Evaluation of no-grid radiography using the digital scattered x-ray removal processing
|
|
- Doris Holmes
- 6 years ago
- Views:
Transcription
1 Evaluation of no-grid radiography using the digital scattered x-ray removal processing Poster No.: C-0416 Congress: ECR 2016 Type: Authors: Scientific Exhibit R. Suzuki 1, T. Goto 1, H. Ogawa 2, N. Amimoto 3 ; 1 Oguni/JP, 2 Sendai/JP, 3 Tokyo/JP Keywords: DOI: Computer applications, Digital radiography, Physics, Artifacts /ecr2016/C-0416 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. Page 1 of 15
2 Aims and objectives In portable exposure using Grids, cut off often uneven image density and contrast reduction caused by grid misalignment to the incident x-ray become a problem. In recent years, Virtual Grid Processing has developed by Fujifilm to enable non-grid exposure in digital Radiography. In Virtual Grid Processing, image acquisition is expected to not be affected by the artifact of the cut-off. We have evaluated and compared the image quality of incorrect alignment exposure images and tilted alignment exposure images for using Grid images and Virtual Grid Processing images. Methods and materials Materials: Diagnostic x-ray high voltage unit KXO-50SS (Toshiba Medical Systems, Japan) Flat Panel Detector System CALNEO C 1417 Wireless SQ and CONSOLE ADVANCE Ver. 8.1 (Fujifilm Co., Japan) Anti-scatter grid Real Grid: MS X-RAY GRID (Mitaya MFG Co., Japan) Grid ratio, strip density, focus distance and interspace material were 8:1, 40 lines/cm, 120 cm, aluminum respectively. Virtual Grid: Virtual Grid was set to the same characteristic parameters with Real Grid. Page 2 of 15
3 Dosimeter RaySafe X2 (Unfors RaySafe, Sweden) Step Phantom Polymethyl methacrylate plates (Fig. 1) Analysing software Image J (National Institutes of Health, USA) Exposure technique: Two Step Phantoms (R, L) were located covering whole width of Flat Panel Detector. The step of Step Phantom is orthogonal to grid line. X-ray exposure was done by correct alignment (0 degree) and titled misalignment (3, 6, 9 degrees of flux angle of incorrect x-ray). (Fig. 2) Tube voltage was set to 80 kv, and mas was set to 0.3 mgy (Incident Air Kerma at the surface of Step Phantom: IAK) for Grid exposure, and 0.15 mgy (IAK) for Virtual Grid processing exposure. Qualitative evaluation: It was visually evaluated for the uneven density and grid lines of the image. Quantitative evaluation: 10 mm x 10 mm ROI was set in the image, and moved every 5 mm in horizontal direction in the image, and average value and standard deviation of pixel value were calculated. (Fig. 3) Density profile curve was made from average value of pixel value. Contrast to Noise Ratio (CNR) was calculated by average value and standard deviation, and profile curve was obtained. Formula for CNR is shown in Fig. 4. Page 3 of 15
4 Images for this section: Fig. 1: Diagram of PMMA Step Phantom Page 4 of 15
5 Fig. 2: Experimental setup for image acquisition Page 5 of 15
6 Fig. 3: ROI for CNR measurement on the Step Phantom Image Page 6 of 15
7 Fig. 4: Formula for Contrast to Noise Ratio (CNR) Page 7 of 15
8 Results Exposure with Grid Correct alignment even uneven density has been confirmed. (Fig. 5) Grid line could be clearly visible in the incident angle 9 degrees. (Fig. 5) It became a density profile of different shapes for each incident angle. (Fig. 6) When the angle increase the contrast is lowered. (Fig. 6) CNR in the incident angle 9 degrees was reduced up to 40 % of correct alignment. (Fig. 7) Exposure with Virtual Grid Processing The density of the region far from the X-ray focal point in accordance with incident angle increases was reduced. (Fig. 8, 9) Image contrast was certain to be independent of the incident angle. (Fig. 9) Page 8 of 15
9 CNR in correct alignment was the same as the exposure with Grid. (Fig. 10) CNR of the distant region from the X-ray focal spot, the incident angle is slightly reduced in accordance with increase. (Fig. 10) Images for this section: Page 9 of 15
10 Fig. 5: Step Phantom Image obtained with Grid technique. Fig. 6: Density profile curve of Step Phantom Image obtained with Grid technique. Page 10 of 15
