Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE)
|
|
- Warren Lewis
- 6 years ago
- Views:
Transcription
1 Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE) Poster No.: C-2459 Congress: ECR 2010 Type: Scientific Exhibit Topic: Neuro Authors: T. Yoneda, H. Arimura, T. Takemaru, Y. Hiai; Kumamoto/JP Keywords: phase, susceptibility, MRI DOI: /ecr2010/C-2459 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. Page 1 of 14
2 Purpose It is important to delineate various nerve tracts in the brain image. Diffusion Weighted Imaging (DWI) and tensor imaging as a application of it are one of the excellent method to delineate them. However, spatial resolution of DWI is relatively low because of technical difficulties to detect them. Therefore it is difficult to detect thin nerve tract on the DWIimage and expected to develop new imaging theory and method. Recently, many study reports that phase information of MRI potentially possibility delineate small structure in the brain such as very small vessels and nerve tracts. Especially, susceptibility weighted imaging (SWI) can beautifully delineate the vein (vessel) in the brain by using phase information [ref. 1, 2]. This is the one of the most excellent result of phase imaging in MRI. In this study we develop new technique "Phase Difference Enhanced Imaging (PADRE)" in order to enhance various tissues in the brain such as OR and vessels in the brain MR-image. This technique selectively enhances the phase (difference) of objective tissue locating even in the white matter in which contrast of different tissues is hardly reconstructed. PADRE is to clearly delineate OR by adjusting several parameters. In this study contrast of tissues (OR and vessels) was measured and fixed such parameters. Additionally, we show an ability of PADRE to illuminate very small vessels. Methods and Materials In this study, we have taken images with 2 sequences and scanning parameters. In order to have much information of phase (differences), we need long echo time (TE) (Fig. 1 on page 3) [ref. 3]. Long TE, however, results in distortion of magnitude image contrast due to the T2* effect. Therefore, whenever we set long TE, we have to take much longer repetition time (TR) for having sufficient signal intensity. But such longer TR automatically means longer scanning time which should be evitable for clinical use. We frequently meet such kind dilemma when we want to use phase information. One of the solution to longer scanning time is using shifted-echo sequence in which TE is longer than TR. In this study we used 3D Principle of Echo Shifting with a Train Observations (PRESTO) as shifted-echo sequence. Since PRESTO is running on Philips MRI system, we have chosen Philips's 3.0 T MRI (Achieve 3.0T, Philips Medical Systems, The Netherlands). Page 2 of 14
3 3D-GE (FFE) acquisition was also used to see a dependence of the contrast of PADREimage. We arranged scanning parameters so as to fix scanning time (10min); TR/TE = 32/51 ms, SENSE factor = 1, FOV = 230x230 mm, Matrix = 512x512, Slice Thickness = 1 mm for PRESTO and TR/TE = 38/30 ms, SENSE factor = 1, FOV = 230x230 mm, Matrix = 512x512, Slice Thickness = 1mm for 3D-GE. Each phase image was enhanced by PADRE and measured the contrast between OR and parenchyma, and also the deep veins as small vessels. Enhancement steps were as below (Fig. 2 on page 4); 1. Low-pass filter with fixed filter size was operated on k-space data and Fourier transform was carried out to create low-pass filtered complex image (data). 2. High-pass filtered complex image was obtained from dividing original image by low-pass filtered image. (We henceforth call filtered image as high-pass filtered complex image.) 3. Filtered phase image (data) was obtained from filtered complex image. We just call this "phase" as shorthand notation. 4. Phase was selected to suit objective tissue [ref. 4]. 5. Selected phase was enhanced by depressing the signal value on the original magnitude. Enhancement achieves by using enhancing function which was designed for adjustable selection phase and enhancement of them. Region of interest (ROI) was located on objective tissue (OR) and parenchyma with fixed area. Obtained data is used to determine the contrast of OR. Here, contrast is to be defined as difference between lowest value of signal in a profile curve of OR and mean value of parenchyma in the ROI. Images for this section: Page 3 of 14
4 Fig. 1: Phase value proportional TE and local magnetic field (magnetic flux density) with susceptibility #. Contrast in phase image should be defined local magnetic field difference between different tissue. Page 4 of 14
5 Fig. 2: Phase (difference) value normally takes very small number, about several ppm. Therefore we must enhance it when we put phase information in the magnitude image. Page 5 of 14
