MOXTEK S NEW ULTRA-LITE X-RAY SOURCES: PERFORMACE CHARACTERIZATIONS
|
|
- Jasper O’Neal’
- 5 years ago
- Views:
Transcription
1 Copyright JCPDS-International Centre for Diffraction Data 2013 ISSN MOXTEK S NEW ULTRA-LITE X-RAY SOURCES: PERFORMACE CHARACTERIZATIONS S. Cornaby, S. Morris, J. Smith, D. Reynolds, K. Kozaczek Moxtek Inc. Orem UT USA ABSTRACT Moxtek is a manufacturer of miniature x-ray sources which are used in portable and handheld XRF instruments. Moxtek is undertaking a substantial effort to fully characterize their mini x- ray sources, with the intent to provide confidence for those using Moxtek s X-ray sources. The characterization test fall into three categories: environmental stress, X-ray flux and spot stability and repeatability, and a multi-settings evaluation. Examples are given of data collected from these characterization tests from evaluation of Moxtek s newest x-ray source, the ULTRA-LITE, which is a 50 kv, 200 µa source with a maximum power of 4 watts and a weight of only 250 grams. INTRODUCTION There have been tremendous advancements in miniature x-ray tube technology over the past ten years, and it has been increasingly critical to predict a miniature x-ray tube s performance in a number of applications. Moxtek is undertaking a substantial effort to fully characterize its mini x-ray sources during their development with the intent to provide confidence for using these X-ray sources in a variety of applications and environments. The basic methodology for testing involves using several detectors and instruments simultaneously measuring and evaluating the x-ray source. The characterization tests fall into three categories: environmental stress, X-ray 46 mm 25 mm 148 mm 94 mm Figure 1. Image of the ULTRA LITE X-ray source in two different orientations. HVPS output kv range 5-50 kv HVPS emission current range µa Anode Material Ag, W, and others HVPS power limit on x-ray tube 4 Watts HVPS power input-voltage range 6-18 V DC Total Weight 250 gm (8.8 oz) Table 1. The ULTRA LITE X-ray source s basic parameters. HVPS =High Voltage Power Supply
2 This document was presented at the Denver X-ray Conference (DXC) on Applications of X-ray Analysis. Sponsored by the International Centre for Diffraction Data (ICDD). This document is provided by ICDD in cooperation with the authors and presenters of the DXC for the express purpose of educating the scientific community. All copyrights for the document are retained by ICDD. Usage is restricted for the purposes of education and scientific research. DXC Website ICDD Website -
3 Copyright JCPDS-International Centre for Diffraction Data 2013 ISSN flux and spot stability and repeatability, and a multi-settings evaluation. Moxtek s most recent source is the ULTRA-LITE, which in development was subjected to Moxtek s new characterization testing (Figure 1 and Table 1). Each of the following sections explain these tests, what their benefits are, and show examples of data collected during the development of the ULTRA-LITE source. ENVIROMENTAL STRESS TEST One of the more critical constraints on portable x-ray sources is the environment. These small x-ray sources need to operate in the field, wherever they are taken. This translates into having a storage temperature range of -40 C to 80 C and an operational range of -10 C to 70 C. While developing a source, the design is considered to be robust when the x-ray source can survive a 120 hour environmental test during which the temperature profile in Figure 2 is repeated twenty times. In the environmental test we can test up to eight x-ray sources at the same time. Temp C ON Storage Temp. Range Enviromental stress test Temp. profile Operational Temp. Range Source on/off state OFF hours 4 5 Figure 2. The temperature profile from a single cycle of the environmental stress test & a profile showing the on/off state of the X-ray source. This cycle is repeated 20 times for the full environmental test. ULTRA-LITE sources were tested with the environmental stress test. Several iterations of this environmental test were performed. Each testing cycle found weaknesses in the design, and the failure points were renovated. This cycle continued until we had an ULTRA-LITE design able to survive this testing. X-RAY FLUX AND SPOT STABILITY AND REPEATABILITY TEST The stability and repeatability test evaluates how the x-ray source operates over time. For the stability test, the x-ray source is left on for the entire test. For the repeatability test, the x-ray source is periodically cycled on and off, typical off cycles are 2 to 10 seconds and typical on cycles are 30 to 120 seconds. The test durations can be from an hour to days. For the stability and repeatability tests, several detectors and instruments are simultaneously measuring and evaluating the x-ray source (Figure 3).
4 Copyright JCPDS-International Centre for Diffraction Data 2013 ISSN For both the stability and repeatability tests, we measure the flux directly from the source, which gives the flux stability (Figure 4), and the high voltage stability is measured using the bremsstrahlung edge (Figure 5). We also measure the flux off a 316 stainless steel secondary target, simulating XRF instruments stability. We measure the flux with a photo-diode in current mode, which provides x-ray flux stability information a third way. We measure a time series of pinhole images of the x-ray tube s anode for x-ray spot stability (Figure 6). In Figure 6, the ULTRA-LITE source generates a spot size of 390x340 µm FWHM, a 70% x-ray spot centered flux diameter of 600 um, and a 70% anode centered flux diameter of 950 um (D. Caruso et al., 2010). We have five thermocouples on different locations of the source tube and high voltage power supply, and we monitor every signal and power line going into and out-of the high voltage power supply. X-ray source XRF target Imaging pinhole 2D X-ray imaging camera Photo-diode XRF detector Direct detector & pinhole Figure 3. A simple schematic of the stability and repeatability test showing all the major instruments, three flux detectors and an imaging detector, simultaneously measuring and evaluating the x-ray source.
