Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2
|
|
- Sybil Moore
- 5 years ago
- Views:
Transcription
1 Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2 Medical Consult, SA;
2
3 Establish a method to correlate image noise with individual demographic characteristics of each patient ; Optimize acquisition parameters without losing diagnostic quality.
4
5 Data standardization in CT guaranties a correct image exposure a fixed amount of radiation attenuation; CT users don t always know when/how to lower mas/kvp.
6 Quality of the exam changes, differing the level of image noise, with patients of different body types; There has been a tendency to increase mas (dose) in order to lower image noise.
7
8 Technical / Demographical Parameters Age; Antero- posterior (AP) and lateral (LAT) diameter; Weight; KVp; mas; Pitch; Slice thickness; Kernel; Height; CTDIvol; DLP; Scan Length; SD; Exam Head Lumbar Spine Chest Abdomen Pelvis Abdomen +Pelvis Sample Weight Slice CTDIvol Noise kvp mas (Kg) (mm) (mgy) (HU) male 38 min female 61 median total 99 max male 27 min female 47 median total 74 max male 15 min female 11 median total 26 max male 9 min female 9 median total 18 max male 10 min female 5 median total 15 max male 9 min female 12 median total 21 max Philips equipments 16 and 64 slice Brilliance / 4 slice Mx8000.
9 Tomografia Computorizada TC 1 (16cortes) Pa#ent ID Região anatómica Sexo Idade Altura (m) Peso (kg) ) (mgy) DLP (mgy*cm) Modulação de Dose kv mas Espessura de corte (mm) Pitch da aquisição (cm) Kernel Espessura (cm) Largura (cm) ROI (SD)
10 The standard deviation, was used as a quantity marker. q One image is chosen, and a ROI is drawn with 2 to 5 cm 2, in a homogeneous region ;
11 The images were digitally manipulated adding different levels of noise, simulating an mas lowering: Adaptation of the used mas (pseudo mas) to reduce the dose: Head Lumbar Spine Chest Abdomen Pelvis
12 Image quality was determined by two Radiologists; q The noise was evaluated without knowledge of the dose and technical parameters. Criteria to evaluate Image Quality Anatomic Criteria Physic Criteria Diagnostic Value General Quality Evaluating system Image Quality Score Unacceptable 1 Sub - Acceptable 2 Acceptable 3 Good Quality 4 High Quality 5
13 Image test with diferent noise levels added, using the Image J software: 350mAs 315 mas (- 10%) 262 (- 25%) mas 210 mas (- 40%) SD +4% SD +10% SD +27% SD
14
15 Percentile 75 of local DRL's: Local DRL s European DRL s Exam CTDIvol CTDIw DLP CTDIw DLP Head Lumbar Spine Chest Abdomen Pelvis Abdomen+Pelvis
16 Quantitative evaluation: relation between image noise with radiation dose and demographic factors (Abdomen) y = x R² = Noise Noise y = x R² = Effective Diameter (cm) Noise y = x R² = IMC Weight (Kg)
17 Noise Quantitative evaluation: relation between image noise with radiation dose and demographic factors (Chest) y = x Noise R² = y = 0.446x R² = Effective Diameter (cm) IMC Ruído y = x R² = Peso (Kg)
18 Quantitative evaluation: relation between image noise with radiation dose and demographic factors (Lumbar Spine) y = x R² = Noise y = x R² = Effective Diameter (cm) y = x R² = Noise Noise IMC Weight (kg)
19 Quantitative evaluation: relation between image noise with radiation dose and demographic factors (Head) 9 y = x R² = Noise Effective Diameter (cm)
20 Relation of image noise with mas from PMMA phantoms SD (HU) mas product (mas) Chest Head Abdomen Lumbar Spine Linear (Chest) Linear (Head)
21 Quality perception in test images with added noise Image Quality (Noise) Score Abdomen CT images Score Pseudo mas Image Quality (Noise) Score Lumbar Spine CT images Score Pseudo mas Image Quality (Noise) Score Chest CT images Score Pseudo mas Def [20-24] cm Def [30-34] cm Def [25-29] cm Def >35 cm Def [20-25] cm Def [31-36] cm Def [26-30] cm Def >37 cm BMI [15-20[ Kg/m^2 BMI [25-30[ Kg/m^2 BMI [20-25[ Kg/m^2 BMI >=30 Kg/m^2
22
23 Test images evaluation with the added noise and mas reduction as proposed Lumbar Spine: Def(cm) Routine mas Target Score Noise Increased Modified mas Dose Reduction % % % % % % > % 435 0% Abdomen : Def(cm) Routine mas Target Score Noise Increased Modified mas Dose Reduction % % % % % % > % % Chest: BMI(kg/m^2) Routine mas Target Score Noise Increased Modified mas Dose Reduction % 75 25% % % % % > % 160 0%
