European Journal of Radiology

Size: px
Start display at page:

Download "European Journal of Radiology"

Transcription

1 European Journal of Radiology 72 (2009) Contents lists available at ScienceDirect European Journal of Radiology journal homepage: Digital radiography: The balance between image quality and required radiation dose Martin Uffmann a,, Cornelia Schaefer-Prokop b a Medical University Vienna, Dept. of Radiology, Waehringer Guertel 18-20, A-1090 Vienna, Austria b Academic Medical Center (AMC), Meibergdreef 9, 1105 AZ Amsterdam, Netherlands article info abstract Article history: Received 7 May 2009 Accepted 7 May 2009 Keywords: Digital radiography Computed radiography Image quality Radiation exposure Dose indicator Although the transition from conventional screen-film imaging to digital image acquisition has been almost completed during the last couple of years, examination parameters, such as tube voltage, tube current, and filtration have been adopted from screen-film technology without further adjustments. Digital systems, however, are characterised by their flexibility: the acquisition dose can be reduced at the expense of image quality and vice versa. The imaging parameters must be optimised according to the best performance of a particular system. The traditional means of dose containment, such as positioning and collimation, are as valid for digital techniques as they were for conventional techniques. Digital techniques increasingly offer options for dose reduction. At the same time, there is a risk of substantially increasing the patient dose, possibly unawares, due to the lack of visual control. Therefore, implementation of dose indicators and dose monitoring is mandatory for digital radiography. The use of image quality classes according to the dose requirements of given clinical indications are a further step toward modern radiation protection Published by Elsevier Ireland Ltd. 1. Introduction The technological development of digital radiography (DR) and computed radiography (CR) is occurring at the same pace as that for cross-sectional imaging modalities like CT or MR imaging. The recent advent of CR detector technology, such as line scanning, dual reading, and needle-crystalline detector materials, have alleviated the differences in DR detectors with regard to dose requirements. Most modern CR and DR systems now effectively offer substantial patient dose reduction compared to screen-film radiography. Unfortunately, the reverse is also possible. There is the risk of substantially increasing the patient dose, possibly without being aware of it, or of decreasing diagnostic information because of deteriorating image quality due to inadequate image processing or suboptimal image display. In the following article, options for radiologists and physicists to assess and control dose and image quality are summarised, and strategies for optimal adjustment of digital systems are discussed. Corresponding author. Tel.: ; fax: address: Martin.uffmann@meduniwien.ac.at (M. Uffmann). 2. Digital detectors compared to conventional screen-film systems Although based on different technical principles, the digital detectors have a great deal in common with screen-film radiography: Image acquisition, image presentation, and image archiving used to be bundled on a single film sheet. With digital detectors, these main functions of radiography have been uncoupled, which is a prerequisite condition for PACS. The dynamic range of digital detectors is about 400-fold compared to film-screen (Fig. 1a, b). The inverse correlation between dose and image contrast is eliminated with digital systems. Image contrast and brightness can be optimised independently. As film blackening at higher doses does not exist with digital systems, there is the hazard of dose creep, meaning an unnoticed increase in exposure over time when using digital systems with manual tube settings. A decrease in diagnostic information through deteriorating image quality due to inadequate image processing or suboptimal image display. The latter includes the regular assessment of adequate softcopy display functions X/$ see front matter 2009 Published by Elsevier Ireland Ltd. doi: /j.ejrad

2 M. Uffmann, C. Schaefer-Prokop / European Journal of Radiology 72 (2009) Fig. 1. a, b: In a phantom series with stepwise progression of dose, the difference in dynamic range between a screen-film system (a) and a flat-panel detector (b) is illustrated (courtesy of Ulrich Neitzel, Philips Medical Systems, Hamburg). 3. Dose requirements and image quality 3.1. ALARA With increasing awareness of the need for radiation protection, a paradigm shift can be observed from the principle of image quality as good as possible to image quality as good as needed. The radiation dose to patients should be as low as reasonably achievable (ALARA) while still providing image quality adequate to enable an accurate diagnosis [1,2]. ALARA does not necessarily mean the lowest radiation dose, nor, when implemented, does it result in the least desirable radiographic image [3]. What, indeed, constitutes adequate image quality is still open for discussion for the various imaging tasks. There is a multitude of studies in the literature comparing the performance of one system with another reference system to define the amount of possible dose reduction that would still achieve an image quality equivalent to that provided by the acknowledged reference. Using this approach, it is possible to survey parameters, such as the detection of artificial lesions or the semi-quantitative assessment of subjective image impression, as a surrogate for image quality and relate these parameters to a reference of dose. To define, however, the minimum level of image quality needed to reliably make a certain type of diagnosis is much more difficult. Individually defining the minimal dose to reliably answer a specific diagnostic question in a prospective manner seems to be impossible, given the vast variety of patient-related and disease-related conditions and the workflow for radiographic examinations. Reduction of patient dose according to the ALARA principle is not only a question of selecting the right detector, but also requires the optimisation of the whole imaging chain and the selection of appropriate imaging parameters Image quality classes To implement the ALARA principle, a number of international work groups introduced the concept of image quality classes (Tables 1 and 2) [1,2,4]. Three levels of image quality (high, medium, and low) and, accordingly, three dose levels (corresponding to speed classes 400, 800, and 1600) were suggested, dependent on the demands of the diagnostic question. Such a pre-examination classification of required image quality would mean that existing

