1 Practical guidelines for color calibration and quality assurance of medical displays Poster No.: C-1140 Congress: ECR 2017 Type: Educational Exhibit Authors: T. Kimpe, J. Rostang, G. Van Hoey, A. Xthona ; Kortrijk/BE, Beaverton, OR/US Keywords: Quality assurance, Image verification, Computer ApplicationsGeneral, Computer Applications-Detection, diagnosis, PACS, Computer applications DOI: /ecr2017/C-1140 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. Page 1 of 22

2 Learning objectives The purpose of this educational exhibit is to: understand and experience why color calibration of medical displays is important understand what the effect is of color calibration on appearance of medical images learn practical guidelines to correctly handle color of medical displays Page 2 of 22

3 Background The last few years color has become more important in medical imaging. Many imaging modalities generate color images and in many cases color information is critical for obtaining a correct diagnosis. Depending on the specific field of medicine, requirements for the representation of colors may vary. For instance, for the interpretation of wound photographs, color is an indicator of the healing state of the wound, correct representation of colors is important and dermatologist are demanding standardization [1]. Other medical modalities use pseudo colors to display numerical/quantitative information on top of grayscale images. The exact color used to visualize quantitative information is less of importance as long as differences are easily perceivable and it is easy for the observer to visually determine what quantitative value is being represented by a specific color. Examples of pseudo color information are fused PET/CT images, doppler ultrasound images and contrast enhanced MRI images. Page 3 of 22

4 Fig. 1: Example of a PET/CT image Image source: clinical/pet_melanoma.htm References: University of Kansas Medical Center Page 4 of 22

5 Fig. 2: Example of a Doppler ultrasound image Image source: References: Siemens Healthcare Page 5 of 22

6 Fig. 3: Example of a contrast enhanced MRI image of the prostate. Image source: References: Massachusetts General Hospital Department of Radiology Greyscale calibration (DICOM GSDF) [2] is not sufficient to guarantee consistent and qualitative color visualization. DICOM GSDF calibration ensures a consistent representation of greyscales, but leaves the color behavior of display systems unaltered. Up till today there is no clear standard that describes how color medical images need to be visualized. This lack of standardization results into large variability of color appearance and makes consistency and quality assurance a challenge. Page 6 of 22

7 Images for this section: Fig. 1: Example of a PET/CT image Image source: clinical/pet_melanoma.htm University of Kansas Medical Center Page 7 of 22

8 Fig. 2: Example of a Doppler ultrasound image Image source: Siemens Healthcare Page 8 of 22

9 Fig. 3: Example of a contrast enhanced MRI image of the prostate. Image source: Massachusetts General Hospital Department of Radiology Page 9 of 22

10 Findings and procedure details Color variability of medical displays If not properly calibrated, display systems have a large variability in color behavior [3]. Not only will there be large differences in how individual display systems visualize color, but also on one and the same display color visualization will change over time as the display system ages [4]. Fig. 4: Example of display aging effect [4]. Original (non-aged) image is at the center with the aged versions around it as follows. From top, clockwise: Chroma aged only at 10,000 hours, chroma aged only at 18,000 hours, chroma and luminance aging 10,000 hours, and chroma and luminance aging 18,000 hours. References: Healthcare Division, Barco - Kortrijk/BE Several scientific publications have confirmed that color variability of uncalibrated displays can be very large and is highly unpredictable [5; 6]. Page 10 of 22

11 Potential clinical implications Inconsistent color visualization can potentially have a large clinical impact. For applications where absolute correct color is important, failing to properly handle the color behavior of the display will simply result into wrong colors being visualized. Color errors can be very large [5]. If absolute color is being used for diagnosis such as eg. in dermatology then this could create risks of wrong diagnosis. For applications where pseudo color are being visualized, failing to properly handle the color behavior of the display can result into inconsistent color visualization and unpredictable visibility of subtle color differences. In practice this means that subtle differences in color may be visible on one display system but not on another display system. Or subtle differences in color may be visible on a particular display but not anymore visible on that same display a few months later as the display ages and changes color behavior. Needless to say that this is an unacceptable situation that could influence clinical decisions, but that also could negatively influence diagnostic confidence of radiologists. Guidelines and practical recommendations The section below describes clear guidelines on how to ensure consistent color visualization. These recommendations cover both the use cases of absolute color as well as pseudo color in medical images. Absolute color For medical images where absolute color correctness is important, it is recommended to use the ICC framework [7]. ICC profiles have been succesfully used in other applications such as prepress, and have proven capable of providing accurate absolute color reproduction. Using an ICC profile as such is not sufficient to guarantee accurate color visualization. There are several types of profiles that can be used and the profile type heavily influences the accuracy of the result. More detailed recommendations on what type of ICC profile should be used and how the ICC profiles should be generated can be found in a recent whitepaper [8] published by the ICC Medical Imaging Working Group (MIWG) [9]. Page 11 of 22

12 Fig. 6: Summary recommendations of the ICC MIWG related to use of ICC profiles for color visualization of medical images References: International Color Consortium Moreover since each display system has its own unique color behavior, it is extremely important that an ICC profile is used that was generated for the specific individual display system being used. Generic ICC profiles that describe the 'typical' behavior of a type of display do not offer the necessary accuracy and will result into still high color errors. Some medical display systems have an integrated front sensor that allows for making automatic measurements that can be used for ICC profile generation. Advanced medical display QA software packages [10] have support for generation of ICC profiles of individual displays. Page 12 of 22

