ScanArray Overview. Principle of Operation. Instrument Components

Size: px
Start display at page:

Download "ScanArray Overview. Principle of Operation. Instrument Components"

Transcription

1 ScanArray Overview The GSI Lumonics ScanArrayÒ Microarray Analysis System is a scanning laser confocal fluorescence microscope that is used to determine the fluorescence intensity of a two-dimensional array of dots on a bio-chip or microarray. Each dot on the microarray consists of target material that is chemically bound to the substrate. A probe, which is complementary material that is labeled with a fluorophore is preferentially bound to the target spots via a hybridization process to generate a fluorescent compound. The scanner can be used in conjunction with QuantArray TM for quantitation, processing and reporting and storage of data. Principle of Operation The fundamental operation of the ScanArrayÒ Microarray Analysis System involves the measurement of the fluorescence intensity of each dot on the sample. The sample dots are typically µm in diameter and a few µm thick on top of the substrate surface. The microarray is scanned (with a user settable resolution) to generate a map or image which indicates the fluorescence intensity for each point or pixel on the array. Typically, two (or more) different fluorophores are used to determine the concentration of the compounds of interest in each sample. This requires that the substrate is scanned two (or more) times with different lasers to provide different excitation wavelength. Each scan is performed on a sequential basis and the data is stored. Instrument Components The instrument includes an Excitation Source, Emission Optics, Detector and a Sample Entry System. Excitation Source Excitation radiation is provided by 1-4 lasers (depending on the system configuration). Most lasers will be mounted inside the scanner, however, large lasers or lasers with a significant power dissipation may be externally mounted. The excitation source will include appropriate beam shaping optics, an excitation filter, a shutter and beam alignment optics. Excitation Source The bio-chip or microarray is placed on the sample stage, which is scanned to collect data from all pixels within a user-defined area. Scanning is performed in both the X and Y direction. In addition, the Y direction is used to insert and eject the microarray into the unit. Emission Optics The emission optics collect the emitted fluorescence and direct it toward the detector. The optics includes emission filters to purify the light by blocking the reflected and scattered excitation radiation. Detector A photomultipler tube detects the emission radiation and converts it into a voltage. The signal is amplified, filtered, digitized and stored.

2 Manually Acquiring a Scan To acquire a scan with the manual sample loader: 1. Click the Acquire command on the Start menu to open the Acquire Image dialog box. 2. Select the desired protocol (and edit, if necessary). 3. Click the Eject command on the Start menu to eject the slide holder from the instrument. 4. Insert the microarray into the scanning position and retract the slide using the Eject command. 5. Focus the instrument (if necessary). 6. Press the Acquire button. Protocol Files When the Acquire Image dialog box is accessed, the user can select the desired protocol (*.sapr) by selecting the file using the Protocol Files field. The protocols are presented in a standard Windows file tree format and the user should click on the desired protocol to select it. Acquisition Area The Acquisition Area field indicates the area that is to be used for data acquisition and the area to be used for the Auto-Range/Auto-Balance function. The region indicated in color on the slide defines the region that contains the calibration spots for the Auto-Range/Auto-Balance operation, while the region indicated in black defines where data will be collected. The acquisition area can be defined by the cursor or by using the X-Position, Y-Position, Width and Height fields. Note: Editing the Acquisition Area for one fluorophore will change the Acquisition Area for all fluorophores. The X-Position, Y-Position, Width and Height fields are used to define the starting coordinates for the scan and the size of the region to be scanned. It is very important to start spotting 1 mm from the top of the slide and to spot 1.5 mm from each side of the slide so that spots are not placed outside the area to be scanned.. To edit the value for any of these fields, click Enter or Tab to move to the next field and enter the desired value. The region for the Auto-Balance Auto-Range function is selected on the Hardware Settings dialog box.

3 Resolution Resolution refers to the pixel (picture element) size, which is the dimension of each small square element that makes up the electronic image. A larger value (e.g. 50 µm) will result in a rapid scan which presents gross features of the sample while a small value (e.g µm) results in a slower scan that allows for the maximum information from the slide. We recommend a pixel size that is roughly 10-12% of the diameter of the microarray spot. Typically, a Quick Scan is used to preview the slide, determine where spots of interest are located and to set the sensitivity. Once the optimum parameters are determined, then the region of the slide that contains spots of interest are re-scanned using a smaller pixel size (and perhaps a slow scan speed) to collect data for quantitative analysis. To indicate the desired resolution, click on the appropriate radio button. If 30 or 50 µm resolution is desired, place a check mark in the box adjacent to Quick Scan. Scans The Scans field is used to indicate the number of scans or passes that are desired in the present data acquisition protocol. The scan number can be edited by clicking the field, typing the desired value and clicking Enter. The Sum and the Average radio buttons are used to indicate the desired data processing option if more than one scan is selected. Summing will increase the measured signal from dim samples, while averaging will reduce the effect of noise (this feature is not yet active). Scan Rate The Scan Rate is set in the Adjust Hardware Settings dialog box. Fluorophore Selection Field The Fluorophore Selection field lists the fluorophores that will be addressed by the current Acquisition protocol and the order in which the samples will be scanned. If the instrument does not include a filter or a laser that is required for a specific fluorophore, the listing in this field will be grayed out and cannot be included in the acquisition. If the operator clicks the Acquire button to indicate that a scan of the non-supported flurophore is desired, a dialog box will be presented that alerting that the scan will not include data from the indicated fluorophore. The Add button is used to access the Choose a Fluorophore dialog box, which lists the various fluorophores that are defined to ScanArray. Fluorophores that are not supported by the instrument that is attached will be indicated in gray. To select a fluorophore in the Selection field, click on the number to the left of each entry (in the # column). The Delete button in the Fluorophore Selection field is used to remove a fluorophore from the Flurophore Selection field. The fluorophore to be deleted should be highlighted via the mouse before the button is clicked. Deleting a fluorophore merely removes the fluorophore from this field; it does not remove it from the listing in the Choose a Fluorophore dialog box.

