An Activity in Computed Tomography

Size: px
Start display at page:

Download "An Activity in Computed Tomography"

Transcription

1 Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum ( nm and nm respectively). As X-rays pass through a person some of the X-rays will be absorbed, meaning fewer X- rays leave the body than entered it. This is called attenuation and is larger (fewer exiting X-rays) for denser material. Bone, for example, attenuates X-rays more than muscle since bone is denser. Images produced with X-rays are like shadows, showing where the X-ray radiation passed through the body and where it was absorbed. Dark regions on an X-ray image correspond to tissue with smaller attenuation coefficients and lighter regions with denser tissue like bone (see Fig. 1.1). The main drawback to X-ray imaging is that intense or prolonged exposure leads to serious health problems such as cancer and radiation burns. High energy X-rays can change the molecular structure of tissue, sometimes causing a person s DNA to mutate faster than normal. Because of this drawback, the goal of all forms of X- ray imaging is to acquire the best possible image while submitting the patient to the lowest possible dose of radiation. Computed Tomography One of the limitations of planar images like chest X-rays are that they are a two dimensional representation of a three dimensional object. This limits the amount of information that can be discerned. For example, an X-ray of a box, in which a cone and sphere are placed, may look something like Fig. 1.2 (a). From this we may be able to say what kind of objects they are but not their relative locations. For example, is the sphere in front of the cone or is the cone in front of the sphere? But if we rotate the X-ray 90 o relative to the box we may get an image like Fig. 1.2 (b). This is the basic idea behind Computed Tomography where multiple X-rays scans from different angles are put together to form an image. Of course this also means that the radiation dose for a CT scan is larger than for a regular single X-ray scan. (a) (b) Fig. 1.1 Cranial CT scan. The white region is the skull and the darker regions are soft tissue. Source: Computed Tomography: From Photon Statistics to Modern Cone-Beam CT Fig. 1.2 Projections of a sphere and a cone. (a) a single image imparts a limited amount of information (b) another image at a different angle can reveal additional information Instructor Note: Question for students: How does an X-ray film look before it is used? Answer: It s all white! Just like the parts that included bone during an exposure. Exposure to X- rays will darken the film. The cone/sphere example could be done as a demonstration during the lecture. Place some kind of screen in front of the objects so the students can t see their positions. Then place a light source in front of the objects to produce Fig. 1.2(a). Ask the students what they can say about the objects. Then ask if there is any way to find out which object is in front. X-ray/CT Topics Covered X-rays are 2D representations of 3D area CT reveals information a single X-ray does not Exposure to X-rays can be dangerous

2 Back Projection Back projection is one of the most common methods for reconstructing an image. The images in Fig. 1.2 are called projections. In a CT scanner, an X-ray source rotates around an object producing area projection data at each angle. That data is spread back onto an image of the area, which is divided into pixels. Pixels are the individual units of the reconstruction. Each pixel has a certain value attached to it which corresponds to the attenuation at that point (Fig. 1.3). When put together, all the pixels reveal an image. As more and more X-ray scans are taken and data is back projected, an image of the area starts to emerge similar to what is shown in Fig It is important to note that because projection data are collected from multiple angles, CT imaging results in a higher dose of radiation than planar imaging. Today patients receiving a CT scan are exposed to the equivalent of 30 to 442 chest x-rays per scan The near IR-light used in a photogate has a larger wavelength (880nm) and lower energy than X- rays. Infrared light is quickly attenuated by water inside the human body and is not suitable for medical imaging of deep tissue. However, it is able to pass through material with low attenuation coefficients like the light filter used in this experiment. The projection data recorded by the photogate will be a digital signal of 0 (unblocked) and 1(blocked), rather than the continuous signal of an X-ray based CT scanner (Fig. 1.3). Photogate scans Back Projection Photogate scans Fig. 1.3 Projection data is smeared back onto a reconstruction of the scanned area. Pixels of higher values correspond to the position of the object. The projection data are the 0s and 1s to the right and bottom of the first grid and represent the unblocked and blocked states of the photogate respectively. Instructor Note: As a demonstration of back projection shine two light sources on a cylinder at a right angle. Explain that just as the two shadows converge at the cylinder s position, back projection data converge at the location of scanned objects. This also emphasizes the correlation between X-rays and shadows. Back Projection Topics Covered Projections Back Projection is spreading the projection onto the grid Back projection grid divided into pixels High values in the reconstruction correspond to the position of the object 2

3 Geometry Once the Photogate is blocked and we have projection data to work with, the computer spreads that data back onto the reconstruction graph. To do this we need to mathematically describe each scan. This requires us to find an equation for the imaginary line from the light source of the photogate to the detector, relative to the scanned area. The only data we have as inputs are (1) the distance between the rotational axis and the origin Focus-Center-Distance (FCD) (see Fig. 2.2), (2) the angle between the rotational axis and the negative x-axis,θ, and (3) the angle between the FCD and the scanning line,φ, (Fig. 1.4). Rotational Axis y φ IR Source Scanning line Focus-Center-Distance (FCD) IR Detector θ x Fig. 1.4 Geometry of the CT scanner, the Scanning line represents the line connecting the light source and detector of the photogate We can use geometry and trigonometry to find the equation of the scanning line. All we need are some trig identities (SOH CAH TOA) and the formula for a line (y=mx+b). There are multiple formulas we use to describe specific scanner geometries. Which formula we choose for the scanning line changes depending on the values ofθ, φ and the FCD. Each possibility is programmed into LabVIEW, which selects the appropriate formula based on the inputs. 3

