Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014
|
|
- Terence Andrews
- 6 years ago
- Views:
Transcription
1 Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER Utility Requirements a. System power is supplied by two 120 VAC/20 A circuits. When doing maintenance to electrical components, disconnect power by toggling the power switch on the back of each console. b. The SEM currently uses two compressed gasses to operate: Nitrogen and Compressed Air. Nitrogen, the vent gas, is house-supplied. This gas is externally regulated to 10 psig. Compressed Air, which is house-supplied, is used to operate system valves. The gas is externally regulated to 80 psig and internally regulated to 60 psig. c. Cooling water for the diffusion pump is provided by a recirculating chiller. The temperature should be kept at 19 C and the operational pressure should be 10 psig. 2. General Precautions a. High Voltage: The SEM employs voltages of up to 30 kv, which may be fatal to users. Do not attempt to defeat the protective interlock systems. b. Sample Loading: Use caution when loading samples in the chamber. If the sample is loaded too high, it may damage the final aperture. When the sample stage is pushed in, be careful that it does not contact the collector screen. Take care when opening and closing the door, as there are pinch points. c. Collector Screen: When positioning the sample for imaging, be sure that it does not come into contact with the collector screen, as this could push it into the final lens, which will cause an electrical short and impede imaging. Use the internal camera system if you think that this could be a problem. d. Contamination: All samples must be completely clean and dry prior to being placed in the chamber to prevent outgassing contamination. No photoresist is allowed in the SEM. Other organic materials need staff approval, require a lower kv setting and a gold coating. No adhesives are allowed in the chamber except for carbon pads, carbon and copper tape, and silver paint (paint must be completely dry.) Other adhesives will contaminate the column and damage the SEM. e. Chamber Vacuum: It is desirable to leave the system under vacuum (STAND BY LED lit) when not in use. This helps to prevent moisture accumulation and will help to desorb contaminants from the inside walls of the chamber. f. Imaging Control Panel: Unless otherwise instructed, the knobs and buttons on the imaging control panel should not be used to control the SEM. These controls are functional and will conflict with the computer software, making it unpredictable and/or inaccurate. Only adept users of the SEM may use these controls. 3. Putting a Sample in the Chamber for Imaging a. Verify that the instrument is ready for use: I. The compressed air supply is at 80 psig and nitrogen is at 10 psig II. The chiller is at 19 C III. The STAND BY, V5, and DIFF PUMP red LEDs and TC1, TC2 and TC3 green LEDs are illuminated on the vacuum control panel (VCP) on the front of the column console If one or more of these conditions are not met, notify cleanroom staff immediately. 1
2 b. To open the chamber for sample loading, press the VENT button on the VCP. After about 15 seconds, the V2 LED will illuminate and chamber venting will be audible. After about another 15 seconds, the TC2 LED will turn off and, soon after, the door will be ready to open. Use caution when opening and shutting the chamber door. There are pinch points between the door and its hinges. c. When the chamber door has been opened, press the STAND BY button on the VCP to close the nitrogen vent valve (V2). d. Place the prepared sample mount into the mount socket and lightly tighten the set screw using the provided driver. Make sure that the sample and/or mount are positioned such that, when the door is closed, they do not contact the collector screen and/or the final lens. e. Close the chamber door and press the EVACUATE button on the VCP to pump down the chamber. Press against the chamber door during the first few seconds of pump-down to ensure that the door seals. After about a minute or two, the chamber will have pumped down and the green READY LED will have illuminated, indicating that the SEM is ready for imaging. 4. Initial Instrument Setup Do not use the buttons or knobs on the Imaging Console control panel unless otherwise told in this SOP to do so. See the SEMTech Solutions SEM Interface Quick Reference Guide for an explanation of the software interface tool bar icons and scan settings. a. Press the green CONSOLE ON button on the right end of the Imaging Console control panel (ICCP). b. Ensure that the FINAL LENS button on the imaging console is on, the CAMERA POWER button is off, the Image Shift, Stigmator and Beam Alignment control knobs on the front panel are set to the 12 o clock position, and that the CONTRAST, BRIGHTNESS, X, Y, SIZE TILT, FOCUS, and GAMMA knobs are all fully counter-clockwise. c. Enter the appropriate information in the LOG BOOK as you proceed. Note the hours on the hour meter on top of the imaging console. If the reading is not the same as what was entered in the log book by the previous user, contact cleanroom staff immediately. d. Start the SEM Interface software (green STS icon.) e. Using the software, set the Gun Voltage to the desired voltage to begin acquiring an image. Application Non-conductive samples Biological specimens, delicate, or heat sensitive samples Non-sensitive, conductive samples Voltage Range 100 to 5000 V 5000 to V to V 2
