1 OMP-0375F Stitching MetroPro Application Stitch.app This booklet is a quick reference; it assumes that you are familiar with MetroPro and the instrument. Information on MetroPro is provided in Getting Started With MetroPro, OMP-0398, and the MetroPro Reference Guide, OMP For information on the instrument, please refer to the applicable Zygo manual. Information herein is applicable as of MetroPro version The Stitching Application The MetroPro Stitching Application is designed to measure and analyze surface areas much larger than possible with a single measurement. Stitching makes several measurements of the test part as it is moved by a motorized stage and then combines or stitches the multiple data sets into one. Effectively, it increases the field of view without compromising lateral or vertical resolution. The graphics and results displayed on screen are based on the entire measured area. After you define the test area, a stage pattern is created that breaks the larger area into a number of smaller sub-regions. Each sub-region is equal to the field of view of the microscope s objective and zoom setting, minus a small percentage for overlap. After software parameters are entered the process is automatic. A number of factors determine the maximum size of the stitched area. These include: the objective magnification, the zoom setting (microscopes with Image Zoom option only), the camera resolution (determined by the Camera Mode control), and the amount of computer RAM. The Stitching Application supports Zygo s microscopes, including the NewView and Maxim, as long as they are equipped with programmable motorized stages. A motorized z-drive (focus) is optional; however, it is required for extended scans or the auto focus option. The stitching application is licensed as part of standard MetroPro. 1

2 Stitch.app To open the application, click on the Stitch.app icon. If the icon is not on the MetroPro base window, then use the Load Application command from the MetroPro Window menu to first load the application icon. Guidelines Use a System Error File - It is recommended that you create a system reference file for the objective in use and then subtract this file from the measurement. Lateral Calibration - The microscope must be calibrated for accurate results; inaccurate pixel size causes misplacement of each measurement with respect to the surrounding measurements. Objectives supplied with an original system are factory calibrated; calibration is required when other objectives are used or when using the free zoom setting. For optimum performance, calibrate the objective and zoom setting before making a stitch sequence. For calibration procedures refer to the MetroPro Reference Guide. Align the X/Y Stage Before Making Measurements - The stage x and y axes must be precisely aligned so each measurement is properly stitched to the surrounding measurements. This is covered in the Stage Alignment Procedure. 2

3 Guidelines (continued) Computer Recommendations - The speed and capability of the Stitching software is linked to computer resources. A fast computer processor, large hard drive, and a large amount of RAM improve performance. For best performance, Zygo recommends to maximize the amount of RAM. Zygo also recommends to set the Windows virtual memory paging file size to the maximum. To do this, point to the Windows Start button and select Settings Control Panel; open the System icon and select the Performance tab; click the Change button under Virtual Memory; set the paging file Maximum Size (MB) as large as possible. MetroPro Control Settings - The speed of the measurement also relates to the Camera Mode and the Scan Length control settings; the larger the Camera Mode and Scan Length settings, the longer each measurement cycle. Note that a very large stitching measurement can take hours to acquire and process data. Stage Alignment Procedure Typical Microscope Controls 1. Power-up the instrument, start MetroPro, and open the application. Click the Objective button so it displays the objective in use and set the Image Zoom control to match the zoom setting of the instrument. Set the Camera Mode control to 640x Click the Home Stage and Home Z Axis buttons to re-establish the motorized stage home positions. 3. Place the test part on the stage under the objective. Position the objective at its working distance from the part. Don t crash the objective into the part or the stage. 4. Put the filter tray to FOCUS. Press F5 to adjust the light level. Focus the microscope for a sharp image; there should be visible fringes. 5. If there are not visible fringes, adjust for tilt in the part. To do this, press the F4 key to open the Light Level window, adjust the stage tip and tilt until the Peak Intensity readout is maximized, and press Enter to close the window. 6. Adjust the stage to center the area of interest in view. Place the filter tray to MEAS, press F5 again to adjust the light level. Fine-tune focus and stage tilt for surface detail, to null or minimize the number of fringes, and to center the darkest fringe. 5. In the Stitch Controls window, set the Type control to Align Stage, then click the Stitch button. The software takes two measurements and then displays the composite image. Examine the area along the intersection of the two measurements. If pixels are offset and there is a noticeable shift greater than ±1 pixel, align the stage as described in step Slightly loosen the three capscrews securing the x-y stage to the theta plate. Slightly rotate the x-y stage clockwise, if the left side of the pixel intersection is higher than the right; rotate it counter-clockwise if the right side of the intersection is higher than the left. 8. Repeat steps 6 and 7 until the intersection of the two images is within one pixel. After proper alignment is achieved, tighten the three capscrews securing the x-y stage. 3

4 Making Measurements 1. Power-up the instrument, start MetroPro, and open the application by clicking on the Stitch.app icon. 2. Make sure the stage is properly aligned. See the previous procedure. 3. Click the Objective button until it displays the objective in use. Place the test part on the stage under the objective. Position the objective at its working distance from the part. Don t crash the objective into the part or the stage 4. Set the Zoom knob to the desired setting; ensure that the software Image Zoom control is set to match. Set the Camera Mode, and Scan Length (NewView) controls within the Measure Cntrl icon to the desired settings. 5. Press the F5 key to adjust the light level. Adjust the microscope for focus and adjust the stage tilt to null or minimize the number of fringes. 6. Make a trial measurement of the test part by clicking the Measure button. The goal is to acquire the maximum number of data points. If necessary to control data loss, adjust measure controls (usually Min Mod (%) and Scan Length) and analyze controls (filtering). 7. Measure the test part in other locations to ensure that data can be obtained at all locations during the stitch measurement. Make software control adjustments if needed. Objective Working Distance 2.5X 5X 10X 20X 50X 10.3 mm 9.3 mm 7.4 mm 4.7 mm 3.4 mm Microscope Controls - NewView 5000 Stitch Control Window (partial) If the microscope has the scan auto focus and tilt option, set the Focus control to On. 8. Select the desired stitching method with the Type control. See the next page for descriptions of the stitching methods. Enter values in the stitching controls if necessary. To start the stitch sequence, click the Stitch button and follow the on-line instructions. 4

