A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers
|
|
- Alexandrina Hubbard
- 5 years ago
- Views:
Transcription
1 A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers Poster No.: C-3011 Congress: ECR 2010 Type: Topic: Authors: Scientific Exhibit Radiographers - Technical Aspects M.-L. Butler 1, P. C. Brennan 2, J. Last 1, L. Rainford 1 ; 1 Dublin/IE, 2 Sydney/AU Keywords: Keywords: DOI: exposure index, radiation dose, manufacturer comparison Technical aspects, Radiation safety /ecr2010/C-3011 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. Page 1 of 13
2 Purpose Medical radiation is the largest source of man-made radiation (1) and it is important to adopt the linear no-threshold model, and assume all exposures carry some risk. Computed radiography (CR) and direct digital radiography (DR) were introduced into clinical departments in the 1980s. The phenomenon of exposure creep in association with the advent of CR and DR has been largely documented (2). In traditional film-screen radiography, over or under exposure to the patient was immediately apparent. In CR and DR, image processing can compensate by up to 100% for under exposure and up to 500% for over exposure, and still produce a clinically acceptable image (3). Manufacturers have responded to this by offering an exposure index (EI) value to clinicians. This value is a measure of radiation dose to the detector and is displayed with the image to the clinician. Different manufacturers use varying types of EI as displayed in Table 1 (4). Manufacturer Agfa Kodak Philips Siemens Fuji Table 1: Exposure Index Values by manufacturer Exposure Indicator lgm- logarithm of the median of the histogram EI- exposure index EI- exposure index EXI- exposure index S (sensitivity) number In a clinical setting a wide range of factors can influence the EI value such as patient size, artifacts, source to image receptor distance, collimation, centering and IP plate size (2). Whilst the EI value has been promoted as a means of quality assurance testing for best clinical practice, its clinical consistency is not apparent in any recent literature. Page 2 of 13
3 The current study aims to assess the consistency of the EI value at constant conditions and to document any fluctuations that may occur. Page 3 of 13
4 Methods and Materials Materials Four manufacturers were used for comparison; Agfa Gaevert CR, Carestream CR, Philips Digital Diagnost DR and Siemens DR Four common, but contrasting, clinical examinations were chosen for the study; skull, chest, abdomen and hand (5). Two anthropomorphic phantoms were used to produce clinically relevant images and simulate typical examination conditions. RANDO Phantom was used for the chest, abdomen and skull projections and PIXY anthropomorphic phantom (Radiology Support Devices, USA) was used for the hand projection. Method For each examination, the phantoms were positioned in line with clinical practice and international guidelines (6). For each system and each projection, the phantom was exposed 20 times. The phantom was not moved between exposures. When using the CR systems, the same imaging plate was used, as was the same digitizer. The imaging plate was processed and the corresponding EI value for each exposure was recorded. Page 4 of 13
5 Page 5 of 13
6 Images for this section: Fig. 0: RANDO Phantom Page 6 of 13
7 Fig. 0: Positioning of Phantom School of Medicine and Medical Science, University College Dublin - Dublin/IE Page 7 of 13
8 Fig. 0: CR System School of Medicine and Medical Science, University College Dublin - Dublin/IE Page 8 of 13
9 Results The results for each manufacturer are displayed in tables 1-4. The ranges of EI values given represent the minimum and the maximum value recorded for the same examination at constant conditions. Examination EI Range Skull Chest Abdomen Hand Table 1: Kodak CR System Examination lgm Range Skull Chest Abdomen Hand Table 2: Agfa CR System Examination EI Range Skull 320 Chest 400 Abdomen 500 Hand 320 Table 3: Philips DR System Examination EI Range Page 9 of 13
10 Skull 487 Chest 475 Abdomen 309 Hand 583 Table 4: Siemens DR System Results show that the EI value is not consistent in CR, and thoroughly consistent in DR. This suggests that only varying clinical conditions such as exposure factor manipulation, collimation, SID and patient size will change the value in DR. The largest fluctuations were seen in the Agfa CR system. It should be noted that at increase of 0.3 in the lgm value, indicates a doubling of dose to the detector. Therefore, in the skull examination, the fluctuations seen would indicate to the clinician that double the exposure was given then is necessary according to manufacturers guidelines. Similarily, in the Kodak system, an increase of 300 EI indicates a doubling of exposure to the plate. In the chest examination, the EI value fluctuated by 100, which would suggest an increase or decrease of one third detector exposure. Page 10 of 13
