CHA ELECTRON BEAM EVAPORATOR SYSTEM OPERATING MANUAL Version: 2.0 August 2016

Size: px
Start display at page:

Download "CHA ELECTRON BEAM EVAPORATOR SYSTEM OPERATING MANUAL Version: 2.0 August 2016"

Transcription

1 CHA ELECTRON BEAM EVAPORATOR SYSTEM OPERATING MANUAL Version: 2.0 August 2016 UNIVERSITY OF TEXAS AT ARLINGTON Nanotechnology Research Center TABLE OF CONTENTS

2 1.0 DESCRIPTION SAFETY REQUIREMENTS/RESTRICTIONS STARTING THE CRYO PUMP LOADING A SAMPLE PROGRAMMING THE XTC THICKNESS MONITOR BRINGING UP THE ELECTRON BEAM FOR EVAPORATION SYSTEM SHUTDOWN

3 1.0 Description The CHA Solution Electron Beam Evaporator uses a six pocket carousel crucible selector. The preapproved list of deposition materials includes Al, Ti, Ni, Ge, Au, Ag, Pt, Cr, Cu, Zn, Pd, In, & Sn. Any other materials will need Nanofab management permission. Single or stack films can be evaporated on samples or wafers up to 100mm in diameter, glass slides and small samples are accommodated as well. 2.0 Safety This machine is connected to HIGH VOLTAGE. If any SYSTEM FAULT lights come ON call NanoFab staff immediately. Be very careful and aware of electrical hazards. If you encounter any electrical malfunctions contact NanoFab staff immediately This machine uses HIGH VOLTAGE. DO NOT operate this machine with any cables disconnected, or component enclosures/panels open This machine uses HIGH VOLTAGE. DO NOT operate this machine with any cables disconnected, or component enclosures/panels open This machine uses water flow to prevent over heating of the cryo pump, electron beam gun source, carousel source crucibles and quartz crystal. If you encounter any water flow malfunctions or leaks notify NanoFab staff immediately This machine has an EMO (Emergency Off) switch/button mounted on the front panel. The EMO switch should be pressed only in an emergency. An emergency would be fire, smoke, electrocution hazards, and an injury to anyone using this particular piece of equipment. If the EMO is pressed notify NanoFab staff immediately Use of sunglasses or a light filter may be required when viewing the material heating/ evaporation process to prevent eye damage 3.0 Requirements/Restrictions You must be a qualified user on the CHA Solution E-beam evaporator Users must supply their own deposition supplies [Crucibles & Deposition Materials] for this system. See the Kurt J Lesker page below (Fig. A) for more details The evaporator can be used to deposit metal or oxide film on 1-4 diameter wafers, glass slides, and small samples by using the substrate holder clips. Extremely small pieces can be processed by using longer holder clips or bonding to larger carrier wafer (bond with drop silver paint, drop of baked photo resist, MUNG II paste or other approved bonding material)

4 4.0 System Checks Before beginning a process verify that the cryo pump is at operational temperature by checking to see if the CTI CONTROL PANEL is displaying that the cryo is ON and the temperature is < 20K (Fig 1). If the cryo is OFF do not proceed. Contact a Nanofab staff member for assistance. (Fig. 1) 5.0 Loading a Sample and Preparing for Depostion Log onto the system by pressing the LOG ON button in the lower right hand corner of the touch screen (Fig. 2). Enter e as the user and cha for the password.

5 (Fig. 2) If the system AUTO MODE indicator is green skip this step. If not then press the AUTO MODE button (See Fig. 2). The system will automatically enter STANDBY mode. Allow the Standby routine to complete Monitoring the touchscreen, once the Auto Mode and Standby indicators are both green press the Start Vent button on the right side of the screen Allow the system to Vent until the Vent Complete (Fig. 3) notification window appears on the screen. The automatic VENT routine will take approximately 5 minutes to finish. Once complete then Press the Standby button located in the VENT COMPLETE window.

6 (Fig. 3) Wait until the Standby indicator in the upper left hand corner of the screen is once again green and then press the Lock button located in the center of the screen (Fig. 4). This will unlock the door and allow you to open the chamber door. (Fig. 4) Move the chamber door to its fully open position and allow the latching mechanism (Fig. 5) to engage and hold the door open while you are loading/unloading the chamber..

7 (Fig. 5) Remove the installed wafer carrier from the chamber. Depending on the prior deposition configuration the system could either be set up with the standard rotating carrier or with the planetary rotating carrier. To remove the standard rotating carrier (Fig. 6a) lift the front most retaining pin and pull the carriage assembly towards you. To remove the planetary deposition carrier (Fig. 6b) grab the main support bar and lift the whole assembly straight up out of the locking mechanism. (Fig. 6a) (Fig. 6b) If the wrong deposition carrier assembly for your process was installed into the chamber contact a Nanofab staff member to set the system up for the alternate deposition carrier that you require Mount your wafers or sample carrier into the desired deposition carrier by installing the wafer face down into one of the wafer holder pockets (Fig. ). It is best to hold the wafer by

8 the flat as it allows you to remove the tweezers without pulling the wafer back out of the carrier pocket. (Fig. ) Latch the two spring loaded wafer retaining clips (Fig. ) by pressing them down and twisting the clip over the wafer. Release the spring loaded clip and the wafer should now be secure in the pocket. (Fig. ) Reinstall the deposition carrier containing your sample or wafer back into the system. Ensure the carrier is fully pushed the system and the locking mechanism has closed behind it.

9 (Fig. ) Now you must install the crucibles containing your evaporative materials into the hearth at the designated positions for those materials. Press the button on the touchscreen labeled SHUTTER. When this button turns green the shutter will be open. (Fig. ) Now to view the current crucible pocket location or rotate to another pocket location press the button Select Pocket button (See. Fig. ) near the center of the touchscreen. This will open the MANUAL POCKET ROTATION window (Fig. ).

10 (Fig. ) Place your crucibles into the pocket that corresponds with the films you will be using. These films are automatically selected by the process you will select from the deposition controller. If you are unsure of which pocket location your materials should go then STOP and ask a Nanofab staff member for assistance. ***Failure to put the materials in the proper pocket to match the process selected can result in damage to the system.*** (Fig. ) Once all of your deposition crucibles have been installed CLOSE the shutter by pressing the shutter button (See. Fig. ) and then ensure the button turns red Release the spring loaded Door Latch by lifting it slightly with your fingers (Fig. ). The door will begin to close on its own once the door has been unlatched. After the door moves slightly remove your fingers from the latch and close the door fully.

