STANDARD OPERATING PROCEDURE: JEOL TEM-2100

Size: px
Start display at page:

Download "STANDARD OPERATING PROCEDURE: JEOL TEM-2100"

Transcription

1 STANDARD OPERATING PROCEDURE: JEOL TEM-2100 Purpose of this Instrument: Essential tool for structural characterization of natural or synthesized nanostructures. Location: WVU - Engineering Sciences Building (EBS) Room: B-63 (basement) Primary Staff Contact: Dr. Marcela Redigolo (304) Office: ESB G75D - (304) The Shared Research Facilities are operated for the benefit of all researchers. If you encounter any problems with this piece of equipment, please contact the staff member listed above immediately. There is never a penalty for asking questions. If the equipment is not behaving exactly the way it should, contact a staff member. 1) INITIAL CHECK This text is a guide to operating the JEOL 2100 TEM. A more detailed and thorough account of these procedures can be found in the JEOL Operating Manual. A glossary for acronyms can be found at the end of this document. 1. Log in your TEM session on the FOM. The computer to the left of the microscope and closest to the wall is connected to the internet for this purpose. (fom.wvu.edu/fom) 2. Check log sheet to see the notes of previous users. This will inform you about anything unusual that occurred before. 3. Write down your name, supervisor, start time (same as FOM) and specimen information on the log sheet. 4. Check the Ion Pump Gauge reading (the blue number in the middle) on the left rack. It is usually around Pa. If the reading exceeds Pa, do not continue and contact SRF personnel immediately. 5. Check the Filament Meter (below the Ion Pump Gauge) and record the filament hours on the log sheet. 6. Check the Cooling Water Flow meters on the back wall. Typical reading should be around (left to right): DP 46 OL 32 CL 14 Gate 42. Small variations (around 2 to 3 units) isn t critical. More than that please contact SRF personnel immediately. 7. Check the HT Tank Gauge reading on the right side of microscope in the back. It should be around MPa. If the reading exceeds 0.02 MPa, do not continue and contact SRF personnel immediately. 2) START-UP 1. Take out the Anti-Contamination Device (ACD) heater from the ACD container if you are the first user of the day. 2. Cover the glass viewing window of the microscope with the protective lid and the eyepieces of the microscope binocular if they are not covered. Place the plastic funnel at the ACD container to make it easier to fill with liquid nitrogen. 3. Get liquid nitrogen from the liquid nitrogen tank. There is always one tank of liquid nitrogen in the TEM room. Electron Microscopy Facility WVU SRF 1

2 IMPORTANT: You are required to wear the apron, gloves and the face shield while handling liquid nitrogen. Failure to do so can result in your being prohibited from using the TEM 4. Fill the ACD container with liquid nitrogen. If necessary, use the rolling ladder in the TEM room. Add more liquid nitrogen after 10 minutes (once the system reaches equilibrium temperature). IMPORTANT: Do not allow liquid nitrogen to contact the window glass, as the glass could crack, causing serious damage to the interior of the microscope column. 5. Liquid nitrogen in the ACD container will last for about four hours. Make sure the ACD container is filled with liquid nitrogen during your entire session. You may need to add more during the session. IMPORTANT: Poor vacuum will occur when the liquid nitrogen in the ACD runs out, which may cause instabilities to the TEM and damage when the filament and high voltage are on. 6. If it s not opened yet, start the software program <TEMCON> on the computer on the right-hand side of the microscope (the closest one to the TEM). The top of the screen when opened should read Controller for JEM2100/HR. 7. Check the stage name and the position on the right of the <TEMCON> program window. It should read <SINGLE TILT HOLDER>. IMPORTANT: Double click the Stage Neutral button if the stage is not neutralized. Pay special attention to the tilt reading TX and TY. Values should be zero. 8. IMPORTANT: Check the switch position of the upper digital camera on the TEM column. It should be in the out position (switch pointing to the right) as shown in Fig. 1. If not, flip the switch to the right to retract the upper digital camera from the column Fig. 1: Upper camera. When switch is to the right, camera is out. 9. Take off the protective lid for the glass viewing window and check the phosphor screen position. The screen should be DOWN. If the screen is not down, press F1 on right control panel to lower down the screen. Cover the lid again. If there are any problems with this step, do not continue and contact SRF personnel immediately. Electron Microscopy Facility WVU SRF 2

3 10. If you find the microscope with any of these start-up steps in different settings than what they should be, report the details in the log sheet and make sure to put the microscope back in the right setting before you start using the equipment. SRF staff will contact the previous user. 11. Check the magnification condition on the right control panel. The system should be in MAG 1 mode (light on) at a magnification of K (reading on the TEM Control software window). If the system is not in MAG 1 mode, press the MAG 1 button (right panel) and set the magnification between 30K and 60K with MAG/CAM knob on right panel. 12. Check to see if Objective Aperture and Selected Area Aperture are inserted. If they are, remove them from the path of the beam by aligning the red dot on knob 1 with the black dot (reference mark) on the column. (See representation on Fig. 3) 3) HIGH VOLTAGE 1. Open the <DIALOGUE> window at the <TEMCON> program, then <HIGH VOLTAGE CONTROL> window. 2. Check the HT reading in the window. It should read 180 kv. If it does not, modify the value to 180 kv with the <DOWN / UP> arrows in the <HT> section of the window. 3. In <HT> section, click <ON> to start high voltage at 180 kv. Wait till HT ON is complete. 4. Check the Beam Current in the <HIGH VOLTAGE CONTROL> window. It should be stabilized at around 92 A. 5. In the <HIGH VOLTAGE CONTROL> window, under the <AUTO HT> section, check the settings: <TARGET> voltage 200 kv, the <STEP> 0.1 kv, and <TIME / STEP> 6 sec. These settings should never be changed! 6. Click on <START> in the <AUTO HT> section. The HT Voltage will gradually increase and will take approximately 20 minutes to stabilize at 200 kv. While the filament voltage is increasing, prepare the specimen holder. Do not alter these settings to raise the HT Voltage faster!! Use this time to prepare your sample on the specimen holder 4) LOADING THE SPECIMEN HOLDER (SH) 1. ALWAYS wear dust-free gloves (IMPORTANT) while handling the specimen holder. 2. Ensure that samples, tweezers and other tools are absolutely clean and free of contamination from contact with fingers, any surface which might not be clean or might have had any contact with fingers or contaminants. 3. NEVER blow on or exhale on samples to dry them! Always use a clean, dry gas source when preparing your samples. 4. Prepare samples in advance, allowing the solvent used to fully dry, before bringing the sample to the microscope. 5. Make sure you always have a sure grip of the SH. Always ask if you have a question. The staff is here to help and guide you. Electron Microscopy Facility WVU SRF 3

4 6. IMPORTANT: Do NOT breathe on the holder. It is very difficult to pump water vapor in TEM! Besides, it will take a long time before the system vacuum gets ready for the measurements. Never touch the O-rings, even with gloves!! Fig. 2: Detail of Specimen Holder (SH) 7. For the single-tilt holder, inset the cartridge-handling tool into the cartridge clamp hole at the end of the SH (see Fig. 2). 8. Open the clamp by tilting the cartridge-handling tool, then using the tweezers remove the cartridge from the holder. 9. Mount the cartridge onto the cartridge support with the guide pin on the support inserted into the guide hole on the cartridge. 10. Loosen the two little screws from the cartridge TWO turns. 11. Rotate the retainer away and mount the sample grid with your sample facing UPWARD. 12. Return the retainer to the original position (make sure of that) and secure it with the two little screws. IMPORTANT: Never overtighten the screws!! 13. Insert the cartridge-handling tool into the cartridge clamp hole at the end of the SH and open the clamp. 14. Attach the cartridge to the holder so that the guide hole on the cartridge aligns with the guide pin on the clamp. 15. Clamp the cartridge by returning the cartridge-handling tool to the clamping position. 16. Turn the holder 180 o to have the sample facing downward. Tap the base of holder rod with gloved finger to ensure that the sample does not fall out. IMPORTANT: This is an important step because if a sample falls inside the microscope, the microscope will have to be opened to remove it, meaning the vacuum must be broken and the column unavoidable contaminated. This requires total shutdown of the machine. 17. Make sure there is no dust and/or lint on the specimen holder O-rings. Use the magnifier lens and lamp at the desk to check. Electron Microscopy Facility WVU SRF 4