24 Dose reduction by exam type mas reduction in standard Abdomen protocol mas reduction in standard Chest protocol 40% 30% 20% 10% 0% [20-25[ [25-30[ [30-35[ 35 25% 20% 15% 10% 5% 0% [20-25[ [25-30[ [30-35[ 35 mas reduction in standard Lumbar Spine protocol 30% 20% 10% 0% [15-19[ [20-24[ [25-29[ >30
25 Results show that the dose can be individually adapted while acquiring a CT exam with good quality wanted noise levels; Dose adjustment depend on size/weight and age of each patient; The image analysis helped the radiologist to determine the dose adjustment that would produce good images.
26
27 The dose must be individually adapted, with the need to standardize protocols and technical parameters adapted to the patient demographic characteristics (BMI and Effective Diameter); Radiologists have different acceptability of noise each department needs to adapt their noise levels, never exceeding the DRL s.
28 THANK YOU
29 REFERENCES
30 CT Image Quality, Wil Reddinger, M.Sc., R.T.(R)(CT); OutSource, Inc; Apr ICRP, International Commission on Radiological Protection. Recommendations of the International Commission on Radiological Protection. ICRP Publication 60; ICRP, International Commission on Radiological Protection. Managing Patient Dose in Multi- Detector Computed Tomography (MDCT). ICRP Publication 102. Annals of the ICRP 37,;Elsevier, IAEA,dose reduction in CT while Maintaining Diagnostic Confidence: A Feasibility/Demonstration Study, Vienna Sep European guidelines on quality criteria for computed tomography. Luxembourg: office for official Publications of the European Communities; X Ray CT, Nick Keat, ImPACT ST. George's Hospital; accessed
31 ACKNOWLEDGMENT
32 Serviço de Imagiologia Hospital de Santa Maria; Medical Consult.
Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London
Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients
More informationInfluence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck
Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Poster No.: C-2205 Congress: ECR 2012 Type: Scientific Paper
More informationPitfalls and Remedies of MDCT Scanners as Quantitative Instruments
intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature
More informationModel Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise
Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise Jonas Rydberg, M.D. Professor of Radiology Indiana University School of Medicine Indianapolis, Indiana Medical
More informationDose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp
Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division
More informationdiagnostic examination
RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011
More informationTeaching Digital Radiography and Fluoroscopic Radiation Protection
Teaching Digital Radiography and Fluoroscopic Radiation Protection WCEC 20 th Student Educator Radiographer Conference Dennis Bowman, RT(R), CRT (R)(F) Community Hospital of the Monterey Peninsula (CHOMP)
More informationTOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions
CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT
More informationWide-Detector CT for TAVR Planning:
Wide-Detector CT for TAVR Planning: Impact on Iodine Dose, Radiation Dose, and Image Quality SCBTMR 2015 Annual Course Thursday, October 8 William P. Shuman MD FSCBTMR Department of Radiology University
More informationHISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD
CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING
More informationOverview of Safety Code 35
Common Quality Control Procedures for All s Quality Control Procedures Film All s Daily Quality Control Tests Equipment Warm-up (D1) According to manufacturers instructions Can include auto calibration(d1)
More informationElectronic Noise in CT Detectors: Impact on Image Noise and Artifacts
Medical Physics and Informatics Original Research Duan et al. Electronic Noise in CT Detectors Medical Physics and Informatics Original Research Xinhui Duan 1 Jia Wang 1,2 Shuai Leng 1 ernhard Schmidt
More informationThe disclaimer on page 1 is an integral part of this document. Copyright November 30, 2017 by AAPM. All rights reserved.