6 Results Original PRESTO and 3D-GE images could slightly show OR tracks due to the T2* effect (Fig. 1 on page 6). From the theoretical point of view, it is expected that PRESTO which has longer TE than 3D-GE sequence could show higher contrast on "phase image". But, in practice, background plays important role to decide enhanced contrast because of its definition, and therefore enhanced contrast do not always increase proportional to TE. Actually, OR is to be seen well in the 3D-GE image compared with PRESTO image (Fig. 1 on page 6). It is also true in the enhanced images. We can see clear contrast of OR on PADRE image reconstructed from 3D-GE image (Fig. 2 on page 7). All acquired images were filtered with 128x128 filter size. Although PADRE-image reconstructed from PRESTO can also delineate OR, its apparent contrast is slightly lower than 3D-GE's one (Fig. 3 on page 8). In order to numerically measure the contrast of OR and parenchyma, we set profile line and ROI to measure the signal of parenchyma (see Fig. 4 on page 9). Obtained data was calculated and plotted on Fig. xx. Imaging parameters were changed so as to enhance for (1) all phase of OR, (2) large phase value of OR, (3) small phase value. PADRE-image very well delineate OR (red arrows in Fig. 2 on page 7, 3 on page 8) compared with magnitude image of PRESTO which is T2* image. Actually, all PADRE-images gave higher contrast value than PRESTO-image (Fig. 5 on page 10). As is in Fig. 5 on page 10, contrast of PADRE-image type 2 and 3 especially show high contrast. But, in those images, background noise looks be enlarged by enhancement and it becomes to be visually difficult to distinguish between noise and signal of tissue in them. By enhancing the phase of vessel, we could have vessel enhanced PADRE-image which is corresponding to SWI-image (Fig. 6 on page 11). In that image, we can see very high contrasted vessels, especially, fine deep veins is very well delineate. Same as OR enhanced images, images type 2 and 3 looked to be noisy one because PADRE simultaneously enhanced phase noise with objective phase. In fact, signal to noise ratio (SNR) of those images shows lower value (Fig. 7 on page 12). Images for this section: Page 6 of 14
7 Fig. 1: Both of these magnitude images show low signal intensity of OR due to T2* effect. Page 7 of 14
8 Fig. 2: All of images show high contrast of OR which is pointed by red arrow. There images were reconstructed by different parameter sets for enhancement. Page 8 of 14
9 Fig. 3: Enhanced Images generated from PRESTO images. These images also show high contrast OR. Additionally, cerebral cortex can be illuminated by PADRE technique. Page 9 of 14
10 Fig. 4: Contrast was defined by lowest signal of transverse profile curve of OR (red line). We obtained signal of parenchyma defined by mean value in the ROI (green circle). Page 10 of 14
11 Fig. 5: All of reconstructed images show high contrast of OR compared with magnitude image. Page 11 of 14
12 Fig. 6: Even small vessels were delineated by PADRE technique. Deep veins (enlarged in the box) can be seen very well. This vessel imnages were generated by minimam intensity projection over 12 slices. Page 12 of 14
13 Fig. 7: Comparison of SNR in the vessel enhanced images. Parameters to generate strong contrast simultaneously enhances signal and phase noise. Therefore, SNR decreases in the case of strong enhancement (image #3). Page 13 of 14
14 Conclusion We had clearly delineated OR and small vessels by using PADRE technique with various imaging parameters. Although all parameter sets lead higher contrast of OR compared with magnitude image, phase noise was also enhanced on the PADRE image so that it is not always good to observe tissues. This situation was also true in the case of vessel enhanced images. Actually, SNR of them varied depending on parameters set. But vessel enhanced PADRE-images shows very clear contrast of very fine vessel. References 1. Haacke E.M., Xu Y., Cheng Y.N., Reichenbach J.R., SusceptibilityWeighted Imaging. Magn. Reson. Med. 52, Reichenbach J.R., Barth M., Haacke E.M., Klarh ofer M., Kaiser W.A., Moser E., High-resolution MR venography at 3.0 Tesla. J. Comput. Assist. Tomogr. 24 (6), E. Mark Haacke, Robert W. Brown, Michael R. Thompson, and Ramesh Venkatesan, Magnetic Resonance Imaging: Physical Principles and Sequence Design. 1999, Wiley-Liss, USA. 4. Zhong K., Leupold J., von Elverfeldt D., Speck O., The molecular basis for gray and white matter contrast in phase imaging. NeuroImage 40, Personal Information Tetsuya YONEDA, PhD, School of Health Sciences, Kumamoto University, Kuhonji , Kumamoto Japan. Page 14 of 14
Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging
Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging Poster No.: C-0709 Congress: ECR 2014 Type: Scientific