13 Pseudo color For medical images that include pseudo colors it is recommended to calibrate the display system such that not only the greyscale behavior is perceptually linearized but also the color behavior of the display is made perceptually linear. This will ensure that subtle differences in color will be easier to perceive and that color images look similar even on different display systems. This is especially important in quantitative imaging applications. Extensive work has been done the last few years [11; 12; 13] on this topic and a proposal has been made to extend the DICOM GSDF standard towards color in order to achieve perceptual linearity of color, while still being compliant with the DICOM GSDF standard. This proposed extension was given the name CSDF (Color Standard Display Function) and is described in a recent scientific publication [14]. This publication also suggest an easy to use visual test pattern that can be used to quickly check whether a display system is properly calibrated or not. Fig. 7: Visual test pattern to verify CSDF calibration. Display is properly calibrated if all "CSDF" letters are equally well visible for each of the colors. References: Healthcare Division, Barco - Kortrijk/BE Page 13 of 22

14 Formal standardization of CSDF has recently been initiated as a possible task for AAPM TG196 [15]. In view of potential future standardization some medical display manufacturers have already incorporated the proposed CSDF standard, albeit under a different (commercial) name [16] into their systems and made it possible to fully automatically calibrate medical displays by means of Quality Assurance and Calibration Software. Page 14 of 22

15 Images for this section: Fig. 5: JND difference per step for different colors Healthcare Division, Barco - Kortrijk/BE Page 15 of 22

16 Fig. 7: Visual test pattern to verify CSDF calibration. Display is properly calibrated if all "CSDF" letters are equally well visible for each of the colors. Healthcare Division, Barco - Kortrijk/BE Fig. 4: Example of display aging effect [4]. Original (non-aged) image is at the center with the aged versions around it as follows. From top, clockwise: Chroma aged only at 10,000 hours, chroma aged only at 18,000 hours, chroma and luminance aging 10,000 hours, and chroma and luminance aging 18,000 hours. Healthcare Division, Barco - Kortrijk/BE Page 16 of 22

17 Fig. 6: Summary recommendations of the ICC MIWG related to use of ICC profiles for color visualization of medical images International Color Consortium Page 17 of 22

18 Conclusion Color has become more and more important in medical imaging. Many imaging modalities generate color images and in many cases color information is critical for coming to a correct diagnosis. Greyscale calibration (DICOM GSDF) is not sufficient to guarantee consistent and qualitative color visualization. A standard that describes how color medical images need to be visualized does not yet exist. Recently a proposal has been made for a "Color Standard Display Function" (CSDF), an extension of DICOM GSDF towards color. This educational exhibit provided clear guidelines and recommendations on how to handle both absolute correct color as well as pseudo color medical images. Following these guidelines makes it possible to achieve accurate and consistent color behavior in medical images. Page 18 of 22

19 Personal information Disclosure: Tom Kimpe is employee of Barco Healthcare, a company developing and selling medical display systems Tom Kimpe received his masters in computer engineering from University of Ghent, Belgium in He also obtained a PhD from the same university on the topic of image quality of medical displays. In 2010 he finalized a Masters in Business Administration at the Vlerick Management School. Since 2001 Tom has been working in Barco's Healthcare Division. He has taken the positions of development engineer, project manager, innovation manager, VP of Technology and Innovation and Chief Technology Officer of Barco's Healthcare Division. Already for several years, he is coordinating all research and innovation activities of Barco's Healthcare Division and is steering an international team of researchers in USA, Belgium and Italy. The last few years Tom has been active in various standardization committees such as ICC MIWG and AAPM TG196 where he is working towards standardization of color in medical imaging. Contact information: tom.kimpe@barco.com Page 19 of 22

20 Images for this section: Fig. 8: Dr. Tom Kimpe, Barco Healthcare Healthcare Division, Barco - Kortrijk/BE Page 20 of 22

21 References [1] A. Marghoob for the International Skin Imaging Collaboration Melanoma Project Working Groups, "Standards in Dermatologic Imaging," JAMA Dermatology, vol. 151, no. 18, pp , [2] NEMA, Digital imaging and communications in medicine (DICOM), part 14: Grayscale Standard Display Function, vol. PS 3.14, National Electrical Manufacturers Association, [3] ucm Last accessed December 16th 2016 [4] Avanaki, Ali RN, et al. "Aging display's effect on interpretation of digital pathology slide." SPIE Medical Imaging. International Society for Optics and Photonics, [5] Badano, Aldo, et al. "Consistency and standardization of color in medical imaging: a consensus report." Journal of digital imaging 28.1 (2015): [6] Krupinski EA, Silverstein LD, Hashmi SF, Graham AR, Weinstein RS, Roehrig H: Observer performance using virtual pathology slides: impact of LCD color reproduction accuracy. Journal of Digital Imaging 25(6): , 2012 [7] P. Green, Color Management : Understanding and Using ICC Profiles - Phil Green, Michael Kriss, P. Green and M. Kriss, Eds., Wiley, 2010, p [8] ICC, "Recommendations for visualization of medical content on color display systems", ICC_White_Paper44_visualization 2016, _of_medical_color_content.pdf [9] [10] QAWeb - Barco healthcare, Last accessed December 16th 2016 Page 21 of 22