3 204 M. Uffmann, C. Schaefer-Prokop / European Journal of Radiology 72 (2009) Table 1 Image quality classes. Examples for different levels dependent on given clinical indications. Image quality High Medium Low Clinical indication Sources: [1,2,5] and - Primary bone tumour - Non-displaced fracture - Lumbar spine in two projections in patients with chronic back pain with no pointers to infection or neoplasm - Control of a known displaced fracture - Control after metal implantation for osteosynthesis - Follow-up of pneumonia in adults - Follow-up in longitudinal studies, e.g., in patients with scoliosis Table 2 Relationship between image quality classes and dose requirements of different radiography systems [4]. Image quality class High Medium Low DR (400) DR (800) DR (1600) CR (200/400) CR (400) CR (800) Film-screen system (200) Film-screen system (400) Film-screen system (800) referral guidelines would require an additional parameter, such as image quality class, although it has to be clearly stated that the responsibility for such a classification should be in the hands of the radiologists. Based on the different dose efficiency of existing digital radiographic equipment, ALARA would also mean that, for the same clinical question, different exposure parameters must be applied, dependent on the radiographic equipment used. Evaluations of phantom images showed that a medium class image quality provided by a film/screen image with a speed class of 400 was achievable by a storage phosphor system (CR systems using a powder storage phosphor plate with a single-sided read-out) with doses equivalent of speed 200 and 400 and a flat panel system (DR system using a CsI/TFT detector) at a dose equivalent of speed 1600 or 25% of the dose [5]. A desirable option to optimally manage this process in a clinical routine in the future may be a kind of examination parameter database that is integrated in the system. For various examination types and image quality classes, the database offers appropriate pre-settings of manual parameters or AEC exposures [6]. Since the responsibility for such a classification should be in the hands of the radiologists, it would require reorganisation of the workflow because, unlike CT and fluoroscopic examinations, ordinary radiographic examinations are usually performed without the radiologist s knowledge or input. The image quality class must be selected beforehand, which requires a close communication between the radiologist and his clinical counterpart Strategies for dose containment In addition to image quality classes, there are a variety of measures before and during the examination that allow a net reduction in patient dose Adaptation of technical acquisition parameters to digital equipment In general, it is important to consider that each change in one of the segments of the imaging chain (generator, tube, filtration, grid, detector, workstation, PACS, monitor, or print-out) will require adaptation of the exposure parameters, and thus, of the threshold values of the automatic exposure control system (AEC). Similarly, as radiologists change the CT acquisition protocols when changing from a 16-slice-scanner to a 64-slice-scanner, they have to adapt the exposure parameters when purchasing new CR or DR equipment. In practice, this process should be integrated in the certification process before starting to work with the new equipment Allocation of examinations to the most suitable radiographic system When radiologists have multiple systems available, knowledge of the different dose requirements of the systems should be reflected in the choice of system according to optimal suitability. Examples include the following: the CsI/TFT flat panel systems produce chest radiographs of higher quality at lower dose levels compared to an a-se/tft-based DR system; charged-coupleddevice (CCD) equipment is better suited for small field applications than chest radiography; and a dual-read-out CR system works more dose efficiently that a single-read-out CR system. The aim of this allocation should be to reduce the net dose for the entire patient group Preservation of dose-saving principles validated for conventional techniques Guidelines implemented for conventional radiography, including appropriate collimation, appropriate source-to-image distance (SID), focal spot size, and patient positioning, are as valid for digital techniques as they were for conventional techniques. This is especially important in an intensive care unit or emergency department setting where many of the above-mentioned parameters are set manually. Unfortunately, there is a tendency to handle these principles less precisely, based on the fact that digital technique is more tolerant to dose variations and offers more options to retrospectively rescue image quality by processing. In conventional screen-film radiography, inadequate images were easily identified. The majority of these images were subsequently re-taken. With digital techniques, however, image processing can compensate for acquisition errors. Under certain circumstances, the source of error is invisible on the processed radiograph. In particular, an unnecessarily large collimation can be masked by electronic shutters. Appropriate collimation of the X-ray beam is important for both radiation protection and image quality in a DR/CR setting [7]. The optimal collimation area depends on the individual patient and is the responsibility of the radiographer, taking into account the patient body size, the diagnostic question, and the individual requirements of the examination type. This becomes especially important when using systems that are not cassette-based (some CR systems, all DR systems) and that use large-area detectors. Continuous training, in-services, oversight, and feedback represent valuable tools to ensure the proper use and radiographic technique in accordance with the ALARA principle Optimisation of tube voltage and beam filtration Traditionally, the limited dynamic range of conventional film/screen systems required the use of high kilovoltage settings to penetrate high attenuating areas, such as the mediastinum, while still maintaining an acceptable level of image contrast in the lungs. In general, digital images are acquired with the same kvp settings as those used for conventional film-screen. This approach, however, was recently questioned. Digital systems have redefined the traditional relationship between tube potential and image contrast because image processing allows the contrast and density of an image to be optimised independently. A second point to challenge the traditional high kvp technique for digital systems is related to the fact that all digital detector mediums have to a different extent, depending on their absorption characteristics higher dose efficiency (DQE) at lower kvp ranges. This,