More informationComputer applications, PACS, Instrumentation, Equipment, Technology assessment /ecr2012/C-0655
The novel medical diagnostic color monitor automatically distinguishing monochrome/color images and displaying these with their own optimal tone curves Poster No.: C-0655 Congress: ECR 2012 Type: Authors:
More informationCorrection of the local intensity nonuniformity artifact in high field MRI
Correction of the local intensity nonuniformity artifact in high field MRI Poster No.: C-0346 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Paper S. Kai, S. Kumazawa, H. Yabuuchi, F. Toyofuku;
More informationEvaluation of no-grid radiography using the digital scattered x-ray removal processing
Evaluation of no-grid radiography using the digital scattered x-ray removal processing Poster No.: C-0416 Congress: ECR 2016 Type: Authors: Scientific Exhibit R. Suzuki 1, T. Goto 1, H. Ogawa 2, N. Amimoto
More informationAnalysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging
Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging Award: Magna Cum Laude Poster No.: C-1988 Congress: ECR 2014 Type: Scientific Exhibit Authors: T. Hamaguchi,
More informationImprovement of CT image quality with iterative reconstruction idose4
Improvement of CT image quality with iterative reconstruction idose4 Poster No.: C-0387 Congress: ECR 2014 Type: Scientific Exhibit Authors: M.-L. Olsson, K. Norrgren, M. Söderberg; Malmö/SE Keywords:
More informationInvestigation of Effective DQE (edqe) parameters for a flat panel detector
Investigation of Effective DQE (edqe) parameters for a flat panel detector Poster No.: C-1892 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit D. Bor 1, S. Cubukcu 1, A. Yalcin 1, O.
More informationDifference in signal-to-noise-ratio (SNR) from vertical to horizontal scanner position using a 0,25 Tesla Weightbearing
Difference in signal-to-noise-ratio (SNR) from vertical to horizontal scanner position using a 0,25 Tesla Weightbearing scanner Poster No.: C-0672 Congress: ECR 2014 Type: Scientific Exhibit Authors: F.
More informationInfluence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck
Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Poster No.: C-2205 Congress: ECR 2012 Type: Scientific Paper
More informationAim. Images for this section: Page 2 of 13
Changes in CT number of high atomic number materials with field of view when using an extended CT number to electron density curve and a metal artifact reduction reconstruction algorithm Poster No.: R-0094
More informationStudies on reduction of exposure dose using digital scattered X-ray removal processing
Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.
More informationStudies on reduction of exposure dose using digital scattered X-ray removal processing
Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.
More informationWorldwide practice of breast MRI: insights from the MIPA study applications - the MIPA study group
Worldwide practice of breast MRI: insights from the MIPA study applications - the MIPA study group Poster No.: B-0757 Congress: ECR 2015 Type: Scientific Paper Authors: G. Di Leo, R. M. Trimboli, I. Ioan,
More informationA study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers
A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers Poster No.: C-3011 Congress: ECR 2010 Type: Topic: Authors: Scientific Exhibit
More informationA comparison between medical-grade liquid crystal display (LCD) and ipad color imaging
A comparison between medical-grade liquid crystal display (LCD) and ipad color imaging Poster No.: C-1377 Congress: ECR 2014 Type: Scientific Exhibit Authors: K. Yoshimura 1, K. Shibata 2, T. Nihashi 1,
More informationMulti-slice computed tomography analysis of bullet trajectory in forensic investigation.
Multi-slice computed tomography analysis of bullet trajectory in forensic investigation. Poster No.: C-0510 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Usui, Y. Kawasumi, Y. Hosokai, A. Nakajima,
More informationDetermining acceptance levels for automatic daily image quality control in magnetic resonance imaging
Determining acceptance levels for automatic daily image quality control in magnetic resonance imaging Poster No.: C-1125 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific Exhibit J. I. Peltonen,
More informationThe impact of increasing SID on patient dose in pa abdomen imaging
The impact of increasing SID on patient dose in pa abdomen imaging Poster No.: C-1074 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific Exhibit T. Starc 1, N. Mekis 2 ; 1 Brezovica/SI, 2 Ljubljana/SI
More informationVisualization of sources of scattered radiation from x-ray equipment used for interventional radiology
Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology Poster No.: C-1190 Congress: ECR 2011 Type: Scientific Exhibit Authors: K. Chida, T. Takahashi, D.
