1 X-ray Sources 0kV 10 Watt MAGNUM X-ray Source Manual Contents MAGNUM X-ray Source Characteristics Initial Inspection and Handling Tube Setup Operating Conditions Operating Precautions and Warnings Operating a MAGNUM without a Moxtek FTC-00 Controller Operating a MAGNUM with a Moxtek FTC-00 Controller 4 1
2 MAGNUM 10 Watt X-ray Source Characteristics The MAGNUM x-ray tube is small, lightweight, and can be packaged into custom enclosures. The MAGNUM x-ray tube package includes an x-ray tube and a high voltage power supply that operates at up to 0 kv and 00 μa. The MAGNUM tube is ideal for battery-powered applications. The tube anode is grounded allowing placement close to the sample. Mechanical Specifications X-ray Tube Characteristics Tube Type: Cathode Type: Operating Temperature: Storage Temperature: Cooling: HV Insulation: Weight: Available Targets: Metal-ceramic Tungsten filament -10 to +0 C -40 to +60 C Air Silicone potting 1.1 lb. (00g) Ag, W, Pd, Rh, Ta, Au HV Polarity: High Voltage: Beam Current: Maximum Power: Window: Input Power: Warranty: Grounded anode -10 to -0 kv 0 to 0.0 ma 10 watts Beryllium, 0. mm thick 0W max (9-1 VDC) One year Drawing 1 MAGNUM X-ray source dimensions (in inches [millimeters])
3 Initial Inspection When a MAGNUM tube is received, it should be unpacked and inspected as soon as possible. A standard MAGNUM tube is packaged in a brass shield. This shield is connected to a high voltage power supply via two white high voltage wires and one green ground wire (Figure 1). Inspect the high voltage power supply and the tube shield assembly for any damage that may have occurred during shipping. If a tube has been damaged, please contact Moxek immediately. The serial number is located on the high voltage power supply. Please do not touch the beryllium window. Do no pull on or bend the high voltage wires. High Voltage Wire High Voltage Power Supply Grounded Wire Max anode shield temperature= 100ºC Source and Shield Assembly Handling Figure 1 MAGNUM Source Figure MAGNUM Source Temperature Care should be taken when handling the tube and the high voltage power supply. Lift the power supply and the tube with both hands to avoid damaging the high voltage wires or potting material. Prevent the tube from striking any surface. Care must be taken to not damage the beryllium window or high voltage wires. Table 1 Pin Assignment Pin Assignment for Ribbon Wire DB9 Connector Function Pin # I/O Value Response Input power N/A +6 to +1 V Input power Input power 1 +6 to +1 V Input power Ground Ground Ground HV control +0. V to +4.0 V -10 to 0 kv HV enable V or + VTTL input Current Monitor 0.0 V or + VTTL 0 Disabled, + Enabled -4 kv to -0 kv Ground 6 Ground 0 Not ready + Ready Current Control Filament ready V or +4 V 0 Disabled, + Enabled V to + V HV monitor V to +4 V Proportional to -4 kv to -0 kv Proportional to 0 to 00 µa Manufacturer: M Manufacturer: OKI Electrical Cable Part Number: Part Number: R000-0-ND Figure : Ribbon Cable (ASM006) Manufacturer: BELDEN Part Number: R16-00 Manufacturer: MOLEX Part Number: Manufacturer: MOLEX Part Number: Figure 4: -Pin Secondary Cable (ASM0064)
4 Source Setup Mechanical The shielded MAGNUM tube assembly may be mounted with the exit collimator facing any direction. Care should be taken not to strain the high voltage wires. Minimum high voltage wire bend radius is 1. inches. The high voltage power supply has three threaded holes on both sides of the metal case. These holes may be used for mounting (see Drawing 1). Cooling The MAGNUM tube does not require forced convection or liquid cooling. The tube should be operated in an environment that allows free convection or conduction to secondary parts. Please refer to page of this manual for ambient and storage temperature specifications. The maximum ambient temperature, as measured at the surface of the power supply enclosure, is +10 C. The maximum operating anode shield temperature is 100 degrees Celsius (see Figure ). Electrical Output power below 4W: Moxtek recommends using the -pin optional secondary cable (ASM0064) while operating the 0kV Magnum Tube (Figure 4). The pin assignment is shown in Drawing 1 and in Table 1. Output power above 4W: -pin secondary cable is required for tube and stability (Figure 4). Attach the pin secondary power cable to the backside of the high voltage power supply case. Refer to Figure and Table 1 for pin identification and electrical specifications. Attach the bare wire leads to a 6 to 10VDC power supply or to a Moxtek FTC-00 x-ray tube controller. The high voltage power supply also has a 10 pin IDC connector. This connector is used to power, control, and monitor the x-ray tube (Figure 6). A 10 lead ribbon cable is available from Moxtek (Figure ). This ribbon cable has a 10 pin IDC connector on one side that connects to the high voltage power supply and a DB9 connector on the other end that connects to the Moxtek FTC-00 controller. The pin assignment for the 10 pin IDC is shown in Drawing 1 and the pin-out for the DB9 connector is shown in Table 1. Safety To assure safe tube operation, a volt TTL signal is sent to the power supply to enable x-ray generation. Moxtek recommends the use of a safety interlock switch when operating the MAGNUM x-ray tube. Connect the high voltage enable pin (see Table 1) to the customer supplied safety interlock switch Figure 6: 10 Pin ICD Connector 1 Moxtek PN: ASM0064 Length: 6ft ± inches Figure : Pin Secondary Power Connector 4
