Zeiss LSM880 Operating Instructions. UTMB Optical Microscopy Core Jan. 16, 2018
|
|
- Blaze Short
- 5 years ago
- Views:
Transcription
1 Zeiss LSM880 Operating Instructions UTMB Optical Microscopy Core Jan. 16,
2 1. Power up the microscope Sing the LOGBOOK Steps below will provide power to the computer and all of the microscope components. Turn the MAIN SWITCH on Wait 2 s Turn the SYSTEMS/PC on (STOP HERE FOR COMPUTER USE ONLY) Wait 2 s Turn the COMPONENTS on 2
3 2. Open the front and top doors of the plexiglas incubator Unless live cell imaging is planned, open 4 doors (2 front and 2 top) to allow easy access to the microscope stage. 3
4 3. Start ZEN software After the windows has been loaded, login by clicking on the LSM User icon. No password is required. On the desktop, click on the ZEN Confocal (black) icon to start confocal acquisition software. (ZEN blue software can be used for controlling the Axiocam and image analysis.) Click on the Start System button. The software will start recognizing the motorized components of the microscope. 4
5 4. Set the objective lens to be used IN ZEN software: In the Locate tab navigate to the Microscope Control window. In that window click on the objective icon to bring up a list of the available objectives (5X, 10X, 20X : AIR; 40X: WATER immersion; 63X: OIL immersion). Select one. ALTERNATIVELY using the touchscreen: Click: Control > Objectives > 20x. When changing from air to immersion objective, or in reverse, a warning to change the immersion media is shown. Click OK, when done. 5
6 5. Apply immersion media IF required Only 63X and 40X objectives require immersion media! 63X Oil immersion objective (has a rubber band around it) 40X Water immersion objective Use Immersol 518 F only (n=1.518). Open the bottle and apply a drop of the immersion oil with the applicator cap to the top of the lens. Use distilled water only (n=1.330). Open the tube and apply a drop of water with a plastic pipette to the top of the lens*. REMEMBER TO WIPE CLEAN THESE OBJECTIVES WITH LENS PAPER ONLY AFTER EACH USE AND ALSO BEFORE USE. *For live-cell experiments at 37C, Immersol W, a viscous water based media must be used to prevent evaporation. Ask the OMC personnel for guidance. 6
7 6. Put objective into the load position First, make sure the objective is in its lowest position (load position). To put the objective into the load position: press on the load button on the right side of the microscope or on the load button on the touchscreen. OR 7
8 7. Place your sample onto the stage Illum. head Adaptor frame Make sure that the stage adaptor frame is not loose/shaky. Tilt the illumination head, if needed. Place your sample coverslip down onto the stage. Do not clamp your slide tight. Pull the illumination head back forward. Using the joystick that controls the motorized sample stage center your specimen above the objective lens. 8
9 8. Bring the immersion objective up IF USING AN AIR OBJECTIVE READ, BUT SKIP DOING THIS STEP! Using the coarse/fine focus knobs on the microscope body bring the objective up (clockwise direction) until the immersion media just touches the coverslip. 9
10 9. Bring your sample into focus In the Locate tab, depending on the fluorophores in your sample, click on any of the four buttons to turn the excitation light on: TL/DIC : white light for brightfield Blue: for near-uv excitation (DAPI, EBFP) Green: for ~ 488 nm excitation (Alexa 488/ EGFP) Red: for ~ 540 nm excitation (Rodamine, Alexa 568/594) The UV source illumination intensity may be changed in the Microscope Control window by clicking on the UV lamp icon and dragging the slider. While looking into the oculars, bring the objective slowly up until you can clearly see your specimen. (If you are using the 40X objective, and the image appears fuzzy. You may then need to adjust the objective s collar ring to the appropriate thickness of your coverslip (default is 0.17mm)). Using the stage joystick to move the sample around find a suitable imaging area. (Minimize sample exposure time to reduce fluorophore bleaching!). Click on Set Work position button on the touchscreen, if present. After finding the spot click on the ALL OFF button to turn the light off. 10
11 10. Load confocal acquisition settings A) If you used this microscope before and saved your images in the.czi format, you may reload your previous settings: At the top of ZEN, click open file. Locate your.czi image and open it. In the Dimensions tab, locate the Reuse button. Click it to load settings. Beware: if the objective in the light path does not match to the one used before, the pinhole setting will not be set properly. See the Info tab and adjust the pinhole manually. B) If you are imaging for the first time or have a new set of dyes, you may then choose from one of the pre-set imaging protocols. In the Acquisition tab, navigate to the Experiment Manager field, click on the open file icon to bring up a list of available pre-set imaging protocols. 11