DISCLAIMER: TO THE EXTENT ALLOWED BY LOCAL LAW, THIS INFORMATION IS PROVIDED TO YOU BY THE AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A NON-PROFIT ORGANIZATION ORGANIZED TO PROMOTE THE APPLICATION
More information1. Patient size AEC. Large Patient High ma. Small Patient Low ma
Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description
More informationQC Testing for Computed Tomography (CT) Scanner
QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including
More informationImprovement of CT image quality with iterative reconstruction idose4
Improvement of CT image quality with iterative reconstruction idose4 Poster No.: C-0387 Congress: ECR 2014 Type: Scientific Exhibit Authors: M.-L. Olsson, K. Norrgren, M. Söderberg; Malmö/SE Keywords:
More informationPractical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S.
Practical Medical Physics Session: TG-151 Dose Monitoring August 5, 2013 Katie Hulme, M.S. Digital Imaging and Dose Creep Images courtesy of Agfa Healthcare Under-Exposed Over-Exposed Freedman et al.,
More informationQC by the MPE in Belgium
Acceptance testing of state-of-the-art CT scanners using a new national protocol: first experience on a large number of scanners of different make and model the working group Radiology of the Belgian Hospital
More informationResearch Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.
Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National
More informationT h e P h a n t o m L a b o r a t o r y
T h e P h a n t o m L a b o r a t o r y 1 CCT228 ATCM Phantom Manual Copyright 2017 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order
More informationWhile digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.
In press 2004 1 2 Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Digital radiology An appropriate analogy that is easy for most people to understand
More informationRadiation Dose Modulation. the Multidetector CT Era: From Basics to Practice 1
Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. EDUCATION EXHIBIT
More informationExposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?
Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago
More informationRadiation Dose Index monitoring (RDIM) systems and establishment of local DRLs
IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, 21017 Radiation Dose Index monitoring (RDIM) systems and establishment of local DRLs Dean Pekarovič UMC Ljubljana, Institute
More informationFOREWORD. Acknowledgements
ΠΑΝΕΠΙΣΗΜΙΟ ΠΑΣΡΩΝ Διαημημαηικό Πρόγραμμα Μεηαπηστιακών ποσδών ζηην Ιαηρική Φσζική Διπλωμαηική εργαζία «ΔΟΙΜΕΣΡΙΑ ΑΘΕΝΩΝ Ε ΕΞΕΣΑΕΙ ΤΠΟΛΟΓΙΣΙΚΗ ΣΟΜΟΓΡΑΦΙΑ ΠΟΛΛΑΠΛΩΝ ΣΟΜΩΝ» ηέλλα Γ. Θαλαζζινού Α.Μ : 1575
More informationAppropriate Inspection Distance of Digital X-Ray Imaging Equipment for Diagnosis
Indian Journal of Science and Technology Vol 8(S8), 380-386, April 2015 ISSN (Print) : 0974-6846 ISSN (Online) : 0974-5645 DOI: 10.17485/ijst/2015/v8iS8/70528 Appropriate Inspection Distance of Digital
More information2012 :15th SESSION of ESMP
2012 :15th SESSION of ESMP Lecture presented in Archamps (Salève Building) by : Elly CASTELLANO (London) Patient dosimetry in x-ray imaging and CT Elly Castellano Objectives measurable dose quantities
More information12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees
Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees
More informationClinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11
Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) Hiroki Kadoya Yukiko Kitagawa MEDIX VOL. 61 P.8 P.11 Clinical Experience Using the Open Bore Multislice CT System Supria
More informationDiagnostic X-Ray Shielding
Diagnostic X-Ray Shielding Multi-Slice CT Scanners Using NCRP 147 Methodology Melissa C. Martin, M.S., FAAPM, FACR Therapy Physics Inc., Bellflower, CA AAPM Annual Meeting, Orlando, FL FL Refresher Course
More informationQuantitation of clinical feedback on image quality differences between two CT scanner models
Received: 4 August 2016 Revised: 4 November 2016 Accepted: 12 December 2016 DOI: 10.1002/acm2.12050 MEDICAL IMAGING Quantitation of clinical feedback on image quality differences between two CT scanner
More informationAsk EuroSafe Imaging Tips & Tricks. Paediatric Imaging Working Group. Dose Management in Digital Radiography
Ask EuroSafe Imaging Tips & Tricks Paediatric Imaging Working Group Dose Management in Digital Radiography Raija Seuri (HUS Medical Imaging Center, FI) Cristina Almeida (Centro Hospitalar de Lisboa Central,
