FEI Helios NanoLab 600 TEM specimen prep recipe Nicholas G. Rudawski (352) (office) (805) (cell) Last updated: 01/19/17

Size: px
Start display at page:

Download "FEI Helios NanoLab 600 TEM specimen prep recipe Nicholas G. Rudawski (352) (office) (805) (cell) Last updated: 01/19/17"

Transcription

1 FEI Helios NanoLab 600 TEM specimen prep recipe Nicholas G. Rudawski (352) (office) (805) (cell) Last updated: 01/19/17 This recipe is based on the methods of Schaffer et al. for preparing relatively soft semiconductor-based specimens for HR-S/TEM analysis as described in detail at: However, depending on the nature of your particular specimens, substantial deviations from this recipe may be necessary including (but not limited to) changes to Pt strap thickness, pattern dimensions (Z in particular), and the method of attachment of the lamella to the grid. You should always consult the scientific literature to see if any FIB-based methods of specimen preparation have been reported for your specific materials and consult staff if you need any recommendations. 1. Grids for in-situ liftouts 1.1. Special grids are needed to perform liftouts. The most effective are made specifically by Omniprobe and are referred to as Omniprobe grids. The grids are made of Cu or Mo and have 3, 4, or 5 fingers with recessed edges to provide points of attachment for samples. In general, it is preferable to use Mo rather than Cu as a grid material. Mo is very stiff, will deform less easily than Cu, and will redeposit less material onto the lamella (though the Mo grids are more expensive with rougher surfaces) When loading a grid into the TEM row holder, it should be loaded so that Omniprobe reads forwards as viewed from the top with the row holder inserted into the loading device (clips side facing up). After loading the grid into the row holder, it should be examined edge-on using the optical microscope to make sure it is basically straight. If it appears bent, it should be removed from the row holder and discarded and another (unbent) grid loaded When loading the TEM row holder into the TEM row holder module in the UMB, orient the row holder so the clips side of the holder faces towards the computer desk (this will properly orient the grid when it is time to attach the lamella, assuming you loaded it properly in the previous step). 2. Stage adjustments for specimen 2.1. When the vacuum is sufficient, turn on the beams; make sure the I-beam is initially set to 30 kv with a low beam current (10 or 30 pa); adjustments to the I-beam voltage and current for subsequent steps will be indicated. 1

2 2.2. If the lamella needs to be prepared so it is aligned with a particular sample direction (e.g. a cleaved edge of a single crystal), first navigate over to the feature to be used for alignment in the live E-beam image and focus the image. Then select Stage from the pull-down menu and then xt Align Feature and follow the instructions; when finished, the stage will rotate to align the feature as specified Navigate to a region of interest on the sample where the lamella is to be made; bring the region of interest to eucentic height, set the stage tilt to T = 52, link the beams (making sure the I-beam scan rotation = 180 ), and set the magnification to 8000 (usually sufficient for lamella preparation). 3. Pt strap deposition 3.1. Set the I-beam to, insert the Pt GIS, and take a snapshot. Draw a rectangle pattern with dimensions X = 8.0, Y = 1.0, and Z = 2.0 µm, respectively, and set the application to Pt dep. Position the pattern as desired (it should remain close to the center of the image) and execute pattern is finished, retract the Pt GIS.. When the 2

3 4. Milling trenches 4.1. Set the I-beam to, tilt the stage to T = 47, and take an I- beam snapshot ; draw a regular cross-section pattern with X = 12.0, Y = 4.0, and Z = 2.0 µm, respectively, and set the application to Si the Pt strap and execute.. Position the pattern so the thick line is just below 4.2. Tilt the stage to T = 57, take an I-beam snapshot, and set the pattern rotation to ; position the pattern so the thick line is just above the Pt strap and execute. 3

4 5. Rough cleaning cuts 5.1. Set the I-beam to, tilt the stage back to T = 47, and take an I-beam snapshot ; set the pattern rotation to and set the pattern dimension to X = 12.0, Y = 1.0, and Z = 1.0 µm, respectively. Position the pattern so the thick line is just below the Pt strap and execute Tilt the stage to T = 57 and taken an I-beam snapshot ; set the pattern rotation to and position so the thick line is just above the Pt strap and execute µm thick after finishing this step.. The lamella should be

5 6. Undercutting and post-undercut cleaning 6.1. Set the I-beam current to, tilt the stage to T = 7, and delete the regular cross-section pattern used up to this point. Take I- beam snapshots and adjust the Y beam shift until the specimen is back within the field of view. Draw three rectangle patterns as shown in the image below; the width (X or Y dimension) of each pattern should be >0.7 µm with Z = 1.0 µm; position the patterns to there is overlap between the corner regions. Make sure the patterns are set for parallel milling and execute. 5

6 6.2. Tilt the stage to T = 57 and delete the rectangle patterns used for the undercut; take snapshot I-beam snapshots and adjust the Y beam shift for the I-beam image until the specimen is back within the field of view. Draw a cleaning cross-section pattern with dimensions X = 12.0 and Z = 1.0 µm, respectively, and manually adjust the Y dimension so it covers the wings generated from the undercut; set the pattern rotation to the Pt strap and execute.. Position the pattern so the thick line is just above 6

7 7. Lifting out the sample 7.1. Set the I-beam current to, tilt the stage to T = 0, and delete the cleaning cross-section pattern just used. Start live I-beam imaging and adjust the Y beam shift until the lamella is back within the field of view. Insert the Omniprobe and use live E- beam and I-beam imaging to approach it to the lamella; when the Omniprobe is a few µm away from the lamella, insert the Pt GIS. Position the Omniprobe so it is over the free end of the lamella as shown below. Draw a rectangle pattern with X and Y dimensions sufficient to cover the point of contact between the Omniprobe and lamella and Z = 0.2 µm and set the application to Pt dep ; take an I-beam snapshot, position the pattern so it is centered over the point of contact, and execute. 7

