QA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion

Size: px
Start display at page:

Download "QA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion"

Transcription

1 Leksell Gamma Knife Perfexion QA Considerations Paula L. Petti, Ph.D. Taylor McAdam Bell Neuroscience Institute Washington Hospital Healthcare System Fremont, CA 1 QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Basic Tests and Measurements: 1) Coincidence of the mechanical isocenter of the patient-positioning system (PPS) with the radiation-focal point (RFP) 2) Agreement of measured beam profiles with Leksell GammaPlan calculations for all collimator sizes in the XY, YZ and XZ planes 3) Measurement of the absolute dose-rate calibration for largest collimator ) Confirmation of the relative output factors (ROFs( ROFs) ) for smaller collimators. 2 Leksell Gamma Knife Perfexion The patient is positioned via precise couch motions A frame adapter attaches the Leksell coordinate frame (affixed to the patient s skull) to the treatment couch Collimation system is built into the unit 3 Collimator sizes: 16-mm, 8-mm and -mm Co sources 60 Co sources not fixed in space: they reside on 8 moveable sectors Collimator Patient Frame Adapter Patient Couch: 3-axis positioning system 3 1

2 8 independent, identical sectors 2 sources per sector Perfexion Sector Design Sectors slide back and forth on outside of collimator Sector Drives Sources are arranged in 5 rings The sources in each sector can be aligned with a different collimator size or blocked completely Coincidence of PPS and RFP for Perfexion No helmets or microswitches to check Collimator settings (, 8 and 16) are independent Therefore, we must check the coincidence of the PPS and RFP for each collimator Elekta uses: Master Diode Tool New Film Holder Tool We use: Diode Test Tool (-mm collimator only) New or old Film Holder 5 Alignment of Patient-Positioning System (PPS) with Radiation Focal Point (RFP) Master Diode Tool: Service instrument used during bi-annual preventative maintenance New Film Holder: Service and Field instrument used for annual QA and acceptance testing Old-Style Film Holder: Field instrument: used for annual QA and acceptance testing Diode Tool: Field instrument used for routine checks (at least monthly, but is usually done 6more often) 2

3 Master Diode Tool: Used by Elekta during commissioning to align the RFP to the PPS Attaches directly to the PPS Accomodates up to 5 diodes There are programmed scannning sequences for all three collimators Diode Test Tool: Used by GK physicist for routine QA Attaches to patient frame adapter, which then attaches to the PPS One central diode There is one programmed scannning sequence for the -mm collimator No scanning sequences for other collimators. 7 Film Holders New Model Old Model Dosimetry adapter More precise than old-style film holder because: 1) Uses the same frame adapter that is used for patient treatments (not shown above) which is machined to very high tolerance 2) Has larger film compartment easier to interpret 16-mm films 1) Requires special dosimetry adapter, which is not machined to the same high tolerances as the patient frame adapter 2) Provides QA for RFP reproducibility, but does not provide QA for patient setup, since patient adapter is not employed 8 EBT GAFCHROMIC Film Manufactured by International Specialty Products (ISP, Wayne, NJ) Excellent white paper available at the manufacturer s website: Main advantage over earlier GAFCHROMIC films for LGK dosimetry: More sensitive to lower radiation doses: EBT films can measure doses in the range of 0.01Gy to 8 Gy Low energy dependence (<5% between MeV and kev photons) Film density growth complete within < 2 hrs 9 3

4 Basic Requirements for EBT Film Analysis Peak absorbance of EBT film is 636 nm, response enhanced by measurement with red light (match absorbance of film with spectral response of scanner) Flatbed scanners: Epson and Microtek scanners are designed to scan color film According to the ISP white paper, Epson scanners produce the optimum results with EBT films 10 Basic Requirements for EBT Film Analysis Software ImageJ: Available for free download at the website: Microsoft Excel More sophisticated software can be employed 11 EBT Film Handling Must keep track of film orientation: Before cutting film into appropriate size pieces, number the same corner on each sub-section of film: When analyzing films with flatbed scanner, the amount of scattered light depends on the orientation of the active particles in the film, which depends on the direction in which the film was coated Film coating for EBT film is parallel to the short edge of film For RFP PPS coincidence measurements, cut film to fit holder For Profile and ROF measurements cut film into pieces between about 6.0 cm and 6.5 cm 12

5 EBT Film Handling (continued) Place film in holder such that the marked corner is in the same place for all films When scanning, place films in the same area of the flatbed scanner with the marked corner of the film in the same place Avoid fingerprints and mechanical stress Avoid long exposures to daylight 13 Recommended Procedures for GK EBT Film Measurements To obtain background subtraction for each piece of cut film, pre-scan each film on flatbed scanner prior to irradiation. Scanning parameters: Scan as positive film Use either 200 or 00 dpi to obtain resolution of the order of 0.1 mm (0.127 mm for 200 dpi and 0.06 mm for 00 dpi) 8-bit color Save as PNG files (or equivalent lossless format) 1 RFP - PPS Coincidence Measurements Not necessary to convert readings to absolute dose (i.e., calibration curve not required), since we are only looking at symmetry about the pin-point. If using ImageJ software to analyze films Extract red channel (Image color RGB split) Invert image (Edit Invert) (so that profile will be positive - personal preference) Using line drawing tool, draw line through center of scan Extract Profile (Analyze Plot Profile) List Profile Data, select it and copy to Excel (or other) worksheet Subtract background 15 5

6 Coincidence of RFP and PPS: Film Analysis with ImageJ Software Raw Data Red Channel Extracted Red Channel Inverted -mm irradiation Old-style film holder XZ Plane Background Reading for Film # 16 Coincidence of RFP and PPS: Film Analysis with ImageJ Software Reading for Background Subtraction Using Ellipse Tool, outline central portion un-irradiated film Under Analyze options, select Histogram Background reading for this film = mean value in central region of unirradiated film 17 Coincidence of RFP and PPS: Film Analysis with ImageJ Software Sup Use Line tool to indicate axis for profile Under Analyze select Plot Profile L R Inf Note: Draw line in the direction of increasing LGK coordinate, in this case from superior to inferior List, select and copy the data to Excel (or other) worksheet 18 6

