RENISHAW RAMAN MICROSCOPE STANDARD OPERATING PROCEDURE

Size: px
Start display at page:

Download "RENISHAW RAMAN MICROSCOPE STANDARD OPERATING PROCEDURE"

Transcription

1 RENISHAW RAMAN MICROSCOPE STANDARD OPERATING PROCEDURE 09DJ00 CHRIS BUXEY (TECHNICIAN) X2715 V5.0 OCTOBER 2008

2 RENISHAW MICRORAMAN STANDARD OPERATING PROCEDURE 1.0 Foreword Chris Buxey: This guide is intended as a reference and a reminder for those already familiar with the microraman. If you have never used the microraman before or are trying to do something complicated or unusual, then please contact me for training and/or assistance. Oh, and please REMEMBER TO BOOK and to FILL IN THE LOG BOOK! Dr Henryk Herman: Warning, this is a delicate and expensive instrument. If it is abused, the user will be banned, and disciplinary proceedings may be instigated. Use the log book. It enables us to keep track of the state of the instrument. Any problems or queries, contact: Chris Buxey, x2715 Dr Richard Curry, x2713, or Dr. Henryk Herman, The Renishaw Raman Microscope (Systems 2000) is a compact laser Raman microprobe/microscope that can collect both Raman spectra and global images. The collection optics are based on a Leica DMLM microscope. The samples are placed underneath the microscope objective (x5, 20, 50) and are excited by the laser directed through the microscope. The scattered light is collected (180 degree backscattering) along the same optical path as the incoming laser. Raman scattered light passes through a notch filter (used as one of the reflective elements for the incident laser beam) into the monochromator that disperses the scattered light over a CCD array detector. The spectrometer is controlled by PC using two software packages, instrument control software (Renishaw WiRE ) and data analysis software (Galactic GRAMS/32 ). Safety Procedure: Both lasers in this system as classified Class 3B, which denotes lasers or laser systems that can produce a hazard if viewed directly. This includes intrabeam viewing or specular reflections. The arrangement is such that hazardous conditions are only present within 2 mm of the focal plane, so in normal operation there will be no material hazard. Servicing, or change of laser, will require access to the spectrometer and laser beams. Under no circumstances should unsupervised, untrained users attempt to gain access. You have been warned!

3 2.0 Switching On Procedure Fig1. Renishaw System 2000 microraman Spectrometer. See schematic in Appendix. Sign and date the microraman instrument log book; The Raman system is normally left switched on. Check the illuminated Raman system power switch; if it is off, switch on. The PC controller is normally left on. Check the power button; if it is off, switch on. The PC flat screen monitor is normally off. Switch on. The software is normally left running. If it isn t, click on the shortcut Renishaw Wire. The Raman software starts. Always Check motors. If this process hangs on Focusing Motor for several minutes, contact Chris Buxey. Whilst the system is going through the switch-on procedure (about 5 minutes) switch on microscope illumination. Decide which laser, 514 or 782 nm, you require. Note that changing from one to the other is an involved process that requires opening and working within the spectrometer. It MUST NOT be carried out by anyone unfamiliar with the necessary alignment procedure and you must have authorisation from R Curry or H Herman. The CCD detector could be destroyed, and your group will be billed. There is a tell-tale on top of the spectrometer. The 782 nm laser (white housing) is associated with LASER1 on the safety interlock box. The 514 nm laser (black housing) is associated with LASER2 on the safety interlock box. Most operations are performed with LASER1.

4 LASER1 switch-on procedure: o Turn key switch on LASER1 quarter-turn clockwise; o There are three lights: INTERLOCK, TEMP and POWER. Under normal switch-on INTERLOCK and POWER are green, and TEMP will turn from red to green once correct internal laser parameters have been established (about three minutes). o If INTERLOCK is red, check The laser interlock switch is set to LASER1 The spectrometer housing is fully closed. o Wait until all lights turn green. If they do not: Log the event Switch off the laser at the key switch Get assistance. o Check LASER1 power by moving the slide on top of LASER1 from LASER to ADJUST. The reading should be 15 to 21 mw. If it is not: Switch-off the laser at the key switch. Wait two minutes. Restart the switch-on procedure. If after several attempts the laser is still not producing power, get assistance. o Maximise the power reading by using the correct Allen key (on top of LASER1 housing through hole A. Engage the Allen key and move gently anti- and clockwise. LASER2 switch-on procedure: o Turn on LASER2 Power at the wall switch o Turn on POWER at the square black power supply o Turn key-switch quarter turn clockwise o Wait 2 minutes o On laser operator panel ensure toggle switches are set for POWER, 50mW and STANDBY o On laser operator panel turn key switch quarter-turn clockwise o Wait 5 minutes o Set toggle to RUN. o No checking is required for LASER2 operation.

