A Multichannel Dosimeter Based on Scintillating Fibers for Medical Applications

Size: px
Start display at page:

Download "A Multichannel Dosimeter Based on Scintillating Fibers for Medical Applications"

Transcription

1 A Multichannel Dosimeter Based on Scintillating Fibers for Medical Applications K.-H. Becks 1, J. Drees 1, K. Goldmann 1, I.M. Gregor 1 2, M. Heintz, Arndt Roeser* 1 Fachbereich Physik, Bergische Universität-Gesamthochschule Wuppertal D Wuppertal, Germany 2 Klinische Strahlenphysik, Universitätsklinikum Essen D Essen, Germany *Klinikum Wuppertal GmbH, Klinik f. Strahlentherapie u. Radio-Onkologie D Wuppertal, Germany ABSTRACT Dosimetry with plastic scintillators is an interesting alternative for the measurement of the absorbed dose. A scintillator does barely disturb the radiation field due to its mass absorption coefficients which is water-equivalent in a wide range of energies. Furthermore, plastic scintillation dosimeters provide a fast and direct reading of the measured value combined with a high spatial resolution. In the set-up described so far, the light produced by the scintillators was transported via light-guides to single-channel or multi-channel photomultipliers to be transformed into an electric current read out by pico-amperemeters. The use of photomultipliers becomes expensive and complicated for a dosimeter system with many parallel channels. For such applications, an image intensifier coupled to a CCD is a simpler approach which can read out some 80 fibers in parallel [1] [2]. In this note, the design and the first successful measurements of a prototype set-up are described. Keywords: dosimetry, multichannel, scintillating fibers, brachy therapy, teletherapy 1 Introduction There exist two radiation methods for cancer treatment, brachytherapy and tele therapy [2]. The common method is the teletherapy, a treatment with radiation from the outside (e.g. 60 Co source, linear accelerator). Detailed mapping of the used radiation fields in 2 or 3 dimensions is desirable for the dosimetry. These measurements are usually done with a single ionisation chamber. In this case it is only possible to measure one field position at a time. For a detailed field mapping too much time is needed. With the developed multichannel dosimeter direct and simultaneous reading of the absorbed doses in several field points becomes possible.

2 Another radiation treatment is the brachytherapy. For example, in the eye plaque treatment an applicator with a radioactive source is applied directly to the eye (see figure 1). By changing the form of the applicator the field of the emitted radiation can be influenced. Therefore high field gradients have to be measured without disturbing the physical conditions. A high spatial resolution of the detector is as well necessary. Plastic scintillators can be used as dosimeters because they emit visible light proportional to the absorbed electron and gamma dose rate [3][5]. A scintillator does not disturb the radiation field due to its mass absorption coefficient and its mass stopping power which are water-equivalent in a wide range of energies. Furthermore, plastic scintillation dosimeters provide a fast and direct reading of the measuring values combined with a high spatial resolution. Depending on the spatial resolution needed, the scintillators can be as small as 1 mm 3 or less. The light produced by the scintillators is transported through clear optical plastic fibers. In a set-up discussed by Flühs et al. [3] and by Beddar et al. [5], these fibers are read out by single-channel or multi-channel photomultipliers. For applications where a small number of channels is required, they are the ideal light-to-current converters. For set-ups with more than channels, however, the electronic expenditure to measure the currents is large. The parallel multi-amperemeter systems required for that purpose are commercially not available. An image intensifier coupled to a CCD is an alternative approach. Such a system is easier to handle. Only some supply electronics and a frame grabber card for the data acquisition by a PC are needed. In the following, we describe a system which allows to read out up to 80 detectors in parallel. Using such a multi-channel dosimeter, all necessary basic functionality tests were performed as well as some studies in a clinical environment. 2 The Set-Up of the Multichannel-Dosimeter The described dosimeter will consist altogether of 80 channels, 40 of these channels with a scintillator (1 mm 3 ) 1 and an optical fiber 2. The remaining 40 channels are without scintillator. They will be used for the suppression of the Čerenkov light background produced in the light guiding fibers. The read-out system (see figure 2) is set up in such a way that the incoming light guides are kept at fixed positions by a mask to ensure a good optical contact between fibers and the image intensifier. The image intensifier transforms the light from all 80 fibers into electrons, amplifies them in parallel, transforms them back into photons and re-emits an amplified image of the incoming light signals. The image intensifier keeps both, spatial resolution, i.e. the ability to distinguish between the individual fibers, and the relative intensity of each fiber s light, i.e. the amplification is linear [1] [2]. The image emitted by the image intensifier is transported via an optical fiber taper to a CCD. This fiber taper is a conical optical light guide consisting of a large number of thin glass fibers. The diameter of these fibers all decrease towards one end of the taper, so an image transmitted by the taper will be reduced in size, but its relative spatial resolution will be maintained. The CCD integrates the light signal of each of the 80 fibers in 1 BC 400 from Bicron Technology 2 CWKF E from Cunz Toray 2

