Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review

Size: px
Start display at page:

Download "Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review"

Transcription

1 MRES Fast Imaging Techniques Module 2 Artefacts and Imaging Optimisation for single shot methods Content: Introduction Phase error Phase bandwidth Chemical shift review Chemical shift in pixels Removing chemical shift artefacts Magnetic susceptibility Reducing geometric distortions Receiver bandwith Ramp sampling Fourier regridding Field of view Slew Rate Frequency Encoding Steps 1/36

2 MRES Fast Imaging Techniques Phase encoding steps Segmentation Advantages and Disadvantages of Segmentation Repetition Time (TR) Nyquist or N2 Ghosting (I) Nyquist or N2 Ghosting (II) Nyquist or N2 Ghosting (III) Reduce Nyquist ghosting Recovering SNR and Resolution Summary 2/36

3 Introduction In Module One, the basic principles FSE and EPI were discussed. When used as single shot methods, a long train of echoes are acquired for a single excitation pulse. As gradients are faster to apply than 180 degree rf pulses, EPI is the faster of the two methods, and has a train of gradient refocussed echoes The oscillating frequency gradient acquires every alternate line in reverse. As a result of the gradient echo train in the EPI readout, the sequence is very susceptible to off resonance effects. Figure 2.1: The traversal of k-space used in EPI. In this module, we shall be discussing the sensitivity of single shot methods to artefacts, particularly EPI. The removal or reduction of these artefacts depends on careful parameter selection and hardware requirements. 3/36

4 Phase error As already mentioned in Module 1, the longer the echo train length, or the longer the echo spacing, the more signal decay there is between echoes due to T2 in FSE or T2* in EPI. This can result in blurring or lower apparent spatial resolution, due to a filtering of k-space or raw data, called the point spread function. There are added complications when off resonance effects are considered. The magnetic field experienced by an individual nucleus can be altered by off resonance effects, namely: chemical shift magnetic susceptibility. These effects are covered in depth in course MRES7001, Module Seven. The result of these effects is an increase signal decay rate, given by T2*. In all Fourier transform imaging, the lines of k-space are filled using phase encoding gradients to increment the position of k-space; as described in MRES7004. If two components have different frequencies due to chemical shift or susceptibility then they will accumulate additional phase and this will translate to different spatial offsets. In SE, each line of k-space is a separate acquisition, so the phase difference is 'reset' to zero at the beginning of each acquisition. In FSE, the 180o rf pulses within the echo train correct the phase error that builds up. That is, signal decay due to chemical shift and magnetic susceptibility effects is refocused. In EPI, with its gradient echo train, such effects accumulate. The time between adjacent points on the ky axis can be quite long and there is no intervening rf pulse to rephase the MR signal. The long time between adjacent ky points gives rise to a relatively low effective bandwidth in the phase direction. The reconstruction cannot determine whether an increase in phase is legitimate or not. That is, any increase in phase is interpreted in the same manner, whether it arises from phase encoding, or from accumulated phase error. Therefore, off resonance spins that have accumulated phase error are misregistered in the phase encode direction. 4/36

5 Phase bandwidth Bandwidth is inversely proportional to the sampling time. In the phase direction, the complete range of frequencies over the image in the phase direction becomes: Equation 2.1 where τpe is the time between two successive phase encoding steps, or time between adjacent ky points. The number of pixels by which the phase error is shifted is dependent upon the phase bandwidth per pixel, or BW/Ny. It is the same for the read encoding direction, where the bandwidth per pixel becomes: Equation 2.2 Where t is the sample time, or time between read encoding points. 5/36

6 Chemical shift review The chemical shift between water and fat is 3.3 ppm. The change in frequency ( ν) generated by this chemical shift can be determined by: Eqn 2.3 where Bσ is δ B0/106. What is ν at 1.5 T? (γ/2π = MHz/T) Solution: Bσ = /106 = T ν = MHz/T T = Hz This change in frequency gives rise to a shift in pixels for the chemical that is off resonance. In EPI, this occurs in both the read and phase directions. The chemical shift in pixels is determined by dividing the change in frequency by the bandwidth per pixel. Note that the bandwidth per pixel is equal to BW/Nx (or BW/Ny in the phase direction). What is the chemical shift in pixels for the above example, for a bandwidth along the read direction of 200 khz, with Nx = 128? Solution: Using the same general formula, what is the chemical shift in pixels in the phase direction for the above example? Use a time of 640 µs between phase encoding steps ( τpe) and Ny of 128 in your calculations. Solution: 6/36

7 7/36

8 Chemical shift in pixels The total number of pixels in the read direction (nread) and phase direction (nphase) shifted due to the chemical shift tissue difference ( B σ = B0 δ/106) is given by: Eqn 2.4 Eqn 2.5 where t is the sampling time, or time between each read encoding point, τpe is the time between phase encoding points, Nx and Ny are the number of read and phase encoding steps. You can see the derivation of these equations in the Extension screen. Note that Ny τpe in EPI is the time to complete the echo train. In other words, the off resonance frequency shift is proportional to the amount of time the spins are accumulating phase error. Watch the animation below to discover the pixel shifts in the read and phase direction for a variety of different chemical shifts in ppm, and imaging parameters. (Note: SW = sweep width which is similiar to Bandwidth) 8/36

9 Simulation 2.1: Calculated nphase and nread for different B0, τ peand SW (see on-screen simulation) 9/36

10 Extension Screen: Derivation of chemical shift in pixels This material is for your information only and is not examinable. To calculate the chemical shift in pixels in the phase direction, the phase bandwidth is used, and the matrix size in the phase direction. 10/36

11 Removing chemical shift artefacts Fat is shifted a significant amount in the phase direction in EPI images. See Figure 2.2(a) below where the subcutaneous fat has been shifted one-eighth of the FOV. All EPI images are fat suppressed to remove this problem (see Figure 2.2(b)). There are a variety of different fat suppression methods available, and a review can be found in MRES7004, module 8. In Module One, a comparison was made between a SE, FSE and EPI image of the brain (Figure 2.3). The missing element was fat, because it had been successfully suppressed. This is why the contrast of SE-EPI or GE-EPI resembles a fat suppressed SE or GE image respectively. (a) 11/36

12 (b)figure 2.2: (a) Non fat suppressed and (b) fat suppressed EPI of brain. Figure 2.3: (a) Spin echo, (b) fast spin echo, and (c) EPI images of the brain (Source: Reproduced by courtesy of GE Medical Systems (1995) Echo Planar Imaging: Principles and Applications, p. 12.) 12/36