11 Fig. 7: CNR profile curve of Step Phantom Image obtained with Grid technique. Page 11 of 15
12 Fig. 8: Step Phantom Image obtained with Virtual Grid Processing. Page 12 of 15
13 Fig. 9: Density profile curve of Step Phantom Image obtained with Virtual Grid Processing. Page 13 of 15
14 Fig. 10: CNR profile curve of Step Phantom Image obtained with Virtual Grid Processing. Page 14 of 15
15 Conclusion In exposure with Virtual Grid Processing, even with half of dose of exposure with grid, equivalent image quality was obtained. The CNR does not change greatly when incident angle of X-ray of FPD become larger, however, gentle gradient of image density can be seen, therefore, in portable exposure, misalignment has to be small as much as possible. Personal information References Michael D. Carlin, Robert M. Nishikawa, Heber MacMahon, and Kunio Doi. The effect of x-ray beam alignment on the performance of antiscatter grids. Med. Phys. 23 (8): (1996). Page 15 of 15
Studies on reduction of exposure dose using digital scattered X-ray removal processing
Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.
More informationStudies on reduction of exposure dose using digital scattered X-ray removal processing
Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.
More informationInvestigation of Effective DQE (edqe) parameters for a flat panel detector
Investigation of Effective DQE (edqe) parameters for a flat panel detector Poster No.: C-1892 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit D. Bor 1, S. Cubukcu 1, A. Yalcin 1, O.
More informationComputer applications, PACS, Instrumentation, Equipment, Technology assessment /ecr2012/C-0655
The novel medical diagnostic color monitor automatically distinguishing monochrome/color images and displaying these with their own optimal tone curves Poster No.: C-0655 Congress: ECR 2012 Type: Authors:
More informationThe impact of increasing SID on patient dose in pa abdomen imaging
The impact of increasing SID on patient dose in pa abdomen imaging Poster No.: C-1074 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific Exhibit T. Starc 1, N. Mekis 2 ; 1 Brezovica/SI, 2 Ljubljana/SI
More informationA study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers
A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers Poster No.: C-3011 Congress: ECR 2010 Type: Topic: Authors: Scientific Exhibit
More informationDose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis
Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis Poster No.: C-0585 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Minehiro,
More informationImprovement of CT image quality with iterative reconstruction idose4
Improvement of CT image quality with iterative reconstruction idose4 Poster No.: C-0387 Congress: ECR 2014 Type: Scientific Exhibit Authors: M.-L. Olsson, K. Norrgren, M. Söderberg; Malmö/SE Keywords:
More informationDevelopment of new dosimeter for measuring dose distribution in CT
Development of new dosimeter for measuring dose distribution in CT Poster No.: C-2925 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology - Without Subtopic Authors: Y. Muramatsu, K.
More informationCorrection of the local intensity nonuniformity artifact in high field MRI
Correction of the local intensity nonuniformity artifact in high field MRI Poster No.: C-0346 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Paper S. Kai, S. Kumazawa, H. Yabuuchi, F. Toyofuku;
More informationFundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging
Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging Poster No.: C-0709 Congress: ECR 2014 Type: Scientific
More informationAim. Images for this section: Page 2 of 13
Changes in CT number of high atomic number materials with field of view when using an extended CT number to electron density curve and a metal artifact reduction reconstruction algorithm Poster No.: R-0094
More informationVisualization of sources of scattered radiation from x-ray equipment used for interventional radiology
Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology Poster No.: C-1190 Congress: ECR 2011 Type: Scientific Exhibit Authors: K. Chida, T. Takahashi, D.
