Overview of Safety Code 35

Size: px
Start display at page:

Download "Overview of Safety Code 35"

Transcription

1 Common Quality Control Procedures for All s Quality Control Procedures Film All s Daily Quality Control Tests Equipment Warm-up (D1) According to manufacturers instructions Can include auto calibration(d1) Meters Operation (D2) Meter, visible and audible indicators should function Equipment Conditions (D3) Visual inspection for loose or broken components, ease of movements Darkroom Cleanliness (D5) (M DAP) Film Processor Function (D6) Overall Visual Assessment of Electronic Display Devices (D7) (M DAP) Display SMPTE or QC pattern for general image quality of all Radiologists workstations Quality Control Procedures All s Weekly Quality Control Tests View box condition (W2) Visual inspection for cleanliness, colour, illumination Print pattern such as SMPTE or PQC Laser Film Printer Operation (W3) Check for 0/5% and 95/100% patch visibility OD of 10% to 90% patches No artifacts or geometrical distortion Monthly Quality Control Tests Darkroom Temperature and Humidity (M2) Temp: C; Humidity 40-60% Darkroom Light Conditions (M3) Visual check for light tightness Film Processor Operation (M4) Temp ±0.5 C; Developer and fixer correct (M DAP) Electronic Display Device Performance (M6) Display pattern such as SMPTE or QC on all image display stations Laser Film Printer Operation (M7) As W3 plus measurement of optical density of the 10% to 90% grey scale

2 Quality Control Procedures All s Quarterly Quality Control Tests Interlocks (Q2) These are not usually used on diagnostic doors Annual Quality Control Tests Safelight Test (Y1) Expose film for 2 min View boxes (Y26) Check luminance, uniformity, homogeneity, ambient light Electronic Display Device All clinical workstations must be calibrated for Performance (Y27) luminance, distortion, resolution and noise Integrity of Protective Equipment Lead aprons, glasses, integral shields (Y28) General Preventive Maintenance As per manufacturer (Y29) Acceptance testing Baseline value determination Device use period Next constancy testing Data evaluation PASS Within the established criteria FAIL Remedy 2

3 Radiographic QC Tests Quality Control Procedures Film CR DR Comments Weekly Quality Control Tests Visual Inspection of Cleanliness of Imaging W1 W1 W1 Inspect screens/cr plates/dr housing s View boxes Condition W2 W2 W2 Cleanliness, luminance Laser Film Printer Operation W3 W3 Print pattern such as SMPTE Monthly Quality Control Tests Cassette, Screen, and Imaging Plate Cleaning M1 M1 Clean screens/cr plates/dr housing Darkroom Temperature and Humidity Conditions (DAP) M2 Temp: C; Humidity 40-60% Darkroom Light Conditions M3 Check for light tightness Film Processor Operation M4 Temp ±0.5 C; Developer and fixer correct Retake Analysis M5 M5 For film, collect discards; for CR use M5 Electronic Display Device Performance M6 M6 M6 Laser Film Printer Operation M7 M7 M7 manufacturers software; DR? Display pattern such as SMPTE on all image display stations Print pattern such as SMPTE Check for 0/5% and 95/100% patch visibility OD of 10% to 90% patches. No artifacts or geometrical distortion Annual QC Tests (1) Safelight Test Y1 Film/Screen Contact Y2 Accuracy of Loading Factors Y3 Y3 Y3 mas Linearity Radiation Output Reproducibility Y4 Y4 Y4 Reproducibility Radiation Output Linearity Y5 Y5 Y5 Output with mas X-ray Beam Filtration Y6 Y6 Y6 HVL Automatic Exposure Control Y7 Y7 Y7 Check AEC for all kvps and thicknesses X-ray Field and Light Field Alignment Y8 Y8 Y8 X-ray Beam Collimation Y9 Y9 Y9 Image Quality Grid Performance Exposure Index Noise, Uniformity and Image Artifacts Congruency of x-ray beam and light field edges Congruency of x-ray beam and light field centers Y10 Y10 Y10 Check uniformity and movement of grid Y12 Y12 Exposure Index versus Dose 1 to 50 mgy For above range of dose measure noise in Y14 Y14 center and each quadrant Check for artifacts 3