4 The Move Down and Move Up buttons are used to change the order of the scanning of the sample for the various fluorophores. To change the order, highlight the desired fluorophore with the mouse and click the appropriate button to lower or raise the fluorophore. Auto-Save Images The Auto-SaveImages check box is used to indicate if the images should be automatically saved at the end of the scan. If the box is not checked, the system will stop after each set of image and the data must be saved manually by accessing the Save dialog box. (Do not use Auto -Save on the BIPL scanner!) Auto Range before Every Scan The Auto Range before every scan check box is used to indicate if the Auto-Range/Auto-Balance feature should be used between each scan. Auto-Range The Auto-Range button is used to perform the Auto-Range/Auto-Balance measurement using the Auto-Range/Auto-Balance (ABAR) settings. Experiment Set Name The Experiment Set Name field is used to indicate the name of the folder in which the acquired images will be stored, if the Auto-save option has been selected. The file name must be Windows compatible (i.e. it cannot contain a slash, a backslash, a dot or any other Windows reserved character). QuantArray Analysis Protocol The QuantArray Analysis Protocol field is used to indicate the QuantArray that should be used to analyze the data. The Browse button is used to access the list of QuantArray analysis protocols, and the desired protocol is selected in the same manner as the acquisition protocol described above. The entire path of the desired QA protocol must be entered, not just the file name. Comments for the Protocol The Comments for the Protocol field is used by the operator to add information that describes the experiment. Any alphanumeric information can be added in this field.

5 The Scan Rate Field The Scan Rate field is used to indicate the speed of scanning of the sample. The radio button selects the desired value. A slow scan (25% or 50 %) will decrease pixel-to-pixel variation (AC noise). We recommend using the maximum speed setting except in unusual cases where the standard deviation or variation of pixel values is critical. Selecting the Auto-Balance, Auto-Range Area The X Position and Y Position fields define the point (in mm) where the Auto-Ranging and Auto- Balancing processes for that fluorophore should begin. The origin reference point is the upper left corner of the slide. The initial X Position and Y Position for Autoranging can be set in the same way as they are set for acquisition. The Width and Height define the size of the region for which Auto-Ranging is desired for that fluorophore. Microarray Spot Size The Microarray Spot Size is selected by adjusting the thumbwheel up or down. The ABAR algorithm uses the spot size to calculate the appropriate quick-scan resolution. If you do not know the diameter of your microarray spots, you can measure them on an image using the Circle cursor tool. Target Intensity The Target Intensity value can be selected by adjusting the thumbwheel or by entering a value. This is the pixel brightness value, which ABAR will attempt to achieve in the brightest part of the ABAR area. This is generally set to 90%, but any other value will work. Adjustment Method Two radio buttons are provided for selecting Adjustment Method. If PMT is selected, the laser power selection will remain fixed and ABAR will adjust the PMT setting to reach the target intensity. If Laser is selected, the PMT Gain selection will remain fixed and ABAR will automatically adjust the laser power to reach the target intensity. The Image File Name Field The Image File Name field defaults to the name of the fluorophore and can be changed (if desired) to simplify reports. Dark Signal Correction Performing a Dark Signal Correction The Dark Signal Correction function is used to compute a dark signal offset so that an appropriate offset adjustment can be made.

6 To perform a Dark Signal Correction on a clean, blank glass slide: 1. Remove the sample (if present) and insert a clean, blank microscope slide in the instrument. 2. Select the x position, y position, width and height to of the region to be used for the measurement of the dark signal. 3. Indicate the Pixel Value Upper Limit, which is the maximum pixel brightness value that will be displayed in the histogram. 4. Click the Dark Signal Correction button to access the Dark Signal Correction dialog box. 5. Press Scan 6. If the dark level, as indicated by the histograms is satisfactory, press OK to accept the correction and return to the Adjust Hardware Settings dialog box. 7. If the dark level is not satisfactory, enter a + or - delta intensity value into the Correction field and scan again. Repeat this process until you are satisfied with the dark level. To perform the dark signal correction with no laser, check the Scan with no Laser check box before initiating the scan. Dark Signal Correction Dialog Box The Dark Signal Correction dialog box is presented when the Dark Signal Correction button is clicked on the Adjust Hardware Settings dialog box. The Dark Signal Correction dialog box contains the following information: The Fluorophores field. The Scan Area field, which is used to define the region for the determination of the dark signal. Select this region by entering the x-position, y-position, width, and height fields by clicking in each box and entering the desired value or by selecting the area on the Preview of Area region with the mouse. The Pixel Value Upper Limit field, which is used to enter the upper limit for the histogram display. The Histogram Scan region presents the data that is obtained when the background scan is run. A histogram with a left-hand tail that goes near zero shows that almost no pixels have a zero intensity value. A histogram that shows a significant population of pixels at zero intensity may indicate that some low-level signals are being hidden by a negative offset to the signal. Your preferences for this setting may depend on the quantitation methods that are used downstream.

