Received 3 December 2016; revised 17 May 2017; editorial decision 21 June 2017; accepted 4 July 2017

Size: px
Start display at page:

Download "Received 3 December 2016; revised 17 May 2017; editorial decision 21 June 2017; accepted 4 July 2017"

Transcription

1 Radiation Protection Dosimetry (2018), Vol. 178, No. 2, pp Advance Access publication 15 July 2017 doi: /rpd/ncx091 NOTE COMPARISON OF PERSONAL DOSE EQUIVALENT Hp(10) IN 137 CS RADIATION BETWEEN THE PRIMARY STANDARDS LABORATORIES OF JAPAN AND AUSTRALIA USING BeO OSL PERSONAL DOSEMETERS D. J. Butler 1, *, T. Kurasawa 2, M. Litwin 1, J Mazaraki 1 and N. Saito 2 1 Australian Radiation Protection and Nuclear Safety Agency (ARPANSA), Yallambie, Victoria 3085, Australia 2 National Measurement Institute of Japan (NMIJ), AIST, Tsukuba Japan *Corresponding author: Duncan.Butler@arpansa.gov.au Received 3 December 2016; revised 17 May 2017; editorial decision 21 June 2017; accepted 4 July 2017 A comparison of personal dose equivalent Hp(10) for 137 Cs radiation was conducted between the primary standards laboratories of Japan and Australia. A set of 120 commercially available passive BeO OSL dosemeters were used (Dosimetrics GmbH, Munich). The aim was to investigate the precision which could be obtained with this technique, and to confirm the personal dose equivalent delivery methods in each standards laboratory. A dose of 5 msv was delivered to 40 dosemeters in each country, and 40 dosemeters were used as controls. The result of the comparison was a ratio of Hp(10) in Japan to Australia of with a combined standard uncertainty of 3.2%. The statistical uncertainty was 0.32% indicating that passive dosemeters can be used for comparisons of high precision. INTRODUCTION Primary and secondary standard dosimetry laboratories regularly perform comparisons to validate and establish equivalence of their standards. The majority of these comparisons are performed as a part of the CIPM Mutual Recognition Arrangement (CIPM MRA) (1). Under this arrangement Japan and Australia both reside in the Asia Pacific Metrology Programme (APMP) and have recently compared standards for air kerma and absorbed dose to water in regional and bilateral comparisons (2 4). The quantity of interest for personal monitoring services, however, is personal dose equivalent. The quantity Hp(d) is defined as the absorbed dose to water at d mm depth in an ICRU phantom in a specific irradiation geometry (5 7). For the purposes of this study we are concerned with Hp(10), the most common quantity used for personal monitoring for penetrating photon radiation. Only a few dosimetry laboratories have primary methods to determine Hp(10). Instead, it is common practice to establish the air kerma rate, free in air, and use a Monte Carlo calculation published in the international standard X and gamma reference radiation for calibrating dosemeters and doserate metres and for determining their response as a function of photon energy Parts 1 3:1996, 1997, 1999 (5 7), to calculate Hp(10). Personal dosemeters may then be irradiated on the surface of a phantom using the calculated Hp(10) value (7). Both ARPANSA and NMIJ provide irradiations for personal monitoring services based on air kerma standards and the application of ISO Hence, their values of Hp(10) should agree because their air kerma standards have been shown to agree to better than 1% (3). Nevertheless, a comparison was considered to be a worthwhile validation of the implementation of ISO 4037 at each laboratory. The Personal Radiation Monitoring Service (PRMS) at ARPANSA has recently changed from a CaS04:Dy TLD system to a commercially available BeO OSL system (Dosimetrics GmbH, Munich, 2015). The BeO OSL Personal Dosimeter meets the IEC Standard and has been type tested in Germany by Helmholtz-Zentrum dosimetry service (AWST Auswertungsstelle) and has been approved by the Physikalish Technischen Bundesanstalt (PTB). The new system has very small Type A uncertainties and the automated readout makes it easy to process large numbers of dosemeters. Therefore a high degree of precision should be possible if these dosemeters are used as the transfer instruments in a comparison. While other transfer instruments such as electronic personal dosemeters can be used (8 10), passive dosemeters are more representative of the majority of instruments used for personal monitoring. Passive dosemeters also make it much easier to conduct a blind comparison, which is important for proficiency testing and the accreditation of monitoring services. The Author Published by Oxford University Press. All rights reserved. For Permissions, please journals.permissions@oup.com

2 PRMS is co-located with (although separate from) the standards laboratory at ARPANSA. Hence, an independent verification of its dosimetry is also of value, as it negates the potential for a claim of conflict of interest regarding the traceability chain. D. J. BUTLER ET AL. METHOD A group of 120 OSL personal dosemeters were prepared by the PRMS at ARPANSA, zeroed, and shipped to Japan. In Japan, 40 dosemeters were selected at random and irradiated to an Hp(10) of 5 msv, the dose being determined and delivered independently by NMIJ. The irradiation details are given in NMIJ irradiations. The dosemeters were returned to ARPANSA where they were received by the primary standards laboratory. Forty of the un-irradiated dosemeters were selected at random and given 5 msv, as determined by ARPANSA, leaving 40 to act as controls. All of the dosemeters were returned to PRMS and read out. The PRMS internal QC procedures confirmed that the readers were operating to specifications for the comparison readings. OSL personal dosemeters One of the personal dosemeters used for the comparison is shown in Figure 1. The personal dosemeters were shipped and irradiated inside blister packs (Figure 1a). The outer dimensions of the pack are mm 3. The light-tight black dosemeter (Figure 1b) has outer dimensions mm 3. The disassembled version (Figure 1c) shows the two active BeO areas. The area on the right is behind a thin window and can be used for Hp(0.07) assessments. For the comparison, only the readings from the active area on the left were used. The fading, linearity, energy dependence and Type A uncertainty associated with reading the dosemeters has already been established (11, 12). When using the dosemeters for a comparison, we decided to minimise the use of corrections by choosing dosemeters zeroed at the same time, and by delivering (as close as possible) the same dose at each laboratory. In reference (11) the fading of these dosemeters is established as being negligible after the first 10 min following irradiation, and so no correction was applied for the 2-week difference between the NMIJ and ARPANSA irradiations. NMIJ irradiations The NMIJ irradiation facility consists of a 34 TBq 137 Cs source with a collimator and iron plate attenuator to reduce the dose rate (Figure 2). A water-filled phantom was up at 400 cm distance from the source. The NMIJ phantom has outer dimensions 32 cm 32 cm cm and is made from PMMA walls 1 cm thick except the front face which is 0.25 cm thick. Figure 1. BeOSL personal dosemeter (Dosimetrics GmbH, Munich) (a) in blister pack, as worn by users and as used for the comparison, (b) removed from blister pack and (c) when removed from the light-tight package, showing the two active BeO elements. The air-kerma rate was determined by the NMIJ primary graphite cavity chamber (13). The irradiation time was about 630 s to deliver 5 msv. The centre of the personal dosemeter was set at the reference point (400 cm). The 2 mm PMMA plate was placed in front of personal dosemeters for the build-up plate. ARPANSA irradiations The ARPANSA irradiation facility consists of a 50 TBq 137 Cs source in a custom-made housing and collimator. A water-filled phantom was set up at 400 cm distance from the source (Figure 3). The ARPANSA phantom has outer dimensions 32 cm 32 cm 16.3 cm and is made from PMMA walls 1 cm thick except the front face which is 0.3 cm thick. The phantom is filled with water. Source timing is done manually. For short exposures, timing errors and source transit effects need to be taken into account. For this reason a transmission monitor chamber (Stealth ChamberTM, IBA Dosimetry GmbH, Schwarzenbruck, Germany) was 236