5 Operating Conditions MAGNUM tube anodes are grounded to a metal shield (Figure 1). The shield is grounded to the power supply via a green wire. Make sure that the green ground wire is securely connected from the tube shield to the grounding terminal on the high voltage power supply case. DO NOT disconnect the green ground wire from the tube shield or the high voltage power supply. It will damage the tube and/ or power supply if the ground cable is loose or damaged; do not operate the tube until the ground cable is secured or replaced. Isolate all grounds for the x-ray tube. When operating the 0kV MAGNUM x-ray tube, adhere to the maximum setting: 0 kv, 0.0 ma max. Failure to adhere to these limits may cause damage the x-ray tube and/or high voltage power supply. Failure to adhere to this parameter will forfeit the tube warranty. When operating the MAGNUM x-ray tube, wait seconds after the tube has been powered off before powering the tube on again. Failure to wait seconds may damage the filament. Failure to adhere to this parameter will forfeit the tube warranty. Operating Precautions and Warnings CAUTION: Verify that the tube and the high voltage power supply are properly grounded before powering on the FTC-00 tube controller. Also verify that the FTC-00 controller is properly grounded to the power outlet. If you are not using the FTC-00 Controller and using a custom controller, make sure the tube and the high voltage power supply are properly grounded. CAUTION: MAGNUM X-ray tubes contain Beryllium. Inhaling Beryllium dust may cause lung disease refer to the MSDS. Under normal use, Beryllium dust is not generated. If the Beryllium window is broken, a very small amount of dust might be generated. If this occurs follow the instructions on the MSDS for cleanup. WARNING: MAGNUM X-ray tubes may become very hot during operation. WARNING: MAGNUM X-ray tubes produce x-ray radiation. MAGNUM tubes are shielded with a metal shield and high-z potting materials. Extra shielding may be required depending on the application. ONLY OPERATE X-RAY TUBES IN PROPERLY SHIELDED ENCLOSURES. It is the responsibility of the operator to ensure that all applicable safety precautions are taken and observed. WARNING: MAGNUM X-ray tubes operate at high voltages. Refer to tube handling instructions on page. Precautions should be taken to protect the operator while applying high voltages to avoid serious injury or death. Operating a MAGNUM Source without a Moxtek FTC-00 Controller The high voltage power supply uses DC voltages and signals to operate the tube. To operate the tube without using a Moxtek FTC- 00 controller, refer to Drawing 1 for the 10 pin high voltage power supply connector pinouts. Operating the MAGNUM X-ray Source with a Moxtek FTC-00 Controller The FTC-00 controller is designed to power, monitor, and control the MAGNUM x-ray tube. This controller contains all of the necessary electronics to operate the high voltage power supply. The FTC-00 controller provides input power to the high voltage power supply. The FTC-00 controls the high voltage and emission current settings on the x-ray tube. The FTC-00 controller includes built-in meters for convenient monitoring and display of the high voltage and emission current. Refer to the FTC-00 Operation Manual for operating instructions. 4 West 160 North / Orem, UT 8407 Phone / Fax info@moxtek.com
DXM 300W-1200W X-RAY GENERATOR MODULE REV. AB
PAGE 1 OF 5 Compact & Lightweight Models from 20kV-160kV, 300W, 600W and 1200W Universal Input, Power Factor Corrected Hot Anode or Hot Cathode X-Ray Tube Capable Standard Digital Interfaces:, Ethernet
More informationMEDIUM-MU AIR-COOLED POWER TRIODE 3CX15,000H3
TECHNICAL DATA MEDIUM-MU AIR-COOLED POWER TRIODE 3CX15,000H3 The EIMAC 3CX15,000H3 is an air cooled, ceramic-metal, medium-mu power triode designed primarily for use in broadcast and industrial radio-frequency
More informationSeries 48 Water Cooled Laser & UC-2000 Quick Start Guide
Important Read all Danger, Warning, Caution terms, symbols, and instructions located in the (Laser Safety Hazard information) sections in the Series 48 Laser Operation Manuals. http://www.synrad.com/synrad/docroot/resources/libraries/manuals
More informationAWOS Owners / Installation / Operation Manual
VAL AVIONICS LTD AWOS 2000 VHF Transmitter Owners / Installation / Operation Manual Revision 7 February 2013 P/N 172210 Table of Contents GENERAL INFORMATION...2 SPECIFICATIONS...3 OPERATION...4 INSTALLATION...5
More informationFirestar vi30/40 Dual Cooled Laser Quick Start Guide
Important See the Firestar vi30/40 Operators Manual for complete installation details and instructions. A PDF version is available online at: http://www.synrad.com/synrad/docroot/resources/libraries/manuals.
