). Click Select file name and save. Make sure the file name and destination are the same as what you used when you saved the.lsm file. a. Turn off all lasers. b. Close Zen software. You should see a warning message that the lasers must cool down for 5 minutes before you can shut down the rest of the system. After the cool-down period, you will hear the cooling fans turn off. c. Shut down the computer. d. Flip the two toggle switches on top of the CPU from to O. e. Turn the black knob on the laser box one quarter turn from ON to OFF. f. Turn off the room lights and make sure the door locks behind you Nils Hasselmo Hall 312 Church St. SE Minneapolis, MN
Zeiss 780 Training Notes
Zeiss 780 Training Notes Turn on Main Switch, System PC and Components Switches 780 Start up sequence Do you need the argon laser (458, 488, 514 nm lines)? Yes Turn on the laser s main power switch and
More informationQuick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s
LSM 5 MP, LSM 510 and LSM 510 META M i c r o s c o p y f r o m C a r l Z e i s s Quick Guide Laser Scanning Microscopes LSM Software ZEN 2007 August 2007 We make it visible. Contents Page Contents... 1
More informationLSM 710 Confocal Microscope Standard Operation Protocol
LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.
More informationMIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL
MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL START-UP On the Switchbox, turn both black switches to the ON position. Wait for the microscope to boot up completely (watch the screen on the side of the microscope).
More informationZeiss LSM 780 Protocol
Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.
More informationZeiss LSM 880 Protocol
Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.
More informationZeiss 880 Training Notes Zen 2.3
Zeiss 880 Training Notes Zen 2.3 1 Turn on the HXP 120V Lamp 2 Turn on Main Power Switch Turn on the Systems PC Switch Turn on the Components Switch. 3 4 5 Turn on the PC and log into your account. Start
More informationMicroscopy from Carl Zeiss
Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path
More informationTraining Guide for Carl Zeiss LSM 510 META Confocal Microscope
Training Guide for Carl Zeiss LSM 510 META Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON Components and System/PC switches
More informationZEISS LSM510META confocal manual
ZEISS LSM510META confocal manual Switching on the system 1) Switch on the Remote Control button located on the table to the right of the microscope. This is the main switch for the whole system including
More informationLSM 510 Training Notes
LSM 510 Training Notes Turning on the system Turn on the arc lamp, found on the bench top left of the microscope. This supplies light for epifluorescence for viewing your samples through the microscope.
More informationLSM 510 Meta Training Notes
LSM 510 Meta Training Notes Turning on the system Turn on X-Cite power supply. This supplies light for epifluorescence for viewing your samples through the microscope. Turn on the remote control switch.
More informationThings to check before start-up.
Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your
More informationLSM 780 Confocal Microscope Standard Operation Protocol
LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch
More informationCell Biology and Bioimaging Core
Cell Biology and Bioimaging Core Leica TCS SP5 Operating Instructions Starting up the instrument 1. First, log in the log book located on the confocal desk. Include your name, your lab s PI, an account
More informationZeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling
Zeiss LSM 510 Confocor III Training Notes Center for Cell Analysis & Modeling Confocor 3 Start Up Go to System Module Turn on Main Switch, System/ PC, and Components Switches Do you need the arc lamp?
More informationZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide
ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide Matyas Molnar, Biovis 2016 Starting the microscpe 1. Check the microscope if everything looks clean and normal. If not, report it in the logbook.