3 A lower voltage setting usually improves the surface image but also adds noise to the detector signal. For maximum resolution and high magnification work, use 20 kv or above, if the sample allows. The higher the voltage the better the resolution, but the greater the heat that is generated on the specimen. f. If necessary, adjust the BIAS VOLTAGE knob (at knee height on the imaging console, behind a sliding door) to bring the Emission Current to within spec: 45 to 50 µa for voltages up to 10 kv; and 90 to 100 µa for voltages from 10 to 30 kv. g. After about 5 minutes, Check the filament current setting by turning the FILAMENT CURRENT knob (at knee height on the imaging console, behind a sliding door) to about 40, and then turning it back up to the saturation point. See the graph to the right as an aid to find the saturation point; be aware of the false peak. h. If necessary, readjust the BIAS VOLTAGE knob to bring the Emission Current to within spec, as defined in step 4f. i. Using the Mag and Mag + buttons, adjust the magnification to about 50X. j. Use the Condenser button to set the spot size to 5.0 (a higher spot size number correlates to a smaller beam diameter.) k. Adjust the X and Y Beam Alignment sliders to get the brightest image. CONTRAST may have to be decreased to optimize beam alignment. l. Manipulate the stage adjustments on the chamber door to bring the sample into view. Be careful not to contact the collector screen with your sample. Use the (Toggle Chamber View) button on the STS software tool bar to view the inside of the chamber. Press the CAMERA POWER button on the ICCP to provide power to the chamber view camera; press it again to turn it off when you are done. m. Focus the image. Press the AUTO FOCUS button on the ICCP if necessary. n. Adjust the Brightness and Contrast to improve the image visibility. o. Degauss the final lens by clicking the DF icon in the tool bar. Degaussing should be done in approximately 15 minute intervals during operation of the SEM. p. Adjust the Condenser to a spot size of 12. q. Adjust the COLL. BIAS knob on the ICCP to a voltage of XXXXX. r. Check alignment of the apertures. I. Adjust the magnification to 10,000X. II. Click the Partial button for a partial field view. III. Click the W icon in the tool bar to wobble the image. If the image is shifting up and down and/ or side to side, contact cleanroom staff immediately if the final aperture alignment seems to be out of adjustment. A pulsing image is considered normal. IV. Click the W icon again to turn off the wobbler. s. Adjust the Condenser to a spot size of 20. t. Adjust the COLL. BIAS knob on the ICCP to a voltage of XXXXX. u. Correct the beam astigmatism. I. Adjust the magnification to 20,000X. Focus on a round feature for best results. 3
4 II. Click the Partial button for a partial field view. III. Increase the Contrast, and decrease the Brightness. Focus if necessary. IV. Coarse adjust the Stig X to get the sharpest image possible. Repeat with fine adjustment. V. Refocus and repeat the previous step with Stig Y. VI. Press the AUTO STIG. button on the ICCP if necessary. 5. Sample Image Final Adjustments and Image Capture a. If necessary, readjust the Gun Voltage. If a significant change in the voltage setting is made, the FILAMENT CURRENT and BIAS VOLTAGE must be readjusted, as described in steps 4g and 4h, respectively. b. Make necessary adjustments to the Stage, Mag +/, Brightness, Contrast, Focus and any of the other features offered in the software, such as Tilt Focus or Line Scan, as needed. No particular order is required. Switching between scan rates other than rapid scan (RS) will make this process easier. Using the Partial field view can also speed up the process of optimizing your final image, especially when using slower scan rates. Additionally, the following ICCP controls may be used to enhance your image (these controls have no software equivalent): Control GAMMA DIFF LOW PASS FILTER INVERT Description Gamma Control A signal processing feature to bring out the information in the dark regions without excessive loss of information in the light regions. Used primarily to view down holes. The knob should be fully counter-clockwise when not in use. Differential Amplifier button activates the differential (derivative) amplifier, which processes the image to give increased