7 Role of the Focus/Line Scan Function Note: It is recommended that the Focusing operation is performed at a low laser power and a higher PMT voltage than might normally be used to minimize photobleaching. The Line Scan function is used to perform a series of scans along the x direction of the sample at a user selected Y position. The function allows the user to view a plot of the intensity as a function of the distance along the scan line and is used to focus the lens to optimize the sensitivity of the ScanArray. The focus for each fluorophore should be maximized each time a protocol is created. Routine scanning using an established protocol should not require regular adjustment to the focus. If the focus needs to be adjusted, it should be done before the sensitivity is adjusted (laser an/or PMT). The Line Scan Dialog Box The Line Scan dialog box is presented when the Focus/Line Scan button is clicked on the Adjust Hardware Settings dialog box or after the scan tool is used to define a line of interest on an image. The Line Scan dialog box contains the following information: The Fluorophores field. Focusing is done separately for each fluorophore. Select the fluorophore of interest, perform the focus operation and then go on to the next fluorophore. The Starting X Position and Size fields are used to indicate the region of the line to be monitored (e.g. each scan is from 4 to 14 mm from the left edge of the microarray). The edit boxes reflect the scan line selected on an image with the line scan tool or can be entered manually here. The Y-Stage field is used to select the position of the sample in the Y direction, and is used to move the microarray to view different regions. This reflects the position selected as an image with the Line Scan tool or can be adjusted here. Gross movements can be made by dragging the slider, fine movements can be made by clicking on the slider and using the up/down arrow keys. The Focus field is used to change the position of the lens to maximize the signal. Gross movements can be made by dragging the slider. Fine (5µm) movements can be made by clicking on the focus slider and using the left/right arrow keys to adjust the focus in increments. Adjust until the signal is maximized. Note: You must locate the line scan in an area on the sample that has a signal in order to determine focus. Nominal focus is at without a cover slip. With a cover slip that does not cover the sample to the edges of the slide, the nominal focus is and with a cover slip that covers the entire slide, it is approximately The scan region in the center of the position presents the data for the present scan. If data for the scan of two (or more) fluorophores are presented, they will appear in different colors. the curve), and is a numerical representation of the signal amplitude.

8 The Auto Focus radio buttons allow the user to select whether the auto-focus algorithm maximizes the signal level or the slope of edge transitions. Edge still works when the signal is saturated, but requires an edge to the scan line. Level works on an area of uniform signal, but requires a signal below saturation. The Current Average Signal field presents the averaged signal over the scanned range (the area under The Graticule field is used to select the appearance of the graticule, which is a pattern of horizontal lines in the display which users can select to suit their preferences. Determining the Focus Position for a Given Line The optimum focus position can be determined by the Auto-Focus command or by manually changing the Focus Position to maximize the intensity for each set of conditions. Note: If the signal approaches 100 %, reduce the Laser power or the PMT gain to avoid saturation of the signal. Automated Determination of the Focus To Manually Determine the Focus: Determining the Focus for the Entire Microarray Automated Determination of the Focus The Automated determination on the focus may be used when you have selected a scan line that includes features that have measurable signal from all of the fluorophores in the protocol. To use the Auto-Focus feature: Set the area parameters. Click Scan with each fluorophore in the protocol to ensure that there are spots at the indicated Y position. If spots are not observed with a reasonable intensity for each fluorophore, move to a new Y position and run another scan. The button will be recessed and the system will collect intensity data along the X position. When the signal is maximized, click the Scan button again to stop the operation. The scan data will be presented in the Scan region and the Current Average Signal field. When a line has been obtained that includes spots of reasonable intensity, press the Auto Focus button. The system will automatically scan the X-position line and move the lens as appropriate to determine the optimum lens position. The operator can stop the Auto-Focus process at any time by pressing the Auto Focus button again. When the optimum lens position is determined, it will be indicated on the Focus Position slide bar. Note: If the signal approaches 100 %, reduce the Laser power or the PMT gain to avoid saturation of the PMT or use AutoFocus in Edge mode.

9 To Manually Determine the Focus: Set the parameters. Click Scan to ensure that there are spots at the indicated Y position. If spots are not observed with a reasonable intensity, move to a new Y position and run another scan. When a line has been obtained that includes spots of reasonable intensity, record the Current Average Signal Change the focus position and continue to monitor the signal. Continue to change the focus position and observe the Current Average Signal until you have determined the position that produces maximum signal. Click the Scan button again to stop the line scan. e) Repeat for each fluorophore. For a good approximation, focus with one fluorophore and enter that focus position for the other fluorophores as well. Determining the Focus for the Entire Microarray The procedure to determines the focus for a given Y position which may be (or may not be representative) of the complete microarray. It is recommended that the above procedure be performed near the center of the array field so that the best focus is held throughout the field. The Auto-Balance, Auto-Range Function Auto-Balancing and Auto-Ranging are used to adjust the PMT gain and laser power to match the brightness of a sample. Role of the Auto-Balance, Auto-Range Function To Auto-Range and Auto-Balance the System Role of the Auto-Balance, Auto-Range Function Auto-ranging is the instrument function whereby the sensitivity is adjusted to match the brightness of a sample. Based on the size of the microarray spots, it selects a quick-scan resolution that acquires at least two scan lines across each spot. From this sample of the data, the part of the region where the 4 brightest adjacent pixels is identified. Lastly, the instrument adjusts either the PMT or the laser setting so that these 4 brightest pixels will produce data at a user-selected intensity level on the next scan. Generally, a user will select that target intensity level to be around 90% so that the dynamic range of the instrument is fully utilized. Auto-balancing is the same process, repeated for each fluorophore. This leaves the instrument with settings that will cause the next scan to produce data where the brightest features in each image will all be at the same user-selected brightness level. This is called balancing the channels. Auto-Balancing and Auto-Ranging (ABAR) would most often be performed on areas that contain microarray control spots, housekeeping or other genes that produce maximum signals.

