1 Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T. Nakamura, N Amimoto, K. Okano, M. Yamada, T. Kawamura, S. Naito ; Osaka/JP, Tokyo/JP, Kanagawa/JP Keywords: Computer applications, Radioprotection / Radiation dose, Anatomy, Digital radiography, Computer Applications-Detection, diagnosis, Computer Applications-General, Radiation effects, Workforce DOI: /ecr2015/C-1834 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. Page 1 of 25
2 Aims and objectives In Japan, x-ray examination of the chest region is carried out with portable equipment on medical wards for the management of the critically-ill patient. These patients whose condition is so severe that they cannot be transferred to an x-ray room, have a portable chest x-ray examination from which the diagnostic information is vital in the patient's treatment. Generally, when performing the portable chest x-ray examination the thicker the subject, the more scattered the x-ray beam will be in the x-ray exposure and the greater its effect on image quality. Some departments take the approach of reducing the kvp used for the portable chest examination thereby reducing the scatter radiation that reaches the image receptor. However, this also reduces the energy of the x-ray photons and the penetration of structures in the chest, resulting in reduced diagnostic information. In some cases, a secondary radiation grid is used with the higher kvp. The higher kvp maintains the penetration of structures in the chest region and the grid reduces the influence of the scattered x-rays, thereby improving image quality and diagnostic information. However, using a grid in a portable environment presents workflow challenges such as carrying the grid, positioning the grid and FPD perpendicular to the x-ray beam. It also creates its own image quality issues in the way of artefacts from the misalignment of grid and x-ray beam (often referred to as "grid cut-off") which can also result in reduced diagnostic information. Due to these reasons, many hospitals do not use a grid in portable chest examinations performed on wards, opting to reduce the kvp used and accept the compromise between image quality (diagnostic value) and acceptable practical workflow. The Osaka General Medical Center currently uses CR for all portable chest x-ray examinations. No grid is used for portable chest x-ray examinations, but a low tube voltage of 65kVp is used to reduce the effect of scattered x-rays. In conjunction with the newly introduced flat panel detector (FPD) made by Fujifilm for portable examination at Osaka General Medical Center a study was undertaken looking at how newly developed scatter removal processing (Virtual Grid/VG) could be used to improve image quality in portable chest examinations with an associated dose reduction when compared to using a standard secondary radiation grid (+). The results of this study are presented here. Scattered x-ray causes a reduction of contrast and increased granularity in an image. Virtual Grid processing has [Contrast Improvement Processing] and [Granularity Improvement Processing] to improve both contrast and granularity (Fig.1). Contrast Improvement Processing estimates scattered x-ray from the input image and reduces the Page 2 of 25
3 scattered x-ray in a ratio similar to a normal secondary radiation grid to improve image contrast. In addition, since scattered x-ray contains almost no information related to the subject structure, Granularity Improvement Processing extracts the granular components not related to the structure of the subject from the image and reduces the extracted components, improving structural definition in the image. An example image with Virtual Grid processing applied is shown in Fig.2. The left image is an image exposed without a grid (-), the middle image is an image exposed with a secondary radiation grid (+) with a grid ratio of 8:1. When these two images are compared, it is obvious that the left image has less contrast due to scattered x-ray. The right image is the same image as shown on the left but with Virtual Grid processing applied that corresponds to a grid ratio of 8:1 (as used in the middle image). As can be seen the contrast of the non-grid (-) image (left) has been improved to the level of that shown in the grid image (middle) by application of Virtual Grid processing. Images for this section: Fig. 1 Page 3 of 25
4 Fig. 2 Page 4 of 25