3 COMPARISON OF PERSONAL DOSE EQUIVALENT HP(10) Figure 2. Irradiation arrangement at the NMIJ. Figure 3. Irradiation arrangement at ARPANSA. positioned over the exit port of the collimator. The charge from the chamber was integrated during each exposure and used to determine the delivered dose. The monitor chamber was calibrated using a Farmer chamber positioned at the reference point (400 cm), free in air, in the absence of the phantom and buildup plate, using exposures of the same duration as required to deliver 5 msv (125 s). Traceability NMIJ maintains the Japanese primary standard for air kerma at 137 Cs and 60 Co (a graphite cavity chamber (13, 14) ). ARPANSA maintains the Australian primary standard for air kerma at 60 Co (a graphite cavity chamber (15) ) and for medium-energy kilovoltage X-rays (a free-air chamber (16) ). The response of a Farmer chamber to 137 Cs gamma rays was determined by linear interpolation between these standards (Table 1). This chamber was then used to determine the air kerma rate in the 137 Cs beam. RESULTS The results of the PRMS readouts of the 120 dosemeters are summarised in Table 2. The average reading of the un-irradiated dosemeters was msv. The average of the dosemeters irradiated in Japan was msv and those in Australia msv. Note that PRMS is traceable to ARPANSA and so the readouts for ARPANSA are expected to average exactly 5 msv above the control. For an individual dosemeter i, wedefine the ratio R i as the ratio of the dose delivered to the dosemeter according to the laboratory (D i,lab ), to the PRMS readout P i less the background signal (Eq. 1). Here the background, B, is determined by averaging the PRMS readouts of the control dosemeters (0.088 msv). R = D /( P B) ( 1) i i,lab i The average, standard deviation and standard deviation of the mean are given in Table 3. For the ARPANSA the average ratio was and for NMIJ 237

4 D. J. BUTLER ET AL. Table 1. Calibration coefficient and traceability for the Farmer chamber used to determine the air kerma rate in the ARPANSA 137 Cs beam. Description Traceability ARPANSA beam code Effective energy (kev) N K (mgy/nc) 60 Co gamma rays Primary standard cavity chamber S-Co kvp X-ray beam Free air chamber NXA Cs gamma rays Interpolation of above S-Cs Table 2. PRMS readouts from the OSL personal dosemeters. Irradiating laboratory Number of dosemeters Mean reading (msv) Standard deviation (msv) Standard deviation of the mean (msv) None (controls) NMIJ ARPANSA Table 3. Average ratios of the delivered dose to the reading, for ARPANSA and NMIJ, and the ratio of NMIJ to ARPANSA calculated from these ratios. Ratio between Number of dosemeters Mean of R i (delivered/reading) Standard deviation in R i (%) Standard deviation of the mean (%) ARPANSA/ PRMS NMIJ/PRMS Ratio of mean ratios NMIJ/ARPANSA a a The combined standard uncertainty evaluated from the Type A uncertainty only (see Table 6). it was 1.015, so that the overall comparison result was a ratio of NMIJ to ARPANSA of The combined statistical uncertainty in this ratio is 0.32%. The comparison result r can be expressed for a general case of n dosemeters irradiated at LAB1 and m dosemeters irradiated at LAB2, as follows: r = m n n Di P,LAB1 m Di,LAB2 B P B j j,lab2 i= 1 i,lab1 = 1 ( 2) where D i, LAB are the doses delivered to each dosemeter according to LAB1 or LAB2. The comparison ratio r is therefore a measure of the agreement between the dose determined by each laboratory. In the case where each laboratory delivers the same target dose, and the same number of dosemeters are used: r = n n 1 1 P i B P B j j i= 1,LAB1 = 1,LAB2 ( 3) Uncertainties Uncertainty budgets for the Hp(10) irradiations in each laboratory are given in Tables 4 and 5. NMIJmaintain a primary standard at 137 Cs, whereas ARPANSA interpolate the response of a Farmer 2571 ionisation chamber calibrated against their primary air kerma standards at 60 Co and medium-energy X-rays. The dominating uncertainty in each case is the conversion factor from air kerma to personal dose equivalent. This value (2% at k = 1) for ARPANSA is given in the ISO 4037 standard (7) and takes into account the different spectra and different geometries which are possible between different sources and using a range of source phantom distances. NMIJ measured a photon energy spectrum to determine this conversion factor using a Ge spectrometer. Then, the uncertainty of this value is 0.9% at k = 1, which is smaller than 2%. In practice this uncertainty is likely to be partially correlated between ARPANSA and NMIJ, especially since each laboratory used the same source detector distance. An additional uncertainty in the comparison ratio arises because of the different reference points used at NMIJ and ARPANSA. NMIJ used the centre 238

5 COMPARISON OF PERSONAL DOSE EQUIVALENT HP(10) Table 4. Uncertainty in the NMIJ irradiations. Quantity Value Standard uncertainty, u (%) Air kerma rate at 4000 mm Gy/s 0.4 Conversion from air kerma to Hp(10) on slab phantom 1.22 Sv/Gy 0.9 Beam uniformity over 10 cm irradiation area Uncertainty for the build-up plate Combined uncertainty in delivered Hp(10) 1.8 Table 5. Uncertainty in the ARPANSA irradiations. Quantity Value Standard uncertainty, u (%) N K of Farmer chamber for 137 Cs a Gy/C 1.2 Calibration of monitor chamber positioned Gy/C 0.5 b at exit port for air kerma at 4000 mm. Distance error between air kerma calibration 5 mm in 4000 mm 0.3 point and dosemeter reference point Beam uniformity over 10 cm irradiation area Timing/source transit error when used 0.1 s in 125 s irradiation 0.1 with monitor chamber Conversion from air kerma to Hp(10) on slab phantom 1.21 Sv/Gy 2.0 Combined uncertainty in delivered Hp(10) 2.6 a Details of the air kerma primary standard uncertainties for ARPANSA are given in references (15, 16). b The overall monitor chamber uncertainty consists of the Farmer chamber uncertainty added in quadrature with the additional uncertainty shown in this cell. Quantity Table 6. Uncertainty in the comparison ratio. Standard uncertainty in r (%) Statistical uncertainty in ratio 0.32 Reader stability (included in the above) Uncertainty due to background <0.1 subtraction a Fading between NMIJ and 0.20 ARPANSA irradiations (14 days) Combined uncertainty in r 0.40 a Includes reader linearity and statistical variation in the background and its readout. of the dosemeter, and ARPANSA used the rear surface. At 4 m the correction for the difference of 5 mm is 0.3%. In future ARPANSA will use the centre of the dosemeter, in accordance with ISO In Table 6 we present the uncertainties in the comparison ratio. The major contributor is the statistical variation in the dosemeter readouts, which yield a combined uncertainty of 0.32% in the ratio r (Table 3). This uncertainty includes the variation in the doses received by the dosemeters (from the laboratory irradiation and otherwise), and the variability of the reader over the course of the readout (about half an hour). It can be shown from equation three that the comparison ratio is insensitive to the background value, provided it issmall,andsocontributionstotheratiouncertainty from the statistical variation in the dosemeters used to evaluate the background, and any reader non-linearity associated with the subtraction of a small background from a larger dose reading, are negligible. In reference (11) the fading is shown to be ~92 96% (with no trend) over periods of days after irradiation, with a standard uncertainty on each measurement of around 2% (Figure 4 of reference (11) ). If we assume that fading, if present, is linear after the first day, and that it is not more than 4% over 300 days, then fading difference over the 14-day interval between the irradiations in our comparison should be <0.2%. DISCUSSION The control dosemeters received an average of msv with a standard deviation of msv. 239

