DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING

Size: px
Start display at page:

Download "DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING"

Transcription

1 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE Rev. 1.0 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE DoseLab users may reference the following instructions to perform Tomotherapy Quality Assurance tests as recommended in AAPM Report of the Task Group Report 148: QA for helical tomotherapy. Each section refers to the corresponding TG-148 heading/text to describe how the tests should be analyzed utilizing the existing DoseLab modules. Familiarity with TG-148 is highly recommended. Prior to completing these tests, the example Tomotherapy preference files should be imported into DoseLab. TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING The machine isocenter must be within 0.5 cm based on the recommended image of the Task Group. Please see Detailed QA Test Descriptions for the instructions to perform this test. 2. V B.1.D. - Y-JAW/GANTRY ROTATION PLANE ALIGNMENT The slope of each line on the recommended image must not vary by more than 5 degrees. Obtain the images as described in the Task Group report and follow these steps to analyze: 1. Open the *.tif image in DoseLab s Single Image Analysis module 2. Crop the image to include only the exposure created at 50 cm SID 3. Open the MLC Strip Test tool via Machine QA MLC Strip test 4. Manually enter an SID of 50 cm 5. In the Tolerances dropdown, select Edit 6. Highlight the V.B.1.d. y-jaw/gantry rotation plane alignment Tolerance set 7. Change the Slope of lines baseline value to match the current slope displayed in the Results window 8. Select OK to save the changes 9. Save these baseline results to the Database (Results Save to PDF or database) and exit the MLC Strip Test tool 10. Re-open the raw, un-cropped image and crop the image again this time only include the exposure at 110 cm SID 11. Manually enter an SID of 110 cm 12. Choose the V.B.1.d. y-jaw/gantry rotation plane alignment tolerance set from the Tolerance dropdown 13. Verify that the slope of the 110 cm SID image is within 5 degrees of the 50 cm SID baseline (set in step 4). 3. V B.1.E. - TREATMENT FIELD CENTERING The center of the described fields must be within 0.5 mm. Please see Detailed QA Test Descriptions for further detail. 4. V.B.1.F. - MLC ALIGNMENT TEST The alignment verification must verify that the MLC offset is less than 1.5 mm and 0.5 degrees at isocenter. Take an image as described in the Task Group Report and follow these instructions to analyze the resulting image: 1. Open the *.tif image in DoseLab s Single Image Analysis 2. Open the MLC Strip Test Tool via Machine QA MLC Strip test 3. Manually define an SID of 85 cm 4. Open the V.B.1.f. MLC alignment test tolerance set, verify the twist slope & strip distances, and save to the Database a. Overall twist is multiplied by a factor of 2 in this case; therefore the result should be divided by 2 for MLC twist. Since this is a baseline value, it is acceptable to monitor the total twist if the user understands this parameter. 1

2 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE Rev V.B.3.A. - GANTRY ANGLE CONSISTENCY The gantry angles must be accurate within 1 degree. For image analysis, follow these steps: 1. Open first image in DoseLab s Single Image Analysis 2. Run the Starshot tool via Machine QA Starshot 3. Under Preferences ensure that Display angle of each spoke is selected 4. Record angles in Database (must be performed manually via Custom Tests, as angle results are not automatically saved) 5. Repeat for any additional starshot images 6. V.B.3.B. - COUCH SPEED UNIFORMITY The dose delivered along the couch s axis of travel must vary by less than 2 % when subjected to couch motion. Delivery shall be based on TG148 and clinically relevant parameters. Follow these instructions to analyze the resulting image: 1. Open the *.tif image in DoseLab s Single Image Analysis 2. Open the Flatness and Symmetry tool via Machine QA Flatness and symmetry 3. Select the V.B.3.b. Couch speed uniformity tolerance set 4. Select the appropriate baseline for comparison 5. Verify the relative optical density of along the couch travel axis does not vary by 2% and save results to the Database 7. V.B.3.C. - COUCH TRANSLATION/GANTRY ROTATION Consistent spacing of exposures during gantry rotation & couch movement is verified with this test. To analyze the resulting image: 1. Open the *.tif image in DoseLab s Single Image Analysis 2. Open the MLC Strip Test Tool via Machine QA MLC Strip test 3. Manually define an SID of 85 cm 4. Open the V.B.3.c. Couch translation/gantry rotation tolerance set, verify the results, and save them to the Database 8. V. B.4.A - INTERRUPTED TREATMENT PROCEDURE Standard and interrupted treatments must differ by no more than 3% in dose and 1 mm in the y direction. Follow these instructions to analyze the image for this test: 1. Convert both films to OD by using the Flatbed Scanners tool 2. Open the Dose Comparison Interface 3. Change gamma criteria to 3%/1 mm 4. Run a Film to Film comparison 9. V.B.4.B. LASER LOCALIZATION The lasers must be within 1 mm of the radiation field. Please see Detailed QA Test Descriptions for further detail. 10. V.I.B.2 IMAGE QUALITY Image quality constancy for the MVCT is verified with this recommended assessment. Analyze the resulting MVCT of this test by: 1. Open MVCT images in the DoseLab CT/CBCT tool 2. Select appropriate slice(s) by manually entering the slice number, verify the results, and save results to the Database 2