5 Stitching Methods Terms used to describe stitching are shown here. There are many ways to define the stitching area. These are summarized below. They are selected with the Type control. Stitching Pattern Terminology Before running a stitch measurement it is recommended to click the Home Stage and Home Z Axis buttons to establish the motorized stage home positions. Method Manual Position Stitch Column & Row Stitch X & Y Size Stitch Start & End Position Stitch Annulus Stitch Run Sequence File Description The user teaches the software the location of all measurement positions. Useful for measuring odd shaped parts, like an S-curve. It is the user s responsibility to ensure that the overlap is sufficient between measurement positions. The pattern starts at the first saved location. The stitched area is determined by the number of rows and columns. Requires entries in N Cols, N Rows, and Overlap (%) controls. The pattern always starts in the upper right. The stitched area is determined by entered dimensions. Useful when the size of the measurement area is known. The number of rows and columns to cover the area is automatically calculated. Requires entries in Size X, Size Y, and Overlap (%) controls. The pattern always starts in the upper right. The user teaches the software the starting position (upper right) and the ending position (lower left) of the stitched area. The number of measurements needed is automatically calculated. Requires an entry in the Overlap (%) control. The pattern always starts in the upper right. The stitched area is circular or ring shaped. The measurement locations are automatically determined. Requires entries in the Inner Radius, Outer Radius, and Overlap (%) controls. The pattern starts at the 12 o clock position on the outer radius. The Sequence file as specified in the Seq File control is used for measurement locations. The pattern starts at the first saved location. Zip Stitch Feature The Zip Stitch function allows you to make two groups of measurements separated by a vertical distance. Note that the lower and upper surfaces are not joined together in the analysis. Zip Stitch is ideal for measuring larger areas at the top and bottom of a vertical step. To make a measurement using Zip Stitch, enter values in the Zip Stitch controls, set the microscope focus and stage position on the lower surface area and press the Zip Stitch button. Zip stitching requires the user to enter x and y dimensions of the lower and upper surface areas. Zip Stitch controls are explained under Software Controls. 5

6 Software Controls Some of the controls of interest for the Stitching Application are listed below. For controls not listed, refer to the MetroPro Reference Guide. Control Function Controls in Measure Cntrl Acquisition Mode Camera Mode Extended Scan Length FDA Res Image Zoom Min Area Size Min Mod (%) Scan Length Subtract Sys Err Sys Err File Selects the technique used to acquire data. Settings are Scan or Phase. Scan is for the NewView. Phase is for the Maxim. Selects the effective camera size or resolution in pixels and the speed of frame captures. Lower settings mean less pixels in each measurement, so larger areas can be stitched. More pixels resolve smaller details, but increase processing time. For NewView with extended scan option only. Specifies the length of the scan. The Scan Length control must be set to Extended. The value entered should be the vertical range of detail in the part plus 10%. NewView. Selects the resolution used during analysis. Settings are Normal or High. Use Normal when measuring rough surfaces; use High when measuring smooth surfaces and steps. For microscopes with image zoom option only. Set the control to match the setting selected on the zoom thumbwheel on the microscope. This ensures that lateral type results are calibrated. Specifies the number of data points in a valid region. Decrease to allow smaller areas. Increase to accept larger areas and reject smaller. Specifies the minimum modulation necessary for a valid data point. Decrease to accept areas with poor fringe contrast or low reflectivity. Increase to exclude unwanted data points. NewView. Selects the vertical scan length. The longer the scan, the greater the height measuring capability, and the longer the time required to make a measurement. Activates a system error correction function to improve measurement accuracy. An error file must exist for the specific objective in use. Specifies the name of a data file to subtract from measurements when the Subtract Sys Err control is On. 6

7 Software Controls (continued) Control Function Controls in Analyze Cntrl Data Fill Filter... Remove Remove Spikes Spike Height (xrms) When On, holes or missing data points are filled, based on the setting of the Fill Max control. When Off (default), missing data points are not filled. The Fill Max control specifies the maximum size of a data hole (in pixels) to fill. For detail on filtering options and controls refer to the MetroPro Reference Guide. Specifies the surface to subtract from measurements to minimize form. Settings are Plane, Piston, Sphere, and Cylinder. When On, spikes are removed from the data based on the setting of the Spike Height control. When Off, data spikes are not removed. For each point, the rms height of the surrounding points is calculated; if the point height is greater than the entered value times the rms it is removed. An entry is ignored unless the Remove Spikes control is On. Trim Controls in Stitch Cntrl Auto Focus Auto Load Masks Auto Save Seq Auto Tilt Average Overlap Regions Diagnostics Specifies the number of pixel layers to remove from edges and from isolated obscurations. The Trim Mode control selects where trimming is applied. Selects how to use auto focus in the stitch sequence. Settings are Off, First Measurement Only, All Measurements. Off disables auto focus. First Measurement Only uses auto focus only on the first measurement of the stitching sequence. All Measurements enables auto focus for all measurements of the stitch sequence. Specifies whether to load a mask after loading the completed stitched image. The mask filename is defined by the Masks File control. Activates the automatic saving of a sequence file. When On, the sequence file named by the Seq File control is saved to file. Selects how to use auto tilt in the stitch sequence. Settings are Off, First Measurement Only. First Measurement Only uses auto tilt only on the first measurement of the stitching sequence. Selects if the stitching algorithm averages overlap regions. Settings are On or Off. It is recommended that this control be On. Averaging the overlap regions results in a stitched image with less discontinuity. Selects if diagnostic information from the stitch algorithm is provided. Settings are On or Off. 7

8 Software Controls (continued) Control Function Controls in Stitch Cntrl (continued) Display Measurements Failure Action Inner Radius Log Reports Masks File Measurement Status N Cols Selects whether individual measurements are displayed on screen during the stitch sequence. Settings are On or Off. Selects the action to take if there is a measurement error during a stitch sequence. Settings are Abort, Retry, or Continue. Abort will stop the stitching sequence. Retry repeats the measurement and aborts the sequence if it fails again. Continue ignores the error and goes to the next location. Specifies the radius of an internal circular area in which there is no data. Applicable when the Type control is set to Annulus. If zero, the entire area as specified by the Outer Radius control is used. When On, the final results of the loaded stitched image are logged or added to the report file. Specifies the name of the mask file to apply to the completed stitched image when Auto Load Masks is On. Displays the measurement number and total number of measurements during a stitch sequence. Attribute only. Specifies the number of vertical columns in the stitch stage pattern. Requires a programmable stage. N Rows Outer Radius Overlap (%) Scanning Seq Seq File Specifies the number of horizontal rows in the stitch stage pattern. Requires a programmable stage. Specifies the radius of a circular area in which there is data. Applicable when the Type control is set to Annulus. Specifies the overlap of the columns and rows in the stitched image. 15 to 20% is recommended, 10% is the minimum. When the Stitch Type is Start & End Pos or the X & Y Dimension, it is the minimum overlap percent. Selects whether to traverse the part rows in a serpentine or raster fashion. Settings are Serp or Rast. It is recommended that Rast be used for more accurate stage movement. Name of the file that contains all relative locations of a stitching sequence. When a stitch sequence is run and the Auto Save Seq control is set to On, the relative locations are saved into the file specified by this control. When the Type control is set to Run Sequence File, the stitching sequence defined by the file specified, is run. This way multiple stitching sequences may be defined, saved, and loaded. 8

