Estimation of signal transfer property for wireless digital detector in different measurement schemes

Size: px
Start display at page:

Download "Estimation of signal transfer property for wireless digital detector in different measurement schemes"

Transcription

1 Estimation of signal transfer property for wireless digital detector in different measurement schemes Anatoli Vladimirov, Kalle Kepler Training Centre of Medical Physics, University of Tartu, Estonia 11 TH International Conference & Workshop Medical Physics in the Baltic Satates, Kaunas, October, 2013

2 Introduction European criteria for acceptability of medical radiological equipment Radiation Protection 162 include and require measurement of signal transfer property (STP). STP refers to a test to be done during the acceptance testing of the computed radiography (CR) or digital radiography (DR) system in order to establish the relationship between detector air kerma (DAK) (or receptor dose) and pixel value (PV).

3 Radiation Protection 162

4 Simple STP ralationship Linear: PV lin = a+b DAK DAK = (PV-a)/b Logarithmic: PV log = a ln(dak)+b DAK = exp((pv-a)/b) Power: PV power = a DAK b +c DAK = ((PV-c)/a) 1/b

5 Linearization of standard deviation The measured standard deviation of the pixel values (σ) cannot simply be corrected to the true standard deviation (σ') using the inverse of the STP. For measurements of several objective parameters of image quality and imaging system: signal-to-noise ratio (SNR), modulation transfer function (MTF), noise power spectrum (NPS), detective quantum efficiency (DQE) the linearized standard deviation can then be estimated by subtracting the linearized mean pixel value: σ'= (PV + σ)' - PV' where ( ) is the linearization function using the inverse of the STP

6 Beam conditions used for DDI in CR Manufacturer Added filter kv Pre-processing Agfa 1.5 mm Cu 75 S=200 / System Diagnosis / Flat Field Fuji 0 mm Al 80 Semi-Auto / L=1 Kodak 0.5 mm Cu + 1 mm Al 80 Pattern mode Konica 0 mm Al 80 Test 1/G =2/F off

7 Beam conditions used for STP Spectrum Added filter kv Nominal HVL (mm Al) Measured HVL (mm Al) IPEM 32 1 mm Cu 70 ~ RQA 5 by IEC RQA 3 by IEC mm Al mm Al PMMA 0 mm Al

8 Materials and methods Scatter geometry (a) X-ray tube (b) Added filter X-ray tube Scatter-free geometry Wide X-ray beam Narrow X-ray beam beam Dose chamber Table PMMA Table Grid Image detector or dose chamber Grid removed Image detector or dose chamber IPEM. Measurement of the Performance Characteristics of Diagnostic X- Ray Systems: Digital Imaging Systems. IPEM Report Number 32, Part VII (2010)

9 Materials and methods Siemens Ysio and Pixium FE 3543pR (Trixell,France) with a CsI scintillator coupled to TFT matrix in a:si -technology, pixel size 155 µm, matrix 34.2 x 43.2 cm, saturation dose: 75 Gy. Barracuda MPD dosimeter (RTI, Sweden), accuracy of 5%. Siemens 2011

10 Setting the amplification factor Detector dose Indicative sensitivity Typical amplification factor 7.1 μgy μgy μgy μgy μgy μgy Siemens 2011

11 Setting the gradation curve (LUT) Siemens 2011

12 Pixel value Results and discussion PV = 54.5 DAK R 2 = PV = 54.0 DAK R 2 = PV = 27.2 DAK R 2 = ,00 10,00 20,00 30,00 40,00 50,00 60,00 Detector air kerm a (μgy) IPEM32 RQA5 RQA3 STP for RQA5, RQA3 and IPEM32 schemes, measured directly (without table, grid)

13 Pixel value Results and discussion PV = DAK R 2 = PV= 55.0 DAK R 2 = PV = 54.4 DAK R 2 = 0, ,00 10,00 20,00 30,00 40,00 50,00 60,00 Detector air kerm a (μgy) IPEM32, w ithout table and grid IPEM32, in Bucky w ithout grid IPEM32, in Bucky w ith grid STP by IPEM32 on table and inside table with and without grid.

14 Pixel value Results and discussion PV = 54.9 DAK R 2 = PV = 55.0 DAK R 2 = ,00 10,00 20,00 30,00 40,00 50,00 60,00 Detector air kerm a (μgy) IPEM32, scatter-free IPEM32, scatter (PMMA) STP in IPEM32 scatter (PMMA) and scatter-free geometry (with dose detector inside the Bucky).