5 Methods and materials For the study, a Caesium Iodide FPD was used for image acquisition. Two Image sets were acquired, one set using a secondary radiation grid (+) exposure technique and the other set using a non-grid (-) exposure technique. The non-grid (-) images were then subjected to Virtual Grid processing. All image sets were evaluated using 3 methods. [1] Physical evaluation A contrast phantom (a 1cm-thick-PMMA block with a 2cm x 2cm square shaped hole) was placed between PMMA blocks with a thickness of 4cm and 4.5cm (9.5cm in total), and 7cm and 6.5cm (14.5cm in total). X-ray tube voltages of 60kVp and 80kVp, were used and a target exposure dose of 1mR at the back of the phantom was set as 100% (Table 1). Image sets were acquired with a grid (+) and without a grid (-) at 100%, 50% and 25% dose levels for both 60kVp and 80kVp. Virtual Grid processing corresponding to grid ratios of 6:1 was applied to the images acquired without a grid (-). The characteristics of contrast (C = D0#D10), the characteristics of noise (N = (SD0#SD10) / 2) and the contrastnoise ratio (CNR = C / N) were calculated setting the average pixel value of the 2cm x 2cm square shaped hole (0cm thick) of the contrast phantom and the area of the same size (1cm thick) next to the hole as D0 and D10, and the standard variation as SD0 and SD10. [2] Visual evaluation As shown in Fig.3, A CDRAD phantom was placed between PMMA blocks with thicknesses of 4cm and 4.5cm (9.5cm in total), and 7cm and 6.5cm (14.5cm in total) (Fig.3). X-ray tube voltages of 60kVp and 80kVp were used and a target exposure dose of 1mR at the back of the phantom was set as 100%. Image sets were acquired with a grid (+) and without a grid (-) at 100%, 50% and 25% dose levels (Table 1) for both 60kVp and 80kVp. (Table.1). Virtual Grid processing corresponding to grid ratio of 6:1 was applied to the images acquired without a grid (-). The CDRAD phantom images were analysed using CDRAD Analyser software ver. 2.1 (manufactured by Artinis Medical Systems). Contrast Detail curves were produced, and the IQFinv calculated for each image (Fig.4). The higher the IQFinv value, the greater the visibility of the details within the image. [3] Chest phantom evaluation Using a chest phantom, an x-ray tube voltage of 80kVp and the exposure dose of 1.8mAs, image sets were acquired with a 6:1 ratio grid (+) and without a Grid (-). Virtual Grid processing corresponding to a grid ratio of 6:1 was applied to the images acquired without a grid (-). Virtual Grid processed images and images exposed with a grid (+) were compared using the Kyoto University method* (one of the representative methods Page 5 of 25
6 in Japan used to assess the anatomical index and the physical factor of the chest region) below: Peripheral side of the lung field (resolution of peripheral blood vessel in S2 at upper right lung and resolution of the peripheral blood vessel of costophrenic region in S8 at bottom right lung). Mediastinum side of lung field (resolution of upper left lung blood vessel and bottom right lung blood vessel, and outline of the bronchial tube). Mediastinum region (Outline of trachea and the main bronchial tubes (left and right), and resolution of each margin of the mediastinum region). Granularity (soft tissue under shoulder blade). Sharpness (trabecular of collar bone and margin of rib bone). Contrast (balance of entire density). For physical and visual evaluation, we used the optimized Virtual Grid parameter to PMMA on account of correcting the different scatter characteristics from human tissue. [Materials] #Diagnostic x-ray high voltage unit: UD150B-40 (manufactured by SHIMADZU) #Ion chamber dosimeter: 2026C (manufactured by TOYO MEDIC) #FPD System: CALNEO C 1417 Wireless SQ (manufactured by FUJIFILM) and CONSOLE ADVANCE Ver.8.0 (manufactured by FUJIFILM) #Phantom: Contrast phantom (scratchbuilt), CDRAD phantom (manufactured by Artinis Medical Systems) and Page 6 of 25
7 Chest phantom (manufactured by KYOTO KAGAKU) #Analysing software: Image J and CDRAD Analyser Version 2.1 (manufactured by Artinis Medical Systems) Images for this section: Table 1 Page 7 of 25
8 Fig. 3 Page 8 of 25
9 Fig. 4 Page 9 of 25
10 Results [1] Physical evaluation The level of image contrast for increasing thickness of PMMA at 60kVp and 100% dose is shown in Fig.5. Images exposed without a grid (-), had the lowest levels of contrast when compared to the contrast of images exposed with a grid (+) and the non-grid (-) images processed by Virtual Grid processing. In addition, it was found that the Virtual Grid processed images had higher contrast levels than images exposed with a grid (+). Although the contrast levels of images exposed with a grid (+) and without a grid (-) remarkably decreased as the thickness of PMMA increased from 9.5cm to 14.5cm, the contrast levels of images processed by Virtual Grid were more stable and did not decrease to the same extent. The level of image contrast for increasing thickness of PMMA at 80kVp and 100% dose is shown in Fig.6. As with the results for 60kVp, the contrast of images exposed with a grid (+) and the non grid (-) images processed by Virtual Grid were improved compared with the images exposed without a grid (-). As well the contrast levels of Virtual Grid processing images were higher and more stable and did not decrease to the same extent over the increased PMMA thickness as the non grid (-) images and grid (+) images. The noise characteristics