11 (Fig. 12) Press the LOCK button near the center of the touchscreen (See Fig. 7) Press the START PUMPDOWN button on the right hand side of the touchscreen (See. Fig. ) You must now wait until the PUMPDOWN COMPLETE notice comes up on the screen (Fig. ). This pumpdown period depends on how long you had the system open and what sample materials you installed into the chamber. Typical pumping times range from minutes until completion. Once the pumpdown is complete you have two options, ABORT which will put the system back in Standby or AUTORUN which will begin an automated deposition process. Do not select AUTORUN unless you have selected and verified your process and film thicknesses from the deposition controller. (Fig. )

12 6.0 Programming the STC-2000 deposition controller The STC-2000 is an automated process controller. It controls all aspects of the evaporation on this system including, crucible selection, film thickness, e-beam gun ramping and shutter open/close, deposition rate. Please refer to the following excerpts from the manufacturers quick start manual for STC-2000 programming information. This guide is a supplement to the STC-2000A User s Manual ( ). WARNING: Improper operation of this equipment could cause equipment damage or personal injury. Read the safety precautions in the user s manual prior to operating this equipment. Terms and definitions: The STC-2000A provides both manual and automatic control of single- or multi-layer film deposition. Manual mode is often used to set up the system in order to run unattended in the fully automatic mode. A film is a set of user-programmable deposition-related parameters (density, Z-ratio, map, etc.) Typically, one material deposition is accomplished per film. The STC-2000A can store up to 50 films. The active film is the film that will run when the start sequence is initiated. A process is a set of multiple films, or layers. The map specifies the configuration of the physical plant (number of sensors, selection of source type, which crystal to read, which gun to use, etc.) Up to 30 maps can be stored to allow multiple system configurations. Each film specifies which map to use.

13 The STC-2000A has two basic process control modes: Sequencing mode and Nonsequencing mode. Non-sequencing mode is used to run a single user-selected film. Sequencing mode is used to run a process (a series of films) for multiple materials or layers. The Runtime or Status Screen is the screen normally displayed during operation. You may return to the Status Screen at any time by pressing the Status button. Screen navigation (using fixed keys and touch screen): Arrow keys on menu screens allow scrolling to unseen list parameters. BYE is a screen or function exit. EDT is edit. GO2 any selects another film, map or process. STATUS exits the menu system and returns the Run-time Screen. STATUS instruments STOP MENU START ZERO HAND CONTROLLER MANUAL STC-2000A THICKNESS / RATE CONTROLLER

14 Navigation of Status Screens: The Status button 1st Status Screen Press STATUS button 2nd Status Screen The 2 nd Status Screen and thickness in a large font for distant viewing. Press STATUS button 3rd Status Screen The 3 rd Status Screen displays gun information. Press STATUS button 4th Status Screen displays sensor information Press STATUS button The 1 st Status Screen displays information related to the current deposition.

15 Status Screen Menu Programming: Sequencing Mode Press fixed MENU key Menu Screen Review Film Number Service Review Source Sensor (SS) Map Executive Menu Next Process Process Accounting Options/ Information/ Diagnostics Review Processes System Configuration I/O Set-up Menu Communications Set-up I/O Set-up: Memory I/O Set-up: Edit I/O Set-up: Operations Mode I/O Set-up: Change

16 Basic Operation Flow Chart: To perform a practice run of X thickness At X rate: Power On REVIEW PROCESSES menu: Enter parameters, select Film Unit power on, connected Press MENU key NEXT Process menu: Enter Process ID no.* REVIEW SOURCE SENSOR (SS) MAP menu: Select output, tooling and sensor values Press "Review SS Map" Set SOURCE MAX PWR to 100%, and CH1 START MODE to "ACTIVE" Press BYE to return to Menu Screen Press "REVIEW PROCESSES" REVIEW FILM Enter Film parameters, select MAP Press BYE to return to Menu Screen Confirm MODE is AUTO, and THICK (KA) is Make back panel connections, check safety interlocks and cooling water Return to Run-time Screen (Status key) FILM" VALUE to 50%, PWR RAMP1 TIME to 00:11, PWR SOAK1 TIME to 00:11, and MAX POWER LIMIT to 100% Press STATUS Run-time Screen Press START key to initiate start sequence Ready to Start * This can only be done when unit is idle.

17 To switch from Sequencing to Non-sequencing or vice-versa: Press the fixed Menu button to view the Menu screen. Press Service on the touchscreen to view the Service menu. Non-sequencing: Sequencing: Press to toggle Press Accept on the touchscreen to accept the change. The STC-2000A will automatically restart after the change is accepted. (Note: Accept button appears only for a status change.)

18 To edit film parameters: Press the fixed Menu button to view the Menu screen. Press Review Film on the touchscreen to view the film parameter edit screen. Select "GO2 Flm" on the touchscreen, and enter the number of the film you want to review/edit (select the number using the touchscreen keypad, then select "Enter".) When the film numbers next to "Edit" and "Active" at the top of the screen match, you are editing the active film. Note: Changes made to the active film will change a running process

19 7.0 Running an Automatic Process Sequence There are two ways to run an automatic process on the system. The first is after you have completed the START PUMPDOWN (Fig. 14) Turn on the interlock key by turning the key from OFF to MAN (Fig. 15), Do not turn the key all the way to AUTO or the system will not function properly. All of the yellow interlock lights will turn on and the green READY light will be on. If all the lights are not on call NanoFab staff. (Fig. 15) Turn SOURCE 1 emission control knob on the front of the Interlock panel (See. Fig. 14) to 0 by turning knob CCW until it stops.