8 7.2. Set the I-beam to and take an I-beam snapshot. Change the application of the rectangle pattern to Si and adjust the X and Y dimensions of the pattern so it covers the remaining point of attachment of the lamella to the sample (on the left in the I-beam image) and set Z = 1.0 µm. Position the pattern over the remaining point of attachment, and then execute it. Take E- beam snapshots stop the pattern frequently to observe the progress of the pattern and once the lamella is released. 8

9 7.3. Set the I-beam to and start live I-beam imaging. Slowly lower the Z stage control to pull the specimen away from the lamella. Once the specimen is sufficiently clear of the specimen surface (a few 10s of µm), retract the Pt GIS ; then place the Omniprobe into the Park position and retract the Omniprobe. 8. Aligning the grid prior to sample attachment 8.1. Navigate over to the position on the grid where the sample is intended to be attached and bring it to eucentric position. The grid may be rotated a few degrees off from being perfectly horizontal in the E-beam image (or more, depending on if the stage was rotated to align a feature on the sample). If that is the case, select Stage from the pull-down menu and then xt Align Feature and follow the instructions to align the grid. 9

10 8.2. The grid may also be tilted slightly away from being perfectly aligned with the E-beam. Adjust the stage tilt (usually T = 1 to 2 ) until the grid appears flat and not tilted; if you have to tilt to T < 2, leave the tilt at T = 2 (the Omniprobe cannot be inserted for T < 2 ). Keep in mind now that the stage tilt corresponding normal incidence for the I-beam will be differ by this amount (e.g = 50 is now the stage tilt for normal incidence for I-beam). Before After 10

11 9. Attaching the lamella to the grid 9.1. Start live I-beam imaging and adjust the Y beam shift until the desired point of attachment is within the field of view. Insert the Omniprobe and use live E-beam and I-beam imaging to approach the lamella; when the lamella is a few µm away from the grid, insert the Pt GIS. Position the lamella so the free end is barely separated from the edge of the grid. Set the I-beam to and draw a rectangle pattern with Z = 0.2 µm and X and Y dimensions sufficient to cover the point of contact between the lamella and grid; set the application to Pt dep. Take an I-beam snapshot contact, and execute., position the pattern so it is centered over the point of 11

12 9.2. Set the I-beam to and take an I-beam snapshot. Change the application of the pattern to Si, set Z = 1.0 µm, and adjust the X and Y dimensions of the pattern so it covers the weld between the lamella and Omniprobe and extends slightly beyond the top and bottom of the Omniprobe. Position the pattern over the weld and execute. Take E-beam snapshots frequently to observe the progress of the pattern and stop the pattern released from the lamella. once the Omniprobe is 9.3. Set the I-beam to and start live I-beam imaging. Move the Omniprobe so it is a few 10s of µm clear of the lamella and then retract the Pt GIS ; then place the Omniprobe in the Park position and retract the Omniprobe If you have additional lamellas to prepare, it is best to lift out and attach each lamella to the grid and then final thin all lamellas at once. 12

13 10. Lamella adjustments before final thinning Set the I-beam to ; bring the lamella to eucentric height, then set the stage tilt to the appropriate value for normal I-beam incidence accounting for the pre-tilt to align the grid (e.g = 50 ), and link the beams Set the I-beam to and take a snapshot. If the lamella appears slightly rotated from horizontal in the I-beam image, select Stage from the pull-down menu and then Scan Rotation Align Feature and follow the instructions to horizontally align the lamella in the I-beam image (the scan rotation for the I-beam image should still be close to 180 ). If the lamella appears rotated during later steps, you can repeat this alignment process again as needed. 13

14 10.3. Draw a rectangle pattern, set Z = 1.0 µm, and adjust the X and Y dimensions of the pattern so it covers the end of the specimen where the Omniprobe was attached. Take an I-beam snapshot pattern over the end of the lamella and execute. and position the 14

15 11. Final thinning: 30 to 8 kv Tilt the stage to +7 relative to normal incidence for the I-beam and take an I-beam snapshot. Delete the rectangle pattern just used and draw a cleaning cross-section pattern with X = 7.0 and Z = 1.0 µm with Y dimension sufficient to cover the upper side of the lamella; set the pattern rotation to. Position the pattern over the upper side of the lamella and execute. 15

16 11.2. Tilt the stage to 7 relative to normal incidence for the I-beam and take an I-beam snapshot. Decrease the X dimension of the pattern to X = 6.5 µm and set the pattern rotation to. Position the pattern over the bottom side of the lamella and execute. 16

17 11.3. Set the I-beam to. Tilt the stage to +4 relative to normal incidence for the I-beam and take an I-beam snapshot. Decrease the X dimension of the pattern to X = 6.0 µm, adjust the Y dimension of the pattern to cover the upper side of the lamella, and set the pattern rotation to upper side of the lamella and execute.. Position the pattern over the 17

18 11.4. Tilt the stage to 4 relative to normal incidence for the I-beam and take an I-beam snapshot. Decrease the X dimension of the pattern to X = 5.5 µm and set the pattern rotation to. Position the pattern over the bottom side of the lamella and execute. 18

19 11.5. Set the I-beam to. Tilt the stage to +2.5 relative to normal incidence for the I-beam and take an I-beam snapshot. Decrease the X dimension of the pattern to X = 5.0 µm, adjust the Y dimension of the pattern to cover the upper side of the lamella, and set the pattern rotation to upper side of the lamella and execute.. Position the pattern over the 19

20 11.6. Tilt the stage to 2.5 relative to normal incidence for the I-beam and take an I-beam snapshot. Decrease the X dimension of the pattern to X = 4.5 µm and set the pattern rotation to. Position the pattern over the bottom side of the lamella and execute. 20