11 Conclusion Given the evidence presented in this study, it is imperative that clinicians are aware of the fluctuations that are occurring in the EI value in CR systems outside of clinical conditions on which they have control. The challenge is posed to the clinician to be able to assess image quality outside of the EI value to ensure optimal clinical practice. The reliability of EI values as a feedback mechanism for CR is also uncertain Page 11 of 13
12 References 1. Mettler FA. Medical radiation exposure in the U.S. in 2006: preliminary results Health Phys 95(5): , (2008). 2. Willis, C. Computed Radiography: a higher dose? Pediatr Radiol 32: (2002). 3. Peters, S.E. and Brennan, P.B. Digital radiography: are the manufacturers' settings too high? Optimization of the Kodak digital radiography system with aid of the computed radiography dose index. Eur Radiol 12: (2002). 4. Willis CE. Strategies for dose reduction in ordinary radiographic examinations using CR and DR. Pediatr. Radiol 34 (Suppl 3): (2004). 5. Hart D and Wall BF. Radiation Exposure of the UK Population from Medical and Dental X-ray examinations. National Radiological Protection Board: Chilton (2002). 6. Swallow RA, Naylor E, Roebuck EJ, Whitney AS. Clark's positioning in radiography. 11 th edition. London: Heinmann Professional Publishing Ltd. (1986) Page 12 of 13
13 Personal Information Ms Butler is a Lecturer and Biological Imaging researcher in University College Dublin, Ireland. Her research interests include radiation dose studies, medical image perception and forensic radiography. Page 13 of 13
The impact of increasing SID on patient dose in pa abdomen imaging
The impact of increasing SID on patient dose in pa abdomen imaging Poster No.: C-1074 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific Exhibit T. Starc 1, N. Mekis 2 ; 1 Brezovica/SI, 2 Ljubljana/SI
More informationComputer applications, PACS, Instrumentation, Equipment, Technology assessment /ecr2012/C-0655
The novel medical diagnostic color monitor automatically distinguishing monochrome/color images and displaying these with their own optimal tone curves Poster No.: C-0655 Congress: ECR 2012 Type: Authors:
More informationImprovement of CT image quality with iterative reconstruction idose4
Improvement of CT image quality with iterative reconstruction idose4 Poster No.: C-0387 Congress: ECR 2014 Type: Scientific Exhibit Authors: M.-L. Olsson, K. Norrgren, M. Söderberg; Malmö/SE Keywords:
More informationEvaluation of no-grid radiography using the digital scattered x-ray removal processing
Evaluation of no-grid radiography using the digital scattered x-ray removal processing Poster No.: C-0416 Congress: ECR 2016 Type: Authors: Scientific Exhibit R. Suzuki 1, T. Goto 1, H. Ogawa 2, N. Amimoto
More informationDose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis
Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis Poster No.: C-0585 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Minehiro,
More informationStudies on reduction of exposure dose using digital scattered X-ray removal processing
Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.
More informationStudies on reduction of exposure dose using digital scattered X-ray removal processing
Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.
More informationInvestigation of Effective DQE (edqe) parameters for a flat panel detector
Investigation of Effective DQE (edqe) parameters for a flat panel detector Poster No.: C-1892 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit D. Bor 1, S. Cubukcu 1, A. Yalcin 1, O.
More informationNovel cassette-sized, flat-panel digital radiography (DR) system: Initial clinical and workflow results versus computed radiography (CR)
Novel cassette-sized, flat-panel digital radiography (DR) system: Initial clinical and workflow results versus computed radiography (CR) Poster No.: C-3027 Congress: ECR 2010 Type: Scientific Exhibit Topic:
More informationVisualization of sources of scattered radiation from x-ray equipment used for interventional radiology
Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology Poster No.: C-1190 Congress: ECR 2011 Type: Scientific Exhibit Authors: K. Chida, T. Takahashi, D.
More informationAnalysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging
Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging Award: Magna Cum Laude Poster No.: C-1988 Congress: ECR 2014 Type: Scientific Exhibit Authors: T. Hamaguchi,
More informationFundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging
Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging Poster No.: C-0709 Congress: ECR 2014 Type: Scientific
More informationDetermining acceptance levels for automatic daily image quality control in magnetic resonance imaging
Determining acceptance levels for automatic daily image quality control in magnetic resonance imaging Poster No.: C-1125 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific Exhibit J. I. Peltonen,
More informationInfluence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck
Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Poster No.: C-2205 Congress: ECR 2012 Type: Scientific Paper
More informationCorrection of the local intensity nonuniformity artifact in high field MRI
Correction of the local intensity nonuniformity artifact in high field MRI Poster No.: C-0346 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Paper S. Kai, S. Kumazawa, H. Yabuuchi, F. Toyofuku;
More informationLumbar disc height assessment: a comparative study between EOS and digital radiography
Lumbar disc height assessment: a comparative study between EOS and digital radiography Poster No.: C-1178 Congress: ECR 2012 Type: Scientific Paper Authors: V. Freire, M. Benhamou, F. Rannou, S. Poiraudeau,
More informationA comparison between medical-grade liquid crystal display (LCD) and ipad color imaging
A comparison between medical-grade liquid crystal display (LCD) and ipad color imaging Poster No.: C-1377 Congress: ECR 2014 Type: Scientific Exhibit Authors: K. Yoshimura 1, K. Shibata 2, T. Nihashi 1,
More informationWorldwide practice of breast MRI: insights from the MIPA study applications - the MIPA study group
Worldwide practice of breast MRI: insights from the MIPA study applications - the MIPA study group Poster No.: B-0757 Congress: ECR 2015 Type: Scientific Paper Authors: G. Di Leo, R. M. Trimboli, I. Ioan,
More informationAim. Images for this section: Page 2 of 13
Changes in CT number of high atomic number materials with field of view when using an extended CT number to electron density curve and a metal artifact reduction reconstruction algorithm Poster No.: R-0094
More informationMulti-slice computed tomography analysis of bullet trajectory in forensic investigation.
Multi-slice computed tomography analysis of bullet trajectory in forensic investigation. Poster No.: C-0510 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Usui, Y. Kawasumi, Y. Hosokai, A. Nakajima,
More informationPractical guidelines for color calibration and quality assurance of medical displays
Practical guidelines for color calibration and quality assurance of medical displays Poster No.: C-1140 Congress: ECR 2017 Type: Educational Exhibit Authors: T. Kimpe, J. Rostang, G. Van Hoey, A. Xthona
More informationClear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE)
Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE) Poster No.: C-2459 Congress: ECR 2010 Type: Scientific Exhibit Topic: Neuro Authors: T. Yoneda,
More informationDifference in signal-to-noise-ratio (SNR) from vertical to horizontal scanner position using a 0,25 Tesla Weightbearing
Difference in signal-to-noise-ratio (SNR) from vertical to horizontal scanner position using a 0,25 Tesla Weightbearing scanner Poster No.: C-0672 Congress: ECR 2014 Type: Scientific Exhibit Authors: F.
More informationPROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution
PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution Poster No.: C-1280 Congress: ECR 2011 Type: Educational Exhibit Authors: A. Monteiro, A.
More informationSuppression of metal artifacts using image-based monoenergetic DECT imaging
Suppression of metal artifacts using image-based monoenergetic DECT imaging Poster No.: C-0519 Congress: ECR 2011 Type: Scientific Paper Authors: B. Krauss, B. Schmidt, M. Sedlmair, T. Flohr; Forchheim/DE
More informationOptimal speed for dual-energy subtraction neck imaging with flat-panel detector radiography
Optimal speed for dual-energy subtraction neck imaging with flat-panel detector radiography Poster No.: C-2938 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology Authors: H. Machida
More informationPROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution
PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution Poster No.: C-1280 Congress: ECR 2011 Type: Educational Exhibit Authors: A. Monteiro, A.