of images acquired at the x-ray tube voltage of 60kVp and 100% exposure dose are shown in Fig.7. The graph shows the noise value (N = (SD0#SD10) / 2) against increasing thickness of PMMA. Because the use of a grid (+) decreases the x-ray dose reaching the FPD, the granularity is greater and the noise value is higher; therefore, images exposed with a grid (+) showed significantly higher noise values than images exposed without a grid (-). The characteristics of the noise in the non-grid (-) images processed with Virtual Grid processing was superior to images exposed with a grid (+) and closer to those values of the non-grid (-) images. At the 9.5cm PMMA thickness, the characteristics of the noise in images processed by Virtual Grid processing were slightly better than images exposed without a grid (-). Fig.8 shows the noise characteristics for images acquired at 80kVp and 100% exposure dose. As with the results for 60kVp and 100% exposure dose, images exposed with a grid (+) showed higher noise values when compared to images exposed without a grid (-). The characteristics of the noise in the non-grid (-) images processed with Virtual Grid processing was superior to images exposed with a grid (+) and again closer to those values of the non-grid (-) images. Page 10 of 25
11 CNR results for images acquired with the x-ray tube voltage of 60kVp and 100% exposure dose are shown in Fig.9. As expected CNR decreases for all image sets as the PMMA thickness increases. However, the CNR of the non-grid (-) images processed by Virtual Grid processing showed significantly higher CNR values compared with images exposed with a grid (+) and without a grid (-). The CNR results for images acquired with the 80kVp x-ray tube voltage and 100% exposure dose are shown in Fig.10. Although the overall CNR values decreased compared with the images acquired at 60kVp and 100% dose, the images processed by Virtual Grid processing still show significantly higher CNR than images exposed with a grid (+) and without a grid (-). Fig.11 shows a comparison of the CNR for images acquired with a grid (+) at 100% dose level, and non-grid images processed with Virtual Grid processing acquired at 3 dose levels (100%, 50% and 25%). The CNR in images processed by Virtual Grid processing even at 25% of the original dose was higher than images exposed with a grid at 100% dose. [2] Visual evaluation The CDRAD IQFinv results for 9.5cm PMMA thickness, x-ray tube voltages of 60kVp and 80kVp are shown in Fig.12. The x axis shows the dose difference before and after transmission through the phantom as absorbed x-ray dose. When images acquired at the x-ray tube voltages of 60kVp and 80kVp were compared with identical x-ray absorption, images processed by Virtual Grid processing showed significantly higher IQFinv values meaning improved detail visibility than in images exposed with a grid (+) and without a grid (-). The CDRAD IQFinv results for 14.5cm PMMA thickness, x-ray tube voltages of 60kVp and 80kVp are shown in Fig.13. Although the thickness of PMMA is increased, non-grid (-) images processed by Virtual Grid processing again showed higher IQFinv values and therefore improved detail visibility than images exposed with a grid (+) and without a grid (-), as observed with the PMMA thickness of 9.5cm. As a result, the visibility of details in non-grid (-) images processed by Virtual Grid processing were regarded as having a higher image visibility than images acquired with a grid (+) and without a grid (-) while reducing the absorbed dose. [3] Chest phantom Fig.14 shows the results of the Chest Phantom images acquired at an x-ray tube voltage of 80kVp and exposure dose of 1.8mAs. Non-grid (-) images processed by Virtual Grid processing were superior to images exposed with a grid (+) in the visualisation of the structures of the peripheral side of lung field, mediastinum side of lung field, mediastinum region and sharpness. Additionally, the levels of granularity and contrast in images processed by Virtual Grid processing were equivalent to images exposed with a grid (+). Page 11 of 25
12 [4] Discussion From the evaluative results for CNR and IQFinv, it is shown that at 100% dose non-grid images processed by Virtual Grid processing showed a level of image contrast equivalent to or greater than images exposed with a conventional secondary radiation grid (+). Even at half the dose non-grid images processed with Virtual Grid Processing showed equivalent detail visibility to images exposed with a grid (+) at 100% dose. Images exposed with a grid (+) have improved contrast, but decreased granularity characteristics compared to images exposed without a grid (-). After having said that both images exposed with a grid (+) and without a grid (-) showed equivalent levels of CNR and IQFiv in terms of results. In contrast, non-grid (-) images processed by Virtual Grid processing showed superior CNR and IQFinv to images