3 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE v. 1.0 DETAILED QA TEST DESCRIPTIONS Prior to performing these tests, the user should manually copy the example Tomotherapy both the MLC Strip Test tolerance.mat and CBCT phantoms.xml files into their Preferences folder (where DoseLab is installed) to populate the preset Tolerance sets and Tomotherapy-related phantoms. These coped files will overwrite the default tolerance set, so if there is any customization already performed by the user it is suggested that they first contact Mobius Medical Systems Customer Support prior to copying these files. Alternatively, advanced users may add these Tolerances sets manually if preferred. For guidance on film scanning with Epson scanners, see Appendix C of the DoseLab User Manual. 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING 1. Take an image as described in the Task Group Report - verification that the center of the described fields is within 0.5 mm can be completed by obtaining two separate vertical profile measurements and subtracting the central position of each respective profile s FWHM 2. Using the scanned image, open the image in Single Image analysis 3. Load the Profiler tool via Analysis Image profiler and export two profiles that are 0.5 cm from the field convergence in the image, demonstrated by the solid and dotted lines on the image below 4. Select the appropriate Profile type based on the image orientation (if image is in the orientation shown below, choose Vertical profile ) 5. Select Save profile to save the data in an Excel spreadsheet format 6. Repeat this for both images 7. Once in Excel, utilizing the Distance and Image value columns from each spreadsheet, normalize the image value column to the maximum value for that column 8. From this conversion, determine the FWHM value and use the appropriate inverse square scaling to determine the offset. For an example of this test results, see sheet labeled V.B.1.C. in DoseLab - TG148 Example.xls. Figure 1: V.B.1.C. Image Display in DoseLab s Image Profiler Module

4 2. V B.1.D. - Y-JAW/GANTRY ROTATION PLANE ALIGNMENT DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE Rev Following the Task Group report, obtain two images: a. One at 50cm SID b. And one at 110cm SID 2. After scanning, manually split the image into two separate images via DoseLab s Single Image Analysis interface: a. One containing only the 50cm SID portion of the image b. And another containing only the 110cm SID portion of the image. 3. Load the 50cm SID image into the MLC Strip Tool Single Image Analysis 4. Open the MLC Strip Tool via Machine QA MLC Strip test 5. Find the slope of the line in the Results window and set this value as the test s Baseline value by selecting V.B.1.d. Y- Jaw/Gantry Rotation Plane Alignment from the Tolerances dropdown 6. Subsequently, load the 110cm SID image and cross-compare the line slope from the 50cm SID baseline value. 7. Show that the line slope does not deviate by more than 0.5 degrees. Note that the warn value should be set to 0.25, as the physical jaw twist is half of the angle between the fields as measured on the film. If utilizing the provided Example Tomotherapy Tolerance Set, this will be preset as the value. Figure 2: V.B.1.D. Image Setup in DoseLab s MLC Strip Test Module 4

5 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE v V B.1.E. - TREATMENT FIELD CENTERING Verifying that the center of the fields is within 0.5 mm can be performed by taking several profile measurements of the TG-148 recommended image and finding the central fit line displayed by a linear regression. A single vertical profile going through each exposure along with a single horizontal profile will be sufficient information. 1. Capture the images as instructed in TG-148 and scan them for analysis 2. Open the exposure in Single Image Analysis 3. Using the Image Profiler tool (Analysis Image profiler), capture a vertical profile though each exposure a. After all profile captures, you should have 8 total: i. 7 vertical profiles, 1 through each of the 7 individual exposures ii. 1 horizontal profile that traverses all exposures at once 4. Select Save profile to save the data in an Excel spreadsheet format one full horizontal profile and one vertical profile for each individual exposure 5. Determine the FWHM of the vertical profiles to find the center of each region 6. Optional: If preferred, the resulting data could be used to calculate a linear regression for overall pass of this test. It would be determined by how close the value of the results is to 1. Failing results would likely be ±0.1 from the nominal value, but this level should be determined by your clinically relevant data. This would be user-calculated. For an example of this test results, see sheet labeled V.B.1.E. in DoseLab - TG148 Example.xls. Figure 3: V.B.1.E. Image Display in DoseLab s Image Profiler Tool

6 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE Rev V.B.1.F. - MLC ALIGNMENT TEST 1. Obtain an image as described in TG the MLC offset should be less than 1.5 mm at the isocenter and the MLC twist should be less than 0.5 (or per vendor specifications, as they may change) 2. After scanning the image, load the scanned image into Single Image Analysis 3. Open the MLC Strip Tool via Machine QA MLC Strip test 4. Once in the MLC Strip tool, proceed to Preferences and uncheck Perform Lorentzian fit to utilize a Gaussian fit 5. Ensure that the provided example Tomotherapy Tolerance Set reflects the appropriate Baseline and Warn/Fail tolerances as described in TG-148, or as is clinically relevant by selecting, V.B.1.f. MLC Alignment Test from the Tolerances dropdown 6. Manually enter the SID 7. The resulting Distance between lines and Slope of lines should supply the necessary data, as recommended in the Task Group Note that the overall twist is magnified by a factor of 2 and should therefore be divided by 2 to obtain true MLC twist results. Uncheck Perform Lorentzian fit Manually define the SID Figure 4: V.B.1.F. Image Setup in DoseLab s MLC Strip Test Module 6

7 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE v V.B.3.A. - GANTRY ANGLE CONSISTENCY The control sinogram for this test is set to open the middle two leaves 32 and 33 at projections centered at 0, 120, and 240. Using a horizontal line marked on the films during setup, the resulting star pattern can be checked for the correct initial angles at the start of treatment and the ability to reproduce this pattern after 24 rotations. 1. Take the images as directed in the Task Group report and scan the image 2. Load the test s recommended image into Single Image Analysis 3. Open the Starshot Tool via Machine QA Starshot 4. Under Preferences ensure that Display angle of each spoke is selected 5. The resulting gantry angles should be reproducible with a tolerance of 1 6. Optional: Enter the results into the V.B.4.B. sheet the DoseLab TG148 Example spreadsheet. Import the spreadsheet values into the Database. For an example of this test built in the Database, see the example in Machine B. Note that the angle values are not directly saved to the database and will need to be entered via a user-created Custom Test. For information on creating custom test, see the DoseLab User Guide. Angles are not automatically saved to the database Figure 5: V.B.3.A. Image Angle Results Display in DoseLab s Starshot Module