4 Quality Control Procedures Film CR DR Comments Spatial Resolution Y15 Y15 Y15 Line-pair or Leeds phantom Contrast Detectability Y16 Y16 Y16 Leeds phantom Digital Detector Residual Take image at 50 mgy then zero; Y17 Y17 Images check for artifacts Measure dose at surface of standard Phantom Dose Measurements Y18 Y18 Y18 phantom e.g. 20 cm PMMA DAP Meter Calibrate Radioscopic QC Tests Quality Control Procedures DF Radioscopic Weekly Quality Control Tests Visual Inspection of Cleanliness of Imaging s W1 W1 Laser Film Printer Operation W3 W3 Digital Subtraction Angiography Performance Use phantom to check consistency (W7) Quarterly Quality Control Tests Protective devices (Q6) Lead skirts, lead glass, etc Table angulation and motion (Q3) Compression devices (Q4) Timer (Q5) Annual Quality Control Tests Safelight Test Y1 Y1 Film/Screen Contact Y2 Y2 Accuracy of Loading Factors Y3 mas Linearity Radiation Output Reproducibility Y4 Reproducibility Radiation Output Linearity Y5 Output with mas X-ray Beam Filtration Y6 HVL X-ray Field and Light Field Alignment Y8 Congruency of x-ray beam and light field edges X-ray Beam Collimation Y9 Congruency of x-ray beam and light field centers Phantom Dose Measurements Y18 Measure dose at surface of standard phantom e.g. 20 cm PMMA Typical Image Intensifier Air Kerma Rate Y19 e.g. 20 cm PMMA, Al or Cu Maximum Image Intensifier Air Kerma Rate Y20 With detector blocked by lead Automatic Intensity Control Y21 Tracking of detector dose with phantom thickness 4

5 Quality Control Procedures DF Radioscopic Annual Quality Control Tests Image Quality Grid Performance Y10 Check uniformity and movement of grid Exposure Index Noise, Uniformity and Image Artifacts Spatial Resolution Contrast Detectability Digital Detector Residual Images Quality Control Procedures Weekly QC Tests CT Number Accuracy CT Noise CT Uniformity Monthly QC Tests Electronic Display Devices Performance CT Tomographic Section Thickness Y12 Exposure Index versus Dose 1 to 50 mgy For above range of dose measure noise in center Y14 and each quadrant Check for artifacts Y15 Line-pair or Leeds phantom Y16 Leeds phantom Take image at 50 mgy then zero; Y17 check for artifacts Computed Tomography QC Tests CT s Check CT number water 0 ±4 HU (W4) Image noise in center of water phantom ±10% from baseline value (W5) Check CT number in center and 4 quadrants ±10% from baseline value (W6) All device used to display digital images use pattern SMPTE Slice thickness should be ±0.5 mm from baseline value (M8) Check CT number water 0 ±4 HU and air 1000 ±10 HU Calibration of CT Number (M9) CT Number Linearity Check CT number over CT range to (M10) Quarterly QC Tests Interlocks (Q2) Error? Interlocks are seldom used on CT scanner doors CT Patient Support Check table movement corresponds to digital display ±1 mm Movement (Q8) CT Spatial Resolution Measure MTF or line pair phantom (Q9) CT Low Contrast Detectability Q10 Semi-annual QC Tests CT Laser Light Accuracy Check laser light vs. X-ray beam with phantom(sy1) CT Accuracy of Automatic Positioning of Tomographic Check localization scan corresponds to digital display ±2 Plane (using the scanned mm (SY2) projection radiograph) CT Accuracy of Gantry Tilt SY3 CT Patient Dose Check CTDI ±20% from baseline values (SY4) 5

6 Quality Control Procedures Annual QC Tests CT Number Dependence on Phantom Position CT Radiation Dose Profile CT Radiation Dose Scout Localisation Image General Preventive Maintenance CT s Check CT number water 0 ±5 HU for possible patient positions in the gantry(y23) Y24 Radiation Dose for Localization Image within 20% of baseline value (Y25) Biomed PMs must be performed (Y29) Tests for Acceptance Testing Acceptance Testing Filmbased CR DR Radioscopic CT 1.0 Identification 1.1 Initial Inspection and Inventory X X X X X 1.2 Inspection of Documentation X X X X X 2.0 Visual and Functional Texts 2.1 Mechanical Properties X X X X X 2.2 Safety s X X X X X 3.0 Performance Evaluation (X-ray Generator) 3.1 X-ray Tube Voltage X X X X X 3.2 Current Time Product X X X 3.3 Loading Time X X X 3.4 Radioscopic Timer and Chronometer X 3.5 Beam Limitation and Indication X X X X X 3.6 X-ray Beam Filtration X X X X 3.7 Automatic Exposure Control X X X 3.8 Automatic Intensity Control X X 3.9 Radiation Output X X X X X 3.10 Maximum Air Kerma Rate X 3.11 Phantom (Entrance) Dose (Rate) X X X X X 4.0 Performance Evaluation (Image Receptor) 4.1 Exposure Index X X 4.2 Dynamic Range X X X 4.3 Spatial Resolution X X X X X 4.4 Contrast Detectability X X X X X 4.5 Uniformity X X X 4.6 Artifacts X X X X X 4.7 Digital Detector Residual Image X X 4.8 Noise X X X 6

