Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging
|
|
- Frederica Ray
- 5 years ago
- Views:
Transcription
1 Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging Poster No.: C-0709 Congress: ECR 2014 Type: Scientific Exhibit Authors: E. Ebihara 1, H. Nagashima 2, N. Hayashi 2, M. Okawara 1, M. Kudo 1, R. Kitazaki 1, S. Imai 3, T. Ogura 2, K. Doi 2 ; 1 Maebashi-city/JP, 2 Maebashi/JP, 3 Maebashi City/JP Keywords: DOI: Osteoporosis, Physics, MR, MR physics /ecr2014/C-0709 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. Page 1 of 14
2 Aims and objectives The zero-filling interpolation (ZIP) technique in magnetic resonance imaging (MRI) is a method that can expand the original matrix size without increasing the scanning time by placing the acquired data in the central regions on k-space, and by filling the data of zero in the outer regions at high spatial frequencies. Figure 1 illustrates ZIP processing in k- space, which has been reported to improve visual sharpness of MR images [1]. On the other hand, it has been reported also that detailed objects were not confirmed clearly by applying ZIP processing [2]. Therefore, we investigated the characteristics of ZIP processing by performing physical and visual evaluations, and examined the usefulness of ZIP. Images for this section: Fig. 1: Illustration of zero-filling interpolation (ZIP) technique in k-space. Page 2 of 14
3 Methods and materials A slit image with matrix size of pixels was prepared by the width of one pixel in the central portion of the image. The shaded slit image was also created by applying a Gaussian filter with a kernel size of 9 9 pixels to simulated slit image. Figure 2 shows the line spread function (LSF) for slit image without and with Gaussian filter. These slit images were reconstructed to square images of 1024 and 2048 pixels as side length by using our ZIP software. Figure 3 shows the overall scheme for the automated application of ZIP processing in MR images. The LSF was determined from these slit images, and the modulation transfer function (MTF) was then obtained by applying the Fourier transform. Furthermore, the uniform images and pin pattern images were obtained by scanning an MRI phantom with use of 1.5 T-MRI scanners (Ingenia, PHILIPS). For the scanning conditions, the fast-spin echo sequence was chosen, and the scan parameters were set as follows: repetition time = 400 ms, echo time = 16 ms, number of signal averages = 1, slice thickness = 1.7 mm, field of view (FOV) = and mm, and matrix size = pixels. These two images were reconstructed to square images of 512, 1024, and 2048 pixels as side length. For comparison between the scanned original images and reconstructed images, the signal-to-noise ratio (SNR) was evaluated by using these uniform images. Moreover, the shape reproducibility and the resolution for the pin pattern images were evaluated visually by using Scheffe's method of paired comparison with ten observers. As a visual evaluation method, various images obtained with and without ZIP processing were displayed side by side on a liquid-crystal display (LCD) monitor. Two different images were then evaluated on a discrete rating scale. Images for this section: Page 3 of 14
4 Fig. 2: Slit images and signal profile curves: (a) without and, (b) with Gaussian filtering with a kernel size of 9 9 pixels. Page 4 of 14
5 Fig. 3: Overall computerized scheme for automated application of ZIP processing in MR images. Page 5 of 14
6 Results Figure 4 shows the signal characteristic of ZIP processing in simulated slit images. On the LSF with ZIP processing, the findings of undershoots and noise appeared. However, the amplitude of these findings was reduced in the LSF of slit image with Gaussian filtering. Figures 5 and 6 show the MTF determined from two different slit images with and without ZIP processing. In Figure 5, the MTF in reconstructed image by ZIP processing without filtering was over 1.0 at low frequencies, because of undershoots appeared on the LSF. As a measure of the spatial resolution, we determined the spatial frequency at the MTF value of 0.50 in figure 6, and the increase in spatial frequency was observed by applying the ZIP processing. These results seem to suggest that ZIP processing can increase slightly sharpness in the MR image. Figure 7 shows the SNRs in original images acquired with different matrix size (pixel size) and reconstructed images with ZIP processing. It is apparent that the SNRs in all conditions were comparable. We conclude that the noise appeared by the ZIP processing was very