4 Computed Tomography Experiment Guide Introduction In this activity students will be given a mystery box, in which 1 to 3 cylinders were placed by the instructor before class. This enclosure should be sealed so that students don t know the locations of the cylinders. The students will use a photogate, rotary motion sensor and a LabVIEW program to make an image of the contents of the enclosure. Using that image, they will determine the position and dimensions of the cylinders inside. Objectives After completing this lab, students should be able to: Describe the basics of how Computed Tomography makes an image through back projection. Demonstrate a practical application of SOH CAH TOA and the line equation y=mx+b. Understand terms important to CT such as image artifacts and windowing Find the location and diameters of objects inside an opaque mystery box Equipment Vernier photogate (with rod), rotary motion sensor (with pulley) and LabPro Clamp to attach the photogate to the rotary sensor Rotating platform and base with angular scale Angular indicator Rotary sensor stand Congo Blue #181 light filter enclosure Cylinders of various diameters (~0.5 2 cm) Fig 2.1 Apparatus setup 4

5 LabPro/Computer Setup 1) Make sure the LabPro is connected to a USB port 2) Plug the photogate into the Dig/Sonic 1 jack and the rotary motion sensor into the Dig/Sonic 2 jack. Apparatus Setup 1) Assemble the apparatus as shown in Fig Make sure that the rotation sensor stand and the rotation table stand are aligned and that the angular indicator points straight down the x-axis. 2) Lower the rotary sensor so that the degree indicator is just resting on the rotating platform. This will help stabilize platform. (Using a heavy rotary sensor stand reduces the chances of moving the rotary sensor during the experiment.) 3) Adjust the apparatus as needed so the photogate is free to pass over the mystery box. 4) Measure the Focus-Center-Distance (the distance in cm between the shaft of the rotation sensor to the center of the platform). Record it in the space below. FCD = (cm) FCD Fig. 2.2 The Focus-Center-Distance (FCD) is the distance between the shaft of the rotary sensor and the center of the rotating platform. 5

6 Apparatus Calibration Open the program Computed Tomography and click on Manual Scanner. Follow the on-screen instructions. It is important that the two stands are aligned and that the angular indicator points straight down the x- axis. If they are not correctly positioned, the resulting image will be distorted. To verify that the apparatus is positioned correctly, perform the following test. 1) Place a cylindrical object at the center of the grid on the platform 2) Enter the FCD you recorded in the Apparatus Setup section. 3) Making sure the photogate is not blocked, click on BEGIN SCAN. Slowly move the photogate over the object placed at the origin. The image in the reconstruction graph should be symmetric about the x-axis (Fig. 2.3). If the image is not symmetric: a. Realign the rotary sensor stand and the rotating platform b. Re-zero the rotating sensor c. Repeat this procedure until the image is symmetric While performing this test, take note of the scanning speed needed to produce a good image. Does a slower or faster scan rate produce better data? (a) Bad Calibration (b) Good Calibration Fig Two scans of the same cylinder positioned at the origin. (a) The image is not symmetric about the x-axis. Make sure the rotating platform base and the base holding the rotary sensor are aligned and re-zero the rotary sensor. (b) The two bases are properly aligned and the rotary sensor was correctly zeroed; therefore the scan of the cylinder is symmetric about the x-axis. 6

7 Data Collection Using a Cylinder at the Origin 1) With the cylinder still at the origin, do multiple scans at differing speeds. Which speed do you think will result in a higher quality scan? There is a certain amount of computer processing time needed to read the data and spread its back projection onto the graph. Moving the photogate too quickly can result in image artifacts. Could slower scan times be dangerous in medical CT? Why? Slower scan times would mean longer exposure to radiation, which means a greater possibility of injury. 2) Change the pixel count by incrementing the Pixel/cm control and clicking SET PIXEL COUNT and scan the cylinder (increment Pixel/cm by at least 10 initially). How do image quality and scan time change with an increased pixel count? Why? An increased pixel count generally results in a higher quality image, but a longer scan time, because more computations are required to create an image. The increased scan time could also result in more image artifacts.lowering the pixel count is the simplest way to handle computers with slower processors or data acquisition units. Using the Mystery Box 1) Place the enclosure on the grid, align the photogate with the x-axis and rotate the platform 360 o to make sure that the enclosure and the photogate won t touch during the scan. In order to ensure the grid inside the enclosure and the grid on the platform align, the instructor may wish to draw marks on the enclosure at the axes. The students can then align the marks with the axes on the platform. 2) With the photogate at the starting position, click BEGIN SCAN. Slowly move the photogate over the enclosure. A single pass should take 15 to 20 seconds. 3) Once the photogate has made a complete pass, rotate the platform, enter the new θ and scan the enclosure again. Repeat for desired number of θ angles. For your first scan try scanning in 20 o increments. An image artifact is a false representation of the actual scanned area. Does your final image show any artifacts? What caused the artifacts? In addition to the actual objects, the reconstructed image shows extra data that resulted from back projecting data. Artifacts may have also been caused by improper scanning techniques. 7

