SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

Size: px
Start display at page:

Download "SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by"

Transcription

1 SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments Oct-06 Draft Igor Gomola Page 1 of 12

2 1. Scope This specification describes the main technical requirements for a kilovoltage X ray calibration system for calibration of equipment used in radiation protection and diagnostic radiology. The procurement of the system is planned for the Secondary Standards Dosimetry Laboratory (SSDL) of Ecuador in the framework of the IAEA Technical Cooperation project ECU6023. Expected outcome of this project is to upgrade calibration services for equipment used in radiation protection to end-users through the SSDL. 2. Applicable documents 1. INTERNATIONAL ATOMIC ENERGY AGENCY, "International Basic Safety Standards for Protection against Ionizing Radiation and for the Safety of Radiation Sources", Safety Series No. 115, IAEA, Vienna (1996), 2. INTERNATIONAL ATOMIC ENERGY AGENCY, Calibration of Radiation Protection Monitoring Instruments, Safety Reports Series No. 16., IAEA Vienna (2000) 3. INTERNATIONAL ATOMIC ENERGY AGENCY, "Dosimetry in diagnostic radiology: An international code of practice", Technical Reports Series No. 457, IAEA, Vienna (2007), 4. INTERNATIONAL ATOMIC ENERGY AGENCY, "Implementation of the International Code of Practice on Dosimetry in Diagnostic Radiology (TRS 457): Review of Test Results", IAEA Human Health Reports No.4, IAEA, Vienna (2011), 5. INTERNATIONAL ELECTROTECHNICAL COMMISSION, Medical diagnostic X-ray equipment - Radiation conditions for use in the determination of characteristics, IEC 61267, Geneva, Switzerland INTERNATIONAL ORGANIZATION FOR STANDARDIZATION, X and gamma reference radiation for calibrating dosimeters and dose rate meters and for determining their response as a function of photon energy, ISO, Geneva, ISO 4037 (1999) 7. The Consultative Committee for Standards of Ionizing Radiations (CCRI), X-ray beam qualities, Bureau International des Poids et Mesures (BIPM), 3. Abbreviations SSDL: HVL: SDD: FS: CBA: Secondary Standards Dosimetry Laboratory of Ecuador Half Value Layer Source to detector distance Field Size Diameter Central Beam Axis Page 2 of 12

3 4. Requirements 4.1. Functional requirements. The new X ray calibration system shall be able to produce desired radiation beam qualities for calibration of equipment used in radiation protection (narrow spectrum series) and diagnostic radiology (conventional diagnostic beams, except mammography). The calibrations benches are already available at the SSDL, for this reason is important to integrate the existing benches with new kv calibration system Performance requirements All performance specifications and tests shall conform to the relevant standards of the IAEA, IEC and ISO, as specified above Technical specification The X ray calibration system shall include the following components: 4.3. Technical specification The X ray calibration system shall include the following components: X ray tube with HV cables with W anode, for (i) diagnostic, (ii) protection level Shielding cabinet for X ray tube assembly High voltage generator System control panel Beam shutter Beam collimation system Filter assembly HVL (Half Value Layer) measurement system Ancillary eqquipment Safety items X ray tube with HV cables Metal ceramic X ray tube X ray tube voltage range, at least: 20 kv kv Continuous rating power at least 3 kw Target material: Tungsten Thickness of the exit window: max 1 mm Material of the exit window: Beryllium Anode angle: at least 20 degrees Focal spot size: at least 4 mm Cooling medium: preferably water Shielding cabinet Page 3 of 12

4 The shielding cabinet shall (i) totally enclose X ray tube with the shutter and cables (ii) provide mechanical support for the remote controlled filter changing device The enclosure shall have a door to access the tubes and the shutter The enclosure door shall be with an interlock that disables the power when opened The shielding cabinet surrounding X ray tube shall provide sufficient shielding from the scatter and leakage-radiation created by the X ray tube in use The design shall ensure that the leakage radiation complies with the International Basic Safety Standards [1] The X ray tube head shall be mounted in a holder allowing geometrical alignment with desired central radiation beam axis (CBA) The height of CBA above the floor level shall be 110 cm High voltage generator Unipolar system, high frequency, constant potential type, the ripple less than 0.5% Output voltage range, at least: 20kV kv (0.1 kv steps). The accuracy of the high voltage shall be better than ± 1% and the repeatability better than 0.05 kv Output current, at least: from 0.5 ma to 40 ma (possible to change in 0.1 ma steps). The accuracy of the current shall be better than ± 1% and the repeatability better than 0.05 ma Output power (at least): 3kW The performance of the generator shall not be affected by ± 10 % fluctuation of the mains input voltage The HV generator shall operate at mains input 110 V single phase (available power line conditions to be confirmed by the counterpart in advance) Cooling unit for X ray system shall have sufficient capacity to allow safe and stable operation of X ray unit at maximum power for several hours. The cooling unit with close circuit for the cooling medium is preferable X ray control panel Preselection of high voltage in steps of 0.1 kv Display actual value of HV (at least 3 digits) Preselection of tube current in steps of 0.1 ma Display actual value of current (at least 3 digits) Preselection of exposure time in steps of 1 sec Display actual value of exposure time (at least 5 digits) Remote control of the beam shutter (On/Off) by selected exposure time Memory for setup of X ray tube parameters Digitally assigned limit data (high voltage, current, anode power dissipation) Warm up programme for X ray tube (manual and automatic mode) Warm up history Emergency stop button Indication X ray beam Off/On Connection to various safety interlocks (door switch, door eye beam, radiation monitor signal) to avoid unintended exposure of staff Page 4 of 12

