Software and Hardware in CCTA. Elly Castellano PhD

Size: px
Start display at page:

Download "Software and Hardware in CCTA. Elly Castellano PhD"

Transcription

1 Software and Hardware in CCTA Elly Castellano PhD

2 Outline technical requirements for coronary CTA the modern cardiac CT scanner ECG-gating technology image reconstruction algorithms 2

3 Technical requirements for coronary CTA 3

4 Technical requirements for coronary CTA ECG-gating capability freezes cardiac motion high temporal resolution reduces blurring due to cardiac motion high contrast and spatial resolution enhances conspicuity of narrow vessels large detector coverage per rotation facilitates short scan times without ECG gating images courtesy of Willi Kalender with ECG gating 4

5 ECG gating two physiological windows of opportunity HR < 65 HR > 65 R-R interval 100% V V t t ~ 65-75% ~ 25-35% end diastole end systole choice in timing and width of acquisition window to capture one / several / all phases of cardiac cycle 5

6 Temporal resolution determined by time taken to acquire projection data rotation time 180º+ or 360º scan conventional CT or DSCT at 60 bpm require < 200 ms possible with 0.4 s rotation time and 180º+ scan 180º+ scan 360º scan 6

7 determined by Spatial resolution in-plane: reconstruction kernel out-of-plane: slice thickness require sharp reconstruction kernels increase spatial resolution, but increase image noise < 1 mm slices increase spatial resolution, but increase image noise Corresponding CT and conventional angiogram images of RCA 7

8 Contrast resolution determined by difference in HU between arteries and background kv setting partial volume effects image noise require contrast enhancement low kv to increase sensitivity to iodine, but image noise and likelihood of photon starvation increase high maximum tube ma for use with lower kv settings < 1 mm slices to reduce partial volume effects, but image noise increases 8

9 determined by Coverage per rotation width of detector array with thin detector elements require > 30 mm of sub-mm detector rows 4-5 rotations to cover heart all data acquisition during optimal contrast enhancement image-stitching artefacts reduced 9

10 The modern cardiac CT scanner 10

11 The basic model: cardiac-enabled 64-DCT scanners* supplier example scanner model detector array rows x element mm z-coverage mm slices per rotation x thinnest slice mm minimum rot s GE Optima CT x x Siemens Definition Edge 64 x x Philips Brilliance x x Toshiba Aquilion CXL 64 x x NB manufacturers with z-ffs or detector resampling double number of slices per rotation * NB scanner specifications are continually changing 11

12 Top MDCT scanners 12

13 Revolution series: Rev CT 512-slice scanner 256 detector rows with resampling improved anatomical coverage 160 mm detector covers heart improved spatial resolution new x-ray tube HD mode using xy-ffs improved contrast resolution Gemstone Clarity detector in 3 sections increased detector efficiency lower detector afterglow image courtesy of GE 13

14 Brilliance ict / ict Elite 256-slice scanner 128 detector rows with z-ffs improved anatomical coverage 80 mm detector faster rotation time 0.27 s with AirGlide technology and imrc tube images courtesy of Philips 14

15 Definition Force 2 imaging chains mounted at ~94º 50 and 35 cm FOV improved temporal resolution faster acquisition of projection data set improved anatomical coverage 96 detector rows, 57.6 mm Turbo Flash mode high pitch helical mode image courtesy of Siemens 15

16 Aquilion One ViSION 640-slice scanner 320 detector rows with double sampling improved anatomical coverage 160 mm detector covers heart volume mode no table feed image courtesy of Toshiba 16

17 Top MDCT scanners: summary* supplier scanner model detector design rows x element mm z-coverage mm slices per rotation x thinnest slice mm minimum rot s GE Siemens Revolution CT Definition Force 256 x x x 0.6 x x Philips ict / Elite 128 x x Toshiba Aquilion One Vision 320 x x NB manufacturers with z-ffs or detector resampling double number of slices per rotation *from manufacturers information. NB scanner specifications are continually changing 17

18 ECG gating technology 18

19 ECG gating of acquisition essentially two types: prospective retrospective different image sets generated different radiation burden to patient 19

20 Prospective gating acquisition triggered by ECG set time after R wave acquisition window matched to minimum acquisition time no padding acquisition image sets in selected phase acquisition window extended beyond minimum acquisition time with padding acquisition at full or reduced ma image sets in 1+ selected phases can adjust for heart rate changes ma ma prescribed mas prescribed mas t t padding mas 20