10 Different ABAR areas can be chosen for different fluorophores. To Auto-Range and Auto-Balance the System Select the fluorophore of interest by clicking on the number in the left column of the Fluorophores box. In the ABAR section of the Adjust Hardware Settings dialog, the area on the sample in which the auto-ranging will be done can then be selected with the mouse on the cartoon of the slide area, or by entering position and size numbers into the edit boxes.

The Zeiss AiryScan System, Confocal Four.

The Zeiss AiryScan System, Confocal Four. The Zeiss AiryScan System, Confocal Four. Overview. The Zeiss AiryScan module is a segmented, radially stacked GaASP detector and collector system designed to subsample the airy disk of a point emission

More information

In our previous lecture, we understood the vital parameters to be taken into consideration before data acquisition and scanning.

In our previous lecture, we understood the vital parameters to be taken into consideration before data acquisition and scanning. Interactomics: Protein Arrays & Label Free Biosensors Professor Sanjeeva Srivastava MOOC NPTEL Course Indian Institute of Technology Bombay Module 7 Lecture No 34 Software for Image scanning and data processing

More information

Leica SPEII confocal microscope. Short Manual

Leica SPEII confocal microscope. Short Manual Leica SPEII confocal microscope Short Manual Switching ON sequence: 1. Turn on the Workstation under the bench (top, far right). 2. Turn on the Supply Unit - Laser box (big green switch first and then

More information

Cell Biology and Bioimaging Core

Cell Biology and Bioimaging Core Cell Biology and Bioimaging Core Leica TCS SP5 Operating Instructions Starting up the instrument 1. First, log in the log book located on the confocal desk. Include your name, your lab s PI, an account

More information

LSM 780 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch

More information

TRAINING MANUAL. Olympus FV1000

TRAINING MANUAL. Olympus FV1000 TRAINING MANUAL Olympus FV1000 September 2014 TABLE OF CONTENTS A. Start-Up Procedure... 1 B. Visual Observation under the Microscope... 1 C. Image Acquisition... 4 A brief Overview of the Settings...

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, Fluorescent Light for the microscope stand. 2. Turn on the Scanner Power (1) on the front

More information

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Verify that main power switches on the

More information

Microscopy from Carl Zeiss

Microscopy from Carl Zeiss Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path

More information

Scanning and Image Processing -by Steve Clough

Scanning and Image Processing -by Steve Clough Scanning and Image Processing -by Steve Clough cdna microarrays use two dyes with well separated emission spectra such as Cy3 and Cy5 to allow direct comparisons on single slide GSI Lumonics Ratio of Expression

More information

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s LSM 5 MP, LSM 510 and LSM 510 META M i c r o s c o p y f r o m C a r l Z e i s s Quick Guide Laser Scanning Microscopes LSM Software ZEN 2007 August 2007 We make it visible. Contents Page Contents... 1

More information

Zeiss 880 Training Notes Zen 2.3

Zeiss 880 Training Notes Zen 2.3 Zeiss 880 Training Notes Zen 2.3 1 Turn on the HXP 120V Lamp 2 Turn on Main Power Switch Turn on the Systems PC Switch Turn on the Components Switch. 3 4 5 Turn on the PC and log into your account. Start

More information

ISCapture User Guide. advanced CCD imaging. Opticstar

ISCapture User Guide. advanced CCD imaging. Opticstar advanced CCD imaging Opticstar I We always check the accuracy of the information in our promotional material. However, due to the continuous process of product development and improvement it is possible

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Leica SP8 TCS Users Manual Follow the procedure for start up and log on as posted in the lab. Please log on with your account only and do not share your password with anyone. We track and confirm usage

More information

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL MIF ZEISS LSM510 CONFOCAL USER PROTOCOL START-UP Turn on the Mercury Bulb Power Supply (if needed). Power-on the Control Box. Turn on the computer. Open the LSM 510 software. Choose Scan New Images and

More information

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide Operation 58. UltraVIEW VoX Getting Started Guide Overview This chapter outlines the basic methods used to operate the UltraVIEW VoX system. About other software Volocity places great demands on the computer

More information

Title: Leica SP5 Confocal User Manual

Title: Leica SP5 Confocal User Manual Title: Leica SP5 Confocal User Manual Date of first issue: 23/10/2015 Date of review: Version: Admin For assistance or to report an issue Office: CG07 or 05 Email: Igmm-imaginghelpdesk@igmm.ed.ac.uk Website:

More information

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA HANDOUT LECTURE-31 MICROARRAY WORK-FLOW: IMAGE SCANNING AND DATA PROCESSING Slide 1: This module contains the summary of the discussion with Mr. Pankaj Khanna, an application specialist from Spinco Biotech,

More information

Scan slides (Axon Genepix 4200AL)

Scan slides (Axon Genepix 4200AL) Page 1 Scan slides (Axon Genepix 4200AL) We need to scan the slides on both channels (Cy3 and Cy5) to obtain a 16-bit grayscale TIFF file for each. Typically these files are about 20-26Mb per channel,

More information

Things to check before start-up.