12 The list has protocols for confocal and airyscan imaging modes. Airyscan mode is not part of this learning module. Confocal protocols start with the word confocal followed by a set of numbers that correspond to the excitation wavelengths and image scan mode (Line/Frame). Line vs. Frame mode: in each of these protocols, images of fluorophores are acquired sequentially by switching on and off lasers in different acquisition tracks. In the line mode, lasers are switched on/off after each image line. In this mode, no hardware movements are allowed (pinhole, emission and excitations settings must remain static for all channels, only lasers can be switched on and off). It is the fastest mode. In the frame mode, lasers are switched on/off after each full image scan. It is the fastest mode if hardware movements are needed (setting pinhole for each track/channel, overlapping detection ranges, etc.). Example: For a sample stained with DAPI (ex. 360 nm), Alexa Fluor 488 (ex. 488 nm) and MitoTracker Red (ex. 569 nm) one may choose either Confocal-561_488_405 Line Zeiss or Confocal-633_561_488_405 Line Zeiss protocol. If the last protocol is chosen, the 633 channel/track can be deactivated or removed after protocol loading. 12
13 After selecting and loading of a protocol, required lasers will be automatically powered on. The status of each laser can be seen in the Laser window. Argon laser needs 5 mins of warmup after start before a stable emission is achieved. THIS IS FOR EDUCATION ONLY, DO NOT CHANGE: In the Imaging Setup window, one can switch between imaging tracks by clicking on their respective buttons (561 is selected in the image) Track switching mode selector line/frame. The bottom of the Imaging Setup window shows the light path configuration for each selected track. 13
14 LIGTH PATH CONFIGURATION (THIS IS FOR EDUCATION ONLY): Currently selected imaging track (561) Delete current track Graphical representation of the activated laser lines (vertical line(s)), theoretical emission spectra of fluorophore(s), and emission wavelengths being collected (shaded area(s)). One can change emission collection by dragging/resizing the bottom slider. Shows activated detectors in the current track (checked), Dye (its selection updates emission graphics above), Color (selects palette to display the image), Detector s name (total of 3, but 1 is used in this track), Range (shows emission collection range shaded area in the above graphics). IMPORTANT: Ranges must not overlap between tracks if track line switching to be uses! Clicking on either Visible or Invisible Light icon on the right one can checkmark laser lines to be turned on during this track activation. Correspondingly, the correct MBS or main dichroic beam splitter on the left must be chosen that would reflect the selected laser light onto the sample. Optionally, one can checkmark the transmission PMT (T- PMT) to also collect brightfield images in this track. 14
15 DETECTORS (THIS IS FOR EDUCATION ONLY): Ch3 Ch2 GaAsP Ch1 The emission light from fluorophore(s) is directed onto the diffraction grating that linearly disperses the light according to its wavelengths (similar to a prism creating a ray of rainbow from sunlight). The resultant rainbow can be cut and bend with sliders and prisms, respectively, to direct its portions to separate photomultipliers (Ch1, Ch2 GaAsP, Ch3) for intensity/photon flux detection. Ch2 GaAsP is a special detector with enhanced sensitivity to the visible light. 15
16 11. Optimize your detection channels In the Channels window, uncheck all but one channel (start with the shortest excitation wavelength channel), highlight the checked channel (by clicking on the line) to update the detector settings below. Important: The goal of channel optimization is to ensure that most of the pixels in your final image are not black or saturated (e.g. for an 8-bit image [0 to 255 intensity gray values; 0 black and 255 white(saturated)]), to accurately quantify differences in fluorescence intensities. At the very beginning of the Acquisition tab click on the Live button to star image scanning from the selected channel. At the bottom of the imaging window, in the Dimensions tab, check the Range Indicator box. This will color all 0 intensity pixels (below detection threshold) in blue and 255 intensity pixels in red (saturated) in the live image. 16
17 In the Display tab click on the Reset button to ensure that you are adjusting your image to the full histogram [0-255]. If you are using the line mode of switching between tracks, you must adjust the pinhole for one track and keep it the same for others. Adjust the pinhole by clicking on the 1AU button in the Channels tab*. *Note: 1AU sets the pinhole diameter to 1 Airy Unit (AU) (diameter ~ ex. wavelength). 1AU is the best compromise between light efficiency collection and optical section thickness. For dim samples pinhole diameter can be set wider (e.g. 2xAU) to increase light collection with a compromise in the axial/lateral resolution). Adjusting channels using Laser Power, Master Gain and Digital Offset sliders: Laser power adjustment slider to brighten pixels (too much laser >10% bleaches sample quicker) Master gain adjustment slider to brighten pixels (keep it >500 but <950 to avoid detector noise) Digital offset adjustment slider (controls background/blue pixels) Too many blue pixels & low signal. Solution: 1. Focus better 2. Increase digital offset to eliminate blue pixels 3. Increase laser power or master gain (to brighten pixels up). Too many blue pixels & high signal. Solution: 1. Focus better 2. Increase digital offset to eliminate blue pixels 3. Decrease laser power or master gain (to desaturate pixels up). Well-adjusted channel: No blue, no red pixels. Pixels are well stretched in the histogram. 17