13 Magnetic susceptibility Adjacent tissues with different magnetic susceptibilities cause disruptions to the local field (see Module 7, MRES7001, Susceptibilities). Some examples of this are bone-tissue interfaces found in the cancellous or 'spongy' bone, or the air-tissue interfaces found in the lung and in the sinuses, such as the ethmoid, maxillary, and mastoid sinuses. Any water protons within this local field inhomogeneity will precess off resonance. The frequency shift of these spins will vary from region to region and patient to patient, but will behave the same as off resonance fat. That is, they are wrongly positioned in the phase encoding direction. Note that very large inhomogeneity in the local field, like those caused by metal braces on the teeth, are still going to cause large distortions even with the most perfectly optimised sequence. Obviously suppressing the water signal as well as fat defeats the purpose. Figure 2.4: A single shot EPI image of the brain with an 256 echoes in the train. The time between each image was 2.7 ms, giving a total train length time of 700 ms. Source: Reproduced by courtesy of GE Medical Systems (1996) EPI Applications Guide, p /36

14 Reducing geometric distortions One way to reduce the impact of the geometric distortion is to swap the phase and read directions.it is possible to minimise the impact by placing the phase direction along a natural axis of symmetry (see figure 2.5) (a) (b) Figure 2.5: Two images of the brain are shown, identical in imaging parameters except for the placement of the phase encoding direction. The phase lies from right to left in (a) and anterior to posterior in (b) 14/36

15 Figure 2.6 To reduce the degree of distortion the echo train length or echo spacing needs to be reduced. The longer the ETL, the more time the spins have to accumulate phase error and the greater the distortions. For FSE, a shorter the echo spacing (or ESP) means the less blurring occurs within the image. In EPI, reducing the echo spacing also decreases the blurring, but, more importantly, also decreases the geometric distortion. Figure 2.6: The read encoding gradient and the sampling time during usual gradient performance (no ramp sampling). There are several methods to reduce the echo spacing. Unfortunately, as with most things in EPI, if one parameter is changed, another aspect is also changed, be it signal to noise, spatial resolution, or scan time (see Table 2.1). Parameter Echo Spacing Slew Rate Receiver Bandwidth Echo Train Length Signal to Spatial Noiseratio Resolution Scan Time _ Geometric Distortion Frequency Encoding Steps Phase Encoding Steps _ Frequency FOV _ Phase FOV _ _ Ramp Sampling _ Field Strength _ Table 2.1: Relationship between parameters in EPI and technical and image components 15/36

16 Receiver bandwidth The rf receiver bandwidth describes how fast the MR signal can be digitised. The higher the receiver bandwidth (RBW) the faster the digitisation can occur. Receiver bandwidth = 1/ t (Eqn 2.6) where t is the time between samples, or the dwell time (see Figure 2.6). Figure 2.7: Data acquisition during a simplified trapezoidal gradient waveform. RBW determines the range of frequencies that will be sampled by the analogue digital converter (ADC), spatially encoded by the frequency or readout gradient. To have enough image data to create and display an image, a balance must be found between the range sampled and the time needed to sample them. For example, to avoid wrap around artefact the receiver must obey the following condition: RBW = γ Gx FOV (Eqn 2.7) 16/36

17 Figure 2.8: The relationship between receiver bandwidth and frequency gradient If the RBW is increased, the amount of time at the flat top of the gradient reduces significantly. (see Figure 2.8) Shortening the duration of the flat top reduces ESP and geometric distortion. However, as RBW increases, more inherent noise is sampled along with the signal, decreasing the signal to noise ratio. Eqn /36

18 Ramp Sampling In a trapezoid gradient waveform, the sampling of spins usually occurs during the application of a constant gradient (the 'flat' portion) (Figure 2.9a). The time duration of the flat top is dependent on the frequency encoding matrix and the dwell time. It is possible to collect or sample the echo while the gradient is rising and falling (ramping in either positive or negative direction), as well as the flat top. This is referred to as ramp sampling (Figure 2.9b). Figure 2.9 (Left): Comparing normal data acquisition (a) with ramp sampling (b). Ramp sampling can be used either to maintain the current spatial resolution while decreasing the echo spacing (Figure 2.10 a), or it can be used to increase the resolution while maintaining the same echo spacing (Figure 2.10 b). Figure 2.10 (Right): Gradient waveforms where ESP is reduced (a) or spatial resolution is increased (b). 18/36

19 Spins being sampled during the ramp experience a non-constant gradient and therefore have different encoding values. The changing value of the magnetic field under the ramp is predictable and consistent, and the data must be repositioned or regridded within k-space before Fourier Transform (see see Extension screen). 19/36

20 Extension Screen: Fourier Regridding Fourier Transform can only occur on k-space data that are linearly spaced along a two-dimensional grid. The position in k-space is determined by: The data points sampled under a ramping gradient experience a nonlinear gradient field. To make sure that k is the same for all points they must be sampled at different times. This requires a variable bandwidth filter to minimise noise. Using a variable bandwidth can lead to reduced signal to noise. Alternatively, it is possible to over-sample in k-space by defining a sample rate based upon the maximum gradient: The Nyquist criteria is satisfied by this sampling rate. The resulting data will still not be uniformly spaced in k-space as the gradient varies with time and the data should be interpolated before reconstruction. The change in gradient field is a known quantity and the raw data can be corrected by convolution with an apodised sinc kernal. This approach is useful if the available hardware is not flexible enough to implement non-uniform sampling times, but the method does require more data storage and processing. Note that the raw data in k-space generated by EPI has to be reordered anyway prior to Fourier transform. Each second line which is encoded from right to left would have to be temporally reversed (back into left to right orientation) prior to reconstruction. 20/36

21 Field of view The field of view (FOV) is proportional to the sampling rate ( t) and gradient strength (Gread): Eqn 2.9 Or in other words, it is inversely proportional to the area under the gradient t Gread frequency encoding matrix). Usually the field of view would be determined by the size of the area of interest. The geometric distortion in EPI is a percentage of the field of view. For example, a geometric distortion that is 10% of the FOV appears less with a 20 cm FOV in the phase encoding direction than with a 40 cm FOV (Figure 2.11). However, the Nyquist ghost is also shifted half the FOV away, and will overlap less with the original image where the larger FOV is in the phase encoding direction. Figure 2.11: (a) 20 cm FOV, (b) 40 cm FOV. In the frequency direction, larger than optimal fields of view might be necessary to reduce geometric distortions. The amplitude of the frequency gradient must increase to obtain smaller field of views. However, for a fixed slew rate, the rise time required will increase as the gradient amplitude increases. This translates to an increased ESP (given that all other parameters remain the same) (Figure 2.12). 21/36

22 Figure 2.12: FOV and gradient shape As for conventional sequences, decreasing FOV results in a loss in signal to noise. Signal to noise is directly proportional to the field of view (SNR approximately equals to FOV) 22/36