similar to the original noise in image acquired by MRI scanner probably because the noise with ZIP processing was embedded in the acquired image. Figure 8 shows results of the visual evaluation using Scheffe's method of paired comparisons. In Figures 8(a) and (b), a significant difference was observed between the original images and the reconstructed images with ZIP processing (P < 0.001). Furthermore, for the original images acquired with FOV of mm in Fig. 8(b), there were significant differences among the images with ZIP processing for different matrix size (P < 0.001). Figure 9 shows the pin pattern images acquired with FOV of mm and three different images reconstructed with ZIP processing. In comparing images of 0.5 mm as a pixel size with and without the ZIP processing, the reproducibility of the object shape among images appeared different visually and improved noticeably by applying ZIP processing. Figure 10 shows the pin pattern images acquired with FOV of mm and reconstructed images with ZIP processing. During the visual evaluation using original images at the pixel size of 1.0 mm, all of signals for pin size of 1.0 mm or more could be identified with ZIP processing. This result shows that it is possible to separate the connected, distorted pin signals by using ZIP processing. However, signals less than 1.0 mm showed no improvement. Therefore, the ZIP processing might be able to separate the connected signals such as adjacent vessels in a clinical image when the size of signals may be larger than the pixel size in scanning. Thus, it seems that ZIP processing would be useful for imaging with a wide FOV, and for improving in sharpness of small objects when the scanning time is reduced. Images for this section: Page 6 of 14
7 Fig. 4: Signal profile curves: (a), (b) without Gaussian filtering; (c), (d) with Gaussian filtering; (a), (c) without ZIP processing; and (b), (d) with ZIP processing at matrix size of pixels. Page 7 of 14
8 Fig. 5: Modulation transfer function (MTF) measured from slit images without Gaussian filtering. Page 8 of 14
9 Fig. 6: MTF measured from slit images with Gaussian filtering. Page 9 of 14
10 Fig. 7: Signal-to-noise ratio (SNR) in original images acquired with different matrix size (pixel size) and reconstructed images with ZIP processing. Page 10 of 14
11 Fig. 8: Results of visual evaluation by Scheffe's method of paired comparisons using original images and reconstructed ZIP images: (a) original image acquired with mm FOV and pixels matrix size, (b) original image acquired with mm FOV and pixels matrix size. Page 11 of 14
12 Fig. 9: Pin pattern images: (a) original image acquired with mm FOV and pixels matrix size, (b); (c); and (d) reconstructed images of matrix size of , , and pixels, respectively, with ZIP processing. Page 12 of 14
13 Fig. 10: Pin pattern images: (a) original image acquired with mm FOV and pixels matrix size, (b); (c); and (d) reconstructed images of matrix size of , , and pixels, respectively, with ZIP processing. Page 13 of 14
14 Conclusion Our results indicated that the application of ZIP processing can improve the sharpness of small objects in clinical MR images. However, further studies for the verification of clinical usefulness with ZIP processing will be required in the future. Personal information References 1. T. Nakamura, S. Naganawa, T. Koshikawa, et al: High-spatial-resolution MR cisternography of the cerebellopontine angle in 90 seconds with a zero-fill interpolated fast recovery 3D fast asymmetric spin-echo sequence, AJNR Am J Neuroradiol, Vol.23, No.8, pp , A. Ogura, F. Maeda, A. Miyai, et al: The property of zero-fill interpolation on slice plane of magnetic resonance imaging, Japanese Journal of Clinical Radiology, Vol.52, No.6, pp , Page 14 of 14
Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE)
Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE) Poster No.: C-2459 Congress: ECR 2010 Type: Scientific Exhibit Topic: Neuro Authors: T. Yoneda,
More informationComputer applications, PACS, Instrumentation, Equipment, Technology assessment /ecr2012/C-0655
The novel medical diagnostic color monitor automatically distinguishing monochrome/color images and displaying these with their own optimal tone curves Poster No.: C-0655 Congress: ECR 2012 Type: Authors:
More informationInvestigation of Effective DQE (edqe) parameters for a flat panel detector
Investigation of Effective DQE (edqe) parameters for a flat panel detector Poster No.: C-1892 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit D. Bor 1, S. Cubukcu 1, A. Yalcin 1, O.