7 Tests for Acceptance Testing Filmbased CR DR Radioscopic CT 4.9 Modulation Transfer Function X X 4.10 Mean CT Numbers (calibration and linearity) X 4.11 Tomographic Section Thickness X 4.12 Axial Patient Positioning Accuracy X 4.13 Positioning of Patient Support X Radiographic Exam Projection Diagnostic Reference Levels DRL Gy.cm 2 Generalized Conversion Coefficients DAP to Effective Dose msv per Gy.cm2 Skull PA Lat AP C Spine PA 0.2 Lat 0.03 Shoulder AP 0.04 Chest PA Lat AP 0.2 T Spine Lat AP L Spine Lat AP LSJ Lat Kidneys PA 0.1 AP 0.2 Abdomen PA AP R hip AP 0.15 L hip AP 0.2 Pelvis AP

8 Complete Exam Fluoroscopy DRL Gy.cm 2 DAP to E Time minutes Barium Swallow Barium Meal Ba follow thru Barium enema Femoral angio Hysterosalpingiography 3 1 IVP MCU Nephrostography Retro Pyelogram Sialography T tube cholangiogram Sinography Small bowel enema Venography leg Coronary angiography Fistulography Complete Exams Fluoroscopy DRL Gy.cm 2 DAP to E Time minutes Biliary Drainage Facet joint injection Hickman line Nephrostomy Esophageal stent Esophageal dilation Pacemaker PTCA single stent

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook DIAGNOSTIC ACCREDITATION PROGRAM Radiology and CT Quality Control Procedures Workbook Quality Control Procedures Radiography/CR/DR Safety Code 35 Summary For more detail about each quality control (QC)

More information

REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS

REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS DEPARTMENT OF HEALTH DIRECTORATE: RADIATION CONTROL Implementation date: 31 March 2009 Contents

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

Test Equipment for Radiology and CT Quality Control Contents

Test Equipment for Radiology and CT Quality Control Contents Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3

More information

NJDEP Medical Physicist s Radiographic QC Survey Registration Number:

NJDEP Medical Physicist s Radiographic QC Survey Registration Number: Facility Name NJDEP ID # NJDEP Medical Physicist s Radiographic QC Survey PLEASE PRINT Facility Information Unit Information Manufacturer Model Console Model # Console serial # Tube serial # Location (room)

More information

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Melissa Martin, M.S., FAAPM, FACR, FACMP AAPM Annual Meeting - Philadelphia July 19, 2010 MO-B-204C-1 Educational Objectives

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

Quality Control of Full Field Digital Mammography Units

Quality Control of Full Field Digital Mammography Units Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc Quality Control for Stereotactic Breast Biopsy Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc. 716-924-0350 Methods of Imaging Guided Breast Biopsy Ultrasound guided, hand-held needle

More information

X-RAYS - NO UNAUTHORISED ENTRY

X-RAYS - NO UNAUTHORISED ENTRY Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-649 CDRH Fluoroscopic Phantom Users Manual March 2005 Manual No. 07-649-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

Aspire HD. Program Manual. 2nd Edition - October 2012

Aspire HD. Program Manual. 2nd Edition - October 2012 Quality Control 1 Aspire HD Quality Control Program Manual 2nd Edition - October 2012 Overview Installation of FDR Mammography QC Program Weekly Test 2 3 4 Quarterly Test 5 Semi-annual Test 6 Annual Test

More information

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS Veterinary Science Preparatory Training for the Veterinary Assistant Floron C. Faries, Jr., DVM, MS Radiology Floron C. Faries, Jr., DVM, MS Objectives Determine the appropriate machine settings for making