23 Slew Rate To make the most of smaller fields of view or thinner slices, the slew rate of the gradient must be able to increase. The gradient slew rate is the rate of ascent to reach the gradient amplitude, or gradient amplitude divided by the rise time. In Figure 2.13, gradient A with the faster slew rate has a shorter ESP and therefore will result in a higher quality image with less geometric distortions. The required time to play out all the gradients will take less time for A resulting in a faster image acquisition, or shorter ETL, or time to acquire more slices during TR. Note that all other parameters, such as signal to noise and resolution, remain the same. Figure 2.13: Gradients A and B, both with the same amplitude but different slew rates. 23/36

24 Frequency Encoding Steps The larger the frequency matrix, the higher the resolution in the frequency direction. The frequency matrix size (Nx) is related to the pixel dimension in the frequency direction ( x) and the field of view by the following: Eqn 2.10 Increasing the number of frequency encoding steps increases the gradient duration and increases ESP (Figure 2.14). It is quite common for EPI images acquired on conventional MR systems to have medium-to-low resolution matrices (from 32 up to 128) due to ESP constraints. To prevent unacceptable increases in geometric distortion, increases in frequency steps should be balanced with increases in receiver bandwidth. Alternatively, the EPI image can be acquired in a number of segments, or over several repetition times instead of acquiring the whole image in a single shot. This is known as multi-shot or segmented EPI and requires a new excitation pulse or TR per segment (see later Screen). Figure 2.14: Interaction between Nx and gradient shape 24/36

25 Phase Encoding Steps Increasing the number of phase encoding steps increases the echo train length in single shot FSE and EPI. The echo spacing does not change, but blurring will increase depending upon the T2 /T2* time of the tissue of interest. In the case of EPI, there is also more time for phase errors to accumulate, giving rise to greater geometric distortion. Increasing the number of phase encoding steps does not directly increase the scan time, providing the slice coverage is maintained. The overall time to acquire the image is increased, but this is only a portion of the 'dead time' or recovery time (see Figure 2.15). The remainder is usually devoted to slice acquisition. Only if TR needs to be increased to maintain the slice coverage does the scan time increase. One way of increasing the resolution without increasing the frequency of phase encoding steps is known as zero filling. For example, a 96 x 96 matrix can be acquired and then interpolated to a 256 x 256 image. The other way of improving the resolution without image blurring or distortion is segmenting to a multi-echo method rather than single shot, parallel imaging, or using the symmetry of k-space (these last two will be discussed in later modules). Figure 2.15: Schematic of acquisition time 25/36

26 Segmentation In a multi-shot FSE or EPI method, several excitation pulses are needed to collect the whole of k-space. Eqn 2.11 The degree of segmentation can be referred to either as the echo train length per excitation or the complete number of 'shots' used to acquire the full image. For example, 256 phase encoding steps acquired in 32 shots would have an echo train length of 8. Segmentation commonly occurs in an interleaved manner, where the extent of k-space is covered (e.g. -kx(max) to +kx(max)), but in reduced detail. Animation 2.2: Interleaved segmentation of a segmented EPI sequence See animation on screen 26/36

27 Advantages and Disadvantages of Segmentation Segmentation provides a number of advantages and disavantages. Segmentation places less stress on the gradients compared with single shot methods, particularly EPI. Phase errors have less time to build up compared with single shot EPI, thus reducing the magnetic susceptibility artefacts (see Figure 2.16). Less blurring occurs in FSE methods as the echo train length reduces. Segmented methods allow for increased resolution without losing image quality. Segmentation allows for T1 weighting to be introduced Segmented methods take longer to perform. The scan time is equal to TR x Ns x NA. Because of the above, segmented methods can be more prone to motion artefacts than single-shot. Segmentation methods are prone to specific artefacts, because of the complex coverage of k-space. EPI is more susceptible to this compared to FSE due to the alternating directions in k-space (L-R then R-L). 27/36

28 Figure 2.16: Demonstration of the reduction of geometric distortion as the number of shots is increased. (a) 2 shots (ETL 128) (b) 4 shots (ETL 64) (c) 8 shots (ETL 32) (d) 16 shots (ETL 16). 28/36

29 Repetition Time (TR) As there is only one TR in single shot methods, the effective TR is infinite, and is only used to determined the space of time allocated for multi-slice selection. In segmented or multi-shot versions of EPI and FSE, the lines of k-space are being filled over several TR periods. In this case, the TR should be selected according to the desired contrast. The longer the repetition time, the less T1 -weighting there is within the image, but the higher the SNR. In multi-shot EPI or FSE, if the TR is shorter than the 5T1 of the tissue, the longitudinal magnetisation will be slowly decreasing with each new excitation. This also occurs for single-shot methods where you are repeatedly exciting the same slices over time (ie following a temporal series of events). Eventually, the spins become saturated and reach steady-state magnetisation. This is where the amount of magnetisation flipped into the transverse plane can fully recovery within the TR. To avoid artefacts, the first few segments with changing SNR are ignored - usually additional excitation pulses are applied but the information discarded. Figure 2.18: Diagram illustrating steady state magnetisation. 29/36

30 Nyquist or N/2 Ghosting (I) Time varying magnetic field gradients result in current induction (eddy currents) in the various conductive components of the rest of the imaging instrument. These, in turn, create magnetic field gradients that may persist after the primary gradients are switched off. While a problem in conventional imaging, they are more severe in EPI. The gradient amplitudes and switching rates in EPI are usually much greater, causing larger eddy currents. Also, the long readout period in EPI results in more opportunity for image distortion from eddy currents. Eddy currents (plus gradient power supply and pre-emphasis problems) transform the trapezoid gradient shape from the idealised version shown so far into a non-linear shape (Figure 2.18). Since the transformation is unknown it cannot be accounted before during acquisition. K-space lines corresponding to positive and negative oscillating gradients are shifted. Figure 2.18: The results of eddy currents on gradient waveforms and in k -space. 30/36

31 Nyquist or N/2 Ghosting (II) The Fourier transform of two shifted signals in k-space, superimposed, has an interferogram effect, as seen in Figure There is a ghost one-half the field of view away in the phase encoding direction. This is referred to as a Nyquist or N/2 ghost. The modulation of the ghost is 180o out of phase relative to the original, which can lead to interference patterns when the images overlap (black lines). Figure 2.19: Examples of the N/2 ghost artefact in EPI. From left to right the degree of shift ( ) is getting smaller. However, when a segmented EPI method is used, the ghosting is increased. Where there was one ghost shifted half the FOV from the original image, there would now be the same number of ghosts as there are number of shots (Ns) in the phase direction. They would be separated by: 31/36

32 Nyquist or N2 Ghosting (III) That is, the spacing between ghosts in segmented EPI are smaller than for single shot EPI, and they are more likely to overlap the original image. See Figure 2.20b a) Figure 2.20: Four shot EPI method showing the shifts in k-space (a) and the resulting ghosting (b). 32/36