21 12. Final thinning: 5 and 2 kv Set the I-beam to. Tilt the stage to +2 relative to normal incidence for the I-beam and take an I-beam snapshot. Decrease the X dimension of the pattern to X = 4.0 µm, adjust the Y dimension of the pattern to cover the upper side of the lamella, and set the pattern rotation to lamella and execute milling is proceeding and stop the pattern reduced in thickness by ~33%.. Position the pattern over the upper side of the ; observe the lamella in the E-beam image while once the Pt strap has 21

22 12.2. Tilt the stage to 2 relative to normal incidence for the I-beam and take an I-beam snapshot. Decrease the X dimension of the pattern to X = 3.5 µm and set the pattern rotation to. Position the pattern over the bottom side of the lamella and execute ; observe the lamella in the E-beam image while milling is proceeding and stop the pattern once the Pt strap has reduced in thickness by ~50% (there should still be some Pt strap left after this step). 22

23 12.3. Set the I-beam to. Tilt the stage to +7 relative to normal incidence for the I-beam and take an I-beam snapshot. Delete the cleaning cross-section previously used and draw a rectangle pattern with X = 3.0 and Z = 1.0 µm, respectively, and Y dimension sufficient to cover the upper side of the lamella. Position the pattern over the upper side of the lamella and execute ; observe the lamella in the E-beam image while milling is proceeding and stop the pattern strap has reduced in thickness by ~50%. once the Pt 23

24 12.4. Tilt the stage to 7 relative to normal incidence for the I-beam and take an I-beam snapshot. Position the rectangle pattern used in the last step over the bottom side of the lamella and execute. As this is the last milling step, it is critical to monitor the lamella in the E-beam image during milling to safely stop the pattern before over-milling occurs. Ideally, this pattern should be stopped while there is still a very small amount of Pt strap left on the lamella (to ensure the surface is still intact and damage-free). However, if additional protective layers are present on the surface (evaporated C, for example), it may be possible (though not always necessary) to continue thinning beyond this point The sample is now finished; do not perform any further imaging of the specimen with the I-beam to prevent any additional specimen damage. Additionally, it is best to minimize E-beam imaging of the completed lamella to prevent the generation of a contamination layer prior to TEM analysis. 24

1. Preliminary sample preparation

1. Preliminary sample preparation FEI Helios NanoLab 600 standard operating procedure Nicholas G. Rudawski ngr@ufl.edu (352) 392 3077 (office) (805) 252-4916 (cell) Last updated: 03/02/18 What this document provides: an overview of basic

More information

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack). FEI Tecnai F20 S/TEM: imaging in TEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 01/21/18 1. Filling the cold trap (if needed) 1.1. Prior to use, the cold trap needs

More information

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the digital camera.

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the digital camera. FEI Tecnai F20 S/TEM: acquiring diffraction patterns Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 10/18/17 This document assumes the user is already familiar with basic

More information

Strata DB235 FESEM FIB

Strata DB235 FESEM FIB Strata DB235 FESEM FIB Standard Operating Procedure Revision: 5.0 Last Updated: August 16/2016, revised by Li Yang Overview This document will provide a detailed operation procedure of the Focused Ion

More information

Deliverable 4.2: TEM cross sections on prototyped Gated Resistors

Deliverable 4.2: TEM cross sections on prototyped Gated Resistors Deliverable 4.2: TEM cross sections on prototyped Gated Resistors Olga G. Varona, Geoff Walsh, Bernie Capraro Intel Ireland 21 June 2011 Abbreviation list D: drain FIB: focused ion-beam HRTEM: high resolution

More information

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier). JEOL 200CX operating procedure Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Specimen loading 1.1. Unlock the TUMI system. 1.2. Load specimen(s) into the holder. If using the double tilt holder, ensure

More information

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface.

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface. FEI Tecnai F20 S/TEM: imaging in STEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 05/10/18 This document assumes the user is already familiar with basic operation

More information

1.1. Log on to the TUMI system (you cannot proceed further until this is done).

1.1. Log on to the TUMI system (you cannot proceed further until this is done). FEI DB235 SEM mode operation Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Sample loading 1.1. Log on to the TUMI system (you cannot proceed further until this is done). 1.2. The FIB software (xp)

More information

The user should already be familiar with operation of the instrument in STEM mode, use of the Microscope Control interface, and TIA.

The user should already be familiar with operation of the instrument in STEM mode, use of the Microscope Control interface, and TIA. FEI Tecnai F20 S/TEM: EDS system operation Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 01/22/18 The user should already be familiar with operation of the instrument in

More information

DualBeam and FIB capability applied to metals research

DualBeam and FIB capability applied to metals research DualBeam and FIB capability applied to metals research The values of DualBeam for metals research The availability of Focused Ion Beam (FIB) capacity on a DualBeam has allowed many researchers to open

More information

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Warning Always adjust your specimen height before closing the chamber door to make sure your specimen will not hit the bottom of the lens;

More information

Basic Users Manual for Tecnai-F20 TEM

Basic Users Manual for Tecnai-F20 TEM Basic Users Manual for Tecnai-F20 TEM NB: This document contains my personal notes on the operating procedure of the Tecnai F20 and may be used as a rough guide for those new to the microscope. It may

More information

1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height.

