Operating Instructions for Zeiss LSM 510

Size: px
Start display at page:

Download "Operating Instructions for Zeiss LSM 510"

Transcription

1 Operating Instructions for Zeiss LSM 510 Location: GNL 6.312q (BSL3) Questions? Contact: Maxim Ivannikov, 1

2 Attend A Complementary Training Before Using The Microscope All future users should be first trained on an identical microscope located in BSB before using this confocal microscope. The training is complementary to BSL3 users. To set up your training and for any other questions, please contact: Maxim Ivannikov, maivanni@utmb.edu. 2

3 1. Powering up the microscope 1.1) Turn on the mercury lamp 1.2) Power the PC 1.3) Turn on the remote SWITCH #1 SWITCH #2 SWITCH #3 3

4 2. Starting Acquisition Image Manager (AIM) 2.1) Windows loads without a password. On the desktop click on the LSM510 icon. 2.2) To load the acquisition mode of AIM, click first on Scan New Images and then on Start Expert Mode. LSM ) The expert mode menu strip is loaded. Click on the Acquire button. 4

5 3. Switching on the lasers to be used 3.1) Click on the Laser button to open the Laser Control window. 3.2) In the Laser Control window select laser(s) based on the excitation wavelengths of fluorophores you will be imaging. Power the lasers on by clicking on either Standby or On button. If the Argon laser is used, bring the output (%) slider to 50%. The 405 laser will power on automatically. Click Close. Tip: DAPI needs 405 nm (violet) laser Alexa 488 & EGFP need 488 nm (blue) laser Alexa & tdtomato need 543 nm (green) laser Alexa & Cy5 need 633 nm (red) laser 5

6 4. Choosing an objective 4.1) Click on the VIS button to enable viewing of your sample with oculars/eyepieces. Then, click on the Micro button to open the Microscope Control window. 4.2) In the open window, click on the objective icon to open a drop-down objective selection menu. Select an objective based on desired detail resolution (objective s NA) and optical magnification (MAGx). Tip: The microscope has 5 objectives, 2 water (10x, 63x) and 1 oil immersion (100x) and 2 air (20x, 40x) objectives. Example: C-Apochromat 63x/1.2W corr: 63x magnification, 1.2 NA, Water immersion objective with a corrective collar (used to adjust to your coverslip thickness). 6

7 5. Focusing on your sample 5.1) If using an immersion objective, apply a drop of either filtered water OR oil to the front lens of an immersion objective. Do not apply anything onto air objectives! 5.2) Click on the Stage icon to bring up the Stage and Focus Control window ) In the Stage and Focus Control window, click on the load button to lower the objective all the way down. Click the Close button. 5.4) Place your sample onto the stage with coverslip down. Bring the objective slowly up using coarse focus knob on the microscope body until the coverslip just touches the immersion media. Stop. Important: If the microscope makes a beeping sound when you try to raise the objective. Keep pressing on the button on the right side of the microscope while moving the focus knob to override the focus lock.

8 5. Focusing on your sample (cont.) 5.5) Navigate to the Microscope Control window (see step 4.1). Based on fluorophores in your sample, choose and click on any of the 3 filter sets (DAPI, FITC, TRITC buttons) to focus on your sample using the oculars. (You should now see the light coming out of the objective) 5.6) While looking into the oculars, rotate the focus knob to bring the objective up until you see your sample in focus. Be careful not to crash the objective into your sample! Click on the OFF button to turn the light off, then Close the Microscope control window. Important: If the microscope makes a beeping sound when you try to focus. Keep pressing on the button on the right side of the microscope while moving the focus knob to override the focus lock. 5.7) Click on the LSM icon to switch to the confocal imaging mode. 8

9 6. Loading imaging settings 6.1) Click on the Config button to open the Configuration Control window ) Switch to the Multi Track mode and then click on the Config diskette button to select and then Apply an imaging configuration. (If required lasers are not on, a warning will be shown, see 3.2) Tip: Choose a protocol based on fluorophores present in your sample. For 4 dyes : 405_488_543_633_Fast_Line_Mode protocol is faster than 405_488_543_633_NoBleed_Slow_Frame_Mode, but might have significant cross-talk between the channels! (test by blanking laser lines)

