Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations

Size: px
Start display at page:

Download "Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations"

Transcription

1 Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations S. Baechler 1, P. Monnin 1, A. Aroua 1, J.F. Valley 1, M. Perrier, P. Trueb 3, F.R. Verdun 1 1 University Institute of Applied Radiation Physics, Grand-Pré 1, CH-1007 Lausanne, Switzerland sebastien.baechler@hospvd.ch University Dental General Hospital of Lausanne, CH-1011 Lausanne, Switzerland 3 Swiss Federal Office of Public Health, CH-3003 Bern, Switzerland Abstract. The purpose of this paper is to present a comparison of doses delivered to the patients when dealing with intra-oral, panoramic and CT dental examinations. The technology progress will be discussed from a dose and an image quality point of view. DAP values were measured for two intra-oral systems using speed D and E/F dental films, and a digital system in two different acquisition modes (CCD technology). For each system the DQE was also assessed. DWP and DAP were measured on three panoramic systems of various generations of units. Finally, CTDI w and DLP were measured using a standard CT (Dentascan) and a new tomographic system (NewTom) in order to assess the dose delivered when planning dental implants. The use of an E/F system instead of the D system allowed to reduce the DAP by a factor of (respectively 56 and mgy cm ) without significant loss of image quality. A further dose reduction by a factor of four was possible with digital systems but with a significant loss of spatial resolution. The DAP measured on OPG systems showed that the oldest system (0 years old) delivered a higher dose than a more recent system (respectively 0. and 0.07 Gy cm ). Finally, DLP values of mgy cm and 55 mgy cm were obtained with respectively the standard CT acquisition and a new tomographic system designed specifically for dental radiology. The introduction of digital detectors has a significant potential in dose reduction. Nevertheless, this will impose to use intra-oral units that handle short exposure times. The new tomographic technology enables to reduce the dose delivered to the patient by a factor of when compared with Dentascan. 1 Introduction Dental radiology belongs to the most common low-dose examinations and affects predominantly young subjects. Optimization efforts are permanently carried out by introducing new technologies in the field of intra-oral digital radiology and tomography by the advent of dedicated system. Thus, comparing different products available in the market enables to evaluate the potential dose reduction using these new devices. For this purpose, D and E/F films are compared with a digital detector in order to assess the dose delivered to the patient as well as the image quality. In addition, dose measurements are performed for extra-oral examinations: panoramic (orthopantomograms - OPG), dedicated and conventional CT systems. In radiodiagnostic, monitoring the dose delivered to the patient is generally performed using dosimetric indicators, such as the entrance skin dose (ESD) and the dose area product (DAP). These values are easily measurable and enable to estimate the effective dose delivered to a «standard» patient. In addition, TLD measurements are used to determine dose distribution in the region under examination. Materials and methods.1 Intra-oral detection systems.1.1 Installation and beam quality Measurements were performed on a conventional Toshiba radiological installation. The beam quality was selected according to the ISO standard 5799:1991(E). The total filtration of the tube was 7.0 ± 0.5 mm of 99.9% pure aluminium (internal filtration of.5 mm Al + additional filtration of.5 mm Al). A high voltage of 50 kv was used so that the half-value layer (HVL) of the beam was 3.0 ± 0. 1

2 mm equivalent aluminium. For Ultra-speed and Insight films, two product tube current time values of 5, respectively 18, were selected. Similarly, two product tube current time values of 10 and.8 were used in high-resolution, respectively high-sensitivity mode of the detector based on a CCD technology..1. Detectors Each detector was characterized by the sensitivity, the Modulated Transfer Function (MTF) and the Detection Quantum Efficiency (DQE). The intra-oral detection systems used in the present study were the followings: Films Two types of films were characterized: - Kodak Ultra Speed DF 58 / batch Class D - Kodak Insight IP 1 / batch Class E or F Film responses were determined at three dose levels leading to optical density (OD) values of 0.8, 1. and.3. The optical density of 1. corresponds to a standard level of film exposure whereas optical densities of 0.8 and.3 correspond respectively to slightly under and over-exposed levels. Digital Detector The RVG Trophy system includes an indirect CCD digital detector having a sensitive area of 36 x 6.5 mm covered by a matrix of 180 x 1360 leading to a pixel size of 19.5-µm. The detector has a dynamic range of 1 bits (096 gray-levels per pixel). The software performs data compression on 8 bits as well as logarithmic image correction. The detector can be used in high-resolution (HR) or highsensitivity (HS) mode. For both acquisition modes, three dose levels leading to background gray levels of 0, 55 and 10 were characterized. These values correspond to a realistic use of the detector (50), as well as under and over-exposed levels..1.3 Dose measurements The ESD and DAP indicators were measured using a Radcal dosimeter (Radcal 3036) connected to a 11 cm 3 ionizing chamber. The conversion factors to determine the effective dose for a skull region exposure, at 70 kv and using a.5-mm thick aluminium filtration, are 0.09, 0.0 and 0.07 msv Gy -1 cm - for ante-posterior, post-anterior and lateral incidences according to Ref. [1].. Extra-oral detection systems..1 Detectors OPG Two conventional OPG installations and one digital OPG system based on the CCD technology have been included in this study: - OPG 1 : Philips Ortho-oralix - monopulse DFF 7 cm (0 years old) - OPG : Soradex Cranex dc monopulse DFF 50 cm (8 years old) - OPG 3 : Planmeca PM00cc Proline monopulse DFF 50 cm (recent installation) Tomographic system Dedicated CT system : Newtom DVT QR This installation does not only enable to take projection radiographs similar to images obtained by OPG, but also tomograms that are used for maxillo-facial surgery planning. The acquisition is performed using a "cone beam" geometry as opposed to the traditional fan beam geometry used in CT. The field dimension in the axial direction is 130 mm with an opening angle of 1. The focus-skin distance is about 30 mm and the distance between the focus and the rotation axis is about 680 mm. The detector uses a /3 CCD sensor (.7 x 7.1 cm ).

