LONG TERM STATISTICS OF X-RAY SPECTROMETERS
|
|
- Nathan Bishop
- 5 years ago
- Views:
Transcription
1 403 LONG TERM STATISTICS OF X-RAY SPECTROMETERS J. F. Dlouhy*, D. Mathieu Department of the Environment, Environmental Technology Center, River Road, Ottawa, Ontario, Canada Kl A OH3 K. N. Stoev Bulgarian Academy of Sciences, Institute for Nuclear Research and Nuclear Energy, Blvd. Tzarigradsko Shousse, Sofia, Bulgaria ABSTRACT Quality assurance and quality control programs (QA/QC) are the essential components of any analytical laboratory. As a part of the QA/QC program in this laboratory (Environmental Technology Center), an extensive set of data was collected over a long period of time (about ten years), which was used to study the long term stability of two x-rays spectrometers. A Rh anode x-ray tube was used for the excitation of the x-ray fluorescence of two thin film x-ray standards. A Si(Li) detector, cooled to liquid nitrogen temperature, was used for recording the x-ray fluorescence spectra. All measurements were done in vacuum. Three measurement conditions were used in order to cover all elements of interest. The energy resolution (full width at half maximum), the peak position, the whole spectrum intensity, the intensity of several x-ray lines, and the concentration of several elements, were monitored and evaluated. The collected data were used to study the stability of the spectrometers, to account for the changes of the spectrometer response (either by using monitor ratios or by performing new calibration), and to determine the source of the problems with the fluctuations in the spectrometer response. The sample of collected data is presented in two tables and illustrated in graphic form. The results are discussed and some conclusions are made about the correctness of the different ways for the elimination of fluctuations of the x-ray spectrometer response. *) The corresponding author. Fax: (613) , Dlouhy.Joe@etc.ec.gc.ca
2 This document was presented at the Denver X-ray Conference (DXC) on Applications of X-ray Analysis. Sponsored by the International Centre for Diffraction Data (ICDD). This document is provided by ICDD in cooperation with the authors and presenters of the DXC for the express purpose of educating the scientific community. All copyrights for the document are retained by ICDD. Usage is restricted for the purposes of education and scientific research. DXC Website ICDD Website -
3 404 Two multichannel pulse height analyzers were or are used as X-ray spectrometers [2] in this laboratory (Kevex, model 770/8000 and model 771/8000). Measurement of the accuracy, precision and stability of the spectrometers is a complex problem. The first task is the check of the differential and integral linearity, which is measured using the precision pulser. That usually only checks the amplifier and the analog-to-digital converter. The measurements are nowadays performed in a different fashion. The primary energy calibration is done using a radioactive source (frequently Fe 55). It has the disadvantage that the X-ray tube and the power supply are not included in the test. The secondary calibration is done by using X-ray lines at the high and low position in the spectrum. The manufacturer recommended the use of an aluminium alloy (NBS 2708) containing cooper. The lines measured are CuKa and AlKa. The measurements were done on the energy scale 10 kev. The desired energy accuracy was achieved using an automatic iterative program. The best energy accuracy was 1 ev, which represents one tenth of a channel under specified conditions. The program resets the amplifier gain (slope of the calibration curve) and the zero offset of the controlling helical potentiometers. The parameters were transferred automatically to other energy scales (20 and 40 kev). The energy and energy resolution calibration procedures were extended in this laboratory by evaluating the spectrum of the NIST thin film X-ray standards NBS1832 and NBS1833 (see Table 1) [3], which are measured with each set of specimens, Table 1 Al Si K Ca Ti V Mn Fe co cu Zn Pb NBS1832 b&w NBS1833 [u&m
4 405 Each specimen spectrum was recalculated to achieve full energy scale agreement for all three measurement conditions. The detection efficiency is the second task. It can be controlled by measuring the integral count rate of the proper standard over a given energy interval. That interval could be the whole measured spectrum or an interval which contains the characteristic excitation interval (without analytical lines) or it can be the interval which is used for the analysis of measured specimens for the given condition. The integral count rate is then used to correct the spectra measured under identical conditions (the same day). There is an easy check of the correctness of the procedure by using thin film X-ray standards (NBS 1832 and NBS1 833, which were already measured to be used for the energy calibration). The corrected spectra are processed and the certified concentrations are calculated and the accuracy is established. The initial setting (Window1 in Table 2) of the first multichannel pulse height analyzer (Kevex, model 770/8000) was selected by late Dave Seilstad (Kevex, Head of the Application Laboratory). Table 2 Window1 Window2 X-ray low high low high tube [kev] [kev] [kevl [kevl WI Condition Condition Condition The X-ray tube current was set as high as possible (<=3.3mA, which is the highest available setting) so as not to exceed 50 % dead time for all standards and specimens. The spectrometer was used for nine years. During that period eight X-ray tubes were used. Full calibration had to be done after each change of the X-ray tube. The tubes did have different parameters including the thickness of the beryllium window.