More informationCorrection of the local intensity nonuniformity artifact in high field MRI
Correction of the local intensity nonuniformity artifact in high field MRI Poster No.: C-0346 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Paper S. Kai, S. Kumazawa, H. Yabuuchi, F. Toyofuku;
More informationAnalysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging
Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging Award: Magna Cum Laude Poster No.: C-1988 Congress: ECR 2014 Type: Scientific Exhibit Authors: T. Hamaguchi,
More informationDetermining acceptance levels for automatic daily image quality control in magnetic resonance imaging
Determining acceptance levels for automatic daily image quality control in magnetic resonance imaging Poster No.: C-1125 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific Exhibit J. I. Peltonen,
More informationEvaluation of no-grid radiography using the digital scattered x-ray removal processing
Evaluation of no-grid radiography using the digital scattered x-ray removal processing Poster No.: C-0416 Congress: ECR 2016 Type: Authors: Scientific Exhibit R. Suzuki 1, T. Goto 1, H. Ogawa 2, N. Amimoto
More informationImprovement of CT image quality with iterative reconstruction idose4
Improvement of CT image quality with iterative reconstruction idose4 Poster No.: C-0387 Congress: ECR 2014 Type: Scientific Exhibit Authors: M.-L. Olsson, K. Norrgren, M. Söderberg; Malmö/SE Keywords:
More informationA comparison between medical-grade liquid crystal display (LCD) and ipad color imaging
A comparison between medical-grade liquid crystal display (LCD) and ipad color imaging Poster No.: C-1377 Congress: ECR 2014 Type: Scientific Exhibit Authors: K. Yoshimura 1, K. Shibata 2, T. Nihashi 1,
More informationStudies on reduction of exposure dose using digital scattered X-ray removal processing
Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.
More informationStudies on reduction of exposure dose using digital scattered X-ray removal processing
Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.
More informationDifference in signal-to-noise-ratio (SNR) from vertical to horizontal scanner position using a 0,25 Tesla Weightbearing
Difference in signal-to-noise-ratio (SNR) from vertical to horizontal scanner position using a 0,25 Tesla Weightbearing scanner Poster No.: C-0672 Congress: ECR 2014 Type: Scientific Exhibit Authors: F.
More informationA study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers
A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers Poster No.: C-3011 Congress: ECR 2010 Type: Topic: Authors: Scientific Exhibit
More informationComparison of high-resolution C-arm cone-beam CT imaging and conventional 3D rotational angiography for the better microvascular visualization
Comparison of high-resolution C-arm cone-beam CT imaging and conventional 3D rotational angiography for the better microvascular visualization Poster No.: C-1040 Congress: ECR 2016 Type: Scientific Exhibit
More informationWorldwide practice of breast MRI: insights from the MIPA study applications - the MIPA study group
Worldwide practice of breast MRI: insights from the MIPA study applications - the MIPA study group Poster No.: B-0757 Congress: ECR 2015 Type: Scientific Paper Authors: G. Di Leo, R. M. Trimboli, I. Ioan,
More informationInfluence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck
Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Poster No.: C-2205 Congress: ECR 2012 Type: Scientific Paper
More informationMulti-slice computed tomography analysis of bullet trajectory in forensic investigation.
Multi-slice computed tomography analysis of bullet trajectory in forensic investigation. Poster No.: C-0510 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Usui, Y. Kawasumi, Y. Hosokai, A. Nakajima,
More informationPractical guidelines for color calibration and quality assurance of medical displays
Practical guidelines for color calibration and quality assurance of medical displays Poster No.: C-1140 Congress: ECR 2017 Type: Educational Exhibit Authors: T. Kimpe, J. Rostang, G. Van Hoey, A. Xthona
More informationAim. Images for this section: Page 2 of 13
Changes in CT number of high atomic number materials with field of view when using an extended CT number to electron density curve and a metal artifact reduction reconstruction algorithm Poster No.: R-0094
More informationThe impact of increasing SID on patient dose in pa abdomen imaging
The impact of increasing SID on patient dose in pa abdomen imaging Poster No.: C-1074 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific Exhibit T. Starc 1, N. Mekis 2 ; 1 Brezovica/SI, 2 Ljubljana/SI
More informationDose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis
Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis Poster No.: C-0585 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Minehiro,
More informationVisualization of sources of scattered radiation from x-ray equipment used for interventional radiology
Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology Poster No.: C-1190 Congress: ECR 2011 Type: Scientific Exhibit Authors: K. Chida, T. Takahashi, D.
