Calibration of KAP meters
|
|
- Millicent Henderson
- 5 years ago
- Views:
Transcription
1 Calibration of KAP meters Alexandr Malusek! Division of Radiological Sciences Department of Medical and Health Sciences Linköping University!
2 Outline 1. KAP meter construction 2. Air kerma-area product 3. Calibration methods (NRPB, tandem, IAEA) 4. IAEA s NK based formalism 5. Calibration corrections 6. Suggestions for improvement 2
3 Introduction A KAP meter is a plane-parallel ionization chamber calibrated to measure the air kerma-area product. Old calibration methods (NRPB) cannot be used as films are no longer used in clinics New calibration methods (IAEA) have been developed but issues with accuracy still exist. Work on improvements continues 3
4 Accuracy of KAP meters Large uncertainties in PKA values measured by KAP meters in clinics are common owing to the strong energy dependence of these chambers. Manufacturers typically guarantee the accuracy of 25% (k=2) required by IEC specifications. In diagnostic radiology, IAEA s [3] and ICRU s [4] specifications require accuracy better than 7% (k=2 ). This holds true for PKA measurements too. 4
5 KAP meters sensitive area ~ 14 cm x 14 cm thickness ~ 1.6 cm conductive coating ~ 10 nm transparent to visible light 1.5 mm 5.9 mm 1.0 mm 5.9 mm 1.5 mm conductive coating outer electrode air cavity inner electrode air cavity outer electrode 5
6 X-ray tubes with collimator housing C-arm x-ray units 6
7 Air kerma-area product, PKA x ray tube Z P KA = A K air da reference plane Α Kair beam axis x ray tube Α PKA does not depend on the position of the reference plane if photons in the beam are neither scattered nor absorbed K air 1/d 2 A d 2 A d is the distance from focus 7
8 Photon scatter and absorption x ray tube KAP meter patient plane some photons registered by the KAP meter may not reach the patient plane owing to scatter or absorption some scattered photons may increase PKA at the patient plane Φ 1 Φ 1 Φ 2 > Scatter increases fluence of particles 8
9 Calibration methods National Radiological Protection Board (NRPB) [1] Tandem calibration (developed at STUK) [2] International Atomic Energy Agency (IAEA) [3] modification of NRPB s method for screen-film systems and computed radiography systems tandem calibration in IAEA s geometry laboratory calibration of the reference KAP meter 9
10 NRPB calibration method focal spot Features:! collimator KAP meter Kair is measured with an ionization chamber at the beam axis beam axis A is determined as the area within 50% of the maximum optical density optical density d m K air ionization chamber film PKA = Kair A Problems:! Kair measured at one point only d m /2 Films are no longer used 10
11 Tandem calibration method focal spot collimator clinical KAP meter Features:! Reference KAP meter measures PKA for incident radiation dcr ~ 30 cm to 40 cm d cr Attenuation in air is neglected Problems:! reference KAP meter A difference between clinical and standards laboratory beam qualities may result in systematic error A KAP meter holder is needed 11
12 IAEA calibration methods 20 cm 10 cm 95 cm 100 cm x ray tube KAP meter diagnostic detector styrofoam couch laboratory calibration 4 6 cm Features! K air is measured with detector at the beam axis Beam size 10 cm x 10 cm at the position of the detector The nominal area is determined as the area contained within 50% of the maximum optical density / pixel value P KA = K air A Alternatively: P KA is measured using a reference KAP meter Problems:! K air is measured in one point only in the beam-area method 12