5 As an option, the control unit shall be equipped with a PC interface allowing remote control X ray beam shutter A shutter shall be mounted inside the shielding cabinet in the front of the exit window of the X ray tube (preferably behind the primary collimator) The shutter must be thick enough to attenuate the transmitted air kerma rate of the primary radiation beam to 0.1% on the CBA for the radiation quality with the highest mean energy The shutter shall be remotely controlled from the X ray control panel The actual shutter status (Open/Close) shall be indicated on the control panel The shutter transit time shall be less than 100 msec The actual value of the shutter transit time together with its uncertainty (Δt, %SD(Δt)) shall be measured in the factory. The report from the measurement shall be provided to the end-user Beam collimation system Primary collimation system shall be made preferably from tungsten in the shape of a conical aperture, which is positioned between the X ray output window and the beam shutter Secondary collimation system is used to define a reference filed size at the given calibration conditions. The secondary collimator shall be placed between the filter assembly and the monitor chamber The diameter of the second conical aperture (collimator) determines the field size at the reference point. There shall be provided a set of easily exchanged collimators with various diameters to obtain field sizes matching various irradiation conditions. At least, the following irradiation conditions shall be possible set-up SDD=200 cm, FS=30 cm SDD=100 cm, FS=10 cm HVL measurements: SDD=100 cm, FS=5 cm The design shall allow easy exchange of the secondary collimators with no need to dismount a monitor chamber Behind the secondary collimation system shall be placed a rigid holder for mounting a monitor ionization chamber (e.g. PTW type 34014). The geometrical centre of the monitoring chamber shall be positioned in the central radiation beam axis Behind the monitor chamber shall be placed the third conical aperture with diameter bigger than radiation field size. The purpose of this aperture is to reduce backscatter radiation The distance between the focal spot of the X ray tube and the end point of the third aperture shall not exceed 40 cm Page 5 of 12

6 Filter assembly The filter assembly is used to modify the photon spectrum of radiation beams produced by the X ray tube to meet the characteristics of radiation beam qualities required for calibration of equipment A set of metal filters shall be provided to produce desired radiation beam qualities characterized by 1 st HVL, as described in Tables 1, 3, 4 and The metal filters (Al, Cu, Pb, Sn) used for the establishment of required radiation beam qualities shall have certified purity of 99.9%, at least The metal filters for all beam qualities shall be placed in removable and remotely controlled carrousels It shall be possible to remotely change filter wheel position to get the required filtration within each series (N series, RQR, RQA, RQT series). It means, for example, that changing the filters to the desired position during the calibration with RQA series (2,3,4,5,6,7,8,9,10) shall be done remotely without need to enter the irradiation room The carrousel to be constructed as a rotational disk with several holes located between the centre of the disk and the radius All holes in the disk shall have the same diameter and shall be designed to hold the metal filters. Each radiation beam quality may require combination of various metal filters therefore the each hole shall allow to install more than one metal filter (see Appendix 1 and 2) The each hole and the each disk shall be uniquely coded to assure correct setup of selected filters during calibration The metal filter (beam code) selected during the irradiation shall be indicated in the control panel of the device The carrousel shall be mounted between the shutter and the secondary collimation system in a way that the centre of the holes in the disk is on the CBA On the radiation beam axis (between the X ray tube output and the primary collimator) shall be included manual or automatic filter insert system allowing to add additional filters into the beam (i.e. N series above 40 kv require inherent filtration adjusted to 4 mm of Aluminium, see Table 1, Appendix 1) The thickness of added filtrations given in Tables 1, 3, 4, 5, are representing the nominal values and shall be adjusted individually with the given X ray tube to obtain desired HVL values within required tolerances [5, 6] HVL measurement system Manual insert system including removable attenuators (sheets of various metals) shall be possible to place on the calibration bench. The centre of the metal sheets shall be located on CBA The minimum distance from the focus of the X ray tube to the position of removable attenuators shall be 50 cm The desired thickness of the attenuator is adjusted by adding different thicknesses of attenuating material until the required attenuation of radiation beam is achieved The set shall include aluminium and copper sheets made of various thicknesses to determine the desired HVLs Each metal filter shall be accompanied with a certificate stating (i) purity of the material (ii) thickness, including the measurement uncertainty The required purity of the material used for attenuation shall be at least % The collimator assuring narrow beam geometry required for measurement of HVLs shall be included Page 6 of 12

7 Optionally, automatic HVL measurement system shall be offered. In this case, the change of filter material thicknesses is remotely controlled so the full set of measurements is performed without entering the irradiation room Ancillary equipment Spare parts and tools used for on-site preventive maintenance of X ray system. More details are required. Spare part and tools list shall be provided Safety items Radiation monitor for workplace monitoring to be installed in the calibration room. The display shall be located outside of the irradiation room (in the control room) and indicate radiation exposure in terms of Sv/hr. A visual and audible alarm signals shall be initiated, when the radiation source is in the irradiation position (exposure rate in the irradiation room is elevated) Video color surveillance system: 2 cameras with wide-angle lenses and wall mounts (each camera to survey 2/3 of the irradiation room) and 2 monitors located in the control room The X ray system shall be connected to the radiation safety infrastructure already installed in the irradiation and control room (see Appendix 3). 5. Accompanying documents 5.1. Performance specifications 5.2. Operating instructions 5.3. Installation instructions 5.4. Preventive maintenance manual 5.5. Service manual with schematic circuit diagrams of electronic circuits of the X ray system 5.6. Installation requirements 5.7. Acceptance testing protocol 6. Marking The X ray system shall have the safety marking as required by international standards given in section 2 Applicable documents. 7. Packing The X ray calibration system shall be packed in accordance with international standards that are applicable for the shipment of this type of equipment. 8. Installation The Contractor shall install and test the X ray calibration system at the SSDL of Ecuador 9. Testing (Factory and Site) and Acceptance 9.1. The X ray calibration system shall be tested in the factory before the shipment (FAT - Factory Acceptance Testing) to assure that the system is in working conditions and fulfil all technical requirements. The FAT results shall be submitted to the IAEA (attn. to Technical Officer of the TC project) before shipping the system to end user. The FAT results shall include at least the following parameters: - radiation beam profiles for all collimators measured at the distance of 1 m with at least one beam quality energy Page 7 of 12