15 Pixel value Results and discussion PV = DAK R 2 = PV = DAK R 2 = PV = 54.4 DAK R 2 = ,00 5,00 10,00 15,00 20,00 25,00 30,00 35,00 40,00 Detector air kerm a (μgy) IPEM32, Ampl.1 IPEM32, Ampl.2 IPEM32, Ampl.4 STP with the amplification factors of 1, 2 and 4.

16 Pixel value Results and discussion PV = Ln(DAK) R 2 = PV = Ln(DAK) R 2 = ,00 5,00 10,00 15,00 20,00 Detector air kerm a (μgy) RQA5 RQA3 STP for RQA5 and RQA3 with typical post-processing with DAK up to 20 µgy.

17 Pixel value Results and discussion RQA5: PV = Ln(DAK ) R 2 = RQA3: PV = Ln(DAK) R 2 = ,00 10,00 20,00 30,00 40,00 50,00 60,00 70,00 Detector air kerm a (μgy) RQA5 RQA3 STP for RQA5 and RQA3 with typical post-processing with DAK up to 70 µgy.

18 Linearized standard diviation Results and discussion 0,10 σ = DAK 0.47 σ = DAK 0.35 σ = DAK 0.48 RQA5 IPEM32 RQA3 0,01 0,10 1,00 10,00 100,00 Detector air kerm a (μgy) Linearized standard deviation of pixel value by RQA5, RQA3 and IPEM32 measured directly on the receptor.

19 Conclusions Several measurement schemes using beam qualities RQA3, RQA5 and IPEM32 were studied along with three different geometries: scatter (PMMA), scatter-free and direct measurement geometry for estimation of STP of a digital wireless detector. For the pre-processed images the more clinically applicable IPEM32 scheme is compared with the IEC RQA5 scheme. The results show that there is no significant difference in the coefficient of correlation R 2 for the STP curve. Also power coefficients for the measured linearized standard deviation dependence from DAK are comparable for these schemes. The different IPEM32 geometries for measurements without table, inside table with and without grid could be interchangeable in practice.

20 Conclusions The post-processed images could be linearizable up to 20 µgy if simple LUT was applied. One conclusion is that if we have to measure STP on non-wireless Bucky, where it is not possible to measure DAK directly on the entrance surface of the receptor, then the dose measurement could be done on the entrance of the table (or vertical stand), and recalculate it on the right distance. If we know transmission of grid and table we could calculate exact DAK.

21 References Bosmans H., Nens J., Delzenne L., et al. Exploration of exposure conditions with a novel wireless detector for bedside digital radiography. Proc. of SPIE, 2012, 8313, p K1-9. Samei E. DQE of wireless digital detectors: Comparative performance with differing filtration schemes. Med. Phys., 2013, 40, p European Commission. Criteria for acceptability of medical radiological equipment used in diagnostic radiology, nuclear medicine and radiotherapy. Rad. Prot Luxembourg: EC, 2012, 84 p. Institute of Physics and Engineering in Medicine. Measurement of the performance characteristics of diagnostic X-ray systems: digital imaging systems. IPEM Rep. 32, Part VII. IPEM, p. Institute of Physics and Engineering in Medicine. Recommended standards for the routine performance testing of diagnostic X-ray imaging systems. IPEM Report 91. IPEM, International Electrotechnical Commission. Medical diagnostic X-ray equipment Radiation conditions for use in the determination of characteristics Edition 2.0. IEC Geneva: IEC (2005). Kepler K., Vladimirov A. Survey of compliance with European acceptability criteria for HVL and AEC. Rad. Prot. Dos., 2013, 153(2), Siemens AG. FLUOROSPOT Compact. Operator manual. Imaging System for Ysio. Erlangen (2011). Mackenzie, A. Validation of correction methods for the non-linear response of digital radiography systems. The Brit. J. of Rad. 81, (2008). Marshall N. W., Mackenzie A., Honey, I. D. Quality control measurements for digital x-ray detectors. Phys. in Med. Biol. 56, (2011).

