Estimation of signal transfer property for wireless digital detector in different measurement schemes
|
|
- Rodney Peters
- 6 years ago
- Views:
Transcription
1 Estimation of signal transfer property for wireless digital detector in different measurement schemes Anatoli Vladimirov, Kalle Kepler Training Centre of Medical Physics, University of Tartu, Estonia 11 TH International Conference & Workshop Medical Physics in the Baltic Satates, Kaunas, October, 2013
2 Introduction European criteria for acceptability of medical radiological equipment Radiation Protection 162 include and require measurement of signal transfer property (STP). STP refers to a test to be done during the acceptance testing of the computed radiography (CR) or digital radiography (DR) system in order to establish the relationship between detector air kerma (DAK) (or receptor dose) and pixel value (PV).
3 Radiation Protection 162
4 Simple STP ralationship Linear: PV lin = a+b DAK DAK = (PV-a)/b Logarithmic: PV log = a ln(dak)+b DAK = exp((pv-a)/b) Power: PV power = a DAK b +c DAK = ((PV-c)/a) 1/b
5 Linearization of standard deviation The measured standard deviation of the pixel values (σ) cannot simply be corrected to the true standard deviation (σ') using the inverse of the STP. For measurements of several objective parameters of image quality and imaging system: signal-to-noise ratio (SNR), modulation transfer function (MTF), noise power spectrum (NPS), detective quantum efficiency (DQE) the linearized standard deviation can then be estimated by subtracting the linearized mean pixel value: σ'= (PV + σ)' - PV' where ( ) is the linearization function using the inverse of the STP
6 Beam conditions used for DDI in CR Manufacturer Added filter kv Pre-processing Agfa 1.5 mm Cu 75 S=200 / System Diagnosis / Flat Field Fuji 0 mm Al 80 Semi-Auto / L=1 Kodak 0.5 mm Cu + 1 mm Al 80 Pattern mode Konica 0 mm Al 80 Test 1/G =2/F off
7 Beam conditions used for STP Spectrum Added filter kv Nominal HVL (mm Al) Measured HVL (mm Al) IPEM 32 1 mm Cu 70 ~ RQA 5 by IEC RQA 3 by IEC mm Al mm Al PMMA 0 mm Al
8 Materials and methods Scatter geometry (a) X-ray tube (b) Added filter X-ray tube Scatter-free geometry Wide X-ray beam Narrow X-ray beam beam Dose chamber Table PMMA Table Grid Image detector or dose chamber Grid removed Image detector or dose chamber IPEM. Measurement of the Performance Characteristics of Diagnostic X- Ray Systems: Digital Imaging Systems. IPEM Report Number 32, Part VII (2010)
9 Materials and methods Siemens Ysio and Pixium FE 3543pR (Trixell,France) with a CsI scintillator coupled to TFT matrix in a:si -technology, pixel size 155 µm, matrix 34.2 x 43.2 cm, saturation dose: 75 Gy. Barracuda MPD dosimeter (RTI, Sweden), accuracy of 5%. Siemens 2011
10 Setting the amplification factor Detector dose Indicative sensitivity Typical amplification factor 7.1 μgy μgy μgy μgy μgy μgy Siemens 2011
11 Setting the gradation curve (LUT) Siemens 2011
12 Pixel value Results and discussion PV = 54.5 DAK R 2 = PV = 54.0 DAK R 2 = PV = 27.2 DAK R 2 = ,00 10,00 20,00 30,00 40,00 50,00 60,00 Detector air kerm a (μgy) IPEM32 RQA5 RQA3 STP for RQA5, RQA3 and IPEM32 schemes, measured directly (without table, grid)
13 Pixel value Results and discussion PV = DAK R 2 = PV= 55.0 DAK R 2 = PV = 54.4 DAK R 2 = 0, ,00 10,00 20,00 30,00 40,00 50,00 60,00 Detector air kerm a (μgy) IPEM32, w ithout table and grid IPEM32, in Bucky w ithout grid IPEM32, in Bucky w ith grid STP by IPEM32 on table and inside table with and without grid.