18 After a successful channel adjustment click on the Stop button at the top of the tab to stop the live scan. In Acquisition mode window: After adjusting all channels one by one, check all of the channels. Uncheck the adjusted channel, check and highlight the next channel to be adjusted. Repeat the adjustments as before, but remember not to change the pinhole! Set image size in pixel by pressing on the X*Y button, typically 1024 by 1024 is sufficient. (Using the Optimal button sets the image size automatically for best possible resolution.) Speed slider adjusts pixel exposure time : the lower the speed the better the signal to noise ratio. Keep it around 7 or 8. Set the number of lines to average for a better signal to noise performance (scans # of lines, calculates and shows the average). To start, keep at 2 lines. Select bit depth (the precision with which fluorescent intensity is determent), typically 8 or 12 bit will suffice. Select bidirectional scan for faster scanning. Using these sliders, one can optically zoom into the sample, shift and rotate the scan area. (Leave it) 18
19 12. Taking a snap Click on the Snap button to take a picture of every channel, which will be displayed as an overlay in the image window. If you do not like the current focus. Click on the Live button, refocus or move the stage, and then click on Snap again. Pixel distribution will appear well stretched between 0 and 255 gray intensity values in the histogram for channels (shown in different colors) that have been optimally adjusted. 13. Saving your images In the upper right corner, locate the Images and Documents window. The current image will be highlighted, click on the Diskette icon to save the image. When it has been saved, the new file name will appear and the exclamation sign will be gone. Save images in the.czi format to be able to reuse your setting later on. 19
20 14. Finding new imaging area Click on the Live button. While scanning your can use the Stage window to move the motorized stage position (in place of the joystick) using inner slices by small steps, outer slices by large steps. Find another area, re-focus and take another snap. Save your Image. Small step Large step 15. Taking a Z-stack of images To start acquiring a z-stack or images: at the top of the Acquisition tab check the Z-stack checkbox. This will activate the Z-Stack window at the bottom of the Acquisition tab. Scroll to the bottom, find the Z- Stack window, click and pull it out to any convenient place on the screen 20
21 In the Z-Stack window click on the First/Last tab. Click on the Live button to start scanning. While scanning, use the fine focus knob to focus on where you would like the image stack to start (such as bottom of your sample) and then, click on the Set Last bottom. Then using the same knob focus on where you would like the image stack to end (such as the top of your sample), click on the Set First button. Stop the live scan. OR After setting your scan range, now you have to set the number of slices to be acquired. For a 3D reconstruction later on: Keep Interval selected, then click on the optimal interval button (number). This sets the interval between the slices at the Nyquist s sampling frequency (50% overlap between adjacent slices) for all channels to be scanned. If no 3D reconstruction is planned, or the number of slices is too many to scan with the optimal interval, you can select Keep Slice and then type the number of slices you would like to acquire into the Slices field. 21
22 After finishing with the Z-stack setting, click on the Start Experiment button. The microscope will begin taking pictures of your image slices. After the scan is done Save your Z-stack the same way as image snaps, see page Microscope shut down 1. Exit ZEN by clicking on the X icon. The software will prompt you to turn the lasers off, select OFF next to every laser. 2. Copy your images to your USB storage device. 3. Clean the objectives and close the incubator doors. 4. Safely unplug your storage device from the computer. 5. Shut down the computer. 6. If 5 mins have not yet passed since turning off the lasers in the software, WAIT for 5 mins. Then, turn all the switches off Components > Systems/PC >Main Switch. 7. Sign the logbook. 22