More informationOperating Instructions for Zeiss LSM 510
Operating Instructions for Zeiss LSM 510 Location: GNL 6.312q (BSL3) Questions? Contact: Maxim Ivannikov, maivanni@utmb.edu 1 Attend A Complementary Training Before Using The Microscope All future users
More informationLEICA TCS SP5 AOBS TANDEM USER MANUAL
LEICA TCS SP5 AOBS TANDEM USER MANUAL STARTING THE SYSTEM...2 THE LAS AF SOFTWARE...3 THE «ACQUIRE» MENU...5 CHOOSE AND CREATE A SETTING...6 THE CONTROL PANEL...8 THE DMI6000B MICROSCOPE...10 ACQUIRE ONE
More informationMicroscope Confocal LSM510 META
Microscope Confocal LSM510 META Welcome to the Zeiss LSM 510 Meta Confocal tutorial. Before using the LSM 510 META, Log off any other computer that is open with your personal login. You will need to put
More informationQuick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center
Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue
More informationLeica SP8 TCS Users Manual
Version : 07/08/0 Leica SP8 TCS Users Manual Start up:. Turn the PC Microscope, Scanner Power, Laser Power, and the Laser Emission key to on (bottom right of desk).. Turn on the fluorescent lamp (top left
More informationContents. Introduction
Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path and lasers... 12 Scanning
More informationLeica SP8 TCS Users Manual
Leica SP8 TCS Users Manual Follow the procedure for start up and log on as posted in the lab. Please log on with your account only and do not share your password with anyone. We track and confirm usage
More informationQuick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center
Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue
More informationLeica Sp5 II Confocal User Guide
Leica Sp5 II Confocal User Guide Turning on the Confocal System (instructions are posted in the room) 1. Turn on Laser Power Button 2. Turn Key to On position 3. Turn on Scanner Power Button 4. Turn on
More informationMIF ZEISS LSM510 CONFOCAL USER PROTOCOL
MIF ZEISS LSM510 CONFOCAL USER PROTOCOL START-UP Turn on the Mercury Bulb Power Supply (if needed). Power-on the Control Box. Turn on the computer. Open the LSM 510 software. Choose Scan New Images and
More informationTRAINING MANUAL. Olympus FV1000
TRAINING MANUAL Olympus FV1000 September 2014 TABLE OF CONTENTS A. Start-Up Procedure... 1 B. Visual Observation under the Microscope... 1 C. Image Acquisition... 4 A brief Overview of the Settings...
More informationREMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data.
1 Use of the Zeiss LSM 510 Inverted Firstly please be aware that this microscope should be treated with respect and care at all times. Rules of use: This Microscope can only be used by Masters by Research
More informationQuick Start Guide. Leica SP5 X
Quick Start Guide Leica SP5 X Please note: Some of the information in this guide was taken from Leica Microsystems Leica TCS SP5 LAS AF Guide for New Users. This work is licensed under the Creative Commons
More informationTraining Guide for Leica SP8 Confocal/Multiphoton Microscope
Training Guide for Leica SP8 Confocal/Multiphoton Microscope LAS AF v3.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON power switch for epifluorescence
More informationPractical work no. 3: Confocal Live Cell Microscopy
Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside
More informationTraining Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope
Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope ZEN 2009 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn Chameleon TiS laser key from Standby
More informationNikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope
Nikon Eclipse Ti A1-A Confocal Operating Manual Start-up 1. Turn on Excite Fluorescent light power supply- metal halide. a. Cool down as for mercury bulb b. Wheel closed liquid light guide 2. Turn on power
More informationTraining Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope
Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Verify that main power switches on the
More informationOperation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009
Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Introduction of Fluoresence Confocal Microscopy The first confocal microscope was invented by Princeton
More informationZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL
ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL START THE SYSTEM... 2 START ZEN SOFTWARE... 3 SET THE TEMPERATURE AND THE CO2 CONTROLLERS... OBSERVATION AT OCULARS... 5 STATIF PRESENTATION... 6 ACQUIRE ONE
More informationTitle: Leica SP5 Confocal User Manual
Title: Leica SP5 Confocal User Manual Date of first issue: 23/10/2015 Date of review: Version: Admin For assistance or to report an issue Office: CG07 or 05 Email: Igmm-imaginghelpdesk@igmm.ed.ac.uk Website:
More informationConfocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:
Confocal imaging on the Leica TCS SP8 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: 4) Do not touch the microscope while the software is initializing. Choose your options: Turn
More informationOperating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.
Smith College August 2010 Operating Checklist for using the Laser Scanning Confocal Microscope Leica TCS SP5. CONTENT, page no. Startup, 1 Initial set-up, 1 Software, 2 Microscope Specimen observation
More informationGuide to Confocal 5. Starting session
Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope
More informationLSM 800 Confocal Microscope Standard Operation Protocol
LSM 800 Confocal Microscope Standard Operation Protocol Turning on the system 1. Switch on the Main switch (labeled 1 and 2 ) mounted on the wall. 2. Turn the Laser Key (labeled 3 ) 90 clockwise for power
More informationTraining Guide for Carl Zeiss LSM 880 with AiryScan FAST
Training Guide for Carl Zeiss LSM 880 with AiryScan FAST ZEN 2.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2018) Power ON Routine 1 2 Turn ON Main Switch from the remote control
More informationOlympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope
Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope Improved Optics More Lasers 405 diode 440 diode 488 Argon 515 Argon 559
More informationSHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014
CIAN LSM1 or LSM2 short instructions, version 1.4, September 2014 page 1 of 6 SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 Before starting To work with LSM1
More informationLeica SPEII confocal microscope. Short Manual
Leica SPEII confocal microscope Short Manual Switching ON sequence: 1. Turn on the Workstation under the bench (top, far right). 2. Turn on the Supply Unit - Laser box (big green switch first and then
More informationb. Turn the power switch and key to on position for blue laser.