21 Prospective gating choice of table motions step-and-shoot slow helical high-pitch helical (Flash) none axial images reconstructed from partial projection data sets MPRs, VR image sets patient dose 1 5 msv CXRs partial projection data set 21

22 Prospective gating Flash helical mode LAD with stent Image courtesy of Universität Ulm, Germany, and Philips images courtesy of Institute of Medical Physics / Erlangen, Germany, and Siemens 22

23 Retrospective gating long acquisition fast rotation time choice of table motion slow helical none pitch typically projections tagged with ECG signal ECG-gated ma pulsing selectable temporal window position and width restricts reconstruction to selected cardiac phase(s) 23

24 Retrospective gating images reconstructed from partial projection data set to improve temporal resolution single or multi-segment recon determines temporal resolution reconstruction window can be adjusted essential for arrhythmic patients typical patient dose msv CXRs single-segment recon multi-segment recon 24

25 Retrospective gating coronary CTA from images reconstructed at one cardiac phase cardiac function from images reconstructed at several cardiac phases Functional Analysis, EF 80% images courtesy of Dr Rybicki, BWH, and Toshiba 25

26 Dealing with irregular heart rates arrhythmia detection algorithms acquisition delayed until rhythm re-established compensation for heart rate changes acquisition window automatically widened to capture required phase(s) may impact on radiation dose to patient 26

27 Image reconstruction algorithms 27

28 Filtered back-projection choice of convolution kernels images courtesy of Willi Kalender 28

29 Iterative reconstruction from Beister et al

30 IR: vendor solutions supplier IR algorithms GE ASiR Veo ASiR-V Siemens SAFIRE ADMIRE Philips idose4 IMR Toshiba AIDR 3D proprietary black boxes! 30

31 IR solutions in raw data and image domains used by Philips (idose), Siemens and Toshiba based on statistical modelling of x-ray detection from Willemink et al

32 IR solutions in raw data domain used by GE, Philips (IMR) based on statistical modelling of x-ray detection geometric modelling of imaging chain (Veo) physics modelling (ASiR-V) from Willemink et al

33 Iterative reconstruction images presented different IR strengths mixed with FBP lower image noise than FBP opportunity for dose reduction FBP SAFIRE images courtesy of Siemens 33

34 Iterative reconstruction slower recon than FBP IR-specific artefacts may be introduced disappearance of hypodense seam from Beister et al

35 Conclusions coronary CT angiography requires cardiacenabled scanners 64+ DCT choice of ECG gating techniques iterative reconstruction algorithms should be carefully validated 35

36 Acknowledgements Dee Mears, RMH Nina Arcuri, RBH Willi Kalender ImPACT group Philips Siemens Toshiba GE 36

Iterative Reconstruction

Iterative Reconstruction RECENT ADVANCES IN CT RADIATION DOSE REDUCTION TECHNIQUES Iterative Reconstruction Kalpana Kanal, PhD, FSCBTMR, FACR, FAAPM Professor and Director, Diagnostic Physics Section University of Washington Seattle,

More information

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National

More information

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature

More information

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose Image Quality and Dose Issues in MSCT Image Quality and Dose Image quality Image noise Spatial resolution Contrast Artefacts Speckle and sharpness S. Edyvean St. George s Hospital London SW17 0QT Radiation

More information

Iterative Reconstruction in Image Space. Answers for life.

Iterative Reconstruction in Image Space. Answers for life. Iterative Reconstruction in Image Space Answers for life. Iterative Reconstruction in Image Space * (IRIS) * Please note: IRIS is used as an abbreviation for Iterative Reconstruction in Image Space throughout

More information

1. Patient size AEC. Large Patient High ma. Small Patient Low ma

1. Patient size AEC. Large Patient High ma. Small Patient Low ma Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description

More information

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

More information

Wide-Detector CT for TAVR Planning:

Wide-Detector CT for TAVR Planning: Wide-Detector CT for TAVR Planning: Impact on Iodine Dose, Radiation Dose, and Image Quality SCBTMR 2015 Annual Course Thursday, October 8 William P. Shuman MD FSCBTMR Department of Radiology University

More information

160-slice CT SCANNER / New Standard for the Future

160-slice CT SCANNER / New Standard for the Future TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

Multislice CT: Current Technology and Future Developments

Multislice CT: Current Technology and Future Developments Multislice CT: Current Technology and Future Developments 1 Stefan Ulzheimer and Thomas Flohr Contents 1.1 Introduction 3 1.2 System Design 5 1.2.1 Gantry 6 1.2.2 X-Ray Tube and Generator 6 1.2.3 MDCT

More information

SAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life.

SAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life. Neuro Thoracic Abdominal Abdominal Cardiovascular Pediatric SAFIRE Sinogram Affirmed Iterative Reconstruction Answers for life. SAFIRE * (Sinogram Affirmed Iterative Reconstruction) * The information

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients

More information

Detector technology in simultaneous spectral imaging

Detector technology in simultaneous spectral imaging Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become

More information

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Poster No.: C-2205 Congress: ECR 2012 Type: Scientific Paper

More information

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT

More information

Inside Biograph mct.

Inside Biograph mct. Inside Biograph mct The technologies behind the world s first molecular CT. www.siemens.com/mi Large 78 cm bore helps reduce claustrophobia and provides more room for RTP positioning devices. 227 kg (500

More information

Maximum Performance, Minimum Space

Maximum Performance, Minimum Space TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

The disclaimer on page 1 is an integral part of this document. Copyright November 30, 2017 by AAPM. All rights reserved.

The disclaimer on page 1 is an integral part of this document. Copyright November 30, 2017 by AAPM. All rights reserved. DISCLAIMER: TO THE EXTENT ALLOWED BY LOCAL LAW, THIS INFORMATION IS PROVIDED TO YOU BY THE AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A NON-PROFIT ORGANIZATION ORGANIZED TO PROMOTE THE APPLICATION

More information

abc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice nww.medical-devices.nhs.

abc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice   nww.medical-devices.nhs. abc September 2004 MHRA 04099 Philips Mx8000 IDT CT scanner technical evaluation Best choice best practice www.mhra.gov.uk nww.medical-devices.nhs.uk About MHRA evaluation reports. What you can expect.

More information

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity) Vascular Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction In recent years, digital cardiovascular

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL

TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL TORNIER BLUEPRINT 3D Planning + PSI SCAN PROTOCOL Contents 3 Introduction 3 Patient preparation 3 Scanning instructions 4 Image instructions 5 Scanning parameters 6 Technical instructions 2 BLUEPRINT 3D

More information

Improvement of CT image quality with iterative reconstruction idose4

Improvement of CT image quality with iterative reconstruction idose4 Improvement of CT image quality with iterative reconstruction idose4 Poster No.: C-0387 Congress: ECR 2014 Type: Scientific Exhibit Authors: M.-L. Olsson, K. Norrgren, M. Söderberg; Malmö/SE Keywords:

More information

Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11

Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11 Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) Hiroki Kadoya Yukiko Kitagawa MEDIX VOL. 61 P.8 P.11 Clinical Experience Using the Open Bore Multislice CT System Supria

More information

Aquilion Precision Ultra-High Resolution CT: Quantifying diagnostic image quality

Aquilion Precision Ultra-High Resolution CT: Quantifying diagnostic image quality Aquilion Precision Ultra-High CT: Quantifying diagnostic image quality Kirsten Boedeker, PhD, DABR Senior Manager, Quantitative Image Quality Canon Medical Systems Corporation Introduction Over the last

More information

PET: New Technologies & Applications, Including Oncology

PET: New Technologies & Applications, Including Oncology PET: New Technologies & Applications, Including Oncology, PhD, FIEEE Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Disclosures Research Contract, GE Healthcare

More information

NeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results

NeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results NeuViz 16 Computed Tomography Elevating routine imaging for exceptional results Essence NeuViz 16 Raising the bar on clinical utility in routine imaging. Get more. More clinical information for patients.