Things to check before start-up. Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your

More information

Practical work no. 3: Confocal Live Cell Microscopy

Practical work no. 3: Confocal Live Cell Microscopy Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside

More information

Leica Sp5 II Confocal User Guide

Leica Sp5 II Confocal User Guide Leica Sp5 II Confocal User Guide Turning on the Confocal System (instructions are posted in the room) 1. Turn on Laser Power Button 2. Turn Key to On position 3. Turn on Scanner Power Button 4. Turn on

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, EL6000 fluorescent light source for the microscope stand. 2. Turn on the Scanner Power

More information

EmbryoCellect. RHS Scanning and Analysis Instructions. for. Genepix Pro Software

EmbryoCellect. RHS Scanning and Analysis Instructions. for. Genepix Pro Software EmbryoCellect RHS Scanning and Analysis Instructions for Genepix Pro Software EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 1 Copyright Reproductive Health

More information

Products - Microarray Scanners - Laser Scanners - InnoScan 900 Series and MAPIX Software

Products - Microarray Scanners - Laser Scanners - InnoScan 900 Series and MAPIX Software Products - Microarray Scanners - Laser Scanners - InnoScan 900 Series and MAPIX Software Arrayit offers the world s only next generation microarray scanning technology, with proprietary rotary motion control,

More information

GenePix Application Note

GenePix Application Note GenePix Application Note Determining the Signal-to-Noise Ratio and Optimal Photomultiplier gain setting in the GenePix 4000B Siobhan Pickett, M.S., Sean Carriedo, Ph.D. and Chang Wang, Ph.D. Axon Instruments,

More information

Operating Instructions for Zeiss LSM 510

Operating Instructions for Zeiss LSM 510 Operating Instructions for Zeiss LSM 510 Location: GNL 6.312q (BSL3) Questions? Contact: Maxim Ivannikov, maivanni@utmb.edu 1 Attend A Complementary Training Before Using The Microscope All future users

More information

LSM 510 Meta Training Notes

LSM 510 Meta Training Notes LSM 510 Meta Training Notes Turning on the system Turn on X-Cite power supply. This supplies light for epifluorescence for viewing your samples through the microscope. Turn on the remote control switch.

More information

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Training Guide for Leica SP8 Confocal/Multiphoton Microscope Training Guide for Leica SP8 Confocal/Multiphoton Microscope LAS AF v3.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON power switch for epifluorescence

More information

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Operations Guide Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Nikon Eclipse 90i Operations Guide STARTUP Startup Powering Up Fluorescence

More information

Point Calibration. July 3, 2012

Point Calibration. July 3, 2012 Point Calibration July 3, 2012 The purpose of the Point Calibration process is to generate a map of voltages (for galvos) or motor positions of the pointing device to the voltages or pixels of the reference

More information

Contents. Introduction

Contents. Introduction Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path and lasers... 12 Scanning

More information

Swept-Field User Guide

Swept-Field User Guide Swept-Field User Guide Note: for more details see the Prairie user manual at http://www.prairietechnologies.com/resources/software/prairieview.html Please report any problems to Julie Last (jalast@wisc.edu)

More information

ThermaViz. Operating Manual. The Innovative Two-Wavelength Imaging Pyrometer

ThermaViz. Operating Manual. The Innovative Two-Wavelength Imaging Pyrometer ThermaViz The Innovative Two-Wavelength Imaging Pyrometer Operating Manual The integration of advanced optical diagnostics and intelligent materials processing for temperature measurement and process control.

More information

Product Information. Introduction

Product Information. Introduction Microarray Scanner Calibration Slide To quantitatively analyze scanners performance and output, adjust and fine-tune scanners, and perform comparative analysis for multiple scanner units. To verify scanners

More information

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes Turn on Main Switch, System PC and Components Switches 780 Start up sequence Do you need the argon laser (458, 488, 514 nm lines)? Yes Turn on the laser s main power switch and

More information

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: Confocal imaging on the Leica TCS SP8 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: 4) Do not touch the microscope while the software is initializing. Choose your options: Turn

More information

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 710 Confocal Microscope Standard Operation Protocol LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.

More information

LSM 510 Training Notes

LSM 510 Training Notes LSM 510 Training Notes Turning on the system Turn on the arc lamp, found on the bench top left of the microscope. This supplies light for epifluorescence for viewing your samples through the microscope.

More information

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling Zeiss LSM 510 Confocor III Training Notes Center for Cell Analysis & Modeling Confocor 3 Start Up Go to System Module Turn on Main Switch, System/ PC, and Components Switches Do you need the arc lamp?