7 Coincidence of RFP and PPS: Film Analysis with ImageJ Software Identify channel containing pin point By definition, this corresponds to X, Y or Z = 100 in LGK coordinate system Assign LGK coordinates to other channels by the formula: LGK Coord = 100 " (C 100 " C) # P Where, C 100 = the channel that contains the pin point C = the channel in question P = conversion factor, e.g., mm/pixel for films scanned at 00 dpi 19 -mm Z Axis: Coincidence of PPS and RFP Counts (Background Subtracted) Central Channel (pin point) defines center of LGK coordinate system (magenta line) Compute the channel value for the center of the FWHM from the endpoints of the horizontal line defining the FWHM (green line) mm Z axis from XZ film PPS Center Center of FWHM FWHM!Z = mm = (center of FWHM) Z Coordinate of PPS (mm) 20 Example of Pin-Point Films to Check RFP - PPS Coincidence for 16-mm Collimator New-Style Film Holder Old-Style Film Holder Y Z X Y 21 7

8 New Film Holder: Example Pin-Point Test of Coincidence of PPS and RFP, New Film Holder 10 Reading (background subtracted) !X = 0.02 mm 16mm X Axis FWHM PPS Center RFP Center X Coordinate of PPS (mm) 22 Old Film Holder: Example mm X Axis (XY Plane): Coincidence of PPS and RFP Counts (Background subtracted) !x = 0.13 mm 16-mm X Axis XY Film PPS Center RFP Center 80% Width X Coordinate of PPS (mm) 23 Test Specifications and Frequency Elekta s specification for the -mm collimator is that Δx, Δy and Δz are all 0.3mm and that Frequency of tests: " r = 2 2 " x + " y + " z! 0. mm 2 1) Master Diode test is done bi-annually by Elekta service engineers as part preventative maintenance 2) Films are usually irradiated annually by the on-site GK physicist 2 8

9 Example of Results using Different Tools to Check PPS and RFP Coincidence for WHHS Perfexion GK Collimator = Master - Diode Tool! r =! x +! y +! z (mm) -mm 8-mm 16-mm New-Style Film Holder Old-Style Film Holder Profile Measurements using Same film handling requirements as for pin-point films Measurements performed in spherical phantom (80 mm radius) Must measure dose versus film response calibration curve 26 Beam Profiles (film): Measurement Tools New-style Spherical Phantom Old-style Spherical Phantom 1) Attaches to patient frame adapter 2) Film positioned between 2 rods 3) Can irradiate a 3D stack of films ) Composed of certified Therapy Grade Solid Water 5) 3 adapters provided for ion chambers or other detectors 1) Attaches to dosimetry adapter 2) Film positioned in central insert 3) Can irradiate only one film at a time ) Presumably composed of polystyrene 5) Additional inserts are supplied for ion chambers and other detectors 27 9

10 EBT Film Calibration Calibration curve obtained by irradiating films at several energy levels between 0.5 Gy and 8 Gy with the 16-mm collimator Pre-scan all films to obtain background subtraction Dose on the horizontal axis equals the dose at (100, 100, 100) plus the transit dose for the 16-mm collimator. Film Reading Calibration Curve Reading minus Background (inverted) Calibration Curve from 16mm Data Dose (Gy) at (100, 100, 100) Film reading in red channel 28 Transit Dose for 16-mm Collimator Sources move to the blocked position at the initiation of a treatment and between shots. Therefore, to achieve a 16-mm shot, the sources must pass over the -mm collimator, and thereby deliver a so-called transit dose. 16 B 8 Blocked source position between - and 8-mm collimators Sources Sector Collimator 29 Transit Dose for 16-mm Collimator One way to measure 16-mm transit dose is to deliver the same prescribed dose via two treatment plans to isocenter in the spherical phantom, one plan consisting of a single 16-mm shot, and the second plan consisting of four 16-mm shots. The transit dose is then: Transit Dose = difference in the number of times the source travels back and forth over the -mm collimator D! D 1 3 difference in measured dose between the two runs Gy/(16-mm shot) 30 10

11 Transit Dose for 16-mm Collimator Using ion-chamber measurements as described on previous slide, transit dose for 16-mm collimator = 0.03 Gy/shot for WHHS Perfexion in July 2008 Note that the transit dose changes as a function of time as the Co-60 sources decay. 31 Determination of Transit Dose by Film Measurements Irradiate 1 st film with 1 shot, dose = 5 Gy Irradiate 2 nd film with 18 shots, dose = 5 Gy Number of shots: large enough s.t. dose difference in measurable, small enough s.t. irradiation times are valid for LGK PFX Subtract background reading determined from unirradiated film Use Dose-to-reading calibration curve to convert readings to dose Transit dose = (D 18 - D 1 )/17 Results (July 2008): 16-mm Collimator: 0.03 Gy/shot (same as result obtained with ion chamber!) 8-mm and -mm Collimators: 0.01 Gy/shot 32 Determining Beam Profiles using ImageJ Using Line Tool, draw line in direction of increasing coordinate value Under Analyze options, select Profile List profile data, Select it, and Copy to Excel (or similar) program for analysis 33 11

12 Beam Profile Analysis Raw Data (from ImageJ,, red channel, inverted) Subtract background Apply dose calibration to data points Identify FWHM Define X (or Y or Z) = 100 at center of FWHM (note that we have already confirmed that the center of the RFP is at (100,100,100) with the pinpoint measurement.) Define X (or Y or Z) coordinate value on horizontal axis (00 dpi) Calculate FWHM in mm mm Y Profile from YZ Film 16 mm Y Profile from YZ film Center FWHM LGP Data 5 Dose (Gy) LGP Y Coordinate (mm) Raw Data Converted to Plot of Dose vs LGP Coordinate (example): Measured FWHM = 21.7 mm, LGP FWHM = 21.3 mm, Diff = 0. mm35 (where LGP stands for Leksell GammaPlan) Simplified Procedure to Check only the FWHM Use Calibration curve to calculate R 1/2, the film reading (with background subtracted) that corresponds to 50% of maximum profile dose Locate R 1/2 on the raw profile curve (background subtracted), and compute: FWHM = (C 2 C 1 ) P Reading minus Background (inverted) R 1/2 Film Reading Calibration Curve ½ Max Dose Dose (Gy) at (100, 100, 100) Calibration Curve from 16mm Data Raw Data from ImageJ R 1/2 Where, P = conversion factor, e.g., mm/pixel for films scanned at 00 dpi C 1 C