More informationComparison of high-resolution C-arm cone-beam CT imaging and conventional 3D rotational angiography for the better microvascular visualization
Comparison of high-resolution C-arm cone-beam CT imaging and conventional 3D rotational angiography for the better microvascular visualization Poster No.: C-1040 Congress: ECR 2016 Type: Scientific Exhibit
More informationDevelopment of new dosimeter for measuring dose distribution in CT
Development of new dosimeter for measuring dose distribution in CT Poster No.: C-2925 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology - Without Subtopic Authors: Y. Muramatsu, K.
More informationAsk EuroSafe Imaging Tips & Tricks. Paediatric Imaging Working Group. Dose Management in Digital Radiography
Ask EuroSafe Imaging Tips & Tricks Paediatric Imaging Working Group Dose Management in Digital Radiography Raija Seuri (HUS Medical Imaging Center, FI) Cristina Almeida (Centro Hospitalar de Lisboa Central,
More information3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development
Teaching Radiographic Technique in a Digital Imaging Paradigm Objectives 1. Discuss the historical development of digital imaging. Dawn Couch Moore, M.M.Sc., RT(R) Assistant Professor and Director Emory
More informationCOMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR
This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura New Technology https://www.youtube.com/watch?v=ptkzznazb 7U COMPUTED
More informationPractical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S.
Practical Medical Physics Session: TG-151 Dose Monitoring August 5, 2013 Katie Hulme, M.S. Digital Imaging and Dose Creep Images courtesy of Agfa Healthcare Under-Exposed Over-Exposed Freedman et al.,
More informationThe role of CT in restoration of a very rare Christ wooden statue: a fundamental contribution.
The role of CT in restoration of a very rare Christ wooden statue: a fundamental contribution. Poster No.: C-1514 Congress: ECR 2015 Type: Scientific Exhibit Authors: M. Disaro', P. Sartori; Venice/IT
More informationExposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?
Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago
More information2217 US Highway 70 East Garner, NC Main: Fax:
Viztek is committed to providing the highest image quality possible in our CR & DR product lines. There are several factors that directly affect the overall quality of CR & DR based images. The eposure
More informationQuality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals
DOI 10.1007/s00247-011-2290-1 ORIGINAL ARTICLE Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals Mervyn D. Cohen & Richard Markowitz
More informationThe feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE
The feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE Poster No.: C-0022 Congress: ECR 2010 Type: Scientific Exhibit Topic: Abdominal Viscera
More informationNew Exposure Indicators for Digital Radiography Simplified for Radiologists and Technologists
Medical Physics and Informatics Technical Innovation Don et al. New Simplified Exposure Indicators Medical Physics and Informatics Technical Innovation Steven Don 1 ruce R. Whiting 2 Lois Jo Rutz 3 ruce
More informationDisclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography
Current Implementation Status of IEC Standard 62494-1: Exposure Index (EI) for Digital Radiography July 31, 2017 Ryan Fisher, PhD, DABR Katie Hulme, MS, DABR None Disclosures Outline Review of IEC Standard
More informationDigital Imaging Considerations Computed Radiography
Digital Imaging Considerations Digital Radiography Computed Radiography o Cassette based Direct or Indirect Digital Radiography o Cassetteless Computed Radiography 1 CR Image Acquisition Most like conventional
More informationThe Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System
The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System Manus
More informationVoice recognition versus transcriptionist
Voice recognition versus transcriptionist Poster No.: C-0263 Congress: ECR 2010 Type: Audit/Professional Issues Topic: Audit/Professional Issues Authors: R. H. Strahan; Clayton South VIC/AU Keywords: Voice
More informationOutline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control
Understanding Digital Modalities: Image Quality and Dose S. Jeff Shepard, M.S. University of Texas M. D. Anderson Cancer Center Houston, Texas Special Acknowledgement: Stephen K. Thompson, M.S. William
More informationComparison of computed radiography and filmõscreen combination using a contrast-detail phantom
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 1, WINTER 2003 Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom Z. F. Lu,* E. L. Nickoloff, J.