More informationLumbar disc height assessment: a comparative study between EOS and digital radiography
Lumbar disc height assessment: a comparative study between EOS and digital radiography Poster No.: C-1178 Congress: ECR 2012 Type: Scientific Paper Authors: V. Freire, M. Benhamou, F. Rannou, S. Poiraudeau,
More informationDose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis
Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis Poster No.: C-0585 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Minehiro,
More informationThe feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE
The feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE Poster No.: C-0022 Congress: ECR 2010 Type: Scientific Exhibit Topic: Abdominal Viscera
More informationPractical guidelines for color calibration and quality assurance of medical displays
Practical guidelines for color calibration and quality assurance of medical displays Poster No.: C-1140 Congress: ECR 2017 Type: Educational Exhibit Authors: T. Kimpe, J. Rostang, G. Van Hoey, A. Xthona
More informationComparison of high-resolution C-arm cone-beam CT imaging and conventional 3D rotational angiography for the better microvascular visualization
Comparison of high-resolution C-arm cone-beam CT imaging and conventional 3D rotational angiography for the better microvascular visualization Poster No.: C-1040 Congress: ECR 2016 Type: Scientific Exhibit
More informationImage quality evaluation of turbo-spin echo diffusion weighted image (TSE-DWI) : A phantom study
Image quality evaluation of turbo-spin echo diffusion weighted image (TSE-DWI) : A phantom study Poster No.: C-0631 Congress: ECR 2016 Type: Scientific Exhibit Authors: T. Yoshida, A. Urikura, K. Shirata,
More informationMRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves
MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves Experiment 1 Goal: Examine the effect caused by changing flip angle on image contrast in a simple gradient echo sequence and derive T1-curves. Image
More informationThe role of CT in restoration of a very rare Christ wooden statue: a fundamental contribution.
The role of CT in restoration of a very rare Christ wooden statue: a fundamental contribution. Poster No.: C-1514 Congress: ECR 2015 Type: Scientific Exhibit Authors: M. Disaro', P. Sartori; Venice/IT
More informationSuppression of metal artifacts using image-based monoenergetic DECT imaging
Suppression of metal artifacts using image-based monoenergetic DECT imaging Poster No.: C-0519 Congress: ECR 2011 Type: Scientific Paper Authors: B. Krauss, B. Schmidt, M. Sedlmair, T. Flohr; Forchheim/DE
More informationDevelopment of new dosimeter for measuring dose distribution in CT
Development of new dosimeter for measuring dose distribution in CT Poster No.: C-2925 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology - Without Subtopic Authors: Y. Muramatsu, K.
More informationNovel cassette-sized, flat-panel digital radiography (DR) system: Initial clinical and workflow results versus computed radiography (CR)
Novel cassette-sized, flat-panel digital radiography (DR) system: Initial clinical and workflow results versus computed radiography (CR) Poster No.: C-3027 Congress: ECR 2010 Type: Scientific Exhibit Topic:
More informationPROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution
PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution Poster No.: C-1280 Congress: ECR 2011 Type: Educational Exhibit Authors: A. Monteiro, A.
More informationPROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution
PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution Poster No.: C-1280 Congress: ECR 2011 Type: Educational Exhibit Authors: A. Monteiro, A.