22 Thank you!

Half value layer and AEC receptor dose compliance survey in Estonia

Half value layer and AEC receptor dose compliance survey in Estonia Half value layer and AEC receptor dose compliance survey in Estonia K. Kepler, A. Vladimirov Training Centre of Medical Physics, University of Tartu Testing Centre of the University of Tartu, Estonia E-mail:

More information

Disclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography

Disclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography Current Implementation Status of IEC Standard 62494-1: Exposure Index (EI) for Digital Radiography July 31, 2017 Ryan Fisher, PhD, DABR Katie Hulme, MS, DABR None Disclosures Outline Review of IEC Standard

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

Investigation of Effective DQE (edqe) parameters for a flat panel detector

Investigation of Effective DQE (edqe) parameters for a flat panel detector Investigation of Effective DQE (edqe) parameters for a flat panel detector Poster No.: C-1892 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit D. Bor 1, S. Cubukcu 1, A. Yalcin 1, O.

More information

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities Journal of Physics: Conference Series PAPER OPEN ACCESS Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities Recent citations - Resolution Properties of a

More information

Learning Objectives: What s my motivation? (unknown screen actor) Workshop Overview

Learning Objectives: What s my motivation? (unknown screen actor) Workshop Overview Practical Medical Physics Adapting Traditional Clinical Medical Physics to Digital Radiography Charles E. Willis, Ph.D., DABR Associate Professor Department of Imaging Physics The University of Texas M.D.

More information

Ask EuroSafe Imaging Tips & Tricks. Paediatric Imaging Working Group. Dose Management in Digital Radiography

Ask EuroSafe Imaging Tips & Tricks. Paediatric Imaging Working Group. Dose Management in Digital Radiography Ask EuroSafe Imaging Tips & Tricks Paediatric Imaging Working Group Dose Management in Digital Radiography Raija Seuri (HUS Medical Imaging Center, FI) Cristina Almeida (Centro Hospitalar de Lisboa Central,

More information

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S.

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S. Practical Medical Physics Session: TG-151 Dose Monitoring August 5, 2013 Katie Hulme, M.S. Digital Imaging and Dose Creep Images courtesy of Agfa Healthcare Under-Exposed Over-Exposed Freedman et al.,

More information

Features and Weaknesses of Phantoms for CR/DR System Testing

Features and Weaknesses of Phantoms for CR/DR System Testing Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal

More information

Tailoring automatic exposure control toward constant detectability in digital mammography

Tailoring automatic exposure control toward constant detectability in digital mammography Tailoring automatic exposure control toward constant detectability in digital mammography Elena Salvagnini a) Department of Imaging and Pathology, Medical Physics and Quality Assessment, KUL, Herestraat

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 62220-1 First edition 2003-10 Medical electrical equipment Characteristics of digital X-ray imaging devices Part 1: Determination of the detective quantum efficiency Appareils

More information

Modeling detector performance in digital mammography using the linear cascaded systems approach

Modeling detector performance in digital mammography using the linear cascaded systems approach 51 Modeling detector performance in digital mammography using the linear cascaded systems approach V. Spyropoulou a,b, N. Kalivas a,b, A. Gaitanis a,b, C. Michail a,b, G. Panayiotakis b, I. Kandarakis

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information

Digital Radiography X-Ray System. X Twin with X Mobil Roesys GmbH [rshsmi] 1/11

Digital Radiography X-Ray System. X Twin with X Mobil Roesys GmbH [rshsmi] 1/11 Digital Radiography X-Ray System 2017 Roesys GmbH [rshsmi] 1/11 1 General specifications The system is intended to be installed with floor mounted components only. It enables spacesaving installation without

More information

Calibration of KAP meters

Calibration of KAP meters Calibration of KAP meters Alexandr Malusek! Division of Radiological Sciences Department of Medical and Health Sciences Linköping University! 2014-04-15 1 Outline 1. KAP meter construction 2. Air kerma-area

More information

Quality Control of Full Field Digital Mammography Units

Quality Control of Full Field Digital Mammography Units Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Melissa Martin, M.S., FAAPM, FACR, FACMP AAPM Annual Meeting - Philadelphia July 19, 2010 MO-B-204C-1 Educational Objectives

More information

A comparison of two methods for the determination of freein-air geometric efficiency in MDCT

A comparison of two methods for the determination of freein-air geometric efficiency in MDCT A comparison of two methods for the determination of freein-air geometric efficiency in MDCT Theocharis Berris *1, Kostas Perisinakis 1,, Antonios E. Papadakis and John Damilakis 1, 1 Department of Medical

More information

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure Accuracy and calibration of integrated radiation output indicators in diagnostic radiology: A report of the AAPM Imaging Physics Committee Task Group 190 Pei-Jan P. Lin a) Virginia Commonwealth University

More information

Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals

Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals DOI 10.1007/s00247-011-2290-1 ORIGINAL ARTICLE Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals Mervyn D. Cohen & Richard Markowitz

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom

Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 1, WINTER 2003 Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom Z. F. Lu,* E. L. Nickoloff, J.

