CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

Size: px
Start display at page:

Download "CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY"

Transcription

1 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to acquire online volumetric and anatomical images (Xing 2006). The successful implementation of CBCT requires a quality assurance (QA) program that not only involves robust and intuitive quality control checks of the equipment, but also thoughtful image-guidance processes based on measurement of positional uncertainties and considering the practicalities of the patient throughput. This chapter discusses about the commissioning of kv-cbct system for the clinical use. It addresses a series of tests to ensure the CBCT system is safe both mechanically and dosimetrically for the patient use. 2.2 METHODS AND MATERIALS On-Board Imaging System The on-board imager (OBI) integrated in a Clinac 2100 C/D medical linear accelerator (Varian Medical Systems, Palo Alto, CA) is used in this work (Figure 2.1). The OBI system consists of a kv X-ray source and an amorphous silicon (asi1000) digital imaging detector mounted on the linear accelerator using robotic arms (Exact ), which are orthogonal to the

2 15 electronic portal imaging device (asi1000, PortalVision ). OBI system provides has capability of two-dimensional (2D) radiographic acquisition, fluoroscopic image acquisition, and CBCT acquisition. Figure 2.1 OBI hardware mounted on a Clinac 2100 C/D linear accelerator A 150 kv x-ray tube with maximum 32 ms pulse length for continuous irradiation and maximum 320 ms pulse length for single pulse is designed for generating high resolution images from a moving gantry. The focal spot of the tube is located at 90 to the mega-voltage (MV) source and 100 cm from the radiation axis of the accelerator. A 39.7 cm 29.8 cm asi1000 image detector acquires digital images with a pixel matrix of

3 16 Using the OBI system, the CBCT data can be acquired in two modes: a full-fan mode and half-fan mode. In the full-fan mode, the beam central axis passes through the detector centre and a full projection of the scanned patient is acquired for each acquirement angle. A total of 650 to 700 projections are taken during the whole 364 gantry rotation with a maximum field of view (FOV) about 25 cm in diameter and 17 cm in length. The data acquisition time is about 60 s and the reconstruction time for 340 slices of CBCT images with a voxel size of 0.5 mm 3 is also about 1 min. The half-fan mode is designed to obtain a larger FOV. In this mode, the detector is shifted laterally by 14.8 cm to take only half of the projection of the scanned patient for each acquirement angle. In half-fan mode, a FOV of 45 cm in the axial plane and 15 cm in the longitudinal direction can be achieved. The effects of X-ray scatter and artifacts are larger in CBCT images than in CT images. A bowtie filter mounted to the X-ray tube improved image quality because it reduces intensity variations across the detector and charge trapped in the detector Commissioning Procedure The test categories, which are introduced in detail in the subsections that follow, were i. System mechanical safety ii. iii. iv. Geometric accuracy (agreement of MV and kv beam isocenter) CBCT image quality Registration and couch shift accuracy, and v. Dose to the patient

4 System mechanical safety The safety QA test checks the safety features built into the OBI system. As per manufacturer s recommendation, tube warm-up is performed before patient treatment to prevent premature failure of the x-ray tube. While the x-ray is on for the tube warm-up, the door interlock, warning lights, and audible warning can be checked. After 20 s of the warm-up, the arms can be retracted using the hand pendant in the room. While retracting the arms, the motion-enable bars of the hand pendant to be released to verify that arms stop moving. The collision detection covers and paddles can also be checked, one at a time, while retracting the arms Geometric accuracy (agreement of MV and kv beam isocenter) Alignment of the isocenters for the kv imaging and the MV treatment system is crucial for accurate patient positioning, because the kv imaging system is used to position the patient with respect to the MV treatment system. The OBI isocenter accuracy quality assurance test evaluates whether the digital graticule generated by the OBI application coincides with the treatment (MV) isocenter. This check is performed using marker block with one fiducial marker at the center and four markers at known locations inside the block. A test patient was created with a plan containing two kv setup fields. The setup field had a digitally reconstructed radiograph (DRR) associated with it so that the entire patient reposition process could be performed. The phantom was placed on the couch using field light cross-hair and the wall lasers so that the center of the phantom was aligned with the treatment isocenter. First, agreement between the center fiducial marker and the isocenter of the MV beam is established by image-supported adjustment of the position of the marker block. Eight MV portal images are taken at the

