Advancing EDS Analysis in the SEM Quantitative XRF. International Microscopy Congress, September 5 th, Outline

Size: px
Start display at page:

Download "Advancing EDS Analysis in the SEM Quantitative XRF. International Microscopy Congress, September 5 th, Outline"

Transcription

1 Advancing EDS Analysis in the SEM with in-situ Quantitative XRF Brian J. Cross (1) & Kenny C. Witherspoon (2) 1) CrossRoads Scientific, El Granada, CA 94018, USA 2) ixrf Systems, Inc., Houston, TX 77059, USA International Microscopy Congress, September 5 th, Outline Why use XRF in the SEM? Low-power x-ray x microtubes External vs. re-entrant entrant designs Optics, collimation and shielding Performance characteristics MDLs Applications Quantitative analysis Point analysis & mapping IMC-16, Sapporo, Japan 2 1

2 Why use XRF in the SEM? Better x-ray x sensitivity at energies >2 kev Improved peak-to to-background ratios EDS count rates are limited more counts in peaks with XRF vs. SEM (high backgrounds) Lower Minimum Detection Limits (MDLs) At higher energies (medium-high Z elements) Analysis in air possible Almost no specimen charging IMC-16, Sapporo, Japan 3 Theoretical Comparison of XRF & SEM Sensitivities SEM-EDS EDS better <2 kev; XRF x 1000x sensitivity >5 kev IMC-16, Sapporo, Japan 4 2

3 Some disadvantages of XRF Spatial resolution usually much worse microns vs. <2 microns for SEM-EDS EDS Sample imaging not possible X-ray mapping with stage usually slower than with electron beam Low energy sensitivities (<1-2 2 kev) are better for SEM-EDS EDS than XRF (w/eds) IMC-16, Sapporo, Japan 5 Advantages of Integrated XRF Only one instrument required Smaller footprint in the lab SEM chambers typically have spare port Less expensive than two separate instruments Use common detector, electronics, stage, chamber, computer and software Use both techniques on same sample In situ analysis, without sample transfer IMC-16, Sapporo, Japan 6 3

4 Modern Transmission-Target Target Tubes Thin film targets deposited on Be window Elements in the range Mo Ag, operating at ~30 kv, key parameters are: Mean electron energy range = ~ microns Integrated path length = ~2-3 3 microns Optimum thickness anode is ~ microns High-energy electrons are mostly stopped in target Self-absorption of low-energy X rays is limited Be window provides final electron barrier & heat conductor for electron energy transfer Typical Be window thickness is ~250 microns IMC-16, Sapporo, Japan 7 fx TM Integrated fx TM Microtube Assembly Assembly ready for mounting on SEM Transmission-target tube with integral high-voltage transformer. Re-entrant entrant design allows close coupling of tube to sample. IMC-16, Sapporo, Japan 8 4

5 fx Tube Mounted on Hitachi SEM Left picture shows fx mounted on Hitachi S3000 SEM Right diagram shows cross-section section diagram Note horizontal mounting in this example, requires vertical stage rotation fx design is re-entrant, entrant, with tube inside SEM IMC-16, Sapporo, Japan 9 X-Ray Optics Enabling technology for microbeam XRF Simple apertures for beam collimation wide bandpass Curved mirrors low bandpass Curved crystals narrow bandpass Formed and graded multilayers rel. narrow bandpass Mono and Polycapillary lenses rel. wide bandpass True microbeams require microspot x-ray x tubes Spot size on tube target must be <100 micron (ideally <10) Obtain analytical spot sizes in the range micron Larger target spot sizes limited to use w/apertures for x-ray x beam sizes in the range ~ mm IMC-16, Sapporo, Japan 10 5

6 fx Tube Optics Transmission-target tube Demountable collimator assembly Collimator assembly contains both front and back apertures Spectrum with and without an initial aperture in front of Be window 1st aperture restricts view to center of target, away from contamination regions 2nd aperture provides collimation of the beam at working distances of mm from the front aperture Total Counts Total Counts Typical Ag Tube Output Spectrum 35 kv, No pinhole aperture 5000 Ni Kα Cu Kα Fe Kα 2000 Ag Kα Ag Kβ Channel Number Typical Ag Tube Output Spectrum 35 kv, with 1.5 mm pinhole in front of window Ag Kα Ag Kβ Channel Number IMC-16, Sapporo, Japan 11 fx Tube Optics - 2 Demountable collimator assembly Collimator assembly contains both front and back apertures 1st aperture (~1 mm) restricts view to target center, away from contamination regions 2nd aperture ( mm) provides beam collimation At working distances of mm from front aperture, beam size is ~2x aperture size High-purity materials provide contamination-free spectra Rear Aperture Sleeve Front Aperture Spring Retainer IMC-16, Sapporo, Japan 12 6

7 X-Beam Polycapillary Optics Multiple total external reflection of X rays Large solid angle capture small focal point Performance gain vs. aperture is typically a factor of >100x under identical conditions Polycapillary provides broadband excitation with falloff at higher energies Courtesy of X-ray Optical Systems, Inc. IMC-16, Sapporo, Japan 13 X-Beam for the SEM Tube is external to SEM chamber Distance offset by polycapillary optic Transports collimated beam of X rays to focal spot at sample. Beam at sample is typically microns in diameter Spot size is energy dependent. Integrated automatic shutter mechanism X rays can remain on for stability. Z adjustment, with XY interface flange adjustment, allows alignment of X-Beam X with e-beam. e Slide adjustment changes spot size. Operates at moderate power (50W) with fan cooling. IMC-16, Sapporo, Japan 14 7

8 Picture shows X-BeamX and fx tubes mounted on Hitachi 3000 SEM with EDS detector. XRF Inside the SEM X-Beam on Hitachi 3000 IMC-16, Sapporo, Japan 15 Scatter Spectra from X-Beam X and fx Tubes fx transmission-target target has more high- & low-energy source radiation vs. X-Beam X w/side window + polycapillary optics IMC-16, Sapporo, Japan 16 8

9 Determination of X-Beam X Spot Size Knife-edge scan of Mo Kα x rays directly from the X-Beam Knife-edge scan Derivative of the knife-edge scan Counts FWHM = 24.8 μ m Knife edge position (mm) X-ray image of a Cu grid (125um pitch and 12 um bars) IMC-16, Sapporo, Japan 17 SEM, X-Beam X and fx Spectra for Glass Standard CH2 4% SiO 2 2% As 3 O 2 0.2% ZrO 2 IMC-16, Sapporo, Japan 18 9

