1 Olympus VS120-L100 Slide Scanner Standard Operating Procedure Startup 1) Red power bar switch (behind monitor) 2) Computer 3) Login: UserVS120 account (no password) 4) Double click: WAIT FOR INITIALIZATION TO COMPLETE Shutdown 1) Transfer data to external drive and delete from computer 2) Remove ALL slides from cassettes 3) Close VS-ASW FL software 4) Shut down computer 5) Switch off power bar MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION CONTENTS BRIGHTFIELD (BF) SLIDE SCANNING SINGLE SLIDE... 2 FLUORESCENCE (FL) slide scanning Single Slide... 4 BRIGHTFIELD (BF) slide scanning Batch Scans... 7 FLUORESCENCE (FL) slide scanning Batch Scans... 9 Troubleshooting FILTER SPECS Multi band (Switch channels before moving XY) Single band (Move XY before switching channels) DAPI-FITC-TRITC-Cy5 CFP-YFP-RFP** DAPI-FITC-TRITC-Cy5 Ex (nm) Em (nm) Ex (nm) Em (nm) Ex (nm) Em (nm) DAPI 391/32 435/30 CFP 436/28 474/24 DAPI 375/28 460/50 FITC 479/33 520/30 YFP 506/21 540/24 FITC 480/30 535/40 TRITC 554/24 594/32 RFP 578/24 642/80 TRITC 540/25 605/55 Cy5 638/31 694/60 Cy5 638/31 694/60 Notes: ** Filter change needed, please contact AMP staff (If you require a filter set suited to mcherry/texas Red, use the CFP/YFP/RFP multiband) Page 1

2 MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION BRIGHTFIELD (BF) SLIDE SCANNING SINGLE SLIDE 1) a. For first time users, click on the Scan Projects tab and choose VS120 Brightfield project. This will populate the parameters with the most common settings (20x detailed scan, VSI file format). You can modify settings during the scan setup. b. For all other users, you may apply your own Scan Project as above or click on: under Single Slide Scan column. 2) If you have more than one slide, carefully load them into the cassettes as instructed by staff. 3) Choose Manual Load or Automatic Load. If Manual Load is chosen, CAREFULLY lift stage access cover NO MORE THAN 90 DEGREES OR YOU WILL HIT THE CAMERA. Pull back round tab and place slide sideways/horizontal with label on the LEFT. 4) If Automatic Load is chosen, the sensor will automatically detect where your slides are loaded in the cassettes. Once the sensor has detected the slide positions, choose the slide you wish to scan. 5) Fill out Slide Properties left column as needed. 6) Set naming and saving parameters in Automatic Naming and Saving middle column. 7) Click Overview 8) Once overview image is completed, choose objective magnification for detailed scan. Z-Scan Mode should be set to Scan one Z-plane. If you require more than one Z-plane, please see a member of staff. 9) Click Edit Scan Area 10) Under Scan Area in the left column. You can either click on Create Scan Area button to allow the program to automatically detect and draw Scan Regions around your samples or you can choose `Remove All Scan Areas` to draw your own. Use mouse wheel to zoom. 11) Under Detailed Scan Settings, make sure Scan whole area and focus sample only (middle button) option is selected. 12) Next, click on to put down a focus map. Choose the Focus Grid Density in the left column. You can manually add more Focus points with a mouse click, move existing focus points or select and delete those not required. NOTE: The more focus points you put down the longer the scan will take. Flat (fold and bubble free) samples with good contrast do not need a large amount of focus points while samples with less contrast and/or uneven surfaces require more points. 13) When done click on 14) If ready to proceed, click on Scan Now. If you require more focusing options, click on Next Page 2

3 a. In the Focus Settings menu, you may choose the Focus Logic. Options are Automatic (Default), Semi-Automatic (user confirms automatic focusing at each point) and Manual (user manually sets each focus point). b. You can now click on the Scan Now button. 15) Once completed you have several options: a. You may Add another Detailed Scan to the existing slide b. Scan a new slide with the current parameters c. Remove/Unload current slide d. Save Project (HIGHLY RECOMMENDED*) e. Or finish the experiment. *To preserve your scan parameters, it is highly recommended that you Save your Project so you do not have to worry about accidentally using a previous users settings. 16) If not saving a projecty, Click Finish and save your scans if you have not done so. 17) To remove the slide, click on and choose Manual Load. Page 3

4 MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION FLUORESCENCE (FL) slide scanning Single Slide 1) a. For first time users, click on the Scan Projects tab and choose one of the following options to populate the parameters with the most common settings (20x detailed scan, VSI file format, typical exposure etc.). You can modify settings during the scan setup. i. VS120 4 Channel Single Band. This option uses separate discrete filters for each channel and moves the stage in XY before switching channels. Less crosstalk but slower. (See page 1 for filter specs) ii. VS120 4 Channel Multi Band. This option uses a single multiband filter for collecting 4 emission bands simultaneously. Best for speed but more crosstalk. (See page 1 for filter specs) iii. VS120 3 Channel FP Multiband. This option uses a single multiband filter for collecting 3 fluorescent protein (CFP/YFP/RFP) emission bands simultaneously. Best for speed but more crosstalk. FILTER CHANGE NEEDED SEE MEMBER OF STAFF. (See page 1 for filter specs) b. For all other users, you may apply your own Scan Project as above or click on: under Single Slide Scan column. 2) If you have more than one slide, carefully load them into the cassettes as instructed by staff. 3) Choose Manual Load or Automatic Load. If Manual Load is chosen, CAREFULLY lift stage access cover NO MORE THAN 90 DEGREES OR YOU WILL HIT THE CAMERA. Pull back round tab and place slide sideways/horizontal with label on the LEFT. 4) If Automatic Load is chosen, the sensor will automatically detect where your slides are loaded in the cassettes. Once the sensor has detected the slide positions, choose the slide you wish to scan. 5) Fill out Slide Properties left column as needed. 6) Set naming and saving parameters in Automatic Naming and Saving middle column. 7) Click Next 8) Under Overview Type, choose the method you want to collect your overview. a. BF (Brightfield): Quickest method but only if your sample has enough contrast. Click drop down arrow to choose Normal, Dim or Faint Sample to correspond to level of contrast. b. Tissue: To scan a fluorescence overview image on tissue samples. Click drop down arrow to choose filter set. c. Cells: To scan a fluorescence overview image on tissue samples. Click drop down arrow to choose filter set. 9) Choose objective lens for overview image (2x is usually sufficient). 10) If you chose option (a) or (b) above, click Start Live button under Exposure Time column to set exposure. Use joystick to navigate to sample. 11) Click Overview button. 12) Once overview image is completed, choose objective magnification for detailed scan. Z-Scan Mode should be set to Scan one Z-plane. If you require more than one Z-plane, please see a member of staff. Click Next button. Page 4