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-647 R/F QC Phantom Operators Manual March 2005 Manual No. 07-647-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. All product names are trademarks of their respective companies

More information

RADIATION SAFETY DIRECTORATE

RADIATION SAFETY DIRECTORATE RADIATION SAFETY DIRECTORATE Pursuant to Article 26-e, paragraph 1, item 13 of the Law on Ionising Radiation Protection and Radiation Safety (Official Gazette of the Republic of Macedonia No. 48/02 and

More information

CR Basics and FAQ. Overview. Historical Perspective

CR Basics and FAQ. Overview. Historical Perspective Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors

More information

Introduction. Digital Mammography QA: Comparing the Manufacturers Recommendations. What is QC and why is it important? Review & compare QC tests

Introduction. Digital Mammography QA: Comparing the Manufacturers Recommendations. What is QC and why is it important? Review & compare QC tests Slide 1 Digital Mammography QA: Comparing the Manufacturers Recommendations Eric A. Berns, Ph.D. Slide 2 Introduction What is QC and why is it important? Review & compare QC tests Key take home points

More information

Photomultiplier Tube

Photomultiplier Tube Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or

More information

Exposure System Selection

Exposure System Selection Principles of Imaging Science II (RAD120) Exposure Systems Exposure System Selection Radiographic exposure is a very complex process Best technique systems manipulate one variable while holding others

More information

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II APPROVED 12/O2/2011 EFFECTIVE SPRING 2013-14 Prefix & Number RAD 150 Course Title: Radiologic Exposure Technique II & Lab Purpose of this submission: New Change/Updated

More information

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS Course Number: RADG 112 Department: Radiography Course Title: Image Production & Eval. Semester: Spring Year: 1997 Objectives/ Unit One: Introduction

More information

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT

More information

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations S. Baechler 1, P. Monnin 1, A. Aroua 1, J.F. Valley 1, M. Perrier, P. Trueb 3, F.R. Verdun 1 1 University

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura New Technology https://www.youtube.com/watch?v=ptkzznazb 7U COMPUTED

More information

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S.

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S. Practical Medical Physics Session: TG-151 Dose Monitoring August 5, 2013 Katie Hulme, M.S. Digital Imaging and Dose Creep Images courtesy of Agfa Healthcare Under-Exposed Over-Exposed Freedman et al.,

More information

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on

More information

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893 Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -

More information

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey Implementation of the 2016 ACR Digital Mammography QC Manual Medical Physicist Mammography Equipment Evaluation and Annual Survey Eric A Berns, PhD, FACR Radiologist Responsibilities Radiologist Responsibilities

More information

Compliance Guidance for RADIOGRAPHIC QUALITY CONTROL (5 th Edition)

Compliance Guidance for RADIOGRAPHIC QUALITY CONTROL (5 th Edition) Compliance Guidance for RADIOGRAPHIC QUALITY CONTROL (5 th Edition) New Jersey Department of Environmental Protection Bureau of X-ray Compliance PO Box 420 Mail Code 25-01 Trenton, NJ 08625-0420 FAX 609-984-5811

More information

Calculati tions - BIR meth thod and NCRP 147

Calculati tions - BIR meth thod and NCRP 147 Calculations l - BIR method and NCRP 147 David Sutton PhD 1. Cath Lab Indicative of all C-arm applications Only secondary radiation Primary Beam is stopped by the image intensifier Geometry BIR 9.5 m Preparation

More information

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation AAPM Annual Clinical Meeting Indianapolis, IN August 5, 2013 Learning Objectives Become familiar with the recommendations and

More information

Essentials of Digital Imaging

Essentials of Digital Imaging Essentials of Digital Imaging Module 7 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 7 Quality 1. ASRT Animation 2. Welcome Welcome to the Essentials of Digital Imaging:

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

Luminos RF Classic. Where value meets performance.