More informationEvaluation of magnetic resonance acoustic noise in 1.5 and 3 Tesla scanners
Evaluation of magnetic resonance acoustic noise in 1.5 and 3 Tesla scanners Poster No.: B-1014 Congress: ECR 2016 Type: Scientific Paper Authors: V. M. F. Silva, I. M. Ramos, J. Moreira, M. Marques; Porto/PT
More informationH 2 O and fat imaging
H 2 O and fat imaging Xu Feng Outline Introduction benefit from the separation of water and fat imaging Chemical Shift definition of chemical shift origin of chemical shift equations of chemical shift
More informationThe SENSE Ghost: Field-of-View Restrictions for SENSE Imaging
JOURNAL OF MAGNETIC RESONANCE IMAGING 20:1046 1051 (2004) Technical Note The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging James W. Goldfarb, PhD* Purpose: To describe a known (but undocumented)
More informationCardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK
Cardiac MR Dr John Ridgway Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Physics for clinicians: Part I Journal of Cardiovascular Magnetic Resonance 2010, 12:71 http://jcmr-online.com/content/12/1/71
More informationDownloaded from by on 02/07/18 from IP address Copyright ARRS. For personal use only; all rights reserved
Downloaded from www.ajronline.org by 46.3.192.5 on 02/07/18 from IP address 46.3.192.5. Copyright RRS. For personal use only; all rights reserved C oil sensitivity encoding (SENSE) is a new technique that
More informationVoice recognition versus transcriptionist
Voice recognition versus transcriptionist Poster No.: C-0263 Congress: ECR 2010 Type: Audit/Professional Issues Topic: Audit/Professional Issues Authors: R. H. Strahan; Clayton South VIC/AU Keywords: Voice
More informationImage Quality/Artifacts Frequency (MHz)
The Larmor Relation 84 Image Quality/Artifacts (MHz) 42 ω = γ X B = 2πf 84 0.0 1.0 2.0 Magnetic Field (Tesla) 1 A 1D Image Magnetic Field Gradients Magnet Field Strength Field Strength / Gradient Coil
More informationISSN X CODEN (USA): PCHHAX. The role of dual spin echo in increasing resolution in diffusion weighted imaging of brain
Available online at www.derpharmachemica.com ISSN 0975-413X CODEN (USA): PCHHAX Der Pharma Chemica, 2016, 8(17):15-20 (http://derpharmachemica.com/archive.html) The role of in increasing resolution in
More information2014 M.S. Cohen all rights reserved
2014 M.S. Cohen all rights reserved mscohen@g.ucla.edu IMAGE QUALITY / ARTIFACTS SYRINGOMYELIA Source http://gait.aidi.udel.edu/res695/homepage/pd_ortho/educate/clincase/syrsco.htm Surgery is usually recommended
More informationSimultaneous Multi-Slice (Slice Accelerated) Diffusion EPI
Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Val M. Runge, MD Institute for Diagnostic and Interventional Radiology Clinics for Neuroradiology and Nuclear Medicine University Hospital Zurich
More informationSupplementary Figure 1
Supplementary Figure 1 Left aspl Right aspl Detailed description of the fmri activation during allocentric action observation in the aspl. Averaged activation (N=13) during observation of the allocentric
More informationOptimal speed for dual-energy subtraction neck imaging with flat-panel detector radiography
Optimal speed for dual-energy subtraction neck imaging with flat-panel detector radiography Poster No.: C-2938 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology Authors: H. Machida
More informationPulse Sequence Design and Image Procedures
Pulse Sequence Design and Image Procedures 1 Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant 2 A pulse sequence is a timing diagram designed with a series of RF pulses, gradients switching, and signal readout
More informationFunctional MRI with variable echo time acquisition
NeuroImage 20 (2003) 2062 2070 www.elsevier.com/locate/ynimg Functional MRI with variable echo time acquisition Nan-kuei Chen, Svetlana Egorova, Charles R.G. Guttmann, and Lawrence P. Panych* Center for
More information3T Unlimited. ipat on MAGNETOM Allegra The Importance of ipat at 3T. medical
3T Unlimited ipat on MAGNETOM Allegra The Importance of ipat at 3T s medical ipat on MAGNETOM Allegra The Importance of ipat at 3T The rise of 3T MR imaging Ultra High Field MR (3T) has flourished during
More informationBackground (~EE369B)
Background (~EE369B) Magnetic Resonance Imaging D. Nishimura Overview of NMR Hardware Image formation and k-space Excitation k-space Signals and contrast Signal-to-Noise Ratio (SNR) Pulse Sequences 13
More informationHETERONUCLEAR IMAGING. Topics to be Discussed:
HETERONUCLEAR IMAGING BioE-594 Advanced MRI By:- Rajitha Mullapudi 04/06/2006 Topics to be Discussed: What is heteronuclear imaging. Comparing the hardware of MRI and heteronuclear imaging. Clinical applications
More informationEncoding of inductively measured k-space trajectories in MR raw data
Downloaded from orbit.dtu.dk on: Apr 10, 2018 Encoding of inductively measured k-space trajectories in MR raw data Pedersen, Jan Ole; Hanson, Christian G.; Xue, Rong; Hanson, Lars G. Publication date:
More informationPassive Tracking Exploiting Local Signal Conservation: The White Marker Phenomenon
Passive Tracking Exploiting Local Signal Conservation: The White Marker Phenomenon Jan-Henry Seppenwoolde,* Max A. Viergever, and Chris J.G. Bakker Magnetic Resonance in Medicine 50:784 790 (2003) This
More informationIR/SR TrueFISP. Works-in-Progress package Version 1.2. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B.