More information

2217 US Highway 70 East Garner, NC Main: Fax:

2217 US Highway 70 East Garner, NC Main: Fax: Viztek is committed to providing the highest image quality possible in our CR & DR product lines. There are several factors that directly affect the overall quality of CR & DR based images. The eposure

More information

Preliminary studies of a new monitor ionization chamber

Preliminary studies of a new monitor ionization chamber 1 Preliminary studies of a new monitor ionization chamber Maíra T. Yoshizumi, Vitor Vivolo and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares (IPEN-CNEN) Comissão Nacional de Energia

More information

Assessment and Optimisation of Digital Radiography Systems for Clinical Use

Assessment and Optimisation of Digital Radiography Systems for Clinical Use Assessment and Optimisation of Digital Radiography Systems for Clinical Use Philip Doyle, BSc. MSc. Submitted in fulfilment of the requirements for the degree of Doctor of Philosophy Department of Clinical

More information

Digital mammography: DQE versus optimized image quality in clinical environment an on site study

Digital mammography: DQE versus optimized image quality in clinical environment an on site study Digital mammography: DQE versus optimized image quality in clinical environment an on site study Nadia Oberhofer* a, Alessandro Fracchetti a, Margareth Springeth a, Ehrenfried Moroder a a Health Service

More information

X-RAY MEDICAL EQUIPMENT

X-RAY MEDICAL EQUIPMENT X-RAY MEDICAL EQUIPMENT CHEST RADIOGRAPHY GENERAL RADIOGRAPHY & FLUOROSCOPY RADIOTHERAPY MOBILE HEALTHCARE MAMMOGRAPHY MAMMOSCAN FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM Biopsy Attachment џ MAMMOSCAN an ADANI

More information

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations S. Baechler 1, P. Monnin 1, A. Aroua 1, J.F. Valley 1, M. Perrier, P. Trueb 3, F.R. Verdun 1 1 University

More information

SAMSUNG SMART DIGITAL RADIOGRAPHY. XGEO GU60A series REDEFINING ERGONOMICS

SAMSUNG SMART DIGITAL RADIOGRAPHY. XGEO GU60A series REDEFINING ERGONOMICS SAMSUNG SMART DIGITAL RADIOGRAPHY XGEO GU60A series REDEFINING ERGONOMICS U-arm Positioner (Fully Automated) The XGEO GU60A/60A-65 is a universal, fully motorized system. Its unique U-arm rotates +120

More information

I. Introduction.

I. Introduction. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Accuracy of measuring half- and quarter-value layers and appropriate aperture width of a convenient method using a lead-covered case

More information

Fully Automatic X-ray meter kvp, dose, rate, HVL and time Extremely easy to use. Unfors ThinX RAD for Radiographic Quality Control

Fully Automatic X-ray meter kvp, dose, rate, HVL and time Extremely easy to use. Unfors ThinX RAD for Radiographic Quality Control Fully Automatic X-ray meter kvp, dose, rate, HVL and time Extremely easy to use Unfors ThinX RAD for Radiographic Quality Control Unfors ThinX RAD No keys! No menus! Just position and expose! Unfors ThinX

More information

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This

More information

Derivation of total filtration thickness for diagnostic x-ray source assembly

Derivation of total filtration thickness for diagnostic x-ray source assembly Physics in Medicine & Biology PAPER OPEN ACCESS Derivation of total filtration thickness for diagnostic x-ray source assembly To cite this article: Michiharu Sekimoto and Yoh Katoh 2016 Phys. Med. Biol.

More information

10/3/2012. Study Harder

10/3/2012. Study Harder This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen

More information

Development of a large-area CMOS-based detector for real-time x-ray imaging

Development of a large-area CMOS-based detector for real-time x-ray imaging Development of a large-area CMOS-based detector for real-time x-ray imaging Sung Kyn Heo a, Sung Kyu Park a, Sung Ha Hwang a, Dong Ak Im a, Jari Kosonen a, Tae Woo Kim a, Seungman Yun b, Ho Kyung Kim b*

More information

10/26/2015. Study Harder

10/26/2015. Study Harder This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen

More information

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist Topics Image Uniformity and Artifacts Image Quality - Detail

More information

A comparative study of several digital flat panel X-ray units: patients doses and image quality in chest radiography