More informationInfluence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck
Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Poster No.: C-2205 Congress: ECR 2012 Type: Scientific Paper
More informationA comparison between medical-grade liquid crystal display (LCD) and ipad color imaging
A comparison between medical-grade liquid crystal display (LCD) and ipad color imaging Poster No.: C-1377 Congress: ECR 2014 Type: Scientific Exhibit Authors: K. Yoshimura 1, K. Shibata 2, T. Nihashi 1,
More informationClear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE)
Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE) Poster No.: C-2459 Congress: ECR 2010 Type: Scientific Exhibit Topic: Neuro Authors: T. Yoneda,
More informationAnalysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging
Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging Award: Magna Cum Laude Poster No.: C-1988 Congress: ECR 2014 Type: Scientific Exhibit Authors: T. Hamaguchi,
More informationDetermining acceptance levels for automatic daily image quality control in magnetic resonance imaging
Determining acceptance levels for automatic daily image quality control in magnetic resonance imaging Poster No.: C-1125 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific Exhibit J. I. Peltonen,
More informationNovel cassette-sized, flat-panel digital radiography (DR) system: Initial clinical and workflow results versus computed radiography (CR)
Novel cassette-sized, flat-panel digital radiography (DR) system: Initial clinical and workflow results versus computed radiography (CR) Poster No.: C-3027 Congress: ECR 2010 Type: Scientific Exhibit Topic:
More informationMulti-slice computed tomography analysis of bullet trajectory in forensic investigation.
Multi-slice computed tomography analysis of bullet trajectory in forensic investigation. Poster No.: C-0510 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Usui, Y. Kawasumi, Y. Hosokai, A. Nakajima,
More informationLumbar disc height assessment: a comparative study between EOS and digital radiography
Lumbar disc height assessment: a comparative study between EOS and digital radiography Poster No.: C-1178 Congress: ECR 2012 Type: Scientific Paper Authors: V. Freire, M. Benhamou, F. Rannou, S. Poiraudeau,
More informationPractical guidelines for color calibration and quality assurance of medical displays
Practical guidelines for color calibration and quality assurance of medical displays Poster No.: C-1140 Congress: ECR 2017 Type: Educational Exhibit Authors: T. Kimpe, J. Rostang, G. Van Hoey, A. Xthona
More informationComparison of high-resolution C-arm cone-beam CT imaging and conventional 3D rotational angiography for the better microvascular visualization
Comparison of high-resolution C-arm cone-beam CT imaging and conventional 3D rotational angiography for the better microvascular visualization Poster No.: C-1040 Congress: ECR 2016 Type: Scientific Exhibit
More informationSuppression of metal artifacts using image-based monoenergetic DECT imaging
Suppression of metal artifacts using image-based monoenergetic DECT imaging Poster No.: C-0519 Congress: ECR 2011 Type: Scientific Paper Authors: B. Krauss, B. Schmidt, M. Sedlmair, T. Flohr; Forchheim/DE
More informationDifference in signal-to-noise-ratio (SNR) from vertical to horizontal scanner position using a 0,25 Tesla Weightbearing
Difference in signal-to-noise-ratio (SNR) from vertical to horizontal scanner position using a 0,25 Tesla Weightbearing scanner Poster No.: C-0672 Congress: ECR 2014 Type: Scientific Exhibit Authors: F.
More informationWorldwide practice of breast MRI: insights from the MIPA study applications - the MIPA study group
Worldwide practice of breast MRI: insights from the MIPA study applications - the MIPA study group Poster No.: B-0757 Congress: ECR 2015 Type: Scientific Paper Authors: G. Di Leo, R. M. Trimboli, I. Ioan,
More informationOptimal speed for dual-energy subtraction neck imaging with flat-panel detector radiography
Optimal speed for dual-energy subtraction neck imaging with flat-panel detector radiography Poster No.: C-2938 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology Authors: H. Machida
More informationThe role of CT in restoration of a very rare Christ wooden statue: a fundamental contribution.