5 13) Click Edit Scan Area 14) Under Scan Area in the left column. You can either click on Create Scan Area button to allow the program to automatically detect and draw Scan Regions around your samples or you can choose `Remove All Scan Areas` to draw your own. Use mouse wheel to zoom. 15) Under Detailed Scan Settings, make sure Scan whole area and focus sample only (middle button) option is selected. 16) Next, click on to put down a focus map. Choose the Focus Grid Density in the left column. You can manually add more Focus points with a mouse click, move existing focus points or select and delete those not required. NOTE: The more focus points you put down the longer the scan will take. Flat (fold and bubble free) samples with good contrast do not need a large amount of focus points while samples with less contrast and/or uneven surfaces require more points. 17) Click Next. 18) If you chose one of the preset Scan Projects, the list of filters under Channel Settings should be filled with the default choices. 19) To add a channel, click and choose to add a single or multiband filter set. To remove a channel, click. 20) Select Channel 1 (this is the channel that will be used for focusing so it should be your brightest and most widespread) and click 21) In the Stage Navigator window in the far right, make sure the yellow square is in the middle of your Detailed Scan area. If not, click inside your region of interest to move the stage. 22) Refer to the histogram in the top right window. Change the Exposure Time so that you achieve at least 1500 counts in the mean intensity reading. The Orca Flash 4.0 fluorescence camera is a 16-bit camera, allowing for a high dynamic range (65536 bit depth). It is recommended to use as much of this range as possible (without excessive exposure times causing photo-bleaching) to capture as much data as possible for potential post-scan analyses. 23) If ready to proceed, click on Scan Now. If you require more focusing options, click on Next 24) In the Focus Settings menu, you may choose the Focus Logic. Options are Automatic (Default), Semi- Automatic (user confirms automatic focusing at each point) and Manual (user manually sets each focus point). 25) You can now click on the Scan Now. Page 5

6 26) Once completed you have several options: a. You may Add another Detailed Scan to the existing slide b. Scan a new slide with the current parameters c. Remove/Unload current slide d. Save Project (HIGHLY RECOMMENDED*) e. Or finish the experiment. *To preserve your scan parameters, it is highly recommended that you Save your Project so you do not have to worry about accidentally using a previous users settings. 27) If not saving a project, Click Finish and save your scans if you have not done so. 28) To remove the slide, click on and choose Manual Load. Page 6

7 MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION BRIGHTFIELD (BF) slide scanning Batch Scans 1) Carefully load slides into the cassettes starting from cassette #1 (closest to you) before populating cassette #2 (furthest from you). Take note of your slide descriptions and positions in the cassettes for your own references. Carefully return cassettes back into the machine in the same order, ensuring that they are seated correctly. 2) To scan multiple brightfield slides, click on: under Batch Scan column. A sensor will now detect the positions of your slides within the cassettes. Once complete the next screen shows the positions of your slides in the cassettes under the Slide Loader column. Use the dropdown menu to choose between cassette #1 and #2. 3) Highlight each slide and fill out its properties in the middle column Slide Properties then click Next when done. A template can be used to pre-fill these properties. Talk to staff if needed. 4) Under the Slide Loader column, you are now able to load one or more previously saved Scan Project(s) or use the provided VS120 Brightfield project for typical parameters. 5) In the Scan Settings column and Magnification and Image Document tab, select desired magnification and Z-scan mode. 6) a) Click the Apply settings to all slides of all cassettes button to save these settings to every slide in every cassette. All slides in the loader should now be highlighted green. b) If you only want to scan a selection of your slides, hold the Ctrl-key and click the positions in the cassette to select only the slides you wish to scan. Once the selection is complete, repeat step 5) and click the Apply settings to selection button to save settings to just the selected slides. The selected slides in the loader should now be highlighted green. c) If you want to scan different slide selections with different parameters/projects, repeat steps 4 6 for each slide selection. 7) Set naming and saving parameters in Automatic Naming and Saving column and click Overview. 8) If for whatever reason you need to scan a single slide NOT loaded in the cassettes, click on Priority Scan to interrupt the batch process. 9) Once overviews are completed, you can review each slide in the Batch Slide Selection column. With a slide selected, click on Edit Scan Area button. 10) Under Scan Area in the left column. You can either click on Create Scan Area button to allow the program to automatically detect and draw Scan Regions around your samples or you can choose `Remove All Scan Areas` to draw your own. Use mouse wheel to zoom. 11) Under Detailed Scan Settings, make sure Scan whole area and focus sample only (middle button) option is selected. Page 7