8 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE Rev V.B.3.B. - COUCH SPEED UNIFORMITY 1. The film is irradiated with couch movement involved as instructed in TG-148 and scanned 2. With the resulting scan, load the image into Single Image Analysis 3. Open the Flatness and Symmetry Tool via Machine QA Flatness and Symmetry 4. Proceed to Preferences and select these options: a. Apply 5x5 Weiner filter b. Only analyze in vertical profile c. Differences from baseline profiles Calculate maximum differences 5. The result is a profile generated along the axis of couch travel 6. The relative optical density (image value) along this line should vary by less than 2% a. Observe the Max vertical diff from base value for this metric Filter for film noise Only relevant profile Compare for constancy. Figure 6: V.B.3.B. Image Display in DoseLab s Flatness and Symmetry Module 8

9 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE v V.B.3.C. - COUCH TRANSLATION/GANTRY ROTATION 1. The resulting film from the Task Group is exposed and scanned 2. Load the image into Single Image Analysis 3. Open the Flatness and Symmetry Tool via Machine QA Flatness and Symmetry 4. Ensure that the provided example Tomotherapy Tolerance Set reflects the appropriate Baseline and Warn/Fail tolerances as described in TG-148, or as is clinically relevant by selecting, V.B.3.c Couch Translation/Gantry Rotation from the Tolerances dropdown 5. The resulting profile should show maxima 5 cm apart to within 1 mm Match exactly with TG148 Figure 7: V.B.3.C. Image Display in DoseLab s Flatness and Symmetry Module

10 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE Rev V.B.4.A - INTERRUPTED TREATMENT PROCEDURE 1. Acquire the necessary films per protocol instruction and scan the films accordingly 2. Convert both scanned films to optical density (OD) by using the Flatbed Scanners tool a. Load an unexposed film image (or select Convert the film to OD without applying uniformity correction if desired) b. Choose the appropriate color channel based on the recommended dose levels c. Select Exposed film image and navigate to the the exposed film d. Choose Process image in the Select Files window to save the file in OD values e. The OD-converted image will save to the same location as the non-converted image was located. For example, if the non-converted film was saved as C:\DoseLab\Measured, the resulting OD film would be named C:\DoseLab\Measured-OD f. Ensure this is repeated for the second film g. Further detail on this step may be found in the Section of the DoseLab User Manual 3. After the image is converted to OD, open the Dose Comparison Interface from DoseLab s main menu 4. Under Preferences Default comparison values, change the gamma criteria to 3%/1mm and select OK 5. In the Dose Comparison Preset dropdown, select Film to film 6. Load one of the converted films as the Calculated distribution and the other as the Measured distribution a. The order of selection is not significant as long as the user is aware of what each film has been labeled 7. Choose Next from the interface to proceed with a film to film comparison a. Details on this type of comparison, which each step entails, as well as all automation provided by the Dose Comparison Interface can be found in Section 4.4 of the DoseLab User Manual 8. When overlaid, the interrupted treatment should differ from the completed procedure by no more that 3% in its delivered dose and the overall length FWHM of the dose distribution in the y-direction should differ by no more than 1 mm. Viewing the Gamma image will provide graphical representation of this comparison. 9. Optional: Enter the results into the V.B.4.A. sheet of the DoseLab TG148 Example spreadsheet. Import the spreadsheet values into the Database. For an example of this test built in the Database, see the example in Machine B. Figure 8: V.B.4.A. Gamma Display in DoseLab s Dose Comparison Module 10

11 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE v V.B.4.B. - LASER LOCALIZATION 1. Obtain images as instructed in TG-148 recommendations 2. Images should be scanned in for analysis if necessary, see Appendix C in the DoseLab User Manual for film scanning recommendations with Epson scanners 3. In order to verify laser localization accuracy within 1 mm, several profiles will need to be taken over the image. These may be obtained using the Image Profiler tool in Single Image analysis (Analysis Image profiler) 4. In each instance, select Save profile to save the data in an Excel spreadsheet format 5. First, open the sagittal laser image and save four vertical profiles: a. Two directly over of the sagittal laser fiducials (as shown with the dotted white line below) b. And two along the exposure just next to the outer fiducials (as shown with the solid white line below) Figure 9a: V.B.4.B. Image Display in DoseLab s Image Profiler Tool with Vertical Profile Markings 6. Next, two horizontal profiles should be taken, one across each of the axial laser fiducials, as shown by the dotted and solid white lines below Figure 9b: V.B.4.B. Image Display in DoseLab s Image Profiler Tool with Horizontal Profile Markings 7. The profiles through the exposure are used to determine the angular displacement of the image itself by performing a linear regression from the center of each respective FWHM. The sagittal marks profiles can then be corrected for any slope to verify the 1mm positioning accuracy. 8. Next, the two horizontal profiles overlaying the axial laser marks will provide information to determine the positioning and demonstrate any angular discrepancy. The value should be within 1 degree of a total 90 degree offset from the sagittal marks. 9. Finally, the distance measurement tool in DoseLab provides the information for the movable laser positioning. To utilize this tool, open the image in Single Image Analysis and select Analysis Distance measurement. For an example of this test results, see sheet labeled V.B.4.E. in DoseLab - TG148 Example.xls.