5 IMPORTANT: Always ensure that the SH O-rings are completely and definitively free of dust or fibers. Check very carefully with the magnifier lens and lamp. Be really sure before introducing the SH into the microscope!!!! 18. While waiting for the voltage to reach 200 kv, add more liquid N2 to the ACD heater to reach the full level. Remember to cover the protective lid for the glass viewing window of the microscope and the binocular eye-pieces whenever you are using liquid nitrogen around the microscope. 5) INSERTING THE SPECIMEN HOLDER (SH) These steps are shown in diagrams attached to the end of this SOP and also taped to the microscope column. 1. Verify that the stage position is neutral (on the right/top corner of <TEMCON> screen) and that the airlock switch is set to AIR position (at the microscope column, below the goniometer). 2. Align the SH guide pin with the guide groove on the microscope column (horizontal left side). 3. Insert the holder into the goniometer until position 1. This will be a long path. When the SH guide pin finally reaches the entrance of the goniometer, you will be reaching position 1. Clicking sounds will be heard immediately after. Never, never turn the SH clockwise before pumping. IMPORTANT: Never use excessive force when inserting the SH, especially sideways!! 4. Set the goniometer switch to PUMP. Yellow light ON. 5. WAIT until the green light on the goniometer is lit. 6. Check the vacuum reading for the specimen chamber on the <TEMCON> program. For that, open <MONITOR>, then <VACUUM SYSTEM>. The window VALVE STATUS will open. Check for the SPECIMEN/PiG 4 reading on the right corner of the window. It should be 36 A or less, and the status should be EVAC READY. a. Note: The system starts evacuating the goniometer. If this process takes too long, the sample inside might be degassing or have some contamination (hand sweat, grease, moisture, etc) that might be preventing the vacuum. This is why it is crucial to have a clean environment when preparing (in your lab) and loading the sample onto the SH! b. If this process takes more than 10 min, contact SRF personnel immediately. 7. Slowly turn the SH clockwise until position 2, while firmly holding it, because it might be pulled into the goniometer to position 3. Be careful!! This will be a very short distance. 8. Carefully HOLD (DO NOT PUSH) the SH until position 3. This will be a very short distance. 9. Slowly turn the SH clockwise until position 4, while firmly holding it. This will be a short distance. This step might require a little bit of more strength, but it will be just a small amount more than before. Do not force the SH in. Electron Microscopy Facility WVU SRF 5

6 10. HOLD (DO NOT PUSH) the SH until it reaches position 5 and black pin is in place. IMPORTANT: All procedures here must be done slowly and carefully. Do not let go of the SH because the vacuum will pull the SH into the column! Do not let the SH hit the end of the goniometer!! Any collision can cause serious damage to the SH and the microscope!! 6) GETTING AN ELECTRON BEAM 1. Select the SH name <SINGLE TILT HOLDER> from the list in the holders on the <TEMCON> screen. (top right position), if it wasn t selected yet. 2. Record the <BEAM CURRENT> reading on HIGH VOLTAGE CONTROL window onto the log sheet. It should be around A. If not, contact SRF personnel immediately. 3. Click on <ON> in the FILAMENT section of the HIGH VOLTAGE CONTROL window. The beam current will increase and stabilize at around 110 A. If not, contact SRF personnel immediately. 4. Wait till the beam current stabilizes. During this time, uncover the lid to observe the beam. Turn OFF the lights of the room and in the left panel click [ROOM LAMP]. 5. Record the <BEAM CURRENT> reading on HIGH VOLTAGE CONTROL window onto the log sheet. 6. Verify that the [TEM] button on the left panel is ON (if not, press it). 7. <SPOT SIZE> must be 1, and <ALPHA> must be 3. They appear on the top of the <TEMCON> software screen. 8. Select coarse sample movements by pressing [CRS] on the SPEC CONTROL (trackball) that is the small controller sitting on the left side of the microscope column. It will light up. 7) FIND THE BEAM AND FOCUS IT Note: If you can t see the electron beam on the phosphor screen, follow the steps below. If you can see the electron beam as a small dot, turn the [BRIGHTNESS] knob to spread it on the screen and use the trackball to move the grid out of view, in case any part of the copper grid is showing up. Go to step 2, if you already see the electron beam. 1. Set magnification to 10K ([MAGNIFICATION] knob on right control panel) and move the sample using the trackball until the beam appears completely on the phosphor screen. IF YOU CAN T SEE THE BEAM YET a. Note: If the beam does not appear on the screen, decrease the magnification or select LOW MAG mode by pressing the [LOW MAG] button on the right control panel. b. Turn the [BRIGHTNESS] knob and observe if the beam appears. c. Check to see if Objective Aperture and Selected Area Aperture are inserted. If they are, remove them from the path of the beam by aligning the red dot on knob 1 with the black dot (reference mark) on the column. (See representation on Fig. 3 below) { d. If you still cannot find the beam, ask SRF personnel for help. Electron Microscopy Facility WVU SRF 6

7 Fig.3: Representation of the knobs for Objective Aperture and for the Selected Area Aperture. 2. Set magnification to 50K for standard corrections. ([MAGNIFICATION] knob on right control panel). 3. Check the Condenser Aperture. (the top one) a. The second biggest size is the standard. It must be always there. If not, align the 2 nd largest white dot on the knob 1 with the small black dot (reference mark) on the column. (See Fig. 3 for reference) b. If you use another aperture during your work, you must always put back the condenser aperture in the column before leaving and LEAVE IT ALIGNED. 4. Press [STANDARD FOCUS] button (right control panel) to bring the beam into focus 5. Focus the beam into the smallest possible spot using the [BRIGHNESS] knob (left control panel). If you need to center the beam, use the [SHIFT X] and [SHIFT Y] buttons on the right and left control panels to do so.. 8) CENTER THE CONDENSER LENS APERTURE (CLA) Note: Change the [BRIGHTNESS] so that the beam is going through the crossover (the smallest size of the beam on the phosphor screen) and see if the beam center is swinging. If yes, the Condenser Lens Aperture needs centering. 1. Focus the beam to crossover on the phosphor screen with [BRIGHNESS] knob and then center it using the [SHIFT X] and [SHIFT Y] knobs on left and right control panels. 2. Adjust the [BRIGHTNESS] knob to expand the beam before and after crossover and see if the beam moves off screen center as you turns. 3. Center the beam using the end and side knobs on the Condenser Lens Aperture. They are knobs 2 and 3, respectively, on Fig. 4 below. IMPORTANT: move the CLA knobs carefully. The amount of movement required to align the aperture is minimum. Fig. 4: Condenser Lens Aperture knobs: (1) aperture size, (2) and (3) alignment. Electron Microscopy Facility WVU SRF 7

8 4. Repeat steps 1 to 3 till there is no beam swinging. Verify the result by turning the [BRIGHNESS] knob back and forth through the crossover and see if the beam expands and contracts coaxially. 9) ALIGNING THE OPTICAL AXIS OF GUN AND CONDENSER SYSTEMS GUN 1 CLA 5 Note: Perform this alignment to correct a shift of beam position when you change the spot size. 1. Select <MAINTENANCE> from the <TEMCON> menu and click on <ALIGNMENT> to open the ALIGNMENT PANEL FOR MAINTENANCE window. 2. Set magnification to 100K. 3. Set the [SPOT SIZE] knob (left panel) to 1 (number appears on the top of the <TEMCON> program) and focus the electron beam to the crossover with the [BRIGHTNESS] knob. 4. Click on <GUN> on DEF SELECT window. 5. Center the electron beam using the [SHIFT X] and [SHIFT Y] knobs. 6. Set the [SPOT SIZE] knob to 5 and focus the beam with the [BRIGHTNESS] knob. 7. Click on <CLA> on DEF SELECT window or press the [BRIGHT TILT] button. 8. Center the electron beam using the [SHIFT X] and [SHIFT Y] knobs. 9. Repeat steps 3 8 until the electron beam stays at the center of the screen when the [SPOT SIZE] knob setting is changed. 10. Keep the [SPOT SIZE] on 1. Whichever is clicked, <CLA> or <GUN>, press it again to turn it OFF. 10) CORRECT THE CONDENSER LENS ASTIGMATISM SHAPE OF THE BEAM Note: Perform condenser astigmatism correction ONLY if the beam is not circular (elongated) when it is open or near crossover. 1. Set magnification to 100K. 2. Focus the beam to crossover using the [BRIGHTNESS] knob and center the beam with [SHIFT X] and [SHIFT Y] knobs. ONLY IF THE BEAM IS NOT CIRCULAR 3. Spread the beam on the phosphor screen and center the Condenser Lens Aperture with the end and side knobs. They are knobs 2 and 3, respectively, on Fig Press the [COND STIG] button (left control panel) or click on the <CL STIG> in STIGMATOR section of the ALIGNMENT PANEL FOR MAINTENANCE window. 5. Slowly turn the [BRIGHTNESS] knob back and forth through the crossover and adjust the [DEF/STIG Y] and [DEF/STIG X] knobs so that the shape of the beam spot becomes round immediately before and after the crossover. 6. Check the beam condition by spreading the beam on the screen and see if it is round and centered. Electron Microscopy Facility WVU SRF 8