8 12) Next, click on to put down a focus map. Choose the Focus Grid Density in the left column. You can manually add more Focus points with a mouse click, move existing focus points or select and delete those not required. NOTE: The more focus points you put down the longer the scan will take. Flat (fold and bubble free) samples with good contrast do not need a large amount of focus points while samples with less contrast and/or uneven surfaces require more points. 13) When done click on 14) Select next slide in the Batch Slide Selection column and repeat steps ) If you do not wish to scan a particular slide(s) in the batch, select the slides in the list and click on Exclude slides from the batch scan button 16) If ready to proceed, click on Scan Now. to exclude from batch scan. Note: If you need to scan a single slide NOT loaded in the cassettes, click on Priority Scan to interrupt the batch process. Page 8

9 MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION FLUORESCENCE (FL) slide scanning Batch Scans 1) Carefully load slides into the cassettes starting from cassette #1 (closest to you) before populating cassette #2 (furthest from you). Take note of your slide descriptions and positions in the cassettes for your own references. Carefully return cassettes back into the machine in the same order, ensuring that they are seated correctly. 2) To scan multiple fluorescence slides, click on: under Batch Scan column. A sensor will now detect the positions of your slides within the cassettes. Once complete the next screen shows the positions of your slides in the cassettes under the Slide Loader column. Use the dropdown menu to choose between cassette #1 and #2. 3) Highlight each slide and fill out its properties in the middle column Slide Properties then click Next when done. A template can be used to pre-fill these properties. Talk to staff if needed. 4) Under the Slide Loader column, you are now able to load one or more previously saved Scan Project(s) or use the provided default projects (outlined on Page 4-1a) for typical parameters. 5) In the Scan Settings column and Magnification and Image Document tab, select desired magnification and Z-scan mode. You may also modify the Overview Settings, Illumination Settings and Focus Settings tabs. However, you will have a chance to modify these settings later. 6) a) Click the Apply settings to all slides of all cassettes button to save these settings to every slide in every cassette. All slides in the loader should now be highlighted green. b) If you only want to scan a selection of your slides, hold the Ctrl-key and click the positions in the cassette to select only the slides you wish to scan. Once the selection is complete, repeat step 5) and click the Apply settings to selection button to save settings to just the selected slides. The selected slides in the loader should now be highlighted green. c) If you want to scan different slide selections with different parameters/projects, repeat steps 4 6 for each slide selection. 7) Set naming and saving parameters in Automatic Naming and Saving column and click Overview. 8) Once overviews are completed, you can review each slide in the Batch Slide Selection column. With a slide selected, click on Edit Scan Area button. 9) Under Scan Area in the left column. You can either click on Create Scan Area button to allow the program to automatically detect and draw Scan Regions around your samples or you can choose `Remove All Scan Areas` to draw your own. Use mouse wheel to zoom. 10) Under Detailed Scan Settings, make sure Scan whole area and focus sample only (middle button) option is selected. Page 9

10 11) Next, click on to put down a focus map. Choose the Focus Grid Density in the left column. You can manually add more Focus points with a mouse click, move existing focus points or select and delete those not required. NOTE: The more focus points you put down the longer the scan will take. Flat (fold and bubble free) samples with good contrast do not need a large amount of focus points while samples with less contrast and/or uneven surfaces require more points. 12) When done click on 13) Select next slide in the Batch Slide Selection column and repeat steps ) In the Scan Settings column, set the parameters required for the Magnification and Image Document tab, Illumination Settings tab and Focus Settings tab if required. 15) If you do not wish to scan a particular slide(s) in the batch, select the slides in the list and click on Exclude slides from the batch scan button 16) If ready to proceed, click on Scan Now. to exclude from batch scan. Note: If you need to scan a single slide NOT loaded in the cassettes, click on Priority Scan to interrupt the batch process. Page 10

11 Troubleshooting Sample not recognized For weakly stained samples with poor contrast (for example, brightfield slides with weakly staining dyes, very sparse stained areas OR brightfield overviews of fluorescently stained samples). During the editing scans dialog you will see the following settings in the left column: Figure 1. To increase the threshold of sample detection, modify the Sample detection sensitivity slider. In the example below (Figure 2) with the slider at 0, you see only a small percentage of the spinal cord tissue sections detected by the system as indicated by the white, highlighted background. With the slider set to +10 (Figure 3), almost all the sections are highlighted. Sample detection is also important when creating focus maps on scan areas. Under Detailed Scan Settings, if the Scan and focus sample only or Scan whole area and focus sample only (first and second button from left) is selected. Focus maps will only be placed on samples that have been detected (as described above). If your sample is still too faint to be detected, you will have to choose the 3 rd button Scan and focus whole area after manually drawing the scan area and/or manually add focus points. There are various options in the Sample Detection Settings and Advanced Sample Detection Settings menu to assist in fine tuning detection of your samples. When your samples are successfully highlighted, you can now click the Create Scan Area button which will automatically put a scan area around the detected sections (Figure 4). This is especially useful if you have many sections on the one slide to avoid manually drawing scan areas. Page 11

12 Figure 2. Brightfield overview image of fluorescently stained mouse spinal cord cross sections, sample detection sensitivity set to Zero. Figure 3. As above with sample detection sensitivity set to +10. Figure 4. As above after clicking on Create Sample Area button. Page 12

13 If you choose Brightfield overview scan for your fluorescent sample, there are various options in the drop down arrow menu. Choosing BF (brightfield) you have a choice of Normal, Dim and Faint samples. Depending on the choice, the program will manipulate the brightness and gamma to improve the contrast of the samples on your slide, giving you a greater chance of sample detection (mentioned above). BF Normal Samples BF Dim Samples BF Faint Samples Page 13