9 Software Controls (continued) Control Function Controls in Stitch Cntrl (continued) Size X Size Y Stitch button Stitch File Stitch Processed Data Store Process Stats Type Specifies the size of the overall stitched image in the x-axis. Requires a programmable stage. Specifies the size of the overall stitched image in the y-axis. Requires a programmable stage. Invokes a stitch measurement sequence based on the settings of the stitch controls. Requires a programmable stage. Specifies the name of the final stitched data set. Selects whether to pass raw data or processed data to the stitch algorithm. Settings are On or Off. It is recommended that this control be Off. The stitching of processed data is only recommended in rare cases. Selects whether to store the stitched image into Process Stats. Settings are On or Off. If On, the final overall stitched measurement results are stored. Selects the type of stitch measurement. Stitching makes several measurements of the test part as it is moved by a motorized stage and then combines the multiple data sets into one. Manual Position. The user teaches the software the location of all measurement positions. Useful for measuring odd shaped parts, like an S- curve. It is advisable to have overlap between measurement positions. Column & Row. The stitched area is determined by the number of rows and columns. Requires entries in N Cols, N Rows, and Overlap (%) controls. The pattern always starts in the upper right. X & Y Size. The stitched area is determined by entered dimensions. Useful when the size of the measurement area is known. The number of rows and columns to cover the area is automatically calculated. Requires entries in Size X, Size Y, and Overlap (%) controls. The pattern always starts in the upper right. Start & End Position. The user teaches the software the starting position (upper right) and the ending position (lower left) of the stitched area. The number of measurements needed is automatically calculated. Requires an entry in the Overlap (%) control. Annulus. The stitched area is circular or ring shaped. The measurement locations are automatically determined. Requires entries in the Inner radius, Outer Radius, and Overlap (%) controls. Run Sequence File. Runs the stitching sequence defined by the Seq File control at the current position. Align Stage. Used when checking stage alignment; two measurements of adjacent areas are taken and stitched together. 9

10 Software Controls (continued) Control Controls in Zip Cntrl Auto Load Masks Display Stitched Surfaces Lower Light Level Offset Lower Overlap (%) Lower Seq File Lower Size X Lower Size Y Masks File Store Process Stats Upper Light Level Offset Upper Offset X Upper Offset Y Upper Overlap (%) Upper Seq File Function Specifies whether to load a mask after loading the completed stitched image. The mask filename is defined by the Masks File control. Specifies whether individual stitched surfaces are displayed. Settings are On, Off. Specifies the percentage to add to the current measurement light level for the lower surface in a zip stitch measurement. Specifies the overlap of the measured sections in the lower measurement region in a zip stitch measurement. 15 to 20% is recommended, 10% is the minimum. Requires a motorized z-axis and a programmable stage. Specifies the filename of the stitching sequence file to use for the lower surface when the Use Lower Seq File control is On. Specifies the size of the lower stitched image in the x-axis, in a zip stitch measurement. Requires a motorized z-axis and a programmable stage. Specifies the size of the lower stitched image in the y-axis, in a zip stitch measurement. Requires a motorized z-axis and a programmable stage. Specifies the name of the mask file to apply to the completed stitched image when Auto Load Masks is On. Selects whether to store the zipped image into Process Stats. Settings are On or Off. If On, the final overall measurement results are stored. Specifies the percentage to add to the current measurement light level for the upper surface in a zip stitch measurement. Specifies the dimension in the x-axis that the upper right corner of the lower stitched area is offset from the upper right corner of the upper stitched area. One of many controls used when making a zip stitch measurement. Specifies the dimension in the y-axis that the upper right corner of the lower stitched area is offset from the upper right corner of the upper stitched area. One of many controls used when making a zip stitch measurement. Specifies the overlap of the measured sections in the upper measurement region, in a zip stitch measurement. 15 to 20% is recommended, 10% is the minimum. Requires a motorized z-axis and a programmable stage. Specifies the filename of the stitching sequence file to use for the lower surface when the Use Upper Seq File control is On. 10

11 Software Controls (continued) Control Function Controls in Zip Cntrl (continued) Upper Size X Upper Size Y Use Lower Seq File Use Upper Seq File Zip Distance Zip File Zip Scan button Zip Stitch button Specifies the size of the upper stitched image in the x-axis, in a zip stitch measurement. Requires a motorized z-axis and a programmable stage. Specifies the size of the upper stitched image in the y-axis, in a zip stitch measurement. Requires a motorized z-axis and a programmable stage. Specifies whether the locations to measure for the lower surface, are specified from a stitching sequence file, or from the Lower X and Y Sizes. Settings are On or Off. Specifies whether the locations to measure for the upper surface, are specified from a stitching sequence file, or from the Lower X and Y Sizes. Settings are On or Off. Specifies the vertical distance between the upper and lower surface areas or measure-ment regions in a zip stitch measure-ment. A zip stitch measurement consists of two groups of measure-ments separated by a vertical distance. When the Phase Res control is set to High, the vertical range is limited to 20 millimeters. Zip Stitch requires a motorized z-axis and a programmable stage. Specifies the name of the final zip stitch data set. Invokes a zip scan measurement sequence. Zip scan is two measurements taken in the same location, without stage movement in the x and y axes, but with vertical movement in the z-axis. The vertical dimension is specified by the Zip Distance control. Invokes a zip stitch measurement sequence based on settings of the zip stitch controls. 11

12 Results The results shown in the main window are explained below. For results not listed, or for greater detail, refer to the MetroPro Reference Guide. Result PV Description The height between the lowest and the highest point on the test part surface. rms Ra Size X Size Y The root-mean-square deviation of all points from a plane fit to the test part surface. The average roughness, or the average deviation, of all points from a plane fit to the test part surface. The dimension of the data set in the X-axis. The dimension of the data set in the Y-axis. Other Operations Operation Save data Print results Save changes made to controls, plots, results, and windows Define a mask Turn off the system How to Click the Save Data button. In the File Handler, click the Current Selection box, enter a name for the file, ending with.dat, and press Enter, then click Done. Click the word zygo in the window you want to print or select the Print command from the window s menu. In the Print Panel, click the Print button. You must save the application under a new name. Select the Save Application command from the Application Window menu. In the File Handler, click the Current Selection box, enter a name for the file, ending with.app and press Enter, then click Done. Click the Mask Data button. Use the Mask Editor to define areas to include or exclude from the test part measurement. See Mask Editor in the MetroPro Reference Guide. Warning! Improper shutdown may damage the instrument. Select the Quit command from the MetroPro menu; turn off power after you have shut down from within Windows. 12