20 DOCUMENT: CHA SOLUTION E-BEAM EVAPORATOR SYSTEM OPERATING PROCEDURES Version: Turn the REMOTE BEAM CONTROLLER (Fig. 16) emission current dial to 0 by turning the EMISSION current potentiometer fully CCW until it stops at approximately 0. (Fig. 16) Turn on MAINS power by moving switch/breaker to ON position for the SEB-15 e- beam power supply (Fig. 16). If any ARC FAULT, OVERLOAD, SYSTEM FAULT lights go on turn OFF MAINS switch/breaker and call NanoFab staff immediately. Only the MAINS ON and INTERLOCK lights should be on Turn on high voltage by pressing the white HV ON push-button switch for the SEB- 15 (Fig. 17). The HV ON light will turn on. If any ARC FAULT, OVERLOAD, or SYSTEM FAULT lights go on press the green HV OFF push-button switch, turn OFF MAINS switch/breaker and call NanoFab staff immediately. 20

21 DOCUMENT: CHA SOLUTION E-BEAM EVAPORATOR SYSTEM OPERATING PROCEDURES Version: 1.0 (Fig. 17) If necessary adjust the OUTPUT VOLTAGE to 9.0 KV by turning output voltage adjust potentiometer (See Fig. 16) CW or CCW until the display shows 9.0 KV While watching the X Y Beam sweep controller Green LED readouts, adjust both the amplitude and position potentiometers on the Remote Beam Controller to be centered (position pot) and approximately 3 to 5 LED s in width (Amplitude pot) Check to ensure the Remote Emission Controller emission current is at 0 by rotating emission potentiometer CCW until it Turn on the Remote Emission Controller by pressing the EMISSION ON pushbutton switch. The ON light will turn red Move the Source shutter switch to its OPEN position (Fig. 18). (Fig. 18) Slowly increase emission current by rotating EMISSION potentiometer CW to a setting of while looking through the source view port window to monitor 21

22 DOCUMENT: CHA SOLUTION E-BEAM EVAPORATOR SYSTEM OPERATING PROCEDURES Version: 1.0 the initial heating process (about 15 downward). If necessary use the filter lens when viewing the material heating/evaporation process to prevent eye damage While looking into the viewport window you should be able to see the blue e-beam spot or the material glowing hot near the center of the crucible. Make fine centering adjustments to the Horizontal and Lateral Position potentiometers On the Remote Beam Controller increase/decrease the Horizontal Amplitude potentiometer CW to generate an optimal e-beam sweep without encroaching on the crucible edge On the Remote Beam Controller increase/decrease the Lateral Amplitude potentiometer CW to generate an optimal e-beam sweep without encroaching on the crucible edge On the Remote Beam Controller increase/decrease the Horizontal and Lateral Frequency potentiometers CW to achieve an optimum sweep frequency/speed Very slowly increase emission current by rotating EMISSION potentiometer CW. When you have the desired evaporation rate (2-3A/sec is typical) then press the ZERO button on the Inficon XTC thickness monitor (See. Fig. 8) Open the Substrate shutter (Fig. 19) to begin the deposition. (Fig. 19) When you have deposited the desired thickness as shown on the Inficon XTC thickness monitor (See Fig. 12), quickly close both the SOURCE AND SUBSTRATE shutter s by moving both switches to their closed position s (See Fig. 18) Begin the emission current shut down by SLOWLY decreasing emission current by rotating EMISSION potentiometer CCW until it 0. This process should take a couple of minutes to allow the crucible and filament sufficient time to cool slowly. 22

23 DOCUMENT: CHA SOLUTION E-BEAM EVAPORATOR SYSTEM OPERATING PROCEDURES Version: Turn off the Remote Emission Controller by pressing the EMISSION OFF pushbutton switch. The OFF light will turn green Turn off high voltage by pressing the green HV OFF push-button switch (SEB-06) and then turn off the Mains power by moving the breaker to the OFF position Then turn off the X Y Beam sweep controller and turn the Interlock key to the OFF position If you will be depositing another film, rotate the SOURCE SELECTOR KNOB (See Fig. 9) to the desired material and then proceed from Step System Shutdown If the substrate was heated then you must wait until the temperature displays < 50 deg before proceeding to the next step Close the HIVAC valve (See Fig. 7) and wait 5 minutes before proceeding to the next step, this allows the crucible sufficient time to cool down Open the VENT valve (See Fig. 7) and monitor the CHAMBER THERMOCOUPLE gauge display (See Fig. 6). Allow the chamber pressure to rise until it indicates approximately ATM (See Fig. 6) Press the RAISE button (See Fig. 7) to raise the bell jar. Once the bell jar has fully raised close the VENT valve Remove the Sample Carrier from the system, unload your sample(s) from the plate and then reinstall the empty sample plate into the system Press the LOWER button (See Fig. 7) to lower the bell jar until it stops moving. Move the Raise/Lower switch back to the middle position Open the CHAMBER ROUGH valve (See Fig. 7) and monitor the CHAMBER THERMOCOUPLE gauge display (See Fig. 11) allowing the chamber pressure to drop to < 100mT before proceeding Close the CHAMBER ROUGH valve and then turn the MECHANICAL PUMP switch off (See Fig. 4) If another user is going to use the CHA E-beam system after you then skip this step. If no other user will be running the CHA E-beam system then proceed. On the Cryo Pump Controller (See Fig. 5) turn the CRYO-PUMP power switch OFF. 23

24 DOCUMENT: CHA SOLUTION E-BEAM EVAPORATOR SYSTEM OPERATING PROCEDURES Version: Rotate the CHAMBER COLD CATHODE (See Fig. 12) controller knob to the OFF position Switch the system controller from MANUAL to the OFF position (see Fig. 1) Pick up your belongings, clean and organize the table and complete your log entry in the system logbook. 24

West Bond Model 7476E-79 Wedge Wire Bonder STANDAR OPERATING PROCEDURE Version: 1.0 March 2014

West Bond Model 7476E-79 Wedge Wire Bonder STANDAR OPERATING PROCEDURE Version: 1.0 March 2014 West Bond Model 7476E-79 Wedge Wire Bonder STANDAR OPERATING PROCEDURE Version: 1.0 March 2014 UNIVERSITY OF TEXAS AT ARLINGTON Nanotechnology Research Center (NRC) TABLE OF CONTENTS 1. Introduction....3

More information

Cressington 108 Auto/SE Sputter Coater Standard Operating Procedures (S.O.P)

Cressington 108 Auto/SE Sputter Coater Standard Operating Procedures (S.O.P) Cressington 108 Auto/SE Sputter Coater Standard Operating Procedures (S.O.P) The Cressington sputter system is designed for only one purpose which is the deposition of gold onto a sample to reduce charging

More information

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Amray 1830 SEM. All users are expected to have read and understood this document.