3 parallel over 20 millisecond. The image is transformed into standard TV format and read out by a personal computer equipped with a frame grabber 3, an extension card for a PC converting a video signal into digital information. To increase the precision, the digital signal can be summed up in the PC over several integration periods of the CCD - typically between a few hundred and a few thousand corresponding to total measurement times of one to five minutes. The dynamic range of the multi-channel dosimeter is determined by two factors. For an individual reading, it is given by the maximum signal the CCD can handle without saturation, and the minimum signal that can be distinguished from the background noise. These limitations give a dynamic range of Furthermore, from one measurement to another, the amplification of the multi-channel dosimeter s image intensifier can be adjusted over a range of The smallest detectable signals are in the order of a few hundred photons per second emerging from a fiber of 1 mm diameter. The amount of radiation equivalent to this signal depends on the size of the scintillator coupled to the fiber, the transmission characteristics of the fiber, and the background noise. 3 Measurements To test the system, depth dose measurements performed with a ionisation chamber were compared to measurements performed with a 10-channel prototype of the described multichannel dosimeter. The used source of radiation was a Siemens linear accelerator at the Clinic for Radiation Therapy, Radio Oncology and Nuclear Medicine in Wuppertal. The acceleration bias was adjustable from 6 MV to 18 MV. Measurements at different acceleration biases were performed. The depth dose distribution is the dose distribution in a body or in a phantom along the central beam. Mostly relative or proportional depth doses are indicated, which refer to the absorbed dose maximum or an indicated reference depth. The so called build up effect describes an increase of the absorbed dose, the passing through a maximum and the following dose decline. For photons this build up effect is due to secondary electrons, which are generated by photoelectric effect, Compton-effect, or pair production depending upon the photon energy. Starting from a photon energy of 0,6 MeV the Comptoneffect outweighs in water. The secondary electrons move in the same direction as the primary beam and increase the absorbed dose up to a depth, which corresponds to the range of the secondary electrons. The dose decline is caused by the attenuation of the photon beams. From a certain depth the curve follows approximately an exponential function. The slope of the dose decline depends on the quality of the radiation, the field size and the divergence of the jet bundle [4]. This progression is shown in figure 3 and figure 4. In the described measurements a field size of 5 cm 5 cm was used. In a plate made of rigid water (RW3 4 ) two out of x10 fibers were placed in a drilling for this purpose. The fibers were positioned in such way, that the tips with the scintillating material were 3 National Instruments PCI a material, which has the same absorption coefficient as water 3

4 situated in the iso center 5. One of the three fibers was a reference fiber without scintillator. Therefore it was possible to measure the background and the signal simultaneously. The background has been subtracted from the measured values afterwards. In order to suppress the Čerenkov light background. Further RW3 plates were subsequently placed onto the previous and measurements were repeated until a depth of 20 cm was achieved. As comparison, the measurements were repeated under same conditions with an ionization chamber, calibrated for the same energy region. The determined values are shown in figure 3. Only two out of ten fibres are shown exemplarily. The build up effect is more strongly pronounced with the fibers than with the ionization chamber, which is to due to the characteristics of the fibers. Figure 3 depicts the depth dose distribution measured at 6MV with an ionisation chamber and the prototype multichannel dosimeter. In larger depths than the maximum of the build up effect, the gradient of the fiber data corresponds to the reference data. In smaller depths the curves of the fibers are clearly steeper than those of the reference data. Due to their physical characteristics the scintillators do not influence the physical environment. However the photons are more influenced by the gas volume of the ionization chamber, the gas and the material of the chamber cause a modification of the environment. The depth dose distribution at an accelerating voltage of 18 MV is shown in figure 4. Here a steeper rise of the measured values for smaller depths is being detected for the same reasons as described above. However the gradient of the fiber values does not correspond to those of the ionization chamber. The depth dose distribution of the fibers are substantially steeper, than those of the ionization chamber. The assumption is that this is due to the coupling of the optical fibers to the opto-electronic read-out. The losses may be explained by a different quality of the optical coupling between fiber ends and the image intensifier. Because of the different manufacturing quality of the fiber ends, the losses during the transition diverse slightly and therefore different Čerenkov light background is registered in the reference fiber as actually occur in the other fibers. In higher energy regions this effect becomes more apparent. Therefore one should take more care when preparing the fibers for the coupling to the image intesifier. A reference fiber for each scintillating fiber would improve the background reduction. Measuring the Čerenkov light background of each fiber before attaching the scintillator would indicate the fluctuations within the manufacturing of the fiber ends. It is possible to reduce the Čerenkov light background further more by using longer light guides. The attenuation of the guides is wavelength depending. The Čerenkov light will be more attenuated in some wavelenth ranges than the information signal. All this will be done for a further prototype before taking it into clinical environment. 4 Conclusion and Future Work In conclusion, we have developed a new parallel read-out system for dosimetry with plastic scintillators, plastic light guides, an image intensifier coupled to a CCD and read out by a PC. This system combines the advantages of the plastic scintillator dosimetry, such as water-equivalence, direct reading and high spatial resolution, with a parallel, computer-controlled read-out of the measured signals. Further advantages over com- 5 intersection of the central beam with the rotation axis of the irradiation device 4