13 Troubleshooting Symptom Application not functional. No data error or graphic displays appear incomplete or with holes. Poor repeatability of results. Plots appear very flat and in one color. Error Message- Error in Stitch Algorithm Possible Cause and Fix MetroPro is not licensed; contact Zygo for licensing information. The hardware key or ID module is not installed. Install the hardware key and restart the computer. The instrument is not properly focused or the fringes are not nulled. Focus the instrument and null the fringes. The Min Mod (%) control is set too high; decrease the value. NewView scan length may be too short. Increase length with Scan length control. Improper light level. Press F4 to open the Light Level window and manually adjust the light level to maximize intensity without saturation. Improper light level. Press F4 to open the Light Level window and manually adjust the light level to maximize intensity without saturation. Parts moving during measurement. Check the support stage and fixturing. Noisy environment. Look at the video monitor for fringe drift or vibration; these signs indicate a noisy environment. Check for air drafts; ensure that the computer is not contributing noise to the vibration isolation system. Improper light level. Press F4 to open the Light Level window and manually adjust the light level to maximize intensity without saturation. Spikes in data. Things to try: raise the Min Mod (%) control to try to remove poorly modulating points; turn on the Remove Spikes control; use fixed scaling in the plots to clip spikes; or use filtering controls. The software has detected that the single measurements do not match up correctly for stitching. Common causes for this error message are: 1) Incorrect zoom or objective setting. This makes the stage move the wrong distance between stitches. 2) Stage not aligned. Select Align Stitch with the Type control and ensure that the stage is aligned. There should be smooth transitions in the overlap area. 3) Lateral calibration not done. Perform calibration. 4) Improper setting for Min Mod (%) makes single measurements on the same area not repeatable. 5) Turret or objective not mounted properly. Remove and reinstall the objective or turret. 6) Manual stitch used and measurements do not overlap. Ensure that the measurements overlap. 7) The part is moving during measurements. Check that the part is properly secured, not moving or flexing. 13

14 Zygo Corporation Phone: (860) OMP-0375F 4/2002 Laurel Brook Road Customer Support: (800) Copyright 2002 Zygo Corporation P.O. Box Zygo and the Zygo logo are Middlefield, CT Web Site: registered trademarks of U.S.A. Zygo Corporation.

Basic Operator Procedure Training NewView 6300 GROWING IN A SHRINKING WORLD

Basic Operator Procedure Training NewView 6300 GROWING IN A SHRINKING WORLD Basic Operator Procedure Training NewView 6300 GROWING IN A SHRINKING WORLD NewView 6200/6300 System Overview Objective Motion Controller / Joystick Live Image Monitor Motorized Stage Application Monitor

More information

Zygo Optical Interferometer SOP

Zygo Optical Interferometer SOP Zygo Optical Interferometer SOP Open MetroPro Click on Mirco7k.app to open measurement application Films7k application Micro7k application Click on Live display window in the task bar. Make sure the objective

More information

Bruker Optical Profilometer SOP Revision 2 01/04/16 Page 1 of 13. Bruker Optical Profilometer SOP

Bruker Optical Profilometer SOP Revision 2 01/04/16 Page 1 of 13. Bruker Optical Profilometer SOP Page 1 of 13 Bruker Optical Profilometer SOP The Contour GT-I, is a versatile bench-top optical surface-profiling system that can measure a wide variety of surfaces and samples. Contour GT optical profilers

More information

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and

More information

Confocal Application Notes Vol. 5 July 2010

Confocal Application Notes Vol. 5 July 2010 Tile Scan Prepared by Myriam Gastard, PhD Application and Technical Support Group, Leica Microsystems, Inc. In this issue of our Confocal Application Notes, proper set up of the Tile function enables you

More information

R I T. Title: Wyko RST Plus. Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 05/23/06 1 SCOPE 2 REFERENCE DOCUMENTS

R I T. Title: Wyko RST Plus. Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 05/23/06 1 SCOPE 2 REFERENCE DOCUMENTS Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Wyko RST Plus. All users are expected to have read and understood this document.

More information

KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017)

KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017) KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017) 1 Introduction You must be trained to operate the Laser-scanning confocal microscope (LSCM) independently.

More information

Alignment of the camera

Alignment of the camera Related topics Detector Alignment, Rotation axis, tilt, Principle Alignment of the detector and the rotation stage is very important to get optimal quality images of a CT scan. In this experiment, the

More information

Operating Rausch ScanCam within POSM.

Operating Rausch ScanCam within POSM. Operating Rausch ScanCam within POSM. POSM (Pipeline Observation System Management) // posmsoftware.com // info@posmsoftware.com // 859-274-0041 RAUSCH USA // www.rauschusa.com // reusa@rauschusa.com //

More information

IncuCyte ZOOM Fluorescent Processing Overview

IncuCyte ZOOM Fluorescent Processing Overview IncuCyte ZOOM Fluorescent Processing Overview The IncuCyte ZOOM offers users the ability to acquire HD phase as well as dual wavelength fluorescent images of living cells producing multiplexed data that

More information

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: )

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: ) University of Minnesota College of Science and Engineering Characterization Facility PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: 2012.10.17) The following instructions

More information

Large Field of View, High Spatial Resolution, Surface Measurements

Large Field of View, High Spatial Resolution, Surface Measurements Large Field of View, High Spatial Resolution, Surface Measurements James C. Wyant and Joanna Schmit WYKO Corporation, 2650 E. Elvira Road Tucson, Arizona 85706, USA jcwyant@wyko.com and jschmit@wyko.com

More information

Advanced Technology and Manufacturing Institute. Zygo ZeScope

Advanced Technology and Manufacturing Institute. Zygo ZeScope Advanced Technology and Manufacturing Institute Zygo ZeScope Created by Andrew Miller ATAMI Oregon State University Revision Date Description Curator 0 8/31/2018 New Document Andrew Miller Zygo ZeScope

More information

Before you start, make sure that you have a properly calibrated system to obtain high-quality images.

Before you start, make sure that you have a properly calibrated system to obtain high-quality images. CONTENT Step 1: Optimizing your Workspace for Acquisition... 1 Step 2: Tracing the Region of Interest... 2 Step 3: Camera (& Multichannel) Settings... 3 Step 4: Acquiring a Background Image (Brightfield)...