12 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE Rev V.I.B.2 - IMAGE QUALITY 1. Obtain the MVCT images as described in TG From DoseLab s main interface, choose CT / CBCT 3. When prompted, navigate to the location of the images for analysis 4. Utilize the CBCT module with manual slice detection by ensuring that Preferences Autodetect slices at startup is unchecked 5. Select Tomo Test from the Phantom dropdown 6. Choose View CT scan to select the appropriate slice and then close the viewer to enter this value into the Slice box 7. Hit Process to obtain results for CNR, Uniformity, Spatial Resolution, Contrast, Scaling Discrepancy, Geometric Distortion. 8. These results may be directly saved to the DoseLab database. Manual Slice selection needed for MVCT Figure 20: V.I.B.2. CBCT Display in DoseLab s CT/CBCT Module with ROIs Overlaid 12

Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film

Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 4, 2016 Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film Eric C. Lobb Department

More information

GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications

GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications I. SCOPE The protocol applies to GafChromic EBT3P and EBT3+P films exposed in GafChromic QuiCk Phantom

More information

Nathan Childress, Ph.D., DABR

Nathan Childress, Ph.D., DABR Nathan Childress, Ph.D., DABR Introduction TG-142 is a comprehensive QA protocol Covers nearly every aspect of machine and safety QA Recommends quantitative results Recommends high testing frequencies

More information

ISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM

ISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM FILMQA STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE Supports all major radiotherapy technologies! FilmQA is optimized for use with Gafchromic film products, including EBT2 and RTQA2. FILMQA helps

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup Huaiqun Guan,

More information

COMPREHENSIVE TG-142 IMAGING AND MACHINE QA

COMPREHENSIVE TG-142 IMAGING AND MACHINE QA QA SOFTWARE COMPREHENSIVE TG-142 IMAGING AND MACHINE QA Automate the analysis of over thirty TG-142 recommended QA tasks The rapid progress of Radiation Therapy has created the need for Quality Assurance

More information

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

3D Diode Array Commissioning: Building Confidence in 3D QA Technology 3D Diode Array Commissioning: Building Confidence in 3D QA Technology Caroline Yount, MS CANCER CENTER 3D QA The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose

More information

GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom

GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom Introduction: These EBT2/EBT3 films, shown in Figure 1a-c, are specially sized and formatted to uniquely fit the Accuray Ball Cube II Phantom. Each

More information

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System.

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System. Use of EPIDs for Non-Routine Linac QA Bin Cai PhD Disclosure Parts of this project received support from Varian Medical System. Learning Objectives Learn the recent development of EPID based Non-routine

More information

PHYSICS QUESTIONNAIRE FORM

PHYSICS QUESTIONNAIRE FORM PHYSICS QUESTIONNAIRE FORM Institution Name: Date: Contact Information (name, address, phone, fax, email): Physicist: Radiation Oncologist: Dosimetrist (if applicable): Study Coordinator (if applicable):

More information

A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine

A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine Iwein Van de Vondel, 1 Koen Tournel, 1 Dirk

More information

IMRT Delivery System QA. IMRT Dose Delivery. Acceptance testing. Why: specific tests for IMRT? Accuracy of leaf positioning (gaps) MLC Alignment

IMRT Delivery System QA. IMRT Dose Delivery. Acceptance testing. Why: specific tests for IMRT? Accuracy of leaf positioning (gaps) MLC Alignment 1 IMRT Delivery System Q Thomas LoSasso, PhD Memorial Sloan Kettering Cancer Center IMRT Dose Delivery cceptance testing Commissioning Quality assurance Verification Q Why: specific tests for IMRT? 2.

More information

Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator

Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator Aleksei E. Zhdanov 1 and Leonid G. Dorosinskiy 1 Ural Federal University named after the first President of Russia B. N.

More information

7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators

7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators Implementing a new digital medical accelerator John Wong Johns Hopkins University AAPM, Austin, 2014 Acknowledgements Yin Zhang, Ken Wang, Kai Ding (Commissioning - JHU) Esteban Velarde, Joe Moore (QA

More information

Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA

Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA Arthur Olch, PhD, FAAPM AAPM Spring Clinical Meeting, March 21, 2017 Or.. What Dose are the Patients Really Getting???

More information

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients

More information

T h e P h a n t o m L a b o r a t o r y

T h e P h a n t o m L a b o r a t o r y T h e P h a n t o m L a b o r a t o r y 1 CCT228 ATCM Phantom Manual Copyright 2017 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order

More information

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools SRS MapCHECK SRS Patient QA, No Film Your Most Valuable QA and Dosimetry Tools SRS Patient QA, No Film With improvements in targeting and localization, stereotactic treatments have become prevalent. To

More information

Nuclear Associates , , CT Head and Body Dose Phantom

Nuclear Associates , , CT Head and Body Dose Phantom Nuclear Associates 76-414,76-414-4150,76-415 CT Head and Body Dose Phantom Users Manual March 2005 Manual No. 76-414-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product

More information

FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May Carl R. Keener, Ph.D., DABMP, DABR

FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May Carl R. Keener, Ph.D., DABMP, DABR FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May 2010 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics, Inc. Physicist's Role in the

More information

Tomophan TSP004 Manual

Tomophan TSP004 Manual T h e P h a n t o m L a b o r a t o r y 1 Tomophan TSP004 Manual Copyright 2016 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order and

More information

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc.