5 Copyright JCPDS-International Centre for Diffraction Data 2013 ISSN Figure 4. An X-ray flux measured directly from ULTRA-LITE source, set at 50 kv and 80 µa (the full 4 watts) during a repeatability test, cycled on for 120 seconds and off for 5 seconds for 2400 cycles. This test measured an x-ray flux repeatability of 0.10% RSD over 85 hours. During the source s warm-up time of minutes, the x-ray flux drifted by 0.45%. Figure 5. A high voltage repeatability test from ULTRA-LITE source from same experiment in Figure 4. The top left figure shows a single X-ray energy spectrum collected directly from the source. The top right figure shows the bremsstrahlung edge of the same spectrum. An algorithm finds the edge, and records it for each of the 2400 spectra taken over 85 hours. This source achieved a voltage of 49.5 kv (set at 50 kv), with a standard deviation of 50 V. 205
6 Copyright JCPDS-International Centre for Diffraction Data 2013 ISSN mm Figure 6. An X-ray spot repeatability test from ULTRA-LITE source, taken simultaneously with information in Figures 4 and 5. To the left is a single pinhole camera image of the X-ray spot, with a spot size of 390x340 µm FWHM. The small white dot in the center of the image represents the physical center of the anode. The graph to the right shows the drift of the spot over time; it drifts ~20 µm as the source warms up over the first 30 to 60 minutes and has a constant position thereafter over 85 hours. Figures 4-6 represent just a few highlights of data collected simultaneously from the x-ray source. Collecting all this information simultaneously allows for finding correlations in the data, which is critical for trouble shooting issues as problems are found. All the detectors and other hardware used are controlled with LabVIEW. In addition, LabVIEW writes a final report for each test, the graphs and images in Figures 4-6 are from this self-generated report. MULTI-SETTINGS TEST In the multi-settings test, we run the X-ray source over its entire range of voltage and emission current settings, to evaluate the source. We go to each setting in a random order, turn the source on and off, and run the source for about five seconds. For example, one series of tests done on the ULTRA-LITE, the source was tested at every setting in 1µA steps (ranging from µa) and 0.5 kev steps (ranging from 5-50 kev). This resulted in testing over 13,000 individual settings, with the entire multi-settings test taking about 24 hours. For this test, the flux is measured with a photo-diode in current mode, Figure 7. This shows the x-ray flux signal measured over time by the photo-diode as the source turns-on. For this setting of 50 kv and 80 µa, to the flux is fully on after 0.51 seconds.
7 Copyright JCPDS-International Centre for Diffraction Data 2013 ISSN Maximum Emission current 5 µa Emission kv ~ µa ~0.7 sec ~ µa ~0.6 sec ~ µa ~0.6 sec ~ µa ~0.5 sec ~ µa ~0.8 sec ~ µa ~1.1 sec ~ µa ~1.4 sec Table 2. This shows the ULTRA-LITE source turning on time over a few selected settings. A turn-on speed of less than 1 second is considered an adequate turn-on time for a handheld XRF instrument. which provides x-ray flux stability information over a short time frame and the turn on speed of the x-ray source. Again we measure as much as we can simultaneously, we have thermocouples on different locations of the source, and we monitor every signal and power line going into and out-of the high voltage power supply. This provides a great deal of good information about the functionality of the source over all the potential settings. Figure 7 and Table 2 gives information about the turn-on speed of the ULTRA-LITE source, which is defined to be the time between time between the high voltage TTL being enabled to the x-ray flux coming on to 99% of its maximum value. Figure 8. An example of data collected from the multi-settings test where the set voltage, and the set current are plotted against the standard deviation of the x-ray flux measured with a photodiode. An unstable flux output region is clearly seen inside the red circle. This issue was resolve in later designs. Figure 8 gives one example of the flux stability at a number of different settings after the source has turned on at each setting. A low standard deviation in the flux signal represents the desired stable flux output, and is represented in blue dots in the plot of Figure 8. A high standard deviation in the flux signal represents the undesired unstable flux output, and is represented in either red or yellow dots in the plot of Figure 8. In Figure 8, settings with unstable flux stability are clearly identified in the region circled in red. This graph is data from an earlier ULTRA- LITE source design; we improved the design of the analog control board to resolve this flux instability issue.