10 Comparison of MDLs for E-beam, E X-Beam X and fx Selected MDLs for Na-Mg Mg- Al borosilicate glass std. E-beam kv Specimen carbon coated X-Beam kv Mo anode, polycap. optics fx tube kv Ag transmission-target target Given E-beam X-beam fx Elt (Wt.%) MDL MDL MDL Mg Al Si P Ca V Fe Ge As Sr Zr Mo Sb La Ce W IMC-16, Sapporo, Japan 19 SEM vs. fx Spectra for Glass Standard CH4 Dots: SEM kv Bars: fx 1% ZrO 2 0.1% PbO 1% CdO IMC-16, Sapporo, Japan 20 10

11 Minimum Detection Limits from Glass Standards E-beam X-Beam fx MDL (3-sig Wt.%) Atomic No. IMC-16, Sapporo, Japan 21 SEM & X-Beam X Spectra for Glass Standard SRM 610 NIST SRM 610 Component Wt% SiO CaO NaO Al2O ppm Co 390 Cu 444 Fe 458 Mn 485 Ni 458 Ag 254 Sr 515 Th 457 Ti 437 Pb 426 K 461 Ru 425 Th 61 U 461 Zn 433 Overlay e-beam spectrum on XRF (X-Beam) Peak fitting to X-beam SRM 610 spectrum IMC-16, Sapporo, Japan 22 11

12 SEM and fx Spectra from Al Alloy 2014 Dots: SEM kv Bars: fx 0.02% V 0.03% Pb IMC-16, Sapporo, Japan 23 Strategies for Quantitative Analysis Analyze same sample by both SEM-EDS EDS and XRF Assume sample is uniform within XRF beam area How to combine results from 2 techniques? Analyze both and select concentrations from each? Many times can t t analyze all elements (not detected)! Fix elements from one analysis while doing other? Does not work - ratio of two element totals is unknown! Iteratively fix some concentrations while analyze others and do vice versa for the other method Difficult and painful, and may not work very well! Solution is do both analyses simultaneously IMC-16, Sapporo, Japan 24 12

13 Combined SEM-XRF Analysis Standardless ZAF for SEM-EDS EDS k = c Z A Fc where, k = I/I p = intensity ratio to pure elt. and, I p = ƒ(p, (p,ω,r,q, )) can be calculated for each element on a relative basis. c is the weight fraction for each element Because absolute count rates are not calculated, assume that c i = 1. Standardless FP for XRF Similar set of equations (but different physics) IMC-16, Sapporo, Japan 25 Combined SEM-XRF Analysis - Solution Use common equation c i = 1 Can t t normalize pure intensities Relationship between ZAF and FP not known. Solution is to normalize the k-ratiosk Done iteratively in the same loop as the concentration normalization. Partial sum done for each method (ZAF, FP). Total of two sums must be 1. Update concentration estimates for both ZAF and FP at the end of each iteration. IMC-16, Sapporo, Japan 26 13

14 Combined SEM-XRF Analysis of AA 2014 Alloy Set both spectra to analyze same elements Choose XRF (FP) or SEM-EDS (ZAF) method for each element, or set to Automatic. Choose appropriate x-ray lines for each method IMC-16, Sapporo, Japan 27 Compare ZAF, FP & Combined Analysis All AA 2014 analyses were done by a standardless method Quantitative method chosen automatically Based on Peak-to to- Background Error ratios Main source of quant error is the spectrum model for the fx tube Can reduce by using FP calibration for XRF Mg 0.45 Ka 1.02 Ka 1.45 Ka ZAF 0.48 Al Ka Ka Ka ZAF Si 0.89 Ka 0.46 Ka 1.21 Ka FP 0.85 Ti Ka 0.07 Ka Ka FP V Ka 0.01 Ka Ka ZAF Cr Ka 0.07 Ka Ka FP Mn 0.77 Ka 0.62 Ka 0.48 Ka ZAF 0.63 Fe 0.48 Ka 0.35 Ka 0.27 Ka ZAF 0.35 Ni 0.04 Ka 0.09 Ka 0.03 Ka FP 0.10 Cu 4.51 Ka 3.74 Ka 1.71 Ka FP 5.34 Zn Ka 0.05 Ka Ka FP Ga Ka 0.07 Ka Ka FP Zr La 0.04 Ka Ka FP Sn La 0.10 Ka Ka FP Pb Ma 0.04 La La FP Bi Ma 0.22 La 0.01 La FP Used automatically selected methods, except V, Mn and Fe changed to SEM spectrum method to avoid diffraction peaks in XRF spectrum IMC-16, Sapporo, Japan 28 14

15 XRF Multilayer Thin-Film Applications Analyze up to 6 layers for thickness & composition Calibrate with single or multi-element thin-film standards Example shows 6-layer stack with 1 component per layer IMC-16, Sapporo, Japan 29 X-Beam Map of Standards Block IMC-16, Sapporo, Japan 30 15

16 X-Beam XRF Mapping of Thin Rock Section Thin-rock section Full map spectra from X-beam X (bars) and e-beam e (dots) XY scans. IMC-16, Sapporo, Japan 31 X-Beam XRF Mapping of Mining Sample Composite of Si, Cl, K & Fe IMC-16, Sapporo, Japan 32 16

17 XRF Elemental Maps of Printed Circuit Board 350x120 Maps Composite XRF Spectrum IMC-16, Sapporo, Japan 33 XRF Composite Maps of Printed Circuit Board Cu Al Pb Pd Ca Zn Ag Sb Mo Incoherent Photo PCB on C Stub Mo Coherent IMC-16, Sapporo, Japan 34 17

18 Chip Area X-Beam X Map and Integral Spectra Bars: XRF Composite Spectrum Overlay: SEM kv IMC-16, Sapporo, Japan 35 Chip Area SEM-EDS EDS X-Map X and Integral Spectrum IMC-16, Sapporo, Japan 36 18

19 Composite X-ray X Maps of Chip Area Composite XRF Map Ca Zn Zr Si Composite SEM-EDS Map Ca Zn O Bi Pd Ag-L Pb Pd Ag Pb Cu Fe Br Si Si Pb Al Pb Si C IMC-16, Sapporo, Japan 37 Summary 1 fx Re-entrant entrant Microtube Integrated re-entrant entrant transmission-target target micro x-ray x tube inside several SEMs. Thin target (1-2 2 microns) important for wide energy range of excitation. Long & small diameter of tube important for getting close coupling with sample. Low power (<5W) allows passive heat conduction no active cooling required. Sample analysis sizes from ~0.5 5 mm. IMC-16, Sapporo, Japan 38 19