5 406 The calibration constants (ng/cmycps) for each element from Na to Bi were measured using thin film X-ray standards (MicroMatter) with element mass from about 5 ug/cm2 to about 50 ug/cm2. The support film is 6.5 urn Mylar. Over one hundred standards were used. That was complemented by the thin film X-ray standards NBS1832 and NBS 1833, NIST2036 and corresponding blanks (Teflon, Mylar and Nuclepore films).the measurement, calculations and evaluation takes about one month. The procedures set at the beginning were used until Different mathematical analysis [l] partially in parallel to the original one was used since That is the reason that the Figure #l is limited to the period. The monitor ratios for the thin film standard NBS1832 are plotted versus time. There is a new x-ray tube at the beginning of each calibration as illustrated in Figure #l. The intervals with missing points indicate the waiting periods for the replacement x-ray tube or the periods when measurements under different conditions were pursued (eg secondary targets excitation, radiation background etc.). In Calibration 5 and 7 are visible additional sudden changes in the monitor ratios, They are caused by the visit of the service personnel and inadvertent change in the close geometry of the analyzer (X-ray tube, filter bar, sample changer, detector collimator and detector). From Calibration 4 a different pair of the thin film X-ray standards NBS 1832 and NBS 1833 was used. The windows for monitor ratios were changed at the same time (Window2 in Table 1) to cover the interval of the spectrum used for the analysis of the measured specimens. FIGURE #l MONITOR RATIOS NBS1832 CONDITION 2 SPECTROMETER 77OBOOO 1166 DATA POINTS p d E z e I 0.5 l/l/87 l/l188. I89 DATE Figure #l contains monitor ratios for Condition 2 only. It illustrates the changes in each calibration and between different calibrations. The difference in monitor ratios between calibrations is caused by the slight difference in the individual x-ray tubes (the i
6 407 geometry of the tube, the thickness of the Be window, the geometry within the spectrometer, the contamination of the Rh anode etc.) and by the need to adjust the x-ray tube current for each tube to fulfill the condition that the dead time for all standards and blanks (NBS, NIST and MicroMatter) will be for all three conditions under 50%. The monitor ratio is expressed as the ratio of the measured value versus value from the mean spectrum of the same standard measured at the time of the same calibration. The mean monitor ratio for Calibration 6 and Condition 1 is 1.O 15 (relative standard deviation is 2.25%) for NBS1832 standard and (relative standard deviation is 2.19%) for the NBS 1833 standard from 387 measurements. The individual ratios between NBS 1832 and NBS1833 monitor ratios give the mean of with 2.36% relative standard deviation which is statistically identical to the direct values of monitor ratios. Some outliers in the Figure were not removed to illustrate possible differences. Most often the outlier has a higher value than other values in the set on account of the failure of the spectrometer to advance to the correct position for the NBS standard. That leads to the inclusion of the massive Al holder line in the spectrum and corresponding increase in the monitor ratio. Same data for the second spectrometer (Kevex, model 771/8000) are presented in Table 3. Table 3 Standard Deviation Monitor Ratio (NBS1832/NBS1833) 0.33 % Condition 3 FWHMRatio (NBS1832/NBS1833) 1.55 % Condition 2 Si Ratio (NBS1832/NBS1833) 0.46 % Condition 3 Cu NBS1832 Ratio 2.07 % Condition l/condition 2 Pb NBS1833 Ratio 0.76 % Condition l/condition % Condition 2 NBS1833 (Fe/Pb) 0.40 % Condition 2 The relative standard deviation of the monitor ratio for NBS1832 standard and condition 3 is 2.04 %. The ratio of the monitor ratios of both standards is 0.33 %. That represents almost an order of magnitude improvement. The Table contains the ratios of different parameters after normalization and proves the stability of the spectrometers together with the stability of the thin film X-ray standards NBS1832 and NBS1833 across different conditions and different parameters. It is not possible to separate the contribution of any individual standard or spectrometer on the basis of available data. The values of the relative standard deviation indicate acceptable quality of data.
7 408 CONCLUSIONS The applicability of two thin film X-ray standards for the work in this laboratory was demonstrated. It reduces the need for frequent full recalibration of the spectrometer. It takes care of short and long term stability of the spectrometer response by using only the linear corrections. It spans the change of the integration intervals for the monitor ratios. The change of the X-ray tube leads usually to the non-linear changes in the spectrometer response and is not presented here. It proves the long term stability of the thin film X-ray standards NBS1832 and NBS1833. LITERATURE [l] K.N.Stoev, J.F.Dlouhy, X-RAY SPECTROMETRY: 23, 112-l 19,1994 [2] K. Siegbahn, Alpha-, Beta- and Gamma-Ray Spectroscopy, Vol. 1, 1969, North Holland Publishing Co. Amsterdam. [3] NIST is National Institute of Standards and Technology, Gaitersburg, MD, USA. Both very important standards are out of stock. More general aspects of x-ray spectrometry are included in E.P.Bertin, Principles and Practice of X-Ray Spectrometric Analysis, 2nd Edition, 1979, Plenum Press, New York. where there is a part on page 323 and ff. on multichannel pulse height analyzers or in R.E.VanGrieken, A. A.Markowicz, Handbook of X-Ray Spectrometry, 1993, Marcel Dekker, New York. Some aspects of QA/QC are covered in F.M. Garfield, Quality Assurance Principles for Analytical Laboratories, 1988, AOAC, Arlington.