8 Copyright JCPDS-International Centre for Diffraction Data 2013 ISSN kv & 80 µa 4 40 kv & 100 ua 4 20 kv & 200 µa 2 40 kv & 50 µa 2 10 kv & 200 µa 1 10 kv & 100 µa 1 5 kv & 200 µa 18 V 10.4 W 9.9 W 10.3 W 7.3 W 7.5 W 5.5 W 6.0 W 12 V 9.2 W 8.8 W 9.1 W 6.0 W 6.3 W 4.1 W 4.6 W 9 V 8.6 W 8.1 W 8.5 W 5.5 W 5.7 W 3.7 W 4.3 W 6 V 8.6 W 8.2 W 8.6 W 5.4 W 5.6 W 3.5W 4.0 W Table 3. This table shows the total power drawn by the ULTRA-LITE x-ray source at different X-ray tube settings (in each column) and at different DC power input settings (in each row) ranging from 6 to 18 Volts DC. One last example of data collected during the multi-settings test is the total power drawn by the source. Table 3 gives information about the total power draw of the source depending on the both the x-ray tube power and the input DC voltage supplied to the ULTRA-LITE source. The multi-settings test has been good at identifying issues with the source, as well as providing how inputs and outputs of the source change over the X-ray source s range of settings. SUMMARY AND CONCLUSIONS Moxtek is undertaking a substantial effort to fully characterize their mini x-ray sources during development with the intent to provide confidence for those using these X-ray sources in a variety of applications and environments. Our methodology for testing involves using several detectors and instruments simultaneously measuring and evaluating the x-ray source. The ULTRA-LITE source has been rigorously tested in development over these three categories of tests: environmental stress tests, X-ray flux and spot stability and repeatability tests, and multi-settings tests. We have found the ULTRA-LITE source to be very stable, with a 0.10% RSD over several hours (Figure 4 gives one example). Also, the ULTRA-LITE has as very stable high voltage, not varying more than 50 V while set at 50 kv, measured using the bremsstrahlung edge. REFERENCES Caruso D., Dinsmore M., and Cornaby S. (2010), Miniature X-ray sources and the effects of spot size on system performance, 58 th Denver X-ray Conference, Advances in X-ray Analysis. 53, pp LabVIEW
MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE
228 MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE D. CARUSO, M. DINSMORE TWX LLC, CONCORD, MA 01742 S. CORNABY MOXTEK, OREM, UT 84057 ABSTRACT Miniature x-ray sources present
More informationSIMULTANEOUS XRD/XRF WITH LOW-POWER X-RAY TUBES
Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 34 SIMULTANEOUS XRD/XRF WITH LOW-POWER X-RAY TUBES S. Cornaby 1, A. Reyes-Mena 1, P. W. Moody 1,
More informationUSING A CHARGE-COUPLED DEVICE (CCD) TO GATHER X-RAY FLUORESCENCE (XRF)AND X-RAY DIFFRACTION (XRD) INFORMATION SIMULTANEOUSLY
Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 343 USING A CHARGE-COUPLED DEVICE (CCD) TO GATHER X-RAY FLUORESCENCE (XRF)AND X-RAY DIFFRACTION (XRD)
More informationMICRO XRF OF LIGHT ELEMENTS USING A POLYCAPILLARY LENS AND AN ULTRA THIN WINDOW SILICON DRIFT DETECTOR INSIDE A VACUUM CHAMBER
Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 229 MICRO XRF OF LIGHT ELEMENTS USING A POLYCAPILLARY LENS AND AN ULTRA THIN WINDOW SILICON DRIFT
More informationDEVELOPMENT OF A WAVELENGTH DISPERSIVE X-RAY FLUORESCENCE SPECTROMETER USING A MULTI-CAPILLARY X-RAY LENS FOR X-RAY DETECTION
Copyright JCPDS - International Centre for Diffraction Data 2003, Advances in X-ray Analysis, Volume 46. 346 DEVELOPMENT OF A WAVELENGTH DISPERSIVE X-RAY FLUORESCENCE SPECTROMETER USING A MULTI-CAPILLARY
More informationCopyright -International Centre for Diffraction Data 2010 ISSN
234 BRIDGING THE PRICE/PERFORMANCE GAP BETWEEN SILICON DRIFT AND SILICON PIN DIODE DETECTORS Derek Hullinger, Keith Decker, Jerry Smith, Chris Carter Moxtek, Inc. ABSTRACT Use of silicon drift detectors
More informationANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER
Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 325 ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER ABSTRACT William Chang, Jonathan Kerner, and Edward
More informationDOUBLE MULTILAYER MONOCHROMATOR WITH FIXED EXIT GEOMETRY. H.Gatterbauer, P.Wobrauschek, F.Hegediis, P.Biini, C.Streli
Copyright (C) JCPDS International Centre for Diffraction Data 1999 379 DOUBLE MULTILAYER MONOCHROMATOR WITH FIXED EXIT GEOMETRY H.Gatterbauer, P.Wobrauschek, F.Hegediis, P.Biini, C.Streli Atominsitut der
More informationWIDE ANGLE GEOMETRY EDXRF SPECTROMETERS WITH SECONDARY TARGET AND DIRECT EXCITATION MODES
Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 11 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42
More informationX-RAY BACKSCATTER IMAGING: PHOTOGRAPHY THROUGH BARRIERS
Copyright JCPDS-International Centre for Diffraction Data 2006 ISSN 1097-0002 X-RAY BACKSCATTER IMAGING: PHOTOGRAPHY THROUGH BARRIERS 13 Joseph Callerame American Science & Engineering, Inc. 829 Middlesex
More informationGUNSHOT RESIDUE INVESTIGATIONS USING TXRF