4 M. Uffmann, C. Schaefer-Prokop / European Journal of Radiology 72 (2009) theoretically, can be translated to an improved signal-to-noise. [8,9]. Most studies were carried out by physicists and were based on SNR measurements or contrast detail experiments. There are few studies that evaluate the effects of low kvp settings in a clinical environment [10,11]. They uniformly describe significantly improved image quality for posterior anterior chest radiographs acquired with 90 kvp (without an increase in effective patient dose). In a recently published study, a reduction of the tube potential proved to be beneficial for both dose reduction and image quality in lumbar spine radiography [12]. Some authors recommend additional beam filtering rather than lowering the tube voltage [13,14]. By using additional filtration to harden the beam, e.g., thin layers of copper, the entrance dose can be lowered by up to 40% for certain body parts [15]. Standard tube filtration in diagnostic radiology, as required by regulations, is 2.5 mm of the aluminium (AL) equivalent. Additional filtration in the X-ray beam can be used to remove the low energy part of the tube spectrum, which is completely absorbed in the patient without contributing to image quality. While additional filters of 1 mm AL plus 0.1 mm or 0.2 mm copper are suggested, and are common practice for paediatric radiography in many countries (parts of the European guidelines for paediatric imaging), this technique is not applicable to imaging of adults due to the considerable prolongation of exposure time (>50%), most likely in many patients beyond the recommended threshold of 20 ms [13,16] Anti-scatter grids Anti-scatter grids are generally used for areas with high absorption and a high level of scattered radiation that leads to deterioration of image quality with respect to signal-to-noise ratio and contrast. Use of an anti-scatter grid is inevitably associated with a higher acquisition dose (factor 2 or 3) compared to the same image obtained without an anti-scatter grid. Based on the fact that CR and DR systems (with the exception of slit-beam technology) represent area detectors that are vulnerable to scatter effects, general practice is that anti-scatter grids are used in applications similar to those in conventional radiography (e.g., upright chest radiography, radiographs of the spine or pelvis, etc.). The use of anti-scatter grids for bedside chest radiography varies among institutions. In favour of anti-scatter grids is the fact that the lower K-absorption edge of the detector material of both CR and DR makes it more sensitive for the absorption of low-energy scattered radiation compared to film-screen systems. On the other hand, the grid can be omitted if the tube voltage is lowered, and therefore, scatter radiation is reduced. The latter is especially applicable for small volumes (e.g., paediatric applications) [3]. With adequate image processing, a sufficient image quality is achievable for most patients, considering the diagnostic requirements in supine chest radiography (Fig. 2 a c). Nevertheless, for patients with a body mass index clearly above average, image acquisition with high voltage and use of a grid is still required Optimising image processing technology There is no doubt that the processing of digital images plays a major role in successfully performing the medical imaging task. Three aspects must be taken into account: (1) The overall impression of a typical radiograph (e.g., chest radiograph) should be similar to conventional radiography performed with screen-film systems in order to recognise the typical features acquired during radiology training and to maintain the option for universal interpretation. (2) There should be a certain consistency in image processing to determine whether something is normal, and for interpretation of studies longitudinally. (3) The modern multi-frequency processing capabilities enhance image contrast differently in different anatomic regions and differently for different image structures. In reader studies, images enhanced with elaborate processing tools are greatly preferred [17]. The question of whether a gain in image quality by processing can be utilised for dose reduction remains to be answered. As image noise can be both amplified and suppressed by processing, the link between image processing and dose requirement is evident Controlling patient dose The inverse correlation between dose and image contrast is eliminated with digital systems. Therefore, film blackening as an indicator of overexposure, no longer exists. Even a 10-fold overexposure cannot be recognised as a too black digital radiograph. In digital radiography, there is a reciprocal relationship between dose and signal-to-noise ratio: a lower acquisition dose is associated with increased image noise and vice versa. The problem is that the subjective visual assessment of image noise is quite insensitive. In addition, increased acquisition dose will be rewarded by increased signal-to-noise. Thus, images will be hardly ever rejected by the radiologist when overexposed. This may consequently lead to a small but continuous increase in the amount of acquisition dose when exposure parameters are set manually. Weatherburn et al. compared acquisition doses in 269 patients admitted to the intensive care unit [18]. Patients were randomly assigned either to CR or conventional radiographic imaging of the chest. The authors found significantly higher surface entry doses, entrance surface air kerma (median 0.21 mgy vs mgy), and higher effective doses (median msv vs msv) for CR compared to conventional radiography. This underscores the supposition that when in doubt the technologists tend to use higher acquisition levels in CR than in conventional radiography Image noise In digital radiography, image noise is inversely related to the amount of detector radiation dose. Thus, visual assessment of image noise is potentially a control mechanism. However, the human visual control mechanism is not only very subjective, but also quite insensitive. Observers complained about the noise in CR images only when they were exposed to considerably less than 50% of the appropriate level [19]. By the time a too high noise level is visualised by the radiologist, there is already an increased risk for loss of diagnostic information Speed-class system Given the very limited visual control and the arbitrarily allocable amount of dose, the question arises of how dose can be controlled in digital systems. Film/screen combinations were characterised by a limited dynamic range, and thus, required a certain (rather limited) range of radiation exposure to obtain an image of optimal diagnostic quality. This dose (also considered detector exposure) was traditionally described in speed or speed class. Multiple studies comparing the performance of film/screen and digital systems described the dose used to acquire film or digital images with speed values (Table 2). This might be understandable in the context of studies that compared the performance of the two

5 206 M. Uffmann, C. Schaefer-Prokop / European Journal of Radiology 72 (2009) Fig. 2. a c: Dose reduction by optimising radiographic technique in supine chest radiography. The old standard included a grid. Parameters were set at 125 kvp and 1.25 mas (a). Substantial dose reduction was achieved by removing the grid and lowering the tube voltage to 90 kvp at 1.0 mas (b). Further dose reduction is feasible by using a needle structured CR system (90 kvp, 0.5 mas, no grid was used) (c). radiographic techniques. However, it is important to understand that the measure of speed is, in fact, not applicable to the detector dose of digital radiographs. This is illustrated by the following. Radiographic speed is formally defined as the inverse of the exposure necessary to produce a net film density (above base and fog) of an optical density (OD) equal to 1. The density of a CR or DR image, however, is entirely arbitrary and dependent on the processing and not completely related to the dose. The film speed is primarily determined by the composition and thickness of the screen and also describes the spatial resolution performance of the film/screen system. Yet, in CR/DR units, the system resolution is determined by other factors. For example, in CR, detector (phosphor) thickness and the laser spot size are determinants of resolution, and, for both CR and DR, pixel size is an important factor. While specification of a speed (e.g., through an arbitrary index term) for a given type of screen-film combination is not directly coupled to the noise level, changing the dose in a digital system will affect the signal-to-noise ratio. Even for screen-film systems, the relation between speed number and image receptor dose is defined only for homogeneous exposures under laboratory conditions, involving specific phantoms and beam qualities. For patient images under clinical conditions, the actual air kerma at the detector may deviate from the value expected from the speed number (e.g., 2.5 Gy at speed 400). It was, therefore, recently recommended, in an editorial by Huda, that authors specify the amount of radiation incident on any digital receptor when characterising the system performance, such as the detector air kerma in Gy [20].