exposed with a grid (+) and without a grid (-) because the processing allows improvement of contrast and does not decrease granularity. From the results obtained in the study it was concluded that visualisation of lung field and mediastinum regions in non-grid (-) chest images processed by Virtual Grid processing was equivalent or greater than in images exposed with a conventional secondary radiation grid (+). Figs.15 and 16 show a chest image of a patient exposed without a grid (-) using the current examination parameters of 65kVp and exposure dose of 5mAs, and an image of the identical patient processed by Virtual Grid processing at 80kVp and the exposure dose of 2mAs. The use of Virtual Grid processing has allowed for improvement of contrast in the chest image normally associated with the use of a grid (+), but in conjunction with a reduction of incident surface dose of 33% while maintaining high quality diagnostic information. Images for this section: Page 12 of 25
13 Fig. 5 Page 13 of 25
14 Fig. 6 Page 14 of 25
15 Fig. 7 Page 15 of 25
16 Fig. 8 Page 16 of 25
17 Fig. 9 Page 17 of 25
18 Fig. 10 Page 18 of 25
19 Fig. 11 Page 19 of 25
20 Fig. 12 Page 20 of 25
21 Fig. 13 Page 21 of 25
22 Fig. 14 Page 22 of 25
23 Fig. 15 Page 23 of 25
24 Fig. 16 Page 24 of 25
25 Conclusion The use of an FPD and new Virtual Grid Processing software was evaluated for mobile chest images. The study determined that the use of an FPD and Virtual Grid processing allowed non-grid (-) images to be acquired with equivalent or greater diagnostic information than that obtained with the use of a conventional grid (+). In addition, non-grid (-) images acquired at a reduced dose and processed with Virtual Grid processing showed equivalent detail visibility to images exposed with a conventional grid (+) at 100% dose. In Osaka General Medical Center, the incident surface dose to the patients was reduced by 33% using the exposure condition (80kVp and 2 mas) with Virtual Grid processing compared to the conventional low tube voltage exposure condition (65kVp and 5 mas). Personal information References * Japanese Society of Radiological Technologies. "Clinical Radiographic Technology Study Handbook, Vol.1", (1996): Page 25 of 25
Studies on reduction of exposure dose using digital scattered X-ray removal processing
Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.
More informationEvaluation of no-grid radiography using the digital scattered x-ray removal processing
Evaluation of no-grid radiography using the digital scattered x-ray removal processing Poster No.: C-0416 Congress: ECR 2016 Type: Authors: Scientific Exhibit R. Suzuki 1, T. Goto 1, H. Ogawa 2, N. Amimoto
More informationInvestigation of Effective DQE (edqe) parameters for a flat panel detector
Investigation of Effective DQE (edqe) parameters for a flat panel detector Poster No.: C-1892 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit D. Bor 1, S. Cubukcu 1, A. Yalcin 1, O.
More informationDose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis
Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis Poster No.: C-0585 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Minehiro,
More informationA study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers
A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers Poster No.: C-3011 Congress: ECR 2010 Type: Topic: Authors: Scientific Exhibit
More informationImprovement of CT image quality with iterative reconstruction idose4
Improvement of CT image quality with iterative reconstruction idose4 Poster No.: C-0387 Congress: ECR 2014 Type: Scientific Exhibit Authors: M.-L. Olsson, K. Norrgren, M. Söderberg; Malmö/SE Keywords:
More informationThe impact of increasing SID on patient dose in pa abdomen imaging
The impact of increasing SID on patient dose in pa abdomen imaging Poster No.: C-1074 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific Exhibit T. Starc 1, N. Mekis 2 ; 1 Brezovica/SI, 2 Ljubljana/SI
More informationComputer applications, PACS, Instrumentation, Equipment, Technology assessment /ecr2012/C-0655
The novel medical diagnostic color monitor automatically distinguishing monochrome/color images and displaying these with their own optimal tone curves Poster No.: C-0655 Congress: ECR 2012 Type: Authors:
More informationInfluence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck
Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Poster No.: C-2205 Congress: ECR 2012 Type: Scientific Paper
More informationFundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging
Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging Poster No.: C-0709 Congress: ECR 2014 Type: Scientific
More informationVisualization of sources of scattered radiation from x-ray equipment used for interventional radiology
Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology Poster No.: C-1190 Congress: ECR 2011 Type: Scientific Exhibit Authors: K. Chida, T. Takahashi, D.