5 monly used systems are its intrinsic simplicity, the high number of channels that can be read out in parallel with negligible cross-talk and in its low cost per channel. Although no long term test of the entire system has been performed under clinical conditions. Yet, the first experiences are very encouraging. Depth dose measurements with a prototye of this dosimeter were performed at different particle energies and compared to measurments with an ionisation chamber. For lower energies the results agree with each other. The uncertainty due to Čerenkov light can be reduced by using reference fibers. At the prototype dosimeter problems with the coupling of the light guides to the read out system appeared. For a further prototype this coupling has to be optimised by cutting the fibers more carful and polishing the ends finer. Also it is possible to reduce the Čerenkov light by using about 10 m long light guides. Due to the wavelength depending attenuation of the light guides the Čerenkov light will be more attenuated in some wavelength ranges than the information signal. When applying the dosimeter in the teletherapy it is also necessary to operate the read out electronic in a sufficient large distance to the radiation source. This can automatically be ensured by the proposed long light guides. Further studies will include long time measurements in the teletherapy with a new prototype, an individual calibration of each channel, and user friendly software tools for the operating of this system. References [1] M. Garg, Konzeption, Aufbau und Test eines Vielkanal-Dosimeters mit szintillierenden Fasern, WUB DIS 96 17, Universität Wuppertal, 1996 [2] I.M. Gregor, Ein Vielkanaldosimeter mit szintillierenden Fasern Klinische Messungen bei Energien oberhalb 1 MeV, WU D 98 14, Universität Wuppertal, 1998 [3] D. Flühs, M. Heintz, F. Indenkämpen, C. Wieczorek, H. Kolanoski and U. Quast, Direct reading measurement of absorbed dose with plastic scintillators - The general concept and applications to ophthalmic plaque dosimetry, Med. Phys. 23 (1996), pages [4] H. Reich, Dosimetrie ionisierender Strahlung, B.G. Teubner, Stuttgart, 1990 [5] A.S. Beddar, T.R. Mackie, and F.H. Attix, Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: 1. Physical characteristics and theoretical considerations, Phys. Med. Biol. 37 (1992), pages

Scintillation Counters

Scintillation Counters PHY311/312 Detectors for Nuclear and Particle Physics Dr. C.N. Booth Scintillation Counters Unlike many other particle detectors, which exploit the ionisation produced by the passage of a charged particle,

More information

A Scintillating Fiber Dosimeter for Radiology and Brachytherapy. with photodiode readout

A Scintillating Fiber Dosimeter for Radiology and Brachytherapy. with photodiode readout A Scintillating Fiber Dosimeter for Radiology and Brachytherapy with photodiode readout Florbela Rêgo a), and Maria da Conceição Abreu 5 Laboratório de Instrumentação e Física Experimental de Partículas,

More information

SCINTILLATING FIBER DOSIMETER ARRAY

SCINTILLATING FIBER DOSIMETER ARRAY SCINTILLATING FIBER DOSIMETER ARRAY FIELD OF THE INVENTION [0001] This invention relates generally to the field of dosimetry and, more particularly, to rapid, high-resolution dosimeters for advanced treatment

More information

Moderne Teilchendetektoren - Theorie und Praxis 2. Dr. Bernhard Ketzer Technische Universität München SS 2013

Moderne Teilchendetektoren - Theorie und Praxis 2. Dr. Bernhard Ketzer Technische Universität München SS 2013 Moderne Teilchendetektoren - Theorie und Praxis 2 Dr. Bernhard Ketzer Technische Universität München SS 2013 7 Signal Processing and Acquisition 7.1 Signals 7.2 Amplifier 7.3 Electronic Noise 7.4 Analog-to-Digital

More information

Australia. Australia. Keywords: Radiation therapy, radiation dosimeters, scintillators, optical fibers, Cerenkov radiation 1.

Australia. Australia. Keywords: Radiation therapy, radiation dosimeters, scintillators, optical fibers, Cerenkov radiation 1. Signal versus noise in fiber coupled radiation dosimeters for medical applications Susan H. Law a, Natalka Suchowerska b,d, Simon C. Fleming a, A. Samad Beddar c, David McKenzie d a Australian Photonics

More information

Monte Carlo study on a new concept of a scanning photon beam system for IMRT

Monte Carlo study on a new concept of a scanning photon beam system for IMRT NUKLEONIKA 2011;56(4):291 297 ORIGINAL PAPER Monte Carlo study on a new concept of a scanning photon beam system for IMRT Anna M. Wysocka-Rabin, Günter H. Hartmann Abstract. Intensity-modulated radiation

More information

Introduction of a Single Chip TLD System for Patient Dosimetry

Introduction of a Single Chip TLD System for Patient Dosimetry Introduction of a Single Chip TLD System for Patient Dosimetry C. Hranitzky a, M. Halda a, G. Müller a, B. Obryk b, H. Stadtmann a* a Austrian Research Centers GmbH ARC, 2444 Seibersdorf, Austria. b Institute

More information

Attenuation length in strip scintillators. Jonathan Button, William McGrew, Y.-W. Lui, D. H. Youngblood

Attenuation length in strip scintillators. Jonathan Button, William McGrew, Y.-W. Lui, D. H. Youngblood Attenuation length in strip scintillators Jonathan Button, William McGrew, Y.-W. Lui, D. H. Youngblood I. Introduction The ΔE-ΔE-E decay detector as described in [1] is composed of thin strip scintillators,