20 Summary 2 External X-Beam X w/optics X-Beam has side-window microbeam x-ray x tube and capillary optics Moderate-power (50W) use requires fan cooling. Polycapillary optic has wide energy range, but less efficient at higher & lower energies. Interlocked integrated shutter mechanism allows tube to remain on for stability. Use stage Z-axis Z and tube XY interface flange to align X-Beam to E-beam. E Typical performance of better than 50 x 90 micron (FWHM) beam at the sample, with 25x45 possible. XY stage scanning allows the acquisition of x-ray x maps and line scans. IMC-16, Sapporo, Japan 39 Conclusions Demonstrated integration of several x-ray x tubes with various SEMs. Advantage of XRF for trace analysis is complementary to SEM-EDS EDS analysis. XRF software added to typical EDS analysis package to allow separate or combined qualitative and quant analysis. Stage automation allows x-ray x chemical mapping using electron or x-ray x beams. IMC-16, Sapporo, Japan 40 20

WIDE ANGLE GEOMETRY EDXRF SPECTROMETERS WITH SECONDARY TARGET AND DIRECT EXCITATION MODES

WIDE ANGLE GEOMETRY EDXRF SPECTROMETERS WITH SECONDARY TARGET AND DIRECT EXCITATION MODES Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 11 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42

More information

ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER

ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 325 ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER ABSTRACT William Chang, Jonathan Kerner, and Edward

More information

MICRO XRF OF LIGHT ELEMENTS USING A POLYCAPILLARY LENS AND AN ULTRA THIN WINDOW SILICON DRIFT DETECTOR INSIDE A VACUUM CHAMBER

MICRO XRF OF LIGHT ELEMENTS USING A POLYCAPILLARY LENS AND AN ULTRA THIN WINDOW SILICON DRIFT DETECTOR INSIDE A VACUUM CHAMBER Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 229 MICRO XRF OF LIGHT ELEMENTS USING A POLYCAPILLARY LENS AND AN ULTRA THIN WINDOW SILICON DRIFT

More information

MICROANALYSIS WITH A POLYCAPILLARY IN A VACUUM CHAMBER

MICROANALYSIS WITH A POLYCAPILLARY IN A VACUUM CHAMBER THE RIGAKU JOURNAL VOL. 20 / NO. 2 / 2003 MICROANALYSIS WITH A POLYCAPILLARY IN A VACUUM CHAMBER CHRISTINA STRELI a), NATALIA MAROSI, PETER WOBRAUSCHEK AND BARBARA FRANK Atominstitut der Österreichischen

More information

S1 TITAN Alloy LE Calibrations (P/N: )

S1 TITAN Alloy LE Calibrations (P/N: ) S1 TITAN 600-800 Alloy LE Calibrations () Low Alloy Si P S Ti V Cr Mn Fe Co Ni Cu Nb Mo W Pb Analysis range, % LLD-2 LLD-0.15 LLD-0.3 LLD - 0.1 0.05-1.8 LLD - 9 0.1-2.0 75-100 LLD - 8 LLD - 5 LLD - 5 LLD-

More information

Qualitative analysis tutorial for Tracer III SD and V+ data

Qualitative analysis tutorial for Tracer III SD and V+ data Qualitative analysis tutorial for Tracer III SD and V+ data Outline What does the spectrum mean? What to watch out for How to normalize Using Artax What does the spectrum mean? Bremsstrahlung radiation

More information

ARTAX. Innovation with Integrity. Portable Micro-XRF Spectrometer. Micro-XRF

ARTAX. Innovation with Integrity. Portable Micro-XRF Spectrometer. Micro-XRF ARTAX Portable Micro-XRF Spectrometer Innovation with Integrity Micro-XRF ARTAX Elemental Analysis for the Art Community and More The ARTAX is the first portable X ray fluorescence (XRF) spectrometer designed

More information

M4 TORNADO PLUS. Innovation with Integrity. Super Light Element Micro-XRF Spectrometer. Micro-XRF

M4 TORNADO PLUS. Innovation with Integrity. Super Light Element Micro-XRF Spectrometer. Micro-XRF M4 TORNADO PLUS Super Light Element Micro-XRF Spectrometer Innovation with Integrity Micro-XRF M4 TORNADO PLUS - A New Era in Micro-XRF M4 TORNADO PLUS is the world's first Micro-XRF spectrometer that

More information

BRUKER ADVANCED X-RAY SOLUTIONS. SPECTROMETRY SOLUTIONS ARTAX mxrf SPECTROMETER

BRUKER ADVANCED X-RAY SOLUTIONS. SPECTROMETRY SOLUTIONS ARTAX mxrf SPECTROMETER BRUKER ADVANCED X-RAY SOLUTIONS SPECTROMETRY SOLUTIONS ARTAX mxrf SPECTROMETER Microanalysis ARTAX Elemental Analysis for the Art Community and More Non-destructive elemental analysis is strictly required

More information

DEVELOPMENT OF A WAVELENGTH DISPERSIVE X-RAY FLUORESCENCE SPECTROMETER USING A MULTI-CAPILLARY X-RAY LENS FOR X-RAY DETECTION

DEVELOPMENT OF A WAVELENGTH DISPERSIVE X-RAY FLUORESCENCE SPECTROMETER USING A MULTI-CAPILLARY X-RAY LENS FOR X-RAY DETECTION Copyright JCPDS - International Centre for Diffraction Data 2003, Advances in X-ray Analysis, Volume 46. 346 DEVELOPMENT OF A WAVELENGTH DISPERSIVE X-RAY FLUORESCENCE SPECTROMETER USING A MULTI-CAPILLARY

More information

Lab Report XRF 441 Elemental distribution analysis on geological samples with the M4 TORNADO

Lab Report XRF 441 Elemental distribution analysis on geological samples with the M4 TORNADO Bruker Nano Spectrum Geological sample M4 TORNADO Quantification Lab Report XRF 441 Elemental distribution analysis on geological samples with the M4 TORNADO Geological samples are inhomogeneous. The distribution

More information

Applications Information

Applications Information Applications Information Window Materials % TRANSMISSION 100 90 80 70 60 50 40 30 20 10 UV Sapphire UV Quartz Pyrex & Glass 100 200 300 400 500 600 700 800 900 Wavelength (nm) Pyrex only In applications