13 The effect of distance 30.0 cm 30.0 cm 100 cm r x ray tube clinical KAP meter P KA,r P KA,p (r) Method:! Monte Carlo simulations of PKA using MCNP Ring detectors were used to score Kair X-ray spectra for 40, 80 and 140 kv filtered with 5 mm Al PKAp(r) PKAr kv 80 kv 140 kv radius / cm From [7] Cylindrical KAP meter based on VacuTec 70157! Results:! Beam attenuation in air Larger beam radius resulted in lower relative difference 13
14 Stray radiation Relative difference in PKA between the two configurations was less than 3% for considered tube voltages. Attenuation in air was the main factor. Stray radiation may be undetected in the IAEA calibration method. or? KAP meter stray radiation 14
15 NK formalism of IAEA s TRS-457 dosimetric quantity calibration coefficient z} { K =(M Q0 {z} z } { Y M 0 {z} ) N K,Q0 i k i {z} dosimeter reading, reference conditions dosimeter reading, no beam correction factors 15
16 Correction for temperature and pressure temperature of air in o C reference pressure P0 = kpa k TP = T T 0 P0 P reference temperature T0 = 20 o C pressure of air in kpa 16
17 Correction for humidity Reference value of relative humidity of air is 50% No correction is needed in the range 30% - 80% (TRS-457) 17
18 Correction for radiation quality unknown calibration coefficient c.c. obtained from standards laboratory K Q = M Q z } { N K,Q = M Q N K,Q N K,Q0 z } { N K,Q0 = M Q N K,Q0 k Q,Q0 {z } Thus N K,Q = N K,Q0 k Q,Q0 beam quality correction factor 18
19 Suggestions for improvement Problem 1: Large energy dependence of KAP meters Calibration coefficients (RQR) Calibration coefficients (RQA) 2000 NKA (Gym 2 C 1 ) nm 15 nm 20 nm KAP 1 KAP 2 KAP 3 NKA (Gym 2 C 1 ) nm 15 nm 20 nm KAP 1 KAP U (kv) U (kv) Figure: Calibration coefficient as a function of tube voltage for RQR and RQA beam qualities. Measured (markers) and simulated (lines) values. 19
20 Suggestions for improvement Problem 2: Beam qualities at clinics differ from beam qualities at standards laboratories.!! As a consequence, transfers of calibration coefficients from the standards laboratory to clinics are associated with uncertainties. 20
21 Our approach Calibration coefficients of built-in KAP meters should be beam-quality specific. This is a task for manufacturers. Before it happens, hospital physicists can determine factors correcting PKA values reported by the built-in KAP meters. To do so, three beam qualities have to be considered: Q0: reference beam quality at standards laboratory Q1: reference beam quality at the clinic Q: any diagnostic beam quality at the clinic 21
22 Our approach P KA = NP ref KA,QMQ ref N ref P KA,Q 1 N P KA,Q ref NP ref KA,Q M 0 Q ref 1 P KA,Q 0 = N ref P KA,Q N ref P KA = kq,q ref k ref 1 Q 1,Q N ref 0 P KA,Q M ref 0 Q where the beam quality correction factors are: k ref Q,Q 1 k ref Q 1,Q 0 Clinical reference beam quality Q1 to diagnostic beam quality Q. E.g. 70 kv and 0.1 mm Cu to 140 kv and 0.3 mm Cu. Standards laboratory reference beam quality Q0 to clinical reference beam quality Q1. E.g. RQR 5 to 70 kv and 0.1 mm Cu. 22
23 Our approach NPKA, Q NPKA, Q mm Cu, 10 nm 0.0 mm Cu, 15 nm 0.0 mm Cu, 20 nm 0.1 mm Cu, 10 nm 0.1 mm Cu, 15 nm 0.1 mm Cu, 20 nm 0.3 mm Cu, 10 nm 0.3 mm Cu, 15 nm 0.3 mm Cu, 20 nm 0.0 mm Cu, measured 0.1 mm Cu, measured 0.3 mm Cu, measured U (kv) Figure: Simulated and measured values of the beam quality correction factor kq,q1. (Q1: 70 kv, 0.1 mm Cu, Q: 0.0, 0.1, and 0.3 mm Cu, any U). 23
24 Our approach Table: Simulated values of the beam quality correction factor kq1,q0 as a function of the KAP meter s conductive coating thickness. Thickness (nm) k
25 Our approach Result for an x-ray stand at the clinic: Manufacturer s calibration was in error by up to 30%. 25