More information

Brightness and Contrast Control Reference Guide

Brightness and Contrast Control Reference Guide innovation Series Scanners Brightness and Contrast Control Reference Guide A-61506 Part No. 9E3722 CAT No. 137 0337 Using the Brightness and Contrast Control This Reference Guide provides information and

More information

Adobe Photoshop. Levels

Adobe Photoshop. Levels How to correct color Once you ve opened an image in Photoshop, you may want to adjust color quality or light levels, convert it to black and white, or correct color or lens distortions. This can improve

More information

Instruction Manual for HyperScan Spectrometer

Instruction Manual for HyperScan Spectrometer August 2006 Version 1.1 Table of Contents Section Page 1 Hardware... 1 2 Mounting Procedure... 2 3 CCD Alignment... 6 4 Software... 7 5 Wiring Diagram... 19 1 HARDWARE While it is not necessary to have

More information

Picture Style Editor Ver Instruction Manual

Picture Style Editor Ver Instruction Manual ENGLISH Picture Style File Creating Software Picture Style Editor Ver. 1.18 Instruction Manual Content of this Instruction Manual PSE stands for Picture Style Editor. In this manual, the windows used in

More information

Picture Style Editor Ver Instruction Manual

Picture Style Editor Ver Instruction Manual ENGLISH Picture Style File Creating Software Picture Style Editor Ver. 1.15 Instruction Manual Content of this Instruction Manual PSE stands for Picture Style Editor. indicates the selection procedure

More information

ThermaViz. Operating Manual. The Innovative Two-Wavelength Imaging Pyrometer

ThermaViz. Operating Manual. The Innovative Two-Wavelength Imaging Pyrometer ThermaViz The Innovative Two-Wavelength Imaging Pyrometer Operating Manual The integration of advanced optical diagnostics and intelligent materials processing for temperature measurement and process control.

More information

User Manual Veterinary

User Manual Veterinary Veterinary Acquisition and diagnostic software Doc No.: Rev 1.0.1 Aug 2013 Part No.: CR-FPM-04-022-EN-S 3DISC, FireCR, Quantor and the 3D Cube are trademarks of 3D Imaging & Simulations Corp, South Korea,

More information

Creo Parametric 2.0: Introduction to Solid Modeling. Creo Parametric 2.0: Introduction to Solid Modeling

Creo Parametric 2.0: Introduction to Solid Modeling. Creo Parametric 2.0: Introduction to Solid Modeling Creo Parametric 2.0: Introduction to Solid Modeling 1 2 Part 1 Class Files... xiii Chapter 1 Introduction to Creo Parametric... 1-1 1.1 Solid Modeling... 1-4 1.2 Creo Parametric Fundamentals... 1-6 Feature-Based...

More information

Picture Style Editor Ver Instruction Manual

Picture Style Editor Ver Instruction Manual ENGLISH Picture Style File Creating Software Picture Style Editor Ver. 1.12 Instruction Manual Content of this Instruction Manual PSE is used for Picture Style Editor. In this manual, the windows used

More information

We recommend downloading the latest core installer for our software from our website. This can be found at:

We recommend downloading the latest core installer for our software from our website. This can be found at: Dusk Getting Started Installing the Software We recommend downloading the latest core installer for our software from our website. This can be found at: https://www.atik-cameras.com/downloads/ Locate and

More information

NIS-Elements: Grid to ND Set Up Interface

NIS-Elements: Grid to ND Set Up Interface NIS-Elements: Grid to ND Set Up Interface This document specifies the set up details of the Grid to ND macro, which is included in material # 97157 High Content Acq. Tools. This documentation assumes some

More information

inphoto ID SLR Automatic ID photography With Canon SLR camera User Guide

inphoto ID SLR Automatic ID photography With Canon SLR camera User Guide inphoto ID SLR Automatic ID photography With Canon SLR camera User Guide 2014 Akond company Phone/fax: +7(812)384-6430 Cell: +7(921)757-8319 e-mail: info@akond.net akondsales@gmail.com http://www.akond.net

More information

Ladybird Project - Vacuum Mould

Ladybird Project - Vacuum Mould - Vacuum Mould Prerequisite Mould drawn and saved as an STL file in SolidWorks Focus of the Lesson On completion of this exercise you will have: Opened an STL file Set Machining Constraints Set up Tools

More information

Photoshop CC Editing Images

Photoshop CC Editing Images Photoshop CC Editing Images Rotate a Canvas A canvas can be rotated 90 degrees Clockwise, 90 degrees Counter Clockwise, or rotated 180 degrees. Navigate to the Image Menu, select Image Rotation and then

More information

Image Processing Tutorial Basic Concepts

Image Processing Tutorial Basic Concepts Image Processing Tutorial Basic Concepts CCDWare Publishing http://www.ccdware.com 2005 CCDWare Publishing Table of Contents Introduction... 3 Starting CCDStack... 4 Creating Calibration Frames... 5 Create

More information

LSM 780 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch

More information

Renishaw InVia Raman microscope

Renishaw InVia Raman microscope Laser Spectroscopy Labs Renishaw InVia Raman microscope Operation instructions 1. Turn On the power switch, system power switch is located towards the back of the system on the right hand side. Wait ~10

More information

ScanGear CS-U 5.3 for CanoScan FB630U/FB636U Color Image Scanner User s Guide

ScanGear CS-U 5.3 for CanoScan FB630U/FB636U Color Image Scanner User s Guide ScanGear CS-U 5.3 for CanoScan FB630U/FB636U Color Image Scanner User s Guide Copyright Notice 1999 Canon Inc. This manual is copyrighted with all rights reserved. Under the copyright laws, this manual

More information

Auto Color Diagnosis Guide

Auto Color Diagnosis Guide Auto Color Diagnosis Guide Using the Auto Color Diagnosis Function For safe and correct use, be sure to read Safety Information before using the machine. TABLE OF CONTENTS 1. Using the Auto Color Diagnosis

More information

TOSHIBA MACHINE CO., LTD.

TOSHIBA MACHINE CO., LTD. User s Manual Product SHAN5 Version 1.12 (V Series Servo Amplifier PC Tool) Model SFV02 July2005 TOSHIBA MACHINE CO., LTD. Introduction This document describes the operation and installation methods of

More information

Typical Interferometer Setups

Typical Interferometer Setups ZYGO s Guide to Typical Interferometer Setups Surfaces Windows Lens Systems Distribution in the UK & Ireland www.lambdaphoto.co.uk Contents Surface Flatness 1 Plano Transmitted Wavefront 1 Parallelism

More information

Using Siemens NX 11 Software. The connecting rod

Using Siemens NX 11 Software. The connecting rod Using Siemens NX 11 Software The connecting rod Based on a Catia tutorial written by Loïc Stefanski. At the end of this manual, you should obtain the following part: 1 Introduction. Start NX 11 and open

More information

Chlorophyll Fluorescence Imaging System

Chlorophyll Fluorescence Imaging System Quick Start Guide Chlorophyll Fluorescence Imaging System Quick Start Guide for Technologica FluorImager software for use with Technlogica CFImager hardware Copyright 2006 2015 TECHNOLOGICA LIMITED. All