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. An Introduction to TG-142 Imaging QA Using Standard Imaging Products Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. Goals Understand the nature and intent of TG 142 imaging

More information

CT Scanner Dose Survey

CT Scanner Dose Survey CT Scanner Dose Survey Measurement Protocol Version 5.0 July 1997 Co-ordinated by ImPACT and The Medical Physics Department,, London SW17, UK. 0181-725-3366 CT Scanner Dose Survey: Measurement Protocol

More information

TECHNICAL NOTE. The following information is provided as a service to our users and customers:

TECHNICAL NOTE. The following information is provided as a service to our users and customers: page 1 of 5 The following information is provided as a service to our users and customers: The star shot is a standard method for measuring the stability of rotation of the linear accelerator (linac) gantry

More information

Planmeca Romexis. quick guide. Viewer EN _2

Planmeca Romexis. quick guide. Viewer EN _2 Planmeca Romexis Viewer quick guide EN 10029550_2 TABLE OF CONTENTS 1 START-UP OF PLANMECA ROMEXIS VIEWER...1 1.1 Selecting the interface language... 1 1.2 Selecting images...1 1.3 Starting the Planmeca

More information

QA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion

QA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion Leksell Gamma Knife Perfexion QA Considerations Paula L. Petti, Ph.D. Taylor McAdam Bell Neuroscience Institute Washington Hospital Healthcare System Fremont, CA 1 QA for LGK Perfexion : : Follow NRC licensing

More information

The most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 BUSINESS UNIT OF ISP

The most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 BUSINESS UNIT OF ISP The most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 GAFCHROMIC EBT dosimetry film Designed and optimized for ALL RT procedures Can be cut into

More information

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity) Vascular Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction In recent years, digital cardiovascular

More information

Using Radiochromic Film Tips and Techniques

Using Radiochromic Film Tips and Techniques Using Radiochromic Film Tips and Techniques David F. Lewis, Ph.D. Senior Science Fellow Advanced Materials Group International Specialty Products dlewis@ispcorp.com October 22, 2010 Topics Scanners and

More information

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017 The Current State of EPID-Based Linear Accelerator Quality Assurance Timothy Ritter, PhD, DABR, FAAPM 1 Disclosures Employed by the Veterans Health Administration Faculty appointment with the University

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

Overview of Safety Code 35

Overview of Safety Code 35 Common Quality Control Procedures for All s Quality Control Procedures Film All s Daily Quality Control Tests Equipment Warm-up (D1) According to manufacturers instructions Can include auto calibration(d1)

More information

Development of new dosimeter for measuring dose distribution in CT

Development of new dosimeter for measuring dose distribution in CT Development of new dosimeter for measuring dose distribution in CT Poster No.: C-2925 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology - Without Subtopic Authors: Y. Muramatsu, K.

More information

User Manual Veterinary

User Manual Veterinary Veterinary Acquisition and diagnostic software Doc No.: Rev 1.0.1 Aug 2013 Part No.: CR-FPM-04-022-EN-S 3DISC, FireCR, Quantor and the 3D Cube are trademarks of 3D Imaging & Simulations Corp, South Korea,

More information

Guidance on Using Scanning Software: Part 5. Epson Scan

Guidance on Using Scanning Software: Part 5. Epson Scan Guidance on Using Scanning Software: Part 5. Epson Scan Version of 4/29/2012 Epson Scan comes with Epson scanners and has simple manual adjustments, but requires vigilance to control the default settings

More information

3D Brachytherapy with Afterloading Machines

3D Brachytherapy with Afterloading Machines 3D Brachytherapy with Afterloading Machines 3D Brachytherapy/MS Page 1 Introduction 3D-Brachytherapy refers to the case when the planning is performed based on a set of CT, MR or UltraSound (US) images.

More information

State of the Art Film Dosimetry

State of the Art Film Dosimetry State of the Art Film Dosimetry Micke A., Lewis D. Advanced Materials Ashland proprietary technology, patents pending Film Dosimetry Radiochromic Film EBT2/EBT3 One-Scan Protocol Multi-channel Film Dosimetry

More information

Evaluation of a diode array for QA measurements on a helical tomotherapy unit

Evaluation of a diode array for QA measurements on a helical tomotherapy unit Evaluation of a diode array for QA measurements on a helical tomotherapy unit K. M. Langen, a S. L. Meeks, D. O. Poole, T. H. Wagner, T. R. Willoughby, O. A. Zeidan, and P. A. Kupelian Department of Radiation

More information

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter Revision Date: January 5th, 2017 www.disc-imaging.com Table of Contents Section A: Preliminary Setup Requirements... 4 Tools

More information

QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING

QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING 1 QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING QualiFormeD Phantoms A selection of test objects facilitating regulatory quality controls in radiation therapy and medical imaging Practical,

More information

Chapter 6: TVA MR and Cardiac Function

Chapter 6: TVA MR and Cardiac Function Chapter 6 Cardiac MR Introduction Chapter 6: TVA MR and Cardiac Function The Time-Volume Analysis (TVA) optional module calculates time-dependent behavior of volumes in multi-phase studies from MR. An

More information

QUALITY CONTROL SOFTWARE FOR RADIOTHERAPY AND MEDICAL IMAGING

QUALITY CONTROL SOFTWARE FOR RADIOTHERAPY AND MEDICAL IMAGING 1 QUALITY CONTROL SOFTWARE FOR RADIOTHERAPY AND MEDICAL IMAGING The QUALIMAGIQ platform and its modules An optimal software solution for automatic analyses and traceability of quality controls in radiation

More information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information Ordering Information Please contact us if you have any questions or if you would like a quote or delivery schedule regarding the Catphan phantom. phone 800-525-1190, or 518-692-1190 fax 518-692-3329 mail

