Detectors for small field dosimetry

Size: px
Start display at page:

Download "Detectors for small field dosimetry"

Transcription

1 Detectors or small ield dosimetry Hugo Palmans MedAustron, Wiener Neustadt, Austria and National Physical Laboratory, Teddington, UK 1

2 Overview Ideal detector Water calorimeter Ionization chamber Alanine Diodes, diamond, ilm, TLD, 2

3 Which detector to use? Read literature: IPEM report 103 Manuacturers manuals, e.g. PTW 3

4 IPEM report 103 4

5 Which detector to use? Read literature: IPEM report 103 Manuacturers manuals, e.g. PTW 5

6 Which detector to use? 6

7 Ideal detector Detector properties Stability Dose linearity Dose rate linearity Dose per pulse linearity Energy response Spatial resolution Size o detector Orientation Recommendation Short-term detector response should be better than 0.1% or a total combined dose o many hundreds o kgy rom multiple exposures. linearity better than 0.1% should be possible over a dose range o at least three orders o magnitude (e.g Gy) Linear accelerators are typically operated at dose rates o 100 Gy/s (instantaneous) and 0.01 Gy/s (averaged) and thereore detectors should be linear to better than 0.1% over this range. A detector s response with changing dose per pulse should remain stable to better than 0.1% ater correction or ion recombination. Essentially water equivalent The choice o suitable detector in terms o spatial resolution is usually based on a trade-o between adequately high signal-to-noise ratio and dosimeter size. The detector size should be such that the volume averaging correction is not larger than 5%. The response o a detector should ideally be independent o the orientation o the detector with 7

8 Calorimetry Radiation energy turns into heat heat is tiny, but measurable our primary standards or absorbed dose are calorimeters 8

9 0.5 mm Water calorimeter - thermistor probes (LSDG) 0.6 mm thermistor

10 Water calorimeter chemical heat deect D w = c w 1. D T. 1 - h 10

11 Water calorimetry why does it not work or small ields Heath low: broad ield small ield 11

12 Problem: dose determination with ion chambers D D s p w = air w, air M W D = air m e air = M r V air W e OF = D D ield re = M ield M re ( s ( s w, air w, air ) ield ) re p p ield re correction actor 12

13 Which one is right? M /M re PTW (0.125cc) PTW (PinPoint) PTW (0.3cc) square ield width / mm Krauss, Vienna, Feb. 2009, 13

14 Ion chambers or small ield dosimetry - Water to air stopping power ratios Eklund&Ahnesjö, PMB 53:4231 (2008) Andreo&Brahme PMB 8:839 (1986) 14

15 Calibration in a 60 Co beam N D, w = D M w, 60 corr, Co 60 Co P 0 D w, 60 Co 15

16 Dosimetry in a high-energy x-ray beam M corr, D w, (P 0 )=M corr, N D,w k Only B! 16

17 OF Dosimetry in a high-energy x-ray beam square ield width / mm M corr, D w, (P 0 )=M corr, N D,w k Only B! 17

18 OF Dosimetry in a high-energy x-ray beam square ield width / mm M corr, D w, (P 0 )=M corr, N D,w k Only B! 18

19 OF Dosimetry in a high-energy x-ray beam square ield width / mm M corr, D w, (P 0 )=M corr, N D,w k Only B! 19

20 OF Dosimetry in a high-energy x-ray beam square ield width / mm M corr, D w, (P 0 )=M corr, N D,w k Only B! 20

21 OF Plane-parallel chamber square ield width / mm 21

22 OF Plane-parallel chamber square ield width / mm 22

23 OF Plane-parallel chamber square ield width / mm 23

24 Ion chambers: recombination 24 LeRoy et al., PMB 56: (2011)

25 Ion chambers: polarity Agostinelli et al., Med Phys 35: (2008) 25

26 Which one is right? M /M re PTW (0.125cc) PTW (PinPoint) PTW (0.3cc) Krauss, Vienna, Feb. 2009, square ield width / mm 26

27 Which one is right? M /M re PTW (electron diode) PTW (photon diode) PTW (MicroLion) PTW (0.125cc) PTW (PinPoint) PTW (0.3cc) Krauss, Vienna, Feb. 2009, square ield width / mm 27

28 Which one is right? M /M re PTW (electron diode) PTW (photon diode) PTW (MicroLion) PTW (0.125cc) PTW (PinPoint) PTW (0.3cc) square ield width / mm Krauss, Vienna, Feb. 2009, 28

29 Diodes or small ield dosimetry Sauer and Wilbert 2007 Med Phys 34:

30 Diamond detectors 1.20 (m en /r) carbon,water volume averaging correction actor k vol Volume averaging correction actor (p vol ) E E E E E photon energy / MeV Side o square ield [cm] Diamond (diameter 2 mm) Diamond (diameter 3.2 mm) Diamond (diameter 4 mm) Diamond (4.3 x 4.3 mm 2 ) Diamond (5 x 3 mm 2 ) side o square ield / cm Lechner 2013 Radiother Oncol 109:356 30

31 Diodes and diamonds 31

32 Organic scintillators Cr. Louis Archambault 32

33 Liquid ionization chambers 33

34 Radiochromic ilm Ralston 2012 Phys Med Biol 57:

35 Alanine Amino acid / readout by ESR Pellets 2.5 mm thick and Ø 5 mm or 2.5 mm Water-equivalent Density 1.2 g cm -3 Bergstrand 2003 Phys Med Biol 48:

36 Real time detectors: diodes, diamond, Al2O3:C and scintillators Passive detectors Some others Table I. Summary o properties o detectors used in this study solid state detectors Detector type Dimensions o sensitive volume [mm] Material Z e Material density [g/cm 3 ] Electron density [e - /g] relative to water TLD chips TLD-100 LiF:Mg,Ti TLD µ-cubes RPLD (GD-302M), Japan Length 6, diameter 1.5 TLD-100 LiF:Mg,Ti Silver activated phosphate glass Alanine, NPL Thickness 2.3, diameter 5 Alanine/Parain Alanine, DTU Thickness 2.8, diameter 4.8 Alanine/Parain IBA SFD diode Thickness 0.06, diameter 0.6 Silicone IBA PFD diode Thickness 0.06, diameter 2 Silicone IBA EFD diode Thickness 0.06, diameter 2 Silicone PTW diamond Thickness Sensitive area Diamond 6 approximately Al 2 O 3 :C, DTU Al 2 O 3 :C Saint Gobain BCF-60 scintillator 1 Length 1, diameter 1 Polystyrene Saint Gobain BCF-60 scintillator 2 Length 2, diameter 0.5 Polystyrene the density varies depending on the purity o the diamond used

37 Field output actors correction actors / components Scott et al Phys. Med. Biol. 57:

38 Other proposals to actorize SF correction actors Bouchard et al Med. Phys. 36:4654 Crop et al Phys. Med. Biol. 54:2951 Czarnecki and Zink 2013 Phys. Med. Biol. 58:2431 Fenwick et al. 2013: Phys. Med. Biol. 58:

39 Correction actors or unshielded diodes Francescon et al 2011, Med Phys 38:6513 Benmakhlou et al 2014, Med Phys 41:

40 Ch 5 Practical implementation dosimetry / availability, re data k, re 40

41 Ch 6 Practical implementation relative dosimetry / correction actors or OF e.g. CyberKnie / iris collimator E 0 =7.0 MeV, FWHM spot 2.1 mm Field size / mm PTW PTW PTW PTW SN Edge Exradin A PTW PTW microlion Francescon et al Med Phys 57:3741 k,, re re 41

42 , Field output actors Monte Carlo calculated correction actors CyberKnie Dierent FWHM primary source k, k k,,,, = = a M M,, b,, = 60 mm collimator,, Francescon et al 2008 Med Phys 35:

43 43 Field output actors: (1) (2) (2) (2) (1) (1) (2) (1),,,,,, rel rel M M M M M M k k = = w w w w k M M M D M D M M D D,,,,,,,, = = = w w M M D D k =,,,,

44 Field output actors - CyberKnie:,, ( 1 ) ratio o correction actors (MC or vol),, ( 6 Point ode ode GE nine D T ilm lymer gel k 2 ) Pantelis et al Med Phys 37:2369 M / M60 k A16 PinPoint Diode Diode EDGE Alanine TLD EBT ilm Polymer gel M / Mre diameter / mm Diode Diode EDGE (M/M 60 ) 2 /(M/M 60 ) diameter / mm TLD ExrA16 PinPoint EBT ilm SHD USD EDGE alanine TLD EBT GEL Polymer gel detector PinPoint Diode Diode EDGE Alanine TLD EBT ilm Polymer gel diameter / mm ( M / M re ) * k, PinPoint Diode Diode EDGE Alanine 44

45 Field output actors correction actors PTW IBA SFD 45

46 Field output actors correction actors PTW PTW PTW

47 47

Physics of small megavoltage photon beam dosimetry

Physics of small megavoltage photon beam dosimetry Physics of small megavoltage photon beam dosimetry P Andreo, Professor of Medical Radiation Physics Karolinska University Hospital, Stockholm, Sweden ICARO2 IAEA, Vienna, 2017 IAEA-AAPM working group IAEA

More information

Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010

Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010 Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010 Indra J. Das, PhD, FACR Department of Radiation Oncology Indiana University of School of Medicine & Midwest Proton Radiation

More information

When small things matter. Small Field Dosimetry Application Guide

When small things matter. Small Field Dosimetry Application Guide R A D I AT I O N T H E R A P Y When small things matter. Small Field Dosimetry Application Guide Contents 1 Introduction 1 Introduction 2 2 The Physics of Small Fields 3 3 Detector Types 10 4 Detector

More information

Cylindrical Ion Chambers

Cylindrical Ion Chambers Cylindrical Ion Chambers Radiation Oncology ON Victoreen Model 550T Series Cylindrical Ion Chambers for use with Model 35040 and Model 560 electrometers Wide range of applications in Diagnostic X-Ray and

More information

Cylindrical Ion Chambers Victoreen Model 550 Series

Cylindrical Ion Chambers Victoreen Model 550 Series Cylindrical Ion Chambers Victoreen Model 550 Series! Cylindrical Ion Chambers for use with Model 35040 and Model 530 electrometers! Wide range of applications in Diagnostic X-Ray and Radiation Oncology

