1 Robert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center Thursday September 8, 2011

2 Quality Control Quality control should be performed because: Detector damage High Voltage drift Electronic component degradation Quality control is necessary to assure accurate results

3 Equipment Requiring QC Gamma Camera Survey Meter Dose Calibrator Other lab equipment

4 Quality Control Procedures- Planar camera Energy Peaking daily Uniformity test daily Linearity and resolution weekly Sensitivity test weekly

5 Energy Peaking Verifying that the pulse height analyzer s energy window over the photopeak energy Best performed 5 UFOVs from the uncollimated camera with a point source Can use a sheet source with collimated camera Most commonly performed with a collimator Often automated must check!

6 Uniformity Test Best performed 5 UFOVs from the uncollimated camera Most commonly performed with a Co-57 sheet source on a collimated camera Also checks the collimator MUCH faster Suggestion Perform extrinsically daily Perform intrinsically weekly

7 Uniformity Test Extrinsic (with collimator) Co-57 radioactive sheet source is placed on top of each detector with the collimator still in place. An image is then taken on each detector for 3-5 million counts. 256x256 matrix A visual assessment (qualitative) of the image is to be made as well as a quantitative assessment.

8 Co-57 Sheet Source Dimensions 17 x 11 cm - 25 x 18 cm 5 20 mci $3000.

9 Uniformity Test Integral uniformity Global parameter measuring uniformity over an extended area 100% x (max-min)/(max+min) Differential uniformity Regional parameter that measures uniformity over a small area Performed using all 5x1-pixel area in both the X and Y direction 100% x Largest deviation (max-min)/(max+min)

10 Performance parameters for detectors- ASNC guidelines PARAMETER STANDARD PREFERRED Integral Uniformity Differential Uniformity <5 <3 <5 <3

11 Uniformity Test

12 Uniformity Test Non-Uniformity caused by Malfunctioning PMT Off-peak/Wrong peak Cracked crystal

13 Uniformity Test ABOVE PEAK BELOW PEAK

14 Resolution and Linearity Performed to Document spatial resolution and its change over time Evaluate ability to image straight lines Image a flood source through a spatial resolution test phantom (using uniformity flood paramaters) If a 4-quadrant phantom is used, it should be rotated 90 degrees so that every 5 th time the pattern repeats. Evaluation How straight are the lines What are the smallest bars discerned

15 Resolution and Linearity

16 Sensitivity Test Documents the system sensitivity as well as the change of sensitivity over time. Cts/pixel/millicurie To measure change in sensitivity: record the time taken to acquire a preset number of counts using a set dose.

17 INSTRUMENTATION ARTIFACTS NON-UNIFORMITY Ring Artifact Uniform Flood Non-Uniform Flood FBP

18 Quality Control Procedures- SPECT camera COR weekly or monthly Uniformity correction manufacturer recommendation Plexiglas phantom quarterly or annually

19 COR Checks for alignment error between the electronic matrix of the detector and the mechanical COR. Always check after camera servicing and software upgrades. Doughnut artifact (360º) or tuning fork artifact (180º) in the transverse images

20 COR COR is a measure of the alignment of opposite views of camera head position Misalignment (Offset camera and center of acquisition matrix) results in donut or tuning fork artifact

21 COR A 1 mci point source of Tc99m is placed on camera table SPECT data is acquired over a 360 (or 180) degree orbit 60 projections per head 5-10 sec/frame

22

23 COR

24 Plexiglass SPECT Phantoms

25 Plexiglass SPECT Phantoms To determine 3D contrast, resolution, uniformity 1. Collimator - LEHR (or LEUHR if used routinely with Tc-99m, whichever is higher resolution) 2. Tc-99m Activity - Between mci 3. Count Rate needs to be below 50 kcps 4. Matrix size = 128x Rotation = 360 degrees 6. Total Views = 120. (Implies 60 steps for dual head, 3-degrees per step) 7. Dual Headed Camera study acquired with camera head in opposed position (180) 8. Acquire for approximately 32 million total counts 9. Radius of Rotation = as close to 20 cm as possible. 10. Orbit = Circular

26 Plexiglass SPECT Phantoms

27 High Count Extrinsic Flood-Field Uniformity Correction In SPECT it is assumed that efficiency of photon detection is uniform across the collimated detector Camera systems store flood-field correction maps to correct variations before reconstruction 30 to 100 million ct images for each collimator 256x256 matrix

28 Image Artifact

29 Survey Meter Used daily to detect radioactive contamination in laboratory Aka Geiger counter, Geiger-Mueller counter, GM counter

30 GM Counter Battery Check Daily Ensures that the meter has adequate electrical power. The meter located in the readout scale will be located within the battery region if the power is adequate.