299 GUNSHOT RESIDUE INVESTIGATIONS USING TXRF Alexander Wastl 1, Bettina Bogner 2, Peter Kregsamer 1, Peter Wobrauschek 1, Christina Streli 1 1 Atominstitut, Vienna University of Technology, Vienna, Austria
More informationLONG TERM STATISTICS OF X-RAY SPECTROMETERS
403 LONG TERM STATISTICS OF X-RAY SPECTROMETERS J. F. Dlouhy*, D. Mathieu Department of the Environment, Environmental Technology Center, River Road, Ottawa, Ontario, Canada Kl A OH3 K. N. Stoev Bulgarian
More informationA MONTE CARLO CODE FOR SIMULATION OF PULSE PILE-UP SPECTRAL DISTORTION IN PULSE-HEIGHT MEASUREMENT
Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 246 A MONTE CARLO CODE FOR SIMULATION OF PULSE PILE-UP SPECTRAL DISTORTION IN PULSE-HEIGHT MEASUREMENT
More informationMINIATURE X-RAY TUBES UTILIZING CARBON-NANOTUBE- BASED COLD CATHODES
Copyright JCPDS - International Centre for Diffraction Data 25, Advances in X-ray Analysis, Volume 48. 24 MINIATURE X-RAY TUBES UTILIZING CARBON-NANOTUBE- BASED COLD CATHODES A. Reyes-Mena, Charles Jensen,
More informationSpectral distribution from end window X-ray tubes
Copyright ISSN (C) 1097-0002, JCPDS-International Advances in X-ray Centre Analysis, for Volume Diffraction 41 Data 1999 393 Spectral distribution from end window X-ray tubes N. Broll 1, P. de Chateaubourg
More informationX-RAY OPTICS FOR TWO-DIMENSIONAL DIFFRACTION
Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 332 ABSTRACT X-RAY OPTICS FOR TWO-DIMENSIONAL DIFFRACTION Bob B. He and Uwe Preckwinkel Bruker
More informationApplications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror
Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror Stephen B. Robie scintag, Inc. 10040 Bubb Road Cupertino, CA 95014 Abstract Corundum
More informationTOWARDS SUB-100 NM X-RAY MICROSCOPY FOR TOMOGRAPHIC APPLICATIONS
Copyright -International Centre for Diffraction Data 2010 ISSN 1097-0002 89 TOWARDS SUB-100 NM X-RAY MICROSCOPY FOR TOMOGRAPHIC APPLICATIONS P. Bruyndonckx, A. Sasov, B. Pauwels Skyscan, Kartuizersweg
More informationMICROFOCUSING X-RAY EQUIPMENT FOR THE LAB DIFFRACTOMETER
29 MICROFOCUSING X-RAY EQUIPMENT FOR THE LAB DIFFRACTOMETER Jörg Wiesmann, 1 Jürgen Graf, 1 Christian Hoffmann, 1 Carsten Michaelsen, 1 Alexandra Oehr, 1 Uwe Preckwinkel, 2 Ning Yang, 2 Holger Cordes,
More informationApplications of Micro XRF for the Analysis of Traditional Japanese "Ainu" Glass Beads and other Artifacts
161 161 Applications of Micro XRF for the Analysis of Traditional Japanese "Ainu" Glass Beads and other Artifacts K.Sugihara 1, M.Satoh 1, Y.Hayakawa 2, A.Saito 3 and T.Sasaki 4 1 Seiko Instruments Inc.,
More informationTOWARDS FAST RECIPROCAL SPACE MAPPING
Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 165 ABSTRACT TOWARDS FAST RECIPROCAL SPACE MAPPING J.F. Woitok and A. Kharchenko PANalytical B.V.,
More informationCA SERIES PRECISION REGULATED, LOW RIPPLE HIGH VOLTAGE DC TO DC CONVERTERS
Proven Reliability CA SERIES PRECISION REGULATED, LOW RIPPLE HIGH DC TO DC CONVERTERS 200V to 2000V @ Watt PRODUCT SELECTION TABLE NOW UL RECOGNIZED PRODUCT DESCRIPTION Precision regulated, very low noise,
More informationRIETVELD REFINEMENT OF POWDER DATA FROM MULTILAYER OPTICS
Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 166 RIETVELD REFINEMENT OF POWDER DATA FROM MULTILAYER OPTICS ABSTRACT Scott T. Misture NYS College
More informationLow Cost Rolled X-ray Prism Lenses to Increase Photon Flux Density in Diffractometry Experiments
Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN 1097-0002 17 Low Cost Rolled X-ray Prism Lenses to Increase Photon Flux Density in Diffractometry Experiments H. Vogt a, A. Last a, J.
More informationABSOLUTE MAXIMUM RATINGS These ratings cannot necessarily be used simultaneously and no individual ratings should be exceeded.
M1621B The M1621B is an electronic frequency tuning pulsed type X-band magnetron, designed to operate at 938 to 944 MHz with a peak output power of 4kW. The oscillation frequency is tuned by applying bias
More informationCA SERIES PRECISION REGULATED, LOW RIPPLE HIGH VOLTAGE DC TO DC CONVERTERS
Proven Reliability CA SERIES PRECISION REGULATED, LOW RIPPLE HIGH VOLTAGE DC TO DC CONVERTERS 200V to 2000V @ 1 Watt PRODUCT SELECTION TABLE NOW UL RECOGNIZED PRODUCT DESCRIPTION Precision regulated, very
More informationFunctions of the SEM subsystems
Functions of the SEM subsystems Electronic column It consists of an electron gun and two or more electron lenses, which influence the path of electrons traveling down an evacuated tube. The base of the
More informationA PORTABLE X-RAY APPARATUS FOR BOTH STRESS MEASUREMENT AND PHASE ANALYSIS UNDER FIELD CONDITIONS.
Copyright(c)JCPDS-International Centre for Diffraction Data 2000,Advances in X-ray Analysis,Vol.43 66 A PORTABLE X-RAY APPARATUS FOR BOTH STRESS MEASUREMENT AND PHASE ANALYSIS UNDER FIELD CONDITIONS. V.