33 Reduce Nyquist ghosting To reduce N/2 ghosting: use appropriately designed coils to minimise eddy current induction (see MRES7002, Module Five) adjust the sampling clock in the hardware to make sure the timing between signal digitisation and gradient activity is calibrated within the software, add an extra timing factor to incorporate an appropriate phase shift in theraw data use the shift parameter ( from Figure 2.18) during reconstruction to construct an artefact free image. In single shot it is possible to calculate a phase shift parameter from regions where the original and the ghosts have not overlapped and use this to reconstruct an artifact free image. This is not possible with interleaved experiments, in which the overlapping of ghosts is almost unavoidable. The solution is to perform a reference or calibration scan without phase gradients to measure the phase shift. From the raw data, phase correction values can be derived. 33/36

34 Recovering SNR and Resolution In trying to decrease the geometric distortion through parameter optimisation, signal to noise is often sacrificed. Keep in mind that different parameters will affect SNR to different extents. There are several ways of recovering SNR: using receiver coils to improve signal to noise increase slice thickness (lose resolution in the slice direction) signal averaging (which increases overall scan time) increase the number of segments change the EPI mode to SE-EPI - this has slightly less dependence upon T2* use magnets with large field strengths. The SNR increases with increasing field strength, but chemical shift and geometric distortions also increase. Resolution is the other area where sacrifices are made, particularly in EPI. Resolution depends upon the maximum gradient amplitude available and the sampling time. To increase resolution: increase gradient amplitude The best resolution is achieved by using the maximum possible gradient amplitude. increase gradient duration...but suffer more geometric distortions use a segmented or multi-shot method Use parallel imaging (see Module 6) Undersample k-space but reconstruct at a higher resolution (zero filling; k-space symmetry; interpolation) 34/36

35 Summary Artefact Image Blurring Chemical Shift Artefact Chemical Shift Artefact Nyquist Ghosting To remove or reduce Reduce ESP Fat Suppression/Saturation Phase encoding along natural axis of symmetry Reduce ETL Reduce ESP Gradient Shielding Adjust sampling clock Apply a phase shift in software; calculated from a reference scan. Reducing Echo Spacing Imaging Parameter Receiver bandwidth Ramp Sampling FOV - frequency direction Slew rate Ny - phase matrix Nx - frequency matrix Effect Large bandwidths mean less distortions but more noise. Use to decrease distortions or improve resolution, but regridding is necessary Increasing will lower geometric distortions if your slew rate is fixed. Faster slew rate means shorter ESP and ETL, less image blurring and less distortions. Increasing will increase resolution, but increase echo train length. Increasing will increase resolution, but increase distortion Review Questions What causes image blurring in single shot sequences? How can blurring be removed or reduced and what impact does that have? What is a chemical shift artefact? Which sequence is most affected by it and why? How can chemical shift artefact be removed or reduced and what impact does that have? What is Nyquist ghosting? Which sequence is most affected by it and why? 35/36

36 How can Nyquist ghosting be removed or reduced and what impact does that have? What is geometric distortion? Which sequence is most affected by it and why? How can geometric distortion be removed or reduced, and what impact does that have? 36/36

Background (~EE369B)

Background (~EE369B) Background (~EE369B) Magnetic Resonance Imaging D. Nishimura Overview of NMR Hardware Image formation and k-space Excitation k-space Signals and contrast Signal-to-Noise Ratio (SNR) Pulse Sequences 13

More information

H 2 O and fat imaging

H 2 O and fat imaging H 2 O and fat imaging Xu Feng Outline Introduction benefit from the separation of water and fat imaging Chemical Shift definition of chemical shift origin of chemical shift equations of chemical shift

More information

Pulse Sequence Design and Image Procedures

Pulse Sequence Design and Image Procedures Pulse Sequence Design and Image Procedures 1 Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant 2 A pulse sequence is a timing diagram designed with a series of RF pulses, gradients switching, and signal readout

More information

Pulse Sequence Design Made Easier

Pulse Sequence Design Made Easier Pulse Sequence Design Made Easier Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant gurumri@gmail.com 1 2 Pulse Sequences generally have the following characteristics: An RF line characterizing RF Pulse applications

More information

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Magnetic Resonance Imaging Spatial

More information

MR Basics: Module 6 Pulse Sequences

MR Basics: Module 6 Pulse Sequences Module 6 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images 10. Phase Cycling and Pulsed Field Gradients 10.1 Introduction to Phase Cycling - Quadrature images The selection of coherence transfer pathways (CTP) by phase cycling or PFGs is the tool that allows the

More information

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves Experiment 1 Goal: Examine the effect caused by changing flip angle on image contrast in a simple gradient echo sequence and derive T1-curves. Image

More information

2014 M.S. Cohen all rights reserved

2014 M.S. Cohen all rights reserved 2014 M.S. Cohen all rights reserved mscohen@g.ucla.edu IMAGE QUALITY / ARTIFACTS SYRINGOMYELIA Source http://gait.aidi.udel.edu/res695/homepage/pd_ortho/educate/clincase/syrsco.htm Surgery is usually recommended

More information

RAD 229: MRI Signals and Sequences

RAD 229: MRI Signals and Sequences RAD 229: MRI Signals and Sequences Brian Hargreaves All notes are on the course website web.stanford.edu/class/rad229 Course Goals Develop Intuition Understand MRI signals Exposure to numerous MRI sequences

More information

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements MR Data for Treatment Planning: Issues, Protocol Optimization, and s (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements TG-117 Use of MRI Data in Treatment Planning

More information

1 Introduction. 2 The basic principles of NMR

1 Introduction. 2 The basic principles of NMR 1 Introduction Since 1977 when the first clinical MRI scanner was patented nuclear magnetic resonance imaging is increasingly being used for medical diagnosis and in scientific research and application

More information

Image Quality/Artifacts Frequency (MHz)

Image Quality/Artifacts Frequency (MHz) The Larmor Relation 84 Image Quality/Artifacts (MHz) 42 ω = γ X B = 2πf 84 0.0 1.0 2.0 Magnetic Field (Tesla) 1 A 1D Image Magnetic Field Gradients Magnet Field Strength Field Strength / Gradient Coil

More information

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Dr John Ridgway Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Physics for clinicians: Part I Journal of Cardiovascular Magnetic Resonance 2010, 12:71 http://jcmr-online.com/content/12/1/71

More information

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University Advanced MSK MRI Protocols at 3.0T Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University Outline Why High Field for MSK? SNR and Relaxation Times Technical Issues Example

More information

MRI Metal Artifact Reduction

MRI Metal Artifact Reduction MRI Metal Artifact Reduction PD Dr. med. Reto Sutter University Hospital Balgrist Zurich University of Zurich OUTLINE Is this Patient suitable for MR Imaging? Metal artifact reduction Is this Patient suitable

More information

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010 (N)MR Imaging Lab Course Script FMP PhD Autumn School Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder Date: November 3rd, 2010 1 Purpose: Understanding the basic principles of MR imaging