A comparative study of several digital flat panel X-ray units: patients doses and image quality in chest radiography A comparative study of several digital flat panel X-ray units: patients doses and image quality in chest radiography Torres Cabrera R. 1, España López M.L. 2 Ruiz Manzano P. 3, Sastre Aguado J.M. 4,, Rivas

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

Mammography: Physics of Imaging

Mammography: Physics of Imaging Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

Digital Mammography: From Physical Performance Evaluation To Image Quality Analysis

Digital Mammography: From Physical Performance Evaluation To Image Quality Analysis Digital Mammography: From Physical Performance Evaluation To Image Quality Analysis Anastasios C. Konstantinidis Department of Medical Physics and Bioengineering, University College London, UK. taskon25@yahoo.gr

More information

VOLUMETRIC computed tomography (CT) based on flatpanel

VOLUMETRIC computed tomography (CT) based on flatpanel IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 3, JUNE 2009 1179 Radiation Effects on the Performance of CMOS Photodiode Array Detectors and the Role of Gain-Offset Corrections Ho Kyung Kim, Min Kook

More information

Evaluation and Comparison of Image Quality for Indirect Flat Panel Systems with CsI and GOS Scintillators

Evaluation and Comparison of Image Quality for Indirect Flat Panel Systems with CsI and GOS Scintillators Evaluation and Comparison of Image Quality for Indirect Flat Panel Systems with CsI and GOS Scintillators Mehmet E. Aksoy a, Mustafa E. Kamasak b, Erkan Akkur c, Ayhan Ucgul d, Muzaffer Basak d, Huseyin

More information

DRAFT Technical evaluation of Philips Microdose SI digital mammography system

DRAFT Technical evaluation of Philips Microdose SI digital mammography system DRAFT Technical evaluation of Philips Microdose SI digital mammography system NHSBSP Equipment Report 1310 August 2013 About the NHS Cancer Screening Programmes The national office of the NHS Cancer Screening

More information

Digital Radiography System

Digital Radiography System www.drgem.co.kr GXR-SD SERIES GXR-SD SERIES Digital Radiography System GXR-SD Series www.drgem.co.kr High Performance and Reliability The Optimal System for Digital Solution The GXR-SD Series is a diagnostic

More information

SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES.

SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES. SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES. Tingberg, Anders Published in: Radiation Protection Dosimetry DOI: 10.1093/rpd/ncs302 Published: 2013-01-01 Link to publication Citation for published

More information

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System Manus

More information

Digital Imaging Considerations Computed Radiography

Digital Imaging Considerations Computed Radiography Digital Imaging Considerations Digital Radiography Computed Radiography o Cassette based Direct or Indirect Digital Radiography o Cassetteless Computed Radiography 1 CR Image Acquisition Most like conventional

More information

An Exposure Indicator for Digital Radiography. Report of AAPM Task Group 116

An Exposure Indicator for Digital Radiography. Report of AAPM Task Group 116 AAPM REPORT NO. 116 An Exposure Indicator for Digital Radiography Report of AAPM Task Group 116 July 2009 DISCLAIMER: This publication is based on sources and information believed to be reliable, but the

More information

Comparison of new digital mammography systems: physical characterisation and image quality evaluation

Comparison of new digital mammography systems: physical characterisation and image quality evaluation Comparison of new digital mammography systems: physical characterisation and image quality evaluation Poster No.: C-2321 Congress: ECR 2011 Type: Scientific Exhibit Authors: N. Oberhofer; Bolzano/IT DOI:

More information

An aging study of the signal and noise characteristics in large-area CMOS detectors

An aging study of the signal and noise characteristics in large-area CMOS detectors An aging study of the signal and noise characteristics in large-area CMOS detectors Jong Chul Han a, Seungman Yun a, Chang Hwy Lim a, Tae Gyun Youm a, Sung Kyn Heo b, Tae Woo Kim b, Ian Cunningham c, Ho

More information

New Exposure Indicators for Digital Radiography Simplified for Radiologists and Technologists

New Exposure Indicators for Digital Radiography Simplified for Radiologists and Technologists Medical Physics and Informatics Technical Innovation Don et al. New Simplified Exposure Indicators Medical Physics and Informatics Technical Innovation Steven Don 1 ruce R. Whiting 2 Lois Jo Rutz 3 ruce