8 - radiation beam output in terms of air kerma rate (mgy/min) measured on CBA at the distance of 1m with at least one beam quality per beam series (i.e. N100, RQR5, RQA5, RQT9). The used kv x ma shall be close to 1.5 kw. - air kerma vs irradiation time (1 sec 60 sec range ) for selected beam quality to verify shutter functionality 9.2. The X ray calibration system shall be tested after installation at the SSDL in accordance with the applicable standards mentioned in chapter 2 by the Contractor together with the staff of the SSDL to demonstrate that: the performance meets the manufacturer s performance specifications and the minimum requirements specified herein as determined by the IAEA the functionalities of the whole system is maintained 9.3. The results of the testing of the X ray calibration system shall be documented by the Contractor in Acceptance Protocol that shall be signed by the representative of the SSDL 10. Training The Contractor shall provide the training in the operation and first line maintenance of the X ray calibration system at the SSDL of Ecuador immediately after the installation. Type of training with the detailed program and expected duration shall be provided. 11. Warranty and service A complete warranty shall be offered for at least 1 year, starting after formal acceptance testing Time of installation, included all the requirements shall be provided Local or regional representation for services shall be provided, CV of the local/ regional representations, answer time, type of penalties in case of no complies shall be provided An option for maintenance contract (starting after the warranty period) shall also be included in the offer Training for in-house engineers on first line maintenance The Contractor shall guarantee that replacement parts are available for at least 10 years after installation of the X ray calibration system The Contractor shall ensure that service support and further upgrades of the system are available As an option, a proposal for service training shall be included in the offer. Page 8 of 12

9 APPENDIX 1 BEAM QUALITIES FOR PROTECTION LEVEL CALIBRATIONS Table 1. Characteristics of narrow-spectrum series ISO-4037 [6]. The minimum range of established beam qualities for the SSDL is in the range from N40 to N200 Page 9 of 12

10 APPENDIX 2 BEAM QUALITIES FOR DIAGNOSTIC LEVEL CALIBRATIONS TABLE 2. RADIATION QUALITIES FOR CALIBRATIONS OF DIAGNOSTIC DOSIMETERS (IAEA TRS- 457) Radiation quality RQR Radiation origin Radiation beam emerging from X ray assembly Material of an additional filter no phantom Application General radiography, fluoroscopy and dental applications (measurements free in air) RQA Radiation beam with an added filter Aluminum Measurements behind the patient (on the image intensifier) RQT Radiation beam with an added filter Copper CT applications (measurements free in air) TABLE 3. CHARACTERIZATION OF RADIATION QUALITY SERIES RQR Radiation quality X ray tube voltage **Added filtration Nominal 1 st HVL (kv) (mm Al) (mm Al) RQR RQR RQR RQR 5* RQR RQR RQR RQR RQR *This value is generally selected as the reference radiation quality for non-attenuated beams for general radiography applications. ** The fixed filtration consists of 1.0 mm Be + monitor chamber. The values of added filtration are informative only and shall be adjusted to achieve desired HVL values within ± 5%. Page 10 of 12

11 APPENDIX 2 CONTD. BEAM QUALITIES FOR DIAGNOSTIC LEVEL CALIBRATIONS TABLE 4. CHARACTERIZATION OF RADIATION QUALITIES SERIES RQA Radiation quality X ray tube voltage **Added filtration (kv) (mm Al) (mm Al) Nominal 1 st HVL RQA 2 40 RQR RQA 3 50 RQR RQA 4 60 RQR RQA 5* 70 RQR RQA 6 80 RQR RQA 7 90 RQR RQA RQR RQA RQR RQA RQR *This value is generally selected as the reference radiation quality for attenuated beams for general radiography applications. ** The fixed filtration consists of 1.0 mm Be + monitor chamber. The values of added filtration are informative only and shall be adjusted to achieve desired HVL values within ± 5%. TABLE 5. CHARACTERIZATION OF RADIATION QUALITY SERIES RQT Radiation quality X ray tube voltage Added filtration Nominal 1 st HVL (kv) (mm Cu) (mm Al) RQT RQR RQT 9* 120 RQR RQT RQR *This value is generally selected as the reference radiation quality for computed tomography. ** The fixed filtration consists of 1.0 mm Be + monitor chamber. The values of added filtration are informative only and shall be adjusted to achieve desired HVL values within ± 5%. Page 11 of 12

12 APPENDIX 3 Layout of the premises at the SSDL of Ecuador Fig.1 Current X-ray system installed at the SSDL of Ecuador Fig.2 X-ray and gamma calibration room at the SSDL of Ecuador Page 12 of 12

Calibration of KAP meters

Calibration of KAP meters Calibration of KAP meters Alexandr Malusek! Division of Radiological Sciences Department of Medical and Health Sciences Linköping University! 2014-04-15 1 Outline 1. KAP meter construction 2. Air kerma-area

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced

More information

XRC X-Ray Calibration System

XRC X-Ray Calibration System XRC X-Ray Calibration System Technical Description Contents 1. Equipment overview 2. X-ray Beam Specifications 3. The Control Console 4. Radiation Safety 5. Filters 6. X-ray Generator & Heat Exchange Equipment