22 [11] L. Silverstein, S. Hashmi, K. Lang and E. Krupinski, "Paradigm for achieving color-reproduction accuracy in LCDs for medical imaging," Journal of the Society for Information Display, vol. 20, pp , January [12] Avanaki, Ali, et al. "Perceptual uniformity of commonly used color spaces." SPIE medical imaging. International Society for Optics and Photonics, [13] Kimpe, Tom, et al. "Does the choice of display system influence perception and visibility of clinically relevant features in digital pathology images?." SPIE Medical Imaging. International Society for Optics and Photonics, [14] Kimpe, Tom, et al. "Color standard display function: A proposed extension of DICOM GSDF." Medical Physics 43.9 (2016): doi/ / /full. Last accessed December 16th [15] Last accessed December 16th [16] Barco Healthcare, SteadyColor calibration algorithm, az vo.msecnd.net/~/media/downloads/technology%20highlights/u/ 14235BARCOtechnology%20highlightSteadyColorHR%20pdf.PDF?la=en&v=2. Last accessed December 16th 2016 Page 22 of 22

Computer applications, PACS, Instrumentation, Equipment, Technology assessment /ecr2012/C-0655

Computer applications, PACS, Instrumentation, Equipment, Technology assessment /ecr2012/C-0655 The novel medical diagnostic color monitor automatically distinguishing monochrome/color images and displaying these with their own optimal tone curves Poster No.: C-0655 Congress: ECR 2012 Type: Authors:

More information

A comparison between medical-grade liquid crystal display (LCD) and ipad color imaging

A comparison between medical-grade liquid crystal display (LCD) and ipad color imaging A comparison between medical-grade liquid crystal display (LCD) and ipad color imaging Poster No.: C-1377 Congress: ECR 2014 Type: Scientific Exhibit Authors: K. Yoshimura 1, K. Shibata 2, T. Nihashi 1,

More information

Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging

Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging Poster No.: C-0709 Congress: ECR 2014 Type: Scientific

More information

Evaluation of no-grid radiography using the digital scattered x-ray removal processing

Evaluation of no-grid radiography using the digital scattered x-ray removal processing Evaluation of no-grid radiography using the digital scattered x-ray removal processing Poster No.: C-0416 Congress: ECR 2016 Type: Authors: Scientific Exhibit R. Suzuki 1, T. Goto 1, H. Ogawa 2, N. Amimoto

More information

Improvement of CT image quality with iterative reconstruction idose4

Improvement of CT image quality with iterative reconstruction idose4 Improvement of CT image quality with iterative reconstruction idose4 Poster No.: C-0387 Congress: ECR 2014 Type: Scientific Exhibit Authors: M.-L. Olsson, K. Norrgren, M. Söderberg; Malmö/SE Keywords:

More information

A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers

A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers Poster No.: C-3011 Congress: ECR 2010 Type: Topic: Authors: Scientific Exhibit

More information

Determining acceptance levels for automatic daily image quality control in magnetic resonance imaging

Determining acceptance levels for automatic daily image quality control in magnetic resonance imaging Determining acceptance levels for automatic daily image quality control in magnetic resonance imaging Poster No.: C-1125 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific Exhibit J. I. Peltonen,

More information

Correction of the local intensity nonuniformity artifact in high field MRI

Correction of the local intensity nonuniformity artifact in high field MRI Correction of the local intensity nonuniformity artifact in high field MRI Poster No.: C-0346 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Paper S. Kai, S. Kumazawa, H. Yabuuchi, F. Toyofuku;

More information

Investigation of Effective DQE (edqe) parameters for a flat panel detector

Investigation of Effective DQE (edqe) parameters for a flat panel detector Investigation of Effective DQE (edqe) parameters for a flat panel detector Poster No.: C-1892 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit D. Bor 1, S. Cubukcu 1, A. Yalcin 1, O.

More information

Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging

Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging Award: Magna Cum Laude Poster No.: C-1988 Congress: ECR 2014 Type: Scientific Exhibit Authors: T. Hamaguchi,

More information

Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE)

Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE) Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE) Poster No.: C-2459 Congress: ECR 2010 Type: Scientific Exhibit Topic: Neuro Authors: T. Yoneda,

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

ABSTRACT 1. PURPOSE 2. METHODS

ABSTRACT 1. PURPOSE 2. METHODS Perceptual uniformity of commonly used color spaces Ali Avanaki a, Kathryn Espig a, Tom Kimpe b, Albert Xthona a, Cédric Marchessoux b, Johan Rostang b, Bastian Piepers b a Barco Healthcare, Beaverton,

More information

Difference in signal-to-noise-ratio (SNR) from vertical to horizontal scanner position using a 0,25 Tesla Weightbearing

Difference in signal-to-noise-ratio (SNR) from vertical to horizontal scanner position using a 0,25 Tesla Weightbearing Difference in signal-to-noise-ratio (SNR) from vertical to horizontal scanner position using a 0,25 Tesla Weightbearing scanner Poster No.: C-0672 Congress: ECR 2014 Type: Scientific Exhibit Authors: F.

More information

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Poster No.: C-2205 Congress: ECR 2012 Type: Scientific Paper

More information

Multi-slice computed tomography analysis of bullet trajectory in forensic investigation.

Multi-slice computed tomography analysis of bullet trajectory in forensic investigation. Multi-slice computed tomography analysis of bullet trajectory in forensic investigation. Poster No.: C-0510 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Usui, Y. Kawasumi, Y. Hosokai, A. Nakajima,

More information

The impact of increasing SID on patient dose in pa abdomen imaging

The impact of increasing SID on patient dose in pa abdomen imaging The impact of increasing SID on patient dose in pa abdomen imaging Poster No.: C-1074 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific Exhibit T. Starc 1, N. Mekis 2 ; 1 Brezovica/SI, 2 Ljubljana/SI

More information

Lumbar disc height assessment: a comparative study between EOS and digital radiography

Lumbar disc height assessment: a comparative study between EOS and digital radiography Lumbar disc height assessment: a comparative study between EOS and digital radiography Poster No.: C-1178 Congress: ECR 2012 Type: Scientific Paper Authors: V. Freire, M. Benhamou, F. Rannou, S. Poiraudeau,