More information

Preclinical X-Ray Computed Tomography

Preclinical X-Ray Computed Tomography Preclinical X-Ray Computed Tomography Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de/ct Is CT a Molecular Imaging Modality? Imaging Modality Molecular Sensitivity

More information

Brilliance in everything Philips CT products and services

Brilliance in everything Philips CT products and services Brilliance in everything Philips CT products and services Ready for anything No one does more than Philips to help you gain the productivity you need with a comprehensive approach to CT that marries significant

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Dr John Ridgway Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Physics for clinicians: Part I Journal of Cardiovascular Magnetic Resonance 2010, 12:71 http://jcmr-online.com/content/12/1/71

More information

PET/CT Instrumentation Basics

PET/CT Instrumentation Basics / Instrumentation Basics 1. Motivations for / imaging 2. What is a / Scanner 3. Typical Protocols 4. Attenuation Correction 5. Problems and Challenges with / 6. Examples Motivations for / Imaging Desire

More information

Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise

Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise Jonas Rydberg, M.D. Professor of Radiology Indiana University School of Medicine Indianapolis, Indiana Medical

More information

diagnostic examination

diagnostic examination RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC? Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM

More information

Translating Protocols Between Scanner Manufacturer and Model

Translating Protocols Between Scanner Manufacturer and Model Translating Protocols Between Scanner Manufacturer and Model Robert J. Pizzutiello, MS, FAAPM, FACMP Sr. Vice-President, Global Physics Solutions President, Upstate Medical Physics Objectives Understand

More information

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology The Fundamentals of... Computed Tomography Computed Tomography (CT) systems use x-rays to produce images of slices through a patient s anatomy. Despite having lower spatial resolution than other x-ray

More information

2D, 3D CT Intervention, and CT Fluoroscopy

2D, 3D CT Intervention, and CT Fluoroscopy 2D, 3D CT Intervention, and CT Fluoroscopy SOMATOM Definition, Definition AS, Definition Flash Answers for life. Siemens CT Vision Siemens CT Vision The justification for the existence of the entire medical

More information

Data. microcat +SPECT

Data. microcat +SPECT Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893 Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -

More information

Dose Reduction in Helical CT: Dynamically Adjustable z-axis X-Ray Beam Collimation

Dose Reduction in Helical CT: Dynamically Adjustable z-axis X-Ray Beam Collimation Medical Physics and Informatics Original Research Christner et al. CT Dose Reduction Medical Physics and Informatics Original Research Downloaded from www.ajronline.org by 8.243.133.8 on 2/26/18 from IP

More information

QC by the MPE in Belgium

QC by the MPE in Belgium Acceptance testing of state-of-the-art CT scanners using a new national protocol: first experience on a large number of scanners of different make and model the working group Radiology of the Belgian Hospital

More information

PARTIAL-DATA INTERPOLATION DURING ARCING OF AN X-RAY TUBE IN A COMPUTED TOMOGRAPHY SCANNER JAISINGH RAJWADE

PARTIAL-DATA INTERPOLATION DURING ARCING OF AN X-RAY TUBE IN A COMPUTED TOMOGRAPHY SCANNER JAISINGH RAJWADE PARTIAL-DATA INTERPOLATION DURING ARCING OF AN X-RAY TUBE IN A COMPUTED TOMOGRAPHY SCANNER JAISINGH RAJWADE Bachelor of Engineering (B.E.) Industrial Electronics University of Pune, India November 1997

More information

CT Basics: Image Quality Module 6

CT Basics: Image Quality Module 6 Module 6 For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions. Any broadcast,

More information

SOMATOM Esprit A Bundle of Energy

SOMATOM Esprit A Bundle of Energy SOMATOM Esprit A Bundle of Energy DATA SOMATOM Esprit An economical CT scanner designed for...... Excellent spiral image quality... A wide range of clinical applications... Value performance and reliabilty

More information

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT) Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in

More information

DOUBLE gated in vivo small animal cone beam micro

DOUBLE gated in vivo small animal cone beam micro Low Dose Phase Correlated Cone Beam Micro CT of Small Animals Stefan Sawall, Frank Bergner, Robert Lapp, Markus Mronz, Marek Karolczak, Andreas Hess, and Marc Kachelrieß, Member, IEEE I. INTRODUCTION DOUBLE

More information

SPECT Reconstruction & Filtering

SPECT Reconstruction & Filtering SPECT Reconstruction & Filtering Goals Understand the basics of SPECT Reconstruction Filtered Backprojection Iterative Reconstruction Make informed choices on filter selection and settings Pre vs. Post

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

Interventional X-ray quality measure based on a psychovisual detectability model

Interventional X-ray quality measure based on a psychovisual detectability model Interventional X-ray quality measure based on a psychovisual detectability model Asli Kumcu, Benhur Ortiz-Jaramillo, Ljiljana Platisa, Bart Goossens, Wilfried Philips iminds-telin-ipi, Ghent University,

More information

WE MAKE RELATIONSHIP FOR LIFE.