More information

Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils

Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils Magn Reson Med Sci doi:10.2463/mrms.tn.2016-0049 Published Online: March 27, 2017 TECHNICAL NOTE Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils

More information

2015 Spin echoes and projection imaging

2015 Spin echoes and projection imaging 1. Spin Echoes 1.1 Find f0, transmit amplitudes, and shim settings In order to acquire spin echoes, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week,

More information

MR Basics: Module 8 Image Quality

MR Basics: Module 8 Image Quality Module 8 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

Echo-Planar Imaging for a 9.4 Tesla Vertical-Bore Superconducting Magnet Using an Unshielded Gradient Coil

Echo-Planar Imaging for a 9.4 Tesla Vertical-Bore Superconducting Magnet Using an Unshielded Gradient Coil Magn Reson Med Sci, Vol. XX, No. X, pp. XXX XXX, 2015 2016 Japanese Society for Magnetic Resonance in Medicine TECHNICAL NOTE by J-STAGE doi:10.2463/mrms.tn.2015-0123 Echo-Planar Imaging for a 9.4 Tesla

More information

System/Imaging Imperfections

System/Imaging Imperfections System/Imaging Imperfections B0 variations: Shim, Susceptibility B1 variations: Transmit, Receive Gradient Imperfections: Non-linearities Delays and Eddy currents Concomitant terms 1 B0 Variations - Off-Resonance

More information

Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling

Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling Coherence editing is used to remove unwanted signals from NMR spectra. For example, in the double quantum filtered COSY experiment,

More information

BOLD fmri: signal source, data acquisition, and interpretation

BOLD fmri: signal source, data acquisition, and interpretation BOLD fmri: signal source, data acquisition, and interpretation Cheryl Olman 4 th year student, Department of Neuroscience and Center for Magnetic Resonance Research Discussion series Week 1: Biological

More information

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP MR Data for Treatment Planning and Stereotactic Procedures: Sources of Distortion, Protocol Optimization, and Assessment (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements

More information

Encoding of inductively measured k-space trajectories in MR raw data

Encoding of inductively measured k-space trajectories in MR raw data Downloaded from orbit.dtu.dk on: Apr 10, 2018 Encoding of inductively measured k-space trajectories in MR raw data Pedersen, Jan Ole; Hanson, Christian G.; Xue, Rong; Hanson, Lars G. Publication date:

More information

Liver imaging beyond expectations with Ingenia

Liver imaging beyond expectations with Ingenia Publication for the Philips MRI Community Issue 47 2012/3 Liver imaging beyond expectations with Ingenia Contributed by John Penatzer, RT, MR clinical product specialist, Cleveland, OH, USA Publication

More information

HETERONUCLEAR IMAGING. Topics to be Discussed:

HETERONUCLEAR IMAGING. Topics to be Discussed: HETERONUCLEAR IMAGING BioE-594 Advanced MRI By:- Rajitha Mullapudi 04/06/2006 Topics to be Discussed: What is heteronuclear imaging. Comparing the hardware of MRI and heteronuclear imaging. Clinical applications

More information

Gradients. Effects of B0 gradients on transverse magnetisation Similar to figure 10 of Sattler review Progr. NMR 34 (1999), 93

Gradients. Effects of B0 gradients on transverse magnetisation Similar to figure 10 of Sattler review Progr. NMR 34 (1999), 93 Gradients 1. What are gradients? Modern high-resolution NMR probes contain -besides the RF coils - additional coils that can be fed a DC current. The coils are built so that a pulse (~1 ms long) of DC

More information

3T Unlimited. ipat on MAGNETOM Allegra The Importance of ipat at 3T. medical

3T Unlimited. ipat on MAGNETOM Allegra The Importance of ipat at 3T. medical 3T Unlimited ipat on MAGNETOM Allegra The Importance of ipat at 3T s medical ipat on MAGNETOM Allegra The Importance of ipat at 3T The rise of 3T MR imaging Ultra High Field MR (3T) has flourished during

More information

MAGNETIC RESONANCE IMAGING

MAGNETIC RESONANCE IMAGING CSEE 4620 Homework 3 Fall 2018 MAGNETIC RESONANCE IMAGING 1. THE PRIMARY MAGNET Magnetic resonance imaging requires a very strong static magnetic field to align the nuclei. Modern MRI scanners require

More information

Lab 8 6.S02 Spring 2013 MRI Projection Imaging

Lab 8 6.S02 Spring 2013 MRI Projection Imaging 1. Spin Echos 1.1 Find f0, TX amplitudes, and shim settings In order to acquire spin echos, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week, but these

More information

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging JOURNAL OF MAGNETIC RESONANCE IMAGING 20:1046 1051 (2004) Technical Note The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging James W. Goldfarb, PhD* Purpose: To describe a known (but undocumented)

More information

MR Advance Techniques. Flow Phenomena. Class II

MR Advance Techniques. Flow Phenomena. Class II MR Advance Techniques Flow Phenomena Class II Flow Phenomena In this class we will explore different phenomenona produced from nuclei that move during the acquisition of data. Flowing nuclei exhibit different

More information

Half-Pulse Excitation Pulse Design and the Artifact Evaluation

Half-Pulse Excitation Pulse Design and the Artifact Evaluation Half-Pulse Excitation Pulse Design and the Artifact Evaluation Phillip Cho. INRODUCION A conventional excitation scheme consists of a slice-selective RF excitation followed by a gradient-refocusing interval

More information

RF pulse design and the Small Tip Angle Approximation

RF pulse design and the Small Tip Angle Approximation RF pulse design and the Small Tip Angle Approximation Dr Shaihan J Malik Lecturer in Imaging Sciences Division of Imaging Sciences & Biomedical Engineering King s College London shaihan.malik@kcl.ac.uk

More information

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Val M. Runge, MD Institute for Diagnostic and Interventional Radiology Clinics for Neuroradiology and Nuclear Medicine University Hospital Zurich

More information

SIEMENS MAGNETOM Skyra syngo MR D13

SIEMENS MAGNETOM Skyra syngo MR D13 Page 1 of 12 SIEMENS MAGNETOM Skyra syngo MR D13 \\USER\CIND\StudyProtocols\PTSA\*ep2d_M0Map_p2_TE15 TA:7.9 s PAT:2 Voxel size:2.5 2.5 3.0 mm Rel. SNR:1.00 :epfid Properties Routine Contrast Prio Recon

More information

SIGNA Explorer Lift revives our MR

SIGNA Explorer Lift revives our MR Seiji Shiotani, MD, PhD Seirei Fuji Hospital in Fuji City, Shizuoka, Japan Masayoshi Sugimura Seirei Fuji Hospital in Fuji City, Shizuoka, Japan SIGN Explorer Lift revives our MR The clinical usefulness