More information

Worldwide practice of breast MRI: insights from the MIPA study applications - the MIPA study group

Worldwide practice of breast MRI: insights from the MIPA study applications - the MIPA study group Worldwide practice of breast MRI: insights from the MIPA study applications - the MIPA study group Poster No.: B-0757 Congress: ECR 2015 Type: Scientific Paper Authors: G. Di Leo, R. M. Trimboli, I. Ioan,

More information

Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis

Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis Poster No.: C-0585 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Minehiro,

More information

Technical Paper CONSISTENT PRESENTATION OF MEDICAL IMAGES

Technical Paper CONSISTENT PRESENTATION OF MEDICAL IMAGES Technical Paper CONSISTENT PRESENTATION OF MEDICAL IMAGES A REPORT ON MEDICAL WORKSTATION CALIBRATION By Tom Schulte In years past, the quality of medical images was strictly controlled during the exposure,

More information

Aim. Images for this section: Page 2 of 13

Aim. Images for this section: Page 2 of 13 Changes in CT number of high atomic number materials with field of view when using an extended CT number to electron density curve and a metal artifact reduction reconstruction algorithm Poster No.: R-0094

More information

Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology

Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology Poster No.: C-1190 Congress: ECR 2011 Type: Scientific Exhibit Authors: K. Chida, T. Takahashi, D.

More information

PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution

PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution Poster No.: C-1280 Congress: ECR 2011 Type: Educational Exhibit Authors: A. Monteiro, A.

More information

PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution

PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution Poster No.: C-1280 Congress: ECR 2011 Type: Educational Exhibit Authors: A. Monteiro, A.

More information

The role of CT in restoration of a very rare Christ wooden statue: a fundamental contribution.

The role of CT in restoration of a very rare Christ wooden statue: a fundamental contribution. The role of CT in restoration of a very rare Christ wooden statue: a fundamental contribution. Poster No.: C-1514 Congress: ECR 2015 Type: Scientific Exhibit Authors: M. Disaro', P. Sartori; Venice/IT

More information

Novel cassette-sized, flat-panel digital radiography (DR) system: Initial clinical and workflow results versus computed radiography (CR)

Novel cassette-sized, flat-panel digital radiography (DR) system: Initial clinical and workflow results versus computed radiography (CR) Novel cassette-sized, flat-panel digital radiography (DR) system: Initial clinical and workflow results versus computed radiography (CR) Poster No.: C-3027 Congress: ECR 2010 Type: Scientific Exhibit Topic:

More information

Development of new dosimeter for measuring dose distribution in CT

Development of new dosimeter for measuring dose distribution in CT Development of new dosimeter for measuring dose distribution in CT Poster No.: C-2925 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology - Without Subtopic Authors: Y. Muramatsu, K.

More information

Comparison of high-resolution C-arm cone-beam CT imaging and conventional 3D rotational angiography for the better microvascular visualization

Comparison of high-resolution C-arm cone-beam CT imaging and conventional 3D rotational angiography for the better microvascular visualization Comparison of high-resolution C-arm cone-beam CT imaging and conventional 3D rotational angiography for the better microvascular visualization Poster No.: C-1040 Congress: ECR 2016 Type: Scientific Exhibit

More information

The feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE

The feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE The feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE Poster No.: C-0022 Congress: ECR 2010 Type: Scientific Exhibit Topic: Abdominal Viscera

More information

Voice recognition versus transcriptionist

Voice recognition versus transcriptionist Voice recognition versus transcriptionist Poster No.: C-0263 Congress: ECR 2010 Type: Audit/Professional Issues Topic: Audit/Professional Issues Authors: R. H. Strahan; Clayton South VIC/AU Keywords: Voice

More information

Suppression of metal artifacts using image-based monoenergetic DECT imaging

Suppression of metal artifacts using image-based monoenergetic DECT imaging Suppression of metal artifacts using image-based monoenergetic DECT imaging Poster No.: C-0519 Congress: ECR 2011 Type: Scientific Paper Authors: B. Krauss, B. Schmidt, M. Sedlmair, T. Flohr; Forchheim/DE

More information

SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES.

SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES. SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES. Tingberg, Anders Published in: Radiation Protection Dosimetry DOI: 10.1093/rpd/ncs302 Published: 2013-01-01 Link to publication Citation for published

More information

White Paper. The Advantages of Hybrid Gamma PXL. 1 Introduction Outline of Hybrid Gamma PXL Study Results...

White Paper. The Advantages of Hybrid Gamma PXL. 1 Introduction Outline of Hybrid Gamma PXL Study Results... White Paper The Advantages of Hybrid Gamma PXL CONTENTS 1 Introduction... 2 2 Outline of Hybrid Gamma PXL... 4 3 Study... 5 4 Results... 7 5 Conclusions... 10 No.17-001 Revision A November 2017 Product

More information

Image Display and Perception

Image Display and Perception Image Display and Perception J. Anthony Seibert, Ph.D. Department of Radiology UC Davis Medical Center Sacramento, California, USA Image acquisition, display, & interpretation X-rays kvp mas Tube filtration

More information

Medical Imaging Display Color Space (mrgb) Teleconference 19 June :00 (EST)

Medical Imaging Display Color Space (mrgb) Teleconference 19 June :00 (EST) Medical Imaging Display Color Space (mrgb) Teleconference 19 June 2014 09:00 (EST) The meeting was called to order at 09:00 am (EST) by Craig Revie, acting chair, with the following attendees: Rich Amador

More information

Increasing the Number of Gray Shades in Medical Display Systems How Much is Enough?