Works-in-Progress package Version 1.2 For the Installation and User s Guide NUMARIS/4VA21B January 22, 2003 Section of Medical Physics, University Hospital Freiburg, Germany Contact: Klaus Scheffler PhD
More informationMR Advance Techniques. Flow Phenomena. Class II
MR Advance Techniques Flow Phenomena Class II Flow Phenomena In this class we will explore different phenomenona produced from nuclei that move during the acquisition of data. Flowing nuclei exhibit different
More informationThe development of the RF-pulse for the low level SAR used by the MRI.
The development of the RF-pulse for the low level SAR used by the MRI. Kojiro Yamaguchi a*, Eizo Umezawa a, Sachiko Ueoku b, Kazuhiro Katada c a Faculty of radiological technology, School of Health Science,
More information1 Introduction. 2 The basic principles of NMR
1 Introduction Since 1977 when the first clinical MRI scanner was patented nuclear magnetic resonance imaging is increasingly being used for medical diagnosis and in scientific research and application
More informationTimTX TrueShape. The parallel transmit architecture of the future. Answers for life.
www.siemens.com/trueshape TimTX TrueShape The parallel transmit architecture of the future. The product/feature (mentioned herein) is not commercially available. Due to regulatory reasons its future availability
More informationSIGNA Pioneer: Ultra High Efficiency Gradient System Advancing the gradient technology curve
GE Healthcare SIGNA Pioneer: Ultra High Efficiency Gradient System Advancing the gradient technology curve NEW TECHNOLOGY 40W Watts spec is irrelevant. 4W LED bulb delivers same brightness as 40W incandescent
More informationACRIN 6686 / RTOG 0825
ACRIN 6686 (RTOG 0825) Advanced MRI Imaging Manual ACRIN 6686 / RTOG 0825 A phase III double blind placebo controlled trial of conventional chemoradiation and adjuvant temozolomide plus bevacizumab vs
More informationPulse Sequence Design Made Easier
Pulse Sequence Design Made Easier Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant gurumri@gmail.com 1 2 Pulse Sequences generally have the following characteristics: An RF line characterizing RF Pulse applications
More informationSIEMENS MAGNETOM Skyra syngo MR D13
Page 1 of 12 SIEMENS MAGNETOM Skyra syngo MR D13 \\USER\CIND\StudyProtocols\PTSA\*ep2d_M0Map_p2_TE15 TA:7.9 s PAT:2 Voxel size:2.5 2.5 3.0 mm Rel. SNR:1.00 :epfid Properties Routine Contrast Prio Recon
More informationAlae Tracker: Tracking of the Nasal Walls in MR-Imaging
Alae Tracker: Tracking of the Nasal Walls in MR-Imaging Katharina Breininger 1, Andreas K. Maier 1, Christoph Forman 1, Wilhelm Flatz 2, Catalina Meßmer 3, Maria Schuster 3 1 Pattern Recognition Lab, Friedrich-Alexander-Universität
More informationM R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia
M R I Physics Course Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Magnetic Resonance Imaging Spatial
More information2015 Spin echoes and projection imaging
1. Spin Echoes 1.1 Find f0, transmit amplitudes, and shim settings In order to acquire spin echoes, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week,
More informationfunctional MRI: A primer
Activation Leads to: functional MRI: A primer CBF Increased +ΔR CBV Increased +ΔR (C+) O Utilization Increased slightly? Venous [O ] Increased -ΔR* Glucose Utilization Increased? Lactate BOLD R=/T R=/T
More informationHigh Field MRI: Technology, Applications, Safety, and Limitations
High Field MRI: Technology, Applications, Safety, and Limitations R. Jason Stafford, Ph.D. The University of Texas M. D. Anderson Cancer Center, Houston, TX Introduction The amount of available signal
More informationInherent Insensitivity to RF Inhomogeneity in FLASH Imaging
Inherent Insensitivity to RF Inhomogeneity in FLASH Imaging Danli Wang, Keith Heberlein, Stephen LaConte, and Xiaoping Hu* Magnetic Resonance in Medicine 52:927 931 (2004) Radiofrequency (RF) field inhomogeneity
More informationPulse Sequences: Rapid Gradient Echo
Pulse Sequences: Rapid Gradient Echo M229 Advanced Topics in MRI Holden H. Wu, Ph.D. 2018.04.17 Department of Radiological Sciences David Geffen School of Medicine at UCLA Class Business Office hours -
More informationRing-Down versus Comet Tail: Two artifacts uncovered