3 Conventional CT system. The 8-slice CT scanner used for the present study was the GE LightSpeed. Measurements were realized considering classical parameters for dental tomographic examinations (Dentascan)... Dose measurements The NRPB proposes the use of the dose width product (DWP) as a dose indicator when dealing with OPG examinations. In this study DWP were measured using a pencil ionization chamber connected to a Radcal dosimeter (Radcal 1515) [3]. The length of the RX beam was measured using a radiological film place in a light tight envelope. The conversion factor of 0.06 msv Gy -1 cm - recommended by the NRPB for OPG systems was used to calculate the effective dose from the DAP []. Entrance distribution was also evaluated using TLD measurements. In this case, bands of 6 TLD chips were positioned on an anthropomorphic skull phantom to determine ESD values. In the present work, dose distributions were determined for the OPG device as well as the Newtom DVT and Dentascan tomographic installations. 3 Results and discussion 3.1 Intra-oral detection systems Characteristic curves Film Characteristic curves of both Ultra-speed and Insight Kodak films are showed in Figure 1 for automatic (warm) and manual (cold) development conditions. Sensitivity and average contrast values extracted from the curves are given in Table 1. 5 Automatic development ULT Film IN Film Manual development ULT Film IN Film Optical density Exposure [µgy] FIG. 1 Characteristic curves of Ultra-speed (ULT) and Insight (IN) Kodak films. Sensitivity values for both films are in good agreement with classes, D respectively E/F, indicated by the manufacturer. Sensitivity loss due to manual development is higher than 30 % for the Insight system but is negligible for the Ultra-speed system. The Insight film is. times more sensitive than the Ultra-speed film. The average contrast of both types of films is nearly identical (.05). 3

4 Table I. Characteristic values of the Ulta-speed and Insight films. * Film Development Base + fog Dose at a net OD of 1 [µgy] Sensitivity [mgy -1 ] Average contrast ULT Manual Automatic IN Manual Automatic * Comparable results were obtained when using a standard intra-oral unit with a high voltage value of 70 kv. Digital detector Detector responses (56 gray-levels for 8 bits images) as function of the dose are showed in Figure for both high resolution and high sensitivity modes. Interpolation curves were used to linearize the images. The 6 characterization points corresponding to three levels of gray at 0, 55 and 10 are also reported on the graph. The sensitivity difference between HS and HR acquisition modes is obvious. The patient exposure is saved by a factor 5 for an image taken under standard conditions (background gray level 50 and 100) using the HS mode RVG Trophy high-resolution mode (HR) high-sensitivity mode (HS) characterization points Grey level Dose [µgy] FIG. Response curves of the RVG Trophy in high-resolution (HR) and high-sensitivity (HS) modes Sensitivity and dose indicators The sensibility was defined here as the inverse dose value at the detector entrance corresponding to a background gray level of 55 (standard clinical gray level). The DAP was calculated by multiplying the ESD and the surface area of the maximum acceptable exposure field corresponding to a diameter of 6 cm. The conversion factor used to determine the effective dose is about 0.03 msv Gy -1 cm -. The Insight film is. times more sensitive than the Ultra-speed. An image taken with the digital detector using the HR mode require a 6 times less exposure than when using an Insight Kodak film. Choosing the HS mode instead of the HR mode makes it possible to save an additional exposure factor of 5 (see Table II) Spatial resolution The Figure 3 shows the MTFs of the detectors under study. Both the 0-µm-pixel size of and the scintillation detection method restrict the resolution of the digital detector compared to the films. At 5 mm -1, the MTF of the films is still close to 90 % while the MTF of the Trophy system has already