More information

User manual for Olympus SD-OSR spinning disk confocal microscope

User manual for Olympus SD-OSR spinning disk confocal microscope User manual for Olympus SD-OSR spinning disk confocal microscope Ved Prakash, PhD. Research imaging specialist Imaging & histology core University of Texas, Dallas ved.prakash@utdallas.edu Once you open

More information

Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal

Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal Todays Goal: Introduce some additional functionalities of the Leica SP8 confocal HyD vs. PMT detectors Dye Assistant Scanning By

More information

ScanGear CS-U 5.6. for CanoScan FB1210U Color Scanner. User s Guide

ScanGear CS-U 5.6. for CanoScan FB1210U Color Scanner. User s Guide ScanGear CS-U 5.6 for CanoScan FB1210U Color Scanner User s Guide Copyright Notice Copyright 2000 Canon Inc. This manual is copyrighted with all rights reserved. Under the copyright laws, this manual may

More information

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and

More information

CHAPTER1: QUICK START...3 CAMERA INSTALLATION... 3 SOFTWARE AND DRIVER INSTALLATION... 3 START TCAPTURE...4 TCAPTURE PARAMETER SETTINGS... 5 CHAPTER2:

CHAPTER1: QUICK START...3 CAMERA INSTALLATION... 3 SOFTWARE AND DRIVER INSTALLATION... 3 START TCAPTURE...4 TCAPTURE PARAMETER SETTINGS... 5 CHAPTER2: Image acquisition, managing and processing software TCapture Instruction Manual Key to the Instruction Manual TC is shortened name used for TCapture. Help Refer to [Help] >> [About TCapture] menu for software

More information

Locating Molecules Using GSD Technology Project Folders: Organization of Experiment Files...1

Locating Molecules Using GSD Technology Project Folders: Organization of Experiment Files...1 .....................................1 1 Project Folders: Organization of Experiment Files.................................1 2 Steps........................................................................2

More information

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope Training Guide for Carl Zeiss LSM 510 META Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON Components and System/PC switches

More information

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL START-UP On the Switchbox, turn both black switches to the ON position. Wait for the microscope to boot up completely (watch the screen on the side of the microscope).

More information

Before you start, make sure that you have a properly calibrated system to obtain high-quality images.

Before you start, make sure that you have a properly calibrated system to obtain high-quality images. CONTENT Step 1: Optimizing your Workspace for Acquisition... 1 Step 2: Tracing the Region of Interest... 2 Step 3: Camera (& Multichannel) Settings... 3 Step 4: Acquiring a Background Image (Brightfield)...

More information

User Guide of ISCapture

User Guide of ISCapture User Guide of ISCapture For Windows2000/XP/Vista(32bit/64bit)/Win7(32bit/64bit) Xintu Photonics Co., Ltd. Version: 2.6 I All the users of Xintu please kindly note that the information and references in

More information

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each lens with a 1 Airy unit pinhole setting. Many water lenses

More information

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Introduction of Fluoresence Confocal Microscopy The first confocal microscope was invented by Princeton

More information

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST Training Guide for Carl Zeiss LSM 880 with AiryScan FAST ZEN 2.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2018) Power ON Routine 1 2 Turn ON Main Switch from the remote control

More information

Axioscan - Startup. 1. Turn on the Axioscan (button to the left) and turn on the computer. 2. Log on and start the ZEN Blue software from the desktop

Axioscan - Startup. 1. Turn on the Axioscan (button to the left) and turn on the computer. 2. Log on and start the ZEN Blue software from the desktop Axioscan - Startup 1. Turn on the Axioscan (button to the left) and turn on the computer 2. Log on and start the ZEN Blue software from the desktop 3. Press ZEN slidescan and Start System 4. Start by changing

More information

Leica SP8 Resonant Confocal. Quick-Start Guide

Leica SP8 Resonant Confocal. Quick-Start Guide Leica SP8 Resonant Confocal Quick-Start Guide Contents Start-up Preparing for Imaging Part 1 On the scope Part 2 Software interface Part 3 Heat & CO2 incubation Part 4 Other hardware options Shut-down

More information

Quick Start Guide. Leica SP5 X

Quick Start Guide. Leica SP5 X Quick Start Guide Leica SP5 X Please note: Some of the information in this guide was taken from Leica Microsystems Leica TCS SP5 LAS AF Guide for New Users. This work is licensed under the Creative Commons

More information

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA LECTURE-31 MICROARRAY WORK-FLOW: IMAGE SACNNING AND DATA PROCESSING TRANSCRIPT Welcome to the proteomics course. In today s lecture we will talk about microarray work-flow the image scanning and processing.

More information

ImagesPlus Basic Interface Operation

ImagesPlus Basic Interface Operation ImagesPlus Basic Interface Operation The basic interface operation menu options are located on the File, View, Open Images, Open Operators, and Help main menus. File Menu New The New command creates a

More information

GXCapture 8.1 Instruction Manual

GXCapture 8.1 Instruction Manual GT Vision image acquisition, managing and processing software GXCapture 8.1 Instruction Manual Contents of the Instruction Manual GXC is the shortened name used for GXCapture Square brackets are used to

More information

Simplified Instructions: Olympus Widefield Microscope S1230

Simplified Instructions: Olympus Widefield Microscope S1230 Contents General Microscope Operation Simple Image Capture Multi-Wavelength Capture Z-Series Timelapse Combining Capture Modes Synopsis of Other Functions Pages 2-23 24-40 41-47 48-56 57-59 60-68 69-83

More information

Chlorophyll Fluorescence Imaging System

Chlorophyll Fluorescence Imaging System Quick Start Guide Chlorophyll Fluorescence Imaging System Quick Start Guide for Technologica FluorImager software for use with Technlogica CFImager hardware Copyright 2006 2015 TECHNOLOGICA LIMITED. All