14 Pixel value Results and discussion PV = 54.9 DAK R 2 = PV = 55.0 DAK R 2 = ,00 10,00 20,00 30,00 40,00 50,00 60,00 Detector air kerm a (μgy) IPEM32, scatter-free IPEM32, scatter (PMMA) STP in IPEM32 scatter (PMMA) and scatter-free geometry (with dose detector inside the Bucky).
15 Pixel value Results and discussion PV = DAK R 2 = PV = DAK R 2 = PV = 54.4 DAK R 2 = ,00 5,00 10,00 15,00 20,00 25,00 30,00 35,00 40,00 Detector air kerm a (μgy) IPEM32, Ampl.1 IPEM32, Ampl.2 IPEM32, Ampl.4 STP with the amplification factors of 1, 2 and 4.
16 Pixel value Results and discussion PV = Ln(DAK) R 2 = PV = Ln(DAK) R 2 = ,00 5,00 10,00 15,00 20,00 Detector air kerm a (μgy) RQA5 RQA3 STP for RQA5 and RQA3 with typical post-processing with DAK up to 20 µgy.
17 Pixel value Results and discussion RQA5: PV = Ln(DAK ) R 2 = RQA3: PV = Ln(DAK) R 2 = ,00 10,00 20,00 30,00 40,00 50,00 60,00 70,00 Detector air kerm a (μgy) RQA5 RQA3 STP for RQA5 and RQA3 with typical post-processing with DAK up to 70 µgy.
18 Linearized standard diviation Results and discussion 0,10 σ = DAK 0.47 σ = DAK 0.35 σ = DAK 0.48 RQA5 IPEM32 RQA3 0,01 0,10 1,00 10,00 100,00 Detector air kerm a (μgy) Linearized standard deviation of pixel value by RQA5, RQA3 and IPEM32 measured directly on the receptor.
19 Conclusions Several measurement schemes using beam qualities RQA3, RQA5 and IPEM32 were studied along with three different geometries: scatter (PMMA), scatter-free and direct measurement geometry for estimation of STP of a digital wireless detector. For the pre-processed images the more clinically applicable IPEM32 scheme is compared with the IEC RQA5 scheme. The results show that there is no significant difference in the coefficient of correlation R 2 for the STP curve. Also power coefficients for the measured linearized standard deviation dependence from DAK are comparable for these schemes. The different IPEM32 geometries for measurements without table, inside table with and without grid could be interchangeable in practice.
20 Conclusions The post-processed images could be linearizable up to 20 µgy if simple LUT was applied. One conclusion is that if we have to measure STP on non-wireless Bucky, where it is not possible to measure DAK directly on the entrance surface of the receptor, then the dose measurement could be done on the entrance of the table (or vertical stand), and recalculate it on the right distance. If we know transmission of grid and table we could calculate exact DAK.
21 References Bosmans H., Nens J., Delzenne L., et al. Exploration of exposure conditions with a novel wireless detector for bedside digital radiography. Proc. of SPIE, 2012, 8313, p K1-9. Samei E. DQE of wireless digital detectors: Comparative performance with differing filtration schemes. Med. Phys., 2013, 40, p European Commission. Criteria for acceptability of medical radiological equipment used in diagnostic radiology, nuclear medicine and radiotherapy. Rad. Prot Luxembourg: EC, 2012, 84 p. Institute of Physics and Engineering in Medicine. Measurement of the performance characteristics of diagnostic X-ray systems: digital imaging systems. IPEM Rep. 32, Part VII. IPEM, p. Institute of Physics and Engineering in Medicine. Recommended standards for the routine performance testing of diagnostic X-ray imaging systems. IPEM Report 91. IPEM, International Electrotechnical Commission. Medical diagnostic X-ray equipment Radiation conditions for use in the determination of characteristics Edition 2.0. IEC Geneva: IEC (2005). Kepler K., Vladimirov A. Survey of compliance with European acceptability criteria for HVL and AEC. Rad. Prot. Dos., 2013, 153(2), Siemens AG. FLUOROSPOT Compact. Operator manual. Imaging System for Ysio. Erlangen (2011). Mackenzie, A. Validation of correction methods for the non-linear response of digital radiography systems. The Brit. J. of Rad. 81, (2008). Marshall N. W., Mackenzie A., Honey, I. D. Quality control measurements for digital x-ray detectors. Phys. in Med. Biol. 56, (2011).