8 4) Adjust the Window slider to highlight important data. Describe why the image is changing. The Window slider sets the low value of the color table. Pixels with values smaller than that of the slider will be black, because black is the lower boundary on the color table. Describe how changing the Window improved the image. Changing the Window slider removes some of the artifacts caused by the reconstruction algorithm. In medical CT scanners, windowing can also be used to improve the contrast between similar media. 5) Right click on the cursor toolbar and click on Create Cursor>Free. Using the cursor(s) determine the position and diameter of each object. To display a cursor, right click on the cursor area below the graph and go to Create Cursor>Free. Click and drag the intersection of the two lines of the cursor to the center of the object. The corresponding grid markers on the left and bottom of the scan area indicate the location. Do the objects in your scan look circular? If not, what could be done to improve your image? Possible methods to improve image quality: Decrease the increment between θ (every 10 o instead of every 20 o ) Increase the pixel count Scan the enclosure more slowly 8

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

PHOTOGATE 0662I WITH PULLEY ATTACHMENT User s Guide

PHOTOGATE 0662I WITH PULLEY ATTACHMENT User s Guide PHOTOGATE 0662I WITH PULLEY ATTACHMENT User s Guide Figure 1. The Photogate with Pulley Attachment CENTRE FOR MICROCOMPUTER APPLICATIONS http://www.cma-science.nl Description The Photogate 0662i is a traditional

More information

Physics 4C Chabot College Scott Hildreth

Physics 4C Chabot College Scott Hildreth Physics 4C Chabot College Scott Hildreth The Inverse Square Law for Light Intensity vs. Distance Using Microwaves Experiment Goals: Experimentally test the inverse square law for light using Microwaves.

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information Ordering Information Please contact us if you have any questions or if you would like a quote or delivery schedule regarding the Catphan phantom. phone 800-525-1190, or 518-692-1190 fax 518-692-3329 mail

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition

More information

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT) Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in

More information

Single Slit Diffraction

Single Slit Diffraction PC1142 Physics II Single Slit Diffraction 1 Objectives Investigate the single-slit diffraction pattern produced by monochromatic laser light. Determine the wavelength of the laser light from measurements

More information

AQA P3 Topic 1. Medical applications of Physics

AQA P3 Topic 1. Medical applications of Physics AQA P3 Topic 1 Medical applications of Physics X rays X-ray properties X-rays are part of the electromagnetic spectrum. X-rays have a wavelength of the same order of magnitude as the diameter of an atom.

More information

AgilEye Manual Version 2.0 February 28, 2007

AgilEye Manual Version 2.0 February 28, 2007 AgilEye Manual Version 2.0 February 28, 2007 1717 Louisiana NE Suite 202 Albuquerque, NM 87110 (505) 268-4742 support@agiloptics.com 2 (505) 268-4742 v. 2.0 February 07, 2007 3 Introduction AgilEye Wavefront

More information

Single-Slit Diffraction. = m, (Eq. 1)

Single-Slit Diffraction. = m, (Eq. 1) Single-Slit Diffraction Experimental Objectives To observe the interference pattern formed by monochromatic light passing through a single slit. Compare the diffraction patterns of a single-slit and a

More information

Distortion Correction in LODOX StatScan X-Ray Images

Distortion Correction in LODOX StatScan X-Ray Images Distortion Correction in LODOX StatScan X-Ray Images Matthew Paul Beets November 27, 2007 X-Ray images produced by the LODOX StatScan machine contain a non-linear distortion in the direction of the beam

More information

Fig. 1

Fig. 1 PhysicsAndMathsTutor.com 1 1. Fig. 1 shows data for the intensity of a parallel beam of X-rays after penetration through varying thicknesses of a material. intensity / MW m 2 thickness / mm 0.91 0.40 0.69

More information

PD233: Design of Biomedical Devices and Systems

PD233: Design of Biomedical Devices and Systems PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/

More information

Laboratory 1: Motion in One Dimension

Laboratory 1: Motion in One Dimension Phys 131L Spring 2018 Laboratory 1: Motion in One Dimension Classical physics describes the motion of objects with the fundamental goal of tracking the position of an object as time passes. The simplest

More information

Name: Date: Block: Light Unit Study Guide Matching Match the correct definition to each term. 1. Waves

Name: Date: Block: Light Unit Study Guide Matching Match the correct definition to each term. 1. Waves Name: Date: Block: Light Unit Study Guide Matching Match the correct definition to each term. 1. Waves 2. Medium 3. Mechanical waves 4. Longitudinal waves 5. Transverse waves 6. Frequency 7. Reflection

More information

X-RAY COMPUTED TOMOGRAPHY

X-RAY COMPUTED TOMOGRAPHY X-RAY COMPUTED TOMOGRAPHY Bc. Jan Kratochvíla Czech Technical University in Prague Faculty of Nuclear Sciences and Physical Engineering Abstract Computed tomography is a powerful tool for imaging the inner