5 18 four cardinal gantry angles and at collimator settings -90 degrees and 90 degrees. Based on image analysis by system software, the deviation of the location of the center fiducial marker from the location of the MV beam isocenter is calculated. The position of the marker is adjusted, and the images are retaken. The process is repeated until the deviation in any directions is below 0.25 mm, the threshold suggested by the manufacturer. Then, the deviation of the fiducial marker from the center of the kv imaging system is determined. Four images are taken with the kv system at four cardinal gantry angles. On each image, pixel locations are used to determine the deviation between the center of the image (automatically marked with digital crosshairs by the software) and the center of the fiducial marker in the image. The horizontal and vertical differences between the observed center of the fiducial marker and the crosshairs are calculated for the image. These measurements and subtractions are performed for each of the four images. Based on other tolerances in the QA process of linear accelerators, the deviation limit for this test was set to 1.04 mm in each direction, which corresponds to 4 pixels. The kv image pixel size corresponds to mm at the isocenter for this high resolution acquisition mode CBCT image quality The image quality achievable with the CBCT imaging system was tested for maximal achievable resolution and ability to display low-contrast objects. A Catphan 504 phantom (The Phantom Laboratory, Salem, NY) was used for image quality measurements. The phantom consists of various cylindrical sections (modules), each of which is designed for a specific test. The Catphan phantom was hung over the end of the couch with the aid of its wooden case and leveled. The phantom was centered so that its physical center was at the treatment isocenter and one full-fan and half-fan CBCT scan of the phantom with scan parameters 125 kvp, 80 ma, 25 ms, 150 cm source

6 19 to imager distance (SID) with a bowtie filter was acquired. The reconstruction matrix was and the slice thickness was 2.5 mm. From the scans, all image quality evaluations were made. (a) Hounsfield Units (HU) linearity The CTP 404 insert contains seven tissue substitute materials of different densities. The materials and the expected HU values are listed in the Catphan phantom manual. A region of interest was selected using the Area profile tool to measure the mean HU value for each material in the CBCT application. The region of interest (ROI) size was adjusted to cover approximately 7 mm 7 mm, which covered about pixels for the full-fan mode and 8 8 pixels for the half-fan mode. The measured HU value for each material should be within ± 40 HU of the value indicated in the Catphan manual. (b) Low contrast resolution In the CTP 515 insert, there are three groups of nine supra-slice disks with diameters ranging from 2-15 mm and subject contrasts of 1%, 0.5%, and 0.3%. The image is visually inspected and the lowest contrast and lowest diameter supra-slice disk that was determined. The recommendation is that the 1%, 7 mm disk should be visible [Varian CBCT Customer Acceptance Procedure]. (c) Spatial resolution (high contrast resolution) The CTP 528 insert contains a spatial resolution rule with bar patterns between 1 to 21 line pairs/cm (lp/cm). The least discernable bar pattern in CBCT images should be selected and compared to the baseline obtained at the time of commissioning and recommendation is that 6 lp/cm [Varian CBCT Customer Acceptance Procedure], which is group 6 should be visible.

7 20 (d) HU uniformity The CTP 486 insert is a uniform disk of 20 cm diameter that has been used to access HU uniformity. In CTP 486, the image slice containing the white markers is selected for the measurement of HU uniformity. Five ROI s one in center and four symmetrically in the upper, lower, left and right peripheral areas in the phantom were selected. Using area profile tool, the HU values in 1 cm 1 cm ROI was measured. The HU values for the all regions should be within ± 40 HU of each other [Varian CBCT Customer Acceptance Procedure]. (e) In slice spatial linearity The geometric accuracy of the CBCT scans is determined by the magnification accuracy as that defines the divergence of the x-ray beam. If the actual divergences of the x-ray beam matches that assumed in the reconstruction, then the CBCT images will be geometrically consistent. The in-slice spatial linearity can be checked using the CTP 404 insert. There are four rods of 3 mm diameter at the corners of a 5 cm square in the CTP 404 module. The distance between pair of rods in the axial view using the distance measuring tool in the CBCT application was measured and averaged. The tolerance of less than 1% of difference (± 0.5 mm) is recommended [Varian CBCT Customer Acceptance Procedure] Registration and couch shift accuracy The CBCT system s ability to correctly register a localization geometry with a reference geometry was tested using anthropomorphic phantom (Rando phantom) that went through the entire patient planning chain. Several methods of performing such tests have been presented in the literature (Letourneau 2005, Oldham 2005 and Sharpe 2006).

8 21 In this test, the algorithm for repositioning the patient at the isocenter was tested to show that the system can precisely detect a preset deviation and determine the appropriate shift back to alignment. The head and neck part of anthropomorphic phantom was used for the test. With the CBCT system, alignment of the phantom on the treatment couch and its position relative to isocenter were first brought into agreement with a reference image (DRR) from a previous CT scan. The phantom was then moved by specific distances in one or more directions relative to the reference position. Each movement of the phantom away from the initial start position was measured by the couch digital readout (1 mm resolution). The movements were each 10 mm in magnitude. The phantom was first moved right, then additionally up and finally in. The ability of the OBI system to accurately describe these moves was analyzed for each step. At the end, the phantom was moved back to isocenter using the repositioning directions generated by the algorithm. Following the repositioning, another CBCT was taken and compared with the reference image Dose to the patient The dose to the patients from a CBCT scan depends significantly on the scanning parameters used. In addition, based on scanning technique (full-fan or half-fan), the dose across the patient may vary. Various authors has already been reported the patterns of dose distribution and the impacts of the various factors on the dose (Letourneau 2005, Islam 2006 and Sawyer 2009). As a part of quality assurance, the dose measurement for a 360 degree CBCT and measurement of the half-value layer of the kv beam was done. Doses were measured at the centre and periphery (1.2 cm from the edge) of a 32 cm diameter cylindrical perspex phantom with 9.5 cm thickness. It should be noted that the limited thickness of computed tomography dose index (CTDI) phantoms can significantly underestimate the contribution of scatter (Perisinakis 2007). A 0.6 cc Farmer type ionization chamber calibrated in the kv energy range was used for the dose measurements in conjunction