13 Specifications and Frequency for Beam Profile measurements Specification: According to Elekta: Measured and LGP values for FWHM should be within ± 1 mm of each other Frequency: Beam profiles should be measured upon acceptance of the GK unit and annually thereafter 37 Results from last Annual QA: All FWHM were between ± 0.1 mm and ± 0. mm of Leksell GammaPlan mm Dose Rate Measurement Currently no official calibration protocol specific to the LGK Charges of AAPM Task Group 178: Suggest a protocol for calibration with ionization chambers calibrated at an ADCL that can be used with all Gamma Stereotactic Radiosurgery (GSR) devices Work with the working group on dosimetry calibration protocol for beams that are not compliant with TG

14 16-mm Dose Rate Measurement Calibration Protocols: Performed in spherical phantom TG-21: Ion chamber calibrated 60 Co in-air Can be used for various phantom materials Can be used for various geometrical setups TG-51: Ion chamber calibrated for 60 Co in water. Designed to facilitate linear accelerator calibration and QA phantom must be water, 10 cm 10 cm field size 0 LGK Dose Rate Measurement: Some Recent Publications R Drzymala R Wood, J Levy: Calibration of the Gamma Knife using a new phantom following AAPM TG51 and TG21 protocols, Med. Phys. 35:51-521;2008 Compared the 2 protocols in the Elekta old-style spherical phantom and in a newly designed water phantom TG-51 in water phantom results were 1.% lower than TG-21 in polystyrene phantom 1 LGK Dose Rate Measurement: Some Recent Publications S Griffin Meltsner and LA DeWerd: Air Kerma based dosimetry calibration for the Leksell Gamma Knife, Med Phys 36: ;2009 Proposes an air-kerma-based dosimetry protocol using either an in-air or in-acrylic phantom measurement Modified version of TG-21 specific to LGK calibration geometry With new protocol, measured dose rates were between 1.5% and 2.9% higher than those used clinically by at 7 LGK sites (Models B and C) 2 1

15 16-mm Dose Rate Measurement: Practical Issues: Choice of new or old-style phantom Old-style phantom requires dosimetry adapter Bhatnagar, et al. (Med Phys 36: ;2009): The dosimetry adapter attenuates some beams in the lateral (3 and 7) sectors of the Perfexion unit, causing the overall 16-mm dose rate to be underestimated by approximately 1%. 3 New spherical phantom Patient Frame Adapter Choice of phantom Does not require dosimetry adapter Attaches to patient adapter More precise Provides better check of entire system 16-mm Dose Rate Measurement: Practical Issues Checking measured dose rate: compare results to TLD: either in-house or outside service (e.g., RPC SRS phantom) EBT GafChromic 5 15

16 Practical Method for Checking GK Dose Rate Calibration using EBT Film Irradiate films in 6-MV linear accelerator beam at 3 dose levels, e.g., 5 and 6 Gy to obtain a mini calibration curve Note that EBT film exhibits very little energy dependence Net Counts in Red Channel (Background subtracted) 6-MV Mini-Calibration MV Mini-Calibration Dose (Gy).0 Gy 5.0 Gy 6.0 Gy Red channel extracted, Background subtracted, film intensity inverted 6 Practical Method for Checking Dose Rate Calibration Irradiate film with 16-mm collimator, 5.0 maximum Net Counts in Red Channel (Background subtracted) 6-MV Mini-Calibration MV Mini-Calibration Dose (Gy) Determine dose at center of peak region from minicalibration curve Compare to expected value (e.g..96 Gy) 7 Expect ± 2% to 3% agreement Relative Output Factors (ROF) for the 8- and -mm Collimators Relative output factor for Gamma Knife is defined as: dd C / dt dd dt 16 / (100,100,100) i.e., ROF = dose rate of collimator C relative to 16- mm collimator, where both are measured at isocenter = (100,100,100) in spherical, 80-mm radius phantom 8 16

17 Unique Features of Perfexion Geometry Co-60 sources distributed in 5 rings for each identical sector Ring Number Number of Sources ROF: Unique Feature of Perfexion Geometry Each collimator within each ring has a different beam geometry Beam Geometry: There are, therefore, 15 distinct beam geometries: 5 rings multiplied by 3 beam-on positions per ring. This is in contrast to previous LGK designs for which all of the beam channels were identical, and there was only one type of beam. Each of the 15 beam types has a different: 1) Virtual source-to-isocenter distance (SAD) 2) Output factor Elekta determined these values by fitting a beam model to Monte Carlo generated data

18 Fitted Values for Relative Output Factor (ROF) and Virtual Source-to-focus distance (R( vsf ) Collimator Size Ring ROF R vsf (mm) Collimator Size Ring ROF R vsf (mm) Relative Output Factor for Each Collimator Size We cannot measure the 15 ROFs individually The ROF for the 8- and -mm collimators relative to the 16-mm collimator is determined from the equation: ROF( c) = 5 " i = 1 5 " ni! OFi ( c) i =, ni n! OF ( c = 16mm) i 1 = i Where the sum is taken over all 5 rings (8,32,0,32,0) n i represents the number of sources in each ring ROF(8mm) = 0.92 ROF(mm) = Two Ways to Determine ROF Measure dose, D c, delivered by each collimator at (100, 100, 100) for: A given treatment time: ROF c = D c /D 16 The same prescription dose: ROF c = D c / T c D 16 / T 16 ROF c " D c no min al # ROF c D 16 T c is the irradiation time for each collimator The approximately equal sign is replaced by an equality if the dose is prescribed to (100,100,100) instead of the point of 5 maximum dose 18