Increasing the Number of Gray Shades in Medical Display Systems How Much is Enough? Increasing the Number of Gray Shades in Medical Display Systems How Much is Enough? Tom Kimpe and Tom Tuytschaever Medical images produced by x-ray detectors, computed tomography (CT) scanners, and other

More information

Barco medical display systems. Product catalog

Barco medical display systems. Product catalog Barco medical display systems Product catalog TECHNOLOGy THAT MAkES A DIFFERENCE Backlight Output Stabilization (BLOS) Backlight Output Stabilization (BLOS) is an image stabilization system driven by a

More information

SIMPLE. VERSATILE. VERSATILE PRINTING ON FILM OR PAPER. DRYVIEW CHROMA Imaging System

SIMPLE. VERSATILE. VERSATILE PRINTING ON FILM OR PAPER. DRYVIEW CHROMA Imaging System DRYVIEW CHROMA Imaging System VERSATILE PRINTING ON FILM OR PAPER One of the most trusted names in medical printing now offers a highly versatile printer that empowers physicians and other healthcare professionals

More information

Things you may want to know

Things you may want to know Things you may want to know 1 / 9 Last updated 18-Nov-05 Why medical monitors are special? Unlike commercial displays, Medical displays are used in the hospital and they are displaying contain life critical

More information

A Module for Visualisation and Analysis of Digital Images in DICOM File Format

A Module for Visualisation and Analysis of Digital Images in DICOM File Format A Module for Visualisation and Analysis of Digital Images in DICOM File Format Rumen Rusev Abstract: This paper deals with design and realisation of software module for visualisation and analysis of digital

More information

A Novel 12-bit Grayscale Topology PACSmate Revolutionary GrayBoost Technology

A Novel 12-bit Grayscale Topology PACSmate Revolutionary GrayBoost Technology A Novel 12-bit Grayscale Topology PACSmate Revolutionary GrayBoost Technology Author: Kenny Jan Date: August 3, 2009 Presented by PACSmate Technology Inc. Table of Contents 1. Introduction... 2 2. GrayBoost

More information

Optimal speed for dual-energy subtraction neck imaging with flat-panel detector radiography

Optimal speed for dual-energy subtraction neck imaging with flat-panel detector radiography Optimal speed for dual-energy subtraction neck imaging with flat-panel detector radiography Poster No.: C-2938 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology Authors: H. Machida

More information

Clinical and management aspects of digital imaging and PACS

Clinical and management aspects of digital imaging and PACS Clinical and management aspects of digital imaging and PACS พญ.จามร เช อเพชระโสภณ นายกร งส ว ทยาสมาคมแห งประเทศไทย chamareec@gmail.com www.radiologythailand.org Digital imaging Abbreviation and terminology

More information

Calibration Slide for Histopathology task force Teleconference 20 February :00 (UK) / 10:00 (EST)

Calibration Slide for Histopathology task force Teleconference 20 February :00 (UK) / 10:00 (EST) Calibration Slide for Histopathology task force Teleconference 20 February 2014 15:00 (UK) / 10:00 (EST) The meeting was called to order at 10:00 am (EST) by Craig Revie, chair of MIWG, with the following

More information

Yagi Digital Microscope Calibration

Yagi Digital Microscope Calibration Yagi Digital Microscope Calibration Method summary, assessment and suggestions for improvement W Craig Revie, International Color Consortium Introduction In the area of pathology, a type of digital microscope

More information

Incorporating novel image processing methods in a hospital-wide PACS

Incorporating novel image processing methods in a hospital-wide PACS International Congress Series 1281 (2005) 1016 1021 www.ics-elsevier.com Incorporating novel image processing methods in a hospital-wide PACS Erwin Bellon a, T, Michel Feron a, Paul Neyens a, Klaas Peeters

More information

Imaging Informatics. Standards Support for Color in Medical Imaging. SAM Imaging Education Course. Photometric Standards Colorimetric Standards

Imaging Informatics. Standards Support for Color in Medical Imaging. SAM Imaging Education Course. Photometric Standards Colorimetric Standards Imaging Informatics SAM Imaging Education Course AAPM Standards Support for Color in Medical Imaging Photometric Standards Colorimetric Standards Michael J. Flynn, PhD Henry Ford Health System, Detroit,

More information

DICOM Conformance. DICOM Detailed Specification for Diagnostic Labs and Radiology Center Connectivity

DICOM Conformance. DICOM Detailed Specification for Diagnostic Labs and Radiology Center Connectivity DICOM Detailed Specification for Diagnostic Labs and Radiology Center Connectivity Authored by Global Engineering Team, Health Gorilla April 10, 2014 Table of Contents About Health Gorilla s Online Healthcare

More information

Medical Imaging Working Group

Medical Imaging Working Group Medical Imaging Working Group FOGRA Einsteinring 1a 85609 Ascheim Munich, Germany 26 February 2018 Craig Revie, MIWG chair, opened the meeting at 08:45 and introduced the agenda as follows: 1. Introductions

More information

Digital Image Management: the Basics

Digital Image Management: the Basics Digital Image Management: the Basics Napapong Pongnapang, Ph.D. Department of Radiological Technology Faculty of Medical Technology Mahidol University Outline From screen/film to digital radiography PACS/Tele

More information

Ring-Down versus Comet Tail: Two artifacts uncovered

Ring-Down versus Comet Tail: Two artifacts uncovered Ring-Down versus Comet Tail: Two artifacts uncovered Poster No.: C-2288 Congress: ECR 2013 Type: Educational Exhibit Authors: S. Méndez Garrido 1, J. Arce Duran 2, A. Talegón Meléndez 1 ; 1 2 Sevilla/ES,

