1 ZEISS EVO SOP May 2017 ELECTRON OPTICS The patented EVO column is the area of the SEM, where electrons are emitted, accelerated, deflected, focused, and scanned. Main characteristics of the EVO optics are the condenser and objective lenses. 1 W or LaB6 cathode 2 Gun alignment and emission display coils 3 Double condenser system 4 Mechanical mid column aperture changer 5 Scanning and stigmator coils 6 Objective lens with through the lens pumping 7 CZ BSD 8 Specimen Gun: The electrons are accelerated by the acceleration voltage (UEHT). The filament is heated by applying the filament current. Electrons are emitted from the heated filament. EHT: The emitted electrons are accelerated by the acceleration voltage (UEHT). Gun Alignment and Display Coils: The electron beam is centered by the gun alignment and emission coils (2). Condensers: A double condenser system (3) allows the continuous regulation of the probe current. Apertures: The electron beam passes through the currently selected aperture in the mechanical mid column aperture changer (4) to ensure optimum quality of beam is maintained. Scanning and Stigmator Coils: The scanning coils (5) move the electron beam in a point-bypoint scan over the specimen surface. The stigmator coils (5) compensate for astigmatism, so that the electron beam becomes rotationally symmetrical. Objective Lens: The electron beam exits the objective lens (6) Signal Detection: When the primary electron beam hits the specimen, certain interaction signals are generated, which are recorded by various detectors, such as the CZ BSD (7).

2 SIGNAL DETECTION The interaction products most frequently used for the generation of images in scanning electron microscopy are secondary electrons (SE) and backscattered electrons (BSE). Standard detectors Detected signals Typical application SE2 detector (Everhart-Thornley) SE2 Topography and surface structure VPSE detector SE On VP systems only Optional detectors Detected signals Typical application Backscattered electron detectors, including 4- quadrant CZBSD Cathodoluminescence (CL) detector Scanning Transmission Electron Microscopy (STEM) detector EPSE (needle and ring) PRINCIPLES OF DETECTION BSE Light photons Transmission electrons SE* *By means of measuring current produced by ionization of gas by SE Electrons generated around When a primary electron (PE) beam hits a specimen, certain electron beam interaction processes occur. Secondary electrons (SEs) and backscattered electrons (BSEs) are then generated. Specific types of detectors are able to detect the SEs and BSEs, and the detector signals can be used to create images and produce information about the properties of the specimen. Secondary Electrons: Secondary electrons are ejected from the outer atomic shell of the specimen material upon impact by the primary electron beam. The three main categories of secondary electrons are based on the origin of the secondary electrons Topographical (crystal orientation), atomic number contrast. Mineralogy Transmission imaging of thin sections in biological and mineralogical examinations Extended pressure applications. Typically when imaging biological specimens in a fully hydrated form

3 and on the distance from the PE impact point, where they leave the specimen. SE1 electrons are generated and leave the surface of the specimen directly at the spot where the primary electron beam impacts on the specimen surface. SE2 electrons are generated after multiple scattering inside the interaction volume, and leave the specimen at a greater distance from the primary beam s impact point. SE3 electrons are generated by backscattered electrons colliding with the chamber walls or the lens system. Secondary electrons have low energy (less than 50 ev). Backscatter Electrons: All electrons with energy higher than 50 ev are known as backscattered electrons (BSEs). BSEs are generated by elastic scattering in a much deeper range of the interaction volume and carry depth information. The backscatter coefficient increases with increasing atomic number of the elements within the specimen. This allows the BSE detector to generate atomic number contrast, or compositional contrast images. ET-SE DETECTOR (SE2 DETECTOR) The ET-SE, or Everhart-Thornley, detector is mounted on the wall of the specimen chamber, and is therefore classed as a chamber detector. Due to its position in the chamber, the SE2 detector views the specimen laterally. The SE2 detector allows detection of secondary electrons with a small backscattered component. Secondary electrons moving towards the detector are attracted by the collector and are directed towards the scintillator accelerated by the scintillator voltage. Operating Principle When the high energy electrons hit the scintillator layer (2), photons are generated inside the scintillator. These photons are directed out of the vacuum system through a light pipe (3) and are transferred to the photomultiplier (5). The photomultiplier multiplies the flashes of light and outputs a signal that can be used for imaging. Collector Bias Voltage The Detectors tab of the SEM Control panel allows you to adjust the collector bias voltage in the range 250 V to +400 V in steps of 1 V. This voltage generates a bias in front of the detector. The bias attracts the low energy SE electrons and accelerates them towards the detector. For all standard applications the collector bias voltage is usually set to 300 V. It is also possible to set the collector bias voltage to a negative value. This generates a field that deflects the secondary electrons, preventing them from reaching the scintillator and contributing to the signal. Backscattered electrons are not affected significantly by the negative bias voltage and reach the scintillator to contribute to image information. This allows the generation of a pseudo- backscattered image, which shows enhanced topographical information. Surface images that show enhanced topographical information can also be generated using BSE detectors, but they do not show the shadows that can be created using the SE2 detector. When using a positive collector bias voltage, surfaces that are tilted in the direction of the detector are emphasized, but there are no shadowing effects.

4 When using a negative collector bias voltage, the image shows enhanced topographical contrast, which arises mainly from the extreme shadowing effects. However, the fine surface details are less visible. Applications The SE2 detector can be used in the complete high-voltage range. The working distance has a significant effect on the efficiency of the SE2 detector. Shadowing effects occur when the working distance is too short and, if the specimen is too close to the final lens, most of the electrons will be deflected by the field of the electrostatic lens or move to the final lens itself. This means they cannot be detected by the SE2 detector. Depending on the specimen material and on the specimen geometry, a minimum working distance of approximately 4 mm should be used. Extreme signal loss is likely to occur if shorter working distances than this are used. Conversely, the SE2 detector is very good when used for imaging at long working distances. This is particularly important for low magnification imaging that is necessary for adjusting the orientation of the specimen holder or locating a specific area on the specimen. Optimal initial settings The following settings provide a good field of view for navigating on the specimen at low magnifications: Initial working distance in the range 10 mm to 20 mm. Acceleration voltage of approximately 10 kv.