More information

Separation of scatter from small MV beams and its effect on detector response

Separation of scatter from small MV beams and its effect on detector response Separation of scatter from small MV beams and its effect on detector response Sonja Wegener a) and Otto A. Sauer Department of Radiation Oncology, University of W urzburg, 97080 W urzburg, Germany (Received

More information

RADIATION THERAPY DETECTORS. For Small Field Dosimetry

RADIATION THERAPY DETECTORS. For Small Field Dosimetry RADIATION THERAPY DETECTORS For Small Field Dosimetry 0.125 cm 3 Semiflex Chamber Type 31010 Standard therapy chamber for scanning systems and for absolute dosimetry Waterproof, semiflexible design for

More information

DETECTORS UNCOMPROMISING QUALITY. The standard in dosimetry measurements for over 40 years. EXRADIN DETECTORS

DETECTORS UNCOMPROMISING QUALITY. The standard in dosimetry measurements for over 40 years. EXRADIN DETECTORS DETECTORS UNCOMPROMISING QUALITY The standard in dosimetry measurements for over 40 years. EXRADIN DETECTORS The Exradin Advantage Better Components Waterproof construction eliminates the need for sleeves

More information

QA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion

QA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion Leksell Gamma Knife Perfexion QA Considerations Paula L. Petti, Ph.D. Taylor McAdam Bell Neuroscience Institute Washington Hospital Healthcare System Fremont, CA 1 QA for LGK Perfexion : : Follow NRC licensing

More information

MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM *

MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM * MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM * C. Kessler and D.T. Burns December 2018 BUREAU INTERNATIONAL DES POIDS ET MESURES

More information

Energy Monitoring Device for Electron Beam Facilities

Energy Monitoring Device for Electron Beam Facilities 1 SM/EB-18 Energy Monitoring Device for Electron Beam Facilities M. Lavalle 1, P.G. Fuochi 1, A. Martelli 1, U. Corda 1, A. Kovács 2, K. Mehta 3, F. Kuntz 4, S. Plumeri 4 1 CNR-ISOF, Via P. Gobetti 101,

More information

IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film

IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film CYRIC Annual Report 2005 IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film Ohuchi H. 1, and Abe K. 2 1 Graduate School of Pharmaceutical Sciences,

More information

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 David F. Lewis Advanced Materials Group Ashland Specialty Ingredients Spain, April 2014 What is Radiochromic Film? A film that instantly changes

More information

Preliminary studies of a new monitor ionization chamber

Preliminary studies of a new monitor ionization chamber 1 Preliminary studies of a new monitor ionization chamber Maíra T. Yoshizumi, Vitor Vivolo and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares (IPEN-CNEN) Comissão Nacional de Energia

More information

DETECTORS. Including Codes of Practice IONIZING RADIATION

DETECTORS. Including Codes of Practice IONIZING RADIATION DETECTORS Including Codes of Practice 2015 IONIZING RADIATION PTW History and General Remarks Looking back on a long history PTW-Freiburg is an internationally operating company, manufacturing and marketing

More information

IONIZING RADIATION 2018 / 2019 DETECTORS. Including Codes of Practice

IONIZING RADIATION 2018 / 2019 DETECTORS. Including Codes of Practice IONIZING RADIATION DETECTORS 2018 / 2019 Including Codes of Practice PTW History and General Remarks Looking back on a long history PTW-Freiburg is an internationally operating company, manufacturing and

More information

Assessment of an Unshielded Electron Field Diode Dosimeter for Beam Scanning in Small- to Medium-Sized 6 MV Photon Fields

Assessment of an Unshielded Electron Field Diode Dosimeter for Beam Scanning in Small- to Medium-Sized 6 MV Photon Fields Iranian Journal of Medical Physics Vol. 10, No. 1-2, Winter & Spring 2013, 51-57 Received: November 22, 2012; Accepted: March 12, 2013 Original Article Assessment of an Unshielded Electron Field Diode

More information

Comparison of measured Varian Clinac 21EX and TrueBeam accelerator electron field characteristics

Comparison of measured Varian Clinac 21EX and TrueBeam accelerator electron field characteristics JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Comparison of measured Varian Clinac 21EX and TrueBeam accelerator electron field characteristics Samantha A.M. Lloyd, 1a Sergei Zavgorodni,

More information

Measurements of MeV Photon Flashes in Petawatt Laser Experiments

Measurements of MeV Photon Flashes in Petawatt Laser Experiments UCRL-JC-131359 PREPRINT Measurements of MeV Photon Flashes in Petawatt Laser Experiments M. J. Moran, C. G. Brown, T. Cowan, S. Hatchett, A. Hunt, M. Key, D.M. Pennington, M. D. Perry, T. Phillips, C.