More information

4.5.1 Mirroring Gain/Offset Registers GPIO CMV Snapshot Control... 14

4.5.1 Mirroring Gain/Offset Registers GPIO CMV Snapshot Control... 14 Thank you for choosing the MityCAM-C8000 from Critical Link. The MityCAM-C8000 MityViewer Quick Start Guide will guide you through the software installation process and the steps to acquire your first

More information

Rapid Array Scanning with the MS2000 Stage

Rapid Array Scanning with the MS2000 Stage Technical Note 124 August 2010 Applied Scientific Instrumentation 29391 W. Enid Rd. Eugene, OR 97402 Rapid Array Scanning with the MS2000 Stage Introduction A common problem for automated microscopy is

More information

APPLICATION NOTE. Understanding the PV Specification. Introduction. Problems with PV

APPLICATION NOTE. Understanding the PV Specification. Introduction. Problems with PV APPLICATION NOTE Understanding the PV Specification Introduction An array of non-standard, arbitrary practices are frequently used in the optics industry to demonstrate conformance of a part to the traditional

More information

Touch Probe Cycles itnc 530

Touch Probe Cycles itnc 530 Touch Probe Cycles itnc 530 NC Software 340 420-xx 340 421-xx User s Manual English (en) 4/2002 TNC Models, Software and Features This manual describes functions and features provided by the TNCs as of

More information

STEM Spectrum Imaging Tutorial

STEM Spectrum Imaging Tutorial STEM Spectrum Imaging Tutorial Gatan, Inc. 5933 Coronado Lane, Pleasanton, CA 94588 Tel: (925) 463-0200 Fax: (925) 463-0204 April 2001 Contents 1 Introduction 1.1 What is Spectrum Imaging? 2 Hardware 3

More information

AgilEye Manual Version 2.0 February 28, 2007

AgilEye Manual Version 2.0 February 28, 2007 AgilEye Manual Version 2.0 February 28, 2007 1717 Louisiana NE Suite 202 Albuquerque, NM 87110 (505) 268-4742 support@agiloptics.com 2 (505) 268-4742 v. 2.0 February 07, 2007 3 Introduction AgilEye Wavefront

More information

Principles and Applications of Microfluidic Devices AutoCAD Design Lab - COMSOL import ready

Principles and Applications of Microfluidic Devices AutoCAD Design Lab - COMSOL import ready Principles and Applications of Microfluidic Devices AutoCAD Design Lab - COMSOL import ready Part I. Introduction AutoCAD is a computer drawing package that can allow you to define physical structures

More information

RENISHAW INVIA RAMAN SPECTROMETER

RENISHAW INVIA RAMAN SPECTROMETER STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw invia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples

More information

WITec Alpha 300R Quick Operation Summary October 2018

WITec Alpha 300R Quick Operation Summary October 2018 WITec Alpha 300R Quick Operation Summary October 2018 This document is frequently updated if you feel information should be added, please indicate that to the facility manager (currently Philip Carubia,

More information

Touch Probe Cycles TNC 426 TNC 430

Touch Probe Cycles TNC 426 TNC 430 Touch Probe Cycles TNC 426 TNC 430 NC Software 280 472-xx 280 473-xx 280 474-xx 280 475-xx 280 476-xx 280 477-xx User s Manual English (en) 6/2003 TNC Model, Software and Features This manual describes

More information

JEOL 6500 User Manual

JEOL 6500 User Manual LOG IN to your session on the computer to the left of the microscope. Starting Conditions 1. Press Ctrl-Alt-Del and log on to the microscope computer. Click on JEOL PC SEM 6500 icon. Click yes if message

More information

Image-Pro Plus 7.0 Product Note

Image-Pro Plus 7.0 Product Note Image-Pro Plus 7.0 Product Note Automated Microscope Configuration Introduction Cameras, software, shutters, stages, objectives, filters, turrets there are a lot of components in an automated microscopy

More information

Lesson 6 2D Sketch Panel Tools

Lesson 6 2D Sketch Panel Tools Lesson 6 2D Sketch Panel Tools Inventor s Sketch Tool Bar contains tools for creating the basic geometry to create features and parts. On the surface, the Geometry tools look fairly standard: line, circle,

More information

ISIS A beginner s guide

ISIS A beginner s guide ISIS A beginner s guide Conceived of and written by Christian Buil, ISIS is a powerful astronomical spectral processing application that can appear daunting to first time users. While designed as a comprehensive

More information

MEASUREMENT CAMERA USER GUIDE

MEASUREMENT CAMERA USER GUIDE How to use your Aven camera s imaging and measurement tools Part 1 of this guide identifies software icons for on-screen functions, camera settings and measurement tools. Part 2 provides step-by-step operating

More information

Leica DMi8A Quick Guide

Leica DMi8A Quick Guide Leica DMi8A Quick Guide 1 Optical Microscope Quick Start Guide The following instructions are provided as a Quick Start Guide for powering up, running measurements, and shutting down Leica s DMi8A Inverted

More information

PASS Sample Size Software

PASS Sample Size Software Chapter 945 Introduction This section describes the options that are available for the appearance of a histogram. A set of all these options can be stored as a template file which can be retrieved later.

More information

Modeling Basic Mechanical Components #1 Tie-Wrap Clip

Modeling Basic Mechanical Components #1 Tie-Wrap Clip Modeling Basic Mechanical Components #1 Tie-Wrap Clip This tutorial is about modeling simple and basic mechanical components with 3D Mechanical CAD programs, specifically one called Alibre Xpress, a freely

More information

LEICA TCS SP5 AOBS TANDEM USER MANUAL

LEICA TCS SP5 AOBS TANDEM USER MANUAL LEICA TCS SP5 AOBS TANDEM USER MANUAL STARTING THE SYSTEM...2 THE LAS AF SOFTWARE...3 THE «ACQUIRE» MENU...5 CHOOSE AND CREATE A SETTING...6 THE CONTROL PANEL...8 THE DMI6000B MICROSCOPE...10 ACQUIRE ONE

More information

BCC Optical Stabilizer Filter

BCC Optical Stabilizer Filter BCC Optical Stabilizer Filter The new Optical Stabilizer filter stabilizes shaky footage. Optical flow technology is used to analyze a specified region and then adjust the track s position to compensate.

More information

Lesson 4 Holes and Rounds

Lesson 4 Holes and Rounds Lesson 4 Holes and Rounds 111 Figure 4.1 Breaker OBJECTIVES Sketch arcs in sections Create a straight hole through a part Complete a Sketched hole Understand the Hole Tool Use Info to extract information

More information

QuiltCAD will be used to create an entire quilt layout. It can be used for single patterns, pantographs, borders, or sashings. There are some options

QuiltCAD will be used to create an entire quilt layout. It can be used for single patterns, pantographs, borders, or sashings. There are some options QuiltCAD will be used to create an entire quilt layout. It can be used for single patterns, pantographs, borders, or sashings. There are some options that only QuiltCAD can do when compared to other portions

More information

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop 1. Log in usage using the SMIF web site 2. Turn power

More information

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 710 Confocal Microscope Standard Operation Protocol LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.