More information

AeroDR X60. Motorized X-ray System

AeroDR X60. Motorized X-ray System AeroDR X60 Motorized X-ray System AeroDR X60 2 MOTORIZED X-RAY SYSTEM The AeroDR X60 is a high quality, cost effective, compact motorized floor mounted X-ray system, built around Konica Minolta s renowned

More information

Evaluation of no-grid radiography using the digital scattered x-ray removal processing

Evaluation of no-grid radiography using the digital scattered x-ray removal processing Evaluation of no-grid radiography using the digital scattered x-ray removal processing Poster No.: C-0416 Congress: ECR 2016 Type: Authors: Scientific Exhibit R. Suzuki 1, T. Goto 1, H. Ogawa 2, N. Amimoto

More information

Digital Detector Array Image Quality for Various GOS Scintillators

Digital Detector Array Image Quality for Various GOS Scintillators Digital Detector Array Image Quality for Various GOS Scintillators More info about this article: http://www.ndt.net/?id=22768 Brian S. White 1, Mark E. Shafer 2, William H. Russel 3, Eric Fallet 4, Jacques

More information

Imaging Technique Optimization of Tungsten Anode FFDM System

Imaging Technique Optimization of Tungsten Anode FFDM System Imaging Technique Optimization of Tungsten Anode FFDM System Biao Chen a*, Andrew P. Smith b, Zhenxue Jing a, Elena Ingal a a Hologic, Inc. 600 Technology Drive, DE 1970 b Hologic, Inc. 35 Crosby Drive,

More information

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter Revision Date: January 5th, 2017 www.disc-imaging.com Table of Contents Section A: Preliminary Setup Requirements... 4 Tools

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

Luminos RF Classic. Where value meets performance.

Luminos RF Classic. Where value meets performance. Luminos RF Classic Where value meets performance www.siemens.com/healthcare What s good value in fluoroscopy? That s easy. Luminos RF Classic. 2 Whether for its handling convenience, outstanding image

More information

STATUS AND PROSPECTS OF DIGITAL DETECTOR TECHNOLOGY FOR CR AND DR Ulrich Neitzel Philips Medical Systems, Röntgenstrasse 24, D Hamburg, Germany

STATUS AND PROSPECTS OF DIGITAL DETECTOR TECHNOLOGY FOR CR AND DR Ulrich Neitzel Philips Medical Systems, Röntgenstrasse 24, D Hamburg, Germany Radiation Protection Dosimetry (2005), Vol. 114, Nos 1-3, pp. 32 38 doi:10.1093/rpd/nch532 INVITED PAPER STATUS AND PROSPECTS OF DIGITAL DETECTOR TECHNOLOGY FOR CR AND DR Ulrich Neitzel Philips Medical

More information

MILADY. Product Data. Page 1 of 8

MILADY. Product Data. Page 1 of 8 Page 1 of 8 The MILADY Mammographic Unit offers the best quality-to-price ratio to our customers worldwide. The unit advanced technology together with the application of industrial production standards,

More information

The importance of radiation quality for optimisation in radiology

The importance of radiation quality for optimisation in radiology Available online at http://www.biij.org/2007/2/e38 doi: 10.2349/biij.3.2.e38 biij Biomedical Imaging and Intervention Journal COMMENTARY The importance of radiation quality for optimisation in radiology

More information

QC in Diagnostic Radiology. Main steps for a QC survey in Diagnostic Radiology

QC in Diagnostic Radiology. Main steps for a QC survey in Diagnostic Radiology EVALUATING X-RAY TUBE AND GENERATOR PERFORMANCE : DEMO for PRACTICAL QUALITY CONTROL (QC) Dr Slavik Tabakov Dept. Medical Eng. & Physics, King's College London slavik.tabakov@kcl.ac.uk QC in Diagnostic

More information

universal Digital Radiography (DR) System

universal Digital Radiography (DR) System Vision Air universal Digital Radiography (DR) System Technical Specification Vision Air is a state-of-the-art, universal, overhead digital radiography system capable of virtually all radiography examinations

More information

Evaluation of a Novel Wafer-Scale CMOS APS X- Ray Detector for Use in Mammography

Evaluation of a Novel Wafer-Scale CMOS APS X- Ray Detector for Use in Mammography Evaluation of a Novel Wafer-Scale CMOS APS X- Ray Detector for Use in Mammography Anastasios C. Konstantinidis, Member, IEEE, Yi Zheng, Alessandro Olivo, Kristina Bliznakova, Mary Yip, Thalis Anaxagoras,