edge contrast. (Slow Scan mode only) LOW PASS FILTER button applies filters to reduce noise from the video signal. (Slow Scan mode only) INVERT button inverts the gray scale of the video signal. c. If necessary, adjust the Condenser (spot size). Generally speaking, use a large spot size (5 or lower) for low magnification work under 1000X magnification. Adjust the Smaller spot size (5 to 20) is used for higher magnifications. d. If necessary, adjust the COLL. BIAS knob on the ICCP. As spot size is reduced (higher number), the collector bias voltage must be increased (and visa versa,) as fewer secondary electrons are generated with smaller spot sizes. e. Select the Scan Rate of your choice. f. Click the Freeze button to prepare for saving your image. The Freeze button will flash during this process and turn solid blue when complete. The time to completion depends on the scan rate. g. Save your image by selecting File Save As... and selecting your preferred save location. Save your images in your group s folder in the Documents folder, preferably in a sub-folder with your name. Images saved to the Desktop or in the Documents top folder will be deleted! 4
5 6. Venting the Chamber for Sample Removal and Putting the SEM in Standby Mode a. Set the Gun Voltage to 0 V. b. Close the SEM Interface software. c. Enter the reading from the hour meter in the LOG BOOK d. Press the red POWER OFF button on the right end of the imaging console. e. Press the VENT button on the VCP. After about 15 seconds, the V2 LED will illuminate and chamber venting will be audible. After about another 15 seconds, the TC2 LED will turn off and, soon after, the door will be ready to open. Use caution when opening and shutting the chamber door. There are pinch points between the door and its hinges. f. When the chamber door has been opened, press the STAND BY button on the VCP to close the nitrogen vent valve (V2). g. Remove the sample mount using the provided driver. h. Close the chamber door and press the EVACUATE button on the VCP to pump down the chamber. Press against the chamber door during the first few seconds of pump-down to ensure that the door seals. After about a minute or two, the chamber will have pumped down and the green READY LED will have illuminated. i. Press the STAND BY button on the VCP to put the SEM in standby mode. 5
Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016
Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Amray 1830 SEM. All users are expected to have read and understood this document.
More informationOPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge
OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE by Doug Bray Department of Biological Sciences University of Lethbridge Revised September, 2000 Note: The terms in bold in this document represent
More informationJEOL 6500 User Manual
LOG IN to your session on the computer to the left of the microscope. Starting Conditions 1. Press Ctrl-Alt-Del and log on to the microscope computer. Click on JEOL PC SEM 6500 icon. Click yes if message
More informationUsing the Hitachi 3400-N VP-SEM
Using the Hitachi 3400-N VP-SEM Opening the Chamber to Load Specimens (This may also be done later using the software) 1. Click the AIR button on the front of the machine: 2. Wait a few minutes until you
More informationSEM Training Notebook
SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside March 8, 2018 (rev. 3.5) 1 Before you begin Complete
More informationJEOL 6700 User Manual 05/18/2009
JEOL 6700 User Manual 05/18/2009 LOG IN to your session on the computer to the right of the microscope. Starting Conditions 1. Click the button and read the Penning Gauge to ensure that the microscope
More informationSEM Training Notebook
SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside December 21, 2017 (rev. 3.4) 1 Before you begin Complete
More informationOperating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.
Smith College August 2009 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check 1 Startup 1 Specimen Insertion 2 Filament Saturation 2 Beam Alignment
More informationBasic Users Manual for Tecnai-F20 TEM
Basic Users Manual for Tecnai-F20 TEM NB: This document contains my personal notes on the operating procedure of the Tecnai F20 and may be used as a rough guide for those new to the microscope. It may
More informationOperating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah
Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Follow the procedures below when you use the Hitachi 7100 TEM. Starting Session 1. Turn on the cold
More informationJSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES
JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES RULES All users must go through a series of standard operation procedure training. For more information contact: Longlong Liao Teaching
More information2. Raise HT to 200kVby following the procedure explained in 1.6.