5 dropped to about 55 % in HR mode. Using the HS mode reduces significantly the resolution in comparison with the HR mode. Table II. Sensitivity and dose area product (DAP) of intra-oral detection systems. Detection system Sensibility [mgy -1 ] Entrance Skin Dose [mgy] Dose Area Product [mgy cm ] Film Kodak Ultra-speed (D) Film Kodak Insight (E/F) RVG Trophy mode HR RVG Trophy mode HS * E=effective dose using the factor 0.03 msv Gy -1 cm -. E * µsv Film IN Film ULT Trophy mode HR Trophy mode HS 0.6 MTF Spatial Frequency [mm -1 ] FIG. 3 Comparison of MTFs for intra-oral detectors Detection quantum efficiency The figure compares the DQE of intra-oral detectors for a usual exposure level (optical density of 1. for films and gray level of 55 for digital images). Since there is no intensifying screen with intraoral dental films, the sensitivity and the DQE are lower compared to the digital detector. With comparable noise levels at low frequencies and. times higher sensitivity than the Ultra-speed film, the Insight film gives a. times better DQE (. % vs. 1%). Because of its high sensitivity, the digital detector provides a DQE definitely higher than films (15.7 % for the HR mode and 1.8 % for the HS mode). The high value obtained for the HS mode is also explained by a more effective detection process. However, due to its lower resolution, the DQE advantage of the digital detector is limited to detection at low and intermediate frequencies (up to 10 mm -1 ). For the same reasons, the DQE improvement using the HS mode compared to the HR mode of the Trophy system is restricted to low frequencies up to.5 mm -1. 5

6 1 8 6 Film ULT Film IN Trophy mode HR Trophy mode HS DQE Spatial Frequency [mm -1 ] FIG. Comparison of DQE for intra-oral dental detectors Extra-oral detection systems 3..1 OPG dose indicator The results obtained for the three OPG are summarized in Table III. The DAP obtained for the two most recent systems are about times smaller than the value of the 0 years old device, and in good agreement with the NRPB reference values. Installations Table III. DWP and DAP values as well as effective doses for the three OPG. Voltage kv Current ma Duration s DWP mgy mm Length/Width Slit (mm) DAP E ** Gy cm µsv OPG 1 (0 years old) / OPG (8 years old) / OPG 3 (recent) / Reference * 65 / /0.1 * Average value given in Ref. [] ** E = effective dose using the factor 0.06 msv Gy -1 cm OPG TLD measurements in phantom The doses distribution obtained for a normal adult when using the conventional OPG are the followings: an average dose of 0.3 mgy to the jawbone (peak dose of 0.8 mgy); a dose of about 0. mgy to the thyroid ; a dose below 0.1 mgy to the crystalline lens CT TLD measurements in phantom and CTDI w DVT Newtom 9000 and Dentascan (GE LightSpeed) To evaluate the dose delivered to the patient, the skull phantom equipped with bands of TLD was scanned using characteristic parameters for an adult. The obtained results are given in Fig. for both DVT Newtom and Dentascan. 6

7 0. mgy 0. mgy 0. mgy.1 mgy.5 mgy 3.5 mgy 1.0 mgy 1 mgy 3.9 mgy 3 mgy a).5 mgy b) 3.5 mgy FIG. 5. Dose distribution using (a) the Newtom DVT and (b) Dentascan technology To extent dose measurements for CT examinations, using CTDI w enables to compare effective doses. The Newtom DVT system yields a CTDI w value of. mgy using the 16-cm diameter phantom. For an irradiation length of about 13 cm, this examination delivers a DLP of 55 mgy cm. Using the CT conversion factors gives an effective dose of 0.13 msv. A note published by Health Physics Society of Michigan gives for this type of installation a CTDI w value of.5 mgy and an effective dose of 0.05 msv. For the Dentascan, the CTDI w was 19 mgy for the lower jawbone examination corresponding to a DLP of 106 mgy cm and an effective dose of 0.5 msv. The same measurement for the upper jawbone gave a CTDI w of 3.5 mgy, a DLP of 118 mgy cm and an effective dose of 0.7 msv. Finally, Dentascan examinations deliver a total effective dose of 0.5 msv. The Health Physics Society of Michigan gives effective dose ranging from 0.15 to 0.58 msv. The Dentascan delivers about times higher doses than the Newtom DVT. 3.3 Conclusion The Insight Kodak film is approximately twice more sensitive than the Ultra-speed film. Thus it is possible to reduce the dose by a factor while keeping similar image quality. Compared to intra-oral films, the RVG Trophy digital detector offers a much higher DQE thanks to the about 7 times higher sensitivity for a comparable level of noise in high-resolution mode. However, this advantage is relevant only if the unit enables to control short exposure time. The dose indicator used for intra-oral radiology or OPG examination is the DAP. For examinations performed using D films, the average DAP is about 0.05 Gy cm. Using digital systems such as Trophy will enable to reduce significantly the dose level in the future. Concerning the OPG, the average DAP was about 0.11 Gy cm, which corresponds approximately to the DAP of intra-oral images. For OPG examination, using digital detector do not reveal a dose reduction at all. An intraoral examination using D films yields an effective dose of about 1.7 µsv while an OPG yields about 7 µsv. Using the dedicated Newtom DVT 9000 CT system reduces the patient exposure by a factor compared to an optimized conventional CT examination (Dentascan). Indeed, the DVT system yields a DLP of 55 mgy cm that corresponds to an effective dose of 0.13 msv (0.5 msv for the conventional CT device). 7

