MICROANALYSIS WITH A POLYCAPILLARY IN A VACUUM CHAMBER

Size: px
Start display at page:

Download "MICROANALYSIS WITH A POLYCAPILLARY IN A VACUUM CHAMBER"

Transcription

1 THE RIGAKU JOURNAL VOL. 20 / NO. 2 / 2003 MICROANALYSIS WITH A POLYCAPILLARY IN A VACUUM CHAMBER CHRISTINA STRELI a), NATALIA MAROSI, PETER WOBRAUSCHEK AND BARBARA FRANK Atominstitut der Österreichischen Universitäten, Vienna University of Technology, Stadionallee 2, A-1020 Vienna, Austria Microanalysis using a polycapillary as a focusing device was performed under vacuum conditions. The polycapillary was implemented in a vacuum chamber so that also low energy radiation could be used for sample excitation. A Mo tube with a 125 mm window was used, offering the possibility to employ Mo-L radiation (2.3 kev) for the excitation of lower Z elements. A Silicon drift detector (KETEK, Munich) with 5 mm 2 active area and a 7 mm Duraberyllium TM window allows the detection of low energy radiation down to Na. The operation in vacuum also reduces the background originating from scattering in air and thereby helps to improve the detection limits for medium Z elements as well. Results from an area scan around a droplet multi-element sample (distribution of elements on the surface) is presented, and measurements on soda lime glass containing Na and Al, as well as a comparison between air and vacuum conditions are shown. Detection limits of 4600 mg/g for Na in soda lime glass have been obtained. 1. Introduction Single and polycapillaries have been used widely in the recent years to produce intensive microbeams for performing microanalysis (EDXRF) with a spatial resolution of about mm [1 5] but with a few exceptions [6] these experiments have been only carried out in air. The detectable elemental range is thereby limited to elements above S (E Ka 2.3 kev), because the air path between sample and detector about 1 cm absorbs most of the fluorescence radiation with lower energies. The advantages of measuring lower Z elements with the spatial resolution obtainable by using a polycapillary initiated the idea to design a microbeam vacuum chamber. The Atominstitut X-ray group has a long experience in building vacuum chambers, however mainly for TXRF [7 10]. This experience was the backing for the actual construction of a vacuum chamber with all important components such as the polycapillary, sample holder and detector situated inside and fully adjustable with proper instrumentation. The main goal was to reduce absorption of the fluorescence radiation between sample and detector as well as to reduce the scattering in air, whereby the spectral background is minimized and the Ar-K line practically removed. The detector used was suitable for the detection of low energy radiation from Na-Ka upwards. a) Electronic mail: streli@ati.ac.at Polycapillaries offer an additional advantage in the low energy region due to the total reflection at the inner walls of the capillaries, whereby the higher energies are absorbed and only the lower energies reflected (cut-off effect), leading to efficient excitation of light elements. 2. Experimental Setup Figure 1 shows the scheme of the experimental setup and Fig. 2 a picture of the vacuum chamber with all components. A long fine focus diffraction Mo-tube (Philips PW 1043/01) of 3 kw maximum power with a point focal spot of mm was used. The thickness of the Be window is only 125 mm, so that Mo-L radiation (E 2.3 kev) is efficiently transmitted. A polycapillary manufactured by XOS [11] was used (borosilicate glass, single capillaries of 9 mm inner diameter) offering the following characteristics: focal distance from source to the entrance: 53 mm, length: 62 mm, diameter: 8mm, focal distance at exit: 13.5 mm, resulting in a total distance from source to sample of 128 mm. The focal spot size as specified for Mo- Ka radiation (17.48 kev) is 32 mm for a source size of mm. The intensity gain in comparison to a pinhole of 50 mm diameter for a source size of mm and the same energy is 39 [11]. The polycapillary was mounted inside the chamber in a brass tube which was fixed at the entrance wall. The entrance window was made of 8 mm Kapton TM, which allows high transmission also of lower energy radiation. Vol. 20 No

2 Fig. 1. Schematic setup with polycapillary in the vacuum chamber. Fig. 2. Experimental setup, view from top. The adjustment was performed without any special devices such as gimbals or rotation or translation stages, simply by moving the X-ray tube together with the adapter unit (which also serves as a radiation protection between tube and chamber) until the irradiated spot on a ZnS screen was at its maximum intensity. This spot was also checked using a CCD camera without optics, just by monitoring the X-rays directly on the chip after passing the glass protection and an Al foil which absorbs the light. After optimization the tube was fixed in its position. The detector used is a silicon drift detector (KETEK GmbH, Munich) [12] with 5 mm2 area and a thickness of 300 m m. The entrance window is 8 m m DuraberylliumTM, offering a transmission of 25% for Na-Ka. The detector was cooled electrically by Peltier effect so that no LN2 is required, which made the detector handsome and versatile to use. The amplification is performed by a Canberra 2025 amplifier with the shaping 26 time set to 0.5 m s. An optical microscope is used to control the position at which the sample is hit by the beam. As the sample is situated inside the chamber and the microscope is mounted outside, the view is performed through a glass window in the chamber wall. The working distance of the microscope is about 186 mm and both, direct view or transfer to a computer via a CCD camera is possible (Fig. 2). The sample is mounted on a slide frame, which can be inserted in a holder which in turn is mounted on an x, y, z stage. This allows scanning of the sample surface, but also moving the sample in the X-ray focal spot, which can be controlled by the focused optical image seen by the optical microscope. The scanning as well as all data handling between MCA and stepper motors involved in the scanning are performed under computer control. The setup was designed and tested within the frame of the diploma thesis of N. Marosi [13]. The Rigaku Journal