More informationHigh Frequency Sinewave Guardian TM Filter
High Frequency Sinewave Guardian TM Filter 380V 480V TECHNICAL REFERENCE MANUAL FORM: SHF-TRM-E REL. April 2015 REV. 001 2015 MTE Corporation Caution Prior to start up; confirm the drive operation mode
More informationInstallation, Operation and Maintenance Manual
TM M723 Rev. C Model 1050 Strain Gage Transmitter Installation, Operation and Maintenance Manual Table of Contents General... 1 Installation... 1 Specifications... 1 Wiring... 2 Setup and Operation...
More information2016 Catalog Oxford Instruments X-Ray Technology
2016 Catalog Oxford Instruments X-Ray Technology Power Supplies, Integrated Sources and X-ray Tubes Shielded X-ray Tubes and Power Supplies Radiation Shielded X-ray Tubes Microfocus X-ray Sources Glass
More informationOEM s #1 Choice for X-Ray Sources and Generators. Focus on Performance, Reliability, Form Factor, and Versatility.
OEM s #1 Choice for X-Ray Sources and Generators Focus on Performance, Reliability, Form Factor, and Versatility www.vjtxray.com Applications Security, Industrial NDT, Food & Pharmaceutical, Quality &
More informationService Instructions. The Conductor Controls. Conductor DC15-A, Enclosed Unit CH15-A, Open Chasis Unit
Service Instructions The Conductor Controls Conductor DC15-A, Enclosed Unit CH15-A, Open Chasis Unit Table of Contents General Section Page Safety Instructions.. 4 Introduction 5 Inspection and Long-Term
More informationExternal RF Driver. Electrical Driver for AOTF-crystal. Instruction Manual
External RF Driver Electrical Driver for AOTF-crystal Instruction Manual Table of Contents 1 General... 3 2 Compliance... 3 3 Labeling... 4 4 Interface... 5 4.1 Front Panel... 5 4.2 Back Panel... 6 4.2.1
More informationUSER'S MANUAL DMR-6700
USER'S MANUAL Multimeter True RMS DMR-6700 CIRCUIT-TEST ELECTRONICS www.circuittest.com Introduction This meter measures AC/DC Voltage, AC/DC Current, Resistance, Capacitance, Frequency (electrical & electronic),
More informationSAGA PRO SERIES STEREO POWER AMPLIFIER OPERATION MANUAL
SAGA PRO SERIES STEREO POWER AMPLIFIER OPERATION MANUAL INSTALLATION Use care in unpacking the amplifier, and be sure to save the carton and packing materials so that you can use them for moving, storing,
More informationSL300 Snow Depth Sensor USL300 SNOW DEPTH SENSOR. Revision User Manual
USL300 SNOW DEPTH SENSOR Revision 1.1.2 User Manual 1 Table of Contents 1. Introduction... 3 2. Operation... 3 2.1. Electrostatic Transducer... 4 2.2. SL300 Analog Board... 4 2.3. SL300 Digital Circuit
More informationMAC-AFRIC USER S MANUAL MODEL: D.C.INVERTER WELDER 160/200 AMP
MAC-AFRIC USER S MANUAL MODEL: D.C.INVERTER WELDER 160/200 AMP Safety precautions! The processes of welding and cutting, involve potential hazards, so please take the necessary precautions when performing
More informationNBU-1012 Square pulse flashlamp driver User Manual
rev 1.07/ 2009 02 03 NBU-1012 Square pulse flashlamp driver User Manual Overview / Applications Cooling NBU-1012 Square Pulse Flashlamp Driver is designed for simplification of solid-state laser systems
More informationPET1606J2F. Pilani Electron Tubes & Devices Pvt. Ltd. Water Cooled Triode. For Industrial RF Heating. Drop in equivalent of BW1606J2F
Water Cooled Triode For Industrial RF Heating Drop in equivalent of BW1606J2F Output Power: 30 kw Anode voltage: 10 kv max Anode dissipation: 15 kw max Frequency up to 30 MHz Manufactured in India, in
More informationHigh Frequency SineWave Guardian TM Filter
High Frequency SineWave Guardian TM Filter 380V 480V TECHNICAL REFERENCE MANUAL WARNING High Voltage! Only a qualified electrician can carry out the electrical installation of this filter. Quick Reference
More informationSeries 48 Air Cooled Laser & UC-2000 Quick Start Guide
Important Read all Danger, Warning, Caution terms, symbols, and instructions located in the (Laser Safety Hazard information) sections in the Series 48 Laser Operation Manuals. http://www.synrad.com/synrad/docroot/resources/libraries/manuals.