31 GM Counter Source check Daily Ensure that the meter responds to a radiation source.

32 GM Counter Calibration Annually Ensures that the meter responds accurately to radiation.

33 Dose Calibrator Used to measure the activity of the radiopharmaceutical

34 Dose Calibrator Standard testing Background Voltage Accuracy Constancy Linearity Geometry

35 Dose Calibrator Background Performed daily Background activity is checked at each setting that will be utilized Contamination Unshielded source in lab

36 Dose Calibrator Voltage Performed daily Ensures that the operating voltage is within manufacturers recommended range

37 Dose Calibrator Accuracy Performed Annually Two sealed radiation sources greater than 50 µci and one with an energy between 100 to 500 kev are assayed Isotopes used: Co-57 (122 kev, 271 days) Cs-137 (662 kev, 30 years) Ns-22 Co-60, Ge-68, or Ba-133.

38 Dose Calibrator The sources must be traceable to The National Institute of Standards and Technology (NIST) Compared to decay corrected calibrated activity Results must be within 5%

39 Dose Calibrator Geometry Performed at installation, recalibration, etc. Fine print (plug-in) Assess accuracy of the dose calibrators ability to measure consistent activity with different volumes A given amount of radioactivity is assayed in a container and the activity is recorded. Saline is incrementally added to include range typically used in lab Measurements must be within 5%.

40 Dose Calibrator Constancy Performed daily Designed to show that a long half-life source yields reproducible readings Compared to previous days reading Within 5% of decay corrected accuracy readings Check each setting (isotope) button Activity used should be greater than 50 µci of Co-57 or Cs-137 (or another gamma emitting sealed source)

41 Dose Calibrator Constancy SETTING MON TUES WED THURS FRI Cs Tc-99m Tl I

42 Dose Calibrator Linearity Performed quarterly Assess ability to read both high and low activity Measured values must be within 5% of calculated values Two methods for linearity Decay method Sleeve method

43 Dose Calibrator Decay method Measure an initial high activity of a short halflife source. Various measurement are taken as it decays. Readings are compared to calculated decay. Elapsed Time (hr) Expected Reading (mci) Measured Reading (mci)

44 Dose Calibrator Sleeve method Decay is simulated by using lead sleeves of various thickness.

45 Dose Calibrator Subpart C--General Technical Requirements Possession, use, and calibration of instruments used to measure the activity of unsealed byproduct material (a) For direct measurements performed in accordance with 35.63, a licensee shall possess and use instrumentation to measure the activity of unsealed byproduct material before it is administered to each patient or human research subject. (b) A licensee shall calibrate the instrumentation required in paragraph (a) of this section in accordance with nationally recognized standards or the manufacturer's instructions. (c) A licensee shall retain a record of each instrument calibration required by this section in accordance with

46 Other Lab Equipment Glucometer Daily calibration Accuracy Defibrillator Battery Test Voltage Test

47 Be Alert All Day! Raw patient data should be carefully reviewed on each patient. All problems do not happen at night Cameras Survey meters Dose calibrator

48 THE END

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT Quality control of Gamma Camera By Dr/ Ibrahim Elsayed Saad 242 NMT WHAT IS QUALITY? The quality of a practice is to fulfill the expectations and demands from: Patient Clinicain Your self Quality assurance

More information

GS Introduction to Medical Physics IV Laboratory 5 Gamma Camera Characteristics

GS Introduction to Medical Physics IV Laboratory 5 Gamma Camera Characteristics GS02 0193 Introduction to Medical Physics IV Laboratory 5 Gamma Camera Characteristics Purpose: To introduce some of the basic characteristics of a gamma camera. This lab will introduce gamma camera QC

More information

NM Module Section 2 6 th Edition Christian, Ch. 3

NM Module Section 2 6 th Edition Christian, Ch. 3 NM 4303 Module Section 2 6 th Edition Christian, Ch. 3 Gas Filled Chamber Voltage Gas filled chamber uses Hand held detectors cutie pie Geiger counter Dose calibrators Cutie pie Chamber voltage in Ionization

More information

Initial Certification

Initial Certification Initial Certification Nuclear Medical Physics (NMP) Study Guide Part 2 Content Guide and Sample Questions The content of all ABR exams is determined by a panel of experts who select the items based on

More information

Acceptance Testing and Annual Physics Survey Recommendations for Gamma Camera, SPECT, and SPECT/CT Systems