5 3.0 Beam Positioning Procedure The Beam Positioning Procedure is required to ensure that the visible and laser beams are collinear and co-focal: o Set microscope turret to x5 OBJ o Put SILICON CALIBRANT slide in position on the microscope stage o Ensure MICROSCOPE MIRROR is set to 2 on the rotary bezel, and MICROSCOPE FILTER is set to 1. This allows the microscope CCD to see the sample o Click on the video tool in the Grams/32 software panel o Using the manual adjusters on the microscope, focus on the silicon surface. Note that the big outer ring on the adjuster is coarse, whilst the inner knob is fine o Adjust level of illumination via ILLUMINATION ADJUST rotary at base of microscope o Set microscope turret to x20 OBJ o Refocus o Set microscope turret to x50 OBJ o Refocus o Allow laser light onto calibrant/sample by setting MICROSCOPE MIRROR to 1 on the rotary bezel, and MICROSCOPE FILTER to 2 o A bright tightly focussed laser beam should be visible. If not contact Chris Buxey. A Warning on the Focusing Motor: Please do not use the microraman control software to adjust the internal laser spot focusing motor. This has already been set correctly, so if your sample is focused optically, the laser spot should also be in focus. Fine adjustments may be made using the fine focus dial on the microscope. If you believe there to be a problem then please contact Chris Buxey on x2715.

6 4.0 Spectrometer Calibration Procedure Spectrometer calibration o In the Toolbox/Experiment Set-Up, ensure Type is Grating, Spectrum and Static. Centre at Set (Detector) Time to 1.00, Gain High, (Misc.) Accum 1, Cos. Ray Rem. Off, Power Obj 50 and Focus 0. Set Descriptor to Silicon Calibration: date. Click OK. o Click the Spectrum Acquire button twice. o Using Windows technique, expand peak so display shows approx. 460 to 550 cm -1. o Click Arithmetic, Curve_Fit. Right click ONCE on peak of spectrum. Then DONE, ITERATE, Start Fit and OK. o Click View Peaks, and Parameters. o The center value should be (+/-0.5). If it is, log it, as well as Height and Width. The Height value should be at least 2000, and the Width, 6 or less. o If the Center Value is not within bounds, note it and quit the curve fit procedure by Parameters, Exit, Quit and Exit. Now Collect, Applications, Calibrate System, Calibrate Grating and Offset Correction. Compute ( Center value) and enter this in Required Shift, then Apply, Done. Repeat the Spectrum Acquire procedure. o If the Height value is low, or the Width is too high, check the beam quality at the sample. If necessary, repeat the Beam Positioning Procedure. The system will operate with these parameters out-of-bounds, but data will be degraded. o Note the values for the Centre Value, Height and Width in the log book.

7 5.0 Sampling Procedure Measuring samples is very similar to the Beam Positioning Procedure. Always start at the lowest magnification and work up. Remember to move mirror and filter wheels to allow laser to fall on sample before acquiring spectrum. If the objective touches the sample, get assistance. o If you can, place sample on Al foil (wrapped round microscope slide) o The optical arrangement of the system is such that only about 400 cm -1 can be acquired in the static mode of the grating. For larger scans, within Toolbox/Experiment Set-Up, ensure Type is Grating, Spectrum and Extended. o This option brings up an extra box, which provides three choices: Continuous this is the favoured option, which synchronises the movement of the grating with the CCD readout speed. It provides the best quality spectra. However, it can only be used for Detector Time of 10 seconds or greater. Step & Repeat acquires spectra in approximately 400 cm -1 blocks, and displays them. Matched Step & Repeat as Step & Repeat but joins the spectra by overlapping them and averaging the overlap region. o This option also changes the Centre option to Range, providing From and To boxes. Note that there is no sensible data below 170 cm -1 with the current apparatus. o Note that the combination of LASER1 (782 nm) and the silicon-based CCD detector means that signals above about 2,500 cm -1 are weak. o The dynamic range of the CCD detector is about 30,000 counts. To obtain the best signal to noise use the highest possible power, and adjust the Detector Time to ensure the dynamic range is filled. o However, be careful about power density. Although the laser only produces a maximum of 13 mw at the sample, the spot size is submicron. If in doubt, choose a small bit of sample, and subject it to the laser for 2 minutes. Check its physical appearance. Especially in dark samples, it will be advantageous to reduce the power and increase the integration time. o The laser power can be adjusted in steps via the Power option 100%, 50%, 25%, 10% and 1%. These values are approximate. o Spectra are NOT automatically saved. Ensure that you save what you want into YOUR folder.