More informationAgilent 5DX System. X-ray Safety Test Procedure. for use with Software Release 8.20 and later
Agilent 5DX System for use with Software Release 8.20 and later Agilent Technologies, Inc. 1996-2008 Agilent Safety and Regulatory Information: Restricted Rights Notice If software is for use in the performance
More informationExperiment 6: Franck Hertz Experiment v1.3
Experiment 6: Franck Hertz Experiment v1.3 Background This series of experiments demonstrates the energy quantization of atoms. The concept was first implemented by James Franck and Gustaf Ludwig Hertz
More informationSINGLE PHASE BUCK & BOOST TRANSFORMERS INSTRUCTION MANUAL
SINGLE PHASE INSTRUCTION MANUAL DIAGRAM D This manual applies to all single-phase buck & boost transformers sold by Larson Electronics. Please refer to the connection diagram on pages 4-6 for properly
More informationINSTALLATION, OPERATION AND MAINTENANCE GUIDE
INSTALLATION, OPERATION AND MAINTENANCE GUIDE FOR INDOOR/OUTDOOR SINGLE PHASE ENCAPSULATED TRANSFORMERS Indoor/Outdoor Encapsulated Transformers The pictures used in this guide are only a representation
More informationSarspec, Lda. - Rua Camilo Castelo Branco, 965 PQ Vila Nova de Gaia Phone:
2 3 IMPORTANT SAFETY NOTE: Before operating this device, please read carefully this User Manual and be familiar with its contents prior to using this equipment. To help avoid potential serious injury to
More informationDEP 001. Operation Manual
DEP 001 Fiber Optic Depolarizer PolaZero Operation Manual June 20, 2003 General Photonics Corp. Ph: (909) 590-5473 5228 Edison Ave. Fax: (909) 902-5536 Chino, CA 91710 USA www.generalphotonics.com Document
More informationFarmPS-1 Table of Contents
Copyright 2004 Grandad s Electronics Seattle, Washington, USA INSTRUCTION MANUAL Model FarmPS-1, 1.5V/90V Farm Radio Power Supply A- A+ FarmPS-1 Table of Contents Section Page Contents 1.0.............................
More informationModel 7000 Low Noise Differential Preamplifier
Model 7000 Low Noise Differential Preamplifier Operating Manual Service and Warranty Krohn-Hite Instruments are designed and manufactured in accordance with sound engineering practices and should give
More informationQ: What does it mean if the RF generator will not put out full power and the reflected power is nearly equal to forward power?
Q: What does it mean if the RF generator will not put out full power and the reflected power is nearly equal to forward power? A: This generally means that the generator is in "reflected foldback." The
More informationPage 1 T O O L S A N D M AT E R I A L S R E Q U I R E D. 1. Screwdriver 2. Carpenter's level 3. Electrical Tape PA R T S L I S T. A.
T O O L S A N D M AT E R I A L S R E Q U I R E D 1. Screwdriver 2. Carpenter's level 3. Electrical Tape PA R T S L I S T A. Cabinet B. Wood Screw X4 C. Plastic wire connector x3 Page 1 I M P O R TA N T
More informationCANARY AUDIO. Power Amplifier CA-309 OWNER S MANUAL. Handcrafted in California MADE IN USA
CANARY AUDIO 300B Push-Pull Parallel Power Amplifier Mono Block Handcrafted in California CA-309 OWNER S MANUAL MADE IN USA Dear Customer: Please allow us to take this opportunity to thank you for purchasing
More informationPET1610F. Pilani Electron Tubes & Devices Pvt. Ltd. Forced-Air Cooled Triode. For Industrial RF Heating. Drop in equivalent of BW1610F
Forced-Air Cooled Triode For Industrial RF Heating Drop in equivalent of BW1610F Output Power: 30 kw Anode voltage: 9 kv max Anode dissipation: 10 kw max Frequency up to 30 MHz Manufactured in India, in
More informationGT-1050A 2 GHz to 50 GHz Microwave Power Amplifier
Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Giga-tronics GT-1050A Microwave Power Amplifier GT-1050A 2 GHz to 50 GHz Microwave Power Amplifier Operation Manual
More informationPULSE DISTRIBUTION AMPLIFIER OPERATING MANUAL
SPECTRADYNAMICS, INC PD5-RM-B PULSE DISTRIBUTION AMPLIFIER OPERATING MANUAL SPECTRADYNAMICS, INC 1849 Cherry St. Unit 2. Louisville, CO 80027 Phone: (303) 665-1852 Fax: (303) 604-6088 www.spectradynamics.com
More informationAcuteB Head. User s Guide
User s Guide SAFETY PRECAUTIONS! Read and follow all safety instructions below carefully to avoid injuries or damages! Make sure that this user manual always accompanies equipment! Profoto products are
More informationSvetlana 4CX7500A Radial Beam Power Tetrode
Svetlana CX7500A Radial Beam Power Tetrode T he Svetlana CX7500A is designed for audio and radio frequency applications. The Svetlana CX7500A has a directly-heated thoriated tungsten mesh filament for
More informationIOM. Model 1050 Strain Gage Transmitter Installation & Operation Manual. (Cage Code #1MU21)
IOM Model 1050 Strain Gage Transmitter Installation & Operation Manual (Cage Code #1MU21) Model 1050 Strain Gage Transmitter Installation & Operation Manual CONTENTS I. GENERAL...1 II. INSTALLATION...2
More informationHAND HELD SAW W MILL
HAND HELD SAW W MILL 92247 ASSEMBLY AND OPERATING INSTRUCTIONS 3491 Mission Oaks Blvd., Camarillo, CA 93011 Visit our Web site at http://www.harborfreight.com Copyright 2004 by Harbor Freight Tools. All
More informationOn-Line Cardio Theater Wireless Digital Transmitter Installation and Instruction Manual
On-Line Cardio Theater Wireless Digital Transmitter Installation and Instruction Manual Full installation instructions accompany your Cardio Theater equipment order. This On-Line version of our Installation/Instruction
More informationDynamo Brushless DC Motor and GreenDriveTM Manual
Dynamo Brushless DC Motor and GreenDriveTM Manual This manual was developed as a guide for use by FIRST Robotics Teams using Controller Part Number 840205-000 in conjunction with the Nidec Dynamo BLDC
More informationDR3535 DR3535-O. Hardware Reference Manual. Document Revision A7 May 16, 2018 MICROKINETICS CORPORATION
-O Hardware Reference Manual Document Revision A7 May 16, 2018 MICROKINETICS CORPORATION 3380 Town Point Drive Suite 330 Kennesaw, GA 30144 Tel: (770) 422-7845 Fax: (770) 422-7854 www.microkinetics.com
More informationThe MDU150 Series provides the same benefits as the MPU150 Series, with nominal 48 volt DC input.