More informationDiffraction-enhanced X-ray Imaging (DEXI) Medical Solutions. More information using less radiation
Diffraction-enhanced X-ray Imaging (DEXI) Medical Solutions More information using less radiation Medical Small Animal Security NDE/NDT Diffraction-Enhanced X-ray Imaging Medical Solutions Safe non-invasive
More informationCHAPTER 11 HPD (Hybrid Photo-Detector)
CHAPTER 11 HPD (Hybrid Photo-Detector) HPD (Hybrid Photo-Detector) is a completely new photomultiplier tube that incorporates a semiconductor element in an evacuated electron tube. In HPD operation, photoelectrons
More informationAG SERIES ISOLATED, PROPORTIONAL DC TO HV DC CONVERTERS
Proven Reliability ISOLATED, PROPORTIONAL DC TO HV DC CONVERTERS 100V to 6000V @ 1.0 and 1.5 Watts Extremely Low Profile: 0.128 inches and volume of < 0.100 cubic inches *5 NOW UL RECOGNIZED PRODUCT DESCRIPTION
More informationXrf Guide READ ONLINE
Xrf Guide READ ONLINE If looking for the book Xrf guide in pdf form, then you have come on to loyal website. We furnish the full edition of this ebook in DjVu, epub, PDF, txt, doc forms. You may reading
More informationABSTRACT. Supported by U.S. DoE grant No. DE-FG02-96ER54375
ABSTRACT A CCD imaging system is currently being developed for T e (,t) and bolometric measurements on the Pegasus Toroidal Experiment. Soft X-rays (E
More informationDESIGN AND MEASUREMENT WITH A NEW PORTABLE X-RAY CAMERA FOR FULL-FIELD FLUORESCENCE IMAGING
14 DESIGN AND MEASUREMENT WITH A NEW PORTABLE X-RAY CAMERA FOR FULL-FIELD FLUORESCENCE IMAGING I. Ordavo 1,2, A. Bjeoumikhov 3, S. Bjeoumikhova 3, G. Buzanich 4, R. Gubzhokov 4, R. Hartmann 1, S. Ihle
More informationPrinciples of operation 5
Principles of operation 5 The following section explains the fundamental principles upon which Solartron Metrology s linear measurement products are based. > Inductive technology (gauging and displacement)
More informationReal Time Pulse Pile-up Recovery in a High Throughput Digital Pulse Processor
Real Time Pulse Pile-up Recovery in a High Throughput Digital Pulse Processor Paul A. B. Scoullar a, Chris C. McLean a and Rob J. Evans b a Southern Innovation, Melbourne, Australia b Department of Electrical
More informationIDENTIFICATION OF PAINTING MATERIALS USED FOR MURAL PAINTINGS BY IMAGE ANALYSIS AND XRF
IDENTIFICATION OF PAINTING MATERIALS USED FOR MURAL PAINTINGS BY IMAGE ANALYSIS AND XRF 213 Seiji SHIRONO, Yasuhiro HAYAKAWA National Research Institute for Cultural Properties, Tokyo, Japan ABSTRACT The
More informationThe Standard for over 40 Years
Light Measurement The Standard for over 40 Years Introduction LI-COR radiation sensors measure the flux of radiant energy the energy that drives plant growth, warms the earth, and lights our world. The
More informationCONFOCAL GRADED d-spacing MULTILAYER BEAM CONDITIONING OPTICS
Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 321 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42
More informationLaser Speckle Reducer LSR-3000 Series
Datasheet: LSR-3000 Series Update: 06.08.2012 Copyright 2012 Optotune Laser Speckle Reducer LSR-3000 Series Speckle noise from a laser-based system is reduced by dynamically diffusing the laser beam. A
More informationMany applications. Mismatched Load Characterization for High-Power RF Amplifiers PA CHARACTERIZATION. This article discusses the
From April 2004 High Frequency Electronics Copyright 2004 Summit Technical Media, LLC Mismatched Load Characterization for High-Power RF Amplifiers By Richard W. Brounley, P.E. Brounley Engineering Many
More informationNano Beam Position Monitor
Introduction Transparent X-ray beam monitoring and imaging is a new enabling technology that will become the gold standard tool for beam characterisation at synchrotron radiation facilities. It allows
More informationX-Ray Spectroscopy with a CCD Detector. Application Note
X-Ray Spectroscopy with a CCD Detector In addition to providing X-ray imaging solutions, including CCD-based cameras that image X-rays using either direct detection (0.5-20 kev) or indirectly using a scintillation
More informationDESIGN OF EDXRF EQUIPMENT FOR THE NONDESTRUCTIVE STUDY OF PRINTS
Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 70 DESIGN OF EDXRF EQUIPMENT FOR THE NONDESTRUCTIVE STUDY OF PRINTS 1,3 M. Ardid, 1 J.L. Ferrero,
More informationLower Hybrid. Ron Parker Alcator C-Mod PAC Meeting January January 2006 Alcator C-Mod PAC Meeting 1
Lower Hybrid Ron Parker Alcator C-Mod PAC Meeting 25-27 January 2006 25-27 January 2006 Alcator C-Mod PAC Meeting 1 Goal of Lower Hybrid Current Drive Experiments Use Lower Hybrid Current Drive to supplement
More informationXRF Instrumentation. Introduction to spectrometer
XRF Instrumentation Introduction to spectrometer AMPTEK, INC., Bedford, MA 01730 Ph: +1 781 275 2242 Fax: +1 781 275 3470 sales@amptek.com 1 Instrument Excitation source Sample X-ray tube or radioisotope
More informationAdvancing EDS Analysis in the SEM Quantitative XRF. International Microscopy Congress, September 5 th, Outline
Advancing EDS Analysis in the SEM with in-situ Quantitative XRF Brian J. Cross (1) & Kenny C. Witherspoon (2) 1) CrossRoads Scientific, El Granada, CA 94018, USA 2) ixrf Systems, Inc., Houston, TX 77059,
More informationMOXTEK. 50kV 10 Watt MAGNUM X-ray Source. X-ray Sources. Contents
X-ray Sources 0kV 10 Watt MAGNUM X-ray Source Manual Contents MAGNUM X-ray Source Characteristics Initial Inspection and Handling Tube Setup Operating Conditions Operating Precautions and Warnings Operating