8 6.0 Exporting Your Work The procedure for exporting your data is as follows. Note that only spectra that have been saved may be exported. o Select the File menu, then Import/Export. o Choose ASCII_XY from the list and click the Export button. o Select the data you wish to export and click OK. o Change the name of the exported file if you wish, then click OK. o Your ASCII data will be exported as a.prn file to the same directory as the saved data file that it was generated from. o You can now copy the exported data to either 3½ floppy disk or zip disk. 7.0 Shutting Down Procedure Shut down procedure: o Switch off laser o Replace samples with the silicon calibrant o Leave software and PC running o Switch off screen only o Sign instrument log-book with time out

9 8.0 Appendix 8.1 Version History Version: 1.0, dated 3 rd September, 2003, Author: Henryk Herman Version: 1.1, dated 19 th November, 2003 Added Beam Positioning Procedure, HH Version 2.0, dated 15 th October, 2005 Modified Sampling Procedure, HH Version: 3.0, dated 24 th October, 2006 Acknowledge changed of location to 09DJ00, HH Version 4.0, dated 25 th October, 2006, HH Inserted additional restriction on changing laser permission must be sought from RJC or HH. Version 5.0, dated 10 th October, 2008, Author: Chris Buxey Minor updates to procedures to bring them into line with current practise. Additional foreword and reformatting added. Section on exporting data files added.

10 8.2 Internal Schematic Microscope Illumination Microscope Manual Z-axis Raman System Power Switch Microscope Mirror Microscope Filter

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and

More information

RENISHAW INVIA RAMAN SPECTROMETER

RENISHAW INVIA RAMAN SPECTROMETER STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw invia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples

More information

Renishaw InVia Raman microscope

Renishaw InVia Raman microscope Laser Spectroscopy Labs Renishaw InVia Raman microscope Operation instructions 1. Turn On the power switch, system power switch is located towards the back of the system on the right hand side. Wait ~10

More information

WITec Alpha 300R Quick Operation Summary October 2018

WITec Alpha 300R Quick Operation Summary October 2018 WITec Alpha 300R Quick Operation Summary October 2018 This document is frequently updated if you feel information should be added, please indicate that to the facility manager (currently Philip Carubia,

More information

Leica SPEII confocal microscope. Short Manual

Leica SPEII confocal microscope. Short Manual Leica SPEII confocal microscope Short Manual Switching ON sequence: 1. Turn on the Workstation under the bench (top, far right). 2. Turn on the Supply Unit - Laser box (big green switch first and then

More information

Renishaw InVia Quick Operation Summary October 2018

Renishaw InVia Quick Operation Summary October 2018 Renishaw InVia Quick Operation Summary October 2018 This document is frequently updated if you feel information should be added, please indicate that to the facility manager (currently Philip Carubia,

More information

CONFOCAL MICROSCOPE CM-1

CONFOCAL MICROSCOPE CM-1 CONFOCAL MICROSCOPE CM-1 USER INSTRUCTIONS Scientific Instruments Dr. J.R. Sandercock Im Grindel 6 Phone: +41 44 776 33 66 Fax: +41 44 776 33 65 E-Mail: info@jrs-si.ch Internet: www.jrs-si.ch 1. Properties

More information

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop 1. Log in usage using the SMIF web site 2. Turn power

More information

Kit for building your own THz Time-Domain Spectrometer

Kit for building your own THz Time-Domain Spectrometer Kit for building your own THz Time-Domain Spectrometer 16/06/2016 1 Table of contents 0. Parts for the THz Kit... 3 1. Delay line... 4 2. Pulse generator and lock-in detector... 5 3. THz antennas... 6

More information

Zeiss LSM 780 Protocol

Zeiss LSM 780 Protocol Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

ENSC 470/894 Lab 3 Version 6.0 (Nov. 19, 2015)

ENSC 470/894 Lab 3 Version 6.0 (Nov. 19, 2015) ENSC 470/894 Lab 3 Version 6.0 (Nov. 19, 2015) Purpose The purpose of the lab is (i) To measure the spot size and profile of the He-Ne laser beam and a laser pointer laser beam. (ii) To create a beam expander

More information

Guide to Confocal 5. Starting session

Guide to Confocal 5. Starting session Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope

More information

Things to check before start-up.