More information

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

Unit thickness. Unit area. σ = NΔX = ΔI / I 0 Unit thickness I 0 ΔI I σ = ΔI I 0 NΔX = ΔI / I 0 NΔX Unit area Δx Average probability of reaction with atom for the incident photons at unit area with the thickness of Delta-X Atom number at unit area

More information

Digital radiography (DR) post processing techniques for pediatric radiology

Digital radiography (DR) post processing techniques for pediatric radiology Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions

More information

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING DIGITAL IMAGE PROCESSING IN X-RAY IMAGING Shalini Kumari 1, Bachan Prasad 2,Aliya Nasim 3 Department of Electronics And Communication Engineering R.V.S College of Engineering & Technology, Jamshedpur,

More information

CT radiation profile width measurement using CR imaging plate raw data

CT radiation profile width measurement using CR imaging plate raw data JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 CT radiation profile width measurement using CR imaging plate raw data Thorarin A Bjarnason, 1,2,3a Chang-Ying Joseph Yang 3,4 Diagnostic

More information

Attikon, Rimini 1, , Athens, Greece , Athens, Greece , Athens, Greece

Attikon, Rimini 1, , Athens, Greece , Athens, Greece , Athens, Greece Radiation Protection Dosimetry (2005), Vol. 117, No. 1 3, pp. 291 297 doi:10.1093/rpd/nci742 Advance Access published on February 7, 2006 QUALITY ASSURANCE (QA) PROCEDURES FOR SOFTWARE: EVALUATION OF AN

More information

Unfors EDD-30 Radiation Protection in Fluoroscopy

Unfors EDD-30 Radiation Protection in Fluoroscopy Unfors EDD-30 Radiation Protection in Fluoroscopy Immediate Warning Decrease Your Dose Interventional radiology procedures are considered to be essential to medical diagnosis and treatment. It is recognized,

More information

AeroDR HD. Wireless Digital Radiography System

AeroDR HD. Wireless Digital Radiography System AeroDR HD Wireless Digital Radiography System AeroDR HD 2 Konica Minolta has developed the AeroDR HD: our most sophisticated detector with the highest resolution and sensitivity enabling the highest image

More information

Radiographic sensitivity improved by optimized high resolution X -ray detector design.

Radiographic sensitivity improved by optimized high resolution X -ray detector design. DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France Radiographic sensitivity improved by optimized high resolution X -ray detector

More information

X-ray detectors in healthcare and their applications

X-ray detectors in healthcare and their applications X-ray detectors in healthcare and their applications Pixel 2012, Inawashiro September 4th, 2012 Martin Spahn, PhD Clinical applications of X-ray imaging Current X-ray detector technology (case study radiography

More information

Testing a wavelet based noise reduction method using computersimulated

Testing a wavelet based noise reduction method using computersimulated Testing a wavelet based noise reduction method using computersimulated mammograms Christoph Hoeschen 1, Oleg Tischenko 1, David R Dance 2, Roger A Hunt 2, Andrew DA Maidment 3, Predrag R Bakic 3 1 GSF-

More information

Iranian Journal of Medical Physics

Iranian Journal of Medical Physics Iranian Journal of Medical Physics ijmp.mums.ac.ir A Phantom Study for the Optimization of Image Quality and Radiation Dose for Common Radiographic Examinations in Digital Radiography Soo-Foon Moey 1*,

More information

Interventional Radiological Equipment selection and installation

Interventional Radiological Equipment selection and installation Interventional Radiological Equipment selection and installation Renato Padovani ICTP Learning objectives To understand the main components of an interventional radiology equipment To understand the relevance

More information

Maximizing clinical outcomes

Maximizing clinical outcomes Maximizing clinical outcomes Digital Tomosynthesis Dual Energy Subtraction Automated Long Length Imaging Improved image quality at a low dose Xray Xray Patented ISS capture technology promotes high sensitivity

More information

Amorphous Selenium Direct Radiography for Industrial Imaging

Amorphous Selenium Direct Radiography for Industrial Imaging DGZfP Proceedings BB 67-CD Paper 22 Computerized Tomography for Industrial Applications and Image Processing in Radiology March 15-17, 1999, Berlin, Germany Amorphous Selenium Direct Radiography for Industrial

More information

Acquisition, Processing and Display

Acquisition, Processing and Display Acquisition, Processing and Display Terri L. Fauber, R.T. (R)(M) Department of Radiation Sciences School of Allied Health Professions Virginia Commonwealth University Topics Image Characteristics Image