9 22 with a NE 2670 electrometer. The measurements were performed for the scanning parameters 125 kv, 80 ma and 25 ms per frame for a 360 degree scan. The half value layer of the kv beam was determined using a stationary beam with aluminum plates and standard procedures (Ma 2001). 2.3 RESULTS System mechanical safety The beam termination at the control console and the door interlock were functional. A not-fully-extended kv source arm and an open door each inhibited engagement of the beam. All touch guards were found to work correctly, in that they inhibited motion when activated Geometric accuracy (agreement of MV and kv beam isocenter) The maximum deviation of the centre fiducial marker from the MV isocenter was below the threshold of 0.25 mm after two iterations of adjustment. Analysis of the subsequent kv images determined that the maximal deviation between the kv isocenter and the center fiducial marker was 1 mm or less in each orthogonal direction. Figure 2.2 shows an example of the fiducial marker in marker phantom aligned with the center of the kv images. Figure 2.2 Orthogonal kilovoltage images of the marker phantom

10 CBCT image quality (a) Hounsfield Units linearity In full-fan mode, the maximum differences in HU value compared with reference CT were less than 20 HU. For the half-fan mode, teflon material shows the maximum HU difference from CT of 40 HU. Figure 2.3 (a) shows the HU profile measured along the Catphan phantom. (b) Contrast resolution Figure 2.3 (b) shows CBCT images of CTP 515 insert, which measures low contrast resolution. In 1% supraslice group, 6 mm and 7 mm disks are clearly discernable in full-fan mode and half-fan mode respectively. The images acquired using a full-fan mode have better contrast resolution than those acquired using half-fan mode. This is because, the volume irradiated in full-fan mode receives a higher dose than the volume irradiated by a half-fan mode. (c) Spatial resolution (high contrast resolution) Figure 2.3 (c) shows a CBCT image of the CTP 528 insert of Catphan phantom. The images obtained using full-fan and half-fan mode scan could able to resolve the 7 lp/cm and 6 lp/cm spatial resolution respectively. As the pixel size in full-fan mode was smaller than that in the half-fan mode, the images acquired using full-fan mode had a better spatial resolution than half-fan mode images. (d) HU uniformity In CTP 486 insert of Catphan phantom, the mean HU values in five regions of interest (ROI: 1 cm 1 cm) were measured to determine HU uniformity. All HU values measured in five ROI for full-fan and half-fan mode acquisition were within the specification of ± 40 HU. The HU values in peripheral regions were found be lesser than that of middle of phantom. The

11 24 images acquired using full-fan mode had more homogeneous HU values than those acquired using half-fan mode. (e) In-slice spatial linearity Figure 2.4 (d) displays an image of the CTP 404 insert, which can be used for the spatial linearity test. The four rods are shown and the two lines generated using the measure distance tool indicate exactly 5 cm separation between the two upper rods and the two right rods. Figure 2.3 Sample images of Catphan 504 phantom imaged using kv-cbct in full-fan mode: (a) HU profile comparison (b) Contrast resolution test (c) Spatial resolution test (d) Spatial linearity test

12 Registration and couch shift accuracy The results of this test are summarized in Table 2.1. It shows the initial positional deviations of the head and neck part of Anthropomorphic phantom about the linear accelerator isocenter after alignment to the DRR of previous CT scan as Step 1. Steps 2 4 are the cumulative shifts of 10 mm each in the lateral direction (Step 2, patient s right), the vertical direction (Step 3, couch up), and the longitudinal direction (Step 5, couch in). The data in the corresponding columns on the right show that the algorithm detected translational shift for every move with an accuracy of better than 1 mm, which is the equivalent of the tolerance of the couch. The rotational error (which should be zero, because no rotations were introduced) stayed within 0.3 degree. To verify the reproducibility of the system, two more scans and image alignments were performed without moving the phantom (Steps 5 and 6). The fluctuation found in this test was within 0.1 mm and 0.1 degree for translation and rotation respectively. In the final step, Step 7, the phantom was moved back to isocenter using the repositioning directions generated by the algorithm. The data show that the phantom was brought within 0.4 mm and 0.4 degree of the starting point (Step 1). Figure 2.5 shows a transverse and sagittal view of the phantom after registration. Table 2.1 Step Position errors reported by the registration following defined moves of the phantom Defined phantom movement 1 Initial position lat: -10 mm (1 cm patient s right) vrt: -10 mm (additional 1 cm up) lng: 10 mm (additional 1 cm in) Position error after 3D-3D match Translation Rotation vrt : 0.4 pitch : 0.2 lng : 0.3 roll : 0.2 lat : 0.3 rtn : 0.3 vrt : 0.3 lng : 0.2 lat : -9.9 vrt : lng : 0.1 lat : vrt : lng : 9.9 lat : -9.9 pitch : 0.3 roll : 0.2 rtn : 0.2 pitch : 0.1 roll : 0.2 rtn : -0.1 pitch : 0.2 roll : 0.2 rtn : -0.1