6 The period between zeroing the dosemeters and reading them out was 30.6 days, so that this dose corresponds to 2.9 μsv/d (1.0 msv/y). This value is consistent with the external component of the natural background in Australia and Japan, and suggests that any additional exposure due to air travel or security screening was not significant. Dosemeters kept at ARPANSA can be expected to accumulate dose at the rate of ~2.3 μsv/d. The background is 1.8% of the 5 msv comparison dose. For high accuracy comparisons, then, the background must always be assessed. If smaller doses are used, or comparisons last longer than 30 days, this requirement becomes even more important. The ratio of ARPANSA to the backgroundcorrected PRMS reading was with a statistical uncertainty of 0.21%. Ideally the ratio would be unity because PRMS is traceable to ARPANSA. However, the difference of 0.9% is consistent with the combined statistical uncertainties of the comparison result and the PRMS system calibration. The value is also well within the acceptable limit for reader stability. The overall comparison result NMIJ/ARPANSA of has a combined standard uncertainty of 3.2%. The major contributions to this value are the statistical uncertainty of 0.32% and the standard uncertainty of each laboratory, 2.6% for ARPANSA and 1.8% for NMIJ. The result is therefore well within the expected uncertainty. That the agreement is significantly better than the combined uncertainty is probably an indication of a high level of correlation in the uncertainty for the conversion factor from air kerma to Hp(10), and should not be taken to imply that personal dose is determined at higher accuracy than the given uncertainty budgets. The purpose of the study was 3-fold: to determine the precision of a comparison using passive dosemeters, to provide validation for the implementation of ISO 4037 in each laboratory and to provide the PRMS with an independent verification of their dosimetry. We have shown that a high level of precision is possible, provided many dosemeters are used. The other aims of the study are satisfied by the close agreement found. The use of passive dosemeters is novel. Usually, international comparisons use electronic personal dosemeters or ion chambers (8 10). However, when many transfer instruments are used we have shown that a similar precision can be achieved. The advantage of using passive dosemeters is that the irradiation service is tested using the devices it is ultimately intended for. Hence, the results support the standards laboratories irradiations for the proficiency testing and accreditation of personal monitoring services. The potential disadvantages are the lack of immediate results, and the need to use many dosemeters in order to get acceptable precision. D. J. BUTLER ET AL. CONCLUSION A comparison of personal dose equivalent using passive personal dosemeters was successfully conducted. The ratio of Hp(10) delivered at ARPANSA and NMIJ was determined with a statistical uncertainty of 0.32%. The high precision is similar that achievable with more commonly used transfer instruments such as ionisation chambers. The ratio of Hp(10) between NMIJ (Japan) and ARPANSA (Australia) was found to be with a combined standard uncertainty of 3.2%. The control dosemeters received a dose consistent with the external source component of natural background, indicating that screening and travel did not add significantly to the dosemeter doses during the comparison. REFERENCES 1. The CIPM MRA (Mutual Recognition Arrangement of national measurement standards and of calibration and measurement certificates issued by national metrology institutes), first signed in Accessed online in 2016: pdf 2. Allisy-Roberts, P. J., Kessler, C. and Burns, D.T. Summary of the BIPM.RI(I)-K5 comparison for air kerma in 137 Cs gamma radiation. Metrologia 50(Tech. Suppl.), (2013). 3. Chun, K. J. et al. Final report on APMP.RI(I)-K1: APMP/TCRI key comparison report of measurement of air kerma for 60 Co gamma-rays. Metrologia 50(Tech. Suppl.), (2013). 4. Allisy-Roberts, P. J. and Burns, D. T. Summary of the BIPM.RI(I)-K4 comparison for absorbed dose to water in 60 Co gamma radiation. Metrologia 42(Tech. Suppl.), (2005). 5. ISO X and gamma reference radiation for calibrating dosemeters and doserate meters and for determining their response as a function of photon energy Part 1: Radiation characteristics and production methods (1996). 6. ISO X and gamma reference radiation for calibrating dosemeters and doserate meters and for determining their response as a function of photon energy Part 2: Dosimetry for radiation protection over the energy ranges from 8 kev to 1.3 MeV and 4 MeV to 9 MeV (1997). 7. ISO X and gamma reference radiation for calibrating dosemeters and doserate meters and for determining their response as a function of photon energy Part 3: Calibration of area and personal dosemeters and the measurement of their response as a function of energy and angle of incidence (1999). 8. Ankerhold, U., Tiefenböck, W., Witzani, J. and Ambrosi, P. Supplementary comparison: EUROMET.RI(I)-S1: personal dose equivalent comparison between the BEV and the PTB. Metrologia39(Tech. Suppl), (2002). 9. Ankerhold, U. and Hupe, O. Supplementary comparison: EURAMET supplementary comparison of the personal dose equivalent quantity for photon radiation: 240