The innovative MPU products incorporate Power Factor Correction (PFC) with a low-profile package designed to meet 1U height constraints. The MPU150-4530G and MPU150-4350G provide high current +3.3 V and
More informationAV30MX-2 Operation Manual
AV30MX-2 Operation Manual 1 Important safety instructions 1. Please read carefully prior to product installation or operation. 2. Read these instructions. 3. Keep these instructions. 4. Heed all warnings.
More informationTECHNICAL DATASHEET #TDAX A DC MOTOR CONTROLLER P/N: AX Variable Speed Control, Onboard I/O CAN SAE J1939, Rugged Packaging
TECHNICAL DATASHEET #TDAX102000 35A DC MOTOR CONTROLLER P/N: AX102000 Variable Speed Control, Onboard I/O CAN SAE J1939, Rugged Packaging with Electronic Assistant Features: Unidirectional or bi-directional
More informationW IDE INPUT CCFL I NVERTER MODULES P R E L I M I N A R Y D ATA S H E E T
LX M1596-01 W IDE INPUT CCFL I NVERTER MODULES T HE I N F I N I T E P O W E R OF I N N O V A T I O N CCFL (cold cathode florescent lamp) Inverter Modules are specifically designed for driving LCD back
More informationTesla K20 GPGPU Card Installation Guide
Dell PowerEdge C80X Tesla K0 GPGPU Card Installation Guide Regulatory Model: B06B Regulatory Type: B06B00 Notes, Cautions, and Warnings NOTE: A NOTE indicates important information that helps you make
More informationThe four pin one (J1) makes possible to supply the board and access to the output voltage easily. It has the following pin-out:
User guide for FHS 40-P Current Transducer FHS 40-P Kit 9 (G2.00.23.108.0) Evaluation PCB Connectors Pin-OUT The board has two single row connectors J1 and J2. The four pin one (J1) makes possible to supply
More informationAdvanced Test Equipment Rentals ATEC (2832)
Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) 6500 Series Loop Antennas User Manual ETS-Lindgren Inc. reserves the right to make changes to any product described
More information3% VOLTAGE DROP. Installation Instructions for 400 Watt 24 Volt DC RGB DMX LED Power Supply SAVE THESE INSTRUCTIONS!
Installation Instructions for 400 Watt 24 Volt DC RGB DMX LED Power Supply Doc # 904-PSB-400W-24VDC_06 1718 W. Fullerton Chicago, IL 60614 Ph: 773.770.1195 Fax: 773.935.5613 www.pureedgelighting.com info@pureedgelighting.com
More informationSPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by
SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This
More information9007 Power Tube. VHF Linear Power Amplifier Tube 33 Kilowatt Peak Sync Output Thru VHF-TV Band
9007 Power Tube VHF Linear Power Amplifier Tube 33 Kilowatt Peak Sync Output Thru VHF-TV Band 14 db Gain High Gain-Bandwidth Products Efficient Forced-Air Cooling Full Input to 400 MHz CERMOLOX Construction
More informationLL-105-R LED Dimmer & 0-10V Bridge Setup Guide
www.i2systems.com i2systems is a registered trademark of Integrated Illumination Systems, Inc. LightLink is a trademark of Integrated Illumination Systems, Inc. 2018 Integrated Illumination Systems, Inc.
More information1-830 SERIES DISPLACEMENT TRANSMITTER
1-830 SERIES DISPLACEMENT TRANSMITTER Operation & Maintenance Manual 746 Arrow Grand Circle Covina, CA 91722 United States of America Tel: (626) 938-0200 Fax: (626) 938-0202 Internet: http://www.cecvp.com
More informationL25 Series Operator Manual
L25 Series Operator Manual L25 899-07001-000 L25P 899-07005-000 L25D 899-07002-000 Access Laser Company 917 134 th St SW Ste A1 Everett, WA 98204 425.582.8674 www.accesslaser.com REVISION J TABLE OF CONTENTS
More informationHIGH-MU POWER TRIODE 3CW40,000A7
TECHNICAL DATA HIGH-MU POWER TRIODE 3CW40,000A7 The EIMAC 3CW40,000A7 is a ceramic/metal power triode designed for use as a high power RF amplifier for industrial and scientific applications. The high-mu
More informationINSTALLATION INSTRUCTIONS HEAVY DUTY TILT WALL MOUNT Model: PPH-2000
INSTALLATION INSTRUCTIONS HEAVY DUTY TILT WALL MOUNT Model: PPH-2000 Specifications: Accomodates Akira and Orion 84" displays without interface bracket; accomodates other large flat panel displays with
More informationPL8877/ 3CX1500A7 High-Mu Power Triode
PL8877/ 3CX1500A7 High-Mu Power Triode The Penta Laboratories PL8877/3CX1500A7 is a rugged ceramic and metal power triode designed for use as cathode driven Class AB2 or Class B amplifi er in audio or
More informationEarth Fault Indicator EFI-BLZ-50 User Manual
Earth Fault Indicator EFI-BLZ-50 User Manual CONTENTS SAFETY NOTICE.................................................................................... 3 1 OVERVIEW......................................................................................