More information

Summer Student project report

Summer Student project report Summer Student project report Mika Väänänen September 1, 2017 Abstract In this report I give a brief overview of my activities during the summer student project. I worked on the scintillating fibre (SciFi)

More information

Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010

Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010 Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010 Indra J. Das, PhD, FACR Department of Radiation Oncology Indiana University of School of Medicine & Midwest Proton Radiation

More information

Post-irradiation colouration of Gafchromic EBT radiochromic film

Post-irradiation colouration of Gafchromic EBT radiochromic film INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 5 (25) N281 N285 PHYSICS IN MEDICINE AND BIOLOGY doi:1.188/31-9155/5/2/n4 NOTE Post-irradiation colouration of Gafchromic EBT radiochromic film Tsang Cheung

More information

SCINTILLATOR WAVELENGTH INFLUENCE IN AN OPTICAL DOSIMETER

SCINTILLATOR WAVELENGTH INFLUENCE IN AN OPTICAL DOSIMETER SCINTILLATOR WAVELENGTH INFLUENCE IN AN OPTICAL DOSIMETER Master of Medical Physics Department of Physics and Astronomy University of Canterbury Christchurch New Zealand 2006 By Tom Chien-Sheng Chen TABLE

More information

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Gena M.A.H 1, Ahmed L.El-Attar 2, Elbadry M. Zahran 3, Hany El-Gamal

More information

Thulium-Doped Fiber Amplifier Development for Power Scaling the 2 Micron Coherent Laser Absorption Instrument for ASCENDS

Thulium-Doped Fiber Amplifier Development for Power Scaling the 2 Micron Coherent Laser Absorption Instrument for ASCENDS Thulium-Doped Fiber Amplifier Development for Power Scaling the 2 Micron Coherent Laser Absorption Instrument for ASCENDS Mark W. Phillips Lockheed Martin Coherent Technologies 135 South Taylor Avenue,

More information

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition

More information

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Takeshi Fujiwara 1, Yuki Mitsuya 2, Hiroyuki Takahashi 2, and Hiroyuki Toyokawa 2 1 National Institute of Advanced Industrial

More information

Radiation Detection Instrumentation

Radiation Detection Instrumentation Radiation Detection Instrumentation Principles of Detection and Gas-filled Ionization Chambers Neutron Sensitive Ionization Chambers Detection of radiation is a consequence of radiation interaction with

More information

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT Quality control of Gamma Camera By Dr/ Ibrahim Elsayed Saad 242 NMT WHAT IS QUALITY? The quality of a practice is to fulfill the expectations and demands from: Patient Clinicain Your self Quality assurance

More information

A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering

A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering Sandra E. Burch Department of Radiology, Medical College of Georgia, Augusta, Georgia 30912 Kimberlee J. Kearfott

More information

Development of Personal Dosimeter Using Electronic Dose Conversion Method

Development of Personal Dosimeter Using Electronic Dose Conversion Method Proceedings of the Korean Nuclear Spring Meeting Gyeong ju, Korea, May 2003 Development of Personal Dosimeter Using Electronic Dose Conversion Method Wanno Lee, Bong Jae Lee, and Chang Woo Lee Korea Atomic

More information

X-rays in medical diagnostics

X-rays in medical diagnostics X-rays in medical diagnostics S.Dolanski Babić 2017/18. History W.C.Röntgen (1845-1923) discovered a new type of radiation Nature, Jan. 23. 1896.; Science, Feb.14. 1896. X- rays: Induced the ionization

More information

High collection efficiency MCPs for photon counting detectors

High collection efficiency MCPs for photon counting detectors High collection efficiency MCPs for photon counting detectors D. A. Orlov, * T. Ruardij, S. Duarte Pinto, R. Glazenborg and E. Kernen PHOTONIS Netherlands BV, Dwazziewegen 2, 9301 ZR Roden, The Netherlands

More information

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering RR Oct 2001 SS Dec 2001 MJ Oct 2009 Physics 34000 Laboratory Scattering of Photons from Electrons: Compton Scattering Objective: To measure the energy of high energy photons scattered from electrons in

More information

The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response

The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response Iran. J. Radiat. Res., 2005; 3 (1): 3-10 The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response M. Mohammadi 1,2,3* and E. Bezak 1,2

More information

Photomultiplier Tube

Photomultiplier Tube Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced

More information

arxiv:hep-ex/ v1 19 Apr 2002

arxiv:hep-ex/ v1 19 Apr 2002 STUDY OF THE AVALANCHE TO STREAMER TRANSITION IN GLASS RPC EXCITED BY UV LIGHT. arxiv:hep-ex/0204026v1 19 Apr 2002 Ammosov V., Gapienko V.,Kulemzin A., Semak A.,Sviridov Yu.,Zaets V. Institute for High

More information

Radiographic sensitivity improved by optimized high resolution X -ray detector design.