19 Relative Output Factors: Measurement Techniques Pin-point ion chamber GafChromic Film TLDs (rods and LiF microcubes) Glass Rods (Perks, et al.) Some References: Mack et al., Precision dosimetry for narrow photon beams used in radiosurgery - determination of Gamma Knife output factors, Med. Phys. 29: ; 2002 Perks et al., Glass rod detectors for small field, stereotactic radiosurgery dosimetric audit, Med. Phys. 32:726-32; 2005 Novotny et al. Measurement of relative output factors for the 8 and mm collimators of the Leksell Gamma Knife Perfexion by film dosimetry, Med Phys. 36: ; ROF Meas: : EBT Film and Fixed Dose Cut and mark Films Scan un-irradiated films to obtain background correction Irradiate 2 films at the 3 dose levels, e.g.,.5, 5.0 and 5.5 Gy for 16 collimator to obtain mini-calibration curve (choose either axial, coronal or sagittal plane). It is reasonable to assume that the calibration curve is piece-wise linear between measured points (6 films) Irradiate 2 films the - and 8-mm collimators to a dose in the middle of the minicalibration range in the same plane as calibration films ( films) Process films with ImageJ: extract red channel, invert intensity values, subtract background 56 Sample intensity values in center of films, convert to dose, calculate ROF My Results: Average of 7 sets of ROF Measurements 8-mm Collimator: ± ( 3.9% lower than Elekta) -mm Collimator: ± ( 2.5% lower than Elekta) Used constant time for 3 sets of measurements, constant dose for sets Used different measurement planes (axial, coronal or sagittal) Did not change values in LGP 57 19

20 Results Reported in Literature: Average of 5 sets of ROF Measurements Collimator 8 mm EDR 2 Film 0.90 ± (-2.1%) EBT Film ± 0.01 (-0.8%) MD-V2-55 Film ± (-2.0%) mm ± (-.5%) ± (+0.6%) ± (+1.7%) Novotny et al., Med. Phys Error levels in ROF Measurement Neglecting transit dose: Neglecting the difference between maximum dose and dose at (100,100,100) 0.03Gy = 0.6% 5Gy < 1% Standard deviation in film pixel values around point of measurement Between 0.5 and 1 count 59 What order of magnitude error in film reading causes a % error in dose? 160 Film Reading Calibration Curve Reading minus Background (inverted) Slope in vicinity of 5 Gy ~ 7 counts/gy 80 Y-intercept ~ 85 counts 60 0 Calibration Curve from 16mm Data Dose (Gy) at (100, 100, 100) 60 20

21 What order of magnitude error in film reading causes a % error in dose? From calibration curve on previous slide: 5 Gy a reading of 120 counts.8 Gy a reading of counts A difference of only 1. counts results in a % dose discrepancy 61 Summary The AAPM is working towards a calibration protocol for GSR units based on ADCL calibrated ionization chambers PPS/RFP coincidence and beam profile measurements are facilitated by using EBT GAFCHROMIC film ROF measurements remain the most challenging aspect of QA for GSR devices 62 Thank you for your attention! 63 21

- Water resistant. - Large size.

- Water resistant. - Large size. GAFCHROMIC EBT product brochure GAFCHROMIC EBT FEATURES GAFCHROMIC EBT dosimetry film has been developed specifically to address the needs of the medical physicist and dosimetrist working in the radiotherapy

More information

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE Rev. 1.0 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE DoseLab users may reference the following instructions to perform Tomotherapy Quality Assurance tests as recommended

More information

ISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM

ISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM FILMQA STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE Supports all major radiotherapy technologies! FilmQA is optimized for use with Gafchromic film products, including EBT2 and RTQA2. FILMQA helps

More information

GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom

GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom Introduction: These EBT2/EBT3 films, shown in Figure 1a-c, are specially sized and formatted to uniquely fit the Accuray Ball Cube II Phantom. Each

More information

Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010

Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010 Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010 Indra J. Das, PhD, FACR Department of Radiation Oncology Indiana University of School of Medicine & Midwest Proton Radiation

More information

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array José A. Bencomo, * Geoffrey Ibbott, Seungsoo Lee, and Joao A. Borges Department of Radiation Physics.

More information

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

3D Diode Array Commissioning: Building Confidence in 3D QA Technology 3D Diode Array Commissioning: Building Confidence in 3D QA Technology Caroline Yount, MS CANCER CENTER 3D QA The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose

More information

State of the Art Film Dosimetry

State of the Art Film Dosimetry State of the Art Film Dosimetry Micke A., Lewis D. Advanced Materials Ashland proprietary technology, patents pending Film Dosimetry Radiochromic Film EBT2/EBT3 One-Scan Protocol Multi-channel Film Dosimetry

More information

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 David F. Lewis Advanced Materials Group Ashland Specialty Ingredients Spain, April 2014 What is Radiochromic Film? A film that instantly changes

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

The most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 BUSINESS UNIT OF ISP

The most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 BUSINESS UNIT OF ISP The most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 GAFCHROMIC EBT dosimetry film Designed and optimized for ALL RT procedures Can be cut into

More information

PHYSICS QUESTIONNAIRE FORM

PHYSICS QUESTIONNAIRE FORM PHYSICS QUESTIONNAIRE FORM Institution Name: Date: Contact Information (name, address, phone, fax, email): Physicist: Radiation Oncologist: Dosimetrist (if applicable): Study Coordinator (if applicable):

More information

END-TO-END TESTING OF THE LEKSELL GAMMA KNIFE ICON SYSTEM

END-TO-END TESTING OF THE LEKSELL GAMMA KNIFE ICON SYSTEM END-TO-END TESTING OF THE LEKSELL GAMMA KNIFE ICON SYSTEM A Thesis Presented to The Academic Faculty by Krista R. Burton In Partial Fulfillment of the Requirements for the Degree Master of Science in Medical

More information

Calibration of KAP meters

Calibration of KAP meters Calibration of KAP meters Alexandr Malusek! Division of Radiological Sciences Department of Medical and Health Sciences Linköping University! 2014-04-15 1 Outline 1. KAP meter construction 2. Air kerma-area

More information

Development of new dosimeter for measuring dose distribution in CT

Development of new dosimeter for measuring dose distribution in CT Development of new dosimeter for measuring dose distribution in CT Poster No.: C-2925 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology - Without Subtopic Authors: Y. Muramatsu, K.