3 3. Results and Discussion First of all the size of the focal spot achieved with this particular polycapillary was determined by moving a knife wedge (razor blade) in 5 m m steps through the beam. Figure 3 shows the resulting intensity as a function of the position of the wedge. Differentiation yields the intensity variation: A Gaussian curve was fitted through the data, leading to a FWHM of 69 m m. This value differs from the value specified by the manufacturer; the reason seems to be the different focal spot sizes of the X-ray tubes. As a test for the capability of the new system the distribution of a multi-element solution dried on a polished TXRF quartz reflector was analyzed. 5 m l of the solution containing 5 m g/g of Cd, Co, Cu, Cr, Mn, Ni, Pb, V, Zn, and Ga were pipetted on a quartz reflector. The dried residue was investigated over an area of mm with a step width of 300 m m horizontally and vertically. The measuring time was 50 s for each scanned point with 30 kv and 20 ma. The distance between sample and detector was about 10 mm for the setting of optimum focusing. The distribution of V, Cr and Mn across the sample can be seen in Fig. 4. It is not homogenous but shows accumulations at specific points forming a ring like structure. However, all elements show their maximum intensity at the same spot. The capability of determining low Z elements with the new system was tested by measuring a soda lime glass (NIST 621 standard reference material). Na and Al are clearly visible; the concentration of Na2O is about 13% (cna 10%) and that of Al2O3 about 3% (cal 2%). The obtained Fig. 3. Vol. 20 No Determination of the focal spot-size. spectrum is shown in Fig. 5. Table 1 lists the detection limits obtained in soda lime under 30 kv and 10 ma. A comparison between operation in air and vacuum was made by measuring a multi-element sample dried on a KaptonTM foil of 8 m m Fig. 4. Distribution of elements of a multi-element solution dried on a TXRF quartz reflector. 27

4 Fig. 5. Spectrum of a soda lime glass (NIST 621) excited with Mo anode X-ray tube, 30 kv, 10 ma, 1000 s counting time. Table 1. Data obtained from NIST 621, standard reference material soda lime. Excitation conditions: 30 kv, 10 ma, measuring time: 1000 s. Index N denotes net counts and index B background counts. Vac: N N Vac: N B LLD (mg/g) Na Al Fig. 6. Multi-element sample on Kapton TM foil, Mo 30 kv, 10 ma, 50 s, (a) under vacuum conditions, (b) in air. 28 The Rigaku Journal

5 Table 2. Net and background counts, N N and N B, for measurement in air and under vacuum conditions as well as the detection limits (LLD) for 1000 s measuring time. Sample: 5 mg/g K and Ca, 100 mg/g Cr. A 5 ml droplet was pipetted on Kapton foil. 30 kv, 10 ma, 50 s. Air: N N Air: N B LLD (mg/g) Vac: N N Vac: N B LLD (mg/g) K Ca Cr thickness. The result is shown in Fig. 6. The intensity of the low energy fluorescence radiation under vacuum conditions is significantly improved. The background changes roughly in proportion to the increasing intensity. Particularly significant is the improvement of the exciting Mo-L radiation from the X-ray tube (as seen by the scatter peak) as well as the fluorescent Mg-K line (1.25 kev), Rb and Sr-L lines (1.83 kev), K and Ca-Ka lines. The Ar-K line is visible only in air. Table 2 shows the differences in net and background counts for K, Ca and Cr. Generally, the net counts are enhanced under vacuum conditions and the peak to background ratio is improved. The dried residue forms roughly a spot of 1 mm diameter; the illuminated area has a diameter of about 70 mm 2. Detection limits obtained under vacuum conditions for K are 8 ng/g. 4. Conclusions The low energy performance of a system operating a polycapillary in a vacuum chamber using a detector suitable for the detection of low energy X-rays was studied. The detection of low energy photons down to Na (1.04 kev) was shown. Detection limits of 4600 mg/g for Na in soda lime glass have been obtained, as well as 8 ng/g for K measured as dried residue on a Kapton by using a Mo-target X-ray tube with a thin window which allowed efficient transmission of Mo-L radiation. References [ 1] Kumakhov: SPIE, 116 (1996), [ 2] Hoffman, Thiel, Bilderback: Nucl. Instr. Meth., A347 (1994), 384. [ 3] B. Vekemans, K. Janssens, G. Vittiglio, F. Adams, L. Andong, Y. Yiming: JCPDS, (1999), [ 4] G. Havrilla: Adv. X-ray Anal., 41 (1999), 234. [ 5] Kumakhov: X-Ray Spectrom., 29 (2000), [6] N. Gao, I. Yu. Ponomarev, Q. F. Xiao, W. M. Gibson, D. A. Carpenter: Enhancement of microbeam x-ray fluorescence (MXRF) analysis using monolithic polycapillary focusing optics., [7] R. Rieder, W. Ladisich, P. Wobrauschek, C. Streli, P. Kregsamer: Nucl. Instr. Meth., A327 (1993), 594. [ 8] C. Streli, P. Wobrauschek, H. Aiginger: Spectrochim. Acta, 48B (1993), 163. [9] C. Streli, P. Wobrauschek, E. Unfried, H. Aiginger: Nucl. Instr. Meth., A334 (1993), 425. [10] C. Streli, P. Wobrauschek, V. Bauer, P. Kregsamer, R. Görgl, P. Pianetta, R. Ryon, S. Pahlke, L. Fabry: Spectrochim Acta, 52B (1997), 861. [11] Specification by XOS. [12] X-ray Optical Systems, Inc., Greenbush, N.Y., USA [13] N. Marosi: Diploma thesis of the Atoministitute, Vienna Univ. of Technology (2002). Vol. 20 No

MICRO XRF OF LIGHT ELEMENTS USING A POLYCAPILLARY LENS AND AN ULTRA THIN WINDOW SILICON DRIFT DETECTOR INSIDE A VACUUM CHAMBER

MICRO XRF OF LIGHT ELEMENTS USING A POLYCAPILLARY LENS AND AN ULTRA THIN WINDOW SILICON DRIFT DETECTOR INSIDE A VACUUM CHAMBER Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 229 MICRO XRF OF LIGHT ELEMENTS USING A POLYCAPILLARY LENS AND AN ULTRA THIN WINDOW SILICON DRIFT

More information

Advancing EDS Analysis in the SEM Quantitative XRF. International Microscopy Congress, September 5 th, Outline

Advancing EDS Analysis in the SEM Quantitative XRF. International Microscopy Congress, September 5 th, Outline Advancing EDS Analysis in the SEM with in-situ Quantitative XRF Brian J. Cross (1) & Kenny C. Witherspoon (2) 1) CrossRoads Scientific, El Granada, CA 94018, USA 2) ixrf Systems, Inc., Houston, TX 77059,