1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height. JEOL 2010F operating procedure Covers operation in STEM mode (See separate procedures for operation in TEM mode and operation of EDS system) Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 NOTE: this operating

More information

FEI Tecnai G 2 F20 Operating Procedures

FEI Tecnai G 2 F20 Operating Procedures FEI Tecnai G 2 F20 Operating Procedures 1. Startup (1) Sign-up in the microscope log-sheet. Please ensure you have written an account number for billing. (2) Log in to the computer: Login to your account

More information

Ion Beam Lithography next generation nanofabrication

Ion Beam Lithography next generation nanofabrication Ion Beam Lithography next generation nanofabrication EFUG Bordeaux 2011 ion beams develop Lloyd Peto IBL sales manager Copyright 2011 by Raith GmbH ionline new capabilities You can now Apply an ion beam

More information

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right Quick Guide to Operating FEI Titan Themis G2 200 (S)TEM: TEM mode Susheng Tan Nanoscale Fabrication and Characterization Facility, University of Pittsburgh Office: M104/B01 Benedum Hall, 412-383-5978,

More information

FEI Falcon Direct Electron Detector. Best Practice Document

FEI Falcon Direct Electron Detector. Best Practice Document FEI Falcon Direct Electron Detector Best Practice Document 2 1. Introduction FEI Falcon Direct Electron Detector Best Practice Application Guide The FEI Falcon Detector is based on direct electron detection

More information

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope The procedures given below were written specifically for the FEI Tecnai G 2 Sphera microscope. Modifications will need to

More information

Introduction of New Products

Introduction of New Products Field Emission Electron Microscope JEM-3100F For evaluation of materials in the fields of nanoscience and nanomaterials science, TEM is required to provide resolution and analytical capabilities that can

More information

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: )

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: ) University of Minnesota College of Science and Engineering Characterization Facility PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: 2012.10.17) The following instructions

More information

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below.

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. http://staff.washington.edu/wpchan/if/cm100_inst.shtml Instructions for the Philips CM100 TEM and peripherals

More information

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Follow the procedures below when you use the Hitachi 7100 TEM. Starting Session 1. Turn on the cold

More information

Electron Microscopy Sciences

Electron Microscopy Sciences Electron Microscopy Sciences INSTRUCTIONAL MANUAL CAT. #7670 FlipScribe P.O. Box 550 s1560 Industry Road s Hatfield PA 19440 1 Overview The FlipScribe enables cleaving through frontside targets with a

More information

Transmission Electron Microscopy 9. The Instrument. Outline

Transmission Electron Microscopy 9. The Instrument. Outline Transmission Electron Microscopy 9. The Instrument EMA 6518 Spring 2009 02/25/09 Outline The Illumination System The Objective Lens and Stage Forming Diffraction Patterns and Images Alignment and Stigmation

More information

Using the Hitachi 3400-N VP-SEM

Using the Hitachi 3400-N VP-SEM Using the Hitachi 3400-N VP-SEM Opening the Chamber to Load Specimens (This may also be done later using the software) 1. Click the AIR button on the front of the machine: 2. Wait a few minutes until you

More information

PANalytical X pert Pro High Resolution Specular and Rocking Curve Scans User Manual (Version: )

PANalytical X pert Pro High Resolution Specular and Rocking Curve Scans User Manual (Version: ) University of Minnesota College of Science and Engineering Characterization Facility PANalytical X pert Pro High Resolution Specular and Rocking Curve Scans User Manual (Version: 2012.10.17) The following

More information

MSE 460 TEM Lab 4: Bright/Dark Field Imaging Operation

MSE 460 TEM Lab 4: Bright/Dark Field Imaging Operation MSE 460 TEM Lab 4: Bright/Dark Field Imaging Operation Last updated on 1/8/2018 Jinsong Wu, jinsong-wu@northwestern.edu Aims: The aim of this lab is to familiarize you with bright/dark field imaging operation.

More information

1. Specimen Holder Removal, Loading, and Insertion

1. Specimen Holder Removal, Loading, and Insertion OPERATION OF THE PHILIPS CM-200 FEG-TEM When not in use, the CM-200 should be in the MICROSCOPE ON configuration with the HIGH TENSION ON (illuminates green when the high tension is on).. The microscope

More information

2. Raise HT to 200kVby following the procedure explained in 1.6.

2. Raise HT to 200kVby following the procedure explained in 1.6. JEOL 2100 MANUAL Quick check list 1. If needed, fill the reservoir with LN2 2. Raise HT to 200kVby following the procedure explained in 1.6. 3. Insert specimen holder into TEM (Insert holder in airlock,

More information

Protective Equipment Nitrile gloves for handling sample holder and safety glasses for filling liquid nitrogen dewar.

Protective Equipment Nitrile gloves for handling sample holder and safety glasses for filling liquid nitrogen dewar. Emergency Information: 1. Medical Emergencies: Contact 911 and McGill Security 514.398.3000 2. Leave TEM as is. Do NOT shut down the vacuum system. 3. If possible, turn off High Tension and Close Column

More information

User Operation of JEOL 1200 EX II

User Operation of JEOL 1200 EX II **Log onto Computer** Open item program Start Up Procedure User Operation of JEOL 1200 EX II 1. If scope is not running, locate an electron microscopy technician (EMT) to find out why not. 2. Turn up brightness

More information

STRUCTURE OF THE MICROSCOPE

STRUCTURE OF THE MICROSCOPE STRUCTURE OF THE MICROSCOPE Use the word list to label the microscope below: Light Source Coarse adjustment knob Diaphragm Stage Clips Objectives Fine Adjustment Knob Base Stage Stage Clips Arm Revolving

More information

TEM SAMPLE-PREPARATION PROCEDURES FOR THIN-FILM MATERIALS

TEM SAMPLE-PREPARATION PROCEDURES FOR THIN-FILM MATERIALS TEM SAMPLE-PREPARATION PROCEDURES FOR THIN-FILM MATERIALS Initial Set-Up: Heat up a hot plate to around 150-200 C Plan view Mounting/Grinding/Dimpling/Polishing: 1) Cleave a square-ish piece of sample.