More information

HEX02 EMBOSSING SYSTEM

HEX02 EMBOSSING SYSTEM HEX02 EMBOSSING SYSTEM LOCATION: Hot Embossing Area PRIMARY TRAINER: 1. Scott Munro (2-4826, smunro@ualberta.ca) OVERVIEW The hot embosser is available to users who require polymer mold fabrication. This

More information

Micro Automation- Model 1006 Dicing Saw Instructions. Serial # Rev 2 ( R.DeVito) Location Chase 1

Micro Automation- Model 1006 Dicing Saw Instructions. Serial # Rev 2 ( R.DeVito) Location Chase 1 Micro Automation- Model 1006 Dicing Saw Instructions Serial # Rev 2 (12-23-05 R.DeVito) Location Chase 1 Dicing Saw Instructions (Revised 8/9/03 - K.J) 1. On the Log Sheet sign in, including Name and Date.

More information

DISCO DICING SAW SOP. April 2014 INTRODUCTION

DISCO DICING SAW SOP. April 2014 INTRODUCTION DISCO DICING SAW SOP April 2014 INTRODUCTION The DISCO Dicing saw is an essential piece of equipment that allows cleanroom users to divide up their processed wafers into individual chips. The dicing saw

More information

Basic Users Manual for Tecnai-F20 TEM

Basic Users Manual for Tecnai-F20 TEM Basic Users Manual for Tecnai-F20 TEM NB: This document contains my personal notes on the operating procedure of the Tecnai F20 and may be used as a rough guide for those new to the microscope. It may

More information

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack). FEI Tecnai F20 S/TEM: imaging in TEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 01/21/18 1. Filling the cold trap (if needed) 1.1. Prior to use, the cold trap needs

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual NanoFactor NVG-200A Silicon Grinder Version 1.1 Page 1 of 18 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness Standard 2.2 Possible Grinding Materials

More information

R-1300-P PROGRAMMABLE ELECTRICAL MUFFLE KILN USER MANUAL

R-1300-P PROGRAMMABLE ELECTRICAL MUFFLE KILN USER MANUAL 1 R-1300-P PROGRAMMABLE ELECTRICAL MUFFLE KILN USER MANUAL PRODUCT INTRODUCTION: The R-1300-P programmable, 1300 C / 2372 F electrical muffle kiln with a 32-segments temperature/time controller is especially

More information

Instructions for Tecnai a brief start up manual

Instructions for Tecnai a brief start up manual Instructions for Tecnai a brief start up manual Version 3.0, 8.12.2015 Manual of Tecnai 12 transmission electron microscope located at Aalto University's Nanomicroscopy Center. More information of Nanomicroscopy

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual Buehler EcoMet TM 300 Polisher Version 1.0 Page 1 of 19 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness Standard 2.2 Possible Polishing Materials 2.3

More information

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE by Doug Bray Department of Biological Sciences University of Lethbridge Revised September, 2000 Note: The terms in bold in this document represent

More information

University of Minnesota Nano Fabrication Center Standard Operating Procedure Equipment Name:

University of Minnesota Nano Fabrication Center Standard Operating Procedure Equipment Name: Equipment Name: Coral Name: Nanoimprinter Revision Number: 1.1 Model: NX-B200 Revisionist: M. Fisher Location: Bay 4 Date: 2/12/2010 1 Description Nanonex NX-B200 nanoimprinter is another method of transfer

More information

Title: Micro Automation Programmable Dicing Saw, Model 1006

Title: Micro Automation Programmable Dicing Saw, Model 1006 Enhanced Sop #53 Revised: March/April 1999 Last Revised: February 2000 Title: Micro Automation Programmable Dicing Saw, Model 1006 Purpose Operational and procedural instructions used to separate individual

More information

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400. Smith College August 2009 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check 1 Startup 1 Specimen Insertion 2 Filament Saturation 2 Beam Alignment

More information

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. Smith College August 2005 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check, 1 Specimen Insertion, 1 Startup, 2 Filament Saturation, 2 Beam Alignment,

More information

Replacing the print cartridges

Replacing the print cartridges http://www.hp.com/support/lj9500 1 2 3 Replacing the print cartridges The printer uses four color print cartridges: yellow, magenta, cyan, and black. Follow this procedure to install the print cartridges.

More information

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Follow the procedures below when you use the Hitachi 7100 TEM. Starting Session 1. Turn on the cold

More information

Series 48 Water Cooled Laser & UC-2000 Quick Start Guide

Series 48 Water Cooled Laser & UC-2000 Quick Start Guide Important Read all Danger, Warning, Caution terms, symbols, and instructions located in the (Laser Safety Hazard information) sections in the Series 48 Laser Operation Manuals. http://www.synrad.com/synrad/docroot/resources/libraries/manuals

More information

2 How to operate the microscope/obtain an image

2 How to operate the microscope/obtain an image Morgagni Operating Instructions 50079 010912 2-1 2 ow to operate the microscope/obtain an image 2.1 Starting the microscope 2.1.1 Starting the microscope with several manually-operated steps 1. Turn on

More information

1. Preliminary sample preparation

1. Preliminary sample preparation FEI Helios NanoLab 600 standard operating procedure Nicholas G. Rudawski ngr@ufl.edu (352) 392 3077 (office) (805) 252-4916 (cell) Last updated: 03/02/18 What this document provides: an overview of basic

More information

Micro Automation Programmable Dicing Saw, Model 1006

Micro Automation Programmable Dicing Saw, Model 1006 Micro Automation Programmable Dicing Saw, Model 1006 Purpose: Operational and procedural instructions used to separate individual die on a silicon wafer or glass substrate. Initial Precautions (1) Edges

More information

Figure 1 The Raith 150 TWO

Figure 1 The Raith 150 TWO RAITH 150 TWO SOP Figure 1 The Raith 150 TWO LOCATION: Raith 150 TWO room, Lithography area, NanoFab PRIMARY TRAINER: SECONDARY TRAINER: 1. OVERVIEW The Raith 150 TWO is an ultra high resolution, low voltage

More information

RENISHAW INVIA RAMAN SPECTROMETER

RENISHAW INVIA RAMAN SPECTROMETER STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw invia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples

More information

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below.