More information

Nathan Childress, Ph.D., DABR

Nathan Childress, Ph.D., DABR Nathan Childress, Ph.D., DABR Introduction TG-142 is a comprehensive QA protocol Covers nearly every aspect of machine and safety QA Recommends quantitative results Recommends high testing frequencies

More information

GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications

GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications I. SCOPE The protocol applies to GafChromic EBT3P and EBT3+P films exposed in GafChromic QuiCk Phantom

More information

Using Radiochromic Film Tips and Techniques

Using Radiochromic Film Tips and Techniques Using Radiochromic Film Tips and Techniques David F. Lewis, Ph.D. Senior Science Fellow Advanced Materials Group International Specialty Products dlewis@ispcorp.com October 22, 2010 Topics Scanners and

More information

NOT FOR DISTRIBUTION JINST_083P_0914 v1

NOT FOR DISTRIBUTION JINST_083P_0914 v1 Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam F. Di Lillo a,b, D. Dreossi c, F. Emiro a,b, C. Fedon d,e, R. Longo d,e, G. Mettivier a,b,*, L. Rigon d,e, P. Russo

More information

Commissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests

Commissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests Acknowledgements David Shepard, Ph.D. Daliang Cao, Ph.D. Muhammad K. N. Afghan, Ph.D. Jinsong Ye, M.S. Tony P. Wong, Ph.D. Fan Chen, Ph.D. Min Rao, Ph.D. Vivek Mehta, M.D. Igor Gomola, Ph.D. David Housley

More information

Analysis of Post-exposure Density Growth in Radiochromic Film with Respect to the Radiation Dose

Analysis of Post-exposure Density Growth in Radiochromic Film with Respect to the Radiation Dose J. Radiat. Res., 53, 301 305 (2012) Analysis of Post-exposure Density Growth in Radiochromic Film with Respect to the Radiation Dose Katsumi SHIMA 1,2, Kunihiko TATEOKA 1 *, Yuichi SAITOH 1,2, Junji SUZUKI

More information

7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators

7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators Implementing a new digital medical accelerator John Wong Johns Hopkins University AAPM, Austin, 2014 Acknowledgements Yin Zhang, Ken Wang, Kai Ding (Commissioning - JHU) Esteban Velarde, Joe Moore (QA

More information

A study of GafChromic XR Type R film response with reflective-type densitometers and economical flatbed scanners

A study of GafChromic XR Type R film response with reflective-type densitometers and economical flatbed scanners JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 A study of GafChromic XR Type R film response with reflective-type densitometers and economical scanners G. Thomas, 1, * R. Y.

More information

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Gena M.A.H 1, Ahmed L.El-Attar 2, Elbadry M. Zahran 3, Hany El-Gamal

More information

A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering

A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering Sandra E. Burch Department of Radiology, Medical College of Georgia, Augusta, Georgia 30912 Kimberlee J. Kearfott

More information

CyberKnife Iris Beam QA using Fluence Divergence

CyberKnife Iris Beam QA using Fluence Divergence CyberKnife Iris Beam QA using Fluence Divergence Ronald Berg, Ph.D., Jesse McKay, M.S. and Brett Nelson, M.S. Erlanger Medical Center and Logos Systems, Scotts Valley, CA Introduction The CyberKnife radiosurgery

More information

When small things matter. Small Field Dosimetry Application Guide

When small things matter. Small Field Dosimetry Application Guide R A D I AT I O N T H E R A P Y When small things matter. Small Field Dosimetry Application Guide Contents 1 Introduction 1 Introduction 2 2 The Physics of Small Fields 3 3 Detector Types 10 4 Detector

More information

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools SRS MapCHECK SRS Patient QA, No Film Your Most Valuable QA and Dosimetry Tools SRS Patient QA, No Film With improvements in targeting and localization, stereotactic treatments have become prevalent. To

More information

Comparison of measured Varian Clinac 21EX and TrueBeam accelerator electron field characteristics

Comparison of measured Varian Clinac 21EX and TrueBeam accelerator electron field characteristics JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Comparison of measured Varian Clinac 21EX and TrueBeam accelerator electron field characteristics Samantha A.M. Lloyd, 1a Sergei Zavgorodni,

More information

Electronic Brachytherapy Sources. Thomas W. Rusch

Electronic Brachytherapy Sources. Thomas W. Rusch Electronic Brachytherapy Sources Thomas W. Rusch Educational Objectives Understand key elements of ebx source construction & operation Understand the rationale and methods for air kerma strength calibration

More information

SCINTILLATING FIBER DOSIMETER ARRAY

SCINTILLATING FIBER DOSIMETER ARRAY SCINTILLATING FIBER DOSIMETER ARRAY FIELD OF THE INVENTION [0001] This invention relates generally to the field of dosimetry and, more particularly, to rapid, high-resolution dosimeters for advanced treatment

More information

Commissioning and validation of BrainLAB cones for 6X FFF and 10X FFF beams on a Varian TrueBeam STx

Commissioning and validation of BrainLAB cones for 6X FFF and 10X FFF beams on a Varian TrueBeam STx JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 6, 2013 Commissioning and validation of BrainLAB cones for 6X FFF and 10X FFF beams on a Varian TrueBeam STx David B. Wiant, a Jonathon A.