More informationLumbar disc height assessment: a comparative study between EOS and digital radiography
Lumbar disc height assessment: a comparative study between EOS and digital radiography Poster No.: C-1178 Congress: ECR 2012 Type: Scientific Paper Authors: V. Freire, M. Benhamou, F. Rannou, S. Poiraudeau,
More informationThe feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE
The feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE Poster No.: C-0022 Congress: ECR 2010 Type: Scientific Exhibit Topic: Abdominal Viscera
More informationThe role of CT in restoration of a very rare Christ wooden statue: a fundamental contribution.
The role of CT in restoration of a very rare Christ wooden statue: a fundamental contribution. Poster No.: C-1514 Congress: ECR 2015 Type: Scientific Exhibit Authors: M. Disaro', P. Sartori; Venice/IT
More informationDevelopment of new dosimeter for measuring dose distribution in CT
Development of new dosimeter for measuring dose distribution in CT Poster No.: C-2925 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology - Without Subtopic Authors: Y. Muramatsu, K.
More informationSuppression of metal artifacts using image-based monoenergetic DECT imaging
Suppression of metal artifacts using image-based monoenergetic DECT imaging Poster No.: C-0519 Congress: ECR 2011 Type: Scientific Paper Authors: B. Krauss, B. Schmidt, M. Sedlmair, T. Flohr; Forchheim/DE
More informationNovel cassette-sized, flat-panel digital radiography (DR) system: Initial clinical and workflow results versus computed radiography (CR)
Novel cassette-sized, flat-panel digital radiography (DR) system: Initial clinical and workflow results versus computed radiography (CR) Poster No.: C-3027 Congress: ECR 2010 Type: Scientific Exhibit Topic:
More informationPROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution
PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution Poster No.: C-1280 Congress: ECR 2011 Type: Educational Exhibit Authors: A. Monteiro, A.
More informationPROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution
PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution Poster No.: C-1280 Congress: ECR 2011 Type: Educational Exhibit Authors: A. Monteiro, A.
More informationImage quality evaluation of turbo-spin echo diffusion weighted image (TSE-DWI) : A phantom study
Image quality evaluation of turbo-spin echo diffusion weighted image (TSE-DWI) : A phantom study Poster No.: C-0631 Congress: ECR 2016 Type: Scientific Exhibit Authors: T. Yoshida, A. Urikura, K. Shirata,
More informationEvaluation of magnetic resonance acoustic noise in 1.5 and 3 Tesla scanners
Evaluation of magnetic resonance acoustic noise in 1.5 and 3 Tesla scanners Poster No.: B-1014 Congress: ECR 2016 Type: Scientific Paper Authors: V. M. F. Silva, I. M. Ramos, J. Moreira, M. Marques; Porto/PT
More informationVoice recognition versus transcriptionist
Voice recognition versus transcriptionist Poster No.: C-0263 Congress: ECR 2010 Type: Audit/Professional Issues Topic: Audit/Professional Issues Authors: R. H. Strahan; Clayton South VIC/AU Keywords: Voice
More informationEncoding of inductively measured k-space trajectories in MR raw data
Downloaded from orbit.dtu.dk on: Apr 10, 2018 Encoding of inductively measured k-space trajectories in MR raw data Pedersen, Jan Ole; Hanson, Christian G.; Xue, Rong; Hanson, Lars G. Publication date:
More informationOptimal speed for dual-energy subtraction neck imaging with flat-panel detector radiography
Optimal speed for dual-energy subtraction neck imaging with flat-panel detector radiography Poster No.: C-2938 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology Authors: H. Machida
More informationC a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information
Ordering Information Please contact us if you have any questions or if you would like a quote or delivery schedule regarding the Catphan phantom. phone 800-525-1190, or 518-692-1190 fax 518-692-3329 mail
More informationTomophan TSP004 Manual
T h e P h a n t o m L a b o r a t o r y 1 Tomophan TSP004 Manual Copyright 2016 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order and
More informationMRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves
MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves Experiment 1 Goal: Examine the effect caused by changing flip angle on image contrast in a simple gradient echo sequence and derive T1-curves. Image
More informationRing-Down versus Comet Tail: Two artifacts uncovered