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. http://staff.washington.edu/wpchan/if/cm100_inst.shtml Instructions for the Philips CM100 TEM and peripherals

More information

INSTRUCTION MANUAL. Force Transducer Output Tube Repair Kit

INSTRUCTION MANUAL. Force Transducer Output Tube Repair Kit INSTRUCTION MANUAL Model 400-TR Force Transducer Output Tube Repair Kit June 4, 2004, Revision 5 Copyright 2004 Aurora Scientific Inc. Aurora Scientific Inc. 360 Industrial Pkwy. S., Unit 4 Aurora, Ontario,

More information

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 1. Utility Requirements a. System power is supplied by two 120 VAC/20 A circuits. When doing maintenance

More information

JEOL 6500 User Manual

JEOL 6500 User Manual LOG IN to your session on the computer to the left of the microscope. Starting Conditions 1. Press Ctrl-Alt-Del and log on to the microscope computer. Click on JEOL PC SEM 6500 icon. Click yes if message

More information

SEM Training Notebook

SEM Training Notebook SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside December 21, 2017 (rev. 3.4) 1 Before you begin Complete

More information

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Warning Always adjust your specimen height before closing the chamber door to make sure your specimen will not hit the bottom of the lens;

More information

CNC Using the FlexiCam CNC and HMI Software. Guldbergsgade 29N, P0 E: T:

CNC Using the FlexiCam CNC and HMI Software. Guldbergsgade 29N, P0 E: T: CNC Using the FlexiCam CNC and HMI Software Guldbergsgade 29N, P0 E: makerlab@kea.dk T: +46 46 03 90 This grey box is the NC controller. Let s start by turning the red switch to the ON position, then press

More information

Repairing Microsoft Wedge Touch Mouse Battery Cover Retaining Clip

Repairing Microsoft Wedge Touch Mouse Battery Cover Retaining Clip Repairing Microsoft Wedge Touch Mouse Battery Cover Retaining Clip Disassembly, repair and reassembly of Wedge Touch mouse when the battery cover will not stay closed. Also is a good guide to repair other

More information

Standard Operating Procedure: TA Instruments Discovery DSC

Standard Operating Procedure: TA Instruments Discovery DSC Core Research Facilities University of Massachusetts Lowell One University Avenue Lowell, MA 01854 Standard Operating Procedure: TA Instruments Discovery DSC APPROVAL REQUIREMENTS TITLE DEPT NAME (PRINTED)

More information

SUSS Mask Aligner. Purpose: To expose photoresist on a wafer using a photomask

SUSS Mask Aligner. Purpose: To expose photoresist on a wafer using a photomask SUSS Mask Aligner Purpose: To expose photoresist on a wafer using a photomask Overview This SOP will go over how to use the machine for basic exposures. This will include commonly used controls and frequently

More information

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES RULES All users must go through a series of standard operation procedure training. For more information contact: Longlong Liao Teaching

More information

USER MANUAL ENGLISH 1450 COIN COUNTER & SORTER

USER MANUAL ENGLISH 1450 COIN COUNTER & SORTER USER MANUAL ENGLISH 1450 COIN COUNTER & SORTER INTRODUCTION ENGLISH Thank you for purchasing the Safescan 1450 coin counter and sorter. For proper use and maintenance, we advise to read this user manual

More information

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and

More information

FEI Tecnai G 2 F20 Operating Procedures

FEI Tecnai G 2 F20 Operating Procedures FEI Tecnai G 2 F20 Operating Procedures 1. Startup (1) Sign-up in the microscope log-sheet. Please ensure you have written an account number for billing. (2) Log in to the computer: Login to your account

More information

PT 335SCC Multi Operator Manual

PT 335SCC Multi Operator Manual PT 335SCC Multi Operator Manual 755 Griffith Court, Burlington, Ontario, Canada L7L 5R9 Tel:+(905) 633-7663 Fax:+(905) 637-4419 www.graphicwhizard.com 1 TABLE OF CONTENTS Important Information General

More information

FEI Helios NanoLab 600 TEM specimen prep recipe Nicholas G. Rudawski (352) (office) (805) (cell) Last updated: 01/19/17

FEI Helios NanoLab 600 TEM specimen prep recipe Nicholas G. Rudawski (352) (office) (805) (cell) Last updated: 01/19/17 FEI Helios NanoLab 600 TEM specimen prep recipe Nicholas G. Rudawski ngr@ufl.edu (352) 392 3077 (office) (805) 252-4916 (cell) Last updated: 01/19/17 This recipe is based on the methods of Schaffer et

More information

SEM Training Notebook

SEM Training Notebook SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside March 8, 2018 (rev. 3.5) 1 Before you begin Complete

More information

TEM SAMPLE-PREPARATION PROCEDURES FOR THIN-FILM MATERIALS

TEM SAMPLE-PREPARATION PROCEDURES FOR THIN-FILM MATERIALS TEM SAMPLE-PREPARATION PROCEDURES FOR THIN-FILM MATERIALS Initial Set-Up: Heat up a hot plate to around 150-200 C Plan view Mounting/Grinding/Dimpling/Polishing: 1) Cleave a square-ish piece of sample.

More information

ESA-300 Full Breakout

ESA-300 Full Breakout Interior View 0 Installation Instructions For use with ESA II Controler DORMA AUTOMATICS, Inc. 94 Sherwood Drive Toll-Free: 877-67-6 DL844-00 Lake Bluff, IL 60044 Fax: 877-4-7999 Rev. /07 Tools Required:

More information

SEM OPERATION IN LOW VACUUM MODE

SEM OPERATION IN LOW VACUUM MODE SEM OPERATION IN LOW VACUUM MODE Instructions for JEOL 5800 LV The EVAC light of the SEM specimen chamber should be already lit when you approach the SEM & the SEM will have been left in the high vacuum

More information

ipad 3 4G Home Button Control Board Replacement

ipad 3 4G Home Button Control Board Replacement ipad 3 4G Home Button Control Board Replacement Replace the home button control board in your ipad 3. Written By: Brett Hartt ifixit CC BY-NC-SA www.ifixit.com Page 1 of 28 INTRODUCTION This guide will

More information

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only Version 1.0 Prepared by D. Turnbull February 21, 2007. Please submit any omissions to the Author Note: This SEM is a recent

More information

ipad 3 4G Home Button Assembly Replacement

ipad 3 4G Home Button Assembly Replacement ipad 3 4G Home Button Assembly Replacement Replace the home button assembly in your ipad 3 4G. Written By: Brett Hartt ifixit CC BY-NC-SA www.ifixit.com Page 1 of 29 INTRODUCTION Use this guide to replace

More information

* * 98/9950WDC. Fire-Rated Devices include these. additional components. 1-Point Latch (LBL)