More information

k y 2k y,max k x 2k x,max

k y 2k y,max k x 2k x,max EE225E/BIOE265 Spring 2012 Principles of MRI Miki Lustig Assignment 5 Due Feb 26, 2012 1. Finish reading Nishimura Ch. 5. 2. For the 16 turn spiral trajectory, plotted below, what is the a) Spatial resolution,

More information

High Field MRI: Technology, Applications, Safety, and Limitations

High Field MRI: Technology, Applications, Safety, and Limitations High Field MRI: Technology, Applications, Safety, and Limitations R. Jason Stafford, Ph.D. The University of Texas M. D. Anderson Cancer Center, Houston, TX Introduction The amount of available signal

More information

SIGNA Pioneer: Ultra High Efficiency Gradient System Advancing the gradient technology curve

SIGNA Pioneer: Ultra High Efficiency Gradient System Advancing the gradient technology curve GE Healthcare SIGNA Pioneer: Ultra High Efficiency Gradient System Advancing the gradient technology curve NEW TECHNOLOGY 40W Watts spec is irrelevant. 4W LED bulb delivers same brightness as 40W incandescent

More information

NMR Basics. Lecture 2

NMR Basics. Lecture 2 NMR Basics Lecture 2 Continuous wave (CW) vs. FT NMR There are two ways of tuning a piano: - key by key and recording each sound (or frequency). - or, kind of brutal, is to hit with a sledgehammer and

More information

NIH Public Access Author Manuscript Magn Reson Med. Author manuscript; available in PMC 2010 July 21.

NIH Public Access Author Manuscript Magn Reson Med. Author manuscript; available in PMC 2010 July 21. NIH Public Access Author Manuscript Published in final edited form as: Magn Reson Med. 2010 April ; 63(4): 1092 1097. doi:10.1002/mrm.22223. Spatially Varying Fat-Water Excitation Using Short 2DRF Pulses

More information

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo Gradient Spoiling Average balanced SSFP magnetization Reduce sensitivity to off-resonance FFE, FISP, GRASS, GRE, FAST, Field Echo 1 Gradient-Spoiled Sequence (GRE, FFE, FISP, GRASS) RF TR G z G y G x Signal

More information

2 Hardware for Magnetic Resonance Imaging

2 Hardware for Magnetic Resonance Imaging Hardware for Magnetic Resonance Imaging 13 2 Hardware for Magnetic Resonance Imaging Kenneth W. Fishbein, Joseph C. McGowan, and Richard G. Spencer CONTENTS 2.1 Introduction 13 2.2 Magnets 13 2.2.1 Permanent

More information

PHY3902 PHY3904. Nuclear magnetic resonance Laboratory Protocol

PHY3902 PHY3904. Nuclear magnetic resonance Laboratory Protocol PHY3902 PHY3904 Nuclear magnetic resonance Laboratory Protocol PHY3902 PHY3904 Nuclear magnetic resonance Laboratory Protocol GETTING STARTED You might be tempted now to put a sample in the probe and try

More information

Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park

Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park Department of Bio and Brain Engineering, Korea Advanced Institute of Science and Technology (KAIST), Daejeon,

More information

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 ACR MRI accreditation program MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics,

More information

Supplementary Figure 1. Scanning Electron Microscopy images of the pristine electrodes. (a) negative electrode and (b) positive electrode.

Supplementary Figure 1. Scanning Electron Microscopy images of the pristine electrodes. (a) negative electrode and (b) positive electrode. a b Supplementary Figure 1. Scanning Electron Microscopy images of the pristine electrodes. (a) negative electrode and (b) positive electrode. Images were performed using a FEI/Philips XL4 microscope with

More information

ISSN X CODEN (USA): PCHHAX. The role of dual spin echo in increasing resolution in diffusion weighted imaging of brain

ISSN X CODEN (USA): PCHHAX. The role of dual spin echo in increasing resolution in diffusion weighted imaging of brain Available online at www.derpharmachemica.com ISSN 0975-413X CODEN (USA): PCHHAX Der Pharma Chemica, 2016, 8(17):15-20 (http://derpharmachemica.com/archive.html) The role of in increasing resolution in

More information

Magnetic Resonance Imaging Principles, Methods, and Techniques

Magnetic Resonance Imaging Principles, Methods, and Techniques Magnetic Resonance Imaging Principles, Methods, and Techniques Perry Sprawls Jr., Emory University Publisher: Medical Physics Publishing Corporation Publication Place: Madison, Wisconsin Publication Date:

More information

磁振影像學 MRI 磁振假影與磁振安全 磁振假影. 本週課程內容 Hardware-related Artifacts 盧家鋒助理教授 磁振假影 磁振安全

磁振影像學 MRI 磁振假影與磁振安全 磁振假影. 本週課程內容   Hardware-related Artifacts 盧家鋒助理教授 磁振假影 磁振安全 本週課程內容 http://www.ym.edu.tw/~cflu 磁振假影 磁振安全 磁振影像學 MRI 磁振假影與磁振安全 盧家鋒助理教授 國立陽明大學生物醫學影像暨放射科學系 alvin4016@ym.edu.tw MRI The Basics (3rd edition) Chapter 18: Artifacts in MRI MRI in Practice, (4th edition) Chapter

More information

Chapter 5: Signal conversion

Chapter 5: Signal conversion Chapter 5: Signal conversion Learning Objectives: At the end of this topic you will be able to: explain the need for signal conversion between analogue and digital form in communications and microprocessors

More information

Chapter 2. The Physics of Magnetic Resonance Imaging

Chapter 2. The Physics of Magnetic Resonance Imaging Chapter 2. The Physics of Magnetic Resonance Imaging 2.1. Introduction The origins of the Nuclear Magnetic Resonance (NMR) signal and how it is manipulated to form images are the subjects of this chapter.

More information

Magnetic Resonance Imaging

Magnetic Resonance Imaging Magnetic Resonance Imaging Principles, Methods, and Techniques Perry Sprawls, Ph.D., FACR, FAAPM, FIOMP Distinguished Emeritus Professor Department of Radiology Emory University Atlanta, Georgia Medical

More information

MRI Anatomy and Positioning Series Module 12: Fat Suppression Techniques

MRI Anatomy and Positioning Series Module 12: Fat Suppression Techniques MRI Anatomy and Positioning Series Module 12: Fat Suppression Techniques 1 Introduction... 3 RF FatSat... 4 HOAST... 4 FatSat... 5 Segment FS... 8 PhaseCycle... 9 Water Excitation... 10 STIR... 12 FatSep...

More information

TimTX TrueShape. The parallel transmit architecture of the future. Answers for life.