Ring-Down versus Comet Tail: Two artifacts uncovered Poster No.: C-2288 Congress: ECR 2013 Type: Educational Exhibit Authors: S. Méndez Garrido 1, J. Arce Duran 2, A. Talegón Meléndez 1 ; 1 2 Sevilla/ES,
More informationCardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK
Cardiac MR Dr John Ridgway Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Physics for clinicians: Part I Journal of Cardiovascular Magnetic Resonance 2010, 12:71 http://jcmr-online.com/content/12/1/71
More informationH 2 O and fat imaging
H 2 O and fat imaging Xu Feng Outline Introduction benefit from the separation of water and fat imaging Chemical Shift definition of chemical shift origin of chemical shift equations of chemical shift
More informationNEMA Standards Publication MS (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging
NEMA Standards Publication MS 1-2008 (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging Published by: National Electrical Manufacturers Association 1300 North
More informationComparison of new digital mammography systems: physical characterisation and image quality evaluation
Comparison of new digital mammography systems: physical characterisation and image quality evaluation Poster No.: C-2321 Congress: ECR 2011 Type: Scientific Exhibit Authors: N. Oberhofer; Bolzano/IT DOI:
More informationSpiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils
Magn Reson Med Sci doi:10.2463/mrms.tn.2016-0049 Published Online: March 27, 2017 TECHNICAL NOTE Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils
More informationSampling Efficiency in Digital Camera Performance Standards
Copyright 2008 SPIE and IS&T. This paper was published in Proc. SPIE Vol. 6808, (2008). It is being made available as an electronic reprint with permission of SPIE and IS&T. One print or electronic copy
More information2015 Spin echoes and projection imaging
1. Spin Echoes 1.1 Find f0, transmit amplitudes, and shim settings In order to acquire spin echoes, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week,
More informationSECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS
RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT
More informationImage Quality/Artifacts Frequency (MHz)
The Larmor Relation 84 Image Quality/Artifacts (MHz) 42 ω = γ X B = 2πf 84 0.0 1.0 2.0 Magnetic Field (Tesla) 1 A 1D Image Magnetic Field Gradients Magnet Field Strength Field Strength / Gradient Coil
More informationDetermining MTF with a Slant Edge Target ABSTRACT AND INTRODUCTION
Determining MTF with a Slant Edge Target Douglas A. Kerr Issue 2 October 13, 2010 ABSTRACT AND INTRODUCTION The modulation transfer function (MTF) of a photographic lens tells us how effectively the lens
More informationSIGNA Explorer Lift revives our MR
Seiji Shiotani, MD, PhD Seirei Fuji Hospital in Fuji City, Shizuoka, Japan Masayoshi Sugimura Seirei Fuji Hospital in Fuji City, Shizuoka, Japan SIGN Explorer Lift revives our MR The clinical usefulness
More informationBlack Science Fiction Script Movie Project 2012
Black Science Fiction Script Movie Project 2012 OFFICIAL 2011 ENTRY FORM AND RELEASE Please make sure to fill in ALL fields to ensure that your entry is accepted! Please attach a separate completed form
More information(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010
(N)MR Imaging Lab Course Script FMP PhD Autumn School Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder Date: November 3rd, 2010 1 Purpose: Understanding the basic principles of MR imaging
More informationA Study of Slanted-Edge MTF Stability and Repeatability
A Study of Slanted-Edge MTF Stability and Repeatability Jackson K.M. Roland Imatest LLC, 2995 Wilderness Place Suite 103, Boulder, CO, USA ABSTRACT The slanted-edge method of measuring the spatial frequency
More informationInfluence of Image Enhancement Processing on SFR of Digital Cameras
IS&T s 998 PICS Conference Copyright 998, IS&T Influence of Image Processing on SFR of Digital Cameras Yukio Okano Sharp Corporation, Information Systems Labs. Yamatokoriyama, Nara, JAPAN Abstract The
More informationModule 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review
MRES 7005 - Fast Imaging Techniques Module 2 Artefacts and Imaging Optimisation for single shot methods Content: Introduction Phase error Phase bandwidth Chemical shift review Chemical shift in pixels
More informationThe SENSE Ghost: Field-of-View Restrictions for SENSE Imaging
JOURNAL OF MAGNETIC RESONANCE IMAGING 20:1046 1051 (2004) Technical Note The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging James W. Goldfarb, PhD* Purpose: To describe a known (but undocumented)
More informationBackground (~EE369B)