5 RECOMMENDED SETTINGS FOR IMAGING WITH THE SE DETECTOR Parameter Value Recommended conditions 100 V to 30 kv In principle suitable for the entire high-voltage range. Acceleration voltage 1 kv to 5 kv 5 kv to 20 kv Low-voltage applications for the compensation of charges and for surface-sensitive imaging. Average voltage range. Suitable for many different applications. 20 kv to 30 kv Voltage range frequently used for analytical purposes. Working distance Collector voltage Mid column aperture Specimen tilt >=5 mm If the working distance is too short, shadowing effects will occur which diminish the efficiency of the detector. Below 20 kv, the SE electrons are absorbed by the field of the electrostatic lens. 4 mm to 6 mm For low-voltage applications (1 kv to 5 kv). 6 mm to 12 mm Useful for the average voltage range (5kV to 20kV). 12 mm to 30 mm Recommended only for low magnifications and to increase the depth of field. 300 V Standard value of the collector voltage. 0 V to 400 V Variation of the collector voltage at high magnifications to obtain the mixed signal V to 0 V For pseudo- BSE images. HV: 20 μm Recommended for high magnification and highresolution imaging. 30 μm The standard aperture is recommended for many applications. VP/EP: 100 μm For VP/EP mode only. Tilting the sample towards the detector increases collection efficiency.

6 OPERATION PROCEDURE!!! Log into the tool by using FOM, it will turn on the monitor. 1. Login to Zeiss Computer, choose SEM user, PW guest 2. Start SMART SEM USER INTERFACE Software. 3. Login to software [User: your SEM user ID, Password: your SEM password]. EM Server will automatically start up. Preparing the specimen holder IMPORTANT Contamination caused by fingerprints can lead to vacuum deterioration or prolonged pumping times. Always wear lint-free gloves when touching specimen, specimen holder or stage. 1. Attach the specimen to the stub by using conductive carbon, adhesive metal or carbon tape. Ensure that the specimen area to be analyzed is in good contact with the stub. This will ensure high electrical conductivity between the specimen and the stub. 2. Use the tweezers to insert the stub into the specimen holder. 3. Use the Allen key to tighten the location screw and secure the specimen. Loading the specimen chamber 1. Click the TV icon in the toolbar or press the Camera button on the keyboard. CAUTION: Risk of damaging the objective lens and/or your specimen Ensure not to hit the objective lens while driving the stage. Change to TV mode to observe the moving stage. 2. Select Tools/Goto Control Panel from the menu, the SEM Control panel opens. 3. Go to the Vacuum tab and click VENT to vent the specimen chamber. The same can be achieved by left-click on VAC indicator in the status bar and click VENT. 4. A message appears asking: Are you sure you want to vent? Confirm by clicking on YES. 5. When vented slowly open the chamber door. IMPORTANT: Keep the chamber door open as short as possible. All specimen holders are fitted with a dovetail fitting so that the position of the specimen holder is exactly defined. 6. Mount the specimen holder: a. Make sure that you place the dovetail fitting in the correct orientation onto the holding device on the specimen stage. b. Make sure that the flat side of the dovetail fitting of the specimen holder is flush with the milled edge of the stage. 7. When closing the chamber door check the chamber scope to ensure the specimen does not hit any components when it is introduced into the specimen chamber.

7 8. Carefully close the chamber door and click Pump in the SEM Control panel. The same can be achieved by left-click on VAC indicator in the status bar and click PUMP. Locating the Specimen 1. Move the specimen by using the dual joystick or by calling the Soft Joystick via Tools/Goto Panel/Soft Joystick. 2. Carefully move the specimen closer to the objective lens. The distance between objective lens and specimen surface should be less than about 10 mm. Switching on the Gun 1. In the Vacuum tab: Check that EHT Vac ready=yes is indicated. Alternatively, check that there is a green tick alongside VAC in the status bar. 2. Click GUN in the status bar and select Beam On from the pop-up menu. The gun is being run up. Switching on the EHT Note: EHT stands for Extra High Tension. This voltage has to be applied to the gun in order to make it emit electrons. 1. Watch the vacuum status messages on the Vacuum tab of the SEM Control panel. When the required vacuum has been reached you will see the message Vac Status = Ready. 2. Go to the Gun tab and set the acceleration voltage: a move the EHT Target slider to the desired value. Alternatively click EHT Target = and type in the desired value. 3. Switch on the EHT: a. Click EHT in the status bar. b. Select EHT On from the pop-up menu. The EHT will run up to the target value. The status bar buttons are merged and the ALL on. Generating an Image 1. Go to the Detectors tab and select SE2 from the Detectors dropdown list. 2. It is recommended that you select the SE2 detector to obtain the first image, as this detector provides a good signal-to-noise ratio even at large working distances. 3. Go to the Scanning tab and select a fast scan speed, e.g. Scan Speed = 1 from the drop-down list. : button appears. Now, the electron beam is

8 The lower the scan speed number, the faster the electron beam is scanned across the specimen. If the image displayed is noisy then change the Scan Speed to 2 or 3 but no higher than 3. This will allow you to move the specimen stage using X or Y controls on the joystick to a suitable location for imaging. 4. Set a low magnification e.g. Mag = 500 x: Click the Magnification/Focus icon in the toolbar. Use appropriate knob on the keyboard to adjust the magnification. The current magnification Mag is indicated in the status bar. 5. Set the focus: Use appropriate knob on the keyboard to adjust the focus. The current working distance (WD) is indicated in the status bar. 6. Adjust contrast and brightness by using the Brightness and Contrast knobs on the keyboard. Alternatively, use the Auto BC =On option for Signal Adjust in the Detectors tab. Click on the Contrast/Brightness icon in the toolbar. The current Brightness and Contrast are indicated in the status bar. 7. Select a detail on the specimen surface. Focus the detail. Use slower scan rate to reduce the signal-to-noise. 8. Adjust contrast and brightness again. Optimizing the Image 1. Set Coarse by toggling the Coarse/Fine button in the status bar. This can also be achieved by pushing <TAB> on the keyboard. 2. Slowly increase the magnification to the value required (e.g. Mag x). It is prudent to refocus the image if and when required whilst increasing the magnification. 3. Click the Reduced Raster icon or press the Reduced button on the keyboard. A small scan frame is shown. The image outside the scan frame is frozen. Use the mouse to change size, shape and position of the reduced raster area. 4. Focus the image in the reduced raster. 5. Aperture alignment (OPTIONAL): a. In the Apertures tab, tick the Focus Wobble checkbox. The Focus Wobble is a function that sweeps the focus of the objective lens backwards and forwards through the focus on the specimen plane. If the aperture is slightly misaligned, a lateral shift in the image will be observed.