7 COMPARISON OF PERSONAL DOSE EQUIVALENT HP(10) EURAMET.RI(I)-S5. Metrologia 49(Tech. Suppl), 0601 (2012). 10. IAEA-TECDOC Intercomparison of Personal Dose Equivalent Measurements by Active Personal Dosimeters (Vienna: IAEA) (2007). 11. Sommer, A. and Henniger, J. Investigation of a BeObased optically stimulated luminescence deosemeter. Radiat. Prot. Dosim. 119(1 4), (2006). 12. Sommer, M., Jahn, A. and Henniger, J. A new personal dosimetry system for Hp(10) and Hp(0.07) photon dose based on OSL-dosimetry of beryllium oxide. Radiati. Meas. 46, (2011). 13. Kessler, C., Saito, N. and Kurosawa, T. Key comparison BIPM.RI(I)-K5 of the air kerma standards of the NMIJ, Japan and the BIPM in 137 Cs gamma radiation. Metrologia 50(Tech. Suppl.), (2013). 14. Kessler, C., Saito, N. and Kurosawa, T. Key comparison BIPM.RI(I)- K1 of the air kerma standards of the NMIJ, Japan and the BIPM in Co-60 gamma radiation. Metrologia 50(Tech. Suppl.), (2013). 15. Kessler, C., Allisy-Roberts, P.J., Lye, J.E. and Oliver, C. Comparison of the standards for air kerma of the ARPANSA and the BIPM for 60 Co gamma radiation. Metrologia 48(Tech. Suppl.), (2011). 16. Burns, D.T., Lye, J.E., Roger, P. and Butler, D.J. Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the ARPANSA, Australia and the BIPM in medium-energy x-rays.metrologia49(tech. Suppl.), (2012). 241

MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM *

MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM * MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM * C. Kessler and D.T. Burns December 2018 BUREAU INTERNATIONAL DES POIDS ET MESURES

More information

Monitoring the eye lens

Monitoring the eye lens Funding TS7e.3 Rolf Behrens Physikalisch-Technische Bundesanstalt Introduction April 2011: Limit of H lens : 150 msv 20 msv per year Protecting and monitoring may be necessary! Main question: Which quantity

More information

Introduction of a Single Chip TLD System for Patient Dosimetry

Introduction of a Single Chip TLD System for Patient Dosimetry Introduction of a Single Chip TLD System for Patient Dosimetry C. Hranitzky a, M. Halda a, G. Müller a, B. Obryk b, H. Stadtmann a* a Austrian Research Centers GmbH ARC, 2444 Seibersdorf, Austria. b Institute

More information

Calibration of KAP meters

Calibration of KAP meters Calibration of KAP meters Alexandr Malusek! Division of Radiological Sciences Department of Medical and Health Sciences Linköping University! 2014-04-15 1 Outline 1. KAP meter construction 2. Air kerma-area

More information

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This

More information

Preliminary studies of a new monitor ionization chamber

Preliminary studies of a new monitor ionization chamber 1 Preliminary studies of a new monitor ionization chamber Maíra T. Yoshizumi, Vitor Vivolo and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares (IPEN-CNEN) Comissão Nacional de Energia

More information

Development of Personal Dosimeter Using Electronic Dose Conversion Method

Development of Personal Dosimeter Using Electronic Dose Conversion Method Proceedings of the Korean Nuclear Spring Meeting Gyeong ju, Korea, May 2003 Development of Personal Dosimeter Using Electronic Dose Conversion Method Wanno Lee, Bong Jae Lee, and Chang Woo Lee Korea Atomic

More information

Update on NIST X-Ray Air-Kerma Standards and Calibrations

Update on NIST X-Ray Air-Kerma Standards and Calibrations CCRI(I)/03-19 Update on NIST X-Ray Air-Kerma Standards and Calibrations Michelle O Brien and Stephen M. Seltzer Ionizing Radiation Division National Institute of Standards and Technology Gaithersburg,

More information

2012 :15th SESSION of ESMP

2012 :15th SESSION of ESMP 2012 :15th SESSION of ESMP Lecture presented in Archamps (Salève Building) by : Elly CASTELLANO (London) Patient dosimetry in x-ray imaging and CT Elly Castellano Objectives measurable dose quantities

More information

Keysight Technologies Optical Power Meter Head Special Calibrations. Brochure

Keysight Technologies Optical Power Meter Head Special Calibrations. Brochure Keysight Technologies Optical Power Meter Head Special Calibrations Brochure Introduction The test and measurement equipment you select and maintain in your production and qualification setups is one of

More information

Ideas for type tests of electronic dosemeters in pulsed fields

Ideas for type tests of electronic dosemeters in pulsed fields Ideas for type tests of electronic dosemeters in pulsed fields Oliver Hupe, Peter Ambrosi Physikalisch-Technische Bundesanstalt Fachbereich 6.3 Strahlenschutzdosimetrie Bundesallee 100, 38116 Braunschweig

More information

CIPM and CCPR What are these organizations and how do they affect my testing results. Maria Nadal Photometry, Surface Color and Appearance NIST

CIPM and CCPR What are these organizations and how do they affect my testing results. Maria Nadal Photometry, Surface Color and Appearance NIST CIPM and CCPR What are these organizations and how do they affect my testing results Maria Nadal Photometry, Surface Color and Appearance NIST CIE USA Annual Meeting October 6-7, 2014 Calibration Laboratory

More information

DETECTORS. Including Codes of Practice IONIZING RADIATION

DETECTORS. Including Codes of Practice IONIZING RADIATION DETECTORS Including Codes of Practice 2015 IONIZING RADIATION PTW History and General Remarks Looking back on a long history PTW-Freiburg is an internationally operating company, manufacturing and marketing

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information

Improved Imaging Plate Dosimetry for X-rays in Interventional Radiology

Improved Imaging Plate Dosimetry for X-rays in Interventional Radiology Improved Imaging Plate Dosimetry for X-rays in Interventional Radiology H. Ohuchi 1, A. Yamadera 2, T.Satoh 3, Y,Eguchi 3 1 Graduate School of Pharmaceutical Sciences, Tohoku University, Aramaki,Aoba-ku,

More information

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS By Aravind Ravichandran arr192@mail.usask.ca University of Saskatchewan Address: 2424 Cumberland

More information

Electronic Brachytherapy Sources. Thomas W. Rusch

Electronic Brachytherapy Sources. Thomas W. Rusch Electronic Brachytherapy Sources Thomas W. Rusch Educational Objectives Understand key elements of ebx source construction & operation Understand the rationale and methods for air kerma strength calibration

More information

IONIZING RADIATION 2018 / 2019 DETECTORS. Including Codes of Practice

IONIZING RADIATION 2018 / 2019 DETECTORS. Including Codes of Practice IONIZING RADIATION DETECTORS 2018 / 2019 Including Codes of Practice PTW History and General Remarks Looking back on a long history PTW-Freiburg is an internationally operating company, manufacturing and

More information

Response characteristics of a tandem ionization chamber in standard X-ray beams

Response characteristics of a tandem ionization chamber in standard X-ray beams Applied Radiation and Isotopes 58 (2003) 495 500 Response characteristics of a tandem ionization chamber in standard X-ray beams Alessandro M. Costa*, Linda V.E. Caldas Instituto de Pesquisas Energ!eticas

More information

DETERMINATION OF INHERENT AND ADDITIONAL FILTRATION IN ORDER TO ESTABLISH RADIATION QUALITIES ACCORDING TO IEC 61267

DETERMINATION OF INHERENT AND ADDITIONAL FILTRATION IN ORDER TO ESTABLISH RADIATION QUALITIES ACCORDING TO IEC 61267 2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 DETERMINATION

More information

Joint ILAC CIPM Communication regarding the. Accreditation of Calibration and Measurement Services. of National Metrology Institutes.