Background (~EE369B) Magnetic Resonance Imaging D. Nishimura Overview of NMR Hardware Image formation and k-space Excitation k-space Signals and contrast Signal-to-Noise Ratio (SNR) Pulse Sequences 13
More informationORDERING A PHOTOGRAPH FOR PERSONAL USE
ORDERING A PHOTOGRAPH FOR PERSONAL USE Thank you for your interest in ordering an item for personal use from The Arizona Republic or azcentral.com. To order a photograph, full-page print, graphic, or Benson
More informationCHAPTER 8 GENERIC PERFORMANCE MEASURES
GENERIC PERFORMANCE MEASURES M.E. DAUBE-WITHERSPOON Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, United States of America 8.1. INTRINSIC AND EXTRINSIC MEASURES 8.1.1.
More informationIntroduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image
Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology
More informationAUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF DIGITAL RADIOGRAPHY IMAGING SYSTEMS
International Workshop SMART MATERIALS, STRUCTURES & NDT in AEROSPACE Conference NDT in Canada 2011 2-4 November 2011, Montreal, Quebec, Canada AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF
More informationHETERONUCLEAR IMAGING. Topics to be Discussed:
HETERONUCLEAR IMAGING BioE-594 Advanced MRI By:- Rajitha Mullapudi 04/06/2006 Topics to be Discussed: What is heteronuclear imaging. Comparing the hardware of MRI and heteronuclear imaging. Clinical applications
More informationDatasheet of SAW Device
Datasheet of SAW Device SAW Single Filter for Band2 / Unbalanced / 5pin /1109 Murata PN: SAFFB1G96AB0F0A Feature GSM1900/Band2 Rx Note : Murata SAW Component is applicable for Cellular /Cordless phone
More informationMulti-spectral acoustical imaging
Multi-spectral acoustical imaging Kentaro NAKAMURA 1 ; Xinhua GUO 2 1 Tokyo Institute of Technology, Japan 2 University of Technology, China ABSTRACT Visualization of object through acoustic waves is generally
More informationObserver Performance of Reduced X-Ray Images on Liquid Crystal Displays
Original Paper Forma, 29, S45 S51, 2014 Observer Performance of Reduced X-Ray Images on Liquid Crystal Displays Akiko Ihori 1, Chihiro Kataoka 2, Daigo Yokoyama 2, Naotoshi Fujita 3, Naruomi Yasuda 4,
More informationTESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist
TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist Topics Image Uniformity and Artifacts Image Quality - Detail
More informationMARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005
ACR MRI accreditation program MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics,
More informationData Sheet of SAW Components
Data Sheet of SAW Components Note : Murata SAW Component is applicable for Cellular /Cordless phone (Terminal) relevant market only. Please also read caution at the end of this document. Package Dimensions
More informationNUF6105FCT1G. 6-Channel EMI Filter with Integrated ESD Protection
6-Channel EMI Filter with Integrated ESD Protection The NUF615FC is a six channel (C R C) Pi style EMI filter array with integrated ESD protection. Its typical component values of R = 1 and C = 27 pf deliver
More informationNUF8001MUT2G. 8-Channel EMI Filter with Integrated ESD Protection
8-Channel EMI Filter with Integrated ESD Protection The NUF8MU is a eight channel (C R C) Pi style EMI filter array with integrated ESD protection. Its typical component values of R = and C = 2 pf deliver
More informationNUF6400MNTBG. 6-Channel EMI Filter with Integrated ESD Protection
6-Channel EMI Filter with Integrated ESD Protection The NUF64MU is a six channel (C R C) Pi style EMI filter array with integrated ESD protection. Its typical component values of R = and C = 5 pf deliver
More informationTI Designs: Biometric Steering Wheel. Amy Ball TIDA-00292
www.ti.com 2 Biometric Steering Wheel - -Revised July 2014 www.ti.com TI Designs: Biometric Steering Wheel - -Revised July 2014 Biometric Steering Wheel 3 www.ti.com 4 Biometric Steering Wheel - -Revised
More informationPulse Sequence Design and Image Procedures
Pulse Sequence Design and Image Procedures 1 Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant 2 A pulse sequence is a timing diagram designed with a series of RF pulses, gradients switching, and signal readout
More informationDatasheet of SAW Device
Datasheet of SAW Device SAW Single Filter for GPS_GLONASS_BEIDOU / Unbalanced / 5pin /1109 Murata PN: SAFFB1G56KB0F0A Feature Low Insertion Loss Note : Murata SAW Component is applicable for Cellular /Cordless
More informationTimTX TrueShape. The parallel transmit architecture of the future. Answers for life.