WIDE ANGLE GEOMETRY EDXRF SPECTROMETERS WITH SECONDARY TARGET AND DIRECT EXCITATION MODES
Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 11 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42
More informationANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER
Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 325 ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER ABSTRACT William Chang, Jonathan Kerner, and Edward
More informationA MONTE CARLO CODE FOR SIMULATION OF PULSE PILE-UP SPECTRAL DISTORTION IN PULSE-HEIGHT MEASUREMENT
Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 246 A MONTE CARLO CODE FOR SIMULATION OF PULSE PILE-UP SPECTRAL DISTORTION IN PULSE-HEIGHT MEASUREMENT
More informationZaidi Embong and Husin Wagiran Physics Department, University Of Technology Malaysia, P.O Box 791, 80990, Johor Baharu
MY9800971 Optimization of a Spectrometry for Energy -Dispersive X-ray Fluorescence Analysis by X-ray Tube in Combination with Secondary Target for Multielements Determination of Sediment Samples. Zaidi
More informationAdvancing EDS Analysis in the SEM Quantitative XRF. International Microscopy Congress, September 5 th, Outline
Advancing EDS Analysis in the SEM with in-situ Quantitative XRF Brian J. Cross (1) & Kenny C. Witherspoon (2) 1) CrossRoads Scientific, El Granada, CA 94018, USA 2) ixrf Systems, Inc., Houston, TX 77059,
More informationCopyright -International Centre for Diffraction Data 2010 ISSN
234 BRIDGING THE PRICE/PERFORMANCE GAP BETWEEN SILICON DRIFT AND SILICON PIN DIODE DETECTORS Derek Hullinger, Keith Decker, Jerry Smith, Chris Carter Moxtek, Inc. ABSTRACT Use of silicon drift detectors
More informationGamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors
Nuclear Physics #1 Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors Introduction: In this experiment you will use both scintillation and semiconductor detectors to study γ- ray energy spectra. The
More informationDEVELOPMENT OF A WAVELENGTH DISPERSIVE X-RAY FLUORESCENCE SPECTROMETER USING A MULTI-CAPILLARY X-RAY LENS FOR X-RAY DETECTION
Copyright JCPDS - International Centre for Diffraction Data 2003, Advances in X-ray Analysis, Volume 46. 346 DEVELOPMENT OF A WAVELENGTH DISPERSIVE X-RAY FLUORESCENCE SPECTROMETER USING A MULTI-CAPILLARY
More informationMOXTEK S NEW ULTRA-LITE X-RAY SOURCES: PERFORMACE CHARACTERIZATIONS
Copyright JCPDS-International Centre for Diffraction Data 2013 ISSN 1097-0002 202 MOXTEK S NEW ULTRA-LITE X-RAY SOURCES: PERFORMACE CHARACTERIZATIONS S. Cornaby, S. Morris, J. Smith, D. Reynolds, K. Kozaczek
More informationUSING A CHARGE-COUPLED DEVICE (CCD) TO GATHER X-RAY FLUORESCENCE (XRF)AND X-RAY DIFFRACTION (XRD) INFORMATION SIMULTANEOUSLY
Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 343 USING A CHARGE-COUPLED DEVICE (CCD) TO GATHER X-RAY FLUORESCENCE (XRF)AND X-RAY DIFFRACTION (XRD)
More informationDOUBLE MULTILAYER MONOCHROMATOR WITH FIXED EXIT GEOMETRY. H.Gatterbauer, P.Wobrauschek, F.Hegediis, P.Biini, C.Streli
Copyright (C) JCPDS International Centre for Diffraction Data 1999 379 DOUBLE MULTILAYER MONOCHROMATOR WITH FIXED EXIT GEOMETRY H.Gatterbauer, P.Wobrauschek, F.Hegediis, P.Biini, C.Streli Atominsitut der
More informationGUNSHOT RESIDUE INVESTIGATIONS USING TXRF
299 GUNSHOT RESIDUE INVESTIGATIONS USING TXRF Alexander Wastl 1, Bettina Bogner 2, Peter Kregsamer 1, Peter Wobrauschek 1, Christina Streli 1 1 Atominstitut, Vienna University of Technology, Vienna, Austria
More informationMICRO XRF OF LIGHT ELEMENTS USING A POLYCAPILLARY LENS AND AN ULTRA THIN WINDOW SILICON DRIFT DETECTOR INSIDE A VACUUM CHAMBER
Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 229 MICRO XRF OF LIGHT ELEMENTS USING A POLYCAPILLARY LENS AND AN ULTRA THIN WINDOW SILICON DRIFT
More informationORTEC. Research Applications. Pulse-Height, Charge, or Energy Spectroscopy. Detectors. Processing Electronics
ORTEC Spectroscopy systems for ORTEC instrumentation produce pulse height distributions of gamma ray or alpha energies. MAESTRO-32 (model A65-B32) is the software included with most spectroscopy systems
More informationSpectral distribution from end window X-ray tubes
Copyright ISSN (C) 1097-0002, JCPDS-International Advances in X-ray Centre Analysis, for Volume Diffraction 41 Data 1999 393 Spectral distribution from end window X-ray tubes N. Broll 1, P. de Chateaubourg
More informationX-RAY OPTICS FOR TWO-DIMENSIONAL DIFFRACTION
Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 332 ABSTRACT X-RAY OPTICS FOR TWO-DIMENSIONAL DIFFRACTION Bob B. He and Uwe Preckwinkel Bruker
More informationR AMP TEK Landed on Mars July 4, 1997 All Solid State Design No Liquid Nitrogen Be Window FET Detector Temperature Monitor Cooler Mounting Stud FEATURES Si-PIN Photodiode Thermoelectric Cooler Beryllium
More informationApplications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror
Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror Stephen B. Robie scintag, Inc. 10040 Bubb Road Cupertino, CA 95014 Abstract Corundum
More informationON THE ENERGY DEPENDENCE OF THE DETECTOR EFFICIENCY OF A Si-PIN DIODE
Copyright(C)JCPDS-International Centre for Diffraction Data 2, Advances in X-ray Analysis, Vol.42 36 Copyright(C)JCPDS-International Centre for Diffraction Data 2, Advances in X-ray Analysis, Vol.42 36
More informationORTEC Experiment 13. Gamma-Gamma Coincidence with Angular Correlation. Equipment Required
ORTEC Experiment 13 Equipment Required Two 905-3 2-in. x 2-in. NaI(Tl) Scintillation Detector Assemblies. Two 266 Photomultiplier Tube Bases. Two 113 Scintillation Preamplifiers. Two 556 High Voltage Power
More informationMINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE
228 MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE D. CARUSO, M. DINSMORE TWX LLC, CONCORD, MA 01742 S. CORNABY MOXTEK, OREM, UT 84057 ABSTRACT Miniature x-ray sources present
More informationON THE DETECTION LIMIT OF TEY (TOTAL ELECTRON YIELD) Maria F. Ebel, Horst Ebel and Robert Svagera
Copyright(C)JCPDS-International Centre for Diffraction Data 2, Advances in X-ray Analysis, Vol.42 91 Copyright(C)JCPDS-International Centre for Diffraction Data 2, Advances in X-ray Analysis, Vol.42 91
More informationApplications of Micro XRF for the Analysis of Traditional Japanese "Ainu" Glass Beads and other Artifacts
161 161 Applications of Micro XRF for the Analysis of Traditional Japanese "Ainu" Glass Beads and other Artifacts K.Sugihara 1, M.Satoh 1, Y.Hayakawa 2, A.Saito 3 and T.Sasaki 4 1 Seiko Instruments Inc.,
More informationXRF Instrumentation. Introduction to spectrometer
XRF Instrumentation Introduction to spectrometer AMPTEK, INC., Bedford, MA 01730 Ph: +1 781 275 2242 Fax: +1 781 275 3470 sales@amptek.com 1 Instrument Excitation source Sample X-ray tube or radioisotope
More informationAbstract. 1. Introduction
941 Monte Carlo Simulation of Pulse Pile Up R. P. Gardner and S. H. Lee Center for Engineering Applications of Radioisotopes Department of Engineering North Carolina State University Raleigh, NC 276957909,
More informationGamma Spectrometer Initial Project Proposal
Gamma Spectrometer Initial Project Proposal Group 9 Aman Kataria Johnny Klarenbeek Dean Sullivan David Valentine Introduction There are currently two main types of gamma radiation detectors used for gamma
More informationEnergy Measurements with a Si Surface Barrier Detector and a 5.5-MeV 241 Am α Source
Energy Measurements with a Si Surface Barrier Detector and a 5.5-MeV 241 Am α Source October 18, 2017 The goals of this experiment are to become familiar with semiconductor detectors, which are widely
More informationPhysics Laboratory Scattering of Photons from Electrons: Compton Scattering
RR Oct 2001 SS Dec 2001 MJ Oct 2009 Physics 34000 Laboratory Scattering of Photons from Electrons: Compton Scattering Objective: To measure the energy of high energy photons scattered from electrons in
More informationSIMULTANEOUS XRD/XRF WITH LOW-POWER X-RAY TUBES
Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 34 SIMULTANEOUS XRD/XRF WITH LOW-POWER X-RAY TUBES S. Cornaby 1, A. Reyes-Mena 1, P. W. Moody 1,
More informationCertificate of Analysis First issued: July 2000 Version: December 2007 MA-2c
Certificate of Analysis First issued: July 2000 Version: December 2007 MA-2c Gold Ore Table l - Certified value for gold and provisional value for silver Element Ag (µg/g) Au (µg/g) Mean 0.51 3.02 Within-laboratory
More informationK 223 Angular Correlation
K 223 Angular Correlation K 223.1 Aim of the Experiment The aim of the experiment is to measure the angular correlation of a γ γ cascade. K 223.2 Required Knowledge Definition of the angular correlation
More informationFAST ELEMENTAL MAPPING WITH MICRO-XRF
286 FAST ELEMENTAL MAPPING WITH MICRO-XRF Haschke, M.; Rossek, U.; Tagle, R.; Waldschläger, U. Bruker Nano GmbH, 12489 Berlin, Schwarzschildstr.12 ABSTRACT X-Ray optics are now in common use for concentrating
More informationS1 TITAN Alloy LE Calibrations (P/N: )
S1 TITAN 600-800 Alloy LE Calibrations () Low Alloy Si P S Ti V Cr Mn Fe Co Ni Cu Nb Mo W Pb Analysis range, % LLD-2 LLD-0.15 LLD-0.3 LLD - 0.1 0.05-1.8 LLD - 9 0.1-2.0 75-100 LLD - 8 LLD - 5 LLD - 5 LLD-
More informationPurpose This experiment will use the coincidence method for time correlation to measure the lifetime in the decay scheme of 57
Equipment Required Two 113 Scintillation Preamplifiers Two 266 Photomultiplier Tube Bases 4001A/4002D Bin and Power Supply 414A Fast Coincidence Two 551 Timing Single-Channel Analyzers 567 Time-to-Amplitude
More informationLYNXEYE XE. Innovation with Integrity. High-Resolution Energy-Dispersive Detector for 0D, 1D, and 2D Diffraction XRD
High-Resolution Energy-Dispersive Detector for 0D, 1D, and 2D Diffraction The is the first energy dispersive 0D, 1D, and 2D detector operating at room temperature for ultra fast X-ray diffraction measurements.