JEOL 2100 MANUAL Quick check list 1. If needed, fill the reservoir with LN2 2. Raise HT to 200kVby following the procedure explained in 1.6. 3. Insert specimen holder into TEM (Insert holder in airlock,
More information1. Preliminary sample preparation
FEI Helios NanoLab 600 standard operating procedure Nicholas G. Rudawski ngr@ufl.edu (352) 392 3077 (office) (805) 252-4916 (cell) Last updated: 03/02/18 What this document provides: an overview of basic
More informationScanning Electron Microscope FEI INSPECT F50. Step by step operation manual
Scanning Electron Microscope FEI INSPECT F50 Step by step operation manual Scanning Electron Microscope, FEI Inspect F50 FE-SEM-F Observation Flow Saving Data And Analysis Specimen preparation Error check
More informationSEM OPERATION IN LOW VACUUM MODE
SEM OPERATION IN LOW VACUUM MODE Instructions for JEOL 5800 LV The EVAC light of the SEM specimen chamber should be already lit when you approach the SEM & the SEM will have been left in the high vacuum
More informationOperating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.
Smith College August 2005 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check, 1 Specimen Insertion, 1 Startup, 2 Filament Saturation, 2 Beam Alignment,
More informationPlease follow these instructions for use of the Philips CM100 TEM. Adopted from website below.
Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. http://staff.washington.edu/wpchan/if/cm100_inst.shtml Instructions for the Philips CM100 TEM and peripherals
More informationJEOL JEM-1400 Transmission Electron Microscope Operating Instructions
JEOL JEM-1400 Transmission Electron Microscope Operating Instructions Anti-contamination device Objective aperture Objective aperture translation knobs Specimen holder Pump/air switch Left hand control
More informationBasic Operating Instructions for Strata Dual Beam 235 FIB/SEM
Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Warning Always adjust your specimen height before closing the chamber door to make sure your specimen will not hit the bottom of the lens;
More informationGeneral information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument.
JEOL JSM-7400F Table of contents General information.. 3 The operation panel. 4 The different sample holders and inserting the samples.. 5 Turning on the beam... 6 Stage map control... 8 Correcting astigmatism...
More information1.1. Log on to the TUMI system (you cannot proceed further until this is done).
FEI DB235 SEM mode operation Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Sample loading 1.1. Log on to the TUMI system (you cannot proceed further until this is done). 1.2. The FIB software (xp)
More informationFE-SEM SU-8020 Operating manual (Preliminary version)
FE-SEM SU-8020 Operating manual (Preliminary version) 2016/04/11 Seimitsu Bunseki sitsu lab. Starting up 1.Turn on the Display switch. Windows OS is starting up 2. Select the user SU-8000. 3. Click the
More informationSOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only
SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only Version 1.0 Prepared by D. Turnbull February 21, 2007. Please submit any omissions to the Author Note: This SEM is a recent
More informationCheck that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course)
JEOL 7000F BASIC OPERATING INSTRUCTIONS-Ver.-2.0 Note: This is minimal operation checklist and does not replace the other reference manuals. Read the manual for Specimen Exchange (JEOL 7000 Specimen Exchange
More informationDickinson College Department of Geology
Dickinson College Department of Geology Title: Equipment: BASIC OPERATION OF THE SCANNING ELECTRON MICROSCOPE (SEM) JEOL JSM-5900 SCANNING ELECTRON MICROSCOPE Revision: 2.2 Effective Date: 1/29/2003 Author(s):
More information1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).
JEOL 200CX operating procedure Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Specimen loading 1.1. Unlock the TUMI system. 1.2. Load specimen(s) into the holder. If using the double tilt holder, ensure
More informationMSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope
MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope Last updated on 1/8/2018 Jinsong Wu, jinsong-wu@northwestern.edu Aims: The aim of this lab is to familiarize you with basic TEM alignment
More informationAmray 3600 FESEM. Standard Operating Procedure. v2.2 modified by Bryan Cord. General Notes...3. Sample Loading...5. System Loading...