More information

Preliminary studies of a new monitor ionization chamber

Preliminary studies of a new monitor ionization chamber 1 Preliminary studies of a new monitor ionization chamber Maíra T. Yoshizumi, Vitor Vivolo and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares (IPEN-CNEN) Comissão Nacional de Energia

More information

X-RAYS - NO UNAUTHORISED ENTRY

X-RAYS - NO UNAUTHORISED ENTRY Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are

More information

TECHNICAL SPECIFICATIONS for 300KV CONSTANT POTENTIAL X RAY EQUIPMENT (for industrial applications) S. No. PARTICULARS BHEL SPECIFICATIONS

TECHNICAL SPECIFICATIONS for 300KV CONSTANT POTENTIAL X RAY EQUIPMENT (for industrial applications) S. No. PARTICULARS BHEL SPECIFICATIONS TECHNICAL SPECIFICATIONS for 300KV CONSTANT POTENTIAL X RAY EQUIPMENT (for industrial applications) BIDDER s OFFER S. No. PARTICULARS BHEL SPECIFICATIONS 1.0 Area of Application The 300kV Constant potential

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

DETERMINATION OF INHERENT AND ADDITIONAL FILTRATION IN ORDER TO ESTABLISH RADIATION QUALITIES ACCORDING TO IEC 61267

DETERMINATION OF INHERENT AND ADDITIONAL FILTRATION IN ORDER TO ESTABLISH RADIATION QUALITIES ACCORDING TO IEC 61267 2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 DETERMINATION

More information

This document is a preview generated by EVS

This document is a preview generated by EVS INTERNATIONAL STANDARD ISO 17636-1 First edition 2013-01-15 Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film Contrôle non destructif des assemblages soudés

More information

Half value layer and AEC receptor dose compliance survey in Estonia

Half value layer and AEC receptor dose compliance survey in Estonia Half value layer and AEC receptor dose compliance survey in Estonia K. Kepler, A. Vladimirov Training Centre of Medical Physics, University of Tartu Testing Centre of the University of Tartu, Estonia E-mail:

More information

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film INTERNATIONAL STANDARD ISO 17636-1 First edition 2013-01-15 Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film Contrôle non destructif des assemblages soudés

More information

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure Accuracy and calibration of integrated radiation output indicators in diagnostic radiology: A report of the AAPM Imaging Physics Committee Task Group 190 Pei-Jan P. Lin a) Virginia Commonwealth University

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 62220-1 First edition 2003-10 Medical electrical equipment Characteristics of digital X-ray imaging devices Part 1: Determination of the detective quantum efficiency Appareils

More information

Test Equipment for Radiology and CT Quality Control Contents

Test Equipment for Radiology and CT Quality Control Contents Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3

More information

Response characteristics of a tandem ionization chamber in standard X-ray beams

Response characteristics of a tandem ionization chamber in standard X-ray beams Applied Radiation and Isotopes 58 (2003) 495 500 Response characteristics of a tandem ionization chamber in standard X-ray beams Alessandro M. Costa*, Linda V.E. Caldas Instituto de Pesquisas Energ!eticas

More information

2012 :15th SESSION of ESMP

2012 :15th SESSION of ESMP 2012 :15th SESSION of ESMP Lecture presented in Archamps (Salève Building) by : Elly CASTELLANO (London) Patient dosimetry in x-ray imaging and CT Elly Castellano Objectives measurable dose quantities

More information

Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision Minnesota Rules, Chapter 4732 X-ray Revision DRAFT INDUSTRIAL X-RAY SYSTEMS DEFINTIONS, 1.0 4732.####. INDUSTRIAL X-RAY SYSTEMS DEFINITIONS. Subpart 1. Scope. For purposes of industrial x-ray systems under

More information

MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM *

MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM * MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM * C. Kessler and D.T. Burns December 2018 BUREAU INTERNATIONAL DES POIDS ET MESURES

More information

Derivation of total filtration thickness for diagnostic x-ray source assembly

Derivation of total filtration thickness for diagnostic x-ray source assembly Physics in Medicine & Biology PAPER OPEN ACCESS Derivation of total filtration thickness for diagnostic x-ray source assembly To cite this article: Michiharu Sekimoto and Yoh Katoh 2016 Phys. Med. Biol.

More information

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information

Electronic Brachytherapy Sources. Thomas W. Rusch

Electronic Brachytherapy Sources. Thomas W. Rusch Electronic Brachytherapy Sources Thomas W. Rusch Educational Objectives Understand key elements of ebx source construction & operation Understand the rationale and methods for air kerma strength calibration

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-649 CDRH Fluoroscopic Phantom Users Manual March 2005 Manual No. 07-649-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

TABLE OF CONTENTS. References

TABLE OF CONTENTS. References ANALYTICAL X-RAY EQUIPMENT USE POLICIES & PROCEDURES Page 1 of 9 Revised: 11/24/2003 TABLE OF CONTENTS 1.0 General 2.0 Purpose 3.0 Scope and Authority 4.0 Equipment Requirements 4.0.1 Safety device 4.0.2

More information

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS By Aravind Ravichandran arr192@mail.usask.ca University of Saskatchewan Address: 2424 Cumberland

More information

RADIATION SAFETY REQUIREMENTS FOR ANALYTICAL X-RAY EQUIPMENT

RADIATION SAFETY REQUIREMENTS FOR ANALYTICAL X-RAY EQUIPMENT 420-3-26-.11 RADIATION SAFETY REQUIREMENTS FOR ANALYTICAL X-RAY EQUIPMENT (1) Purpose and Scope. This rule 420-3-26-.11 provides special requirements for analytical x- ray equipment; provided, however,