* * 98/9950WDC. Fire-Rated Devices include these. additional components. 1-Point Latch (LBL) *24739468* 24739468 98/9950WDC Installation Instructions 1-Point Latch (LBL) Fire-Rated Devices include these additional components Metal Edge Wrap required for 60 and 90-minute applications 2-Point Latch

More information

OKI C911 / 931 / 941

OKI C911 / 931 / 941 OKI C911 / 931 / 941 Disassembly & Reassembly Procedures Distributed at the Oki Data C900 Series Service Classes Oki Data Training Department REV. 1.60 2000 Bishops Gate Blvd. Mt. Laurel, NJ 08054 Copyright

More information

Blue Point Engineering

Blue Point Engineering Blue Point Engineering Instruction I www.bpesolutions.com Pointing the Way to Solutions! Animatronic Wizard - 3 Board (BPE No. WAC-0030) Version 3.0 2009 Controller Page 1 The Wizard 3 Board will record

More information

JEOL 6700 User Manual 05/18/2009

JEOL 6700 User Manual 05/18/2009 JEOL 6700 User Manual 05/18/2009 LOG IN to your session on the computer to the right of the microscope. Starting Conditions 1. Click the button and read the Penning Gauge to ensure that the microscope

More information

Lindberg/Blue 3-Zone Thermal Oxidation/Anneal Furnace (Model STF55366) Operating Instructions

Lindberg/Blue 3-Zone Thermal Oxidation/Anneal Furnace (Model STF55366) Operating Instructions Lindberg/Blue 3-Zone Thermal Oxidation/Anneal Furnace (Model STF55366) Operating Instructions This tool is to be used by authorized personnel only. For training & consultation contact: Manager, Omid Mahdavi,

More information

Model SU3500 Scanning Electron Microscope

Model SU3500 Scanning Electron Microscope Model SU3500 Scanning Electron Microscope Modified and Parts taken from Hitachi Easy Operation Guide. Before using the Model SU3500 SEM, be sure to read the [GENERAL SAFETY GUIDELINES] in the instruction

More information

How to remove Jammed Paper

How to remove Jammed Paper How to remove Jammed Paper C301/C310/C321/C330/ C331/C510/C511/C530/ C531/ES5430/ES5431 Rev.1 Paper Jams Paper jams occasionally occur due to paper misfeeding from a paper tray or at any point on the paper

More information

05/20/14 1. Philips CM200T. Standby Condition

05/20/14 1. Philips CM200T. Standby Condition 05/20/14 1 Philips CM200T Standby Condition HT and filament off, HT setting at 200kV. RESET HOLDER, center sample tilt knobs, and remove sample. Mag ~ 5-10kX Objective and SA apertures out, C2 aperture

More information

1.1. Log on to the TUMI system (you cannot proceed further until this is done).

1.1. Log on to the TUMI system (you cannot proceed further until this is done). FEI DB235 SEM mode operation Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Sample loading 1.1. Log on to the TUMI system (you cannot proceed further until this is done). 1.2. The FIB software (xp)

More information

User Operation of JEOL 1200 EX II

User Operation of JEOL 1200 EX II **Log onto Computer** Open item program Start Up Procedure User Operation of JEOL 1200 EX II 1. If scope is not running, locate an electron microscopy technician (EMT) to find out why not. 2. Turn up brightness

More information

LASER ENHANCED REVOLVER GRIP OWNER S MANUAL RED LASER GREEN LASER

LASER ENHANCED REVOLVER GRIP OWNER S MANUAL RED LASER GREEN LASER LASER ENHANCED RED LASER GREEN LASER REVOLVER GRIP OWNER S MANUAL LASER ENHANCED GRIP Installation Instructions Caution... 3 Safety Labels... 4 Installation...5-7 Programming...8-10 Batteries (Red Laser)...

More information

How to remove Jammed Paper

How to remove Jammed Paper How to remove Jammed Paper C801/C810/C821/C830/ C8600/C8800/ES2632a3/ ES8430 Rev.1 Paper Jams Paper jams occasionally occur due to paper misfeeding from a paper tray or at any point on the paper path through

More information

Standard Operating Procedure

Standard Operating Procedure RIT MULTIDISCIPLINARY SENIOR DESIGN 2010 Standard Operating Procedure Baja Water Propulsion Test Stand This SOP specifies how to assemble, use, troubleshoot, and disassemble the water propulsion system

More information

FE-SEM SU-8020 Operating manual (Preliminary version)

FE-SEM SU-8020 Operating manual (Preliminary version) FE-SEM SU-8020 Operating manual (Preliminary version) 2016/04/11 Seimitsu Bunseki sitsu lab. Starting up 1.Turn on the Display switch. Windows OS is starting up 2. Select the user SU-8000. 3. Click the

More information

Series 70 Servo NXT - Modulating Controller Installation, Operation and Maintenance Manual

Series 70 Servo NXT - Modulating Controller Installation, Operation and Maintenance Manual THE HIGH PERFORMANCE COMPANY Series 70 Hold 1 sec. Hold 1 sec. FOR MORE INFORMATION ON THIS PRODUCT AND OTHER BRAY PRODUCTS PLEASE VISIT OUR WEBSITE www.bray.com Table of Contents 1. Definition of Terms.........................................2

More information

Heidelberg µpg 101 Laser Writer

Heidelberg µpg 101 Laser Writer Heidelberg µpg 101 Laser Writer Standard Operating Procedure Revision: 3.0 Last Updated: Aug.1/2012, Revised by Nathanael Sieb Overview This document will provide a detailed operation procedure of the

More information

WARNINGS & SAFETY INFORMATION: PLEASE READ CAREFULLY AND KEEP A COPY ON THE PROPERTY FOR FUTURE REFERENCE!

WARNINGS & SAFETY INFORMATION: PLEASE READ CAREFULLY AND KEEP A COPY ON THE PROPERTY FOR FUTURE REFERENCE! WARNINGS & SAFETY INFORMATION: PLEASE READ CAREFULLY AND KEEP A COPY ON THE PROPERTY FOR FUTURE REFERENCE! General Safety Information To reduce the risk of death, personal injury or property damage from

More information

Print Head Installation Guide

Print Head Installation Guide Print Head Installation Guide MCS Raptor 6 (MCS Eagle AMS Software) is copyright of MCS Incorporated. 2015 MCS Incorporated. 1 Contents Tools... 4 Warnings... 4 Introduction... 4 Section One - Pillar Installation...