22 Thank you!
Half value layer and AEC receptor dose compliance survey in Estonia
Half value layer and AEC receptor dose compliance survey in Estonia K. Kepler, A. Vladimirov Training Centre of Medical Physics, University of Tartu Testing Centre of the University of Tartu, Estonia E-mail:
More informationDisclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography
Current Implementation Status of IEC Standard 62494-1: Exposure Index (EI) for Digital Radiography July 31, 2017 Ryan Fisher, PhD, DABR Katie Hulme, MS, DABR None Disclosures Outline Review of IEC Standard
More informationExposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?
Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago
More informationInvestigation of Effective DQE (edqe) parameters for a flat panel detector
Investigation of Effective DQE (edqe) parameters for a flat panel detector Poster No.: C-1892 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit D. Bor 1, S. Cubukcu 1, A. Yalcin 1, O.
More informationPredicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities
Journal of Physics: Conference Series PAPER OPEN ACCESS Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities Recent citations - Resolution Properties of a
More informationLearning Objectives: What s my motivation? (unknown screen actor) Workshop Overview
Practical Medical Physics Adapting Traditional Clinical Medical Physics to Digital Radiography Charles E. Willis, Ph.D., DABR Associate Professor Department of Imaging Physics The University of Texas M.D.
More informationAsk EuroSafe Imaging Tips & Tricks. Paediatric Imaging Working Group. Dose Management in Digital Radiography
Ask EuroSafe Imaging Tips & Tricks Paediatric Imaging Working Group Dose Management in Digital Radiography Raija Seuri (HUS Medical Imaging Center, FI) Cristina Almeida (Centro Hospitalar de Lisboa Central,
More informationPractical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S.
Practical Medical Physics Session: TG-151 Dose Monitoring August 5, 2013 Katie Hulme, M.S. Digital Imaging and Dose Creep Images courtesy of Agfa Healthcare Under-Exposed Over-Exposed Freedman et al.,
More informationFeatures and Weaknesses of Phantoms for CR/DR System Testing
Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal
More informationTailoring automatic exposure control toward constant detectability in digital mammography
Tailoring automatic exposure control toward constant detectability in digital mammography Elena Salvagnini a) Department of Imaging and Pathology, Medical Physics and Quality Assessment, KUL, Herestraat
More informationJoint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009
2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced
More informationINTERNATIONAL STANDARD
INTERNATIONAL STANDARD IEC 62220-1 First edition 2003-10 Medical electrical equipment Characteristics of digital X-ray imaging devices Part 1: Determination of the detective quantum efficiency Appareils
More informationModeling detector performance in digital mammography using the linear cascaded systems approach
51 Modeling detector performance in digital mammography using the linear cascaded systems approach V. Spyropoulou a,b, N. Kalivas a,b, A. Gaitanis a,b, C. Michail a,b, G. Panayiotakis b, I. Kandarakis
More informationY11-DR Digital Radiography (DR) Image Quality
Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures
More informationDigital Radiography X-Ray System. X Twin with X Mobil Roesys GmbH [rshsmi] 1/11
Digital Radiography X-Ray System 2017 Roesys GmbH [rshsmi] 1/11 1 General specifications The system is intended to be installed with floor mounted components only. It enables spacesaving installation without
More informationCalibration of KAP meters
Calibration of KAP meters Alexandr Malusek! Division of Radiological Sciences Department of Medical and Health Sciences Linköping University! 2014-04-15 1 Outline 1. KAP meter construction 2. Air kerma-area
More informationQuality Control of Full Field Digital Mammography Units
Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of
More informationIntroduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image
Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology
More informationPractical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities
Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Melissa Martin, M.S., FAAPM, FACR, FACMP AAPM Annual Meeting - Philadelphia July 19, 2010 MO-B-204C-1 Educational Objectives
More informationA comparison of two methods for the determination of freein-air geometric efficiency in MDCT
A comparison of two methods for the determination of freein-air geometric efficiency in MDCT Theocharis Berris *1, Kostas Perisinakis 1,, Antonios E. Papadakis and John Damilakis 1, 1 Department of Medical
More informationKey words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure
Accuracy and calibration of integrated radiation output indicators in diagnostic radiology: A report of the AAPM Imaging Physics Committee Task Group 190 Pei-Jan P. Lin a) Virginia Commonwealth University
More informationQuality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals
DOI 10.1007/s00247-011-2290-1 ORIGINAL ARTICLE Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals Mervyn D. Cohen & Richard Markowitz
More informationI. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA
FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE
More informationComparison of computed radiography and filmõscreen combination using a contrast-detail phantom
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 1, WINTER 2003 Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom Z. F. Lu,* E. L. Nickoloff, J.