Joint ILAC CIPM Communication regarding the. Accreditation of Calibration and Measurement Services. of National Metrology Institutes. Joint ILAC CIPM Communication regarding the Accreditation of Calibration and Measurement Services of National Metrology Institutes 7 March 2012 Authorship This document was prepared by the International

More information

First EURADOS intercomparison exercise of eye lens dosemeters for medical applications

First EURADOS intercomparison exercise of eye lens dosemeters for medical applications First EURADOS intercomparison exercise of eye lens dosemeters for medical applications I. Clairand 1, M. Ginjaume 2, F. Vanhavere 3, E. Carinou 4, J. Daures 5, M. Denoziere 5, E. Honorio da Silva 3, M.

More information

Determination of the tube voltage from clinic mammographic system using two types of detectors

Determination of the tube voltage from clinic mammographic system using two types of detectors BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 03-1A (2015) 01-06 Determination of the tube voltage from clinic mammographic system using two types of detectors J. S. Barreira a ; V. Vivolo a a Gerência

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

Test Equipment for Radiology and CT Quality Control Contents

Test Equipment for Radiology and CT Quality Control Contents Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3

More information

Cylindrical Ion Chambers Victoreen Model 550 Series

Cylindrical Ion Chambers Victoreen Model 550 Series Cylindrical Ion Chambers Victoreen Model 550 Series! Cylindrical Ion Chambers for use with Model 35040 and Model 530 electrometers! Wide range of applications in Diagnostic X-Ray and Radiation Oncology

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced

More information

Ensuring Shielding adequacy in Lead shielded spent fuel transportation casks using gamma scanning

Ensuring Shielding adequacy in Lead shielded spent fuel transportation casks using gamma scanning Ensuring Shielding adequacy in Lead shielded spent fuel transportation casks using gamma scanning More info about this article: http://www.ndt.net/?id=21208 M.Ravichandra 1, P.Raghavendra 1, Dhiren Kothari

More information

Radiation Protection Instruments PTB-A 23.2

Radiation Protection Instruments PTB-A 23.2 These PTB Requirements are originally written in German. In case of any conflict between the German language version the English translation of it, the German version shall prevail. PTB Requirements Physikalisch-

More information

Barcelona, January 2011 Workshop on optimization of Radiation Protection of Medical Staff

Barcelona, January 2011 Workshop on optimization of Radiation Protection of Medical Staff USE OF ACTIVE PERSONAL DOSEMETERS IN INTERVENTIONAL RADIOLOGY/CARDIOLOGY: TESTS WITH CONTINUOUS AND PULSED FIELDS IN LABORATORY CONDITIONS ORAMED PROJECT I. Clairand 1, J-M. Bordy 2, E. Carinou 3, J. Daures

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

TECHNICAL DATA. benefits

TECHNICAL DATA. benefits benefits > Instant & direct, non-destructive reading of radiation dose > Zero or very low power consumption > Large dynamic range > Smallest active volume of all dosimeters > Easily integrated into an

More information

A STUDY OF THE RESPONSE OF DEPLETED TYPE p-mosfets TO PHOTON AND ELECTRON DOSE

A STUDY OF THE RESPONSE OF DEPLETED TYPE p-mosfets TO PHOTON AND ELECTRON DOSE A STUDY OF THE RESPONSE OF DEPLETED TYPE p-mosfets TO PHOTON AND ELECTRON DOSE M. Fragopoulou 1, S. Stoulos 1, and M. Zamani 1, 1 School of Physics, Nuclear and Elementary Particle Section Aristotle University

More information

Cylindrical Ion Chambers

Cylindrical Ion Chambers Cylindrical Ion Chambers Radiation Oncology ON Victoreen Model 550T Series Cylindrical Ion Chambers for use with Model 35040 and Model 560 electrometers Wide range of applications in Diagnostic X-Ray and

More information

ORAMED S MEASUREMENT AND SIMULATION CAMPAIGN FOR EXREMITY AND EYE LENS DOSES OF MEDICAL STAFF INVOLVED IN INTERVENTIONAL RADIOLOGY AND CARDIOLOGY

ORAMED S MEASUREMENT AND SIMULATION CAMPAIGN FOR EXREMITY AND EYE LENS DOSES OF MEDICAL STAFF INVOLVED IN INTERVENTIONAL RADIOLOGY AND CARDIOLOGY ORAMED S MEASUREMENT AND SIMULATION CAMPAIGN FOR EXREMITY AND EYE LENS DOSES OF MEDICAL STAFF INVOLVED IN INTERVENTIONAL RADIOLOGY AND CARDIOLOGY Carinou E., Brodecki M.,Domienik J., Donadille L., Ferrari

More information

Bilateral Comparison EURAMET.AUV.V-K1.2. (vibration acceleration) Swiss Federal Office of Metrology METAS Christian Hof

Bilateral Comparison EURAMET.AUV.V-K1.2. (vibration acceleration) Swiss Federal Office of Metrology METAS Christian Hof Final report 2011 04 20 Bilateral Comparison EURAMET.AUV.V-K1.2 (vibration acceleration) Swiss Federal Office of Metrology METAS Christian Hof Abstract This report describes the results obtained in a bilateral

More information

DETECTORS UNCOMPROMISING QUALITY. The standard in dosimetry measurements for over 40 years. EXRADIN DETECTORS

DETECTORS UNCOMPROMISING QUALITY. The standard in dosimetry measurements for over 40 years. EXRADIN DETECTORS DETECTORS UNCOMPROMISING QUALITY The standard in dosimetry measurements for over 40 years. EXRADIN DETECTORS The Exradin Advantage Better Components Waterproof construction eliminates the need for sleeves

More information

熊本大学学術リポジトリ. Kumamoto University Repositor

熊本大学学術リポジトリ. Kumamoto University Repositor 熊本大学学術リポジトリ Kumamoto University Repositor Title Monte Carlo calculations of the rep correction factor, Ρ_, for cy chamber cav Author(s) Araki, Fujio CitationRadiological Physics and Technology Issue

More information

Wide beam CT dosimetry. Elly Castellano

Wide beam CT dosimetry. Elly Castellano Wide beam CT dosimetry Elly Castellano Outline revision: CT dose indices wide-beam CT: the end of the road for CTDI? the IEC rescue plan for CTDI 100 the american way AAPM report 111 better estimates of

More information

Detectors for small field dosimetry

Detectors for small field dosimetry Detectors or small ield dosimetry Hugo Palmans MedAustron, Wiener Neustadt, Austria and National Physical Laboratory, Teddington, UK 1 080915 Overview Ideal detector Water calorimeter Ionization chamber