9 Intensity of wobble can be adjusted by using the Wobble Amplitude scroll bar. Set Wobble Amplitude to between 60 % and 70 %. Wobble speed can be accelerated by ticking the Wobble Fast checkbox. b. Align the mid column aperture by carefully adjusting the X and Y micrometer gauges to eliminate lateral image shift. The specimen should be seen to pulsate and not to shift. c. Untick the Focus Wobble checkbox. 6. In the Scanning tab, set Scan Speed = Bring the image into focus at a higher magnification than the desired final magnification. Typically double the desired magnification. 8. Toggle to Fine in the status bar. Use Coarse and Fine mode of adjustment where appropriate. 9. Correct astigmatism: a. Select a detail (e.g. a mark or an edge) on the specimen surface. b. Click the Reduced Raster icon. Ensure the selected detail is in the raster c. Use Stigmator X and Stigmator Y knobs on the keyboard to obtain the sharpest possible image. This is similar to focusing. In order to obtain a well-focused image, it is prudent to adjust focus and stigmation a number of times before moving to the next stages of image recording. 10. Deactivate the reduced raster and reduce the magnification to the required value. Saving the Image 1. Once a steady contrast and brightness level has been obtained, stop the scan by clicking Photo. A red dot at the right bottom of the image area indicates that the image is frozen. OR You can skip that step and save the image by selecting File/Save Image or left-click on the image and choose Send to. You would have a choice to save you image in tiff, jpg, or bmp format. Choose tiff format since it also saves metadata. NOTE: Available image resolution - 512*384 - rapid scan mode *768 - normal operating mode * * High Resolution mode If desired image resolution could be changed. Goto Scanning ->Store resolution and select one from the drop down list. Finishing the Work Session 1. To finish your work session, switch off the EHT: a. Click the All: button in the status bar. b. Select EHT Off from the pop-up menu.

10 IMPORTANT: Tungsten Filament: It is recommended that when using a Tungsten filament, the gun is turned off when the system is not being used for any significant period of time. LaB6 Filament: It is recommended that when using a LaB6 filament that the gun is not turned off during its service life time. To obtain the maximum life time of a LaB6 emitter the EHT must be switched off when not in use and for long periods of time the gun should be switched to STANDBY. STANDBY (filament stays at 1 st peak I=1.4 A) EHT OFF GUN SHUTDOWN recommended for switching between users recommended for switching between samples recommended for long periods of time 2. Bring the stage down and go to Stage Initialization, wait until stage is stops at initial position. 3. Go to the Vacuum tab and click VENT to vent the specimen chamber. The same can be achieved by left-click on VAC indicator in the status bar and click VENT. A message appears asking: Are you sure you want to vent? Confirm by clicking on YES. 4. When vented roll out chamber door slowly, careful and watch out for gun. 5. Unload the sample holder. ALWAYS, use gloves to remove your samples. 6. Carefully close the chamber door and click Pump in the SEM Control panel. The same can be achieved by left-click on VAC indicator in the status bar and click PUMP. 7. Wait until VAC is ready. 8. Close the SmartSEM user interface by selecting File/Log Off from the menu. A window appears asking for confirmation to close the session. 9. Confirm by clicking on the Yes button. 10. Exit EM Server. IMPORTANT: Don t forget to log out from tool by using your FOM account.

11 RECOMMENDED SETTING FOR DIFFERENT TYPES OF SPECIMEN AND DETECTION MODES Conductive Specimens: General Microscopy of conductive specimens EHT =20 kv, Iprobe = 200 pa, WD = 15 mm Filament I set to 1 st peak for magnifications < 10 kx (gives longer filament life) Filament I set to 2 nd peak for magnifications > 10 kx (for better resolution) Detector = SE with Collector Bias > V Aperture = 30 µm Cycle time to reduce noise = 20 sec EDX of conductive specimens EHT = 20 kv for metals and minerals EHT = 7.5 kv for semiconductors and organic materials Filament I set to 1 st peak for Qualitative analysis Filament I set to 2 nd peak for Quantitative analysis Iprobe = 1000 pa or adjust for 30 % dead time WD = 8.5 mm for a 35 take off (elevation) angle Detector = BSD Aperture = 30 µm Cycle time = 20 sec to reduce noise or longer for X-ray mapping High resolution of conductive specimens EHT=30 kv, Iprobe = 10 pa, WD = 5 mm Filament I set to 2 nd peak Aperture = 30 µm Detector = SE with Collector Bias V Cycle time to reduce noise =1.3 min or longer Non-conductive Specimens: High Vacuum (HV) mode of non-conductive specimens EHT = 1 kv, Iprobe = 10 pa, WD = 5 rnm Filament I set to 2 nd peak Aperture = 30 µm

12 Detector = SE with Collector Bias V Use Scan Speed 3 with frames to average ~ 30 to reduce noise. Variable Pressure (VP) mode of non-conductive specimens EHT = 25 kv, Iprobe = 250 pa, WD = 8.5 mm Filament I set to 2 nd peak Aperture = 750 µm PLA = 100 µm Detector = BSD or VPSE Cycle time = 20 sec or longer Chamber pressure = 10 Pa for BSD detector, 40 Pa for VPSE detector or adjust to eliminate charge disturbances. XVP vacuum mode for hydrated samples EHT = 30 kv, Iprobe = 300 pa, WD = 8.5 mm Filament I set to 2 nd peak Aperture = 750 µm PLA = 100 µm Beam Sleeve = 1000 µm Detector = BSD or VPSE Cycle time = 20 sec or longer Chamber pressure = 600 Pa to slow dehydration Temperature of specimen to retain water at 600 Pa = 1 C. EP vacuum mode for hydrated samples EHT = 30 kv, Iprobe = 300 pa, WD = 5 mm Filament I set to 2nd peak Aperture = 750 µm PLA = 100 µm Beam Sleeve = 500 pm Detector = BSD or VPSE Cycle time = 20 sec or longer Chamber pressure = 2000 Pa to retain water at 20 C.