More information2217 US Highway 70 East Garner, NC Main: Fax:
Viztek is committed to providing the highest image quality possible in our CR & DR product lines. There are several factors that directly affect the overall quality of CR & DR based images. The eposure
More informationPreliminary studies of a new monitor ionization chamber
1 Preliminary studies of a new monitor ionization chamber Maíra T. Yoshizumi, Vitor Vivolo and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares (IPEN-CNEN) Comissão Nacional de Energia
More informationAssessment and Optimisation of Digital Radiography Systems for Clinical Use
Assessment and Optimisation of Digital Radiography Systems for Clinical Use Philip Doyle, BSc. MSc. Submitted in fulfilment of the requirements for the degree of Doctor of Philosophy Department of Clinical
More informationDigital mammography: DQE versus optimized image quality in clinical environment an on site study
Digital mammography: DQE versus optimized image quality in clinical environment an on site study Nadia Oberhofer* a, Alessandro Fracchetti a, Margareth Springeth a, Ehrenfried Moroder a a Health Service
More informationX-RAY MEDICAL EQUIPMENT
X-RAY MEDICAL EQUIPMENT CHEST RADIOGRAPHY GENERAL RADIOGRAPHY & FLUOROSCOPY RADIOTHERAPY MOBILE HEALTHCARE MAMMOGRAPHY MAMMOSCAN FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM Biopsy Attachment џ MAMMOSCAN an ADANI
More informationExposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations
Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations S. Baechler 1, P. Monnin 1, A. Aroua 1, J.F. Valley 1, M. Perrier, P. Trueb 3, F.R. Verdun 1 1 University
More informationSAMSUNG SMART DIGITAL RADIOGRAPHY. XGEO GU60A series REDEFINING ERGONOMICS
SAMSUNG SMART DIGITAL RADIOGRAPHY XGEO GU60A series REDEFINING ERGONOMICS U-arm Positioner (Fully Automated) The XGEO GU60A/60A-65 is a universal, fully motorized system. Its unique U-arm rotates +120
More informationI. Introduction.
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Accuracy of measuring half- and quarter-value layers and appropriate aperture width of a convenient method using a lead-covered case
More informationFully Automatic X-ray meter kvp, dose, rate, HVL and time Extremely easy to use. Unfors ThinX RAD for Radiographic Quality Control
Fully Automatic X-ray meter kvp, dose, rate, HVL and time Extremely easy to use Unfors ThinX RAD for Radiographic Quality Control Unfors ThinX RAD No keys! No menus! Just position and expose! Unfors ThinX
More informationSPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by
SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This
More informationDerivation of total filtration thickness for diagnostic x-ray source assembly
Physics in Medicine & Biology PAPER OPEN ACCESS Derivation of total filtration thickness for diagnostic x-ray source assembly To cite this article: Michiharu Sekimoto and Yoh Katoh 2016 Phys. Med. Biol.