WE MAKE RELATIONSHIP FOR LIFE. 2018 WE MAKE RELATIONSHIP FOR LIFE. www.sanrad.in Contents Corporate Profile 02 Company Profile Rich in heritage and synonymous with healthcare, the field of medicine is deeply ingrained in Sanrad s roots

More information

Architecture of Quality Imaging Mary K. Henne, MS, CNMT, RDMS, RVT Ultrasound Education Specialist GE Healthcare

Architecture of Quality Imaging Mary K. Henne, MS, CNMT, RDMS, RVT Ultrasound Education Specialist GE Healthcare Architecture of Quality Imaging Mary K. Henne, MS, CNMT, RDMS, RVT Ultrasound Education Specialist GE Healthcare 2 DOC1292532 Architecture of Quality Imaging Agile Acoustic Architecture E-Series and XDclear

More information

Truly flexible to meet your clinical needs

Truly flexible to meet your clinical needs Truly flexible to meet your clinical needs 2 Adapting to meet your needs Flexible Fast and responsive Excellent image quality Designed with ergonomic efficiency Equipped with dose management tools 3 Three

More information

NIH Public Access Author Manuscript Int J Cardiovasc Imaging. Author manuscript; available in PMC 2008 May 26.

NIH Public Access Author Manuscript Int J Cardiovasc Imaging. Author manuscript; available in PMC 2008 May 26. NIH Public Access Author Manuscript Published in final edited form as: Int J Cardiovasc Imaging. 2001 August ; 17(4): 287 296. A comparison of prospective and retrospective respiratory navigator gating

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

INNOVATION BY DESIGN. Toshiba A History of Leadership REMOTE CONTROL R/F SYSTEM

INNOVATION BY DESIGN. Toshiba A History of Leadership REMOTE CONTROL R/F SYSTEM INNOVATION BY DESIGN For over 130 years, Toshiba has led the world in developing technology to improve the quality of life. This Made for Life TM commitment is reflected in our family of leading-edge imaging

More information

CT Basics: Data Acquisition Module 3

CT Basics: Data Acquisition Module 3 Module 3 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

Digital Radiographic Inspection replacing traditional RT and 3D RT Development

Digital Radiographic Inspection replacing traditional RT and 3D RT Development Digital Radiographic Inspection replacing traditional RT and 3D RT Development Iploca Novel Construction Meeting 27&28 March 2014 Geneva By Jan van der Ent Technical Authority International Contents Introduction

More information

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation Image Quality Review I: Basics and Image Quality TH-A-16A-1 Thursday 7:30AM - 9:30AM Room: 16A J. Anthony

More information

Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution

Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution Cynthia H. McCollough, PhD, DABR, FAAPM, FACR Director, CT Clinical Innovation Center Professor of Medical Physics and

More information

Optimized CT metal artifact reduction using the Metal Deletion Technique (MDT)

Optimized CT metal artifact reduction using the Metal Deletion Technique (MDT) Optimized CT metal artifact reduction using the Metal Deletion Technique (MDT) F Edward Boas, Roland Bammer, and Dominik Fleischmann Extended abstract for RSNA 2012 Purpose CT metal streak artifacts are

More information

Multi-Access Biplane Lab

Multi-Access Biplane Lab Multi-Access Biplane Lab Advanced technolo gies deliver optimized biplane imaging Designed in concert with leading physicians, the Infinix VF-i/BP provides advanced, versatile patient access to meet the

More information

Photomultiplier Tube

Photomultiplier Tube Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or

More information

of sufficient quality and quantity

of sufficient quality and quantity of sufficient quality and quantity The patient s body attenuates the beam as it passes though the body More energy is deposited in organs located near the entry of the beam than near the exit of the beam

More information

MR Advance Techniques. Flow Phenomena. Class II

MR Advance Techniques. Flow Phenomena. Class II MR Advance Techniques Flow Phenomena Class II Flow Phenomena In this class we will explore different phenomenona produced from nuclei that move during the acquisition of data. Flowing nuclei exhibit different

More information

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Val M. Runge, MD Institute for Diagnostic and Interventional Radiology Clinics for Neuroradiology and Nuclear Medicine University Hospital Zurich