26 Primary references [1] NRPB. National Protocol for Patient Dose Measurements in Diagnostic Radiology. Chilton: National Radiological Protection Board, [2] Toroi, P, T Komppa, and A Kosunen. A Tandem Calibration Method for Kerma area Product Meters. Physics in Medicine and Biology 53 (September 21, 2008): [3] International Atomic Energy Agency. Dosimetry in Diagnostic Radiology: An International Code of Practice. Vienna: International Atomic Energy Agency,
27 Secondary references [4] ICRU. Patient Dosimetry for X Rays Used in Medical Imaging (Report 74). International Comission on Radiation Units & Measurements, [5] Toroi, P, T Komppa, A Kosunen, and M Tapiovaara. Effects of Radiation Quality on the Calibration of Kerma-Area Product Meters in X-Ray Beams. Physics in Medicine and Biology 53 (September 21, 2008): [6] Malusek, A, J P Larsson, and G Alm Carlsson. Monte Carlo Study of the Dependence of the KAP-Meter Calibration Coefficient on Beam Aperture, X- Ray Tube Voltage and Reference Plane. Physics in Medicine and Biology 52 (February 21, 2007): [7] Malusek, Alexandr, and Gudrun Alm Carlsson. Analysis of the Tandem Calibration Method for Kerma Area Product Meters via Monte Carlo Simulations. In Standards, Applications and Quality Assurance in Medical Radiation Dosimetry (IDOS), 1: Vienna: IAEA, [8] Larsson, J P, J Persliden, and G Alm Carlsson. Ionization Chambers for Measuring Air Kerma Integrated over Beam Area. Deviations in Calibration Values Using Simplified Calibration Methods. Physics in Medicine and Biology 43 (March 1, 1998):
Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009
2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced
More informationSPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by
SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This
More information2012 :15th SESSION of ESMP
2012 :15th SESSION of ESMP Lecture presented in Archamps (Salève Building) by : Elly CASTELLANO (London) Patient dosimetry in x-ray imaging and CT Elly Castellano Objectives measurable dose quantities
More informationPreliminary studies of a new monitor ionization chamber
1 Preliminary studies of a new monitor ionization chamber Maíra T. Yoshizumi, Vitor Vivolo and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares (IPEN-CNEN) Comissão Nacional de Energia
More informationDose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp
Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division
More informationTechnical aspects on DAP calibration and CT calibration
EURADOS Report 2015-03 European Radiation Dosimetry Group e. V. Braunschweig, August 2015 Technical aspects on DAP calibration and CT calibration H. Järvinen, O. Turak, M. Ginjaume, J. Daures, C. Hourdakis,
More informationMEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM *
MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM * C. Kessler and D.T. Burns December 2018 BUREAU INTERNATIONAL DES POIDS ET MESURES
More informationDETERMINATION OF INHERENT AND ADDITIONAL FILTRATION IN ORDER TO ESTABLISH RADIATION QUALITIES ACCORDING TO IEC 61267
2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 DETERMINATION
More informationI. Introduction.
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Accuracy of measuring half- and quarter-value layers and appropriate aperture width of a convenient method using a lead-covered case
More informationTest Equipment for Radiology and CT Quality Control Contents
Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3
More informationIntroduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image
Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology
More informationKey words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure
Accuracy and calibration of integrated radiation output indicators in diagnostic radiology: A report of the AAPM Imaging Physics Committee Task Group 190 Pei-Jan P. Lin a) Virginia Commonwealth University
More informationExposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?
Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago
More informationResponse characteristics of a tandem ionization chamber in standard X-ray beams
Applied Radiation and Isotopes 58 (2003) 495 500 Response characteristics of a tandem ionization chamber in standard X-ray beams Alessandro M. Costa*, Linda V.E. Caldas Instituto de Pesquisas Energ!eticas
More informationCylindrical Ion Chambers Victoreen Model 550 Series
Cylindrical Ion Chambers Victoreen Model 550 Series! Cylindrical Ion Chambers for use with Model 35040 and Model 530 electrometers! Wide range of applications in Diagnostic X-Ray and Radiation Oncology
More informationUpdate on NIST X-Ray Air-Kerma Standards and Calibrations
CCRI(I)/03-19 Update on NIST X-Ray Air-Kerma Standards and Calibrations Michelle O Brien and Stephen M. Seltzer Ionizing Radiation Division National Institute of Standards and Technology Gaithersburg,
More informationHalf value layer and AEC receptor dose compliance survey in Estonia
Half value layer and AEC receptor dose compliance survey in Estonia K. Kepler, A. Vladimirov Training Centre of Medical Physics, University of Tartu Testing Centre of the University of Tartu, Estonia E-mail:
More information熊本大学学術リポジトリ. Kumamoto University Repositor
熊本大学学術リポジトリ Kumamoto University Repositor Title Monte Carlo calculations of the rep correction factor, Ρ_, for cy chamber cav Author(s) Araki, Fujio CitationRadiological Physics and Technology Issue
More informationORAMED S MEASUREMENT AND SIMULATION CAMPAIGN FOR EXREMITY AND EYE LENS DOSES OF MEDICAL STAFF INVOLVED IN INTERVENTIONAL RADIOLOGY AND CARDIOLOGY
ORAMED S MEASUREMENT AND SIMULATION CAMPAIGN FOR EXREMITY AND EYE LENS DOSES OF MEDICAL STAFF INVOLVED IN INTERVENTIONAL RADIOLOGY AND CARDIOLOGY Carinou E., Brodecki M.,Domienik J., Donadille L., Ferrari
More informationEstimation of signal transfer property for wireless digital detector in different measurement schemes
Estimation of signal transfer property for wireless digital detector in different measurement schemes Anatoli Vladimirov, Kalle Kepler Training Centre of Medical Physics, University of Tartu, Estonia 11
More informationBeam-Restricting Devices
Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the
More informationMC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE
MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE A. Miceli ab, R. Thierry a, A. Flisch a, U. Sennhauser a, F. Casali b a Empa - Swiss Federal Laboratories for
More informationDOSE EVALUATION IN UROGRAPHY RADIOLOGICAL PROCEDURES
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 DOSE EVALUATION IN
More informationElectronic Brachytherapy Sources. Thomas W. Rusch
Electronic Brachytherapy Sources Thomas W. Rusch Educational Objectives Understand key elements of ebx source construction & operation Understand the rationale and methods for air kerma strength calibration
More informationCylindrical Ion Chambers
Cylindrical Ion Chambers Radiation Oncology ON Victoreen Model 550T Series Cylindrical Ion Chambers for use with Model 35040 and Model 560 electrometers Wide range of applications in Diagnostic X-Ray and
More informationPD233: Design of Biomedical Devices and Systems
PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/
More informationQC Testing for Computed Tomography (CT) Scanner
QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including
More informationThe importance of radiation quality for optimisation in radiology
Available online at http://www.biij.org/2007/2/e38 doi: 10.2349/biij.3.2.e38 biij Biomedical Imaging and Intervention Journal COMMENTARY The importance of radiation quality for optimisation in radiology
More informationEffect of patient support pads on image quality and dose in fluoroscopy a)
Effect of patient support pads on image quality and dose in fluoroscopy a) William R. Geiser, Walter Huda, b) and Nikolaos A. Gkanatsios Department of Radiology, University of Florida, Gainesville, Florida
More informationDETECTORS. Including Codes of Practice IONIZING RADIATION
DETECTORS Including Codes of Practice 2015 IONIZING RADIATION PTW History and General Remarks Looking back on a long history PTW-Freiburg is an internationally operating company, manufacturing and marketing
More informationNew Detectors for X-Ray Metal Thickness Measuring
ECNDT 2006 - Poster 132 New Detectors for X-Ray Metal Thickness Measuring Boris V. ARTEMIEV, Alexander I. MASLOV, Association SPEKTR- GROUP, Moscow, Russia Abstract. X-ray thickness measuring instruments
More informationThe Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017
The Current State of EPID-Based Linear Accelerator Quality Assurance Timothy Ritter, PhD, DABR, FAAPM 1 Disclosures Employed by the Veterans Health Administration Faculty appointment with the University