Amray 3600 FESEM Standard Operating Procedure v2.2 modified 5.13.13 by Bryan Cord Contents General Notes...3 Sample Loading...5 System Loading...8 Imaging...12 Saving Data...16 System Unloading...18 Troubleshooting...20
More informationModel SU3500 Scanning Electron Microscope
Model SU3500 Scanning Electron Microscope Modified and Parts taken from Hitachi Easy Operation Guide. Before using the Model SU3500 SEM, be sure to read the [GENERAL SAFETY GUIDELINES] in the instruction
More informationSTEM alignment procedures
STEM alignment procedures Step 1. ASID alignment mode 1. Write down STD for TEM, and then open the ASID control window from dialogue. Also, start Simple imager viewer program on the Desktop. 2. Click on
More informationFull-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right
Quick Guide to Operating FEI Titan Themis G2 200 (S)TEM: TEM mode Susheng Tan Nanoscale Fabrication and Characterization Facility, University of Pittsburgh Office: M104/B01 Benedum Hall, 412-383-5978,
More informationFEI Tecnai G 2 F20 Operating Procedures
FEI Tecnai G 2 F20 Operating Procedures 1. Startup (1) Sign-up in the microscope log-sheet. Please ensure you have written an account number for billing. (2) Log in to the computer: Login to your account
More informationLEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center
LEO 912 TEM Short Manual Prepared/copyrighted by RH Berg Danforth Plant Science Center Specimen holder [1] Never touch the holder (outside of the O-ring, double-headed arrow) because finger oils will contaminate
More informationZEISS EVO SOP. May 2017 ELECTRON OPTICS
ZEISS EVO SOP May 2017 ELECTRON OPTICS The patented EVO column is the area of the SEM, where electrons are emitted, accelerated, deflected, focused, and scanned. Main characteristics of the EVO optics
More informationInstructions for Tecnai a brief start up manual
Instructions for Tecnai a brief start up manual Version 3.0, 8.12.2015 Manual of Tecnai 12 transmission electron microscope located at Aalto University's Nanomicroscopy Center. More information of Nanomicroscopy
More information1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).
FEI Tecnai F20 S/TEM: imaging in TEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 01/21/18 1. Filling the cold trap (if needed) 1.1. Prior to use, the cold trap needs
More informationUser Operation of JEOL 1200 EX II
**Log onto Computer** Open item program Start Up Procedure User Operation of JEOL 1200 EX II 1. If scope is not running, locate an electron microscopy technician (EMT) to find out why not. 2. Turn up brightness
More informationFEI Quanta 200 ESEM Basic instructions
FEI Quanta 200 ESEM Basic instructions Desktop and then start the UI. If the computer has restarted and you need to login, Username: supervisor and Password: supervisor Log-in to the Microscope using the
More informationProtective Equipment Nitrile gloves for handling sample holder and safety glasses for filling liquid nitrogen dewar.
Emergency Information: 1. Medical Emergencies: Contact 911 and McGill Security 514.398.3000 2. Leave TEM as is. Do NOT shut down the vacuum system. 3. If possible, turn off High Tension and Close Column
More informationZeiss LSM 880 Protocol
Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.
More informationJEOL 5DII. Operation introduction. By Serge Charlebois
JEOL 5DII Operation introduction By Serge Charlebois July 2003 General procedure Loading the cassette in the load lock Selecting EOS mode, table and aperture Setting and maximising current Observation
More informationZeiss LSM 780 Protocol
Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.
More information2 How to operate the microscope/obtain an image
Morgagni Operating Instructions 50079 010912 2-1 2 ow to operate the microscope/obtain an image 2.1 Starting the microscope 2.1.1 Starting the microscope with several manually-operated steps 1. Turn on
More informationOperation Guide. Hitachi S-3400N. Variable Pressure Scanning Electron Microscope. with. Deben Peltier Coolstage
Operation Guide Hitachi S-3400N Variable Pressure Scanning Electron Microscope with Deben Peltier Coolstage www.deben.co.uk www.taltos.stanford.edu www.hitachi-hta.com Index Main Unit 3 Electron Optical
More information1. Specimen Holder Removal, Loading, and Insertion
OPERATION OF THE PHILIPS CM-200 FEG-TEM When not in use, the CM-200 should be in the MICROSCOPE ON configuration with the HIGH TENSION ON (illuminates green when the high tension is on).. The microscope
More informationTecnai T12 Operating Procedures
Tecnai T12 Operating Procedures I. Initial Procedures 1 II. Accelerating Voltage 3 III. Specimen Loading and Holder Insertion/Removal 3 IV. Emission Current 7 V. Alignment 7 VI. Camera Control and Imaging
More informationMSE 460 TEM Lab 4: Bright/Dark Field Imaging Operation
MSE 460 TEM Lab 4: Bright/Dark Field Imaging Operation Last updated on 1/8/2018 Jinsong Wu, jinsong-wu@northwestern.edu Aims: The aim of this lab is to familiarize you with bright/dark field imaging operation.