9 7. Click on the <CL STIG> in STIGMATOR section of the ALIGNMENT PANEL FOR MAINTENANCE window to turn it OFF. 11) CONDENSER LENS DEFLECTION COIL TILT X / Y ADJUSTMENT TILT X / TILT Y Note: This is to make sure that beam tilt and shift are independent, that is, the beam can stay at the same place while it is being tilted. 1. Set magnification to 100K. 2. Focus the electron beam to crossover using the [BRIGHNESS] knob. 3. Press [BRIGHT TILT] button (left panel) or click on <CLA> in DEF SELECT section of the ALIGNMENT PANEL FOR MAINTENANCE window. 4. Center the electron beam using the [SHIFT X] and [SHIFT Y] knobs. 5. Click on <TILT> in the COMPENSATOR section of the ALIGNMENT PANEL FOR MAINTENANCE window. 6. Click on <TILT-X> in the WOBBLER section of the ALIGNMENT PANEL FOR MAINTENANCE window. 7. Using the [DEF/STIG X] knob (left panel), bring the two splitting beam spots together. a. Note: If the beam is going too far out the screen, turn OFF the TILT Y and TILT. Repeat this procedure from step Turn OFF by clicking again the <TILT-X> in the WOBBLER section of the ALIGNMENT PANEL FOR MAINTENANCE window. { 9. Click on <TILT> in the COMPENSATOR section of the ALIGNMENT PANEL FOR MAINTENANCE window to turn it OFF. 10. If the beam moves off center as you turn the [BRIGHNESS] knob, press the [BRIGHT TILT] button or click on <CLA> in DEF SELECT section, and use the [SHIFT X] and [SHIFT Y] knobs to center the beam. 11. Click on <TILT> in the COMPENSATOR section of the ALIGNMENT PANEL FOR MAINTENANCE window. 12. Click on the <TILT-Y> in the WOBBLER section of the ALIGNMENT PANEL FOR MAINTENANCE window. 13. Using the [DEF/STIG Y] knob (right panel), bring the two splitting beam spots together. a. Note: If the beam is going too far out the screen, turn OFF the TILT Y and TILT. Repeat this procedure from step Turn OFF by clicking again the <TILT-Y> in the WOBBLER section of the ALIGNMENT PANEL FOR MAINTENANCE window. { 15. Click on <TILT> in the COMPENSATOR section of the ALIGNMENT PANEL FOR MAINTENANCE window to turn it OFF. Electron Microscopy Facility WVU SRF 9

10 16. If the beam moves off center as you turn the [BRIGHNESS] knob, press the [BRIGHT TILT] button or click on <CLA> in DEF SELECT section, and use the [SHIFT X] and [SHIFT Y] knobs to center the beam. 17. Remember to turn OFF the [BRIGHT TILT] button or click on <CLA> in DEF SELECT section. 12) CENTERING CURRENT AXIS (OBJECTIVE LENS ALIGNEMENT) OBJ Note: Perform current axis adjustment if you observe a shift of image when you change the focus (i.e., changing the objective lens current). 1. Obtain an image on the phosphor screen at over 100K magnification. To focus the image, first make sure you are in the proper sample height by pushing the [UP/DOWN Z] buttons (right control panel) to adjust the Z height until the image is in focus. 2. For further focus adjustments, use [OBJ FOCUS] knobs. 3. Spread the beam fully across the phosphor screen, and move an object in the image to the center of the screen. 4. Click on <OBJ> in WOBBLER section of the ALIGNMENT PANEL FOR MAINTENANCE window. The image will expand and contract periodically. 5. Press the [BRIGHT TILT] button (left panel) or click on <CLA> in DEF SELECT section. 6. Adjust [DEF/STIG X] and [DEF/STIG Y] knobs so that the image expands and contracts around the center of the phosphor screen, i.e., the object in the center do not slide while vibrating. 7. Click on <OBJ> in WOBBLER section again to turn it OFF. 8. Verify the result by turning the [OBJ FOCUS] knob to go in and out of focus and see if the image shifts. 9. Click on <CLA> in DEF SELECT section again to turn it OFF. 13) CENTERING VOLTAGE AXIS (OBJECTIVE LENS ALIGNEMENT) HT Note: Adjust the TEM optical axis so that the image stays in the center of the screen even when the accelerating voltage is changed. This is to ensure that the beam passes through the optical axis of the objective lens. 1. Obtain a focused image on the phosphor screen at 300K. 2. Focus the image using [OBJ FOCUS] knobs. 3. Spread the beam fully across the phosphor screen, and move an object in the image to the center of the screen. 4. Press the [HT WOBB] button (right panel) or click on <HT> in WOBBLER section of the ALIGNMENT PANEL FOR MAINTENANCE window. The image will expand and contract periodically. 5. Press the [BRIGHT TILT] button (left panel) or click on <CLA> in <DEF Select> section. 6. Adjust [DEF/STIG X] and [DEF/STIG Y] knobs so that the image expands and contracts around the center of the phosphor screen, i.e., the object in the center do not vibrate. Electron Microscopy Facility WVU SRF 10

11 7. Press the [HT WOBB] button or click on <HT> in WOBBLER section again to turn it OFF. 8. Click on <CLA> in DEF SELECT section again to turn it OFF. 14) OBJECTIVE LENS ASTIGMATISM CORRECTION Note: Perform Objective Lens astigmatism correction if you observe asymmetrical Fresnel fringes appear along the rim of a hole. If the correction of objective lens astigmatism is inadequate, it is very difficult to obtain an in-focus image no matter how you try to focus. Method 1: Fresnel Fringe Method 1. Set magnification between x20k and x100k. 2. Adjust the [BRIGHNESS] knob for optimum brightness of the image. CHOOSE ONE OF THE METHODS 3. Press [BRIGHT TILT] button (left panel) or click on <CLA> in DEF SELECT section. 4. Adjust the [SHIFT] knobs to center the beam. 5. Focus the image using [OBJ FOCUS] knob. 6. Move the image of a small hole or a curved edge on the sample to the center of screen. 7. Press [OBJ STIG] (left panel) or click on <OL STIG> in DEF SELECT section of the ALIGNMENT PANEL FOR MAINTENANCE window. 8. Adjust the [OBJ FOCUS] knobs so that the Fresnel fringes appear along the rim of the hole or edge. The Fresnel fringes are the white lines seen along the edge in a slightly over focused condition. If there is astigmatism in the objective lens, these fringes will be asymmetrical. 9. Adjust the [DEF/STIG X] and [DEF/STIG Y] knobs so that Fresnel fringes appear to be symmetric (have same contrast and same width in all directions). 10. Press again [OBJ STIG] (left panel) or click on <OL STIG> in the DEF SELECT section of the ALIGNMENT PANEL FOR MAINTENANCE window to turn it OFF. Method 2: Granularity of an amorphous sample The supporting film itself is amorphous and can be used for astigmatism correction at high magnifications. 1. Obtain an image at your desired magnification. (over x100k) 2. Adjust the [BRIGHNESS] knob for optimum brightness of the image. 3. Press [BRIGHT TILT] button (left panel) or click on <CLA> in DEF SELECT section of the ALIGNMENT PANEL FOR MAINTENANCE. 4. Adjust the [SHIFT X] and [SHIFT Y] knobs to center the beam. 5. Focus the image using [OBJ FOCUS] knobs. Electron Microscopy Facility WVU SRF 11