14 Focusing Issues After setting focus maps on your samples and clicking Scan Now, the system will go to each focus point to set the focus. If the focus is found, the focus point square will change colour to green. If the focus could not be found, the focus point square will change to grey. The scan will proceed if there is at least ONE green focus point square. Various focusing issues and possible solutions: 1. Samples are weakly stained (either brightfield or fluorescent). o In the case of weakly stained brightfield samples, go into Live mode before proceeding with the scan to manually set exposure times. o In the case of weak fluorescent signals, increase the exposure time of the focusing channel (channel on the top of the list) so the mean intensity on the histogram is at least 1500 counts (Page 5). Always choose the strongest and most widespread fluorescent channel to be the focusing channel. 2. Samples are too thick or too thin. o Thin samples <5 microns will have a smaller volume of fluorescent signal. Ensure that the stage is close to the focus by going into Live mode before proceeding with the scan. o Thick samples >50 microns will have many points of focus. The algorithm will usually choose the top of the sample. 3. Stage is too far away from sample or focusing algorithm gets lost. o Pertaining to both points 1 and 2, double check in Live mode to make sure the stage is at least close to the focus of your sample. o Avoid putting focus maps on empty areas where there is no sample. Focus maps on a glass slide can often bring the stage far from the focus. Page 14

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope ZEN 2012 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 If you require fluorescence imaging,

More information

Using Olympus dotslide for polarising microscopy

Using Olympus dotslide for polarising microscopy Using Olympus dotslide for polarising microscopy Background Because the dotslide system is built using a conventional microscope frame it lends itself to acquiring scans in different imaging modes whose

More information

Axioscan - Startup. 1. Turn on the Axioscan (button to the left) and turn on the computer. 2. Log on and start the ZEN Blue software from the desktop

Axioscan - Startup. 1. Turn on the Axioscan (button to the left) and turn on the computer. 2. Log on and start the ZEN Blue software from the desktop Axioscan - Startup 1. Turn on the Axioscan (button to the left) and turn on the computer 2. Log on and start the ZEN Blue software from the desktop 3. Press ZEN slidescan and Start System 4. Start by changing

More information

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017. Nikon AZ100 Laser Scanning Macro Confocal Microscope Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin May 2017 Contents 1 Introduction 2 2 Hardware - Startup 2 3 Software/Operation 4 3.1 Multidimensional

More information

Center for Microscopy and Image Analysis Axio Scan.Z1 Operating Manual

Center for Microscopy and Image Analysis Axio Scan.Z1 Operating Manual No index entries found. Center for Microscopy and Image Analysis Axio Scan.Z1 Operating Manual Table of contents 1. Starting procedure... 3 1.1. Turn on hardware... 3 1.2. Starting ZEN blue... 4 2. Load

More information

Title: Leica SP5 Confocal User Manual

Title: Leica SP5 Confocal User Manual Title: Leica SP5 Confocal User Manual Date of first issue: 23/10/2015 Date of review: Version: Admin For assistance or to report an issue Office: CG07 or 05 Email: Igmm-imaginghelpdesk@igmm.ed.ac.uk Website:

More information

Zeiss AxioImager.Z2 Brightfield Protocol

Zeiss AxioImager.Z2 Brightfield Protocol Zeiss AxioImager.Z2 Brightfield Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge

More information

Using the Nikon TE2000 Inverted Microscope

Using the Nikon TE2000 Inverted Microscope Wellcome Trust Centre for Human Genetics Molecular Cytogenetics and Microscopy Core Using the Nikon TE2000 Inverted Microscope Fluorescence image acquisition using Scanalytic s IPLab software and the B&W

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Version : 07/08/0 Leica SP8 TCS Users Manual Start up:. Turn the PC Microscope, Scanner Power, Laser Power, and the Laser Emission key to on (bottom right of desk).. Turn on the fluorescent lamp (top left

More information

ZEISS LSM510META confocal manual

ZEISS LSM510META confocal manual ZEISS LSM510META confocal manual Switching on the system 1) Switch on the Remote Control button located on the table to the right of the microscope. This is the main switch for the whole system including

More information

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope Improved Optics More Lasers 405 diode 440 diode 488 Argon 515 Argon 559

More information

Leica SPEII confocal microscope. Short Manual

Leica SPEII confocal microscope. Short Manual Leica SPEII confocal microscope Short Manual Switching ON sequence: 1. Turn on the Workstation under the bench (top, far right). 2. Turn on the Supply Unit - Laser box (big green switch first and then

More information

Zeiss 880 Training Notes Zen 2.3

Zeiss 880 Training Notes Zen 2.3 Zeiss 880 Training Notes Zen 2.3 1 Turn on the HXP 120V Lamp 2 Turn on Main Power Switch Turn on the Systems PC Switch Turn on the Components Switch. 3 4 5 Turn on the PC and log into your account. Start

More information

Nasmyth Ultraview Vox User Protocol

Nasmyth Ultraview Vox User Protocol Nasmyth Ultraview Vox User Protocol Switch on all wall sockets labelled Nasmyth, switch camera on (power supply located on table behind monitor), switch on laser switch in laser rack, switch computer on

More information

Everest System / Slidebook Operating Procedures

Everest System / Slidebook Operating Procedures Everest System / Slidebook Operating Procedures NOTICE: This guide is meant to supplement training, not replace it. All users must be trained first hand by a core employee. Training of others in your lab