More information

Volumetric Modulated Arc Therapy. David Shepard Swedish Cancer Institute Seattle, WA

Volumetric Modulated Arc Therapy. David Shepard Swedish Cancer Institute Seattle, WA Volumetric Modulated Arc Therapy David Shepard Swedish Cancer Institute Seattle, WA Disclaimer Our VMAT work has been sponsored in part by Elekta. Outline David Shepard VMAT Basics and VMAT Plan Quality

More information

Sub-mm accuracy of accelerators: How manufacturers achieve it, how physicists verify it

Sub-mm accuracy of accelerators: How manufacturers achieve it, how physicists verify it Sub-mm accuracy of accelerators: How manufacturers achieve it, how physicists verify it Presented by Ivan A. Brezovich, PhD, at the 2015 Annual meeting of the SEAAPM, Raleigh, NC, April 24, 2015 Purpose

More information

A Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017

A Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017 A Fast Monolithic System for Proton Imaging Fritz DeJongh ProtonVDA Inc October 2017 Disclosures I am a cofounder and co-owner of ProtonVDA Inc We hold intellectual property rights on our proton imaging

More information

MOVING LASER SYSTEMS IN RT

MOVING LASER SYSTEMS IN RT MOVING LASER SYSTEMS IN RT ESSENTIAL FOR PRECISE PATIENT ALIGNMENT NEXT GENERATION LAP IS YOUR IDEAL PARTNER FOR YOUR EXTERNAL LASER SYSTEMS OUR MOVING LASER SYSTEMS ARE RECOMMENDED BY LEADING CT SCANNER

More information

SKF Shaft Alignment Tool Horizontal machines app

SKF Shaft Alignment Tool Horizontal machines app SKF Shaft Alignment Tool Horizontal machines app Short flex couplings Instructions for use Table of contents 1. Using the Horizontal shaft alignment app... 2 1.1 How to change the app language...2 1.2

More information

SCANNING IMAGES - USER S GUIDE. Scanning Images with Epson Smart Panel and PhotoShop [for Epson 1670 scanners]

SCANNING IMAGES - USER S GUIDE. Scanning Images with Epson Smart Panel and PhotoShop [for Epson 1670 scanners] University of Arizona Information Commons Training 1 SCANNING IMAGES - USER S GUIDE Scanning Images with Epson Smart Panel and PhotoShop [for Epson 1670 scanners] p.2 Introduction and Overview Differences

More information

Laser Installation Engineers, CT Room Designers and Contractors

Laser Installation Engineers, CT Room Designers and Contractors Page 1 of 10 Title Room Preparation For LS Laser System Installation in CT Environments Scope Centralite LS Laser Systems Target Group Laser Installation Engineers, CT Room Designers and Contractors Page

More information

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 1, WINTER 2010 A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA Abdul Qadir Jangda,

More information

Disclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography

Disclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography Current Implementation Status of IEC Standard 62494-1: Exposure Index (EI) for Digital Radiography July 31, 2017 Ryan Fisher, PhD, DABR Katie Hulme, MS, DABR None Disclosures Outline Review of IEC Standard

More information

Imaging system QA of a medical accelerator, Novalis Tx, for IGRT per TG 142: our 1 year experience

Imaging system QA of a medical accelerator, Novalis Tx, for IGRT per TG 142: our 1 year experience JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 4, 2012 Imaging system QA of a medical accelerator, Novalis Tx, for IGRT per TG 142: our 1 year experience Zheng Chang, a James Bowsher, Jing

More information

SolidWorks 95 User s Guide

SolidWorks 95 User s Guide SolidWorks 95 User s Guide Disclaimer: The following User Guide was extracted from SolidWorks 95 Help files and was not originally distributed in this format. All content 1995, SolidWorks Corporation Contents

More information

ThermaViz. Operating Manual. The Innovative Two-Wavelength Imaging Pyrometer

ThermaViz. Operating Manual. The Innovative Two-Wavelength Imaging Pyrometer ThermaViz The Innovative Two-Wavelength Imaging Pyrometer Operating Manual The integration of advanced optical diagnostics and intelligent materials processing for temperature measurement and process control.

More information

IQM Detector Characteristics: Signal reproducibility

IQM Detector Characteristics: Signal reproducibility The Integral Quality Monitor (IQM) System is a real-time beam verification system that monitors the accuracy of radiation delivery throughout each patient treatment without any user interaction. IQM continuously

More information

Stability of the Helical TomoTherapy Hi Art II detector for treatment beam irradiations

Stability of the Helical TomoTherapy Hi Art II detector for treatment beam irradiations JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 6, 2014 Stability of the Helical TomoTherapy Hi Art II detector for treatment beam irradiations Karin Schombourg, François Bochud, Raphaël

More information

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools SRS MapCHECK SRS Patient QA, No Film Your Most Valuable QA and Dosimetry Tools SRS Patient QA, No Film As clinics strive to treat with the smallest field possible, the demand for SRS/SBRT QA grows as well.

More information

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT Quality control of Gamma Camera By Dr/ Ibrahim Elsayed Saad 242 NMT WHAT IS QUALITY? The quality of a practice is to fulfill the expectations and demands from: Patient Clinicain Your self Quality assurance

More information

ArcCHECK. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools

ArcCHECK. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools ArcCHECK The Ultimate 4D QA Solution A 4D isotropical cylindrical detector array for arc delivery QA and Dosimetry U.S.Patent No. 8,044,359; 6,125,335 Compatible with: FFF Beams VMAT RapidArc TomoTherapy

More information

LSM 780 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch

More information

Instruction Manual for HyperScan Spectrometer

Instruction Manual for HyperScan Spectrometer August 2006 Version 1.1 Table of Contents Section Page 1 Hardware... 1 2 Mounting Procedure... 2 3 CCD Alignment... 6 4 Software... 7 5 Wiring Diagram... 19 1 HARDWARE While it is not necessary to have

More information

!! Select Professional Mode. !! Click on Other then Select Scratch.