More informationHomework Set 3.5 Sensitive optoelectronic detectors: seeing single photons
Homework Set 3.5 Sensitive optoelectronic detectors: seeing single photons Due by 12:00 noon (in class) on Tuesday, Nov. 7, 2006. This is another hybrid lab/homework; please see Section 3.4 for what you
More informationZ-LASER Optoelektronik GmbH Stemmer 3d Technologietag Useful information on Z-Lasers for Vision
Z-LASER Optoelektronik GmbH Stemmer 3d Technologietag - 24.2.2011 Useful information on Z-Lasers for Vision The Company Core Competences How to Build a Z-LASER Electronics and Modulation Wavelength and
More informationAbstract. 1. Introduction
941 Monte Carlo Simulation of Pulse Pile Up R. P. Gardner and S. H. Lee Center for Engineering Applications of Radioisotopes Department of Engineering North Carolina State University Raleigh, NC 276957909,
More informationCHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT
CHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT In this chapter, the experimental results for fine-tuning of the laser wavelength with an intracavity liquid crystal element
More informationSilicon Drift Detector. with On- Chip Ele ctronics for X-Ray Spectroscopy. KETEK GmbH Am Isarbach 30 D O berschleißheim GERMANY
KETEK GmbH Am Isarbach 30 D-85764 O berschleißheim GERMANY Silicon Drift Detector Phone +49 (0)89 315 57 94 Fax +49 (0)89 315 58 16 with On- Chip Ele ctronics for X-Ray Spectroscopy high energy resolution
More informationA SERIES ISOLATED, PROPORTIONAL DC TO HV DC CONVERTERS
Proven Reliability A SERIES ISOLATED, PROPORTIONAL DC TO HV DC CONVERTERS 100V to 6000V @ 1.0 and 1.5 Watts Extremely Low Profile: 0.25 inches and volume of < 0.100 cubic inches *5 NOW UL RECOGNIZED PRODUCT
More informationTemperature Dependent Dark Reference Files: Linear Dark and Amplifier Glow Components
Instrument Science Report NICMOS 2009-002 Temperature Dependent Dark Reference Files: Linear Dark and Amplifier Glow Components Tomas Dahlen, Elizabeth Barker, Eddie Bergeron, Denise Smith July 01, 2009
More informationAgilOptics mirrors increase coupling efficiency into a 4 µm diameter fiber by 750%.
Application Note AN004: Fiber Coupling Improvement Introduction AgilOptics mirrors increase coupling efficiency into a 4 µm diameter fiber by 750%. Industrial lasers used for cutting, welding, drilling,
More informationCARBON-NANOTUBE FIELD EMISSION X-RAY TUBE FOR SPACE EXPLORATION XRD/XRF INSTRUMENT.
Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 232 CARBON-NANOTUBE FIELD EMISSION X-RAY TUBE FOR SPACE EXPLORATION XRD/XRF INSTRUMENT. P. Sarrazin
More informationInterpixel crosstalk in a 3D-integrated active pixel sensor for x-ray detection
Interpixel crosstalk in a 3D-integrated active pixel sensor for x-ray detection The MIT Faculty has made this article openly available. Please share how this access benefits you. Your story matters. Citation
More informationCHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES
CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES The current multiplication mechanism offered by dynodes makes photomultiplier tubes ideal for low-light-level measurement. As explained earlier, there
More informationA SERIES ISOLATED, PROPORTIONAL DC TO HV DC CONVERTERS
Proven Reliability ISOLATED, PROPORTIONAL DC TO HV DC CONVERTERS 100V to 6000V @ 1.0 and 1.5 Watts Volum Extremely Low Profile: 0.25 inches and volume of < 0.100 cubic inches *5 NOW UL Recognized PRODUCT
More informationAutoMATE II. Micro-area X-ray stress measurement system. Highly accurate micro area residual stress
AutoMATE II Micro-area X-ray stress measurement system Highly accurate micro area residual stress The accuracy of an R&D diffractom dedicated residua In the past, if you wanted to make highly accurate
More informationOPTIMIZING THE ELEMENTAL SENSITIVITY AND FOCAL SPOT SIZE OF A MONOLITHIC POLYCAPILLARY OPTIC USING MICRO-X-RAY FLUORESCENCE
, Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 26 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42
More informationMobile digital radiography system for nondestructive testing of large diameter pipelines
18th World Conference on Nondestructive Testing, 16-20 April 2012, Durban, South Africa Mobile digital radiography system for nondestructive testing of large diameter pipelines Vasily A. KLIMENOV, Aleksey
More informationON THE ENERGY DEPENDENCE OF THE DETECTOR EFFICIENCY OF A Si-PIN DIODE
Copyright(C)JCPDS-International Centre for Diffraction Data 2, Advances in X-ray Analysis, Vol.42 36 Copyright(C)JCPDS-International Centre for Diffraction Data 2, Advances in X-ray Analysis, Vol.42 36
More informationInvestigation of Effect of Chatter Amplitude on Surface Roughness during End Milling of Medium Carbon Steel
Proceedings of the 2010 International Conference on Industrial Engineering and Operations Management Dhaka, Bangladesh, January 9 10, 2010 Investigation of Effect of Chatter Amplitude on Surface Roughness
More informationQuality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT
Quality control of Gamma Camera By Dr/ Ibrahim Elsayed Saad 242 NMT WHAT IS QUALITY? The quality of a practice is to fulfill the expectations and demands from: Patient Clinicain Your self Quality assurance
More informationCO2 laser heating system for thermal compensation of test masses in high power optical cavities. Submitted by: SHUBHAM KUMAR to Prof.