More informationDose Reduction in Helical CT: Dynamically Adjustable z-axis X-Ray Beam Collimation
Medical Physics and Informatics Original Research Christner et al. CT Dose Reduction Medical Physics and Informatics Original Research Downloaded from www.ajronline.org by 8.243.133.8 on 2/26/18 from IP
More informationBreast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School
Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis
More informationEuropean Journal of Radiology
European Journal of Radiology 72 (2009) 202 208 Contents lists available at ScienceDirect European Journal of Radiology journal homepage: www.elsevier.com/locate/ejrad Digital radiography: The balance
More informationReducing Radiation Exposure from Survey CT Scans
Reducing Survey CT Scan Exposure Pediatric Imaging Original Research Jennifer C. O Daniel 1 Donna M. Stevens 2 Dianna D. Cody 2 O Daniel JC, Stevens DM, Cody DD Received July 28, 2004; accepted after revision
More informationThe importance of radiation quality for optimisation in radiology
Available online at http://www.biij.org/2007/2/e38 doi: 10.2349/biij.3.2.e38 biij Biomedical Imaging and Intervention Journal COMMENTARY The importance of radiation quality for optimisation in radiology
More informationDigital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of
Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of radiation to the population due to Medical Imaging
More informationREQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS
REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS DEPARTMENT OF HEALTH DIRECTORATE: RADIATION CONTROL Implementation date: 31 March 2009 Contents
More informationCHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY
14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to
More information10 kvp rule an anthropomorphic pelvis phantom imaging study using a CR system : impact on image quality and effective dose using AEC and manual mode
10 kvp rule an anthropomorphic pelvis phantom imaging study using a CR system : impact on image quality and effective dose using AEC and manual mode Lanca, L, Franco, L, Ahmed, A and Hogg, P http://dx.doi.org/10.1016/j.radi.2014.04.007
More informationTranslating Protocols Between Scanner Manufacturer and Model
Translating Protocols Between Scanner Manufacturer and Model Robert J. Pizzutiello, MS, FAAPM, FACMP Sr. Vice-President, Global Physics Solutions President, Upstate Medical Physics Objectives Understand
More informationAquilion Precision Ultra-High Resolution CT: Quantifying diagnostic image quality
Aquilion Precision Ultra-High CT: Quantifying diagnostic image quality Kirsten Boedeker, PhD, DABR Senior Manager, Quantitative Image Quality Canon Medical Systems Corporation Introduction Over the last
More informationSlide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?
Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM
More informationX-RAYS - NO UNAUTHORISED ENTRY
Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are
More informationIterative Reconstruction in Image Space. Answers for life.
Iterative Reconstruction in Image Space Answers for life. Iterative Reconstruction in Image Space * (IRIS) * Please note: IRIS is used as an abbreviation for Iterative Reconstruction in Image Space throughout
More information1. Queries are issued to the image archive for information about computed tomographic (CT)
Appendix E1 Exposure Extraction Method examinations. 1. Queries are issued to the image archive for information about computed tomographic (CT) 2. Potential dose report screen captures (hereafter, dose
More informationPATIENT EFFECTIVE DOSES IN DIAGNOSTIC RADIOLOGY, NA
Title of Paper: Patient effective doses in diagnostic radiology Authors: N.A. Gkanatsios, and W. Huda * Corresponding Author: Department of Radiology, University of Florida, P.O. Box 100374, Gainesville,
More informationImage Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose
Image Quality and Dose Issues in MSCT Image Quality and Dose Image quality Image noise Spatial resolution Contrast Artefacts Speckle and sharpness S. Edyvean St. George s Hospital London SW17 0QT Radiation
More informationabc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice nww.medical-devices.nhs.
abc September 2004 MHRA 04099 Philips Mx8000 IDT CT scanner technical evaluation Best choice best practice www.mhra.gov.uk nww.medical-devices.nhs.uk About MHRA evaluation reports. What you can expect.