12 6. Move a suitable image of an amorphous region on the sample (or support film) to the center of the screen. 7. Press [OBJ STIG] (left panel) or click on <OL STIG> in the DEF SELECT section of the ALIGNMENT PANEL FOR MAINTENANCE window. 8. Slowly turn [OBJ FOCUS] knob back and forth through the focus position to observe the variation of phase contrast. If there is astigmatism, the granularity of the amorphous film will be observed with preferred directionality. 9. Adjust [DEF/STIG X] and [DEF/STIG Y] so that the granularity presents no preferred directionality. Under perfect adjustment, once focused, the support film should almost disappear in the image. 10. Press again [OBJ STIG] (left panel) or click on <OL STIG> in the DEF SELECT section of the ALIGNMENT PANEL FOR MAINTENANCE window to turn it OFF. Method 3: Fast Fourier Transfor (FFT) DO NOT INSERT THE CAMERA UNTIL 1. Obtain an image at high magnification (around x300k), on the phosphor screen. Preferably of the carbon film on the grid. 2. Start the Gatan <DIGITAL MICROGRAPH> software on the dedicated computer on the right-hand side of the TEM console (the closest to the door). You will hear two beeps. 3. If the software asks whether to insert the camera, click <YES>. 4. Go to <CAMERA> menu, under <CAMERA> sub-menu, select <ES500W> camera. 5. Turn the [BRIGHTNESS] knob (left panel) to reduce the intensity of the electron beam on the phosphor screen. (Important: The image/beam should at least cover the whole phosphor screen.) 6. If <CAMERA VIEW> is opened on the right side of the screen, click on <START VIEW> on to open the view window. 7. If <CAMERA VIEW> isn t opened, click on <WINDOW>, <FLOATING WINDOWS>, <CAMERA VIEW> to open it. Click then on <START VIEW>. Then, click on <START VIEW> on to open the view window 8. Change the Exposure Time from 1 second to 0.3 second. 9. On <CAMERA VIEW>, click the symbol (settings) on the lower right corner of this window. (Fig. 5) Fig. 5: Camera view window and settings. 10. Adjust the settings to read: Top: 0, Left: 0, Bottom: 1024, Right: Click <OK> and close this window. Electron Microscopy Facility WVU SRF 12

13 11. Insert the upper digital camera by setting the switch to the left. (Fig. 6) HERE!!!! 12. Go to <PROCESS> on the Digital Micrograph software. 13. Select <LIVE> and <FFT>. Fig. 6: Upper camera. When switch is to the right, camera is out. 14. The new window that opens will have the FFT of the chosen image. 15. Press [OBJ STIG] (left panel) or click on <OL STIG> in the DEF SELECT section of the ALIGNMENT PANEL FOR MAINTENANCE window. 16. Adjust [DEF/STIG X] and [DEF/STIG Y] so that the FFT will be as round as possible. 17. Press again [OBJ STIG] (left panel) or click on <OL STIG> in the DEF SELECT section of the ALIGNMENT PANEL FOR MAINTENANCE window to turn it OFF. 18. Close the FFT window. 19. Click on <STOP VIEW> on the CAMERA VIEW window. 20. Retract the camera by setting the switch to the right. 15) IMAGE RECORDING WITH THE UPPER DIGITAL CAMERA 1. Start the Gatan <DIGITAL MICROGRAPH> software on the dedicated computer on the right-hand side of the TEM console (the closest to the door) is it is not already opened. 2. Go to <CAMERA> menu, under <CAMERA> sub-menu, select <ES500W> camera. 3. Obtain the image of interest on the phosphor screen. 4. Turn the [BRIGHTNESS] knob (left panel) to reduce the intensity of the electron beam on the phosphor screen. (Important: The image should cover the whole phosphor screen.) 5. If <CAMERA VIEW> is opened on the right side of the screen, click on <START VIEW> on to open the view window. 6. If <CAMERA VIEW> isn t opened, click on <WINDOW>, <FLOATING WINDOWS>, <CAMERA VIEW> to open it. Then, click on <START VIEW>. Electron Microscopy Facility WVU SRF 13

14 7. Insert the upper digital camera by setting the switch to the left. (Fig. 1) 8. Click on <STOP VIEW> to freeze the image. 9. Click on <START ACQUIRE> at the <CAMERA ACQUIRE> window, on the right side of the DIGITAL MICROGRAPH software screen. If this window isn t opened, click on <WINDOW>, <FLOATING WINDOWS>, <CAMERA ACQUIRE> to open it. 10. Once the image is acquired, click on <FILE>, <SAVE AS> and save the image in the.dm3 format. 11. When you finish recording the image, retract the camera by setting the switch to the right, even if you want to look for another region on your sample. 12. At the end of your session, save all your data to a memory stick and take it with you. Do not store data on the TEM computer. 13. If you need to convert images to a different format, at the end of the session, click on <FILE>, <BATCH CONVERT>. 14. Choose the folder in which your data is stored in the new window. 15. The files to be converted should be *.dm3. Choose between converting to.bmp,.jpg or.tiff formats. 16. Click <OK>. 18) SAMPLE EXCHANGE 1. If you were using the Gatan DigitalMicrograph software, click <STOP VIEW> at the <CAMERA VIEW> window. Make sure the camera is retracted from the microscope column by setting the camera switch to the right. 2. Set the Condenser Lens Aperture at position 2 (2 nd largest aperture). 3. Take the Objective Lens Aperture and Selected Area Aperture out of the column by aligning the red dot on the knob with the small black dot on the column. 4. Set [SPOT SIZE] to 1 and [ALPHA] to 3. Reduce [MAG] to 30K 60K. Using the [BRIGHTNESS] knob, bring the beam to the crossover and center it using the [SHIFT X] and [SHIFT Y] knobs on the control panels. 5. Turn off Filament by clicking on <OFF> in the FILAMENT section of the HIGH VOLTAGE CONTROL window. The system might open a new window to ask confirmation to turn off the filament. In that case, select <YES>. a. Never remove a sample with the Filament ON. b. Wait to remove the sample untill the beam current stabilizes! 6. Put the stage in neutral position by double clicking the <Stage Neutral> button on the <TEMCON> window. 7. Remove the specimen holder from the goniometer following the proper procedure below (also can be found on Appendix B). Observe the representation of these steps in the diagrams attached to the end of this SOP and also taped to the microscope column for your convenience. Electron Microscopy Facility WVU SRF 14

15 8. Be careful!! Slowly pull out the SH until it stops (position 1). Be sure to firmly hold the SH while doing this because it might be pulled back into the goniometer due to the vacuum. 9. Turn it counterclockwise until it stops (position 2). DO NOT PULL IT, JUST TURN IT. It s a short distance. 10. Slowly pull out the SH until it stops (position 3). It s a very short distance. 11. Turn it counterclockwise until it stops (position 4). DO NOT PULL IT, JUST TURN IT. It s a very short distance. Never, never pull it before the chamber is filled with air!!! 12. Set the goniometer switch to <AIR>. 13. Wait about 30 seconds (sound of air filling the chamber will be heard briefly). 14. Remove the SH carefully, out of the goniometer. IMPORTANT: All procedures here must be done slowly and carefully. Do not let go of the SH because the vacuum will pull the SH back into the goniometer. Collision can cause serious damage to the SH and the microscope!! Do not use force to remove the SH from the goniometer, just hold it firmly. DO NOT pull it out with speed and strength! 15. Reinsert the SH into the goniometer by following procedure described in section 5 of this SOP or in Appendix A. 16. Click on <ON> in the FILAMENT section of the HIGH VOLTAGE CONTROL window. The beam current will increase and stabilize at around 110 A. If not, contact SRF personnel immediately. 17. Find the electron beam and repeat all processes starting from section 7 of this SOP. 19) SHUT DOWN PROCEDURE 1. If you were using the Gatan DigitalMicrograph software, click <STOP VIEW> at the <CAMERA VIEW> window. Make sure the camera is retracted from the microscope column by setting the camera switch to the right. 2. Set the Condenser Lens Aperture at position 2 (2 nd largest aperture). 3. Remove the Objective Lens Aperture and Selected Area Aperture from the column by aligning the red dot on the knob with the small black dot on the column. 4. Set [SPOT SIZE] to 1, [ALPHA] to 3 and reduce [MAG] to 30,000, center the beam to crossover and center it using [SHIFT X] and [SHIFT Y]. 5. Keep [MAG] between 30K 60K to keep the lenses warm and stable, and prevent any water vapor condensation. 6. Turn off Filament by clicking on <OFF> in the FILAMENT section of the HIGH VOLTAGE CONTROL window. The system might open a new window to ask confirmation to turn off the filament. In that case, select <YES>. 7. Wait till the process is finished! Never remove a sample when the filament is ON. Electron Microscopy Facility WVU SRF 15