More informationApplication Note. Monitoring the Release of Radioactive Noble Gases Through the Stack of a Nuclear Power Plant (NPP): Stack Monitor System
Application Note Monitoring the Release of Radioactive Noble Gases Through the Stack of a Nuclear Power Plant (NPP): Stack Monitor System Based on the German KTA 1503.1 and respective international regulations,
More informationORTEC. AN34 Experiment 14 Nuclear Lifetimes and the Coincidence Method. Equipment Needed from ORTEC. Equipment Required from Other Manufacturers
Equipment Needed from ORTEC Two 113 Scintillation Preamplifiers Two 266 Photomultiplier Tube Bases 4001A/4002D Bin and Power Supply 414A Fast Coincidence Two 551 Timing Single-Channel Analyzers 567 Time-to-Amplitude
More informationSPECTROMETRIC DETECTION PROBE Model 310. Operator's manual
SPECTROMETRIC DETECTION PROBE Model 310 Operator's manual CONTENTS 1. INTRODUCTION... 3 2. SPECIFICATIONS... 4 3. DESIGN FEATURES... 6 4. INSTALLATION... 10 5. SAFETY AND PRECAUTIONS... 13 6. THEORY OF
More informationLYNXEYE XE-T. < 380 ev. Innovation with Integrity. Energy. Resolution. High-Resolution Position Sensitive Detector with Superb Energy Resolution XRD
Energy < 380 ev Resolution High-Resolution Position Sensitive Detector with Superb Energy Resolution The is the next generation "Compound Silicon Strip" detector with superb energy resolution for ultrafast
More informationIN-SITU ENVIRONMENTAL XRF
Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 137 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42
More informationM4 TORNADO PLUS. Innovation with Integrity. Super Light Element Micro-XRF Spectrometer. Micro-XRF
M4 TORNADO PLUS Super Light Element Micro-XRF Spectrometer Innovation with Integrity Micro-XRF M4 TORNADO PLUS - A New Era in Micro-XRF M4 TORNADO PLUS is the world's first Micro-XRF spectrometer that
More informationPHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002
PHYSICS 334 - ADVANCED LABORATORY I COMPTON SCATTERING Spring 00 Purposes: Demonstrate the phenomena associated with Compton scattering and the Klein-Nishina formula. Determine the mass of the electron.
More informationPartial Replication of Storms/Scanlan Glow Discharge Radiation
Partial Replication of Storms/Scanlan Glow Discharge Radiation Rick Cantwell and Matt McConnell Coolescence, LLC March 2008 Introduction The Storms/Scanlan paper 1 presented at the 8 th international workshop
More informationAmptek Silicon Drift Diode (SDD) at High Count Rates
Amptek Silicon Drift Diode (SDD) at High Count Rates A silicon drift diode (SDD) is functionally similar to a SiPIN photodiode but its unique electrode structure reduces the electronic noise at short peaking
More informationNew Detectors for X-Ray Metal Thickness Measuring
ECNDT 2006 - Poster 132 New Detectors for X-Ray Metal Thickness Measuring Boris V. ARTEMIEV, Alexander I. MASLOV, Association SPEKTR- GROUP, Moscow, Russia Abstract. X-ray thickness measuring instruments
More informationLECTURE 10. Dr. Teresa D. Golden University of North Texas Department of Chemistry
LECTURE 10 Dr. Teresa D. Golden University of North Texas Department of Chemistry Components for the source include: -Line voltage supply -high-voltage generator -x-ray tube X-ray source requires -high
More informationEvaluating the Performance of a Commercial Silicon Drift Detector for X-ray Microanalysis
Evaluating the Performance of a Commercial Silicon Drift Detector for X-ray Microanalysis Edward A. Kenik Materials Science & Technology Division, Oak Ridge National Laboratory, Oak Ridge, TN 37831 kenikea@ornl.gov
More informationCHAPTER-2 INSTRUMENTATION AND METHODS OF DATA ANALYSIS
CHAPTER-2 INSTRUMENTATION AND METHODS OF DATA ANALYSIS 2.1 INTRODUCTION 21 2.2 PHOTON SOURCES FOR XRF 21 2.2.1 Radioactive Source 21 2.2.2 X-ray Tubes 22 2.2.3 Synchrotron Radiation 23 2.3 X-RAY DETECTION
More informationUse of Back Scattered Ionizing Radiation for Measurement of Thickness of the Catalytic Agent Active Material
18th World Conference on Nondestructive Testing, 16- April 1, Durban, South Africa Use of Back Scattered Ionizing Radiation for Measurement of Thickness of the Catalytic Agent Active Material Boris V.