More information

DOUBLE MULTILAYER MONOCHROMATOR WITH FIXED EXIT GEOMETRY. H.Gatterbauer, P.Wobrauschek, F.Hegediis, P.Biini, C.Streli

DOUBLE MULTILAYER MONOCHROMATOR WITH FIXED EXIT GEOMETRY. H.Gatterbauer, P.Wobrauschek, F.Hegediis, P.Biini, C.Streli Copyright (C) JCPDS International Centre for Diffraction Data 1999 379 DOUBLE MULTILAYER MONOCHROMATOR WITH FIXED EXIT GEOMETRY H.Gatterbauer, P.Wobrauschek, F.Hegediis, P.Biini, C.Streli Atominsitut der

More information

Zaidi Embong and Husin Wagiran Physics Department, University Of Technology Malaysia, P.O Box 791, 80990, Johor Baharu

Zaidi Embong and Husin Wagiran Physics Department, University Of Technology Malaysia, P.O Box 791, 80990, Johor Baharu MY9800971 Optimization of a Spectrometry for Energy -Dispersive X-ray Fluorescence Analysis by X-ray Tube in Combination with Secondary Target for Multielements Determination of Sediment Samples. Zaidi

More information

Silicon Drift Detector. with On- Chip Ele ctronics for X-Ray Spectroscopy. KETEK GmbH Am Isarbach 30 D O berschleißheim GERMANY

Silicon Drift Detector. with On- Chip Ele ctronics for X-Ray Spectroscopy. KETEK GmbH Am Isarbach 30 D O berschleißheim GERMANY KETEK GmbH Am Isarbach 30 D-85764 O berschleißheim GERMANY Silicon Drift Detector Phone +49 (0)89 315 57 94 Fax +49 (0)89 315 58 16 with On- Chip Ele ctronics for X-Ray Spectroscopy high energy resolution

More information

research papers First results from a macromolecular crystallography system with a polycapillary collimating optic and a microfocus X-ray generator

research papers First results from a macromolecular crystallography system with a polycapillary collimating optic and a microfocus X-ray generator Journal of Applied Crystallography ISSN 0021-8898 First results from a macromolecular crystallography system with a polycapillary collimating optic and a microfocus X-ray generator Received 7 September

More information

ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER

ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 325 ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER ABSTRACT William Chang, Jonathan Kerner, and Edward

More information

BRUKER ADVANCED X-RAY SOLUTIONS. SPECTROMETRY SOLUTIONS ARTAX mxrf SPECTROMETER

BRUKER ADVANCED X-RAY SOLUTIONS. SPECTROMETRY SOLUTIONS ARTAX mxrf SPECTROMETER BRUKER ADVANCED X-RAY SOLUTIONS SPECTROMETRY SOLUTIONS ARTAX mxrf SPECTROMETER Microanalysis ARTAX Elemental Analysis for the Art Community and More Non-destructive elemental analysis is strictly required

More information

OPTIMIZING THE ELEMENTAL SENSITIVITY AND FOCAL SPOT SIZE OF A MONOLITHIC POLYCAPILLARY OPTIC USING MICRO-X-RAY FLUORESCENCE

OPTIMIZING THE ELEMENTAL SENSITIVITY AND FOCAL SPOT SIZE OF A MONOLITHIC POLYCAPILLARY OPTIC USING MICRO-X-RAY FLUORESCENCE , Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 26 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42

More information

Multi-Element Analysis by Portable Total Reflection X-ray Fluorescence Spectrometer

Multi-Element Analysis by Portable Total Reflection X-ray Fluorescence Spectrometer ANALYTICAL SCIENCES AUGUST 2013, VOL. 29 793 2013 The Japan Society for Analytical Chemistry Multi-Element Analysis by Portable Total Reflection X-ray Fluorescence Spectrometer Ying LIU, Susumu IMASHUKU,

More information

Spectral distribution from end window X-ray tubes

Spectral distribution from end window X-ray tubes Copyright ISSN (C) 1097-0002, JCPDS-International Advances in X-ray Centre Analysis, for Volume Diffraction 41 Data 1999 393 Spectral distribution from end window X-ray tubes N. Broll 1, P. de Chateaubourg

More information

ARTAX. Innovation with Integrity. Portable Micro-XRF Spectrometer. Micro-XRF

ARTAX. Innovation with Integrity. Portable Micro-XRF Spectrometer. Micro-XRF ARTAX Portable Micro-XRF Spectrometer Innovation with Integrity Micro-XRF ARTAX Elemental Analysis for the Art Community and More The ARTAX is the first portable X ray fluorescence (XRF) spectrometer designed

More information

Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4

Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4 Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4 S.V. Roth, R. Döhrmann, M. Dommach, I. Kröger, T. Schubert, R. Gehrke Definition of the upgrade The wiggler beamline BW4 is dedicated to

More information

WIDE ANGLE GEOMETRY EDXRF SPECTROMETERS WITH SECONDARY TARGET AND DIRECT EXCITATION MODES

WIDE ANGLE GEOMETRY EDXRF SPECTROMETERS WITH SECONDARY TARGET AND DIRECT EXCITATION MODES Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 11 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42

More information

DEVELOPMENT OF A WAVELENGTH DISPERSIVE X-RAY FLUORESCENCE SPECTROMETER USING A MULTI-CAPILLARY X-RAY LENS FOR X-RAY DETECTION

DEVELOPMENT OF A WAVELENGTH DISPERSIVE X-RAY FLUORESCENCE SPECTROMETER USING A MULTI-CAPILLARY X-RAY LENS FOR X-RAY DETECTION Copyright JCPDS - International Centre for Diffraction Data 2003, Advances in X-ray Analysis, Volume 46. 346 DEVELOPMENT OF A WAVELENGTH DISPERSIVE X-RAY FLUORESCENCE SPECTROMETER USING A MULTI-CAPILLARY

More information

Preliminary Comparison of Monolithic and Aperture Optics for XRMF. George J. Havrilla CST-8 Christopher G. Worley CST-8