CO2 laser heating system for thermal compensation of test masses in high power optical cavities. Submitted by: SHUBHAM KUMAR to Prof. DAVID BLAIR Abstract This report gives a description of the setting
More informationPractical Quadrupole Theory: Graphical Theory
Extrel Application Note RA_21A Practical Quadrupole Theory: Graphical Theory Randall E. Pedder ABB Inc., Analytical-QMS Extrel Quadrupole Mass Spectrometry, 575 Epsilon Drive, Pittsburgh, PA 15238 (Poster
More informationPhotons and solid state detection
Photons and solid state detection Photons represent discrete packets ( quanta ) of optical energy Energy is hc/! (h: Planck s constant, c: speed of light,! : wavelength) For solid state detection, photons
More informationTHE CRYSTAL OSCILLATOR CHARACTERIZATION FACILITY AT THE AEROSPACE CORPORATION
THE CRYSTAL OSCILLATOR CHARACTERIZATION FACILITY AT THE AEROSPACE CORPORATION S. Karuza, M. Rolenz, A. Moulthrop, A. Young, and V. Hunt The Aerospace Corporation El Segundo, CA 90245, USA Abstract At the
More informationLow Contrast Dielectric Metasurface Optics. Arka Majumdar 1,2,+ 8 pages, 4 figures S1-S4
Low Contrast Dielectric Metasurface Optics Alan Zhan 1, Shane Colburn 2, Rahul Trivedi 3, Taylor K. Fryett 2, Christopher M. Dodson 2, and Arka Majumdar 1,2,+ 1 Department of Physics, University of Washington,
More informationANALYSIS OF ELECTRON CURRENT INSTABILITY IN E-BEAM WRITER. Jan BOK, Miroslav HORÁČEK, Stanislav KRÁL, Vladimír KOLAŘÍK, František MATĚJKA
ANALYSIS OF ELECTRON CURRENT INSTABILITY IN E-BEAM WRITER Jan BOK, Miroslav HORÁČEK, Stanislav KRÁL, Vladimír KOLAŘÍK, František MATĚJKA Institute of Scientific Instruments of the ASCR, v. v.i., Královopolská
More informationCharacterization of Cd primary and secondary reference materials
Fischer Traceability Report JL-2013-10-Cd Sindelfingen, 10/09/2013 Characterization of Cd primary and secondary reference materials Abstract: Three new Cd primary reference standards (Cd foils) have been
More informationMid-Infrared (MIR) Light-Emitting Diode
Power, µw Current, ma Sensitivity, a.u. Sensitivity, a.u. Mid-Infrared (MIR) Light-Emitting Diode Lms34LEDhp series Device parameters Symbol Value Units Storage temperature Operating temperature Soldering
More informationDCS laser for Thomson scattering diagnostic applications
DCS laser for Thomson scattering diagnostic applications Authors Jason Zweiback 10/6/2015 jzweiback@logostech.net 1 Summary Motivation DCS laser Laser for Thomson scattering diagnostics 2 What is the Dynamic
More informationSTATE-OF-THE-ART SILICON DETECTORS FOR X-RAY SPECTROSCOPY
Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 47 STATE-OF-THE-ART SILICON DETECTORS FOR X-RAY SPECTROSCOPY P. Lechner* 1, R. Hartmann* 1, P. Holl*
More informationANALYSIS OF PIGMENTS USED IN SCROLL PAINTINGS OF A NATIONAL TREASURE "TALE OF GENJI" USING A PORTABLE X-RAY FLUORESCENCE SPECTROMETER
Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 432 ANALYSIS OF PIGMENTS USED IN SCROLL PAINTINGS OF A NATIONAL TREASURE "TALE OF GENJI" USING A PORTABLE X-RAY FLUORESCENCE SPECTROMETER
More informationSiC Power Schottky Diodes in Power Factor Correction Circuits
SiC Power Schottky Diodes in Power Factor Correction Circuits By Ranbir Singh and James Richmond Introduction Electronic systems operating in the -12 V range currently utilize silicon (Si) PiN diodes,
More informationHYGROCHIP 1/6 DIGITAL HUMIDITY SENSOR HYT-271
HYGROCHIP 1/6 Characteristic Features Measuring range 0 100% rh, -40 125 C Low drift Stable at high humidity I 2 C protocol for humidity and temperature (address 0x28 or alternative address) Accuracy ±1.8%
More informationCoherent Laser Measurement and Control Beam Diagnostics
Coherent Laser Measurement and Control M 2 Propagation Analyzer Measurement and display of CW laser divergence, M 2 (or k) and astigmatism sizes 0.2 mm to 25 mm Wavelengths from 220 nm to 15 µm Determination
More informationElectric Circuit I Lab Manual Session # 2
Electric Circuit I Lab Manual Session # 2 Name: ----------- Group: -------------- 1 Breadboard and Wiring Objective: The objective of this experiment is to be familiar with breadboard and connection made
More informationON THE DETECTION LIMIT OF TEY (TOTAL ELECTRON YIELD) Maria F. Ebel, Horst Ebel and Robert Svagera
Copyright(C)JCPDS-International Centre for Diffraction Data 2, Advances in X-ray Analysis, Vol.42 91 Copyright(C)JCPDS-International Centre for Diffraction Data 2, Advances in X-ray Analysis, Vol.42 91
More informationC SERIES MINIATURE, REGULATED HIGH VOLTAGE DC TO DC CONVERTERS
Proven Reliability C SERIES MINIATURE, REGULATED HIGH DC TO DC CONVERTERS V to 8kV at W PRODUCT SELECTION TABLE NOW UL RECOGNIZED PRODUCT DESCRIPTION Miniature, well-regulated high voltage power supplies
More informationElectron-Bombarded CMOS