More information

National 3 Physics Waves and Radiation. 1. Wave Properties

National 3 Physics Waves and Radiation. 1. Wave Properties 1. Wave Properties What is a wave? Waves are a way of transporting energy from one place to another. They do this through some form of vibration. We see waves all the time, for example, ripples on a pond

More information

X rays X-ray properties Denser material = more absorption = looks lighter on the x-ray photo X-rays CT Scans circle cross-sectional images Tumours

X rays X-ray properties Denser material = more absorption = looks lighter on the x-ray photo X-rays CT Scans circle cross-sectional images Tumours X rays X-ray properties X-rays are part of the electromagnetic spectrum. X-rays have a wavelength of the same order of magnitude as the diameter of an atom. X-rays are ionising. Different materials absorb

More information

Explain what is meant by a photon and state one of its main properties [2]

Explain what is meant by a photon and state one of its main properties [2] 1 (a) A patient has an X-ray scan taken in hospital. The high-energy X-ray photons interact with the atoms inside the body of the patient. Explain what is meant by a photon and state one of its main properties....

More information

Computer Tools for Data Acquisition

Computer Tools for Data Acquisition Computer Tools for Data Acquisition Introduction to Capstone You will be using a computer to assist in taking and analyzing data throughout this course. The software, called Capstone, is made specifically

More information

Principles of CT scan

Principles of CT scan Related topics Detector calibration, saturation, CT acquisition, CT reconstruction Principle X-ray computed tomography consists of using X-rays that are converted to a digital signal by a detector and

More information

Kit for building your own THz Time-Domain Spectrometer

Kit for building your own THz Time-Domain Spectrometer Kit for building your own THz Time-Domain Spectrometer 16/06/2016 1 Table of contents 0. Parts for the THz Kit... 3 1. Delay line... 4 2. Pulse generator and lock-in detector... 5 3. THz antennas... 6

More information

Longitudinal No, Mechanical wave ~340 m/s (in air) 1,100 feet per second More elastic/denser medium = Greater speed of sound

Longitudinal No, Mechanical wave ~340 m/s (in air) 1,100 feet per second More elastic/denser medium = Greater speed of sound Type of wave Travel in Vacuum? Speed Speed vs. Medium Light Sound vs. Sound Longitudinal No, Mechanical wave ~340 m/s (in air) 1,100 feet per second More elastic/denser medium = Greater speed of sound

More information

Instruction manual for T3DS software. Tool for THz Time-Domain Spectroscopy. Release 4.0

Instruction manual for T3DS software. Tool for THz Time-Domain Spectroscopy. Release 4.0 Instruction manual for T3DS software Release 4.0 Table of contents 0. Setup... 3 1. Start-up... 5 2. Input parameters and delay line control... 6 3. Slow scan measurement... 8 4. Fast scan measurement...

More information

1 Diffraction of Microwaves

1 Diffraction of Microwaves 1 Diffraction of Microwaves 1.1 Purpose In this lab you will investigate the coherent scattering of electromagnetic waves from a periodic structure. The experiment is a direct analog of the Bragg diffraction

More information

Nuclear Associates , , CT Head and Body Dose Phantom

Nuclear Associates , , CT Head and Body Dose Phantom Nuclear Associates 76-414,76-414-4150,76-415 CT Head and Body Dose Phantom Users Manual March 2005 Manual No. 76-414-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product

More information

used to diagnose and treat medical conditions. State the precautions necessary when X ray machines and CT scanners are used.

used to diagnose and treat medical conditions. State the precautions necessary when X ray machines and CT scanners are used. Page 1 State the properties of X rays. Describe how X rays can be used to diagnose and treat medical conditions. State the precautions necessary when X ray machines and CT scanners are used. What is meant

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

Lab 12 Microwave Optics.

Lab 12 Microwave Optics. b Lab 12 Microwave Optics. CAUTION: The output power of the microwave transmitter is well below standard safety levels. Nevertheless, do not look directly into the microwave horn at close range when the

More information

COMPUTATIONAL IMAGING. Berthold K.P. Horn

COMPUTATIONAL IMAGING. Berthold K.P. Horn COMPUTATIONAL IMAGING Berthold K.P. Horn What is Computational Imaging? Computation inherent in image formation What is Computational Imaging? Computation inherent in image formation (1) Computing is getting

More information

Related topics Beam hardening, cupping effect, Beam hardening correction, metal artefacts, photon starvation

Related topics Beam hardening, cupping effect, Beam hardening correction, metal artefacts, photon starvation Beam hardening and metal artefacts TEP Related topics Beam hardening, cupping effect, Beam hardening correction, metal artefacts, photon starvation Principle X-ray sources produce a polychromatic spectrum

More information

LECTURE 1 The Radiographic Image

LECTURE 1 The Radiographic Image LECTURE 1 The Radiographic Image Prepared by:- KAMARUL AMIN ABDULLAH @ ABU BAKAR UiTM Faculty of Health Sciences Medical Imaging Department 11/23/2011 KAMARUL AMIN (C) 1 Lesson Objectives At the end of

More information

LECTURE 20 ELECTROMAGNETIC WAVES. Instructor: Kazumi Tolich

LECTURE 20 ELECTROMAGNETIC WAVES. Instructor: Kazumi Tolich LECTURE 20 ELECTROMAGNETIC WAVES Instructor: Kazumi Tolich Lecture 20 2 25.6 The photon model of electromagnetic waves 25.7 The electromagnetic spectrum Radio waves and microwaves Infrared, visible light,