16 8. Turn OFF the HT VOLTAGE by clicking on <OFF> in HT section of the HIGH VOLTAGE CONTROL window. The system might open a new window to ask confirmation to turn off the HT. In that case, select <YES>. 9. Wait till the beam current reaches zero! 10. Reduce HT value to 180 kv by clicking on the <DOWN/UP> buttons in the HT section. 11. Put the stage in neutral position by double clicking the <Stage Neutral> button on the <TEMCON> window. 12. Remove the specimen holder from the goniometer following the proper procedure below (also can be found on Appendix B). Observe the representation of these steps in the diagrams attached to the end of this SOP and also taped to the microscope column for your convenience. 13. Be careful!! Slowly pull out the SH until it stops (position 1). Be sure to firmly hold the SH while doing this because it might be pulled back into the goniometer due to the vacuum. 14. Turn it counterclockwise until it stops (position 2). DO NOT PULL IT, JUST TURN IT. It s a short distance. 15. Slowly pull out the SH until it stops (position 3). It s a very short path. 16. Turn it counterclockwise until it stops (position 4). DO NOT PULL IT, JUST TURN IT. It s a very short distance. 17. Set the goniometer switch to <AIR>. Never, never pull it before the chamber is filled with air!!! 18. Wait about 30 seconds (sound of air filling the chamber will be heard briefly). 19. Remove the SH carefully from the goniometer. IMPORTANT: All procedures here must be done slowly and carefully. Do not let go of the SH because the vacuum will pull in the SH back into the goniometer. Any collision can cause serious damage to the SH and the microscope!! Do not use force to remove the SH from the goniometer, just hold it firmly. DO NOT pull it out with speed and strength! Continue from # 20 after removing your sample! 20. Cover the glass viewing window of the microscope with the protective lid and the eyepieces of the microscope binocular if they weren t covered yet. 21. Log in on <FOM> to check if there is any user reserved a session after you today. 22. Log OFF on the FOM from your TEM session. 23. Sign out on the log sheet. Report any problems in the comments section and notify the SRF staff. Procedure 1: There will be another user AFTER you If someone has reserved a session after you: Electron Microscopy Facility WVU SRF 16

17 a. Save all your images and record them to your memory stick. b. Make sure that the room is clean and organized before you leave. c. Switch off the room lights. Procedure 2: You are the LAST user of the day If no one has reserved the instrument for use during the next 4 hrs, follow the procedure below: a. Remove the ACD container fill cap. Use face shield for protection from liquid nitrogen spills! b. Insert the ACD heater into the ACD container. IMPORTANT: Use protection clothes/face shield! Direct the plastic pipe AWAY from your face!! Liquid nitrogen will come out through the plastic pipe when inserting the heater! c. Insert the heater plug into the HTR socket on the connector box. d. Select <Maintenance> from the <TEMCON> menu bar. e. Click on <ACD & BAKE> to open the <BAKE OUT /ACD HEAT> window. f. Click on the <ACD HEAT> tab to open the <ACD HEAT> window. g. Click on the ACD Heat <ON> button. h. Click <OK>. Note: The liquid nitrogen in the ACD has to be burnt off at the end of the day. If this is not done, there is a danger of contaminating the column and adversely affecting microscope performance. i. Save all your images and copy them to your memory stick. j. Make sure the room is clean and tidy before you leave. k. Switch off the room lights. Electron Microscopy Facility WVU SRF 17

18 GLOSSARY Glossary of terms and acronyms used in this SOP. If you have any suggestions to be added here, please let us know. Thanks. 1) ACD = Anti-Contamination Device 2) CL = Condenser Lens 3) CLA = Condenser Lens Aperture 4) COND = Condenser 5) Crossover = the smallest size of the beam on the phosphor screen 6) CRS = Coarse 7) DEF = Deflection 8) DP = Diffusion Pump 9) FFT = Fast Fourier Transform 10) FOM = Facility Online Manager 11) HT = High Tension 12) MAG = Magnification 13) OBJ = Objective 14) OL = Objective Lens 15) SA = Selected Area (for diffraction) 16) SH = Specimen Holder 17) SOP = Standard Operating Procedure 18) SRF = Shared Research Facilities 19) STIG = Stigmatism 20) TEM = Transmission Electron Microscope 21) TEMCON = TEM Control software Electron Microscopy Facility WVU SRF 18

19 INTRODUCING THE SPECIMENT HOLDER (SH) APPENDIX A 1. Align the SH guide pin with the guide groove on the microscope column (see photo below). 2. Insert the holder into the goniometer until position 1. Clicking sounds will be heard right after. 3. Set the goniometer switch to <PUMP>. Yellow light ON. 4. WAIT until the green light on the goniometer is lit. a. Note: If this process takes more than 10 min / green light does not turn ON, contact SRF personnel immediately. 5. Slowly turn the SH clockwise until position 2, while firmly holding it, because it might be pulled into the goniometer to position 3. Be careful!! This will be a very short path. 6. Carefully HOLD (DO NOT PUSH) the SH until position 3. This will be a very short path. 7. Slowly turn the SH clockwise until position 4, while firmly holding it. This will be a short path. a. This step might require a little bit of more strength, but still it will be just a bit more than before. 8. HOLD (DO NOT PUSH) the SH until it reaches position 5 and black pin is in place. Electron Microscopy Facility WVU SRF 19

20 REMOVING THE SPECIMEN HOLDER (SH) APPENDIX B Always place the stage in neutral position before removing it!!! 1. Be careful!! Slowly pull out the SH until it stops (position 1). 2. Turn it anticlockwise until it stops (position 2). DO NOT PULL IT, JUST TURN IT. 3. Slowly pull out the SH until it stops (position 3). 4. Turn it anticlockwise until it stops (position 4). DO NOT PULL IT, JUST TURN IT. 5. Set the goniometer switch to <AIR>. 6. Wait about 30 seconds (sound of air filling the chamber will be heard briefly). Electron Microscopy Facility WVU SRF 20

21 EMERGENCY OPERATING PROCEDURES If you have any questions, even if you are just slightly unsure, ASK someone who knows and can help. There are no penalties for asking for help but there may be for not reporting damage to the equipment that may delay or prevent others from working. If, at any time, you need to contact someone for help, call or locate the following staff of the Shared Research Facility (SRF): Marcela Redigolo Office: ESB G75D Phone: (304) Cell: (304) This microscope is protected by various safety devices. If none of the above-listed individuals is available, the user must: Turn the filament OFF. Turn HT OFF. If possible, the user should stay with the microscope while trying to contact the above individuals. If it becomes necessary to leave the microscope then the user should leave a large, legible note on both the microscope and at least one of the above individuals offices, stating: The problem (describe what happened and steps taken) When it occurred (date and time) User name and phone number Prof. Xueyan Song, faculty expert in TEM, is also available for assistance if the contact above is not available. Her office is located at ESB 537 and her office phone number is (304) If a dangerous situation is evident (smoke, fire, sparks, etc), ONLY if it is safe to do so, the user should press EM STOP switch (on the control panel under the left side of the microscope) to turn OFF power to the entire microscope and notify the proper emergency personnel. In any case, the user should leave the facility and contact emergency personnel as soon as possible from a safe place. Marcela Redigolo Office: ESB G75D Phone: (304) Cell: (304) Electron Microscopy Facility WVU SRF 21

2. Raise HT to 200kVby following the procedure explained in 1.6.

2. Raise HT to 200kVby following the procedure explained in 1.6. JEOL 2100 MANUAL Quick check list 1. If needed, fill the reservoir with LN2 2. Raise HT to 200kVby following the procedure explained in 1.6. 3. Insert specimen holder into TEM (Insert holder in airlock,

More information

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Follow the procedures below when you use the Hitachi 7100 TEM. Starting Session 1. Turn on the cold

More information

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope Last updated on 1/8/2018 Jinsong Wu, jinsong-wu@northwestern.edu Aims: The aim of this lab is to familiarize you with basic TEM alignment

More information

Basic Users Manual for Tecnai-F20 TEM

Basic Users Manual for Tecnai-F20 TEM Basic Users Manual for Tecnai-F20 TEM NB: This document contains my personal notes on the operating procedure of the Tecnai F20 and may be used as a rough guide for those new to the microscope. It may

More information

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack). FEI Tecnai F20 S/TEM: imaging in TEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 01/21/18 1. Filling the cold trap (if needed) 1.1. Prior to use, the cold trap needs

More information

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier). JEOL 200CX operating procedure Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Specimen loading 1.1. Unlock the TUMI system. 1.2. Load specimen(s) into the holder. If using the double tilt holder, ensure

More information

JEOL 6500 User Manual

JEOL 6500 User Manual LOG IN to your session on the computer to the left of the microscope. Starting Conditions 1. Press Ctrl-Alt-Del and log on to the microscope computer. Click on JEOL PC SEM 6500 icon. Click yes if message

More information

JEOL JEM-1400 Transmission Electron Microscope Operating Instructions

JEOL JEM-1400 Transmission Electron Microscope Operating Instructions JEOL JEM-1400 Transmission Electron Microscope Operating Instructions Anti-contamination device Objective aperture Objective aperture translation knobs Specimen holder Pump/air switch Left hand control

More information

Instructions for Tecnai a brief start up manual

Instructions for Tecnai a brief start up manual Instructions for Tecnai a brief start up manual Version 3.0, 8.12.2015 Manual of Tecnai 12 transmission electron microscope located at Aalto University's Nanomicroscopy Center. More information of Nanomicroscopy

More information

User Operation of JEOL 1200 EX II

User Operation of JEOL 1200 EX II **Log onto Computer** Open item program Start Up Procedure User Operation of JEOL 1200 EX II 1. If scope is not running, locate an electron microscopy technician (EMT) to find out why not. 2. Turn up brightness

More information

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below.

