USING A CHARGE-COUPLED DEVICE (CCD) TO GATHER X-RAY FLUORESCENCE (XRF)AND X-RAY DIFFRACTION (XRD) INFORMATION SIMULTANEOUSLY

Size: px
Start display at page:

Download "USING A CHARGE-COUPLED DEVICE (CCD) TO GATHER X-RAY FLUORESCENCE (XRF)AND X-RAY DIFFRACTION (XRD) INFORMATION SIMULTANEOUSLY"

Transcription

1 Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol USING A CHARGE-COUPLED DEVICE (CCD) TO GATHER X-RAY FLUORESCENCE (XRF)AND X-RAY DIFFRACTION (XRD) INFORMATION SIMULTANEOUSLY A. Reyes-Mena 1, S. Cornaby 2, H. K. Pew 1, P. W. Moody 1, T. Hughes 2, A. Stradling 2, and L. V. Knight 2 1 MOXTEK Inc., 452 West 1260 North, Orem, UT Department of Physics and Astronomy, Brigham Young University, Provo, UT ABSTRACT A breadboard setup constructed at MOXTEK, Inc. is capable of capturing both XRF and XRD information using a charge-coupled device (CCD) as the x-ray detector. This preliminary setup will lead to a prototype simultaneous XRD/XRF instrument. NASA is funding the instrument s construction because of its capabilities and small size; it could be used for future Mars missions for rock analysis. The instrument uses a CCD to capture both the energy information of an incoming x-ray and the position where the x-ray hit. A powdered sample of material is placed in front of the CCD and then a collimated x-ray beam bombards the sample. The energy information is used to accumulate the x-ray events caused by diffraction, thus enabling us to construct x-ray diffraction patterns. We are also able to discard split events and use all the single x-ray events to create an x-ray fluorescence spectrum from the data. The focus of this paper is how a front-sideilluminated (FSI) CCD and back-side-illuminated (BSI) CCD gather information and how this information is sorted. A split event comparison is made between the two types of CCDs. INTRODUCTION A first prototype XRF/XRD instrument was developed at Ames National Laboratory and is currently in use at Los Alamos National Laboratory [1,2]. The intended applications for this instrument are planetary exploration and a portable instrument for terrestrial use. NASA funding for the instrument s construction is based on its possible use in future Mars missions. Mars missions have previously only used methods that give the elemental composition, which are subject to sizable uncertainties for mineral identification. XRF gives the elemental composition and suggests certain minerals, but the traditional well-tested method of definitive mineral identification is by XRD [3]. With a small, compact XRF/XRD instrument, both the diffraction and the fluorescence information can be practically gathered on Mars as well as on Earth. The purpose of the breadboard setup constructed at MOXTEK is to incorporate and test some of the various key components, including some of MOXTEK s products, involved in an XRF/XRD instrument design. The information gained from this breadboard will lead to a prototype instrument which will be small and portable when a final design is completed.

2 This document was presented at the Denver X-ray Conference (DXC) on Applications of X-ray Analysis. Sponsored by the International Centre for Diffraction Data (ICDD). This document is provided by ICDD in cooperation with the authors and presenters of the DXC for the express purpose of educating the scientific community. All copyrights for the document are retained by ICDD. Usage is restricted for the purposes of education and scientific research. DXC Website ICDD Website -

3 Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol EXPERIMENTAL SETUP The breadboard setup for simultaneously capturing both XRF and XRD consists of a commercial CCD camera, a high brilliance Rotaflex Rigaku rotating copper anode x-ray generator (model RU-200), a beam collimating system, a shutter and a sample support (Fig. 1). Both a Princeton Instruments FSI CCD (which consists of an EEV 576 x 384 CCD with 22 x 22-µm pixels) and a BSI CCD (which consists of an EEV 512 x 512 CCD with 25 x 25-µm pixels) were used in the breadboard instrument to compare split events. The x-ray source is run at 30 kv and 30 ma to generate x rays. To capture the XRD information the x rays are collimated with a 100-µm tungsten and a 300-µm lead pinholes. The collimated x-ray beam irradiates the sample, and the CCD camera is used to collect the scattered x rays. The sample is mounted outside the CCD camera. A vacuum environment is maintained inside the camera with the use of an x-ray transparent window mounted on the camera flange. MOXTEK s DuraBeryllium TM and UltraThin TM polymer AP1.3 windows are used in the test setup (Fig. 2) [4]. The information is collected and sorted using algorithms. X-RAY DETECTION USING A CCD Figure 1. Components of the XRF/XRD breadboard instrument. Figure 2. Transparency of different x-ray windows. There is an ideal situation for obtaining both the XRF and the XRD information simultaneously. This situation is to know both the energy of each incoming x ray and the location where each x ray hits. When a sample is exposed to a collimated x-ray beam, both XRF and XRD can be observed. By knowing the energy of the incoming x-rays, the XRD and XRF events can be distinguished. By creating a histogram of all the energies of x-rays events, a fluorescent spectra is observed. By filtering out only the Cu K events and displaying the location of these events spatially, a diffraction image can be observed. A CCD is a detector that has these capabilities. It can be used as an energy-dispersive detector that will give the energy of an x ray and the location where an x ray hit. Each pixel on the CCD can be used as an individual energy-dispersive detector. Each pixel also has a different location in space allowing spatial information to be captured.

4 Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol Each pixel on a CCD can be used as an energy dispersive detector because in the soft x-ray range the generation of electron-hole pairs is dominated by the direct band gap. The direct band gap in silicon has an energy gap of 3.65 ev [5]. In the x-ray range the number of electron-hole pairs created in silicon when an x ray is absorbed is proportional to the energy of the x ray. In order to use each pixel as an energy dispersive detector, the pixels on the CCD must work in single-event count mode. In order to acquire the energy information from an x ray, only one x ray must be captured in a pixel. If more than one x ray is deposited in a pixel, the pixel will contain the energy information from both of the events and there is no way to separate the energy. As long as one x-ray event is captured in one pixel, then the CCD can be used as an energy dispersive detector. split events single events Figure 3. Single and split events. (a) total events (b) single events (c) split events, two or more Not all of the x-ray events that hit the CCD are collected into a single pixel. There are a number (d) split events, three or more of events that are collected into two or more adjacent pixels; these events are called split events (Fig. 3). The split events do not give the Figure 4. Number of counts per exposure. energy information needed to know the energy of the x ray and are therefore eliminated from the data set. In order to get the XRF and the XRD information, we keep only the single events. A single event is a pixel with a count of electrons above the noise level and has all of the eight neighboring pixels with an electron count at the noise level. Also, in order to get a large number of both the XRF and XRD events, multiple exposures need to be taken by the CCD. The number of x-ray events that we typically capture per exposure for the FSI CCD is shown in Figure 4. Figure 4(a) shows the total number of events that hit the CCD per exposure; 4(b) represents the total number of single events; 4(c) shows the total number of events split between two or more pixels; 4(d) shows the number of events split between three or more pixels. With other samples taken there have been between 400 to 700 single events captured on a single CCD exposure. The CCD exposure time presently runs between five to eight seconds. To get approximately 300,000 events (enough events to start to view Cu K diffraction rings), 400 to 600 exposures need to be taken. Minimizing the number of split events is crucial in making the CCD an effective energy-dispersive detector.