More information

Visibility of Detail

Visibility of Detail Visibility of Detail Radiographic Quality Quality radiographic images represents the, and information is for diagnosis. The of the anatomic structures and the accuracy of their ( ) determine the overall

More information

Educational Spectrophotometer Accessory Kit and System OS-8537 and OS-8539

Educational Spectrophotometer Accessory Kit and System OS-8537 and OS-8539 GAIN 1 10 Instruction Manual with Experiment Guide and Teachers Notes 012-06575C *012-06575* Educational Spectrophotometer Accessory Kit and System OS-8537 and OS-8539 100 CI-6604A LIGHT SENSOR POLARIZER

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

CPSC 4040/6040 Computer Graphics Images. Joshua Levine

CPSC 4040/6040 Computer Graphics Images. Joshua Levine CPSC 4040/6040 Computer Graphics Images Joshua Levine levinej@clemson.edu Lecture 04 Displays and Optics Sept. 1, 2015 Slide Credits: Kenny A. Hunt Don House Torsten Möller Hanspeter Pfister Agenda Open

More information

Modern Physics Laboratory MP4 Photoelectric Effect

Modern Physics Laboratory MP4 Photoelectric Effect Purpose MP4 Photoelectric Effect In this experiment, you will investigate the photoelectric effect and determine Planck s constant and the work function. Equipment and components Photoelectric Effect Apparatus

More information

GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom

GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom Introduction: These EBT2/EBT3 films, shown in Figure 1a-c, are specially sized and formatted to uniquely fit the Accuray Ball Cube II Phantom. Each

More information

TSBB09 Image Sensors 2018-HT2. Image Formation Part 1

TSBB09 Image Sensors 2018-HT2. Image Formation Part 1 TSBB09 Image Sensors 2018-HT2 Image Formation Part 1 Basic physics Electromagnetic radiation consists of electromagnetic waves With energy That propagate through space The waves consist of transversal

More information

A Study of Slanted-Edge MTF Stability and Repeatability

A Study of Slanted-Edge MTF Stability and Repeatability A Study of Slanted-Edge MTF Stability and Repeatability Jackson K.M. Roland Imatest LLC, 2995 Wilderness Place Suite 103, Boulder, CO, USA ABSTRACT The slanted-edge method of measuring the spatial frequency

More information

Test procedures Page: 1 of 5

Test procedures Page: 1 of 5 Test procedures Page: 1 of 5 1 Scope This part of document establishes uniform requirements for measuring the numerical aperture of optical fibre, thereby assisting in the inspection of fibres and cables

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

ECEN 4606, UNDERGRADUATE OPTICS LAB

ECEN 4606, UNDERGRADUATE OPTICS LAB ECEN 4606, UNDERGRADUATE OPTICS LAB Lab 2: Imaging 1 the Telescope Original Version: Prof. McLeod SUMMARY: In this lab you will become familiar with the use of one or more lenses to create images of distant

More information

DIGITAL RADIOGRAPHY ARTIFACTS

DIGITAL RADIOGRAPHY ARTIFACTS IMAGING LAB MPHY 487 DIGITAL RADIOGRAPHY ARTIFACTS Mohammad Esmael Alsulimane B.Sc, M.Sc Medical Physics Lecturer - Physics Department All Rights Reserved: Some information and figures in this presentation

More information

CyberKnife Iris Beam QA using Fluence Divergence

CyberKnife Iris Beam QA using Fluence Divergence CyberKnife Iris Beam QA using Fluence Divergence Ronald Berg, Ph.D., Jesse McKay, M.S. and Brett Nelson, M.S. Erlanger Medical Center and Logos Systems, Scotts Valley, CA Introduction The CyberKnife radiosurgery

More information

LIGHT BOX & OPTICAL SET CAT NO. PH0615

LIGHT BOX & OPTICAL SET CAT NO. PH0615 LIGHT BOX & OPTICAL SET CAT NO. PH0615 Experiment Guide ACTIVITIES INCLUDED: Diffraction Angle of Reflection Using a Plane Mirror Refraction of Different Shaped Prisms Refraction (Snell's Law) Index of

More information

Electromagnetic Waves

Electromagnetic Waves Electromagnetic Waves What is an Electromagnetic Wave? An EM Wave is a disturbance that transfers energy through a field. A field is a area around an object where the object can apply a force on another

More information

X-rays in medical diagnostics

X-rays in medical diagnostics X-rays in medical diagnostics S.Dolanski Babić 2017/18. History W.C.Röntgen (1845-1923) discovered a new type of radiation Nature, Jan. 23. 1896.; Science, Feb.14. 1896. X- rays: Induced the ionization

More information

SYLLABUS. 1. Identification of Subject:

SYLLABUS. 1. Identification of Subject: SYLLABUS Date/ Revision : 30 January 2017/1 Faculty : Life Sciences Approval : Dean, Faculty of Life Sciences SUBJECT : Biophysics 1. Identification of Subject: Name of Subject : Biophysics Code of Subject