23 Confocal Imaging Suggestions Channel adjustment must be performed for every series of samples stained under identical conditions. It should be done at the onset of imaging and then the settings to be reused during later imaging sessions for similar/same samples, provided that you do not change staining conditions. Channels need ideally to be adjusted to the brightest sample/area in the series. Alternatively, channels may be adjusted to 50-80% of the maximum gray value to leave room for a higher intensity signal to prevent saturation. Example: Sample set: BHK 21 cells treated with increasing concentrations of Bengamide B and untreated controls. Cells are fixed with 4% PFA, permeabilized, blocked and stained with a primary antibody against NF-kB followed by a conjugated secondary antibody. It is advisable, in addition to treated and untreated BHK 21 cells to have these controls for imaging: unstained BHK 21 cells, BHK 21 cells incubated with only secondary antibody, and if more than one target/fluorophore is used, BHK 21 cells stained only for one target/fluorophore. IMPORTANT: Any change in the staining protocol (fixation method or duration, non-specific blocker change, change in antibody dilutions or incubation times) may require a readjustment of imaging channels, which may prohibit pooling with the previous results. 23
Operating Instructions for Zeiss LSM 510
Operating Instructions for Zeiss LSM 510 Location: GNL 6.312q (BSL3) Questions? Contact: Maxim Ivannikov, maivanni@utmb.edu 1 Attend A Complementary Training Before Using The Microscope All future users
More informationLSM 710 Confocal Microscope Standard Operation Protocol
LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.
More informationLSM 780 Confocal Microscope Standard Operation Protocol
LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch
More informationZeiss 880 Training Notes Zen 2.3
Zeiss 880 Training Notes Zen 2.3 1 Turn on the HXP 120V Lamp 2 Turn on Main Power Switch Turn on the Systems PC Switch Turn on the Components Switch. 3 4 5 Turn on the PC and log into your account. Start
More informationLSM 800 Confocal Microscope Standard Operation Protocol
LSM 800 Confocal Microscope Standard Operation Protocol Turning on the system 1. Switch on the Main switch (labeled 1 and 2 ) mounted on the wall. 2. Turn the Laser Key (labeled 3 ) 90 clockwise for power
More informationTraining Guide for Carl Zeiss LSM 880 with AiryScan FAST
Training Guide for Carl Zeiss LSM 880 with AiryScan FAST ZEN 2.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2018) Power ON Routine 1 2 Turn ON Main Switch from the remote control
More informationZeiss 780 Training Notes
Zeiss 780 Training Notes Turn on Main Switch, System PC and Components Switches 780 Start up sequence Do you need the argon laser (458, 488, 514 nm lines)? Yes Turn on the laser s main power switch and
More informationMicroscopy from Carl Zeiss
Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path
More informationZeiss LSM 880 Protocol
Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.
More informationLSM 510 Training Notes
LSM 510 Training Notes Turning on the system Turn on the arc lamp, found on the bench top left of the microscope. This supplies light for epifluorescence for viewing your samples through the microscope.
More informationTraining Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope
Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Verify that main power switches on the
More informationTraining Guide for Leica SP8 Confocal/Multiphoton Microscope
Training Guide for Leica SP8 Confocal/Multiphoton Microscope LAS AF v3.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON power switch for epifluorescence
More informationZeiss LSM 780 Protocol
Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.
More informationTraining Guide for Carl Zeiss LSM 510 META Confocal Microscope
Training Guide for Carl Zeiss LSM 510 META Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON Components and System/PC switches
More informationPractical work no. 3: Confocal Live Cell Microscopy
Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside
More informationZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide
ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide Matyas Molnar, Biovis 2016 Starting the microscpe 1. Check the microscope if everything looks clean and normal. If not, report it in the logbook.
More informationMIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL
MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL START-UP On the Switchbox, turn both black switches to the ON position. Wait for the microscope to boot up completely (watch the screen on the side of the microscope).
More informationLeica SP8 TCS Users Manual
Version : 07/08/0 Leica SP8 TCS Users Manual Start up:. Turn the PC Microscope, Scanner Power, Laser Power, and the Laser Emission key to on (bottom right of desk).. Turn on the fluorescent lamp (top left
More informationQuick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center
Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue
More informationLSM 510 Meta Training Notes
LSM 510 Meta Training Notes Turning on the system Turn on X-Cite power supply. This supplies light for epifluorescence for viewing your samples through the microscope. Turn on the remote control switch.