More information

Rotating Anode X-ray Tube

Rotating Anode X-ray Tube Datasheet RADII KL74-1.0/2.0-125 Rotating Anode X-ray Tube Rotating anode X-ray tube for the purpose of general diagnostic X-ray procedures. Specially processed Tungsten faced molybdenum target of 74 mm

More information

Disclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography

Disclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography Current Implementation Status of IEC Standard 62494-1: Exposure Index (EI) for Digital Radiography July 31, 2017 Ryan Fisher, PhD, DABR Katie Hulme, MS, DABR None Disclosures Outline Review of IEC Standard

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

Update on NIST X-Ray Air-Kerma Standards and Calibrations

Update on NIST X-Ray Air-Kerma Standards and Calibrations CCRI(I)/03-19 Update on NIST X-Ray Air-Kerma Standards and Calibrations Michelle O Brien and Stephen M. Seltzer Ionizing Radiation Division National Institute of Standards and Technology Gaithersburg,

More information

Nuclear Associates , , CT Head and Body Dose Phantom

Nuclear Associates , , CT Head and Body Dose Phantom Nuclear Associates 76-414,76-414-4150,76-415 CT Head and Body Dose Phantom Users Manual March 2005 Manual No. 76-414-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS 7:28-14.1 Scope (a) This subchapter covers therapeutic installations used in the healing arts. These therapeutic installations include x-ray, accelerator and teletherapy

More information

Technical Specifications SECTION C

Technical Specifications SECTION C Page 1 of 12 INSTITUTE FOR PLASMA RESEARCH Technical Specifications SECTION C Design, Fabrication, assembly, testing and supply of Filter polychromators & associated components and demonstration of performance

More information

RULES OF TENNESSEE DEPARTMENT OF ENVIRONMENT AND CONSERVATION DIVISION OF RADIOLOGICAL HEALTH CHAPTER USE OF X-RAY APPARATUS

RULES OF TENNESSEE DEPARTMENT OF ENVIRONMENT AND CONSERVATION DIVISION OF RADIOLOGICAL HEALTH CHAPTER USE OF X-RAY APPARATUS RULES OF TENNESSEE DEPARTMENT OF ENVIRONMENT AND CONSERVATION DIVISION OF RADIOLOGICAL HEALTH CHAPTER 0400-20-06 USE OF X-RAY APPARATUS TABLE OF CONTENTS 0400-20-06-.01 Purpose 0400-20-06-.06 Veterinary

More information

NEMA XR X-ray Equipment for Interventional Procedures User Quality Control Mode

NEMA XR X-ray Equipment for Interventional Procedures User Quality Control Mode NEMA XR 27-2012 X-ray Equipment for Interventional Procedures User Quality Control Mode Published by: National Electrical Manufacturers Association 1300 North 17th Street, Suite 1752 Rosslyn, Virginia

More information

I. Introduction.

I. Introduction. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Accuracy of measuring half- and quarter-value layers and appropriate aperture width of a convenient method using a lead-covered case

More information

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage

More information

Ch. 223 VETERINARY MEDICINE CHAPTER 223. VETERINARY MEDICINE GENERAL PROVISIONS X-RAYS RADIOACTIVE MATERIAL. Authority

Ch. 223 VETERINARY MEDICINE CHAPTER 223. VETERINARY MEDICINE GENERAL PROVISIONS X-RAYS RADIOACTIVE MATERIAL. Authority Ch. 223 VETERINARY MEDICINE 25 223.1 CHAPTER 223. VETERINARY MEDICINE Sec. 223.1. Purpose and scope. 223.2. [Reserved]. 223.2a. Definitions. 223.3 223.6. [Reserved]. 223.7. Structural shielding. 223.8.

More information

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R Building 3, No.9, Jiuxianqiaodong Road, Chaoyang District, Beijing 100015, P.R. China E-mail:

More information

MAN Revision 002. InSight Mini C-arm Imaging System Technical Reference Manual

MAN Revision 002. InSight Mini C-arm Imaging System Technical Reference Manual MAN-00668 Revision 002 InSight Mini C-arm Imaging System Technical Reference Manual April 2007 The information contained in this Manual is confidential and proprietary to Hologic, Inc. This information

More information

ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone

ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY 12477 (800)VIP-XRAY (845)247-9028 Fax or 800-847-9729 local phone 845-246-9013 http://www.eccxray.com sales@eccxray.com INSTRUCTION MANUAL

More information

Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision Minnesota Rules, Chapter 4732 X-ray Revision PROPOSED DRAFT GAUGING X-RAY SYSTEMS, 1.0 4732.#### GAUGING X-RAY SYSTEMS. Subpart 1. Applicability. A registrant s gauging x-ray system must comply with the

More information

GE Sensing & Inspection Technologies. The ISOVOLT Titan E X-ray Generator

GE Sensing & Inspection Technologies. The ISOVOLT Titan E X-ray Generator GE Sensing & Inspection Technologies The Titan E X-ray Generator Robust, Reliable and Highly Accurate Stationary X-ray Generators for the Widest Range of Applications. The Reference Class for X-ray generators

More information

35080M kvp Divider. Operators Manual

35080M kvp Divider. Operators Manual 35080M kvp Divider Operators Manual July 2006 Manual No. 1210001000 Rev. 3 2004, 2005, 2006 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective

More information

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE A. Miceli ab, R. Thierry a, A. Flisch a, U. Sennhauser a, F. Casali b a Empa - Swiss Federal Laboratories for