8 References [1] NRPB Guidelines on patient dose to promote the optimisation of protection for diagnostic medical exposure. NRPB, Chilton UK, (1) (1999). [] Hart D., Jones D.G., Wall B.F., Estimation of effective dose in diagnostic radiology from entrance surface dose and dose area product measurements, NRPB, Chilton UK R6. [3] Isoadi P., Ropolo R., Measurements do pose width product in panoramic dental radiology. BJR, 76:19-131, (003). 8

2012 :15th SESSION of ESMP

2012 :15th SESSION of ESMP 2012 :15th SESSION of ESMP Lecture presented in Archamps (Salève Building) by : Elly CASTELLANO (London) Patient dosimetry in x-ray imaging and CT Elly Castellano Objectives measurable dose quantities

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients

More information

Introduction of a Single Chip TLD System for Patient Dosimetry

Introduction of a Single Chip TLD System for Patient Dosimetry Introduction of a Single Chip TLD System for Patient Dosimetry C. Hranitzky a, M. Halda a, G. Müller a, B. Obryk b, H. Stadtmann a* a Austrian Research Centers GmbH ARC, 2444 Seibersdorf, Austria. b Institute

More information

Overview of Safety Code 35

Overview of Safety Code 35 Common Quality Control Procedures for All s Quality Control Procedures Film All s Daily Quality Control Tests Equipment Warm-up (D1) According to manufacturers instructions Can include auto calibration(d1)

More information

Calculati tions - BIR meth thod and NCRP 147

Calculati tions - BIR meth thod and NCRP 147 Calculations l - BIR method and NCRP 147 David Sutton PhD 1. Cath Lab Indicative of all C-arm applications Only secondary radiation Primary Beam is stopped by the image intensifier Geometry BIR 9.5 m Preparation

More information

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Half value layer and AEC receptor dose compliance survey in Estonia

Half value layer and AEC receptor dose compliance survey in Estonia Half value layer and AEC receptor dose compliance survey in Estonia K. Kepler, A. Vladimirov Training Centre of Medical Physics, University of Tartu Testing Centre of the University of Tartu, Estonia E-mail:

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced

More information

Mammography: Physics of Imaging

Mammography: Physics of Imaging Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

Wide beam CT dosimetry. Elly Castellano

Wide beam CT dosimetry. Elly Castellano Wide beam CT dosimetry Elly Castellano Outline revision: CT dose indices wide-beam CT: the end of the road for CTDI? the IEC rescue plan for CTDI 100 the american way AAPM report 111 better estimates of

More information

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

Unit thickness. Unit area. σ = NΔX = ΔI / I 0 Unit thickness I 0 ΔI I σ = ΔI I 0 NΔX = ΔI / I 0 NΔX Unit area Δx Average probability of reaction with atom for the incident photons at unit area with the thickness of Delta-X Atom number at unit area

More information

Development of new dosimeter for measuring dose distribution in CT

Development of new dosimeter for measuring dose distribution in CT Development of new dosimeter for measuring dose distribution in CT Poster No.: C-2925 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology - Without Subtopic Authors: Y. Muramatsu, K.