More informationCell Biology and Bioimaging Core
Cell Biology and Bioimaging Core Leica TCS SP5 Operating Instructions Starting up the instrument 1. First, log in the log book located on the confocal desk. Include your name, your lab s PI, an account
More informationQuick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s
LSM 5 MP, LSM 510 and LSM 510 META M i c r o s c o p y f r o m C a r l Z e i s s Quick Guide Laser Scanning Microscopes LSM Software ZEN 2007 August 2007 We make it visible. Contents Page Contents... 1
More informationThings to check before start-up.
Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your
More informationZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL
ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL START THE SYSTEM... 2 START ZEN SOFTWARE... 3 SET THE TEMPERATURE AND THE CO2 CONTROLLERS... OBSERVATION AT OCULARS... 5 STATIF PRESENTATION... 6 ACQUIRE ONE
More informationZeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling
Zeiss LSM 510 Confocor III Training Notes Center for Cell Analysis & Modeling Confocor 3 Start Up Go to System Module Turn on Main Switch, System/ PC, and Components Switches Do you need the arc lamp?
More informationQuick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center
Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue
More informationThe Zeiss AiryScan System, Confocal Four.
The Zeiss AiryScan System, Confocal Four. Overview. The Zeiss AiryScan module is a segmented, radially stacked GaASP detector and collector system designed to subsample the airy disk of a point emission
More informationZEISS LSM510META confocal manual
ZEISS LSM510META confocal manual Switching on the system 1) Switch on the Remote Control button located on the table to the right of the microscope. This is the main switch for the whole system including
More informationContents. Introduction
Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path and lasers... 12 Scanning
More informationLeica TCS SP8 Quick Start Guide
Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, Fluorescent Light for the microscope stand. 2. Turn on the Scanner Power (1) on the front
More informationConfocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:
Confocal imaging on the Leica TCS SP8 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: 4) Do not touch the microscope while the software is initializing. Choose your options: Turn
More informationLeica TCS SP8 Quick Start Guide
Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, EL6000 fluorescent light source for the microscope stand. 2. Turn on the Scanner Power
More informationLeica SP8 TCS Users Manual
Leica SP8 TCS Users Manual Follow the procedure for start up and log on as posted in the lab. Please log on with your account only and do not share your password with anyone. We track and confirm usage
More informationTraining Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope
Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope ZEN 2009 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn Chameleon TiS laser key from Standby
More informationGuide to Confocal 5. Starting session
Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope
More informationQuick Start Guide. Leica SP5 X
Quick Start Guide Leica SP5 X Please note: Some of the information in this guide was taken from Leica Microsystems Leica TCS SP5 LAS AF Guide for New Users. This work is licensed under the Creative Commons
More informationSupplemental Method Information Zeiss LSM710
Supplemental Method Information Zeiss LSM710 1 Under the Light Path window set up the confocal for imaging a green dye (Alexa488-EGFP). For example, set up the light path as shown here using the 488 nm
More informationLeica SP8 Resonant Confocal. Quick-Start Guide
Leica SP8 Resonant Confocal Quick-Start Guide Contents Start-up Preparing for Imaging Part 1 On the scope Part 2 Software interface Part 3 Heat & CO2 incubation Part 4 Other hardware options Shut-down
More informationTRAINING MANUAL. Olympus FV1000
TRAINING MANUAL Olympus FV1000 September 2014 TABLE OF CONTENTS A. Start-Up Procedure... 1 B. Visual Observation under the Microscope... 1 C. Image Acquisition... 4 A brief Overview of the Settings...
More informationTitle: Leica SP5 Confocal User Manual
Title: Leica SP5 Confocal User Manual Date of first issue: 23/10/2015 Date of review: Version: Admin For assistance or to report an issue Office: CG07 or 05 Email: Igmm-imaginghelpdesk@igmm.ed.ac.uk Website:
More informationOperating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.