Acceptance Testing and Annual Physics Survey Recommendations for Gamma Camera, SPECT, and SPECT/CT Systems AAPM REPORT NO. 177 Acceptance Testing and Annual Physics Survey Recommendations for Gamma Camera, SPECT, and SPECT/CT Systems The Report of AAPM Task Group 177 February 2019 DISCLAIMER: This publication

More information

Investigation of Multiple Head Registration / Center of Rotation for SPECT Gamma Cameras

Investigation of Multiple Head Registration / Center of Rotation for SPECT Gamma Cameras Egyptian J. Nucl. Med., Vol 2, No. 2, Dec. 2009 82 PHYSICS, Original Artical Investigation of Multiple Head Registration / Center of Rotation for SPECT Gamma Cameras Abdelsattar, M.B. Ph.D.; BuHumaid,

More information

Parameters Affecting on Intrinsic Uniformity Test For MEDISO

Parameters Affecting on Intrinsic Uniformity Test For MEDISO ISPUB.COM The Internet Journal of Nuclear Medicine Volume 5 Number 2 Parameters Affecting on Intrinsic Uniformity Test For MEDISO S Zobly, A Osman Citation S Zobly, A Osman. Parameters Affecting on Intrinsic

More information

Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM

Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM Preamplifiers and amplifiers The current from PMT must be further amplified before it can be processed and counted (the number of electrons yielded

More information

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT) Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in

More information

Nuclear Associates , &

Nuclear Associates , & Nuclear Associates 76-810, 76-814 76-815 & 76-818 Bar Phantoms and Test Patterns Operators Manual March 2005 Manual No. 76-810-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A.

More information

Results of the Measurement of the Collimator Hole Angulation for Different Collimators of SPECT with Adaptive Quality Control Phantom

Results of the Measurement of the Collimator Hole Angulation for Different Collimators of SPECT with Adaptive Quality Control Phantom Modern Instrumentation, 2012, 1, 4953 http://dx.doi.org/10.4236/mi.2012.14007 Published Online October 2012 (http://www.scirp.org/journal/mi) Results of the Measurement of the Collimator Hole Angulation

More information

Nuclear medicine is critically dependent on the accurate,

Nuclear medicine is critically dependent on the accurate, CONTINUING EDUCATION Routine Quality Control of Clinical Nuclear Medicine Instrumentation: A Brief Review* Pat Zanzonico Departments of Medical Physics and Radiology, Memorial Sloan-Kettering Cancer Center,

More information

2/14/2019. Nuclear Medicine Artifacts. Symmetric energy windows

2/14/2019. Nuclear Medicine Artifacts. Symmetric energy windows Nuclear Medicine Artifacts SCPMG Medical Imaging Technology & Informatics Medical Physics Group Brian Helbig, MS, DABR 1 2 Symmetric energy windows 3 1 Dynamic clinical study Energy peak shift Electrical

More information

T h e P h a n t o m L a b o r a t o r y

T h e P h a n t o m L a b o r a t o r y T h e P h a n t o m L a b o r a t o r y 1 ECTphan Phantom SMR330 M a n u a l Copyright 2015 WARNING The use of this phantom requires radioactive fill solutions. Only people trained in the safe handling

More information

Intrinsic and Tomographic Evaluation of Siemens e.cam SPECT System at the Korle-Bu Teaching Hospital (Ghana)

Intrinsic and Tomographic Evaluation of Siemens e.cam SPECT System at the Korle-Bu Teaching Hospital (Ghana) Research Journal of Applied Sciences, Engineering and Technology 3(10): 1152-1158, 2011 ISSN: 2040-7467 Maxwell Scientific Organization, 2011 Submitted: July 17, 2011 Accepted: September 05, 2011 Published:

More information

Contrast Evaluation for a Dual Head SPECT System with Different Energy Peaking- Drift

Contrast Evaluation for a Dual Head SPECT System with Different Energy Peaking- Drift PHYSICS, ORIGINAL ARTICLE Contrast Evaluation for a Dual Head SPECT System with Different Energy Peaking- Drift Saad, I. *, Mattar, E. **, Ashour, A. *** * Department of Clinical oncology and Nuclear Medicine,

More information

How Gamma Camera s Head-Tilts Affect Image Quality of a Nuclear Scintigram?

How Gamma Camera s Head-Tilts Affect Image Quality of a Nuclear Scintigram? November 2014, Volume 1, Number 4 How Gamma Camera s Head-Tilts Affect Image Quality of a Nuclear Scintigram? Hojjat Mahani 1,2, Alireza Kamali-Asl 3, *, Mohammad Reza Ay 2, 4 1. Radiation Application

More information

Factors Affecting the resolution of SPECT Imaging. h.