!! Select Professional Mode. !! Click on Other then Select Scratch. SCANNING BASICS FOR EPSON SCANNERS When using EpsonScan software with any of the Epson flatbed scanners in Labs 1, 2, 3 or 4, the following steps must be followed in order to properly save files. Select

More information

- Water resistant. - Large size.

- Water resistant. - Large size. GAFCHROMIC EBT product brochure GAFCHROMIC EBT FEATURES GAFCHROMIC EBT dosimetry film has been developed specifically to address the needs of the medical physicist and dosimetrist working in the radiotherapy

More information

DOE Project: Resist Characterization

DOE Project: Resist Characterization DOE Project: Resist Characterization GOAL To achieve high resolution and adequate throughput, a photoresist must possess relatively high contrast and sensitivity to exposing radiation. The objective of

More information

Module 1G: Creating a Circle-Based Cylindrical Sheet-metal Lateral Piece with an Overlaying Lateral Edge Seam And Dove-Tail Seams on the Top Edge

Module 1G: Creating a Circle-Based Cylindrical Sheet-metal Lateral Piece with an Overlaying Lateral Edge Seam And Dove-Tail Seams on the Top Edge Inventor (10) Module 1G: 1G- 1 Module 1G: Creating a Circle-Based Cylindrical Sheet-metal Lateral Piece with an Overlaying Lateral Edge Seam And Dove-Tail Seams on the Top Edge In Module 1A, we have explored

More information

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook DIAGNOSTIC ACCREDITATION PROGRAM Radiology and CT Quality Control Procedures Workbook Quality Control Procedures Radiography/CR/DR Safety Code 35 Summary For more detail about each quality control (QC)

More information

CyberKnife Iris Beam QA using Fluence Divergence

CyberKnife Iris Beam QA using Fluence Divergence CyberKnife Iris Beam QA using Fluence Divergence Ronald Berg, Ph.D., Jesse McKay, M.S. and Brett Nelson, M.S. Erlanger Medical Center and Logos Systems, Scotts Valley, CA Introduction The CyberKnife radiosurgery

More information

Collimation Assessment Using GAFCHROMIC XR-M2

Collimation Assessment Using GAFCHROMIC XR-M2 Collimation Assessment Using GAFCHROMIC XR-M2 I. Introduction A method of collimation assessment for GE Senographe full-field digital mammography (FFDM) systems is described that uses a self-developing

More information

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation Image Quality Review I: Basics and Image Quality TH-A-16A-1 Thursday 7:30AM - 9:30AM Room: 16A J. Anthony

More information

Principles and Applications of Microfluidic Devices AutoCAD Design Lab - COMSOL import ready

Principles and Applications of Microfluidic Devices AutoCAD Design Lab - COMSOL import ready Principles and Applications of Microfluidic Devices AutoCAD Design Lab - COMSOL import ready Part I. Introduction AutoCAD is a computer drawing package that can allow you to define physical structures

More information

Impact of energy variation on Cone Ratio, PDD10, TMR20 10 and IMRT doses for flattening filter free (FFF) beam of TomoTherapy Hi-Art TM machines

Impact of energy variation on Cone Ratio, PDD10, TMR20 10 and IMRT doses for flattening filter free (FFF) beam of TomoTherapy Hi-Art TM machines JBUON 2014; 19(4): 1105-1110 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com ORIGINAL ARTICLE Impact of energy variation on Cone Ratio, PDD10, TMR20 10 and IMRT

More information

Physics 2310 Lab #5: Thin Lenses and Concave Mirrors Dr. Michael Pierce (Univ. of Wyoming)

Physics 2310 Lab #5: Thin Lenses and Concave Mirrors Dr. Michael Pierce (Univ. of Wyoming) Physics 2310 Lab #5: Thin Lenses and Concave Mirrors Dr. Michael Pierce (Univ. of Wyoming) Purpose: The purpose of this lab is to introduce students to some of the properties of thin lenses and mirrors.

More information

Measurement of table feed speed in modern CT

Measurement of table feed speed in modern CT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 3, 2014 Measurement of table feed speed in modern CT Atsushi Fukuda, 1,2a Pei-Jan P. Lin, 3 Kosuke Matsubara, 2 Tosiaki Miyati 2 Department

More information

Phantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging

Phantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging in partnership with Phantoms in Medical Physics (RT) U. Oelfke Division of Radiotherapy & Imaging uwe.oelfke@icr.ac.uk Making the discoveries that defeat cancer 1. Introduction What is a phantom? Wiki:

More information

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE Takeyuki Hashimoto 1), Morio Onoe 2), Hiroshi Nakamura 3), Tamon Inouye 4), Hiromichi Jumonji 5), Iwao Takahashi 6); 1)Yokohama Soei

More information

Initial setup and subsequent temporal position monitoring using implanted RF transponders

Initial setup and subsequent temporal position monitoring using implanted RF transponders Initial setup and subsequent temporal position monitoring using implanted RF transponders James Balter, Ph.D. University of Michigan Has financial interest in Calypso Medical Technologies Acknowledgements

More information

Importing and processing gel images

Importing and processing gel images BioNumerics Tutorial: Importing and processing gel images 1 Aim Comprehensive tools for the processing of electrophoresis fingerprints, both from slab gels and capillary sequencers are incorporated into

More information

Agilent 10774A Short Range Straightness Optics and Agilent 10775A Long Range Straightness Optics

Agilent 10774A Short Range Straightness Optics and Agilent 10775A Long Range Straightness Optics 7Y Agilent 10774A Short Range Straightness Optics and Agilent 10775A Long Range Straightness Optics Introduction Introduction Straightness measures displacement perpendicular to the axis of intended motion

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

Be aware that there is no universal notation for the various quantities.