6 M. Uffmann, C. Schaefer-Prokop / European Journal of Radiology 72 (2009) Fig. 3. List of terms for exposure indices for various digital systems and their relationship to traditional dose measure (in Gy). In the second column, the proposal for an international standardisation is detailed (courtesy of Ulrich Neitzel, Philips Medical Systems, Hamburg) Exposure index In an attempt to give the user feedback about the actual detector dose level of a clinical image, most digital systems provide what is called an exposure index. It is important to note that the definition of this exposure index was only recently standardised [21] and for systems now in use the numbers given for the exposure index still refer to different dose quantities for different systems. In all digital systems, the exposure index is derived from signals in the acquired image itself, and thus, describes the detector dose. The mathematical definition and calibration of the exposure index is different for the different manufacturers of digital systems (Fig. 3). The exposure index (EI) value provides a composite measure of patient-related features and exposure. Different combinations of patient body constitutions and exposure can result in the same detected signal. Variation of the EI may occur due to varying imaging content (e.g., with or without pneumonia in a chest radiograph), even if the same exposure setting was used and patient entrance exposure was identical. This means that the exposure index in digital systems can only be used as a surrogate for dose management; EI does not represent an equivalent for patient entrance exposure. Interpretation of these EI values is further complicated by factors such as the collimation detection, the manufacturer-specific calibration, and the examination-specific image processing. The main purpose of the exposure index is to allow for a longitudinal (time) comparison of system operation during instillation and practice, which depends on a properly calibrated system. Furthermore, the exposure index is influenced by the accuracy of the collimation detection algorithm, the collimation, the time delay between acquisition and read-out, and the reproducibility of the exposure system. Thus, for proper interpretation, it is necessary to have an idea about the degree of variation of EI attributable to varying clinical and acquisition conditions. This range then may have a variable effect on dose. At present, a broad acceptance of the exposure index is hindered by the multitude of different vendor-specific variants (Fig. 2) Automated dose control Modern digital equipment usually provides the necessary data interface for automatic data collection and evaluation [22,23]. Integrated in a suitable RIS/PACS environment, dose control can be established in clinical routine as part of an overall quality control Fig. 4. a, b: Automated data collection and evaluation for longitudinal assessment of dose, illustrated by anterior-posterior lumbar spine radiographs (a). Even small alterations of the automated exposure control are indicated (arrow). Automated assessment of the kerma-area-product in posterior anterior chest radiographs (b). The majority of exposures are below the diagnostic reference level (red line) (Courtesy of Ulrich Neitzel, Philips Medical Systems, Hamburg). program. With longitudinal assessment of dose-relevant parameters, even small deviations can be identified (Fig. 4 a, b) Diagnostic reference dose levels Diagnostic reference levels (DRL) are defined as dose levels for typical examinations for groups of standard-sized patients or standard phantoms for broadly defined types of equipment. They are specified as entrance skin air kerma (ESAK, measured in air without backscatter) or as entrance skin dose (ESD, measured in specified material with backscatter, more commonly used in the United States). The concept of DRL was introduced by the International Commission of Radiological Protection (ICRP) in the 1990s [2]. DRL are typically set at the third quartile (75% value) of the dose distribution, derived from a suitable patient dose survey. The DRL specified are not to be exceeded with routine practice (Fig. 4b). The reference levels are periodically reviewed and, if necessary and possible, modified on the basis of knowledge of current practice. In Table 3, some examples of DRL for radiographic examinations in adults for different countries are given. Specific values have also been set for paediatric examinations in different age groups. The DRL are advisory. That is, they do not distinguish between acceptable and unacceptable practice. It should be noted that the reference levels derived from these surveys represent the state of practice and not the state of the art, and should be considered as such. They do not take into account the options provided by most

7 208 M. Uffmann, C. Schaefer-Prokop / European Journal of Radiology 72 (2009) Table 3 Examples of diagnostic reference levels (ESD = entrance surface dose, ESAK = entrance-surface air kerma). US 1999 ESAK [mgy] UK 2000 ESD [mgy] AP/PA skull 3 5 Lat skull PA chest Lat chest AP abdomen AP pelvis 4 10 AP C-spine 1.25 AP L-spine Lat L-spine Sources: [22,24,25]. Germany 2003 ESD [mgy] modern CR and DR detector technologies, and allow for flexibility to accommodate individual variation in dose reduction and resultant image quality. However, because digital systems have this greater freedom in setting the dose level without overexposing, adherence to reference levels is even more important to avoid dose levels to the patient that do not contribute to the clinical purpose of a medical imaging task Quality assurance National regulations exist for the approval and operation of radiography equipment to assure that sufficient image quality is achieved with a reasonable amount of radiation dose. Technical details are specified in national or international standards. The image quality of the imaging systems is verified with technical phantoms at regular intervals. A typical phantom for this purpose comes with a copper step wedge to control the dynamic range, different low-contrast objects for assessment of the low contrast resolution, and a lead bar pattern for determination of the spatial resolution Perspective The advances in digital radiography have revolutionised the way radiologists interpret medical images and communicate the findings. Modern image processing algorithms provide optimised image quality at a reasonable dose level. Beyond this, a variety of elaborate post-processing techniques have evolved, which increasingly assist the radiologist in the differentiation between anatomic noise and abnormal findings. These diagnostic enhancements, such as dual energy subtraction, temporary subtraction, digital tomosynthesis, and computer-aided diagnosis, are described in detail in other articles of this issue. 4. Summary The balance between the necessary image quality and the amount of radiation dose invested is much more flexible for digital detectors compared to screen-film systems. Digital detectors are based on different technical principles, and therefore, require different levels of dose. However, the dose requirement differences between DR and CR are shrinking. Optimising image quality and lowering the dose requires the optimisation of the whole imaging chain (detector, acquisition, processing, and display). Guidelines implemented for conventional radiography, including appropriate collimation, appropriate source-to-image distance (SID), focal spot size, and patient positioning, are as valid for digital techniques as they were for conventional techniques. In contrast to options for dose reduction with digital systems, there is the hazard of dose creep, meaning an unnoticed increase in exposure with time. This is most problematic in portable examinations using manual tube settings, paediatric applications, and fluoroscopic studies. With longitudinal assessment of dose-relevant parameters, dose control can be established in the clinical routine as part of an overall quality control program. As part of such quality control, diagnostic reference levels should be maintained. In addition, implementing image quality classes represents active dose containment. However, this requires reorganisation of the routine workflow. References [1] Managing patient dose in digital radiology. ICRP Publication 93. Elsevier: ICRP; [2] 1990 Recommendation of the international commission on radiological protection users edition. ICRP Publication 60. Oxford: ICRP: Pergamon Press; [3] Seibert JA. Tradeoffs between image quality and dose. Pediatr Radiol 2004;34(suppl. 3):S [discussion S ]. [4] Busch HP, Busch S, Decker C, Schilz C. Image quality and exposure dose in digital projection radiography. Rofo 2003;175:32 7. [5] Busch HP, Faulkner K. Image quality and dose management in digital radiography: a new paradigm for optimisation. Radiat Prot Dosimetry 2005;117: [6] Neitzel U. Management of pediatric radiation dose using Philips digital radiography. Pediatr Radiol 2004;34(suppl. 3):S [discussion S ]. [7] MacMahon H. Digital chest radiography: practical issues. J Thorac Imaging 2003;18: [8] Launders J, Cowen A, Bury R, Hawkridge P. Towards image quality, beam energy and effective dose optimisation in digital thoracic radiography. Eur Radiol 2001;11: [9] Honey ID, Mackenzie A, Evans DS. Investigation of optimum energies for chest imaging using film-screen and computed radiography. Br J Radiol 2005;78: [10] Metz S, Damoser P, Hollweck R, et al. Chest radiography with a digital flat-panel detector: experimental receiver operating characteristic analysis. Radiology 2005;234: [11] Uffmann M, Neitzel U, Prokop M, et al. Flat-panel-detector chest radiography: effect of tube voltage on image quality. Radiology 2005;235: [12] Geijer H, Norrman E, Persliden J. Optimizing the tube potential for lumbar spine radiography with a flat-panel digital detector. Br J Radiol 2009;82:62 8. [13] Samei E, Dobbins III JT, Lo JY, Tornai MP. A framework for optimising the radiographic technique in digital X-ray imaging. Radiat Prot Dosimetry 2005;114: [14] Hamer OW, Volk M, Zorger N, et al. Contrast-detail phantom study for X-ray spectrum optimization regarding chest radiography using a cesium iodideamorphous silicon flat-panel detector. Invest Radiol 2004;39: [15] Schaetzing R. Management of pediatric radiation dose using Agfa computed radiography. Pediatr Radiol 2004;34(suppl. 3):S [discussion S ]. [16] Dobbins III JT, Samei E, Chotas HG, et al. Chest radiography: optimization of x-ray spectrum for cesium iodide-amorphous silicon flat-panel detector. Radiology 2003;226: [17] Niemann T, Reisinger C, Rau P, Schwarz J, Ruis-Lopez L, Bongartz G. Image quality in conventional chest radiography. Evaluation using the postprocessing tool Diamond View ((R)). Eur J Radiol [18] Weatherburn GC, Bryan S, Davies JG. Comparison of doses for bedside examinations of the chest with conventional screen-film and computed radiography: results of a randomized controlled trial. Radiology 2000;217: [19] Willis CE. Computed radiography: a higher dose? Pediatr Radiol 2002;32: [discussion ]. [20] Huda W. The current concept of speed should not be used to describe digital imaging systems. Radiology 2005;234: [21] Medical electrical equipment exposure index of digital X-ray imaging systems Part 1: definitions and requirements for general radiography. Geneva, Switzerland: International Electrotechnical Commission (IEC), International Standard IEC : ; [22] Schuncke A, Neitzel U. Retrospective patient dose analysis of a digital radiography system in routine clinical use. Radiat Prot Dosimetry 2005;114: [23] Ward M, Hughes D, Connolly P, Moores BM. Central dose data management and analysis in IT-driven radiation protection strategies. Radiat Prot Dosimetry 2005;114: [24] Gray JE, Archer BR, Butler PF, et al. Reference values for diagnostic radiology: application and impact. Radiology 2005;235: [25] Wall BF. Diagnostic reference levels in the X-ray department. Eur Radiol Syllabus 2004;14:66 73.