More information

LONG TERM STATISTICS OF X-RAY SPECTROMETERS

LONG TERM STATISTICS OF X-RAY SPECTROMETERS 403 LONG TERM STATISTICS OF X-RAY SPECTROMETERS J. F. Dlouhy*, D. Mathieu Department of the Environment, Environmental Technology Center, River Road, Ottawa, Ontario, Canada Kl A OH3 K. N. Stoev Bulgarian

More information

Rigaku Innovative Technologies Europe (RITE) Presented by: Dr.Peter Oberta

Rigaku Innovative Technologies Europe (RITE) Presented by: Dr.Peter Oberta Rigaku Innovative Technologies Europe (RITE) Presented by: Dr.Peter Oberta 1 Introducing Rigaku Since its inception in Japan in 1951, Rigaku has been at the forefront of analytical and industrial instrumentation

More information

Bruker Nano. M4 tornado. High performance micro-xrf spectrometer. think forward

Bruker Nano. M4 tornado. High performance micro-xrf spectrometer. think forward Bruker Nano M4 tornado High performance micro-xrf spectrometer think forward µ-xrf M4 TORNADO setting standards in µ-xrf µ-xrf is the method of choice for highly sensitive and non-destructive elemental

More information

X-Ray Spectroscopy with a CCD Detector. Application Note

X-Ray Spectroscopy with a CCD Detector. Application Note X-Ray Spectroscopy with a CCD Detector In addition to providing X-ray imaging solutions, including CCD-based cameras that image X-rays using either direct detection (0.5-20 kev) or indirectly using a scintillation

More information

AutoMATE II. Micro-area X-ray stress measurement system. Highly accurate micro area residual stress

AutoMATE II. Micro-area X-ray stress measurement system. Highly accurate micro area residual stress AutoMATE II Micro-area X-ray stress measurement system Highly accurate micro area residual stress The accuracy of an R&D diffractom dedicated residua In the past, if you wanted to make highly accurate

More information

XRF Instrumentation. Introduction to spectrometer

XRF Instrumentation. Introduction to spectrometer XRF Instrumentation Introduction to spectrometer AMPTEK, INC., Bedford, MA 01730 Ph: +1 781 275 2242 Fax: +1 781 275 3470 sales@amptek.com 1 Instrument Excitation source Sample X-ray tube or radioisotope

More information

Applications of Micro XRF for the Analysis of Traditional Japanese "Ainu" Glass Beads and other Artifacts

Applications of Micro XRF for the Analysis of Traditional Japanese Ainu Glass Beads and other Artifacts 161 161 Applications of Micro XRF for the Analysis of Traditional Japanese "Ainu" Glass Beads and other Artifacts K.Sugihara 1, M.Satoh 1, Y.Hayakawa 2, A.Saito 3 and T.Sasaki 4 1 Seiko Instruments Inc.,

More information

Zaidi Embong and Husin Wagiran Physics Department, University Of Technology Malaysia, P.O Box 791, 80990, Johor Baharu

Zaidi Embong and Husin Wagiran Physics Department, University Of Technology Malaysia, P.O Box 791, 80990, Johor Baharu MY9800971 Optimization of a Spectrometry for Energy -Dispersive X-ray Fluorescence Analysis by X-ray Tube in Combination with Secondary Target for Multielements Determination of Sediment Samples. Zaidi

More information

NanoSpective, Inc Progress Drive Suite 137 Orlando, Florida

NanoSpective, Inc Progress Drive Suite 137 Orlando, Florida TEM Techniques Summary The TEM is an analytical instrument in which a thin membrane (typically < 100nm) is placed in the path of an energetic and highly coherent beam of electrons. Typical operating voltages

More information

DTU DANCHIP an open access micro/nanofabrication facility bridging academic research and small scale production

DTU DANCHIP an open access micro/nanofabrication facility bridging academic research and small scale production DTU DANCHIP an open access micro/nanofabrication facility bridging academic research and small scale production DTU Danchip National Center for Micro- and Nanofabrication DTU Danchip DTU Danchip is Denmark

More information

LYNXEYE XE. Innovation with Integrity. High-Resolution Energy-Dispersive Detector for 0D, 1D, and 2D Diffraction XRD

LYNXEYE XE. Innovation with Integrity. High-Resolution Energy-Dispersive Detector for 0D, 1D, and 2D Diffraction XRD High-Resolution Energy-Dispersive Detector for 0D, 1D, and 2D Diffraction The is the first energy dispersive 0D, 1D, and 2D detector operating at room temperature for ultra fast X-ray diffraction measurements.

More information

Introduction of New Products

Introduction of New Products Field Emission Electron Microscope JEM-3100F For evaluation of materials in the fields of nanoscience and nanomaterials science, TEM is required to provide resolution and analytical capabilities that can

More information

ZSX Primus IV Sequential wavelength dispersive X-ray fluorescence. ZSX Primus IV Sequential wavelength dispersive X-ray fluorescence.

ZSX Primus IV Sequential wavelength dispersive X-ray fluorescence. ZSX Primus IV Sequential wavelength dispersive X-ray fluorescence. Sequential wavelength dispersive X-ray fluorescence Sequential wavelength dispersive X-ray fluorescence www.rigaku.com Elemental analysis by WDXRF spectroscopy employing advanced Guidance expert system

More information

FIRST Newsletter March 2013, Issue 20. Elemental Distribution Analysis of a Meteorite Sample from the Rochechouart Structure with the µ-xrf M4 TORNADO

FIRST Newsletter March 2013, Issue 20. Elemental Distribution Analysis of a Meteorite Sample from the Rochechouart Structure with the µ-xrf M4 TORNADO FIRST Newsletter March 2013, Issue 20 Elemental Distribution Analysis of a Meteorite Sample from the Rochechouart Structure with the µ-xrf M4 TORNADO By Dr. Roald Tagle, Ulrich Waldschlager, Dr. Michael

More information

Development of X-ray Tool For Critical- Dimension Metrology

Development of X-ray Tool For Critical- Dimension Metrology Development of X-ray Tool For Critical- Dimension Metrology Boris Yokhin, Alexander Krokhmal, Alexander Dikopoltsev, David Berman, Isaac Mazor Jordan Valley Semiconductors Ltd., Ramat Gabriel Ind. Zone,