More information

Evaluation of a commercial flatbed document scanner and radiographic film scanner for radiochromic EBT film dosimetry

Evaluation of a commercial flatbed document scanner and radiographic film scanner for radiochromic EBT film dosimetry JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 2, SPRING 2010 Evaluation of a commercial flatbed document scanner and radiographic film scanner for radiochromic EBT film dosimetry Jason

More information

Cylindrical Ion Chambers

Cylindrical Ion Chambers Cylindrical Ion Chambers Radiation Oncology ON Victoreen Model 550T Series Cylindrical Ion Chambers for use with Model 35040 and Model 560 electrometers Wide range of applications in Diagnostic X-Ray and

More information

Effective energy measurement using radiochromic film: application of a mobile scanner

Effective energy measurement using radiochromic film: application of a mobile scanner Polish Journal of Medical Physics and Engineering 2016;22(4):85-92 December 2016 doi: 10.1515/pjmpe-2016-0015 ISSN 1898-0309 Scientific Paper Effective energy measurement using radiochromic film: application

More information

Cylindrical Ion Chambers Victoreen Model 550 Series

Cylindrical Ion Chambers Victoreen Model 550 Series Cylindrical Ion Chambers Victoreen Model 550 Series! Cylindrical Ion Chambers for use with Model 35040 and Model 530 electrometers! Wide range of applications in Diagnostic X-Ray and Radiation Oncology

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information

Post-irradiation colouration of Gafchromic EBT radiochromic film

Post-irradiation colouration of Gafchromic EBT radiochromic film INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 5 (25) N281 N285 PHYSICS IN MEDICINE AND BIOLOGY doi:1.188/31-9155/5/2/n4 NOTE Post-irradiation colouration of Gafchromic EBT radiochromic film Tsang Cheung

More information

GAFCHROMIC. Therapy Dosimetry Media Models to

GAFCHROMIC. Therapy Dosimetry Media Models to GAFCHROMIC Therapy Dosimetry Media Models 37-040 to 37-045! Superior uniformity and sensitivity! Dose rate and fractionation independent! Maps dose distribution! Provides quantitative measurements (via

More information

IQM Detector Characteristics: Signal reproducibility

IQM Detector Characteristics: Signal reproducibility The Integral Quality Monitor (IQM) System is a real-time beam verification system that monitors the accuracy of radiation delivery throughout each patient treatment without any user interaction. IQM continuously

More information

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017 The Current State of EPID-Based Linear Accelerator Quality Assurance Timothy Ritter, PhD, DABR, FAAPM 1 Disclosures Employed by the Veterans Health Administration Faculty appointment with the University

More information

The Ultimate 4D QA Solution A 4D isotropic cylindrical detector array for arc delivery QA and Dosimetry.

The Ultimate 4D QA Solution A 4D isotropic cylindrical detector array for arc delivery QA and Dosimetry. The Ultimate 4D QA Solution A 4D isotropic cylindrical detector array for arc delivery QA and Dosimetry. U.S.Patent No. 8,044,359; 6,125,335 Your Most Valuable QA and Dosimetry Tools 2 Y o u r M o s t

More information

IMRT Delivery System QA. IMRT Dose Delivery. Acceptance testing. Why: specific tests for IMRT? Accuracy of leaf positioning (gaps) MLC Alignment

IMRT Delivery System QA. IMRT Dose Delivery. Acceptance testing. Why: specific tests for IMRT? Accuracy of leaf positioning (gaps) MLC Alignment 1 IMRT Delivery System Q Thomas LoSasso, PhD Memorial Sloan Kettering Cancer Center IMRT Dose Delivery cceptance testing Commissioning Quality assurance Verification Q Why: specific tests for IMRT? 2.

More information

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System.

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System. Use of EPIDs for Non-Routine Linac QA Bin Cai PhD Disclosure Parts of this project received support from Varian Medical System. Learning Objectives Learn the recent development of EPID based Non-routine

More information

Monte Carlo study on a new concept of a scanning photon beam system for IMRT

Monte Carlo study on a new concept of a scanning photon beam system for IMRT NUKLEONIKA 2011;56(4):291 297 ORIGINAL PAPER Monte Carlo study on a new concept of a scanning photon beam system for IMRT Anna M. Wysocka-Rabin, Günter H. Hartmann Abstract. Intensity-modulated radiation

More information

Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film

Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 4, 2016 Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film Eric C. Lobb Department

More information

ArcCHECK. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools

ArcCHECK. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools ArcCHECK The Ultimate 4D QA Solution A 4D isotropical cylindrical detector array for arc delivery QA and Dosimetry U.S.Patent No. 8,044,359; 6,125,335 Compatible with: FFF Beams VMAT RapidArc TomoTherapy

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

Introduction of a Single Chip TLD System for Patient Dosimetry

Introduction of a Single Chip TLD System for Patient Dosimetry Introduction of a Single Chip TLD System for Patient Dosimetry C. Hranitzky a, M. Halda a, G. Müller a, B. Obryk b, H. Stadtmann a* a Austrian Research Centers GmbH ARC, 2444 Seibersdorf, Austria. b Institute

More information

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup Huaiqun Guan,

More information

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter MR-guided RT in Heidelberg Funded by the German Research Foundation (DFG) Heidelberg consortium received

More information

GAFCHROMIC HD-810 Radiochromic Dosimetry Film Configuration, Specifications and Performance Data

GAFCHROMIC HD-810 Radiochromic Dosimetry Film Configuration, Specifications and Performance Data GAFCHROMIC HD-810 Radiochromic Dosimetry Film Configuration, Specifications and Performance Data Description GAFCHROMIC HD-810 dosimetry film is designed for the measurement of absorbed dose of high-energy

More information

ArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung

ArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung ArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung Treffen des Arbeitskreises IMRT der DGMP Würzburg, 26 + 27.03.2009 Salih Arican Sun Nuclear Corporation QA Challenge for Rotational Beams

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

BRANDON RICE UNIVERSITY OF FLORIDA

BRANDON RICE UNIVERSITY OF FLORIDA METHODS FOR PRODUCING OFF-AXIS RATIO TABLES FROM MINI-MULTILEAF COLLIMATOR SHAPED CIRCULAR FIELDS FOR INPUT INTO A STEREOTACTIC RADIOSURGERY TREATMENT PLANNING SYSTEM By BRANDON RICE A THESIS PRESENTED

More information

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools SRS MapCHECK SRS Patient QA, No Film Your Most Valuable QA and Dosimetry Tools SRS Patient QA, No Film As clinics strive to treat with the smallest field possible, the demand for SRS/SBRT QA grows as well.