More information10/3/2012. Study Harder
This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen
More informationDevelopment of a large-area CMOS-based detector for real-time x-ray imaging
Development of a large-area CMOS-based detector for real-time x-ray imaging Sung Kyn Heo a, Sung Kyu Park a, Sung Ha Hwang a, Dong Ak Im a, Jari Kosonen a, Tae Woo Kim a, Seungman Yun b, Ho Kyung Kim b*
More information10/26/2015. Study Harder
This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen
More informationTESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist
TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist Topics Image Uniformity and Artifacts Image Quality - Detail
More informationA comparative study of several digital flat panel X-ray units: patients doses and image quality in chest radiography
A comparative study of several digital flat panel X-ray units: patients doses and image quality in chest radiography Torres Cabrera R. 1, España López M.L. 2 Ruiz Manzano P. 3, Sastre Aguado J.M. 4,, Rivas
More informationAcceptance Testing of a Digital Breast Tomosynthesis Unit
Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures
More informationMammography: Physics of Imaging
Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness
More informationSTEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS
STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%
More informationDigital Mammography: From Physical Performance Evaluation To Image Quality Analysis
Digital Mammography: From Physical Performance Evaluation To Image Quality Analysis Anastasios C. Konstantinidis Department of Medical Physics and Bioengineering, University College London, UK. taskon25@yahoo.gr
More informationVOLUMETRIC computed tomography (CT) based on flatpanel
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 3, JUNE 2009 1179 Radiation Effects on the Performance of CMOS Photodiode Array Detectors and the Role of Gain-Offset Corrections Ho Kyung Kim, Min Kook
More informationEvaluation and Comparison of Image Quality for Indirect Flat Panel Systems with CsI and GOS Scintillators
Evaluation and Comparison of Image Quality for Indirect Flat Panel Systems with CsI and GOS Scintillators Mehmet E. Aksoy a, Mustafa E. Kamasak b, Erkan Akkur c, Ayhan Ucgul d, Muzaffer Basak d, Huseyin
More informationDRAFT Technical evaluation of Philips Microdose SI digital mammography system
DRAFT Technical evaluation of Philips Microdose SI digital mammography system NHSBSP Equipment Report 1310 August 2013 About the NHS Cancer Screening Programmes The national office of the NHS Cancer Screening
More informationDigital Radiography System
www.drgem.co.kr GXR-SD SERIES GXR-SD SERIES Digital Radiography System GXR-SD Series www.drgem.co.kr High Performance and Reliability The Optimal System for Digital Solution The GXR-SD Series is a diagnostic
More informationSUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES.
SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES. Tingberg, Anders Published in: Radiation Protection Dosimetry DOI: 10.1093/rpd/ncs302 Published: 2013-01-01 Link to publication Citation for published
More informationThe Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System
The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System Manus
More informationDigital Imaging Considerations Computed Radiography
Digital Imaging Considerations Digital Radiography Computed Radiography o Cassette based Direct or Indirect Digital Radiography o Cassetteless Computed Radiography 1 CR Image Acquisition Most like conventional
More informationAn Exposure Indicator for Digital Radiography. Report of AAPM Task Group 116
AAPM REPORT NO. 116 An Exposure Indicator for Digital Radiography Report of AAPM Task Group 116 July 2009 DISCLAIMER: This publication is based on sources and information believed to be reliable, but the
More informationComparison of new digital mammography systems: physical characterisation and image quality evaluation
Comparison of new digital mammography systems: physical characterisation and image quality evaluation Poster No.: C-2321 Congress: ECR 2011 Type: Scientific Exhibit Authors: N. Oberhofer; Bolzano/IT DOI:
More informationAn aging study of the signal and noise characteristics in large-area CMOS detectors
An aging study of the signal and noise characteristics in large-area CMOS detectors Jong Chul Han a, Seungman Yun a, Chang Hwy Lim a, Tae Gyun Youm a, Sung Kyn Heo b, Tae Woo Kim b, Ian Cunningham c, Ho
More informationNew Exposure Indicators for Digital Radiography Simplified for Radiologists and Technologists
Medical Physics and Informatics Technical Innovation Don et al. New Simplified Exposure Indicators Medical Physics and Informatics Technical Innovation Steven Don 1 ruce R. Whiting 2 Lois Jo Rutz 3 ruce
More informationAeroDR X60. Motorized X-ray System
AeroDR X60 Motorized X-ray System AeroDR X60 2 MOTORIZED X-RAY SYSTEM The AeroDR X60 is a high quality, cost effective, compact motorized floor mounted X-ray system, built around Konica Minolta s renowned
More informationEvaluation of no-grid radiography using the digital scattered x-ray removal processing
Evaluation of no-grid radiography using the digital scattered x-ray removal processing Poster No.: C-0416 Congress: ECR 2016 Type: Authors: Scientific Exhibit R. Suzuki 1, T. Goto 1, H. Ogawa 2, N. Amimoto
More informationDigital Detector Array Image Quality for Various GOS Scintillators
Digital Detector Array Image Quality for Various GOS Scintillators More info about this article: http://www.ndt.net/?id=22768 Brian S. White 1, Mark E. Shafer 2, William H. Russel 3, Eric Fallet 4, Jacques
More informationImaging Technique Optimization of Tungsten Anode FFDM System
Imaging Technique Optimization of Tungsten Anode FFDM System Biao Chen a*, Andrew P. Smith b, Zhenxue Jing a, Elena Ingal a a Hologic, Inc. 600 Technology Drive, DE 1970 b Hologic, Inc. 35 Crosby Drive,
More informationDISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter
DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter Revision Date: January 5th, 2017 www.disc-imaging.com Table of Contents Section A: Preliminary Setup Requirements... 4 Tools
More informationDose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp
Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division
More informationLuminos RF Classic. Where value meets performance.
Luminos RF Classic Where value meets performance www.siemens.com/healthcare What s good value in fluoroscopy? That s easy. Luminos RF Classic. 2 Whether for its handling convenience, outstanding image
More informationSTATUS AND PROSPECTS OF DIGITAL DETECTOR TECHNOLOGY FOR CR AND DR Ulrich Neitzel Philips Medical Systems, Röntgenstrasse 24, D Hamburg, Germany
Radiation Protection Dosimetry (2005), Vol. 114, Nos 1-3, pp. 32 38 doi:10.1093/rpd/nch532 INVITED PAPER STATUS AND PROSPECTS OF DIGITAL DETECTOR TECHNOLOGY FOR CR AND DR Ulrich Neitzel Philips Medical
More informationMILADY. Product Data. Page 1 of 8
Page 1 of 8 The MILADY Mammographic Unit offers the best quality-to-price ratio to our customers worldwide. The unit advanced technology together with the application of industrial production standards,
More informationThe importance of radiation quality for optimisation in radiology
Available online at http://www.biij.org/2007/2/e38 doi: 10.2349/biij.3.2.e38 biij Biomedical Imaging and Intervention Journal COMMENTARY The importance of radiation quality for optimisation in radiology
More informationQC in Diagnostic Radiology. Main steps for a QC survey in Diagnostic Radiology
EVALUATING X-RAY TUBE AND GENERATOR PERFORMANCE : DEMO for PRACTICAL QUALITY CONTROL (QC) Dr Slavik Tabakov Dept. Medical Eng. & Physics, King's College London slavik.tabakov@kcl.ac.uk QC in Diagnostic
More informationuniversal Digital Radiography (DR) System
Vision Air universal Digital Radiography (DR) System Technical Specification Vision Air is a state-of-the-art, universal, overhead digital radiography system capable of virtually all radiography examinations
More informationEvaluation of a Novel Wafer-Scale CMOS APS X- Ray Detector for Use in Mammography
Evaluation of a Novel Wafer-Scale CMOS APS X- Ray Detector for Use in Mammography Anastasios C. Konstantinidis, Member, IEEE, Yi Zheng, Alessandro Olivo, Kristina Bliznakova, Mary Yip, Thalis Anaxagoras,