More informationHalf value layer and AEC receptor dose compliance survey in Estonia
Half value layer and AEC receptor dose compliance survey in Estonia K. Kepler, A. Vladimirov Training Centre of Medical Physics, University of Tartu Testing Centre of the University of Tartu, Estonia E-mail:
More informationLearning Objectives: What s my motivation? (unknown screen actor) Workshop Overview
Practical Medical Physics Adapting Traditional Clinical Medical Physics to Digital Radiography Charles E. Willis, Ph.D., DABR Associate Professor Department of Imaging Physics The University of Texas M.D.
More informationAcquisition, Processing and Display
Acquisition, Processing and Display Terri L. Fauber, R.T. (R)(M) Department of Radiation Sciences School of Allied Health Professions Virginia Commonwealth University Topics Image Characteristics Image
More informationHistory of digital imaging
CR/QA RADCHEX History of digital imaging Early, crude digital detectors were developed in the 1970 s Image quality was problematic Processing time of digital images was untenable Viewing, transfer and
More informationRAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II
RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II APPROVED 12/O2/2011 EFFECTIVE SPRING 2013-14 Prefix & Number RAD 150 Course Title: Radiologic Exposure Technique II & Lab Purpose of this submission: New Change/Updated
More informationWhile digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.
In press 2004 1 2 Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Digital radiology An appropriate analogy that is easy for most people to understand
More informationof sufficient quality and quantity
of sufficient quality and quantity The patient s body attenuates the beam as it passes though the body More energy is deposited in organs located near the entry of the beam than near the exit of the beam
More information- KiloVoltage. Technique 101: Getting Back to Basics
Why do I need to know technique? Technique 101: Getting Back to Basics Presented by: Thomas G. Sandridge, M.S., M.Ed., R.T.(R) Program Director Northwestern Memorial Hospital School of Radiography Chicago,
More informationDigital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of
Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of radiation to the population due to Medical Imaging
More informationAppropriate Inspection Distance of Digital X-Ray Imaging Equipment for Diagnosis
Indian Journal of Science and Technology Vol 8(S8), 380-386, April 2015 ISSN (Print) : 0974-6846 ISSN (Online) : 0974-5645 DOI: 10.17485/ijst/2015/v8iS8/70528 Appropriate Inspection Distance of Digital
More informationTeaching Digital Radiography and Fluoroscopic Radiation Protection
Teaching Digital Radiography and Fluoroscopic Radiation Protection WCEC 20 th Student Educator Radiographer Conference Dennis Bowman, RT(R), CRT (R)(F) Community Hospital of the Monterey Peninsula (CHOMP)
More informationThe effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography
Journal of Physics: Conference Series OPEN ACCESS The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography To cite this article: N A A Daud et al 2014 J. Phys.:
More informationIranian Journal of Medical Physics
Iranian Journal of Medical Physics ijmp.mums.ac.ir A Phantom Study for the Optimization of Image Quality and Radiation Dose for Common Radiographic Examinations in Digital Radiography Soo-Foon Moey 1*,
More informationA Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations. Shirley Weddle, RT(R)(M), CIIP, BBA
A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations Shirley Weddle, RT(R)(M), CIIP, BBA OBJECTIVES Define Computed Radiography (CR) Discuss CR vendor
More informationBeam-Restricting Devices
Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the