More information

Zeiss LSM 880 Protocol

Zeiss LSM 880 Protocol Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

User Manual for HoloStudio M4 2.5 with HoloMonitor M4. Phase Holographic Imaging

User Manual for HoloStudio M4 2.5 with HoloMonitor M4. Phase Holographic Imaging User Manual for HoloStudio M4 2.5 with HoloMonitor M4 Phase Holographic Imaging 1 2 HoloStudio M4 2.5 Software instruction manual 2013 Phase Holographic Imaging AB 3 Contact us: Phase Holographic Imaging

More information

RENISHAW INVIA RAMAN SPECTROMETER

RENISHAW INVIA RAMAN SPECTROMETER STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw invia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples

More information

Zeiss LSM 780 Protocol

Zeiss LSM 780 Protocol Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017. Nikon AZ100 Laser Scanning Macro Confocal Microscope Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin May 2017 Contents 1 Introduction 2 2 Hardware - Startup 2 3 Software/Operation 4 3.1 Multidimensional

More information

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5. Smith College August 2010 Operating Checklist for using the Laser Scanning Confocal Microscope Leica TCS SP5. CONTENT, page no. Startup, 1 Initial set-up, 1 Software, 2 Microscope Specimen observation

More information

GenePix Application Note

GenePix Application Note GenePix Application Note Biological Relevance of GenePix Results Shawn Handran, Ph.D. and Jack Y. Zhai, Ph.D. Axon Instruments, Inc. 3280 Whipple Road, Union City, CA 94587 Last Updated: Aug 22, 2003.

More information

ScanGear CS-U 6.0. for CanoScan D646U Color Scanner. User s Guide

ScanGear CS-U 6.0. for CanoScan D646U Color Scanner. User s Guide ScanGear CS-U 6.0 for CanoScan D646U Color Scanner User s Guide Copyright Notice Copyright 2000 Canon Inc. This manual is copyrighted with all rights reserved. Under the copyright laws, this manual may

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Version : 07/08/0 Leica SP8 TCS Users Manual Start up:. Turn the PC Microscope, Scanner Power, Laser Power, and the Laser Emission key to on (bottom right of desk).. Turn on the fluorescent lamp (top left

More information

Quick Guide. NucleoCounter NC-3000

Quick Guide. NucleoCounter NC-3000 Quick Guide NucleoCounter NC-3000 Table of contents Setting up the FlexiCyte Protocol 2 Editing Image Capture and Analysis Parameters 3 Optimizing Exposure Time 4 Compensation for Spectral Overlap 6 Creating

More information

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging Bi/BE 227 Winter 2016 Assignment #3 Adding the third dimension: 3D Confocal Imaging Schedule: Jan 20: Assignment Jan 20-Feb 8: Work on assignment Feb 10: Student PowerPoint presentations. Goals for this

More information

MC3 Motion Control System Shutter Stream Quickstart

MC3 Motion Control System Shutter Stream Quickstart MC3 Motion Control System Shutter Stream Quickstart Revised 7/6/2016 Carousel USA 6370 N. Irwindale Rd. Irwindale, CA 91702 www.carousel-usa.com Proprietary Information Carousel USA has proprietary rights

More information

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope ZEN 2012 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 If you require fluorescence imaging,

More information

Optika ISview. Image acquisition and processing software. Instruction Manual

Optika ISview. Image acquisition and processing software. Instruction Manual Optika ISview Image acquisition and processing software Instruction Manual Key to the Instruction Manual IS is shortened name used for OptikaISview Square brackets are used to indicate items such as menu

More information

CHROMACAL User Guide (v 1.1) User Guide

CHROMACAL User Guide (v 1.1) User Guide CHROMACAL User Guide (v 1.1) User Guide User Guide Notice Hello and welcome to the User Guide for the Datacolor CHROMACAL Color Calibration System for Optical Microscopy, a cross-platform solution that

More information

Photoshop CS2. Step by Step Instructions Using Layers. Adobe. About Layers:

Photoshop CS2. Step by Step Instructions Using Layers. Adobe. About Layers: About Layers: Layers allow you to work on one element of an image without disturbing the others. Think of layers as sheets of acetate stacked one on top of the other. You can see through transparent areas

More information

AgilEye Manual Version 2.0 February 28, 2007

AgilEye Manual Version 2.0 February 28, 2007 AgilEye Manual Version 2.0 February 28, 2007 1717 Louisiana NE Suite 202 Albuquerque, NM 87110 (505) 268-4742 support@agiloptics.com 2 (505) 268-4742 v. 2.0 February 07, 2007 3 Introduction AgilEye Wavefront

More information

Introduction to Parametric Modeling AEROPLANE. Design & Communication Graphics 1

Introduction to Parametric Modeling AEROPLANE. Design & Communication Graphics 1 AEROPLANE Design & Communication Graphics 1 Object Analysis sheet Design & Communication Graphics 2 Aeroplane Assembly The part files for this assembly are saved in the folder titled Aeroplane. Open an

More information

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope ZEN 2009 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn Chameleon TiS laser key from Standby

More information

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000 The ideal K-12 science microscope solution User Guide for use with the Nova5000 NovaScope User Guide Information in this document is subject to change without notice. 2009 Fourier Systems Ltd. All rights

More information

Guide to Confocal 5. Starting session

Guide to Confocal 5. Starting session Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS Zeiss LSM 510 M eta Confocal M icroscope OPERATING INSTRUCTIONS Starting the System: 1. Turn the black knob on the laser box one-quarter turn from Off to On. You will hear the laser cooling mechanisms

More information

Fire CR Calibration Guide

Fire CR Calibration Guide 1 Fire CR Calibration Guide This reference guide will guide you through the steps to complete the calibration for the Fire CR.. Getting Started: 1. Click on the Opal Icon on the Desktop. Figure 1 2. Once

More information

Opterra II Multipoint Scanning Confocal Microscope. Innovation with Integrity

Opterra II Multipoint Scanning Confocal Microscope. Innovation with Integrity Opterra II Multipoint Scanning Confocal Microscope Enabling 4D Live-Cell Fluorescence Imaging through Speed, Sensitivity, Viability and Simplicity Innovation with Integrity Fluorescence Microscopy The

More information

Kigamo Scanback which fits in your view camera in place of conventional film.