TimTX TrueShape.  The parallel transmit architecture of the future. Answers for life. www.siemens.com/trueshape TimTX TrueShape The parallel transmit architecture of the future. The product/feature (mentioned herein) is not commercially available. Due to regulatory reasons its future availability

More information

Works-in-Progress package Version 1.0. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B. January 22, 2003

Works-in-Progress package Version 1.0. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B. January 22, 2003 Works-in-Progress package Version 1.0 For the Installation and User s Guide NUMARIS/4VA21B January 22, 2003 Section of Medical Physics, University Hospital Freiburg, Germany Contact: Klaus Scheffler PhD,

More information

RF Pulse Design. Multi-dimensional Excitation II. M229 Advanced Topics in MRI Kyung Sung, Ph.D Class Business

RF Pulse Design. Multi-dimensional Excitation II. M229 Advanced Topics in MRI Kyung Sung, Ph.D Class Business RF Pulse Design Multi-dimensional Excitation II M229 Advanced Topics in MRI Kyung Sung, Ph.D. 2018.04.12 Class Business - Homework 1 will be due on 4/26 - Office hours Instructors: Fri 10-12 noon TAs:

More information

IR/SR TrueFISP. Works-in-Progress package Version 1.2. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B.

IR/SR TrueFISP. Works-in-Progress package Version 1.2. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B. Works-in-Progress package Version 1.2 For the Installation and User s Guide NUMARIS/4VA21B January 22, 2003 Section of Medical Physics, University Hospital Freiburg, Germany Contact: Klaus Scheffler PhD

More information

RF Pulse Toolkit: Application Specific Design

RF Pulse Toolkit: Application Specific Design RF Pulse Toolkit: Application Specific Design William A Grissom Department of Biomedical Engineering, Vanderbilt University, Nashville, TN, USA will.grissom@vanderbilt.edu Introduction RF excitation is

More information

MRI SYSTEM COMPONENTS Module One

MRI SYSTEM COMPONENTS Module One MRI SYSTEM COMPONENTS Module One 1 MAIN COMPONENTS Magnet Gradient Coils RF Coils Host Computer / Electronic Support System Operator Console and Display Systems 2 3 4 5 Magnet Components 6 The magnet The

More information

Supplementary Material

Supplementary Material Supplementary Material Orthogonal representation of sound dimensions in the primate midbrain Simon Baumann, Timothy D. Griffiths, Li Sun, Christopher I. Petkov, Alex Thiele & Adrian Rees Methods: Animals

More information

Principles of MRI EE225E / BIO265. Lecture 21. Instructor: Miki Lustig UC Berkeley, EECS. M. Lustig, EECS UC Berkeley

Principles of MRI EE225E / BIO265. Lecture 21. Instructor: Miki Lustig UC Berkeley, EECS. M. Lustig, EECS UC Berkeley Principles of MRI Lecture 21 EE225E / BIO265 Instructor: Miki Lustig UC Berkeley, EECS Question What is the difference between the images? Answer Both T1-weighted spin-echo gradient-echo Lower SNR Meniscus

More information

functional MRI: A primer

functional MRI: A primer Activation Leads to: functional MRI: A primer CBF Increased +ΔR CBV Increased +ΔR (C+) O Utilization Increased slightly? Venous [O ] Increased -ΔR* Glucose Utilization Increased? Lactate BOLD R=/T R=/T

More information

EE469B: Assignment 4 Solutions

EE469B: Assignment 4 Solutions EE469B Fall 26-7 RF Pulse Design for MRI EE469B: Assignment 4 Solutions Due Thursday Oct 27. True Null/Flyback Spectral-Spatial Pulses True null and flyback designs are very closely related. In this problem

More information

Chapter 2 Analog-to-Digital Conversion...

Chapter 2 Analog-to-Digital Conversion... Chapter... 5 This chapter examines general considerations for analog-to-digital converter (ADC) measurements. Discussed are the four basic ADC types, providing a general description of each while comparing

More information

H Micro-Imaging. Tuning and Matching. i. Open any 1H data set and type wobb.

H Micro-Imaging. Tuning and Matching. i. Open any 1H data set and type wobb. - 1-1 H Micro-Imaging The NMR-specific properties of the objects are visualized as multidimensional images. Translational motion can be observed and spectroscopic information can be spatially resolved.

More information

NEMA Standards Publication MS (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging

NEMA Standards Publication MS (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging NEMA Standards Publication MS 1-2008 (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging Published by: National Electrical Manufacturers Association 1300 North

More information

MRI Systems and Coil Technology

MRI Systems and Coil Technology MRI for Technologists MRI Systems and Coil Technology PROGRAM INFORMATION MRI for Technologists is a training program designed to meet the needs of radiologic technologists entering or working in the field

More information

In a typical biological sample the concentration of the solute is 1 mm or less. In many situations,

In a typical biological sample the concentration of the solute is 1 mm or less. In many situations, Water suppression n a typical biological sample the concentration of the solute is 1 mm or less. n many situations, the signals of interest are those of amide protons that exchange with the solvent water.

More information

Supplementary Figures

Supplementary Figures 1 Supplementary Figures a) f rep,1 Δf f rep,2 = f rep,1 +Δf RF Domain Optical Domain b) Aliasing region Supplementary Figure 1. Multi-heterdoyne beat note of two slightly shifted frequency combs. a Case

More information

MRI: From Signal to Image

MRI: From Signal to Image MRI: From Signal to Image Johannes Koch physics654 2013-05-06 1 / 27 Tomography Magnetic Resonance Tomography Tomography: tomos: section graphein: to write Signal measured as function of space 2 / 27 Tomography

More information

Your first NMR measurement

Your first NMR measurement Your first NMR measurement Introduction Select 10mM water in D2O as NMR sample. The NMR spectrum of such sample consists of only two signals: the water signal and the peak of the reference (TSP). Follow

More information

EE469B: Assignment 1 Solutions

EE469B: Assignment 1 Solutions EE469B Fall 26-7 RF Pulse Design for MRI EE469B: Assignment Solutions Due Thursday Oct 6 Introduction This assignment concerns typical Fourier transform designs of excitation pulses. This includes designing

More information

Radar-Verfahren und -Signalverarbeitung

Radar-Verfahren und -Signalverarbeitung Radar-Verfahren und -Signalverarbeitung - Lesson 2: RADAR FUNDAMENTALS I Hon.-Prof. Dr.-Ing. Joachim Ender Head of Fraunhoferinstitut für Hochfrequenzphysik and Radartechnik FHR Neuenahrer Str. 20, 53343

More information

Improving high-field MRI using parallel excitation

Improving high-field MRI using parallel excitation review Improving high-field MRI using parallel excitation MRI at high magnetic field strengths promises to deliver clearer images of the body s structure and function. However, high-field MRI currently

More information

Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging

Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging Poster No.: C-0709 Congress: ECR 2014 Type: Scientific

More information

I am very pleased to teach this class again, after last year s course on electronics over the Summer Term. Based on the SOLE survey result, it is clear that the format, style and method I used worked with

More information

1.5T HIGH FIELD SMALL ANIMAL MRI

1.5T HIGH FIELD SMALL ANIMAL MRI 1.5T HIGH FIELD SMALL ANIMAL MRI Designed Specifically for Veterinarians TECHNICAL GUIDE ADVANCING THE ART AND SCIENCE OF VETERINARY MRI The PetVet is the only high-field MRI system designed specifically

More information

Jitter Analysis Techniques Using an Agilent Infiniium Oscilloscope

Jitter Analysis Techniques Using an Agilent Infiniium Oscilloscope Jitter Analysis Techniques Using an Agilent Infiniium Oscilloscope Product Note Table of Contents Introduction........................ 1 Jitter Fundamentals................. 1 Jitter Measurement Techniques......