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. http://staff.washington.edu/wpchan/if/cm100_inst.shtml Instructions for the Philips CM100 TEM and peripherals

More information

1. Specimen Holder Removal, Loading, and Insertion

1. Specimen Holder Removal, Loading, and Insertion OPERATION OF THE PHILIPS CM-200 FEG-TEM When not in use, the CM-200 should be in the MICROSCOPE ON configuration with the HIGH TENSION ON (illuminates green when the high tension is on).. The microscope

More information

STEM alignment procedures

STEM alignment procedures STEM alignment procedures Step 1. ASID alignment mode 1. Write down STD for TEM, and then open the ASID control window from dialogue. Also, start Simple imager viewer program on the Desktop. 2. Click on

More information

2 How to operate the microscope/obtain an image

2 How to operate the microscope/obtain an image Morgagni Operating Instructions 50079 010912 2-1 2 ow to operate the microscope/obtain an image 2.1 Starting the microscope 2.1.1 Starting the microscope with several manually-operated steps 1. Turn on

More information

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right Quick Guide to Operating FEI Titan Themis G2 200 (S)TEM: TEM mode Susheng Tan Nanoscale Fabrication and Characterization Facility, University of Pittsburgh Office: M104/B01 Benedum Hall, 412-383-5978,

More information

JEOL 6700 User Manual 05/18/2009

JEOL 6700 User Manual 05/18/2009 JEOL 6700 User Manual 05/18/2009 LOG IN to your session on the computer to the right of the microscope. Starting Conditions 1. Click the button and read the Penning Gauge to ensure that the microscope

More information

1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height.

1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height. JEOL 2010F operating procedure Covers operation in STEM mode (See separate procedures for operation in TEM mode and operation of EDS system) Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 NOTE: this operating

More information

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I MSE 595T Basic Transmission Electron Microscopy TEM Imaging - I Purpose The purpose of this lab is to: 1. Make fine adjustments to the microscope alignment 2. Obtain a diffraction pattern 3. Obtain an

More information

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 1. Utility Requirements a. System power is supplied by two 120 VAC/20 A circuits. When doing maintenance

More information

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE by Doug Bray Department of Biological Sciences University of Lethbridge Revised September, 2000 Note: The terms in bold in this document represent

More information

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope The procedures given below were written specifically for the FEI Tecnai G 2 Sphera microscope. Modifications will need to

More information

FEI Tecnai G 2 F20 Operating Procedures

FEI Tecnai G 2 F20 Operating Procedures FEI Tecnai G 2 F20 Operating Procedures 1. Startup (1) Sign-up in the microscope log-sheet. Please ensure you have written an account number for billing. (2) Log in to the computer: Login to your account

More information

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES RULES All users must go through a series of standard operation procedure training. For more information contact: Longlong Liao Teaching

More information

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. Smith College August 2005 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check, 1 Specimen Insertion, 1 Startup, 2 Filament Saturation, 2 Beam Alignment,

More information

Section 1: TEM parts and functions... 2

Section 1: TEM parts and functions... 2 Introduction The set of instructions below are written by Charlie Sanabria within the first few sessions of his TEM training process, and are intended for anyone interested in viewing the TEM operation

More information

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center LEO 912 TEM Short Manual Prepared/copyrighted by RH Berg Danforth Plant Science Center Specimen holder [1] Never touch the holder (outside of the O-ring, double-headed arrow) because finger oils will contaminate

More information

FE-SEM SU-8020 Operating manual (Preliminary version)

FE-SEM SU-8020 Operating manual (Preliminary version) FE-SEM SU-8020 Operating manual (Preliminary version) 2016/04/11 Seimitsu Bunseki sitsu lab. Starting up 1.Turn on the Display switch. Windows OS is starting up 2. Select the user SU-8000. 3. Click the

More information

How to use the Jeol 1010 TEM of GI (Liesbeth own GI version)

How to use the Jeol 1010 TEM of GI (Liesbeth own GI version) How to use the Jeol 1010 TEM of GI (Liesbeth own GI version) 1.Load the specimen Load a grid into the rod holder: USE ONLY THE TOP POSITION (blue arrow), Specimen selection on 1 (The rear one is only a

More information

05/20/14 1. Philips CM200T. Standby Condition

05/20/14 1. Philips CM200T. Standby Condition 05/20/14 1 Philips CM200T Standby Condition HT and filament off, HT setting at 200kV. RESET HOLDER, center sample tilt knobs, and remove sample. Mag ~ 5-10kX Objective and SA apertures out, C2 aperture

More information

Protective Equipment Nitrile gloves for handling sample holder and safety glasses for filling liquid nitrogen dewar.

Protective Equipment Nitrile gloves for handling sample holder and safety glasses for filling liquid nitrogen dewar. Emergency Information: 1. Medical Emergencies: Contact 911 and McGill Security 514.398.3000 2. Leave TEM as is. Do NOT shut down the vacuum system. 3. If possible, turn off High Tension and Close Column

More information

JEOL JEM 2010 TRAINING TRANSMISSION ELECTRON MICROSCOPE USER MANUAL

JEOL JEM 2010 TRAINING TRANSMISSION ELECTRON MICROSCOPE USER MANUAL JEOL JEM 2010 TRAINING TRANSMISSION ELECTRON MICROSCOPE USER MANUAL Version 5.1 EM Facility CMSE-SEF Massachusetts Institution of Technology TABLE OF CONTENTS 1. Specifications...2 1.1 Performance...2

More information

COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual

COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual 1 COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual ABOUT THIS MICROSCOPE (room HG01.240) The JEM-1400Flash

More information

Model SU3500 Scanning Electron Microscope

Model SU3500 Scanning Electron Microscope Model SU3500 Scanning Electron Microscope Modified and Parts taken from Hitachi Easy Operation Guide. Before using the Model SU3500 SEM, be sure to read the [GENERAL SAFETY GUIDELINES] in the instruction

More information

1. Preliminary sample preparation

1. Preliminary sample preparation FEI Helios NanoLab 600 standard operating procedure Nicholas G. Rudawski ngr@ufl.edu (352) 392 3077 (office) (805) 252-4916 (cell) Last updated: 03/02/18 What this document provides: an overview of basic

More information

SEM OPERATION IN LOW VACUUM MODE

SEM OPERATION IN LOW VACUUM MODE SEM OPERATION IN LOW VACUUM MODE Instructions for JEOL 5800 LV The EVAC light of the SEM specimen chamber should be already lit when you approach the SEM & the SEM will have been left in the high vacuum

More information

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Warning Always adjust your specimen height before closing the chamber door to make sure your specimen will not hit the bottom of the lens;

More information

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Amray 1830 SEM. All users are expected to have read and understood this document.

More information

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400. Smith College August 2009 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check 1 Startup 1 Specimen Insertion 2 Filament Saturation 2 Beam Alignment

More information

Jeol JEM Responsible personell: Endy ( ) Online booking is compulsory!

Jeol JEM Responsible personell: Endy ( ) Online booking is compulsory! Jeol JEM 1230 Responsible personell: Endy (45279377) Online booking is compulsory! After training you will have access to working alone on the instrument. All insertion of samples is done by responsible

More information

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the digital camera.

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the digital camera. FEI Tecnai F20 S/TEM: acquiring diffraction patterns Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 10/18/17 This document assumes the user is already familiar with basic

More information

Using the Hitachi 3400-N VP-SEM

Using the Hitachi 3400-N VP-SEM Using the Hitachi 3400-N VP-SEM Opening the Chamber to Load Specimens (This may also be done later using the software) 1. Click the AIR button on the front of the machine: 2. Wait a few minutes until you

More information

JEOL 2010 FasTEM & DigitalMicrograph User's Guide

JEOL 2010 FasTEM & DigitalMicrograph User's Guide JEOL 2010 FasTEM & DigitalMicrograph User's Guide Electron Microscopy Laboratory Instititute of Materials Science University of Connecticut The purpose of this manual is to remind you of the essential

More information

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course)

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course) JEOL 7000F BASIC OPERATING INSTRUCTIONS-Ver.-2.0 Note: This is minimal operation checklist and does not replace the other reference manuals. Read the manual for Specimen Exchange (JEOL 7000 Specimen Exchange

More information

INSTRUCTIONS JEM-2010F FIELD-EMISSION TRANSMISSION ELECTRON MICROSCOPE WITH STEM CAPABILITY

INSTRUCTIONS JEM-2010F FIELD-EMISSION TRANSMISSION ELECTRON MICROSCOPE WITH STEM CAPABILITY INSTRUCTIONS JEM-2010F FIELD-EMISSION TRANSMISSION ELECTRON MICROSCOPE WITH STEM CAPABILITY August 2011 PRELIMINARIES OPERATION 1. Ensure that EMISSION and HT are on: The HT READY and FEG READY lights

More information

Therefore, all descriptions and illustrations in this instruction manual, including all specifications are subject to change without notice.