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them. In press 2004 1 2 Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Digital radiology An appropriate analogy that is easy for most people to understand

More information

Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals

Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals DOI 10.1007/s00247-011-2290-1 ORIGINAL ARTICLE Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals Mervyn D. Cohen & Richard Markowitz

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

2217 US Highway 70 East Garner, NC Main: Fax:

2217 US Highway 70 East Garner, NC Main: Fax: Viztek is committed to providing the highest image quality possible in our CR & DR product lines. There are several factors that directly affect the overall quality of CR & DR based images. The eposure

More information

Ask EuroSafe Imaging Tips & Tricks. Paediatric Imaging Working Group. Dose Management in Digital Radiography

Ask EuroSafe Imaging Tips & Tricks. Paediatric Imaging Working Group. Dose Management in Digital Radiography Ask EuroSafe Imaging Tips & Tricks Paediatric Imaging Working Group Dose Management in Digital Radiography Raija Seuri (HUS Medical Imaging Center, FI) Cristina Almeida (Centro Hospitalar de Lisboa Central,

More information

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S.

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S. Practical Medical Physics Session: TG-151 Dose Monitoring August 5, 2013 Katie Hulme, M.S. Digital Imaging and Dose Creep Images courtesy of Agfa Healthcare Under-Exposed Over-Exposed Freedman et al.,

More information

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura New Technology https://www.youtube.com/watch?v=ptkzznazb 7U COMPUTED

More information

Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom

Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 1, WINTER 2003 Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom Z. F. Lu,* E. L. Nickoloff, J.

More information

Digital radiography (DR) post processing techniques for pediatric radiology

Digital radiography (DR) post processing techniques for pediatric radiology Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions

More information

Teaching Digital Radiography and Fluoroscopic Radiation Protection

Teaching Digital Radiography and Fluoroscopic Radiation Protection Teaching Digital Radiography and Fluoroscopic Radiation Protection WCEC 20 th Student Educator Radiographer Conference Dennis Bowman, RT(R), CRT (R)(F) Community Hospital of the Monterey Peninsula (CHOMP)

More information

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of radiation to the population due to Medical Imaging

More information

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography Journal of Physics: Conference Series OPEN ACCESS The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography To cite this article: N A A Daud et al 2014 J. Phys.:

More information

CR Basics and FAQ. Overview. Historical Perspective

CR Basics and FAQ. Overview. Historical Perspective Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors

More information

The importance of radiation quality for optimisation in radiology

The importance of radiation quality for optimisation in radiology Available online at http://www.biij.org/2007/2/e38 doi: 10.2349/biij.3.2.e38 biij Biomedical Imaging and Intervention Journal COMMENTARY The importance of radiation quality for optimisation in radiology

More information

Acquisition, Processing and Display

Acquisition, Processing and Display Acquisition, Processing and Display Terri L. Fauber, R.T. (R)(M) Department of Radiation Sciences School of Allied Health Professions Virginia Commonwealth University Topics Image Characteristics Image

More information

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development Teaching Radiographic Technique in a Digital Imaging Paradigm Objectives 1. Discuss the historical development of digital imaging. Dawn Couch Moore, M.M.Sc., RT(R) Assistant Professor and Director Emory