More informationTube Facing Tool.
www.swagelok.com Tube Facing Tool This manual contains important information for the safe and effective operation of the Swagelok TF72 series tube facing tool. Users should read and understand its contents
More informationFLOATING OUTPUT DRIVE, CCFL INVERTER MODULES
Not Recommended for New Design T HE I N F I N I T E P O W E R OF I N N O V A TION The floating output drive CCFL (Cold Cathode Fluorescent Lamp) Inverter Module is specifically designed to drive large
More informationDiode Pointer Quick Start Guide
Important Note: Read all Danger, Warning, Caution terms, symbols, and instructions located on our website. The diode pointer provides a safety measure by enabling operators to monitor the location of the
More informationOPERATION & SERVICE MANUAL FOR FC 110 AC POWER SOURCE
OPERATION & SERVICE MANUAL FOR FC 100 SERIES AC POWER SOURCE FC 110 AC POWER SOURCE VERSION 1.3, April 2001. copyright reserved. DWG No. FC00001 TABLE OF CONTENTS CHAPTER 1 INTRODUCTION... 1 1.1 GENERAL...
More informationAmplifier Series BASIC. Installation & Operations Manual
Amplifier Series BASIC Installation & Operations Manual Bittner-Audio 200 Power Amplifier Series BASIC Bittner - Audio September 200 200 Bittner-Audio. All Rights Reserved. Bittner-Audio reserves specification
More informationAgilent 5DX Series 2L, 3, and 5000 X-ray Safety Test Report
Agilent 5DX Series 2L, 3, and 5000 X-ray Safety Test Report To be completed by the EMG RADIATION SAFETY & REGULATORY ENGINEER: System SN: Test Date: Month Day Year Survey Report Type: Net X-ray emission:
More informationUser Manual Digital Multimeter
User Manual Digital Multimeter model no.: MSR-R500 Questions or Concerns? support@etekcity.com visit etekcity.com for more products Safe and Proper Usage Thank you for purchasing the Etekcity MSR-R500
More informationCommercial Microwave Technical Information. CAUTION All safety information must be followed as provided in Service Manual
Commercial Microwave Technical Information 0 V, 0 Hz Models RCS0MPA P00M RCS0A P00M RCS0PBDA P00M RCS0MPSA P00M RFSMPSA P00M RFSSWA P00M RCS0SWA P0M RCS0MPA P0M Due to possibility of personal injury or
More informationPERFORMANCE SPECIFICATION SHEET ELECTRON TUBE, MAGNETRON TYPE DOD-011
INCH-POUND PERFORMANCE SPECIFICATION SHEET 10 May 2018 SUPERSEDING MIL-PRF-1/1719C 22 June 2012 ELECTRON TUBE, MAGNETRON TYPE DOD-011 This specification is approved for use by all Departments and Agencies
More informationIDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: MODEL: Multimeter Service Information
IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: 61-340 MODEL: 61-342 Multimeter Service Information The Service Information provides the following information: Precautions and safety information Specifications
More informationBroadcast Concepts Inc NW 102 Road Suite 4 Medley FL Tel: : Fax Model P50FM42MH-SMA2 FM Pallet Amplifier Module
Model P50FM42MH-SMA2 FM Pallet Amplifier Module This amplifier module is ideal for driver stages in FM Broadcast transmitters. 86 110MHz 28Volts Pout: 50W CW minimum 40dB Gain Class AB MACOM MRF173 Mosfet
More informationTITANIUM SERIES R 51 PART NUMBER SELECTION. High Performance Multiport Switches. Our Most Important Connection is with You.