More information

Simplified Instructions: Olympus Widefield Microscope S1230

Simplified Instructions: Olympus Widefield Microscope S1230 Contents General Microscope Operation Simple Image Capture Multi-Wavelength Capture Z-Series Timelapse Combining Capture Modes Synopsis of Other Functions Pages 2-23 24-40 41-47 48-56 57-59 60-68 69-83

More information

Zeiss Axio Imager.A1 manual

Zeiss Axio Imager.A1 manual Zeiss Axio Imager.A1 manual Power-up protocol 1. Mercury lamp 2. Power strip on shelf 3. Computer The Mercury lamp should always be first-on and last-off. This prevents any electrical surges caused by

More information

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software September 2007 Check website for most current User Guide

More information

Quick Start Guide. Leica SP5 X

Quick Start Guide. Leica SP5 X Quick Start Guide Leica SP5 X Please note: Some of the information in this guide was taken from Leica Microsystems Leica TCS SP5 LAS AF Guide for New Users. This work is licensed under the Creative Commons

More information

Nikon SIM-E & A1-R System

Nikon SIM-E & A1-R System Nikon SIM-E & A1-R System USER GUIDE LSU Health Sciences Center Shreveport Research Core Facility June 01 2017 Chaowei Shang 1 Table of Content 1. Start Up the System... Page 3 Hardware and microscope

More information

EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY

EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY TURN ON THE FOLLOWING EQUIPMENT The fluorescent light (if needed) The power strip for the microscope and accessories The CoolSNAP HQ camera on the right (Turn

More information

Operating Instructions for Zeiss LSM 510

Operating Instructions for Zeiss LSM 510 Operating Instructions for Zeiss LSM 510 Location: GNL 6.312q (BSL3) Questions? Contact: Maxim Ivannikov, maivanni@utmb.edu 1 Attend A Complementary Training Before Using The Microscope All future users

More information

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL START-UP On the Switchbox, turn both black switches to the ON position. Wait for the microscope to boot up completely (watch the screen on the side of the microscope).

More information

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Operations Guide Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Nikon Eclipse 90i Operations Guide STARTUP Startup Powering Up Fluorescence

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Leica SP8 TCS Users Manual Follow the procedure for start up and log on as posted in the lab. Please log on with your account only and do not share your password with anyone. We track and confirm usage

More information

Before you start, make sure that you have a properly calibrated system to obtain high-quality images.

Before you start, make sure that you have a properly calibrated system to obtain high-quality images. CONTENT Step 1: Optimizing your Workspace for Acquisition... 1 Step 2: Tracing the Region of Interest... 2 Step 3: Camera (& Multichannel) Settings... 3 Step 4: Acquiring a Background Image (Brightfield)...

More information

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 Leica DMI AF6000LX Table of contents START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 ACQUIRE MODULE 6 SETTING THE LIGHTPATH 6

More information

Zeiss LSM 880 Protocol

Zeiss LSM 880 Protocol Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue

More information

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Verify that main power switches on the

More information

personal DELTAVISION (pdv)

personal DELTAVISION (pdv) GUIDELINES AND HINTS Version 1.3 (March 2015) personal DELTAVISION (pdv) Epifluorescence microscope from Applied Precision Inc.: The microscope can be found in room 1.320. For details see the architectural

More information

Standard Operating Procedure (SOP) for Shared Equipment: Spinning Disk Confocal Microscope

Standard Operating Procedure (SOP) for Shared Equipment: Spinning Disk Confocal Microscope Standard Operating Procedure (SOP) for Shared Equipment: Spinning Disk Confocal Microscope This document is to be used as a supplementary guide and not as a replacement for formal training. DO NOT operate

More information

Leica Sp5 II Confocal User Guide

Leica Sp5 II Confocal User Guide Leica Sp5 II Confocal User Guide Turning on the Confocal System (instructions are posted in the room) 1. Turn on Laser Power Button 2. Turn Key to On position 3. Turn on Scanner Power Button 4. Turn on

More information

Guide to Confocal 5. Starting session

Guide to Confocal 5. Starting session Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope

More information

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Introduction of Fluoresence Confocal Microscopy The first confocal microscope was invented by Princeton

More information

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope ZEN 2009 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn Chameleon TiS laser key from Standby

More information

Olympus xcellence Software - basic user guide

Olympus xcellence Software - basic user guide Olympus xcellence Software - basic user guide This is a basic overview of setting up time lapse experiments using Olympus's xcellence software on BIU's IX81 inverted phase contrast system - the software

More information

SPINNING DISK CSU-X1 USER MANUAL

SPINNING DISK CSU-X1 USER MANUAL SPINNING DISK CSU-X1 USER MANUAL Starting the temperature controller... 2 Starting the CO2 controller... 3 Start the spinning disk... 4 Sample observation with the oculars... 5 Spatial sampling, Pixel

More information

LEICA TCS SP5 AOBS TANDEM USER MANUAL

LEICA TCS SP5 AOBS TANDEM USER MANUAL LEICA TCS SP5 AOBS TANDEM USER MANUAL STARTING THE SYSTEM...2 THE LAS AF SOFTWARE...3 THE «ACQUIRE» MENU...5 CHOOSE AND CREATE A SETTING...6 THE CONTROL PANEL...8 THE DMI6000B MICROSCOPE...10 ACQUIRE ONE

More information

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE University of Virginia ITC Academic Computing Health Sciences CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE Introduction The Olympus BH-2 microscope in ACHS s microscope lab has objectives from

More information

b. Turn the power switch and key to on position for blue laser.

b. Turn the power switch and key to on position for blue laser. OLYMPUS FLUOVIEW 300 CONFOCAL MICOSCOPE OPERATION PROCEDURE 1. Turn ON microscope in this order: 1) Turn on mercury lamp (Note: once the mercury lamp is turned off, DO NOT turn it back on for at least

More information

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue

More information

Microscopy from Carl Zeiss

Microscopy from Carl Zeiss Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path

More information

DSU Spinning Disk Confocal and Slidebook 4.2 Quick Guide Light Microscopy Core Facility University of Chicago Vers 0.