13 EVO SPECIFICATION Essential specifications EVO MA 10 and LS 10 Resolution Acceleration Voltage Field of View X-ray Analysis Pressure Range Chamber 5-Axes Motorised Specimen Stage Maximum Specimen Height Image Framestore 3 nm (2 30 kv - SE and W (LaB6) kv - BSD (VP mode) kv - 1nA, LaB6 20 nm (15 1 kv - SE and W (LaB6) 10 3 kv - SE < 7-1,000,000 x 6 mm at the Analytical Working Distance (AWD) 8.5 mm AWD and 35 take-off angle Pa (MA Series) Pa (LS Series) 310 mm (Ø) x 220 mm (h) X = 80mm Y = 100mm Z = 35mm T = 0-90 R = 360 (continuous) Stage control by mouse or optional joystick and control panel 100 mm max x2304 pixel, signal acquisition by integrating and averaging

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual Scanning Electron Microscope FEI INSPECT F50 Step by step operation manual Scanning Electron Microscope, FEI Inspect F50 FE-SEM-F Observation Flow Saving Data And Analysis Specimen preparation Error check

More information

JEOL 6500 User Manual

JEOL 6500 User Manual LOG IN to your session on the computer to the left of the microscope. Starting Conditions 1. Press Ctrl-Alt-Del and log on to the microscope computer. Click on JEOL PC SEM 6500 icon. Click yes if message

More information

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Warning Always adjust your specimen height before closing the chamber door to make sure your specimen will not hit the bottom of the lens;

More information

SEM Training Notebook

SEM Training Notebook SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside December 21, 2017 (rev. 3.4) 1 Before you begin Complete

More information

SEM Training Notebook

SEM Training Notebook SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside March 8, 2018 (rev. 3.5) 1 Before you begin Complete

More information

Using the Hitachi 3400-N VP-SEM

Using the Hitachi 3400-N VP-SEM Using the Hitachi 3400-N VP-SEM Opening the Chamber to Load Specimens (This may also be done later using the software) 1. Click the AIR button on the front of the machine: 2. Wait a few minutes until you

More information

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES RULES All users must go through a series of standard operation procedure training. For more information contact: Longlong Liao Teaching

More information

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 1. Utility Requirements a. System power is supplied by two 120 VAC/20 A circuits. When doing maintenance

More information

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE by Doug Bray Department of Biological Sciences University of Lethbridge Revised September, 2000 Note: The terms in bold in this document represent

More information

Model SU3500 Scanning Electron Microscope

Model SU3500 Scanning Electron Microscope Model SU3500 Scanning Electron Microscope Modified and Parts taken from Hitachi Easy Operation Guide. Before using the Model SU3500 SEM, be sure to read the [GENERAL SAFETY GUIDELINES] in the instruction

More information

1.1. Log on to the TUMI system (you cannot proceed further until this is done).

1.1. Log on to the TUMI system (you cannot proceed further until this is done). FEI DB235 SEM mode operation Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Sample loading 1.1. Log on to the TUMI system (you cannot proceed further until this is done). 1.2. The FIB software (xp)

More information

JEOL 6700 User Manual 05/18/2009

JEOL 6700 User Manual 05/18/2009 JEOL 6700 User Manual 05/18/2009 LOG IN to your session on the computer to the right of the microscope. Starting Conditions 1. Click the button and read the Penning Gauge to ensure that the microscope

More information

2. Raise HT to 200kVby following the procedure explained in 1.6.

2. Raise HT to 200kVby following the procedure explained in 1.6. JEOL 2100 MANUAL Quick check list 1. If needed, fill the reservoir with LN2 2. Raise HT to 200kVby following the procedure explained in 1.6. 3. Insert specimen holder into TEM (Insert holder in airlock,

More information

FE-SEM SU-8020 Operating manual (Preliminary version)

FE-SEM SU-8020 Operating manual (Preliminary version) FE-SEM SU-8020 Operating manual (Preliminary version) 2016/04/11 Seimitsu Bunseki sitsu lab. Starting up 1.Turn on the Display switch. Windows OS is starting up 2. Select the user SU-8000. 3. Click the

More information

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. Smith College August 2005 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check, 1 Specimen Insertion, 1 Startup, 2 Filament Saturation, 2 Beam Alignment,

More information

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400. Smith College August 2009 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check 1 Startup 1 Specimen Insertion 2 Filament Saturation 2 Beam Alignment

More information

Dickinson College Department of Geology

Dickinson College Department of Geology Dickinson College Department of Geology Title: Equipment: BASIC OPERATION OF THE SCANNING ELECTRON MICROSCOPE (SEM) JEOL JSM-5900 SCANNING ELECTRON MICROSCOPE Revision: 2.2 Effective Date: 1/29/2003 Author(s):

More information

1. Preliminary sample preparation

1. Preliminary sample preparation FEI Helios NanoLab 600 standard operating procedure Nicholas G. Rudawski ngr@ufl.edu (352) 392 3077 (office) (805) 252-4916 (cell) Last updated: 03/02/18 What this document provides: an overview of basic

More information

SECONDARY ELECTRON DETECTION

SECONDARY ELECTRON DETECTION SECONDARY ELECTRON DETECTION CAMTEC Workshop Presentation Haitian Xu June 14 th 2010 Introduction SEM Raster scan specimen surface with focused high energy e- beam Signal produced by beam interaction with

More information

SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS

SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS Robert Edward Lee Electron Microscopy Center Department of Anatomy and Neurobiology Colorado State University P T R Prentice Hall, Englewood Cliffs,

More information

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Amray 1830 SEM. All users are expected to have read and understood this document.