More informationScanning Electron Microscopy SEM. Warren Straszheim, PhD MARL, 23 Town Engineering
Scanning Electron Microscopy SEM Warren Straszheim, PhD MARL, 23 Town Engineering wesaia@iastate.edu 515-294-8187 How it works Create a focused electron beam Accelerate it Scan it across the sample Map
More informationProcedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope
Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope The procedures given below were written specifically for the FEI Tecnai G 2 Sphera microscope. Modifications will need to
More informationThis document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface.
FEI Tecnai F20 S/TEM: imaging in STEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 05/10/18 This document assumes the user is already familiar with basic operation
More informationZeiss AxioImager.Z2 Brightfield Protocol
Zeiss AxioImager.Z2 Brightfield Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge
More informationMSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I
MSE 595T Basic Transmission Electron Microscopy TEM Imaging - I Purpose The purpose of this lab is to: 1. Make fine adjustments to the microscope alignment 2. Obtain a diffraction pattern 3. Obtain an
More informationHow to use the Jeol 1010 TEM of GI (Liesbeth own GI version)
How to use the Jeol 1010 TEM of GI (Liesbeth own GI version) 1.Load the specimen Load a grid into the rod holder: USE ONLY THE TOP POSITION (blue arrow), Specimen selection on 1 (The rear one is only a
More informationSAMUEL ROBERTS NOBLE ELECTRON MICROSCOPY LABORATORY. Operating Procedures for the Zeiss 9 S-2. Transmission Electron Microscope
1 SAMUEL ROBERTS NOBLE ELECTRON MICROSCOPY LABORATORY Operating Procedures for the Zeiss 9 S-2 Transmission Electron Microscope Prepared by Dr. Scott D. Russell Department of Botany and Microbiology September,
More informationSCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS
SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS Robert Edward Lee Electron Microscopy Center Department of Anatomy and Neurobiology Colorado State University P T R Prentice Hall, Englewood Cliffs,
More informationHoriba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer
Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and
More informationSTANDARD OPERATING PROCEDURE: JEOL TEM-2100
STANDARD OPERATING PROCEDURE: JEOL TEM-2100 Purpose of this Instrument: Essential tool for structural characterization of natural or synthesized nanostructures. Location: WVU - Engineering Sciences Building
More informationLSM 710 Confocal Microscope Standard Operation Protocol
LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.
More informationCAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE
University of Virginia ITC Academic Computing Health Sciences CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE Introduction The Olympus BH-2 microscope in ACHS s microscope lab has objectives from
More informationRAITH e-line OPERATING INSTRUCTIONS
RAITH e-line OPERATING INSTRUCTIONS 1) LOADING A SAMPLE a. Start the system i. On the Column PC (Right side monitor [R]), select the SmartSEM icon to on the desktop to begin the column software. ii. On
More informationFigure 1 The Raith 150 TWO
RAITH 150 TWO SOP Figure 1 The Raith 150 TWO LOCATION: Raith 150 TWO room, Lithography area, NanoFab PRIMARY TRAINER: SECONDARY TRAINER: 1. OVERVIEW The Raith 150 TWO is an ultra high resolution, low voltage
More informationSCIENTIFIC INSTRUMENT NEWS. Introduction. Design of the FlexSEM 1000
SCIENTIFIC INSTRUMENT NEWS 2017 Vol. 9 SEPTEMBER Technical magazine of Electron Microscope and Analytical Instruments. Technical Explanation The FlexSEM 1000: A Scanning Electron Microscope Specializing
More informationSOP: EDAX Eagle III Microspot XRF
SOP: EDAX Eagle III Microspot XRF Page 1 of 6 SOP: EDAX Eagle III Microspot XRF 1. Scope 1.1 This document describes the standard operating procedure (SOP) for the EDAX Eagle III Microspot XRF. This X-ray
More informationThis procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface.