More information

Beam Commissioning and Annual QA Phantoms, Detectors & Accessories. iba-dosimetry.com

Beam Commissioning and Annual QA Phantoms, Detectors & Accessories. iba-dosimetry.com Beam Commissioning and Annual QA Phantoms, Detectors & Accessories iba-dosimetry.com Benefit from efficient and smartly designed beam commissioning & annual QA tools that make your life EASIEST. For MORE

More information

Test Equipment for Radiology and CT Quality Control Contents

Test Equipment for Radiology and CT Quality Control Contents Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3

More information

Homogeneity of GAFCHROMIC EBT2 film among different lot numbers

Homogeneity of GAFCHROMIC EBT2 film among different lot numbers JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 4, 2012 Homogeneity of GAFCHROMIC EBT2 film among different lot numbers Hirokazu Mizuno, 1,2a Yutaka Takahashi, 3 Atsushi Tanaka, 1 Takamitsu

More information

ArcCHECKTM. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools. VMAT RapidArc TomoTherapy Pinnacle 3 SmartArc Conventional IMRT

ArcCHECKTM. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools. VMAT RapidArc TomoTherapy Pinnacle 3 SmartArc Conventional IMRT TM The Ultimate 4D QA Solution A 4D isotropical cylindrical detector array for arc delivery QA and Dosimetry U.S.Patent No. 8,044,359 What is? The world s first true 4D detector array The world s first

More information

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images Aiping Ding, Bin Han, Lei Wang, Lei Xing Department of Radiation Oncology, Stanford University School of

More information

IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film

IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film CYRIC Annual Report 2005 IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film Ohuchi H. 1, and Abe K. 2 1 Graduate School of Pharmaceutical Sciences,

More information

COMPREHENSIVE TG-142 IMAGING AND MACHINE QA

COMPREHENSIVE TG-142 IMAGING AND MACHINE QA QA SOFTWARE COMPREHENSIVE TG-142 IMAGING AND MACHINE QA Automate the analysis of over thirty TG-142 recommended QA tasks The rapid progress of Radiation Therapy has created the need for Quality Assurance

More information

Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam

Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam Home Search Collections Journals About Contact us My IOPscience Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam This content has been downloaded from IOPscience.

More information

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS 7:28-14.1 Scope (a) This subchapter covers therapeutic installations used in the healing arts. These therapeutic installations include x-ray, accelerator and teletherapy

More information

Ludlum Medical Physics

Ludlum Medical Physics Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to

More information

Detectors for small field dosimetry

Detectors for small field dosimetry Detectors or small ield dosimetry Hugo Palmans MedAustron, Wiener Neustadt, Austria and National Physical Laboratory, Teddington, UK 1 080915 Overview Ideal detector Water calorimeter Ionization chamber

More information

Prototype electron phantom for radiographic and radiochromic film dosimetry

Prototype electron phantom for radiographic and radiochromic film dosimetry Louisiana State University LSU Digital Commons LSU Master's Theses Graduate School 2010 Prototype electron phantom for radiographic and radiochromic film dosimetry Chad Joseph Robertson Louisiana State

More information

I. Introduction.

I. Introduction. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Accuracy of measuring half- and quarter-value layers and appropriate aperture width of a convenient method using a lead-covered case

More information

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc.

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. An Introduction to TG-142 Imaging QA Using Standard Imaging Products Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. Goals Understand the nature and intent of TG 142 imaging

More information

Evaluation of a diode array for QA measurements on a helical tomotherapy unit

Evaluation of a diode array for QA measurements on a helical tomotherapy unit Evaluation of a diode array for QA measurements on a helical tomotherapy unit K. M. Langen, a S. L. Meeks, D. O. Poole, T. H. Wagner, T. R. Willoughby, O. A. Zeidan, and P. A. Kupelian Department of Radiation

More information

Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy

Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy Thomas LoSasso, a) Chen-Shou Chui, and C. Clifton Ling Department

More information

A proposed method for linear accelerator photon beam steering using EPID

A proposed method for linear accelerator photon beam steering using EPID Received: 13 January 2018 Revised: 11 May 2018 Accepted: 29 June 2018 DOI: 10.1002/acm2.12419 RADIATION ONCOLOGY PHYSICS A proposed method for linear accelerator photon beam steering using EPID Michael

More information

Aim. Images for this section: Page 2 of 13

Aim. Images for this section: Page 2 of 13 Changes in CT number of high atomic number materials with field of view when using an extended CT number to electron density curve and a metal artifact reduction reconstruction algorithm Poster No.: R-0094

More information

Commissioning an Elekta Versa HD linear accelerator

Commissioning an Elekta Versa HD linear accelerator JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 1, 2016 Commissioning an Elekta Versa HD linear accelerator Ganesh Narayanasamy, 1,2 Daniel Saenz, 1 Wilbert Cruz, 1,3 Chul S. Ha, 1 Niko

More information

SUN NUCLEAR. EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm. corporation. Your Most Valuable QA and Dosimetry Tools

SUN NUCLEAR. EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm. corporation. Your Most Valuable QA and Dosimetry Tools EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm SUN NUCLEAR corporation Your Most Valuable QA and Dosimetry Tools introduction Pre-treatment dose QA is an important process required

More information

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients

More information

Monica Kishore. Medical Physics Graduate Program Duke University. Approved: Jennifer O Daniel, Co-Supervisor. Fang-Fang Yin, Co-Supervisor

Monica Kishore. Medical Physics Graduate Program Duke University. Approved: Jennifer O Daniel, Co-Supervisor. Fang-Fang Yin, Co-Supervisor Accuracy of Planar Dosimetry for Volumetric Modulated Arc Therapy Quality Assurance by Monica Kishore Medical Physics Graduate Program Duke University Date: Approved: Jennifer O Daniel, Co-Supervisor Fang-Fang