More information

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course)

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course) JEOL 7000F BASIC OPERATING INSTRUCTIONS-Ver.-2.0 Note: This is minimal operation checklist and does not replace the other reference manuals. Read the manual for Specimen Exchange (JEOL 7000 Specimen Exchange

More information

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Follow the procedures below when you use the Hitachi 7100 TEM. Starting Session 1. Turn on the cold

More information

Operation Guide. Hitachi S-3400N. Variable Pressure Scanning Electron Microscope. with. Deben Peltier Coolstage

Operation Guide. Hitachi S-3400N. Variable Pressure Scanning Electron Microscope. with. Deben Peltier Coolstage Operation Guide Hitachi S-3400N Variable Pressure Scanning Electron Microscope with Deben Peltier Coolstage www.deben.co.uk www.taltos.stanford.edu www.hitachi-hta.com Index Main Unit 3 Electron Optical

More information

SCIENTIFIC INSTRUMENT NEWS. Introduction. Design of the FlexSEM 1000

SCIENTIFIC INSTRUMENT NEWS. Introduction. Design of the FlexSEM 1000 SCIENTIFIC INSTRUMENT NEWS 2017 Vol. 9 SEPTEMBER Technical magazine of Electron Microscope and Analytical Instruments. Technical Explanation The FlexSEM 1000: A Scanning Electron Microscope Specializing

More information

Schottky Emission VP FE-SEM

Schottky Emission VP FE-SEM Schottky Emission VP FE-SEM Variable Pressure The Scanning Electron Microscope (SEM) has played an important role for many years for research and development of advanced materials in the leading edge of

More information

Scanning Electron Microscopy SEM. Warren Straszheim, PhD MARL, 23 Town Engineering

Scanning Electron Microscopy SEM. Warren Straszheim, PhD MARL, 23 Town Engineering Scanning Electron Microscopy SEM Warren Straszheim, PhD MARL, 23 Town Engineering wesaia@iastate.edu 515-294-8187 How it works Create a focused electron beam Accelerate it Scan it across the sample Map

More information

RAITH e-line OPERATING INSTRUCTIONS

RAITH e-line OPERATING INSTRUCTIONS RAITH e-line OPERATING INSTRUCTIONS 1) LOADING A SAMPLE a. Start the system i. On the Column PC (Right side monitor [R]), select the SmartSEM icon to on the desktop to begin the column software. ii. On

More information

Scanning electron microscope

Scanning electron microscope Scanning electron microscope 6 th CEMM workshop Maja Koblar, Sc. Eng. Physics Outline The basic principle? What is an electron? Parts of the SEM Electron gun Electromagnetic lenses Apertures Chamber and

More information

Functions of the SEM subsystems

Functions of the SEM subsystems Functions of the SEM subsystems Electronic column It consists of an electron gun and two or more electron lenses, which influence the path of electrons traveling down an evacuated tube. The base of the

More information

General information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument.

General information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument. JEOL JSM-7400F Table of contents General information.. 3 The operation panel. 4 The different sample holders and inserting the samples.. 5 Turning on the beam... 6 Stage map control... 8 Correcting astigmatism...

More information

Scanning electron microscope

Scanning electron microscope Scanning electron microscope 5 th CEMM workshop Maja Koblar, Sc. Eng. Physics Outline The basic principle? What is an electron? Parts of the SEM Electron gun Electromagnetic lenses Apertures Detectors

More information

Scanning Electron Microscope. Instructions for Use

Scanning Electron Microscope. Instructions for Use Scanning Electron Microscope Instructions for Use The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders are liable for damages.

More information

Secondary Electron Detector

Secondary Electron Detector Secondary Electron Detector Fig. 17 Everhart-Thornley Detector (Fig. 7-9, p. 215, Bozzola and Russell) Secondary electrons (SE) are attracted to Faraday cage because of its positive charge. Detector surface

More information

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only Version 1.0 Prepared by D. Turnbull February 21, 2007. Please submit any omissions to the Author Note: This SEM is a recent

More information

SEM OPERATION IN LOW VACUUM MODE

SEM OPERATION IN LOW VACUUM MODE SEM OPERATION IN LOW VACUUM MODE Instructions for JEOL 5800 LV The EVAC light of the SEM specimen chamber should be already lit when you approach the SEM & the SEM will have been left in the high vacuum

More information

FEI Quanta 200 ESEM Basic instructions

FEI Quanta 200 ESEM Basic instructions FEI Quanta 200 ESEM Basic instructions Desktop and then start the UI. If the computer has restarted and you need to login, Username: supervisor and Password: supervisor Log-in to the Microscope using the

More information

Scanning Electron Microscopy Basics and Applications

Scanning Electron Microscopy Basics and Applications Scanning Electron Microscopy Basics and Applications Dr. Julia Deuschle Stuttgart Center for Electron Microscopy MPI for Solid State Research Room: 1E15, phone: 0711/ 689-1193 email: j.deuschle@fkf.mpg.de

More information

MODULE I SCANNING ELECTRON MICROSCOPE (SEM)

MODULE I SCANNING ELECTRON MICROSCOPE (SEM) MODULE I SCANNING ELECTRON MICROSCOPE (SEM) Scanning Electron Microscope (SEM) Initially, the plan of SEM was offered by H. Stintzing in 1927 (a German patent application). His suggested procedure was

More information

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below.