DSU Spinning Disk Confocal and Slidebook 4.2 Quick Guide Light Microscopy Core Facility University of Chicago Vers 0. DSU Spinning Disk Confocal and Slidebook 4.2 Quick Guide Light Microscopy Core Facility University of Chicago Vers 0.8, February 2007 By: Vytas Bindokas, Ph.D. Core Director 2 The DSU system is optimized

More information

Zeiss Deconvolution Microscope: A Quick Guide

Zeiss Deconvolution Microscope: A Quick Guide Zeiss Deconvolution Microscope: A Quick Guide Start-up Uncover microscope. Do not put dust cover on the floor. Plug in both cameras. The default camera is the AxioCam HRm (monochrome camera) for fluorescence

More information

Things to check before start-up.

Things to check before start-up. Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your

More information

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 710 Confocal Microscope Standard Operation Protocol LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.

More information

Microscope Confocal Sp2 Upright.

Microscope Confocal Sp2 Upright. Microscope Confocal Sp2 Upright. Welcome to the Leica Sp2 Confocal Upright tutorial. Before using the Sp2 Invert, You will need to put down your name on the reservation system = http://svintranet.epfl.ch/index.php?optio

More information

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions Starting up: Schematic of Olympus BX-61. 1. Turn on Olympus microscope power box (left of microscope) with toggle switch on

More information

Fixed cell DSU Spinning Disk Confocal and Slidebook 5.0 Quick Guide. Light Microscopy Core Facility University of Chicago Vers 1.

Fixed cell DSU Spinning Disk Confocal and Slidebook 5.0 Quick Guide. Light Microscopy Core Facility University of Chicago Vers 1. Fixed cell DSU Spinning Disk Confocal and Slidebook 5.0 Quick Guide Light Microscopy Core Facility University of Chicago Vers 1.0, February 2009 By: Vytas Bindokas, Ph.D., Core Director The fixed cell

More information

QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Imperial College London

QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Imperial College London Imperial College London Facility for Imaging by Light Microscopy QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Observing Life As It Happens Startup procedure...

More information

Nikon E800 Operating Instructions.

Nikon E800 Operating Instructions. Nikon E800 Operating Instructions. You can request electronic copies of this manual by contacting lshats@jhsph.edu Copies are also available on the JHU MMI Department web site. Please send your comments

More information

Olympus IX71 Microscope and DP71 Camera Instructions

Olympus IX71 Microscope and DP71 Camera Instructions Olympus IX71 Microscope and DP71 Camera Instructions Microscopy in Medicine (MiM) Core Emory University Department of Medicine 1 Olympus IX71 Image Capture Procedure 2 3 1. STARTING-UP PROCEDURE: Remove

More information

10.2. Scanning Document Camera Scoring. Page 1 of 5. How do I score answer sheets using a document camera? STEP 1

10.2. Scanning Document Camera Scoring. Page 1 of 5. How do I score answer sheets using a document camera? STEP 1 Step by Step How do I score answer sheets using a document camera? STEP 1 Click on the Assessment icon in the top navigation bar. STEP 2 To locate your assessment in an assessment list, first select the

More information

Cell Biology and Bioimaging Core

Cell Biology and Bioimaging Core Cell Biology and Bioimaging Core Leica TCS SP5 Operating Instructions Starting up the instrument 1. First, log in the log book located on the confocal desk. Include your name, your lab s PI, an account

More information

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol Nikon Eclipse Ti-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol Please sign on the log sheet before switching on system. Turn on system Turn on A only if confocal mode or laser

More information

Zeiss AxioObserver with ApoTome

Zeiss AxioObserver with ApoTome Zeiss AxioObserver with ApoTome Quick Start User Guide LSU Health Sciences Center-Shreveport Research Core Facility (RCF) Microscopy Table of Contents 1 Start up the system.. Page 3 2 Touch screen controller

More information

LSM 780 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch

More information

Batch Counting of Foci

Batch Counting of Foci Batch Counting of Foci Getting results from Z stacks of images. 1. First it is necessary to determine suitable CHARM parameters to be used for batch counting. First drag a stack of images taken with the

More information

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: Confocal imaging on the Leica TCS SP8 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: 4) Do not touch the microscope while the software is initializing. Choose your options: Turn

More information

User Guide for TWAIN / DirectX interface for GRYPHAX USB 3.0 cameras

User Guide for TWAIN / DirectX interface for GRYPHAX USB 3.0 cameras User Guide for TWAIN / DirectX interface for GRYPHAX USB 3.0 cameras The TWAIN & DirectX driver for PROGRES GRYPHAX USB 3.0 cameras enables user to operate with TWAIN and DirectX supported 3 rd party software

More information

Nikon TE300 Eclipse Wide-Field Microscope

Nikon TE300 Eclipse Wide-Field Microscope Nikon TE300 Eclipse Wide-Field Microscope User Guide LSU Health Science Center-Shreveport Research Core Facility 1 User manual for Nikon Elements software Equipment: Nikon TE300 Eclipse microscope Photometrics

More information

TRAINING MANUAL. Olympus FV1000

TRAINING MANUAL. Olympus FV1000 TRAINING MANUAL Olympus FV1000 September 2014 TABLE OF CONTENTS A. Start-Up Procedure... 1 B. Visual Observation under the Microscope... 1 C. Image Acquisition... 4 A brief Overview of the Settings...