5 Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol Figure 5. XRF spectrum collected with the FSI CCD. Figure 6. XRF spectrum collected with the BSI CCD. CCD COMPARISON In order to determine the optimum CCD detector for gathering the XRF/XRD information for the instrument, we compared the performance of both the FSI CCD and the BSI CCD. We found that the FSI CCD had a fewer number of split events then the BSI CCD. Figure 5 shows a typical energy spectrum collected with the FSI CCD with 277-five second exposures while doing simultaneous diffraction. The upper function in this graph represents the energy spectrum with both the single and the split events present. The lower function shows the spectrum with only the single events remaining. Eliminating the split events reduced the background noise in the energy spectrum. Figure 6 shows a typical energy spectrum collected by a BSI CCD in a single fivesecond exposure. The upper function shows the spectrum with all the events included and the lower function includes only single events. It is clear that the FSI CCD has fewer split events than the BSI CCD. Since the FSI captures more single events per exposure than the BSI CCD, fewer exposures are needed to capture the same number of single events with the FSI CCD. For this reason we have chosen to use the FSI CCD for the XRF/XRD instrument. RESULTS AND DISCUSSION The energy information for each x ray is captured, along with the location where the event happened. Using this energy information as a filter for x-ray events, we can construct the diffraction pattern from a sample. Since all the x-ray events are captured, this allows the acquisition of multiple diffraction patterns. Figures 7 and 8 show diffraction patterns collected simultaneously from a KCl sample. Figure 7 is constructed from the K events, and Figure 8 is constructed from the K events. The nondiffracted beam that irradiates the sample is blocked with a beam stop to keep the CCD from being damaged. The shadow of the beam stop is the vertical line where there are few x-ray events (Figs. 7 and 8). Once these images are created they can be radially integrated to give the diffraction intensities as a function of the 2 angle. Figure 9 shows the intensity versus 2 plot for the K events, and Figure 10 shows the plot for the K events. Figure 11 shows the XRF for the KCl sample. The camera was fitted with an AP1.3 window when data for this sample were taken. This causes a Si K peak because of the silicon

6 Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol rib structure of the x-ray window on the CCD. This rib structure also affects the XRD images and can be seen in the spotty change of intensity in the diffraction rings (Figs. 7 and 8). If the rib structure was not present, then each diffraction ring would have a uniform intensity. Data collected with a beryllium window do not have a Si K peak present and the diffraction rings have uniform intensities. With the breadboard XRF/XRD instrument, all the information in Figures 7-11 was gathered simultaneously. Figure 7. Cu K events for KCl sample. Figure 8. Cu K events for KCl sample. Figure 9. Cu K events in a 2 plot for a KCl sample. Figure 10. Cu K events in a 2 plot for a KCl sample. CONCLUSIONS Both the XRF and XRD information can be captured simultaneously using a CCD as the detector. A CCD can be used to gather both the diffraction information and the fluorescence information when the CCD is able to capture single events within a single pixel. Also, in comparing FSI and BSI CCDs it was found that the FSI CCD had fewer split events than the BSI CCD. This allows the FSI CCD to capture more single Si K Cl K K K Cu K Cu K Figure 11. XRF for the KCl sample.

7 Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol events for a given exposure time. The breadboard instrument has shown adequately that the construction of a small, portable XRF/XRD instrument is feasible for terrestrial applications and possibly for planetary missions to Mars. ACKNOWLEDGMENTS This work is being supported by NASA under SBIR contract number NAS We thank Brigham Young University for student partial-financial support and for use of equipment. REFERENCES [1] Blake, D.F. et al., Proceedings of the LPSC XXIII, 1992, [2] Blake, D.F. et al. X-ray Diffraction Apparatus, 1993, U.S. Patent No. 5,491,738. [3] Vaniman, D.; Bish, D.; Blake, D.F.; Elliott, S.T.; Sarrazin, P.; Collins, S.A.; Chipera, S., J. Geophys. Res., 1998, 103(E13), 31,477-31,489. [4] MOXTEK, Inc., UltraThin Soft X-ray Window Data Sheet, [5] Sweedler, J.V.; Ratzlaff, K.L; Denton, M.B.; eds. Charge-Transfer Devices in Spectroscopy, VCH Publishers Inc.: New York, 1994, 12.

SIMULTANEOUS XRD/XRF WITH LOW-POWER X-RAY TUBES

SIMULTANEOUS XRD/XRF WITH LOW-POWER X-RAY TUBES Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 34 SIMULTANEOUS XRD/XRF WITH LOW-POWER X-RAY TUBES S. Cornaby 1, A. Reyes-Mena 1, P. W. Moody 1,

More information

MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE

MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE 228 MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE D. CARUSO, M. DINSMORE TWX LLC, CONCORD, MA 01742 S. CORNABY MOXTEK, OREM, UT 84057 ABSTRACT Miniature x-ray sources present

More information

ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER

ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 325 ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER ABSTRACT William Chang, Jonathan Kerner, and Edward

More information

MOXTEK S NEW ULTRA-LITE X-RAY SOURCES: PERFORMACE CHARACTERIZATIONS

MOXTEK S NEW ULTRA-LITE X-RAY SOURCES: PERFORMACE CHARACTERIZATIONS Copyright JCPDS-International Centre for Diffraction Data 2013 ISSN 1097-0002 202 MOXTEK S NEW ULTRA-LITE X-RAY SOURCES: PERFORMACE CHARACTERIZATIONS S. Cornaby, S. Morris, J. Smith, D. Reynolds, K. Kozaczek