Things to check before start-up. Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your

More information

Zeiss AxioImager.Z2 Brightfield Protocol

Zeiss AxioImager.Z2 Brightfield Protocol Zeiss AxioImager.Z2 Brightfield Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge

More information

Zeiss LSM 880 Protocol

Zeiss LSM 880 Protocol Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Confocal Raman Microscopy (WITec Alpha 300R)

Confocal Raman Microscopy (WITec Alpha 300R) Confocal Raman Microscopy (WITec Alpha 300R) Please refer to Witec Alpha300R Confocal Raman Microscope User Manual for the details of the operating procedure. Sample preparation 1. Attach your sample on

More information

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier). JEOL 200CX operating procedure Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Specimen loading 1.1. Unlock the TUMI system. 1.2. Load specimen(s) into the holder. If using the double tilt holder, ensure

More information

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120)

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120) Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120) Please contact Dr. Amanda Henkes for training requests and assistance: 979-862-5959, amandahenkes@tamu.edu Hardware LN 2 FTIR FTIR camera 1

More information

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO TRAINING MANUAL Multiphoton Microscopy LSM 510 META-NLO September 2010 Multiphoton Microscopy Training Manual Multiphoton microscopy is only available on the LSM 510 META-NLO system. This system is equipped

More information

Microscopy. The dichroic mirror is an important component of the fluorescent scope: it reflects blue light while transmitting green light.

Microscopy. The dichroic mirror is an important component of the fluorescent scope: it reflects blue light while transmitting green light. Microscopy I. Before coming to lab Read this handout and the background. II. Learning Objectives In this lab, you'll investigate the physics of microscopes. The main idea is to understand the limitations

More information

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: )

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: ) University of Minnesota College of Science and Engineering Characterization Facility PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: 2012.10.17) The following instructions

More information

Standard Operating Procedure

Standard Operating Procedure Standard Operating Procedure Nanosurf Atomic Force Microscopy Operation Facility NCCRD Nanotechnology Center for Collaborative Research and Development Department of Chemistry and Engineering Physics The

More information

Instructions for the Experiment

Instructions for the Experiment Instructions for the Experiment Excitonic States in Atomically Thin Semiconductors 1. Introduction Alongside with electrical measurements, optical measurements are an indispensable tool for the study of

More information

SOP: Polyvar Met Light Microscope

SOP: Polyvar Met Light Microscope SOP Polyvar Met Light Microscope Page 1 of 8 SOP: Polyvar Met Light Microscope 1. Scope 1.1 This document describes the standard operating procedure (SOP) for the Polyvar Met Light Microscope. 2. Table

More information

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 CIAN LSM1 or LSM2 short instructions, version 1.4, September 2014 page 1 of 6 SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 Before starting To work with LSM1

More information

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 710 Confocal Microscope Standard Operation Protocol LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Version : 07/08/0 Leica SP8 TCS Users Manual Start up:. Turn the PC Microscope, Scanner Power, Laser Power, and the Laser Emission key to on (bottom right of desk).. Turn on the fluorescent lamp (top left

More information

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center LEO 912 TEM Short Manual Prepared/copyrighted by RH Berg Danforth Plant Science Center Specimen holder [1] Never touch the holder (outside of the O-ring, double-headed arrow) because finger oils will contaminate

More information

Using the Nikon TE2000 Inverted Microscope

Using the Nikon TE2000 Inverted Microscope Wellcome Trust Centre for Human Genetics Molecular Cytogenetics and Microscopy Core Using the Nikon TE2000 Inverted Microscope Fluorescence image acquisition using Scanalytic s IPLab software and the B&W

More information

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right Quick Guide to Operating FEI Titan Themis G2 200 (S)TEM: TEM mode Susheng Tan Nanoscale Fabrication and Characterization Facility, University of Pittsburgh Office: M104/B01 Benedum Hall, 412-383-5978,

More information

Guide to SPEX Optical Spectrometer

Guide to SPEX Optical Spectrometer Guide to SPEX Optical Spectrometer GENERAL DESCRIPTION A spectrometer is a device for analyzing an input light beam into its constituent wavelengths. The SPEX model 1704 spectrometer covers a range from

More information

Operating Instructions for Zeiss LSM 510

Operating Instructions for Zeiss LSM 510 Operating Instructions for Zeiss LSM 510 Location: GNL 6.312q (BSL3) Questions? Contact: Maxim Ivannikov, maivanni@utmb.edu 1 Attend A Complementary Training Before Using The Microscope All future users

More information

SOP-P051. Scanning of Optical Filters With USB2000. Objective: To determine the spectral transmittance properties of an optical filter.

SOP-P051. Scanning of Optical Filters With USB2000. Objective: To determine the spectral transmittance properties of an optical filter. Purdue University Cytometry Laboratories SOP-P051 Scanning of Optical Filters With USB2000 Objective: To determine the spectral transmittance properties of an optical filter. Procedure: 1. Ensure that

More information

Spectroscopy of Ruby Fluorescence Physics Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018

Spectroscopy of Ruby Fluorescence Physics Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018 1 Spectroscopy of Ruby Fluorescence Physics 3600 - Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018 I. INTRODUCTION The laser was invented in May 1960 by Theodor Maiman.