More information

New Technology in Nuclear Medicine

New Technology in Nuclear Medicine New Technology in Nuclear Medicine Reed G. Selwyn, PhD, DABR Vice Chair of Research & Imaging Sciences Associate Professor and Chief, Medical Physics Dept. of Radiology, University of New Mexico Objectives

More information

High Field MRI: Technology, Applications, Safety, and Limitations

High Field MRI: Technology, Applications, Safety, and Limitations High Field MRI: Technology, Applications, Safety, and Limitations R. Jason Stafford, Ph.D. The University of Texas M. D. Anderson Cancer Center, Houston, TX Introduction The amount of available signal

More information

Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques

Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques Wataru FUKUDA* Junya MORITA* and Masahiko YAMADA* Abstract Tomosynthesis is a three-dimensional imaging technology

More information

January 2012 February 1, Release Date: Expiration Date:

January 2012 February 1, Release Date: Expiration Date: CT for Technologists is a training program designed to meet the needs of radiologic technologists entering or working in the field of computed tomography (CT). This series is designed to augment classroom

More information

Wide beam CT dosimetry. Elly Castellano

Wide beam CT dosimetry. Elly Castellano Wide beam CT dosimetry Elly Castellano Outline revision: CT dose indices wide-beam CT: the end of the road for CTDI? the IEC rescue plan for CTDI 100 the american way AAPM report 111 better estimates of

More information

CT Basics: Computed Tomography Fundamentals Module 1

CT Basics: Computed Tomography Fundamentals Module 1 Module 1 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

Distributed source x-ray tube technology for tomosynthesis imaging

Distributed source x-ray tube technology for tomosynthesis imaging Distributed source x-ray tube technology for tomosynthesis imaging Authors: F. Sprenger a*, X. Calderon-Colon b, Y. Cheng a, K. Englestad a, J. Lu b, J. Maltz c, A. Paidi c, X. Qian b, D. Spronk a, S.

More information

Quantitation of clinical feedback on image quality differences between two CT scanner models

Quantitation of clinical feedback on image quality differences between two CT scanner models Received: 4 August 2016 Revised: 4 November 2016 Accepted: 12 December 2016 DOI: 10.1002/acm2.12050 MEDICAL IMAGING Quantitation of clinical feedback on image quality differences between two CT scanner

More information

Evaluation Report. Single Slice CT Scanner Comparison Report Version (Free to the NHS) ImPACT report. MDA Evaluation Report MDA 03022

Evaluation Report. Single Slice CT Scanner Comparison Report Version (Free to the NHS) ImPACT report. MDA Evaluation Report MDA 03022 February 2003 Evaluation Report NUMBER MDA 03022 Single Slice CT Scanner Comparison Report Version 8 ImPACT report MDA Evaluation Report MDA 03022 Crown Copyright 100 (Free to the NHS) WHAT YOU CAN EXPECT

More information

P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s

P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s AN ALL AROUND PERFECT PICTURE. The perfect combination of image quality, efficiency and design. P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s Air Techniques experts in digital imaging ProVecta

More information

SOMATOM Sensation 4 Computed Tomography System for Multislice Spiral Scanning

SOMATOM Sensation 4 Computed Tomography System for Multislice Spiral Scanning SOMATOM Sensation 4 Computed Tomography System for Multislice Spiral Scanning SOMATOM Sensation 4 Computed Tomography System for Multislice Spiral Scanning Siemens has been a global player in CT for more

More information

X-RAY COMPUTED TOMOGRAPHY

X-RAY COMPUTED TOMOGRAPHY X-RAY COMPUTED TOMOGRAPHY Bc. Jan Kratochvíla Czech Technical University in Prague Faculty of Nuclear Sciences and Physical Engineering Abstract Computed tomography is a powerful tool for imaging the inner

More information

X-ray detectors in healthcare and their applications

X-ray detectors in healthcare and their applications X-ray detectors in healthcare and their applications Pixel 2012, Inawashiro September 4th, 2012 Martin Spahn, PhD Clinical applications of X-ray imaging Current X-ray detector technology (case study radiography

More information

P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s

P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s AN ALL AROUND PERFECT PICTURE. The perfect combination of image quality, efficiency and design. P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s Air Techniques experts in digital imaging ProVecta

More information

CT Basics: Equipment and Instrumentation Module 2

CT Basics: Equipment and Instrumentation Module 2 Module 2 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational in-house or online educational course use only in educational and corporate institutions.