More information

SCINTILLATING FIBER DOSIMETER ARRAY

SCINTILLATING FIBER DOSIMETER ARRAY SCINTILLATING FIBER DOSIMETER ARRAY FIELD OF THE INVENTION [0001] This invention relates generally to the field of dosimetry and, more particularly, to rapid, high-resolution dosimeters for advanced treatment

More information

Calibration of KAP meters

Calibration of KAP meters Calibration of KAP meters Alexandr Malusek! Division of Radiological Sciences Department of Medical and Health Sciences Linköping University! 2014-04-15 1 Outline 1. KAP meter construction 2. Air kerma-area

More information

熊本大学学術リポジトリ. Kumamoto University Repositor

熊本大学学術リポジトリ. Kumamoto University Repositor 熊本大学学術リポジトリ Kumamoto University Repositor Title Monte Carlo calculations of the rep correction factor, Ρ_, for cy chamber cav Author(s) Araki, Fujio CitationRadiological Physics and Technology Issue

More information

Post-irradiation colouration of Gafchromic EBT radiochromic film

Post-irradiation colouration of Gafchromic EBT radiochromic film INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 5 (25) N281 N285 PHYSICS IN MEDICINE AND BIOLOGY doi:1.188/31-9155/5/2/n4 NOTE Post-irradiation colouration of Gafchromic EBT radiochromic film Tsang Cheung

More information

Development of Personal Dosimeter Using Electronic Dose Conversion Method

Development of Personal Dosimeter Using Electronic Dose Conversion Method Proceedings of the Korean Nuclear Spring Meeting Gyeong ju, Korea, May 2003 Development of Personal Dosimeter Using Electronic Dose Conversion Method Wanno Lee, Bong Jae Lee, and Chang Woo Lee Korea Atomic

More information

Progress Report from the HDR Brachytherapy Working Party. Margaret Bidmead

Progress Report from the HDR Brachytherapy Working Party. Margaret Bidmead Progress Report from the HDR Brachytherapy Working Party Margaret Bidmead HDR Ir-192 calibration service provided by NPL since May 2004 Primary standard irradiated with a Nucletron classic HDR IR-192 source

More information

LaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging

LaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging 8th International Workshop on Radiation Imaging Detectors Pisa 2-6 July 2006 LaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging Roberto Pani On behalf of SCINTIRAD Collaboration

More information

NOT FOR DISTRIBUTION JINST_083P_0914 v1

NOT FOR DISTRIBUTION JINST_083P_0914 v1 Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam F. Di Lillo a,b, D. Dreossi c, F. Emiro a,b, C. Fedon d,e, R. Longo d,e, G. Mettivier a,b,*, L. Rigon d,e, P. Russo

More information

Effective energy measurement using radiochromic film: application of a mobile scanner

Effective energy measurement using radiochromic film: application of a mobile scanner Polish Journal of Medical Physics and Engineering 2016;22(4):85-92 December 2016 doi: 10.1515/pjmpe-2016-0015 ISSN 1898-0309 Scientific Paper Effective energy measurement using radiochromic film: application

More information

Characterization, Commissioning and Evaluation of Delta 4 IMRT QA System. Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX.

Characterization, Commissioning and Evaluation of Delta 4 IMRT QA System. Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX. Characterization, Commissioning and Evaluation of Delta 4 IMRT QA System Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX. 1 Acknowledgements Collaborators: Jose Bencomo, Rafael. M. Landrove, Peter

More information

Monte Carlo study on a new concept of a scanning photon beam system for IMRT

Monte Carlo study on a new concept of a scanning photon beam system for IMRT NUKLEONIKA 2011;56(4):291 297 ORIGINAL PAPER Monte Carlo study on a new concept of a scanning photon beam system for IMRT Anna M. Wysocka-Rabin, Günter H. Hartmann Abstract. Intensity-modulated radiation

More information

Electronic Brachytherapy Sources. Thomas W. Rusch

Electronic Brachytherapy Sources. Thomas W. Rusch Electronic Brachytherapy Sources Thomas W. Rusch Educational Objectives Understand key elements of ebx source construction & operation Understand the rationale and methods for air kerma strength calibration

More information

A Gel Lung-Equivalent Material for 3D Dosimetry

A Gel Lung-Equivalent Material for 3D Dosimetry A Gel Lung-Equivalent Material for 3D Dosimetry Joao A. Borges a), M.S., José A. Bencomo, Ph.D. 1), Geoffrey S. Ibbott, Ph.D. 2), and Jingfei Ma, Ph.D. 3) a), 1), 2) Departments of Radiation Physics, 3)

More information

Commissioning and validation of BrainLAB cones for 6X FFF and 10X FFF beams on a Varian TrueBeam STx

Commissioning and validation of BrainLAB cones for 6X FFF and 10X FFF beams on a Varian TrueBeam STx JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 6, 2013 Commissioning and validation of BrainLAB cones for 6X FFF and 10X FFF beams on a Varian TrueBeam STx David B. Wiant, a Jonathon A.

More information

I. Introduction.

I. Introduction. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Accuracy of measuring half- and quarter-value layers and appropriate aperture width of a convenient method using a lead-covered case

More information

SCINTILLATOR DETECTORS FOR THE ESS HIGH ENERGY WIRE SCANNER

SCINTILLATOR DETECTORS FOR THE ESS HIGH ENERGY WIRE SCANNER MOPL8 Proceedings of HB6, Malmö, Sweden SCINTILLATOR DETECTORS FOR THE ESS HIGH ENERGY WIRE SCANNER B. Cheymol, European Spallation Source, Lund, Sweden Abstract In the ESS linac [], during commissioning

More information

X-Ray Spectroscopy with a CCD Detector. Application Note

X-Ray Spectroscopy with a CCD Detector. Application Note X-Ray Spectroscopy with a CCD Detector In addition to providing X-ray imaging solutions, including CCD-based cameras that image X-rays using either direct detection (0.5-20 kev) or indirectly using a scintillation