More information

Technical aspects on DAP calibration and CT calibration

Technical aspects on DAP calibration and CT calibration EURADOS Report 2015-03 European Radiation Dosimetry Group e. V. Braunschweig, August 2015 Technical aspects on DAP calibration and CT calibration H. Järvinen, O. Turak, M. Ginjaume, J. Daures, C. Hourdakis,

More information

Post-irradiation colouration of Gafchromic EBT radiochromic film

Post-irradiation colouration of Gafchromic EBT radiochromic film INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 5 (25) N281 N285 PHYSICS IN MEDICINE AND BIOLOGY doi:1.188/31-9155/5/2/n4 NOTE Post-irradiation colouration of Gafchromic EBT radiochromic film Tsang Cheung

More information

JRPR. Microwave Dielectric Absorption Spectroscopy Aiming at Novel Dosimetry Using DNAs. Original Research. Introduction

JRPR. Microwave Dielectric Absorption Spectroscopy Aiming at Novel Dosimetry Using DNAs. Original Research. Introduction Journal of Radiation Protection and Research 2017;42(1):21-25 pissn 2508-1888 eissn 2466-2461 Microwave Dielectric Absorption Spectroscopy Aiming at Novel Dosimetry Using DNAs Yoshinobu Izumi 1, *, Makoto

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 16371-1 First edition 2011-10-01 Non-destructive testing Industrial computed radiography with storage phosphor imaging plates Part 1: Classification of systems Essais non destructifs

More information

Evaluation of X-Ray Beam Quality Based on Measurements and Estimations Using SpekCalc and Ipem78 Models

Evaluation of X-Ray Beam Quality Based on Measurements and Estimations Using SpekCalc and Ipem78 Models Original Article Evaluation of X-Ray Beam Quality Based on Measurements and Estimations Using SpekCalc and Ipem78 Models Suk Chiang Chen, Wei Loong Jong, Ahmad Zaky Harun Submitted: 20 Aug 2011 Accepted:

More information

Borderline between CIPM MRA and testing activities

Borderline between CIPM MRA and testing activities Borderline between CIPM MRA and testing activities Michela Sega, EURAMET TC-MC Chair INRIM-Italy 7 th Meeting of the Focus Group on Facilitating National Metrology Infrastructure Development Zagreb, Croatia,

More information

DESIGN AND OPERATION OF A WIDE RANGE SEGMENTED GAMMA RAY SCANNING ASSAY INSTRUMENT FOR THE MEASUREMENT OF BOTH LOW AND INTERMEDIATE LEVEL WASTE

DESIGN AND OPERATION OF A WIDE RANGE SEGMENTED GAMMA RAY SCANNING ASSAY INSTRUMENT FOR THE MEASUREMENT OF BOTH LOW AND INTERMEDIATE LEVEL WASTE 11-A-424-INMM DESIGN AND OPERATION OF A WIDE RANGE SEGMENTED GAMMA RAY SCANNING ASSAY INSTRUMENT FOR THE MEASUREMENT OF BOTH LOW AND INTERMEDIATE LEVEL WASTE John A. Mason, Marc R. Looman, Robert A. Price

More information

33 rd Asia Pacific Metrology Programme General Assembly 30 November 1 December 2017 New Delhi, India

33 rd Asia Pacific Metrology Programme General Assembly 30 November 1 December 2017 New Delhi, India 33 rd Asia Pacific Metrology Programme General Assembly 30 November 1 December 2017 New Delhi, India Laboratory Report National Metrology Institute of Japan (NMIJ) National Institute of Advanced Industrial

More information

Version 1.1 (01/2015) EURAMET e.v. Bundesallee 100 D Braunschweig Germany. Phone:

Version 1.1 (01/2015) EURAMET e.v. Bundesallee 100 D Braunschweig Germany.   Phone: Authorship and Imprint This document was developed by the EURAMET e.v. Authors: W. Schmid (EURAMET), M. Sega (INRIM, Italy, EURAMET TC-MC Chair), J. Drnovsek (MIRS/UL-FE/LMK, Slovenia, EURAMET Vice-Chairperson

More information

Energy Monitoring Device for Electron Beam Facilities

Energy Monitoring Device for Electron Beam Facilities 1 SM/EB-18 Energy Monitoring Device for Electron Beam Facilities M. Lavalle 1, P.G. Fuochi 1, A. Martelli 1, U. Corda 1, A. Kovács 2, K. Mehta 3, F. Kuntz 4, S. Plumeri 4 1 CNR-ISOF, Via P. Gobetti 101,

More information

2-5 Frequency Calibration

2-5 Frequency Calibration 2-5 Frequency SAITO Haruo, IWAMA Tsukasa, TSUCHIYA Shigeru, and KOYAMA Yasuhiro The Japan Standard Time (JST) and the Coordinated Universal Time (UTC(NICT)), which are constructed by National Institute

More information

Effect of pressure, temperature and humidity in air on photon fluence and air kerma values at low photon energies

Effect of pressure, temperature and humidity in air on photon fluence and air kerma values at low photon energies ARTICLE IN PRESS Radiation Physics and Chemistry 68 (2003) 707 720 Effect of pressure, temperature and humidity in air on photon fluence and air kerma values at low photon energies M. Assiamah, D. Mavunda,

More information

GAFCHROMIC HD-810 Radiochromic Dosimetry Film Configuration, Specifications and Performance Data

GAFCHROMIC HD-810 Radiochromic Dosimetry Film Configuration, Specifications and Performance Data GAFCHROMIC HD-810 Radiochromic Dosimetry Film Configuration, Specifications and Performance Data Description GAFCHROMIC HD-810 dosimetry film is designed for the measurement of absorbed dose of high-energy

More information

ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone

ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY 12477 (800)VIP-XRAY (845)247-9028 Fax or 800-847-9729 local phone 845-246-9013 http://www.eccxray.com sales@eccxray.com INSTRUCTION MANUAL

More information

Bureau International des Poids et Mesures. International Recognition of NMI Calibration and Measurement Capabilities: The CIPM MRA

Bureau International des Poids et Mesures. International Recognition of NMI Calibration and Measurement Capabilities: The CIPM MRA Bureau International des Poids et Mesures International Recognition of NMI Calibration and Measurement Capabilities: The CIPM MRA Prof. Michael Kühne International School of Physics Enrico Fermi Metrology

More information

SMD ENS Metrology - National Standards. Short overview. Hugo Pirée. World Metrology Day 2018 Brussels.