More information

3B SCIENTIFIC PHYSICS

3B SCIENTIFIC PHYSICS B SCIENTIFIC PHYSICS Cavendish Torsion Balance 007 Operating instructions 06/8 ALF. Description The Cavendish torsion balance is for demonstrating the gravitational attraction between two masses and determining

More information

PANalytical X pert Pro High Resolution Specular and Rocking Curve Scans User Manual (Version: )

PANalytical X pert Pro High Resolution Specular and Rocking Curve Scans User Manual (Version: ) University of Minnesota College of Science and Engineering Characterization Facility PANalytical X pert Pro High Resolution Specular and Rocking Curve Scans User Manual (Version: 2012.10.17) The following

More information

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands Differential Interference Contrast (DIC) imaging is a technique used to increase contrast in brightfield images. In confocal systems,

More information

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Introduction of Fluoresence Confocal Microscopy The first confocal microscope was invented by Princeton

More information

MSE 460 TEM Lab 4: Bright/Dark Field Imaging Operation

MSE 460 TEM Lab 4: Bright/Dark Field Imaging Operation MSE 460 TEM Lab 4: Bright/Dark Field Imaging Operation Last updated on 1/8/2018 Jinsong Wu, jinsong-wu@northwestern.edu Aims: The aim of this lab is to familiarize you with bright/dark field imaging operation.

More information

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017. Nikon AZ100 Laser Scanning Macro Confocal Microscope Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin May 2017 Contents 1 Introduction 2 2 Hardware - Startup 2 3 Software/Operation 4 3.1 Multidimensional

More information

UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY

UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY References: http://virlab.virginia.edu/vl/home.htm (University of Virginia virtual lab. Click on the AFM link) An atomic force

More information

FEI Tecnai G 2 F20 Operating Procedures

FEI Tecnai G 2 F20 Operating Procedures FEI Tecnai G 2 F20 Operating Procedures 1. Startup (1) Sign-up in the microscope log-sheet. Please ensure you have written an account number for billing. (2) Log in to the computer: Login to your account

More information

IBIL setup operation manual for SynerJY software version

IBIL setup operation manual for SynerJY software version IBIL setup operation manual for SynerJY software version 1.8.5.0 Manual version 1.0, 31/10/2008 Author: Carlos Marques Equipment Managers: Carlos Marques, +351219946084, cmarques@itn.pt Luís Alves, +351219946112,

More information

Practical work no. 3: Confocal Live Cell Microscopy

Practical work no. 3: Confocal Live Cell Microscopy Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside

More information

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands Differential Interference Contrast (DIC) imaging is a technique used to increase contrast in brightfield images. In confocal systems,

More information

Parameter Selection and Spectral Optimization Using the RamanStation 400

Parameter Selection and Spectral Optimization Using the RamanStation 400 Parameter Selection and Spectral Optimization Using the RamanStation 400 RAMAN SPECTROSCOPY A P P L I C A T I O N N O T E In modern dispersive Raman spectroscopy, good quality spectra can be obtained from

More information

Technische Universität Chemnitz Physikalische Chemie. Laser Safety

Technische Universität Chemnitz Physikalische Chemie. Laser Safety Technische Universität Chemnitz Physikalische Chemie Laser Safety. Lasers and laser systems are classified by their ability to cause biological damage to the eye or skin during use. Class I: Lasers or

More information

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope Nikon Eclipse Ti A1-A Confocal Operating Manual Start-up 1. Turn on Excite Fluorescent light power supply- metal halide. a. Cool down as for mercury bulb b. Wheel closed liquid light guide 2. Turn on power

More information

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE by Doug Bray Department of Biological Sciences University of Lethbridge Revised September, 2000 Note: The terms in bold in this document represent

More information

Jeol JEM Responsible personell: Endy ( ) Online booking is compulsory!