Factors Affecting the resolution of SPECT Imaging. h. Factors Affecting the resolution of SPECT Imaging H. E. Mostafa *1, H. A. Ayoub 2 and Sh.Magraby 1 1 Kasr El-Ini Center for Oncology, Cairo University, 2 Faculty of Science, Suez Canal University hayamayoub@yahoo.com

More information

J of Nuclear Medicine Technology, first published online July 27, 2011 as doi: /jnmt

J of Nuclear Medicine Technology, first published online July 27, 2011 as doi: /jnmt J of Nuclear Medicine Technology, first published online July 27, 2011 as doi:10.2967/jnmt.110.084814 Extrinsic Versus Intrinsic Uniformity Correction for g-cameras Randy Bolstad 1, Jody Brown, RT(N),

More information

QUALITY CONTROL OF IMAGING DEVICES. P.S. Soni

QUALITY CONTROL OF IMAGING DEVICES. P.S. Soni XA9847605 Chapter 8 P.S. Soni Quality Assurance and Quality Control in Nuclear Medicine Quality assurance in nuclear medicine refers collectively to all aspects of a nuclear medicine service. It would

More information

... In vivo imaging in Nuclear Medicine. 1957: Anger camera (X;Y) X Y

... In vivo imaging in Nuclear Medicine. 1957: Anger camera (X;Y) X Y József Varga, PhD EMISSION IMAGING BASICS OF QUANTIFICATION Imaging devices Aims of image processing Reconstruction University of Debrecen Department of Nuclear Medicine. In vivo imaging in Nuclear Medicine

More information

CHAPTER 8 GENERIC PERFORMANCE MEASURES

CHAPTER 8 GENERIC PERFORMANCE MEASURES GENERIC PERFORMANCE MEASURES M.E. DAUBE-WITHERSPOON Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, United States of America 8.1. INTRINSIC AND EXTRINSIC MEASURES 8.1.1.

More information

Appendix I: Artefacts and Trouble-shooting

Appendix I: Artefacts and Trouble-shooting Appendix I: Artefacts and Trouble-shooting Slide set of 101 slides based on the chapter authored by E. Busemann Sokole, N.J. Forwood of the publication (ISBN 978 92 0 143810 2): Nuclear Medicine Physics:

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

ACR Update in Nuclear Medicine Accreditation

ACR Update in Nuclear Medicine Accreditation Disclaimer ACR Update in Nuclear Medicine Accreditation Beth A. Harkness, MS, DABR, FACR Henry Ford Health System Detroit, MI ACR physics subcommittee for nuclear medicine accreditation. My facility is

More information

Quality control in dual head γ- cameras: Comparison between methods and softwares used for image analysis

Quality control in dual head γ- cameras: Comparison between methods and softwares used for image analysis Quality control in dual head γ- cameras: Comparison between methods and softwares used for image analysis Abdalrhman Nayl Edam 1, *, Maria Rosa Fornasier 2, Mario de Denaro 2 Abdelmoneim Sulieman 3, Mohammed

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information Ordering Information Please contact us if you have any questions or if you would like a quote or delivery schedule regarding the Catphan phantom. phone 800-525-1190, or 518-692-1190 fax 518-692-3329 mail

More information

LaBr 3 :Ce, the latest crystal for nuclear medicine

LaBr 3 :Ce, the latest crystal for nuclear medicine 10th Topical Seminar on Innovative Particle and Radiation Detectors 1-5 October 2006 Siena, Italy LaBr 3 :Ce, the latest crystal for nuclear medicine Roberto Pani On behalf of SCINTIRAD Collaboration INFN

More information

CHAPTER 15 DEVICES FOR EVALUATING IMAGING SYSTEMS

CHAPTER 15 DEVICES FOR EVALUATING IMAGING SYSTEMS DEVICES FOR EVALUATING IMAGING SYSTEMS O. DEMIRKAYA, R. AL-MAZROU Department of Biomedical Physics, King Faisal Specialist Hospital and Research Centre, Riyadh, Saudi Arabia 15.1. DEVELOPING A QUALITY

More information

The future of nuclear imaging is clear

The future of nuclear imaging is clear Cardius X-ACT The future of nuclear imaging is clear Increased regulations, growing competition, and concerns about radiation exposure are just a sampling of the current challenges facing the nuclear medicine

More information

Reconstruction Filtering in Industrial gamma-ray CT Application

Reconstruction Filtering in Industrial gamma-ray CT Application Reconstruction Filtering in Industrial gamma-ray CT Application Lakshminarayana Yenumula *, Rajesh V Acharya, Umesh Kumar, and Ashutosh Dash Industrial Tomography and Instrumentation Section, Isotope Production

More information

International Journal of Scientific & Engineering Research, Volume 4, Issue 9, September ISSN

International Journal of Scientific & Engineering Research, Volume 4, Issue 9, September ISSN International Journal of Scientific & Engineering Research, Volume 4, Issue 9, September-013 06 Evaluating the effect of acquisition parameters on image quality and acquisition time with SPECT using collimator

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

PHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002

PHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002 PHYSICS 334 - ADVANCED LABORATORY I COMPTON SCATTERING Spring 00 Purposes: Demonstrate the phenomena associated with Compton scattering and the Klein-Nishina formula. Determine the mass of the electron.