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis

Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis Poster No.: C-0585 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Minehiro,

More information

The importance of radiation quality for optimisation in radiology

The importance of radiation quality for optimisation in radiology Available online at http://www.biij.org/2007/2/e38 doi: 10.2349/biij.3.2.e38 biij Biomedical Imaging and Intervention Journal COMMENTARY The importance of radiation quality for optimisation in radiology

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

X-RAYS - NO UNAUTHORISED ENTRY

X-RAYS - NO UNAUTHORISED ENTRY Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

Estimation of signal transfer property for wireless digital detector in different measurement schemes

Estimation of signal transfer property for wireless digital detector in different measurement schemes Estimation of signal transfer property for wireless digital detector in different measurement schemes Anatoli Vladimirov, Kalle Kepler Training Centre of Medical Physics, University of Tartu, Estonia 11

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

Nuclear Associates , , CT Head and Body Dose Phantom

Nuclear Associates , , CT Head and Body Dose Phantom Nuclear Associates 76-414,76-414-4150,76-415 CT Head and Body Dose Phantom Users Manual March 2005 Manual No. 76-414-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product

More information

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities Journal of Physics: Conference Series PAPER OPEN ACCESS Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities Recent citations - Resolution Properties of a

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

PATIENT EFFECTIVE DOSES IN DIAGNOSTIC RADIOLOGY, NA

PATIENT EFFECTIVE DOSES IN DIAGNOSTIC RADIOLOGY, NA Title of Paper: Patient effective doses in diagnostic radiology Authors: N.A. Gkanatsios, and W. Huda * Corresponding Author: Department of Radiology, University of Florida, P.O. Box 100374, Gainesville,

More information

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose Image Quality and Dose Issues in MSCT Image Quality and Dose Image quality Image noise Spatial resolution Contrast Artefacts Speckle and sharpness S. Edyvean St. George s Hospital London SW17 0QT Radiation

More information

Reducing Radiation Exposure from Survey CT Scans

Reducing Radiation Exposure from Survey CT Scans Reducing Survey CT Scan Exposure Pediatric Imaging Original Research Jennifer C. O Daniel 1 Donna M. Stevens 2 Dianna D. Cody 2 O Daniel JC, Stevens DM, Cody DD Received July 28, 2004; accepted after revision

More information

Applied Sciences at Winterthur)

Applied Sciences at Winterthur) Comparison of image quality between a digital panorama X-ray unit with a CdTe-CMOS detector and panorama X- ray units with other types of digital detectors Stephan Scheidegger 1 1 Zürcher Hochschule für

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 62220-1 First edition 2003-10 Medical electrical equipment Characteristics of digital X-ray imaging devices Part 1: Determination of the detective quantum efficiency Appareils

More information

Improved Imaging Plate Dosimetry for X-rays in Interventional Radiology

Improved Imaging Plate Dosimetry for X-rays in Interventional Radiology Improved Imaging Plate Dosimetry for X-rays in Interventional Radiology H. Ohuchi 1, A. Yamadera 2, T.Satoh 3, Y,Eguchi 3 1 Graduate School of Pharmaceutical Sciences, Tohoku University, Aramaki,Aoba-ku,

More information

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook DIAGNOSTIC ACCREDITATION PROGRAM Radiology and CT Quality Control Procedures Workbook Quality Control Procedures Radiography/CR/DR Safety Code 35 Summary For more detail about each quality control (QC)

More information

Test Equipment for Radiology and CT Quality Control Contents

Test Equipment for Radiology and CT Quality Control Contents Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3

More information

10/26/2015. Study Harder

10/26/2015. Study Harder This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen

More information

Unfors EDD-30 Radiation Protection in Fluoroscopy

Unfors EDD-30 Radiation Protection in Fluoroscopy Unfors EDD-30 Radiation Protection in Fluoroscopy Immediate Warning Decrease Your Dose Interventional radiology procedures are considered to be essential to medical diagnosis and treatment. It is recognized,

More information

Beam-Restricting Devices

Beam-Restricting Devices Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the

More information

IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film

IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film CYRIC Annual Report 2005 IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film Ohuchi H. 1, and Abe K. 2 1 Graduate School of Pharmaceutical Sciences,

More information

Version 1.0. TechnicVR. Student Guide

Version 1.0. TechnicVR. Student Guide Version 1.0 TechnicVR s h a d e r w a r e. c o m Student Guide TechnicVR s h a d e r w a r e. c o m Student Guide shaderware 2008 PO Box 103 Saltburn Cleveland TS12 1WP w w w. s h a d e r w a r e. c o

More information

ProX Intraoral X-ray. PLANMECA is proud to introduce a new intraoral X-ray unit to its comprehensive collection of imaging products- the ProX.

ProX Intraoral X-ray. PLANMECA is proud to introduce a new intraoral X-ray unit to its comprehensive collection of imaging products- the ProX. The premium intraoral X-ray unit... ProX Intraoral X-ray PLANMECA is proud to introduce a new intraoral X-ray unit to its comprehensive collection of imaging products- the ProX. This advanced unit provides

More information

DIGITAL RADIOGRAPHY. Digital radiography is a film-less technology used to record radiographic images.