More information

GAFCHROMIC. Therapy Dosimetry Media Models to

GAFCHROMIC. Therapy Dosimetry Media Models to GAFCHROMIC Therapy Dosimetry Media Models 37-040 to 37-045! Superior uniformity and sensitivity! Dose rate and fractionation independent! Maps dose distribution! Provides quantitative measurements (via

More information

Analysis of Post-exposure Density Growth in Radiochromic Film with Respect to the Radiation Dose

Analysis of Post-exposure Density Growth in Radiochromic Film with Respect to the Radiation Dose J. Radiat. Res., 53, 301 305 (2012) Analysis of Post-exposure Density Growth in Radiochromic Film with Respect to the Radiation Dose Katsumi SHIMA 1,2, Kunihiko TATEOKA 1 *, Yuichi SAITOH 1,2, Junji SUZUKI

More information

SCINTILLATION FIBER DETECTOR FOR DOSE VERIFICATION IN PROTON THERAPY

SCINTILLATION FIBER DETECTOR FOR DOSE VERIFICATION IN PROTON THERAPY SCINTILLATION FIBER DETECTOR FOR DOSE VERIFICATION IN PROTON THERAPY Suhyun Lee Nuclear Physics Lab. Korea University CONTENTS Introduction Detector - Scintillation-fiber array - Electronics Beam Test

More information

Recent Development and Study of Silicon Solid State Photomultiplier (MRS Avalanche Photodetector)

Recent Development and Study of Silicon Solid State Photomultiplier (MRS Avalanche Photodetector) Recent Development and Study of Silicon Solid State Photomultiplier (MRS Avalanche Photodetector) Valeri Saveliev University of Obninsk, Russia Vienna Conference on Instrumentation Vienna, 20 February

More information

Commissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests

Commissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests Acknowledgements David Shepard, Ph.D. Daliang Cao, Ph.D. Muhammad K. N. Afghan, Ph.D. Jinsong Ye, M.S. Tony P. Wong, Ph.D. Fan Chen, Ph.D. Min Rao, Ph.D. Vivek Mehta, M.D. Igor Gomola, Ph.D. David Housley

More information

A Sort-of Tissue Equivalent Proportional Counter (STEPC) for Space Radiation Dosimetry Applications

A Sort-of Tissue Equivalent Proportional Counter (STEPC) for Space Radiation Dosimetry Applications A Sort-of Tissue Equivalent Proportional Counter (STEPC) for Space Radiation Dosimetry Applications Eric Benton, Tyler Collums, and Art Lucas E. V. Benton Radiation Physics Laboratory Oklahoma State University

More information

Received 3 December 2016; revised 17 May 2017; editorial decision 21 June 2017; accepted 4 July 2017

Received 3 December 2016; revised 17 May 2017; editorial decision 21 June 2017; accepted 4 July 2017 Radiation Protection Dosimetry (2018), Vol. 178, No. 2, pp. 235 241 Advance Access publication 15 July 2017 doi:10.1093/rpd/ncx091 NOTE COMPARISON OF PERSONAL DOSE EQUIVALENT Hp(10) IN 137 CS RADIATION

More information

The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response

The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response Iran. J. Radiat. Res., 2005; 3 (1): 3-10 The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response M. Mohammadi 1,2,3* and E. Bezak 1,2

More information

Radiation transmission, leakage and beam penumbra measurements of a micro-multileaf collimator using GafChromic EBT film

Radiation transmission, leakage and beam penumbra measurements of a micro-multileaf collimator using GafChromic EBT film JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 3, SUMMER 2008 Radiation transmission, leakage and beam penumbra measurements of a micro-multileaf collimator using GafChromic EBT film Olivia

More information

Summer Student project report

Summer Student project report Summer Student project report Mika Väänänen September 1, 2017 Abstract In this report I give a brief overview of my activities during the summer student project. I worked on the scintillating fibre (SciFi)

More information

2012 :15th SESSION of ESMP

2012 :15th SESSION of ESMP 2012 :15th SESSION of ESMP Lecture presented in Archamps (Salève Building) by : Elly CASTELLANO (London) Patient dosimetry in x-ray imaging and CT Elly Castellano Objectives measurable dose quantities

More information

RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS

RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS S-94,316 PATENTS-US-A96698 BEAM POSITION MONITOR RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS CONTRACTUAL ORIGIN OF THE INVENTION The United States Government has rights in this invention pursuant

More information

A STUDY OF THE RESPONSE OF DEPLETED TYPE p-mosfets TO PHOTON AND ELECTRON DOSE

A STUDY OF THE RESPONSE OF DEPLETED TYPE p-mosfets TO PHOTON AND ELECTRON DOSE A STUDY OF THE RESPONSE OF DEPLETED TYPE p-mosfets TO PHOTON AND ELECTRON DOSE M. Fragopoulou 1, S. Stoulos 1, and M. Zamani 1, 1 School of Physics, Nuclear and Elementary Particle Section Aristotle University

More information

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array José A. Bencomo, * Geoffrey Ibbott, Seungsoo Lee, and Joao A. Borges Department of Radiation Physics.