More informationX-rays in medical diagnostics
X-rays in medical diagnostics S.Dolanski Babić 2017/18. History W.C.Röntgen (1845-1923) discovered a new type of radiation Nature, Jan. 23. 1896.; Science, Feb.14. 1896. X- rays: Induced the ionization
More informationSTEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS
STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%
More informationY11-DR Digital Radiography (DR) Image Quality
Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures
More informationINTERNATIONAL STANDARD
INTERNATIONAL STANDARD IEC 62220-1 First edition 2003-10 Medical electrical equipment Characteristics of digital X-ray imaging devices Part 1: Determination of the detective quantum efficiency Appareils
More informationABSORBED DOSE DISTRIBUTIONS USING THE ISODENSITOMETRIC METHOD FOR EXPOSURES WITH FILTER EMPLOYED FOR MAMMOGRAPHIES
Romanian Reports in Physics, Vol. 65, No. 1, P. 168 177, 213 ABSORBED DOSE DISTRIBUTIONS USING THE ISODENSITOMETRIC METHOD FOR EXPOSURES WITH FILTER EMPLOYED FOR MAMMOGRAPHIES F. SCARLAT 1, A. SCARISOREANU
More informationDetermination of the tube voltage from clinic mammographic system using two types of detectors
BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 03-1A (2015) 01-06 Determination of the tube voltage from clinic mammographic system using two types of detectors J. S. Barreira a ; V. Vivolo a a Gerência
More informationTitle: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS
Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS By Aravind Ravichandran arr192@mail.usask.ca University of Saskatchewan Address: 2424 Cumberland
More information8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey
Implementation of the 2016 ACR Digital Mammography QC Manual Medical Physicist Mammography Equipment Evaluation and Annual Survey Eric A Berns, PhD, FACR Radiologist Responsibilities Radiologist Responsibilities
More informationIONIZING RADIATION 2018 / 2019 DETECTORS. Including Codes of Practice
IONIZING RADIATION DETECTORS 2018 / 2019 Including Codes of Practice PTW History and General Remarks Looking back on a long history PTW-Freiburg is an internationally operating company, manufacturing and
More informationLudlum Medical Physics
Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to
More informationX-RAYS - NO UNAUTHORISED ENTRY
Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are
More informationX-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah
X-RAY IMAGING EE 472 F2017 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Stewart C. Bushong, Radiologic Science for Technologists: Physics, Biology, and Protection, 10 th ed., Mosby,
More informationDisclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography
Current Implementation Status of IEC Standard 62494-1: Exposure Index (EI) for Digital Radiography July 31, 2017 Ryan Fisher, PhD, DABR Katie Hulme, MS, DABR None Disclosures Outline Review of IEC Standard
More informationX-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage
X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage
More informationEffect of pressure, temperature and humidity in air on photon fluence and air kerma values at low photon energies
ARTICLE IN PRESS Radiation Physics and Chemistry 68 (2003) 707 720 Effect of pressure, temperature and humidity in air on photon fluence and air kerma values at low photon energies M. Assiamah, D. Mavunda,
More informationDosepix Detector as kvp-meter in Radiology and Mammography: First steps
Dosepix Detector as kvp-meter in Radiology and Mammography: First steps F.Bisello, I.Ritter, F.Tennert, A.Zang MediPix Collaboration Meeting, 19th February 2014, CERN Protect, Enhance, and Save Lives -
More informationAsk EuroSafe Imaging Tips & Tricks. Paediatric Imaging Working Group. Dose Management in Digital Radiography
Ask EuroSafe Imaging Tips & Tricks Paediatric Imaging Working Group Dose Management in Digital Radiography Raija Seuri (HUS Medical Imaging Center, FI) Cristina Almeida (Centro Hospitalar de Lisboa Central,
More informationMinnesota Rules, Chapter 4732 X-ray Revision
Minnesota Rules, Chapter 4732 X-ray Revision DRAFT FLUOROSCOPIC X-RAY SYSTEMS, 1.0 Subpart 1. Applicability. Subpart 2. Limitation of the useful beam. Subpart 3. Measuring compliance; primary protective
More informationDerivation of total filtration thickness for diagnostic x-ray source assembly
Physics in Medicine & Biology PAPER OPEN ACCESS Derivation of total filtration thickness for diagnostic x-ray source assembly To cite this article: Michiharu Sekimoto and Yoh Katoh 2016 Phys. Med. Biol.