More information

IBEX MATERIALS DETECTION TECHNOLOGY

IBEX MATERIALS DETECTION TECHNOLOGY WHITE PAPER: IBEX MATERIALS DETECTION TECHNOLOGY IBEX Innovations Ltd. Registered in England and Wales: 07208355 Address: Discovery 2, NETPark, William Armstrong Way, Sedgefield, TS21 3FH, UK Patents held

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

BLACK BODY LIGHT SOURCE FOR THE OS-8539 EDUCATIONAL SPECTROPHOTOMETER

BLACK BODY LIGHT SOURCE FOR THE OS-8539 EDUCATIONAL SPECTROPHOTOMETER Includes Teacher's Notes and Typical Experiment Results Instruction Manual and Experiment Guide for the PASCO scientific Model OS-8542 012-07105B BLACK BODY LIGHT SOURCE FOR THE OS-8539 EDUCATIONAL SPECTROPHOTOMETER

More information

Introduction of a Single Chip TLD System for Patient Dosimetry

Introduction of a Single Chip TLD System for Patient Dosimetry Introduction of a Single Chip TLD System for Patient Dosimetry C. Hranitzky a, M. Halda a, G. Müller a, B. Obryk b, H. Stadtmann a* a Austrian Research Centers GmbH ARC, 2444 Seibersdorf, Austria. b Institute

More information

Technical Notes. Integrating Sphere Measurement Part II: Calibration. Introduction. Calibration

Technical Notes. Integrating Sphere Measurement Part II: Calibration. Introduction. Calibration Technical Notes Integrating Sphere Measurement Part II: Calibration This Technical Note is Part II in a three part series examining the proper maintenance and use of integrating sphere light measurement

More information

I look forward to seeing you on August 24!!

I look forward to seeing you on August 24!! AP Physics 1 Summer Assignment Packet Welcome to AP Physics 1! Your summer assignment is below. You are to complete the entire packet and bring it with you on the first day of school (Monday August 24,

More information

Experiment 1: Fraunhofer Diffraction of Light by a Single Slit

Experiment 1: Fraunhofer Diffraction of Light by a Single Slit Experiment 1: Fraunhofer Diffraction of Light by a Single Slit Purpose 1. To understand the theory of Fraunhofer diffraction of light at a single slit and at a circular aperture; 2. To learn how to measure

More information

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

Unit thickness. Unit area. σ = NΔX = ΔI / I 0 Unit thickness I 0 ΔI I σ = ΔI I 0 NΔX = ΔI / I 0 NΔX Unit area Δx Average probability of reaction with atom for the incident photons at unit area with the thickness of Delta-X Atom number at unit area

More information

Experiment P55: Light Intensity vs. Position (Light Sensor, Motion Sensor)

Experiment P55: Light Intensity vs. Position (Light Sensor, Motion Sensor) PASCO scientific Vol. 2 Physics Lab Manual: P55-1 Experiment P55: (Light Sensor, Motion Sensor) Concept Time SW Interface Macintosh file Windows file illuminance 30 m 500/700 P55 Light vs. Position P55_LTVM.SWS

More information

Spectroscopy Lab 2. Reading Your text books. Look under spectra, spectrometer, diffraction.

Spectroscopy Lab 2. Reading Your text books. Look under spectra, spectrometer, diffraction. 1 Spectroscopy Lab 2 Reading Your text books. Look under spectra, spectrometer, diffraction. Consult Sargent Welch Spectrum Charts on wall of lab. Note that only the most prominent wavelengths are displayed

More information

Communication Graphics Basic Vocabulary

Communication Graphics Basic Vocabulary Communication Graphics Basic Vocabulary Aperture: The size of the lens opening through which light passes, commonly known as f-stop. The aperture controls the volume of light that is allowed to reach the

More information

Fast Optical Form Measurements of Rough Cylindrical and Conical Surfaces in Diesel Fuel Injection Components

Fast Optical Form Measurements of Rough Cylindrical and Conical Surfaces in Diesel Fuel Injection Components Fast Optical Form Measurements of Rough Cylindrical and Conical Surfaces in Diesel Fuel Injection Components Thomas J. Dunn, Robert Michaels, Simon Lee, Mark Tronolone, and Andrew Kulawiec; Corning Tropel

More information

Alignment of the camera

Alignment of the camera Related topics Detector Alignment, Rotation axis, tilt, Principle Alignment of the detector and the rotation stage is very important to get optimal quality images of a CT scan. In this experiment, the

More information

EDUCATIONAL SPECTROPHOTOMETER ACCESSORY KIT AND EDUCATIONAL SPECTROPHOTOMETER SYSTEM

EDUCATIONAL SPECTROPHOTOMETER ACCESSORY KIT AND EDUCATIONAL SPECTROPHOTOMETER SYSTEM GAIN 1 10 100 Instruction Manual and Experiment Guide for the PASCO scientific Model OS-8537 and OS-8539 012-06575A 3/98 EDUCATIONAL SPECTROPHOTOMETER ACCESSORY KIT AND EDUCATIONAL SPECTROPHOTOMETER SYSTEM

More information

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE A. Miceli ab, R. Thierry a, A. Flisch a, U. Sennhauser a, F. Casali b a Empa - Swiss Federal Laboratories for

More information

The physics of ultrasound. Dr Graeme Taylor Guy s & St Thomas NHS Trust

The physics of ultrasound. Dr Graeme Taylor Guy s & St Thomas NHS Trust The physics of ultrasound Dr Graeme Taylor Guy s & St Thomas NHS Trust Physics & Instrumentation Modern ultrasound equipment is continually evolving This talk will cover the basics What will be covered?