More information

First Applications of the YAPPET Small Animal Scanner

First Applications of the YAPPET Small Animal Scanner First Applications of the YAPPET Small Animal Scanner Guido Zavattini Università di Ferrara CALOR2 Congress, Annecy - FRANCE YAP-PET scanner Scintillator: YAP:Ce Size: matrix of 2x2 match like crystals

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 05-582 PRIMA 7 Transport Index Radiation Monitor Operators Manual March 2005 Manual No. 135011 Rev. 3 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product

More information

Master of Science Thesis. SIMIND Based Pinhole Imaging

Master of Science Thesis. SIMIND Based Pinhole Imaging Master of Science Thesis SIMIND Based Pinhole Imaging * Development and Validation Kurt Sundin Supervisor: Michael Ljungberg, PhD Medical Radiation Physics Clinical Sciences, Lund Lund University, 2006

More information

2010 Philips BrightView XCT SPECT/CT

2010 Philips BrightView XCT SPECT/CT 2010 Philips BrightView XCT SPECT/CT Unit was purchased from Philips training center in 2015. Installed but never been used by the current facility. (Scroll for pictures) BrightView XCT Camera with PinPoint

More information

Nuclear Associates , &

Nuclear Associates , & Nuclear Associates 76-823, 76-824 & 76-825 PET/SPECT Phantom Source Tank, Phantom Inserts and Cardiac Insert Users Manual March 2005 Manual No. 76-823-1 Rev. 2 2004, 2005 Fluke Corporation, All rights

More information

DESIGN AND OPERATION OF A WIDE RANGE SEGMENTED GAMMA RAY SCANNING ASSAY INSTRUMENT FOR THE MEASUREMENT OF BOTH LOW AND INTERMEDIATE LEVEL WASTE

DESIGN AND OPERATION OF A WIDE RANGE SEGMENTED GAMMA RAY SCANNING ASSAY INSTRUMENT FOR THE MEASUREMENT OF BOTH LOW AND INTERMEDIATE LEVEL WASTE 11-A-424-INMM DESIGN AND OPERATION OF A WIDE RANGE SEGMENTED GAMMA RAY SCANNING ASSAY INSTRUMENT FOR THE MEASUREMENT OF BOTH LOW AND INTERMEDIATE LEVEL WASTE John A. Mason, Marc R. Looman, Robert A. Price

More information

SPECT Reconstruction & Filtering

SPECT Reconstruction & Filtering SPECT Reconstruction & Filtering Goals Understand the basics of SPECT Reconstruction Filtered Backprojection Iterative Reconstruction Make informed choices on filter selection and settings Pre vs. Post

More information

Application Note. ipix A Gamma imager to support various applications. Introduction. An easy to carry and deploy instrument

Application Note. ipix A Gamma imager to support various applications. Introduction. An easy to carry and deploy instrument Application Note ipix A Gamma imager to support various applications Introduction ipix is a unique gamma imager that quickly locates low level radioactive sources from a distance and estimates the dose

More information

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard.

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard. Chiara Secco PET Performance measurements of the new LSO-Based Whole Body PET/CT Scanner biograph 16 HI-REZ using the NEMA NU 2-2001 Standard. INTRODUCTION Since its introduction, CT has become a fundamental

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-647 R/F QC Phantom Operators Manual March 2005 Manual No. 07-647-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. All product names are trademarks of their respective companies

More information

PET Detectors. William W. Moses Lawrence Berkeley National Laboratory March 26, 2002

PET Detectors. William W. Moses Lawrence Berkeley National Laboratory March 26, 2002 PET Detectors William W. Moses Lawrence Berkeley National Laboratory March 26, 2002 Step 1: Inject Patient with Radioactive Drug Drug is labeled with positron (β + ) emitting radionuclide. Drug localizes

More information

Industry Breakthrough

Industry Breakthrough Industry Breakthrough Dynamic SPECT Acquisition Quantifying Myocardial Blood Flow D-S P EC T Cardiac Imaging System Nuclear Cardiology in the 21st Century In the 21st century, most nuclear cameras are