More information

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center LEO 912 TEM Short Manual Prepared/copyrighted by RH Berg Danforth Plant Science Center Specimen holder [1] Never touch the holder (outside of the O-ring, double-headed arrow) because finger oils will contaminate

More information

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope Last updated on 1/8/2018 Jinsong Wu, jinsong-wu@northwestern.edu Aims: The aim of this lab is to familiarize you with basic TEM alignment

More information

Bruker Dimension Icon AFM Quick User s Guide

Bruker Dimension Icon AFM Quick User s Guide Bruker Dimension Icon AFM Quick User s Guide March 3, 2015 GLA Contacts Jingjing Jiang (jjiang2@caltech.edu 626-616-6357) Xinghao Zhou (xzzhou@caltech.edu 626-375-0855) Bruker Tech Support (AFMSupport@bruker-nano.com

More information

Instructions for Tecnai a brief start up manual

Instructions for Tecnai a brief start up manual Instructions for Tecnai a brief start up manual Version 3.0, 8.12.2015 Manual of Tecnai 12 transmission electron microscope located at Aalto University's Nanomicroscopy Center. More information of Nanomicroscopy

More information

NANO 703-Notes. Chapter 9-The Instrument

NANO 703-Notes. Chapter 9-The Instrument 1 Chapter 9-The Instrument Illumination (condenser) system Before (above) the sample, the purpose of electron lenses is to form the beam/probe that will illuminate the sample. Our electron source is macroscopic

More information

2. Refraction and Reflection

2. Refraction and Reflection 2. Refraction and Reflection In this lab we will observe the displacement of a light beam by a parallel plate due to refraction. We will determine the refractive index of some liquids from the incident

More information

JEOL JEM-1400 Transmission Electron Microscope Operating Instructions

JEOL JEM-1400 Transmission Electron Microscope Operating Instructions JEOL JEM-1400 Transmission Electron Microscope Operating Instructions Anti-contamination device Objective aperture Objective aperture translation knobs Specimen holder Pump/air switch Left hand control

More information

(12) Patent Application Publication (10) Pub. No.: US 2011/ A1

(12) Patent Application Publication (10) Pub. No.: US 2011/ A1 (19) United States US 20110017922A1 (12) Patent Application Publication (10) Pub. No.: US 2011/0017922 A1 Amador (43) Pub. Date: Jan. 27, 2011 (54) VARIABLE-TILTTEM SPECIMENHOLDER (52) U.S. Cl.... 250/442.11

More information

SCIENTIFIC INSTRUMENT NEWS. Introduction. Design of the FlexSEM 1000

SCIENTIFIC INSTRUMENT NEWS. Introduction. Design of the FlexSEM 1000 SCIENTIFIC INSTRUMENT NEWS 2017 Vol. 9 SEPTEMBER Technical magazine of Electron Microscope and Analytical Instruments. Technical Explanation The FlexSEM 1000: A Scanning Electron Microscope Specializing

More information

Advanced Technology and Manufacturing Institute. Zygo ZeScope

Advanced Technology and Manufacturing Institute. Zygo ZeScope Advanced Technology and Manufacturing Institute Zygo ZeScope Created by Andrew Miller ATAMI Oregon State University Revision Date Description Curator 0 8/31/2018 New Document Andrew Miller Zygo ZeScope

More information

MICROSCOPE LAB. Resolving Power How well specimen detail is preserved during the magnifying process.

MICROSCOPE LAB. Resolving Power How well specimen detail is preserved during the magnifying process. AP BIOLOGY Cells ACTIVITY #2 MICROSCOPE LAB OBJECTIVES 1. Demonstrate proper care and use of a compound microscope. 2. Identify the parts of the microscope and describe the function of each part. 3. Compare

More information

Setting up and Using Digital Micrometer Controlled Lapping Fixtures

Setting up and Using Digital Micrometer Controlled Lapping Fixtures Setting up and Using Digital Micrometer Controlled Lapping Fixtures Purpose polishing fixtures are commonly used in materials preparation labs around the world. Lapping fixtures provide stability, precision,

More information

JEOL 6500 User Manual

JEOL 6500 User Manual LOG IN to your session on the computer to the left of the microscope. Starting Conditions 1. Press Ctrl-Alt-Del and log on to the microscope computer. Click on JEOL PC SEM 6500 icon. Click yes if message

More information

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual Scanning Electron Microscope FEI INSPECT F50 Step by step operation manual Scanning Electron Microscope, FEI Inspect F50 FE-SEM-F Observation Flow Saving Data And Analysis Specimen preparation Error check

More information

Standard Operating Procedure

Standard Operating Procedure Standard Operating Procedure Nanosurf Atomic Force Microscopy Operation Facility NCCRD Nanotechnology Center for Collaborative Research and Development Department of Chemistry and Engineering Physics The

More information

University of Washington Molecular Analysis Facility

University of Washington Molecular Analysis Facility University of Washington Molecular Analysis Facility Apreo-S (Variable Pressure) is a Schottky Field Emission Scanning Electron Microscope (FESEM) that combines high- and low-voltage ultra-high resolution

More information

Triple Beam FIB-SEM-Ar(Xe) Combined System NX2000

Triple Beam FIB-SEM-Ar(Xe) Combined System NX2000 SCIENTIFIC INSTRUMENT NEWS 2017 Vol. 8 M A R C H Technical magazine of Electron Microscope and Analytical Instruments. Technical Explanation Triple Beam FIB-SEM-Ar(Xe) Combined System NX2000 Masahiro Kiyohara

More information

RAITH e-line OPERATING INSTRUCTIONS

RAITH e-line OPERATING INSTRUCTIONS RAITH e-line OPERATING INSTRUCTIONS 1) LOADING A SAMPLE a. Start the system i. On the Column PC (Right side monitor [R]), select the SmartSEM icon to on the desktop to begin the column software. ii. On