More information

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Gena M.A.H 1, Ahmed L.El-Attar 2, Elbadry M. Zahran 3, Hany El-Gamal

More information

Evaluation of X-Ray Beam Quality Based on Measurements and Estimations Using SpekCalc and Ipem78 Models

Evaluation of X-Ray Beam Quality Based on Measurements and Estimations Using SpekCalc and Ipem78 Models Original Article Evaluation of X-Ray Beam Quality Based on Measurements and Estimations Using SpekCalc and Ipem78 Models Suk Chiang Chen, Wei Loong Jong, Ahmad Zaky Harun Submitted: 20 Aug 2011 Accepted:

More information

AIDA-2020 Advanced European Infrastructures for Detectors at Accelerators. Milestone Report

AIDA-2020 Advanced European Infrastructures for Detectors at Accelerators. Milestone Report AIDA-2020-MS15 AIDA-2020 Advanced European Infrastructures for Detectors at Accelerators Milestone Report Design specifications of test stations for irradiated silicon sensors and LHC oriented front-end

More information

P ILC A. Calcaterra (Resp.), L. Daniello (Tecn.), R. de Sangro, G. Finocchiaro, P. Patteri, M. Piccolo, M. Rama

P ILC A. Calcaterra (Resp.), L. Daniello (Tecn.), R. de Sangro, G. Finocchiaro, P. Patteri, M. Piccolo, M. Rama P ILC A. Calcaterra (Resp.), L. Daniello (Tecn.), R. de Sangro, G. Finocchiaro, P. Patteri, M. Piccolo, M. Rama Introduction and motivation for this study Silicon photomultipliers ), often called SiPM

More information

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26 Today s Outline - January 25, 2018 C. Segre (IIT) PHYS 570 - Spring 2018 January 25, 2018 1 / 26 Today s Outline - January 25, 2018 HW #2 C. Segre (IIT) PHYS 570 - Spring 2018 January 25, 2018 1 / 26 Today

More information

Diamond X-ray Rocking Curve and Topograph Measurements at CHESS

Diamond X-ray Rocking Curve and Topograph Measurements at CHESS Diamond X-ray Rocking Curve and Topograph Measurements at CHESS G. Yang 1, R.T. Jones 2, F. Klein 3 1 Department of Physics and Astronomy, University of Glasgow, Glasgow, UK G12 8QQ. 2 University of Connecticut

More information

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Takeshi Fujiwara 1, Yuki Mitsuya 2, Hiroyuki Takahashi 2, and Hiroyuki Toyokawa 2 1 National Institute of Advanced Industrial

More information

Status of the PRad Experiment (E )

Status of the PRad Experiment (E ) Status of the PRad Experiment (E12-11-106) NC A&T State University Outline Experimental apparatus, current status Installation plan Draft run plan Summary PRad Experimental Setup Main detectors and elements:

More information

TNT DoseMate Dosimeter Detector

TNT DoseMate Dosimeter Detector TNT 12000 DoseMate Dosimeter Detector Quick Reference Guide Introduction The TNT 12000 DoseMate makes accurate dose and dose rate measurements quickly. Figure 1 and Table 1 shows the typical TNT 12000

More information

Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam

Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam Home Search Collections Journals About Contact us My IOPscience Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam This content has been downloaded from IOPscience.

More information

) was derived from 50% of the central axis dose created by nominal light field using geometry and mathematical methods. Leaf position (X mlc.

) was derived from 50% of the central axis dose created by nominal light field using geometry and mathematical methods. Leaf position (X mlc. JOURNAL O APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 6, 2012 A light field-based method to adjust rounded leaf end MLC position for split shape dose calculation correction in a radiation therapy

More information

A novel and fast method for proton range verification using a step wedge and 2D scintillator

A novel and fast method for proton range verification using a step wedge and 2D scintillator A novel and fast method for proton range verification using a step wedge and 2D scintillator Jiajian Shen, a) Bryce C. Allred, Daniel G. Robertson, Wei Liu, and Terence T. Sio Department of Radiation Oncology,

More information

Prompt-Gamma Based Range Verification in Particle Therapy: New prospects (also) for 4D?

Prompt-Gamma Based Range Verification in Particle Therapy: New prospects (also) for 4D? Prompt-Gamma Based Range Verification in Particle Therapy: New prospects (also) for 4D? Guntram Pausch 1-3 1 OncoRay National Center for Radiation Research in Oncology, Dresden 2 Technische Universität

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

arxiv: v2 [physics.ins-det] 14 Jan 2009

arxiv: v2 [physics.ins-det] 14 Jan 2009 Study of Solid State Photon Detectors Read Out of Scintillator Tiles arxiv:.v2 [physics.ins-det] 4 Jan 2 A. Calcaterra, R. de Sangro [], G. Finocchiaro, E. Kuznetsova 2, P. Patteri and M. Piccolo - INFN,

More information

Commissioning an Elekta Versa HD linear accelerator

Commissioning an Elekta Versa HD linear accelerator JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 1, 2016 Commissioning an Elekta Versa HD linear accelerator Ganesh Narayanasamy, 1,2 Daniel Saenz, 1 Wilbert Cruz, 1,3 Chul S. Ha, 1 Niko