DIGITAL RADIOGRAPHY. Digital radiography is a film-less technology used to record radiographic images. DIGITAL RADIOGRAPHY Digital radiography is a film-less technology used to record radiographic images. 1 The purpose of digital imaging is to generate images that can be used in the diagnosis and assessment

More information

CT Scanner Dose Survey

CT Scanner Dose Survey CT Scanner Dose Survey Measurement Protocol Version 5.0 July 1997 Co-ordinated by ImPACT and The Medical Physics Department,, London SW17, UK. 0181-725-3366 CT Scanner Dose Survey: Measurement Protocol

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

1. Patient size AEC. Large Patient High ma. Small Patient Low ma

1. Patient size AEC. Large Patient High ma. Small Patient Low ma Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 62220-1 First edition 2003-10 Medical electrical equipment Characteristics of digital X-ray imaging devices Part 1: Determination of the detective quantum efficiency Appareils

More information

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National

More information

biij Optimisation in general radiography CJ Martin, PhD, FIPEM, FioP Biomedical Imaging and Intervention Journal REVIEW PAPER

biij Optimisation in general radiography CJ Martin, PhD, FIPEM, FioP Biomedical Imaging and Intervention Journal REVIEW PAPER Available online at http://www.biij.org/2007/2/e18 doi: 10.2349/biij.3.2.e18 biij Biomedical Imaging and Intervention Journal REVIEW PAPER Optimisation in general radiography CJ Martin, PhD, FIPEM, FioP

More information

10/3/2012. Study Harder

10/3/2012. Study Harder This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen

More information

RaySafe X2. Effortless measurements of X-ray

RaySafe X2. Effortless measurements of X-ray RaySafe X2 Effortless measurements of X-ray At your fingertips We ve grown accustomed to intuitive interactions with our devices. After all, it s not the device that s most important, but what you can

More information

Digital Detector Array Image Quality for Various GOS Scintillators

Digital Detector Array Image Quality for Various GOS Scintillators Digital Detector Array Image Quality for Various GOS Scintillators More info about this article: http://www.ndt.net/?id=22768 Brian S. White 1, Mark E. Shafer 2, William H. Russel 3, Eric Fallet 4, Jacques

More information

Q3D. Speak to a 3D Specialist. CBCT 3D / Panoramic Imaging GENERAL DIMENSIONS. Suni Imaging Product Lines GET.

Q3D. Speak to a 3D Specialist. CBCT 3D / Panoramic Imaging GENERAL DIMENSIONS. Suni Imaging Product Lines GET. GENERAL Q3D Q3D Ceph Exposure Time FOV Voxel Size Focal Spot Target Angle Tube Voltage Tube Current Line Voltage Warranty Panoramic CT 9 to 17 sec 9 to 17 sec 4 to 12 sec 7.7/14.5 sec 7.7/14.5 sec 4 x

More information

Veraviewepocs 2D High Speed Panoramic X-Ray Crystal Clear Images with Reduced Radiation

Veraviewepocs 2D High Speed Panoramic X-Ray Crystal Clear Images with Reduced Radiation Diagnostic and Imaging Equipment Treatment Units Handpieces and Instruments Endodontic Systems Laser Equipment Laboratory Devices Veraviewepocs 2D High Speed Panoramic X-Ray Crystal Clear Images with Reduced

More information

AnySensor (1.0 /1.5)

AnySensor (1.0 /1.5) HQ-20080217-MM-kcg Dental Digital Imaging Radiography System AnySensor (1.0 /1.5) Cutting-edge CMOS Technology CMOS? CCD? CCD CMOS Past - Particular Process -High Quality & High Cost - General Process

More information

I. Introduction.

I. Introduction. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Accuracy of measuring half- and quarter-value layers and appropriate aperture width of a convenient method using a lead-covered case

More information

Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution

Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution Cynthia H. McCollough, PhD, DABR, FAAPM, FACR Director, CT Clinical Innovation Center Professor of Medical Physics and

More information

NOT FOR DISTRIBUTION JINST_083P_0914 v1

NOT FOR DISTRIBUTION JINST_083P_0914 v1 Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam F. Di Lillo a,b, D. Dreossi c, F. Emiro a,b, C. Fedon d,e, R. Longo d,e, G. Mettivier a,b,*, L. Rigon d,e, P. Russo

More information

Amorphous Selenium Direct Radiography for Industrial Imaging

Amorphous Selenium Direct Radiography for Industrial Imaging DGZfP Proceedings BB 67-CD Paper 22 Computerized Tomography for Industrial Applications and Image Processing in Radiology March 15-17, 1999, Berlin, Germany Amorphous Selenium Direct Radiography for Industrial

More information

abc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice nww.medical-devices.nhs.

abc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice   nww.medical-devices.nhs. abc September 2004 MHRA 04099 Philips Mx8000 IDT CT scanner technical evaluation Best choice best practice www.mhra.gov.uk nww.medical-devices.nhs.uk About MHRA evaluation reports. What you can expect.