The role of CT in restoration of a very rare Christ wooden statue: a fundamental contribution. Poster No.: C-1514 Congress: ECR 2015 Type: Scientific Exhibit Authors: M. Disaro', P. Sartori; Venice/IT
More informationPROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution
PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution Poster No.: C-1280 Congress: ECR 2011 Type: Educational Exhibit Authors: A. Monteiro, A.
More informationPROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution
PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution Poster No.: C-1280 Congress: ECR 2011 Type: Educational Exhibit Authors: A. Monteiro, A.
More informationComparison of new digital mammography systems: physical characterisation and image quality evaluation
Comparison of new digital mammography systems: physical characterisation and image quality evaluation Poster No.: C-2321 Congress: ECR 2011 Type: Scientific Exhibit Authors: N. Oberhofer; Bolzano/IT DOI:
More informationY11-DR Digital Radiography (DR) Image Quality
Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures
More informationThe feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE
The feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE Poster No.: C-0022 Congress: ECR 2010 Type: Scientific Exhibit Topic: Abdominal Viscera
More informationVoice recognition versus transcriptionist
Voice recognition versus transcriptionist Poster No.: C-0263 Congress: ECR 2010 Type: Audit/Professional Issues Topic: Audit/Professional Issues Authors: R. H. Strahan; Clayton South VIC/AU Keywords: Voice
More informationYour worldwide source of grids. What a difference a grid makes
Your worldwide source of grids What a difference a grid makes The right assembly to meet your every need Trust the experts A high degree of expertise is required to design and manufacture quality X-ray
More informationEvaluation of magnetic resonance acoustic noise in 1.5 and 3 Tesla scanners
Evaluation of magnetic resonance acoustic noise in 1.5 and 3 Tesla scanners Poster No.: B-1014 Congress: ECR 2016 Type: Scientific Paper Authors: V. M. F. Silva, I. M. Ramos, J. Moreira, M. Marques; Porto/PT
More informationAcceptance Testing of a Digital Breast Tomosynthesis Unit
Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures
More informationTest Equipment for Radiology and CT Quality Control Contents
Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3
More informationQC Testing for Computed Tomography (CT) Scanner
QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including
More informationEstimation of signal transfer property for wireless digital detector in different measurement schemes
Estimation of signal transfer property for wireless digital detector in different measurement schemes Anatoli Vladimirov, Kalle Kepler Training Centre of Medical Physics, University of Tartu, Estonia 11
More informationMaximizing clinical outcomes
Maximizing clinical outcomes Digital Tomosynthesis Dual Energy Subtraction Automated Long Length Imaging Improved image quality at a low dose Xray Xray Patented ISS capture technology promotes high sensitivity
More informationTruly flexible to meet your clinical needs
Truly flexible to meet your clinical needs 2 Adapting to meet your needs Flexible Fast and responsive Excellent image quality Designed with ergonomic efficiency Equipped with dose management tools 3 Three
More informationBreast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School
Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis
More informationImage quality evaluation of turbo-spin echo diffusion weighted image (TSE-DWI) : A phantom study
Image quality evaluation of turbo-spin echo diffusion weighted image (TSE-DWI) : A phantom study Poster No.: C-0631 Congress: ECR 2016 Type: Scientific Exhibit Authors: T. Yoshida, A. Urikura, K. Shirata,
More informationHalf value layer and AEC receptor dose compliance survey in Estonia
Half value layer and AEC receptor dose compliance survey in Estonia K. Kepler, A. Vladimirov Training Centre of Medical Physics, University of Tartu Testing Centre of the University of Tartu, Estonia E-mail:
More informationDerivation of total filtration thickness for diagnostic x-ray source assembly
Physics in Medicine & Biology PAPER OPEN ACCESS Derivation of total filtration thickness for diagnostic x-ray source assembly To cite this article: Michiharu Sekimoto and Yoh Katoh 2016 Phys. Med. Biol.