Preliminary Comparison of Monolithic and Aperture Optics for XRMF. George J. Havrilla CST-8 Christopher G. Worley CST-8 Title: Author(s): Submitted to: 1998327 63 LosAlamos NATIONAL LABORATORY Preliminary Comparison of Monolithic and Aperture Optics for XRMF George J. Havrilla CST-8 Christopher G. Worley CST-8 RECEIVED

More information

DESIGN AND MEASUREMENT WITH A NEW PORTABLE X-RAY CAMERA FOR FULL-FIELD FLUORESCENCE IMAGING

DESIGN AND MEASUREMENT WITH A NEW PORTABLE X-RAY CAMERA FOR FULL-FIELD FLUORESCENCE IMAGING 14 DESIGN AND MEASUREMENT WITH A NEW PORTABLE X-RAY CAMERA FOR FULL-FIELD FLUORESCENCE IMAGING I. Ordavo 1,2, A. Bjeoumikhov 3, S. Bjeoumikhova 3, G. Buzanich 4, R. Gubzhokov 4, R. Hartmann 1, S. Ihle

More information

X-ray generation by femtosecond laser pulses and its application to soft X-ray imaging microscope

X-ray generation by femtosecond laser pulses and its application to soft X-ray imaging microscope X-ray generation by femtosecond laser pulses and its application to soft X-ray imaging microscope Kenichi Ikeda 1, Hideyuki Kotaki 1 ' 2 and Kazuhisa Nakajima 1 ' 2 ' 3 1 Graduate University for Advanced

More information

M4 TORNADO PLUS. Innovation with Integrity. Super Light Element Micro-XRF Spectrometer. Micro-XRF

M4 TORNADO PLUS. Innovation with Integrity. Super Light Element Micro-XRF Spectrometer. Micro-XRF M4 TORNADO PLUS Super Light Element Micro-XRF Spectrometer Innovation with Integrity Micro-XRF M4 TORNADO PLUS - A New Era in Micro-XRF M4 TORNADO PLUS is the world's first Micro-XRF spectrometer that

More information

PSPC/MDG 2000 X-RAY MICRODIFFRACTOMETER. Product Information

PSPC/MDG 2000 X-RAY MICRODIFFRACTOMETER. Product Information THE RIGAKU JOURNAL VOL. 11 I NO.2 I 1994 Product Information X-RAY MICRODIFFRACTOMETER PSPC/MDG 2000 1. Introduction The analysis of X-ray diffraction patterns is well known as an effective means of obtaining

More information

GUNSHOT RESIDUE INVESTIGATIONS USING TXRF

GUNSHOT RESIDUE INVESTIGATIONS USING TXRF 299 GUNSHOT RESIDUE INVESTIGATIONS USING TXRF Alexander Wastl 1, Bettina Bogner 2, Peter Kregsamer 1, Peter Wobrauschek 1, Christina Streli 1 1 Atominstitut, Vienna University of Technology, Vienna, Austria

More information

RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS

RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS S-94,316 PATENTS-US-A96698 BEAM POSITION MONITOR RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS CONTRACTUAL ORIGIN OF THE INVENTION The United States Government has rights in this invention pursuant

More information

Supplementary Figure S1. Schematic representation of different functionalities that could be

Supplementary Figure S1. Schematic representation of different functionalities that could be Supplementary Figure S1. Schematic representation of different functionalities that could be obtained using the fiber-bundle approach This schematic representation shows some example of the possible functions

More information

Polycapillary optic source combinations for protein crystallography

Polycapillary optic source combinations for protein crystallography Journal of Applied Crystallography ISSN 0021-8898 Polycapillary optic source combinations for protein crystallography F. A. Hofmann, W. M. Gibson, C. A. MacDonald, D. A. Carter, J. X. Ho and J. R. Ruble

More information

X-RAY OPTICS FOR TWO-DIMENSIONAL DIFFRACTION

X-RAY OPTICS FOR TWO-DIMENSIONAL DIFFRACTION Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 332 ABSTRACT X-RAY OPTICS FOR TWO-DIMENSIONAL DIFFRACTION Bob B. He and Uwe Preckwinkel Bruker

More information

Supplementary Figure 1

Supplementary Figure 1 Supplementary Figure 1 Technical overview drawing of the Roadrunner goniometer. The goniometer consists of three main components: an inline sample-viewing microscope, a high-precision scanning unit for

More information

SIMULTANEOUS XRD/XRF WITH LOW-POWER X-RAY TUBES

SIMULTANEOUS XRD/XRF WITH LOW-POWER X-RAY TUBES Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 34 SIMULTANEOUS XRD/XRF WITH LOW-POWER X-RAY TUBES S. Cornaby 1, A. Reyes-Mena 1, P. W. Moody 1,

More information

X-Ray Spectroscopy with a CCD Detector. Application Note

X-Ray Spectroscopy with a CCD Detector. Application Note X-Ray Spectroscopy with a CCD Detector In addition to providing X-ray imaging solutions, including CCD-based cameras that image X-rays using either direct detection (0.5-20 kev) or indirectly using a scintillation

More information

High Rep-Rate KrF Laser Development and Intense Pulse Interaction Experiments for IFE*

High Rep-Rate KrF Laser Development and Intense Pulse Interaction Experiments for IFE* High Rep-Rate KrF Laser Development and Intense Pulse Interaction Experiments for IFE* Y. Owadano, E. Takahashi, I. Okuda, I. Matsushima, Y. Matsumoto, S. Kato, E. Miura and H.Yashiro 1), K. Kuwahara 2)

More information

XRF Instrumentation. Introduction to spectrometer

XRF Instrumentation. Introduction to spectrometer XRF Instrumentation Introduction to spectrometer AMPTEK, INC., Bedford, MA 01730 Ph: +1 781 275 2242 Fax: +1 781 275 3470 sales@amptek.com 1 Instrument Excitation source Sample X-ray tube or radioisotope

More information

USING A CHARGE-COUPLED DEVICE (CCD) TO GATHER X-RAY FLUORESCENCE (XRF)AND X-RAY DIFFRACTION (XRD) INFORMATION SIMULTANEOUSLY