PART NUMBER SELECTION Models: 2: Without 50 Ω terminations 4: With 50 Ω terminations RF Connectors: 3: SMA up to 6 GHz 4: SMA up to 20 GHz F: SMA up to 26.5 GHz 8: SMA 2.9 up to 40 GHz (1) Radiall s TITANIUM
More informationHITEK POWER OL1K SERIES
HIGH VOLTAGE POWER SUPPLY The HiTek Power OL1K Series range of single output high voltage power supplies meets the exacting requirements found in electron and ion beam systems, ion implantation and X-ray
More informationHITEK POWER XR KW X-RAY POWER SUPPLY MODULE
HITEK POWER XR1000 1 KW X-RAY POWER SUPPLY MODULE The HiTek Power XR1000 series is a component power supply for industrial x-ray systems, elemental analysis equipment, x-ray diffraction spectrometers,
More informationINSTRUCTION MANUAL. Specifications Electrical. Front-To-Back Ratio VSWR at Resonance Less than 1.5:1 Nominal Impedance. Mechanical
300 Industrial Park Road, Starkville, MS 39759 Ph: (662) 323-8538 FAX: (662) 323-6551 TH-3JRS Tri-band HF 3 Elements Beam Covers 10, 15 and 20 Meters INSTRUCTION MANUAL WARNING Installation of this product
More informationGuide to the Risø XHV-1 x-ray unit
Guide to the Risø XHV-1 x-ray unit DTU Nutech Radiation Research Division Risø Campus, Building. 201 Technical University of Denmark 4000 Roskilde, Denmark Revision: May 20, 2016 2 Contents 1 Introduction
More informationDCM w Power Amplifier
DCM500 500w Power Amplifier HI MAINS TEMP ON 24 21 18 15 12 9 6 3 0 +3 FAIL DCM 500 Operating Manual DCM500, 500 Watt Power Amplifier Product Description The DCM500 is a 500 watt power amplifier designed
More informationRotating Anode X-Ray Tube Housing Assembly. General Data
Display Devices & Components Company TECHNICAL DATA ROTANODE TM E7252X Rotating Anode X-Ray Tube Housing Assembly! High speed rotating anode X-ray tube housing assembly for high energy radiographic and
More informationUltrasonic Displacement Sensor E4DA
Sensing Supply voltage Output Ultrasonic beam Ultrasonic Displacement Sensor Sensor s Narrow Ultrasonic Beam Accurately Detects Small Objects, Provides Linear Analog Output For Inspection, Measurement
More informationAL30 Operator Manual
AL30 Operator Manual Pictured AL30 Air Cool Access Laser Company 917 134 th St SW, Suite A1 Everett, WA 98204 425.582.8674 www.accesslaser.com AL30 Series Operator Manual_Rev_I.Docx Section TABLE OF CONTENTS
More informationDirected Energy, Inc Oakridge Dr., Suite 100, Fort Collins, CO
PCO-7121 Laser Diode Driver Module Operation Manual Directed Energy, Inc. 1609 Oakridge Dr., Suite 100, Fort Collins, CO 80525 sales@ixyscolorado.com www.ixyscolorado.com Contents Contents... 3 Safety...
More informationHITEK POWER PSM10 SERIES
HITEK POWER PSM10 SERIES PRECISION SCIENTIFIC POWER SUPPLY MODULES The HiTek Power PSM10 series is a range of versatile high voltage component power supply modules equally suited to both laboratory and
More informationSafety Precautions. Copyright / Trademarks
Safety Precautions Observe the following notices to ensure personal safety or to prevent accidents. To ensure that you use this product correctly, read this User s Manual thoroughly before use. Make sure
More informationMONOBLOCK MEDICAL INDUSTRIAL & COMMERCIAL X-RAY ANALYSIS. spellmanhv.com. The Integrated X-Ray Source DRIVING TOMORROW S TECHNOLOGY
2013 The Integrated X-Ray Source DRIVING TOMORROW S TECHNOLOGY MEDICAL INDUSTRIAL & COMMERCIAL X-RAY ANALYSIS spellmanhv.com MONOBLOCK The Integrated X-Ray Source POWER SUPPLY & X-RAY TUBE IN AN INTEGRATED
More informationL4 and L4i 915/940 nm Fiber- Coupled Lasers
L4 and L4i 915/940 nm Fiber- Coupled Lasers wwwlumentumcom Data Sheet Lumentum L4-series diode lasers offer up to 10 W of power from a 105 μm fiber The L4 is a revolutionary platform based on a long history
More informationWATERFLUX 3000 Quick Start
WATERFLUX 3000 Quick Start Electromagnetic flowmeter The documentation is only complete when used in combination with the relevant documentation for the signal converter. KROHNE CONTENTS WATERFLUX 3000
More informationType 297, High-Voltage Mica Capacitors Corona-free Mica Coupling Capacitors for Medium-Voltage PDA s
Designed for Partial Discharge Analyzers (PDA s) monitoring rotating machinery or other medium-voltage equipment from 1 to 35 kvac RMS at power-line frequencies of 10 Hz to 1 khz, Mica Capacitor Type 297
More informationDUAL OUTPUT AC CURRENT/VOLTAGE TRANSDUCER
OPERATOR S MANUAL DUAL OUTPUT AC CURRENT/VOLTAGE TRANSDUCER Masibus Automation & Instrumentation Pvt. Ltd. B/30, GIDC Electronics Estate, Sector-25, Gandhinagar-382044, Gujarat, India Web Site: www..com