OLYMPUS FLUOVIEW 300 CONFOCAL MICOSCOPE OPERATION PROCEDURE 1. Turn ON microscope in this order: 1) Turn on mercury lamp (Note: once the mercury lamp is turned off, DO NOT turn it back on for at least
More informationNikon SIM-E & A1-R System
Nikon SIM-E & A1-R System USER GUIDE LSU Health Sciences Center Shreveport Research Core Facility June 01 2017 Chaowei Shang 1 Table of Content 1. Start Up the System... Page 3 Hardware and microscope
More informationTRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO
TRAINING MANUAL Multiphoton Microscopy LSM 510 META-NLO September 2010 Multiphoton Microscopy Training Manual Multiphoton microscopy is only available on the LSM 510 META-NLO system. This system is equipped
More information1 Co Localization and Working flow with the lsm700
1 Co Localization and Working flow with the lsm700 Samples -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ BrDU with alexa 488. -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ no BrDU (but with
More informationSimplified Instructions: Olympus Widefield Microscope S1230
Contents General Microscope Operation Simple Image Capture Multi-Wavelength Capture Z-Series Timelapse Combining Capture Modes Synopsis of Other Functions Pages 2-23 24-40 41-47 48-56 57-59 60-68 69-83
More informationInternal Medicine Imaging Core Emory University Department of Medicine
Internal Medicine Imaging Core Emory University Department of Medicine 1 OPERATION OF THE ZEISS LSM 510 META YOU MUST SIGN UP TO USE THE MICROSCOPE OR COMPUTER EVERY TIME NO EXCEPTIONS Before attempting
More informationZeiss LSM 510 Multiphoton Confocal Microscope
Zeiss LSM 510 Multiphoton Confocal Microscope Quick Start User Guide LSU Health Sciences Research Core Facility Table of Contents 1 Safety... Page 3 2 Turn On the System... Page 4 3 Start Up the ZEN Software.
More informationLeica TCS SP8 Quick Start Guide
Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, EL6000 fluorescent light source for the microscope stand. 2. Turn on the Scanner Power
More informationBi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging
Bi/BE 227 Winter 2016 Assignment #3 Adding the third dimension: 3D Confocal Imaging Schedule: Jan 20: Assignment Jan 20-Feb 8: Work on assignment Feb 10: Student PowerPoint presentations. Goals for this
More informationNikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.
Nikon AZ100 Laser Scanning Macro Confocal Microscope Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin May 2017 Contents 1 Introduction 2 2 Hardware - Startup 2 3 Software/Operation 4 3.1 Multidimensional
More informationLeica TCS SL Confocal Training. Neuroscience Imaging Core Staff. Core Director. Facility Manager
Leica TCS SL Confocal Training Neuroscience Imaging Facility The Ohio State University Rightmire Hall 614-292-1367 Staff Core Director Anthony Brown, Ph. D. 060 Rightmire Hall 614-292-1205 brown.2302@osu.edu
More informationZeiss Axio Imager.A1 manual
Zeiss Axio Imager.A1 manual Power-up protocol 1. Mercury lamp 2. Power strip on shelf 3. Computer The Mercury lamp should always be first-on and last-off. This prevents any electrical surges caused by
More informationCONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2)
Wellcome Trust Centre for Human Genetics Molecular Cytogenetics and Microscopy Core CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2) 1) STARTING THE SYSTEM Abridged INSTRUCTIONS Switch on the mercury bulb
More informationZeiss LSM 510 Multiphoton Confocal Microscope
Zeiss LSM 510 Multiphoton Confocal Microscope User Guide LSU Health Sciences Center-Shreveport Research Core Facility Table of Contents 1 Safety... Page 3 2 Turn On the System... Page 4 3 Start Up the
More informationZeiss LSM880 Operating Instructions. UTMB Optical Microscopy Core Jan. 16, 2018
Zeiss LSM880 Operating Instructions UTMB Optical Microscopy Core Jan. 16, 2018 1 1. Power up the microscope Sing the LOGBOOK Steps below will provide power to the computer and all of the microscope components.