Therefore, all descriptions and illustrations in this instruction manual, including all specifications are subject to change without notice. We are constantly endeavouring to improve our instruments and to adapt them to the requirements of modern research techniques and testing methods. This involves modification to the mechanical structure

More information

SEM Training Notebook

SEM Training Notebook SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside December 21, 2017 (rev. 3.4) 1 Before you begin Complete

More information

General information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument.

General information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument. JEOL JSM-7400F Table of contents General information.. 3 The operation panel. 4 The different sample holders and inserting the samples.. 5 Turning on the beam... 6 Stage map control... 8 Correcting astigmatism...

More information

FEI Titan Image Corrected STEM

FEI Titan Image Corrected STEM 05/03/16 1 FEI Titan 60-300 Image Corrected STEM Standby Condition HT setting at 300kV, Col. Valves Closed RESET Holder and remove sample. Mag ~ 5-10kX Objective and SA apertures out, C2 aperture at 150µm

More information

SEM Training Notebook

SEM Training Notebook SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside March 8, 2018 (rev. 3.5) 1 Before you begin Complete

More information

CM20 USER GUIDE. Duncan Alexander, CIME 2010

CM20 USER GUIDE. Duncan Alexander, CIME 2010 CM20 USER GUIDE Duncan Alexander, CIME 2010 CM20 START UP AND CHECK LIST 2 SPECIMEN EXCHANGE 5 - REMOVING SAMPLE HOLDER 6 - INSERTING SAMPLE HOLDER 7 TURNING ON HT 8 STARTING THE FILAMENT 9 GUN TILT ALIGNMENT

More information

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com User Manual Digital Compound Binocular LED Microscope Model MD82ES10 MicroscopeNet.com Table of Contents i. Caution... 1 ii. Care and Maintenance... 2 1. Components Illustration... 3 2. Installation...

More information

Transmission Electron Microscopy 9. The Instrument. Outline

Transmission Electron Microscopy 9. The Instrument. Outline Transmission Electron Microscopy 9. The Instrument EMA 6518 Spring 2009 02/25/09 Outline The Illumination System The Objective Lens and Stage Forming Diffraction Patterns and Images Alignment and Stigmation

More information

DISCO DICING SAW SOP. April 2014 INTRODUCTION

DISCO DICING SAW SOP. April 2014 INTRODUCTION DISCO DICING SAW SOP April 2014 INTRODUCTION The DISCO Dicing saw is an essential piece of equipment that allows cleanroom users to divide up their processed wafers into individual chips. The dicing saw

More information

Tecnai T12 Operating Procedures

Tecnai T12 Operating Procedures Tecnai T12 Operating Procedures I. Initial Procedures 1 II. Accelerating Voltage 3 III. Specimen Loading and Holder Insertion/Removal 3 IV. Emission Current 7 V. Alignment 7 VI. Camera Control and Imaging

More information

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual Scanning Electron Microscope FEI INSPECT F50 Step by step operation manual Scanning Electron Microscope, FEI Inspect F50 FE-SEM-F Observation Flow Saving Data And Analysis Specimen preparation Error check

More information

ML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X

ML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X JAPAN DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30 (a) Field Iris Control Lever (c) Filter Slots EYEPIECES, KHW10X ANALYZER CONTROL LEVER (b) Aperture Iris Control Lever LIGHT SOURCE HOUSING VERTICAL

More information

Dickinson College Department of Geology

Dickinson College Department of Geology Dickinson College Department of Geology Title: Equipment: BASIC OPERATION OF THE SCANNING ELECTRON MICROSCOPE (SEM) JEOL JSM-5900 SCANNING ELECTRON MICROSCOPE Revision: 2.2 Effective Date: 1/29/2003 Author(s):

More information

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface.

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface. FEI Tecnai F20 S/TEM: imaging in STEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 05/10/18 This document assumes the user is already familiar with basic operation

More information

MSE 460 TEM Lab 4: Bright/Dark Field Imaging Operation

MSE 460 TEM Lab 4: Bright/Dark Field Imaging Operation MSE 460 TEM Lab 4: Bright/Dark Field Imaging Operation Last updated on 1/8/2018 Jinsong Wu, jinsong-wu@northwestern.edu Aims: The aim of this lab is to familiarize you with bright/dark field imaging operation.

More information

User instructions Compound laboratory microscope

User instructions Compound laboratory microscope KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-mail: info@kern-sohn.com User instructions Compound laboratory microscope Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com KERN

More information

Manual for BMS E1 eplan series, compound microscope

Manual for BMS E1 eplan series, compound microscope Manual for BMS E1 eplan series, compound microscope The compound microscope allows it to study, at cell level, structures of textures of botanical and zoological nature. (e.g. slides of roots, leaves and

More information

Zoom Stereo Microscope NYMCS-360 Instruction Manual

Zoom Stereo Microscope NYMCS-360 Instruction Manual Zoom Stereo Microscope NYMCS-60 Instruction Manual This manual is written for stereo microscope NYMCS-60. To ensure the safety, obtain optimum performance and to familiarize yourself fully with the use

More information

User instructions Compound laboratory microscope

User instructions Compound laboratory microscope KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-mail: info@kern-sohn.com User instructions Compound laboratory microscope Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com KERN

More information

SAMUEL ROBERTS NOBLE ELECTRON MICROSCOPY LABORATORY. Operating Procedures for the Zeiss 9 S-2. Transmission Electron Microscope

SAMUEL ROBERTS NOBLE ELECTRON MICROSCOPY LABORATORY. Operating Procedures for the Zeiss 9 S-2. Transmission Electron Microscope 1 SAMUEL ROBERTS NOBLE ELECTRON MICROSCOPY LABORATORY Operating Procedures for the Zeiss 9 S-2 Transmission Electron Microscope Prepared by Dr. Scott D. Russell Department of Botany and Microbiology September,

More information

PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry

PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry Purpose PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry In this experiment, you will study the principles and applications of interferometry. Equipment and components PASCO

More information

RENISHAW INVIA RAMAN SPECTROMETER

RENISHAW INVIA RAMAN SPECTROMETER STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw invia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples

More information

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE University of Virginia ITC Academic Computing Health Sciences CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE Introduction The Olympus BH-2 microscope in ACHS s microscope lab has objectives from

More information

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only Version 1.0 Prepared by D. Turnbull February 21, 2007. Please submit any omissions to the Author Note: This SEM is a recent

More information

CATHODOLUMINESCENCE IMAGING UNIVERSITY OF ST ANDREWS

CATHODOLUMINESCENCE IMAGING UNIVERSITY OF ST ANDREWS CATHODOLUMINESCENCE IMAGING AT THE UNIVERSITY OF ST ANDREWS Instruction Guide by Adrian Finch 2 Noddy s Guide to Using the CL at St Andrews (SHORT version) 1. Switch on plugs at Wall (Microscope, Luminoscope

More information

RAITH e-line OPERATING INSTRUCTIONS

RAITH e-line OPERATING INSTRUCTIONS RAITH e-line OPERATING INSTRUCTIONS 1) LOADING A SAMPLE a. Start the system i. On the Column PC (Right side monitor [R]), select the SmartSEM icon to on the desktop to begin the column software. ii. On

More information

Introduction: Why electrons?