More informationMössbauer ~ Spectrometer. Following to our long-term experiences, we offer complete Mössbauer spectroscopy instrumental support
www.mossbauer-spectrometers.com Mössbauer ~ Spectrometer Following to our long-term experiences, we offer complete Mössbauer spectroscopy instrumental support Mössbauer ~ Spectrometer > Mössbauer spectroscopy
More informationRADIONUCLIDES IN AGRICULTURAL SOIL IN VOJVODINA REGION
RADIONUCLIDES IN AGRICULTURAL SOIL IN VOJVODINA REGION N. Zikic-Todorovic 1), I. Bikit 1), J. Slivka 1), M. Veskovic 1), Lj. Conkic 1), E. Varga 1), S. Curcic 1), D. Mrdja 1) 1) Department of Physics,
More informationTOWARDS FAST RECIPROCAL SPACE MAPPING
Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 165 ABSTRACT TOWARDS FAST RECIPROCAL SPACE MAPPING J.F. Woitok and A. Kharchenko PANalytical B.V.,
More informationCitation X-Ray Spectrometry (2011), 40(4): 2. Right final form at
TitleSi PIN X-ray photon counter Author(s) Nakaye, Yasukazu; Kawai, Jun Citation X-Ray Spectrometry (2011), 40(4): 2 Issue Date 2011-03-24 URL http://hdl.handle.net/2433/197743 This is the peer reviewed
More informationComponents of Optical Instruments
Components of Optical Instruments General Design of Optical Instruments Sources of Radiation Wavelength Selectors (Filters, Monochromators, Interferometers) Sample Containers Radiation Transducers (Detectors)
More informationDETECTORS Important characteristics: 1) Wavelength response 2) Quantum response how light is detected 3) Sensitivity 4) Frequency of response
DETECTORS Important characteristics: 1) Wavelength response 2) Quantum response how light is detected 3) Sensitivity 4) Frequency of response (response time) 5) Stability 6) Cost 7) convenience Photoelectric
More informationMWPC Gas Gain with Argon-CO 2 80:20 Gas Mixture
IMA Journal of Mathematical Control and Information Page 1 of 10 doi:10.1093/imamci/dri000 1. Principles of Operation MWPC Gas Gain with Argon-CO 2 80:20 Gas Mixture Michael Roberts A multi-wire proportional
More informationX-Ray Spectroscopy with a CCD Detector. Application Note
X-Ray Spectroscopy with a CCD Detector In addition to providing X-ray imaging solutions, including CCD-based cameras that image X-rays using either direct detection (0.5-20 kev) or indirectly using a scintillation
More informationRIETVELD REFINEMENT OF POWDER DATA FROM MULTILAYER OPTICS
Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 166 RIETVELD REFINEMENT OF POWDER DATA FROM MULTILAYER OPTICS ABSTRACT Scott T. Misture NYS College
More informationSilicon Drift Detector. with On- Chip Ele ctronics for X-Ray Spectroscopy. KETEK GmbH Am Isarbach 30 D O berschleißheim GERMANY
KETEK GmbH Am Isarbach 30 D-85764 O berschleißheim GERMANY Silicon Drift Detector Phone +49 (0)89 315 57 94 Fax +49 (0)89 315 58 16 with On- Chip Ele ctronics for X-Ray Spectroscopy high energy resolution
More informationIDENTIFICATION OF PAINTING MATERIALS USED FOR MURAL PAINTINGS BY IMAGE ANALYSIS AND XRF
IDENTIFICATION OF PAINTING MATERIALS USED FOR MURAL PAINTINGS BY IMAGE ANALYSIS AND XRF 213 Seiji SHIRONO, Yasuhiro HAYAKAWA National Research Institute for Cultural Properties, Tokyo, Japan ABSTRACT The
More informationChemistry 985. Some constants: q e 1.602x10 19 Coul, ɛ x10 12 F/m h 6.626x10 34 J-s, c m/s, 1 atm = 760 Torr = 101,325 Pa
Chemistry 985 Fall, 2o17 Distributed: Mon., 17 Oct. 17, 8:30AM Exam # 1 OPEN BOOK Due: 17 Oct. 17, 10:00AM Some constants: q e 1.602x10 19 Coul, ɛ 0 8.854x10 12 F/m h 6.626x10 34 J-s, c 299 792 458 m/s,
More informationOnline Gauging As We Know It! by Udo Skarke Erhardt-Leimer Inc.
Online Gauging As We Know It! by Udo Skarke Erhardt-Leimer Inc. For many decades online measurement has made continuous processes better, faster, more accurate and more reliable. Since their introduction,
More informationANALYSIS OF LEAD IN CANDLE PARTICULATE EMISSIONS BY XRF USING UNIQUANT 4
Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 539 ANALYSIS OF LEAD IN CANDLE PARTICULATE EMISSIONS BY XRF USING UNIQUANT 4 Shirley J. Wasson
More informationThe SS6000 Gold Mate Series For analyzing all precious metals and other elements from Mg to U
The SS6000 Gold Mate Series For analyzing all precious metals and other elements from Mg to U Portable desk top EDXRF analyzers Responsive, bright, color touch screen display Uses Silicon Drift or Silicon
More informationNEEP 427 PROPORTIONAL COUNTERS. Knoll, Chapters 6 & 14 Sect. I & II
NEEP 427 PROPORTIONAL COUNTERS References: Knoll, Chapters 6 & 14 Sect. I & II a proportional counter the height of the output pulse is proportional to the number of ion pairs produced in the counter gas.