More information

Methods. Experimental Stimuli: We selected 24 animals, 24 tools, and 24

Methods. Experimental Stimuli: We selected 24 animals, 24 tools, and 24 Methods Experimental Stimuli: We selected 24 animals, 24 tools, and 24 nonmanipulable object concepts following the criteria described in a previous study. For each item, a black and white grayscale photo

More information

Pulse Sequences: Rapid Gradient Echo

Pulse Sequences: Rapid Gradient Echo Pulse Sequences: Rapid Gradient Echo M229 Advanced Topics in MRI Holden H. Wu, Ph.D. 2018.04.17 Department of Radiological Sciences David Geffen School of Medicine at UCLA Class Business Office hours -

More information

Gradient hysteresis in MRI and NMR experiments

Gradient hysteresis in MRI and NMR experiments Journal of Magnetic Resonance 177 (2005) 336 340 Communication Gradient hysteresis in MRI and NMR experiments Brian J. Nieman a,b, *, Jonathan Bishop a, R. Mark Henkelman a,b a Mouse Imaging Centre, Hospital

More information

Signals and Systems. Lecture 13 Wednesday 6 th December 2017 DR TANIA STATHAKI

Signals and Systems. Lecture 13 Wednesday 6 th December 2017 DR TANIA STATHAKI Signals and Systems Lecture 13 Wednesday 6 th December 2017 DR TANIA STATHAKI READER (ASSOCIATE PROFFESOR) IN SIGNAL PROCESSING IMPERIAL COLLEGE LONDON Continuous time versus discrete time Continuous time

More information

FLASH rf gun. beam generated within the (1.3 GHz) RF gun by a laser. filling time: typical 55 μs. flat top time: up to 800 μs

FLASH rf gun. beam generated within the (1.3 GHz) RF gun by a laser. filling time: typical 55 μs. flat top time: up to 800 μs The gun RF control at FLASH (and PITZ) Elmar Vogel in collaboration with Waldemar Koprek and Piotr Pucyk th FLASH Seminar at December 19 2006 FLASH rf gun beam generated within the (1.3 GHz) RF gun by

More information

6.S02 MRI Lab Acquire MR signals. 2.1 Free Induction decay (FID)

6.S02 MRI Lab Acquire MR signals. 2.1 Free Induction decay (FID) 6.S02 MRI Lab 1 2. Acquire MR signals Connecting to the scanner Connect to VMware on the Lab Macs. Download and extract the following zip file in the MRI Lab dropbox folder: https://www.dropbox.com/s/ga8ga4a0sxwe62e/mit_download.zip

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

Introduction: The FFT emission measurement method

Introduction: The FFT emission measurement method Introduction: The FFT emission measurement method Tim Williams Elmac Services C o n s u l t a n c y a n d t r a i n i n g i n e l e c t r o m a g n e t i c c o m p a t i b i l i t y Wareham, Dorset, UK

More information

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT) Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in

More information

Laboratory Assignment 5 Amplitude Modulation

Laboratory Assignment 5 Amplitude Modulation Laboratory Assignment 5 Amplitude Modulation PURPOSE In this assignment, you will explore the use of digital computers for the analysis, design, synthesis, and simulation of an amplitude modulation (AM)

More information

EE469B: Assignment 2 Solutions

EE469B: Assignment 2 Solutions EE469B Fall 26-7 RF Pulse Design for MRI EE469B: Assignment 2 s Due Thursday Oct 3 Introduction This assignment concerns the design of small-tip-angle 2D excitation pulses based on spiral k-space trajectories.

More information

Time Matters How Power Meters Measure Fast Signals

Time Matters How Power Meters Measure Fast Signals Time Matters How Power Meters Measure Fast Signals By Wolfgang Damm, Product Management Director, Wireless Telecom Group Power Measurements Modern wireless and cable transmission technologies, as well

More information

A. SPECIFIC AIMS: phase graph (EPG) algorithms to cover a wide range of MRI

A. SPECIFIC AIMS: phase graph (EPG) algorithms to cover a wide range of MRI A. SPECIFIC AIMS: A.. Overview: The promise of improved MRI results at high field strength is compromised by the difficulties encountered at high field, including: i) Non-uniform excitation, due to the

More information

MRI at a Glance. Catherine Westbrook. Blackwell Science

MRI at a Glance. Catherine Westbrook. Blackwell Science MRI at a Glance Catherine Westbrook Blackwell Science MRI at a Glance MRI at a Glance CATHERINE WESTBROOK MSC DCRR CTC Director of Training and Education Lodestone Patient Care Ltd Blackwell Science 2002

More information

Localization of microscale devices in vivo using addressable transmitters operated as magnetic spins

Localization of microscale devices in vivo using addressable transmitters operated as magnetic spins SUPPLEMENTARY INFORMATION Articles DOI: 10.1038/s41551-017-0129-2 In the format provided by the authors and unedited. Localization of microscale devices in vivo using addressable transmitters operated

More information

Multiplexed Echo Planar Imaging for Sub-Second Whole Brain FMRI and Fast Diffusion Imaging

Multiplexed Echo Planar Imaging for Sub-Second Whole Brain FMRI and Fast Diffusion Imaging Multiplexed Echo Planar Imaging for Sub-Second Whole Brain FMRI and Fast Diffusion Imaging David A. Feinberg 1,2,3 *, Steen Moeller 4, Stephen M. Smith 5, Edward Auerbach 4, Sudhir Ramanna 1,MattF. Glasser

More information

A Spread Spectrum Network Analyser

A Spread Spectrum Network Analyser A Spread Spectrum Network Analyser Author: Cornelis Jan Kikkert Associate Professor Head of Electrical and Computer Engineering James Cook University Townsville, Queensland, 4811 Phone 07-47814259 Fax

More information

Mapping the Flip Angle in Magnetic Resonance Imaging Using the Accelerated 3D Look-Locker Sequence

Mapping the Flip Angle in Magnetic Resonance Imaging Using the Accelerated 3D Look-Locker Sequence Western University Scholarship@Western Electronic Thesis and Dissertation Repository January 2011 Mapping the Flip Angle in Magnetic Resonance Imaging Using the Accelerated 3D Look-Locker Sequence Trevor

More information