More information

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+)

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+) Standard Operating Procedure of Atomic Force Microscope (Anasys afm+) The Anasys Instruments afm+ system incorporates an Atomic Force Microscope which can scan the sample in the contact mode and generate

More information

Unit 5: Designing with Data (Safer Buildings) Contents

Unit 5: Designing with Data (Safer Buildings) Contents Contents 1. Working with Balsa Wood and Foam Core. 1 a. Cutting & Safety 1 b. Joining 2 c. Aligning & Squaring... 3 2. Building the Base 3 3. Building the Roof Load Holder.. 4 1. Working with Balsa Wood

More information

ML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X

ML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X JAPAN DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30 (a) Field Iris Control Lever (c) Filter Slots EYEPIECES, KHW10X ANALYZER CONTROL LEVER (b) Aperture Iris Control Lever LIGHT SOURCE HOUSING VERTICAL

More information

UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY

UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY References: http://virlab.virginia.edu/vl/home.htm (University of Virginia virtual lab. Click on the AFM link) An atomic force

More information

CAD tutorial for the drinking straw support

CAD tutorial for the drinking straw support CAD tutorial for the drinking straw support Having tried a number of different designs, this one worked best on the greatest variety of glasses straight sided glass, angled glass and even a champagne flute.

More information

Match the microscope structures given in the left column with the statements in the right column that identify or describe them.

Match the microscope structures given in the left column with the statements in the right column that identify or describe them. 49 Prelab for Name Match the microscope structures given in the left column with the statements in the right column that identify or describe them. Key: a. coarse adjustment knob f. turret or nosepiece

More information

FEI Quanta 200 ESEM Basic instructions

FEI Quanta 200 ESEM Basic instructions FEI Quanta 200 ESEM Basic instructions Desktop and then start the UI. If the computer has restarted and you need to login, Username: supervisor and Password: supervisor Log-in to the Microscope using the

More information

Basic P-XRD instructions for Operating the Instrument

Basic P-XRD instructions for Operating the Instrument Basic P-XRD instructions for Operating the Instrument Instrument Parts Incident Beam Optics (left arm) 1) X-ray source (Cu) i. Rest settings: 45 kv, 20mA ii. Run settings: 45 kv, 40mA 2) Monochromator

More information

STEM Spectrum Imaging Tutorial

STEM Spectrum Imaging Tutorial STEM Spectrum Imaging Tutorial Gatan, Inc. 5933 Coronado Lane, Pleasanton, CA 94588 Tel: (925) 463-0200 Fax: (925) 463-0204 April 2001 Contents 1 Introduction 1.1 What is Spectrum Imaging? 2 Hardware 3

More information

E N G L I S H GARDEN SHED. Assembly Instructions. Suitable for Models WITH VARYING DEPTHS

E N G L I S H GARDEN SHED. Assembly Instructions. Suitable for Models WITH VARYING DEPTHS GARDEN SHED Assembly Instructions Suitable for Models 6' Wide 8' Wide 0' Wide WITH VARYING DEPTHS GI0003 November 0 INSTALLATION ADVICE It's Not That Difficult! The construction of your shed isn't as complicated

More information

Instructions XRD. 1 Choose your setup , Sami Suihkonen. General issues

Instructions XRD. 1 Choose your setup , Sami Suihkonen. General issues Instructions XRD 28.10.2016, Sami Suihkonen General issues Be very gentle when closing the doors Always use Cu attenuator when count rate exceeds 500 000 c/s Do not over tighten optical modules or attach

More information

Intro to One Point Perspective

Intro to One Point Perspective Intro to One Point Perspective Horizon Line - The horizon line in perspective drawing is a horizontal line across the picture. It is always at eye level - its placement determines where we seem to be looking

More information

2014 HTD-E with options

2014 HTD-E with options with options The HT7700 : a user-friendly, ergonomic digital TEM with options User-Friendly r end Design Ambient light operation. Multiple automated functions for alignment, focus and stigmation as standard

More information

JEOL 6700 User Manual 05/18/2009

JEOL 6700 User Manual 05/18/2009 JEOL 6700 User Manual 05/18/2009 LOG IN to your session on the computer to the right of the microscope. Starting Conditions 1. Click the button and read the Penning Gauge to ensure that the microscope

More information

Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE

Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE Prior to lab: 1) Read these instructions (p 1-6) 2) Go through the online tutorial, the microscopy pre-lab

More information

This procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface.

This procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface. Tescan MIRA3 SEM: EDS using EDAX TEAM Nicholas G. Rudawski ngr@ufl.edu Cell: (805) 252-4916 Office: (352) 392-3077 Last updated: 12/04/17 This procedure assumes the user is already familiar with basic

More information

ConstructDXF DXF Drawing Requirements User Guide <Virtual Environment> 6.0

ConstructDXF DXF Drawing Requirements User Guide <Virtual Environment> 6.0 ConstructDXF DXF Drawing Requirements User Guide 6.0 Page 1 of 8 Contents 1 Introduction... 3 1.1 Purpose of this Guide... 3 1.2 What is ConstructDXF?... 3 2 Preparation of the DXF

More information

IN SITU BIASING & HEATING SOLUTIONS FOR TEM PLATFORMS

IN SITU BIASING & HEATING SOLUTIONS FOR TEM PLATFORMS IN SITU BIASING & HEATING SOLUTIONS FOR TEM PLATFORMS Solutions for In Situ Microscopy THE LIGHTNING SERIES LIGHTNING FEATURES Observe the real-time dynamics of materials under electrical and heating stimuli.