Smith College August 2010 Operating Checklist for using the Laser Scanning Confocal Microscope Leica TCS SP5. CONTENT, page no. Startup, 1 Initial set-up, 1 Software, 2 Microscope Specimen observation
More informationLEICA TCS SP5 AOBS TANDEM USER MANUAL
LEICA TCS SP5 AOBS TANDEM USER MANUAL STARTING THE SYSTEM...2 THE LAS AF SOFTWARE...3 THE «ACQUIRE» MENU...5 CHOOSE AND CREATE A SETTING...6 THE CONTROL PANEL...8 THE DMI6000B MICROSCOPE...10 ACQUIRE ONE
More informationNikon SIM-E & A1-R System
Nikon SIM-E & A1-R System USER GUIDE LSU Health Sciences Center Shreveport Research Core Facility June 01 2017 Chaowei Shang 1 Table of Content 1. Start Up the System... Page 3 Hardware and microscope
More informationLeica SPEII confocal microscope. Short Manual
Leica SPEII confocal microscope Short Manual Switching ON sequence: 1. Turn on the Workstation under the bench (top, far right). 2. Turn on the Supply Unit - Laser box (big green switch first and then
More informationMIF ZEISS LSM510 CONFOCAL USER PROTOCOL
MIF ZEISS LSM510 CONFOCAL USER PROTOCOL START-UP Turn on the Mercury Bulb Power Supply (if needed). Power-on the Control Box. Turn on the computer. Open the LSM 510 software. Choose Scan New Images and
More informationBi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging
Bi/BE 227 Winter 2016 Assignment #3 Adding the third dimension: 3D Confocal Imaging Schedule: Jan 20: Assignment Jan 20-Feb 8: Work on assignment Feb 10: Student PowerPoint presentations. Goals for this
More informationNikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope
Nikon Eclipse Ti A1-A Confocal Operating Manual Start-up 1. Turn on Excite Fluorescent light power supply- metal halide. a. Cool down as for mercury bulb b. Wheel closed liquid light guide 2. Turn on power
More informationNikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.
Nikon AZ100 Laser Scanning Macro Confocal Microscope Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin May 2017 Contents 1 Introduction 2 2 Hardware - Startup 2 3 Software/Operation 4 3.1 Multidimensional
More informationOperation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009
Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Introduction of Fluoresence Confocal Microscopy The first confocal microscope was invented by Princeton
More informationOPERATING INSTRUCTIONS
Zeiss LSM 510 M eta Confocal M icroscope OPERATING INSTRUCTIONS Starting the System: 1. Turn the black knob on the laser box one-quarter turn from Off to On. You will hear the laser cooling mechanisms
More informationLeica Sp5 II Confocal User Guide
Leica Sp5 II Confocal User Guide Turning on the Confocal System (instructions are posted in the room) 1. Turn on Laser Power Button 2. Turn Key to On position 3. Turn on Scanner Power Button 4. Turn on
More informationNikon C1si Spectral Laser Scanning Confocal Microscope. User Guide
Nikon C1si Spectral Laser Scanning Confocal Microscope User Guide Contents: C1Si Turn-On/ShutDown Procedures... 2 Overview... 4 Setup for epi-illumination to view through the eyepieces:... 5 Setup for
More informationTopics. - How to calibrate the LSM scanner. - How to clean the microscope. - How to adjust the pinhole alignment. - How to adjust the Collimator
Topics - How to calibrate the LSM scanner - How to measure the PSF - How to clean the microscope - How to adjust the pinhole alignment - How to adjust the Collimator How to calibrate the LSM scanner The
More informationZeiss LSM 510 Multiphoton Confocal Microscope
Zeiss LSM 510 Multiphoton Confocal Microscope Quick Start User Guide LSU Health Sciences Research Core Facility Table of Contents 1 Safety... Page 3 2 Turn On the System... Page 4 3 Start Up the ZEN Software.
More informationTraining Guide for Carl Zeiss AxioZoom V16 Stereo Microscope
Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope ZEN 2012 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 If you require fluorescence imaging,
More informationTRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO
TRAINING MANUAL Multiphoton Microscopy LSM 510 META-NLO September 2010 Multiphoton Microscopy Training Manual Multiphoton microscopy is only available on the LSM 510 META-NLO system. This system is equipped
More informationSupplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each
Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each lens with a 1 Airy unit pinhole setting. Many water lenses
More information3 Choose the Channels button and set the Channel Settings. Set the Pinhole to 1 Airy unit.
1 Set up the confocal light path for imaging a green dye (e.g. Alexa488-EGFP). For example, under the Configuration Control window the light path could be set up as shown here using the 488 nm LASER (found
More informationMicroscope Confocal LSM510 META
Microscope Confocal LSM510 META Welcome to the Zeiss LSM 510 Meta Confocal tutorial. Before using the LSM 510 META, Log off any other computer that is open with your personal login. You will need to put
More informationZeiss Deconvolution Microscope: A Quick Guide
Zeiss Deconvolution Microscope: A Quick Guide Start-up Uncover microscope. Do not put dust cover on the floor. Plug in both cameras. The default camera is the AxioCam HRm (monochrome camera) for fluorescence
More information1 Co Localization and Working flow with the lsm700
1 Co Localization and Working flow with the lsm700 Samples -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ BrDU with alexa 488. -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ no BrDU (but with
More information1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the
1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the light path as shown here using the 488 nm LASER (Laser Unit 1) reflecting off of the 405/488 nm Dichroic mirror
More informationZeiss LSM 510 Multiphoton Confocal Microscope
Zeiss LSM 510 Multiphoton Confocal Microscope User Guide LSU Health Sciences Center-Shreveport Research Core Facility Table of Contents 1 Safety... Page 3 2 Turn On the System... Page 4 3 Start Up the
More informationb. Turn the power switch and key to on position for blue laser.