More information

A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers

A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers Poster No.: C-3011 Congress: ECR 2010 Type: Topic: Authors: Scientific Exhibit

More information

of sufficient quality and quantity

of sufficient quality and quantity of sufficient quality and quantity The patient s body attenuates the beam as it passes though the body More energy is deposited in organs located near the entry of the beam than near the exit of the beam

More information

10/3/2012. Study Harder

10/3/2012. Study Harder This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen

More information

Appropriate Inspection Distance of Digital X-Ray Imaging Equipment for Diagnosis

Appropriate Inspection Distance of Digital X-Ray Imaging Equipment for Diagnosis Indian Journal of Science and Technology Vol 8(S8), 380-386, April 2015 ISSN (Print) : 0974-6846 ISSN (Online) : 0974-5645 DOI: 10.17485/ijst/2015/v8iS8/70528 Appropriate Inspection Distance of Digital

More information

Digital Image Management: the Basics

Digital Image Management: the Basics Digital Image Management: the Basics Napapong Pongnapang, Ph.D. Department of Radiological Technology Faculty of Medical Technology Mahidol University Outline From screen/film to digital radiography PACS/Tele

More information

IMAGE QUALITY AND DOSE MANAGEMENT IN DIGITAL RADIOGRAPHY: A NEW PARADIGM FOR OPTIMISATION H. P. Busch 1 and K. Faulkner 2,

IMAGE QUALITY AND DOSE MANAGEMENT IN DIGITAL RADIOGRAPHY: A NEW PARADIGM FOR OPTIMISATION H. P. Busch 1 and K. Faulkner 2, Radiation Protection Dosimetry (2005), Vol. 117, No. 1 3, pp. 143 147 doi:10.1093/rpd/nci728 Advance Access published on February 3, 2006 IMAGE QUALITY AND DOSE MANAGEMENT IN DIGITAL RADIOGRAPHY: A NEW

More information

Half value layer and AEC receptor dose compliance survey in Estonia

Half value layer and AEC receptor dose compliance survey in Estonia Half value layer and AEC receptor dose compliance survey in Estonia K. Kepler, A. Vladimirov Training Centre of Medical Physics, University of Tartu Testing Centre of the University of Tartu, Estonia E-mail:

More information

10/26/2015. Study Harder

10/26/2015. Study Harder This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen

More information

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II APPROVED 12/O2/2011 EFFECTIVE SPRING 2013-14 Prefix & Number RAD 150 Course Title: Radiologic Exposure Technique II & Lab Purpose of this submission: New Change/Updated

More information

A comparative study of several digital flat panel X-ray units: patients doses and image quality in chest radiography

A comparative study of several digital flat panel X-ray units: patients doses and image quality in chest radiography A comparative study of several digital flat panel X-ray units: patients doses and image quality in chest radiography Torres Cabrera R. 1, España López M.L. 2 Ruiz Manzano P. 3, Sastre Aguado J.M. 4,, Rivas

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 16371-1 First edition 2011-10-01 Non-destructive testing Industrial computed radiography with storage phosphor imaging plates Part 1: Classification of systems Essais non destructifs

More information

Do you have any other questions? Please call us at (Toll Free) or , or

Do you have any other questions? Please call us at (Toll Free) or , or INSTRUCTIONS Read the appropriate course/ textbook. This is an open book test. A score of 75% or higher is needed to receive CE credit. You will have a maximum of three attempts to pass this course. Please

More information

Digital Imaging Considerations Computed Radiography

Digital Imaging Considerations Computed Radiography Digital Imaging Considerations Digital Radiography Computed Radiography o Cassette based Direct or Indirect Digital Radiography o Cassetteless Computed Radiography 1 CR Image Acquisition Most like conventional

More information

Grid-like contrast enhancement for bedside chest radiographs acquired without anti-scatter grid

Grid-like contrast enhancement for bedside chest radiographs acquired without anti-scatter grid -like contrast enhancement for bedside chest radiographs acquired without anti-scatter grid Philips Detlef Mentrup, PhD, Image Processing Specialist Ulrich Neitzel, PhD, Clinical Scientist Sascha Jockel,

More information

Test Equipment for Radiology and CT Quality Control Contents

Test Equipment for Radiology and CT Quality Control Contents Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3

More information

Attikon, Rimini 1, , Athens, Greece , Athens, Greece , Athens, Greece

Attikon, Rimini 1, , Athens, Greece , Athens, Greece , Athens, Greece Radiation Protection Dosimetry (2005), Vol. 117, No. 1 3, pp. 291 297 doi:10.1093/rpd/nci742 Advance Access published on February 7, 2006 QUALITY ASSURANCE (QA) PROCEDURES FOR SOFTWARE: EVALUATION OF AN

More information

Overview of Safety Code 35

Overview of Safety Code 35 Common Quality Control Procedures for All s Quality Control Procedures Film All s Daily Quality Control Tests Equipment Warm-up (D1) According to manufacturers instructions Can include auto calibration(d1)

More information

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 1 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT RADT 3463 COMPUTERIZED IMAGING Section I: Chapter

More information

Beam-Restricting Devices

Beam-Restricting Devices Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the

More information

Outline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control

Outline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control Understanding Digital Modalities: Image Quality and Dose S. Jeff Shepard, M.S. University of Texas M. D. Anderson Cancer Center Houston, Texas Special Acknowledgement: Stephen K. Thompson, M.S. William

More information

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System Manus

More information

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS Veterinary Science Preparatory Training for the Veterinary Assistant Floron C. Faries, Jr., DVM, MS Radiology Floron C. Faries, Jr., DVM, MS Objectives Determine the appropriate machine settings for making

More information

Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2

Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2 Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2 Medical Consult, SA; Establish a method to correlate image

More information

biij Optimisation in general radiography CJ Martin, PhD, FIPEM, FioP Biomedical Imaging and Intervention Journal REVIEW PAPER

biij Optimisation in general radiography CJ Martin, PhD, FIPEM, FioP Biomedical Imaging and Intervention Journal REVIEW PAPER Available online at http://www.biij.org/2007/2/e18 doi: 10.2349/biij.3.2.e18 biij Biomedical Imaging and Intervention Journal REVIEW PAPER Optimisation in general radiography CJ Martin, PhD, FIPEM, FioP

More information

Artefacts found in computed radiography

Artefacts found in computed radiography The British Journal of Radiology, 74 (2001), 195 202 E 2001 The British Institute of Radiology Pictorial review Artefacts found in computed radiography L J CESAR, RT(R)(QM), B A SCHUELER, PhD, F E ZINK,

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

Learning Objectives: What s my motivation? (unknown screen actor) Workshop Overview

Learning Objectives: What s my motivation? (unknown screen actor) Workshop Overview Practical Medical Physics Adapting Traditional Clinical Medical Physics to Digital Radiography Charles E. Willis, Ph.D., DABR Associate Professor Department of Imaging Physics The University of Texas M.D.