More informationRENISHAW INVIA RAMAN SPECTROMETER
STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw invia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples
More informationSupplemental Method Information Zeiss LSM710
Supplemental Method Information Zeiss LSM710 1 Under the Light Path window set up the confocal for imaging a green dye (Alexa488-EGFP). For example, set up the light path as shown here using the 488 nm
More informationLeica TCS SP8 Quick Start Guide
Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, Fluorescent Light for the microscope stand. 2. Turn on the Scanner Power (1) on the front
More informationUser Guide to the IBIF Leica TCS SP8 MP Confocal Microscope
User Guide to the IBIF Leica TCS SP8 MP Confocal Microscope This version: 7.24.14. Introduction The IBIF confocal microscope is made available on a fee-for-use-hour basis to all users who have been trained.
More informationRenishaw InVia Raman microscope
Laser Spectroscopy Labs Renishaw InVia Raman microscope Operation instructions 1. Turn On the power switch, system power switch is located towards the back of the system on the right hand side. Wait ~10
More informationNikon A1Rsi Confocal Start-Up Sequence
1. Turn the key on the Nikon LUN-V Laser Launch. Nikon A1Rsi Confocal Start-Up Sequence 2. Press the button the left side of the A1Rsi Controller unit. 3. Turn on the power strip underneath the microscope.
More informationConfocal 510 Tutorial
Confocal 510 Tutorial You will have to log on in order to use the microscope. You will be charged according the time you are logged in, so please don t forget to log out after you are done. If you don
More informationSTART-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7
Leica DMI AF6000LX Table of contents START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 ACQUIRE MODULE 6 SETTING THE LIGHTPATH 6
More informationNikon A1R. Multi-Photon & Laser Scanning Confocal Microscope. Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin. April 2017.
Nikon A1R Multi-Photon & Laser Scanning Confocal Microscope Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin April 2017 Contents 1 Introduction 2 2 Start-Up 2 3 Imaging 4 3.1 Sample Alignment...........................................
More informationContents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING
Operations Guide Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Nikon Eclipse 90i Operations Guide STARTUP Startup Powering Up Fluorescence
More informationZeiss LSM 510 META Guide
Zeiss LSM 50 META Guide Ge#ng Started. Ensure the main power is switched on (black knob on the far le; of the laser module box).. Turn on the System/PC and Components switches (located on the grey switch
More informationHoriba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer
Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and
More informationUser manual for Olympus SD-OSR spinning disk confocal microscope
User manual for Olympus SD-OSR spinning disk confocal microscope Ved Prakash, PhD. Research imaging specialist Imaging & histology core University of Texas, Dallas ved.prakash@utdallas.edu Once you open
More informationZeiss AxioImager.Z2 Brightfield Protocol
Zeiss AxioImager.Z2 Brightfield Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge
More informationMicroscope Confocal Sp2 Upright.
Microscope Confocal Sp2 Upright. Welcome to the Leica Sp2 Confocal Upright tutorial. Before using the Sp2 Invert, You will need to put down your name on the reservation system = http://svintranet.epfl.ch/index.php?optio
More informationNikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON
N-SIM guide NIKON IMAGING CENTRE @ KING S COLLEGE LONDON Starting-up / Shut-down The NSIM hardware is calibrated after system warm-up occurs. It is recommended that you turn-on the system for at least
More informationZEN 2012 SP5 black edition Hotfix 12
Information about the software ZEN 2012 SP5 black edition Hotfix 12 Software name: ZEN 2012 Service Pack 5 black edition Hotfix 12 Software version: The software version in ZEN Help About changes to 14.0.12.201
More informationOlympus xcellence Software - basic user guide
Olympus xcellence Software - basic user guide This is a basic overview of setting up time lapse experiments using Olympus's xcellence software on BIU's IX81 inverted phase contrast system - the software
More informationKEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017)
KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017) 1 Introduction You must be trained to operate the Laser-scanning confocal microscope (LSCM) independently.
More informationSupplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each
Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each lens with a 1 Airy unit pinhole setting. Many water lenses
More informationZeiss Deconvolution Microscope: A Quick Guide
Zeiss Deconvolution Microscope: A Quick Guide Start-up Uncover microscope. Do not put dust cover on the floor. Plug in both cameras. The default camera is the AxioCam HRm (monochrome camera) for fluorescence
More informationDIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands
DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands Differential Interference Contrast (DIC) imaging is a technique used to increase contrast in brightfield images. In confocal systems,
More information1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the
1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the light path as shown here using the 488 nm LASER (Laser Unit 1) reflecting off of the 405/488 nm Dichroic mirror
More informationZeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)
University of Chicago Integrated Light Microscopy Core Dr. Vytas Bindokas, Director http://digital.bsd.uchicago.edu By: Christine Labno, Assistant Director Room: AB-129 Phone: 4-9040 Zeiss Axiovert 135
More informationDIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands
DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands Differential Interference Contrast (DIC) imaging is a technique used to increase contrast in brightfield images. In confocal systems,
More informationUsing the Nikon TE2000 Inverted Microscope
Wellcome Trust Centre for Human Genetics Molecular Cytogenetics and Microscopy Core Using the Nikon TE2000 Inverted Microscope Fluorescence image acquisition using Scanalytic s IPLab software and the B&W
More informationTopics. - How to calibrate the LSM scanner. - How to clean the microscope. - How to adjust the pinhole alignment. - How to adjust the Collimator
Topics - How to calibrate the LSM scanner - How to measure the PSF - How to clean the microscope - How to adjust the pinhole alignment - How to adjust the Collimator How to calibrate the LSM scanner The
More informationAxioVision 4.5 Brightfield Image Capture Procedure
AxioVision 4.5 Brightfield Image Capture Procedure 1. STARTING-UP PROCEDURE: Remove blue dust cover and place on shelf under microscope. Turn on the halogen lamp by pushing the switch at the back right
More informationTitle: Nikon A1R Confocal User Manual
Title: Nikon A1R Confocal User Manual Date of first issue: 23/10/2015 Date of review: Version: Admin For assistance or to report an issue Office: CG.07 or CG.05 Email: Igmm-imaginghelpdesk@igmm.ed.ac.uk
More informationConfocal Application Notes Vol. 5 July 2010
Tile Scan Prepared by Myriam Gastard, PhD Application and Technical Support Group, Leica Microsystems, Inc. In this issue of our Confocal Application Notes, proper set up of the Tile function enables you
More informationEverest System / Slidebook Operating Procedures
Everest System / Slidebook Operating Procedures NOTICE: This guide is meant to supplement training, not replace it. All users must be trained first hand by a core employee. Training of others in your lab
More informationThe Zeiss AiryScan System, Confocal Four.
The Zeiss AiryScan System, Confocal Four. Overview. The Zeiss AiryScan module is a segmented, radially stacked GaASP detector and collector system designed to subsample the airy disk of a point emission
More informationOverview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide
Operation 58. UltraVIEW VoX Getting Started Guide Overview This chapter outlines the basic methods used to operate the UltraVIEW VoX system. About other software Volocity places great demands on the computer
More informationWhy and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005
Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev Why use confocal microscopy? Principles of the laser scanning confocal microscope. Image resolution. Manipulating the
More informationNikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol
Nikon Eclipse Ti-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol Please sign on the log sheet before switching on system. Turn on system Turn on A only if confocal mode or laser
More informationZeiss Axioskop II. The AIF's "routine" light microscope. (Installed 8/24/04)AxioCam installed July 11th 2005
Zeiss Axioskop II The AIF's "routine" light microscope. (Installed 8/24/04)AxioCam installed July 11th 2005 Featuring: Phase Contrast Darkfield DIC/Nomarski Brightfield Fluorescent filters for Dapi, FITC,Rhodamine
More informationCMI STANDARD OPERATING PROCEDURE. Fluoview 300 laser scanning confocal microscope
CMI STANDARD OPERATING PROCEDURE Fluoview 300 laser scanning confocal microscope CMI documentid:sop001 CONTACT INFORMATION: Peter Owens: 091 494036 (office) Peter.owens@nuigalway.ie Kerry Thompson: 091
More informationWidefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software
Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software September 2007 Check website for most current User Guide
More informationNIS-Elements C (For CONFOCAL MICROSCOPE A1) Instructions (Ver. 4.40)
M487E 15.4.NF.17 (1/4) *M487EN17* NIS-Elements C (For CONFOCAL MICROSCOPE A1) Instructions (Ver. 4.40) Preface Thank you for purchasing the Nikon products. This instruction manual has been prepared for
More information