13 26 Table 2.1 (Continued) Step Defined phantom movement No movement (Reproducibility check) No movement (Reproducibility check) vrt: 10 mm, lng: -9.9 mm, lat: Position error after 3D-3D match vrt : lng : 9.9 lat : 10.0 vrt : lng : 9.9 lat : 10.1 vrt : 0.1 lng : -0.1 lat : 0.2 pitch : 0.2 roll : 0.3 rtn : 0.0 pitch : 0.2 roll : 0.2 rtn : -0.1 pitch : 0.2 roll : 0.1 rtn : -0.1 Figure 2.4 A transverse and sagittal view of the phantom after registration. The selected display type is chess view, showing alternating panels of reconstructed and reference image Dose to the Patient For full-fan mode with the scanning parameters 125 kv, 80 ma, 40 ms per frame, for a 360 degree scan (~650 projections), the dose in the CTDI phantom was 23.4 mgy at the isocentre and 53.1 mgy at the periphery. This can be compared with a value of 54 mgy reported for the same procedure in

14 27 the literature (Song 2008). With bowtie filter, the half-value layer of the beam was found to be 6.1 mm Al. 2.4 DISCUSSION The clinical introduction of kv-cbct based IGRT systems necessitates formal commissioning of hardware and image-guided processes to be used. Because, positioning of the patient is based directly on data from the CBCT systems, their performance has a major effect on the outcome of treatment. The accuracy and reliability of CBCT systems therefore need to be tested and related to familiar tolerances of the treatment process. This study details about the commissioning of kv-cbct integrated into Clinac 2100 C/D linear accelerator. The mechanical safety of the OBI system was verified. The significant point of concern is the agreement of the kv imaging line isocenter and the MV treatment line isocenter. In the commissioning, this agreement was found to be within 1 mm. The importance of this agreement has already been reported by others (Sharpe 2006 and Oldham 2005). The test of CBCT image quality only used the Catphan phantom. However, the CBCT images of a large patient can result in more artifacts due to phenomenon such as x-ray scatter and charge trapping than would occur with the Catphan phantom. Therefore it is advisable to acquire the CBCT images of an anthropomorphic phantom on a routine basis to detect changes in image quality that might not be detected by the Catphan measurements. Regarding image registration and couch shift accuracy, we have performed a simple QA using Rando phantom where the image analysis and position correction algorithm was found to be within the resolution of the patient couch motion (1 mm and 1 degree). The dose to the patient from a

15 28 CBCT scan and its relation to image quality is an interesting field of research. If the primary use for the images is to enable reliable patient alignment before treatment, image quality need be sufficient only for that task. Optimal settings of parameters like kilovolt, milliampere, field size and other parameters specific to each anatomic area will be found through clinical use and protocol development. The impact of target motion in CBCT imaging, use of CBCT scans for dose recalculation based on the patient s anatomy on the day of treatment and CBCT based adaptive radiotherapy were studied in detail in later chapters. 2.5 CONCLUSION The development of volumetric imaging in the treatment room and the opportunity for off-line and on-line guidance of radiotherapy offer strategies to increase the precision and accuracy of dose delivery. The above described procedures allow for safe clinical implementation of CBCT systems in a modern radiotherapy department.

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup Huaiqun Guan,

More information

CHAPTER 6 QUALITY ASSURANCE OF VARIAN ON-BOARD IMAGER

CHAPTER 6 QUALITY ASSURANCE OF VARIAN ON-BOARD IMAGER 127 CHAPTER 6 QUALITY ASSURANCE OF VARIAN ON-BOARD IMAGER 6.1 INTRODUCTION Accurate and repeatable setup of patients is a requisite in radiotherapy. In the treatment of head-and-neck tumors, accurate setup

More information

Imaging system QA of a medical accelerator, Novalis Tx, for IGRT per TG 142: our 1 year experience

Imaging system QA of a medical accelerator, Novalis Tx, for IGRT per TG 142: our 1 year experience JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 4, 2012 Imaging system QA of a medical accelerator, Novalis Tx, for IGRT per TG 142: our 1 year experience Zheng Chang, a James Bowsher, Jing

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING

QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING 1 QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING QualiFormeD Phantoms A selection of test objects facilitating regulatory quality controls in radiation therapy and medical imaging Practical,

More information

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc.