More information

RaySafe X2. Effortless measurements of X-ray

RaySafe X2. Effortless measurements of X-ray RaySafe X2 Effortless measurements of X-ray At your fingertips We ve grown accustomed to intuitive interactions with our devices. After all, it s not the device that s most important, but what you can

More information

APPENDIX K UNF RADIATION GENERATING DEVICE SAFETY PROCEDURES

APPENDIX K UNF RADIATION GENERATING DEVICE SAFETY PROCEDURES APPENDIX K UNF RADIATION GENERATING DEVICE SAFETY PROCEDURES Policy and Purpose This policy provides administrative control over the use of radiation generating devices and is designed to ensure that such

More information

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah X-RAY IMAGING EE 472 F2017 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Stewart C. Bushong, Radiologic Science for Technologists: Physics, Biology, and Protection, 10 th ed., Mosby,

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

THE PANTAK THERAPAX SXT 150

THE PANTAK THERAPAX SXT 150 THE PANTAK THERAPAX SXT 150 combines the safety and ease-of-use of photon radiation therapy with state-of-the-art technology, including microprocessor control and solid-state electronics. Using Therapax,

More information

Ludlum Medical Physics

Ludlum Medical Physics Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to

More information

QC in Diagnostic Radiology. Main steps for a QC survey in Diagnostic Radiology

QC in Diagnostic Radiology. Main steps for a QC survey in Diagnostic Radiology EVALUATING X-RAY TUBE AND GENERATOR PERFORMANCE : DEMO for PRACTICAL QUALITY CONTROL (QC) Dr Slavik Tabakov Dept. Medical Eng. & Physics, King's College London slavik.tabakov@kcl.ac.uk QC in Diagnostic

More information

(a) This subchapter applies to installations using analytical x-ray equipment and establishes requirements for their use.

(a) This subchapter applies to installations using analytical x-ray equipment and establishes requirements for their use. SUBCHAPTER 21. ANALYTICAL X-RAY INSTALLATIONS 7:28-21.1 Scope (a) This subchapter applies to installations using analytical x-ray equipment and establishes requirements for their use. (b) The provisions

More information

Miniflex. Rigaku/ Miniflex X-ray Diffractometer System. Rigaku Corporation

Miniflex. Rigaku/ Miniflex X-ray Diffractometer System. Rigaku Corporation Miniflex Rigaku/ Miniflex X-ray Diffractometer System Rigaku Corporation Rigaku/ Miniflex X-ray Diffractometer System 1. Introduction Rigaku s general purpose X-ray diffractometer systems are broadly classified

More information

GE Inspection Technologies. The ISOVOLT Titan E X-ray Generator

GE Inspection Technologies. The ISOVOLT Titan E X-ray Generator GE Inspection Technologies The Titan E X-ray Generator Robust, Reliable and Highly Accurate Stationary X-ray Generators for the Widest Range of Applications. The Reference Class for X-ray generators is

More information

Determination of the tube voltage from clinic mammographic system using two types of detectors

Determination of the tube voltage from clinic mammographic system using two types of detectors BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 03-1A (2015) 01-06 Determination of the tube voltage from clinic mammographic system using two types of detectors J. S. Barreira a ; V. Vivolo a a Gerência

More information

By: Georgia Institute of Technology

By: Georgia Institute of Technology Date: / / Page 1 of 19 1.0 PURPOSE 2.0 SCOPE To provide instructions for the registration, use, operation, control, and accountability of radiation machines at the Georgia Institute of Technology (Georgia

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 62220-1 First edition 2003-10 Medical electrical equipment Characteristics of digital X-ray imaging devices Part 1: Determination of the detective quantum efficiency Appareils

More information

CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85

CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85 PD53-012 p Rotating Anode X-ray tube Assembly 0.3/0.8P32&0.6/1.2P32DK-85 GENERAL The Shimadzu 0.3/0.8P32DK-85 & 0.6/1.2P32DK-85, Rotating Anode X-ray tube assemblies are rated to 150kV and feature a 100mm

More information

Diagnostic x-ray equipment compliance and facility survey

Diagnostic x-ray equipment compliance and facility survey Canada Health Canada Canada CA9600871 CA9600871 Diagnostic x-ray equipment compliance and facility survey Diagnostic x-ray equipment compliance and facility survey Recommended procedures for equipment

More information

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT Quality control of Gamma Camera By Dr/ Ibrahim Elsayed Saad 242 NMT WHAT IS QUALITY? The quality of a practice is to fulfill the expectations and demands from: Patient Clinicain Your self Quality assurance

More information

Technical data CAMARGUE CS-VH50/300. VARIABLE Height Bucky Table With Ceiling Suspension

Technical data CAMARGUE CS-VH50/300. VARIABLE Height Bucky Table With Ceiling Suspension Technical data VARIABLE Height Bucky Table With Ceiling Suspension Model Variations CAMARGUE FH (Fixed Height) CAMARGUE FH Tomo CAMARGUE FH Ceiling suspension CAMARGUE VH (Variable Height) CAMARGUE VH

More information

MILADY. Product Data. Page 1 of 8

MILADY. Product Data. Page 1 of 8 Page 1 of 8 The MILADY Mammographic Unit offers the best quality-to-price ratio to our customers worldwide. The unit advanced technology together with the application of industrial production standards,

More information

MASINZA, ALWYN STANSLAUS. BSc (Hons.), PGDSc

MASINZA, ALWYN STANSLAUS. BSc (Hons.), PGDSc By MASINZA, ALWYN STANSLAUS BSc (Hons.), PGDSc A thesis submitted in partial fulfilment for the award of degree of Master of Scienc (Physics), University o f Nairobi. December, 2012 ! Declaration This