More information

MICRO XRF OF LIGHT ELEMENTS USING A POLYCAPILLARY LENS AND AN ULTRA THIN WINDOW SILICON DRIFT DETECTOR INSIDE A VACUUM CHAMBER

MICRO XRF OF LIGHT ELEMENTS USING A POLYCAPILLARY LENS AND AN ULTRA THIN WINDOW SILICON DRIFT DETECTOR INSIDE A VACUUM CHAMBER Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 229 MICRO XRF OF LIGHT ELEMENTS USING A POLYCAPILLARY LENS AND AN ULTRA THIN WINDOW SILICON DRIFT

More information

WIDE ANGLE GEOMETRY EDXRF SPECTROMETERS WITH SECONDARY TARGET AND DIRECT EXCITATION MODES

WIDE ANGLE GEOMETRY EDXRF SPECTROMETERS WITH SECONDARY TARGET AND DIRECT EXCITATION MODES Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 11 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42

More information

X-RAY OPTICS FOR TWO-DIMENSIONAL DIFFRACTION

X-RAY OPTICS FOR TWO-DIMENSIONAL DIFFRACTION Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 332 ABSTRACT X-RAY OPTICS FOR TWO-DIMENSIONAL DIFFRACTION Bob B. He and Uwe Preckwinkel Bruker

More information

DEVELOPMENT OF A WAVELENGTH DISPERSIVE X-RAY FLUORESCENCE SPECTROMETER USING A MULTI-CAPILLARY X-RAY LENS FOR X-RAY DETECTION

DEVELOPMENT OF A WAVELENGTH DISPERSIVE X-RAY FLUORESCENCE SPECTROMETER USING A MULTI-CAPILLARY X-RAY LENS FOR X-RAY DETECTION Copyright JCPDS - International Centre for Diffraction Data 2003, Advances in X-ray Analysis, Volume 46. 346 DEVELOPMENT OF A WAVELENGTH DISPERSIVE X-RAY FLUORESCENCE SPECTROMETER USING A MULTI-CAPILLARY

More information

Copyright -International Centre for Diffraction Data 2010 ISSN

Copyright -International Centre for Diffraction Data 2010 ISSN 234 BRIDGING THE PRICE/PERFORMANCE GAP BETWEEN SILICON DRIFT AND SILICON PIN DIODE DETECTORS Derek Hullinger, Keith Decker, Jerry Smith, Chris Carter Moxtek, Inc. ABSTRACT Use of silicon drift detectors

More information

Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror

Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror Stephen B. Robie scintag, Inc. 10040 Bubb Road Cupertino, CA 95014 Abstract Corundum

More information

LONG TERM STATISTICS OF X-RAY SPECTROMETERS

LONG TERM STATISTICS OF X-RAY SPECTROMETERS 403 LONG TERM STATISTICS OF X-RAY SPECTROMETERS J. F. Dlouhy*, D. Mathieu Department of the Environment, Environmental Technology Center, River Road, Ottawa, Ontario, Canada Kl A OH3 K. N. Stoev Bulgarian

More information

GUNSHOT RESIDUE INVESTIGATIONS USING TXRF

GUNSHOT RESIDUE INVESTIGATIONS USING TXRF 299 GUNSHOT RESIDUE INVESTIGATIONS USING TXRF Alexander Wastl 1, Bettina Bogner 2, Peter Kregsamer 1, Peter Wobrauschek 1, Christina Streli 1 1 Atominstitut, Vienna University of Technology, Vienna, Austria

More information

X-Ray Spectroscopy with a CCD Detector. Application Note

X-Ray Spectroscopy with a CCD Detector. Application Note X-Ray Spectroscopy with a CCD Detector In addition to providing X-ray imaging solutions, including CCD-based cameras that image X-rays using either direct detection (0.5-20 kev) or indirectly using a scintillation

More information

DOUBLE MULTILAYER MONOCHROMATOR WITH FIXED EXIT GEOMETRY. H.Gatterbauer, P.Wobrauschek, F.Hegediis, P.Biini, C.Streli

DOUBLE MULTILAYER MONOCHROMATOR WITH FIXED EXIT GEOMETRY. H.Gatterbauer, P.Wobrauschek, F.Hegediis, P.Biini, C.Streli Copyright (C) JCPDS International Centre for Diffraction Data 1999 379 DOUBLE MULTILAYER MONOCHROMATOR WITH FIXED EXIT GEOMETRY H.Gatterbauer, P.Wobrauschek, F.Hegediis, P.Biini, C.Streli Atominsitut der

More information

CARBON-NANOTUBE FIELD EMISSION X-RAY TUBE FOR SPACE EXPLORATION XRD/XRF INSTRUMENT.

CARBON-NANOTUBE FIELD EMISSION X-RAY TUBE FOR SPACE EXPLORATION XRD/XRF INSTRUMENT. Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 232 CARBON-NANOTUBE FIELD EMISSION X-RAY TUBE FOR SPACE EXPLORATION XRD/XRF INSTRUMENT. P. Sarrazin

More information

DESIGN AND MEASUREMENT WITH A NEW PORTABLE X-RAY CAMERA FOR FULL-FIELD FLUORESCENCE IMAGING

DESIGN AND MEASUREMENT WITH A NEW PORTABLE X-RAY CAMERA FOR FULL-FIELD FLUORESCENCE IMAGING 14 DESIGN AND MEASUREMENT WITH A NEW PORTABLE X-RAY CAMERA FOR FULL-FIELD FLUORESCENCE IMAGING I. Ordavo 1,2, A. Bjeoumikhov 3, S. Bjeoumikhova 3, G. Buzanich 4, R. Gubzhokov 4, R. Hartmann 1, S. Ihle

More information

A PORTABLE NON-INVASIVE XRD-XRF INSTRUMENT FOR THE STUDY OF ART OBJECTS.

A PORTABLE NON-INVASIVE XRD-XRF INSTRUMENT FOR THE STUDY OF ART OBJECTS. 175 A PORTABLE NON-INVASIVE XRD-XRF INSTRUMENT FOR THE STUDY OF ART OBJECTS. P. Sarrazin 1, G. Chiari 2, M. Gailhanou 3 1 inxitu, Inc. 2551 Casey ave. Suite A, Mountain View, CA 94043, USA 2 Getty Conservation

More information

A PORTABLE X-RAY APPARATUS FOR BOTH STRESS MEASUREMENT AND PHASE ANALYSIS UNDER FIELD CONDITIONS.