Be aware that there is no universal notation for the various quantities. Fourier Optics v2.4 Ray tracing is limited in its ability to describe optics because it ignores the wave properties of light. Diffraction is needed to explain image spatial resolution and contrast and

More information

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Gena M.A.H 1, Ahmed L.El-Attar 2, Elbadry M. Zahran 3, Hany El-Gamal

More information

Improving bar code quality

Improving bar code quality Improving bar code quality The guidance documented here is intended to help packaging designers and printers achieve good quality printed bar codes on their packaging and products. This advice is particularly

More information

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 710 Confocal Microscope Standard Operation Protocol LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.

More information

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter MR-guided RT in Heidelberg Funded by the German Research Foundation (DFG) Heidelberg consortium received

More information

CHROMACAL User Guide (v 1.1) User Guide

CHROMACAL User Guide (v 1.1) User Guide CHROMACAL User Guide (v 1.1) User Guide User Guide Notice Hello and welcome to the User Guide for the Datacolor CHROMACAL Color Calibration System for Optical Microscopy, a cross-platform solution that

More information

ArcCHECKTM. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools. VMAT RapidArc TomoTherapy Pinnacle 3 SmartArc Conventional IMRT

ArcCHECKTM. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools. VMAT RapidArc TomoTherapy Pinnacle 3 SmartArc Conventional IMRT TM The Ultimate 4D QA Solution A 4D isotropical cylindrical detector array for arc delivery QA and Dosimetry U.S.Patent No. 8,044,359 What is? The world s first true 4D detector array The world s first

More information

VTube-LASER Quick Start Guide

VTube-LASER Quick Start Guide VTube-LASER Quick Start Guide This guide shows how to import a STEP file and then MEASURE and qualify demo tube 4 using the standard UNISCAN method of measuring. The steps in this workflow are from version

More information

Sensitivity study of an automated system for daily patient QA using EPID exit dose images

Sensitivity study of an automated system for daily patient QA using EPID exit dose images Received: 27 June 2017 Revised: 8 December 2017 Accepted: 27 January 2018 DOI: 10.1002/acm2.12303 RADIATION ONCOLOGY PHYSICS Sensitivity study of an automated system for daily patient QA using EPID exit

More information

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 David F. Lewis Advanced Materials Group Ashland Specialty Ingredients Spain, April 2014 What is Radiochromic Film? A film that instantly changes

More information

The Ultimate 4D QA Solution A 4D isotropic cylindrical detector array for arc delivery QA and Dosimetry.

The Ultimate 4D QA Solution A 4D isotropic cylindrical detector array for arc delivery QA and Dosimetry. The Ultimate 4D QA Solution A 4D isotropic cylindrical detector array for arc delivery QA and Dosimetry. U.S.Patent No. 8,044,359; 6,125,335 Your Most Valuable QA and Dosimetry Tools 2 Y o u r M o s t

More information

An Introduction to Dimensioning Dimension Elements-

An Introduction to Dimensioning Dimension Elements- An Introduction to Dimensioning A precise drawing plotted to scale often does not convey enough information for builders to construct your design. Usually you add annotation showing object measurements

More information

Diagnostic X-Ray Shielding

Diagnostic X-Ray Shielding Diagnostic X-Ray Shielding Multi-Slice CT Scanners Using NCRP 147 Methodology Melissa C. Martin, M.S., FAAPM, FACR Therapy Physics Inc., Bellflower, CA AAPM Annual Meeting, Orlando, FL FL Refresher Course

More information

Commissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests

Commissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests Acknowledgements David Shepard, Ph.D. Daliang Cao, Ph.D. Muhammad K. N. Afghan, Ph.D. Jinsong Ye, M.S. Tony P. Wong, Ph.D. Fan Chen, Ph.D. Min Rao, Ph.D. Vivek Mehta, M.D. Igor Gomola, Ph.D. David Housley

More information

User Manual for HoloStudio M4 2.5 with HoloMonitor M4. Phase Holographic Imaging

User Manual for HoloStudio M4 2.5 with HoloMonitor M4. Phase Holographic Imaging User Manual for HoloStudio M4 2.5 with HoloMonitor M4 Phase Holographic Imaging 1 2 HoloStudio M4 2.5 Software instruction manual 2013 Phase Holographic Imaging AB 3 Contact us: Phase Holographic Imaging

More information

A Beam-Level Delivery Accuracy Study of the Robotic Image Guided Radiosurgery System Using a Scintillator/CCD Phantom

A Beam-Level Delivery Accuracy Study of the Robotic Image Guided Radiosurgery System Using a Scintillator/CCD Phantom A Beam-Level Delivery Accuracy Study of the Robotic Image Guided Radiosurgery System Using a Scintillator/CCD Phantom Lei Wang 1, Shi Liu 1, Brett Nelson 2 1. Department of Radiation Oncology, Stanford

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

QC by the MPE in Belgium

QC by the MPE in Belgium Acceptance testing of state-of-the-art CT scanners using a new national protocol: first experience on a large number of scanners of different make and model the working group Radiology of the Belgian Hospital

More information