OLYMPUS FLUOVIEW 300 CONFOCAL MICOSCOPE OPERATION PROCEDURE 1. Turn ON microscope in this order: 1) Turn on mercury lamp (Note: once the mercury lamp is turned off, DO NOT turn it back on for at least
More informationRenishaw InVia Raman microscope
Laser Spectroscopy Labs Renishaw InVia Raman microscope Operation instructions 1. Turn On the power switch, system power switch is located towards the back of the system on the right hand side. Wait ~10
More informationInternal Medicine Imaging Core Emory University Department of Medicine
Internal Medicine Imaging Core Emory University Department of Medicine 1 OPERATION OF THE ZEISS LSM 510 META YOU MUST SIGN UP TO USE THE MICROSCOPE OR COMPUTER EVERY TIME NO EXCEPTIONS Before attempting
More informationSPINNING DISK CSU-X1 USER MANUAL
SPINNING DISK CSU-X1 USER MANUAL Starting the temperature controller... 2 Starting the CO2 controller... 3 Start the spinning disk... 4 Sample observation with the oculars... 5 Spatial sampling, Pixel
More informationDiskovery Spinning Disk Guide
Diskovery Spinning Disk Guide qbi.microscopy@uq.edu.au Getting started The microscope and its peripherals (Fig. 1a) should always be turned on, but if they are not, turn them on in the following way: 1.
More informationHoriba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer
Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and
More informationContents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING
Operations Guide Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Nikon Eclipse 90i Operations Guide STARTUP Startup Powering Up Fluorescence
More informationCONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2)
Wellcome Trust Centre for Human Genetics Molecular Cytogenetics and Microscopy Core CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2) 1) STARTING THE SYSTEM Abridged INSTRUCTIONS Switch on the mercury bulb
More informationSHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014
CIAN LSM1 or LSM2 short instructions, version 1.4, September 2014 page 1 of 6 SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 Before starting To work with LSM1
More informationUsing the Nikon TE2000 Inverted Microscope
Wellcome Trust Centre for Human Genetics Molecular Cytogenetics and Microscopy Core Using the Nikon TE2000 Inverted Microscope Fluorescence image acquisition using Scanalytic s IPLab software and the B&W
More informationQUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Imperial College London
Imperial College London Facility for Imaging by Light Microscopy QUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Observing Life As It Happens
More informationNikon A1R. Multi-Photon & Laser Scanning Confocal Microscope. Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin. April 2017.
Nikon A1R Multi-Photon & Laser Scanning Confocal Microscope Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin April 2017 Contents 1 Introduction 2 2 Start-Up 2 3 Imaging 4 3.1 Sample Alignment...........................................
More information3. are adherent cells (ie. cells in suspension are too far away from the coverslip)
Before you begin, make sure your sample... 1. is seeded on #1.5 coverglass (thickness = 0.17) 2. is an aqueous solution (ie. fixed samples mounted on a slide will not work - not enough difference in refractive
More informationQUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Imperial College London
Imperial College London Facility for Imaging by Light Microscopy QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Observing Life As It Happens Startup procedure...
More informationRENISHAW INVIA RAMAN SPECTROMETER
STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw invia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples
More informationConfocal 510 Tutorial
Confocal 510 Tutorial You will have to log on in order to use the microscope. You will be charged according the time you are logged in, so please don t forget to log out after you are done. If you don
More informationTravel to New Dimensions- LSM 880. The Resolution of a Microscope is limited. The Resolution of a Microscope is limited. Image. Image. Object.
Travel to New Dimensions- LSM 880 LSM 880: The Power of Sensitivity Our Latest Member of the LSM 880 with GaAsP Detectors Sensitivity, and Ease of Use Innovative High-End Laser Scanning Microscopes from
More informationZeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)
University of Chicago Integrated Light Microscopy Core Dr. Vytas Bindokas, Director http://digital.bsd.uchicago.edu By: Christine Labno, Assistant Director Room: AB-129 Phone: 4-9040 Zeiss Axiovert 135
More informationBX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions
BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions Starting up: Schematic of Olympus BX-61. 1. Turn on Olympus microscope power box (left of microscope) with toggle switch on
More informationREMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data.