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. http://staff.washington.edu/wpchan/if/cm100_inst.shtml Instructions for the Philips CM100 TEM and peripherals

More information

Figure 1 The Raith 150 TWO

Figure 1 The Raith 150 TWO RAITH 150 TWO SOP Figure 1 The Raith 150 TWO LOCATION: Raith 150 TWO room, Lithography area, NanoFab PRIMARY TRAINER: SECONDARY TRAINER: 1. OVERVIEW The Raith 150 TWO is an ultra high resolution, low voltage

More information

Standard Operating Procedure Hitachi UHR CFE SU8230 SEM

Standard Operating Procedure Hitachi UHR CFE SU8230 SEM Standard Operating Procedure Hitachi UHR CFE SU8230 Yale West Campus Materials Characterization Core ywcmatsci.yale.edu ESC II, Room E119F 810 West Campus Drive West Haven, CT 06516 Version 1.1, May 2016

More information

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right Quick Guide to Operating FEI Titan Themis G2 200 (S)TEM: TEM mode Susheng Tan Nanoscale Fabrication and Characterization Facility, University of Pittsburgh Office: M104/B01 Benedum Hall, 412-383-5978,

More information

ALTURA EDS. Rev. 0915

ALTURA EDS. Rev. 0915 ALTURA EDS Rev. 0915 Enable the Oxford PC Enable the Altura-EDS under Dual Beam Tools in Coral. Or enter your NETID and password directly into the Oxford PC. Warning: Ion-milling, GISs/microprobe, and

More information

Instructions for Tecnai a brief start up manual

Instructions for Tecnai a brief start up manual Instructions for Tecnai a brief start up manual Version 3.0, 8.12.2015 Manual of Tecnai 12 transmission electron microscope located at Aalto University's Nanomicroscopy Center. More information of Nanomicroscopy

More information

Oct. 30th- Nov. 1st, 2017

Oct. 30th- Nov. 1st, 2017 Thomas LaGrange, Ph.D. Faculty Lecturer and Senior Staff Scientist Electron Sources, Optics and Detectors SEM Doctoral Course MS-636 Oct. 30th- Nov. 1st, 2017 Summary Electron propagation is only possible

More information

Basic Users Manual for Tecnai-F20 TEM

Basic Users Manual for Tecnai-F20 TEM Basic Users Manual for Tecnai-F20 TEM NB: This document contains my personal notes on the operating procedure of the Tecnai F20 and may be used as a rough guide for those new to the microscope. It may

More information

Transmission Electron Microscopy 9. The Instrument. Outline

Transmission Electron Microscopy 9. The Instrument. Outline Transmission Electron Microscopy 9. The Instrument EMA 6518 Spring 2009 02/25/09 Outline The Illumination System The Objective Lens and Stage Forming Diffraction Patterns and Images Alignment and Stigmation

More information

Oct. 30th- Nov. 1st, 2017

Oct. 30th- Nov. 1st, 2017 Thomas LaGrange, Ph.D. Faculty Lecturer and Senior Staff Scientist Electron Sources, Optics and Detectors SEM Doctoral Course MS-636 Oct. 30th- Nov. 1st, 2017 Summary Electron propagation is only possible

More information

FEI Tecnai G 2 F20 Operating Procedures

FEI Tecnai G 2 F20 Operating Procedures FEI Tecnai G 2 F20 Operating Procedures 1. Startup (1) Sign-up in the microscope log-sheet. Please ensure you have written an account number for billing. (2) Log in to the computer: Login to your account

More information

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope Last updated on 1/8/2018 Jinsong Wu, jinsong-wu@northwestern.edu Aims: The aim of this lab is to familiarize you with basic TEM alignment

More information

University of Washington Molecular Analysis Facility

University of Washington Molecular Analysis Facility University of Washington Molecular Analysis Facility Apreo-S (Variable Pressure) is a Schottky Field Emission Scanning Electron Microscope (FESEM) that combines high- and low-voltage ultra-high resolution

More information

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack). FEI Tecnai F20 S/TEM: imaging in TEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 01/21/18 1. Filling the cold trap (if needed) 1.1. Prior to use, the cold trap needs

More information

Module 4B7: VLSI Design, Technology, and CAD. Scanning Electron Microscopical Examination of CMOS Integrated Circuit

Module 4B7: VLSI Design, Technology, and CAD. Scanning Electron Microscopical Examination of CMOS Integrated Circuit Engineering Tripos Part IIB FOURTH YEAR Module 4B7: VLSI Design, Technology, and CAD Laboratory Experiment Dr D Holburn and Mr B Breton Scanning Electron Microscopical Examination of CMOS Integrated Circuit

More information

(Refer Slide Time: 00:10)

(Refer Slide Time: 00:10) Fundamentals of optical and scanning electron microscopy Dr. S. Sankaran Department of Metallurgical and Materials Engineering Indian Institute of Technology, Madras Module 03 Unit-6 Instrumental details

More information

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier). JEOL 200CX operating procedure Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Specimen loading 1.1. Unlock the TUMI system. 1.2. Load specimen(s) into the holder. If using the double tilt holder, ensure

More information

Scanning Electron Microscope in Our Facility

Scanning Electron Microscope in Our Facility SEM Training Scanning Electron Microscope in Our Facility Specifications Table SEM ESEM FE-SEM-F FE-SEM-J FE-SEM-H FE-SEM-CZ Device name TM3030 Inspect S50 Inspect F50 JSM-7600 S-4700 Marlin compact Company

More information

This procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface.

This procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface. Tescan MIRA3 SEM: EDS using EDAX TEAM Nicholas G. Rudawski ngr@ufl.edu Cell: (805) 252-4916 Office: (352) 392-3077 Last updated: 12/04/17 This procedure assumes the user is already familiar with basic

More information

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface.

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface. FEI Tecnai F20 S/TEM: imaging in STEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 05/10/18 This document assumes the user is already familiar with basic operation

More information

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and

More information

Introduction of New Products

Introduction of New Products Field Emission Electron Microscope JEM-3100F For evaluation of materials in the fields of nanoscience and nanomaterials science, TEM is required to provide resolution and analytical capabilities that can

More information

Standard Operating Procedure Hitachi UHR CFE SU8230 SEM

Standard Operating Procedure Hitachi UHR CFE SU8230 SEM Standard Operating Procedure Hitachi UHR CFE SU8230 Yale West Campus Materials Characterization Core ywcmatsci.yale.edu ESC II, Room E119F 810 West Campus Drive West Haven, CT 06516 Version 1.5, February

More information

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center LEO 912 TEM Short Manual Prepared/copyrighted by RH Berg Danforth Plant Science Center Specimen holder [1] Never touch the holder (outside of the O-ring, double-headed arrow) because finger oils will contaminate

More information

Amray 3600 FESEM. Standard Operating Procedure. v2.2 modified by Bryan Cord. General Notes...3. Sample Loading...5. System Loading...