SMD ENS Metrology - National Standards. Short overview. Hugo Pirée. World Metrology Day 2018 Brussels. SMD ENS Metrology - National Standards Short overview Hugo Pirée World Metrology Day 2018 Brussels Federal government FPS ECONOMY Gen. Direction QUALITY & SAFETY Division METROLOGY 14 persons National

More information

ILAC input to CIPM MRA Review Workshop October 2015

ILAC input to CIPM MRA Review Workshop October 2015 ILAC input to CIPM MRA Review Workshop 13-14 October 2015 BIPM Sevres 13 October 2015 By Erik Oehlenschlaeger ILAC (DANAK) ILAC History ILAC first started as a conference in 1977 (Copenhagen) with the

More information

FREQUENCY COMPARISON AT 633 NM WAVELENGTH: DETERMINATION OF DIAGONAL ELEMENTS OF MATRIX MEASUREMENTS BY USING A MASTER-SLAVE He-Ne LASER SYSTEM

FREQUENCY COMPARISON AT 633 NM WAVELENGTH: DETERMINATION OF DIAGONAL ELEMENTS OF MATRIX MEASUREMENTS BY USING A MASTER-SLAVE He-Ne LASER SYSTEM Journal of Optoelectronics and Advanced Materials Vol. 2, No. 3, September 2000, p. 267-273 FREQUENCY COMPARISON AT 633 NM WAVELENGTH: DETERMINATION OF DIAGONAL ELEMENTS OF MATRIX MEASUREMENTS BY USING

More information

Measurements of MeV Photon Flashes in Petawatt Laser Experiments

Measurements of MeV Photon Flashes in Petawatt Laser Experiments UCRL-JC-131359 PREPRINT Measurements of MeV Photon Flashes in Petawatt Laser Experiments M. J. Moran, C. G. Brown, T. Cowan, S. Hatchett, A. Hunt, M. Key, D.M. Pennington, M. D. Perry, T. Phillips, C.

More information

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS 7:28-14.1 Scope (a) This subchapter covers therapeutic installations used in the healing arts. These therapeutic installations include x-ray, accelerator and teletherapy

More information

Metrology at the service of the economy, society and citizens

Metrology at the service of the economy, society and citizens Bureau International des Poids et Mesures Metrology at the service of the economy, society and citizens Dr Martin Milton Director, BIPM 26 th Sept 2013 Bureau International des Poids et Mesures Established

More information

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

3D Diode Array Commissioning: Building Confidence in 3D QA Technology 3D Diode Array Commissioning: Building Confidence in 3D QA Technology Caroline Yount, MS CANCER CENTER 3D QA The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose

More information

COOMET Pilot Comparison 473/RU-a/09: Comparison of hydrophone calibrations in the frequency range 250 Hz to 200 khz

COOMET Pilot Comparison 473/RU-a/09: Comparison of hydrophone calibrations in the frequency range 250 Hz to 200 khz COOMET Pilot Comparison 473/RU-a/09: Comparison of hydrophone calibrations in the frequency range 250 Hz to 200 khz Chen Yi 1, A E Isaev 2, Wang Yuebing 1, A M Enyakov 2, Fei Teng 1 and A N Matveev 2 1

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

Report on the CCPR Pilot Comparison. Spectral Responsivity 10 nm to 20 nm

Report on the CCPR Pilot Comparison. Spectral Responsivity 10 nm to 20 nm Page 1 of 30 Report on the CCPR Pilot Comparison Spectral Responsivity 10 nm to 20 nm reported by Frank Scholze Physikalisch-Technische Bundesanstalt Berlin Germany October 2009 Page 2 of 30 Page 3 of

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

SCINTILLATING FIBER DOSIMETER ARRAY

SCINTILLATING FIBER DOSIMETER ARRAY SCINTILLATING FIBER DOSIMETER ARRAY FIELD OF THE INVENTION [0001] This invention relates generally to the field of dosimetry and, more particularly, to rapid, high-resolution dosimeters for advanced treatment

More information

IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film

IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film CYRIC Annual Report 2005 IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film Ohuchi H. 1, and Abe K. 2 1 Graduate School of Pharmaceutical Sciences,

More information

Traceability measurement results of accurate time and frequency in Bosnia and Herzegovina

Traceability measurement results of accurate time and frequency in Bosnia and Herzegovina INFOTEH-JAHORINA Vol. 11, March 2012. Traceability measurement results of accurate time and frequency in Bosnia and Herzegovina Osman Šibonjić, Vladimir Milojević, Fatima Spahić Institute of Metrology

More information

MWPC Gas Gain with Argon-CO 2 80:20 Gas Mixture

MWPC Gas Gain with Argon-CO 2 80:20 Gas Mixture IMA Journal of Mathematical Control and Information Page 1 of 10 doi:10.1093/imamci/dri000 1. Principles of Operation MWPC Gas Gain with Argon-CO 2 80:20 Gas Mixture Michael Roberts A multi-wire proportional

More information

Dose conversion coefficients for electron exposure of the human eye lens

Dose conversion coefficients for electron exposure of the human eye lens Physics in Medicine & Biology CORRIGENDUM Dose conversion coefficients for electron exposure of the human eye lens To cite this article: R Behrens et al 2010 Phys. Med. Biol. 55 3937 View the article online

More information

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

Unit thickness. Unit area. σ = NΔX = ΔI / I 0 Unit thickness I 0 ΔI I σ = ΔI I 0 NΔX = ΔI / I 0 NΔX Unit area Δx Average probability of reaction with atom for the incident photons at unit area with the thickness of Delta-X Atom number at unit area

More information

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Gena M.A.H 1, Ahmed L.El-Attar 2, Elbadry M. Zahran 3, Hany El-Gamal

More information

Continuous development of the national standard for vibration

Continuous development of the national standard for vibration Continuous development of the national standard for vibration by Ian Veldman, National Metrology Laboratory, CSIR In this fast technologically-advancing world, higher measurement accuracy is part of the

More information

When small things matter. Small Field Dosimetry Application Guide

When small things matter. Small Field Dosimetry Application Guide R A D I AT I O N T H E R A P Y When small things matter. Small Field Dosimetry Application Guide Contents 1 Introduction 1 Introduction 2 2 The Physics of Small Fields 3 3 Detector Types 10 4 Detector

More information

The Physikalisch-Technische Bundesanstalt,

The Physikalisch-Technische Bundesanstalt, Physikalisch-Technische Bundesanstalt the National Metrology Institute of Germany The Physikalisch-Technische Bundesanstalt, Germany snational metrologyinstitute, isa scientific and technical higher federal

More information

Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology

Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology Poster No.: C-1190 Congress: ECR 2011 Type: Scientific Exhibit Authors: K. Chida, T. Takahashi, D.

More information

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY 12 th A-PCNDT 2006 Asia-Pacific Conference on NDT, 5 th 10 th Nov 2006, Auckland, New Zealand PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY Rajashekar

More information

Prompt-Gamma Based Range Verification in Particle Therapy: New prospects (also) for 4D?