More information

Exercise 4-1. Chaff Clouds EXERCISE OBJECTIVE

Exercise 4-1. Chaff Clouds EXERCISE OBJECTIVE Exercise 4-1 Chaff Clouds EXERCISE OBJECTIVE To demonstrate chaff as a method of denying target information to a radar. To verify whether MTI processing is an effective anti-chaff processing technique

More information

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893 Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -

More information

I = I 0 cos 2 θ (1.1)

I = I 0 cos 2 θ (1.1) Chapter 1 Faraday Rotation Experiment objectives: Observe the Faraday Effect, the rotation of a light wave s polarization vector in a material with a magnetic field directed along the wave s direction.

More information

Phase Imaging Using Focused Polycapillary Optics

Phase Imaging Using Focused Polycapillary Optics Phase Imaging Using Focused Polycapillary Optics Sajid Bashir, Sajjad Tahir, Jonathan C. Petruccelli, C.A. MacDonald Dept. of Physics, University at Albany, Albany, New York Abstract Contrast in conventional

More information

Activity P35: Light Intensity in Double-Slit and Single-Slit Diffraction Patterns (Light Sensor, Rotary Motion Sensor)

Activity P35: Light Intensity in Double-Slit and Single-Slit Diffraction Patterns (Light Sensor, Rotary Motion Sensor) Name Class Date Activity P35: Light Intensity in Double-Slit and Single-Slit Diffraction Patterns (Light Sensor, Rotary Motion Sensor) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Interference

More information

Photomultiplier Tube

Photomultiplier Tube Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or

More information

T h e P h a n t o m L a b o r a t o r y

T h e P h a n t o m L a b o r a t o r y T h e P h a n t o m L a b o r a t o r y 1 CCT228 ATCM Phantom Manual Copyright 2017 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order

More information

KULLIYYAH OF ENGINEERING

KULLIYYAH OF ENGINEERING KULLIYYAH OF ENGINEERING DEPARTMENT OF ELECTRICAL & COMPUTER ENGINEERING ANTENNA AND WAVE PROPAGATION LABORATORY (ECE 4103) EXPERIMENT NO 3 RADIATION PATTERN AND GAIN CHARACTERISTICS OF THE DISH (PARABOLIC)

More information

- Water resistant. - Large size.

- Water resistant. - Large size. GAFCHROMIC EBT product brochure GAFCHROMIC EBT FEATURES GAFCHROMIC EBT dosimetry film has been developed specifically to address the needs of the medical physicist and dosimetrist working in the radiotherapy

More information

Information & Instructions

Information & Instructions KEY FEATURES 1. USB 3.0 For the Fastest Transfer Rates Up to 10X faster than regular USB 2.0 connections (also USB 2.0 compatible) 2. High Resolution 4.2 MegaPixels resolution gives accurate profile measurements

More information

11Beamage-3. CMOS Beam Profiling Cameras

11Beamage-3. CMOS Beam Profiling Cameras 11Beamage-3 CMOS Beam Profiling Cameras Key Features USB 3.0 FOR THE FASTEST TRANSFER RATES Up to 10X faster than regular USB 2.0 connections (also USB 2.0 compatible) HIGH RESOLUTION 2.2 MPixels resolution

More information

ENSC 470/894 Lab 3 Version 6.0 (Nov. 19, 2015)

ENSC 470/894 Lab 3 Version 6.0 (Nov. 19, 2015) ENSC 470/894 Lab 3 Version 6.0 (Nov. 19, 2015) Purpose The purpose of the lab is (i) To measure the spot size and profile of the He-Ne laser beam and a laser pointer laser beam. (ii) To create a beam expander

More information

Learning Intentions: P3 Revision. Basically everything in the unit of Physics 3

Learning Intentions: P3 Revision. Basically everything in the unit of Physics 3 Learning Intentions: P3 Revision Basically everything in the unit of Physics 3 P3.1 Medical applications of physics Physics has many applications in the field of medicine. These include the uses of X-rays

More information

Section Electromagnetic Waves and the Electromagnetic Spectrum

Section Electromagnetic Waves and the Electromagnetic Spectrum Section 18.6 Electromagnetic Waves and the Electromagnetic Spectrum Electromagnetic Waves Electromagnetic Waves Electromagnetic waves are transverse waves produced by the motion of electrically charged

More information

Using the Advanced Sharpen Transformation

Using the Advanced Sharpen Transformation Using the Advanced Sharpen Transformation Written by Jonathan Sachs Revised 10 Aug 2014 Copyright 2002-2014 Digital Light & Color Introduction Picture Window Pro s Advanced Sharpen transformation is a

More information

INTRODUCTION. 5. Electromagnetic Waves

INTRODUCTION. 5. Electromagnetic Waves INTRODUCTION An electric current produces a magnetic field, and a changing magnetic field produces an electric field Because of such a connection, we refer to the phenomena of electricity and magnetism