More information

inphoto ID PS Automatic ID photography With Canon PowerShot camera User Guide

inphoto ID PS Automatic ID photography With Canon PowerShot camera User Guide inphoto ID PS Automatic ID photography With Canon PowerShot camera User Guide 2018 Akond company Phone/fax: +7(812)384-6430 Cell: +7(921)757-8319 e-mail: info@akond.net akondsales@gmail.com http://www.akond.net

More information

Micro-Image Capture 8 Installation Instructions & User Guide

Micro-Image Capture 8 Installation Instructions & User Guide Micro-Image Capture 8 Installation Instructions & User Guide Software installation: Micro-Image Capture Software 1. Load Micro-Image Capture software CD onto host PC. Auto Run should start driver/software

More information

Learning Guide. ASR Automated Systems Research Inc. # Douglas Crescent, Langley, BC. V3A 4B6. Fax:

Learning Guide. ASR Automated Systems Research Inc. # Douglas Crescent, Langley, BC. V3A 4B6. Fax: Learning Guide ASR Automated Systems Research Inc. #1 20461 Douglas Crescent, Langley, BC. V3A 4B6 Toll free: 1-800-818-2051 e-mail: support@asrsoft.com Fax: 604-539-1334 www.asrsoft.com Copyright 1991-2013

More information

User s Guide. Windows Lucis Pro Plug-in for Photoshop and Photoshop Elements

User s Guide. Windows Lucis Pro Plug-in for Photoshop and Photoshop Elements User s Guide Windows Lucis Pro 6.1.1 Plug-in for Photoshop and Photoshop Elements The information contained in this manual is subject to change without notice. Microtechnics shall not be liable for errors

More information

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Operations Guide Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Nikon Eclipse 90i Operations Guide STARTUP Startup Powering Up Fluorescence

More information

JUST EMBROIDER IT! ALL YOU NEED TO KNOW ABOUT HOOP 'N' BUDDYZ

JUST EMBROIDER IT! ALL YOU NEED TO KNOW ABOUT HOOP 'N' BUDDYZ BERNINA ebook Series JUST EMBROIDER IT! ALL YOU NEED TO KNOW ABOUT HOOP 'N' BUDDYZ Written by: Judy Hahner BERNINA Educator, Professional & Home Embroidery Specialist Hoop 'N' Buddyz Embroidery Hoop and

More information

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Follow the procedures below when you use the Hitachi 7100 TEM. Starting Session 1. Turn on the cold

More information

Contents Technical background II. RUMBA technical specifications III. Hardware connection IV. Set-up of the instrument Laboratory set-up

Contents Technical background II. RUMBA technical specifications III. Hardware connection IV. Set-up of the instrument Laboratory set-up RUMBA User Manual Contents I. Technical background... 3 II. RUMBA technical specifications... 3 III. Hardware connection... 3 IV. Set-up of the instrument... 4 1. Laboratory set-up... 4 2. In-vivo set-up...

More information

SKF Shaft Alignment Tool Horizontal machines app

SKF Shaft Alignment Tool Horizontal machines app SKF Shaft Alignment Tool Horizontal machines app Short flex couplings Instructions for use Table of contents 1. Using the Horizontal shaft alignment app... 2 1.1 How to change the app language...2 1.2

More information

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE University of Virginia ITC Academic Computing Health Sciences CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE Introduction The Olympus BH-2 microscope in ACHS s microscope lab has objectives from

More information

PASS Sample Size Software. These options specify the characteristics of the lines, labels, and tick marks along the X and Y axes.

PASS Sample Size Software. These options specify the characteristics of the lines, labels, and tick marks along the X and Y axes. Chapter 940 Introduction This section describes the options that are available for the appearance of a scatter plot. A set of all these options can be stored as a template file which can be retrieved later.

More information

Flip for User Guide. Inches. When Reliability Matters

Flip for User Guide. Inches. When Reliability Matters Flip for User Guide Inches by When Reliability Matters Mastercam HSM Performance Pack Tutorial 1 Mastercam HSM Performance Pack Tutorial Tutorial I... 2 Getting started... 2 Tools used... 2 Roughing...

More information

Edge Blender Controller

Edge Blender Controller Edge Blender Controller Calibration Manual Version 2017 V1 Contents Contents INTRODUCTION 1 FEATURES 1 OS AND HARDWARE 1 INSTALLATION 2 INSTALLATION 2 UNINSTALLATION 4 INTERFACE 6 COMMUNICATION 7 CONFIG

More information

Nikon SIM-E & A1-R System

Nikon SIM-E & A1-R System Nikon SIM-E & A1-R System USER GUIDE LSU Health Sciences Center Shreveport Research Core Facility June 01 2017 Chaowei Shang 1 Table of Content 1. Start Up the System... Page 3 Hardware and microscope

More information

Difrotec Product & Services. Ultra high accuracy interferometry & custom optical solutions

Difrotec Product & Services. Ultra high accuracy interferometry & custom optical solutions Difrotec Product & Services Ultra high accuracy interferometry & custom optical solutions Content 1. Overview 2. Interferometer D7 3. Benefits 4. Measurements 5. Specifications 6. Applications 7. Cases

More information

Extreme Makeovers: Photoshop Retouching Techniques

Extreme Makeovers: Photoshop Retouching Techniques Extreme Makeovers: Table of Contents About the Workshop... 1 Workshop Objectives... 1 Getting Started... 1 Photoshop Workspace... 1 Retouching Tools... 2 General Steps... 2 Resolution and image size...