More information

Scalable Data Storage Analysis and Solution for Picture Archiving and Communication Systems (PACS)

Scalable Data Storage Analysis and Solution for Picture Archiving and Communication Systems (PACS) Scalable Data Storage Analysis and Solution for Picture Archiving and Communication Systems (PACS) Ayhan Akbal and Erhan Akbal II. PIURE ARCHIVING AND COMMUNICATION SYSTEM Abstract Information technologies

More information

Black Science Fiction Script Movie Project 2012

Black Science Fiction Script Movie Project 2012 Black Science Fiction Script Movie Project 2012 OFFICIAL 2011 ENTRY FORM AND RELEASE Please make sure to fill in ALL fields to ensure that your entry is accepted! Please attach a separate completed form

More information

MEDICAL DISPLAYS. NEC 90 Series. NEC MD Series

MEDICAL DISPLAYS. NEC 90 Series. NEC MD Series MEDICAL DISPLAYS NEC MD Series Are you really sure your display is as accurate as your diagnosis? In medical environments such as radiology and emergency rooms, the right equipment and technology can help

More information

Evaluating clinical performance of the new Coronis Fusion 6MP DL display

Evaluating clinical performance of the new Coronis Fusion 6MP DL display White Paper Evaluating clinical performance of the new Coronis Fusion 6MP DL display What s inside? Clinical performance of Coronis Fusion 6MP DL for lung nodule screening ROC curves for Coronis Fusion

More information

Outline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control

Outline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control Understanding Digital Modalities: Image Quality and Dose S. Jeff Shepard, M.S. University of Texas M. D. Anderson Cancer Center Houston, Texas Special Acknowledgement: Stephen K. Thompson, M.S. William

More information

Image quality evaluation of turbo-spin echo diffusion weighted image (TSE-DWI) : A phantom study

Image quality evaluation of turbo-spin echo diffusion weighted image (TSE-DWI) : A phantom study Image quality evaluation of turbo-spin echo diffusion weighted image (TSE-DWI) : A phantom study Poster No.: C-0631 Congress: ECR 2016 Type: Scientific Exhibit Authors: T. Yoshida, A. Urikura, K. Shirata,

More information

Assessment of Color Displays. Michael Flynn Radiology Research Henry Ford Health System Detroit, MI

Assessment of Color Displays. Michael Flynn Radiology Research Henry Ford Health System Detroit, MI Assessment of Color Displays Learning Objectives 1. Human color vision 2. CIE color spaces 3. ICC color management 4. Color measurements 5. Medical standards for color Michael Flynn Radiology Research

More information

IHE Radiology Technical Framework Supplement. Stereotactic Mammography Image (SMI) Trial Implementation

IHE Radiology Technical Framework Supplement. Stereotactic Mammography Image (SMI) Trial Implementation Integrating the Healthcare Enterprise 5 IHE Radiology Technical Framework Supplement 10 Stereotactic Mammography Image (SMI) 15 Trial Implementation 20 25 Date: June 11, 2013 Author: IHE Radiology Technical

More information

ICC Medical Imaging Working Group Webinar August 1, 2013

ICC Medical Imaging Working Group Webinar August 1, 2013 ICC Medical Imaging Working Group Webinar August 1, 2013 The meeting was called to order at 10:00 am (EDT) by Craig Revie, acting chair with the following in attendance: Aldo Badano Vipul Baxi James Chang

More information

Review Paper on. Quantitative Image Quality Assessment Medical Ultrasound Images

Review Paper on. Quantitative Image Quality Assessment Medical Ultrasound Images Review Paper on Quantitative Image Quality Assessment Medical Ultrasound Images Kashyap Swathi Rangaraju, R V College of Engineering, Bangalore, Dr. Kishor Kumar, GE Healthcare, Bangalore C H Renumadhavi

More information

Infrared Screening. with TotalVision anatomy software

Infrared Screening. with TotalVision anatomy software Infrared Screening with TotalVision anatomy software Unlimited possibilities with our high-quality infrared screening systems Energetic Health Systems leads the fi eld in infrared screening and is the

More information

Chin J Radiol 2003; 28:

Chin J Radiol 2003; 28: Chin J Radiol 2003; 28: 175-180 175 modality console Image acquisition Quality assurance ISO Quality control PACS Picture Archiving and Communication System - Communication in Medicine [1] service WorkList

More information

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC? Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM

More information

TI Designs: Biometric Steering Wheel. Amy Ball TIDA-00292

TI Designs: Biometric Steering Wheel. Amy Ball TIDA-00292 www.ti.com 2 Biometric Steering Wheel - -Revised July 2014 www.ti.com TI Designs: Biometric Steering Wheel - -Revised July 2014 Biometric Steering Wheel 3 www.ti.com 4 Biometric Steering Wheel - -Revised

More information

High Intensity Red Low Current Seven Segment Display

High Intensity Red Low Current Seven Segment Display High Intensity Red Low Current Seven Segment Display DESCRIPTION 19237 This series defines a new standard for low current displays. It is a single digit 7-segment LED display utilizing AllnGaP technology

More information

ORDERING A PHOTOGRAPH FOR PERSONAL USE

ORDERING A PHOTOGRAPH FOR PERSONAL USE ORDERING A PHOTOGRAPH FOR PERSONAL USE Thank you for your interest in ordering an item for personal use from The Arizona Republic or azcentral.com. To order a photograph, full-page print, graphic, or Benson