More information

LaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging

LaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging 8th International Workshop on Radiation Imaging Detectors Pisa 2-6 July 2006 LaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging Roberto Pani On behalf of SCINTIRAD Collaboration

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

Primer on molecular imaging technology

Primer on molecular imaging technology Primer on molecular imaging technology Craig S. Levin Division of Nuclear Medicine, Department of Radiology and Molecular Imaging Program at Stanford (MIPS), Stanford University School of Medicine, 300

More information

Positron Emission Tomography

Positron Emission Tomography Positron Emission Tomography UBC Physics & Astronomy / PHYS 409 1 Introduction Positron emission tomography (PET) is a non-invasive way to produce the functional 1 image of a patient. It works by injecting

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

Photomultiplier Tube

Photomultiplier Tube Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or

More information

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC? Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM

More information

Testing and Performance Validation of a Shielded Waste Segregation and Clearance Monitor Designed for the Measurement of Low Level Waste-13043

Testing and Performance Validation of a Shielded Waste Segregation and Clearance Monitor Designed for the Measurement of Low Level Waste-13043 Testing and Performance Validation of a Shielded Waste Segregation and Clearance Monitor Designed for the Measurement of Low Level Waste-13043 John A. Mason*, Graham Beaven**, Kevin J. Burke*, Robert Spence**

More information

Pinhole collimator design for nuclear survey system

Pinhole collimator design for nuclear survey system Annals of Nuclear Energy 29 (2002) 2029 2040 www.elsevier.com/locate/anucene Pinhole collimator design for nuclear survey system Wanno Lee*, Gyuseong Cho Department of Nuclear Engineering, Korea Advanced

More information

NON-UNIFORM ATTENUATION CORRECTION USING SIMULTANEOUS TRANSMISSION AND EMISSION CONVERGING TOMOGRAPHY

NON-UNIFORM ATTENUATION CORRECTION USING SIMULTANEOUS TRANSMISSION AND EMISSION CONVERGING TOMOGRAPHY 1134 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 39, NO. 4,1992 NON-UNIFORM ATTENUATION CORRECTION USING SIMULTANEOUS TRANSMISSION AND EMISSION CONVERGING TOMOGRAPHY C-H Tung, G. T. Gullberg, G. L. Zeng,

More information

Nuclear Associates , , CT Head and Body Dose Phantom

Nuclear Associates , , CT Head and Body Dose Phantom Nuclear Associates 76-414,76-414-4150,76-415 CT Head and Body Dose Phantom Users Manual March 2005 Manual No. 76-414-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product

More information

Industry Breakthrough

Industry Breakthrough Industry Breakthrough Dynamic SPECT Acquisition Quantifying Myocardial Blood Flow Nuclear Cardiology in the 21st Century In the 21st century, most nuclear cameras are still relying on a technology invented

More information

CZT Technology: Fundamentals and Applications

CZT Technology: Fundamentals and Applications GE Healthcare CZT Technology: Fundamentals and Applications White Paper Abstract Nuclear Medicine traces its technology roots to the 1950 s, and while it has continued to evolve since the invention of

More information

IAEA Coordinated Research Project on Development of Harmonized QA/QC Procedures for Maintenance and Repair of Nuclear Instruments

IAEA Coordinated Research Project on Development of Harmonized QA/QC Procedures for Maintenance and Repair of Nuclear Instruments PROCEDURE: TEST PROCEDURE FOR GEIGER-MUELLER RADIATION DETECTORS Nº: MRNI-501 DECEMBER 2008 PAGE: 1 OF: 17 IAEA Coordinated Research Project on Development of Harmonized QA/QC Procedures for Maintenance

More information

- Nuclear Radiation Sensor - Detects Beta and Gamma Radiation and X-Rays

- Nuclear Radiation Sensor - Detects Beta and Gamma Radiation and X-Rays - Nuclear Radiation Sensor - Detects Beta and Gamma Radiation and X-Rays Description The function of the RD2014 radiation sensor is based on an array of customized PIN diodes. The integrated pulse discriminator

More information

3/8 or 1 Free-Geometry Dual-Detector Cameras

3/8 or 1 Free-Geometry Dual-Detector Cameras Page 1 System Description Infinia is a family of premium, highperformance all-purpose dual detector imaging systems. The Infinia cameraline s foundation is the new advanced, all-digital Elite detector