More information

Written By: Brett Hartt

Written By: Brett Hartt Replacing the Wi-Fi antenna in the third generation ipad Written By: Brett Hartt ifixit CC BY-NC-SA www.ifixit.com Page 1 of 31 INTRODUCTION Wireless internet is awesome. A third generation ipad without

More information

С 800 CASSIDA C 800 HIGH SPEED COIN COUNTER

С 800 CASSIDA C 800 HIGH SPEED COIN COUNTER С 800 CASSIDA C 800 HIGH SPEED COIN COUNTER This manual contains important information on safety measures and operational features. Please read it carefully before operating your coin counter, and keep

More information

Lexmark T65x Clearing Jams Guide

Lexmark T65x Clearing Jams Guide Lexmark T65x Clearing Jams Guide Clearing jams Avoiding jams The following hints can help you avoid jams. Paper tray recommendations Make sure the paper lies flat in the tray. Do not remove trays while

More information

Series 48 Air Cooled Laser & UC-2000 Quick Start Guide

Series 48 Air Cooled Laser & UC-2000 Quick Start Guide Important Read all Danger, Warning, Caution terms, symbols, and instructions located in the (Laser Safety Hazard information) sections in the Series 48 Laser Operation Manuals. http://www.synrad.com/synrad/docroot/resources/libraries/manuals.

More information

Analog Laser Sensor. Installation Instructions. Specifications. Summary of Changes. Description. Catalog Numbers 45BPD-8LTB1-D5, 45BPD-8LTB2-D5

Analog Laser Sensor. Installation Instructions. Specifications. Summary of Changes. Description. Catalog Numbers 45BPD-8LTB1-D5, 45BPD-8LTB2-D5 Installation Instructions Original Instructions Analog Laser Sensor Catalog Numbers 45BPD-8LTB1-D5, 45BPD-8LTB2-D5 Topic IMPORTANT Summary of Changes This manual contains an update to the sensing beam

More information

A3 Pro INSTRUCTION MANUAL. Oct 25, 2017 Revision IMPORTANT NOTES

A3 Pro INSTRUCTION MANUAL. Oct 25, 2017 Revision IMPORTANT NOTES A3 Pro INSTRUCTION MANUAL Oct 25, 2017 Revision IMPORTANT NOTES 1. Radio controlled (R/C) models are not toys! The propellers rotate at high speed and pose potential risk. They may cause severe injury

More information

INSTRUCTION MANUAL INF Fax: (503)

INSTRUCTION MANUAL INF Fax: (503) INSTRUCTION MANUAL INF151 1-800-547-5740 Fax: (503) 643-6322 www.ueiautomotive.com email: info@ueitest.com Introduction Congratulations on your purchase of the INF151 infrared thermometer. Like all UEi

More information

Lexmark X65x Clearing Jams Guide

Lexmark X65x Clearing Jams Guide Lexmark X65x Clearing Jams Guide Clearing jams Avoiding jams The following hints can help you avoid jams. Paper tray recommendations Make sure the paper lies flat in the tray. Do not remove trays while

More information

Nikon D7100 Camera Kit. -Checklist and Operations Manual-

Nikon D7100 Camera Kit. -Checklist and Operations Manual- Airborne Digital Reconnaissance System (ADRS) Nikon D7100 Camera Kit -Checklist and Operations Manual- V4.2 October 21, 2014 National Headquarters, Civil Air Patrol 2 1.0 Equipment Pre-Mission Check 1.1

More information

8000 SERIES PRECISION MULTIMETER VERIFICATION AND ADJUSTMENT GUIDE

8000 SERIES PRECISION MULTIMETER VERIFICATION AND ADJUSTMENT GUIDE 8000 SERIES PRECISION MULTIMETER VERIFICATION AND ADJUSTMENT GUIDE TRANSMILLE LTD. Version 1.1 : Apr 2015 TABLE OF CONTENTS PREPARING FOR CALIBRATION... 4 INTRODUCTION... 4 CALIBRATION INTERVAL SELECTION...

More information

Strata DB235 FESEM FIB

Strata DB235 FESEM FIB Strata DB235 FESEM FIB Standard Operating Procedure Revision: 5.0 Last Updated: August 16/2016, revised by Li Yang Overview This document will provide a detailed operation procedure of the Focused Ion

More information

MM540 Installation Instructions IMPORTANT SAFETY INSTRUCTIONS - SAVE THESE INSTRUCTIONS

MM540 Installation Instructions IMPORTANT SAFETY INSTRUCTIONS - SAVE THESE INSTRUCTIONS MM50 Installation Instructions IMPORTANT SAFETY INSTRUCTIONS - SAVE THESE INSTRUCTIONS Please read this entire manual before you begin. Do not unpack any contents until you verify all requirements on PAGE.

More information

Written By: Walter Galan

Written By: Walter Galan Replace the logic board in your ipad 3 Wi-Fi. Written By: Walter Galan ifixit CC BY-NC-SA www.ifixit.com Page 1 of 29 INTRODUCTION Use this guide to replace the logic board. TOOLS: iopener (1) Phillips

More information

Written By: Brett Hartt

Written By: Brett Hartt Replace the battery in your ipad 3 4G. Written By: Brett Hartt ifixit CC BY-NC-SA www.ifixit.com Page 1 of 36 INTRODUCTION When your ipad can't stay awake for longer than a few hours, it is time to replace

More information

Installing the detector filter assembly (G )

Installing the detector filter assembly (G ) s1 Installing the detector filter assembly (G1600-62700) The detector filter transmits light of a defined wavelength range. Figure 1 shows a typical transmission spectrum. Maximum transmission is between

More information

QL55 Series Luminescence Sensor

QL55 Series Luminescence Sensor Self-contained, microprocessor-based luminescence sensor Features Self-contained design in a robust, compact metal housing High sensitivity Microprocessor-controlled Senses luminescent marks, even on luminescent

More information

MobileRadio. Owner'sManual

MobileRadio. Owner'sManual EMH MobileRadio Owner'sManual TABLE OF CONTENTS Introduction... 1 Basic Operation... 2 Code Guard Operation... 3 EMH Radio Controls... 4 Button Functions... 4 Built-in Features... 7 Keypad Microphone Operation...