More informationEvaluation of magnetic resonance acoustic noise in 1.5 and 3 Tesla scanners
Evaluation of magnetic resonance acoustic noise in 1.5 and 3 Tesla scanners Poster No.: B-1014 Congress: ECR 2016 Type: Scientific Paper Authors: V. M. F. Silva, I. M. Ramos, J. Moreira, M. Marques; Porto/PT
More informationAttikon, Rimini 1, , Athens, Greece , Athens, Greece , Athens, Greece
Radiation Protection Dosimetry (2005), Vol. 117, No. 1 3, pp. 291 297 doi:10.1093/rpd/nci742 Advance Access published on February 7, 2006 QUALITY ASSURANCE (QA) PROCEDURES FOR SOFTWARE: EVALUATION OF AN
More informationRADIOGRAPHIC EXPOSURE
RADIOGRAPHIC EXPOSURE Receptor Exposure Receptor Exposure the that interacts with the receptor. Computed Radiography ( ) requires a. Direct Digital Radiography (DR) requires a. Exposure Indicators Exposure
More informationExposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations
Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations S. Baechler 1, P. Monnin 1, A. Aroua 1, J.F. Valley 1, M. Perrier, P. Trueb 3, F.R. Verdun 1 1 University
More informationFeatures and Weaknesses of Phantoms for CR/DR System Testing
Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal
More informationRing-Down versus Comet Tail: Two artifacts uncovered
Ring-Down versus Comet Tail: Two artifacts uncovered Poster No.: C-2288 Congress: ECR 2013 Type: Educational Exhibit Authors: S. Méndez Garrido 1, J. Arce Duran 2, A. Talegón Meléndez 1 ; 1 2 Sevilla/ES,
More informationX-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto
X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring
More informationX-RAYS - NO UNAUTHORISED ENTRY
Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are
More informationEuropean Journal of Radiology
European Journal of Radiology 72 (2009) 202 208 Contents lists available at ScienceDirect European Journal of Radiology journal homepage: www.elsevier.com/locate/ejrad Digital radiography: The balance
More informationIMAGE QUALITY AND DOSE MANAGEMENT IN DIGITAL RADIOGRAPHY: A NEW PARADIGM FOR OPTIMISATION H. P. Busch 1 and K. Faulkner 2,
Radiation Protection Dosimetry (2005), Vol. 117, No. 1 3, pp. 143 147 doi:10.1093/rpd/nci728 Advance Access published on February 3, 2006 IMAGE QUALITY AND DOSE MANAGEMENT IN DIGITAL RADIOGRAPHY: A NEW
More informationX-RAY MEDICAL EQUIPMENT
X-RAY MEDICAL EQUIPMENT CHEST RADIOGRAPHY GENERAL RADIOGRAPHY & FLUOROSCOPY RADIOTHERAPY MOBILE HEALTHCARE MAMMOGRAPHY MAMMOSCAN FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM Biopsy Attachment џ MAMMOSCAN an ADANI
More informationComparison of new digital mammography systems: physical characterisation and image quality evaluation
Comparison of new digital mammography systems: physical characterisation and image quality evaluation Poster No.: C-2321 Congress: ECR 2011 Type: Scientific Exhibit Authors: N. Oberhofer; Bolzano/IT DOI:
More informationInstant DR in Jordan
Hashemite University leads the way with first Instant DR in Jordan DR Retrofit supports research and education goals of the Faculty of Allied Health Sciences, while enhancing care for staff and students
More informationOpen. the Digitized world. Fuji Computed Radiography
Open the Digitized world Fuji Computed Radiography If just one of these applies to you... Managing developing fluid is hard and darkroom work is a hassle... Images are not stable... Isn t digitalization
More informationddr Compact Series Setting a new benchmark in digital radiography.
ddr Compact Series Setting a new benchmark in digital radiography. ddrcompact When productivity and exceptional value come together. With the introduction of its newest DR system, the ddrcompact, Swissray
More informationLuminos RF Classic. Where value meets performance.