More informationINSTALLATION INSTRUCTIONS ATV WINCH MULTI-MOUNT Multi-Mount Mounting-Kit: PN Application: YAMAHA BIGBEAR 00+
INSTALLATION INSTRUCTIONS ATV WINCH MULTI-MOUNT Multi-Mount Mounting-Kit: PN 61025 Application: YAMAHA BIGBEAR 00+ Your safety, and the safety of others, is very important. To help you make informed decisions
More informationADAE portable X ray machine Operation manual
ADAE portable X ray machine Operation manual CONTENTS ABOUT THE SYMBOL----------------------------------------------------------3 INTRODUCTION-----------------------------------------------------------------4
More informationETX Powered Loudspeaker Accessories
ETX Powered Loudspeaker Accessories ETX-BRKT10, ETX-BRKT12, ETX-BRKT15, ETX-TCA-S, ETX-TCA-L, and ETX-BRKT35 en Installation Guide en 3 Table of contents 1 Safety 4 2 Installation 6 2.1 Wall mount bracket
More informationSDI SPECTRADYNAMICS, INC. HIGH PERFORMANCE DISTRIBUTION AMPLIFIER OPERATING MANUAL
SPECTRADYNAMICS, INC. HIGH PERFORMANCE DISTRIBUTION AMPLIFIER HPDA15RMC OPERATING MANUAL SPECTRADYNAMICS, INC 1849 Cherry St. Unit 2. Louisville, CO 80027 Phone: (303) 6651852 Fax: (303) 6046088 www.spectradynamics.com
More informationPSW-002. Fiber Optic Polarization Switch. User Guide
PSW-002 Fiber Optic Polarization Switch User Guide Version: 1.0 Date: May 30, 2014 General Photonics, Incorporated is located in Chino California. For more information visit the company's website at: www.generalphotonics.com
More informationTECHNICAL SPECIFICATIONS for 300KV CONSTANT POTENTIAL X RAY EQUIPMENT (for industrial applications) S. No. PARTICULARS BHEL SPECIFICATIONS
TECHNICAL SPECIFICATIONS for 300KV CONSTANT POTENTIAL X RAY EQUIPMENT (for industrial applications) BIDDER s OFFER S. No. PARTICULARS BHEL SPECIFICATIONS 1.0 Area of Application The 300kV Constant potential
More information500 kva Dry-type Isolation Transformer V-Delta to 208Y/120V 3PH - NEMA 1/Indoor
500 kva Dry-type Isolation Transformer - 13200V-Delta to 208Y/120V 3PH - NEMA 1/Indoor Part #: MT-DOE16-3P-13200D-500KVA-208Y.120-N1 Page: 1 The MT-DOE16-3P-13200D-500KVA-208Y.120-N1 Dry-type Isolation
More informationSK-PW7 Servo. Power Bus. User Manual and Setup Guide. Rev 3.00 SK-PW7 User Manual & Setup Guide
SK-PW7 Servo Power Bus User Manual and Setup Guide Rev 3.00 SK-PW7 User Manual & Setup Guide Table of Contents 1 Introduction...4 2 Package Contents...5 3 Safety...6 4 Principle of Operation...7 5 Connections...8
More informationCable I.D. Live. Primary or Secondary In or Out of the Trench TX-Former to TX-Former TX-Former to Meter Energized or Grounded.
Cable I.D. Live Primary or Secondary In or Out of the Trench TX-Former to TX-Former TX-Former to Meter Energized or Grounded www.aquatronics.com i ii TABLE OF CONTENTS Warning Description Battery Test
More informationOrtel (Lucent) Fiber Optic Interface
Ortel (Lucent) Fiber Optic Interface 144-701-1 Serial Number January 31, 2013 Part number 098-00120-000 Revision C ORTEL (LUCENT) FIBER OPTIC INTERFACE 1. GENERAL INFORMATION...1 1.1. SCOPE OF OPTION...1
More informationQRS14 User s Guide. MEMS GYROSCOPE Model QRS14. Systron Donner Inertial Sales and Customer Service. Phone: Fax:
RELEASED DOCUMENT DATE: 0-25-206 QRS4 User s Guide MEMS GYROSCOPE Model QRS4 Systron Donner Inertial Sales and Customer Service Phone: + 925.979.4500 Fax: + 925.349.366 E-Mail: sales@systron.com www.systron.com
More informationREAD THIS BEFORE ATTEMPTING TO COMBINE
HANDBOOK FM SERIES 2000W POWER COMBINER READ THIS BEFORE ATTEMPTING TO COMBINE 4 s PTEK Contents Theory of operation...2 Specifications...3 Safety Notice...4 Rear Panel Layout...5 Installation...8 Schematic
More informationTECHNIlCAL DATA. Amperes Cathode Heating Time..300
TECHNIlCAL DATA TRIODE The BIMAC Y-810 tube has been specifically designed for high voltage series regulator or switch tube (modulator) service. The compact, rugged design has very low internal inductance
More informationA53106 SERIES DC-TO-DC CONVERTER
INSTALLATION & MAINTENANCE A53106 SERIES DC-TO-DC CONVERTER AUGUST 2011, REVISED AUGUST 2014 DOCUMENT NO. COM-00-04-20 VERSION C.1 Siemens Industry, Inc., Rail Automation 9568 Archibald Ave., Suite 100,
More information2014 CINDE Toronto Portable X-ray & X-Ray Production
2014 CINDE Toronto Portable X-ray & X-Ray Production Nils Hase, Business Development Manager Portable X-Ray business unit Copenhagen and USA Technology with Passion 3 COMET Headquarters Flamatt Switzerland
More information