More information

Image Quality Artifacts in Digital Imaging

Image Quality Artifacts in Digital Imaging MAHIDOL UNIVERSITY Wisdom of the Land Image Quality Artifacts in Digital Imaging Napapong Pongnapang, Ph.D. Department of Radiological Technology Faculty of Medical Technology Mahidol University, Bangkok,

More information

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology CODE: RADT 156 INSTITUTE: Health Science TITLE: Equipment Operation I DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of equipment operation and maintenance of radiographic

More information

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations S. Baechler 1, P. Monnin 1, A. Aroua 1, J.F. Valley 1, M. Perrier, P. Trueb 3, F.R. Verdun 1 1 University

More information

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter Revision Date: January 5th, 2017 www.disc-imaging.com Table of Contents Section A: Preliminary Setup Requirements... 4 Tools

More information

Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1

Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1 Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1 Gham Hur, M.D., Yoon Joon Hwang, M.D., Soon Joo Cha, M.D., Su Young Kim, M.D., Yong Hoon Kim, M.D.

More information

Radiographic Techniques, Contrast, and Noise in X-Ray Imaging

Radiographic Techniques, Contrast, and Noise in X-Ray Imaging Residents Section Physics Minimodule Huda and Abrahams Techniques, Contrast, and Noise in Radiography Residents Section Physics Minimodule Residents inradiology Walter Huda 1 R. Brad Abrahams 2 Huda W,

More information

A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations. Shirley Weddle, RT(R)(M), CIIP, BBA

A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations. Shirley Weddle, RT(R)(M), CIIP, BBA A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations Shirley Weddle, RT(R)(M), CIIP, BBA OBJECTIVES Define Computed Radiography (CR) Discuss CR vendor

More information

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. RA110 test 3 Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. 1. An object 35 cm in width is radiographed at 100 cm SID and at a 50 cm SOD. What

More information

Outline ASRT Changes Impact on current curriculum Potential new courses WECM Changes Last update Resources and needs

Outline ASRT Changes Impact on current curriculum Potential new courses WECM Changes Last update Resources and needs Change nd Annual Blinn College 2 nd Educator s Workshop For Radiologic Sciences July 28, 2007 Christi Carter, MSRS, RT(R) Outline ASRT Changes Impact on current curriculum Potential new courses WECM Changes

More information

Disclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography

Disclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography Current Implementation Status of IEC Standard 62494-1: Exposure Index (EI) for Digital Radiography July 31, 2017 Ryan Fisher, PhD, DABR Katie Hulme, MS, DABR None Disclosures Outline Review of IEC Standard

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

Visibility of Detail

Visibility of Detail Visibility of Detail Radiographic Quality Quality radiographic images represents the, and information is for diagnosis. The of the anatomic structures and the accuracy of their ( ) determine the overall

More information

LECTURE 1 The Radiographic Image

LECTURE 1 The Radiographic Image LECTURE 1 The Radiographic Image Prepared by:- KAMARUL AMIN ABDULLAH @ ABU BAKAR UiTM Faculty of Health Sciences Medical Imaging Department 11/23/2011 KAMARUL AMIN (C) 1 Lesson Objectives At the end of

More information

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients

More information

Mammography: Physics of Imaging

Mammography: Physics of Imaging Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness

More information

History of digital imaging

History of digital imaging CR/QA RADCHEX History of digital imaging Early, crude digital detectors were developed in the 1970 s Image quality was problematic Processing time of digital images was untenable Viewing, transfer and

More information

MUSICA Nerve Center. Artificial Intelligence. Intelligent tools for your Digital Radiography workflow. Fluoroscopy. Workflow Optimization

MUSICA Nerve Center. Artificial Intelligence. Intelligent tools for your Digital Radiography workflow. Fluoroscopy. Workflow Optimization Image Quality Bariatric Abdomen Pediatric Imaging Diagnostic Confidence Fluoroscopy Neonatal Imaging Scatter Suppression Dental Full Leg Full Spine Exposure Control Index Artificial Intelligence General

More information

X-RAYS - NO UNAUTHORISED ENTRY

X-RAYS - NO UNAUTHORISED ENTRY Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

Essentials of Digital Imaging

Essentials of Digital Imaging Essentials of Digital Imaging Module 6 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 6 Dose Reduction and Patient Safety 1. ASRT Animation 2. Welcome Welcome to Essentials

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information

radiography detector

radiography detector Clinical evaluation of a full field digital projection radiography detector Gary S. Shaber'1, Denny L. Leeb, Jeffrey Belib, Gregory Poweii1', Andrew D.A. Maidment'1 a Thomas Jefferson University Hospital,

More information

DRX Plus Detectors: Going from Good to Great

DRX Plus Detectors: Going from Good to Great DRX Plus Detectors: Going from Good to Great Authors: Karin Töpfer, Tim Wojcik Introduction Carestream s introduction in 2009 of the world s first portable, wireless, cassette-sized detector the CARESTREAM

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality Digital radiography: Digital radiography is set to become the most common form of processing radiographic images in the next 10 years. This is due to a number of practical and image quality issues. Practical

More information

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016 Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016 Current technology in digital image production (CR/DR and other modalities) 2/ Overview Digital

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

RADIOGRAPHIC EXPOSURE

RADIOGRAPHIC EXPOSURE RADIOGRAPHIC EXPOSURE Receptor Exposure Receptor Exposure the that interacts with the receptor. Computed Radiography ( ) requires a. Direct Digital Radiography (DR) requires a. Exposure Indicators Exposure