More informationUnit thickness. Unit area. σ = NΔX = ΔI / I 0
Unit thickness I 0 ΔI I σ = ΔI I 0 NΔX = ΔI / I 0 NΔX Unit area Δx Average probability of reaction with atom for the incident photons at unit area with the thickness of Delta-X Atom number at unit area
More informationDigital radiography (DR) post processing techniques for pediatric radiology
Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions
More informationDIGITAL IMAGE PROCESSING IN X-RAY IMAGING
DIGITAL IMAGE PROCESSING IN X-RAY IMAGING Shalini Kumari 1, Bachan Prasad 2,Aliya Nasim 3 Department of Electronics And Communication Engineering R.V.S College of Engineering & Technology, Jamshedpur,
More informationCT radiation profile width measurement using CR imaging plate raw data
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 CT radiation profile width measurement using CR imaging plate raw data Thorarin A Bjarnason, 1,2,3a Chang-Ying Joseph Yang 3,4 Diagnostic
More informationAttikon, Rimini 1, , Athens, Greece , Athens, Greece , Athens, Greece
Radiation Protection Dosimetry (2005), Vol. 117, No. 1 3, pp. 291 297 doi:10.1093/rpd/nci742 Advance Access published on February 7, 2006 QUALITY ASSURANCE (QA) PROCEDURES FOR SOFTWARE: EVALUATION OF AN
More informationUnfors EDD-30 Radiation Protection in Fluoroscopy
Unfors EDD-30 Radiation Protection in Fluoroscopy Immediate Warning Decrease Your Dose Interventional radiology procedures are considered to be essential to medical diagnosis and treatment. It is recognized,
More informationAeroDR HD. Wireless Digital Radiography System
AeroDR HD Wireless Digital Radiography System AeroDR HD 2 Konica Minolta has developed the AeroDR HD: our most sophisticated detector with the highest resolution and sensitivity enabling the highest image
More informationRadiographic sensitivity improved by optimized high resolution X -ray detector design.
DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France Radiographic sensitivity improved by optimized high resolution X -ray detector
More informationX-ray detectors in healthcare and their applications
X-ray detectors in healthcare and their applications Pixel 2012, Inawashiro September 4th, 2012 Martin Spahn, PhD Clinical applications of X-ray imaging Current X-ray detector technology (case study radiography
More informationTesting a wavelet based noise reduction method using computersimulated
Testing a wavelet based noise reduction method using computersimulated mammograms Christoph Hoeschen 1, Oleg Tischenko 1, David R Dance 2, Roger A Hunt 2, Andrew DA Maidment 3, Predrag R Bakic 3 1 GSF-
More informationIranian Journal of Medical Physics
Iranian Journal of Medical Physics ijmp.mums.ac.ir A Phantom Study for the Optimization of Image Quality and Radiation Dose for Common Radiographic Examinations in Digital Radiography Soo-Foon Moey 1*,
More informationInterventional Radiological Equipment selection and installation
Interventional Radiological Equipment selection and installation Renato Padovani ICTP Learning objectives To understand the main components of an interventional radiology equipment To understand the relevance
More informationMaximizing clinical outcomes
Maximizing clinical outcomes Digital Tomosynthesis Dual Energy Subtraction Automated Long Length Imaging Improved image quality at a low dose Xray Xray Patented ISS capture technology promotes high sensitivity
More informationAmorphous Selenium Direct Radiography for Industrial Imaging
DGZfP Proceedings BB 67-CD Paper 22 Computerized Tomography for Industrial Applications and Image Processing in Radiology March 15-17, 1999, Berlin, Germany Amorphous Selenium Direct Radiography for Industrial
More informationAcquisition, Processing and Display
Acquisition, Processing and Display Terri L. Fauber, R.T. (R)(M) Department of Radiation Sciences School of Allied Health Professions Virginia Commonwealth University Topics Image Characteristics Image
More informationAspire HD. Program Manual. 2nd Edition - October 2012
Quality Control 1 Aspire HD Quality Control Program Manual 2nd Edition - October 2012 Overview Installation of FDR Mammography QC Program Weekly Test 2 3 4 Quarterly Test 5 Semi-annual Test 6 Annual Test