More information

X-RAY OPTICS FOR TWO-DIMENSIONAL DIFFRACTION

X-RAY OPTICS FOR TWO-DIMENSIONAL DIFFRACTION Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 332 ABSTRACT X-RAY OPTICS FOR TWO-DIMENSIONAL DIFFRACTION Bob B. He and Uwe Preckwinkel Bruker

More information

The SS6000 Gold Mate Series For analyzing all precious metals and other elements from Mg to U

The SS6000 Gold Mate Series For analyzing all precious metals and other elements from Mg to U The SS6000 Gold Mate Series For analyzing all precious metals and other elements from Mg to U Portable desk top EDXRF analyzers Responsive, bright, color touch screen display Uses Silicon Drift or Silicon

More information

(Refer Slide Time: 00:10)

(Refer Slide Time: 00:10) Fundamentals of optical and scanning electron microscopy Dr. S. Sankaran Department of Metallurgical and Materials Engineering Indian Institute of Technology, Madras Module 03 Unit-6 Instrumental details

More information

LECTURE 10. Dr. Teresa D. Golden University of North Texas Department of Chemistry

LECTURE 10. Dr. Teresa D. Golden University of North Texas Department of Chemistry LECTURE 10 Dr. Teresa D. Golden University of North Texas Department of Chemistry Components for the source include: -Line voltage supply -high-voltage generator -x-ray tube X-ray source requires -high

More information

research papers First results from a macromolecular crystallography system with a polycapillary collimating optic and a microfocus X-ray generator

research papers First results from a macromolecular crystallography system with a polycapillary collimating optic and a microfocus X-ray generator Journal of Applied Crystallography ISSN 0021-8898 First results from a macromolecular crystallography system with a polycapillary collimating optic and a microfocus X-ray generator Received 7 September

More information

Bandpass Interference Filters

Bandpass Interference Filters Precise control of center wavelength and bandpass shape Wide selection of stock wavelengths from 250 nm-1550 nm Selection of bandwidths Available in 1/2 and 1 sizes High peak transmission values Excellent

More information

QUANTAX FlatQUAD. Innovation with Integrity. EDS for SEM with the XFlash FlatQUAD EDS

QUANTAX FlatQUAD. Innovation with Integrity. EDS for SEM with the XFlash FlatQUAD EDS QUANTAX FlatQUAD EDS for SEM with the XFlash FlatQUAD Innovation with Integrity EDS Maximum Efficiency in X-ray Detection... QUANTAX FlatQUAD is the EDS microanalysis system based on the revolutionary

More information

photolithographic techniques (1). Molybdenum electrodes (50 nm thick) are deposited by

photolithographic techniques (1). Molybdenum electrodes (50 nm thick) are deposited by Supporting online material Materials and Methods Single-walled carbon nanotube (SWNT) devices are fabricated using standard photolithographic techniques (1). Molybdenum electrodes (50 nm thick) are deposited

More information

Cr, Co, Cu, Mo, Ag (others on request) Mean Reflectivity: R > 70%

Cr, Co, Cu, Mo, Ag (others on request) Mean Reflectivity: R > 70% PARALLEL BEAM X-RAY OPTICS y Mirror length L Θ = f(x) b p/2 λ = 2d eff (x) sin Θ(x) eff x m Parallel beam width b=f(p,λ,l,,l,x m ) x Fabrication of high precision 6 mm parallel beam optics both on prefigured

More information

plasmonic nanoblock pair

plasmonic nanoblock pair Nanostructured potential of optical trapping using a plasmonic nanoblock pair Yoshito Tanaka, Shogo Kaneda and Keiji Sasaki* Research Institute for Electronic Science, Hokkaido University, Sapporo 1-2,

More information

State-of-the-art thin film X-ray optics for synchrotrons and FEL sources. Frank Hertlein Incoatec GmbH Geesthacht, Germany

State-of-the-art thin film X-ray optics for synchrotrons and FEL sources. Frank Hertlein Incoatec GmbH Geesthacht, Germany State-of-the-art thin film X-ray optics for synchrotrons and FEL sources Frank Hertlein Incoatec GmbH Geesthacht, Germany Incoatec: Innovative Coating Technologies Incoatec is founded with Bruker AXS in

More information

Various beams for RBS at IFIN-HH

Various beams for RBS at IFIN-HH OVERWIEW RBS E sc : function of K and de/dx (mass and depth of target nucleus); Measured spectrum is the sum for all contribution of constitutive elements; Specific Analysis Performed COMPOSITION ANALYSIS:

More information

FAST ELEMENTAL MAPPING WITH MICRO-XRF

FAST ELEMENTAL MAPPING WITH MICRO-XRF 286 FAST ELEMENTAL MAPPING WITH MICRO-XRF Haschke, M.; Rossek, U.; Tagle, R.; Waldschläger, U. Bruker Nano GmbH, 12489 Berlin, Schwarzschildstr.12 ABSTRACT X-Ray optics are now in common use for concentrating

More information

Material analysis by infrared mapping: A case study using a multilayer

Material analysis by infrared mapping: A case study using a multilayer Material analysis by infrared mapping: A case study using a multilayer paint sample Application Note Author Dr. Jonah Kirkwood, Dr. John Wilson and Dr. Mustafa Kansiz Agilent Technologies, Inc. Introduction

More information

Lesson 2 Diffractometers

Lesson 2 Diffractometers Lesson 2 Diffractometers Nicola Döbelin RMS Foundation, Bettlach, Switzerland January 14 16, 2015, Bern, Switzerland Repetition: Generation of X-rays / Diffraction SEM: BSE detector, BSED / SAED detector

More information

A Framed Monochromatic X-Ray Microscope for ICF

A Framed Monochromatic X-Ray Microscope for ICF A Framed Monochromatic X-Ray Microscope for ICF The Laser Fusion Experiments Groups from the Laboratory for Laser Energetics (LLE) and the Los Alamos National Laboratory (LANL) have jointly developed an

More information

Basic P-XRD instructions for Operating the Instrument

Basic P-XRD instructions for Operating the Instrument Basic P-XRD instructions for Operating the Instrument Instrument Parts Incident Beam Optics (left arm) 1) X-ray source (Cu) i. Rest settings: 45 kv, 20mA ii. Run settings: 45 kv, 40mA 2) Monochromator

More information

:... resolution is about 1.4 μm, assumed an excitation wavelength of 633 nm and a numerical aperture of 0.65 at 633 nm.