More information

Digital radiography. bucky table and wall stand as a dual detector or wireless system. Amadeo DR Systems

Digital radiography. bucky table and wall stand as a dual detector or wireless system. Amadeo DR Systems Amadeo DR Systems with dicom PACS DX-R Software X-ray Systems for the Future Digital radiography with Amadeo R-DR including bucky table and wall stand as a dual detector or wireless system High contrast

More information

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on

More information

Instant DR in Jordan

Instant DR in Jordan Hashemite University leads the way with first Instant DR in Jordan DR Retrofit supports research and education goals of the Faculty of Allied Health Sciences, while enhancing care for staff and students

More information

Maximizing clinical outcomes

Maximizing clinical outcomes Maximizing clinical outcomes Digital Tomosynthesis Dual Energy Subtraction Automated Long Length Imaging Improved image quality at a low dose Xray Xray Patented ISS capture technology promotes high sensitivity

More information

REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS

REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS DEPARTMENT OF HEALTH DIRECTORATE: RADIATION CONTROL Implementation date: 31 March 2009 Contents

More information

1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013).

1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013). CODE: RADT 151 INSTITUTE: Health Science TITLE: Radiographic Exposure DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of radiographic exposure selection and manipulation

More information

IBEX MATERIALS DETECTION TECHNOLOGY

IBEX MATERIALS DETECTION TECHNOLOGY WHITE PAPER: IBEX MATERIALS DETECTION TECHNOLOGY IBEX Innovations Ltd. Registered in England and Wales: 07208355 Address: Discovery 2, NETPark, William Armstrong Way, Sedgefield, TS21 3FH, UK Patents held

More information

Ludlum Medical Physics

Ludlum Medical Physics Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

Optimization of chest radiographic imaging parameters: a comparison of image quality and entrance skin dose for digital chest radiography systems,

Optimization of chest radiographic imaging parameters: a comparison of image quality and entrance skin dose for digital chest radiography systems, Clinical Imaging 36 (2012) 279 286 Optimization of chest radiographic imaging parameters: a comparison of image quality and entrance skin dose for digital chest radiography systems, Zhonghua Sun a,, Chenghsun

More information

WHITE PAPER. Moving from CR to DR. Optimizing Image Quality and Dose

WHITE PAPER. Moving from CR to DR. Optimizing Image Quality and Dose Moving from CR to DR Optimizing Image Quality and Dose Table of contents 1. Introduction...3 2. CR and DR: similarities and differences...4 3. Image capture technologies...5 3.1 DR panel technology...5

More information

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS Course Number: RADG 112 Department: Radiography Course Title: Image Production & Eval. Semester: Spring Year: 1997 Objectives/ Unit One: Introduction

More information

PLD5600A High Frequency Digital Gastrointestinal &DR System(630mA)

PLD5600A High Frequency Digital Gastrointestinal &DR System(630mA) PLD5600A High Frequency Digital Gastrointestinal &DR System(630mA) Application: Full support perspective, gastrointestinal spot film, GI (barium meal, barium enema), orthopedic photography, pediatrics

More information

ddr Compact Series Setting a new benchmark in digital radiography.

ddr Compact Series Setting a new benchmark in digital radiography. ddr Compact Series Setting a new benchmark in digital radiography. ddrcompact When productivity and exceptional value come together. With the introduction of its newest DR system, the ddrcompact, Swissray

More information

Essentials of Digital Imaging

Essentials of Digital Imaging Essentials of Digital Imaging Module 2 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 2 Processing 1. ASRT Animation 2. Welcome Welcome to Essentials of Digital Imaging

More information

Estimation of signal transfer property for wireless digital detector in different measurement schemes

Estimation of signal transfer property for wireless digital detector in different measurement schemes Estimation of signal transfer property for wireless digital detector in different measurement schemes Anatoli Vladimirov, Kalle Kepler Training Centre of Medical Physics, University of Tartu, Estonia 11

More information

Features and Weaknesses of Phantoms for CR/DR System Testing

Features and Weaknesses of Phantoms for CR/DR System Testing Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal

More information

Get more from your images with Symphony Image Processing

Get more from your images with Symphony Image Processing DIRECT RADIOGRAPHY The user-friendly DelWorks image acquisition and processing software provides a wide range of tools for a variety of image enhancements. Its user interface simplifies every step of the

More information

New Exposure Indicators for Digital Radiography Simplified for Radiologists and Technologists

New Exposure Indicators for Digital Radiography Simplified for Radiologists and Technologists Medical Physics and Informatics Technical Innovation Don et al. New Simplified Exposure Indicators Medical Physics and Informatics Technical Innovation Steven Don 1 ruce R. Whiting 2 Lois Jo Rutz 3 ruce

More information

DR _ solutions. We understand that customers don t need just products, they want. solutions

DR _ solutions. We understand that customers don t need just products, they want. solutions DR _ solutions We understand that customers don t need just products, they want solutions index company profile 1974-2005 2006-2007 - 2008 ITALRAY Srl was founded in 1974 as the production branch of Marzocchi

More information

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc Quality Control for Stereotactic Breast Biopsy Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc. 716-924-0350 Methods of Imaging Guided Breast Biopsy Ultrasound guided, hand-held needle

More information

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring

More information

Redefining Ergonomics

Redefining Ergonomics Samsung Electronics Co., Ltd. inspires the world and shapes the future with transformative ideas and technologies, redefining the worlds of TVs, smartphones, wearable devices, tablets, cameras, digital

More information

Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis

Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis Poster No.: C-0585 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Minehiro,

More information

Detector technology in simultaneous spectral imaging

Detector technology in simultaneous spectral imaging Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become

More information

Objective Evaluation of Radiographic Contrast- Enhancement Masks

Objective Evaluation of Radiographic Contrast- Enhancement Masks Chapter 8 Objective Evaluation of Radiographic Contrast- Enhancement Masks The development and application of radiographic contrast-enhancement masks (RCMs) in digital radiography (DR) were discussed in

More information

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

Unit thickness. Unit area. σ = NΔX = ΔI / I 0 Unit thickness I 0 ΔI I σ = ΔI I 0 NΔX = ΔI / I 0 NΔX Unit area Δx Average probability of reaction with atom for the incident photons at unit area with the thickness of Delta-X Atom number at unit area

More information