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. An Introduction to TG-142 Imaging QA Using Standard Imaging Products Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. Goals Understand the nature and intent of TG 142 imaging

More information

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE Rev. 1.0 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE DoseLab users may reference the following instructions to perform Tomotherapy Quality Assurance tests as recommended

More information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information Ordering Information Please contact us if you have any questions or if you would like a quote or delivery schedule regarding the Catphan phantom. phone 800-525-1190, or 518-692-1190 fax 518-692-3329 mail

More information

Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film

Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 4, 2016 Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film Eric C. Lobb Department

More information

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017 The Current State of EPID-Based Linear Accelerator Quality Assurance Timothy Ritter, PhD, DABR, FAAPM 1 Disclosures Employed by the Veterans Health Administration Faculty appointment with the University

More information

PHYSICS QUESTIONNAIRE FORM

PHYSICS QUESTIONNAIRE FORM PHYSICS QUESTIONNAIRE FORM Institution Name: Date: Contact Information (name, address, phone, fax, email): Physicist: Radiation Oncologist: Dosimetrist (if applicable): Study Coordinator (if applicable):

More information

Image Quality and Dose of an Accelerator-Integrated kv CBCT Systems

Image Quality and Dose of an Accelerator-Integrated kv CBCT Systems Image Quality and Dose of an Accelerator-Integrated kv CBCT Systems by Amani Shaaer A thesis submitted in partial fulfillment of the requirements for the degree of Master of Science in Physics School of

More information

T h e P h a n t o m L a b o r a t o r y

T h e P h a n t o m L a b o r a t o r y T h e P h a n t o m L a b o r a t o r y 1 CCT228 ATCM Phantom Manual Copyright 2017 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order

More information

CT Scanner Dose Survey

CT Scanner Dose Survey CT Scanner Dose Survey Measurement Protocol Version 5.0 July 1997 Co-ordinated by ImPACT and The Medical Physics Department,, London SW17, UK. 0181-725-3366 CT Scanner Dose Survey: Measurement Protocol

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

specifications TrueBeam System

specifications TrueBeam System specifications TrueBeam System TrueBeam System s The TrueBeam system specifications in this document are identified as belonging to two categories, performance specifications and descriptive specifications.

More information

Phantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging

Phantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging in partnership with Phantoms in Medical Physics (RT) U. Oelfke Division of Radiotherapy & Imaging uwe.oelfke@icr.ac.uk Making the discoveries that defeat cancer 1. Introduction What is a phantom? Wiki:

More information

Sub-mm accuracy of accelerators: How manufacturers achieve it, how physicists verify it

Sub-mm accuracy of accelerators: How manufacturers achieve it, how physicists verify it Sub-mm accuracy of accelerators: How manufacturers achieve it, how physicists verify it Presented by Ivan A. Brezovich, PhD, at the 2015 Annual meeting of the SEAAPM, Raleigh, NC, April 24, 2015 Purpose

More information

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook DIAGNOSTIC ACCREDITATION PROGRAM Radiology and CT Quality Control Procedures Workbook Quality Control Procedures Radiography/CR/DR Safety Code 35 Summary For more detail about each quality control (QC)

More information

specifications TrueBeam STx System

specifications TrueBeam STx System specifications TrueBeam STx System TrueBeam STx s The TrueBeam STx system specifications in this document are identified as belonging to two categories, performance specifications and descriptive specifications.

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

Nathan Childress, Ph.D., DABR

Nathan Childress, Ph.D., DABR Nathan Childress, Ph.D., DABR Introduction TG-142 is a comprehensive QA protocol Covers nearly every aspect of machine and safety QA Recommends quantitative results Recommends high testing frequencies

More information

Test Equipment for Radiology and CT Quality Control Contents

Test Equipment for Radiology and CT Quality Control Contents Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3

More information

Wide beam CT dosimetry. Elly Castellano

Wide beam CT dosimetry. Elly Castellano Wide beam CT dosimetry Elly Castellano Outline revision: CT dose indices wide-beam CT: the end of the road for CTDI? the IEC rescue plan for CTDI 100 the american way AAPM report 111 better estimates of

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Gena M.A.H 1, Ahmed L.El-Attar 2, Elbadry M. Zahran 3, Hany El-Gamal

More information

A Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017

A Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017 A Fast Monolithic System for Proton Imaging Fritz DeJongh ProtonVDA Inc October 2017 Disclosures I am a cofounder and co-owner of ProtonVDA Inc We hold intellectual property rights on our proton imaging

More information

COMPREHENSIVE TG-142 IMAGING AND MACHINE QA

COMPREHENSIVE TG-142 IMAGING AND MACHINE QA QA SOFTWARE COMPREHENSIVE TG-142 IMAGING AND MACHINE QA Automate the analysis of over thirty TG-142 recommended QA tasks The rapid progress of Radiation Therapy has created the need for Quality Assurance

More information

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients

More information

Overview of Safety Code 35

Overview of Safety Code 35 Common Quality Control Procedures for All s Quality Control Procedures Film All s Daily Quality Control Tests Equipment Warm-up (D1) According to manufacturers instructions Can include auto calibration(d1)

More information

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System.