www.siemens.com/trueshape TimTX TrueShape The parallel transmit architecture of the future. The product/feature (mentioned herein) is not commercially available. Due to regulatory reasons its future availability
More informationDigital images. Digital Image Processing Fundamentals. Digital images. Varieties of digital images. Dr. Edmund Lam. ELEC4245: Digital Image Processing
Digital images Digital Image Processing Fundamentals Dr Edmund Lam Department of Electrical and Electronic Engineering The University of Hong Kong (a) Natural image (b) Document image ELEC4245: Digital
More informationNUF8401MNT4G. 8-Channel EMI Filter with Integrated ESD Protection
8-Channel EMI Filter with Integrated ESD Protection The NUF841MN is an eight channel (C R C) Pi style EMI filter array with integrated ESD protection. Its typical component values of R = 1 and C = 12 pf
More information2014 M.S. Cohen all rights reserved
2014 M.S. Cohen all rights reserved mscohen@g.ucla.edu IMAGE QUALITY / ARTIFACTS SYRINGOMYELIA Source http://gait.aidi.udel.edu/res695/homepage/pd_ortho/educate/clincase/syrsco.htm Surgery is usually recommended
More informationImage Processing Final Test
Image Processing 048860 Final Test Time: 100 minutes. Allowed materials: A calculator and any written/printed materials are allowed. Answer 4-6 complete questions of the following 10 questions in order
More informationAlae Tracker: Tracking of the Nasal Walls in MR-Imaging
Alae Tracker: Tracking of the Nasal Walls in MR-Imaging Katharina Breininger 1, Andreas K. Maier 1, Christoph Forman 1, Wilhelm Flatz 2, Catalina Meßmer 3, Maria Schuster 3 1 Pattern Recognition Lab, Friedrich-Alexander-Universität
More informationM R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia
M R I Physics Course Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Magnetic Resonance Imaging Spatial
More informationImproved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques
Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques Wataru FUKUDA* Junya MORITA* and Masahiko YAMADA* Abstract Tomosynthesis is a three-dimensional imaging technology
More informationSIEMENS MAGNETOM Skyra syngo MR D13
Page 1 of 12 SIEMENS MAGNETOM Skyra syngo MR D13 \\USER\CIND\StudyProtocols\PTSA\*ep2d_M0Map_p2_TE15 TA:7.9 s PAT:2 Voxel size:2.5 2.5 3.0 mm Rel. SNR:1.00 :epfid Properties Routine Contrast Prio Recon
More informationData Sheet of SAW Components
Data Sheet of SAW Components Note : Murata SAW Component is applicable for Cellular /Cordless phone (Terminal) relevant market only. Please also read caution at the end of this document. 5-0.325±0.050
More information1 Introduction. 2 The basic principles of NMR
1 Introduction Since 1977 when the first clinical MRI scanner was patented nuclear magnetic resonance imaging is increasingly being used for medical diagnosis and in scientific research and application
More informationNUF4401MNT1G. 4-Channel EMI Filter with Integrated ESD Protection
4-Channel EMI Filter with Integrated ESD Protection The is a four channel (C R C) Pi style EMI filter array with integrated ESD protection. Its typical component values of R = 200 and C = 5 pf deliver
More informationIntroduction to digital image processing
Introduction to digital image processing Chapter1 Digital images Visible light is essentially electromagnetic radiation with wavelengths between 400 and 700 nm. Each wavelength corresponds to a different
More informationBWI MARSHALL AIRPORT ART COMPETITION