More information

Dual N-Channel 30 V (D-S) MOSFET with Schottky Diode

Dual N-Channel 30 V (D-S) MOSFET with Schottky Diode SPICE Device Model SiR77DP Dual N-Channel 3 V (D-S) MOSFET with Schottky Diode DESCRIPTION The attached SPICE model describes the typical electrical characteristics of the n-channel vertical DMOS. The

More information

Single Phase Rectifier Bridge, 2 A

Single Phase Rectifier Bridge, 2 A Single Phase Rectifier Bridge, 2 A 2KBP Series FEATURES Suitable for printed circuit board mounting Compact construction RoHS COMPLIANT D-44 PRODUCT SUMMARY I O V RRM 2 A 50 to 1000 V High surge current

More information

Nathan Childress, Ph.D., DABR

Nathan Childress, Ph.D., DABR Nathan Childress, Ph.D., DABR Introduction TG-142 is a comprehensive QA protocol Covers nearly every aspect of machine and safety QA Recommends quantitative results Recommends high testing frequencies

More information

Medical Imaging Working Group

Medical Imaging Working Group Medical Imaging Working Group BenQ HQ No. 16 Ji-Hu Road, Nei-Hu District Taipei, Taiwan 4 May 2016 Craig Revie, MIWG chair, opened the meeting at 14:00 and after self-introductions and a sound check for

More information

Preliminary Assessment of High Dynamic Range Displays for Pathology Detection Tasks. CIS/Kodak New Collaborative Proposal

Preliminary Assessment of High Dynamic Range Displays for Pathology Detection Tasks. CIS/Kodak New Collaborative Proposal Preliminary Assessment of High Dynamic Range Displays for Pathology Detection Tasks CIS/Kodak New Collaborative Proposal CO-PI: Karl G. Baum, Center for Imaging Science, Post Doctoral Researcher CO-PI:

More information

Improving Color Image Quality in Medical Photography

Improving Color Image Quality in Medical Photography White Paper #46 Level: Intermediate Date: April 2017 Improving Color Image Quality in Medical Photography Introduction Medical images have played an important role in the development of modern medicine.

More information

Evaluation of magnetic resonance acoustic noise in 1.5 and 3 Tesla scanners

Evaluation of magnetic resonance acoustic noise in 1.5 and 3 Tesla scanners Evaluation of magnetic resonance acoustic noise in 1.5 and 3 Tesla scanners Poster No.: B-1014 Congress: ECR 2016 Type: Scientific Paper Authors: V. M. F. Silva, I. M. Ramos, J. Moreira, M. Marques; Porto/PT

More information

Single Phase Rectifier Bridge, 2 A

Single Phase Rectifier Bridge, 2 A Single Phase Rectifier Bridge, 2 A 2KBP Series FEATURES Suitable for printed circuit board mounting Compact construction RoHS COMPLIANT D-44 PRODUCT SUMMARY I O V RRM 2 A 50 to 1000 V High surge current

More information

DR _ solutions. We understand that customers don t need just products, they want. solutions

DR _ solutions. We understand that customers don t need just products, they want. solutions DR _ solutions We understand that customers don t need just products, they want solutions index company profile 1974-2005 2006-2007 - 2008 ITALRAY Srl was founded in 1974 as the production branch of Marzocchi

More information

HP DREAMCOLOR USE CASES

HP DREAMCOLOR USE CASES HP DREAMCOLOR USE CASES ACHIEVING ACCURATE COLORS ACROSS DIVERSE WORKFLOWS TECHNOLOGy SPOTLIGHT HP PROFESSIONAL INNOVATIONS FOR business NOTEbOOKS WHEN COLOR IS EVERyTHING For creative professionals, consistent

More information

Ambient Light Sensor

Ambient Light Sensor Ambient Light Sensor DESCRIPTION 2118 ambient light sensor is a silicon NPN epitaxial planar phototransistor in a T-1¾ package. It is sensitive to visible light much like the human eye and has peak sensitivity

More information

Medical Imaging Displays: Psychophysics and Quality Assurance: Psychophysics and the Human Visual System

Medical Imaging Displays: Psychophysics and Quality Assurance: Psychophysics and the Human Visual System Medical Imaging Displays: Psychophysics and Quality Assurance: Psychophysics and the Human Visual System Rationale Display options present significant challenge - which use, under what circumstances, how

More information

IR Sensor Module for Remote Control Systems

IR Sensor Module for Remote Control Systems IR Sensor Module for Remote Control Systems MECHANICAL DATA Pinning: 1 = Carrier OUT, 2 = GND, 3 = V S 19026 FEATURES Photo detector and preamplifier in one package AC coupled response from 20 khz to 60

More information

High Intensity LED, Ø 5 mm Untinted Non-Diffused

High Intensity LED, Ø 5 mm Untinted Non-Diffused TLHE5., TLHG5., TLHK5. High Intensity LED, Ø 5 mm Untinted Non-Diffused FEATURES Untinted non diffused lens Choice of four colors TLH.5 for cost effective design Medium viewing angle Compliant to RoHS

More information

Little Giant Comics OLD SCHOOL COMIC SHOW

Little Giant Comics OLD SCHOOL COMIC SHOW Little Giant Comics OLD SCHOOL COMIC SHOW Saturday 4/28/18 at the Everett Arena, 15 Loudon Rd. Concord NH Show Hours: Saturday 10am 5pm, SET UP Friday 12pm 8pm, Sat 8am-10am Vendor Info: PLEASE PRINT CLEARLY