Prompt-Gamma Based Range Verification in Particle Therapy: New prospects (also) for 4D? Prompt-Gamma Based Range Verification in Particle Therapy: New prospects (also) for 4D? Guntram Pausch 1-3 1 OncoRay National Center for Radiation Research in Oncology, Dresden 2 Technische Universität

More information

Enhanced Low Dose Rate Sensitivity (ELDRS) Radiation Testing of the RH1814MW Quad Op Amp for Linear Technology

Enhanced Low Dose Rate Sensitivity (ELDRS) Radiation Testing of the RH1814MW Quad Op Amp for Linear Technology Enhanced Low Dose Rate Sensitivity (ELDRS) Radiation Testing of the RH1814MW Quad Op Amp for Linear Technology Customer: Linear Technology (PO 57472L) RAD Job Number: 10-417 Part Type Tested: Linear Technology

More information

Use of the BVD for traceability of bipolar DC voltage scale from 1 mv up to 1200 V

Use of the BVD for traceability of bipolar DC voltage scale from 1 mv up to 1200 V Use of the BVD for traceability of bipolar DC voltage scale from 1 mv up to 1200 V Speaker: Roman Honig, MI-Europe, Druzstevni 845, 686 05 Uherske Hradiste, Czech Republic, Tel.: #420 731 440 665, Fax:

More information

CIE 220:2016 Characterization and Calibration Method of UV Radiometers

CIE 220:2016 Characterization and Calibration Method of UV Radiometers CIE 220:2016 Characterization and Calibration Method of UV Radiometers Anton Gugg-Helminger Gigahertz-Optik GmbH, Germany www.gigahertz-optik.de Editor s note: This article has been reprinted from UV News,

More information

Enhanced Low Dose Rate Sensitivity (ELDRS) Radiation Testing of the RH118W Op-Amp for Linear Technology

Enhanced Low Dose Rate Sensitivity (ELDRS) Radiation Testing of the RH118W Op-Amp for Linear Technology Enhanced Low Dose Rate Sensitivity (ELDRS) Radiation Testing of the RH118W Op-Amp for Linear Technology Customer: Linear Technology, PO# 60225L RAD Job Number: 11-351 Part Type Tested: RH118W Op-Amp, RH118

More information

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT Quality control of Gamma Camera By Dr/ Ibrahim Elsayed Saad 242 NMT WHAT IS QUALITY? The quality of a practice is to fulfill the expectations and demands from: Patient Clinicain Your self Quality assurance

More information

Derivation of total filtration thickness for diagnostic x-ray source assembly

Derivation of total filtration thickness for diagnostic x-ray source assembly Physics in Medicine & Biology PAPER OPEN ACCESS Derivation of total filtration thickness for diagnostic x-ray source assembly To cite this article: Michiharu Sekimoto and Yoh Katoh 2016 Phys. Med. Biol.

More information

A Multichannel Dosimeter Based on Scintillating Fibers for Medical Applications

A Multichannel Dosimeter Based on Scintillating Fibers for Medical Applications A Multichannel Dosimeter Based on Scintillating Fibers for Medical Applications K.-H. Becks 1, J. Drees 1, K. Goldmann 1, I.M. Gregor 1 2, M. Heintz, Arndt Roeser* 1 Fachbereich Physik, Bergische Universität-Gesamthochschule

More information

New Detectors for X-Ray Metal Thickness Measuring

New Detectors for X-Ray Metal Thickness Measuring ECNDT 2006 - Poster 132 New Detectors for X-Ray Metal Thickness Measuring Boris V. ARTEMIEV, Alexander I. MASLOV, Association SPEKTR- GROUP, Moscow, Russia Abstract. X-ray thickness measuring instruments

More information

Monte Carlo study on a new concept of a scanning photon beam system for IMRT

Monte Carlo study on a new concept of a scanning photon beam system for IMRT NUKLEONIKA 2011;56(4):291 297 ORIGINAL PAPER Monte Carlo study on a new concept of a scanning photon beam system for IMRT Anna M. Wysocka-Rabin, Günter H. Hartmann Abstract. Intensity-modulated radiation

More information

Active Tissue Equivalent Dosimeter (ATED): a tissue equivalent proportional counter currently operating aboard ISS

Active Tissue Equivalent Dosimeter (ATED): a tissue equivalent proportional counter currently operating aboard ISS Active Tissue Equivalent Dosimeter (ATED): a tissue equivalent proportional counter currently operating aboard ISS Eric Benton 1, Oliver Causey 1, Bryan Hayes 1, Arthur Lucas 1, Satoshi Kodaira 2, Hisashi

More information

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage

More information

THE DOSIMETRY OF A HIGHLY-COLLIMATED BREMSSTRAHLUNG SOURCE IN AIR

THE DOSIMETRY OF A HIGHLY-COLLIMATED BREMSSTRAHLUNG SOURCE IN AIR THE DOSIMETRY OF A HIGHLY-COLLIMATED BREMSSTRAHLUNG SOURCE IN AIR A Dissertation Presented to The Academic Faculty by Michael Paul Shannon In Partial Fulfillment of the Requirements for the Degree Doctor

More information

- Water resistant. - Large size.

- Water resistant. - Large size. GAFCHROMIC EBT product brochure GAFCHROMIC EBT FEATURES GAFCHROMIC EBT dosimetry film has been developed specifically to address the needs of the medical physicist and dosimetrist working in the radiotherapy

More information

APMP.AUV.V-K1 Regional Key Comparison of Standard Accelerometer. Final Report. Project Number: APMP-IC Shing Chen and Hans-Jürgen von Martens

APMP.AUV.V-K1 Regional Key Comparison of Standard Accelerometer. Final Report. Project Number: APMP-IC Shing Chen and Hans-Jürgen von Martens Regional Key Comparison of Standard Accelerometer Final Report Project Number: APMP-IC-4-95 Shing Chen and Hans-Jürgen von Martens January 2001 Abstract In project APMP-IC-4-95 (BIPM comparison identifier:

More information

Physikalisch Technische Bundesanstalt

Physikalisch Technische Bundesanstalt EURAMET Intercomparison: Involute Gear Artifacts 1 Physikalisch Technische Bundesanstalt EURAMET Intercomparison Involute Gear Artifacts Technical Protocol Rev 3 EURAMET Intercomparison: Involute Gear

More information

Robert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center. Thursday September 8, 2011

Robert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center. Thursday September 8, 2011 Robert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center Thursday September 8, 2011 Quality Control Quality control should be performed because:

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

QA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion

QA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion Leksell Gamma Knife Perfexion QA Considerations Paula L. Petti, Ph.D. Taylor McAdam Bell Neuroscience Institute Washington Hospital Healthcare System Fremont, CA 1 QA for LGK Perfexion : : Follow NRC licensing

More information

Gamex CR 2.0 Program description and operating manual

Gamex CR 2.0 Program description and operating manual Gamex CR 2.0 Program description and operating manual Issue No. : 2.0 Date of Issue : Jan. 2013 Z.U.T. NDT SOFT http://www.ndtsoft.eu Copyright (c) 2013 by Z.U.T. NDT SOFT All Rights Reserved Disclaimer

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-649 CDRH Fluoroscopic Phantom Users Manual March 2005 Manual No. 07-649-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

Certificate of Calibration

Certificate of Calibration Certificate of Calibration ISO/IEC 17025:2005 and ANSI/NCSL Z540.1-1994 Model Number Manufacturer Description Serial Number Customer Asset No. 11667A Keysight Technologies Inc DC-18 GHz power splitter,

More information