Jeol JEM Responsible personell: Endy ( ) Online booking is compulsory! Jeol JEM 1230 Responsible personell: Endy (45279377) Online booking is compulsory! After training you will have access to working alone on the instrument. All insertion of samples is done by responsible

More information

LEICA TCS SP5 AOBS TANDEM USER MANUAL

LEICA TCS SP5 AOBS TANDEM USER MANUAL LEICA TCS SP5 AOBS TANDEM USER MANUAL STARTING THE SYSTEM...2 THE LAS AF SOFTWARE...3 THE «ACQUIRE» MENU...5 CHOOSE AND CREATE A SETTING...6 THE CONTROL PANEL...8 THE DMI6000B MICROSCOPE...10 ACQUIRE ONE

More information

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL MIF ZEISS LSM510 CONFOCAL USER PROTOCOL START-UP Turn on the Mercury Bulb Power Supply (if needed). Power-on the Control Box. Turn on the computer. Open the LSM 510 software. Choose Scan New Images and

More information

Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation

Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Rotation By: Michael Case and Roy Grayzel, Acton Research Corporation Introduction The majority of modern spectrographs and scanning

More information

Zeiss Axio Imager.A1 manual

Zeiss Axio Imager.A1 manual Zeiss Axio Imager.A1 manual Power-up protocol 1. Mercury lamp 2. Power strip on shelf 3. Computer The Mercury lamp should always be first-on and last-off. This prevents any electrical surges caused by

More information

CONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application

CONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application CONFIGURING Your Spectroscopy System For PEAK PERFORMANCE A guide to selecting the best Spectrometers, s, and s for your application Spectral Measurement System Spectral Measurement System Spectrograph

More information

Instructions XRD. 1 Choose your setup , Sami Suihkonen. General issues

Instructions XRD. 1 Choose your setup , Sami Suihkonen. General issues Instructions XRD 28.10.2016, Sami Suihkonen General issues Be very gentle when closing the doors Always use Cu attenuator when count rate exceeds 500 000 c/s Do not over tighten optical modules or attach

More information

Instructions for Making On-Line Reservations for Microscopes in NB11-204

Instructions for Making On-Line Reservations for Microscopes in NB11-204 Instructions for Making On-Line Reservations for Microscopes in NB11-204 1. Log into Mail using Mail.swmed.edu 2. Log in using your university id and password. 3. Click the Calendar Tab at the top right

More information

KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017)

KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017) KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017) 1 Introduction You must be trained to operate the Laser-scanning confocal microscope (LSCM) independently.

More information

Ocean Optics R-2000 Raman Spectrometer Setup and Operating Instructions Arlen Viste and Deanna Donohoue April 2000 Update 2003, DEW

Ocean Optics R-2000 Raman Spectrometer Setup and Operating Instructions Arlen Viste and Deanna Donohoue April 2000 Update 2003, DEW Ocean Optics R-2000 Raman Spectrometer Setup and Operating Instructions Arlen Viste and Deanna Donohoue April 2000 Update 2003, DEW References Raman Systems R-2000 Operating Manual, Version 1.6, Ocean

More information

LSM 800 Confocal Microscope Standard Operation Protocol

LSM 800 Confocal Microscope Standard Operation Protocol LSM 800 Confocal Microscope Standard Operation Protocol Turning on the system 1. Switch on the Main switch (labeled 1 and 2 ) mounted on the wall. 2. Turn the Laser Key (labeled 3 ) 90 clockwise for power

More information

Laser Telemetric System (Metrology)

Laser Telemetric System (Metrology) Laser Telemetric System (Metrology) Laser telemetric system is a non-contact gauge that measures with a collimated laser beam (Refer Fig. 10.26). It measure at the rate of 150 scans per second. It basically

More information

The user should already be familiar with operation of the instrument in STEM mode, use of the Microscope Control interface, and TIA.

The user should already be familiar with operation of the instrument in STEM mode, use of the Microscope Control interface, and TIA. FEI Tecnai F20 S/TEM: EDS system operation Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 01/22/18 The user should already be familiar with operation of the instrument in

More information

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES RULES All users must go through a series of standard operation procedure training. For more information contact: Longlong Liao Teaching

More information

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Warning Always adjust your specimen height before closing the chamber door to make sure your specimen will not hit the bottom of the lens;

More information

Be aware that there is no universal notation for the various quantities.

Be aware that there is no universal notation for the various quantities. Fourier Optics v2.4 Ray tracing is limited in its ability to describe optics because it ignores the wave properties of light. Diffraction is needed to explain image spatial resolution and contrast and

More information

Stitching MetroPro Application

Stitching MetroPro Application OMP-0375F Stitching MetroPro Application Stitch.app This booklet is a quick reference; it assumes that you are familiar with MetroPro and the instrument. Information on MetroPro is provided in Getting

More information

Fast Laser Raman Microscope RAMAN

Fast Laser Raman Microscope RAMAN Fast Laser Raman Microscope RAMAN - 11 www.nanophoton.jp Fast Raman Imaging A New Generation of Raman Microscope RAMAN-11 developed by Nanophoton was created by combining confocal laser microscope technology

More information

1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height.