:... resolution is about 1.4 μm, assumed an excitation wavelength of 633 nm and a numerical aperture of 0.65 at 633 nm. PAGE 30 & 2008 2007 PRODUCT CATALOG Confocal Microscopy - CFM fundamentals :... Over the years, confocal microscopy has become the method of choice for obtaining clear, three-dimensional optical images

More information

Stop Worrying About Interferences With These ICP-OES Solutions

Stop Worrying About Interferences With These ICP-OES Solutions ASTS 2013 Agilent Science & Technology Symposium Stop Worrying About Interferences With These ICP-OES Solutions Steve Wall Agilent Technologies Page 1 Agilent ICP-OES The world's most productive high performance

More information

Processing of MA(or µ)-xrf Data with the M6 software

Processing of MA(or µ)-xrf Data with the M6 software Processing of MA(or µ)-xrf Data with the M6 software Roald Tagle, Max Bügler, Falk Reinhardt, and Ulrich Waldschläger Bruker Nano Berlin Innovation with Integrity Outline 1. Introduction 2. From the object

More information

Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror

Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror Stephen B. Robie scintag, Inc. 10040 Bubb Road Cupertino, CA 95014 Abstract Corundum

More information

OPTIMIZING THE ELEMENTAL SENSITIVITY AND FOCAL SPOT SIZE OF A MONOLITHIC POLYCAPILLARY OPTIC USING MICRO-X-RAY FLUORESCENCE

OPTIMIZING THE ELEMENTAL SENSITIVITY AND FOCAL SPOT SIZE OF A MONOLITHIC POLYCAPILLARY OPTIC USING MICRO-X-RAY FLUORESCENCE , Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 26 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42

More information

Scanning Electron Microscopy SEM. Warren Straszheim, PhD MARL, 23 Town Engineering

Scanning Electron Microscopy SEM. Warren Straszheim, PhD MARL, 23 Town Engineering Scanning Electron Microscopy SEM Warren Straszheim, PhD MARL, 23 Town Engineering wesaia@iastate.edu 515-294-8187 How it works Create a focused electron beam Accelerate it Scan it across the sample Map

More information

Sources & Beam Line Optics

Sources & Beam Line Optics SSRL Scattering Workshop May 16, 2006 Sources & Beam Line Optics Thomas Rabedeau SSRL Beam Line Development Objective/Scope Objective - develop a better understanding of the capabilities and limitations

More information

MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE

MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE 228 MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE D. CARUSO, M. DINSMORE TWX LLC, CONCORD, MA 01742 S. CORNABY MOXTEK, OREM, UT 84057 ABSTRACT Miniature x-ray sources present

More information

sp1 sp2 sp3 sp4 sp5 TAP LPET LPET TAP LLIF Na Kα (albite) Ca Kα (anorthite) K Kα (orthoclase) Mg Kα (forsterite) Mn Kα (rhodonite)

sp1 sp2 sp3 sp4 sp5 TAP LPET LPET TAP LLIF Na Kα (albite) Ca Kα (anorthite) K Kα (orthoclase) Mg Kα (forsterite) Mn Kα (rhodonite) These are the analytical routines I've set up on the University of Arizona Cameca SX100 microprobe for the analysis of silicates, oxides, and other oxy-anion minerals in my FKM thin sections. A pair of

More information

Devices & Services Company

Devices & Services Company Devices & Services Company 10290 Monroe Drive, Suite 202 - Dallas, Texas 75229 USA - Tel. 214-902-8337 - Fax 214-902-8303 Web: www.devicesandservices.com Email: sales@devicesandservices.com D&S Technical

More information

Fluorescence X-ray Spectrometer System ZSX Series

Fluorescence X-ray Spectrometer System ZSX Series The Rigaku Journal Vol. 16/ number 2/ 1999 Product Information Fluorescence X-ray Spectrometer System ZSX Series The ZSX: Innovative XRF Technology-Accelerated. 1 Introduction The ZSX is a revolutionary

More information

Quick and simple installation and no maintenance needed. 3 Times More affordable Than a normal SEM. Obtaining results in less than 4 minutes

Quick and simple installation and no maintenance needed. 3 Times More affordable Than a normal SEM. Obtaining results in less than 4 minutes INTRODUCTION We believe that every laboratory working in the field of nanotechnology needs an SEM, therefore we would like to introduce to you our IEM series of SEM. In short space of time, our device

More information

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Takeshi Fujiwara 1, Yuki Mitsuya 2, Hiroyuki Takahashi 2, and Hiroyuki Toyokawa 2 1 National Institute of Advanced Industrial

More information

X-ray generation by femtosecond laser pulses and its application to soft X-ray imaging microscope

X-ray generation by femtosecond laser pulses and its application to soft X-ray imaging microscope X-ray generation by femtosecond laser pulses and its application to soft X-ray imaging microscope Kenichi Ikeda 1, Hideyuki Kotaki 1 ' 2 and Kazuhisa Nakajima 1 ' 2 ' 3 1 Graduate University for Advanced

More information

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on

More information

Data Collection with. VÅNTEC-2000 Detector

Data Collection with. VÅNTEC-2000 Detector Data Collection with IµS Source and VÅNTEC-2000 Detector D8 System Configuration for Reflection Microfocus Source IµS Optics with Housing 2D Detector (VÅNTEC-2000) DHS 900 Heating Stage Sample Stage Bruker

More information

M6 JETSTREAM. Innovation with Integrity. Large Area Micro X-ray Fluorescence Spectrometer. Micro-XRF

M6 JETSTREAM. Innovation with Integrity. Large Area Micro X-ray Fluorescence Spectrometer. Micro-XRF M6 JETSTREAM Large Area Micro X-ray Fluorescence Spectrometer Innovation with Integrity Micro-XRF Spatially Resolved Elemental Analysis of Large Objects The Bruker M6 JETSTREAM is designed for the nondestructive

More information

USING A CHARGE-COUPLED DEVICE (CCD) TO GATHER X-RAY FLUORESCENCE (XRF)AND X-RAY DIFFRACTION (XRD) INFORMATION SIMULTANEOUSLY

USING A CHARGE-COUPLED DEVICE (CCD) TO GATHER X-RAY FLUORESCENCE (XRF)AND X-RAY DIFFRACTION (XRD) INFORMATION SIMULTANEOUSLY Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 343 USING A CHARGE-COUPLED DEVICE (CCD) TO GATHER X-RAY FLUORESCENCE (XRF)AND X-RAY DIFFRACTION (XRD)