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System. Use of EPIDs for Non-Routine Linac QA Bin Cai PhD Disclosure Parts of this project received support from Varian Medical System. Learning Objectives Learn the recent development of EPID based Non-routine

More information

abc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice nww.medical-devices.nhs.

abc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice   nww.medical-devices.nhs. abc September 2004 MHRA 04099 Philips Mx8000 IDT CT scanner technical evaluation Best choice best practice www.mhra.gov.uk nww.medical-devices.nhs.uk About MHRA evaluation reports. What you can expect.

More information

Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging

Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2008 Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging Dean

More information

Aim. Images for this section: Page 2 of 13

Aim. Images for this section: Page 2 of 13 Changes in CT number of high atomic number materials with field of view when using an extended CT number to electron density curve and a metal artifact reduction reconstruction algorithm Poster No.: R-0094

More information

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity) Vascular Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction In recent years, digital cardiovascular

More information

QC by the MPE in Belgium

QC by the MPE in Belgium Acceptance testing of state-of-the-art CT scanners using a new national protocol: first experience on a large number of scanners of different make and model the working group Radiology of the Belgian Hospital

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

Data. microcat +SPECT

Data. microcat +SPECT Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for

More information

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT Quality control of Gamma Camera By Dr/ Ibrahim Elsayed Saad 242 NMT WHAT IS QUALITY? The quality of a practice is to fulfill the expectations and demands from: Patient Clinicain Your self Quality assurance

More information

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT

More information

Maximizing clinical outcomes

Maximizing clinical outcomes Maximizing clinical outcomes Digital Tomosynthesis Dual Energy Subtraction Automated Long Length Imaging Improved image quality at a low dose Xray Xray Patented ISS capture technology promotes high sensitivity

More information

X-RAYS - NO UNAUTHORISED ENTRY

X-RAYS - NO UNAUTHORISED ENTRY Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are

More information

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose Image Quality and Dose Issues in MSCT Image Quality and Dose Image quality Image noise Spatial resolution Contrast Artefacts Speckle and sharpness S. Edyvean St. George s Hospital London SW17 0QT Radiation

More information

ISO Cube Daily QA Package

ISO Cube Daily QA Package ISO Cube Daily QA Package Model 023-05 AFFORDABLE TURNKEY SOLUTION FOR DAILY MACHINE QA POWERED BY AQUILAB 2428 Almeda Avenue Suite 316 Norfolk, Virginia 23513 USA Tel: 757-855-2765 WWW.CIRSINC.COM CAPABILITIES

More information

Introduction of a Single Chip TLD System for Patient Dosimetry

Introduction of a Single Chip TLD System for Patient Dosimetry Introduction of a Single Chip TLD System for Patient Dosimetry C. Hranitzky a, M. Halda a, G. Müller a, B. Obryk b, H. Stadtmann a* a Austrian Research Centers GmbH ARC, 2444 Seibersdorf, Austria. b Institute

More information

Initial setup and subsequent temporal position monitoring using implanted RF transponders

Initial setup and subsequent temporal position monitoring using implanted RF transponders Initial setup and subsequent temporal position monitoring using implanted RF transponders James Balter, Ph.D. University of Michigan Has financial interest in Calypso Medical Technologies Acknowledgements

More information

SUN NUCLEAR. EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm. corporation. Your Most Valuable QA and Dosimetry Tools

SUN NUCLEAR. EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm. corporation. Your Most Valuable QA and Dosimetry Tools EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm SUN NUCLEAR corporation Your Most Valuable QA and Dosimetry Tools introduction Pre-treatment dose QA is an important process required

More information

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

3D Diode Array Commissioning: Building Confidence in 3D QA Technology 3D Diode Array Commissioning: Building Confidence in 3D QA Technology Caroline Yount, MS CANCER CENTER 3D QA The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose

More information

X-RAD SmART (Small Animal RadioTherapy) IMAGE GUIDED BIOLOGICAL IRRADIATOR Technical Specifications (updated March 2015)

X-RAD SmART (Small Animal RadioTherapy) IMAGE GUIDED BIOLOGICAL IRRADIATOR Technical Specifications (updated March 2015) X-RAD SmART (Small Animal RadioTherapy) IMAGE GUIDED BIOLOGICAL IRRADIATOR Technical Specifications (updated March 2015) The X-RAD SmART Small Animal IGRT research system provides a high accuracy cone

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

X3D in Radiation Therapy Procedure Planning. Felix G. Hamza-Lup, Ph.D. Computer Science Armstrong Atlantic State University Savannah, Georgia USA

X3D in Radiation Therapy Procedure Planning. Felix G. Hamza-Lup, Ph.D. Computer Science Armstrong Atlantic State University Savannah, Georgia USA X3D in Radiation Therapy Procedure Planning Felix G. Hamza-Lup, Ph.D. Computer Science Armstrong Atlantic State University Savannah, Georgia USA Outline 1. What is radiation therapy? 2. Treatment planning

More information

A proposed method for linear accelerator photon beam steering using EPID

A proposed method for linear accelerator photon beam steering using EPID Received: 13 January 2018 Revised: 11 May 2018 Accepted: 29 June 2018 DOI: 10.1002/acm2.12419 RADIATION ONCOLOGY PHYSICS A proposed method for linear accelerator photon beam steering using EPID Michael