Amray 3600 FESEM. Standard Operating Procedure. v2.2 modified by Bryan Cord. General Notes...3. Sample Loading...5. System Loading... Amray 3600 FESEM Standard Operating Procedure v2.2 modified 5.13.13 by Bryan Cord Contents General Notes...3 Sample Loading...5 System Loading...8 Imaging...12 Saving Data...16 System Unloading...18 Troubleshooting...20

More information

STEM alignment procedures

STEM alignment procedures STEM alignment procedures Step 1. ASID alignment mode 1. Write down STD for TEM, and then open the ASID control window from dialogue. Also, start Simple imager viewer program on the Desktop. 2. Click on

More information

Introduction: Why electrons?

Introduction: Why electrons? Introduction: Why electrons? 1 Radiations Visible light X-rays Electrons Neutrons Advantages Not very damaging Easily focused Eye wonderful detector Small wavelength (Angstroms) Good penetration Small

More information

S200 Course LECTURE 1 TEM

S200 Course LECTURE 1 TEM S200 Course LECTURE 1 TEM Development of Electron Microscopy 1897 Discovery of the electron (J.J. Thompson) 1924 Particle and wave theory (L. de Broglie) 1926 Electromagnetic Lens (H. Busch) 1932 Construction

More information

SOP: EDAX Eagle III Microspot XRF

SOP: EDAX Eagle III Microspot XRF SOP: EDAX Eagle III Microspot XRF Page 1 of 6 SOP: EDAX Eagle III Microspot XRF 1. Scope 1.1 This document describes the standard operating procedure (SOP) for the EDAX Eagle III Microspot XRF. This X-ray

More information

Scanning Electron Microscopy. EMSE-515 F. Ernst

Scanning Electron Microscopy. EMSE-515 F. Ernst Scanning Electron Microscopy EMSE-515 F. Ernst 1 2 Scanning Electron Microscopy Max Knoll Manfred von Ardenne Manfred von Ardenne Principle of Scanning Electron Microscopy 3 Principle of Scanning Electron

More information

User Operation of JEOL 1200 EX II

User Operation of JEOL 1200 EX II **Log onto Computer** Open item program Start Up Procedure User Operation of JEOL 1200 EX II 1. If scope is not running, locate an electron microscopy technician (EMT) to find out why not. 2. Turn up brightness

More information

Zeiss LSM 880 Protocol

Zeiss LSM 880 Protocol Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height.

1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height. JEOL 2010F operating procedure Covers operation in STEM mode (See separate procedures for operation in TEM mode and operation of EDS system) Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 NOTE: this operating

More information

Quick and simple installation and no maintenance needed. 3 Times More affordable Than a normal SEM. Obtaining results in less than 4 minutes

Quick and simple installation and no maintenance needed. 3 Times More affordable Than a normal SEM. Obtaining results in less than 4 minutes INTRODUCTION We believe that every laboratory working in the field of nanotechnology needs an SEM, therefore we would like to introduce to you our IEM series of SEM. In short space of time, our device

More information

PERFORMANCE IN NANOSPACE PRODUCT OVERVIEW

PERFORMANCE IN NANOSPACE PRODUCT OVERVIEW PERFORMANCE IN NANOSPACE PRODUCT OVERVIEW TESCAN, a.s. is a Czech joint-stock company focused on research, development and manufacture of scientific instruments and laboratory equipment such as: scanning

More information

Introduction to Scanning Electron Microscopy

Introduction to Scanning Electron Microscopy Introduction to Scanning Electron Microscopy By: Brandon Cheney Ant s Leg Integrated Circuit Nano-composite This document was created as part of a Senior Project in the Materials Engineering Department

More information

5. The Scanning Electron Microscope

5. The Scanning Electron Microscope Physical Principles of Electron Microscopy 5. The Scanning Electron Microscope Ray Egerton University of Alberta and National Institute of Nanotechnology Edmonton, Canada www.tem-eels.ca regerton@ualberta.ca

More information

Topics 3b,c Electron Microscopy

Topics 3b,c Electron Microscopy Topics 3b,c Electron Microscopy 1.0 Introduction and History 1.1 Characteristic Information 2.0 Basic Principles 2.1 Electron-Solid Interactions 2.2 Electromagnetic Lenses 2.3 Breakdown of an Electron

More information

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I MSE 595T Basic Transmission Electron Microscopy TEM Imaging - I Purpose The purpose of this lab is to: 1. Make fine adjustments to the microscope alignment 2. Obtain a diffraction pattern 3. Obtain an

More information

Low Voltage Electron Microscope

Low Voltage Electron Microscope LVEM5 Low Voltage Electron Microscope Nanoscale from your benchtop LVEM5 Delong America DELONG INSTRUMENTS COMPACT BUT POWERFUL The LVEM5 is designed to excel across a broad range of applications in material

More information

2 How to operate the microscope/obtain an image

2 How to operate the microscope/obtain an image Morgagni Operating Instructions 50079 010912 2-1 2 ow to operate the microscope/obtain an image 2.1 Starting the microscope 2.1.1 Starting the microscope with several manually-operated steps 1. Turn on

More information

INTRODUCTION We believe that every laboratory working in the field of nanotechnology needs an SEM, therefore we would like to introduce to you our IEM

INTRODUCTION We believe that every laboratory working in the field of nanotechnology needs an SEM, therefore we would like to introduce to you our IEM INTRODUCTION We believe that every laboratory working in the field of nanotechnology needs an SEM, therefore we would like to introduce to you our IEM series of SEM. In short space of time, our device