More information

Practical work no. 3: Confocal Live Cell Microscopy

Practical work no. 3: Confocal Live Cell Microscopy Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside

More information

Quick Operation Guide

Quick Operation Guide Quick Operation Guide Power ON Mounting specimens Set the specimen on the sample holder, and install the sample holder to the holder frame. Attach the holder frame to the XY stage. Type of holder Main

More information

Image Analysis for Fluorescence

Image Analysis for Fluorescence Image Analysis for Fluorescence Terminology Table Image Analysis Macro Colocalization Intensity Dye AFI The extraction of meaningful information from digital images by means of digital image processing

More information

Zeiss LSM 780 Protocol

Zeiss LSM 780 Protocol Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS Zeiss LSM 510 M eta Confocal M icroscope OPERATING INSTRUCTIONS Starting the System: 1. Turn the black knob on the laser box one-quarter turn from Off to On. You will hear the laser cooling mechanisms

More information

LSM 800 Confocal Microscope Standard Operation Protocol

LSM 800 Confocal Microscope Standard Operation Protocol LSM 800 Confocal Microscope Standard Operation Protocol Turning on the system 1. Switch on the Main switch (labeled 1 and 2 ) mounted on the wall. 2. Turn the Laser Key (labeled 3 ) 90 clockwise for power

More information

Image Viewing. with ImageScope

Image Viewing. with ImageScope Image Viewing with ImageScope ImageScope Components Use ImageScope to View These File Types: ScanScope Virtual Slides.SVS files created when the ScanScope scanner scans glass microscope slides. JPEG files

More information

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide Nikon C1si Spectral Laser Scanning Confocal Microscope User Guide Contents: C1Si Turn-On/ShutDown Procedures... 2 Overview... 4 Setup for epi-illumination to view through the eyepieces:... 5 Setup for

More information

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 CIAN LSM1 or LSM2 short instructions, version 1.4, September 2014 page 1 of 6 SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 Before starting To work with LSM1

More information

Swept-Field User Guide

Swept-Field User Guide Swept-Field User Guide Note: for more details see the Prairie user manual at http://www.prairietechnologies.com/resources/software/prairieview.html Please report any problems to Julie Last (jalast@wisc.edu)

More information

Figure 1 The Raith 150 TWO

Figure 1 The Raith 150 TWO RAITH 150 TWO SOP Figure 1 The Raith 150 TWO LOCATION: Raith 150 TWO room, Lithography area, NanoFab PRIMARY TRAINER: SECONDARY TRAINER: 1. OVERVIEW The Raith 150 TWO is an ultra high resolution, low voltage

More information

Microscope Confocal LSM510 META

Microscope Confocal LSM510 META Microscope Confocal LSM510 META Welcome to the Zeiss LSM 510 Meta Confocal tutorial. Before using the LSM 510 META, Log off any other computer that is open with your personal login. You will need to put

More information

IncuCyte ZOOM Fluorescent Processing Overview

IncuCyte ZOOM Fluorescent Processing Overview IncuCyte ZOOM Fluorescent Processing Overview The IncuCyte ZOOM offers users the ability to acquire HD phase as well as dual wavelength fluorescent images of living cells producing multiplexed data that

More information

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes Turn on Main Switch, System PC and Components Switches 780 Start up sequence Do you need the argon laser (458, 488, 514 nm lines)? Yes Turn on the laser s main power switch and

More information

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5. Smith College August 2010 Operating Checklist for using the Laser Scanning Confocal Microscope Leica TCS SP5. CONTENT, page no. Startup, 1 Initial set-up, 1 Software, 2 Microscope Specimen observation

More information

AxioVision 4.5 Brightfield Image Capture Procedure

AxioVision 4.5 Brightfield Image Capture Procedure AxioVision 4.5 Brightfield Image Capture Procedure 1. STARTING-UP PROCEDURE: Remove blue dust cover and place on shelf under microscope. Turn on the halogen lamp by pushing the switch at the back right

More information

SHORT GUIDE TO LASER MICRODISSECTION USING THE PALM COMBI SYSTEM

SHORT GUIDE TO LASER MICRODISSECTION USING THE PALM COMBI SYSTEM SHORT GUIDE TO LASER MICRODISSECTION USING THE PALM COMBI SYSTEM Turning ON the PALM DuoFlex Combi system 1. Turn on the three power point switches on the wall. From right to left: mercury lamp, microscope

More information

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s LSM 5 MP, LSM 510 and LSM 510 META M i c r o s c o p y f r o m C a r l Z e i s s Quick Guide Laser Scanning Microscopes LSM Software ZEN 2007 August 2007 We make it visible. Contents Page Contents... 1

More information

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Training Guide for Leica SP8 Confocal/Multiphoton Microscope Training Guide for Leica SP8 Confocal/Multiphoton Microscope LAS AF v3.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON power switch for epifluorescence

More information

Last updated: May 2014 Y.DeGraaf

Last updated: May 2014 Y.DeGraaf FLINDERS MICROSCOPY BIOMEDICAL SERVICES AVAILABLE MICROSCOPES AND SPECIFICATIONS & INFORMATION REGARDING TRAINING FOR NEW USERS Last updated: May 2014 Y.DeGraaf If you have new staff or students (Honours/Masters

More information

Axioscan - Startup. 1. Turn on the Axioscan on (button to the left) and turn on the computer

Axioscan - Startup. 1. Turn on the Axioscan on (button to the left) and turn on the computer Axioscan - Startup 1. Turn on the Axioscan on (button to the left) and turn on the computer 2. Log on and start the ZEN Blue software from the desktop 3. Press ZEN slidescan and Start System 4. Start by

More information

Nikon A1R. Multi-Photon & Laser Scanning Confocal Microscope. Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin. April 2017.