Radiographic sensitivity improved by optimized high resolution X -ray detector design. DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France Radiographic sensitivity improved by optimized high resolution X -ray detector

More information

5. Scintillation counters

5. Scintillation counters 5. Scintillation counters to detect radiation by means of scintillation is among oldest methods of particle detection historical example: particle impinging on ZnS screen -> emission of light flash principle

More information

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26 Today s Outline - January 25, 2018 C. Segre (IIT) PHYS 570 - Spring 2018 January 25, 2018 1 / 26 Today s Outline - January 25, 2018 HW #2 C. Segre (IIT) PHYS 570 - Spring 2018 January 25, 2018 1 / 26 Today

More information

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

3D Diode Array Commissioning: Building Confidence in 3D QA Technology 3D Diode Array Commissioning: Building Confidence in 3D QA Technology Caroline Yount, MS CANCER CENTER 3D QA The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose

More information

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors Nuclear Physics #1 Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors Introduction: In this experiment you will use both scintillation and semiconductor detectors to study γ- ray energy spectra. The

More information

5. Scintillation counters

5. Scintillation counters 5. Scintillation counters to detect radiation by means of scintillation is among oldest methods of particle detection particle impinging on ZnS screen -> emission of light flash principle of scintillation

More information

Atomic and Nuclear Physics

Atomic and Nuclear Physics Atomic and Nuclear Physics Nuclear physics -spectroscopy LEYBOLD Physics Leaflets Detecting radiation with a scintillation counter Objects of the experiments Studying the scintillator pulses with an oscilloscope

More information

Preliminary studies of a new monitor ionization chamber

Preliminary studies of a new monitor ionization chamber 1 Preliminary studies of a new monitor ionization chamber Maíra T. Yoshizumi, Vitor Vivolo and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares (IPEN-CNEN) Comissão Nacional de Energia

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Chemistry 985. Some constants: q e 1.602x10 19 Coul, ɛ x10 12 F/m h 6.626x10 34 J-s, c m/s, 1 atm = 760 Torr = 101,325 Pa

Chemistry 985. Some constants: q e 1.602x10 19 Coul, ɛ x10 12 F/m h 6.626x10 34 J-s, c m/s, 1 atm = 760 Torr = 101,325 Pa Chemistry 985 Fall, 2o17 Distributed: Mon., 17 Oct. 17, 8:30AM Exam # 1 OPEN BOOK Due: 17 Oct. 17, 10:00AM Some constants: q e 1.602x10 19 Coul, ɛ 0 8.854x10 12 F/m h 6.626x10 34 J-s, c 299 792 458 m/s,

More information

Gamma Spectrometer Initial Project Proposal

Gamma Spectrometer Initial Project Proposal Gamma Spectrometer Initial Project Proposal Group 9 Aman Kataria Johnny Klarenbeek Dean Sullivan David Valentine Introduction There are currently two main types of gamma radiation detectors used for gamma

More information

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array José A. Bencomo, * Geoffrey Ibbott, Seungsoo Lee, and Joao A. Borges Department of Radiation Physics.

More information

Cylindrical Ion Chambers Victoreen Model 550 Series

Cylindrical Ion Chambers Victoreen Model 550 Series Cylindrical Ion Chambers Victoreen Model 550 Series! Cylindrical Ion Chambers for use with Model 35040 and Model 530 electrometers! Wide range of applications in Diagnostic X-Ray and Radiation Oncology

More information

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare GE Healthcare Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare There is excitement across the industry regarding the clinical potential of a hybrid

More information

Optical fibre. Principle and applications

Optical fibre. Principle and applications Optical fibre Principle and applications Circa 2500 B.C. Earliest known glass Roman times-glass drawn into fibers Venice Decorative Flowers made of glass fibers 1609-Galileo uses optical telescope 1626-Snell

More information

A Sort-of Tissue Equivalent Proportional Counter (STEPC) for Space Radiation Dosimetry Applications

A Sort-of Tissue Equivalent Proportional Counter (STEPC) for Space Radiation Dosimetry Applications A Sort-of Tissue Equivalent Proportional Counter (STEPC) for Space Radiation Dosimetry Applications Eric Benton, Tyler Collums, and Art Lucas E. V. Benton Radiation Physics Laboratory Oklahoma State University

More information

Beam Production, characteristics and shaping

Beam Production, characteristics and shaping Kantonsspital Luzern Beam Production, characteristics and shaping Dr. Manfred Sassowsky Cantonal Hospital Lucerne (KSL) Institute for Radio-Oncology 3.9.2007 X-ray production 60 Co units Linear Accelerators

More information

Calibration of Scintillator Tiles with SiPM Readout

Calibration of Scintillator Tiles with SiPM Readout EUDET Calibration of Scintillator Tiles with SiPM Readout N. D Ascenzo, N. Feege,, B. Lutz, N. Meyer,, A. Vargas Trevino December 18, 2008 Abstract We report the calibration scheme for scintillator tiles

More information

I. Introduction.

I. Introduction. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Accuracy of measuring half- and quarter-value layers and appropriate aperture width of a convenient method using a lead-covered case

More information

Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector

Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector ORIGINAL ARTICLES Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector Gopiraj ANNAMALAI 1, Ramasubramanian VELAYUDHAM 2 ABSTRACT Received: 7.07.2009 Accepted: 2.11.2009

More information

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

Unit thickness. Unit area. σ = NΔX = ΔI / I 0 Unit thickness I 0 ΔI I σ = ΔI I 0 NΔX = ΔI / I 0 NΔX Unit area Δx Average probability of reaction with atom for the incident photons at unit area with the thickness of Delta-X Atom number at unit area