10 7. Adjusting channels and image parameters 7.1) Click on the Scan button to open the Scan Control window alongside the Configuration Control window 10

11 7. Adjusting channels and image parameters (cont.) 7.2) Adjusting scan area, image size, pixel depth and scan speed in the Scan Control window. Make sure the Mode button is on Scanning images/2d arrays of pixels - Frame button on. Shows and selects an objective (if new selected refocusing will have to be redone). Mind Immersion! Chose pixel dimensions of your final images by pressing on the preset buttons (1:1 aspect ratio) or by typing in. Start with 512 by 512 or 1024 by Pixel scan speed, slower the better the signal to noise. Set to 7-9. Pixel depth (# of shades of gray in your image). Choose: 8bit Image averaging for a better signal to noise ratio. The more the longer the scan. Choose: Line Mean 2 or 4 As required: Zoom into the area & rotate the scan area As required: Move the scan area (to center the specimen) 11

12 7. Adjusting channels and image parameters (cont.) 7.3) Adjusting individual channels for optimal resolution and to minimize under- and overexposed pixels using the Scan Control window. In the Configuration Control window, uncheck all but one track. In the Scan Control window, click on the Channels button Click on the Fast XY button to start fast continuous image scan. An image window will automatically open. Drag to minimize the image window if it overflows the screen. From the menu strip on the right click on the Palette button. In the Color Palette window select Range Indicator and then Close the window. Overexposed pixels will be shown in the image in red, and underexposed in blue. 12

13 7. Adjusting channels and image parameters (cont.) 7.3) Continued in the Scan Control window Set Pinhole to 1 airy unit by clicking on the 1 button. Estimated Z axis image thickness (Optical slice) will be displayed below. At this point adjust the focus (fine focus knob) to find the brightest focal plane. If necessary, move the stage to find another area. If there are many blue pixels, increase the Offset. If there are no blue pixels decrease the Offset until you see very few blue pixels. If the image is dim (dark gray/black) increase Detector Gain and/or Laser Power (next to the checked box) until you see few red pixels. Sometimes, if staining is too weak, image brightness can be increased by slightly opening the pinhole. If the image has red pixels reduce Laser Power (next to the checked box) and/or Detector Gain. Important: the Laser Power slider for the 405 diode laser is not operational, adjust Detector Gain only! SEE LAST PAGE FOR MORE Click on the Stop button to stop the scan. Too many blue pixels & low signal. Solution: 1. Focus better 2. Increase amplifier offset to eliminate blue pixels 3. Increase laser power or detector gain (to brighten pixels up). Too many blue pixels & high signal. Solution: 1. Focus better 2. Increase offset to eliminate blue pixels 3. Decrease laser power or detector gain (to desaturate pixels up). Well-adjusted channel: No blue, no red pixels. Pixels are well stretched in the histogram. 13

14 8. Taking a snap 8.1) Go back to the beginning of 7.3 step to adjust the remaining tracks one by one. 14 In the Configuration Control window, uncheck the adjusted track and check the next on to be adjusted. 8.2) After all of the tracks are adjusted. Check all of the tracks. In the Configuration Control window check all of the tracks. 8.3) To take a snap of your sample with all channel. Click on the Single button in the Scan Control window. Save the image in your folder. (tip: if images are grey in color change the palette to no palette, see p.12) 8.4) In the image window: To show a merge of all channels click on the xy button, to show all channels individually click on Split xy button. To draw a scale bar, click on the Overlay button, select the scale bar and draw it on the image. To adjust image contrast and brightness click on the Contr button. You can reuse all of your settings next time you come to use the microscope by opening a past image and clicking on the reuse button!