Luminos RF Classic. Where value meets performance. Luminos RF Classic Where value meets performance www.siemens.com/healthcare What s good value in fluoroscopy? That s easy. Luminos RF Classic. 2 Whether for its handling convenience, outstanding image

More information

Half value layer and AEC receptor dose compliance survey in Estonia

Half value layer and AEC receptor dose compliance survey in Estonia Half value layer and AEC receptor dose compliance survey in Estonia K. Kepler, A. Vladimirov Training Centre of Medical Physics, University of Tartu Testing Centre of the University of Tartu, Estonia E-mail:

More information

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter Revision Date: January 5th, 2017 www.disc-imaging.com Table of Contents Section A: Preliminary Setup Requirements... 4 Tools

More information

Diagnostic X-Ray Shielding

Diagnostic X-Ray Shielding Diagnostic X-Ray Shielding Multi-Slice CT Scanners Using NCRP 147 Methodology Melissa C. Martin, M.S., FAAPM, FACR Therapy Physics Inc., Bellflower, CA AAPM Annual Meeting, Orlando, FL FL Refresher Course

More information

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org INTRODUCTION

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced

More information

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology CODE: RADT 156 INSTITUTE: Health Science TITLE: Equipment Operation I DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of equipment operation and maintenance of radiographic

More information

MILADY. Product Data. Page 1 of 8

MILADY. Product Data. Page 1 of 8 Page 1 of 8 The MILADY Mammographic Unit offers the best quality-to-price ratio to our customers worldwide. The unit advanced technology together with the application of industrial production standards,

More information

Ansur TNT Users Manual. Plug-In

Ansur TNT Users Manual. Plug-In Ansur TNT 12000 Plug-In Users Manual August 2009, Rev. 2, 12/09 2009 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their

More information

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring

More information

Teaching Digital Radiography and Fluoroscopic Radiation Protection

Teaching Digital Radiography and Fluoroscopic Radiation Protection Teaching Digital Radiography and Fluoroscopic Radiation Protection WCEC 20 th Student Educator Radiographer Conference Dennis Bowman, RT(R), CRT (R)(F) Community Hospital of the Monterey Peninsula (CHOMP)

More information

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging The Physics of Radiation Oncology X-ray Imaging Charles E. Willis, Ph.D. DABR Associate Professor Department of Imaging Physics The University of Texas M.D. Anderson Cancer Center Houston, Texas Overview

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-644 Grid Alignment Test Tool Users Manual March 2005 Manual No. 07-644-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

MAMMOGRAPHY - HIGH LEVEL TROUBLESHOOTING

MAMMOGRAPHY - HIGH LEVEL TROUBLESHOOTING MAMMOGRAPHY - HIGH LEVEL TROUBLESHOOTING Maynard High New York Medical College SS2001-M.High 1 Objectives: Review MQSA and ACR annual QC tests as opportunities for troubleshooting before a significant

More information

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development Teaching Radiographic Technique in a Digital Imaging Paradigm Objectives 1. Discuss the historical development of digital imaging. Dawn Couch Moore, M.M.Sc., RT(R) Assistant Professor and Director Emory

More information

LECTURE 1 The Radiographic Image

LECTURE 1 The Radiographic Image LECTURE 1 The Radiographic Image Prepared by:- KAMARUL AMIN ABDULLAH @ ABU BAKAR UiTM Faculty of Health Sciences Medical Imaging Department 11/23/2011 KAMARUL AMIN (C) 1 Lesson Objectives At the end of

More information

Renato Padovani, University Hospital, Udine, Italy

Renato Padovani, University Hospital, Udine, Italy DATAMED II Training Course Sofia, May 19-20, 2011 Patient dose assessment in fluoroscopy and interventional procedures Renato Padovani, University Hospital, Udine, Italy Content Fluoroscopy guided diagnostic

More information

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016 Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016 Current technology in digital image production (CR/DR and other modalities) 2/ Overview Digital

More information

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. RA110 test 3 Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. 1. An object 35 cm in width is radiographed at 100 cm SID and at a 50 cm SOD. What

More information

NATIONWIDE EVALUATION OF X-RAY TRENDS (NEXT) TABULATION AND GRAPHICAL SUMMARY OF THE 1999 DENTAL RADIOGRAPHY SURVEY

NATIONWIDE EVALUATION OF X-RAY TRENDS (NEXT) TABULATION AND GRAPHICAL SUMMARY OF THE 1999 DENTAL RADIOGRAPHY SURVEY CRCPD Publication E-3-6-a Available Online at No Charge $15. for a Computer-Generated Copy NATIONWIDE EVALUATION OF X-RAY TRENDS (NEXT) TABULATION AND GRAPHICAL SUMMARY OF THE 1999 DENTAL RADIOGRAPHY SURVEY

More information

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

More information

diagnostic examination

diagnostic examination RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011

More information

Features and Weaknesses of Phantoms for CR/DR System Testing

Features and Weaknesses of Phantoms for CR/DR System Testing Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal

More information

Unfors EDD-30 Radiation Protection in Fluoroscopy

Unfors EDD-30 Radiation Protection in Fluoroscopy Unfors EDD-30 Radiation Protection in Fluoroscopy Immediate Warning Decrease Your Dose Interventional radiology procedures are considered to be essential to medical diagnosis and treatment. It is recognized,

More information

Acquisition, Processing and Display

Acquisition, Processing and Display Acquisition, Processing and Display Terri L. Fauber, R.T. (R)(M) Department of Radiation Sciences School of Allied Health Professions Virginia Commonwealth University Topics Image Characteristics Image

More information

Learning Objectives: What s my motivation? (unknown screen actor) Workshop Overview

Learning Objectives: What s my motivation? (unknown screen actor) Workshop Overview Practical Medical Physics Adapting Traditional Clinical Medical Physics to Digital Radiography Charles E. Willis, Ph.D., DABR Associate Professor Department of Imaging Physics The University of Texas M.D.

More information

Interventional Radiological Equipment selection and installation

Interventional Radiological Equipment selection and installation Interventional Radiological Equipment selection and installation Renato Padovani ICTP Learning objectives To understand the main components of an interventional radiology equipment To understand the relevance

More information

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist Topics Image Uniformity and Artifacts Image Quality - Detail

More information

Visibility of Detail

Visibility of Detail Visibility of Detail Radiographic Quality Quality radiographic images represents the, and information is for diagnosis. The of the anatomic structures and the accuracy of their ( ) determine the overall

More information

QC by the MPE in Belgium

QC by the MPE in Belgium Acceptance testing of state-of-the-art CT scanners using a new national protocol: first experience on a large number of scanners of different make and model the working group Radiology of the Belgian Hospital

More information

Ludlum Medical Physics

Ludlum Medical Physics Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to

More information

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System Manus

More information

Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2

Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2 Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2 Medical Consult, SA; Establish a method to correlate image

More information

CHAPTER 6 QC Test For Fluoroscopic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

CHAPTER 6 QC Test For Fluoroscopic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC CHAPTER 6 QC Test For Fluoroscopic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done. Identify QC

More information

Beam-Restricting Devices

Beam-Restricting Devices Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the

More information

T h e P h a n t o m L a b o r a t o r y

T h e P h a n t o m L a b o r a t o r y T h e P h a n t o m L a b o r a t o r y 1 CCT228 ATCM Phantom Manual Copyright 2017 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order

More information

Nuclear Associates , , CT Head and Body Dose Phantom

Nuclear Associates , , CT Head and Body Dose Phantom Nuclear Associates 76-414,76-414-4150,76-415 CT Head and Body Dose Phantom Users Manual March 2005 Manual No. 76-414-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product

More information

Contrast. Contrast: the difference in density on adjacent areas of a radiograph or other image receptor. Subjective. Long Scale (Low Contrast)

Contrast. Contrast: the difference in density on adjacent areas of a radiograph or other image receptor. Subjective. Long Scale (Low Contrast) Contrast Contrast: the difference in density on adjacent areas of a radiograph or other image receptor. Subject Subjective Radiographic Long Scale (Low Contrast) Short Scale (High Contrast) Factors affecting

More information

A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations. Shirley Weddle, RT(R)(M), CIIP, BBA

A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations. Shirley Weddle, RT(R)(M), CIIP, BBA A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations Shirley Weddle, RT(R)(M), CIIP, BBA OBJECTIVES Define Computed Radiography (CR) Discuss CR vendor

More information

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. RA202 image production class two Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. 1. What removes excess chemistry from the film prior to it

More information

PATIENT EFFECTIVE DOSES IN DIAGNOSTIC RADIOLOGY, NA

PATIENT EFFECTIVE DOSES IN DIAGNOSTIC RADIOLOGY, NA Title of Paper: Patient effective doses in diagnostic radiology Authors: N.A. Gkanatsios, and W. Huda * Corresponding Author: Department of Radiology, University of Florida, P.O. Box 100374, Gainesville,

More information

Mammography: Physics of Imaging

Mammography: Physics of Imaging Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness

More information

Digital Mammography Quality Control for the Mammographic Technologist

Digital Mammography Quality Control for the Mammographic Technologist Ontario Breast Screening Program Digital Mammography Quality Control for the Mammographic Technologist Authors: G.E. Mawdsley, A.K. Bloomquist, M.J. Yaffe October 2011 Revision 3.1 Mammographic Physics