More information

Aspire HD. Program Manual. 2nd Edition - October 2012

Aspire HD. Program Manual. 2nd Edition - October 2012 Quality Control 1 Aspire HD Quality Control Program Manual 2nd Edition - October 2012 Overview Installation of FDR Mammography QC Program Weekly Test 2 3 4 Quarterly Test 5 Semi-annual Test 6 Annual Test

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

CT-XGEO GC FTW EN

CT-XGEO GC FTW EN Samsung Electronics Co., Ltd. is a global leader in technology, opening new possibilities for people everywhere. Through relentless innovation and discovery, we are transforming the worlds of televisions,

More information

Introduction of Computed Radiography in Two Mammography Services: Image Quality and Dose Analysis

Introduction of Computed Radiography in Two Mammography Services: Image Quality and Dose Analysis Introduction of Computed Radiography in Two Mammography Services: Image Quality and Dose Analysis Rosangela Requi Jakubiak* a, Humberto Remigio Gamba a, Maria Manuela Ramos a, Gislene Gabrielle Faversani

More information

WHITE PAPER. Moving from CR to DR. Optimizing Image Quality and Dose

WHITE PAPER. Moving from CR to DR. Optimizing Image Quality and Dose Moving from CR to DR Optimizing Image Quality and Dose Table of contents 1. Introduction...3 2. CR and DR: similarities and differences...4 3. Image capture technologies...5 3.1 DR panel technology...5

More information

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices Moving from film to digital: A study of digital x-ray benefits, challenges and best practices H.U. Pöhler 1 and N. D Ademo 2 DÜRR NDT GmbH & Co. KG, Höpfigheimer Straße 22, Bietigheim-Bissingen, 74321,

More information

Response characteristics of a tandem ionization chamber in standard X-ray beams

Response characteristics of a tandem ionization chamber in standard X-ray beams Applied Radiation and Isotopes 58 (2003) 495 500 Response characteristics of a tandem ionization chamber in standard X-ray beams Alessandro M. Costa*, Linda V.E. Caldas Instituto de Pesquisas Energ!eticas

More information

RaySafe Xi. Specifications

RaySafe Xi. Specifications RaySafe Xi Specifications raysafe Xi General EMC tested Exposure needed Reset Temp. range According to EN 61000-6-1:2007 and en 61000-6-3:2007 One Automatic 15 35 C (59 95 F) DetECtor cable length 2 and

More information

REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS

REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS DEPARTMENT OF HEALTH DIRECTORATE: RADIATION CONTROL Implementation date: 31 March 2009 Contents

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 62220-1 First edition 2003-10 Medical electrical equipment Characteristics of digital X-ray imaging devices Part 1: Determination of the detective quantum efficiency Appareils

More information

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura New Technology https://www.youtube.com/watch?v=ptkzznazb 7U COMPUTED

More information

DRX Plus Detectors: Going from Good to Great

DRX Plus Detectors: Going from Good to Great DRX Plus Detectors: Going from Good to Great Authors: Karin Töpfer, Tim Wojcik Introduction Carestream s introduction in 2009 of the world s first portable, wireless, cassette-sized detector the CARESTREAM

More information

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah X-RAY IMAGING EE 472 F2017 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Stewart C. Bushong, Radiologic Science for Technologists: Physics, Biology, and Protection, 10 th ed., Mosby,

More information

Determination of the tube voltage from clinic mammographic system using two types of detectors

Determination of the tube voltage from clinic mammographic system using two types of detectors BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 03-1A (2015) 01-06 Determination of the tube voltage from clinic mammographic system using two types of detectors J. S. Barreira a ; V. Vivolo a a Gerência

More information

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation Image Quality Review I: Basics and Image Quality TH-A-16A-1 Thursday 7:30AM - 9:30AM Room: 16A J. Anthony

More information

Outline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control

Outline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control Understanding Digital Modalities: Image Quality and Dose S. Jeff Shepard, M.S. University of Texas M. D. Anderson Cancer Center Houston, Texas Special Acknowledgement: Stephen K. Thompson, M.S. William

More information

Image Quality Artifacts in Digital Imaging

Image Quality Artifacts in Digital Imaging MAHIDOL UNIVERSITY Wisdom of the Land Image Quality Artifacts in Digital Imaging Napapong Pongnapang, Ph.D. Department of Radiological Technology Faculty of Medical Technology Mahidol University, Bangkok,

More information