More information

SPECTRAL SCANNER. Recycling

SPECTRAL SCANNER. Recycling SPECTRAL SCANNER The Spectral Scanner, produced on an original project of DV s.r.l., is an instrument to acquire with extreme simplicity the spectral distribution of the different wavelengths (spectral

More information

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and

More information

Part 1: Standing Waves - Measuring Wavelengths

Part 1: Standing Waves - Measuring Wavelengths Experiment 7 The Microwave experiment Aim: This experiment uses microwaves in order to demonstrate the formation of standing waves, verifying the wavelength λ of the microwaves as well as diffraction from

More information

Fire CR Calibration Guide

Fire CR Calibration Guide 1 Fire CR Calibration Guide This reference guide will guide you through the steps to complete the calibration for the Fire CR.. Getting Started: 1. Click on the Opal Icon on the Desktop. Figure 1 2. Once

More information

Ph 3455 The Photoelectric Effect

Ph 3455 The Photoelectric Effect Ph 3455 The Photoelectric Effect Required background reading Tipler, Llewellyn, section 3-3 Prelab Questions 1. In this experiment you will be using a mercury lamp as the source of photons. At the yellow

More information

diagnostic examination

diagnostic examination RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011

More information

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

3D Diode Array Commissioning: Building Confidence in 3D QA Technology 3D Diode Array Commissioning: Building Confidence in 3D QA Technology Caroline Yount, MS CANCER CENTER 3D QA The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose

More information

Collecting Live Data in Fathom NCTM 2013

Collecting Live Data in Fathom NCTM 2013 Collecting Live Data in Fathom NCTM 2013 Tyler Pulis hpulis@ncsu.edu Blake Whitley kbwhitle@ncsu.edu North Carolina State University, Raleigh, NC During this session, we will explore how to collect data

More information

Projectile Launcher (Order Code VPL)

Projectile Launcher (Order Code VPL) Projectile Launcher (Order Code VPL) The Vernier Projectile Launcher allows students to investigate important concepts in two-dimensional kinematics. Sample experiments include: Investigate projectile

More information

Concepts of Physics Lab 1: Motion

Concepts of Physics Lab 1: Motion THE MOTION DETECTOR Concepts of Physics Lab 1: Motion Taner Edis and Peter Rolnick Fall 2018 This lab is not a true experiment; it will just introduce you to how labs go. You will perform a series of activities

More information

INTRODUCTION TO FLEXIBLE BRONCHOSCOPY. Fluoroscopy Synopsis HENRI G COLT MD SECOND EDITION THE BRONCHOSCOPY EDUCATION PROJECT SERIES

INTRODUCTION TO FLEXIBLE BRONCHOSCOPY. Fluoroscopy Synopsis HENRI G COLT MD SECOND EDITION THE BRONCHOSCOPY EDUCATION PROJECT SERIES SECOND EDITION INTRODUCTION TO FLEXIBLE BRONCHOSCOPY Fluoroscopy Synopsis HENRI G COLT MD With contributions from Dr. S. Murgu THE BRONCHOSCOPY EDUCATION PROJECT SERIES FLUOROSCOPY SYNOPSIS The purpose

More information

CHM 109 Excel Refresher Exercise adapted from Dr. C. Bender s exercise

CHM 109 Excel Refresher Exercise adapted from Dr. C. Bender s exercise CHM 109 Excel Refresher Exercise adapted from Dr. C. Bender s exercise (1 point) (Also see appendix II: Summary for making spreadsheets and graphs with Excel.) You will use spreadsheets to analyze data

More information

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: )

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: ) University of Minnesota College of Science and Engineering Characterization Facility PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: 2012.10.17) The following instructions

More information

Introductory Physics, High School Learning Standards for a Full First-Year Course

Introductory Physics, High School Learning Standards for a Full First-Year Course Introductory Physics, High School Learning Standards for a Full First-Year Course I. C ONTENT S TANDARDS 4.1 Describe the measurable properties of waves (velocity, frequency, wavelength, amplitude, period)

More information

BEAMAGE-3.0 KEY FEATURES BEAM DIAGNOSTICS AVAILABLE MODELS MAIN FUNCTIONS SEE ALSO ACCESSORIES. CMOS Beam Profiling Cameras

BEAMAGE-3.0 KEY FEATURES BEAM DIAGNOSTICS AVAILABLE MODELS MAIN FUNCTIONS SEE ALSO ACCESSORIES. CMOS Beam Profiling Cameras BEAM DIAGNOSTICS BEAM DIAGNOSTICS SPECIAL PRODUCTS OEM DETECTORS THZ DETECTORS PHOTO DETECTORS HIGH POWER DETECTORS POWER DETECTORS ENERGY DETECTORS MONITORS CMOS Beam Profiling Cameras AVAILABLE MODELS

More information

MicroLab 500-series Getting Started

MicroLab 500-series Getting Started MicroLab 500-series Getting Started 2 Contents CHAPTER 1: Getting Started Connecting the Hardware....6 Installing the USB driver......6 Installing the Software.....8 Starting a new Experiment...8 CHAPTER

More information