More information

RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS

RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS S-94,316 PATENTS-US-A96698 BEAM POSITION MONITOR RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS CONTRACTUAL ORIGIN OF THE INVENTION The United States Government has rights in this invention pursuant

More information

Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4

Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4 Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4 S.V. Roth, R. Döhrmann, M. Dommach, I. Kröger, T. Schubert, R. Gehrke Definition of the upgrade The wiggler beamline BW4 is dedicated to

More information

PSPC/MDG 2000 X-RAY MICRODIFFRACTOMETER. Product Information

PSPC/MDG 2000 X-RAY MICRODIFFRACTOMETER. Product Information THE RIGAKU JOURNAL VOL. 11 I NO.2 I 1994 Product Information X-RAY MICRODIFFRACTOMETER PSPC/MDG 2000 1. Introduction The analysis of X-ray diffraction patterns is well known as an effective means of obtaining

More information

A Laser-Based Thin-Film Growth Monitor

A Laser-Based Thin-Film Growth Monitor TECHNOLOGY by Charles Taylor, Darryl Barlett, Eric Chason, and Jerry Floro A Laser-Based Thin-Film Growth Monitor The Multi-beam Optical Sensor (MOS) was developed jointly by k-space Associates (Ann Arbor,

More information

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the ECEN 4606 Lab 8 Spectroscopy SUMMARY: ROBLEM 1: Pedrotti 3 12-10. In this lab, you will design, build and test an optical spectrum analyzer and use it for both absorption and emission spectroscopy. The

More information

Observational Astronomy

Observational Astronomy Observational Astronomy Instruments The telescope- instruments combination forms a tightly coupled system: Telescope = collecting photons and forming an image Instruments = registering and analyzing the

More information

Bandpass Edge Dichroic Notch & More

Bandpass Edge Dichroic Notch & More Edmund Optics BROCHURE Filters COPYRIGHT 217 EDMUND OPTICS, INC. ALL RIGHTS RESERVED 1/17 Bandpass Edge Dichroic Notch & More Contact us for a Stock or Custom Quote Today! USA: +1-856-547-3488 EUROPE:

More information

LYNXEYE XE-T. < 380 ev. Innovation with Integrity. Energy. Resolution. High-Resolution Position Sensitive Detector with Superb Energy Resolution XRD

LYNXEYE XE-T. < 380 ev. Innovation with Integrity. Energy. Resolution. High-Resolution Position Sensitive Detector with Superb Energy Resolution XRD Energy < 380 ev Resolution High-Resolution Position Sensitive Detector with Superb Energy Resolution The is the next generation "Compound Silicon Strip" detector with superb energy resolution for ultrafast

More information

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES The current multiplication mechanism offered by dynodes makes photomultiplier tubes ideal for low-light-level measurement. As explained earlier, there

More information

attocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G

attocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G APPLICATION NOTE M01 attocfm I for Surface Quality Inspection Confocal microscopes work by scanning a tiny light spot on a sample and by measuring the scattered light in the illuminated volume. First,

More information

CONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application

CONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application CONFIGURING Your Spectroscopy System For PEAK PERFORMANCE A guide to selecting the best Spectrometers, s, and s for your application Spectral Measurement System Spectral Measurement System Spectrograph

More information

By using patented polycapillary optics this diffractometer obviates the need for monochromators and collimators for linear projection of X-Rays.

By using patented polycapillary optics this diffractometer obviates the need for monochromators and collimators for linear projection of X-Rays. XRD X-Ray Diffractometer Innovative, Integrated, Multifunctional By using patented polycapillary optics this diffractometer obviates the need for monochromators and collimators for linear projection of

More information

Confocal Imaging Through Scattering Media with a Volume Holographic Filter

Confocal Imaging Through Scattering Media with a Volume Holographic Filter Confocal Imaging Through Scattering Media with a Volume Holographic Filter Michal Balberg +, George Barbastathis*, Sergio Fantini % and David J. Brady University of Illinois at Urbana-Champaign, Urbana,

More information

Measurement and alignment of linear variable filters

Measurement and alignment of linear variable filters Measurement and alignment of linear variable filters Rob Sczupak, Markus Fredell, Tim Upton, Tom Rahmlow, Sheetal Chanda, Gregg Jarvis, Sarah Locknar, Florin Grosu, Terry Finnell and Robert Johnson Omega

More information

Development of ultra-fine structure t metrology system using coherent EUV source

Development of ultra-fine structure t metrology system using coherent EUV source 2009 International Workshop On EUV Lithography, July 13-17,2009 Development of ultra-fine structure t metrology system using coherent EUV source University of Hyogo 1, Hiroo Kinoshita 1,3, Tetuo Harada

More information

CERTIFICATE SD SAMPLE PREPARATION ANALYTICAL PROCEDURES. Signature: Colin Ramshaw, Vancouver Laboratory Manager

CERTIFICATE SD SAMPLE PREPARATION ANALYTICAL PROCEDURES. Signature: Colin Ramshaw, Vancouver Laboratory Manager Page: 1 CERTIFICATE SD16032427 This report is for 1 Rock sample submitted to our lab in Sudbury, ON, Canada on 4-MAR-2016. The following have access to data associated with this certificate: TOM VANDRUNEN

More information

Evaluating the Performance of a Commercial Silicon Drift Detector for X-ray Microanalysis

Evaluating the Performance of a Commercial Silicon Drift Detector for X-ray Microanalysis Evaluating the Performance of a Commercial Silicon Drift Detector for X-ray Microanalysis Edward A. Kenik Materials Science & Technology Division, Oak Ridge National Laboratory, Oak Ridge, TN 37831 kenikea@ornl.gov

More information

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage

More information

Add CLUE to your SEM. High-efficiency CL signal-collection. Designed for your SEM and application. Maintains original SEM functionality

Add CLUE to your SEM. High-efficiency CL signal-collection. Designed for your SEM and application. Maintains original SEM functionality Add CLUE to your SEM Designed for your SEM and application The CLUE family offers dedicated CL systems for imaging and spectroscopic analysis suitable for most SEMs. In addition, when combined with other

More information

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy Bi177 Lecture 5 Adding the Third Dimension Wide-field Imaging Point Spread Function Deconvolution Confocal Laser Scanning Microscopy Confocal Aperture Optical aberrations Alternative Scanning Microscopy

More information

Possibilities for Thick, Simple- Structure Silicon X-Ray Detectors Operated by Peltier Cooling