Luminos RF Classic Where value meets performance www.siemens.com/healthcare What s good value in fluoroscopy? That s easy. Luminos RF Classic. 2 Whether for its handling convenience, outstanding image
More information10/3/2012. Study Harder
This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen
More informationImage quality evaluation of turbo-spin echo diffusion weighted image (TSE-DWI) : A phantom study
Image quality evaluation of turbo-spin echo diffusion weighted image (TSE-DWI) : A phantom study Poster No.: C-0631 Congress: ECR 2016 Type: Scientific Exhibit Authors: T. Yoshida, A. Urikura, K. Shirata,
More informationCR Basics and FAQ. Overview. Historical Perspective
Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors
More informationDigital radiography (DR) post processing techniques for pediatric radiology
Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions
More information10/26/2015. Study Harder
This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen
More informationDose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp
Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division
More informationAutomated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London
Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients
More informationNuclear Associates EZ CR-DIN Phantoms
Nuclear Associates 07-605-7777 EZ CR-DIN Phantoms Users Manual August 2006 Manual No. 07-605-7777-1 Rev. 4 2006 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks
More informationLearning Objectives. Outline. Getting Started with CR. Converting the Radiology Department from Film-Screen to Digital: Making the Transition
Converting the Radiology Department from Film-Screen to Digital: Making the Transition S. Jeff Shepard, MS, DABR University of Texas M. D. Anderson Cancer Center Houston, Texas jshepard@di.mdacc.tmc.edu
More informationIntroduction. Sam R. Kottamasu Lawrence R. Kuhns
Pediatr Radiol (1997) 27: 119 123 Springer-Verlag 1997 Sam R. Kottamasu Lawrence R. Kuhns Musculoskeletal computed radiography in children: scatter reduction and improvement in bony trabecular sharpness
More informationEssentials of Digital Imaging
Essentials of Digital Imaging Module 2 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 2 Processing 1. ASRT Animation 2. Welcome Welcome to Essentials of Digital Imaging
More informationInvestigation of the line-pair pattern method for evaluating mammographic focal spot performance
Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann
More information10 kvp rule an anthropomorphic pelvis phantom imaging study using a CR system : impact on image quality and effective dose using AEC and manual mode
10 kvp rule an anthropomorphic pelvis phantom imaging study using a CR system : impact on image quality and effective dose using AEC and manual mode Lanca, L, Franco, L, Ahmed, A and Hogg, P http://dx.doi.org/10.1016/j.radi.2014.04.007
More informationDedicated Veterinary Imaging Solutions Digital, CR and Analog Imaging Solutions for any size patient and any size budget.
by Dedicated Veterinary Imaging Solutions Digital, CR and Analog Imaging Solutions for any size patient and any size budget. Serving the Veterinary Profession for Over 75 Years. ... We See Things Differently
More informationRedefining Ergonomics
Samsung Electronics Co., Ltd. inspires the world and shapes the future with transformative ideas and technologies, redefining the worlds of TVs, smartphones, wearable devices, tablets, cameras, digital
More informationImage Quality Artifacts in Digital Imaging
MAHIDOL UNIVERSITY Wisdom of the Land Image Quality Artifacts in Digital Imaging Napapong Pongnapang, Ph.D. Department of Radiological Technology Faculty of Medical Technology Mahidol University, Bangkok,
More informationSYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology
CODE: RADT 156 INSTITUTE: Health Science TITLE: Equipment Operation I DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of equipment operation and maintenance of radiographic
More informationDo you have any other questions? Please call us at (Toll Free) or , or
INSTRUCTIONS Read the appropriate course/ textbook. This is an open book test. A score of 75% or higher is needed to receive CE credit. You will have a maximum of three attempts to pass this course. Please
More informationDigital Image Management: the Basics
Digital Image Management: the Basics Napapong Pongnapang, Ph.D. Department of Radiological Technology Faculty of Medical Technology Mahidol University Outline From screen/film to digital radiography PACS/Tele
More informationOptimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1
Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1 Gham Hur, M.D., Yoon Joon Hwang, M.D., Soon Joo Cha, M.D., Su Young Kim, M.D., Yong Hoon Kim, M.D.
More informationArtefacts found in computed radiography
The British Journal of Radiology, 74 (2001), 195 202 E 2001 The British Institute of Radiology Pictorial review Artefacts found in computed radiography L J CESAR, RT(R)(QM), B A SCHUELER, PhD, F E ZINK,
More informationY11-DR Digital Radiography (DR) Image Quality
Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures
More informationVersion 1.0. TechnicVR. Student Guide
Version 1.0 TechnicVR s h a d e r w a r e. c o m Student Guide TechnicVR s h a d e r w a r e. c o m Student Guide shaderware 2008 PO Box 103 Saltburn Cleveland TS12 1WP w w w. s h a d e r w a r e. c o
More information9/10/2012. Computed Radiography Chapter 3 Physics and Technology. What is Computed Radiography?
Computed Radiography Chapter 3 Physics and Technology This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Today s Humor:
More information