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

STANDARD OPERATING PROCEDURES SOP:: 2057 PAGE: 1 of 6 REV: 0.0 DATE: 07/11/03

STANDARD OPERATING PROCEDURES SOP:: 2057 PAGE: 1 of 6 REV: 0.0 DATE: 07/11/03 PAGE: 1 of 6 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc Quality Control for Stereotactic Breast Biopsy Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc. 716-924-0350 Methods of Imaging Guided Breast Biopsy Ultrasound guided, hand-held needle

More information

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering RR Oct 2001 SS Dec 2001 MJ Oct 2009 Physics 34000 Laboratory Scattering of Photons from Electrons: Compton Scattering Objective: To measure the energy of high energy photons scattered from electrons in

More information

Fundamentals of Positron Emission Tomography (PET)

Fundamentals of Positron Emission Tomography (PET) Fundamentals of Positron Emission Tomography (PET) NPRE 435, Principles of Imaging with Ionizing Radiation, Fall 2017 Content Fundamentals of PET Camera & Detector Design Real World Considerations Performance

More information

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors Nuclear Physics #1 Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors Introduction: In this experiment you will use both scintillation and semiconductor detectors to study γ- ray energy spectra. The

More information

LUDLUM MODEL MODEL AND MODEL GAMMA SCINTILLATORS. June 2017

LUDLUM MODEL MODEL AND MODEL GAMMA SCINTILLATORS. June 2017 LUDLUM MODEL 44-20 MODEL 44-20-1 AND MODEL 44-20-3 GAMMA SCINTILLATORS June 2017 LUDLUM MODEL 44-20 MODEL 44-20-1 AND MODEL 44-20-3 GAMMA SCINTILLATORS June 2017 STATEMENT OF WARRANTY Ludlum Measurements,

More information

Changing the Shape of Nuclear Medicine

Changing the Shape of Nuclear Medicine TRUTH IN IMAGING Changing the Shape of Nuclear Medicine Multi-Purpose SPECT Scanner Nothing Gets Closer Introducing 360 Body Contour Scanning With 360 degree detector coverage, and unique proximity sensors

More information

ipix Gamma Imager Product Introduction Steve Laskos Product Management Director

ipix Gamma Imager Product Introduction Steve Laskos Product Management Director ipix Gamma Imager Product Introduction Steve Laskos Product Management Director ipix: The Next Generation Gamma Imaging System ipix the new generation of gamma camera Simple, easy to use for experts and

More information

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation AAPM Annual Clinical Meeting Indianapolis, IN August 5, 2013 Learning Objectives Become familiar with the recommendations and

More information

Nathan Childress, Ph.D., DABR

Nathan Childress, Ph.D., DABR Nathan Childress, Ph.D., DABR Introduction TG-142 is a comprehensive QA protocol Covers nearly every aspect of machine and safety QA Recommends quantitative results Recommends high testing frequencies

More information

TABLE OF CONTENTS. References

TABLE OF CONTENTS. References ANALYTICAL X-RAY EQUIPMENT USE POLICIES & PROCEDURES Page 1 of 9 Revised: 11/24/2003 TABLE OF CONTENTS 1.0 General 2.0 Purpose 3.0 Scope and Authority 4.0 Equipment Requirements 4.0.1 Safety device 4.0.2

More information

Gamma-ray spectral imaging using a single-shutter radiation camera

Gamma-ray spectral imaging using a single-shutter radiation camera Nuclear Instruments and Methods in Physics Research A299 (1990) 495-500 North-Holland 495 Gamma-ray spectral imaging using a single-shutter radiation camera T.A. DeVol, D.K. Wehe and G.F. Knoll The University

More information

New Technology in Nuclear Medicine

New Technology in Nuclear Medicine New Technology in Nuclear Medicine Reed G. Selwyn, PhD, DABR Vice Chair of Research & Imaging Sciences Associate Professor and Chief, Medical Physics Dept. of Radiology, University of New Mexico Objectives

More information

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY 12 th A-PCNDT 2006 Asia-Pacific Conference on NDT, 5 th 10 th Nov 2006, Auckland, New Zealand PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY Rajashekar

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter Revision Date: January 5th, 2017 www.disc-imaging.com Table of Contents Section A: Preliminary Setup Requirements... 4 Tools

More information

ISO-CART-85. Mobile Low-Level Waste Assay System

ISO-CART-85. Mobile Low-Level Waste Assay System Mobile Low-Level Waste Assay System Complete In-Situ NDA Gamma-Ray Analysis Solutions for a Wide Variety of Samples, including Free-Release Decommissioning Waste. ISO-CART -85 : A Complete Turnkey Solution

More information

16 Instrumentation and Data Acquisition

16 Instrumentation and Data Acquisition Instrumentation and Data Acquisition 275 16 Instrumentation and Data Acquisition Sibylle I. Ziegler and Magnus Dahlbom CONTENTS 16.1 Detectors and Imaging Systems 275 16.1.1 Principles of Scintillation