More informationAim. Images for this section: Page 2 of 13
Changes in CT number of high atomic number materials with field of view when using an extended CT number to electron density curve and a metal artifact reduction reconstruction algorithm Poster No.: R-0094
More informationCorrection of the local intensity nonuniformity artifact in high field MRI
Correction of the local intensity nonuniformity artifact in high field MRI Poster No.: C-0346 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Paper S. Kai, S. Kumazawa, H. Yabuuchi, F. Toyofuku;
More informationClear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE)
Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE) Poster No.: C-2459 Congress: ECR 2010 Type: Scientific Exhibit Topic: Neuro Authors: T. Yoneda,
More informationNovel cassette-sized, flat-panel digital radiography (DR) system: Initial clinical and workflow results versus computed radiography (CR)
Novel cassette-sized, flat-panel digital radiography (DR) system: Initial clinical and workflow results versus computed radiography (CR) Poster No.: C-3027 Congress: ECR 2010 Type: Scientific Exhibit Topic:
More informationA comparison between medical-grade liquid crystal display (LCD) and ipad color imaging
A comparison between medical-grade liquid crystal display (LCD) and ipad color imaging Poster No.: C-1377 Congress: ECR 2014 Type: Scientific Exhibit Authors: K. Yoshimura 1, K. Shibata 2, T. Nihashi 1,
More informationAnalysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging
Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging Award: Magna Cum Laude Poster No.: C-1988 Congress: ECR 2014 Type: Scientific Exhibit Authors: T. Hamaguchi,
More informationLumbar disc height assessment: a comparative study between EOS and digital radiography
Lumbar disc height assessment: a comparative study between EOS and digital radiography Poster No.: C-1178 Congress: ECR 2012 Type: Scientific Paper Authors: V. Freire, M. Benhamou, F. Rannou, S. Poiraudeau,
More informationDetermining acceptance levels for automatic daily image quality control in magnetic resonance imaging
Determining acceptance levels for automatic daily image quality control in magnetic resonance imaging Poster No.: C-1125 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific Exhibit J. I. Peltonen,
More informationMulti-slice computed tomography analysis of bullet trajectory in forensic investigation.
Multi-slice computed tomography analysis of bullet trajectory in forensic investigation. Poster No.: C-0510 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Usui, Y. Kawasumi, Y. Hosokai, A. Nakajima,
More informationOptimal speed for dual-energy subtraction neck imaging with flat-panel detector radiography
Optimal speed for dual-energy subtraction neck imaging with flat-panel detector radiography Poster No.: C-2938 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology Authors: H. Machida
More informationDevelopment of new dosimeter for measuring dose distribution in CT
Development of new dosimeter for measuring dose distribution in CT Poster No.: C-2925 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology - Without Subtopic Authors: Y. Muramatsu, K.
More informationDifference in signal-to-noise-ratio (SNR) from vertical to horizontal scanner position using a 0,25 Tesla Weightbearing
Difference in signal-to-noise-ratio (SNR) from vertical to horizontal scanner position using a 0,25 Tesla Weightbearing scanner Poster No.: C-0672 Congress: ECR 2014 Type: Scientific Exhibit Authors: F.
More informationSuppression of metal artifacts using image-based monoenergetic DECT imaging
Suppression of metal artifacts using image-based monoenergetic DECT imaging Poster No.: C-0519 Congress: ECR 2011 Type: Scientific Paper Authors: B. Krauss, B. Schmidt, M. Sedlmair, T. Flohr; Forchheim/DE
More informationPractical guidelines for color calibration and quality assurance of medical displays
Practical guidelines for color calibration and quality assurance of medical displays Poster No.: C-1140 Congress: ECR 2017 Type: Educational Exhibit Authors: T. Kimpe, J. Rostang, G. Van Hoey, A. Xthona
More informationComparison of high-resolution C-arm cone-beam CT imaging and conventional 3D rotational angiography for the better microvascular visualization
Comparison of high-resolution C-arm cone-beam CT imaging and conventional 3D rotational angiography for the better microvascular visualization Poster No.: C-1040 Congress: ECR 2016 Type: Scientific Exhibit
More informationWorldwide practice of breast MRI: insights from the MIPA study applications - the MIPA study group
Worldwide practice of breast MRI: insights from the MIPA study applications - the MIPA study group Poster No.: B-0757 Congress: ECR 2015 Type: Scientific Paper Authors: G. Di Leo, R. M. Trimboli, I. Ioan,
More informationPROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution
PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution Poster No.: C-1280 Congress: ECR 2011 Type: Educational Exhibit Authors: A. Monteiro, A.
More informationY11-DR Digital Radiography (DR) Image Quality
Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures
More informationPROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution
PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution Poster No.: C-1280 Congress: ECR 2011 Type: Educational Exhibit Authors: A. Monteiro, A.
More informationEnhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)
Vascular Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction In recent years, digital cardiovascular
More informationBeam-Restricting Devices
Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the
More informationThe importance of radiation quality for optimisation in radiology
Available online at http://www.biij.org/2007/2/e38 doi: 10.2349/biij.3.2.e38 biij Biomedical Imaging and Intervention Journal COMMENTARY The importance of radiation quality for optimisation in radiology
More informationThe effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography
Journal of Physics: Conference Series OPEN ACCESS The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography To cite this article: N A A Daud et al 2014 J. Phys.:
More informationThe role of CT in restoration of a very rare Christ wooden statue: a fundamental contribution.
The role of CT in restoration of a very rare Christ wooden statue: a fundamental contribution. Poster No.: C-1514 Congress: ECR 2015 Type: Scientific Exhibit Authors: M. Disaro', P. Sartori; Venice/IT
More informationCOMPUTED TOMOGRAPHY 1
COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the
More informationIBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY
WHITE PAPER: IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY IBEX Innovations Ltd. Registered in England and Wales: 07208355 Address: Discovery 2, NETPark, William Armstrong Way, Sedgefield, UK Patents:
More informationRedefining Ergonomics
Samsung Electronics Co., Ltd. inspires the world and shapes the future with transformative ideas and technologies, redefining the worlds of TVs, smartphones, wearable devices, tablets, cameras, digital
More informationX-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto
X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring
More informationDigital radiography (DR) post processing techniques for pediatric radiology
Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions
More informationTest Equipment for Radiology and CT Quality Control Contents
Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3
More informationTeaching Digital Radiography and Fluoroscopic Radiation Protection
Teaching Digital Radiography and Fluoroscopic Radiation Protection WCEC 20 th Student Educator Radiographer Conference Dennis Bowman, RT(R), CRT (R)(F) Community Hospital of the Monterey Peninsula (CHOMP)
More informationIntroduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image
Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology
More informationDigital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of
Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of radiation to the population due to Medical Imaging
More informationThe feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE
The feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE Poster No.: C-0022 Congress: ECR 2010 Type: Scientific Exhibit Topic: Abdominal Viscera
More informationLECTURE 1 The Radiographic Image
LECTURE 1 The Radiographic Image Prepared by:- KAMARUL AMIN ABDULLAH @ ABU BAKAR UiTM Faculty of Health Sciences Medical Imaging Department 11/23/2011 KAMARUL AMIN (C) 1 Lesson Objectives At the end of
More informationVisibility of Detail
Visibility of Detail Radiographic Quality Quality radiographic images represents the, and information is for diagnosis. The of the anatomic structures and the accuracy of their ( ) determine the overall
More informationRAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II
RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II APPROVED 12/O2/2011 EFFECTIVE SPRING 2013-14 Prefix & Number RAD 150 Course Title: Radiologic Exposure Technique II & Lab Purpose of this submission: New Change/Updated
More informationCOMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR
This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura New Technology https://www.youtube.com/watch?v=ptkzznazb 7U COMPUTED
More informationExposure System Selection
Principles of Imaging Science II (RAD120) Exposure Systems Exposure System Selection Radiographic exposure is a very complex process Best technique systems manipulate one variable while holding others
More informationCR Basics and FAQ. Overview. Historical Perspective
Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors
More informationyou can Portable DR with High Sensitivity. Less is more. CXDI-501C DIGITAL RADIOGRAPHY SYSTEM
you can Portable DR with High Sensitivity. Less is more. CXDI-501C DIGITAL RADIOGRAPHY SYSTEM High quality DR technology from a dose-efficient, thin and lightweight portable Flat Panel Detector Time Reduction
More informationFCR XL-2 and FCR XC-2. High-quality digital x-ray that's perfect for your private practice.
F U J I F I L M D I G I T A L X - R A Y FCR XL-2 and FCR XC-2. High-quality digital x-ray that's perfect for your private practice. So small, yet so powerful. With a compact footprint of less than 2.5
More informationWhile digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.
In press 2004 1 2 Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Digital radiology An appropriate analogy that is easy for most people to understand
More informationComparison of computed radiography and filmõscreen combination using a contrast-detail phantom
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 1, WINTER 2003 Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom Z. F. Lu,* E. L. Nickoloff, J.
More informationComparison of new digital mammography systems: physical characterisation and image quality evaluation
Comparison of new digital mammography systems: physical characterisation and image quality evaluation Poster No.: C-2321 Congress: ECR 2011 Type: Scientific Exhibit Authors: N. Oberhofer; Bolzano/IT DOI:
More informationAcquisition, Processing and Display
Acquisition, Processing and Display Terri L. Fauber, R.T. (R)(M) Department of Radiation Sciences School of Allied Health Professions Virginia Commonwealth University Topics Image Characteristics Image
More informationDose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp
Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division
More informationHalf value layer and AEC receptor dose compliance survey in Estonia
Half value layer and AEC receptor dose compliance survey in Estonia K. Kepler, A. Vladimirov Training Centre of Medical Physics, University of Tartu Testing Centre of the University of Tartu, Estonia E-mail:
More informationDigital radiography. bucky table and wall stand as a dual detector or wireless system. Amadeo DR Systems
Amadeo DR Systems with dicom PACS DX-R Software X-ray Systems for the Future Digital radiography with Amadeo R-DR including bucky table and wall stand as a dual detector or wireless system High contrast
More informationExposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?
Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago
More informationFeatures and Weaknesses of Phantoms for CR/DR System Testing
Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal
More informationRADIOGRAPHIC EXPOSURE
RADIOGRAPHIC EXPOSURE Receptor Exposure Receptor Exposure the that interacts with the receptor. Computed Radiography ( ) requires a. Direct Digital Radiography (DR) requires a. Exposure Indicators Exposure
More informationImproved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques
Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques Wataru FUKUDA* Junya MORITA* and Masahiko YAMADA* Abstract Tomosynthesis is a three-dimensional imaging technology
More informationIBEX MATERIALS DETECTION TECHNOLOGY
WHITE PAPER: IBEX MATERIALS DETECTION TECHNOLOGY IBEX Innovations Ltd. Registered in England and Wales: 07208355 Address: Discovery 2, NETPark, William Armstrong Way, Sedgefield, TS21 3FH, UK Patents held
More information3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development
Teaching Radiographic Technique in a Digital Imaging Paradigm Objectives 1. Discuss the historical development of digital imaging. Dawn Couch Moore, M.M.Sc., RT(R) Assistant Professor and Director Emory
More informationMulti-Access Biplane Lab
Multi-Access Biplane Lab Advanced technolo gies deliver optimized biplane imaging Designed in concert with leading physicians, the Infinix VF-i/BP provides advanced, versatile patient access to meet the
More informationradiography detector
Clinical evaluation of a full field digital projection radiography detector Gary S. Shaber'1, Denny L. Leeb, Jeffrey Belib, Gregory Poweii1', Andrew D.A. Maidment'1 a Thomas Jefferson University Hospital,
More informationX-RAYS - NO UNAUTHORISED ENTRY
Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are
More informationObjective Evaluation of Radiographic Contrast- Enhancement Masks
Chapter 8 Objective Evaluation of Radiographic Contrast- Enhancement Masks The development and application of radiographic contrast-enhancement masks (RCMs) in digital radiography (DR) were discussed in
More informationMedical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging
Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition
More informationYour worldwide source of grids. What a difference a grid makes
Your worldwide source of grids What a difference a grid makes The right assembly to meet your every need Trust the experts A high degree of expertise is required to design and manufacture quality X-ray
More informationMultiple Choice Identify the letter of the choice that best completes the statement or answers the question.
RA110 test 3 Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. 1. An object 35 cm in width is radiographed at 100 cm SID and at a 50 cm SOD. What
More informationQC Testing for Computed Tomography (CT) Scanner
QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including
More informationX-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah
X-RAY IMAGING EE 472 F2017 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Stewart C. Bushong, Radiologic Science for Technologists: Physics, Biology, and Protection, 10 th ed., Mosby,
More informationVeterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS
Veterinary Science Preparatory Training for the Veterinary Assistant Floron C. Faries, Jr., DVM, MS Radiology Floron C. Faries, Jr., DVM, MS Objectives Determine the appropriate machine settings for making
More informationEvaluation and Comparison of Image Quality for Indirect Flat Panel Systems with CsI and GOS Scintillators
Evaluation and Comparison of Image Quality for Indirect Flat Panel Systems with CsI and GOS Scintillators Mehmet E. Aksoy a, Mustafa E. Kamasak b, Erkan Akkur c, Ayhan Ucgul d, Muzaffer Basak d, Huseyin
More informationVoice recognition versus transcriptionist
Voice recognition versus transcriptionist Poster No.: C-0263 Congress: ECR 2010 Type: Audit/Professional Issues Topic: Audit/Professional Issues Authors: R. H. Strahan; Clayton South VIC/AU Keywords: Voice
More informationSmartRAD. Advanced Digital Radiography System
SmartRAD Advanced Digital Radiography System SmartRAD Expanding The Horizons Of Digital Radiography CMT introduces the SmartRAD Digital Radiography system, featuring an integrated flat panel digital detector
More informationSTEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS
STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%
More informationRADspeed Pro. EDGEpackage C501-E041C
RADspeed Pro EDGEpackage C501-E041C 2 Some of the FPDs may be not available in your country. Please contact us to check the availability in your country. 3 Tomosynthesis in the Standing Position Tomosynthesis