More information

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This

More information

Introduction of a Single Chip TLD System for Patient Dosimetry

Introduction of a Single Chip TLD System for Patient Dosimetry Introduction of a Single Chip TLD System for Patient Dosimetry C. Hranitzky a, M. Halda a, G. Müller a, B. Obryk b, H. Stadtmann a* a Austrian Research Centers GmbH ARC, 2444 Seibersdorf, Austria. b Institute

More information

Use of Back Scattered Ionizing Radiation for Measurement of Thickness of the Catalytic Agent Active Material

Use of Back Scattered Ionizing Radiation for Measurement of Thickness of the Catalytic Agent Active Material 18th World Conference on Nondestructive Testing, 16- April 1, Durban, South Africa Use of Back Scattered Ionizing Radiation for Measurement of Thickness of the Catalytic Agent Active Material Boris V.

More information

A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering

A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering Sandra E. Burch Department of Radiology, Medical College of Georgia, Augusta, Georgia 30912 Kimberlee J. Kearfott

More information

Monitoring the eye lens

Monitoring the eye lens Funding TS7e.3 Rolf Behrens Physikalisch-Technische Bundesanstalt Introduction April 2011: Limit of H lens : 150 msv 20 msv per year Protecting and monitoring may be necessary! Main question: Which quantity

More information

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring

More information

Characterization of an in vivo diode dosimetry system for clinical use

Characterization of an in vivo diode dosimetry system for clinical use JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 2, SPRING 2003 Characterization of an in vivo diode dosimetry system for clinical use Kai Huang, 1, * William S. Bice, Jr., 2, and Oscar Hidalgo-Salvatierra

More information

Title detector with operating temperature.

Title detector with operating temperature. Title Radiation measurements by a detector with operating temperature cryogen Kanno, Ikuo; Yoshihara, Fumiki; Nou Author(s) Osamu; Murase, Yasuhiro; Nakamura, Masaki Citation REVIEW OF SCIENTIFIC INSTRUMENTS

More information

Muon detection in security applications and monolithic active pixel sensors

Muon detection in security applications and monolithic active pixel sensors Muon detection in security applications and monolithic active pixel sensors Tracking in particle physics Gaseous detectors Silicon strips Silicon pixels Monolithic active pixel sensors Cosmic Muon tomography

More information

CyberKnife Iris Beam QA using Fluence Divergence

CyberKnife Iris Beam QA using Fluence Divergence CyberKnife Iris Beam QA using Fluence Divergence Ronald Berg, Ph.D., Jesse McKay, M.S. and Brett Nelson, M.S. Erlanger Medical Center and Logos Systems, Scotts Valley, CA Introduction The CyberKnife radiosurgery

More information

ADVANCED THERAPY DOSIMETER MODEL 35040

ADVANCED THERAPY DOSIMETER MODEL 35040 ADVANCED THERAPY DOSIMETER MODEL 35040 TECHNICAL SPECIFICATIONS Designed for ultra long-term stability error of approximately 0.1% per five years. Virtually removes effects of system leakage during measurement.

More information

Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector

Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector ORIGINAL ARTICLES Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector Gopiraj ANNAMALAI 1, Ramasubramanian VELAYUDHAM 2 ABSTRACT Received: 7.07.2009 Accepted: 2.11.2009

More information

Improved spatial resolution by MOSFET dosimetry of an x-ray microbeam

Improved spatial resolution by MOSFET dosimetry of an x-ray microbeam University of Wollongong Research Online Faculty of Engineering - Papers (Archive) Faculty of Engineering and Information Sciences 2000 Improved spatial resolution by MOSFET dosimetry of an x-ray microbeam

More information

X-Ray Transport, Diagnostic, & Commissioning Plans. LCLS Diagnostics and Commissioning Workshop

X-Ray Transport, Diagnostic, & Commissioning Plans. LCLS Diagnostics and Commissioning Workshop X-Ray Transport, Diagnostic, & Commissioning Plans LCLS Diagnostics and Commissioning Workshop *This work was performed under the auspices of the U.S. Department of Energy by the University of California,

More information

Application of spherical diodes for megavoltage photon beams dosimetry

Application of spherical diodes for megavoltage photon beams dosimetry Application of spherical diodes for megavoltage photon beams dosimetry Benigno Barbés, Juan D. Azcona, Javier Burguete, and Josep M. Martí-Climent Citation: Medical Physics 41, 012102 (2014); doi: 10.1118/1.4837178

More information

- Water resistant. - Large size.

- Water resistant. - Large size. GAFCHROMIC EBT product brochure GAFCHROMIC EBT FEATURES GAFCHROMIC EBT dosimetry film has been developed specifically to address the needs of the medical physicist and dosimetrist working in the radiotherapy

More information

Chemical Engineering 412

Chemical Engineering 412 Chemical Engineering 412 Introductory Nuclear Engineering Lecture 25 Radiation Detection & Measurement Spiritual Thought 2 I realize that there are some, perhaps many, [who] feel overwhelmed by the lack

More information

Photomultiplier Tube

Photomultiplier Tube Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or

More information

A Multichannel Dosimeter Based on Scintillating Fibers for Medical Applications

A Multichannel Dosimeter Based on Scintillating Fibers for Medical Applications A Multichannel Dosimeter Based on Scintillating Fibers for Medical Applications K.-H. Becks 1, J. Drees 1, K. Goldmann 1, I.M. Gregor 1 2, M. Heintz, Arndt Roeser* 1 Fachbereich Physik, Bergische Universität-Gesamthochschule