A PORTABLE X-RAY APPARATUS FOR BOTH STRESS MEASUREMENT AND PHASE ANALYSIS UNDER FIELD CONDITIONS. Copyright(c)JCPDS-International Centre for Diffraction Data 2000,Advances in X-ray Analysis,Vol.43 66 A PORTABLE X-RAY APPARATUS FOR BOTH STRESS MEASUREMENT AND PHASE ANALYSIS UNDER FIELD CONDITIONS. V.

More information

X-RAY BACKSCATTER IMAGING: PHOTOGRAPHY THROUGH BARRIERS

X-RAY BACKSCATTER IMAGING: PHOTOGRAPHY THROUGH BARRIERS Copyright JCPDS-International Centre for Diffraction Data 2006 ISSN 1097-0002 X-RAY BACKSCATTER IMAGING: PHOTOGRAPHY THROUGH BARRIERS 13 Joseph Callerame American Science & Engineering, Inc. 829 Middlesex

More information

Spectral distribution from end window X-ray tubes

Spectral distribution from end window X-ray tubes Copyright ISSN (C) 1097-0002, JCPDS-International Advances in X-ray Centre Analysis, for Volume Diffraction 41 Data 1999 393 Spectral distribution from end window X-ray tubes N. Broll 1, P. de Chateaubourg

More information

A MONTE CARLO CODE FOR SIMULATION OF PULSE PILE-UP SPECTRAL DISTORTION IN PULSE-HEIGHT MEASUREMENT

A MONTE CARLO CODE FOR SIMULATION OF PULSE PILE-UP SPECTRAL DISTORTION IN PULSE-HEIGHT MEASUREMENT Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 246 A MONTE CARLO CODE FOR SIMULATION OF PULSE PILE-UP SPECTRAL DISTORTION IN PULSE-HEIGHT MEASUREMENT

More information

Improving the Collection Efficiency of Raman Scattering

Improving the Collection Efficiency of Raman Scattering PERFORMANCE Unparalleled signal-to-noise ratio with diffraction-limited spectral and imaging resolution Deep-cooled CCD with excelon sensor technology Aberration-free optical design for uniform high resolution

More information

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region Feature Article JY Division I nformation Optical Spectroscopy Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region Raymond Pini, Salvatore Atzeni Abstract Multichannel

More information

MICROANALYSIS WITH A POLYCAPILLARY IN A VACUUM CHAMBER

MICROANALYSIS WITH A POLYCAPILLARY IN A VACUUM CHAMBER THE RIGAKU JOURNAL VOL. 20 / NO. 2 / 2003 MICROANALYSIS WITH A POLYCAPILLARY IN A VACUUM CHAMBER CHRISTINA STRELI a), NATALIA MAROSI, PETER WOBRAUSCHEK AND BARBARA FRANK Atominstitut der Österreichischen

More information

Development of X-ray Tool For Critical- Dimension Metrology

Development of X-ray Tool For Critical- Dimension Metrology Development of X-ray Tool For Critical- Dimension Metrology Boris Yokhin, Alexander Krokhmal, Alexander Dikopoltsev, David Berman, Isaac Mazor Jordan Valley Semiconductors Ltd., Ramat Gabriel Ind. Zone,

More information

Lab Report XRF 441 Elemental distribution analysis on geological samples with the M4 TORNADO

Lab Report XRF 441 Elemental distribution analysis on geological samples with the M4 TORNADO Bruker Nano Spectrum Geological sample M4 TORNADO Quantification Lab Report XRF 441 Elemental distribution analysis on geological samples with the M4 TORNADO Geological samples are inhomogeneous. The distribution

More information

FIRST Newsletter March 2013, Issue 20. Elemental Distribution Analysis of a Meteorite Sample from the Rochechouart Structure with the µ-xrf M4 TORNADO

FIRST Newsletter March 2013, Issue 20. Elemental Distribution Analysis of a Meteorite Sample from the Rochechouart Structure with the µ-xrf M4 TORNADO FIRST Newsletter March 2013, Issue 20 Elemental Distribution Analysis of a Meteorite Sample from the Rochechouart Structure with the µ-xrf M4 TORNADO By Dr. Roald Tagle, Ulrich Waldschlager, Dr. Michael

More information

RIETVELD REFINEMENT OF POWDER DATA FROM MULTILAYER OPTICS

RIETVELD REFINEMENT OF POWDER DATA FROM MULTILAYER OPTICS Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 166 RIETVELD REFINEMENT OF POWDER DATA FROM MULTILAYER OPTICS ABSTRACT Scott T. Misture NYS College

More information

Interpixel crosstalk in a 3D-integrated active pixel sensor for x-ray detection

Interpixel crosstalk in a 3D-integrated active pixel sensor for x-ray detection Interpixel crosstalk in a 3D-integrated active pixel sensor for x-ray detection The MIT Faculty has made this article openly available. Please share how this access benefits you. Your story matters. Citation

More information

Advancing EDS Analysis in the SEM Quantitative XRF. International Microscopy Congress, September 5 th, Outline

Advancing EDS Analysis in the SEM Quantitative XRF. International Microscopy Congress, September 5 th, Outline Advancing EDS Analysis in the SEM with in-situ Quantitative XRF Brian J. Cross (1) & Kenny C. Witherspoon (2) 1) CrossRoads Scientific, El Granada, CA 94018, USA 2) ixrf Systems, Inc., Houston, TX 77059,

More information

Lesson 2 Diffractometers

Lesson 2 Diffractometers Lesson 2 Diffractometers Nicola Döbelin RMS Foundation, Bettlach, Switzerland January 14 16, 2015, Bern, Switzerland Repetition: Generation of X-rays / Diffraction SEM: BSE detector, BSED / SAED detector

More information

Applications of Micro XRF for the Analysis of Traditional Japanese "Ainu" Glass Beads and other Artifacts

Applications of Micro XRF for the Analysis of Traditional Japanese Ainu Glass Beads and other Artifacts 161 161 Applications of Micro XRF for the Analysis of Traditional Japanese "Ainu" Glass Beads and other Artifacts K.Sugihara 1, M.Satoh 1, Y.Hayakawa 2, A.Saito 3 and T.Sasaki 4 1 Seiko Instruments Inc.,

More information

Renishaw InVia Raman microscope

Renishaw InVia Raman microscope Laser Spectroscopy Labs Renishaw InVia Raman microscope Operation instructions 1. Turn On the power switch, system power switch is located towards the back of the system on the right hand side. Wait ~10