More informationC501-E029C. RADspeed. GENERAL RADIOGRAPHIC SYSTEM Automatic
C501-E029C RADspeed GENERAL RADIOGRAPHIC SYSTEM Automatic High-Performance General Radiographic System Improves Workflow and Achieves Dose Reduction GENERAL RADIOGRAPHIC SYSTEM Automatic This state-of-the-art
More informationImage Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose
Image Quality and Dose Issues in MSCT Image Quality and Dose Image quality Image noise Spatial resolution Contrast Artefacts Speckle and sharpness S. Edyvean St. George s Hospital London SW17 0QT Radiation
More informationLudlum Medical Physics
Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to
More informationCXDI-70C WIRELESS SPECIFICATIONS
CXDI-70C WIRELESS SPECIFICATIONS Purpose Method Sensor Scintillator Pixel Pitch Pixels Image Size A/D Grayscale Wireless Standard Preview Image Access Time High Resolution Image Display Time Cycle Time
More informationDo you have any other questions? Please call us at (Toll Free) or , or
INSTRUCTIONS Read the appropriate course/ textbook. This is an open book test. A score of 75% or higher is needed to receive CE credit. You will have a maximum of three attempts to pass this course. Please
More informationDR General Radiography System. DR Series
DR General Radiography System ACH049_Shimadzu A4 Medical Bro 1 DR Series 6/5/08 2:41:22 PM High Productivity and Reliability: Hallmarks of Shimadzu X-ray Systems! With a near century of experience in manufacturing
More informationAutomated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London
Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients
More informationInvestigation of the line-pair pattern method for evaluating mammographic focal spot performance
Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann
More informationAppropriate Inspection Distance of Digital X-Ray Imaging Equipment for Diagnosis
Indian Journal of Science and Technology Vol 8(S8), 380-386, April 2015 ISSN (Print) : 0974-6846 ISSN (Online) : 0974-5645 DOI: 10.17485/ijst/2015/v8iS8/70528 Appropriate Inspection Distance of Digital
More informationFilm Replacement in Radiographic Weld Inspection The New ISO Standard
BAM Berlin Film Replacement in Radiographic Weld Inspection The New ISO Standard 17636-2 Uwe Ewert, Uwe Zscherpel, Mirko Jechow Requests and information to: uwez@bam.de 1 Outline - The 3 essential parameters
More informationSONIALVISION G4 Multi-purpose Digital R/F System C506-E075
SONIALVISION G4 Multi-purpose Digital R/F System C506-E075 Selecting the best for every examination environment BEST in CLASS Multi-purpose Digital R/F System 2 With the Sonialvision G4, Shimadzu offers
More informationSONIALVISION G4 Multi-purpose Digital R/F System C506-E075
SONIALVISION G4 Multi-purpose Digital R/F System C506-E075 Selecting the best for every examination environment BEST in CLASS Multi-purpose Digital R/F System With the Sonialvision G4, Shimadzu offers
More informationOpen. the Digitized world. Fuji Computed Radiography
Open the Digitized world Fuji Computed Radiography If just one of these applies to you... Managing developing fluid is hard and darkroom work is a hassle... Images are not stable... Isn t digitalization
More information4. Contrast is the. There must The function of contrast is to:. The types of contrast are.
RADIOGRAPHIC VISIBILITY OF DETAIL STUDY QUESTIONS 1. What is visibility of detail? It is controlled by properties. What are the two factors that affect it? 2. What is sharpness of detail? It is controlled
More informationJoint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009
2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced
More informationSONIALVISION G4 Multi-purpose Digital R/F System
C506-E075A SONIALVISION G4 Multi-purpose Digital R/F System Founded in 1875, Shimadzu Corporation, a leader in the development of advanced technologies, has a distinguished history of innovation built
More informationDigital Quality Control Phantom for X-ray Mammography Image
Digital Quality Control Phantom for X-ray Mammography Image Yuichi Nagai, Atsuko Takada National: Cancer Hospital East Department of Radiology Collaborated by Nobuyuki Niwa : Kyoto Kagaku Co., Ltd. Introduction
More informationAmorphous Selenium Direct Radiography for Industrial Imaging
DGZfP Proceedings BB 67-CD Paper 22 Computerized Tomography for Industrial Applications and Image Processing in Radiology March 15-17, 1999, Berlin, Germany Amorphous Selenium Direct Radiography for Industrial
More informationMammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over
Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on
More information