Nikon A1R. Multi-Photon & Laser Scanning Confocal Microscope. Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin. April 2017. Nikon A1R Multi-Photon & Laser Scanning Confocal Microscope Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin April 2017 Contents 1 Introduction 2 2 Start-Up 2 3 Imaging 4 3.1 Sample Alignment...........................................

More information

Proudly serving laboratories worldwide since 1979 SPECIFICATIONS

Proudly serving laboratories worldwide since 1979 SPECIFICATIONS www.ietltd.com Proudly serving laboratories worldwide since 1979 SPECIFICATIONS Scan RDI Specifications System Components Main analytical console Laser Module CRT Printer Data Manager Motorized stage (option)

More information

Motic Live Imaging Module. Windows OS User Manual

Motic Live Imaging Module. Windows OS User Manual Motic Live Imaging Module Windows OS User Manual Motic Live Imaging Module Windows OS User Manual CONTENTS (Linked) Introduction 05 Menus, bars and tools 06 Title bar 06 Menu bar 06 Status bar 07 FPS 07

More information

CHAPTER1: QUICK START...3 CAMERA INSTALLATION... 3 SOFTWARE AND DRIVER INSTALLATION... 3 START TCAPTURE...4 TCAPTURE PARAMETER SETTINGS... 5 CHAPTER2:

CHAPTER1: QUICK START...3 CAMERA INSTALLATION... 3 SOFTWARE AND DRIVER INSTALLATION... 3 START TCAPTURE...4 TCAPTURE PARAMETER SETTINGS... 5 CHAPTER2: Image acquisition, managing and processing software TCapture Instruction Manual Key to the Instruction Manual TC is shortened name used for TCapture. Help Refer to [Help] >> [About TCapture] menu for software

More information

Zoom Set Too Tight Zoom Set Correctly Zoom Set Too Wide

Zoom Set Too Tight Zoom Set Correctly Zoom Set Too Wide The ISG-E300 AutoCam Elite offers special features that increase capture efficiency and enhance image quality. By following the procedures outlined in this document, the ISG-E300 Elite can be used to its

More information

Olympus Digital Microscope Camera (DP70) checklist

Olympus Digital Microscope Camera (DP70) checklist Smith College - July 2005 Olympus Digital Microscope Camera (DP70) checklist CONTENT, page no. Camera Information, 1 Startup, 1 Retrieve an Image, 2 Microscope Setup, 2 Capture, 3 Preview. 3 Color Balans,

More information

PIXERA InStudio Software Package

PIXERA InStudio Software Package PIXERA InStudio Software Package Background InStudio is a comprehensive imaging capture and manipulation software program from Pixera. It combines the features of the original ViewfinderPro and StudioPro

More information

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL START THE SYSTEM... 2 START ZEN SOFTWARE... 3 SET THE TEMPERATURE AND THE CO2 CONTROLLERS... OBSERVATION AT OCULARS... 5 STATIF PRESENTATION... 6 ACQUIRE ONE

More information

IncuCyte ZOOM Scratch Wound Processing Overview

IncuCyte ZOOM Scratch Wound Processing Overview IncuCyte ZOOM Scratch Wound Processing Overview The IncuCyte ZOOM Scratch Wound assay utilizes the WoundMaker-IncuCyte ZOOM-ImageLock Plate system to analyze both 2D-migration and 3D-invasion in label-free,

More information

Quick Guide. NucleoCounter NC-3000

Quick Guide. NucleoCounter NC-3000 Quick Guide NucleoCounter NC-3000 Table of contents Setting up the FlexiCyte Protocol 2 Editing Image Capture and Analysis Parameters 3 Optimizing Exposure Time 4 Compensation for Spectral Overlap 6 Creating

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, EL6000 fluorescent light source for the microscope stand. 2. Turn on the Scanner Power

More information

RENISHAW INVIA RAMAN SPECTROMETER

RENISHAW INVIA RAMAN SPECTROMETER STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw invia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples

More information

SHORT GUIDE TO LASER MICRODISSECTION USING THE PALM COMBI SYSTEM

SHORT GUIDE TO LASER MICRODISSECTION USING THE PALM COMBI SYSTEM SHORT GUIDE TO LASER MICRODISSECTION USING THE PALM COMBI SYSTEM Turning ON the PALM DuoFlex Combi system 1. Turn on the three power point switches on the wall. From right to left: mercury lamp, microscope

More information

1. Preliminary sample preparation

1. Preliminary sample preparation FEI Helios NanoLab 600 standard operating procedure Nicholas G. Rudawski ngr@ufl.edu (352) 392 3077 (office) (805) 252-4916 (cell) Last updated: 03/02/18 What this document provides: an overview of basic

More information

GXCapture 8.1 Instruction Manual

GXCapture 8.1 Instruction Manual GT Vision image acquisition, managing and processing software GXCapture 8.1 Instruction Manual Contents of the Instruction Manual GXC is the shortened name used for GXCapture Square brackets are used to

More information

Bruker Optical Profilometer SOP Revision 2 01/04/16 Page 1 of 13. Bruker Optical Profilometer SOP

Bruker Optical Profilometer SOP Revision 2 01/04/16 Page 1 of 13. Bruker Optical Profilometer SOP Page 1 of 13 Bruker Optical Profilometer SOP The Contour GT-I, is a versatile bench-top optical surface-profiling system that can measure a wide variety of surfaces and samples. Contour GT optical profilers

More information