1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height. JEOL 2010F operating procedure Covers operation in STEM mode (See separate procedures for operation in TEM mode and operation of EDS system) Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 NOTE: this operating

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, EL6000 fluorescent light source for the microscope stand. 2. Turn on the Scanner Power

More information

Point Calibration. July 3, 2012

Point Calibration. July 3, 2012 Point Calibration July 3, 2012 The purpose of the Point Calibration process is to generate a map of voltages (for galvos) or motor positions of the pointing device to the voltages or pixels of the reference

More information

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions Starting up: Schematic of Olympus BX-61. 1. Turn on Olympus microscope power box (left of microscope) with toggle switch on

More information

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Training Guide for Leica SP8 Confocal/Multiphoton Microscope Training Guide for Leica SP8 Confocal/Multiphoton Microscope LAS AF v3.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON power switch for epifluorescence

More information

UltraGraph Optics Design

UltraGraph Optics Design UltraGraph Optics Design 5/10/99 Jim Hagerman Introduction This paper presents the current design status of the UltraGraph optics. Compromises in performance were made to reach certain product goals. Cost,

More information

Supplementary Materials

Supplementary Materials Supplementary Materials In the supplementary materials of this paper we discuss some practical consideration for alignment of optical components to help unexperienced users to achieve a high performance

More information

Observational Astronomy

Observational Astronomy Observational Astronomy Instruments The telescope- instruments combination forms a tightly coupled system: Telescope = collecting photons and forming an image Instruments = registering and analyzing the

More information

LSM 780 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch

More information

CCD User s Guide SBIG ST7E CCD camera and Macintosh ibook control computer with Meade flip mirror assembly mounted on LX200

CCD User s Guide SBIG ST7E CCD camera and Macintosh ibook control computer with Meade flip mirror assembly mounted on LX200 Massachusetts Institute of Technology Department of Earth, Atmospheric, and Planetary Sciences Handout 8 /week of 2002 March 18 12.409 Hands-On Astronomy, Spring 2002 CCD User s Guide SBIG ST7E CCD camera

More information

Guide to Operating the Bruker FT-IR Microscope I. Basic Spectrum Acquisition with Vertex 70

Guide to Operating the Bruker FT-IR Microscope I. Basic Spectrum Acquisition with Vertex 70 Guide to Operating the Bruker FT-IR Microscope I. Basic Spectrum Acquisition with Vertex 70 Susheng Tan, Ph.D. NanoScale Fabrication and Characterization Facility, University of Pittsburgh SB60 Benedum

More information

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 Leica DMI AF6000LX Table of contents START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 ACQUIRE MODULE 6 SETTING THE LIGHTPATH 6

More information

SpectraPro 2150 Monochromators and Spectrographs

SpectraPro 2150 Monochromators and Spectrographs SpectraPro 215 Monochromators and Spectrographs SpectraPro 215 15 mm imaging spectrographs and monochromators from are the industry standard for researchers who demand the highest quality data. Acton monochromators

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, Fluorescent Light for the microscope stand. 2. Turn on the Scanner Power (1) on the front

More information

Leica SP8 Resonant Confocal. Quick-Start Guide

Leica SP8 Resonant Confocal. Quick-Start Guide Leica SP8 Resonant Confocal Quick-Start Guide Contents Start-up Preparing for Imaging Part 1 On the scope Part 2 Software interface Part 3 Heat & CO2 incubation Part 4 Other hardware options Shut-down

More information

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I MSE 595T Basic Transmission Electron Microscopy TEM Imaging - I Purpose The purpose of this lab is to: 1. Make fine adjustments to the microscope alignment 2. Obtain a diffraction pattern 3. Obtain an

More information

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5. Smith College August 2010 Operating Checklist for using the Laser Scanning Confocal Microscope Leica TCS SP5. CONTENT, page no. Startup, 1 Initial set-up, 1 Software, 2 Microscope Specimen observation

More information

Fastest high definition Raman imaging. Fastest Laser Raman Microscope RAMAN

Fastest high definition Raman imaging. Fastest Laser Raman Microscope RAMAN Fastest high definition Raman imaging Fastest Laser Raman Microscope RAMAN - 11 www.nanophoton.jp Observation A New Generation in Raman Observation RAMAN-11 developed by Nanophoton was newly created by

More information

A510S Operation Manual

A510S Operation Manual A510S Operation Manual REV 1.1 1 Table of Contents 1 General Information 1-1 Description 1-2 Potential Operational Hazards 1-3 Technical Specifications 1-4 Instrument Overview 1-5 Function Summary 2 How

More information

Leica DMi8A Quick Guide

Leica DMi8A Quick Guide Leica DMi8A Quick Guide 1 Optical Microscope Quick Start Guide The following instructions are provided as a Quick Start Guide for powering up, running measurements, and shutting down Leica s DMi8A Inverted