More information

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering RR Oct 2001 SS Dec 2001 MJ Oct 2009 Physics 34000 Laboratory Scattering of Photons from Electrons: Compton Scattering Objective: To measure the energy of high energy photons scattered from electrons in

More information

IDENTIFICATION OF PAINTING MATERIALS USED FOR MURAL PAINTINGS BY IMAGE ANALYSIS AND XRF

IDENTIFICATION OF PAINTING MATERIALS USED FOR MURAL PAINTINGS BY IMAGE ANALYSIS AND XRF IDENTIFICATION OF PAINTING MATERIALS USED FOR MURAL PAINTINGS BY IMAGE ANALYSIS AND XRF 213 Seiji SHIRONO, Yasuhiro HAYAKAWA National Research Institute for Cultural Properties, Tokyo, Japan ABSTRACT The

More information

Author s Accepted Manuscript

Author s Accepted Manuscript Author s Accepted Manuscript The X-ray quantum efficiency measurement of high resistivity CCDs Neil J. Murray, Andrew D. Holland, David R. Smith, Jason P. Gow, Peter J. Pool, David J. Burt PII: S0168-9002(09)00147-8

More information

Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4

Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4 Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4 S.V. Roth, R. Döhrmann, M. Dommach, I. Kröger, T. Schubert, R. Gehrke Definition of the upgrade The wiggler beamline BW4 is dedicated to

More information

HR2000+ Spectrometer. User-Configured for Flexibility. now with. Spectrometers

HR2000+ Spectrometer. User-Configured for Flexibility. now with. Spectrometers Spectrometers HR2000+ Spectrometer User-Configured for Flexibility HR2000+ One of our most popular items, the HR2000+ Spectrometer features a high-resolution optical bench, a powerful 2-MHz analog-to-digital

More information

M4 TORNADO PLUS. Innovation with Integrity. Super Light Element Micro-XRF Spectrometer. Micro-XRF

M4 TORNADO PLUS. Innovation with Integrity. Super Light Element Micro-XRF Spectrometer. Micro-XRF M4 TORNADO PLUS Super Light Element Micro-XRF Spectrometer Innovation with Integrity Micro-XRF M4 TORNADO PLUS - A New Era in Micro-XRF M4 TORNADO PLUS is the world's first Micro-XRF spectrometer that

More information

ABSTRACT. Supported by U.S. DoE grant No. DE-FG02-96ER54375

ABSTRACT. Supported by U.S. DoE grant No. DE-FG02-96ER54375 ABSTRACT A CCD imaging system is currently being developed for T e (,t) and bolometric measurements on the Pegasus Toroidal Experiment. Soft X-rays (E

More information

ikon-m 934 DO Series

ikon-m 934 DO Series X-Ray diffraction pattern of the Synchrotron Soleil,France. Courtesy of Vincent Jacques Low Light Imaging X-Ray Cameras Features and Benefits Open front end 1 CF152 flange and knife-edge sealing provided

More information

TOWARDS FAST RECIPROCAL SPACE MAPPING

TOWARDS FAST RECIPROCAL SPACE MAPPING Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 165 ABSTRACT TOWARDS FAST RECIPROCAL SPACE MAPPING J.F. Woitok and A. Kharchenko PANalytical B.V.,

More information

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION Revised November 15, 2017 INTRODUCTION The simplest and most commonly described examples of diffraction and interference from two-dimensional apertures

More information

ON THE ENERGY DEPENDENCE OF THE DETECTOR EFFICIENCY OF A Si-PIN DIODE

ON THE ENERGY DEPENDENCE OF THE DETECTOR EFFICIENCY OF A Si-PIN DIODE Copyright(C)JCPDS-International Centre for Diffraction Data 2, Advances in X-ray Analysis, Vol.42 36 Copyright(C)JCPDS-International Centre for Diffraction Data 2, Advances in X-ray Analysis, Vol.42 36

More information

DESIGN OF EDXRF EQUIPMENT FOR THE NONDESTRUCTIVE STUDY OF PRINTS

DESIGN OF EDXRF EQUIPMENT FOR THE NONDESTRUCTIVE STUDY OF PRINTS Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 70 DESIGN OF EDXRF EQUIPMENT FOR THE NONDESTRUCTIVE STUDY OF PRINTS 1,3 M. Ardid, 1 J.L. Ferrero,

More information

ARTAX. Innovation with Integrity. Portable Micro-XRF Spectrometer. Micro-XRF

ARTAX. Innovation with Integrity. Portable Micro-XRF Spectrometer. Micro-XRF ARTAX Portable Micro-XRF Spectrometer Innovation with Integrity Micro-XRF ARTAX Elemental Analysis for the Art Community and More The ARTAX is the first portable X ray fluorescence (XRF) spectrometer designed

More information

FAST ELEMENTAL MAPPING WITH MICRO-XRF

FAST ELEMENTAL MAPPING WITH MICRO-XRF 286 FAST ELEMENTAL MAPPING WITH MICRO-XRF Haschke, M.; Rossek, U.; Tagle, R.; Waldschläger, U. Bruker Nano GmbH, 12489 Berlin, Schwarzschildstr.12 ABSTRACT X-Ray optics are now in common use for concentrating

More information

RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS

RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS S-94,316 PATENTS-US-A96698 BEAM POSITION MONITOR RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS CONTRACTUAL ORIGIN OF THE INVENTION The United States Government has rights in this invention pursuant

More information

CALIBRATION OF X-RAY IMAGING DEVICES FOR ACCURATE INTENSITY MEASUREMENT

CALIBRATION OF X-RAY IMAGING DEVICES FOR ACCURATE INTENSITY MEASUREMENT DOE/NV/25946--1306 23 CALIBRATION OF X-RAY IMAGING DEVICES FOR ACCURATE INTENSITY MEASUREMENT Michael J. Haugh*, Michael Charest*, Patrick Ross*, Joshua Lee*, Marilyn Schneider, Nathan Palmer, and Alan

More information

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES The current multiplication mechanism offered by dynodes makes photomultiplier tubes ideal for low-light-level measurement. As explained earlier, there

More information

Pinhole collimator design for nuclear survey system

Pinhole collimator design for nuclear survey system Annals of Nuclear Energy 29 (2002) 2029 2040 www.elsevier.com/locate/anucene Pinhole collimator design for nuclear survey system Wanno Lee*, Gyuseong Cho Department of Nuclear Engineering, Korea Advanced