More informationRedefining Ergonomics
Samsung Electronics Co., Ltd. inspires the world and shapes the future with transformative ideas and technologies, redefining the worlds of TVs, smartphones, wearable devices, tablets, cameras, digital
More informationHalf value layer and AEC receptor dose compliance survey in Estonia
Half value layer and AEC receptor dose compliance survey in Estonia K. Kepler, A. Vladimirov Training Centre of Medical Physics, University of Tartu Testing Centre of the University of Tartu, Estonia E-mail:
More informationTest Equipment for Radiology and CT Quality Control Contents
Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3
More informationPATIENT EXPOSURE AND DOSE GUIDE
CRCPD Publication E-03-2 $15.00 PATIENT EXPOSURE AND DOSE GUIDE - 2003 January 2003/Republished April 2003 Published by Conference of Radiation Control Program Directors, Inc. [This page is intentionally
More information2217 US Highway 70 East Garner, NC Main: Fax:
Viztek is committed to providing the highest image quality possible in our CR & DR product lines. There are several factors that directly affect the overall quality of CR & DR based images. The eposure
More informationThe effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography
Journal of Physics: Conference Series OPEN ACCESS The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography To cite this article: N A A Daud et al 2014 J. Phys.:
More informationDetector technology in simultaneous spectral imaging
Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become
More informationbiij Optimisation in general radiography CJ Martin, PhD, FIPEM, FioP Biomedical Imaging and Intervention Journal REVIEW PAPER
Available online at http://www.biij.org/2007/2/e18 doi: 10.2349/biij.3.2.e18 biij Biomedical Imaging and Intervention Journal REVIEW PAPER Optimisation in general radiography CJ Martin, PhD, FIPEM, FioP
More informationDevelopment of new dosimeter for measuring dose distribution in CT
Development of new dosimeter for measuring dose distribution in CT Poster No.: C-2925 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology - Without Subtopic Authors: Y. Muramatsu, K.
More informationTISSUE EQUIVALENT PHANTOMS FOR EVALUATING IN-PLANE TUBE CURRENT MODULATED CT DOSE AND IMAGE QUALITY
TISSUE EQUIVALENT PHANTOMS FOR EVALUATING IN-PLANE TUBE CURRENT MODULATED CT DOSE AND IMAGE QUALITY By RYAN F. FISHER A THESIS PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY OF FLORIDA IN PARTIAL FULFILLMENT
More informationof sufficient quality and quantity
of sufficient quality and quantity The patient s body attenuates the beam as it passes though the body More energy is deposited in organs located near the entry of the beam than near the exit of the beam
More informationAutomated Technique to Measure Noise in Clinical CT Examinations
Medical Physics and Informatics Original Research Christianson et al. Noise Measurement in Clinical CT Medical Physics and Informatics Original Research Downloaded from www.ajronline.org by 46.3.195.17
More informationWide beam CT dosimetry. Elly Castellano
Wide beam CT dosimetry Elly Castellano Outline revision: CT dose indices wide-beam CT: the end of the road for CTDI? the IEC rescue plan for CTDI 100 the american way AAPM report 111 better estimates of
More informationDigital radiography (DR) post processing techniques for pediatric radiology
Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions
More informationEC Guideline 109: Guidance on Diagnostic Reference Levels REFERENCE LEVELS AT EUROPEAN LEVEL FOR CARDIAC INTERVENTIONAL PROCEDURES
SENTINEL WORKSHOP Delft, - April 7 EC Guideline : Guidance on Diagnostic Reference Levels REFERENCE LEVELS AT EUROPEAN LEVEL FOR CARDIAC INTERVENTIONAL PROCEDURES R. Padovani, E. Vano, A. Trianni, C. Bokou,
More informationFocal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution
Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution Cynthia H. McCollough, PhD, DABR, FAAPM, FACR Director, CT Clinical Innovation Center Professor of Medical Physics and
More informationDose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis
Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis Poster No.: C-0585 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Minehiro,
More informationFeatures and Weaknesses of Phantoms for CR/DR System Testing
Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal
More informationA comparative study of several digital flat panel X-ray units: patients doses and image quality in chest radiography
A comparative study of several digital flat panel X-ray units: patients doses and image quality in chest radiography Torres Cabrera R. 1, España López M.L. 2 Ruiz Manzano P. 3, Sastre Aguado J.M. 4,, Rivas
More informationTESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist
TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist Topics Image Uniformity and Artifacts Image Quality - Detail
More informationAim. Images for this section: Page 2 of 13
Changes in CT number of high atomic number materials with field of view when using an extended CT number to electron density curve and a metal artifact reduction reconstruction algorithm Poster No.: R-0094
More informationSoftware and Hardware in CCTA. Elly Castellano PhD
Software and Hardware in CCTA Elly Castellano PhD Outline technical requirements for coronary CTA the modern cardiac CT scanner ECG-gating technology image reconstruction algorithms 2 Technical requirements
More informationA comparison of two methods for the determination of freein-air geometric efficiency in MDCT
A comparison of two methods for the determination of freein-air geometric efficiency in MDCT Theocharis Berris *1, Kostas Perisinakis 1,, Antonios E. Papadakis and John Damilakis 1, 1 Department of Medical
More information160-slice CT SCANNER / New Standard for the Future
TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.
More informationAbdominal Radiology ISSN X. Abdom Radiol DOI /s
Assessment of 70-keV virtual monoenergetic spectral images in abdominal CT imaging: A comparison study to conventional polychromatic 120- kvp images Negin Rassouli, Hamid Chalian, Prabhakar Rajiah, Amar
More informationSTEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS
STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%
More informationQuality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals
DOI 10.1007/s00247-011-2290-1 ORIGINAL ARTICLE Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals Mervyn D. Cohen & Richard Markowitz
More informationNATIONWIDE EVALUATION OF X-RAY TRENDS (NEXT) TABULATION AND GRAPHICAL SUMMARY OF 2001 SURVEY OF ADULT CHEST RADIOGRAPHY
CRCPD Publication E-05-2 Available Online at No Charge $15.00 for a Computer-Generated Copy NATIONWIDE EVALUATION OF X-RAY TRENDS (NEXT) TABULATION AND GRAPHICAL SUMMARY OF 2001 SURVEY OF ADULT CHEST RADIOGRAPHY
More informationNeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results
NeuViz 16 Computed Tomography Elevating routine imaging for exceptional results Essence NeuViz 16 Raising the bar on clinical utility in routine imaging. Get more. More clinical information for patients.
More informationSAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life.
Neuro Thoracic Abdominal Abdominal Cardiovascular Pediatric SAFIRE Sinogram Affirmed Iterative Reconstruction Answers for life. SAFIRE * (Sinogram Affirmed Iterative Reconstruction) * The information
More informationDiffraction-enhanced X-ray Imaging (DEXI) Medical Solutions. More information using less radiation
Diffraction-enhanced X-ray Imaging (DEXI) Medical Solutions More information using less radiation Medical Small Animal Security NDE/NDT Diffraction-Enhanced X-ray Imaging Medical Solutions Safe non-invasive
More informationDigital Radiography X-Ray System. X Twin with X Mobil Roesys GmbH [rshsmi] 1/11
Digital Radiography X-Ray System 2017 Roesys GmbH [rshsmi] 1/11 1 General specifications The system is intended to be installed with floor mounted components only. It enables spacesaving installation without
More informationDisclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography
Current Implementation Status of IEC Standard 62494-1: Exposure Index (EI) for Digital Radiography July 31, 2017 Ryan Fisher, PhD, DABR Katie Hulme, MS, DABR None Disclosures Outline Review of IEC Standard
More informationNew spectral benefi ts, proven low dose
New spectral benefi ts, proven low dose Philips MicroDose mammography SI, technical data sheet Philips MicroDose SI with single-shot spectral imaging is a fullfi eld digital mammography solution that delivers
More informationCOCIR SELF-REGULATORY INITIATIVE FOR MEDICAL IMAGING EQUIPMENT COMPUTED TOMOGRAPHY MEASUREMENT OF ENERGY CONSUMPTION
COCIR SELF-REGULATORY INITIATIVE FOR MEDICAL IMAGING EQUIPMENT COMPUTED TOMOGRAPHY MEASUREMENT OF ENERGY CONSUMPTION Revision: 1 Date: June 2015 Approved: June 2015 TABLE OF CONTENT 1. INTRODUCTION...