More informationPERFORMANCE MEASUREMENT OF MEDICAL IMAGING SYSTEMS BASED ON MUTUAL INFORMATION METRIC
XIX IMEKO World Congress Fundamental and Applied Metrology September 6 11, 2009, Lisbon, Portugal PERFORMANCE MEASUREMENT OF MEDICAL IMAGING SYEMS BASED ON MUTUAL INFORMATION METRIC Eri Matsuyama 1, Du-Yih
More informationIBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY
WHITE PAPER: IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY IBEX Innovations Ltd. Registered in England and Wales: 07208355 Address: Discovery 2, NETPark, William Armstrong Way, Sedgefield, UK Patents:
More informationImaging Technique Optimization of Tungsten Anode FFDM System
Imaging Technique Optimization of Tungsten Anode FFDM System Biao Chen a*, Andrew P. Smith b, Zhenxue Jing a, Elena Ingal a a Hologic, Inc. 600 Technology Drive, DE 1970 b Hologic, Inc. 35 Crosby Drive,
More informationGrid-like contrast enhancement for bedside chest radiographs acquired without anti-scatter grid
-like contrast enhancement for bedside chest radiographs acquired without anti-scatter grid Philips Detlef Mentrup, PhD, Image Processing Specialist Ulrich Neitzel, PhD, Clinical Scientist Sascha Jockel,
More informationComputed Radiography
BAM Berlin Computed Radiography --INDE 2007, Kalpakkam, India -- Uwe Zscherpel, Uwe Ewert BAM Berlin, Division VIII.3 Requests Requests and and information information to: to: Dr. Dr. U. U. Zscherpel Zscherpel
More informationRaySafe X2. Effortless measurements of X-ray
RaySafe X2 Effortless measurements of X-ray At your fingertips We ve grown accustomed to intuitive interactions with our devices. After all, it s not the device that s most important, but what you can
More informationAppropriate Inspection Distance of Digital X-Ray Imaging Equipment for Diagnosis
Indian Journal of Science and Technology Vol 8(S8), 380-386, April 2015 ISSN (Print) : 0974-6846 ISSN (Online) : 0974-5645 DOI: 10.17485/ijst/2015/v8iS8/70528 Appropriate Inspection Distance of Digital
More informationuniversal Digital Radiography (DR) System
Vision Air universal Digital Radiography (DR) System Technical Specification Vision Air is a state-of-the-art, universal, overhead digital radiography system capable of virtually all radiography examinations
More informationDISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter
DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter Revision Date: January 5th, 2017 www.disc-imaging.com Table of Contents Section A: Preliminary Setup Requirements... 4 Tools
More informationCollimation Assessment Using GAFCHROMIC XR-M2
Collimation Assessment Using GAFCHROMIC XR-M2 I. Introduction A method of collimation assessment for GE Senographe full-field digital mammography (FFDM) systems is described that uses a self-developing
More informationDose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp
Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division
More informationR/F. Comparison of Long View Radiography Systems. 1. Introduction. 2. Methods of Long View Radiography
R/F Comparison of Long View Radiography Systems Department of Radiology, Tokyo Women's Medical University Medical Center East 1 Department of Central Radiology, Tokyo Women's Medical University Hospital
More informationX-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto
X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring
More informationDelftDI Mobile DR. The next generation mobile X-ray system. JARNÍ SYMPOZIUM BRNO 2016 David Navrátil
DelftDI Mobile DR The next generation mobile X-ray system JARNÍ SYMPOZIUM BRNO 2016 David Navrátil DelftDI systems Mobile DR current market 3 Mobile DR- current situation Mobility Competitors Comparison
More informationRAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II
RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II APPROVED 12/O2/2011 EFFECTIVE SPRING 2013-14 Prefix & Number RAD 150 Course Title: Radiologic Exposure Technique II & Lab Purpose of this submission: New Change/Updated
More informationExposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?
Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago
More informationPLD5600A High Frequency Digital Gastrointestinal &DR System(630mA)
PLD5600A High Frequency Digital Gastrointestinal &DR System(630mA) Application: Full support perspective, gastrointestinal spot film, GI (barium meal, barium enema), orthopedic photography, pediatrics
More informationAyaz Jhanorwala, Rishikesh Kumar, Satish Tilva. L & T MHPS Boilers Pvt Ltd., Hazira, Surat , India
National Seminar & Exhibition on Non-Destructive Evaluation, NDE 2014, Pune, December 4-6, 2014 (NDE-India 2014) Vol.20 No.6 (June 2015) - The e-journal of Nondestructive Testing - ISSN 1435-4934 www.ndt.net/?id=17842
More informationIBEX MATERIALS DETECTION TECHNOLOGY
WHITE PAPER: IBEX MATERIALS DETECTION TECHNOLOGY IBEX Innovations Ltd. Registered in England and Wales: 07208355 Address: Discovery 2, NETPark, William Armstrong Way, Sedgefield, TS21 3FH, UK Patents held
More informationTailoring automatic exposure control toward constant detectability in digital mammography
Tailoring automatic exposure control toward constant detectability in digital mammography Elena Salvagnini a) Department of Imaging and Pathology, Medical Physics and Quality Assessment, KUL, Herestraat
More informationJoint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009
2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced
More informationX-RAYS - NO UNAUTHORISED ENTRY
Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are
More informationinspexio SMX-225CT FPD HR
Microfocus X-Ray CT System C251-E029A Advanced Operability and Excellent Image Quality That Overturns Conventional Assumptions Microfocus X-Ray CT System The is a high-performance microfocus X-ray CT system
More informationPredicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities
Journal of Physics: Conference Series PAPER OPEN ACCESS Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities Recent citations - Resolution Properties of a
More informationCOMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR
This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura New Technology https://www.youtube.com/watch?v=ptkzznazb 7U COMPUTED
More informationCHAPTER 6 QC Test For Fluoroscopic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC
CHAPTER 6 QC Test For Fluoroscopic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done. Identify QC
More informationMoving from film to digital: A study of digital x-ray benefits, challenges and best practices
Moving from film to digital: A study of digital x-ray benefits, challenges and best practices H.U. Pöhler 1 and N. D Ademo 2 DÜRR NDT GmbH & Co. KG, Höpfigheimer Straße 22, Bietigheim-Bissingen, 74321,
More informationRing-Down versus Comet Tail: Two artifacts uncovered
Ring-Down versus Comet Tail: Two artifacts uncovered Poster No.: C-2288 Congress: ECR 2013 Type: Educational Exhibit Authors: S. Méndez Garrido 1, J. Arce Duran 2, A. Talegón Meléndez 1 ; 1 2 Sevilla/ES,
More informationPERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY
12 th A-PCNDT 2006 Asia-Pacific Conference on NDT, 5 th 10 th Nov 2006, Auckland, New Zealand PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY Rajashekar
More informationDigital radiography. bucky table and wall stand as a dual detector or wireless system. Amadeo DR Systems
Amadeo DR Systems with dicom PACS DX-R Software X-ray Systems for the Future Digital radiography with Amadeo R-DR including bucky table and wall stand as a dual detector or wireless system High contrast
More informationCT radiation profile width measurement using CR imaging plate raw data
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 CT radiation profile width measurement using CR imaging plate raw data Thorarin A Bjarnason, 1,2,3a Chang-Ying Joseph Yang 3,4 Diagnostic
More informationImproved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques
Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques Wataru FUKUDA* Junya MORITA* and Masahiko YAMADA* Abstract Tomosynthesis is a three-dimensional imaging technology
More informationOpen. the Digitized world. Fuji Computed Radiography
Open the Digitized world Fuji Computed Radiography If just one of these applies to you... Managing developing fluid is hard and darkroom work is a hassle... Images are not stable... Isn t digitalization
More informationSEAMS DUE TO MULTIPLE OUTPUT CCDS
Seam Correction for Sensors with Multiple Outputs Introduction Image sensor manufacturers are continually working to meet their customers demands for ever-higher frame rates in their cameras. To meet this
More informationUnfors EDD-30 Radiation Protection in Fluoroscopy
Unfors EDD-30 Radiation Protection in Fluoroscopy Immediate Warning Decrease Your Dose Interventional radiology procedures are considered to be essential to medical diagnosis and treatment. It is recognized,
More informationI. Introduction.