More information

Reflective Optical Sensor with Transistor Output

Reflective Optical Sensor with Transistor Output TCRT0, TCRT1010 Reflective Optical Sensor with Transistor Output 21836 TCRT0 A C E C TCRT1010 19155_1 FEATURES Package type: leaded Detector type: phototransistor Dimensions (L x W x H in mm): 7 x 4 x

More information

ASSIGNMENT OF RIGHTS

ASSIGNMENT OF RIGHTS ASSIGNMENT OF RIGHTS BETWEEN: Fondation Louis Vuitton, a corporate foundation whose incorporation was authorised by order of the Prefect of Paris, published in the French Official Journal on November 18

More information

FFIFF 2018 Shoot Your Short SCREENPLAY COMPETITION RULES

FFIFF 2018 Shoot Your Short SCREENPLAY COMPETITION RULES FFIFF 2018 Shoot Your Short SCREENPLAY COMPETITION RULES General Eligibility All writers of submitted material, including co-authors, must be 18 years or older. Screenplays written by teams of two or more

More information

Silicon PIN Photodiode

Silicon PIN Photodiode Silicon PIN Photodiode DESCRIPTION is a silicon PIN photodiode with high radiant sensitivity in clear, T-1 plastic package. It is sensitive to visible and near infrared radiation. FEATURES Package type:

More information

Reflective Optical Sensor with Transistor Output

Reflective Optical Sensor with Transistor Output Reflective Optical Sensor with Transistor Output CNY7 2835 DESCRIPTION The CNY7 is a reflective sensor that includes an infrared emitter and phototransistor in a leaded package which blocks visible light.

More information

UP-D74XRD. DICOM Digital Imager.

UP-D74XRD. DICOM Digital Imager. DICOM Digital Imager www.sonybiz.net/healthcare Supports both 8x10inch Blue Thermal Film and Thermal Printing Paper. Delivers photo-quality fi lm and paper prints with 512-step grayscale. Ensures minimal

More information

Best and Worst Practices and Software Development Tools DICOM

Best and Worst Practices and Software Development Tools DICOM DICOMWeb 2015 Conference & Hands-on Workshop Best and Worst Practices and Software Development Tools DICOM David Clunie (dclunie@dclunie.com) PixelMed Publishing Background & Disclosures l Owner, PixelMed

More information

IR Sensor Module for Remote Control Systems

IR Sensor Module for Remote Control Systems IR Sensor Module for Remote Control Systems MECHANICAL DATA Pinning: 1 = Carrier OUT, 2 = GND, 3 = V S 19026 FEATURES Photo detector and preamplifier in one package AC coupled response from 30 khz to 50

More information

3D and Sequential Representations of Spatial Relationships among Photos

3D and Sequential Representations of Spatial Relationships among Photos 3D and Sequential Representations of Spatial Relationships among Photos Mahoro Anabuki Canon Development Americas, Inc. E15-349, 20 Ames Street Cambridge, MA 02139 USA mahoro@media.mit.edu Hiroshi Ishii

More information

Model-Based Design Toolbox

Model-Based Design Toolbox Model-Based Design Toolbox License Installation & Management Manual An Embedded Target for S32K1xx Family of Processors Version 3.0.0 Target Based Automatic Code Generation Tools For MATLAB /Simulink /Stateflow

More information

The Quantitative Aspects of Color Rendering for Memory Colors

The Quantitative Aspects of Color Rendering for Memory Colors The Quantitative Aspects of Color Rendering for Memory Colors Karin Töpfer and Robert Cookingham Eastman Kodak Company Rochester, New York Abstract Color reproduction is a major contributor to the overall

More information

Instant DR in Jordan

Instant DR in Jordan Hashemite University leads the way with first Instant DR in Jordan DR Retrofit supports research and education goals of the Faculty of Allied Health Sciences, while enhancing care for staff and students

More information

Introduction to Grayscale Calibration and Related Aspects of Medical Imaging Grade Liquid Crystal Displays

Introduction to Grayscale Calibration and Related Aspects of Medical Imaging Grade Liquid Crystal Displays Introduction to Grayscale Calibration and Related Aspects of Medical Imaging Grade Liquid Crystal Displays Kenneth A. Fetterly, 1 Hartwig R. Blume, 2 Michael J. Flynn, 3 and Ehsan Samei 4,5,6 Consistent

More information

imagespectrum ADVANCED DIGITAL IMAGE MANAGEMENT SYSTEM Get a Better Handle on the Big Picture

imagespectrum ADVANCED DIGITAL IMAGE MANAGEMENT SYSTEM Get a Better Handle on the Big Picture ADVANCED DIGITAL IMAGE MANAGEMENT SYSTEM Get a Better Handle on the Big Picture SECURELY STREAMLINE YOUR PRACTICE WORKFLOW imagespectrum enables eye care practices, clinics, and even entire hospital departments

More information

IR Sensor Module for Remote Control Systems

IR Sensor Module for Remote Control Systems TSOP98238 IR Sensor Module for Remote MECHANICAL DATA Pinning: 1 = Carrier OUT, 2 = GND, 3 = V S 19026 FEATURES Photo detector and preamplifier in one package AC coupled response from 30 khz to 50 khz,

More information

Ten Turns Servo or Bushing Mount Hall Effect Sensor in Size 09 (22.2 mm)

Ten Turns Servo or Bushing Mount Hall Effect Sensor in Size 09 (22.2 mm) Ten Turns Servo or Bushing Mount Hall Effect Sensor in Size 09 (22.2 mm) ELECTRICAL SPECIFICATIONS FEATURES All electrical angles available up to: 3600 Accurate linearity down to: ± 0.5 % Very long life:

More information