More information

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and

More information

(12) Patent Application Publication (10) Pub. No.: US 2007/ A1

(12) Patent Application Publication (10) Pub. No.: US 2007/ A1 (19) United States US 2007.00030 12A1 (12) Patent Application Publication (10) Pub. No.: US 2007/0003012 A1 Taguchi et al. (43) Pub. Date: Jan. 4, 2007 (54) X-RAY DIFFRACTION APPARATUS (75) Inventors:

More information

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26 Today s Outline - January 25, 2018 C. Segre (IIT) PHYS 570 - Spring 2018 January 25, 2018 1 / 26 Today s Outline - January 25, 2018 HW #2 C. Segre (IIT) PHYS 570 - Spring 2018 January 25, 2018 1 / 26 Today

More information

Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM

Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM Preamplifiers and amplifiers The current from PMT must be further amplified before it can be processed and counted (the number of electrons yielded

More information

MINIATURE X-RAY TUBES UTILIZING CARBON-NANOTUBE- BASED COLD CATHODES

MINIATURE X-RAY TUBES UTILIZING CARBON-NANOTUBE- BASED COLD CATHODES Copyright JCPDS - International Centre for Diffraction Data 25, Advances in X-ray Analysis, Volume 48. 24 MINIATURE X-RAY TUBES UTILIZING CARBON-NANOTUBE- BASED COLD CATHODES A. Reyes-Mena, Charles Jensen,

More information

TOWARDS SUB-100 NM X-RAY MICROSCOPY FOR TOMOGRAPHIC APPLICATIONS

TOWARDS SUB-100 NM X-RAY MICROSCOPY FOR TOMOGRAPHIC APPLICATIONS Copyright -International Centre for Diffraction Data 2010 ISSN 1097-0002 89 TOWARDS SUB-100 NM X-RAY MICROSCOPY FOR TOMOGRAPHIC APPLICATIONS P. Bruyndonckx, A. Sasov, B. Pauwels Skyscan, Kartuizersweg

More information

Optical Coherence: Recreation of the Experiment of Thompson and Wolf

Optical Coherence: Recreation of the Experiment of Thompson and Wolf Optical Coherence: Recreation of the Experiment of Thompson and Wolf David Collins Senior project Department of Physics, California Polytechnic State University San Luis Obispo June 2010 Abstract The purpose

More information

OPTIMIZING THE ELEMENTAL SENSITIVITY AND FOCAL SPOT SIZE OF A MONOLITHIC POLYCAPILLARY OPTIC USING MICRO-X-RAY FLUORESCENCE

OPTIMIZING THE ELEMENTAL SENSITIVITY AND FOCAL SPOT SIZE OF A MONOLITHIC POLYCAPILLARY OPTIC USING MICRO-X-RAY FLUORESCENCE , Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 26 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42

More information

Water-Window Microscope Based on Nitrogen Plasma Capillary Discharge Source

Water-Window Microscope Based on Nitrogen Plasma Capillary Discharge Source 2015 International Workshop on EUV and Soft X-Ray Sources Water-Window Microscope Based on Nitrogen Plasma Capillary Discharge Source T. Parkman 1, M. F. Nawaz 2, M. Nevrkla 2, M. Vrbova 1, A. Jancarek

More information

STEM Spectrum Imaging Tutorial

STEM Spectrum Imaging Tutorial STEM Spectrum Imaging Tutorial Gatan, Inc. 5933 Coronado Lane, Pleasanton, CA 94588 Tel: (925) 463-0200 Fax: (925) 463-0204 April 2001 Contents 1 Introduction 1.1 What is Spectrum Imaging? 2 Hardware 3

More information

BRUKER ADVANCED X-RAY SOLUTIONS. SPECTROMETRY SOLUTIONS ARTAX mxrf SPECTROMETER

BRUKER ADVANCED X-RAY SOLUTIONS. SPECTROMETRY SOLUTIONS ARTAX mxrf SPECTROMETER BRUKER ADVANCED X-RAY SOLUTIONS SPECTROMETRY SOLUTIONS ARTAX mxrf SPECTROMETER Microanalysis ARTAX Elemental Analysis for the Art Community and More Non-destructive elemental analysis is strictly required

More information

DETECTION OF VISIBLE AND LATENT FINGERPRINTS BY MICRO-X-RAY FLUORESCENCE

DETECTION OF VISIBLE AND LATENT FINGERPRINTS BY MICRO-X-RAY FLUORESCENCE DETECTION OF VISIBLE AND LATENT FINGERPRINTS BY MICRO-X-RAY FLUORESCENCE 363 Christopher G. Worley, Sara S. Wiltshire, Thomasin C. Miller, George J. Havrilla, and Vahid Majidi Los Alamos National Laboratory,

More information

Low Cost Rolled X-ray Prism Lenses to Increase Photon Flux Density in Diffractometry Experiments

Low Cost Rolled X-ray Prism Lenses to Increase Photon Flux Density in Diffractometry Experiments Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN 1097-0002 17 Low Cost Rolled X-ray Prism Lenses to Increase Photon Flux Density in Diffractometry Experiments H. Vogt a, A. Last a, J.

More information

ON THE DETECTION LIMIT OF TEY (TOTAL ELECTRON YIELD) Maria F. Ebel, Horst Ebel and Robert Svagera

ON THE DETECTION LIMIT OF TEY (TOTAL ELECTRON YIELD) Maria F. Ebel, Horst Ebel and Robert Svagera Copyright(C)JCPDS-International Centre for Diffraction Data 2, Advances in X-ray Analysis, Vol.42 91 Copyright(C)JCPDS-International Centre for Diffraction Data 2, Advances in X-ray Analysis, Vol.42 91

More information

NanoSpective, Inc Progress Drive Suite 137 Orlando, Florida

NanoSpective, Inc Progress Drive Suite 137 Orlando, Florida TEM Techniques Summary The TEM is an analytical instrument in which a thin membrane (typically < 100nm) is placed in the path of an energetic and highly coherent beam of electrons. Typical operating voltages

More information

X-Ray diffraction pattern of the Synchrotron Soleil,France. Courtesy of Vincent Jacques. Direct Detection X-Ray Imaging. Specifications Summary

X-Ray diffraction pattern of the Synchrotron Soleil,France. Courtesy of Vincent Jacques. Direct Detection X-Ray Imaging. Specifications Summary X-Ray diffraction pattern of the Synchrotron Soleil,France. Courtesy of Vincent Jacques Low Light Imaging X-Ray Cameras Features and Benefits Open front end 1 CF152 flange and knife-edge sealing provided