More information

Total Absorption Dual Readout Calorimetry R&D

Total Absorption Dual Readout Calorimetry R&D Available online at www.sciencedirect.com Physics Procedia 37 (2012 ) 309 316 TIPP 2011 - Technology and Instrumentation for Particle Physics 2011 Total Absorption Dual Readout Calorimetry R&D B. Bilki

More information

Calibration of KAP meters

Calibration of KAP meters Calibration of KAP meters Alexandr Malusek! Division of Radiological Sciences Department of Medical and Health Sciences Linköping University! 2014-04-15 1 Outline 1. KAP meter construction 2. Air kerma-area

More information

GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom

GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom Introduction: These EBT2/EBT3 films, shown in Figure 1a-c, are specially sized and formatted to uniquely fit the Accuray Ball Cube II Phantom. Each

More information

NM Module Section 2 6 th Edition Christian, Ch. 3

NM Module Section 2 6 th Edition Christian, Ch. 3 NM 4303 Module Section 2 6 th Edition Christian, Ch. 3 Gas Filled Chamber Voltage Gas filled chamber uses Hand held detectors cutie pie Geiger counter Dose calibrators Cutie pie Chamber voltage in Ionization

More information

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017 The Current State of EPID-Based Linear Accelerator Quality Assurance Timothy Ritter, PhD, DABR, FAAPM 1 Disclosures Employed by the Veterans Health Administration Faculty appointment with the University

More information

Cylindrical Ion Chambers

Cylindrical Ion Chambers Cylindrical Ion Chambers Radiation Oncology ON Victoreen Model 550T Series Cylindrical Ion Chambers for use with Model 35040 and Model 560 electrometers Wide range of applications in Diagnostic X-Ray and

More information

ABSORBED DOSE DISTRIBUTIONS USING THE ISODENSITOMETRIC METHOD FOR EXPOSURES WITH FILTER EMPLOYED FOR MAMMOGRAPHIES

ABSORBED DOSE DISTRIBUTIONS USING THE ISODENSITOMETRIC METHOD FOR EXPOSURES WITH FILTER EMPLOYED FOR MAMMOGRAPHIES Romanian Reports in Physics, Vol. 65, No. 1, P. 168 177, 213 ABSORBED DOSE DISTRIBUTIONS USING THE ISODENSITOMETRIC METHOD FOR EXPOSURES WITH FILTER EMPLOYED FOR MAMMOGRAPHIES F. SCARLAT 1, A. SCARISOREANU

More information

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images Aiping Ding, Bin Han, Lei Wang, Lei Xing Department of Radiation Oncology, Stanford University School of

More information

Electronic Brachytherapy Sources. Thomas W. Rusch

Electronic Brachytherapy Sources. Thomas W. Rusch Electronic Brachytherapy Sources Thomas W. Rusch Educational Objectives Understand key elements of ebx source construction & operation Understand the rationale and methods for air kerma strength calibration

More information

Product Range Electronic Units

Product Range Electronic Units Pyramid Technical Consultants, Inc. 1050 Waltham Street Suite 200 Lexington, MA 02421 TEL: +1 781 402-1700 TEL (UK): +44 1273 492001 FAX: (781) 402-1750 EMAIL: SUPPORT@PTCUSA.COM Product Range Electronic

More information

Chemistry Instrumental Analysis Lecture 10. Chem 4631

Chemistry Instrumental Analysis Lecture 10. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 10 Types of Instrumentation Single beam Double beam in space Double beam in time Multichannel Speciality Types of Instrumentation Single beam Requires stable

More information

Uses of Electromagnetic Waves

Uses of Electromagnetic Waves Uses of Electromagnetic Waves 1 of 42 Boardworks Ltd 2016 Uses of Electromagnetic Waves 2 of 42 Boardworks Ltd 2016 What are radio waves? 3 of 42 Boardworks Ltd 2016 The broadcast of every radio and television

More information

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES The current multiplication mechanism offered by dynodes makes photomultiplier tubes ideal for low-light-level measurement. As explained earlier, there

More information

Digital Radiography : Flat Panel

Digital Radiography : Flat Panel Digital Radiography : Flat Panel Flat panels performances & operation How does it work? - what is a sensor? - ideal sensor Flat panels limits and solutions - offset calibration - gain calibration - non

More information

Muon detection in security applications and monolithic active pixel sensors

Muon detection in security applications and monolithic active pixel sensors Muon detection in security applications and monolithic active pixel sensors Tracking in particle physics Gaseous detectors Silicon strips Silicon pixels Monolithic active pixel sensors Cosmic Muon tomography

More information

Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM

Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM Preamplifiers and amplifiers The current from PMT must be further amplified before it can be processed and counted (the number of electrons yielded

More information

Chapter 3 OPTICAL SOURCES AND DETECTORS

Chapter 3 OPTICAL SOURCES AND DETECTORS Chapter 3 OPTICAL SOURCES AND DETECTORS 3. Optical sources and Detectors 3.1 Introduction: The success of light wave communications and optical fiber sensors is due to the result of two technological breakthroughs.