More information

Radiation Detection by Cerenkov Emission in. Optical Fibers at TTF

Radiation Detection by Cerenkov Emission in. Optical Fibers at TTF Tesla-Report 2-27 Radiation Detection by Cerenkov Emission in Optical Fibers at TTF by E. Janata 1, M. Körfer 2 1 Hahn-Meitner-Institut Berlin, Bereich Solarenergieforschung, D-1419 Berlin 2 Deutsches

More information

Radiation Detection Instrumentation

Radiation Detection Instrumentation Radiation Detection Instrumentation Principles of Detection and Gas-filled Ionization Chambers Neutron Sensitive Ionization Chambers Detection of radiation is a consequence of radiation interaction with

More information

High granularity scintillating fiber trackers based on Silicon Photomultiplier

High granularity scintillating fiber trackers based on Silicon Photomultiplier High granularity scintillating fiber trackers based on Silicon Photomultiplier A. Papa Paul Scherrer Institut, Villigen, Switzerland E-mail: angela.papa@psi.ch Istituto Nazionale di Fisica Nucleare Sez.

More information

X-ray tube with needle-like anode

X-ray tube with needle-like anode NUKLEONIKA 2002;47(3):101 105 ORIGINAL PAPER X-ray tube with needle-like anode Mieczys aw S apa, W odzimierz StraÊ, Marek Traczyk, Jerzy Dora, Miros aw Snopek, Ryszard Gutowski, Wojciech Drabik Abstract

More information

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC R. Bellazzini a,b, G. Spandre a*, A. Brez a, M. Minuti a, M. Pinchera a and P. Mozzo b a INFN Pisa

More information

How Does One Obtain Spectral/Imaging Information! "

How Does One Obtain Spectral/Imaging Information! How Does One Obtain Spectral/Imaging Information! How do we measure the position, energy, and arrival time of! an X-ray photon?! " What we observe depends on the instruments that one observes with!" In

More information

GAFCHROMIC HD-810 Radiochromic Dosimetry Film Configuration, Specifications and Performance Data

GAFCHROMIC HD-810 Radiochromic Dosimetry Film Configuration, Specifications and Performance Data GAFCHROMIC HD-810 Radiochromic Dosimetry Film Configuration, Specifications and Performance Data Description GAFCHROMIC HD-810 dosimetry film is designed for the measurement of absorbed dose of high-energy

More information

LaBr 3 :Ce, the latest crystal for nuclear medicine

LaBr 3 :Ce, the latest crystal for nuclear medicine 10th Topical Seminar on Innovative Particle and Radiation Detectors 1-5 October 2006 Siena, Italy LaBr 3 :Ce, the latest crystal for nuclear medicine Roberto Pani On behalf of SCINTIRAD Collaboration INFN

More information

CMS Conference Report

CMS Conference Report Available on CMS information server CMS CR 2004/067 CMS Conference Report 20 Sptember 2004 The CMS electromagnetic calorimeter M. Paganoni University of Milano Bicocca and INFN, Milan, Italy Abstract The

More information

The physical characteristics of a SLIC-EPID for transmitted dosimetry

The physical characteristics of a SLIC-EPID for transmitted dosimetry Iran. J. Radiat. Res., 2005; 2 (4): 175-183 The physical characteristics of a SLIC-EPID for transmitted dosimetry M. Mohammadi 1,2,3* and E. Bezak 1,2 1 School of Chemistry and Physics, The University

More information

Digital Radiography : Flat Panel

Digital Radiography : Flat Panel Digital Radiography : Flat Panel Flat panels performances & operation How does it work? - what is a sensor? - ideal sensor Flat panels limits and solutions - offset calibration - gain calibration - non

More information

Characterization of Laser Eyewear Using Varying Pulse Conditions and Wavelengths

Characterization of Laser Eyewear Using Varying Pulse Conditions and Wavelengths Characterization of Laser Eyewear Using Varying Pulse Conditions and Wavelengths Michael D. Thomas, Andrew Griffin Spica Technologies Inc. 18 Clinton Dr. #3 Hollis, NH 03049 Bonnie Simmons Kentek Corporation

More information

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter MR-guided RT in Heidelberg Funded by the German Research Foundation (DFG) Heidelberg consortium received

More information

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage

More information

ABSORBED DOSE DISTRIBUTIONS USING THE ISODENSITOMETRIC METHOD FOR EXPOSURES WITH FILTER EMPLOYED FOR MAMMOGRAPHIES

ABSORBED DOSE DISTRIBUTIONS USING THE ISODENSITOMETRIC METHOD FOR EXPOSURES WITH FILTER EMPLOYED FOR MAMMOGRAPHIES Romanian Reports in Physics, Vol. 65, No. 1, P. 168 177, 213 ABSORBED DOSE DISTRIBUTIONS USING THE ISODENSITOMETRIC METHOD FOR EXPOSURES WITH FILTER EMPLOYED FOR MAMMOGRAPHIES F. SCARLAT 1, A. SCARISOREANU

More information