More information

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II APPROVED 12/O2/2011 EFFECTIVE SPRING 2013-14 Prefix & Number RAD 150 Course Title: Radiologic Exposure Technique II & Lab Purpose of this submission: New Change/Updated

More information

The influence of bowtie filtration on cone-beam CT image quality

The influence of bowtie filtration on cone-beam CT image quality The influence of bowtie filtration on cone-beam CT image quality N. Mail Radiation Medicine Program, Princess Margaret Hospital, Toronto, Ontario M5G 2M9, Canada and Ontario Cancer Institute, University

More information

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This

More information

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure Accuracy and calibration of integrated radiation output indicators in diagnostic radiology: A report of the AAPM Imaging Physics Committee Task Group 190 Pei-Jan P. Lin a) Virginia Commonwealth University

More information

8/3/2016. The EPID Strikes Back. Novel Applications for Current EPID Technology. Joerg Rottmann, PhD. Disclosures and acknowledgements

8/3/2016. The EPID Strikes Back. Novel Applications for Current EPID Technology. Joerg Rottmann, PhD. Disclosures and acknowledgements The EPID Strikes Back Joerg Rottmann Brigham and Women s Hospital / Dana-Farber Cancer Institute Harvard Medical School Disclosures and acknowledgements Disclosures Varian MRA grant Acknowledgements Boston

More information

1. Patient size AEC. Large Patient High ma. Small Patient Low ma

1. Patient size AEC. Large Patient High ma. Small Patient Low ma Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description

More information

Nuclear Associates , , CT Head and Body Dose Phantom

Nuclear Associates , , CT Head and Body Dose Phantom Nuclear Associates 76-414,76-414-4150,76-415 CT Head and Body Dose Phantom Users Manual March 2005 Manual No. 76-414-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product

More information

CyberKnife Iris Beam QA using Fluence Divergence

CyberKnife Iris Beam QA using Fluence Divergence CyberKnife Iris Beam QA using Fluence Divergence Ronald Berg, Ph.D., Jesse McKay, M.S. and Brett Nelson, M.S. Erlanger Medical Center and Logos Systems, Scotts Valley, CA Introduction The CyberKnife radiosurgery

More information

QUALITY CONTROL SOFTWARE FOR RADIOTHERAPY AND MEDICAL IMAGING

QUALITY CONTROL SOFTWARE FOR RADIOTHERAPY AND MEDICAL IMAGING 1 QUALITY CONTROL SOFTWARE FOR RADIOTHERAPY AND MEDICAL IMAGING The QUALIMAGIQ platform and its modules An optimal software solution for automatic analyses and traceability of quality controls in radiation

More information

New spectral benefi ts, proven low dose

New spectral benefi ts, proven low dose New spectral benefi ts, proven low dose Philips MicroDose mammography SI, technical data sheet Philips MicroDose SI with single-shot spectral imaging is a fullfi eld digital mammography solution that delivers

More information

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

More information

Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method

Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method Markus Wendling, Robert J. W. Louwe, a Leah N. McDermott, Jan-Jakob Sonke, Marcel van Herk, and Ben J. Mijnheer

More information

Amorphous Selenium Direct Radiography for Industrial Imaging

Amorphous Selenium Direct Radiography for Industrial Imaging DGZfP Proceedings BB 67-CD Paper 22 Computerized Tomography for Industrial Applications and Image Processing in Radiology March 15-17, 1999, Berlin, Germany Amorphous Selenium Direct Radiography for Industrial

More information

A Beam-Level Delivery Accuracy Study of the Robotic Image Guided Radiosurgery System Using a Scintillator/CCD Phantom

A Beam-Level Delivery Accuracy Study of the Robotic Image Guided Radiosurgery System Using a Scintillator/CCD Phantom A Beam-Level Delivery Accuracy Study of the Robotic Image Guided Radiosurgery System Using a Scintillator/CCD Phantom Lei Wang 1, Shi Liu 1, Brett Nelson 2 1. Department of Radiation Oncology, Stanford

More information

inspexio SMX-225CT FPD HR

inspexio SMX-225CT FPD HR Microfocus X-Ray CT System C251-E029A Advanced Operability and Excellent Image Quality That Overturns Conventional Assumptions Microfocus X-Ray CT System The is a high-performance microfocus X-ray CT system

More information

GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications

GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications I. SCOPE The protocol applies to GafChromic EBT3P and EBT3+P films exposed in GafChromic QuiCk Phantom

More information

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 1, WINTER 2010 A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA Abdul Qadir Jangda,

More information

P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s

P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s AN ALL AROUND PERFECT PICTURE. The perfect combination of image quality, efficiency and design. P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s Air Techniques experts in digital imaging ProVecta

More information

P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s

P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s AN ALL AROUND PERFECT PICTURE. The perfect combination of image quality, efficiency and design. P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s Air Techniques experts in digital imaging ProVecta

More information

NeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results

NeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results NeuViz 16 Computed Tomography Elevating routine imaging for exceptional results Essence NeuViz 16 Raising the bar on clinical utility in routine imaging. Get more. More clinical information for patients.