More information

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Poster No.: C-2205 Congress: ECR 2012 Type: Scientific Paper

More information

A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers

A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers Poster No.: C-3011 Congress: ECR 2010 Type: Topic: Authors: Scientific Exhibit

More information

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

More information

Detector technology in simultaneous spectral imaging

Detector technology in simultaneous spectral imaging Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become

More information

PD233: Design of Biomedical Devices and Systems

PD233: Design of Biomedical Devices and Systems PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/

More information

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography Journal of Physics: Conference Series OPEN ACCESS The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography To cite this article: N A A Daud et al 2014 J. Phys.:

More information

Dose Reduction in Helical CT: Dynamically Adjustable z-axis X-Ray Beam Collimation

Dose Reduction in Helical CT: Dynamically Adjustable z-axis X-Ray Beam Collimation Medical Physics and Informatics Original Research Christner et al. CT Dose Reduction Medical Physics and Informatics Original Research Downloaded from www.ajronline.org by 8.243.133.8 on 2/26/18 from IP

More information

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality Digital radiography: Digital radiography is set to become the most common form of processing radiographic images in the next 10 years. This is due to a number of practical and image quality issues. Practical

More information

NEWTOM GO 2D GREAT.VISION

NEWTOM GO 2D GREAT.VISION CEFLA s.c. Via Selice Provinciale 23/a 40026 Imola Italy t. +39 045 8202727 045 583500 info@newtom.it newtom.it 05/2018 NGO2GB181S00 According to the standards in force, in extra-eu areas the availability

More information

X-Ray/ Digital Imaging

X-Ray/ Digital Imaging Chapter 20 X-Ray/ Digital Imaging Carestream - Intraoral Cameras 852 Intraoral X-Rays 855 Digital Sensors 857 Intraoral Imaging Plate System 860 Extraoral Imaging System 862 Intraoral Scanners 868 CBCT

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-649 CDRH Fluoroscopic Phantom Users Manual March 2005 Manual No. 07-649-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

Computed Radiography

Computed Radiography BAM Berlin Computed Radiography --INDE 2007, Kalpakkam, India -- Uwe Zscherpel, Uwe Ewert BAM Berlin, Division VIII.3 Requests Requests and and information information to: to: Dr. Dr. U. U. Zscherpel Zscherpel

More information

X-ray Tube and Generator Basic principles and construction

X-ray Tube and Generator Basic principles and construction X-ray Tube and Generator Basic principles and construction Dr Slavik Tabakov - Production of X-rays OBJECTIVES - X-ray tube construction - Anode - types, efficiency - X-ray tube working characteristics

More information

RADIATION SAFETY DIRECTORATE

RADIATION SAFETY DIRECTORATE RADIATION SAFETY DIRECTORATE Pursuant to Article 26-e, paragraph 1, item 13 of the Law on Ionising Radiation Protection and Radiation Safety (Official Gazette of the Republic of Macedonia No. 48/02 and

More information

A COMPARATIVE STUDY ON THE PERFORMANCE OF DIGITAL DETECTOR SYSTEMS FOR HIGH ENERGY APPLICATIONS

A COMPARATIVE STUDY ON THE PERFORMANCE OF DIGITAL DETECTOR SYSTEMS FOR HIGH ENERGY APPLICATIONS 11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic More Info at Open Access Database www.ndt.net/?id=16394 A COMPARATIVE STUDY ON THE PERFORMANCE

More information

ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone

ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY 12477 (800)VIP-XRAY (845)247-9028 Fax or 800-847-9729 local phone 845-246-9013 http://www.eccxray.com sales@eccxray.com INSTRUCTION MANUAL

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

A comparison of two methods for the determination of freein-air geometric efficiency in MDCT

A comparison of two methods for the determination of freein-air geometric efficiency in MDCT A comparison of two methods for the determination of freein-air geometric efficiency in MDCT Theocharis Berris *1, Kostas Perisinakis 1,, Antonios E. Papadakis and John Damilakis 1, 1 Department of Medical

More information

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS By Aravind Ravichandran arr192@mail.usask.ca University of Saskatchewan Address: 2424 Cumberland

More information

RaySafe Xi. Specifications

RaySafe Xi. Specifications RaySafe Xi Specifications raysafe Xi General EMC tested Exposure needed Reset Temp. range According to EN 61000-6-1:2007 and en 61000-6-3:2007 One Automatic 15 35 C (59 95 F) DetECtor cable length 2 and