More information

2217 US Highway 70 East Garner, NC Main: Fax:

2217 US Highway 70 East Garner, NC Main: Fax: Viztek is committed to providing the highest image quality possible in our CR & DR product lines. There are several factors that directly affect the overall quality of CR & DR based images. The eposure

More information

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of radiation to the population due to Medical Imaging

More information

Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision Minnesota Rules, Chapter 4732 X-ray Revision DRAFT FLUOROSCOPIC X-RAY SYSTEMS, 1.0 Subpart 1. Applicability. Subpart 2. Limitation of the useful beam. Subpart 3. Measuring compliance; primary protective

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 76-700 Digital Subtraction Angiography Phantom Users Manual March 2005 Manual No. 76-700-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names

More information

biij Optimisation in general radiography CJ Martin, PhD, FIPEM, FioP Biomedical Imaging and Intervention Journal REVIEW PAPER

biij Optimisation in general radiography CJ Martin, PhD, FIPEM, FioP Biomedical Imaging and Intervention Journal REVIEW PAPER Available online at http://www.biij.org/2007/2/e18 doi: 10.2349/biij.3.2.e18 biij Biomedical Imaging and Intervention Journal REVIEW PAPER Optimisation in general radiography CJ Martin, PhD, FIPEM, FioP

More information

Image Display and Perception

Image Display and Perception Image Display and Perception J. Anthony Seibert, Ph.D. Department of Radiology UC Davis Medical Center Sacramento, California, USA Image acquisition, display, & interpretation X-rays kvp mas Tube filtration

More information

of sufficient quality and quantity

of sufficient quality and quantity of sufficient quality and quantity The patient s body attenuates the beam as it passes though the body More energy is deposited in organs located near the entry of the beam than near the exit of the beam

More information

CHAPTER 6 QUALITY ASSURANCE OF VARIAN ON-BOARD IMAGER

CHAPTER 6 QUALITY ASSURANCE OF VARIAN ON-BOARD IMAGER 127 CHAPTER 6 QUALITY ASSURANCE OF VARIAN ON-BOARD IMAGER 6.1 INTRODUCTION Accurate and repeatable setup of patients is a requisite in radiotherapy. In the treatment of head-and-neck tumors, accurate setup

More information

The importance of radiation quality for optimisation in radiology

The importance of radiation quality for optimisation in radiology Available online at http://www.biij.org/2007/2/e38 doi: 10.2349/biij.3.2.e38 biij Biomedical Imaging and Intervention Journal COMMENTARY The importance of radiation quality for optimisation in radiology

More information

STUDENT REVIEW QUESTION SET I RANDOM CONTENT AREA RADT 2913 COMPREHENSIVE REVIEW

STUDENT REVIEW QUESTION SET I RANDOM CONTENT AREA RADT 2913 COMPREHENSIVE REVIEW STUDENT REVIEW QUESTION SET I RANDOM CONTENT AREA RADT 2913 COMPREHENSIVE REVIEW 1 What is the most probable source for the radiographic artifact which appears as high-density artifacts resembling lightning

More information

Open. the Digitized world. Fuji Computed Radiography

Open. the Digitized world. Fuji Computed Radiography Open the Digitized world Fuji Computed Radiography If just one of these applies to you... Managing developing fluid is hard and darkroom work is a hassle... Images are not stable... Isn t digitalization

More information

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah X-RAY IMAGING EE 472 F2017 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Stewart C. Bushong, Radiologic Science for Technologists: Physics, Biology, and Protection, 10 th ed., Mosby,

More information

FOUR CATEGORIES OF SAFETY

FOUR CATEGORIES OF SAFETY OCTOBER 2013 FOUR CATEGORIES OF SAFETY DOSIMETRY PERSONAL SAFETY EQUIPMENT EQUIPMENT KNOWLEDGE PHYSICAL SAFETY DOSIMETRY THERMAL LUMINISCENT DEVICES AND FILM BADGES CNSC PERMISSIBLE DOSES WHOLE BODY DOSE

More information

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

Unit thickness. Unit area. σ = NΔX = ΔI / I 0 Unit thickness I 0 ΔI I σ = ΔI I 0 NΔX = ΔI / I 0 NΔX Unit area Δx Average probability of reaction with atom for the incident photons at unit area with the thickness of Delta-X Atom number at unit area

More information