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

Exprerimental Evaluation of a Dedicated Pinhole SPECT System for Small Animal Imaging and Scintimammography

Exprerimental Evaluation of a Dedicated Pinhole SPECT System for Small Animal Imaging and Scintimammography ETASR - Engineering, Technology & Applied Science Research Vol. 1, o. 1, 211, 17-22 17 Exprerimental Evaluation of a Dedicated Pinhole SPECT System for Small Animal Imaging and Scintimammography S. David

More information

Test Equipment for Radiology and CT Quality Control Contents

Test Equipment for Radiology and CT Quality Control Contents Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3

More information

Technical Description and Operator s Manual. Survey Meter SM 5 D. Sensortechnik und Elektronik Pockau GmbH. Siedlungsstraße 5-7 D Pockau

Technical Description and Operator s Manual. Survey Meter SM 5 D. Sensortechnik und Elektronik Pockau GmbH. Siedlungsstraße 5-7 D Pockau Technical Description and Operator s Manual Survey Meter SM 5 D Sensortechnik und Elektronik Pockau GmbH Siedlungsstraße 5-7 D 09509 Pockau Content: Descriptions 1 Safety Instructions 2 Design and control

More information

Victoreen , H, A Gamma Scintillation Detector

Victoreen , H, A Gamma Scintillation Detector Victoreen 943-35, 943-36 943-36H, 943-37 943-237A Gamma Scintillation Detector Operators Manual March 2005 Manual No. 943-35-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A.

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

FRCR Nuclear Medicine

FRCR Nuclear Medicine FRCR Nuclear Medicine FRCR LECTURES Lecture I 20/09/2016: Nuclear Medicine and Image Formation Lecture II 22/09/2016: Positron Emission Tomography & QA Lecture III 27/09/2016: Radiation Detectors - Radiation

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information

236 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 1, FEBRUARY 2012

236 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 1, FEBRUARY 2012 236 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 1, FEBRUARY 2012 Characterization of the H3D ASIC Readout System and 6.0 cm 3-D Position Sensitive CdZnTe Detectors Feng Zhang, Cedric Herman, Zhong

More information

As the role of gamma cameras expands in positron

As the role of gamma cameras expands in positron BASIC SCIENCE INVESTIGATIONS Feasibility of a High-Speed Gamma-Camera Design Using the High-Yield-Pileup-Event- Recovery Method Wai-Hoi Wong, Hongdi Li, Jorge Uribe, Hossain Baghaei, Yu Wang, and Shigeru

More information

Ensuring Shielding adequacy in Lead shielded spent fuel transportation casks using gamma scanning

Ensuring Shielding adequacy in Lead shielded spent fuel transportation casks using gamma scanning Ensuring Shielding adequacy in Lead shielded spent fuel transportation casks using gamma scanning More info about this article: http://www.ndt.net/?id=21208 M.Ravichandra 1, P.Raghavendra 1, Dhiren Kothari

More information

Development of the LBNL Positron Emission Mammography Camera

Development of the LBNL Positron Emission Mammography Camera Development of the LBNL Positron Emission Mammography Camera J.S. Huber, Member, IEEE, W.S. Choong, Member, IEEE, J. Wang, Member, IEEE, J.S. Maltz, Member, IEEE, J. Qi, Member, IEEE, E. Mandelli, Member,

More information

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Melissa Martin, M.S., FAAPM, FACR, FACMP AAPM Annual Meeting - Philadelphia July 19, 2010 MO-B-204C-1 Educational Objectives

More information

Cartogam Real-Time Portable Gamma-Ray Imaging System

Cartogam Real-Time Portable Gamma-Ray Imaging System Features Complete tool for in situ cartography, saving time, cost and dose Real-time acquisition and immediate display Two dimensional gamma mapping and dose range Dose rate estimation of hot spots Excellent

More information

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc.

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. An Introduction to TG-142 Imaging QA Using Standard Imaging Products Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. Goals Understand the nature and intent of TG 142 imaging

More information

T h e P h a n t o m L a b o r a t o r y

T h e P h a n t o m L a b o r a t o r y T h e P h a n t o m L a b o r a t o r y 1 CCT228 ATCM Phantom Manual Copyright 2017 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order

More information

Chemical Engineering 412

Chemical Engineering 412 Chemical Engineering 412 Introductory Nuclear Engineering Lecture 25 Radiation Detection & Measurement Spiritual Thought 2 I realize that there are some, perhaps many, [who] feel overwhelmed by the lack

More information