More information

Version 1.0. TechnicVR. Student Guide

Version 1.0. TechnicVR. Student Guide Version 1.0 TechnicVR s h a d e r w a r e. c o m Student Guide TechnicVR s h a d e r w a r e. c o m Student Guide shaderware 2008 PO Box 103 Saltburn Cleveland TS12 1WP w w w. s h a d e r w a r e. c o

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

Rotating Anode X-ray Tube Assembly

Rotating Anode X-ray Tube Assembly RADII H18X H18Y Datasheet Rotating Anode X-ray Tube Assembly The tube unit is so constructed that an X-ray tube is sealed in a diagnostic type protective tube housing of electric shockproof, radiation

More information

RaySafe Xi. User Manual

RaySafe Xi. User Manual RaySafe Xi User Manual 2013.02 Unfors RaySafe 5000071-L All rights are reserved. Reproduction or transmission in whole or in part, in any form or by any means, electronic, mechanical or otherwise, is prohibited

More information

Dosepix Detector as kvp-meter in Radiology and Mammography: First steps

Dosepix Detector as kvp-meter in Radiology and Mammography: First steps Dosepix Detector as kvp-meter in Radiology and Mammography: First steps F.Bisello, I.Ritter, F.Tennert, A.Zang MediPix Collaboration Meeting, 19th February 2014, CERN Protect, Enhance, and Save Lives -

More information

Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision Minnesota Rules, Chapter 4732 X-ray Revision DRAFT NON-MEDICAL HAND-HELD X-RAY SYSTEMS, 2.0 4732.#### NON-MEDICAL HAND-HELD X-RAY SYSTEMS. Summary of Changes MDH made a number of changes to the Non-Medical

More information

Corsix R 32 Accu PRODUCT DATA

Corsix R 32 Accu PRODUCT DATA PRODUCT DATA Product data Rev. 1 (April 2011) Page Description and Configurations 3 Technical Features 4 Radiological characteristics 4 Operating characteristics 5 Mechanical characteristics 6 Size and

More information

X-ray Tube and Generator Basic principles and construction

X-ray Tube and Generator Basic principles and construction X-ray Tube and Generator Basic principles and construction Dr Slavik Tabakov - Production of X-rays and Patient Dose OBJECTIVES - X-ray tube construction - Anode - types, efficiency - Classical X-ray generator

More information

X-ray Tube and Generator Basic principles and construction

X-ray Tube and Generator Basic principles and construction X-ray Tube and Generator Basic principles and construction Dr Slavik Tabakov - Production of X-rays OBJECTIVES - X-ray tube construction - Anode - types, efficiency - X-ray tube working characteristics

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-706 Patient Phantom/Penetrometer System Users Manual March 2005 Manual No. 07-706-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are

More information

Technical aspects on DAP calibration and CT calibration

Technical aspects on DAP calibration and CT calibration EURADOS Report 2015-03 European Radiation Dosimetry Group e. V. Braunschweig, August 2015 Technical aspects on DAP calibration and CT calibration H. Järvinen, O. Turak, M. Ginjaume, J. Daures, C. Hourdakis,

More information

Estimation of signal transfer property for wireless digital detector in different measurement schemes

Estimation of signal transfer property for wireless digital detector in different measurement schemes Estimation of signal transfer property for wireless digital detector in different measurement schemes Anatoli Vladimirov, Kalle Kepler Training Centre of Medical Physics, University of Tartu, Estonia 11

More information

Maximizing clinical outcomes

Maximizing clinical outcomes Maximizing clinical outcomes Digital Tomosynthesis Dual Energy Subtraction Automated Long Length Imaging Improved image quality at a low dose Xray Xray Patented ISS capture technology promotes high sensitivity

More information

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW Digital RF System - HF51-5

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW Digital RF System - HF51-5 China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW Digital RF System - HF51-5 #3, No.9, Jiuxianqiaodong Road, Chaoyang District, Beijing 100015, P.R. China E-mail: international@wandong.com.cn

More information

MOXTEK. 50kV 10 Watt MAGNUM X-ray Source. X-ray Sources. Contents

MOXTEK. 50kV 10 Watt MAGNUM X-ray Source. X-ray Sources. Contents X-ray Sources 0kV 10 Watt MAGNUM X-ray Source Manual Contents MAGNUM X-ray Source Characteristics Initial Inspection and Handling Tube Setup Operating Conditions Operating Precautions and Warnings Operating

More information

Industrial X-ray Tube KL

Industrial X-ray Tube KL Description stationary anode x-ray tube is a highly popular and proven side window x-ray tube, which has focus 1.0 mm, is designed for non-destructive x-ray inspection. stationary anode x-ray tube can

More information

Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision Minnesota Rules, Chapter 4732 X-ray Revision PROPOSED DRAFT NON-MEDICAL HAND-HELD X-RAY SYSTEMS, 1.0 4732.#### NON-MEDICAL HAND-HELD X-RAY SYSTEMS. Subpart 1. Applicability. A registrant s non-medical

More information

Radiological Safety Analysis Document for the CLAS12 Engineering and the first physics run of Run Group A

Radiological Safety Analysis Document for the CLAS12 Engineering and the first physics run of Run Group A Radiological Safety Analysis Document for the CLAS12 Engineering and the first physics run of Run Group A This Radiological Safety Analysis Document (RSAD) will identify the general conditions associated

More information

New spectral benefi ts, proven low dose

New spectral benefi ts, proven low dose New spectral benefi ts, proven low dose Philips MicroDose mammography SI, technical data sheet Philips MicroDose SI with single-shot spectral imaging is a fullfi eld digital mammography solution that delivers