More information

Development of JEM-2800 High Throughput Electron Microscope

Development of JEM-2800 High Throughput Electron Microscope Development of JEM-2800 High Throughput Electron Microscope Mitsuhide Matsushita, Shuji Kawai, Takeshi Iwama, Katsuhiro Tanaka, Toshiko Kuba and Noriaki Endo EM Business Unit, JEOL Ltd. Electron Optics

More information

Bruker Dimension Icon AFM Quick User s Guide

Bruker Dimension Icon AFM Quick User s Guide Bruker Dimension Icon AFM Quick User s Guide August 8 2014 GLA Contacts Jingjing Jiang (jjiang2@caltech.edu 626-616-6357) Xinghao Zhou (xzzhou@caltech.edu 626-375-0855) Bruker Tech Support (AFMSupport@bruker-nano.com

More information

PROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, Edited by

PROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, Edited by X - R A Y M I C R O S C O P Y A N D M I C R O R A D I O G R A P H Y PROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, 1956 Edited by V. E. COSSLETT Cavendish Laboratory, University

More information

2 How to operate the microscope/obtain an image

2 How to operate the microscope/obtain an image Morgagni Operating Instructions 50079 010912 2-1 2 ow to operate the microscope/obtain an image 2.1 Starting the microscope 2.1.1 Starting the microscope with several manually-operated steps 1. Turn on

More information

Video. Part I. Equipment

Video. Part I. Equipment 1 of 7 11/8/2013 11:32 AM There are two parts to this lab that can be done in either order. In Part I you will study the Laws of Reflection and Refraction, measure the index of refraction of glass and

More information

Model SU3500 Scanning Electron Microscope

Model SU3500 Scanning Electron Microscope Model SU3500 Scanning Electron Microscope Modified and Parts taken from Hitachi Easy Operation Guide. Before using the Model SU3500 SEM, be sure to read the [GENERAL SAFETY GUIDELINES] in the instruction

More information

Ox-RAC-08 Ribbon Angled Fiber Cleaver User Manual

Ox-RAC-08 Ribbon Angled Fiber Cleaver User Manual Ox-RAC-08 Ribbon Angled Fiber Cleaver User Manual Issue 2.0 Contents Introduction... 2 Contents of Cleaving Kit & Unpacking... 3 Cleaving Problems... 8 Blade damage:... 9 Cleaver Maintenance... 10 Cleaning

More information

DISCO DICING SAW SOP. April 2014 INTRODUCTION

DISCO DICING SAW SOP. April 2014 INTRODUCTION DISCO DICING SAW SOP April 2014 INTRODUCTION The DISCO Dicing saw is an essential piece of equipment that allows cleanroom users to divide up their processed wafers into individual chips. The dicing saw

More information

Outline. 1 Introduction. 2 Basic IC fabrication processes. 3 Fabrication techniques for MEMS. 4 Applications. 5 Mechanics issues on MEMS MDL NTHU

Outline. 1 Introduction. 2 Basic IC fabrication processes. 3 Fabrication techniques for MEMS. 4 Applications. 5 Mechanics issues on MEMS MDL NTHU Outline 1 Introduction 2 Basic IC fabrication processes 3 Fabrication techniques for MEMS 4 Applications 5 Mechanics issues on MEMS 2.2 Lithography Reading: Runyan Chap. 5, or 莊達人 Chap. 7, or Wolf and

More information

Training CAD/ Part Designer: Designing with Angled Parts

Training CAD/ Part Designer: Designing with Angled Parts Training CAD/ Part Designer: Designing with Angled Parts We have attempted to keep the content of this document complete, accurate and under permanent review. However, due to the continuous development,

More information

STEM alignment procedures

STEM alignment procedures STEM alignment procedures Step 1. ASID alignment mode 1. Write down STD for TEM, and then open the ASID control window from dialogue. Also, start Simple imager viewer program on the Desktop. 2. Click on

More information

COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual

COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual 1 COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual ABOUT THIS MICROSCOPE (room HG01.240) The JEM-1400Flash

More information

Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation

Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation 238 Hitachi Review Vol. 65 (2016), No. 7 Featured Articles Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation AFM5500M Scanning Probe Microscope Satoshi Hasumura

More information

CHA ELECTRON BEAM EVAPORATOR SYSTEM OPERATING MANUAL Version: 2.0 August 2016

CHA ELECTRON BEAM EVAPORATOR SYSTEM OPERATING MANUAL Version: 2.0 August 2016 CHA ELECTRON BEAM EVAPORATOR SYSTEM OPERATING MANUAL Version: 2.0 August 2016 UNIVERSITY OF TEXAS AT ARLINGTON Nanotechnology Research Center TABLE OF CONTENTS 1.0 DESCRIPTION.... 3 2.0 SAFETY....... 3

More information

Advanced Optical Communications Prof. R. K. Shevgaonkar Department of Electrical Engineering Indian Institute of Technology, Bombay

Advanced Optical Communications Prof. R. K. Shevgaonkar Department of Electrical Engineering Indian Institute of Technology, Bombay Advanced Optical Communications Prof. R. K. Shevgaonkar Department of Electrical Engineering Indian Institute of Technology, Bombay Lecture No. # 39 Laboratory Experiment - 1 Let us now conduct some experiments

More information

Tool 3-1 A2 Bushing 6 Check if the bushing supports (angle iron) are properly welded with the upper part of the structure pipes. welded bushing suppor

Tool 3-1 A2 Bushing 6 Check if the bushing supports (angle iron) are properly welded with the upper part of the structure pipes. welded bushing suppor Rope Pump MANUFACTURING checklist for quality control A Welding General 1 Check if welding jigs are used for welding of the main parts (wheel, structure frame and bushings) 2 Check if the pump parts are

More information

PhE102-VASE. PHE102 Variable Angle Spectroscopic Ellipsometer. Angstrom Advanced Inc. Angstrom Advanced. Angstrom Advanced