Tescan MIRA3 SEM: EDS using EDAX TEAM Nicholas G. Rudawski ngr@ufl.edu Cell: (805) 252-4916 Office: (352) 392-3077 Last updated: 12/04/17 This procedure assumes the user is already familiar with basic
More information05/20/14 1. Philips CM200T. Standby Condition
05/20/14 1 Philips CM200T Standby Condition HT and filament off, HT setting at 200kV. RESET HOLDER, center sample tilt knobs, and remove sample. Mag ~ 5-10kX Objective and SA apertures out, C2 aperture
More informationLeica SP8 TCS Users Manual
Version : 07/08/0 Leica SP8 TCS Users Manual Start up:. Turn the PC Microscope, Scanner Power, Laser Power, and the Laser Emission key to on (bottom right of desk).. Turn on the fluorescent lamp (top left
More informationThings to check before start-up.
Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your
More informationJEOL JEM 2010 TRAINING TRANSMISSION ELECTRON MICROSCOPE USER MANUAL
JEOL JEM 2010 TRAINING TRANSMISSION ELECTRON MICROSCOPE USER MANUAL Version 5.1 EM Facility CMSE-SEF Massachusetts Institution of Technology TABLE OF CONTENTS 1. Specifications...2 1.1 Performance...2
More informationRENISHAW INVIA RAMAN SPECTROMETER
STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw invia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples
More information1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height.
JEOL 2010F operating procedure Covers operation in STEM mode (See separate procedures for operation in TEM mode and operation of EDS system) Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 NOTE: this operating
More informationTransmission Electron Microscopy 9. The Instrument. Outline
Transmission Electron Microscopy 9. The Instrument EMA 6518 Spring 2009 02/25/09 Outline The Illumination System The Objective Lens and Stage Forming Diffraction Patterns and Images Alignment and Stigmation
More informationBX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions
BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions Starting up: Schematic of Olympus BX-61. 1. Turn on Olympus microscope power box (left of microscope) with toggle switch on
More informationStrata DB235 FESEM FIB
Strata DB235 FESEM FIB Standard Operating Procedure Revision: 5.0 Last Updated: August 16/2016, revised by Li Yang Overview This document will provide a detailed operation procedure of the Focused Ion
More informationS200 Course LECTURE 1 TEM
S200 Course LECTURE 1 TEM Development of Electron Microscopy 1897 Discovery of the electron (J.J. Thompson) 1924 Particle and wave theory (L. de Broglie) 1926 Electromagnetic Lens (H. Busch) 1932 Construction
More informationALTURA EDS. Rev. 0915
ALTURA EDS Rev. 0915 Enable the Oxford PC Enable the Altura-EDS under Dual Beam Tools in Coral. Or enter your NETID and password directly into the Oxford PC. Warning: Ion-milling, GISs/microprobe, and
More informationCOMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual
1 COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual ABOUT THIS MICROSCOPE (room HG01.240) The JEM-1400Flash
More informationCM20 USER GUIDE. Duncan Alexander, CIME 2010
CM20 USER GUIDE Duncan Alexander, CIME 2010 CM20 START UP AND CHECK LIST 2 SPECIMEN EXCHANGE 5 - REMOVING SAMPLE HOLDER 6 - INSERTING SAMPLE HOLDER 7 TURNING ON HT 8 STARTING THE FILAMENT 9 GUN TILT ALIGNMENT
More informationTRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO
TRAINING MANUAL Multiphoton Microscopy LSM 510 META-NLO September 2010 Multiphoton Microscopy Training Manual Multiphoton microscopy is only available on the LSM 510 META-NLO system. This system is equipped
More informationIntroduction: Why electrons?