More informationSeminar 8. Radiology S8 1
Seminar 8 Radiology Medical imaging. X-ray image formation. Energizing and controlling the X-ray tube. Image detectors. The acquisition of analog and digital images. Digital image processing. Selected
More informationMASINZA, ALWYN STANSLAUS. BSc (Hons.), PGDSc
By MASINZA, ALWYN STANSLAUS BSc (Hons.), PGDSc A thesis submitted in partial fulfilment for the award of degree of Master of Scienc (Physics), University o f Nairobi. December, 2012 ! Declaration This
More informationUpdate on Fluoroscopy Physics AAPM MO-A-210A-1 Stephen Balter, Ph.D.
Update on Fluoroscopy Physics Stephen Balter, PhD Columbia University Draft for MO-A-210A-1 2009 AAPM Educational objectives Understand dosimetric concepts relating to interventional fluoroscopy Characterize
More informationA Guide to Radiochromic Film Dosimetry with EBT2 and EBT3
A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 David F. Lewis Advanced Materials Group Ashland Specialty Ingredients Spain, April 2014 What is Radiochromic Film? A film that instantly changes
More informationWide beam CT dosimetry. Elly Castellano
Wide beam CT dosimetry Elly Castellano Outline revision: CT dose indices wide-beam CT: the end of the road for CTDI? the IEC rescue plan for CTDI 100 the american way AAPM report 111 better estimates of
More informationAssessment of Beam Alignment, Collimation and Half Value Layer of Some Selected X-Ray Machines in Plateau State, Nigeria
International Journal of Innovative Scientific & Engineering Technologies Research 5(4):-5, Oct.-Dec., 07 SEAHI PUBLICATIONS, 07 www.seahipaj.org ISSN: 60-896X Assessment of Beam Alignment, Collimation
More informationX-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto
X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring
More informationA Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images
A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images Aiping Ding, Bin Han, Lei Wang, Lei Xing Department of Radiation Oncology, Stanford University School of
More informationIntroduction of a Single Chip TLD System for Patient Dosimetry
Introduction of a Single Chip TLD System for Patient Dosimetry C. Hranitzky a, M. Halda a, G. Müller a, B. Obryk b, H. Stadtmann a* a Austrian Research Centers GmbH ARC, 2444 Seibersdorf, Austria. b Institute
More informationStudies on reduction of exposure dose using digital scattered X-ray removal processing
Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.
More informationMammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over
Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on
More informationStudies on reduction of exposure dose using digital scattered X-ray removal processing
Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.
More informationQA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion
Leksell Gamma Knife Perfexion QA Considerations Paula L. Petti, Ph.D. Taylor McAdam Bell Neuroscience Institute Washington Hospital Healthcare System Fremont, CA 1 QA for LGK Perfexion : : Follow NRC licensing
More informationIBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY
WHITE PAPER: IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY IBEX Innovations Ltd. Registered in England and Wales: 07208355 Address: Discovery 2, NETPark, William Armstrong Way, Sedgefield, UK Patents:
More informationDigital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of
Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of radiation to the population due to Medical Imaging
More information1-1. GENERAL 1-2. DISCOVERY OF X-RAYS
1-1. GENERAL Radiography is a highly technical field, indispensable to the modern dental practice, but presenting many potential hazards. The dental radiographic specialist must be thoroughly familiar
More informationX-ray Tube and Generator Basic principles and construction
X-ray Tube and Generator Basic principles and construction Dr Slavik Tabakov - Production of X-rays and Patient Dose OBJECTIVES - X-ray tube construction - Anode - types, efficiency - Classical X-ray generator
More informationSummer Student project report
Summer Student project report Mika Väänänen September 1, 2017 Abstract In this report I give a brief overview of my activities during the summer student project. I worked on the scintillating fibre (SciFi)
More informationRadiographic sensitivity improved by optimized high resolution X -ray detector design.
DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France Radiographic sensitivity improved by optimized high resolution X -ray detector
More informationUse of Back Scattered Ionizing Radiation for Measurement of Thickness of the Catalytic Agent Active Material
18th World Conference on Nondestructive Testing, 16- April 1, Durban, South Africa Use of Back Scattered Ionizing Radiation for Measurement of Thickness of the Catalytic Agent Active Material Boris V.
More informationReport on the study and optimisation of chest and lumbar spine X-ray imaging
Institutionen för medicin och vård Avdelningen för radiofysik Hälsouniversitetet Report on the study and optimisation of chest and lumbar spine X-ray imaging Michael Sandborg, Graham McVey, David R Dance
More informationMaster of Science Thesis. SIMIND Based Pinhole Imaging
Master of Science Thesis SIMIND Based Pinhole Imaging * Development and Validation Kurt Sundin Supervisor: Michael Ljungberg, PhD Medical Radiation Physics Clinical Sciences, Lund Lund University, 2006
More informationYour worldwide source of grids. What a difference a grid makes
Your worldwide source of grids What a difference a grid makes The right assembly to meet your every need Trust the experts A high degree of expertise is required to design and manufacture quality X-ray
More informationMedical Imaging: A Look inside. Medical Imaging. Medical Imaging. Visible Human Project
Medical Imaging: A Look inside Medical Imaging Allows physicians to see what had previously been unseeable: internal organs and tissues, bones, a beating heart, etc. Allows physicians to: detect brain
More informationQUALITY CONTROL TESTS IN SOME DIAGNOSTIC X-RAY UNITS IN BANGLADESH
Bangladesh Journal of Medical Physics Vol. 4, No.1, 2011 QUALITY CONTROL TESTS IN SOME DIAGNOSTIC X-RAY UNITS IN BANGLADESH M. Begum 1, A. S. Mollah 2, M. A. Zaman 3 and A. K. M. M. Rahman 4 1 Health Physics
More informationOptimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation
Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.125 pissn 2508-4445, eissn 2508-4453 Optimization of Energy Modulation Filter for Dual Energy CBCT
More informationCR Basics and FAQ. Overview. Historical Perspective
Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors
More informationReceived 3 December 2016; revised 17 May 2017; editorial decision 21 June 2017; accepted 4 July 2017
Radiation Protection Dosimetry (2018), Vol. 178, No. 2, pp. 235 241 Advance Access publication 15 July 2017 doi:10.1093/rpd/ncx091 NOTE COMPARISON OF PERSONAL DOSE EQUIVALENT Hp(10) IN 137 CS RADIATION
More informationChina Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R
China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R Building 3, No.9, Jiuxianqiaodong Road, Chaoyang District, Beijing 100015, P.R. China E-mail:
More informationEffect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging
The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2008 Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging Dean
More informationI. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA
FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE
More informationMaximizing clinical outcomes
Maximizing clinical outcomes Digital Tomosynthesis Dual Energy Subtraction Automated Long Length Imaging Improved image quality at a low dose Xray Xray Patented ISS capture technology promotes high sensitivity
More informationRenato Padovani, University Hospital, Udine, Italy
DATAMED II Training Course Sofia, May 19-20, 2011 Patient dose assessment in fluoroscopy and interventional procedures Renato Padovani, University Hospital, Udine, Italy Content Fluoroscopy guided diagnostic
More informationEvaluation of X-Ray Beam Quality Based on Measurements and Estimations Using SpekCalc and Ipem78 Models