15 9. Taking a Z-stack Source for the slide: Zeiss 510 Meta Tour Presentation 15

16 10. Shutting the microscope down 1. Save and close all of your images 2. Click on the laser button and turn all of the lasers OFF 3. Close the AIM program 4. Copy your data to your USB storage device 5. Turn off the computer 6. Remove your sample from the microscome stage and wipe the objective clean with LENS PAPER 7. Cover the microscope with the dust cover 8. If 5 minutes have not yet passed since turning off the lasers, wait, then turn the remote control OFF (switch #3) 9. Turn the mercury lamp OFF (switch #1) 16

17 Only If needed: the power of the 405 diode laser can be changed by the adjustment knob on the laser driver box as shown below: 17

ZEISS LSM510META confocal manual

ZEISS LSM510META confocal manual ZEISS LSM510META confocal manual Switching on the system 1) Switch on the Remote Control button located on the table to the right of the microscope. This is the main switch for the whole system including

More information

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL MIF ZEISS LSM510 CONFOCAL USER PROTOCOL START-UP Turn on the Mercury Bulb Power Supply (if needed). Power-on the Control Box. Turn on the computer. Open the LSM 510 software. Choose Scan New Images and

More information

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 710 Confocal Microscope Standard Operation Protocol LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.

More information

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL START-UP On the Switchbox, turn both black switches to the ON position. Wait for the microscope to boot up completely (watch the screen on the side of the microscope).

More information

LSM 510 Meta Training Notes

LSM 510 Meta Training Notes LSM 510 Meta Training Notes Turning on the system Turn on X-Cite power supply. This supplies light for epifluorescence for viewing your samples through the microscope. Turn on the remote control switch.

More information

LSM 510 Training Notes

LSM 510 Training Notes LSM 510 Training Notes Turning on the system Turn on the arc lamp, found on the bench top left of the microscope. This supplies light for epifluorescence for viewing your samples through the microscope.

More information

Zeiss LSM880 Operating Instructions. UTMB Optical Microscopy Core Jan. 16, 2018

Zeiss LSM880 Operating Instructions. UTMB Optical Microscopy Core Jan. 16, 2018 Zeiss LSM880 Operating Instructions UTMB Optical Microscopy Core Jan. 16, 2018 1 1. Power up the microscope Sing the LOGBOOK Steps below will provide power to the computer and all of the microscope components.

More information

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling Zeiss LSM 510 Confocor III Training Notes Center for Cell Analysis & Modeling Confocor 3 Start Up Go to System Module Turn on Main Switch, System/ PC, and Components Switches Do you need the arc lamp?

More information

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Verify that main power switches on the

More information

Cell Biology and Bioimaging Core

Cell Biology and Bioimaging Core Cell Biology and Bioimaging Core Leica TCS SP5 Operating Instructions Starting up the instrument 1. First, log in the log book located on the confocal desk. Include your name, your lab s PI, an account

More information

LSM 780 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch

More information

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope Training Guide for Carl Zeiss LSM 510 META Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON Components and System/PC switches

More information

Things to check before start-up.

Things to check before start-up. Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your

More information

Practical work no. 3: Confocal Live Cell Microscopy

Practical work no. 3: Confocal Live Cell Microscopy Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside

More information

b. Turn the power switch and key to on position for blue laser.

b. Turn the power switch and key to on position for blue laser. OLYMPUS FLUOVIEW 300 CONFOCAL MICOSCOPE OPERATION PROCEDURE 1. Turn ON microscope in this order: 1) Turn on mercury lamp (Note: once the mercury lamp is turned off, DO NOT turn it back on for at least

More information

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: Confocal imaging on the Leica TCS SP8 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: 4) Do not touch the microscope while the software is initializing. Choose your options: Turn

More information

Zeiss 880 Training Notes Zen 2.3

Zeiss 880 Training Notes Zen 2.3 Zeiss 880 Training Notes Zen 2.3 1 Turn on the HXP 120V Lamp 2 Turn on Main Power Switch Turn on the Systems PC Switch Turn on the Components Switch. 3 4 5 Turn on the PC and log into your account. Start

More information

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue

More information

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue

More information

CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2)

CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2) Wellcome Trust Centre for Human Genetics Molecular Cytogenetics and Microscopy Core CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2) 1) STARTING THE SYSTEM Abridged INSTRUCTIONS Switch on the mercury bulb

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Version : 07/08/0 Leica SP8 TCS Users Manual Start up:. Turn the PC Microscope, Scanner Power, Laser Power, and the Laser Emission key to on (bottom right of desk).. Turn on the fluorescent lamp (top left

More information

LSM 800 Confocal Microscope Standard Operation Protocol

LSM 800 Confocal Microscope Standard Operation Protocol LSM 800 Confocal Microscope Standard Operation Protocol Turning on the system 1. Switch on the Main switch (labeled 1 and 2 ) mounted on the wall. 2. Turn the Laser Key (labeled 3 ) 90 clockwise for power

More information

Quick Start Guide. Leica SP5 X

Quick Start Guide. Leica SP5 X Quick Start Guide Leica SP5 X Please note: Some of the information in this guide was taken from Leica Microsystems Leica TCS SP5 LAS AF Guide for New Users. This work is licensed under the Creative Commons

More information

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope ZEN 2009 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn Chameleon TiS laser key from Standby

More information

Internal Medicine Imaging Core Emory University Department of Medicine

Internal Medicine Imaging Core Emory University Department of Medicine Internal Medicine Imaging Core Emory University Department of Medicine 1 OPERATION OF THE ZEISS LSM 510 META YOU MUST SIGN UP TO USE THE MICROSCOPE OR COMPUTER EVERY TIME NO EXCEPTIONS Before attempting

More information

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes Turn on Main Switch, System PC and Components Switches 780 Start up sequence Do you need the argon laser (458, 488, 514 nm lines)? Yes Turn on the laser s main power switch and

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Leica SP8 TCS Users Manual Follow the procedure for start up and log on as posted in the lab. Please log on with your account only and do not share your password with anyone. We track and confirm usage

More information

LEICA TCS SP5 AOBS TANDEM USER MANUAL

LEICA TCS SP5 AOBS TANDEM USER MANUAL LEICA TCS SP5 AOBS TANDEM USER MANUAL STARTING THE SYSTEM...2 THE LAS AF SOFTWARE...3 THE «ACQUIRE» MENU...5 CHOOSE AND CREATE A SETTING...6 THE CONTROL PANEL...8 THE DMI6000B MICROSCOPE...10 ACQUIRE ONE

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS Zeiss LSM 510 M eta Confocal M icroscope OPERATING INSTRUCTIONS Starting the System: 1. Turn the black knob on the laser box one-quarter turn from Off to On. You will hear the laser cooling mechanisms

More information

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope Nikon Eclipse Ti A1-A Confocal Operating Manual Start-up 1. Turn on Excite Fluorescent light power supply- metal halide. a. Cool down as for mercury bulb b. Wheel closed liquid light guide 2. Turn on power

More information

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s LSM 5 MP, LSM 510 and LSM 510 META M i c r o s c o p y f r o m C a r l Z e i s s Quick Guide Laser Scanning Microscopes LSM Software ZEN 2007 August 2007 We make it visible. Contents Page Contents... 1

More information

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Introduction of Fluoresence Confocal Microscopy The first confocal microscope was invented by Princeton

More information

Zeiss LSM 880 Protocol

Zeiss LSM 880 Protocol Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Zeiss LSM 780 Protocol

Zeiss LSM 780 Protocol Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope Improved Optics More Lasers 405 diode 440 diode 488 Argon 515 Argon 559

More information

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 CIAN LSM1 or LSM2 short instructions, version 1.4, September 2014 page 1 of 6 SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 Before starting To work with LSM1

More information

Microscope Confocal LSM510 META

Microscope Confocal LSM510 META Microscope Confocal LSM510 META Welcome to the Zeiss LSM 510 Meta Confocal tutorial. Before using the LSM 510 META, Log off any other computer that is open with your personal login. You will need to put

More information

Zeiss LSM 510 Multiphoton Confocal Microscope

Zeiss LSM 510 Multiphoton Confocal Microscope Zeiss LSM 510 Multiphoton Confocal Microscope Quick Start User Guide LSU Health Sciences Research Core Facility Table of Contents 1 Safety... Page 3 2 Turn On the System... Page 4 3 Start Up the ZEN Software.