Introduction: Why electrons? 1 Radiations Visible light X-rays Electrons Neutrons Advantages Not very damaging Easily focused Eye wonderful detector Small wavelength (Angstroms) Good penetration Small
More informationModule 4B7: VLSI Design, Technology, and CAD. Scanning Electron Microscopical Examination of CMOS Integrated Circuit
Engineering Tripos Part IIB FOURTH YEAR Module 4B7: VLSI Design, Technology, and CAD Laboratory Experiment Dr D Holburn and Mr B Breton Scanning Electron Microscopical Examination of CMOS Integrated Circuit
More informationIntroduction to Scanning Electron Microscopy
Introduction to Scanning Electron Microscopy By: Brandon Cheney Ant s Leg Integrated Circuit Nano-composite This document was created as part of a Senior Project in the Materials Engineering Department
More informationLSM 800 Confocal Microscope Standard Operation Protocol
LSM 800 Confocal Microscope Standard Operation Protocol Turning on the system 1. Switch on the Main switch (labeled 1 and 2 ) mounted on the wall. 2. Turn the Laser Key (labeled 3 ) 90 clockwise for power
More informationFunctions of the SEM subsystems
Functions of the SEM subsystems Electronic column It consists of an electron gun and two or more electron lenses, which influence the path of electrons traveling down an evacuated tube. The base of the
More informationJEOL 2010 FasTEM & DigitalMicrograph User's Guide
JEOL 2010 FasTEM & DigitalMicrograph User's Guide Electron Microscopy Laboratory Instititute of Materials Science University of Connecticut The purpose of this manual is to remind you of the essential
More informationTraining Guide for Leica SP8 Confocal/Multiphoton Microscope
Training Guide for Leica SP8 Confocal/Multiphoton Microscope LAS AF v3.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON power switch for epifluorescence
More informationMicroscopy from Carl Zeiss
Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path
More informationStandard Operating Procedure Hitachi UHR CFE SU8230 SEM
Standard Operating Procedure Hitachi UHR CFE SU8230 Yale West Campus Materials Characterization Core ywcmatsci.yale.edu ESC II, Room E119F 810 West Campus Drive West Haven, CT 06516 Version 1.1, May 2016
More informationQuick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center
Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue
More informationJeol JEM Responsible personell: Endy ( ) Online booking is compulsory!
Jeol JEM 1230 Responsible personell: Endy (45279377) Online booking is compulsory! After training you will have access to working alone on the instrument. All insertion of samples is done by responsible
More informationLSM 780 Confocal Microscope Standard Operation Protocol
LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch
More informationGuide to Confocal 5. Starting session
Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope
More informationMicro-Image Capture 8 Installation Instructions & User Guide
Micro-Image Capture 8 Installation Instructions & User Guide Software installation: Micro-Image Capture Software 1. Load Micro-Image Capture software CD onto host PC. Auto Run should start driver/software
More informationSecondary Electron Detector
Secondary Electron Detector Fig. 17 Everhart-Thornley Detector (Fig. 7-9, p. 215, Bozzola and Russell) Secondary electrons (SE) are attracted to Faraday cage because of its positive charge. Detector surface
More informationWITec Alpha 300R Quick Operation Summary October 2018
WITec Alpha 300R Quick Operation Summary October 2018 This document is frequently updated if you feel information should be added, please indicate that to the facility manager (currently Philip Carubia,
More informationR I T. Title: Wyko RST Plus. Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 05/23/06 1 SCOPE 2 REFERENCE DOCUMENTS
Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Wyko RST Plus. All users are expected to have read and understood this document.
More informationMarine Invertebrate Zoology Microscope Introduction
Marine Invertebrate Zoology Microscope Introduction Introduction A laboratory tool that has become almost synonymous with biology is the microscope. As an extension of your eyes, the microscope is one
More informationDISCO DICING SAW SOP. April 2014 INTRODUCTION
DISCO DICING SAW SOP April 2014 INTRODUCTION The DISCO Dicing saw is an essential piece of equipment that allows cleanroom users to divide up their processed wafers into individual chips. The dicing saw
More informationBruker Dimension Icon AFM Quick User s Guide
Bruker Dimension Icon AFM Quick User s Guide March 3, 2015 GLA Contacts Jingjing Jiang (jjiang2@caltech.edu 626-616-6357) Xinghao Zhou (xzzhou@caltech.edu 626-375-0855) Bruker Tech Support (AFMSupport@bruker-nano.com
More informationUser Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com
User Manual Digital Compound Binocular LED Microscope Model MD82ES10 MicroscopeNet.com Table of Contents i. Caution... 1 ii. Care and Maintenance... 2 1. Components Illustration... 3 2. Installation...
More informationOperating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.
Smith College August 2010 Operating Checklist for using the Laser Scanning Confocal Microscope Leica TCS SP5. CONTENT, page no. Startup, 1 Initial set-up, 1 Software, 2 Microscope Specimen observation
More information