More informationComputed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology
The Fundamentals of... Computed Tomography Computed Tomography (CT) systems use x-rays to produce images of slices through a patient s anatomy. Despite having lower spatial resolution than other x-ray
More informationNuclear Associates
Nuclear Associates 07-706 Patient Phantom/Penetrometer System Users Manual March 2005 Manual No. 07-706-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are
More informationNuclear Associates EZ CR-DIN Phantoms
Nuclear Associates 07-605-7777 EZ CR-DIN Phantoms Users Manual August 2006 Manual No. 07-605-7777-1 Rev. 4 2006 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks
More informationExposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations
Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations S. Baechler 1, P. Monnin 1, A. Aroua 1, J.F. Valley 1, M. Perrier, P. Trueb 3, F.R. Verdun 1 1 University
More informationMedical Imaging. Latest generation bucky room. system. radiology ahead
Medical Imaging Latest generation bucky room system radiology ahead system it grows efficiency, the productivity is on the up Moviplan ic System is the latest-generation X-ray equipment, combining all
More informationMUSICA Nerve Center. Artificial Intelligence. Intelligent tools for your Digital Radiography workflow. Fluoroscopy. Workflow Optimization
Image Quality Bariatric Abdomen Pediatric Imaging Diagnostic Confidence Fluoroscopy Neonatal Imaging Scatter Suppression Dental Full Leg Full Spine Exposure Control Index Artificial Intelligence General
More informationStudies on reduction of exposure dose using digital scattered X-ray removal processing
Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.
More informationStudies on reduction of exposure dose using digital scattered X-ray removal processing
Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.
More informationMetal Artifact Reduction for Orthopedic Implants (O-MAR)
Metal Artifact Reduction for Orthopedic Implants () Summary Since the inception of CT, numerous methods have been proposed to suppress metal artifacts with varying degrees of success. 1-4 (Metal Artifact
More informationSIGNA Explorer Lift revives our MR
Seiji Shiotani, MD, PhD Seirei Fuji Hospital in Fuji City, Shizuoka, Japan Masayoshi Sugimura Seirei Fuji Hospital in Fuji City, Shizuoka, Japan SIGN Explorer Lift revives our MR The clinical usefulness
More informationRaySafe X2. Effortless measurements of X-ray
RaySafe X2 Effortless measurements of X-ray At your fingertips We ve grown accustomed to intuitive interactions with our devices. After all, it s not the device that s most important, but what you can
More informationPerformance and care. all in one
Performance and care all in one INNOVATION IS WHAT DRIVES US THINKING ABOUT THE FUTURE Preventive diagnostics remains an essential weapon in defeating breast cancer. Metaltronica s forward-thinking design
More informationCT Basics: Data Acquisition Module 3
Module 3 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.
More informationImproved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques
Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques Wataru FUKUDA* Junya MORITA* and Masahiko YAMADA* Abstract Tomosynthesis is a three-dimensional imaging technology
More informationObjective Evaluation of Radiographic Contrast- Enhancement Masks
Chapter 8 Objective Evaluation of Radiographic Contrast- Enhancement Masks The development and application of radiographic contrast-enhancement masks (RCMs) in digital radiography (DR) were discussed in
More information