PhE102-VASE. PHE102 Variable Angle Spectroscopic Ellipsometer. Angstrom Advanced Inc. Angstrom Advanced. Angstrom Advanced Angstrom Advanced PhE102-VASE PHE102 Variable Angle Spectroscopic Ellipsometer Angstrom Advanced Instruments for Thin Film and Semiconductor Applications sales@angstromadvanced.com www.angstromadvanced.com

More information

High Rise Sit-Stand Desk Converter

High Rise Sit-Stand Desk Converter High Rise Sit-Stand Desk Converter Assembly Instructions for Model DC300 Patent Pending PRE-ASSEMBLY Please read all instructions before beginning assembly. We strongly recommend you watch the video at

More information

Radial Polarization Converter With LC Driver USER MANUAL

Radial Polarization Converter With LC Driver USER MANUAL ARCoptix Radial Polarization Converter With LC Driver USER MANUAL Arcoptix S.A Ch. Trois-portes 18 2000 Neuchâtel Switzerland Mail: info@arcoptix.com Tel: ++41 32 731 04 66 Principle of the radial polarization

More information

The microscope is useful in making observations and collecting data in scientific experiments. Microscopy involves three basic concepts:

The microscope is useful in making observations and collecting data in scientific experiments. Microscopy involves three basic concepts: AP BIOLOGY Chapter 6 NAME DATE Block MICROSCOPE LAB PART I: COMPOUND MICROSCOPE OBJECTIVES: After completing this exercise you should be able to: Demonstrate proper care and use of a compound microscope.

More information

Celerity Fiber Termination Kit

Celerity Fiber Termination Kit Celerity Fiber Termination Kit User Guide Product Overview The Celerity Fiber Termination Kit (CT-FTK) is a professional tool set for splicing and terminating fiber optic cables in the field or at an assembly

More information

Choosing the Right Accelerating Voltage for SEM (An Introduction for Beginners)

Choosing the Right Accelerating Voltage for SEM (An Introduction for Beginners) Microscopy101 Choosing the Right Accelerating Voltage for SEM (An Introduction for Beginners) V.M. Dusevich*, J.H. Purk, and J.D. Eick University of Missouri Kansas City, School of Dentistry, 650 E. 25

More information

Supporting Information. for. Visualization of Electrode-Electrolyte Interfaces in LiPF 6 /EC/DEC Electrolyte for Lithium Ion Batteries via In-Situ TEM

Supporting Information. for. Visualization of Electrode-Electrolyte Interfaces in LiPF 6 /EC/DEC Electrolyte for Lithium Ion Batteries via In-Situ TEM Supporting Information for Visualization of Electrode-Electrolyte Interfaces in LiPF 6 /EC/DEC Electrolyte for Lithium Ion Batteries via In-Situ TEM Zhiyuan Zeng 1, Wen-I Liang 1,2, Hong-Gang Liao, 1 Huolin

More information

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. Smith College August 2005 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check, 1 Specimen Insertion, 1 Startup, 2 Filament Saturation, 2 Beam Alignment,

More information

Phase plates for cryo-em

Phase plates for cryo-em Max Planck Institute of Biochemistry Martinsried, Germany MAX PLANCK SOCIETY Phase plates for cryo-em Rado Danev Max Planck Institute of Biochemistry, Martinsried, Germany. EMBO course 2017, London, UK

More information

Z-Truck Up-and-Down Motion. Y-Truck Side-to-Side Motion. Head. Squaring Plate. Sliding Plate FIGURE 1: THE CARVEWRIGHT MACHINE

Z-Truck Up-and-Down Motion. Y-Truck Side-to-Side Motion. Head. Squaring Plate. Sliding Plate FIGURE 1: THE CARVEWRIGHT MACHINE Setup and use of CarveWright CO2 Powered Dragster Jig The CO 2 powered Dragster Jig will arrive from the factory fully assembled, calibrated, and squared. In order to get the best results, your CarveWright

More information

Supplementary Information

Supplementary Information Supplementary Information Supplementary Figure 1 Two blue-rayed limpets in their natural habitat on the stipe of a macroalgae. The surrounding light-yellow regions show exposed algal tissue eaten away

More information

Log Hauler. Senior Design Project. Department of Mechanical Engineering Trevor Kline 4/19/2011

Log Hauler. Senior Design Project. Department of Mechanical Engineering Trevor Kline 4/19/2011 Log Hauler Senior Design Project Department of Mechanical Engineering Trevor Kline 4/19/2011 Table of Contents BACKGROUND:... 3 INSPIRATION:... 3 PROPOSAL AND REQUIREMENTS:... 4 DESIGN:... 4 LOG SIZE:...

More information

Tensioning Instructions

Tensioning Instructions Tensioning Instructions S I G N E X T R U S I O N S A N D S Y S T E M S S I G N E X T R U S I O N S A N D S Y S T E M S Before beginning, thoroughly read all tensioning instructions. Tools Required: 1.

More information

Aeon ALR Series. EDGE FREE Ambient Light Rejecting Fixed Frame Screen USER S GUIDE

Aeon ALR Series. EDGE FREE Ambient Light Rejecting Fixed Frame Screen USER S GUIDE Aeon ALR Series EDGE FREE Ambient Light Rejecting Fixed Frame Screen USER S GUIDE Product Description: The Aeon ALR Series is a fixed frame projection screen with an EDGE FREE design The EDGE FREE design

More information

Christian Boit TUB Berlin University of Technology Sect. Semiconductor Devices. 1

Christian Boit TUB Berlin University of Technology Sect. Semiconductor Devices. 1 Semiconductor Device & Analysis Center Berlin University of Technology Christian Boit TUB Berlin University of Technology Sect. Semiconductor Devices Christian.Boit@TU-Berlin.DE 1 Semiconductor Device

More information