More information

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION Revised November 15, 2017 INTRODUCTION The simplest and most commonly described examples of diffraction and interference from two-dimensional apertures

More information

Instructions for Tecnai a brief start up manual

Instructions for Tecnai a brief start up manual Instructions for Tecnai a brief start up manual Version 3.0, 8.12.2015 Manual of Tecnai 12 transmission electron microscope located at Aalto University's Nanomicroscopy Center. More information of Nanomicroscopy

More information

Olympus LEXT OLS 4000 Confocal Laser Microscope

Olympus LEXT OLS 4000 Confocal Laser Microscope Olympus LEXT OLS 4000 Confocal Laser Microscope The Olympus LEXT OLS4000 is a confocal microscope capable of taking high-resolution 3D images. The magnification (Optical and Digital) of this microscope

More information

FPPO 1000 Fiber Laser Pumped Optical Parametric Oscillator: FPPO 1000 Product Manual

FPPO 1000 Fiber Laser Pumped Optical Parametric Oscillator: FPPO 1000 Product Manual Fiber Laser Pumped Optical Parametric Oscillator: FPPO 1000 Product Manual 2012 858 West Park Street, Eugene, OR 97401 www.mtinstruments.com Table of Contents Specifications and Overview... 1 General Layout...

More information

ECEN 4606, UNDERGRADUATE OPTICS LAB

ECEN 4606, UNDERGRADUATE OPTICS LAB ECEN 4606, UNDERGRADUATE OPTICS LAB Lab 2: Imaging 1 the Telescope Original Version: Prof. McLeod SUMMARY: In this lab you will become familiar with the use of one or more lenses to create images of distant

More information

NAME SECTION PERFORMANCE TASK # 3. Part I. Qualitative Relationships

NAME SECTION PERFORMANCE TASK # 3. Part I. Qualitative Relationships NAME SECTION PARTNERS DATE PERFORMANCE TASK # 3 You must work in teams of three or four (ask instructor) and will turn in ONE report. Answer all questions. Write in complete sentences. You must hand this

More information

1. Specimen Holder Removal, Loading, and Insertion

1. Specimen Holder Removal, Loading, and Insertion OPERATION OF THE PHILIPS CM-200 FEG-TEM When not in use, the CM-200 should be in the MICROSCOPE ON configuration with the HIGH TENSION ON (illuminates green when the high tension is on).. The microscope

More information

X-ray generation by femtosecond laser pulses and its application to soft X-ray imaging microscope

X-ray generation by femtosecond laser pulses and its application to soft X-ray imaging microscope X-ray generation by femtosecond laser pulses and its application to soft X-ray imaging microscope Kenichi Ikeda 1, Hideyuki Kotaki 1 ' 2 and Kazuhisa Nakajima 1 ' 2 ' 3 1 Graduate University for Advanced

More information

STEM Spectrum Imaging Tutorial

STEM Spectrum Imaging Tutorial STEM Spectrum Imaging Tutorial Gatan, Inc. 5933 Coronado Lane, Pleasanton, CA 94588 Tel: (925) 463-0200 Fax: (925) 463-0204 April 2001 Contents 1 Introduction 1.1 What is Spectrum Imaging? 2 Hardware 3

More information

Phys 531 Lecture 9 30 September 2004 Ray Optics II. + 1 s i. = 1 f

Phys 531 Lecture 9 30 September 2004 Ray Optics II. + 1 s i. = 1 f Phys 531 Lecture 9 30 September 2004 Ray Optics II Last time, developed idea of ray optics approximation to wave theory Introduced paraxial approximation: rays with θ 1 Will continue to use Started disussing

More information

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire Nikon Ti-E Microscope Manual Rightmire Hall Ohio State University Director: Tony Brown Rightmire 060 292-1205 brown.2302@osu.edu Facility Manager: Paula Monsma Rightmire 062 293-0939 292-1367 monsma.1@osu.edu

More information

Instruction manual for T3DS software. Tool for THz Time-Domain Spectroscopy. Release 4.0

Instruction manual for T3DS software. Tool for THz Time-Domain Spectroscopy. Release 4.0 Instruction manual for T3DS software Release 4.0 Table of contents 0. Setup... 3 1. Start-up... 5 2. Input parameters and delay line control... 6 3. Slow scan measurement... 8 4. Fast scan measurement...

More information

JEOL 6500 User Manual

JEOL 6500 User Manual LOG IN to your session on the computer to the left of the microscope. Starting Conditions 1. Press Ctrl-Alt-Del and log on to the microscope computer. Click on JEOL PC SEM 6500 icon. Click yes if message

More information