BWI MARSHALL AIRPORT ART COMPETITION Baltimore/Washington International Thurgood Marshall Airport (BWI Marshall Airport) is currently constructing the D/E Connector and Security Checkpoint. This project
More informationDatasheet of SAW Device
Datasheet of SAW Device SAW Single Filter for Band30 / Unbalanced / 5pin /1109 Murata PN: SAFFB2G35KA1F0A Feature For Diversity Rx High Attenuation Note : Murata SAW Component is applicable for Cellular
More informationPublic Art Network Best Practice Goals and Guidelines
Public Art Network Best Practice Goals and Guidelines The Public Art Network (PAN) Council of Americans for the Arts appreciates the need to identify best practice goals and guidelines for the field. The
More informationEdge-Raggedness Evaluation Using Slanted-Edge Analysis
Edge-Raggedness Evaluation Using Slanted-Edge Analysis Peter D. Burns Eastman Kodak Company, Rochester, NY USA 14650-1925 ABSTRACT The standard ISO 12233 method for the measurement of spatial frequency
More informationW H I T E P A P E R. EMC Countermeasure Techniques in Hardware. Introduction
W H I T E P A P E R Shusaku Suzuki, Techniques for EMC countermeasure in hardware Cypress Semiconductor Corp. EMC Countermeasure Techniques in Hardware Abstract This white paper presents the techniques
More informationDatasheet of SAW Device
Datasheet of SAW Device SAW Single Filter for Band41 / Unbalanced / 5pin /1411 Murata PN: SAFEA2G53MA1F0A Feature Band41 Lower side 2496-2566MHz Note : Murata SAW Component is applicable for Cellular /Cordless
More informationPanoramic imaging. Ixyzϕθλt. 45 degrees FOV (normal view)
Camera projections Recall the plenoptic function: Panoramic imaging Ixyzϕθλt (,,,,,, ) At any point xyz,, in space, there is a full sphere of possible incidence directions ϕ, θ, covered by 0 ϕ 2π, 0 θ
More informationDatasheet of SAW Device
Datasheet of SAW Device SAW Dual Filter for Band1_Band13 / 1in4out Balanced / LH /1511 Murata PN: SAWFD751MCAFA Feature Diplex dual SAW filter Small size 1511 mm 1 ohm balanced Note : Murata SAW Component
More informationAnswer: TGC is needed to amplify echoes from deeper structures so that they appear as bright as similar structures located at more shallow depths.
Q47. When performing a sonogram why the sonographer needs to use the TGC? TGC is needed to amplify echoes from deeper structures so that they appear as bright as similar structures located at more shallow
More informationADC Guide, Part 1 The Ideal ADC
ADC Guide, Part 1 The Ideal ADC By Sachin Gupta and Akshay Phatak, Cypress Semiconductor Analog to Digital Converters (ADCs) are one of the most commonly used blocks in embedded systems. Applications of
More informationTech Paper. Anti-Sparkle Film Distinctness of Image Characterization
Tech Paper Anti-Sparkle Film Distinctness of Image Characterization Anti-Sparkle Film Distinctness of Image Characterization Brian Hayden, Paul Weindorf Visteon Corporation, Michigan, USA Abstract: The
More informationBeam Profiling. Introduction. What is Beam Profiling? by Michael Scaggs. Haas Laser Technologies, Inc.
Beam Profiling by Michael Scaggs Haas Laser Technologies, Inc. Introduction Lasers are ubiquitous in industry today. Carbon Dioxide, Nd:YAG, Excimer and Fiber lasers are used in many industries and a myriad
More informationNoise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner
Noise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner Kisung Lee, Member, IEEE, Paul E. Kinahan, Senior Member, Robert S. Miyaoka, Member, IEEE, Jeffrey A. Fessler,
More information