More information

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5. Smith College August 2010 Operating Checklist for using the Laser Scanning Confocal Microscope Leica TCS SP5. CONTENT, page no. Startup, 1 Initial set-up, 1 Software, 2 Microscope Specimen observation

More information

TRAINING MANUAL. Olympus FV1000

TRAINING MANUAL. Olympus FV1000 TRAINING MANUAL Olympus FV1000 September 2014 TABLE OF CONTENTS A. Start-Up Procedure... 1 B. Visual Observation under the Microscope... 1 C. Image Acquisition... 4 A brief Overview of the Settings...

More information

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL START THE SYSTEM... 2 START ZEN SOFTWARE... 3 SET THE TEMPERATURE AND THE CO2 CONTROLLERS... OBSERVATION AT OCULARS... 5 STATIF PRESENTATION... 6 ACQUIRE ONE

More information

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Training Guide for Leica SP8 Confocal/Multiphoton Microscope Training Guide for Leica SP8 Confocal/Multiphoton Microscope LAS AF v3.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON power switch for epifluorescence

More information

Zeiss LSM 510 Multiphoton Confocal Microscope

Zeiss LSM 510 Multiphoton Confocal Microscope Zeiss LSM 510 Multiphoton Confocal Microscope User Guide LSU Health Sciences Center-Shreveport Research Core Facility Table of Contents 1 Safety... Page 3 2 Turn On the System... Page 4 3 Start Up the

More information

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO TRAINING MANUAL Multiphoton Microscopy LSM 510 META-NLO September 2010 Multiphoton Microscopy Training Manual Multiphoton microscopy is only available on the LSM 510 META-NLO system. This system is equipped

More information

Guide to Confocal 5. Starting session

Guide to Confocal 5. Starting session Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope

More information

Zeiss Axio Imager.A1 manual

Zeiss Axio Imager.A1 manual Zeiss Axio Imager.A1 manual Power-up protocol 1. Mercury lamp 2. Power strip on shelf 3. Computer The Mercury lamp should always be first-on and last-off. This prevents any electrical surges caused by

More information

Leica SPEII confocal microscope. Short Manual

Leica SPEII confocal microscope. Short Manual Leica SPEII confocal microscope Short Manual Switching ON sequence: 1. Turn on the Workstation under the bench (top, far right). 2. Turn on the Supply Unit - Laser box (big green switch first and then

More information

Contents. Introduction

Contents. Introduction Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path and lasers... 12 Scanning

More information

Microscopy from Carl Zeiss

Microscopy from Carl Zeiss Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path

More information

3 Choose the Channels button and set the Channel Settings. Set the Pinhole to 1 Airy unit.

3 Choose the Channels button and set the Channel Settings. Set the Pinhole to 1 Airy unit. 1 Set up the confocal light path for imaging a green dye (e.g. Alexa488-EGFP). For example, under the Configuration Control window the light path could be set up as shown here using the 488 nm LASER (found

More information

Supplemental Method Information Zeiss LSM710

Supplemental Method Information Zeiss LSM710 Supplemental Method Information Zeiss LSM710 1 Under the Light Path window set up the confocal for imaging a green dye (Alexa488-EGFP). For example, set up the light path as shown here using the 488 nm

More information

REMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data.

REMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data. 1 Use of the Zeiss LSM 510 Inverted Firstly please be aware that this microscope should be treated with respect and care at all times. Rules of use: This Microscope can only be used by Masters by Research

More information

Zeiss LSM 510 META Guide

Zeiss LSM 510 META Guide Zeiss LSM 50 META Guide Ge#ng Started. Ensure the main power is switched on (black knob on the far le; of the laser module box).. Turn on the System/PC and Components switches (located on the grey switch

More information

Topics. - How to calibrate the LSM scanner. - How to clean the microscope. - How to adjust the pinhole alignment. - How to adjust the Collimator

Topics. - How to calibrate the LSM scanner. - How to clean the microscope. - How to adjust the pinhole alignment. - How to adjust the Collimator Topics - How to calibrate the LSM scanner - How to measure the PSF - How to clean the microscope - How to adjust the pinhole alignment - How to adjust the Collimator How to calibrate the LSM scanner The

More information

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017. Nikon AZ100 Laser Scanning Macro Confocal Microscope Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin May 2017 Contents 1 Introduction 2 2 Hardware - Startup 2 3 Software/Operation 4 3.1 Multidimensional

More information

Nikon A1R. Multi-Photon & Laser Scanning Confocal Microscope. Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin. April 2017.

Nikon A1R. Multi-Photon & Laser Scanning Confocal Microscope. Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin. April 2017. Nikon A1R Multi-Photon & Laser Scanning Confocal Microscope Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin April 2017 Contents 1 Introduction 2 2 Start-Up 2 3 Imaging 4 3.1 Sample Alignment...........................................

More information

Title: Leica SP5 Confocal User Manual

Title: Leica SP5 Confocal User Manual Title: Leica SP5 Confocal User Manual Date of first issue: 23/10/2015 Date of review: Version: Admin For assistance or to report an issue Office: CG07 or 05 Email: Igmm-imaginghelpdesk@igmm.ed.ac.uk Website:

More information

Leica Sp5 II Confocal User Guide

Leica Sp5 II Confocal User Guide Leica Sp5 II Confocal User Guide Turning on the Confocal System (instructions are posted in the room) 1. Turn on Laser Power Button 2. Turn Key to On position 3. Turn on Scanner Power Button 4. Turn on

More information

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST Training Guide for Carl Zeiss LSM 880 with AiryScan FAST ZEN 2.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2018) Power ON Routine 1 2 Turn ON Main Switch from the remote control

More information

Leica SP8 Resonant Confocal. Quick-Start Guide

Leica SP8 Resonant Confocal. Quick-Start Guide Leica SP8 Resonant Confocal Quick-Start Guide Contents Start-up Preparing for Imaging Part 1 On the scope Part 2 Software interface Part 3 Heat & CO2 incubation Part 4 Other hardware options Shut-down

More information

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging Bi/BE 227 Winter 2016 Assignment #3 Adding the third dimension: 3D Confocal Imaging Schedule: Jan 20: Assignment Jan 20-Feb 8: Work on assignment Feb 10: Student PowerPoint presentations. Goals for this

More information

Nikon SIM-E & A1-R System

Nikon SIM-E & A1-R System Nikon SIM-E & A1-R System USER GUIDE LSU Health Sciences Center Shreveport Research Core Facility June 01 2017 Chaowei Shang 1 Table of Content 1. Start Up the System... Page 3 Hardware and microscope

More information

1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the

1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the 1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the light path as shown here using the 488 nm LASER (Laser Unit 1) reflecting off of the 405/488 nm Dichroic mirror

More information

Using the Nikon TE2000 Inverted Microscope

Using the Nikon TE2000 Inverted Microscope Wellcome Trust Centre for Human Genetics Molecular Cytogenetics and Microscopy Core Using the Nikon TE2000 Inverted Microscope Fluorescence image acquisition using Scanalytic s IPLab software and the B&W

More information

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide Matyas Molnar, Biovis 2016 Starting the microscpe 1. Check the microscope if everything looks clean and normal. If not, report it in the logbook.

More information

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 Leica DMI AF6000LX Table of contents START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 ACQUIRE MODULE 6 SETTING THE LIGHTPATH 6

More information

The Zeiss AiryScan System, Confocal Four.