More information

Observational Astronomy

Observational Astronomy Observational Astronomy Instruments The telescope- instruments combination forms a tightly coupled system: Telescope = collecting photons and forming an image Instruments = registering and analyzing the

More information

Beam Production, Characteristics and Shaping

Beam Production, Characteristics and Shaping Beam Production, Characteristics and Shaping Dr. Manfred Sassowsky Outline X-ray production 60 Co units Linear Accelerators Beam characteristics Beam shaping Beam Production, Characteristics and Shaping

More information

X-ray tube with needle-like anode

X-ray tube with needle-like anode NUKLEONIKA 2002;47(3):101 105 ORIGINAL PAPER X-ray tube with needle-like anode Mieczys aw S apa, W odzimierz StraÊ, Marek Traczyk, Jerzy Dora, Miros aw Snopek, Ryszard Gutowski, Wojciech Drabik Abstract

More information

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices Moving from film to digital: A study of digital x-ray benefits, challenges and best practices H.U. Pöhler 1 and N. D Ademo 2 DÜRR NDT GmbH & Co. KG, Höpfigheimer Straße 22, Bietigheim-Bissingen, 74321,

More information

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters 12 August 2011-08-12 Ahmad Darudi & Rodrigo Badínez A1 1. Spectral Analysis of the telescope and Filters This section reports the characterization

More information

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS By Aravind Ravichandran arr192@mail.usask.ca University of Saskatchewan Address: 2424 Cumberland

More information

Derivation of total filtration thickness for diagnostic x-ray source assembly

Derivation of total filtration thickness for diagnostic x-ray source assembly Physics in Medicine & Biology PAPER OPEN ACCESS Derivation of total filtration thickness for diagnostic x-ray source assembly To cite this article: Michiharu Sekimoto and Yoh Katoh 2016 Phys. Med. Biol.

More information

Real-Time Scanning Goniometric Radiometer for Rapid Characterization of Laser Diodes and VCSELs

Real-Time Scanning Goniometric Radiometer for Rapid Characterization of Laser Diodes and VCSELs Real-Time Scanning Goniometric Radiometer for Rapid Characterization of Laser Diodes and VCSELs Jeffrey L. Guttman, John M. Fleischer, and Allen M. Cary Photon, Inc. 6860 Santa Teresa Blvd., San Jose,

More information

Positron Emission Tomography

Positron Emission Tomography Positron Emission Tomography UBC Physics & Astronomy / PHYS 409 1 Introduction Positron emission tomography (PET) is a non-invasive way to produce the functional 1 image of a patient. It works by injecting

More information

熊本大学学術リポジトリ. Kumamoto University Repositor

熊本大学学術リポジトリ. Kumamoto University Repositor 熊本大学学術リポジトリ Kumamoto University Repositor Title Monte Carlo calculations of the rep correction factor, Ρ_, for cy chamber cav Author(s) Araki, Fujio CitationRadiological Physics and Technology Issue

More information

The Light Amplifier Concept

The Light Amplifier Concept The Light Amplifier Concept Daniel Ferenc 1 Eckart Lorenz 1,2 Daniel Kranich 1 Alvin Laille 1 (1) Physics Department, University of California Davis (2) Max Planck Institute, Munich Work supported partly

More information

Testing the Electronics for the MicroBooNE Light Collection System

Testing the Electronics for the MicroBooNE Light Collection System Testing the Electronics for the MicroBooNE Light Collection System Kathleen V. Tatem Nevis Labs, Columbia University & Fermi National Accelerator Laboratory August 3, 2012 Abstract This paper discusses

More information

Measurements of MeV Photon Flashes in Petawatt Laser Experiments

Measurements of MeV Photon Flashes in Petawatt Laser Experiments UCRL-JC-131359 PREPRINT Measurements of MeV Photon Flashes in Petawatt Laser Experiments M. J. Moran, C. G. Brown, T. Cowan, S. Hatchett, A. Hunt, M. Key, D.M. Pennington, M. D. Perry, T. Phillips, C.

More information

Electronic Instrumentation for Radiation Detection Systems

Electronic Instrumentation for Radiation Detection Systems Electronic Instrumentation for Radiation Detection Systems January 23, 2018 Joshua W. Cates, Ph.D. and Craig S. Levin, Ph.D. Course Outline Lecture Overview Brief Review of Radiation Detectors Detector

More information

Ph 3324 The Scintillation Detector and Gamma Ray Spectroscopy

Ph 3324 The Scintillation Detector and Gamma Ray Spectroscopy Ph 3324 The Scintillation Detector and Gamma Ray Spectroscopy Required background reading Attached are several pages from an appendix on the web for Tipler-Llewellyn Modern Physics. Read the section on

More information

Study of monitoring system of a calibration laser for the itop detector at Belle II

Study of monitoring system of a calibration laser for the itop detector at Belle II Università degli Studi di Padova Dipartimento di Fisica e Astronomia G. Galilei Corso di laurea in Fisica Tesi di Laurea Study of monitoring system of a calibration laser for the itop detector at Belle

More information

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring

More information

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This

More information

MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM *

MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM * MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM * C. Kessler and D.T. Burns December 2018 BUREAU INTERNATIONAL DES POIDS ET MESURES