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Accuracy of measuring half- and quarter-value layers and appropriate aperture width of a convenient method using a lead-covered case
More informationHardware for High Energy Applications 30 October 2009
Paper No. 003 09 Hardware for High Energy Applications 30 October 2009 This document was created by the Federal Working Group on Industrial Digital Radiography. Reproduction is authorized. Federal Working
More informationBeam-Restricting Devices
Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the
More informationInvestigation of the line-pair pattern method for evaluating mammographic focal spot performance
Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann
More informationDosepix Detector as kvp-meter in Radiology and Mammography: First steps
Dosepix Detector as kvp-meter in Radiology and Mammography: First steps F.Bisello, I.Ritter, F.Tennert, A.Zang MediPix Collaboration Meeting, 19th February 2014, CERN Protect, Enhance, and Save Lives -
More informationMeasurement of table feed speed in modern CT
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 3, 2014 Measurement of table feed speed in modern CT Atsushi Fukuda, 1,2a Pei-Jan P. Lin, 3 Kosuke Matsubara, 2 Tosiaki Miyati 2 Department
More informationIntroduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image
Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology
More informationMultiple Choice Identify the letter of the choice that best completes the statement or answers the question.
RA110 test 3 Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. 1. An object 35 cm in width is radiographed at 100 cm SID and at a 50 cm SOD. What
More informationDRAFT Technical evaluation of Philips Microdose SI digital mammography system
DRAFT Technical evaluation of Philips Microdose SI digital mammography system NHSBSP Equipment Report 1310 August 2013 About the NHS Cancer Screening Programmes The national office of the NHS Cancer Screening
More informationRADspeed Pro. EDGEpackage C501-E041C
RADspeed Pro EDGEpackage C501-E041C 2 Some of the FPDs may be not available in your country. Please contact us to check the availability in your country. 3 Tomosynthesis in the Standing Position Tomosynthesis
More informationMXHF-1500RF is controlled by Digital key panel console that displays KV, ma and mas with APR menu programmed.
R/F TV X-RAY SYSTEM DIAGNOSTIC RADIOGRAPHIC FLUOROSCOPIC TV SYSTEM MXHF-1500RF SYSTEM OUTLINE Product Data No. 041021-01 MXHF-1500RF is controlled by Digital key panel console that displays KV, ma and
More informationCHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY
14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to
More informationThe effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography
Journal of Physics: Conference Series OPEN ACCESS The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography To cite this article: N A A Daud et al 2014 J. Phys.:
More informationNOT FOR DISTRIBUTION JINST_083P_0914 v1
Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam F. Di Lillo a,b, D. Dreossi c, F. Emiro a,b, C. Fedon d,e, R. Longo d,e, G. Mettivier a,b,*, L. Rigon d,e, P. Russo
More informationFeatures and Weaknesses of Phantoms for CR/DR System Testing
Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal
More informationFilm Replacement in Radiographic Weld Inspection The New ISO Standard
BAM Berlin Film Replacement in Radiographic Weld Inspection The New ISO Standard 17636-2 Uwe Ewert, Uwe Zscherpel, Mirko Jechow Requests and information to: uwez@bam.de 1 Outline - The 3 essential parameters
More informationCXDI-70C WIRELESS SPECIFICATIONS
CXDI-70C WIRELESS SPECIFICATIONS Purpose Method Sensor Scintillator Pixel Pitch Pixels Image Size A/D Grayscale Wireless Standard Preview Image Access Time High Resolution Image Display Time Cycle Time
More informationCalibration of KAP meters
Calibration of KAP meters Alexandr Malusek! Division of Radiological Sciences Department of Medical and Health Sciences Linköping University! 2014-04-15 1 Outline 1. KAP meter construction 2. Air kerma-area
More informationAUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF DIGITAL RADIOGRAPHY IMAGING SYSTEMS
International Workshop SMART MATERIALS, STRUCTURES & NDT in AEROSPACE Conference NDT in Canada 2011 2-4 November 2011, Montreal, Quebec, Canada AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF
More information