More information

Visibility of Detail

Visibility of Detail Visibility of Detail Radiographic Quality Quality radiographic images represents the, and information is for diagnosis. The of the anatomic structures and the accuracy of their ( ) determine the overall

More information

Single Photon Interference Katelynn Sharma and Garrett West University of Rochester, Institute of Optics, 275 Hutchison Rd. Rochester, NY 14627

Single Photon Interference Katelynn Sharma and Garrett West University of Rochester, Institute of Optics, 275 Hutchison Rd. Rochester, NY 14627 Single Photon Interference Katelynn Sharma and Garrett West University of Rochester, Institute of Optics, 275 Hutchison Rd. Rochester, NY 14627 Abstract: In studying the Mach-Zender interferometer and

More information

CONFOCAL GRADED d-spacing MULTILAYER BEAM CONDITIONING OPTICS

CONFOCAL GRADED d-spacing MULTILAYER BEAM CONDITIONING OPTICS Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 321 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42

More information

How Does One Obtain Spectral/Imaging Information! "

How Does One Obtain Spectral/Imaging Information! How Does One Obtain Spectral/Imaging Information! How do we measure the position, energy, and arrival time of! an X-ray photon?! " What we observe depends on the instruments that one observes with!" In

More information

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE A. Miceli ab, R. Thierry a, A. Flisch a, U. Sennhauser a, F. Casali b a Empa - Swiss Federal Laboratories for

More information

LYNXEYE XE-T. < 380 ev. Innovation with Integrity. Energy. Resolution. High-Resolution Position Sensitive Detector with Superb Energy Resolution XRD

LYNXEYE XE-T. < 380 ev. Innovation with Integrity. Energy. Resolution. High-Resolution Position Sensitive Detector with Superb Energy Resolution XRD Energy < 380 ev Resolution High-Resolution Position Sensitive Detector with Superb Energy Resolution The is the next generation "Compound Silicon Strip" detector with superb energy resolution for ultrafast

More information

THE CALIBRATION OF THE OPTICAL IMAGER FOR THE HOKU KEA TELESCOPE. Jamie L. H. Scharf Physics & Astronomy, University of Hawai i at Hilo Hilo, HI 96720

THE CALIBRATION OF THE OPTICAL IMAGER FOR THE HOKU KEA TELESCOPE. Jamie L. H. Scharf Physics & Astronomy, University of Hawai i at Hilo Hilo, HI 96720 THE CALIBRATION OF THE OPTICAL IMAGER FOR THE HOKU KEA TELESCOPE Jamie L. H. Scharf Physics & Astronomy, University of Hawai i at Hilo Hilo, HI 96720 ABSTRACT I have been calibrating the science CCD camera

More information

X-ray backscattering: Variable irradiation geometry facilitates new insights

X-ray backscattering: Variable irradiation geometry facilitates new insights 18 th World Conference of Non Destructive Testing, 16-20 April 2012, Durban, South Africa X-ray backscattering: Variable irradiation geometry facilitates new insights Norma WROBEL 1, Kurt OSTERLOH 1, Mirko

More information

photolithographic techniques (1). Molybdenum electrodes (50 nm thick) are deposited by

photolithographic techniques (1). Molybdenum electrodes (50 nm thick) are deposited by Supporting online material Materials and Methods Single-walled carbon nanotube (SWNT) devices are fabricated using standard photolithographic techniques (1). Molybdenum electrodes (50 nm thick) are deposited

More information

MICROFOCUSING X-RAY EQUIPMENT FOR THE LAB DIFFRACTOMETER

MICROFOCUSING X-RAY EQUIPMENT FOR THE LAB DIFFRACTOMETER 29 MICROFOCUSING X-RAY EQUIPMENT FOR THE LAB DIFFRACTOMETER Jörg Wiesmann, 1 Jürgen Graf, 1 Christian Hoffmann, 1 Carsten Michaelsen, 1 Alexandra Oehr, 1 Uwe Preckwinkel, 2 Ning Yang, 2 Holger Cordes,

More information

QE65000 Spectrometer. Scientific-Grade Spectroscopy in a Small Footprint. now with. Spectrometers

QE65000 Spectrometer. Scientific-Grade Spectroscopy in a Small Footprint. now with. Spectrometers QE65000 Spectrometer Scientific-Grade Spectroscopy in a Small Footprint QE65000 The QE65000 Spectrometer is the most sensitive spectrometer we ve developed. Its Hamamatsu FFT-CCD detector provides 90%

More information

Orbit Stability Challenges for Storage Rings. Glenn Decker Advanced Photon Source Beam Diagnostics March 8, 2012

Orbit Stability Challenges for Storage Rings. Glenn Decker Advanced Photon Source Beam Diagnostics March 8, 2012 Orbit Stability Challenges for Storage Rings Glenn Decker Advanced Photon Source Beam Diagnostics March 8, 2012 Outline Beam stability requirements RF beam position monitor technology NSLS II developments

More information

Fastest high definition Raman imaging. Fastest Laser Raman Microscope RAMAN

Fastest high definition Raman imaging. Fastest Laser Raman Microscope RAMAN Fastest high definition Raman imaging Fastest Laser Raman Microscope RAMAN - 11 www.nanophoton.jp Observation A New Generation in Raman Observation RAMAN-11 developed by Nanophoton was newly created by

More information

AutoMATE II. Micro-area X-ray stress measurement system. Highly accurate micro area residual stress

AutoMATE II. Micro-area X-ray stress measurement system. Highly accurate micro area residual stress AutoMATE II Micro-area X-ray stress measurement system Highly accurate micro area residual stress The accuracy of an R&D diffractom dedicated residua In the past, if you wanted to make highly accurate

More information

ME 6406 MACHINE VISION. Georgia Institute of Technology

ME 6406 MACHINE VISION. Georgia Institute of Technology ME 6406 MACHINE VISION Georgia Institute of Technology Class Information Instructor Professor Kok-Meng Lee MARC 474 Office hours: Tues/Thurs 1:00-2:00 pm kokmeng.lee@me.gatech.edu (404)-894-7402 Class