More information

Zeiss LSM 780 Protocol

Zeiss LSM 780 Protocol Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

1. Specimen Holder Removal, Loading, and Insertion

1. Specimen Holder Removal, Loading, and Insertion OPERATION OF THE PHILIPS CM-200 FEG-TEM When not in use, the CM-200 should be in the MICROSCOPE ON configuration with the HIGH TENSION ON (illuminates green when the high tension is on).. The microscope

More information

Scanning Electron Microscopy

Scanning Electron Microscopy Scanning Electron Microscopy For the semiconductor industry A tutorial Titel Vorname Nachname Titel Jobtitle, Bereich/Abteilung Overview Scanning Electron microscopy Scanning Electron Microscopy (SEM)

More information

COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual

COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual 1 COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual ABOUT THIS MICROSCOPE (room HG01.240) The JEM-1400Flash

More information

Atomic Force Microscopy (Bruker MultiMode Nanoscope IIIA)

Atomic Force Microscopy (Bruker MultiMode Nanoscope IIIA) Atomic Force Microscopy (Bruker MultiMode Nanoscope IIIA) This operating procedure intends to provide guidance for general measurements with the AFM. For more advanced measurements or measurements with

More information

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope The procedures given below were written specifically for the FEI Tecnai G 2 Sphera microscope. Modifications will need to

More information

PERFORMANCE IN NANOSPACE PRODUCT OVERVIEW

PERFORMANCE IN NANOSPACE PRODUCT OVERVIEW PERFORMANCE IN NANOSPACE PRODUCT OVERVIEW TESCAN Brno, s.r.o. was established as subsidiary of a multi-national company TESCAN ORSAY HOLDING after the merger (August 2013) of Czech company TESCAN, a global

More information

05/20/14 1. Philips CM200T. Standby Condition

05/20/14 1. Philips CM200T. Standby Condition 05/20/14 1 Philips CM200T Standby Condition HT and filament off, HT setting at 200kV. RESET HOLDER, center sample tilt knobs, and remove sample. Mag ~ 5-10kX Objective and SA apertures out, C2 aperture

More information

Nanoscale Fabrication & Characterization Facility. Raith e-line EBL Users Guide (updated:aug 2 nd, 2017)

Nanoscale Fabrication & Characterization Facility. Raith e-line EBL Users Guide (updated:aug 2 nd, 2017) Nanoscale Fabrication & Characterization Facility Raith e-line EBL Users Guide (updated:aug 2 nd, 2017) Overview: The Raith e-line EBL system is designed to write features with critical dimensions as small

More information

NANO 703-Notes. Chapter 9-The Instrument

NANO 703-Notes. Chapter 9-The Instrument 1 Chapter 9-The Instrument Illumination (condenser) system Before (above) the sample, the purpose of electron lenses is to form the beam/probe that will illuminate the sample. Our electron source is macroscopic

More information

Low Voltage Electron Microscope. Nanoscale from your benchtop LVEM5. Delong America

Low Voltage Electron Microscope. Nanoscale from your benchtop LVEM5. Delong America LVEM5 Low Voltage Electron Microscope Nanoscale from your benchtop LVEM5 Delong America DELONG INSTRUMENTS COMPACT BUT POWERFUL The LVEM5 is designed to excel across a broad range of applications in material

More information

2.Components of an electron microscope. a) vacuum systems, b) electron guns, c) electron optics, d) detectors. Marco Cantoni, 021/

2.Components of an electron microscope. a) vacuum systems, b) electron guns, c) electron optics, d) detectors. Marco Cantoni, 021/ 2.Components of an electron microscope a) vacuum systems, b) electron guns, c) electron optics, d) detectors Marco Cantoni, 021/693.48.16 Centre Interdisciplinaire de Microscopie Electronique CIME MSE-603

More information

JEOL 5800LV SCANNING ELECTRON MICROSCOPE OPERATOR'S MANUAL

JEOL 5800LV SCANNING ELECTRON MICROSCOPE OPERATOR'S MANUAL JEOL 5800LV SCANNING ELECTRON MICROSCOPE OPERATOR'S MANUAL Michael N. Spilde Christopher Adcock Original version: October 1999 Revised: January 2006 Department of Earth and Planetary Sciences and Institute

More information

Olympus LEXT OLS 4000 Confocal Laser Microscope

Olympus LEXT OLS 4000 Confocal Laser Microscope Olympus LEXT OLS 4000 Confocal Laser Microscope The Olympus LEXT OLS4000 is a confocal microscope capable of taking high-resolution 3D images. The magnification (Optical and Digital) of this microscope

More information

How to use the Jeol 1010 TEM of GI (Liesbeth own GI version)

How to use the Jeol 1010 TEM of GI (Liesbeth own GI version) How to use the Jeol 1010 TEM of GI (Liesbeth own GI version) 1.Load the specimen Load a grid into the rod holder: USE ONLY THE TOP POSITION (blue arrow), Specimen selection on 1 (The rear one is only a

More information

Zeiss AxioImager.Z2 Brightfield Protocol

Zeiss AxioImager.Z2 Brightfield Protocol Zeiss AxioImager.Z2 Brightfield Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge

More information

JEOL 5DII. Operation introduction. By Serge Charlebois

JEOL 5DII. Operation introduction. By Serge Charlebois JEOL 5DII Operation introduction By Serge Charlebois July 2003 General procedure Loading the cassette in the load lock Selecting EOS mode, table and aperture Setting and maximising current Observation

More information

Strata DB235 FESEM FIB

Strata DB235 FESEM FIB Strata DB235 FESEM FIB Standard Operating Procedure Revision: 5.0 Last Updated: August 16/2016, revised by Li Yang Overview This document will provide a detailed operation procedure of the Focused Ion

More information

JEOL JEM-1400 Transmission Electron Microscope Operating Instructions

JEOL JEM-1400 Transmission Electron Microscope Operating Instructions JEOL JEM-1400 Transmission Electron Microscope Operating Instructions Anti-contamination device Objective aperture Objective aperture translation knobs Specimen holder Pump/air switch Left hand control

More information