USING A CHARGE-COUPLED DEVICE (CCD) TO GATHER X-RAY FLUORESCENCE (XRF)AND X-RAY DIFFRACTION (XRD) INFORMATION SIMULTANEOUSLY Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 343 USING A CHARGE-COUPLED DEVICE (CCD) TO GATHER X-RAY FLUORESCENCE (XRF)AND X-RAY DIFFRACTION (XRD)

More information

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage

More information

Phase Imaging Using Focused Polycapillary Optics

Phase Imaging Using Focused Polycapillary Optics Phase Imaging Using Focused Polycapillary Optics Sajid Bashir, Sajjad Tahir, Jonathan C. Petruccelli, C.A. MacDonald Dept. of Physics, University at Albany, Albany, New York Abstract Contrast in conventional

More information

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Takeshi Fujiwara 1, Yuki Mitsuya 2, Hiroyuki Takahashi 2, and Hiroyuki Toyokawa 2 1 National Institute of Advanced Industrial

More information

LECTURE 10. Dr. Teresa D. Golden University of North Texas Department of Chemistry

LECTURE 10. Dr. Teresa D. Golden University of North Texas Department of Chemistry LECTURE 10 Dr. Teresa D. Golden University of North Texas Department of Chemistry Components for the source include: -Line voltage supply -high-voltage generator -x-ray tube X-ray source requires -high

More information

Spectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer :

Spectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer : Spectrophotometer An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer : Spectrophotometer components Excitation sources Deuterium Lamp Tungsten

More information

Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror

Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror Stephen B. Robie scintag, Inc. 10040 Bubb Road Cupertino, CA 95014 Abstract Corundum

More information

Bandpass Interference Filters

Bandpass Interference Filters Precise control of center wavelength and bandpass shape Wide selection of stock wavelengths from 250 nm-1550 nm Selection of bandwidths Available in 1/2 and 1 sizes High peak transmission values Excellent

More information

Observation of X-rays generated by relativistic electrons in waveguide target mounted inside a betatron

Observation of X-rays generated by relativistic electrons in waveguide target mounted inside a betatron Observation of X-rays generated by relativistic electrons in waveguide target mounted inside a betatron V.V.Kaplin (1), V.V.Sohoreva (1), S.R.Uglov (1), O.F.Bulaev (2), A.A.Voronin (2), M.Piestrup (3),

More information

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION Revised November 15, 2017 INTRODUCTION The simplest and most commonly described examples of diffraction and interference from two-dimensional apertures

More information

SILICON DRIFT DETECTORS (SDDs) [1] with integrated. Preliminary Results on Compton Electrons in Silicon Drift Detector

SILICON DRIFT DETECTORS (SDDs) [1] with integrated. Preliminary Results on Compton Electrons in Silicon Drift Detector Preliminary Results on Compton Electrons in Silicon Drift Detector T. Çonka-Nurdan, K. Nurdan, K. Laihem, A. H. Walenta, C. Fiorini, B. Freisleben, N. Hörnel, N. A. Pavel, and L. Strüder Abstract Silicon

More information

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120)

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120) Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120) Please contact Dr. Amanda Henkes for training requests and assistance: 979-862-5959, amandahenkes@tamu.edu Hardware LN 2 FTIR FTIR camera 1

More information

Evaluating the Performance of a Commercial Silicon Drift Detector for X-ray Microanalysis

Evaluating the Performance of a Commercial Silicon Drift Detector for X-ray Microanalysis Evaluating the Performance of a Commercial Silicon Drift Detector for X-ray Microanalysis Edward A. Kenik Materials Science & Technology Division, Oak Ridge National Laboratory, Oak Ridge, TN 37831 kenikea@ornl.gov

More information

CHAPTER 7. Components of Optical Instruments

CHAPTER 7. Components of Optical Instruments CHAPTER 7 Components of Optical Instruments From: Principles of Instrumental Analysis, 6 th Edition, Holler, Skoog and Crouch. CMY 383 Dr Tim Laurens NB Optical in this case refers not only to the visible

More information

M6 JETSTREAM. Innovation with Integrity. Large Area Micro X-ray Fluorescence Spectrometer. Micro-XRF

M6 JETSTREAM. Innovation with Integrity. Large Area Micro X-ray Fluorescence Spectrometer. Micro-XRF M6 JETSTREAM Large Area Micro X-ray Fluorescence Spectrometer Innovation with Integrity Micro-XRF Spatially Resolved Elemental Analysis of Large Objects The Bruker M6 JETSTREAM is designed for the nondestructive

More information

StarBright XLT Optical Coatings

StarBright XLT Optical Coatings StarBright XLT Optical Coatings StarBright XLT is Celestron s revolutionary optical coating system that outperforms any other coating in the commercial telescope market. Our most popular Schmidt-Cassegrain

More information

Assembly and Experimental Characterization of Fiber Collimators for Low Loss Coupling

Assembly and Experimental Characterization of Fiber Collimators for Low Loss Coupling Assembly and Experimental Characterization of Fiber Collimators for Low Loss Coupling Ruby Raheem Dept. of Physics, Heriot Watt University, Edinburgh, Scotland EH14 4AS, UK ABSTRACT The repeatability of

More information

Full-Field X-ray Fluorescence Microscopy Using a Color X-ray Camera (CXC)

Full-Field X-ray Fluorescence Microscopy Using a Color X-ray Camera (CXC) Full-Field X-ray Fluorescence Microscopy Using a Color X-ray Camera (CXC) Ursula Elisabeth Adriane Fittschen, 1 * Oliver Scharf, 2 and Magnus Menzel 3 1 Washington State University, Chemistry Department,

More information

Author s Accepted Manuscript

Author s Accepted Manuscript Author s Accepted Manuscript The X-ray quantum efficiency measurement of high resistivity CCDs Neil J. Murray, Andrew D. Holland, David R. Smith, Jason P. Gow, Peter J. Pool, David J. Burt PII: S0168-9002(09)00147-8

More information

Introduction to X-ray Detectors for Synchrotron Radiation Applications

Introduction to X-ray Detectors for Synchrotron Radiation Applications Introduction to X-ray Detectors for Synchrotron Radiation Applications Pablo Fajardo Instrumentation Services and Development Division ESRF, Grenoble EIROforum School on Instrumentation (ESI 2011) Outline