Introduction: Why electrons? Introduction: Why electrons? 1 Radiations Visible light X-rays Electrons Neutrons Advantages Not very damaging Easily focused Eye wonderful detector Small wavelength (Angstroms) Good penetration Small

More information

Last updated 6/12/18. F20 User Manual at the Simons Electron Microscopy Center

Last updated 6/12/18. F20 User Manual at the Simons Electron Microscopy Center F20 User Manual at the Simons Electron Microscopy Center 1 Table of Contents F20 Information Sheet 2 F20 User Guide (starting your session) 3 F20 User Guide (ending your session) 5 Cryo Screening with

More information

User Manual. Trinocular Metallurgical Microscope. MicroscopeNet.com

User Manual. Trinocular Metallurgical Microscope. MicroscopeNet.com User Manual Trinocular Metallurgical Microscope Model M83MPTR MicroscopeNet.com Table of Contents i. Caution.. 1 ii. Care and Maintenance... 2 1. Components Illustration..... 3 2. Installation...4 3. Operation

More information

TABLE OF CONTENTS. Safety notes i. Care and Maintenance. ii. 1. Components Illustration Installation of Components.. 4

TABLE OF CONTENTS. Safety notes i. Care and Maintenance. ii. 1. Components Illustration Installation of Components.. 4 TABLE OF CONTENTS Safety notes i Care and Maintenance. ii 1. Components Illustration... 1 2. Installation of Components.. 4 2.1 Installation Diagram... 4 2.2 Installation Procedures 5 3. Operation...11

More information

Eyepieces KHW10X. Diopter Adjustment Ring. Binocular Body Inclined 30. Binocular Clamp Screw. Analyzer control Lever. Reflected Light Illuminator

Eyepieces KHW10X. Diopter Adjustment Ring. Binocular Body Inclined 30. Binocular Clamp Screw. Analyzer control Lever. Reflected Light Illuminator JAPAN Eyepieces KHW10X Diopter Adjustment Ring Binocular Body Inclined 30 Binocular Clamp Screw Analyzer control Lever Reflected Light Illuminator Ball-Bearing Objective Nosepiece Objectives Large Scan

More information

User instructions Metallurgical inverted microscope

User instructions Metallurgical inverted microscope KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-mail: info@kern-sohn.com Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com User instructions Metallurgical inverted microscope

More information

Standard Operating Procedure Hitachi UHR CFE SU8230 SEM

Standard Operating Procedure Hitachi UHR CFE SU8230 SEM Standard Operating Procedure Hitachi UHR CFE SU8230 Yale West Campus Materials Characterization Core ywcmatsci.yale.edu ESC II, Room E119F 810 West Campus Drive West Haven, CT 06516 Version 1.1, May 2016

More information

ZEISS EVO SOP. May 2017 ELECTRON OPTICS

ZEISS EVO SOP. May 2017 ELECTRON OPTICS ZEISS EVO SOP May 2017 ELECTRON OPTICS The patented EVO column is the area of the SEM, where electrons are emitted, accelerated, deflected, focused, and scanned. Main characteristics of the EVO optics

More information

OM FL400. Reflected Light Fluorescence Microscope. Instruction Manual. Please read instructions carefully before using microscope.

OM FL400. Reflected Light Fluorescence Microscope. Instruction Manual. Please read instructions carefully before using microscope. OM FL400 Reflected Light Fluorescence Microscope Instruction Manual Please read instructions carefully before using microscope. Contents Safety ---------------------------------------------- 2 Parts List

More information

OMM300. Inverted Metallurgical Microscope

OMM300. Inverted Metallurgical Microscope OMM300 Inverted Metallurgical Microscope Instruction Manual Please read the instructions carefully before operating CONTENTS Safety 2 Parts List 2 Features 3 Assembly 5 Operation 7 Maintenance 9 Specifications

More information

RIGAKU VariMax Dual Part 0 Startup & Shutdown Manual

RIGAKU VariMax Dual Part 0 Startup & Shutdown Manual i RIGAKU VariMax Dual Part 0 Startup & Shutdown Manual X-ray Laboratory, Nano-Engineering Research Center, Institute of Engineering Innovation, School of Engineering, The University of Tokyo Figure 0:

More information

Instruction Manual T Binocular Acromat Research Scope T Trinocular Acromat Research Scope

Instruction Manual T Binocular Acromat Research Scope T Trinocular Acromat Research Scope Research Scope Instruction Manual T-29031 Binocular Acromat Research Scope T-29041 Trinocular Acromat Research Scope T-29032 Binocular Semi-Plan Research Scope T-29042 Trinocular Semi-Plan Research Scope

More information

Introduction of New Products

Introduction of New Products Field Emission Electron Microscope JEM-3100F For evaluation of materials in the fields of nanoscience and nanomaterials science, TEM is required to provide resolution and analytical capabilities that can

More information

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment FAST SHIPPING AND DELIVERY TENS OF THOUSANDS OF IN-STOCK ITEMS EQUIPMENT DEMOS HUNDREDS OF MANUFACTURERS SUPPORTED

More information

XTEM. --Software for Complex Transmission Electron Microscopy. Version 1.0

XTEM. --Software for Complex Transmission Electron Microscopy. Version 1.0 XTEM --Software for Complex Transmission Electron Microscopy Version 1.0 1. Introduction XTEM is the software for complex microscopy on JEOL 3100 electron microscopes. The XTEM software consists of a suite

More information

CNC Using the FlexiCam CNC and HMI Software. Guldbergsgade 29N, P0 E: T:

CNC Using the FlexiCam CNC and HMI Software. Guldbergsgade 29N, P0 E: T: CNC Using the FlexiCam CNC and HMI Software Guldbergsgade 29N, P0 E: makerlab@kea.dk T: +46 46 03 90 This grey box is the NC controller. Let s start by turning the red switch to the ON position, then press

More information

Operating F20/F30 with SerialEM

Operating F20/F30 with SerialEM Chen Xu xuchen@brandeis.ede $BrandeisEM: ~emdoc-xml/en_us.iso8859-1/articles/operating-f20-or-f30/article.xml, 1 2013-01-19 01:42:20 xuchen Exp$ This is a quick check list for the Tecnai F20 or Tecnai

More information

User instructions Metallurgical microscope

User instructions Metallurgical microscope KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com User instructions Metallurgical microscope Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com KERN OKM-1

More information

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and

More information

Biological Microscope Manual

Biological Microscope Manual Version No.: V1.2 Series Biological Microscope Manual This manual expatiates the using method, troubleshooting and maintenance about MT-50 series biological microscope. Please study this manual thoroughly

More information

Zeiss AxioImager.Z2 Brightfield Protocol

Zeiss AxioImager.Z2 Brightfield Protocol Zeiss AxioImager.Z2 Brightfield Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge

More information

User Manual. Trinocular Infinity Compound LED Microscope. MicroscopeNet.com

User Manual. Trinocular Infinity Compound LED Microscope. MicroscopeNet.com User Manual Trinocular Infinity Compound LED Microscope Model M8333Z series MicroscopeNet.com Table of Contents i. Caution... 1 ii. Care and Maintenance... 2 1. Components Illustration... 3 2. Installation...

More information

Service Manual for XLE/XLT Series Laser Engravers

Service Manual for XLE/XLT Series Laser Engravers Service Manual for XLE/XLT Series Laser Engravers Table of Contents Maintenance...1 Beam alignment...3 Auto focus alignment...8 Bridge alignment...10 Electronics panel replacement...11 X motor change...12

More information

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer Brightfield Microscopy and Image Acquisition on Spotcam1 by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer Log onto the computer. Enter your username and password to log onto the server.

More information

Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE

Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE Prior to lab: 1) Read these instructions (p 1-6) 2) Go through the online tutorial, the microscopy pre-lab

More information

AxioVision 4.5 Brightfield Image Capture Procedure

AxioVision 4.5 Brightfield Image Capture Procedure AxioVision 4.5 Brightfield Image Capture Procedure 1. STARTING-UP PROCEDURE: Remove blue dust cover and place on shelf under microscope. Turn on the halogen lamp by pushing the switch at the back right

More information

S200 Course LECTURE 1 TEM

S200 Course LECTURE 1 TEM S200 Course LECTURE 1 TEM Development of Electron Microscopy 1897 Discovery of the electron (J.J. Thompson) 1924 Particle and wave theory (L. de Broglie) 1926 Electromagnetic Lens (H. Busch) 1932 Construction

More information

ELECTRON MICROSCOPY. 13:10 16:00, Oct. 6, 2008 Institute of Physics, Academia Sinica. Tung Hsu

ELECTRON MICROSCOPY. 13:10 16:00, Oct. 6, 2008 Institute of Physics, Academia Sinica. Tung Hsu ELECTRON MICROSCOPY 13:10 16:00, Oct. 6, 2008 Institute of Physics, Academia Sinica Tung Hsu Department of Materials Science and Engineering National Tsing Hua University Hsinchu 300, TAIWAN Tel. 03-5742564

More information

SOP: Polyvar Met Light Microscope

SOP: Polyvar Met Light Microscope SOP Polyvar Met Light Microscope Page 1 of 8 SOP: Polyvar Met Light Microscope 1. Scope 1.1 This document describes the standard operating procedure (SOP) for the Polyvar Met Light Microscope. 2. Table

More information

Scanning electron microscope

Scanning electron microscope Scanning electron microscope 6 th CEMM workshop Maja Koblar, Sc. Eng. Physics Outline The basic principle? What is an electron? Parts of the SEM Electron gun Electromagnetic lenses Apertures Chamber and

More information