The Zeiss AiryScan System, Confocal Four. The Zeiss AiryScan System, Confocal Four. Overview. The Zeiss AiryScan module is a segmented, radially stacked GaASP detector and collector system designed to subsample the airy disk of a point emission

More information

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope ZEN 2012 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 If you require fluorescence imaging,

More information

Confocal 510 Tutorial

Confocal 510 Tutorial Confocal 510 Tutorial You will have to log on in order to use the microscope. You will be charged according the time you are logged in, so please don t forget to log out after you are done. If you don

More information

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each lens with a 1 Airy unit pinhole setting. Many water lenses

More information

KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017)

KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017) KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017) 1 Introduction You must be trained to operate the Laser-scanning confocal microscope (LSCM) independently.

More information

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Operations Guide Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Nikon Eclipse 90i Operations Guide STARTUP Startup Powering Up Fluorescence

More information

Standard Operating Procedure (SOP) for Shared Equipment: Spinning Disk Confocal Microscope

Standard Operating Procedure (SOP) for Shared Equipment: Spinning Disk Confocal Microscope Standard Operating Procedure (SOP) for Shared Equipment: Spinning Disk Confocal Microscope This document is to be used as a supplementary guide and not as a replacement for formal training. DO NOT operate

More information

Microscope Confocal Sp2 Upright.

Microscope Confocal Sp2 Upright. Microscope Confocal Sp2 Upright. Welcome to the Leica Sp2 Confocal Upright tutorial. Before using the Sp2 Invert, You will need to put down your name on the reservation system = http://svintranet.epfl.ch/index.php?optio

More information

Confocal Laser Scanning Microscopy

Confocal Laser Scanning Microscopy Name of the Core Facility: Confocal Laser Scanning Microscopy CORE Forschungszentrum Immunologie Mainz Welcome to the CSLM Core Facility: The CLSM Core Facility enables working groups to incorporate high

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, EL6000 fluorescent light source for the microscope stand. 2. Turn on the Scanner Power

More information

Leica TCS SL Confocal Training. Neuroscience Imaging Core Staff. Core Director. Facility Manager

Leica TCS SL Confocal Training. Neuroscience Imaging Core Staff. Core Director. Facility Manager Leica TCS SL Confocal Training Neuroscience Imaging Facility The Ohio State University Rightmire Hall 614-292-1367 Staff Core Director Anthony Brown, Ph. D. 060 Rightmire Hall 614-292-1205 brown.2302@osu.edu

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, Fluorescent Light for the microscope stand. 2. Turn on the Scanner Power (1) on the front

More information

Diskovery Spinning Disk Guide

Diskovery Spinning Disk Guide Diskovery Spinning Disk Guide qbi.microscopy@uq.edu.au Getting started The microscope and its peripherals (Fig. 1a) should always be turned on, but if they are not, turn them on in the following way: 1.

More information

Zeiss AxioObserver with ApoTome

Zeiss AxioObserver with ApoTome Zeiss AxioObserver with ApoTome Quick Start User Guide LSU Health Sciences Center-Shreveport Research Core Facility (RCF) Microscopy Table of Contents 1 Start up the system.. Page 3 2 Touch screen controller

More information

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide Nikon C1si Spectral Laser Scanning Confocal Microscope User Guide Contents: C1Si Turn-On/ShutDown Procedures... 2 Overview... 4 Setup for epi-illumination to view through the eyepieces:... 5 Setup for

More information

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005 Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev Why use confocal microscopy? Principles of the laser scanning confocal microscope. Image resolution. Manipulating the

More information

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands Differential Interference Contrast (DIC) imaging is a technique used to increase contrast in brightfield images. In confocal systems,

More information

LSM 510 META in Chang Gung University

LSM 510 META in Chang Gung University Content LSM 510 META in Chang ung University LSM 510 META 路 理 The features and applications of LSM 510 META 01-09 Introduction of the hardware 10-12 Fluorescence observation in conventional microscope

More information

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION Olympus VS120-L100 Slide Scanner Standard Operating Procedure Startup 1) Red power bar switch (behind monitor) 2) Computer 3) Login: UserVS120 account (no password) 4) Double click: WAIT FOR INITIALIZATION

More information

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON N-SIM guide NIKON IMAGING CENTRE @ KING S COLLEGE LONDON Starting-up / Shut-down The NSIM hardware is calibrated after system warm-up occurs. It is recommended that you turn-on the system for at least

More information

Nikon E800 Operating Instructions.