Original Article Evaluation of X-Ray Beam Quality Based on Measurements and Estimations Using SpekCalc and Ipem78 Models Suk Chiang Chen, Wei Loong Jong, Ahmad Zaky Harun Submitted: 20 Aug 2011 Accepted:
More informationThe effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography
Journal of Physics: Conference Series OPEN ACCESS The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography To cite this article: N A A Daud et al 2014 J. Phys.:
More informationBreast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School
Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis
More informationISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film
INTERNATIONAL STANDARD ISO 17636-1 First edition 2013-01-15 Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film Contrôle non destructif des assemblages soudés
More informationNOT FOR DISTRIBUTION JINST_083P_0914 v1
Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam F. Di Lillo a,b, D. Dreossi c, F. Emiro a,b, C. Fedon d,e, R. Longo d,e, G. Mettivier a,b,*, L. Rigon d,e, P. Russo
More informationWhen small things matter. Small Field Dosimetry Application Guide
R A D I AT I O N T H E R A P Y When small things matter. Small Field Dosimetry Application Guide Contents 1 Introduction 1 Introduction 2 2 The Physics of Small Fields 3 3 Detector Types 10 4 Detector
More informationTeaching Digital Radiography and Fluoroscopic Radiation Protection
Teaching Digital Radiography and Fluoroscopic Radiation Protection WCEC 20 th Student Educator Radiographer Conference Dennis Bowman, RT(R), CRT (R)(F) Community Hospital of the Monterey Peninsula (CHOMP)
More informationFeatures and Weaknesses of Phantoms for CR/DR System Testing
Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal
More informationMonte Carlo study on a new concept of a scanning photon beam system for IMRT
NUKLEONIKA 2011;56(4):291 297 ORIGINAL PAPER Monte Carlo study on a new concept of a scanning photon beam system for IMRT Anna M. Wysocka-Rabin, Günter H. Hartmann Abstract. Intensity-modulated radiation
More informationExposure System Selection
Principles of Imaging Science II (RAD120) Exposure Systems Exposure System Selection Radiographic exposure is a very complex process Best technique systems manipulate one variable while holding others
More informationClinical Experiences with a Patient Skin Dose Monitoring and Tracking Program
Clinical Experiences with a Patient Skin Dose Monitoring and Tracking Program Allen R. Goode, MS, DABR Chief Diagnostic Medical Physicist Department of Radiology & Medical Imaging University of Virginia
More informationDevelopment of Personal Dosimeter Using Electronic Dose Conversion Method
Proceedings of the Korean Nuclear Spring Meeting Gyeong ju, Korea, May 2003 Development of Personal Dosimeter Using Electronic Dose Conversion Method Wanno Lee, Bong Jae Lee, and Chang Woo Lee Korea Atomic
More informationRADIOGRAPHIC EXPOSURE
RADIOGRAPHIC EXPOSURE Receptor Exposure Receptor Exposure the that interacts with the receptor. Computed Radiography ( ) requires a. Direct Digital Radiography (DR) requires a. Exposure Indicators Exposure
More informationCyberKnife Iris Beam QA using Fluence Divergence
CyberKnife Iris Beam QA using Fluence Divergence Ronald Berg, Ph.D., Jesse McKay, M.S. and Brett Nelson, M.S. Erlanger Medical Center and Logos Systems, Scotts Valley, CA Introduction The CyberKnife radiosurgery
More informationPredicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities
Journal of Physics: Conference Series PAPER OPEN ACCESS Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities Recent citations - Resolution Properties of a
More informationTechnical data CAMARGUE CS-VH50/300. VARIABLE Height Bucky Table With Ceiling Suspension
Technical data VARIABLE Height Bucky Table With Ceiling Suspension Model Variations CAMARGUE FH (Fixed Height) CAMARGUE FH Tomo CAMARGUE FH Ceiling suspension CAMARGUE VH (Variable Height) CAMARGUE VH
More information