Kigamo Scanback which fits in your view camera in place of conventional film. What's included Kigamo Scanback which fits in your view camera in place of conventional film. SCSI Cable to connect your Scanback to the host computer. A 3-meter SCSI cable is standard. Kigamo also has

More information

Quick Operation Guide

Quick Operation Guide Quick Operation Guide Power ON Mounting specimens Set the specimen on the sample holder, and install the sample holder to the holder frame. Attach the holder frame to the XY stage. Type of holder Main

More information

Renishaw InVia Raman microscope

Renishaw InVia Raman microscope Laser Spectroscopy Labs Renishaw InVia Raman microscope Operation instructions 1. Turn On the power switch, system power switch is located towards the back of the system on the right hand side. Wait ~10

More information

SPECTRALIS Training Guide

SPECTRALIS Training Guide SPECTRALIS Training Guide SPECTRALIS Diagram 1 SPECTRALIS Training Guide Table of Contents 1. Entering Patient Information & Aligning the Patient a. Start Up/Shut Down the System... 4 b. Examine a New

More information

ScanGear CS-U 5.3 for CanoScan FB630U/FB636U Color Image Scanner User s Guide

ScanGear CS-U 5.3 for CanoScan FB630U/FB636U Color Image Scanner User s Guide ScanGear CS-U 5.3 for CanoScan FB630U/FB636U Color Image Scanner User s Guide Copyright Notice 1999 Canon Inc. This manual is copyrighted with all rights reserved. Under the copyright laws, this manual

More information

Adobe Studio on Adobe Photoshop CS2 Enhance scientific and medical images. 2 Hide the original layer.

Adobe Studio on Adobe Photoshop CS2 Enhance scientific and medical images. 2 Hide the original layer. 1 Adobe Studio on Adobe Photoshop CS2 Light, shadow and detail interact in wild and mysterious ways in microscopic photography, posing special challenges for the researcher and educator. With Adobe Photoshop

More information

Instructions for Howto Scan µarrays

Instructions for Howto Scan µarrays Instructions for Howto Scan µarrays Introduction After probing the µarray slides with samples, one is now ready to scan them. To scan a µarrays slide is too convert the biological information trapped on

More information

LSM 800 Confocal Microscope Standard Operation Protocol

LSM 800 Confocal Microscope Standard Operation Protocol LSM 800 Confocal Microscope Standard Operation Protocol Turning on the system 1. Switch on the Main switch (labeled 1 and 2 ) mounted on the wall. 2. Turn the Laser Key (labeled 3 ) 90 clockwise for power

More information

LA-T LED ANALYSER EVALUATION KIT INSTRUCTION MANUAL. rev

LA-T LED ANALYSER EVALUATION KIT INSTRUCTION MANUAL. rev LA-T LED ANALYSER EVALUATION KIT INSTRUCTION MANUAL rev. 300117 TABLE OF CONTENTS General Information 3 Application 3 Design 3 Features 3 Operation conditions 3 Operation instructions 4-7 2 GENERAL INFORMATION

More information

Quick Start for Autodesk Inventor

Quick Start for Autodesk Inventor Quick Start for Autodesk Inventor Autodesk Inventor Professional is a 3D mechanical design tool with powerful solid modeling capabilities and an intuitive interface. In this lesson, you use a typical workflow

More information

Topics. - How to calibrate the LSM scanner. - How to clean the microscope. - How to adjust the pinhole alignment. - How to adjust the Collimator

Topics. - How to calibrate the LSM scanner. - How to clean the microscope. - How to adjust the pinhole alignment. - How to adjust the Collimator Topics - How to calibrate the LSM scanner - How to measure the PSF - How to clean the microscope - How to adjust the pinhole alignment - How to adjust the Collimator How to calibrate the LSM scanner The

More information

GlassSpection User Guide

GlassSpection User Guide i GlassSpection User Guide GlassSpection User Guide v1.1a January2011 ii Support: Support for GlassSpection is available from Pyramid Imaging. Send any questions or test images you want us to evaluate

More information

Photoshop CC Editing Images

Photoshop CC Editing Images Photoshop CC Editing Images Rotate a Canvas A canvas can be rotated 90 degrees Clockwise, 90 degrees Counter Clockwise, or rotated 180 degrees. Navigate to the Image Menu, select Image Rotation and then

More information

Batch Counting of Foci

Batch Counting of Foci Batch Counting of Foci Getting results from Z stacks of images. 1. First it is necessary to determine suitable CHARM parameters to be used for batch counting. First drag a stack of images taken with the

More information

MicroLab 500-series Getting Started

MicroLab 500-series Getting Started MicroLab 500-series Getting Started 2 Contents CHAPTER 1: Getting Started Connecting the Hardware....6 Installing the USB driver......6 Installing the Software.....8 Starting a new Experiment...8 CHAPTER

More information