More information

Senior Swing. Caution

Senior Swing. Caution *740132* 740132 2800 Overhead Concealed Series 9500 Surface Applied Series Senior Swing Installation Instructions Caution LCN Senior Swing The Senior Swing Power Operator System is a low energy product

More information

2004-(Z Series) HDILA Lamp Replacement Procedure Quick Step Guide

2004-(Z Series) HDILA Lamp Replacement Procedure Quick Step Guide 2004-(Z Series) HDILA Lamp Replacement Procedure Quick Step Guide How To Determine If The Lamp Requires Replacement? After 30 seconds of turning on the TV The POWER LED and LAMP LED rapidly blink together

More information

CW-Type Wall Mounted Cabinets Installation Instructions

CW-Type Wall Mounted Cabinets Installation Instructions Instruction Sheet 860446194 Issue 2, March 2013 CW-Type Wall Mounted Cabinets Installation Instructions General CommScope wall mount cabinets are available in 24-inch (610mm), 36-inch (914mm) and 48-inch

More information

ABM International, Inc.

ABM International, Inc. ABM International, Inc. Lightning Stitch required 1 1.0: Parts List head and motor assembly (Qty. 1) Reel stand (Qty. 1) Needle bar frame clamp (Qty. 1) Motor drive (Qty. 1) 2 Cable harness with bracket

More information

Standard Operating Procedure for Cutting Stainless Steel on the Rofin Laser Machining Station

Standard Operating Procedure for Cutting Stainless Steel on the Rofin Laser Machining Station Standard Operating Procedure for Cutting Stainless Steel on the Rofin Laser Machining Station Version 1.1, August 15, 2007, by Steven Sergenese Reviewed and up-dated 02-02-2010, by D.Turnbull. Introduction

More information

BATTERY GRIP INSTRUCTION MANUAL

BATTERY GRIP INSTRUCTION MANUAL BG-D700 BG-D90 BATTERY GRIP INSTRUCTION MANUAL Product Diagram 1- Contact Cap 2- Holder for Nikon dslr Signal Contact Cover 3- Signal Contacts 4- Auto-Focus Button 5- Main Control Dial 6- Multi-Selector

More information

Z-Truck Up-and-Down Motion. Y-Truck Side-to-Side Motion. Head. Squaring Plate. Sliding Plate FIGURE 1: THE CARVEWRIGHT MACHINE

Z-Truck Up-and-Down Motion. Y-Truck Side-to-Side Motion. Head. Squaring Plate. Sliding Plate FIGURE 1: THE CARVEWRIGHT MACHINE Setup and use of CarveWright CO2 Powered Dragster Jig The CO 2 powered Dragster Jig will arrive from the factory fully assembled, calibrated, and squared. In order to get the best results, your CarveWright

More information

From the home screen, navigate to:

From the home screen, navigate to: A4 LTR LGL LTR A4 LGL Quick Reference Loading paper and specialty media This section explains how to load the 550-sheet tray, an optional,000-sheet tray, a multipurpose feeder, and an optional 550-sheet

More information

R PROFLAME Instruction Book Collection

R PROFLAME Instruction Book Collection 9.956.028 R00 584 PROFLAME Instruction Book Collection 4-17 18-29 584 PROFLAME System 30-39 Appendix: DIP SWITCH NUMBER (0=ON 1=OFF) 40-41 4-17 Fig. 1 The SIT is a device that allows, in conjunction with

More information

Quick Reference Guide

Quick Reference Guide 9-Pin Dot Matrix Printer Quick Reference Guide Contents Loading Continuous Paper With the Front Push Tractor.................. 2 Loading Continuous Paper With the Rear Push Tractor.................. 4

More information

Basic P-XRD instructions for Operating the Instrument

Basic P-XRD instructions for Operating the Instrument Basic P-XRD instructions for Operating the Instrument Instrument Parts Incident Beam Optics (left arm) 1) X-ray source (Cu) i. Rest settings: 45 kv, 20mA ii. Run settings: 45 kv, 40mA 2) Monochromator

More information

Operating Guide. HT25 Multi Side Tabber & Stamp Affixer. HASLER America s better choice. Mailing Systems And Solutions

Operating Guide. HT25 Multi Side Tabber & Stamp Affixer. HASLER America s better choice. Mailing Systems And Solutions Operating Guide Mailing Systems And Solutions HASLER America s better choice HT25 Multi Side Tabber & Stamp Affixer An ISO 9001 Quality System Certified company Rev. 8/25/2010 Please record the following

More information

Written By: Walter Galan

Written By: Walter Galan Xbox 360 CPU Heat Sink Replacement CPU heat sink replacement. Written By: Walter Galan ifixit CC BY-NC-SA www.ifixit.com Page 1 of 27 INTRODUCTION Use this guide to remove the CPU heat sink from your Xbox

More information

MOXTEK. 50kV 10 Watt MAGNUM X-ray Source. X-ray Sources. Contents

MOXTEK. 50kV 10 Watt MAGNUM X-ray Source. X-ray Sources. Contents X-ray Sources 0kV 10 Watt MAGNUM X-ray Source Manual Contents MAGNUM X-ray Source Characteristics Initial Inspection and Handling Tube Setup Operating Conditions Operating Precautions and Warnings Operating

More information

7878 K940. Checkpoint Antenna. Kit Instructions. Issue B

7878 K940. Checkpoint Antenna. Kit Instructions. Issue B 7878 K940 Checkpoint Antenna Kit Instructions Issue B Revision Record Issue Date Remarks A July 7, 2009 First issue B Nov2013 Revised the Checkpoint installation procedures for 7878 and 7874 scanners Added

More information

1. Turn off or disconnect power to unit (machine). 2. Push IN the release bar on the quick change base plate. Locking latch will pivot downward.

1. Turn off or disconnect power to unit (machine). 2. Push IN the release bar on the quick change base plate. Locking latch will pivot downward. Figure 1 Miniature Quick Change Applicators, of the end feed type, are designed to crimp end feed strip terminals to prestripped wires. Each applicator is set up to accept the strip form of certain specific

More information

7-4-2 (4) Set up mode (Play fee, BGM, Tilt sensor)

7-4-2 (4) Set up mode (Play fee, BGM, Tilt sensor) 7-4-2 (4) Set up mode (Play fee, BGM, Tilt sensor) A B Maintenance door Display A shows which mode you are in. Display B shows which content you are selecting. 1 Open the maintenance door that is locked

More information