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

v tome x m microfocus CT

v tome x m microfocus CT GE Inspection Technologies v tome x m microfocus CT Uniting premium 3D metrology and inspection with quality and speed. gemeasurement.com/ct x plore precision CT line Inspect with precision, power, and

More information

Teaching Digital Radiography and Fluoroscopic Radiation Protection

Teaching Digital Radiography and Fluoroscopic Radiation Protection Teaching Digital Radiography and Fluoroscopic Radiation Protection WCEC 20 th Student Educator Radiographer Conference Dennis Bowman, RT(R), CRT (R)(F) Community Hospital of the Monterey Peninsula (CHOMP)

More information

7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators

7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators Implementing a new digital medical accelerator John Wong Johns Hopkins University AAPM, Austin, 2014 Acknowledgements Yin Zhang, Ken Wang, Kai Ding (Commissioning - JHU) Esteban Velarde, Joe Moore (QA

More information

Clinical Use of Electronic Portal Imaging : Report of AAPM Radiation Therapy Committee Task Group 58

Clinical Use of Electronic Portal Imaging : Report of AAPM Radiation Therapy Committee Task Group 58 Clinical Use of Electronic Portal Imaging : Report of AAPM Radiation Therapy Committee Task Group 58 AAPM Refresher Course Salt Lake City July 2001 Michael G. Herman Division of Radiation Oncology, Mayo

More information

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard.

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard. Chiara Secco PET Performance measurements of the new LSO-Based Whole Body PET/CT Scanner biograph 16 HI-REZ using the NEMA NU 2-2001 Standard. INTRODUCTION Since its introduction, CT has become a fundamental

More information

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation Image Quality Review I: Basics and Image Quality TH-A-16A-1 Thursday 7:30AM - 9:30AM Room: 16A J. Anthony

More information

CT Basics: Data Acquisition Module 3

CT Basics: Data Acquisition Module 3 Module 3 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

diagnostic examination

diagnostic examination RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

Distributed source x-ray tube technology for tomosynthesis imaging

Distributed source x-ray tube technology for tomosynthesis imaging Distributed source x-ray tube technology for tomosynthesis imaging Authors: F. Sprenger a*, X. Calderon-Colon b, Y. Cheng a, K. Englestad a, J. Lu b, J. Maltz c, A. Paidi c, X. Qian b, D. Spronk a, S.

More information

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter Revision Date: January 5th, 2017 www.disc-imaging.com Table of Contents Section A: Preliminary Setup Requirements... 4 Tools

More information

Collimation Assessment Using GAFCHROMIC XR-M2

Collimation Assessment Using GAFCHROMIC XR-M2 Collimation Assessment Using GAFCHROMIC XR-M2 I. Introduction A method of collimation assessment for GE Senographe full-field digital mammography (FFDM) systems is described that uses a self-developing

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

Electronic Brachytherapy Sources. Thomas W. Rusch

Electronic Brachytherapy Sources. Thomas W. Rusch Electronic Brachytherapy Sources Thomas W. Rusch Educational Objectives Understand key elements of ebx source construction & operation Understand the rationale and methods for air kerma strength calibration

More information

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National

More information

Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA

Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA Arthur Olch, PhD, FAAPM AAPM Spring Clinical Meeting, March 21, 2017 Or.. What Dose are the Patients Really Getting???

More information

2D, 3D CT Intervention, and CT Fluoroscopy

2D, 3D CT Intervention, and CT Fluoroscopy 2D, 3D CT Intervention, and CT Fluoroscopy SOMATOM Definition, Definition AS, Definition Flash Answers for life. Siemens CT Vision Siemens CT Vision The justification for the existence of the entire medical

More information

Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator

Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator Aleksei E. Zhdanov 1 and Leonid G. Dorosinskiy 1 Ural Federal University named after the first President of Russia B. N.

More information

A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine

A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine Iwein Van de Vondel, 1 Koen Tournel, 1 Dirk

More information

Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision Minnesota Rules, Chapter 4732 X-ray Revision DRAFT FLUOROSCOPIC X-RAY SYSTEMS, 1.0 Subpart 1. Applicability. Subpart 2. Limitation of the useful beam. Subpart 3. Measuring compliance; primary protective

More information

Exposure System Selection

Exposure System Selection Principles of Imaging Science II (RAD120) Exposure Systems Exposure System Selection Radiographic exposure is a very complex process Best technique systems manipulate one variable while holding others

More information

CR Basics and FAQ. Overview. Historical Perspective

CR Basics and FAQ. Overview. Historical Perspective Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors

More information

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on

More information

make it easy, with Ray

make it easy, with Ray Lower dose - Quick scan times - Pulsed X-ray technology - Multiple scan modes 3 Dedicated detectors - Reliable performance - No damage - Long life span Easy upgrade - Ready to upgrade CBCT & Cephalometric

More information