More information

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This

More information

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann

More information

Film Replacement in Radiographic Weld Inspection The New ISO Standard

Film Replacement in Radiographic Weld Inspection The New ISO Standard BAM Berlin Film Replacement in Radiographic Weld Inspection The New ISO Standard 17636-2 Uwe Ewert, Uwe Zscherpel, Mirko Jechow Requests and information to: uwez@bam.de 1 Outline - The 3 essential parameters

More information

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Takeshi Fujiwara 1, Yuki Mitsuya 2, Hiroyuki Takahashi 2, and Hiroyuki Toyokawa 2 1 National Institute of Advanced Industrial

More information

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE Takeyuki Hashimoto 1), Morio Onoe 2), Hiroshi Nakamura 3), Tamon Inouye 4), Hiromichi Jumonji 5), Iwao Takahashi 6); 1)Yokohama Soei

More information

REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS

REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS DEPARTMENT OF HEALTH DIRECTORATE: RADIATION CONTROL Implementation date: 31 March 2009 Contents

More information

Technical aspects on DAP calibration and CT calibration

Technical aspects on DAP calibration and CT calibration EURADOS Report 2015-03 European Radiation Dosimetry Group e. V. Braunschweig, August 2015 Technical aspects on DAP calibration and CT calibration H. Järvinen, O. Turak, M. Ginjaume, J. Daures, C. Hourdakis,

More information

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC R. Bellazzini a,b, G. Spandre a*, A. Brez a, M. Minuti a, M. Pinchera a and P. Mozzo b a INFN Pisa

More information

Calibration of KAP meters

Calibration of KAP meters Calibration of KAP meters Alexandr Malusek! Division of Radiological Sciences Department of Medical and Health Sciences Linköping University! 2014-04-15 1 Outline 1. KAP meter construction 2. Air kerma-area

More information

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016 Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016 Current technology in digital image production (CR/DR and other modalities) 2/ Overview Digital

More information

Effective energy measurement using radiochromic film: application of a mobile scanner

Effective energy measurement using radiochromic film: application of a mobile scanner Polish Journal of Medical Physics and Engineering 2016;22(4):85-92 December 2016 doi: 10.1515/pjmpe-2016-0015 ISSN 1898-0309 Scientific Paper Effective energy measurement using radiochromic film: application

More information

Quality Control of Full Field Digital Mammography Units

Quality Control of Full Field Digital Mammography Units Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of

More information

Welcome to the smiling world of SOREDEX CRANEX. Novus e. Easy as a breeze. soredex.com

Welcome to the smiling world of SOREDEX CRANEX. Novus e. Easy as a breeze. soredex.com Welcome to the smiling world of SOREDEX CRANEX Novus e Easy as a breeze soredex.com Benefits at a glance Essential part of your daily patient care Excellent Fast scanning times are convenient for the patient

More information

Diagnostic X-Ray Shielding

Diagnostic X-Ray Shielding Diagnostic X-Ray Shielding Multi-Slice CT Scanners Using NCRP 147 Methodology Melissa C. Martin, M.S., FAAPM, FACR Therapy Physics Inc., Bellflower, CA AAPM Annual Meeting, Orlando, FL FL Refresher Course

More information

P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s

P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s AN ALL AROUND PERFECT PICTURE. The perfect combination of image quality, efficiency and design. P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s Air Techniques experts in digital imaging ProVecta

More information

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 David F. Lewis Advanced Materials Group Ashland Specialty Ingredients Spain, April 2014 What is Radiochromic Film? A film that instantly changes

More information

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology CODE: RADT 156 INSTITUTE: Health Science TITLE: Equipment Operation I DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of equipment operation and maintenance of radiographic

More information

Iterative Reconstruction

Iterative Reconstruction RECENT ADVANCES IN CT RADIATION DOSE REDUCTION TECHNIQUES Iterative Reconstruction Kalpana Kanal, PhD, FSCBTMR, FACR, FAAPM Professor and Director, Diagnostic Physics Section University of Washington Seattle,

More information

PIXSCAN CT scanner for Small Animal Imaging Based on hybrid pixel detectors

PIXSCAN CT scanner for Small Animal Imaging Based on hybrid pixel detectors PIXSCAN CT scanner for Small Animal Imaging Based on hybrid pixel detectors Centre de Physique des Particules de M arseille (CPPM -IN2P3), France S. B a s o lo, A. B o n is s e n t, P. B re u g n o n,

More information

Improvement of CT image quality with iterative reconstruction idose4

Improvement of CT image quality with iterative reconstruction idose4 Improvement of CT image quality with iterative reconstruction idose4 Poster No.: C-0387 Congress: ECR 2014 Type: Scientific Exhibit Authors: M.-L. Olsson, K. Norrgren, M. Söderberg; Malmö/SE Keywords:

More information