More information

Water-Window Microscope Based on Nitrogen Plasma Capillary Discharge Source

Water-Window Microscope Based on Nitrogen Plasma Capillary Discharge Source 2015 International Workshop on EUV and Soft X-Ray Sources Water-Window Microscope Based on Nitrogen Plasma Capillary Discharge Source T. Parkman 1, M. F. Nawaz 2, M. Nevrkla 2, M. Vrbova 1, A. Jancarek

More information

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region Feature Article JY Division I nformation Optical Spectroscopy Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region Raymond Pini, Salvatore Atzeni Abstract Multichannel

More information

CONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application

CONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application CONFIGURING Your Spectroscopy System For PEAK PERFORMANCE A guide to selecting the best Spectrometers, s, and s for your application Spectral Measurement System Spectral Measurement System Spectrograph

More information

FAST ELEMENTAL MAPPING WITH MICRO-XRF

FAST ELEMENTAL MAPPING WITH MICRO-XRF 286 FAST ELEMENTAL MAPPING WITH MICRO-XRF Haschke, M.; Rossek, U.; Tagle, R.; Waldschläger, U. Bruker Nano GmbH, 12489 Berlin, Schwarzschildstr.12 ABSTRACT X-Ray optics are now in common use for concentrating

More information

POLYCAPILLARY OPTICS AND X-RAY ANALYTICAL TECHNIQUES

POLYCAPILLARY OPTICS AND X-RAY ANALYTICAL TECHNIQUES Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 298 POLYCAPILLARY OPTICS AND X-RAY ANALYTICAL TECHNIQUES Yiming Yan a,b and Walter M. Gibson a,c

More information

LOS 1 LASER OPTICS SET

LOS 1 LASER OPTICS SET LOS 1 LASER OPTICS SET Contents 1 Introduction 3 2 Light interference 5 2.1 Light interference on a thin glass plate 6 2.2 Michelson s interferometer 7 3 Light diffraction 13 3.1 Light diffraction on a

More information

Full characterization tests of Micromegas with elongated pillars

Full characterization tests of Micromegas with elongated pillars University of Würzburg Full characterization tests of Micromegas with elongated pillars B. Alvarez1 Gonzalez, L. Barak1, J. Bortfeldt1, F. Dubinin3, G. Glonti1, F. Kuger1,2, P. Iengo1, E. Oliveri1, J.

More information

Components of Optical Instruments

Components of Optical Instruments Components of Optical Instruments General Design of Optical Instruments Sources of Radiation Wavelength Selectors (Filters, Monochromators, Interferometers) Sample Containers Radiation Transducers (Detectors)

More information

Unpolarized Cluster, Jet and Pellet Targets

Unpolarized Cluster, Jet and Pellet Targets Unpolarized Cluster, Jet and Pellet Targets Intense Electron Beams Workshop Cornell University, June 17-19, 2015 Institut für Kernphysik Typical Requirements on Internal Targets Target material: H 2, D

More information

Charge Loss Between Contacts Of CdZnTe Pixel Detectors

Charge Loss Between Contacts Of CdZnTe Pixel Detectors Charge Loss Between Contacts Of CdZnTe Pixel Detectors A. E. Bolotnikov 1, W. R. Cook, F. A. Harrison, A.-S. Wong, S. M. Schindler, A. C. Eichelberger Space Radiation Laboratory, California Institute of

More information

AutoMATE II. Micro-area X-ray stress measurement system. Highly accurate micro area residual stress

AutoMATE II. Micro-area X-ray stress measurement system. Highly accurate micro area residual stress AutoMATE II Micro-area X-ray stress measurement system Highly accurate micro area residual stress The accuracy of an R&D diffractom dedicated residua In the past, if you wanted to make highly accurate

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION Optically reconfigurable metasurfaces and photonic devices based on phase change materials S1: Schematic diagram of the experimental setup. A Ti-Sapphire femtosecond laser (Coherent Chameleon Vision S)

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, Fluorescent Light for the microscope stand. 2. Turn on the Scanner Power (1) on the front

More information

Lecture 21. Wind Lidar (3) Direct Detection Doppler Lidar

Lecture 21. Wind Lidar (3) Direct Detection Doppler Lidar Lecture 21. Wind Lidar (3) Direct Detection Doppler Lidar Overview of Direct Detection Doppler Lidar (DDL) Resonance fluorescence DDL Fringe imaging DDL Scanning FPI DDL FPI edge-filter DDL Absorption

More information

Partial Replication of Storms/Scanlan Glow Discharge Radiation

Partial Replication of Storms/Scanlan Glow Discharge Radiation Partial Replication of Storms/Scanlan Glow Discharge Radiation Rick Cantwell and Matt McConnell Coolescence, LLC March 2008 Introduction The Storms/Scanlan paper 1 presented at the 8 th international workshop

More information

Bruker Nano. M4 tornado. High performance micro-xrf spectrometer. think forward

Bruker Nano. M4 tornado. High performance micro-xrf spectrometer. think forward Bruker Nano M4 tornado High performance micro-xrf spectrometer think forward µ-xrf M4 TORNADO setting standards in µ-xrf µ-xrf is the method of choice for highly sensitive and non-destructive elemental

More information

Possibilities for Thick, Simple- Structure Silicon X-Ray Detectors Operated by Peltier Cooling

Possibilities for Thick, Simple- Structure Silicon X-Ray Detectors Operated by Peltier Cooling Possibilities for Thick, Simple- Structure Silicon X-Ray Detectors Operated by Peltier Cooling Hideharu Matsuura 1, Derek Hullinger 2, Ryota Okada 1, Seigo Kitanoya 1, Seiji Nishikawa 1, and Keith Decker

More information

1. Introduction X-ray absorption fine structure (XAFS) is an element-specific powerful technique for chemical analysis. In general, XAFS spectra are o

1. Introduction X-ray absorption fine structure (XAFS) is an element-specific powerful technique for chemical analysis. In general, XAFS spectra are o Installation of the soft X-ray quick XAFS system in the SR Center of Ritsumeikan University Masashi Yoshimura 1, Kohji Nakanishi 1, Kei Mitsuhara 2, Toshiaki Ohta 1 1) The SR Center, Ritsumeikan University,