More information

Master of Science Thesis. SIMIND Based Pinhole Imaging

Master of Science Thesis. SIMIND Based Pinhole Imaging Master of Science Thesis SIMIND Based Pinhole Imaging * Development and Validation Kurt Sundin Supervisor: Michael Ljungberg, PhD Medical Radiation Physics Clinical Sciences, Lund Lund University, 2006

More information

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 1, WINTER 2010 A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA Abdul Qadir Jangda,

More information

Rotational total skin electron irradiation with a linear accelerator

Rotational total skin electron irradiation with a linear accelerator JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 4, FALL 2008 Rotational total skin electron irradiation with a linear accelerator Eric P. Reynard, 1,a Michael D.C. Evans, 1 Slobodan Devic,

More information

A novel and fast method for proton range verification using a step wedge and 2D scintillator

A novel and fast method for proton range verification using a step wedge and 2D scintillator A novel and fast method for proton range verification using a step wedge and 2D scintillator Jiajian Shen, a) Bryce C. Allred, Daniel G. Robertson, Wei Liu, and Terence T. Sio Department of Radiation Oncology,

More information

Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method

Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method Markus Wendling, Robert J. W. Louwe, a Leah N. McDermott, Jan-Jakob Sonke, Marcel van Herk, and Ben J. Mijnheer

More information

Characterization of an in vivo diode dosimetry system for clinical use

Characterization of an in vivo diode dosimetry system for clinical use JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 2, SPRING 2003 Characterization of an in vivo diode dosimetry system for clinical use Kai Huang, 1, * William S. Bice, Jr., 2, and Oscar Hidalgo-Salvatierra

More information

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT Quality control of Gamma Camera By Dr/ Ibrahim Elsayed Saad 242 NMT WHAT IS QUALITY? The quality of a practice is to fulfill the expectations and demands from: Patient Clinicain Your self Quality assurance

More information

GS Introduction to Medical Physics IV Laboratory 5 Gamma Camera Characteristics

GS Introduction to Medical Physics IV Laboratory 5 Gamma Camera Characteristics GS02 0193 Introduction to Medical Physics IV Laboratory 5 Gamma Camera Characteristics Purpose: To introduce some of the basic characteristics of a gamma camera. This lab will introduce gamma camera QC

More information

A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine

A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine Iwein Van de Vondel, 1 Koen Tournel, 1 Dirk

More information

) was derived from 50% of the central axis dose created by nominal light field using geometry and mathematical methods. Leaf position (X mlc.

) was derived from 50% of the central axis dose created by nominal light field using geometry and mathematical methods. Leaf position (X mlc. JOURNAL O APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 6, 2012 A light field-based method to adjust rounded leaf end MLC position for split shape dose calculation correction in a radiation therapy

More information

Radiation transmission, leakage and beam penumbra measurements of a micro-multileaf collimator using GafChromic EBT film

Radiation transmission, leakage and beam penumbra measurements of a micro-multileaf collimator using GafChromic EBT film JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 3, SUMMER 2008 Radiation transmission, leakage and beam penumbra measurements of a micro-multileaf collimator using GafChromic EBT film Olivia

More information

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure Accuracy and calibration of integrated radiation output indicators in diagnostic radiology: A report of the AAPM Imaging Physics Committee Task Group 190 Pei-Jan P. Lin a) Virginia Commonwealth University

More information

DETECTORS UNCOMPROMISING QUALITY. The standard in dosimetry measurements for over 40 years. EXRADIN DETECTORS

DETECTORS UNCOMPROMISING QUALITY. The standard in dosimetry measurements for over 40 years. EXRADIN DETECTORS DETECTORS UNCOMPROMISING QUALITY The standard in dosimetry measurements for over 40 years. EXRADIN DETECTORS The Exradin Advantage Better Components Waterproof construction eliminates the need for sleeves

More information

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This

More information

Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector

Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector ORIGINAL ARTICLES Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector Gopiraj ANNAMALAI 1, Ramasubramanian VELAYUDHAM 2 ABSTRACT Received: 7.07.2009 Accepted: 2.11.2009

More information

A comparison of two methods for the determination of freein-air geometric efficiency in MDCT

A comparison of two methods for the determination of freein-air geometric efficiency in MDCT A comparison of two methods for the determination of freein-air geometric efficiency in MDCT Theocharis Berris *1, Kostas Perisinakis 1,, Antonios E. Papadakis and John Damilakis 1, 1 Department of Medical

More information

The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response

The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response Iran. J. Radiat. Res., 2005; 3 (1): 3-10 The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response M. Mohammadi 1,2,3* and E. Bezak 1,2

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information Ordering Information Please contact us if you have any questions or if you would like a quote or delivery schedule regarding the Catphan phantom. phone 800-525-1190, or 518-692-1190 fax 518-692-3329 mail

More information

Wide beam CT dosimetry. Elly Castellano

Wide beam CT dosimetry. Elly Castellano Wide beam CT dosimetry Elly Castellano Outline revision: CT dose indices wide-beam CT: the end of the road for CTDI? the IEC rescue plan for CTDI 100 the american way AAPM report 111 better estimates of

More information

Energy dependence of radiochromic dosimetry films for use in radiotherapy verification

Energy dependence of radiochromic dosimetry films for use in radiotherapy verification reports of practical oncology and radiotherapy 1 5 (2 0 1 0) 40 46 available at www.sciencedirect.com journal homepage: http://www.rpor.eu/ Original article Energy dependence of radiochromic dosimetry

More information

Diagnostic X-Ray Shielding

Diagnostic X-Ray Shielding Diagnostic X-Ray Shielding Multi-Slice CT Scanners Using NCRP 147 Methodology Melissa C. Martin, M.S., FAAPM, FACR Therapy Physics Inc., Bellflower, CA AAPM Annual Meeting, Orlando, FL FL Refresher Course

More information