1 Use of the Zeiss LSM 510 Inverted Firstly please be aware that this microscope should be treated with respect and care at all times. Rules of use: This Microscope can only be used by Masters by Research
More informationImaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal
Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal Todays Goal: Introduce some additional functionalities of the Leica SP8 confocal HyD vs. PMT detectors Dye Assistant Scanning By
More informationBoulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement
Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement CONFOCAL MICROSCOPY BioVis Uppsala, 2017 Jeremy Adler Matyas Molnar Dirk Pacholsky Widefield & Confocal Microscopy
More informationOlympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope
Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope Improved Optics More Lasers 405 diode 440 diode 488 Argon 515 Argon 559
More informationZeiss Axio Imager.A1 manual
Zeiss Axio Imager.A1 manual Power-up protocol 1. Mercury lamp 2. Power strip on shelf 3. Computer The Mercury lamp should always be first-on and last-off. This prevents any electrical surges caused by
More informationOverview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide
Operation 58. UltraVIEW VoX Getting Started Guide Overview This chapter outlines the basic methods used to operate the UltraVIEW VoX system. About other software Volocity places great demands on the computer
More informationZEN 2012 SP5 black edition Hotfix 12
Information about the software ZEN 2012 SP5 black edition Hotfix 12 Software name: ZEN 2012 Service Pack 5 black edition Hotfix 12 Software version: The software version in ZEN Help About changes to 14.0.12.201
More informationWhy and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005
Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev Why use confocal microscopy? Principles of the laser scanning confocal microscope. Image resolution. Manipulating the
More informationConfocal Laser Scanning Microscopy
Name of the Core Facility: Confocal Laser Scanning Microscopy CORE Forschungszentrum Immunologie Mainz Welcome to the CSLM Core Facility: The CLSM Core Facility enables working groups to incorporate high
More informationWidefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software
Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software September 2007 Check website for most current User Guide
More informationBasics of confocal imaging (part I)
Basics of confocal imaging (part I) Swiss Institute of Technology (EPFL) Faculty of Life Sciences Head of BIOIMAGING AND OPTICS BIOP arne.seitz@epfl.ch Lateral resolution BioImaging &Optics Platform Light
More informationOlympus Digital Microscope Camera (DP70) checklist
Smith College - July 2005 Olympus Digital Microscope Camera (DP70) checklist CONTENT, page no. Camera Information, 1 Startup, 1 Retrieve an Image, 2 Microscope Setup, 2 Capture, 3 Preview. 3 Color Balans,
More informationNikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol
Nikon Eclipse Ti-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol Please sign on the log sheet before switching on system. Turn on system Turn on A only if confocal mode or laser
More informationSimplified Instructions: Olympus Widefield Microscope S1230
Contents General Microscope Operation Simple Image Capture Multi-Wavelength Capture Z-Series Timelapse Combining Capture Modes Synopsis of Other Functions Pages 2-23 24-40 41-47 48-56 57-59 60-68 69-83
More informationLeica TCS SL Confocal Training. Neuroscience Imaging Core Staff. Core Director. Facility Manager
Leica TCS SL Confocal Training Neuroscience Imaging Facility The Ohio State University Rightmire Hall 614-292-1367 Staff Core Director Anthony Brown, Ph. D. 060 Rightmire Hall 614-292-1205 brown.2302@osu.edu
More informationUser Guide to the IBIF Leica TCS SP8 MP Confocal Microscope
User Guide to the IBIF Leica TCS SP8 MP Confocal Microscope This version: 7.24.14. Introduction The IBIF confocal microscope is made available on a fee-for-use-hour basis to all users who have been trained.
More informationANSWER KEY Lab 2 (IGB): Bright Field and Fluorescence Optical Microscopy and Sectioning
Phys598BP Spring 2016 University of Illinois at Urbana-Champaign ANSWER KEY Lab 2 (IGB): Bright Field and Fluorescence Optical Microscopy and Sectioning Location: IGB Core Microscopy Facility Microscope:
More informationDIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands
DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands Differential Interference Contrast (DIC) imaging is a technique used to increase contrast in brightfield images. In confocal systems,
More information