More information

Technical Note How to Compensate Lateral Chromatic Aberration

Technical Note How to Compensate Lateral Chromatic Aberration Lateral Chromatic Aberration Compensation Function: In JAI color line scan cameras (3CCD/4CCD/3CMOS/4CMOS), sensors and prisms are precisely fabricated. On the other hand, the lens mounts of the cameras

More information

User manual Automatic Material Alignment Beta 2

User manual Automatic Material Alignment Beta 2 www.cnccamera.nl User manual Automatic Material Alignment For integration with USB-CNC Beta 2 Table of Contents 1 Introduction... 4 1.1 Purpose... 4 1.2 OPENCV... 5 1.3 Disclaimer... 5 2 Overview... 6

More information

User Manual for HoloStudio M4 2.5 with HoloMonitor M4. Phase Holographic Imaging

User Manual for HoloStudio M4 2.5 with HoloMonitor M4. Phase Holographic Imaging User Manual for HoloStudio M4 2.5 with HoloMonitor M4 Phase Holographic Imaging 1 2 HoloStudio M4 2.5 Software instruction manual 2013 Phase Holographic Imaging AB 3 Contact us: Phase Holographic Imaging

More information

All projected images must be visible from the camera point of view. The content exists in 2D - an "unwrapped" view of the content in the aspect ratio

All projected images must be visible from the camera point of view. The content exists in 2D - an unwrapped view of the content in the aspect ratio How do I calibrate 360 panoramas? You can calibrate cylindrical panoramas using Vioso technology just with one single camera. This can be done by placing the camera with fisheye lens in the center of the

More information

ivu Series TG Image Sensor

ivu Series TG Image Sensor Quick Start Guide Introduction The ivu Series Image Sensor is used to monitor labels, parts, and packaging for type, size, orientation, shape, and location. The sensor has an integrated color touch screen

More information

Advanced Diploma in. Photoshop. Summary Notes

Advanced Diploma in. Photoshop. Summary Notes Advanced Diploma in Photoshop Summary Notes Suggested Set Up Workspace: Essentials or Custom Recommended: Ctrl Shift U Ctrl + T Menu Ctrl + I Ctrl + J Desaturate Free Transform Filter options Invert Duplicate

More information

The Revolve Feature and Assembly Modeling

The Revolve Feature and Assembly Modeling The Revolve Feature and Assembly Modeling PTC Clock Page 52 PTC Contents Introduction... 54 The Revolve Feature... 55 Creating a revolved feature...57 Creating face details... 58 Using Text... 61 Assembling

More information

Figure 1 The Raith 150 TWO

Figure 1 The Raith 150 TWO RAITH 150 TWO SOP Figure 1 The Raith 150 TWO LOCATION: Raith 150 TWO room, Lithography area, NanoFab PRIMARY TRAINER: SECONDARY TRAINER: 1. OVERVIEW The Raith 150 TWO is an ultra high resolution, low voltage

More information

Be aware that there is no universal notation for the various quantities.

Be aware that there is no universal notation for the various quantities. Fourier Optics v2.4 Ray tracing is limited in its ability to describe optics because it ignores the wave properties of light. Diffraction is needed to explain image spatial resolution and contrast and

More information

SUGAR fx. LightPack 3 User Manual

SUGAR fx. LightPack 3 User Manual SUGAR fx LightPack 3 User Manual Contents Installation 4 Installing SUGARfx 4 What is LightPack? 5 Using LightPack 6 Lens Flare 7 Filter Parameters 7 Main Setup 8 Glow 11 Custom Flares 13 Random Flares

More information

TM5. Guide Book. Hardware Version: 2.00 Software Version: 1.62

TM5. Guide Book. Hardware Version: 2.00 Software Version: 1.62 TM5 Guide Book Hardware Version: 2.00 Software Version: 1.62 ii Release Date : 2017-07-10 The information contained herein is the property of Techman Robot Corporation (hereinafter referred to as the Corporation).

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, EL6000 fluorescent light source for the microscope stand. 2. Turn on the Scanner Power

More information

Experiment P55: Light Intensity vs. Position (Light Sensor, Motion Sensor)

Experiment P55: Light Intensity vs. Position (Light Sensor, Motion Sensor) PASCO scientific Vol. 2 Physics Lab Manual: P55-1 Experiment P55: (Light Sensor, Motion Sensor) Concept Time SW Interface Macintosh file Windows file illuminance 30 m 500/700 P55 Light vs. Position P55_LTVM.SWS

More information

Chapter 6: TVA MR and Cardiac Function

Chapter 6: TVA MR and Cardiac Function Chapter 6 Cardiac MR Introduction Chapter 6: TVA MR and Cardiac Function The Time-Volume Analysis (TVA) optional module calculates time-dependent behavior of volumes in multi-phase studies from MR. An

More information

AutoCAD 2D I. Module 16. Isometric and Dimensioning. IAT Curriculum Unit PREPARED BY. January 2011

AutoCAD 2D I. Module 16. Isometric and Dimensioning. IAT Curriculum Unit PREPARED BY. January 2011 AutoCAD 2D I Module 16 Isometric and Dimensioning PREPARED BY IAT Curriculum Unit January 2011 Institute of Applied Technology, 2011 Module 16 Auto CAD Self-paced Learning Modules AutoCAD 2D Isometric

More information

IncuCyte ZOOM Scratch Wound Processing Overview

IncuCyte ZOOM Scratch Wound Processing Overview IncuCyte ZOOM Scratch Wound Processing Overview The IncuCyte ZOOM Scratch Wound assay utilizes the WoundMaker-IncuCyte ZOOM-ImageLock Plate system to analyze both 2D-migration and 3D-invasion in label-free,

More information

Using the Advanced Sharpen Transformation

Using the Advanced Sharpen Transformation Using the Advanced Sharpen Transformation Written by Jonathan Sachs Revised 10 Aug 2014 Copyright 2002-2014 Digital Light & Color Introduction Picture Window Pro s Advanced Sharpen transformation is a

More information

Minolta Scanner Plugin

Minolta Scanner Plugin Minolta Scanner Plugin For a list of Minolta digitizers and Geomagic software products with which this plugin is compatible, see Release Notes for Geomagic Minolta Plugin 7.6.0.3. Copyright 2005, Raindrop

More information

BacklightFly Manual.

BacklightFly Manual. BacklightFly Manual http://www.febees.com/ Contents Start... 3 Installation... 3 Registration... 7 BacklightFly 1-2-3... 9 Overview... 10 Layers... 14 Layer Container... 14 Layer... 16 Density and Design

More information

LAB I. INTRODUCTION TO LAB EQUIPMENT

LAB I. INTRODUCTION TO LAB EQUIPMENT 1. OBJECTIVE LAB I. INTRODUCTION TO LAB EQUIPMENT In this lab you will learn how to properly operate the oscilloscope Agilent MSO6032A, the Keithley Source Measure Unit (SMU) 2430, the function generator

More information

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 Leica DMI AF6000LX Table of contents START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 ACQUIRE MODULE 6 SETTING THE LIGHTPATH 6

More information

Advance Steel. Tutorial

Advance Steel. Tutorial Advance Steel Tutorial Table of contents About this tutorial... 7 How to use this guide...9 Lesson 1: Creating a building grid...10 Step 1: Creating an axis group in the X direction...10 Step 2: Creating

More information

Figure 1: NC EDM menu

Figure 1: NC EDM menu Click To See: How to Use Online Documents SURFCAM Online Documents 685)&$0Ã5HIHUHQFHÃ0DQXDO 6 :,5(('0 6.1 INTRODUCTION SURFCAM s Wire EDM mode is used to produce toolpaths for 2 Axis and 4 Axis EDM machines.

More information