More informationQualitative analysis tutorial for Tracer III SD and V+ data
Qualitative analysis tutorial for Tracer III SD and V+ data Outline What does the spectrum mean? What to watch out for How to normalize Using Artax What does the spectrum mean? Bremsstrahlung radiation
More informationEnergy Dispersive Spectrometry Hardware
from: http://www.x-raymicroanalysis.com/pages/tutorial1/system1.htm Energy Dispersive Spectrometry Hardware Introduction The Microanalysis System Ease of use has become a major focus in the selection of
More informationBRUKER ADVANCED X-RAY SOLUTIONS. SPECTROMETRY SOLUTIONS ARTAX mxrf SPECTROMETER
BRUKER ADVANCED X-RAY SOLUTIONS SPECTROMETRY SOLUTIONS ARTAX mxrf SPECTROMETER Microanalysis ARTAX Elemental Analysis for the Art Community and More Non-destructive elemental analysis is strictly required
More informationChemical Engineering 412
Chemical Engineering 412 Introductory Nuclear Engineering Lecture 25 Radiation Detection & Measurement Spiritual Thought 2 I realize that there are some, perhaps many, [who] feel overwhelmed by the lack
More informationBruker Nano. M4 tornado. High performance micro-xrf spectrometer. think forward
Bruker Nano M4 tornado High performance micro-xrf spectrometer think forward µ-xrf M4 TORNADO setting standards in µ-xrf µ-xrf is the method of choice for highly sensitive and non-destructive elemental
More informationPSPC/MDG 2000 X-RAY MICRODIFFRACTOMETER. Product Information
THE RIGAKU JOURNAL VOL. 11 I NO.2 I 1994 Product Information X-RAY MICRODIFFRACTOMETER PSPC/MDG 2000 1. Introduction The analysis of X-ray diffraction patterns is well known as an effective means of obtaining
More information-_.-~ Sample. HIGH SENSITIVITY TYPE TOTAL REFLECTION X-RAY SPECTROMETER SYSTEM Wafer Surface Analysis System --
THE RIGAKU JOURNAL VOl. 8 / NO. 1 / 1991 HIGH SENSITIVITY TYPE TOTAL REFLECTION X-RAY SPECTROMETER SYSTEM 3726 --Wafer Surface Analysis System -- 1. Introduction System 3726 utilizes the total reflection
More informationTOWARDS SUB-100 NM X-RAY MICROSCOPY FOR TOMOGRAPHIC APPLICATIONS
Copyright -International Centre for Diffraction Data 2010 ISSN 1097-0002 89 TOWARDS SUB-100 NM X-RAY MICROSCOPY FOR TOMOGRAPHIC APPLICATIONS P. Bruyndonckx, A. Sasov, B. Pauwels Skyscan, Kartuizersweg
More informationDESIGN OF EDXRF EQUIPMENT FOR THE NONDESTRUCTIVE STUDY OF PRINTS
Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 70 DESIGN OF EDXRF EQUIPMENT FOR THE NONDESTRUCTIVE STUDY OF PRINTS 1,3 M. Ardid, 1 J.L. Ferrero,
More informationISO-CART-85. Mobile Low-Level Waste Assay System
Mobile Low-Level Waste Assay System Complete In-Situ NDA Gamma-Ray Analysis Solutions for a Wide Variety of Samples, including Free-Release Decommissioning Waste. ISO-CART -85 : A Complete Turnkey Solution
More informationVirtual Laboratory of Nuclear Fission Virtual practicum in the framework of the project Virtual Laboratory of Nuclear Fission
Virtual Laboratory of Nuclear Fission Virtual practicum in the framework of the project Virtual Laboratory of Nuclear Fission Khanyisa Sowazi, University of the Western Cape JINR SAR, September 2015 INDEX
More informationCalibration of ARM Spectral Shortwave Radiometers
Calibration of ARM Spectral Shortwave Radiometers J. J. Michalsky, J. L. Berndt, P. W. Kiedron, and L. C. Harrison Atmospheric Sciences Research Center State University of New York at Albany Albany, New
More informationKeyser, Ronald M., Twomey, Timothy R., and Bingham, Russell D. ORTEC, 801 South Illinois Avenue, Oak Ridge, TN 37831s
Improved Performance in Germanium Detector Gamma Spectrometers based on Digital Signal Processing Keyser, Ronald M., Twomey, Timothy R., and Bingham, Russell D. ORTEC, 801 South Illinois Avenue, Oak Ridge,
More informationFully Automated Auto Scanning System VPD-ICPMS Expert
Fully Automated Auto Scanning System VPD-ICPMS Expert www.vpdicpms.com An indispensable tool for analysis of metallic impurities in Si wafer Developed in cooperation with Samsung Electronics Co., Ltd.