More information

Report on BLP Spectroscopy Experiments Conducted on October 6, 2017: M. Nansteel

Report on BLP Spectroscopy Experiments Conducted on October 6, 2017: M. Nansteel Report on BLP Spectroscopy Experiments Conducted on October 6, 2017: M. Nansteel Summary Several spectroscopic measurements were conducted on October 6, 2017 at BLP to characterize the radiant power of

More information

Supplementary Information for. Surface Waves. Angelo Angelini, Elsie Barakat, Peter Munzert, Luca Boarino, Natascia De Leo,

Supplementary Information for. Surface Waves. Angelo Angelini, Elsie Barakat, Peter Munzert, Luca Boarino, Natascia De Leo, Supplementary Information for Focusing and Extraction of Light mediated by Bloch Surface Waves Angelo Angelini, Elsie Barakat, Peter Munzert, Luca Boarino, Natascia De Leo, Emanuele Enrico, Fabrizio Giorgis,

More information

Design and Construction of a High Energy, High Average Power Nd:Glass Slab Amplifier. Dale Martz Department of Electrical & Computer Engineering

Design and Construction of a High Energy, High Average Power Nd:Glass Slab Amplifier. Dale Martz Department of Electrical & Computer Engineering Design and Construction of a High Energy, High Average Power Nd:Glass Slab Amplifier Dale Martz Department of Electrical & Computer Engineering 7/19/2006 Outline Introduction Nd:Glass Slab Nd:Glass Material

More information

Improving the Collection Efficiency of Raman Scattering

Improving the Collection Efficiency of Raman Scattering PERFORMANCE Unparalleled signal-to-noise ratio with diffraction-limited spectral and imaging resolution Deep-cooled CCD with excelon sensor technology Aberration-free optical design for uniform high resolution

More information

Spectroscopy of Ruby Fluorescence Physics Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018

Spectroscopy of Ruby Fluorescence Physics Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018 1 Spectroscopy of Ruby Fluorescence Physics 3600 - Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018 I. INTRODUCTION The laser was invented in May 1960 by Theodor Maiman.

More information

DESIGN OF SINGLE-BOUNCE MONOCAPILLARY X-RAY OPTICS

DESIGN OF SINGLE-BOUNCE MONOCAPILLARY X-RAY OPTICS 194 DESIGN OF SINGLE-BOUNCE MONOCAPILLARY X-RAY OPTICS Sterling Cornaby 2, Thomas Szebenyi 1, Rong Huang 3 and Donald H Bilderback 1,2 ABSTRACT 1 Cornell High Energy Synchrotron Source (CHESS) 2 School

More information

FIRST INDIRECT X-RAY IMAGING TESTS WITH AN 88-mm DIAMETER SINGLE CRYSTAL

FIRST INDIRECT X-RAY IMAGING TESTS WITH AN 88-mm DIAMETER SINGLE CRYSTAL FERMILAB-CONF-16-641-AD-E ACCEPTED FIRST INDIRECT X-RAY IMAGING TESTS WITH AN 88-mm DIAMETER SINGLE CRYSTAL A.H. Lumpkin 1 and A.T. Macrander 2 1 Fermi National Accelerator Laboratory, Batavia, IL 60510

More information

IBIL setup operation manual for SynerJY software version

IBIL setup operation manual for SynerJY software version IBIL setup operation manual for SynerJY software version 1.8.5.0 Manual version 1.0, 31/10/2008 Author: Carlos Marques Equipment Managers: Carlos Marques, +351219946084, cmarques@itn.pt Luís Alves, +351219946112,

More information

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES The current multiplication mechanism offered by dynodes makes photomultiplier tubes ideal for low-light-level measurement. As explained earlier, there

More information

Active Pixel Matrix for X-ray Satellite Missions

Active Pixel Matrix for X-ray Satellite Missions Active Pixel Matrix for X-ray Satellite Missions P. Holl 1,*, P. Fischer 2, P. Klein 3, G. Lutz 4, W. Neeser 2, L. Strüder 5, N. Wermes 2 1 Ketek GmbH, Am Isarbach 30, D-85764 Oberschleißheim, Germany

More information

Citation X-Ray Spectrometry (2011), 40(4): 2. Right final form at

Citation X-Ray Spectrometry (2011), 40(4): 2.   Right final form at TitleSi PIN X-ray photon counter Author(s) Nakaye, Yasukazu; Kawai, Jun Citation X-Ray Spectrometry (2011), 40(4): 2 Issue Date 2011-03-24 URL http://hdl.handle.net/2433/197743 This is the peer reviewed

More information

ON THE ENERGY DEPENDENCE OF THE DETECTOR EFFICIENCY OF A Si-PIN DIODE

ON THE ENERGY DEPENDENCE OF THE DETECTOR EFFICIENCY OF A Si-PIN DIODE Copyright(C)JCPDS-International Centre for Diffraction Data 2, Advances in X-ray Analysis, Vol.42 36 Copyright(C)JCPDS-International Centre for Diffraction Data 2, Advances in X-ray Analysis, Vol.42 36

More information

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26 Today s Outline - January 25, 2018 C. Segre (IIT) PHYS 570 - Spring 2018 January 25, 2018 1 / 26 Today s Outline - January 25, 2018 HW #2 C. Segre (IIT) PHYS 570 - Spring 2018 January 25, 2018 1 / 26 Today

More information

Administrative details:

Administrative details: Administrative details: Anything from your side? www.photonics.ethz.ch 1 What are we actually doing here? Optical imaging: Focusing by a lens Angular spectrum Paraxial approximation Gaussian beams Method

More information

Guide to SPEX Optical Spectrometer

Guide to SPEX Optical Spectrometer Guide to SPEX Optical Spectrometer GENERAL DESCRIPTION A spectrometer is a device for analyzing an input light beam into its constituent wavelengths. The SPEX model 1704 spectrometer covers a range from

More information

Lab Report XRF 441 Elemental distribution analysis on geological samples with the M4 TORNADO

Lab Report XRF 441 Elemental distribution analysis on geological samples with the M4 TORNADO Bruker Nano Spectrum Geological sample M4 TORNADO Quantification Lab Report XRF 441 Elemental distribution analysis on geological samples with the M4 TORNADO Geological samples are inhomogeneous. The distribution