More information

Chemical Engineering 412

Chemical Engineering 412 Chemical Engineering 412 Introductory Nuclear Engineering Lecture 25 Radiation Detection & Measurement Spiritual Thought 2 I realize that there are some, perhaps many, [who] feel overwhelmed by the lack

More information

CyberKnife Iris Beam QA using Fluence Divergence

CyberKnife Iris Beam QA using Fluence Divergence CyberKnife Iris Beam QA using Fluence Divergence Ronald Berg, Ph.D., Jesse McKay, M.S. and Brett Nelson, M.S. Erlanger Medical Center and Logos Systems, Scotts Valley, CA Introduction The CyberKnife radiosurgery

More information

International Journal of Advance Engineering and Research Development

International Journal of Advance Engineering and Research Development Scientific Journal of Impact Factor (SJIF): 5.71 International Journal of Advance Engineering and Research Development Volume 5, Issue 06, June -2018 e-issn (O): 2348-4470 p-issn (P): 2348-6406 STUDY OF

More information

FIBER OPTICS. Prof. R.K. Shevgaonkar. Department of Electrical Engineering. Indian Institute of Technology, Bombay. Lecture: 18.

FIBER OPTICS. Prof. R.K. Shevgaonkar. Department of Electrical Engineering. Indian Institute of Technology, Bombay. Lecture: 18. FIBER OPTICS Prof. R.K. Shevgaonkar Department of Electrical Engineering Indian Institute of Technology, Bombay Lecture: 18 Optical Sources- Introduction to LASER Diodes Fiber Optics, Prof. R.K. Shevgaonkar,

More information

Response characteristics of a tandem ionization chamber in standard X-ray beams

Response characteristics of a tandem ionization chamber in standard X-ray beams Applied Radiation and Isotopes 58 (2003) 495 500 Response characteristics of a tandem ionization chamber in standard X-ray beams Alessandro M. Costa*, Linda V.E. Caldas Instituto de Pesquisas Energ!eticas

More information

Development of Photon Detectors at UC Davis Daniel Ferenc Eckart Lorenz Alvin Laille Physics Department, University of California Davis

Development of Photon Detectors at UC Davis Daniel Ferenc Eckart Lorenz Alvin Laille Physics Department, University of California Davis Development of Photon Detectors at UC Davis Daniel Ferenc Eckart Lorenz Alvin Laille Physics Department, University of California Davis Work supported partly by DOE, National Nuclear Security Administration

More information

Goal of the project. TPC operation. Raw data. Calibration

Goal of the project. TPC operation. Raw data. Calibration Goal of the project The main goal of this project was to realise the reconstruction of α tracks in an optically read out GEM (Gas Electron Multiplier) based Time Projection Chamber (TPC). Secondary goal

More information

Laser Scanning for Surface Analysis of Transparent Samples - An Experimental Feasibility Study

Laser Scanning for Surface Analysis of Transparent Samples - An Experimental Feasibility Study STR/03/044/PM Laser Scanning for Surface Analysis of Transparent Samples - An Experimental Feasibility Study E. Lea Abstract An experimental investigation of a surface analysis method has been carried

More information

Detectors for microscopy - CCDs, APDs and PMTs. Antonia Göhler. Nov 2014

Detectors for microscopy - CCDs, APDs and PMTs. Antonia Göhler. Nov 2014 Detectors for microscopy - CCDs, APDs and PMTs Antonia Göhler Nov 2014 Detectors/Sensors in general are devices that detect events or changes in quantities (intensities) and provide a corresponding output,

More information

X-ray light valve (XLV): a novel detectors technology for digital mammography*

X-ray light valve (XLV): a novel detectors technology for digital mammography* X-ray light valve (XLV): a novel detectors technology for digital mammography* Sorin Marcovici, Vlad Sukhovatkin, Peter Oakham XLV Diagnostics Inc., Thunder Bay, ON P7A 7T1, Canada ABSTRACT A novel method,

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

Supplementary Figure S1. Schematic representation of different functionalities that could be

Supplementary Figure S1. Schematic representation of different functionalities that could be Supplementary Figure S1. Schematic representation of different functionalities that could be obtained using the fiber-bundle approach This schematic representation shows some example of the possible functions

More information

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING DIGITAL IMAGE PROCESSING IN X-RAY IMAGING Shalini Kumari 1, Bachan Prasad 2,Aliya Nasim 3 Department of Electronics And Communication Engineering R.V.S College of Engineering & Technology, Jamshedpur,

More information

Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4

Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4 Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4 S.V. Roth, R. Döhrmann, M. Dommach, I. Kröger, T. Schubert, R. Gehrke Definition of the upgrade The wiggler beamline BW4 is dedicated to

More information

Applications of Optics

Applications of Optics Nicholas J. Giordano www.cengage.com/physics/giordano Chapter 26 Applications of Optics Marilyn Akins, PhD Broome Community College Applications of Optics Many devices are based on the principles of optics

More information

PHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002

PHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002 PHYSICS 334 - ADVANCED LABORATORY I COMPTON SCATTERING Spring 00 Purposes: Demonstrate the phenomena associated with Compton scattering and the Klein-Nishina formula. Determine the mass of the electron.

More information