More informationMinnesota Rules, Chapter 4732 X-ray Revision
Minnesota Rules, Chapter 4732 X-ray Revision DRAFT INDUSTRIAL X-RAY SYSTEMS DEFINTIONS, 1.0 4732.####. INDUSTRIAL X-RAY SYSTEMS DEFINITIONS. Subpart 1. Scope. For purposes of industrial x-ray systems under
More informationCALIBRATION OF X-RAY IMAGING DEVICES FOR ACCURATE INTENSITY MEASUREMENT
DOE/NV/25946--1306 23 CALIBRATION OF X-RAY IMAGING DEVICES FOR ACCURATE INTENSITY MEASUREMENT Michael J. Haugh*, Michael Charest*, Patrick Ross*, Joshua Lee*, Marilyn Schneider, Nathan Palmer, and Alan
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 16371-1 First edition 2011-10-01 Non-destructive testing Industrial computed radiography with storage phosphor imaging plates Part 1: Classification of systems Essais non destructifs
More informationOPTIMIZING THE ELEMENTAL SENSITIVITY AND FOCAL SPOT SIZE OF A MONOLITHIC POLYCAPILLARY OPTIC USING MICRO-X-RAY FLUORESCENCE
, Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 26 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42
More informationCONFOCAL GRADED d-spacing MULTILAYER BEAM CONDITIONING OPTICS
Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 321 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42
More informationToday s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26
Today s Outline - January 25, 2018 C. Segre (IIT) PHYS 570 - Spring 2018 January 25, 2018 1 / 26 Today s Outline - January 25, 2018 HW #2 C. Segre (IIT) PHYS 570 - Spring 2018 January 25, 2018 1 / 26 Today
More informationWebinar Organizers. Ryan Shea. Don Miller. Joe Ryan. Support Specialist. Applications Specialist. Product Manager. Precision Digital Corporation
Webinar Organizers Joe Ryan Product Manager Precision Digital Corporation Ryan Shea Applications Specialist Precision Digital Corporation Don Miller Support Specialist Precision Digital Corporation Agenda,
More informationMethod for digital particle spectrometry Khryachkov Vitaly
Method for digital particle spectrometry Khryachkov Vitaly Institute for physics and power engineering (IPPE) Obninsk, Russia The goals of Analog Signal Processing Signal amplification Signal filtering
More informationX-ray Fluorescence of Some Egyptian Coins
International Journal of Pure and Applied Physics. ISSN 0973-1776 Volume 8, Number 2 (2012), pp. 69-78 Research India Publications http://www.ripublication.com/ijpap.htm X-ray Fluorescence of Some Egyptian
More informationX-RAY BACKSCATTER IMAGING: PHOTOGRAPHY THROUGH BARRIERS
Copyright JCPDS-International Centre for Diffraction Data 2006 ISSN 1097-0002 X-RAY BACKSCATTER IMAGING: PHOTOGRAPHY THROUGH BARRIERS 13 Joseph Callerame American Science & Engineering, Inc. 829 Middlesex
More informationTitle detector with operating temperature.
Title Radiation measurements by a detector with operating temperature cryogen Kanno, Ikuo; Yoshihara, Fumiki; Nou Author(s) Osamu; Murase, Yasuhiro; Nakamura, Masaki Citation REVIEW OF SCIENTIFIC INSTRUMENTS
More informationCHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES
CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES The current multiplication mechanism offered by dynodes makes photomultiplier tubes ideal for low-light-level measurement. As explained earlier, there
More informationCharge Loss Between Contacts Of CdZnTe Pixel Detectors
Charge Loss Between Contacts Of CdZnTe Pixel Detectors A. E. Bolotnikov 1, W. R. Cook, F. A. Harrison, A.-S. Wong, S. M. Schindler, A. C. Eichelberger Space Radiation Laboratory, California Institute of
More informationTHERMAL NOISE. Advanced Laboratory, Physics 407, University of Wisconsin. Madison, Wisconsin 53706
(revised 1/25/07) THERMAL NOISE Advanced Laboratory, Physics 407, University of Wisconsin Madison, Wisconsin 53706 Abstract The aim of this experiment is to observe the thermal noise in a resistor, to
More informationRadiation Detection Instrumentation
Radiation Detection Instrumentation Principles of Detection and Gas-filled Ionization Chambers Neutron Sensitive Ionization Chambers Detection of radiation is a consequence of radiation interaction with
More informationX-ray fluorescence Analyzer
X-ray Fluorescence Spectrometer New Energy Dispersive X-ray fluorescence Analyzer www.matsusada.com Measures for compliance with the RoHS / ELV and other regulations Lead content control in the solder
More informationAnalog-to-Digital-Converter User Manual
7070 Analog-to-Digital-Converter User Manual copyright FAST ComTec GmbH Grünwalder Weg 28a, D-82041 Oberhaching Germany Version 2.0, July 7, 2005 Software Warranty FAST ComTec warrants proper operation
More informationELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone
ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY 12477 (800)VIP-XRAY (845)247-9028 Fax or 800-847-9729 local phone 845-246-9013 http://www.eccxray.com sales@eccxray.com INSTRUCTION MANUAL
More informationARTAX. Innovation with Integrity. Portable Micro-XRF Spectrometer. Micro-XRF
ARTAX Portable Micro-XRF Spectrometer Innovation with Integrity Micro-XRF ARTAX Elemental Analysis for the Art Community and More The ARTAX is the first portable X ray fluorescence (XRF) spectrometer designed
More informationANALYSIS OF PIGMENTS USED IN SCROLL PAINTINGS OF A NATIONAL TREASURE "TALE OF GENJI" USING A PORTABLE X-RAY FLUORESCENCE SPECTROMETER
Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 432 ANALYSIS OF PIGMENTS USED IN SCROLL PAINTINGS OF A NATIONAL TREASURE "TALE OF GENJI" USING A PORTABLE X-RAY FLUORESCENCE SPECTROMETER
More informationLUDLUM MODEL MODEL AND MODEL GAMMA SCINTILLATORS. June 2017
LUDLUM MODEL 44-20 MODEL 44-20-1 AND MODEL 44-20-3 GAMMA SCINTILLATORS June 2017 LUDLUM MODEL 44-20 MODEL 44-20-1 AND MODEL 44-20-3 GAMMA SCINTILLATORS June 2017 STATEMENT OF WARRANTY Ludlum Measurements,
More information