Possibilities for Thick, Simple- Structure Silicon X-Ray Detectors Operated by Peltier Cooling Possibilities for Thick, Simple- Structure Silicon X-Ray Detectors Operated by Peltier Cooling Hideharu Matsuura 1, Derek Hullinger 2, Ryota Okada 1, Seigo Kitanoya 1, Seiji Nishikawa 1, and Keith Decker

More information

PAD Correlator Computer

PAD Correlator Computer ALIGNMENT OF CONVENTIONAL ROATING ARM INSTRUMENT GENERAL PRINCIPLES The most important thing in aligning the instrument is ensuring that the beam GOES OVER THE CENTER OF THE TABLE. The particular direction

More information

Lecture 25. Wind Lidar (3) Direct Detection Doppler Lidar

Lecture 25. Wind Lidar (3) Direct Detection Doppler Lidar Lecture 25. Wind Lidar (3) Direct Detection Doppler Lidar Overview of Direct Detection Doppler Lidar (DDL) Fringe imaging DDL Scanning FPI DDL FPI edge-filter DDL Iodine absorption-line edge-filter DDL

More information

Introduction to Electron Microscopy

Introduction to Electron Microscopy Introduction to Electron Microscopy Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built the first electron microscope in 1931 (Nobel Prize to Ruska in 1986)

More information

Certificate of Analysis First issued: July 2000 Version: December 2007 MA-2c

Certificate of Analysis First issued: July 2000 Version: December 2007 MA-2c Certificate of Analysis First issued: July 2000 Version: December 2007 MA-2c Gold Ore Table l - Certified value for gold and provisional value for silver Element Ag (µg/g) Au (µg/g) Mean 0.51 3.02 Within-laboratory

More information

Performance of Microchannel Plates Fabricated Using Atomic Layer Deposition

Performance of Microchannel Plates Fabricated Using Atomic Layer Deposition Performance of Microchannel Plates Fabricated Using Atomic Layer Deposition Andrey Elagin on behalf of the LAPPD collaboration Introduction Performance (timing) Conclusions Large Area Picosecond Photo

More information

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: )

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: ) University of Minnesota College of Science and Engineering Characterization Facility PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: 2012.10.17) The following instructions

More information

3 Analytical report of glass beads from Hoa Diem site, Khanh Hoa, Viet Nam.

3 Analytical report of glass beads from Hoa Diem site, Khanh Hoa, Viet Nam. 3 Analytical report of glass beads from Hoa Diem site, Khanh Hoa, Viet Nam. Yoshiyuki Iizuka (Institute of Earth Sciences, Academia Sinica) Studied glass beads are listed and shown in Table 1 and Figure

More information

High Performance Thin Film Optical Coatings Technical Reference Document 09/13. Coatings Capabilities. Heat Control - Hot Mirror Filters

High Performance Thin Film Optical Coatings Technical Reference Document 09/13. Coatings Capabilities. Heat Control - Hot Mirror Filters Heat Control - Hot Mirror Filters A hot mirror is in essence a thin film coating applied to substrates in an effort to reflect infra-red radiation either as a means to harness the reflected wavelengths

More information

Scanning Electron Microscopy. EMSE-515 F. Ernst

Scanning Electron Microscopy. EMSE-515 F. Ernst Scanning Electron Microscopy EMSE-515 F. Ernst 1 2 Scanning Electron Microscopy Max Knoll Manfred von Ardenne Manfred von Ardenne Principle of Scanning Electron Microscopy 3 Principle of Scanning Electron

More information

Optical Coherence: Recreation of the Experiment of Thompson and Wolf

Optical Coherence: Recreation of the Experiment of Thompson and Wolf Optical Coherence: Recreation of the Experiment of Thompson and Wolf David Collins Senior project Department of Physics, California Polytechnic State University San Luis Obispo June 2010 Abstract The purpose

More information

SCIENTIFIC INSTRUMENT NEWS. Introduction. Design of the FlexSEM 1000

SCIENTIFIC INSTRUMENT NEWS. Introduction. Design of the FlexSEM 1000 SCIENTIFIC INSTRUMENT NEWS 2017 Vol. 9 SEPTEMBER Technical magazine of Electron Microscope and Analytical Instruments. Technical Explanation The FlexSEM 1000: A Scanning Electron Microscope Specializing

More information

This procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface.

This procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface. Tescan MIRA3 SEM: EDS using EDAX TEAM Nicholas G. Rudawski ngr@ufl.edu Cell: (805) 252-4916 Office: (352) 392-3077 Last updated: 12/04/17 This procedure assumes the user is already familiar with basic

More information

EUV Multilayer Fabrication

EUV Multilayer Fabrication EUV Multilayer Fabrication Rigaku Innovative Technologies Inc. Yuriy Platonov, Michael Kriese, Jim Rodriguez ABSTRACT: In this poster, we review our use of tools & methods such as deposition flux simulation

More information

ScanArray Overview. Principle of Operation. Instrument Components

ScanArray Overview. Principle of Operation. Instrument Components ScanArray Overview The GSI Lumonics ScanArrayÒ Microarray Analysis System is a scanning laser confocal fluorescence microscope that is used to determine the fluorescence intensity of a two-dimensional

More information

Spatial resolution. Spatial resolution

Spatial resolution. Spatial resolution 11/05/00 Refraction Compound refractive lenses (concave) Snigirev et al, NATURE 199 patents: Tomie 1995 x-rays: n = 1 - δ - i β < 1 www.accel.de Chromatic lenses Prod.: Lengeler @RWTH Aachen, D need of

More information

Spectroscopy Lab 2. Reading Your text books. Look under spectra, spectrometer, diffraction.

Spectroscopy Lab 2. Reading Your text books. Look under spectra, spectrometer, diffraction. 1 Spectroscopy Lab 2 Reading Your text books. Look under spectra, spectrometer, diffraction. Consult Sargent Welch Spectrum Charts on wall of lab. Note that only the most prominent wavelengths are displayed

More information

Water-Window Microscope Based on Nitrogen Plasma Capillary Discharge Source

Water-Window Microscope Based on Nitrogen Plasma Capillary Discharge Source 2015 International Workshop on EUV and Soft X-Ray Sources Water-Window Microscope Based on Nitrogen Plasma Capillary Discharge Source T. Parkman 1, M. F. Nawaz 2, M. Nevrkla 2, M. Vrbova 1, A. Jancarek

More information