New Megapixel Single Photon Position Sensitive HPD: Electron-Bombarded CMOS University of Lyon / CNRS-IN2P3 in collaboration with J. Baudot, E. Chabanat, P. Depasse, W. Dulinski, N. Estre, M. Winter N56:
More informationLow-energy Electron Diffractive Imaging for Three dimensional Light-element Materials
Low-energy Electron Diffractive Imaging for Three dimensional Light-element Materials Hitachi Review Vol. 61 (2012), No. 6 269 Osamu Kamimura, Ph. D. Takashi Dobashi OVERVIEW: Hitachi has been developing
More information3D light microscopy techniques
3D light microscopy techniques The image of a point is a 3D feature In-focus image Out-of-focus image The image of a point is not a point Point Spread Function (PSF) 1D imaging 1 1 2! NA = 0.5! NA 2D imaging
More informationDavid P. Eartly, Robert H. Lee*, Fermi National Laboratory, Batavia, IL 60510, USA. Abstract
Available on CMS information server CMS NOTE 1998/004 January 26,1998 CMS Endcap Muon System Long Term Resolution Tests of Max Planck Institute Transparent Amorphous Silicon Optical Beam Position Sensors
More informationTesting Aspheric Lenses: New Approaches
Nasrin Ghanbari OPTI 521 - Synopsis of a published Paper November 5, 2012 Testing Aspheric Lenses: New Approaches by W. Osten, B. D orband, E. Garbusi, Ch. Pruss, and L. Seifert Published in 2010 Introduction
More informationAtomic and nuclear physics
Atomic and nuclear physics X-ray physics Physics of the atomic shell LEYBOLD Physics Leaflets Investigating the energy spectrum of an x-ray tube as a function of the high voltage and the emission current
More informationCHARACTERIZATION OF A PORTABLE X-RAY DEVICE FOR RESIDUAL STRESS MEASUREMENTS
CHARACTERIZATION OF A PORTABLE X-RAY DEVICE FOR RESIDUAL STRESS MEASUREMENTS 153 Jingjing Ling and Seung-Yub Lee Applied Physics and Applied Mathematics, Columbia University, New York, NY 10027 ABSTRACT
More informationThe Development of Miniature Electron Multipliers for Use In Portable Mass Spectrometers
The Development of Miniature Electron Multipliers for Use In Portable Mass Spectrometers *Bruce N. Laprade, Lenny Erickson William G. Dunn and Reginald Farr BURLE Electro-Optics Sturbridge MA Paper 10400-2100
More informationUse of Back Scattered Ionizing Radiation for Measurement of Thickness of the Catalytic Agent Active Material
18th World Conference on Nondestructive Testing, 16- April 1, Durban, South Africa Use of Back Scattered Ionizing Radiation for Measurement of Thickness of the Catalytic Agent Active Material Boris V.
More informationCALIBRATION OF X-RAY IMAGING DEVICES FOR ACCURATE INTENSITY MEASUREMENT
DOE/NV/25946--1306 23 CALIBRATION OF X-RAY IMAGING DEVICES FOR ACCURATE INTENSITY MEASUREMENT Michael J. Haugh*, Michael Charest*, Patrick Ross*, Joshua Lee*, Marilyn Schneider, Nathan Palmer, and Alan
More informationFabry Perot Resonator (CA-1140)
Fabry Perot Resonator (CA-1140) The open frame Fabry Perot kit CA-1140 was designed for demonstration and investigation of characteristics like resonance, free spectral range and finesse of a resonator.
More informationGas scintillation Glass GEM detector for high-resolution X-ray imaging and CT
Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Takeshi Fujiwara 1, Yuki Mitsuya 2, Hiroyuki Takahashi 2, and Hiroyuki Toyokawa 2 1 National Institute of Advanced Industrial
More informationMB Scientific AB Extremely High Energy Resolution Photoelectron Analyser
December 2014 Product catalogue MB Scientific AB Extremely High Energy Resolution Photoelectron Analyser Model : MBS A-1 and Lens4 A-1 with Lens4 Outlook photo* *System might deviate from the picture depending
More informationElectron Linacs for Cargo Inspection and Other Industrial Applications
Electron Linacs for Cargo Inspection and Other Industrial Applications Chuanxiang Tang 1, Huaibi Chen 1,Yaohong Liu 2 Tang.xuh@tsinghua.edu.cn 1 Department of Engineering Physics, Tsinghua U., Beijing
More informationTECHNICAL SPECIFICATIONS for 300KV CONSTANT POTENTIAL X RAY EQUIPMENT (for industrial applications) S. No. PARTICULARS BHEL SPECIFICATIONS
TECHNICAL SPECIFICATIONS for 300KV CONSTANT POTENTIAL X RAY EQUIPMENT (for industrial applications) BIDDER s OFFER S. No. PARTICULARS BHEL SPECIFICATIONS 1.0 Area of Application The 300kV Constant potential
More informationC SERIES MINIATURE, REGULATED HIGH VOLTAGE DC TO DC CONVERTERS
Proven Reliability C SERIES MINIATURE, REGULATED HIGH DC TO DC CONVERTERS V to 8kV at W PRODUCT SELECTION TABLE NOW UL RECOGNIZED PRODUCT DESCRIPTION Miniature, well-regulated high voltage power supplies
More informationToday s wireless. Best Practices for Making Accurate WiMAX Channel- Power Measurements. WiMAX MEASUREMENTS. fundamental information
From August 2008 High Frequency Electronics Copyright Summit Technical Media, LLC Best Practices for Making Accurate WiMAX Channel- Power Measurements By David Huynh and Bob Nelson Agilent Technologies
More information