More information

Distributed by. Ecotron Anyview Series. The Intelligent C-Arm System

Distributed by. Ecotron Anyview Series. The Intelligent C-Arm System Distributed by Ecotron Anyview Series The Intelligent C-Arm System The Ecotron Anyview Series is a new standard for C-arm imaging Excellent radiographic imaging at an economical price Advanced X-ray Generator

More information

Unfors EDD-30 Radiation Protection in Fluoroscopy

Unfors EDD-30 Radiation Protection in Fluoroscopy Unfors EDD-30 Radiation Protection in Fluoroscopy Immediate Warning Decrease Your Dose Interventional radiology procedures are considered to be essential to medical diagnosis and treatment. It is recognized,

More information

GE AMX 4+ Portable X-Ray

GE AMX 4+ Portable X-Ray GE AMX 4+ Portable X-Ray Typical Manufacturer s Picture GE Healthcare s AMX-4+ analog X-ray system provides high-performance in a compact, easy-to-maneuver package. The rotating arm and tube simplify positioning

More information

Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision Minnesota Rules, Chapter 4732 X-ray Revision DRAFT FLUOROSCOPIC X-RAY SYSTEMS, 1.0 Subpart 1. Applicability. Subpart 2. Limitation of the useful beam. Subpart 3. Measuring compliance; primary protective

More information

Cylindrical Ion Chambers Victoreen Model 550 Series

Cylindrical Ion Chambers Victoreen Model 550 Series Cylindrical Ion Chambers Victoreen Model 550 Series! Cylindrical Ion Chambers for use with Model 35040 and Model 530 electrometers! Wide range of applications in Diagnostic X-Ray and Radiation Oncology

More information

ABOVETABLE X-RAY SOURCE FLUOROSCOPIC AND SPOT-FILM SYSTEMS

ABOVETABLE X-RAY SOURCE FLUOROSCOPIC AND SPOT-FILM SYSTEMS PART VI ABOVETABLE X-RAY SOURCE FLUOROSCOPIC AND SPOT-FILM SYSTEMS FORM FDA 3069 REPRINTED April 2000 ROUTINE COMPLIANCE TESTING ABOVETABLE X-RAY SOURCE FLUOROSCOPIC AND SPOT-FILM SYSTEMS (Test Procedure

More information

Minnesota Rules, Chapter 4732 X-ray Revision PROPOSED NON-MEDICAL CABINET X-RAY SYSTEMS, 1.0

Minnesota Rules, Chapter 4732 X-ray Revision PROPOSED NON-MEDICAL CABINET X-RAY SYSTEMS, 1.0 Minnesota Rules, Chapter 4732 X-ray Revision PROPOSED NON-MEDICAL CABINET X-RAY SYSTEMS, 1.0 4732.#### NON-MEDICAL CABINET X-RAY SYSTEMS Subpart 1. Applicability. A registrant s non-medical cabinet x-ray

More information

Nuclear Associates &

Nuclear Associates & Nuclear Associates 06-526 & 06-526-2200 RAD-CHECK PLUS Operators Manual March 2005 Manual No. 136201 Rev. 5 2003, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

PART H RADIATION SAFETY REQUIREMENTS FOR NON-HEALING ARTS RADIATION GENERATING DEVICES (RGD)

PART H RADIATION SAFETY REQUIREMENTS FOR NON-HEALING ARTS RADIATION GENERATING DEVICES (RGD) Sec. H.1-H.4 SSRCR Volume I - July 2016 PART H RADIATION SAFETY REQUIREMENTS FOR NON-HEALING ARTS RADIATION GENERATING DEVICES (RGD) Sec. H.1 - Purpose. This Part provides special requirements for non-healing

More information

CyberKnife Iris Beam QA using Fluence Divergence

CyberKnife Iris Beam QA using Fluence Divergence CyberKnife Iris Beam QA using Fluence Divergence Ronald Berg, Ph.D., Jesse McKay, M.S. and Brett Nelson, M.S. Erlanger Medical Center and Logos Systems, Scotts Valley, CA Introduction The CyberKnife radiosurgery

More information

Digital Panoramic X- Ray unit. Product Data

Digital Panoramic X- Ray unit. Product Data Digital Panoramic X- Ray unit Product Data Standard examination programs Standard panoramic: adult/child panoramic exam TMJ open/close mouth: 4 slices are taken in the same image (left/right condyle, open/close

More information

CT Scanner Dose Survey

CT Scanner Dose Survey CT Scanner Dose Survey Measurement Protocol Version 5.0 July 1997 Co-ordinated by ImPACT and The Medical Physics Department,, London SW17, UK. 0181-725-3366 CT Scanner Dose Survey: Measurement Protocol

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

DETECTORS. Including Codes of Practice IONIZING RADIATION

DETECTORS. Including Codes of Practice IONIZING RADIATION DETECTORS Including Codes of Practice 2015 IONIZING RADIATION PTW History and General Remarks Looking back on a long history PTW-Freiburg is an internationally operating company, manufacturing and marketing

More information

THE KENYA POWER AND LIGHTING CO. LTD. SPECIFICATIONS for IMPULSE VOLTAGE TEST EQUIPMENT

THE KENYA POWER AND LIGHTING CO. LTD. SPECIFICATIONS for IMPULSE VOLTAGE TEST EQUIPMENT THE KENYA POWER AND LIGHTING CO. LTD SPECIFICATIONS for IMPULSE VOLTAGE TEST EQUIPMENT NAME DESIGNATION SIGNATURE DATE Compiled by Ephantus Nyagah Ag. Senior Engineer Checked by Eng. Thagichu Kiiru Deputy

More information

ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER

ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 325 ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER ABSTRACT William Chang, Jonathan Kerner, and Edward

More information