More informationBreast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School
Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis
More informationCT-XGEO GC FTW EN
Samsung Electronics Co., Ltd. is a global leader in technology, opening new possibilities for people everywhere. Through relentless innovation and discovery, we are transforming the worlds of televisions,
More informationIntroduction of Computed Radiography in Two Mammography Services: Image Quality and Dose Analysis
Introduction of Computed Radiography in Two Mammography Services: Image Quality and Dose Analysis Rosangela Requi Jakubiak* a, Humberto Remigio Gamba a, Maria Manuela Ramos a, Gislene Gabrielle Faversani
More informationWHITE PAPER. Moving from CR to DR. Optimizing Image Quality and Dose
Moving from CR to DR Optimizing Image Quality and Dose Table of contents 1. Introduction...3 2. CR and DR: similarities and differences...4 3. Image capture technologies...5 3.1 DR panel technology...5
More informationMoving from film to digital: A study of digital x-ray benefits, challenges and best practices
Moving from film to digital: A study of digital x-ray benefits, challenges and best practices H.U. Pöhler 1 and N. D Ademo 2 DÜRR NDT GmbH & Co. KG, Höpfigheimer Straße 22, Bietigheim-Bissingen, 74321,
More informationResponse characteristics of a tandem ionization chamber in standard X-ray beams
Applied Radiation and Isotopes 58 (2003) 495 500 Response characteristics of a tandem ionization chamber in standard X-ray beams Alessandro M. Costa*, Linda V.E. Caldas Instituto de Pesquisas Energ!eticas
More informationRaySafe Xi. Specifications
RaySafe Xi Specifications raysafe Xi General EMC tested Exposure needed Reset Temp. range According to EN 61000-6-1:2007 and en 61000-6-3:2007 One Automatic 15 35 C (59 95 F) DetECtor cable length 2 and
More informationREQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS
REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS DEPARTMENT OF HEALTH DIRECTORATE: RADIATION CONTROL Implementation date: 31 March 2009 Contents
More informationINTERNATIONAL STANDARD
INTERNATIONAL STANDARD IEC 62220-1 First edition 2003-10 Medical electrical equipment Characteristics of digital X-ray imaging devices Part 1: Determination of the detective quantum efficiency Appareils
More informationCOMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR
This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura New Technology https://www.youtube.com/watch?v=ptkzznazb 7U COMPUTED
More informationDRX Plus Detectors: Going from Good to Great
DRX Plus Detectors: Going from Good to Great Authors: Karin Töpfer, Tim Wojcik Introduction Carestream s introduction in 2009 of the world s first portable, wireless, cassette-sized detector the CARESTREAM
More informationX-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah
X-RAY IMAGING EE 472 F2017 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Stewart C. Bushong, Radiologic Science for Technologists: Physics, Biology, and Protection, 10 th ed., Mosby,
More informationDetermination of the tube voltage from clinic mammographic system using two types of detectors
BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 03-1A (2015) 01-06 Determination of the tube voltage from clinic mammographic system using two types of detectors J. S. Barreira a ; V. Vivolo a a Gerência
More information7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline
Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation Image Quality Review I: Basics and Image Quality TH-A-16A-1 Thursday 7:30AM - 9:30AM Room: 16A J. Anthony
More informationOutline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control
Understanding Digital Modalities: Image Quality and Dose S. Jeff Shepard, M.S. University of Texas M. D. Anderson Cancer Center Houston, Texas Special Acknowledgement: Stephen K. Thompson, M.S. William
More informationImage Quality Artifacts in Digital Imaging
MAHIDOL UNIVERSITY Wisdom of the Land Image Quality Artifacts in Digital Imaging Napapong Pongnapang, Ph.D. Department of Radiological Technology Faculty of Medical Technology Mahidol University, Bangkok,
More information