More information

TOWARDS SUB-100 NM X-RAY MICROSCOPY FOR TOMOGRAPHIC APPLICATIONS

TOWARDS SUB-100 NM X-RAY MICROSCOPY FOR TOMOGRAPHIC APPLICATIONS Copyright -International Centre for Diffraction Data 2010 ISSN 1097-0002 89 TOWARDS SUB-100 NM X-RAY MICROSCOPY FOR TOMOGRAPHIC APPLICATIONS P. Bruyndonckx, A. Sasov, B. Pauwels Skyscan, Kartuizersweg

More information

Lesson 2 Diffractometers

Lesson 2 Diffractometers Lesson 2 Diffractometers Nicola Döbelin RMS Foundation, Bettlach, Switzerland January 14 16, 2015, Bern, Switzerland Repetition: Generation of X-rays / Diffraction SEM: BSE detector, BSED / SAED detector

More information

R AMP TEK Landed on Mars July 4, 1997 All Solid State Design No Liquid Nitrogen Be Window FET Detector Temperature Monitor Cooler Mounting Stud FEATURES Si-PIN Photodiode Thermoelectric Cooler Beryllium

More information

BEAM HALO OBSERVATION BY CORONAGRAPH

BEAM HALO OBSERVATION BY CORONAGRAPH BEAM HALO OBSERVATION BY CORONAGRAPH T. Mitsuhashi, KEK, TSUKUBA, Japan Abstract We have developed a coronagraph for the observation of the beam halo surrounding a beam. An opaque disk is set in the beam

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 76-700 Digital Subtraction Angiography Phantom Users Manual March 2005 Manual No. 76-700-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names

More information

Optical Module in KM3NeT

Optical Module in KM3NeT Photo-Sensors for a Multi-PMT. Optical Module in KM3NeT PMT tests improving collection efficiency (Winston cone), Q. Dorosti-Hasankiadeh, H. Löhner representing the KM3NeT Consortium, The Netherlands o.kavatsyuk@rug.nl

More information

Introduction to TOTEM T2 DCS

Introduction to TOTEM T2 DCS Introduction to TOTEM T2 DCS Leszek Ropelewski CERN PH-DT2 DT2-ST & TOTEM Single Wire Proportional Chamber Electrons liberated by ionization drift towards the anode wire. Electrical field close to the

More information

BIFOCAL MINIATURE TOROIDAL SHAPED X-RAY MIRRORS

BIFOCAL MINIATURE TOROIDAL SHAPED X-RAY MIRRORS 1 BIFOCAL MINIATURE TOROIDAL SHAPED X-RAY MIRRORS Sterling Cornaby 1,2, Detlef-M. Smilgies 2, and Donald H. Bilderback 1,2 1 Cornell High Energy Synchrotron Source (CHESS) 2 School of Applied and Engineering

More information

ADVANCED OPTICS LAB -ECEN Basic Skills Lab

ADVANCED OPTICS LAB -ECEN Basic Skills Lab ADVANCED OPTICS LAB -ECEN 5606 Basic Skills Lab Dr. Steve Cundiff and Edward McKenna, 1/15/04 Revised KW 1/15/06, 1/8/10 Revised CC and RZ 01/17/14 The goal of this lab is to provide you with practice

More information

Data Collection with. VÅNTEC-2000 Detector

Data Collection with. VÅNTEC-2000 Detector Data Collection with IµS Source and VÅNTEC-2000 Detector D8 System Configuration for Reflection Microfocus Source IµS Optics with Housing 2D Detector (VÅNTEC-2000) DHS 900 Heating Stage Sample Stage Bruker

More information

EUVL Activities in China. Xiangzhao Wang Shanghai Inst. Of Opt. and Fine Mech. Of CAS. (SIOM) Shanghai, China.

EUVL Activities in China. Xiangzhao Wang Shanghai Inst. Of Opt. and Fine Mech. Of CAS. (SIOM) Shanghai, China. EUVL Activities in China Xiangzhao Wang Shanghai Inst. Of Opt. and Fine Mech. Of CAS. (SIOM) Shanghai, China. wxz26267@siom.ac.cn Projection Optics Imaging System Surface Testing Optical Machining ML Coating

More information

Status of the PRad Experiment (E )

Status of the PRad Experiment (E ) Status of the PRad Experiment (E12-11-106) NC A&T State University Outline Experimental apparatus, current status Installation plan Draft run plan Summary PRad Experimental Setup Main detectors and elements:

More information

Optoelectronic Search Systems

Optoelectronic Search Systems ECNDT 2006 - Poster 56 Optoelectronic Search Systems Sergey V. SADKOV, Mikhail V. PONOMAREV, Alexey A. KOVALEV, Alexey V. KOVALEV, Association SPEKTR-GROUP, Moscow, Russia Abstract. One of the hottest

More information

Multianode Photo Multiplier Tubes as Photo Detectors for Ring Imaging Cherenkov Detectors

Multianode Photo Multiplier Tubes as Photo Detectors for Ring Imaging Cherenkov Detectors Multianode Photo Multiplier Tubes as Photo Detectors for Ring Imaging Cherenkov Detectors F. Muheim a edin]department of Physics and Astronomy, University of Edinburgh Mayfield Road, Edinburgh EH9 3JZ,

More information

Radial Polarization Converter With LC Driver USER MANUAL

Radial Polarization Converter With LC Driver USER MANUAL ARCoptix Radial Polarization Converter With LC Driver USER MANUAL Arcoptix S.A Ch. Trois-portes 18 2000 Neuchâtel Switzerland Mail: info@arcoptix.com Tel: ++41 32 731 04 66 Principle of the radial polarization

More information

Phase Contrast Imaging with X-ray tube

Phase Contrast Imaging with X-ray tube Phase Contrast Imaging with X-ray tube Institute for Roentgen Optics /IRO/, Moscow Vladimir Shovkun and Muradin Kumakhov Proc. SPIE v.5943, 2005 Institute for Roentgen Optics. Vladimir Ya. Shovkun. E-mail:

More information