More information

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging JOURNAL OF MAGNETIC RESONANCE IMAGING 20:1046 1051 (2004) Technical Note The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging James W. Goldfarb, PhD* Purpose: To describe a known (but undocumented)

More information

Measurement of table feed speed in modern CT

Measurement of table feed speed in modern CT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 3, 2014 Measurement of table feed speed in modern CT Atsushi Fukuda, 1,2a Pei-Jan P. Lin, 3 Kosuke Matsubara, 2 Tosiaki Miyati 2 Department

More information

Industry Breakthrough

Industry Breakthrough Industry Breakthrough Dynamic SPECT Acquisition Quantifying Myocardial Blood Flow Nuclear Cardiology in the 21st Century In the 21st century, most nuclear cameras are still relying on a technology invented

More information

Philips XPER FD10C R7.0.4

Philips XPER FD10C R7.0.4 Philips XPER FD10C R7.0.4 Reconditioned 2005 System- Upgraded to R7 in Oct 2010 The Allura Xper FD10 (Ceiling) single-plane cardiovascular system is comprised of a ceiling mounted C-arm stand and digital

More information

1. Queries are issued to the image archive for information about computed tomographic (CT)

1. Queries are issued to the image archive for information about computed tomographic (CT) Appendix E1 Exposure Extraction Method examinations. 1. Queries are issued to the image archive for information about computed tomographic (CT) 2. Potential dose report screen captures (hereafter, dose

More information

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016 Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016 Current technology in digital image production (CR/DR and other modalities) 2/ Overview Digital

More information

Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2

Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2 Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2 Medical Consult, SA; Establish a method to correlate image

More information

Resona 6 Premium Ultrasound System

Resona 6 Premium Ultrasound System The system Resona 6 is the newly developed, unique result of the mergence of leading companies Mindray Bio-medical Electronics Co. Ltd. and ZONARE Medical Systems, Inc.. By additions to the core competencies

More information

William Hallet - PRIMA IV 1

William Hallet - PRIMA IV 1 Quantitative and application specific imaging PET: the measurement process,reconstruction, calibration, quantification Dr William Hallett Centre for Imaging Sciences Imperial College Hammersmith Hospital

More information

Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing. Koichi Hirokawa MEDIX VOL. 56 P.43 P.46

Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing. Koichi Hirokawa MEDIX VOL. 56 P.43 P.46 Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing Taiga Goto Koichi Hirokawa Hisashi Takahashi MEDIX VOL. 56 P.43 P.46 Advanced Noise Reduction Processing for X-ray CT System

More information

23 CP Clarify Enhanced US Volume Image and Frame Type Values 3 and 4

23 CP Clarify Enhanced US Volume Image and Frame Type Values 3 and 4 23 CP-1463 - Clarify Enhanced US Volume Image and Frame Type Values 3 and 4 Page 1 1 Status Finale Text 2 Date of Last Update 2015/11/10 3 Person Assigned David Clunie 4 mailto:dclunie@dclunie.com 5 Submitter

More information

Computerized Medical Imaging and Graphics

Computerized Medical Imaging and Graphics Computerized Medical Imaging and Graphics 36 (2012) 387 395 Contents lists available at SciVerse ScienceDirect Computerized Medical Imaging and Graphics jo ur n al homep age : www.elsevier.com/locate/compmedimag

More information

23 CP Clarify Enhanced US Volume Image and Frame Type Values 3 and 4

23 CP Clarify Enhanced US Volume Image and Frame Type Values 3 and 4 23 CP-1463 - Clarify Enhanced US Volume Image and Frame Type Values 3 and 4 Page 1 1 Status Letter Ballot 2 Date of Last Update 2015/09/16 3 Person Assigned David Clunie 4 mailto:dclunie@dclunie.com 5

More information

GE Healthcare. Discovery MI. with LightBurst Digital 4-Ring Detector

GE Healthcare. Discovery MI. with LightBurst Digital 4-Ring Detector GE Healthcare Discovery MI with LightBurst Digital 4-Ring Detector Discovery TM MI is a PET/CT system conceptualized with lofty goals, equal only to your own. It was created to help you diagnose and stage

More information

The future of nuclear imaging is clear

The future of nuclear imaging is clear Cardius X-ACT The future of nuclear imaging is clear Increased regulations, growing competition, and concerns about radiation exposure are just a sampling of the current challenges facing the nuclear medicine

More information