Nikon E800 Operating Instructions. Nikon E800 Operating Instructions. You can request electronic copies of this manual by contacting lshats@jhsph.edu Copies are also available on the JHU MMI Department web site. Please send your comments

More information

1 Co Localization and Working flow with the lsm700

1 Co Localization and Working flow with the lsm700 1 Co Localization and Working flow with the lsm700 Samples -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ BrDU with alexa 488. -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ no BrDU (but with

More information

ScanArray Overview. Principle of Operation. Instrument Components

ScanArray Overview. Principle of Operation. Instrument Components ScanArray Overview The GSI Lumonics ScanArrayÒ Microarray Analysis System is a scanning laser confocal fluorescence microscope that is used to determine the fluorescence intensity of a two-dimensional

More information

Using the AmScope Microscope Cameras

Using the AmScope Microscope Cameras Using the AmScope Microscope Cameras Part 1 Setup. In order to use the camera, you will need: a) the camera system; b) a computer running the camera software. The camera system is contained in a Pelican

More information

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands Differential Interference Contrast (DIC) imaging is a technique used to increase contrast in brightfield images. In confocal systems,

More information

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09) University of Chicago Integrated Light Microscopy Core Dr. Vytas Bindokas, Director http://digital.bsd.uchicago.edu By: Christine Labno, Assistant Director Room: AB-129 Phone: 4-9040 Zeiss Axiovert 135

More information

CCAM Microscope Objectives

CCAM Microscope Objectives CCAM Microscope Objectives Things to consider when selecting an objective Magnification Numerical Aperture (NA) resolving power and light intensity of the objective Working Distance distance between the

More information

User Guide to the IBIF Leica TCS SP8 MP Confocal Microscope

User Guide to the IBIF Leica TCS SP8 MP Confocal Microscope User Guide to the IBIF Leica TCS SP8 MP Confocal Microscope This version: 7.24.14. Introduction The IBIF confocal microscope is made available on a fee-for-use-hour basis to all users who have been trained.

More information

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide Operation 58. UltraVIEW VoX Getting Started Guide Overview This chapter outlines the basic methods used to operate the UltraVIEW VoX system. About other software Volocity places great demands on the computer

More information

Olympus Confocal Microscope User Guide Last updated

Olympus Confocal Microscope User Guide Last updated 1 Olympus Confocal Microscope User Guide Last updated 11-22-2017 Contents: Laser Safety Training Requirement... 1 Mercury Lamp Precautions... 1 Biosafety requirements and rules for work in the MIC... 2

More information

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions Starting up: Schematic of Olympus BX-61. 1. Turn on Olympus microscope power box (left of microscope) with toggle switch on

More information

CCAM s Selection of. Zeiss Microscope Objectives

CCAM s Selection of. Zeiss Microscope Objectives CCAM s Selection of Zeiss Microscope Objectives 1. Magnification Image scale 2. Resolution The minimum separation distance between two points that are clearly resolved. The resolution of an objective is

More information

Brief manual how to start and close the Leica sp2 Confocal. (TCS SP2 AOBS system mounted on a DM IRE2)

Brief manual how to start and close the Leica sp2 Confocal. (TCS SP2 AOBS system mounted on a DM IRE2) Brief manual how to start and close the Leica sp2 Confocal (TCS SP2 AOBS system mounted on a DM IRE2) A. Switching on hardware B. Acquiring and saving images C. Switching off the microscope D. Good working

More information

Use of the HSW5 Spinning Disk Confocal Microscope Updated last May 25, 2010 OK

Use of the HSW5 Spinning Disk Confocal Microscope Updated last May 25, 2010 OK Use of the HSW5 Spinning Disk Confocal Microscope Updated last May 25, 2010 OK Getting Started: 2 Starting Micromanager and Loading a Configuration 3 The Main Micromanager GUI 3 Configuration Settings

More information