More information

Miniflex. Rigaku/ Miniflex X-ray Diffractometer System. Rigaku Corporation

Miniflex. Rigaku/ Miniflex X-ray Diffractometer System. Rigaku Corporation Miniflex Rigaku/ Miniflex X-ray Diffractometer System Rigaku Corporation Rigaku/ Miniflex X-ray Diffractometer System 1. Introduction Rigaku s general purpose X-ray diffractometer systems are broadly classified

More information

Abstract. 1. Introduction

Abstract. 1. Introduction 941 Monte Carlo Simulation of Pulse Pile Up R. P. Gardner and S. H. Lee Center for Engineering Applications of Radioisotopes Department of Engineering North Carolina State University Raleigh, NC 276957909,

More information

FIRST INDIRECT X-RAY IMAGING TESTS WITH AN 88-mm DIAMETER SINGLE CRYSTAL

FIRST INDIRECT X-RAY IMAGING TESTS WITH AN 88-mm DIAMETER SINGLE CRYSTAL FERMILAB-CONF-16-641-AD-E ACCEPTED FIRST INDIRECT X-RAY IMAGING TESTS WITH AN 88-mm DIAMETER SINGLE CRYSTAL A.H. Lumpkin 1 and A.T. Macrander 2 1 Fermi National Accelerator Laboratory, Batavia, IL 60510

More information

IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY

IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY WHITE PAPER: IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY IBEX Innovations Ltd. Registered in England and Wales: 07208355 Address: Discovery 2, NETPark, William Armstrong Way, Sedgefield, UK Patents:

More information

By using patented polycapillary optics this diffractometer obviates the need for monochromators and collimators for linear projection of X-Rays.

By using patented polycapillary optics this diffractometer obviates the need for monochromators and collimators for linear projection of X-Rays. XRD X-Ray Diffractometer Innovative, Integrated, Multifunctional By using patented polycapillary optics this diffractometer obviates the need for monochromators and collimators for linear projection of

More information

A PORTABLE X-RAY APPARATUS FOR BOTH STRESS MEASUREMENT AND PHASE ANALYSIS UNDER FIELD CONDITIONS.

A PORTABLE X-RAY APPARATUS FOR BOTH STRESS MEASUREMENT AND PHASE ANALYSIS UNDER FIELD CONDITIONS. Copyright(c)JCPDS-International Centre for Diffraction Data 2000,Advances in X-ray Analysis,Vol.43 66 A PORTABLE X-RAY APPARATUS FOR BOTH STRESS MEASUREMENT AND PHASE ANALYSIS UNDER FIELD CONDITIONS. V.

More information

Experiment #12 X-Ray Diffraction Laboratory

Experiment #12 X-Ray Diffraction Laboratory Physics 360/460 Experiment #12 X-Ray Diffraction Laboratory Introduction: To determine crystal lattice spacings, as well as identify unknown substances, a x-ray diffractometer is used to replace the traditional

More information

University of Wisconsin Chemistry 524 Spectroscopic Components *

University of Wisconsin Chemistry 524 Spectroscopic Components * University of Wisconsin Chemistry 524 Spectroscopic Components * In journal articles, presentations, and textbooks, chemical instruments are often represented as block diagrams. These block diagrams highlight

More information

Multianode Photo Multiplier Tubes as Photo Detectors for Ring Imaging Cherenkov Detectors

Multianode Photo Multiplier Tubes as Photo Detectors for Ring Imaging Cherenkov Detectors Multianode Photo Multiplier Tubes as Photo Detectors for Ring Imaging Cherenkov Detectors F. Muheim a edin]department of Physics and Astronomy, University of Edinburgh Mayfield Road, Edinburgh EH9 3JZ,

More information

R-AXIS RAPID. X-ray Single Crystal Structure Analysis System. Product Information

R-AXIS RAPID. X-ray Single Crystal Structure Analysis System. Product Information The Rigaku Journal Vol. 15/ number 2/ 1998 Product Information X-ray Single Crystal Structure Analysis System R-AXIS RAPID 1. Introduction X-ray single crystal structure analysis is known as the easiest

More information

Charged-Coupled Devices

Charged-Coupled Devices Charged-Coupled Devices Charged-Coupled Devices Useful texts: Handbook of CCD Astronomy Steve Howell- Chapters 2, 3, 4.4 Measuring the Universe George Rieke - 3.1-3.3, 3.6 CCDs CCDs were invented in 1969

More information

Initial Results from a Cryogenic Proton Irradiation of a p-channel CCD

Initial Results from a Cryogenic Proton Irradiation of a p-channel CCD Centre for Electronic Imaging Initial Results from a Cryogenic Proton Irradiation of a p-channel CCD Jason Gow Daniel Wood, David Hall, Ben Dryer, Simeon Barber, Andrew Holland and Neil Murray Jason P.

More information

Experimental Competition

Experimental Competition 37 th International Physics Olympiad Singapore 8 17 July 2006 Experimental Competition Wed 12 July 2006 Experimental Competition Page 2 List of apparatus and materials Label Component Quantity Label Component

More information

Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA

Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA Abstract: Speckle interferometry (SI) has become a complete technique over the past couple of years and is widely used in many branches of

More information

Simultaneous Infrared-Visible Imager/Spectrograph a Multi-Purpose Instrument for the Magdalena Ridge Observatory 2.4-m Telescope

Simultaneous Infrared-Visible Imager/Spectrograph a Multi-Purpose Instrument for the Magdalena Ridge Observatory 2.4-m Telescope Simultaneous Infrared-Visible Imager/Spectrograph a Multi-Purpose Instrument for the Magdalena Ridge Observatory 2.4-m Telescope M.B. Vincent *, E.V. Ryan Magdalena Ridge Observatory, New Mexico Institute

More information

UltraGraph Optics Design

UltraGraph Optics Design UltraGraph Optics Design 5/10/99 Jim Hagerman Introduction This paper presents the current design status of the UltraGraph optics. Compromises in performance were made to reach certain product goals. Cost,

More information

Oriel MS260i TM 1/4 m Imaging Spectrograph

Oriel MS260i TM 1/4 m Imaging Spectrograph Oriel MS260i TM 1/4 m Imaging Spectrograph MS260i Spectrograph with 3 Track Fiber on input and InstaSpec CCD on output. The MS260i 1 4 m Imaging Spectrographs are economical, fully automated, multi-grating

More information

Instructions for the Experiment

Instructions for the Experiment Instructions for the Experiment Excitonic States in Atomically Thin Semiconductors 1. Introduction Alongside with electrical measurements, optical measurements are an indispensable tool for the study of

More information