Ring-Down versus Comet Tail: Two artifacts uncovered Poster No.: C-2288 Congress: ECR 2013 Type: Educational Exhibit Authors: S. Méndez Garrido 1, J. Arce Duran 2, A. Talegón Meléndez 1 ; 1 2 Sevilla/ES,
More informationBOLD fmri: signal source, data acquisition, and interpretation
BOLD fmri: signal source, data acquisition, and interpretation Cheryl Olman 4 th year student, Department of Neuroscience and Center for Magnetic Resonance Research Discussion series Week 1: Biological
More informationNEMA Standards Publication MS (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging
NEMA Standards Publication MS 1-2008 (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging Published by: National Electrical Manufacturers Association 1300 North
More informationMARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005
ACR MRI accreditation program MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics,
More informationModule 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review
MRES 7005 - Fast Imaging Techniques Module 2 Artefacts and Imaging Optimisation for single shot methods Content: Introduction Phase error Phase bandwidth Chemical shift review Chemical shift in pixels
More informationSECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS
RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT
More informationThe Usefulness of Simultaneously Excited Magnetic Resonance Signals from Diffusion Tensor Image
Journal of Magnetics 23(3), 370-374 (2018) ISSN (Print) 1226-1750 ISSN (Online) 2233-6656 https://doi.org/10.4283/jmag.2018.23.3.370 The Usefulness of Simultaneously Excited Magnetic Resonance Signals
More informationField Simulation Software to Improve Magnetic Resonance Imaging
Field Simulation Software to Improve Magnetic Resonance Imaging a joint project with the NRI in South Korea CST Usergroup Meeting 2010 Darmstadt Institute for Biometry and Medicine Informatics J. Mallow,
More informationExperience in implementing continuous arterial spin labeling on a commercial MR scanner
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 1, WINTER 2005 Experience in implementing continuous arterial spin labeling on a commercial MR scanner Theodore R. Steger and Edward F. Jackson
More informationTesting a wavelet based noise reduction method using computersimulated
Testing a wavelet based noise reduction method using computersimulated mammograms Christoph Hoeschen 1, Oleg Tischenko 1, David R Dance 2, Roger A Hunt 2, Andrew DA Maidment 3, Predrag R Bakic 3 1 GSF-
More informationTomophan TSP004 Manual
T h e P h a n t o m L a b o r a t o r y 1 Tomophan TSP004 Manual Copyright 2016 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order and
More informationSUPPLEMENTARY INFORMATION
a b STS IOS IOS STS c "#$"% "%' STS posterior IOS dorsal anterior ventral d "( "& )* e f "( "#$"% "%' "& )* Supplementary Figure 1. Retinotopic mapping of the non-lesioned hemisphere. a. Inflated 3D representation
More informationAdvanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University
Advanced MSK MRI Protocols at 3.0T Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University Outline Why High Field for MSK? SNR and Relaxation Times Technical Issues Example
More informationApplication Guide & Release Notes
Application Guide & Release Notes Inner-volume-imaging (IVI) EPI C2P Release 002a 1 September 2015 TMII Translational and Molecular Imaging Institute Conditions of Use This package is provided to support
More information12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees
Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees
More informationDiffusion and Functional MRI of the Spinal Cord Methods and Clinical Applications
Diffusion and Functional MRI of the Spinal Cord Methods and Clinical Applications Susceptibility artifacts in DTI of the spinal cord J. Cohen-Adad Q-space imaging and axon diameter measurements Functional
More informationMRI Metal Artifact Reduction
MRI Metal Artifact Reduction PD Dr. med. Reto Sutter University Hospital Balgrist Zurich University of Zurich OUTLINE Is this Patient suitable for MR Imaging? Metal artifact reduction Is this Patient suitable
More informationMulti-spectral acoustical imaging
Multi-spectral acoustical imaging Kentaro NAKAMURA 1 ; Xinhua GUO 2 1 Tokyo Institute of Technology, Japan 2 University of Technology, China ABSTRACT Visualization of object through acoustic waves is generally
More informationSpiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils
Magn Reson Med Sci doi:10.2463/mrms.tn.2016-0049 Published Online: March 27, 2017 TECHNICAL NOTE Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils
More informationConfocal Imaging Through Scattering Media with a Volume Holographic Filter
Confocal Imaging Through Scattering Media with a Volume Holographic Filter Michal Balberg +, George Barbastathis*, Sergio Fantini % and David J. Brady University of Illinois at Urbana-Champaign, Urbana,
More informationWorks-in-Progress package Version 1.0. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B. January 22, 2003
Works-in-Progress package Version 1.0 For the Installation and User s Guide NUMARIS/4VA21B January 22, 2003 Section of Medical Physics, University Hospital Freiburg, Germany Contact: Klaus Scheffler PhD,
More information(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010
(N)MR Imaging Lab Course Script FMP PhD Autumn School Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder Date: November 3rd, 2010 1 Purpose: Understanding the basic principles of MR imaging
More informationSIGNA Explorer Lift revives our MR
Seiji Shiotani, MD, PhD Seirei Fuji Hospital in Fuji City, Shizuoka, Japan Masayoshi Sugimura Seirei Fuji Hospital in Fuji City, Shizuoka, Japan SIGN Explorer Lift revives our MR The clinical usefulness
More informationHigh-Resolution, Spin-Echo BOLD, and CBF fmri at 4and7T
Magnetic Resonance in Medicine 48:589 593 (2002) High-Resolution, Spin-Echo BOLD, and CBF fmri at 4and7T Timothy Q. Duong,* Essa Yacoub, Gregory Adriany, Xiaoping Hu, Kamil Ugurbil, J. Thomas Vaughan,
More informationEnhancing Gray-to-White Matter Contrast in 3T T1 Spin-Echo Brain Scans by Optimizing Flip Angle
AJNR Am J Neuroradiol 26:2000 2004, September 2005 Enhancing Gray-to-White Matter Contrast in 3T T1 Spin-Echo Brain Scans by Optimizing Flip Angle Bernd L. Schmitz, Georg Grön, Florian Brausewetter, Martin
More informationNumerical Evaluation of an 8-element Phased Array Torso Coil for Magnetic Resonance Imaging
Numerical Evaluation of an 8-element Phased Array Torso Coil for Magnetic Resonance Imaging Feng Liu, Joe Li, Ian Gregg, Nick Shuley and Stuart Crozier School of Information Technology and Electrical Engineering,
More informationTissue classification based on relaxation environments
Rochester Institute of Technology RIT Scholar Works Theses Thesis/Dissertation Collections 1998 Tissue classification based on relaxation environments Jordan Guinn Follow this and additional works at:
More informationNoninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park
Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park Department of Bio and Brain Engineering, Korea Advanced Institute of Science and Technology (KAIST), Daejeon,
More informationThe ACR magnetic resonance accreditation phantom (ACR MRAP) has been designed to examine a broad range of instrument parameters.
OVERVIEW OF THE ACR MRI ACCREDITATION PHANTOM Geoffrey D. Clarke, Ph.D. University of Texas Southwestern Medical Center at Dallas email: GEOFFREY.CLARKE@MAIL.SWMED.EDU INTRODUCTION The ACR magnetic resonance
More informationChapter 2. The Physics of Magnetic Resonance Imaging
Chapter 2. The Physics of Magnetic Resonance Imaging 2.1. Introduction The origins of the Nuclear Magnetic Resonance (NMR) signal and how it is manipulated to form images are the subjects of this chapter.
More informationGE Healthcare. Discovery MR T. Simply powerful. Powerfully simple.
GE Healthcare Discovery MR750 3.0T Simply powerful. Powerfully simple. Break free. The breast images you need in only two sequences. A complete liver study in a 15-minute time slot. Routine fmri with shorter
More informationHalf-Pulse Excitation Pulse Design and the Artifact Evaluation
Half-Pulse Excitation Pulse Design and the Artifact Evaluation Phillip Cho. INRODUCION A conventional excitation scheme consists of a slice-selective RF excitation followed by a gradient-refocusing interval
More informationImprove Image Quality of Transversal Relaxation Time PROPELLER and FLAIR on Magnetic Resonance Imaging
Journal of Physics: Conference Series PAPER OPEN ACCESS Improve Image Quality of Transversal Relaxation Time PROPELLER and FLAIR on Magnetic Resonance Imaging To cite this article: N Rauf et al 2018 J.
More information